As in every field, technology moves fast in the area of electromedical equipment.
Both electronically and medically, immense progress has been made over
the last couple of years.
Page 1
It goes without saying that comprehensive technical knowledge of the
product is needed more than ever.
This manual aims to meet this ever increasing need by giving you a
complete and accurate picture of the pulson 330, an advanced microcomputer controlled ultrasound apparatus.
In doing so, it will hopefully help you to reach your goal: to form a
correct diagnosis and to solve the client’s problem as thoroughly as
possible.
If you have any questions or if you need additional information about
this manual or about the use of the pulson 330, please do not hesitate
to contact us.
Page 2
3. Introduction
The Pulson 330 is a modern micro-controlled apparatus, intended for
ultrasound therapy.
It is a member of the Gymna family, and thus built according to the
newest requirements concerning the European security standards.
The unit is portable and can therefore be used for in-house treatments
as well as for domestic use.
The unit provides following possibilities: continuous and pulsed
ultrasound at frequencies of 1 Mhz or 3 Mhz.
Combination therapy is possible and useful with following Gymna
apparatus:
Interdia 200
Interdia NST
Duo 400
Duo 410
Two different or equal treatment heads can be connected at the same
time, with the possibility to switch between.
Following heads are available:
1 Mhz head 5 cm²
3 Mhz head 0.8 cm²
The heads are insulated against parasitic side radiation and
waterproof, which makes them suitable for underwater treatments.
Thanks to a built in efficiency resistor, the heads are fully
interchangeable, without the need of recalibration of the output
stage.
The treatment heads are also interchangeable with the Combi 410 unit .
Page 3
4. Important remarks
4.1 Safety aspects
To understand and practice all procedures described in this manual a
good technical background is a must.
Gymna will not be held responsible for any actions executed on the
unit by unauthorized persons, or for executing any procedures not
prescribed in this manual.
All information in this manual has a Gymna copyright.
4.2 Data registration.
The distributor must be able to provide the following data for each
unit:
• Instrument data: Part and serial number.
Gymna has the original configuration of each unit.
The Pulson 330 has following critical parts:
Main board: part number: 10555
Display board: part number: 10578
Output stage board: part number:10558
Main power transformer: part number:9677
Inlet filter: part number:8552
These parts have a serial number which has to be filed in case of a
configuration
change.
• Customer data: Name, full address and date of delivery.
Service activities: All service activities must be filed.
If any critical parts are changed, we also expect the distributor
to file
the new serial numbers of these part(s).
To file this data, a data registration document is available on next
page.
Page 4
Data registration document
Information
Distributor name Distributor address
Customer name Customer address
Instrument name: Pulson 330 Serial number:
Date of delivery:
Service activities
Date:
Error complaint:
Service action
Critical spare parts exchange list
Critical part name Old serial number New serial number
PCB Display.
PCB Main.
PCB.Ultrasound
Power transformer.
Ac inlet.
Date:
Error complaint:
Service action
Critical spare parts exchange list
Critical part name Old serial number New serial number
PCB Display.
PCB Main.
PCB.Ultrasound
Power transformer.
Ac inlet.
Page 5
5. General technical data.
• Treatment unit
Mains voltage: 115/230 Vac +- 10%
Frequency: 50/60 Hz
Mains fuses: 1A type L
Current consumption: maximum 500 mA
Dimensions 413 ✕ 280 ✕ 100 mm
Weight: 5 kg.
Labels Hazard warning label
Serial number label
Type identification label
Fuse value label
Bauartzulassungs Nr.
Manufacturer, mains, Voltage
Frequency and power consumption
Class according to IEC601-1: I
Type according to IEC601-1: BF
Limits test values according to IEC 601-1:
Earth leakage current: IEC requirement < 500 µA
Earth leakage single fault: IEC requirement < 1 mA
Enclosure leakage current normal: IEC requirement < 100 µA
Enclosure leakage single fault: IEC requirement < 500 µA
Patient leakage normal: IEC requirement < 100 UA
Patient leakage single fault: IEC requirement < 500 µA
Patient leakage mains on applied part sfc: IEC requirement < 5
1. US- power display. The intensity is displayed in W/cm².
2. Treatment time display.
Displayed in Min & Sec. Maximum treatment time is 30 minutes.
The timer runs only when the contact control is sufficient.
3. Intensity up key. Maximum intensity is 2 W/cm•.
4. Intensity down key. Minimum intensity is 0.1 W/cm•
5. Output mode switch.
There are 4 output modes
1 continuous
3 pulsed
The modulation frequency is fixed at 100 Hz for all pulsed USoutputs.
There is only a difference in duty cycle.
Page 8
6. 1/3 MHz head selection key.
Is a key to switch from the 1MHz to the 3MHz head and back.
For a good understanding, some important notes have to be taken into
consideration:
The right side panel has 2 US-output connections, CON2 and CON3.
The pulson 330 is designed in such a way that it always detects
which head is
connected to which output gate.
In other words, both heads can be connected to both outputs, and it
is of no
importance whether one ore two heads are connected.
The following examples may illustrate this:
If both heads are connected and you switch on the instrument, it
will always
select the 1MHz head. If only the 3MHz head is connected, there
is no other choice,
so this one will be selected.
If you want to switch from 1MHz to 3MHz, there are two
possibilities:
The 3MHz is connected.
Press the 1/3 MHz and the switch is made.
The 3MHz is not connected.
The instrument doesnít find the other head. The 3MHz led
flashes for 10 sec.
after which the 1MHz is selected again.
If only the 1 MHz is connected, and you pull out this head, both
leds start flashing
for about 10 sec., after which they extinguish.
If two heads are connected and you pull ot the 1MHz while using it
for a treatment,
the instrument will immediately switch off the US-power and
select the 3MHz head.
7. Timer down switch.
8. Timer up switch.
The treatment time can be changed in steps of a min. if time > 1
min. If the treatment
time < 1 min. the timer can be changed in steps of a sec.
Page 9
8. Right side panel.
7. Output connector 1 (CON 1). Can be connected to a 1/3 Mhz. head.
8. Output connector 2 (CON 2). Can be connected to a 3/1 Mhz head.
9. Mains entrance.
The pulson 330 must be connected to a wall socket with a earth
contact.
Use only the original cable that is supplied with the apparatus in
the
accessories package.
Check whether the required supply voltage, mentioned on the serial
number plate,
corresponds with the available mains supply ratings.
10. Serial number plate.
On this plate the serial number, as well as the required supply
voltage and
frequency are mentioned.
Also the mains fuses ratings are mentioned.
11. Plug for combination therapy.
Via this plug, the pulson 330 can be linked with another GYMNA
apparatus
in order to perform combination therapy.
12. Power switch.
Will light up when the power is switched on.
13. Mains fuses. 2 Fuses of 1 A type L.
Page 10
9. Power supply.
Main fuses are inserted in the main entrance ( F1 and F2, 1A fast ).
The transformer contains a thermic fuse . It will cut off the primary
winding, when the inside temperature exceeds 120 °C. To switch on
again, transformer temperature has to go below about 60 °C.
For this reason, there is no need for secondary fuses anymore.
The ringcore transformer can be fed either with an incoming voltage of
230V/50Hz ( CON 6 ) or with 115V/60Hz ( CON 5).
There are 2 secondary windings. Output voltage of winding 1 is 22V,
winding 2 produces 8V.
Power supply output voltages:
1. VCC: Is a 5V stabilised voltage, with a maximum output current of
700 mA.
D22 protects U12 during switch off.
2. VEE: Is a -5V stabilised voltage, with a maximum output current of
100 mA.
D21 protects U13 during switch off.
3. V27: Is a non stabilised power supply of approximately 27V.
VLK3: analog ground, connected to the metal cover-shield of pcb 2.
VLK4: analog ground, not connected and can be used for service.
RV1: Varistor. Filters out spikes on the incoming voltage.
Page 11
10. Microcontroller circuits.
10.1 The microprocessor (µp).
The pulson 330 uses a Z86E21, a One Time Programmable (OTP)
microcontroller which controls all sequences necessary for the
apparatus to function.
It has on board ROM (8K), and RAM. Clock speed is defined by an
external crystal, which runs at 12 MHz.
10.2 Microprocessor guard components.
To monitor the proper functioning of the µp, some external components
are added.
1. U2 (MAX1232): µp supervisor IC.
Does constant monitoring of VCC. When it goes below 4.75 V, a
pulse with a fixed length is produced, to assure a good reset
of the µp.
R7 is a pull-up resistance and C3 filters out burst signals
which may possibly disturb the µp.
U2 also monitors the watchdog output of the µp. In case of
processor hang-up,
the same reset pulse is given.
2. U6 (74HC4538): output circuit protection IC.
In case of processor hang-up, the output stage has to be
switched off
immediately.
When the output VUS_EN goes low/high, Q10 switches and AMP_US
goes to
zero. This IC is also controlled by the watchdog output of the
µp.
10.3 I≤C bus
The combination of outputs SCK and SCD forms a serial bus which
follows the I•C protocol.
A maximum of 10 components, each having their own address, can be
connected.
In this case 3 components are connected.
1. U10 (24C16)
This is a 2K eeprom, which can be written during the special
service routine.
It contains the calibration data of the output stage and a
copy of the checksum from the µp ROM. Each time the instrument
is switched on, the µp calculates
the ROM checksum and compares it to the value in the eeprom.
In case of difference an error code will be displayed. Refer
to error codes:
2. U15 (PCF8591)
ADDA IC. Data coming from and going to this IC are sent by the
I•C bus.
For more information refer to US power circuits, page
3. Display.
Connection to the display pcb is made by CON7.
All display data arrive via this bus.
Page 13
10.4 Relay 4 (RLY4).
After the instrument has been switched on , or the 1/3 MHz-key has
been pressed, the µp checks the head sizes of all heads connected.
RLY4 closes the HEAD_SIZE line, and reads the size of all heads
connected to the apparatus.
After reading, RLY4 opens again. This is done in order to obtain a
galvanic separation of
the outputs in case of combination therapy.
10.5 Microprocessor (µp) pin description.
• USD_UP_KEY Input signal. Switches to logic low level when
the ultrasonic density up key is pressed.
• USD_DOWN_KEY Input signal. Switches to logic low level when
the ultrasonic density down key is pressed.
• SELECT_LEFT_KEY Input signal. Switches to logic low level when
the left output mode key is pressed.
• SELECT_RIGHT_KEY Input signal. Switches to logic low level when
the right output mode key is pressed.
• N.U. Not used.
• FREQ_ 1_3_MHZ_KEY Input signal. Switches to logic low level when
the 1/3 MHz. key is pressed.
• TIME_DOWN_KEY Input signal. Switches to logic low level when
the time down key is pressed.
• TIME_UP_KEY Input signal. Switches to logic low level when
the time µp key is pressed.
• WATCH_DOG Output signal. In case of µp hang up, this
signal is at logic high level.
• KEY_SELECT Output signal. Logic low enables hardware
information to
be read in the µp. Logic high level disables
hardware
information and connects the keyboard to the µp.
• VUS_EN US voltage enable. Low/high active. Disables
output stage
in case of µp hang up.
• SCK Serial clock. Clock signal for I•C- bus.
• SDA Serial data. Data signal for I•C- bus.
• SCK_ADDA Clock signal for I•C -bus, connected only to U15
(ADDA)
• SDA_ADDA Data signal for I•C-bus, connected only to
U15(ADDA)
• HEAD_SIZE Input signal. This line is read during
initialising, to let the µp
know the head sizes of the heads connected.
• SERVICE_REQUEST High active. When a jumper is placed on CON1 ,
this line
switches to logic high level and the service
routine is activated.
• ON/OFF Output signal: When this signal is at logic high
level, the
1/3 MHz signals are cut off, which results in 0
W output. This signal will switch on/off at 100 Hz in
case of combination therapy.
• 1/3 MHZ Output signal. When this signal is at logic high
level, the
3 MHz base frequency is selected.
Page 14
• SIZE_DET Output signal, high active. Activates RLY4
during start-up in order to detect head size of all heads
connected to the
instrument.
• US 1/3 Output signal. In case of logic high level, RLY1
is
activated. At this time, RLY 3 , which is
always in opposite position, is inactivated.
• CON 1/2 Output signal. High logic level selects output
connector 2
( CON2 ).
•
RESET Low active. In case of µp hang-up, or VCCfailure, U2 creates a reset pulse.
• BUZZER Buzzer sounds when the output is at logic high
level.
Page 15
Loading...
+ 34 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.