GymnaUniphy Myaction Service manual

SERVICE INFORMATION Myaction FIT/SPORT/MED
V01 (non-controlled version)
Myaction FIT/SPORT/MED service manual V01 Page 1
1. TABLE OF CONTENTS
1. TABLE OF CONTENTS................................................................................................2
2.1 Introduction. ............................................................................................................................3
3. Safety aspects..................................................................................................................4
3.1.1 Safety............................................................................................................................4
3.1.2 Installation. ...................................................................................................................4
3.2 Mains voltage and the use of the rechargeable battery............................................................4
4. Technical specifications..................................................................................................5
5. Most important information............................................................................................ 6
5.1 Display definitions... .... ....... .... ... .... ... .... .... ... .... ... .... ... .... .... ... .... .... ... .... ... ........ ... .... ... .... ... . .......6
5.2 Output connectors....................................................................................................................6
5.3 Keyboard definitions ..................................................... .... ... .... .... ... .... ... .... ... .... .... ... .... ... .... ....7
6. Functional test Myaction FIT, Myaction SPORT and Myaction MED..........................8
6.1 Amplitude................................................................................................................................8
6.2 Pulse duration ..........................................................................................................................8
6.3 Equipment test.........................................................................................................................8
6.4 Equipment check .....................................................................................................................9
6.5 Cables and output socket .........................................................................................................9
6.6 Charging ..................................................................................................................................9
7. Software versions..........................................................................................................10
8. Exchanging boards........................................................................................................11
8.1 Introduction ...........................................................................................................................11
8.2 Opening the unit and replacing the main PCB ......................................................................11
Myaction FIT/SPORT/MED service manual V01 Page 2
2. GENERAL INFORMATION
2.1 Introduction.
The Uniphy Myaction series are portable electrotherapy devices. As a member of the Uniphy family, they are built according to the latest requirements of the Medical Device Directive.
There are 3 different devices: Myaction FIT Myaction SPORT Myaction MED
This service manual gives a complete and accurate picture of the Myaction series. In doing so, it will help you to reach your goal: to form a correct diagnosis and to solve the client’s problem as thoroughly as possible.
If you have any questions, or if you need additional information about this manual or the use of the machine, feel free to contact us.
Bilzen, November 2004
Myaction FIT/SPORT/MED service manual V01 Page 3
3. Safety aspects.
3.1.1 Safety.
See operation manual, chapter: SAFETY
3.1.2 Installation.
Check that the equipment have not been damaged during transportation. In the event of damage and/or defect you should inform your supplier immediately. Do Not use the equipment if it is damaged or defective.
3.2 Mains voltage and the use of the rechargeable battery.
The equipment is battery powered and is therefore provided with an AC/DC adapter. The batteries are Nickel Metal hydrate type with peak efficiency without memory effect. When the battery indicator lights up on the display, the battery needs to be recharged.
Switch off the stimulator and disconnect the electrodes. Connect the originally supplied charger to the battery plug (photo below)
The LED on the charger indicates that the connection is correctly executed. The first charge must be completed without any interruption. A complete recharge takes about 10 hours (overnight). However, the battery will not be damaged when the charger would stay connected for several days.
Only use the original batteries supplied by GymnaUniphy nv.
During the charging, the unit MAY NOT be connected to the electrodes. If the charger is connected with the stimulator switched on, it will automatically be switched off.