Haag-Streit Octopus 300 Perimeter Service manual

4 (1)

OCTOPUS® 300

Service Manual

HS No. 7220084

8. Edition / 2014-08

HAAG-STREIT AG, Switzerland, Phone: +41 31 978 01 11, Fax: +41 31 978 02 82, info@haag-streit.com

HAAG-STREIT DEUTSCHLAND GmbH, Germany, Phone: +49 4103 709 02, Fax: +49 4103 709 370, info@haag-streit.de HAAG-STREIT FRANCE, France, Phone +33 4 5009 0033, Fax +33 4 5009 7190, info@haag-streit.fr

HAAG-STREIT UK, United Kingdom, Phone +44 1279 414969, Fax +44 1279 635232, info@haag-streit.uk HAAG-STREIT USA, INC., USA, Phone: +1 513 336 6858, Fax: +1 513 336 7828, octopus@haag-streit-usa.com

OCTOPUS 300 Service Manual

HAAG STREIT No. 7220084

 

 

Tradition and Innovation

TABLE OF CONTENTS

 

 

1

PARTS DELIVERED ......................................................................................................................................................................

 

3

2

SAFETY INSTRUCTIONS..............................................................................................................................................................

 

4

 

2.1

General................................................................................................................................................

 

4

 

2.2

Instrument Transportation ...................................................................................................................

 

4

 

2.3

Symbols...............................................................................................................................................

 

5

 

Special Notices in the Text ............................................................................................................................

 

5

3

INTRODUCTION ............................................................................................................................................................................

 

6

 

3.1

OCTOPUS Perimeter 300 ...................................................................................................................

 

6

 

3.2

Instrument Table .................................................................................................................................

 

7

 

3.3

Installation ...........................................................................................................................................

 

8

4

SOFTWARE .................................................................................................................................................................................

 

10

 

4.1

Installation, Exchange, Upgrade .......................................................................................................

 

10

 

4.2

Releasing Program Options ..............................................................................................................

 

10

5

PRINCIPLE OF OPERATION ......................................................................................................................................................

 

12

 

5.1

Functional Diagram ...........................................................................................................................

 

12

 

5.2

Stimulus Intensity Control .................................................................................................................

 

12

 

5.3

Background Intensity Control ............................................................................................................

 

14

 

5.4

Reference Point.................................................................................................................................

 

14

 

5.5

Optics Module ...................................................................................................................................

 

15

6

CARE AND MAINTENANCE .......................................................................................................................................................

 

16

 

6.1

Cleaning ............................................................................................................................................

 

16

7

MAINTENANCE AND REPAIR ....................................................................................................................................................

 

17

 

7.1

Housing (1805001)............................................................................................................................

 

17

 

7.2

Optical Unit (1802669) ......................................................................................................................

 

18

 

7.3

Z-Shaft (1802190) .............................................................................................................................

 

19

 

7.4

Fixation Target (1802206) .................................................................................................................

 

20

 

7.5

Correction Lens Holder (1007496)....................................................................................................

 

22

 

7.6

Background Module, Lamp house (1802247) ...................................................................................

 

26

 

7.7

Brightness Regulator Module (1802248) ..........................................................................................

 

27

 

7.8

Stimulus Projector OCTOPUS 301(1802249)/ Stimulus Projector OCTOPUS 300 (1802544)........

31

 

7.9

Touch Screen Module (new version with TFT display) – Part No. 1802870.....................................

 

32

 

7.10

Lift-/Swivel Unit (1802241) ................................................................................................................

 

34

 

7.11

Headrest Assembly ...........................................................................................................................

 

36

 

7.12

Power Supply (1802199)...................................................................................................................

 

38

 

7.13

Processor Board I (1802431), Processor Board II (1802654) ..........................................................

 

38

 

7.14

SDRAM Module (1802152) ...............................................................................................................

 

44

 

7.15

Interface Board (1802224), Ethernet Interface Board (1802671) .....................................................

 

47

 

7.16

Connector Board (1802213)..............................................................................................................

 

48

8

SERVICE PROGRAMS ................................................................................................................................................................

 

49

 

8.1

Subassembly Test.............................................................................................................................

 

49

 

8.2

Service/Diagnostic.............................................................................................................................

 

50

 

8.3

Printer ................................................................................................................................................

 

56

 

8.4

Dongle Codes....................................................................................................................................

 

56

 

8.5

Light Intensity Calibration ..................................................................................................................

 

56

 

8.6

Entering Parameters .........................................................................................................................

 

56

 

8.7

Verifying Calibration Parameters ......................................................................................................

 

57

 

8.8

Instrument Information ......................................................................................................................

 

58

9

SYSTEM MESSAGES..................................................................................................................................................................

 

59

 

9.1

Messages ..........................................................................................................................................

 

59

 

9.2

Errors.................................................................................................................................................

 

59

10

APPENDIX....................................................................................................................................................................................

 

62

 

10.1

Technical Data ..................................................................................................................................

 

62

 

 

Page 2 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

1

PARTS DELIVERED

 

1 OCTOPUS perimeter 300

1802238

1

Set of accessories containing:

1802250

 

1

Patient response button

1802032

 

1

Dust cover

1802304

 

1

User's Manual

1802737

 

1

Ocular cover

1800339

 

3

Touch pens

1802303

 

1

Eye occluder (2 pieces)

1800339

 

2

FuesT3.15 A / 250 V

1801326

 

1

Allen (hexagon) wrench 2.5mm

1802338

 

1

Screwdriver

1802345

 

1

USB printer cable

1802347

 

1

Octopus VF Digest

1800092

 

1

FTP Server Installation

1802742

OCTOPUS 300 Service Manual

Page 3 / 64

Tradition and Innovation

2 SAFETY INSTRUCTIONS

2.1 General

This instrument can only be used for the purpose as described in this manual.

The examination of the patients, the operation of the instrument and the interpretation of the results can only be performed by persons who are experienced and who have been trained accordingly.

The instrument is to be installed on the height-adjustable table and employed in a dimly lit room in a medical area.

Keep this User's Manual there where persons who operate the instrument can consult it at all times. Guarantee claims can only be valid when the instructions in the User's Manual are followed.

The doctor or operator is to instruct the patient in the safety instructions and check that they are observed.

Make sure that the instrument is connected only to the power source as defined on the perimeter rating plate. The on/off switch does not separate the perimeter from the power lines. Before maintenance or cleaning work is performed, the power cord must be removed from the wall socket.

Removal of the housings and repairs to the instrument are permitted only by trained and authorized technicians. Considerable danger for the operator and patients can be caused by improper repairs.

Only original replacement parts and original accessories are permitted for repairs.

2.2 Instrument Transportation

Transport the instrument over large distances in the original packing. For short distances the instrument can be lifted using the two lower housing sides. Two ribbed grips are provided on the left and right sides which prevent from slipping sideways.

Do not use the forehead rest of the perimeter as a carrying handle.

This plastic part is not adequate for the weight and can therefore break.

Page 4 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

2.3 Symbols

2.3.1 Messages, Errors

Messages, information

Error messages

Warning

Warning against hazard

Special Notices in the Text

Further safety instructions are put into the text as needed. They vary in importance and should help preclude risks for the patient and operator, hinder damage to the perimeter and insure optimal functioning.

ATTENTION: Instruction which must be followed to preclude risks to the patient and operator.

WARNING: Instruction which must be followed to hinder damage to the perimeter.

TIP: Tip for insuring optimal functioning of the perimeter.

OCTOPUS 300 Service Manual

Page 5 / 64

Tradition and Innovation

3 INTRODUCTION

3.1 OCTOPUS Perimeter 300

The OCTOPUS 300 s a direct projection perimeter for examinations of the central visual field (30°). It is a stand-alone system, which means, the examination and control units are integrated into the instrument.

6

1

9

2

 

10

3

 

7

4

8

5

 

11

Figure 1

12

13

14

1Optical unit with upper housing

2Lower housing (left and right)

3Grip locations (left and right)

4Rotation and height adjustable instrument column for fine positioning

5Headrest with

6Forehead rest with integrated sensors

7Chin rest

8Turning knob for chin rest positioning (rough positioning)

9Ocular

10Trial lens holder with IR eye illumination

11Connector for Patient response button

12Turning knob for image focusing

13Operating unit with LCD monitor and Touch module

14Base with Connector panel and Power Supply

Figure 2

Page 6 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

16

17

18

19

20

21

22

15

Figure 3

15Connector panel with

16Power switch

17Power connector

18Printer connector (USB)

19Serial interface (RS 232)

20Ethernet connector (OCTOPUS 311)

21Ethernet control lights

22Contrast LCD monitor

ATTENTION:

All external devices connected to the connector panel must comply with the corresponding safety

standards.

 

 

 

3.2 Instrument Table

An electrically adjustable instrument table (optional) makes it possible to comfortably adjust the height of the instrument to the individual size of the patient.

1

 

 

 

1

Table top with holder for patient response button

2

2

Electronics box

3

Patient response button

 

3

4

Electrical column

 

5

Table foot

4

 

 

5

 

 

Figure 4

Up / Down switch on the electronics box

ATTENTION: With patients in wheelchairs, care must be taken when lowering the table that the table plate does not come into contact with their legs.

Figure 5

OCTOPUS 300 Service Manual

Page 7 / 64

Tradition and Innovation

3.3 Installation

3.3.1 Instrument Table

The instrument table is delivered in a separate package. Utilize the instructions included with the table to put the table together and take care to select the correct voltage before connecting the power cord.

6

7

Figure 6

6Voltage selection switch on the electronics box

7Power connector for attaching the OCTOPUS 300

WARNING: Be sure to set the correct voltage before inserting the power cord connector.

Power connector socket with fuse holder on the electronics box

Figure 7

3.3.2 OCTOPUS 300

Handle the instrument using both lower housing halves to lift it out of the packing. Two ribbed grips hinder sideways slipping.

Do not use the forehead rest as a carrying handle during unpacking or transportation!

Stirnstütze beim Auspacken oder

Transportieren nicht als Traggriff benutzen!

Ne pas utiliser l'appui-tête comme poignée pendant le déballage ou pour transporter!

Since the OCTOPUS 300 works without a cupola, a fully darkened room is not required. In order, though, to make the examination conditions pleasant for the patient and for obtaining reliable results, the instrument is to be placed in the room so that no direct light falls on the instrument or the patient.

The positioning between the patient and the operator or the operating panel can be so chosen that the room conditions are optimally used.

Page 8 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

The patient sits across from the operator

Figure 8

The patient sits on the operator's left

The patient sits on the operator's right

Loosen the three Allen head screws roughly 2 turns counterclockwise on the perimeter of the headrest holder (Allen wrench is included with the accessories) and turn the headrest together with the optical unit into the desired position. Tighten the three Allen head screws again. A complete rotation around the column is blocked by a built in limit.

Place the instrument on the table so that the opening of the table foot and the headrest are in the same direction.

Connect the ‘Patient response button’ to the connector located on the support of the headrest.

 

The connector housing for the ‘Patient response

 

button’ is accessible from below on the headrest

 

support. The catches on the Patient response button

 

connector are oriented in the direction of the

 

instrument column.

 

 

 

ATTENTION: Besides the ‘Patient response

 

button’ no other cable should be

 

attached to the RJ11 connector.

 

 

 

Headrest support with connector

 

Instrument column

Figure 9

Connector with catches

 

Stick the connector plug so far into the connector housing that the catches audibly click. To remove the ‘Patient response button’, press the catches in the direction of the headrest and pull out the cable downwards.

If the examination data are to be transmitted to a PC, connect the OCTOPUS 300 and the PC with the serial connection cable.

Plug the power cord in. The built in ‘Power supply’ works with the voltage specified in Chapter 10.1 'Technical Data'. A change of voltage on the instrument is not required. If an instrument table was delivered with the unit, the OCTOPUS 300 can be connected using the power socket in the electronics box of the instrument table.

OCTOPUS 300 Service Manual

Page 9 / 64

Tradition and Innovation

4 SOFTWARE

4.1 Installation, Exchange, Upgrade

For application software installation and upgrade refer to Chapter 7.14.1 'SIM Flash Module, Application Software (1802222).

4.2 Releasing Program Options

Some of the functions of the OCTOPUS 300 are offered as options and are not accessible in the basic version of the software. Basically all the functions are present in the software, but some can only be made accessible with a code. This code is delivered by HAAG STREIT and is then entered by the customer or the service technician into the instrument. Information about such additional functions can be obtained directly from HAAG STREIT or from your representative.

4.2.1 Procedure

Dongle codes

Read out software master dataActual dongle code

Serial no.

Software version

Enter new dongle code. The keyboard is only released, when the serial number of the instrument is stored in the memory.

To print out the complete instrument information see Chapter 8.8 'Instrument Information'.

Procedure to make additional functions accessible in the perimeter.

 

 

IMPORTANT:

The 'Dongle code' function is only used, when additional program options have been purchased. To avoid

 

 

 

 

 

free access and faulty manipulations the function is protected. The 'Dongle' button is only released after the

 

 

 

 

 

two headrest sensors and the 'Dongle' button are pressed at the same time. As from Serial No. 1031 the

 

 

 

 

 

head rest sensors have been replaced with non-contact head sensors. In this case hold your hand (close) in

 

 

 

 

 

front of the head rest and simultaneously press the ‘Dongle’ button.

 

 

 

 

 

 

 

 

 

 

 

Customer

 

Sends the serial no., software version and the complete dongle code to HAAG STREIT or the

 

 

 

 

 

 

 

 

 

 

 

representative and orders the desired supplementary function(s).

 

 

 

 

HAAG STREIT AG

 

Based on the serial no. and the present and ordered option(s), the new dongle code is computed

 

 

 

 

 

 

and given to the customer.

 

 

 

 

Customer

 

Enters the complete new dongle code into the OCTOPUS 300.

 

 

 

 

 

 

 

 

 

 

 

Open dongle code tab.

 

 

 

 

 

 

Click on 'Dongle code' entry field (numerical keyboard will be displayed).

 

 

 

 

 

 

Enter complete dongle code into the three entry fields.

 

 

 

 

 

 

Close keyboard with 'Esc'.

 

 

 

 

 

 

Save new dongle code by clicking on

 

 

 

 

 

 

Before the instrument saves the setting, the parameters are verified by the system. A faulty

 

 

 

 

 

 

dongle code will not be saved. The message 'Dongle Code not accepted' is displayed on the

 

 

 

 

 

 

screen. Delete wrong code and enter it the correct way.

 

 

 

 

OCTOPUS 300

 

The new additional functions are accessed via the graphical user interface.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 10 / 64

 

OCTOPUS 300 Service Manual

 

 

Tradition and Innovation

4.2.2 Entering the Dongle Code

 

D o n g l e

 

c o d e

 

 

 

 

 

 

 

 

 

 

1234567890

 

1234

 

 

123

 

 

 

 

 

 

 

 

 

4.2.3 The Results of Code Mistakes

The complete dongle code consists of a multi-digit three part series of numbers (see example on the left). Entering the code is done via 'Setup' – 'Service' – 'Dongle'

IMPORTANT: The additional functions are only made accessible when the dongle code is entered without any mistakes into the instrument it was created for.

Every dongle code is unique and only valid for a specific instrument. Mistakes made while entering it or the entering of it into the false instrument (false serial no.) have the consequence that only the basic functions are then accessible. Changing to the old code is possible at all times.

OCTOPUS 300 Service Manual

Page 11 / 64

Tradition and Innovation

5 PRINCIPLE OF OPERATION

HAAG-STEIT is the only company offering perimeters with direct projection systems (OCTOPUS 300, 1-2-3). By means of the

‘Optical unit’, which is replacing the cupola, stimuli are directly projected into the patient eye. This technique does not require dark room condition to take a visual field.

5.1 Functional Diagram

 

5

6

4

 

 

 

1

2

 

3

 

 

 

 

 

7

 

 

 

8

14

 

13

12

 

 

 

 

 

9

15

 

 

11

 

 

 

 

 

 

10

 

1

Patient eye

9

Beam splitter 45°

 

2

Trial lens holder

10

Background illumination

 

3

Ocular

11

Electronics

 

4

Image plane

12

Fixation targets

 

5

Beam splitter 37°

13

Display/Touch module

 

6

Stimulus projection unit

14

External PC

Figure 5-1: OCTOPUS 300

7

Stimulus LED

15

External printer with USB interface

Functional diagram

8

Condenser lens

 

 

5.2 Stimulus Intensity Control

The required dynamic range of the stimulus intensity of 40 dB (1:10'000) is achieved with a combined control mode:

a)Two point current control (I)

b)Pulse-width modulation (S)

Figure 5-2:

Principle of

 

stimulus generation

I (LED)

S (MOD)

Stimulus LED

Page 12 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

5.2.1 Current Control

Two current reference values (operating points) are ascertained during the stimulus calibration procedure:

 

Operating point OP1:

0dB

(max. stimulus intensity)

 

Operating point OP2:

20dB

(max. stimulus intensity/100)

Thus the entire dynamic range is divided into two segments:

 

Segment 1 / operating point OP1:

20dB ...

0dB

 

Segment 2 / operating point OP2:

40dB ...

20dB

I (LED)

I (LED) OP1

Operating point 1

I (LED) OP2

Operating point 2

L (asb)

Lmin = 40dB L max/100 = 20dB L max = 0dB

Figure 5-3: Stimulus LED current control

5.2.2 Pulse-width Modulation

A typical stimulus of 100ms duration is formed by a burst of 20 5ms pulses. The width (ratio LED switch-on/LED switch-off time) of the 20 pulses can be increased or decreased by processor control in steps of 0.4 s (12'500 steps for 5ms). With a longer switch-on time the LED is emitting light for a longer time period, resulting to the human eye, which has an integrating characteristic, in an impact of higher brightness. Pulse-width and light emission are proportional. Pulse-width modulation and stimulus duration are controlled by the microprocessor software through logic timer signals.

OP1

OP2

I (LED)

Stimulus duration = 100ms (20 pulses)

Pulse of 5ms

S (MOD) = on

S (MOD) = off

 

 

 

darker

 

brighter

 

darker

 

brighter

 

darker

 

brighter

 

 

 

 

 

 

 

 

 

 

 

 

Pulse 1

 

 

Pulse 2

 

 

Pulse 20

 

 

t

OCTOPUS 300 Service Manual

Page 13 / 64

Tradition and Innovation

Figure 5-4: Stimulus LED pulse-width modulation

5.3 Background Intensity Control

The light source for the background illumination consists of a number of white LED’s. During initialization of the perimeter the background brightness is measured by a photo sensor and controlled by a DAC value. Pulse-width modulation is not used for background illumination.

5.4 Reference Point

During initialization of the perimeter the stimulus projection unit is centered to the optical system (position x/y = 0/0). The photo sensor for background illumination calibration is used as reference point.

Page 14 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

5.5 Optics Module

 

1

Fixation targets

 

 

 

 

Light measurement

2

 

3

 

 

 

 

CCD camera

 

 

4

 

 

17

5

Stimulus projection unit

7

8

9

6

10

11

Ocular

Background illumination

13

12

14

16

15

 

 

1

Fixation targets

10

Stimulus LED

 

 

2

Fixation target objective lens

11

Condenser lens

 

 

3

CCD camera chip

12

Deflection mirror

 

 

4

Camera objective lens

13

Deflection mirror

 

 

5

Ocular lenses

14

Scatter pane

 

 

6

Beam splitter 37°

15

Background LED’s

 

 

7

Field lens

16

Compensation filter

 

 

8

Beam splitter 45°

17

Light measurement unit

Figure 5-5:

Optics module

9

Stimulus objective lens

 

 

 

 

 

 

OCTOPUS 300 Service Manual

Page 15 / 64

Tradition and Innovation

6 CARE AND MAINTENANCE

6.1 Cleaning

It suffices to dust the instrument periodically with a soft cloth. More obstinate dirt particles can be removed using a soft cloth dampened slightly with water or alcohol.

ATTENTION: Avoid making the instrument wet and never use any solvents.

6.1.1 Patient Response Button, Chin and Headrest, Turning Knobs and Eye Occluder

All of these parts are made from plastic which can be cleaned with no problems. In order to keep them hygienically clean, they should be cleaned periodically with a cloth or cotton dampened with alcohol.

6.1.2 Ocular

Finger prints and dust can be removed with a moist soft cloth.

6.1.3 Monitor Screen, Touch Module

Finger prints and dust can be removed with a moist soft cloth.

Page 16 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

7 MAINTENANCE AND REPAIR

WARNING: Before opening the instrument make sure that the mains cable is separated from the instrument.

WARNING: Some of the modules need alignment after being exchanged. Therefore it is important to go through all the steps described in exchanging the module that needs to be replaced.

WARNING: Electronic circuits of the OCTOPUS 300, particularly the 'Processor board', the 'Interface board' and the LEDs in the 'Background LED module' are all very sensitive to electrostatic charges. Keep all modules in static shielding bags for storage, transportation and shipping.

Besides an ordinary set of regular tools the following tools and instruments are recommended:

Set of Allan- (hexagonal) wrenches, metric 1.3 to 8mm

Set of Torx screwdrivers (the sizes in this manual refer to the screwdriver size)

Universal digital voltmeter, range 200mV to 250V

Cleaning Materials

Can of compressed air

Lens tissue

Lens cleaner (diluted alcohol)

Cotton swabs

Whenever possible, field repairs of the OCTOPUS 300 should be restricted to module exchange. The sub module (part) causing a problem should be localized with the help of the service manual and error list. It should then be replaced with an exchange module.

7.1 Housing (1805001)

Housing left

Housing right

Housing top

NOTE: If the 'Housing' is removed, the perimeter will only initialize and calibrate correctly if it is placed in an entirely dark environment. The covers can temporarily be put back in place for its operation and to run the 'Service programs' such as the 'General test' in a semi-dark room.

Removal

Remove the 'Housing top' by removing the four Phillips (M4) screws in each outer corner from the bottom part of the 'Housing left' and 'Housing right' (see Figure 7-1).

OCTOPUS 300 Service Manual

Page 17 / 64

Haag-Streit Octopus 300 Perimeter Service manual

Tradition and Innovation

Housing top

Housing right

Screws to remove 'Housing top'

(in the same position on both sides)

Screws to remove Housing right

Housing left

Figure 7-1

Once the screws have been removed, simply lift the 'Housing top' and place it on a clean surface.

Remove the 'Housings left' and 'Housing right' by removing the 4 Phillips (M4) screws in each inner corner from the bottom of the 'Housings left-/right' (see Figure 7-1). The Optical unit can be swung in a left/right movement for easy access. Remove the 'Housing left-/right' by gently lifting both parts out of the instrument. Place both parts on a clean surface.

Reassembly

Reassemble in reverse order to removal.

Check

Once reassembled check that the screws have been tightened and that the 'Housing' is clean.

7.2 Optical Unit (1802669)

Removal

Remove the 'Housing' (see Chapter 7.1).

Remove the bottom 5 connectors of the 'Interface board'. Remove the screw and washer holding the 'Grounding cable' connected to the 'Optical unit'.

On the 'Processor board' disconnect the cable 'Switches' (6 pin) and the cable 'LCD' (50 pin).

Connector 'Switches' (6 pin)

SD RAM module

Position of the cable 'LCD' (50 pin)

Figure 7-2: Processor board

Page 18 / 64

OCTOPUS 300 Service Manual

Tradition and Innovation

SIM Flash module

Processor board

LCD cable

Fixation target

Interface board

Swivel ring

Figure 7-3

Cut the tie-wraps, holding the cables to the 'Optical unit'.

Remove the 'Touch screen module' (see Chapter 7.9), then remove the 'Power connector' from TB2 (see Chapter 7.12).

Remove the 4 nuts holding the 'Optical unit' to the 'Z-shaft'. See Figure 7.4.

Nuts holding the 'Optical unit' to the 'Z-shaft'

Figure 7-4

Gently lift out the 'Optical unit', at the same time glide the 'Power connector' through the slot in the 'Z-shaft'. Be careful not to damage any wires. Place the unit on a flat, dust-free surface.

Reassembly

Reassemble in reverse order to removal.

Check

Are all cables connected?

Are the four nuts holding the 'Optical unit' to the 'Z-shaft' back in place and tightened correctly.

Have all the cut tie-wraps been replaced by new ones.

After reassembly, run 'General test' to verify faultless operation. Regarding the 'Service Programs', refer to chapter 8.

7.3 Z-Shaft (1802190)

OCTOPUS 300 Service Manual Page 19 / 64

Tradition and Innovation

Removal

Remove 'Housing' and 'Optical unit' as described in Chapter 7.1 and 7.2.

Remove the 3 torx screws (size 25) which are fastened down onto the 'Lift-/ swivel unit' from the top. See Figure 7-5.

3 torx screws (size 25).

Figure 7-5

Gently lift up the 'Z-shaft', at the same time thread the connectors that were loosened in Chapter 7.1 and 7.2 through the slots in the 'Z-shaft'.

Place the 'Z-shaft' on a clean surface.

Reassembly

Reassemble in reverse order to removal.

Check

Before mounting the 'Optical unit' be sure that all the connectors and cables have been passed through the slot in the 'Z-shaft'.

7.4 Fixation Target (1802206)

7.4.1 Fixation Board (1802207)

NOTE: If the 'Fixation board' is defective there is no need to replace the entire 'Fixation target', one can simply remove the 'Fixation board' from the 'Fixation target' by following the steps below.

Removal

Remove the housing (see Chapter 7.1)

Remove the 10 pin flat cable from the 'Fixation board'.

Remove the torx screw (size 15) holding the 'Fixation target' to the 'Optical unit'. Simply lift the 'Fixation target' out of the two guide pins.

Remove the two hexagon socket screws (wrench size 2,5) and washer holding the 'Fixation board' to the 'Fixation target'.

The 'Fixation board' can simply be removed.

Reassembly

Reassemble in reverse order. Before remounting the 'Housing', go through the check procedure below.

Page 20 / 64

OCTOPUS 300 Service Manual

Loading...
+ 44 hidden pages