User Manual Combi 200L
Device for electrotherapy, ultrasound therapy, combination
therapy and laser therapy
The Combi 200L is intended solely for medical applications. You can use the
Combi 200L for electrotherapy, ultrasound therapy, combination therapy
and laser therapy. The device is suited for continuous use.
1.2Safety instructions
1.2.1General
• Only qualified people who are trained in the application of the
therapies may use the appliance.
• Only a technician authorised by GymnaUniphy N.V. may open
the equipment or the accessories.
• Follow the instructions and directions in these user
instructions.
• Place the equipment on a horizontal and stable base.
• Keep the ventilation openings at the bottom and rear of the
equipment free.
• Do not place any objects on the equipment.
• Do not place the equipment in the sun or above a heat source.
• Do not use the equipment in a damp area.
• Do not let any liquid flow into the equipment.
• Do not disinfect or sterilise the equipment. Clean the
equipment with a dry or moistened cloth. See §5.
• Only treat patients with electrical implants (pacemaker) after
obtaining medical advice.
• The 'Directive on Medical Devices' from the European
Commission (93/42/EEG) requires that safe devices are used.
It is recommended to perform a yearly technical safety
inspection. See §5.1.2.
• For optimum treatment, a patient investigation must first be
performed. On the basis of the findings of the investigation, a
treatment plan with objectives will be formulated. Follow the
treatment plan during the therapy. This will limit possible risks,
related to the treatment, to a minimum.
• Always keep these user instructions with the equipment.
9
Combi 200L
1.2.2Electrical safety
• Only use the equipment in an area with facilities that meet the
applicable legal regulations.
• Connect the equipment to an outlet with a protective earth
terminal. The outlet must meet the locally applicable
requirements for medical areas.
1.2.3Prevention of explosion
• Do not use the equipment in an area where combustible
gases or vapours are present.
• Switch off the equipment when it is not used.
1.2.4Electro Magnetic Compatibility
• Medical electrical equipment requires special precautions for
Electro Magnetic Compatibility (EMC). Follow the instructions
for the installation of the equipment. See §2.
• Do not use mobile telephones or other radio, shortwave, or
microwave equipment in the vicinity of the equipment. This
kind of equipment can cause disturbances.
• Only use the accompanying accessories that are supplied by
GymnaUniphy. See §7.
Other accessories can lead to an increased emission or a
reduced immunity.
10
1.2.5Electrotherapy
• Do not use the equipment simultaneously with high frequency
surgical equipment. This combination can cause burning of
the skin under the electrodes.
• Do not use adhesive electrodes with currents that have a
galvanic component, such as galvanic, diadynamic, MF
rectangular, pulsed rectangular and triangular currents. With
these currents, etching of the skin can occur.
• Check the electrode cables and the electrodes at least once a
month. Check whether the insulation is still intact. See §5.1.
• The safety standards for electrical stimulation advise not to
exceed the current density of 2.0 mA
However, with iontophoresis treatments, we advise a
maximum current density of 0.25 mÂ/cm
the MF rectangular current. Exceeding this value can result in
skin irritation and burns.
• Always use sterilised gauze with iontophoresis treatments.
1.2.6Ultrasound therapy
• Move the US head evenly over the skin during the treatment.
This prevents internal burns.
• The US treatment heads are exchangeable. The device
detects the characteristics and supplies the right power at the
right frequency.
• Handle the US heads carefully. With rough handling, the
characteristics can change. Test the US head if it falls on the
ground or knocks against something. See §5.1.1.
• Check the US head at least once a month. During the check,
look for dents, cracks and other damage that could allow
liquids to ingress. Check whether the insulation of the cable is
still intact. Check whether all pins are present and straight in
the connectors. Replace the US head if the head, the cable or
the connector is damaged. See §5.1.
Combi 200L
/cm2.
rms
2
, because of using
11
Combi 200L
1.2.7Laser therapy
The laser is a class 3b laser product and has an invisible beam.
• Make sure the laser warning sign is clearly visible outside the
entrance to the therapy room.
• The radiation of a laser probe can cause a physiological effect.
• Use the laser therapy only for therapeutic purposes.
• Use of controls or adjustments or performance of procedures
other then those specified in this manual may result in
hazardous radiation exposure.
• Start a laser therapy only when all persons in the room wear
laser goggles for eye protection. If you do not obey this
warning, you can cause blindness. Use goggles with at least
the characteristics: I 100 - 1000 L2 and with a clear view of
the control, the display and the signal lights. See §7.10.
• Do not look into the laser beam during a laser therapy.
• Do not point the laser beam into eyes.
• Do not use the laser near flammable materials or liquids.
• Do not use the equipment if any damage shows.
• Regulary check the output of the laser probe with the test
facility. See §4.9.8.
• Check the laser probe at least once a month. During the
check, look for dents, cracks and other damage. Check
whether the insulation of the cable is still intact. Check
whether all pins are present and straight in the connectors.
Replace the laser probe if the laser, the cable or the connector
is damaged. See §5.1.
• Place the laser probe in the holder when the laser is not used.
• Remove the key of the laser lock when the laser therapy is not
used.
1.3Medical Devices Directive
The device complies with the essential requirements of the Medical Device
Directive of the European Committee (93/42/EEC) as most recently
changed.
12
Combi 200L
1.4Liability
The manufacturer cannot be held liable for injury to the therapist, the
patient or third parties, or for damage to or by the equipment used, if for
example:
•an incorrect diagnosis is made;
•the equipment or the accessories are used incorrectly;
•the user instructions are wrongly interpreted or ignored;
•the equipment is badly maintained;
•maintenance or repairs are performed by people or organisations that
are not authorised to do so by GymnaUniphy.
Neither the manufacturer nor the local GymnaUniphy dealer can be held
liable, in any way whatsoever, for the transfer of infections via the vaginal,
anal and rectal probes and/or other accessories.
13
Combi 200L
14
Combi 200L
2INSTALLATION
2.1Receipt
1.Check whether the equipment has been damaged during transport.
2.Check whether the accessories are intact and complete. See §7.
•Inform your supplier of any damage or defects by no later than
within 3 working days after receipt. Report the damage by
telephone, fax, e-mail or letter.
•Do not use the equipment if it is damaged or defective.
2.2Placing and connection
1.Place the equipment on a horizontal and stable base.
•Keep the ventilation openings at the bottom and rear of the
equipment free.
•Do not place the equipment in the sun or above a heat source.
•Do not use the equipment in a wet area.
2.Check whether the mains voltage that is stated on the rear of the
equipment corresponds with the voltage of your mains supply. The
equipment is suited for a nominal mains voltage from 100 V to 240 VAC
/ 50-60 Hz.
3.Connect the device to an outlet with protective earth terminal.
2.3Performing the functional test
1.Switch the equipment on with the switch at the rear of the equipment.
2.When the equipment is switched on, it automatically performs a test.
Check whether the indicator lamps next to and light briefly
during the test.
3.If the lamps do not light up: See §6.
A
B
2.4Setting contrast and selecting language
1.Press for 5 seconds. The System setting menu appears. See §4.9.
2.Press next to Contrast, 1
3.If necessary, change the contrast with and .
4.Press next to Language.
5.If necessary, change the language with and .
6.Press to return to the start menu.
st
key from the top.
15
Combi 200L
2.5Transport and storage
Take account of the following matters if the equipment has to be
transported or stored:
•Transport or store the equipment in the original packaging.
•The maximum period for transport or storage is: 15 weeks.
•Temperature: -20 °C to +60 °C.
•Relative humidity: 10% to 100%.
•Atmospheric pressure: 200 hPa to 1060 hPa.
2.6Reselling
This medical equipment must be traceable. The equipment, the US head
and some other accessories have a unique serial number. Provide the
dealer with the name and address of the new owner.
16
Combi 200L
3DESCRIPTIONOFTHEEQUIPMENT
3.1Combi 200L and accessories
12
11
10
9
8
7
1.Combi 200L. See §3.2.
2.Power cord
3.Laser goggles. Optional
accessory. See §7.10.
4.Laser probe. Optional
accessory. See §7.10.
5.US head
6.Contact gel
1
2
6
5
7.EL sponges for rubber
electrode (4 pieces)
8.Elastic fixation straps (4 pieces)
9.VAS score card
10. Rubber electrodes (4 pieces)
11. Two-ply electrode cable (2
pieces)
12. Test connector
3
4
17
Combi 200L
3.2Components of Combi 200L
1
2
3
4
5
6
7
8
A
B
AB
12
11
91510
13 14
1.Display. See §3.3.
2.Electrotherapy
3.Ultrasound therapy
4.Combination therapy
5.Laser therapy
6.Memory
7.Start menu
8.Channel selection: A or B
9.Stop
10. Intensity of channel A
11. Pause
12. Laser test eye
13. Return to previous menu
14. Intensity of channel B
15. Enter
16. Indication: Read manual
17. Down
18. Up
20
19
21 22 2324
18
17
16
!
25
A
292631
303227 28
B
19. Select parameter or menu
20. Connector for US head
21. On/off switch
22. Connection to mains supply
23. Type plate
24. Ventilation opening
25. Fuse holder
26. Indication: Floating patient
circuit
27. Laser lock
28. Connector for laser probe
29. Connector for electrotherapy,
channel A
30. Indicator lamp for channel A
31. Connector for electrotherapy,
channel B
32. Indicator lamp for channel B
18
3.3Display
Combi 200L
1234
10
11
9
12
13
1.Channel
2.Therapy
3.Current shape
4.Title of the screen
5.Program number
6.Parameters with selection
knobs
7.Use to go to the next
parameters
15:00
set
red+
0
.
0
0
0
.
0
US+Conv
Pulse time
Pulseform
CV
Min.
Max.
Sweepmode
Treatment
Perceptibe, comfortable
5
.
TENS
frequency
frequency
time
8
8.Explanation or recommendation
9.Screen for channel A (here,
electrotherapy). See §4.3.5.
10. Remaining treatment time
11. Polarity
12. Set intensity
13. Screen for channel B (here,
ultrasound therapy). See §4.4.2.
40 s
80
100
15:00
Hz
Hz
34
6
7
19
3.4Display symbols
3.4.1General
Combi 200L
SEQ
ElectrotherapyChannel A
Ultrasound therapyChannel B
Combination therapy
Laser therapyTreatment time
Sequential current
shapes
A
B
A + B
0:00
Channel A and B
simultaneously
Treatment completed
3.4.2Current shape groups
Unidirectional currents2-pole medium frequency
Diadynamic4-pole Interferential
TENS currents
NMES currentsDiagnostic programs
4-pole interferential with
vector
3.5Symbols for current shapes in memory menu
Medium frequency
unidirectional current
Unidirectional rectangular
current
Unidirectional triangular
current
Conventional TENSBiphasic surge current
Low frequency TENS
Random TENS
20
Burst TENS
Rectangular surge
current
Triangular surge current
Intrapulse interval surge
current
2-pole medium frequency
surge current
Combi 200L
CP
DF
LP
MF
CP (diadynamic)2-pole medium frequency
DF (diadynamic)Dipole vector field
LP (diadynamic)Rheobase and chronaxie
MF (diadynamic)Rheobase and AQ
3.6Parameter symbols
3.6.1Electrotherapy
Red+
Red-
+
Sweep mode
Polarity indicationConstant Current
Alternating polarityConstant Voltage
-
Biphasic pulse shape,
symmetrical
Biphasic pulse shape,
asymmetrical
CHR
AQ
CC
CV
mA
V
mA peak
Volt peak
12
12
6
12s/12s1s/5s -1s/5s
6
6s/6s1s/1s
3.6.2Ultrasound therapy
1
2
3
4
5
10ms
10ms
10ms
10ms
10ms
US duty cycle 10%US duty cycle 100%
10%100%
US duty cycle 20%Set US intensity
20%
US duty cycle 30%Peak US output power
30%
US duty cycle 40%
40%
US duty cycle 50%
50%
11
W cm
21
5
5
1
1
set
P
pk
Unit of the set US
2
/
intensity
3.6.3Laser therapy
Combi 200L
P
E
Set average power
set
Energy per pulseWear goggles
p
MonoprobeClusterprobe
3.7Current shapes
3.7.1Unidirectional currents
Rectangular pulse currentTriangular pulse current
2-5 current (UltraReiz)
5 ms
2 ms
3.7.2Diadynamic currents
MFCP
DFLP
3.7.3Interferential currents
E
tot
energy
Medium frequency
rectangular current
DF
MF
DF
MF
Total administered
2-pole medium frequency
4-pole Interferential
3.7.4TENS currents
Conventional TENS,
asymmetrical
Conventional TENS,
alternating asymmetrical
Conventional TENS,
symmetrical
4-pole interferential with
rotating vector
Conventional TENS,
alternating symmetrical
TENS burst
TENS burst, alternating
22
3.7.5NMES currents
Combi 200L
Rectangular surge
current
Triangular surge current
Medium frequency surge
current (2- and 4-pole)
Biphasic surge current
Intrapulse interval surge
current
23
Combi 200L
24
Combi 200L
4OPERATION
4.1Therapy selection
You can select a therapy in different ways, with the therapy keys or with the
parameters in the Start menu:
•Therapy keys: Quickly select a therapy with therapy keys , , and
. See §4.1.1.
•Objectives: Select a therapy on the basis of an objective. See §4.1.2.
•Indication list: Select a therapy on the basis of a medical indication.
See §4.1.3.
•Program number: Select a certain program number or a program
number that you previously saved. See §4.1.4.
•Diagnostic programs: Perform a diagnosis, for example to determine
the rheobase and the chronaxie. See §4.1.5.
•Contra indications: Display an overview with contra indications for the
different therapies. See §4.1.6.
Besides this, you can change the system settings. See §4.9.
4.1.1Therapy keys
Electrotherapy selection
1.Press : Electrotherapy.
2.Select the current shape group
with .
3.Select the current shape with
.
Ultrasound therapy, direct selection
1.Press : Ultrasound therapy. The Ultrasound screen appears.
Combination therapy, direct selection
1.Press : Combination therapy.
2.Select the current shape with . See §4.5.1.
Laser therapy, direct selection
1.Press : Laser therapy. The Laser therapy screen appears.
25
Combi 200L
4.1.2Therapy selection via objectives
1.Press to go to the start menu.
2.Select Objectives.
3.Select Electrotherapy,
Ultrasound therapy or Laser
therapy.
4.Select the desired treatment
with .
4.1.3Therapy selection via indication list
1.Press to go to the start menu.
2.Select Indication list.
3.Go to the following indications
with or . See §9.1.4.
4.Select the desired indication
with .
•ET: Electrotherapy
•US: Ultrasound therapy
•LA: Laser therapy
•CO: Combination therapy
4.1.4Program number selection
1.Press to go to the start menu.
2.Select Program number.
3.Select the desired program with
or . See §9.1.
4.Press . See §4.8.
26
Combi 200L
4.1.5Diagnostic program selection
With the diagnostic programs, you can localise and treat pain points, and
look for stress fractures, etc.
1.Press to go to the start menu.
2.Select Diagnostic programs.
3.Select the desired diagnosis
with . See §4.7.
4.1.6Contra indication selection
1.Press to go to the start menu.
2.Select Contra indications.
3.Select the therapy for which you
want to see the contra
indications.
27
Combi 200L
4.2Performing therapy
4.2.1Set and start therapy
1.Press to go to the start menu.
2.Select the desired menu with
until the treatment appears.
3.Select the desired parameters
with . You can only change the
outlined parameters.
4.Set the Treatment time as
follows: Press once on to set
the minutes, press twice on
to set the seconds.
5.Change the value of the parameter
with and . The setting range of the parameter is shown at the
bottom of the screen. You can change the parameter as long as the
parameter has a black background.
6.Start the therapy:
•Electrotherapy or ultrasound therapy: Rotate intensity knob A or B
to start the treatment and to set the desired intensity. The set
intensity is displayed in the screen.
•Laser therapy: See §4.6.1.
4.2.2Set channels A and B
The Combi 200L has two separated electrotherapy channels A and B. The
only restriction is that both are in the CC mode or the CV mode.
The channels A and B can be used independently. You can treat two
different indications simultaneously with two different treatments.
1.Press for 5 seconds. The System setting menu appears. See §4.9.
2.If necessary, change the parameter Copy parameters to OFF.
3.The selected channel has a black background. If desired, press to
A
B
change the first channel.
4.Select a treatment. See §4.1.
5.Set the parameters for the first channel. See §4.2.1.
6.Press to select the other channel.
A
B
7.Select a treatment for the second channel. See §4.1.
8.Set the parameters for the second channel. See §4.2.1.
Both channels are selected simultaneously and automatically in case of:
•4-pole current shapes
•Combination therapy
28
Combi 200L
Copy channel
On the second channel, you can set the same parameters for
electrotherapy as for the first set channel.
1.Press for 5 seconds. The System setting menu appears. See §4.9.
2.If necessary, change the parameter Copy parameters to ON.
3.Select a treatment. See §4.1.
4.Set the parameters for the first channel. See §4.2.1.
5.Press to select the other channel. The treatment including the
A
B
settings are copied to the other channel.
6.If desired you can change the parameters or the treatment of the
selected channel.
Clear channel
1.Make sure that the intensity is set to zero.
2.Press to select the channel that you want to clear.
A
B
3.Press . The channel is cleared.
4.2.3Opening the intensity screen
1.Set and start the treatment. See
§4.2.1.
2.Press . The intensity screen
appears.
The left part of the screen
shows channel A. The right part
of the screen shows channel B.
3.Press to return to the setting
menu.
4.2.4Temporary interruption of treatment
1.If the other channel has to pause: Select this channel with .
2.Press during the treatment. The treatment time of the selected
channel is stopped. Pause appears on the screen. The parameter
settings are retained.
3.Press on again to restart the treatment. The intensity now increases
gradually to the set level and the treatment time continues again.
A
B
4.2.5Immediately stop treatment
1.Press . All active treatments are stopped immediately. Stop appears
on the screen. The parameter settings are retained.
2.Set the intensity of the channel again to continue the treatment.
29
Combi 200L
4.3Electrotherapy
4.3.1Performing electrotherapy with electrodes
1.Select the desired electrotherapy program.
2.Place the electrodes. See page 30: Placing rubber electrodes and page 31: Placing the adhesive electrodes. With some treatments, the
Electrode placing parameter refers to the number in the placing
diagrams.
3.Rotate intensity knob A or B to start the electrotherapy and to set the
desired intensity. See §4.2.1.
4.Check the patient's reaction. Repeat this check regularly during the
treatment.
5.The equipment stops the treatment and indicates that the treatment is
completed. Remove the electrodes.
Placing rubber electrodes
1.Moisten two EL sponges. Use
water with a saline solution to
improve the conductivity of the EL
sponges.
2.Slide a rubber electrode into each
sponge.
3.Place the sponges on the part of
the body that must be treated.
4.Fasten the sponges to the part of
the body with the elastic fixation
straps.
5.Connect the rubber electrode with the red connector to the red
connector of the two-ply electrode cable.
6.Connect the rubber electrode with the black connector to the black
connector of the two-ply electrode cable.
7.Connect the two-ply cable to connector or of the Combi 200L.
A
B
30