GW Instek PSH-A User Manual

Programmable Power Supply
PSH series
USER MANUAL
GW INSTEK PART NO. XXXX-XXXXXX
This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.
Table of Contents
Table of Contents
PSH User Manual
Buzzer sound Setting ................................51
Remote Operation.........................................52
Safety Instructions ........................................ 5
Safety Symbols ......................................... 6
Safety Guidelines....................................... 6
Power cord for the United Kingdom .............. 9
PSH Overview .............................................. 11
Main Features ..........................................12
PSH Series Lineup.....................................13
Package Contents .....................................14
Front Panel ..............................................16
Rear Panel ...............................................18
Display....................................................19
Setup ........................................................... 21
AC Power Cable Assembly..........................22
Remote Sensing and Local Sensing .............24
Load / Remote Sensing Wire Selection.........25
Load Configuration....................................26
Load Wire Assembly..................................30
Remote Sensing Wire Assembly ..................32
Functionality Check...................................34
Rack Mounting (Optional) ..........................40
Interface Selection....................................53
Command Syntax .....................................57
Command Set ..........................................58
Calibration ....................................................61
Calibration Preparation ..............................62
Entering calibration mode ..........................63
Output Voltage calibration..........................64
Output Current calibration .........................66
OVP Calibration ........................................68
FAQ...............................................................69
Appendix.......................................................71
Specifications...........................................71
Declaration of Conformity ..........................75
Index ............................................................76
Panel Operation ........................................... 43
Menu Key Overview ..................................44
Constant Voltage/ Constant Current Crossover
Characteristic...........................................45
Output Voltage Setting ..............................46
Output Current Setting ..............................47
OVP (Output Voltage Protection) Setting ......48
OCP (Output Current Protection) Setting ......49
Display Contrast Setting ............................50
3
4
Safety Instructions
PSH User Manual
Safety Symbols
Safety Guidelines
Power cord
Safety Instructions
This chapter contains important safety instructions that must be followed when operating PSH and when keeping it in storage. Read the following before any operation to insure safety and to keep the best condition for PSH.
Safety Symbols .........................................6
Safety Guidelines....................................... 6
Power Supply ............................................ 7
Fuse ........................................................7
Cleaning PSH ............................................7
Operation Environment............................... 8
Storage Environment ................................. 8
Power cord for the United Kingdom .............. 9
Safety Symbols
These safety symbols may appear in this manual or on PSH.
Warning: Identifies conditions or practices that
WARNING
could result in injury or loss of life.
Caution: Identifies conditions or practices that
CAUTION
could result in damage to PSH or to other properties.
DANGER High Voltage
Attention Refer to Manual
Protective Conductor Terminal
Earth (ground) Terminal
Safety Guidelines
General Guideline
CAUTION
Do not place any heavy object on PSH.
Avoid severe impacts or rough handling that
leads to damaging PSH.
Do not discharge static electricity to PSH.
Do not block or obstruct cooling fan vent
opening.
Do not perform measurements at power source
and building installation site (Note below).
Do not disassemble PSH unless you are
qualified as service personnel.
(Note) EN 61010-1:2001 specifies the measurement categories and their requirements as follows. PSH falls under category II. Measurement category IV is for measurement performed
5
6
Safety Instructions
PSH User Manual
Power Supply
WARNING
Fuse
WARNING
Cleaning PSH
at the source of low-voltage installation. Measurement category III is for measurement performed in the building installation. Measurement category II is for measurement performed on the circuits directly connected to the low voltage installation.
Input voltage: 90 ~ 250 V AC, 50/60 Hz
The power supply voltage should not fluctuate
more than 10%.
Connect the protective grounding conductor of
the power cord to earth ground, to avoid electrical shock.
PSH-1036A
6.3A/250V x2 PSH-2018A PSH-3610A PSH-6006A
PSH-1070A PSH-2035A PSH-3620A
6.3A/250V x2
0.5A/250V x1
15A/250V x1
PSH-6012A PSH-10100A
PSH-2050A PSH-3630A
6.3A/250V x3
0.5A/250V x1
20A/250V x1
PSH-6018A
Make sure the correct type of fuse is installed
before powering up.
Replace the fuse with the specified type and
rating only, for continued fire protection.
Disconnect the power cord before fuse
replacement.
Make sure the cause of the fuse blowout is
fixed before fuse replacement.
Disconnect the power cord before cleaning.
Operation Environment
WARNING
Storage Environment
xylene, and acetone.
Location: Indoor, no direct sunlight, dust free, almost non-conductive pollution (Note below)
Relative Humidity: < 85%
Altitude: < 2000m
Temperature: 0°C to 40°C
Input Breaker Capacity: Over 20A (PSH-10100A, 2050A, 3630A, 6018A)
This is a Class A product which may cause radio interference in a domestic environment. In such case, take adequate measures.
(Note) EN 61010-1:2001 specifies the pollution degrees and their requirements as follows. PSH falls under degree
2. Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”. Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. Pollution degree 2: Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected. Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Location: Indoor
Relative Humidity: < 70%
Temperature: 10°C to 70°C
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid into PSH.
Do not use chemicals or cleaners containing
harsh materials such as benzene, toluene,
7
8
Safety Instructions
Power cord for the United Kingdom
When using PSH in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead / appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the following code: Green/ Yellow: Earth Blue: Neutral Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth symbol or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.
or coloured Green
9
PSH Overview
PSH User Manual
Main features
Series lineup
Package contents
Panel descriptions
PSH Overview
PSH Series are modular-type programmable switching power supplies designed for broad range of applications. The series consists of 12 models, output ranging from 360W to 1000W. Switching technology and built-in PFC control give PSH higher power efficiency, power density, and power factor compared with other linear power supplies. Protection mechanisms keep the output voltage, current, and temperature within limit. Remote control interface equipped with SCPI command set and Lab-View Driver facilitates ATE software development.
This chapter describes PSH series features and appearances in a nutshell.
Main Features ......................................... 12
360W..................................................... 13
720W..................................................... 13
1000/ 1080W.......................................... 13
Main unit ................................................ 14
AC input cable kit .................................... 14
Output connection kit............................... 15
Manual................................................... 15
Front Panel ............................................. 16
Rear Panel .............................................. 18
Main Features
Performance
Operation
Protection
Interface
Optional items
12 models with wide output range: 6~100A,
10~60V, and 360~1000W.
High power factor with PFC control
High efficiency power conversion
Compact size, light weight
Constant voltage operation
Constant current operation
Output On/Off control
Built-in buzzer
Self test and calibration
LCD display
Over voltage protection (OVP)
Over current protection (OCP)
Over temperature protection (OTP)
RS-232 (standard) / GPIB (optional) interface
IEEE 488.2/SCPI compatible command set
LabView driver
GPIB remote control interface
19 inch standard rack mounting
Display descriptions
11
Default display ........................................ 19
Menu mode display .................................. 20
12
PSH Overview
PSH User Manual
PSH Series Lineup
PSH series consist of the following 12 models with various output voltage and current ratings. For the detailed specifications, see page71.
360W (Mainframe only)
720W (Mainframe + one slave module)
PSH-1036A
10V, 36A
PSH-2018A
20V, 18A
PSH-3610A
36V, 10A
PSH-6006A
60V, 6A
PSH-1070A
10V, 70A
PSH-2035A
20V, 35A
PSH-3620A
36V, 20A
PSH-6012A
60V, 12A
Package Contents
Check the contents before using PSH series. Contact your local dealer in case there is a missing item.
Main unit
AC input cable kit
Cable gland
Terminal cover
1000/ 1080W (Mainframe + two slave modules)
PSH-10100A
10V, 100A
PSH-2050A
20V, 50A
PSH-3630A
AC power input cord
36V, 30A
PSH-6018A
60V, 18A
13
14
PSH Overview
PSH User Manual
Package Contents (cont.)
Output connection kit
Manual
Terminal cover
Output cable screw
User manual (this document)
Programming manual
Front Panel
1 LCD Display
2 Power
Switch
3 Wheel knob
4 Output
Switch
Shows the output and the configuration status. See page19 for details.
Sets parameters. Turn right: increase, turn left: decrease.
Turns the output On or Off. When On, the “OUT” sign appears on the display.
On Off
Out On
OCP OUT
CV 2.58V 2.01A
15
16
Out Off
OCP
SET 2.58V 2.01A
PSH Overview
5 Vset/ Iset/
Enter key
PSH User Manual
Rear Panel
Switches between voltage setting mode and current setting mode, or confirm the entered value in the menu mode (see page44).
Vset (edit Voltage)
OCP
.58V 2.01A
SET 2
6 F/C (Fine/
Coarse) key
Switches the editing location and resolution: before (coarse) or after (fine) the decimal point.
Fine (after decimal)
SET 2.58
7 Menu key
OCP
V 2.01A
Enters into the menu mode. For details, see page44.
Default mode
OCP OUT
CV 2.58V 2.01A
8 Local key
Switches from remote control mode to local operation mode (page56), OR releases OVP/OCP error messages and go back to normal operation (page37), OR enters the calibration mode when pressed for more than 5 seconds (page61).
Error message example
OCP Error! Press “Local” to reset
Iset (edit Current)
SET 2.58V 2
OCP
Coarse (before decimal)
SET 2
OCP
.58V 2.01A
Menu mode (OVP setting)
Set OVP
* 21.10V
Calibration mode
Calibration
Voltage
.01A
1 Ground
Terminal
2 Output
Terminal
3 Sense Terminal
Connect the output line shield (page25) and the remote sensing line shield (page32).
Connect DUT (Device Under Test). For details, see page24.
Connect the feedback line to compensate for cable loss. For details, see page24 (theory),
4 RS232/ GPIB
Terminal
page32 (cable connection).
Connect the remote control line. For remote control details, see page52. GPIB requires an optional module. For installation details, see the service manual. Note: Only one interface module (RS232 or GPIB) can be installed at a time.
5 AC input
Terminal
Connect the AC power input cable. For details, see page22.
17
18
PSH Overview
PSH User Manual
Display
Default display
Display mode
A B C D
RMT ADRS OCP OUT
CV 10.10V 36.10A
E F G
Editing mode
A B C D
SET 10.10V 36.10A
E F G
OCP
A RMT: remote control mode
(Nothing): panel operation mode Not available in editing mode
B ADRS: RS-232 or GPIB address (available
in remote control) Not available in editing mode
C OCP: Output Current Protection On
(Nothing): Output Current Protection Off
D OUT: Output On
(Nothing): Output Off Not available in editing mode
E CV/CC: Current and Voltage display
mode (Constant Voltage/ Constant Current) SET: Current and Voltage editing mode
F Output Voltage readback value (display
mode) Output Voltage setting value (editing mode)
G Output Current readback value (display
mode) Output Current setting value (editing mode)
Display (cont.)
Menu mode display
The following displays appear when pressing the
Menu key press the Menu key repeatedly. When inactive for more than 5 seconds, the display goes back to the default mode.
MENU
. To move to the next configuration,
OCP OUT
CV 2.58V 2.01A
Set OVP
21.10V
Set OCP ON (OFF)
Set Contrast
70%
Set Buzzer ON (OFF)
Interface RS232 Baud Rate 9600
(Default display)
Output Voltage Protection setting. For details, see page48.
Output Current Protection setting. For details, see page49.
Display contrast setting. For details, see page50.
Buzzer setting. For details, see page51.
Remote control interface setting. For details, see page52.
19
20
Setup
PSH User Manual
AC power cable assembly
Load configuration
Wire assembly
Setup
This chapter describes load configurations and setup procedures. Follow these instructions to properly install PSH series.
AC Power Cable Assembly......................... 22
AC power cable requirement ..................... 23
Remote Sensing and Local Sensing ............ 24
Load / Remote Sensing Wire Selection........ 25
Single Load + Local Sensing...................... 26
Single Load + Remote Sensing .................. 26
Multiple Loads + Local Sensing .................. 27
Multiple Loads + Remote Sensing .............. 27
Series Operation + Local Sensing............... 28
Series Operation + Remote Sensing ........... 29
Load Wire Assembly................................. 30
AC Power Cable Assembly
1 Cable gland + Terminal Cover
2 Cable wire + Terminal
CAUTION
Put the power cable through the cable gland and the terminal cover, screw them together.
Screw the wire onto the AC input terminal. Note the wire color: Neutral (white), GND (green), and Line (black).
Remote Sensing Wire Assembly ................. 32
Functionality Check
Preparation............................................. 34
Output Voltage & OVP Check ..................... 35
Output Current Check .............................. 37
3 Terminal cover + Terminal
Screw the terminal cover onto the terminal.
OCP Check.............................................. 38
Rack mounting (optional)
21
Rack mounting kit contents....................... 40
Rack mounting assembly .......................... 41
22
Setup
PSH User Manual
AC power cable requirement
Here is the AC power cable specification, in case of using cables other than the attached one.
Cable length
Cable gland
Cable type (recommended)
Model: H05 VV-F
3m
KSS or PG-2013
Model: SJT Type: 3 x 14 AWG stranded copper Rating: 60°C min, 300V Diameter: 9.143~10.03 mm
Type: 3G 1.5mm
2
stranded copper Rating: 300V/500V Diameter: 8.5 ± 0.2 mm
Remote Sensing and Local Sensing
Remote sensing compensates the cable loss between PSH and load, up to 0.5V. Use remote sensing whenever the load voltage has to be accurate.
Local Sensing (default)
VSET = VOUT V
OUT > VLOAD
Remote Sensing
VSET = VLOAD VOUT > VSET
VSET
PSH Load
Output Setting
|VSET–VOUT|
The sense terminal is internally connected to the PSH output terminal. The delta between the voltage setting level (V level (V voltage (V
OUT) is compensated. The load terminal
LOAD) might become lower than the
output due to cable loss.
VSET
PSH Load
Output Setting
|VSET–VLOAD|
VOUT VLOAD
VOUT
Sense
SET) and the actual output
OUT VLOAD
V
VLOAD
Sense
The sense terminal is connected to the load input terminal. The delta between the voltage setting level (V compensated. The output voltage (V
SET) and the actual load voltage (VLOAD) is
OUT) might
become higher than the setting due to the compensation.
23
24
Setup
PSH User Manual
Load / Remote Sensing Wire Selection
The following instructions apply to both load wire and remote sensing wire, unless noted.
Wire size (FOR LOAD ONLY)
Wire length
Wire shielding
Load wires must have enough current capacity to minimize cable loss and load line impedance. Voltage drop across a wire should not excess
0.5V. The following list is the wire current rating at 450A/cm
Wire size
(AWG)
2
.
Max.
current (A)
Wire size
(AWG)
Max.
current (A)
20 2.5 6 61 18 4 4 97 16 6 2 155 14 10 1 192 12 16 1/0 247 10 21 2/0 303
8 36
To avoid excessive cable loss, shorten the distance between PSH and load as much as possible. Remote sensing compensates cable loss up to 0.5V.
To minimize noise effect and load line impedance, use shielded pair wiring. Twisted wires are more effective, especially for remote sensing.
Load Configuration
Select the appropriate configuration for the target application. For local sensing and remote sensing explanation, see the previous page. For connection guideline, see page25 (wire selection), page30 (load connection), page32 (remote sensing connection).
Single PSH + single load
Single Load + Local Sensing
Single Load + Remote Sensing
—OUT
PSH
+OUT
—Sense
+Sense
Connect the output wire to the load and the sense terminal to the PSH local output monitor.
PSH
+OUT
—Sense
+Sense
—IN
Load
+IN
—IN—OUT
Load
+IN
Connect both the output wire and the remote sensing wire to the load.
Wire shield should be connected to the rear panel ground connector.
25
26
Setup
PSH User Manual
Single PSH + multiple Load
Condition
Multiple Loads + Local Sensing
Multiple Loads + Remote Sensing
Output current for each load follows the load
requirement.
When the sum of the load current surpasses the
rating, PSH automatically switches to CC (Constant Current) mode.
—OUT
PSH
+OUT
—Sense
+Sense
All loads share one PSH output.
Distribution
—OUT
PSH
+OUT
—Sense
+Sense
Create a distribution point between PSH and the loads. PSH compensates the cable loss up to the distribution point. To maximize compensation, make sure the distribution point is closer to the load side.
Point
—IN
+IN
—IN
+IN
—IN
+IN
—IN
+IN
Load #1
Load #2
Load #1
Load #2
Multiple PSH + single load (series operation)
Condition
Series Operation + Local Sensing
Up to four PSH series (with identical output
ratings) can be cascaded.
Output voltage is the sum of the cascaded PSH.
Output current is the same as a single PSH.
—IN—OUT
PSH #1
+OUT
+IN
Load
—Sense
+Sense
—OUT
PSH #2
+OUT
—Sense
+Sense
First PSH: Negative output Negative load input Positive output The next PSH negative output
Intermediate PSH: Negative output The previous PSH positive output Positive output The next PSH negative output
Last PSH: Negative output The previous PSH positive output Positive output Positive load input
27
28
Setup
PSH User Manual
Series Operation (cont.)
Series Operation + Remote Sensing
PSH #1
+OUT
—Sense
+Sense
—OUT
PSH #2
+OUT
—Sense
+Sense
First PSH: Negative output Negative load input Positive output The next PSH negative output Negative sense Negative load input Positive sense Positive output monitor
Intermediate PSH: Negative output The previous PSH positive output Positive output The next PSH negative output Negative sense The previous PSH positive output Positive sense Positive output monitor
—IN—OUT
+IN
Load
Load Wire Assembly
1. Wire selection
2. Terminal screw replacement
3. Terminal connection
Select the appropriate wire according to the guideline on page25.
Replace the two screws on the output terminal with the hex screw in the output connection kit.
Screw load wires to the terminal.
Last PSH: Negative output The previous PSH positive output Positive output → Positive load input Negative sense The previous PSH positive output Positive sense Positive load input
29
30
Setup
PSH User Manual
Load Wire Assembly (cont.)
4. Terminal cover assembly
5. Wire shield connection (recommended)
Screw the output terminal cover to the rear panel.
Connect the load wire shield to the rear panel ground connector.
Remote Sensing Wire Assembly
Select the appropriate wire according to the guideline on page25.
Local sensing
The sense terminal is connected to the output monitor terminal with bare wires.
Remote Sensing
1. Take off the wire jumpers.
2. Screw wires to the sense (S) side.
Do not screw remote sensing wires to the
CAUTION
monitor (M) side.
M+, M are for output monitoring only. NEVER
screw load wires to the remote sensing terminal.
31
32
Setup
PSH User Manual
Remote Sensing Wire Assembly (cont.)
5. Wire shield connection
CAUTION
To minimize noise effect, we recommend covering the remote sensing wire with ground shield and connect it to the ground terminal.
An open remote sense circuit leads to output level overshoot. Make sure the line is securely connected.
Functionality Check
Check the PSH basic functionalities before operation.
Preparation
Check items
Equipment
When there is a problem…
Output Voltage
OVP functionality
Output Current
OCP functionality
Digital
DC Voltage Accuracy: <±0.1%
Multimeter
Recommended model:
GDM-8245, GDM-8246
Current Shunt Current range: >100A
Accuracy: <±0.1%
PSH – Current
Voltage rating: >70V
Shunt cable
Current rating: >100A
Multimeter –
N/A
Current Shunt cable
Run calibrations (page61). If this does not solve the issue, refer to the service center.
33
34
Setup
PSH User Manual
Output Voltage & OVP Check
Connection
PSH Series
Digital Multimeter
V
A
Checking step
1. Power On PSH and connect the Multimeter Voltage
measurement terminal.
2. Set Output Voltage and Output Current to the rating value.
Vset/Iset/Enter key
VSET/ ISET
ENTER
PSH-2018A (20V, 18A)
switches the cursor between Voltage and Current.
F/C key
F/C
switches the cursor between before and after the decimal point.
+
SET 20.00V 18.00A
4. Make sure the Multimeter shows no value (No output).
5. Press the Output key
OUTPUT
and turn On the output.
The display changes into CV
(Constant Voltage) mode and shows the OUT sign on the top right corner.
CV 20.00V 18.00A
OUT
6. Make sure the Multimeter reading and the PSH display show
approximately the same Voltage.
7. Increase the PSH output voltage beyond the OVP value.
8. Make sure the display shows an error message.
Press the Local key back to the normal display.
LOCAL
to get
OVP Error! Press “Local” to reset
9. Make sure the output is now turned Off in the display and the
Multimeter shows no value (No output).
Wheel knob changes the value.
3. Press the Menu key once and set the OVP value to rating
voltage + 100mV.
Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
to confirm the OVP
Set OVP
20.10V
setting.
35
36
Setup
PSH User Manual
Output Current Check
Connection
Digital Multimeter PSH Series
V
A
Checking step
1. Power On PSH and connect the Multimeter/ Current Shunt
terminal.
2. Set the Output Voltage and Current value to the rating.
Vset/Iset/Enter key
VSET/ ISET
ENTER
switches the cursor between Voltage and Current.
F/C key
F/C
switches the cursor
before and after the decimal point.
Wheel knob changes the value.
3. Make sure the Multimeter/Current Shunt shows no output.
4. Press the Output key
OUTPUT
5. Make sure the Multimeter/Current Shunt reading and the
PSH display show approximately the same Current.
Current
Shunt
PSH-2018A (20V, 18A)
+
SET 20.00V 18.00A
and turn On the output.
OCP Check
Connection
Current
Shunt
PSH-2018A (20V, 18A)
SET 20.00V 18.00A
Set OCP
ON
VSET/ ISET
ENTER
to confirm the setting.
+
Digital Multimeter PSH Series
V
A
Checking step
1. Power On PSH and connect the Multimeter / Current Shunt
terminal.
2. Set the Output Voltage and Current value to the rating.
Vset/Iset/Enter key
VSET/ ISET
ENTER
switches the cursor between Voltage and Current.
F/C key
F/C
switches the cursor
before and after the decimal point.
Wheel knob changes the value.
3. Press the Menu key twice to view the OCP setting.
Turn the wheel knob
to change
OCP setting to ON.
Press the Vset/Iset/Enter key
37
38
Setup
PSH User Manual
The OCP sign appears on the
upper side of the display.
4. Press the Output key
CV 20.00V 18.00A
OUTPUT
and turn On the output.
OCP
5. Make sure the display shows the error message.
Press the Local key back to the normal display.
LOCAL
to get
OCP Error! Press “Local” to reset
6. Make sure the output is now turned Off in the display and the
Multimeter / Current Shunt shows no value (No output).
Rack Mounting (Optional)
PSH can be mounted on standard 19 inch rack using GRA-403 rack mounting kit.
Rack mounting kit contents
Main bracket x 1
Front panel x 1
Large decoration board x 1
Small decoration board x 2
Medium decoration board x 2
Binding plate x 6
Handle x 2
Screw M4 * 0.7 L=10mm x 4
Screw M3 * 0.5 L = 8mm x 18
39
40
Setup
PSH User Manual
Rack mounting assembly
1. Rack mounting layout
Confirm the rack mount layout. Make sure there is a space between each PSH. Below is the example of rack mounting layout.
2. Decoration board assembly
Once the layout is fixed, screw the decoration boards to the front panel using M3 screws.
4. Main bracket assembly
Fix the PSH to the main bracket using M3 screws.
Fix the front panel and the handle to the main bracket using M4 screws.
3. Binding plate assembly
41
Take off two M3 screws from PSH side panels, front and rear. Hold the binding plate between PSH and drive the same screws in.
42
Panel Operation
PSH User Manual
Menu key overview
CV/ CC characteristic
Output Voltage setting
Output Current setting
OVP setting
OCP setting
Display contrast setting
Buzzer sound setting
Panel Operation
This chapter describes the manual operations done at the front panel, together with the constant voltage/ constant current crossover characteristics.
Menu Key Overview ................................. 44
Constant Voltage/ Constant Current Crossover
Characteristic.......................................... 45
Output Voltage Setting ............................. 46
Output Current Setting ............................. 47
Note ...................................................... 47
Set the OVP value.................................... 48
Clear OVP error ....................................... 48
Turn the OCP On/Off ................................ 49
Clear OCP error ....................................... 49
Display Contrast Setting ........................... 50
Buzzer sound Setting ............................... 51
Buzzer condition...................................... 51
Menu Key Overview
Press the Menu key Menu key repeatedly.
CV 2.58V 2.01A
Set OVP
* 21.10V
Set OCP
* ON (OFF)
Set Contrast
* 70%
Set Buzzer
* ON (OFF)
Interface RS232 Baud Rate 9600
MENU
OCP OUT
. To move to the next item, press the
Default display
Output Voltage Protection setting
To set the value, use the wheel
knob To select the editing point (before or after the decimal point), press the
F/C key
Output Current Protection setting
To enable or disable OCP, use the
wheel knob “OCP” appears on the default display.
Display contrast setting
To set the display contrast, use the
wheel knob Turn right: increases contrast Turn left: decreases contrast
Buzzer setting
To enable or disable buzzer sound, use the wheel knob. The buzzer sounds when trying to set a value beyond the specification.
Remote control interface setting
RS232 (standard): To set the baud
rate, use the wheel knob GPIB (optional): To set the GPIB
address, use the wheel knob
.
F/C
repeatedly.
. When enabled,
.
.
.
43
44
Panel Operation
PSH User Manual
Constant Voltage/ Constant Current Crossover Characteristic
PSH series automatically switch between constant voltage mode and constant current mode, according to the load change.
When the load current is smaller than the limit (I
SET),
PSH operates in Constant Voltage mode, changing the current level according to the load but keeping the Voltage level at the limit (V
SET).
When the load current is the same as the limit (I
SET), PSH
operates in Constant Current mode, changing the Voltage level according to the load but keeping the Current level at the limit (I
SET).
Let’s take recharging a 12V battery as an example. PSH output setting is 13.8V, 1A. An empty battery puts a heavy current load on the power supply. PSH starts running at Constant Current mode, supplying full 1A current but keeping the voltage lower than 13.8V. As the battery becomes charged, the load also becomes smaller. PSH then switches to Constant Voltage mode, supplying less than 1A current but supplying full 13.8V.
V
VSET
Constant
Voltage
Constant
Current
ISET
VSET = Output Voltage setting I
SET = Output Current setting
Output Voltage Setting
Operation
1. Press the Vset/ Iset key
repeatedly to move the underline to the Voltage side.
2. Press the F/C key
underline before or after the decimal point.
3. Use the wheel knob
I
Range
Step
0.00V ~ rating voltage
10mV (rating voltage < 36V) 20mV (rating voltage 36V)
VSET/ ISET
SET 2
F/C
to move the
SET 2
to change the output voltage.
OCP
.58V 2.01A
OCP
.58V2.01A
Display
The display indicates CV (Constant Voltage) or CC (Constant Current) on the left side.
Constant Voltage
OCP OUT
CV 2.58V 1.99A
45
Constant Current
OUT
CC 1.33V 2.00A
46
Panel Operation
PSH User Manual
Output Current Setting
Operation
1. Press the Vset/ Iset key
repeatedly to move the underline to the Current side.
2. Press the F/C key
underline before or after the decimal point.
3. Use the wheel knob
Range
Step
Note
0.00A ~ rating current
10mA
When the output current exceeds the setting value, PSH automatically switches from CV (Constant Voltage) to CC (Constant Current) mode.
VSET/ ISET
F/C
to move the
to change the output current.
SET 2.58V 2
OCP
OCP
SET 2.58V 2
.01A
.01A
OVP (Output Voltage Protection) Setting
Set the OVP value
1. Press the Menu key
repeatedly until the OVP menu appears.
2. Press the F/C key
underline before or after the decimal point.
3. Use the wheel knob
4. Press the Iset/Vset/Enter key
Clear OVP error
1. When the output voltage exceeds
OVP value, the output is shut off and an OVP error message appears on the display.
MENU
F/C
to move the
to change the value.
VSET/ ISET
ENTER
OCP OUT
CV 2.58V 2.01A
Set OVP
* 21.10V
Set OVP
.10V
* 21
to confirm the value.
OVP Error! Press “Local” to reset
LOCAL
2. Press the Local key
to clear the error message (the output is still Off).
Range
Step
0.00V ~ rating voltage
10mV (rating voltage < 36V) 20mV (rating voltage 36V)
47
48
Panel Operation
PSH User Manual
OCP (Output Current Protection) Setting
Turn the OCP On/Off
1. Press the Menu key
repeatedly until the OCP menu appears.
2. Use the wheel knob
the status to ON (or OFF).
3. Press the Iset/Vset/Enter key
Clear OCP error
1. When the output current exceeds
the current setting, the output is shut off and an OCP error message appears on the display.
MENU
to change
CV 2.58V 2.01A
Set OCP
OFF
Set OCP
ON
VSET/ ISET
ENTER
to confirm the setting.
OCP Error! Press “Local” to reset
OCP OUT
Display Contrast Setting
Operation
1. Press the Menu key
repeatedly until the Contrast menu appears.
2. Use the wheel knob
increase, Counterclockwise: decrease.
Range
Step
5% ~ 95%
6%
MENU
CV 2.58V 2.01A
OCP OUT
Set Contrast
71%
to change the contrast. Clockwise:
2. Press the Local key
LOCAL
to clear the error message (the output is still Off).
Range
49
According to the output current
50
Panel Operation
PSH User Manual
Buzzer sound Setting
Operation
1. Press the Menu key
repeatedly until the Buzzer menu appears.
2. Use the wheel knob
the buzzer setting to ON (or OFF).
3. Press the Iset/Vset/Enter key to
confirm the value.
Buzzer condition
When “Set Buzzer ON”, the buzzer sounds under the following conditions. * When pressing the panel keys * When the value setting exceeds the limit
MENU
to change
OCP OUT
CV 2.58V 2.01A
Set Buzzer
OFF
Set Buzzer
ON
Remote Operation
This chapter describes the IEEE 488.2 based remote control configuration, command syntax, and command set overview. For further descriptions and examples, refer to the programming manual.
Interface configuration
Command set
RS232 (Standard).................................... 53
GPIB (Optional) ....................................... 55
Interface functionality check ..................... 56
Command Syntax .................................... 57
General commands .................................. 58
Status commands .................................... 59
Miscellaneous commands.......................... 60
51
52
Remote Operation
PSH User Manual
Interface Selection
RS232 (Standard)
Baud rate setting
1. Press the Menu key
Interface menu appears.
CV 2.58V 2.01A
Interface RS232 Baud Rate 9600
2. Use the wheel knob to select the baud
rate.
3. Press the Iset/Vset/Enter key
confirm the setting.
OCP OUT
MENU
repeatedly until the
VSET/ ISET
ENTER
to
PC RS-232 pin assignment
PSH – PC connection
1 2 3 4 5
Pin 2: TxD Pin 3: RxD
Pin 5: GND
6 7 8 9
Pin 1, 4, 6 ~ 9: No Connection
Null-modem connection, in which transmit (TxD) and receive (RxD) lines are crosslinked, is required.
PSH PC
RxD
Pin2
Pin3
Pin5
TxD
GND
TxD
RxD
GND
Pin2
Pin3
Pin5
4. Connect the RS232 cable between your PC
and PSH rear panel. PSH switches to remote control mode as soon as the connection is established.
RS-232 interface protocol
PSH RS-232 pin assignment
Baud rate: 9600, 4800, 2400, 1200 (selectable) Parity: None (fixed) Stop bit: 1 (fixed) Data bit: 8 (fixed)
1 2 3 4 5
Pin 2: RxD Pin 3: TxD
Pin 5: GND
6 7 8 9
53
Pin 1, 4, 6 ~ 9: No Connection
54
Remote Operation
PSH User Manual
GPIB (Optional)
Refer to the service manual for GPIB module installation. Note: The RS232 module has to be replaced with the GPIB module: they cannot be used together.
Address setting
1. Press the Menu key
Interface menu appears.
OCP OUT
CV 2.58V 2.01A
Interface GPIB Address 8
2. Use the wheel knob
GPIB address: 1 ~ 30
3. Press the Iset/Vset/Enter key
confirm the setting.
MENU
repeatedly until the
to select the address.
VSET/ ISET
ENTER
to
Interface functionality check
Remote mode display
Functionality check
In the remote control mode, the display shows “RMT” on the top left corner. The panel operation is disabled.
RMT OCP CV 0.05V 1.05A
To enable panel operation, press the Local
LOCAL
key
. This also disconnects the remote
control.
Run this query command via the terminal.
*idn?
This should return the Manufacturer, Model number, Serial number, and Firmware version in the following format.
GW.Inc,PSH-2018A,12345678,FW1.00
4. Connect the GPIB cable between your PC and
PSH rear panel. PSH switches to remote control mode as soon as the connection is established.
GPIB constraints
Address range: 1 ~ 30
Altogether less than 15 devices, less than 20m
total cable length, maximum 2m between each device
Unique address assigned for each device
At least 2/3 of the GPIB devices turned On
No loop or parallel structure
55
56
Remote Operation
PSH User Manual
Command Syntax
The commands are fully compatible with IEEE488.2 (1992) standard and partially compatible with SCPI (1994) standard.
Example command
Command Header
Parameter example
Message Terminator
:chan1:prot:curr <0/1> LF
4321
1: Command Header 2: Single space 3: Parameter 4: Message Terminator OR Message Separator
Several command header elements (nodes) can be concatenated to form a complex command. The above example can be separated into: :chan1: (root node) + prot: + curr:
<0/1> 0 or 1.
<1~4> Integer between 1, 2, 3, or 4.
<0.01~5> Decimal number between 0.01 and 5.
Marks the end of a command line. Any of the following is acceptable, in accordance with IEEE488.2 standard.
LF^END Line feed code (hexadecimal 0A)
with END message
LF Line feed code
<dab>^END Last data byte with END message
Command Set
Commands are non-case sensitive. For more detailed information, refer to the Programming Manual which is downloadable from www.gwinstek.com.tw
General commands
:chan1:curr? Returns the output current (unit: A).
Range: 0.01~rating current :chan1:curr <0.01~rating> :chan1:volt? Returns the output voltage (unit: V).
:chan1:volt <0.01~rating> :chan1:meas:cu rr? :chan1:meas:vo lt? :chan1:prot:cur r? :chan1:prot:cur r <0/1> :chan1:prot:volt? Returns the Over Voltage Protection value.
:chan1:prot:volt <0.01~rating> :chan1:prot:cle Clears the OCP & OVP protection message from
:outp:stat? Returns the output status.
:outp:stat <0/1>
Sets the output current (unit: A).
Range: 0.01~rating current
Range: 0.01~rating voltage
Returns the output voltage (unit: V).
Range: 0.01~rating current
Returns the actual output load current (unit: A).
Returns the actual output load voltage (unit: V).
Returns the Over Current Protection status.
Range: 0 (Off), 1 (On)
Sets the Over Current Protection.
Range: 0 (Off), 1 (On)
Range: 0.01~rating (unit: V)
Sets the Over Voltage Protection value.
Range: 0.01~rating (unit: V)
the display.
Range: 0 (Output Off), 1 (Output On)
Sets the output status.
Range: 0 (Output Off), 1 (Output On)
.
Message Separator
57
; (semicolon) Command separator.
58
Remote Operation
PSH User Manual
Status commands
*cls Clears all event status registers (Output Queue,
Operation Event Status, Questionable Event Status, Standard Event Status)
*ese? Returns the ESER (Event Status Enable
Register) contents. Example: 130means ESER=10000010
*ese <0~255> Sets the ESER contents.
Example: *ese 65sets ESER to 01000001
*esr? Returns and clears the SESR (Standard Event
Status Register) contents. Example: 198means SESR=11000110
*sre? Returns the SRER (Service Request Enable
Register) contents. Example: 3means SRER=00000011
*sre <0~255> Sets the SRER contents.
Example: *SRE 7→SRER=00000111
*stb? Returns the SBR (Status Byte Register) contents.
Example: 81means SBR=01010001 :stat:oper:cond? Returns the Operation register contents. :stat:oper:enab? Returns the mask conditions of the event
register. Range: 0~32767
Example: 32767 means all 15bits are set to 1. :stat:oper:enab <0~32767>
:stat:oper:even? Returns the operation register contents.
:stat:pres Sets the operation register and questionable
:stat:ques:enab? Returns the mask conditions of the event
:stat:ques:enab <0~32767>
:stat:oper:even? Returns the questionable register contents.
Returns the mask conditions of the event
register. Range: 0~32767
Example: 32767 sets all 15bits to 1.
Example: 32767 means all 15bits are set to 1.
enable register to zero.
register. Range: 0~32767
Example: 32767 means all 15bits are set to 1.
Returns the mask conditions of the event
register. Range: 0~32767
Example: 32767 sets all 15bits to 1.
Example: 32767 means all 15bits are set to 1.
Miscellaneous commands
*idn? Returns the power supply ID as Manufacturer,
Model No, Serial No, Firmware version. Example: GW.Inc,PSH-2018A,12345678,FW1.00
*opc? “1” is placed in the output queue when all the
pending operations are completed.
*opc Sets the operation complete bit (bit0) in SERS
(Standard Event Status Register) when all the pending operations are completed.
*rcl <1~99> Recalls the panel setup from internal memory.
Example: *RCL 1recalls setup from memory1 *rst Recalls the default panel setup (reset the device). *sav <1~99> Saves the panel setup to internal memory.
Example: *SAV 1saves setup to memory1 *tst? Run the self-test for RAM and ROM and returns
0 (successful) or -300 (unsuccessful). *wai Prevents execution of further commands until all
the pending operations are completed. :syst:err? Returns the error number followed by the error
message from the queue.
Example: 0, “No error” :syst:vers? Returns the SCPI version.
Example: 1994.0
59
60
Calibration
PSH User Manual
Setup
Calibration
Calibration
Run calibration when the output exceeds the specification, or when the functionality check (page34) fails. To get the optimal result, make sure PSH is powered for at least 30 minutes before the calibration.
Calibration Preparation ............................. 62
Entering calibration mode ......................... 63
Output Voltage calibration......................... 64
Output Current calibration......................... 66
OVP Calibration ....................................... 68
Calibration Preparation
Calibration condition
Calibration items
Calibration Equipment
At least 30 minutes of warm-up time
Temperature: 23 ± 5 °C
Relative Humidity: 80%
Minimum/ Maximum Output Voltage
Minimum/ Maximum Output Current
OVP functionality
Digital
DC Voltage Accuracy: <±0.1%
Multimeter
Recommended model:
Current Shunt Current range: >100A
Accuracy: <±0.1%
PSH – Current
Voltage rating: >70V
Shunt cable
Current rating: >100A
Multimeter –
N/A
Current Shunt
cable
GDM-8245, GDM-8246
61
62
Calibration
PSH User Manual
Entering calibration mode
1. Press the Local key for 5 seconds.
The password entry menu appears.
2. Enter the model number as the
password.
Wheel knob
F/C key
changes the digit.
F/C
moves the cursor to
the next digit.
3. Press the Vset/Iset/Enter key
4. The calibration menu appears.
Wheel knob
Vset/Iset/Enter key
Minimum/Maximum output
selects the item.
VSET/ ISET
ENTER
confirms the selection.
voltage calibration
Minimum/Maximum output
current calibration
(PSH-2018A)
VSET/ ISET
ENTER
as confirmation.
Please enter the
Password: 0
Please enter the Password: 2018
Calibration
Voltage
Calibration
Current
Output Voltage calibration
Connection
Digital Multimeter
V
Calibration step
A
1. Enter the Voltage calibration
menu. See page63 for details.
2. Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
and enter Minimum
Voltage calibration mode.
PSH Series
+
Calibration
Voltage
Calibration
Voltmin 0.00V
Over Voltage Protection calibration
Save the calibration result and exit
the calibration menu
Exit calibration menu without
saving the calibration result
Calibration
OVP
Calibration
Save
Calibration
Exit
3. Adjust the display value to the
Multimeter reading.
Wheel knob
F/C key
changes the value.
F/C
moves the cursor
before and after the decimal point.
4. Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
to confirm the setting. The display enters Maximum Voltage calibration mode.
Multimeter reading: 0.02V
Calibration
Voltmin 0.02
V
Calibration
Voltmax 20.00V
63
64
Calibration
PSH User Manual
5. Adjust the display value to the
Multimeter reading.
Wheel knob
F/C key
changes the value.
F/C
moves the cursor
before and after the decimal point.
6. Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
to confirm the setting. The display goes back to the default calibration menu.
7. Save the result and exit the
calibration mode (Save), OR Exit the calibration mode without saving the result (Exit).
Multimeter reading:
20.03V
Calibration
Voltmax 20.03
Calibration
Voltage
Calibration
Save
Calibration
Exit
Output Current calibration
Connection
V
Digital Multimeter PSH Series
V
A
Calibration step
1. Enter the Current calibration
menu. See page63 for details.
2. Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
and enter Maximum
Current calibration mode.
Current
Shunt
+
Calibration
Current
Calibration CurrMax 18.00A
65
3. Adjust the display value to the
Multimeter / Current Shunt reading.
Wheel knob
F/C key
changes the value.
F/C
moves the cursor
before and after the decimal point.
4. Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
to confirm the setting. The display enters Minimum Current calibration mode.
66
Multimeter reading :18.02A
Calibration CurrMax 18.02
A
Calibration
CurrMin 0.00A
Calibration
PSH User Manual
5. Adjust the display value to the
Multimeter / Current Shunt reading.
Wheel knob
F/C key
changes the value.
F/C
moves the cursor
before and after the decimal point.
6. Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
to confirm the setting. The display goes back to the default calibration menu.
7. Save the result and exit the
calibration mode (Save), OR Exit the calibration mode without saving the result (Exit).
Multimeter reading: 0.02A
Calibration
CurrMin 0.02
A
Calibration
Current
Calibration
Save
Calibration
Exit
OVP Calibration
Connection
No connection required
Calibration step
1. Enter the Voltage calibration
menu. See page63 for details.
2. Press the Vset/Iset/Enter
VSET/ ISET
ENTER
key
. PSH calibrates OVP
automatically.
3. The display goes back to the
default calibration menu.
4. Save the result and exit the
calibration mode (Save), OR Exit the calibration mode without saving the result (Exit).
Calibration
OVP
Calibration
OVP
Calibration
OVP
Calibration
Save
100%
67
Calibration
Exit
68
FAQ
FAQ
The OVP value/ OCP status has not been changed.
After editing OVP value and OCP status, press
VSET/ ISET
the Vset/Iset/Enter key Otherwise the setting does not change.
The front panel does not respond.
When in remote control mode, PSH does not respond to panel operations. Terminate the
remote operation or press the Local key on the front panel. The “RMT” message disappears from the display and panel operation is resumed.
The actual output is lower than the setting.
The cable loss might be causing the load voltage lower than the setting. Upgrade the output cable to a better capacity and/or use the remote sense to compensate for the voltage drop. See page25 (theory) and page25 (cable assembly).
The actual output is much higher than the setting.
The remote sensing (page24) compensates for the cable loss for up to 0.5V. If the output level goes much higher, the remote sensing terminal might be disconnected. Make sure the connection is secure. See page32 for details.
The load wire does not fit into the terminal.
ENTER
to confirm.
LOCAL
Make sure you are NOT connecting the load wire to the remote sensing terminal. Refer to page30 for connection details.
If there is still a problem, please contact your local dealer or GWInstek at www.gwinstek.com.tw
/ marketing@goodwill.com.tw.
69
Appendix
PSH User Manual
Appendix
Specifications
The specifications apply under the following conditions: PSH is powered on for at least 30 minutes, within +20°C~+30°C.
Common specification Recovery Time CV mode 2ms (50% step load change from
25%~75%)
Response Time
Temperature Coefficient Protection Over Voltage Protection, Over Current Protection, Over
Output On/Off Available AC Power Input 100~230V, 50/ 60 Hz Operation Condition
Storage Condition Accessories User manual, Programming manual, Cable gland, AC
Voltage Up 150ms (10%~90% 95% rating
load)
Voltage Down 150ms (90%~10% 10% rating
load)
Voltage (mVrms) 10mVrms,100mVpp,20Hz~20MHz Ripple & Noise
Current (mArms) 0.2% + 40mA
Voltage
(25±5°C)
Heat Protection, Inrush Current Protection
Location Indoor
Altitude 2000m
Ambient
temperature
Relative
Humidity
Installation Category II (for details, see page6)
Pollution Degree 2 (for details, see page8)
10°C~70°C, 70% RH (maximum)
power cord, AC input cover, O/P terminal cover
100ppm/ °C
Specification: 10°C~35°C (50°F~95°F)
Operation: 0°C~40°C (32°F~104°F)
85% RH (maximum), non condensing
PSH-1036A/ 2018A/ 3610A/ 6006A Output PSH-1036A 10V, 36A
PSH-2018A 20V, 18A PSH-3610A 36V, 10A PSH-6006A 60V, 6A Regulation (C.V.) Line 0.05% + 5mV Regulation (C.C.) Ripple & Noise
Program Accuracy
Program Resolution
Readback (Meter) Accuracy Current 0.2% + 30mA Readback (Meter) Resolution Fuse 6.3A/ 250V AC x 2 Dimensions 108 x 141 x 388 mm Weight Approx. 3.3kg
Load 0.1% + 5mV
Load 0.2% + 5mA
Line 0.2% + 5mA
Voltage (mVrms) 10mVrms, 100mVpp,
20Hz~20MHz
Current (mArms) 0.2%
Volt age 0.05% + 25mV (rating 36V)
0.05% + 50mV (rating > 36V)
OVP 0.1% + 50mV (rating 36V)
0.1% + 100mV (rating > 36V)
Current 0.2% + 30mA
Voltage & OVP 10mV (rating 36V)
20mV (rating > 36V)
Current 10mA
Volt age 0.05% + 25mV (rating 36V)
0.05% + 50mV (rating > 36V)
Volt age 10mV (rating ≤ 36V)
20mV (rating > 36V)
Current 10mA
71
72
Appendix
PSH User Manual
PSH-1070A/ 2035A/ 3620A/ 6012A Output PSH-1070A 10V, 70A
PSH-2035A 20V, 35A PSH-3620A 36V, 20A PSH-6012A 60V, 12A Regulation (C.V.) Line 0.05% + 5mV Regulation (C.C.) Ripple & Noise
Program Accuracy
Program Resolution
Readback (Meter) Accuracy Current 0.2% + 60mA (rating 10V)
Readback (Meter) Resolution
Fuse 6.3A/ 250VAC x 2, 0.5A/ 250VAC x 1, 15A/ 250VAC x 1 Dimensions 188 x 141 x 388 mm Weight Approx. 6.2kg
Load 0.1% + 5mV
Load 0.2% + 10mA
Line 0.2% + 10mA
Voltage (mVrms) 10mVrms, 100mVpp,
20Hz~20MHz
Current (mArms) 0.2% + 20mA
Voltage 0.05% + 25mV (rating 36V)
0.05% + 50mV (rating > 36V)
OVP 0.1% + 50mV (rating 36V)
0.1% + 100mV (rating > 36V)
Current 0.2% + 60mA (rating 10V)
0.2% + 30mA (rating 60V)
Voltage & OVP 10mV (rating 36V)
20mV (rating > 36V)
Current 20mA (rating ≤ 10V)
10mA (rating 60V)
Voltage 0.05% + 25mV (rating 36V)
0.05% + 50mV (rating > 36V)
0.2% + 30Ma (rating 60V)
Voltage 10mV (rating ≤ 36V)
20mV (rating > 36V)
Current 20mA (rating ≤ 10V)
10mA (rating 60V)
PSH-10100A/ 2050A/ 3630A/ 6018A Output PSH-10100A 10V, 100A
PSH-2050A 20V, 50A PSH-3630A 36V, 30A PSH-6018A 60V, 18A Regulation (C.V.) Line 0.05% + 5mV Regulation (C.C.) Ripple & Noise
Program Accuracy
Program Resolution
Readback (Meter) Accuracy Current 0.2% + 90mA (rating 10V)
Readback (Meter) Resolution
Fuse 6.3A/ 250VAC x 3, 0.5A/ 250VAC x 1, 20A/ 250VAC x 1 Dimensions 268 x 141 x 388 mm Weight Approx. 9.3kg
Load 0.1% + 5mV
Load 0.2% + 15mA
Line 0.2% + 15mA
Voltage (mVrms) 10mVrms, 100mVpp,
20Hz~20MHz
Current (mArms) 0.2% + 40mA
Volt age 0.05% + 25mV (rating 36V)
0.05% + 50mV (rating > 36V)
OVP 0.1% + 50mV (rating 36V)
0.1% + 100mV (rating > 36V)
Current 0.2% + 90mA (rating 10V)
0.2% + 60mA (rating 20V)
0.2% + 30mA (rating 60V)
Voltage & OVP 10mV (rating 36V)
20mV (rating > 36V)
Current 30mA (rating ≤ 10V)
20mA (rating 20V)
10mA (rating 60V)
Volt age 0.05% + 25mV (rating 36V)
0.05% + 50mV (rating > 36V)
0.2% + 60mA (rating 20V)
0.2% + 30Ma (rating 60V)
Volt age 10mV (rating ≤ 36V)
20mV (rating > 36V)
Current 30mA (rating ≤ 10V)
20mA (rating 20V)
10mA (rating 60V)
73
74
Appendix
PSH User Manual
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
(1) No.7-1, Jhongsing Rd., Tucheng City, Taipei County, Taiwan (2) No. 69, Lu San Road, Suzhou City (Xin Qu), Jiangsu Sheng, China declare, that the below mentioned product
Type of Product: Power Supply Model Number: PSH-1036A/ 2018A/ 3610A/ 6006A PSH-1070A/ 2035A/ 3620A/ 6012A PSH-10100A/ 2050A/ 3630A/ 6018A
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (89/336/EEC, 92/31/EEC, 93/68/EEC) and Low Voltage Directive (73/23/EEC, 93/68/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Directive, the following standards were applied:
EMC
EN 61326-1: Electrical equipment for measurement, control and laboratory use –– EMC requirements (1997 + A1:1998 + A2:2001 + A3:2003) Conducted Emission Radiated Emission EN 55011: Class A 1998 + A1:1999 + A2:2002 Current Harmonics EN 61000-3-2: 2000 + A2:2005 Voltage Fluctuations EN 61000-3-3: 1995 + A1:2001
------------------------- Surge Immunity
------------------------- Conducted Susceptibility
------------------------- Power Frequency Magnetic Field
------------------------- Voltage Dip/ Interruption
Safety
Low Voltage Equipment Directive 73/23/EEC Safety Requirements
IEC/EN 61010-1: 2001
Electrostatic Discharge EN 61000-4-2: 1995 + A1:1998 + A2:2001
Radiated Immunity EN 61000-4-3: 2002 + A1:2002 Electrical Fast Transients EN 61000-4-4: 2004
EN 61000-4-5: 1995 + A1:2001
EN 61000-4-6: 1996 + A1:2001
EN 61000-4-8: 1993 + A1:2001
EN 61000-4-11: 2004
Index
A
AC input
cable assembly................................... 22
cord requirement............................... 23
terminal............................................. 18
C
calibration
certificate........................................... 15
CC
characteristic..................................... 45
cleaning PSH ....................................... 5, 7
current
OCP setting ....................................... 49
output setting.................................... 47
CV
characteristic..................................... 45
D
display
contrast setting ................................. 50
default mode...................................... 19
fine/coarse.......................................... 17
menu mode ...................................20, 44
output On/Off .................................... 16
remote control mode.......................... 17
E
EN 55011 ............................................... 75
EN 61000 ............................................... 75
EN 61010 ........................................6, 8, 75
F
faq .......................................................... 69
feature list ............................................. 12
front panel.............................................. 16
faq ...................................................... 69
functionality check
OCP.................................................... 38
output current ................................... 37
output voltage.................................... 35
OVP.................................................... 35
setup .................................................. 34
fuse
rating ................................. 7, 72, 73, 74
safety instruction................................. 7
G
GPIB
interface setting................................. 55
I
IEEE remote control
general commands............................. 58
miscellaneous commands.................. 60
status commands...............................59
syntax ................................................ 57
O
OCP
check .................................................. 38
operation environment
safety instruction............................. 5, 8
output
cable assembly................................... 30
cable selection.................................... 25
current check ..................................... 37
current setting................................... 47
faq ...................................................... 69
load configuration.............................. 26
terminal ............................................. 18
75
76
Index
voltage check ......................................35
voltage setting..............................43, 46
OVP
check...................................................35
P
package contents ....................................14
power supply
safety instruction ................................ 7
protection
OCP setting ........................................49
OVP setting ........................................48
PSH lineup .............................................13
R
rack mounting
assembly method................................41
parts list .............................................40
rear panel ...............................................18
remote control
display mode.......................................17
terminal..............................................18
remote sensing
cable assembly....................................32
cable selection ....................................25
configuration ..................................... 26
theory................................................. 24
S
safety
guidelines............................................. 6
symbol.................................................. 6
UK power cord..................................... 9
sound setting ......................................... 51
specification ........................................... 71
storage environment
safety instruction ............................ 5, 8
T
table of contents....................................... 3
V
verification
equipment.......................................... 64
output voltage.................................... 66
setup .................................................. 63
voltage
output setting .................................... 46
OVP setting ....................................... 48
77
Loading...