This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the instrument or to
other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Frame or Chassis Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to ensure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
4
SAFETY INSTRUCTIONS
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
General
Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening.
Do not disassemble the GSB-01/GSB-02 unless
you are qualified.
(Measurement categories) EN 61010-1:2010 specifies the
measurement categories and their requirements as follows. The
GSB-01/GSB-02 falls under category II.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Power Supply
WARNING
AC Input voltage range: 100 - 240VAC ±10%
Frequency: 50Hz/60Hz
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
Safety Guidelines
5
GSB-01/GSB-02 Series User Manual
Fuse
WARNING
Fuse Type: T 2A/250V
To ensure fire protection, replace the fuse only
with them specified type and rating.
Disconnect the power cord before replacing the
fuse.
Make sure the cause of the fuse blowout is fixed
before replacing the fuse.
Cleaning the
GSB-01/GSB-02
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees
and their requirements as follows. The GSB-01/GSB-02 falls under
degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
6
SAFETY INSTRUCTIONS
Storage
environment
Location: Indoor
Temperature: -10°C to 70°C
Relative Humidity: ≤ 85% (no condensation)
Disposal
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
7
GSB-01/GSB-02 Series User Manual
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the scanner box in the United Kingdom, make sure the
power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
As the colours of the wires in main leads may not correspond with
the coloured marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to
the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cable
of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any exposed wiring from a cable, plug or connection that is
engaged in a live socket is extremely hazardous. If a cable or plug is
deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be
immediately destroyed and replaced in accordance to the above
standard.
8
GETTING STARTED
Multiplex Scanner Box
GSB-01
CAUTION
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL
POWER
1 2 3 4 5 6 7 8
PASS
FAIL
HIGH VOLTAGE
5.0 kVAC MAX.
6.0 kVDC MAX.
RETURN
GSB-01 and GSB-02 Overview .......................................... 10
This chapter describes the scanner box in a
nutshell, including its main features and front /
rear panel introduction. After going through the
overview, please read the safety considerations.
9
GSB-01/GSB-02 Series User Manual
Model name
ACW
DCW
IR
GB
Outputs
GSB-01
8 x HV
GSB-02
6 x HV, 2 x GB
GSB-01 and GSB-02 Overview
Lineup
The aim of these scanner boxes is to allow multiple DUTs to be
tested either concurrently or in sequence using the GPT-9800, GPT9900 or GPT-9900A safety testers. The scanner boxes are
particularly well suited for multi-point safety testing as well for
volume testing on factory floors.
The GSB-01 has connections for ACW, DCW and IR testing, while
the GSB-02 also includes support for GB testing.
Firmware Note
Please make sure the firmware is up to date before using the
scanner boxes. Please see the user manual to check the firmware
version.
GPT-9800: firmware version V3.0 or above
GPT-9900/9900A: firmware version V2.0 or above
Note: Throughout this user manual, the terms scanner box or GSB
will refer to either model (GSB-01, GSB-02) unless specifically
stated otherwise. GPT-9000 will refer to any of the GPT-9800, GPT9900 or GPT-9900A safety testers, unless stated otherwise. HV and
H will refer to High Voltage terminals, while LO and L will refer to
the return terminal.
Model Overview
10
GETTING STARTED
Performance
8 HV outputs (6 for GSB-02)
2 GB outputs (GSB-02 only)
ACW: 5kV AC
DCW: 6kV DC
IR: 1kV DC
GB: 40A (GSB-02 only)
Features
PASS/FAIL LEDs
HI LO LEDs
Up to 4 scanner boxes can be connected
Interface
RS-232 interface.
Standard
Accessories
Part Number
Description
N/A
GSB-01/02 unit
N/A
Quick start guide
N/A
User manual CD
Region dependent
Power cord
GHT-108 x1
High voltage wiring leads
GHT-109 x1
GB wiring leads (GSB-02 only)
GHT-116R
x8(GSB-01), x6(GSB-02)
HV test leads for scanner box
outputs
GHT-116B x1
Return test lead for scanner
box
Main Features
Accessories and Package Contents
11
GSB-01/GSB-02 Series User Manual
GTL-116R x2
GB sense/source H test lead
(GSB-02 only)
GTL-116B x1
GB sense/source L test lead
(GSB-02 only)
GTL-235
RS232C Cable
Note
Keep the packaging, including the box, polystyrene
foam and plastic envelopes should the need arise to
return the unit to GW Instek.
12
GETTING STARTED
Multiplex Scanner Box
GSB-01
CAUTION
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL
POWER
12345678
PASS
FAIL
HIGH VOLTAGE
5.0 kVAC MAX.
6.0 kVDC MAX.
RETURN
HI / LO Indicators
HIGH VOLTAGE
Indicator
POWER Button
Channel Pass/Fail Indicators
Return Terminal
SOURCE HSOURCE L
SENSE H SENSE L
RETURN
HI-POT
IR
GSB-02
Multiplex Scanner Box
GB Rx
POWER
H I
1CHANNEL
PASS
FAIL
LO2LO
3
H I
4LO5LO6
H I7H I
8
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
LOLOLOLO
H IH IH IH I
HI / LO Indicators
HIGH VOLTAGE
Indicator
POWER Button
Channel Pass/Fail Indicators
Return
Terminal/
SENSE L
SOURCE L
SOURCE HSENSE H
Appearance
GSB-01 Front Panel
GSB-02 Front Panel
13
GSB-01/GSB-02 Series User Manual
POWER switch
POWER
Turns the unit on or off. It is
recommended that the unit is
powered up before the connected
safety tester is turned on.
HI LO Indicators
The HI/LO Indicators indicate whether a channel
is set to the HI or LO output. If neither HI nor LO
is lit, it indicates that that channel is disabled.
Channel
Indicators
The Channel Indicators will become lit green on a
pass judgment or red on a fail judgment.
HIGH VOLTAGE
output terminal
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
The HIGH VOLTAGE
terminal output is used
for outputting the
testing voltage. The
terminal is recessed for
safety. This terminal is
used in conjunction
with the RETURN
terminal.
WARNING
USE EXTREME CAUTION.
Do not touch the HIGH VOLTAGE terminal
during testing.
RETURN terminal
RETURN
The RETURN terminal is
used for IR, DCW and ACW
tests.
14
GETTING STARTED
RETURN, SENSE
and SOURCE
terminals
GSB-02 Only
SOURCE HSOURCE L
SENSE HSENSE L
RETURN
GB Rx
The RETURN terminal
is used for IR, DCW
and ACW tests.
The SOURCE H,
SOURCE L, SENSE H
and SENSE L terminals
are used for GB tests.
15
GSB-01/GSB-02 Series User Manual
HV1HV2
RETURN
CH1
CH2
CH3
CH4CH6
CH5CH7
CH8
GND
ENSURE THE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T 2A 250V
RS232/IN
RS232/OUT
SER. NO. LB
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
HV1HV2
CH1
Return
CH3CH5
CH2CH4
CH7
RS232/INRS232/
OUT
CH6CH8GND
Line voltage input & fuse
HV1HV2
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
CH3
CH4CH6
CH5
CH7
GND
ENSURE THE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T 2A 250V
RS232/IN
RS232/OUT
CH8
CH1
CH2
GB Rx
SENSE H
SOURCE H
SENSE L
SOURCE L
HI-POT
IR
RETURN
SER. NO. LB
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
HV1HV2
CH1
Return/ SENSE L
CH3CH5
CH2CH4
CH7
RS232/INRS232/
OUT
CH6CH8GND
Line voltage input & fuse
SOURCE H SOURCE L
SENSE H
GSB-01 Rear Panel
GSB-02 Rear Panel
16
GETTING STARTED
HV1, HV2
The HV1 input terminal is used as
the primary high voltage input for
the scanner box. The HV2 output
terminal is used to daisy chain the
high voltage output to the next
scanner when multiple scanner
boxes are used together.
CH1 ~ CH8
These channels are connected to the
DUTs that are to be tested. The
output state is configured from the
GPT-9000 safety tester master.
Note: For the GSB-02 CH7 and CH8
are used for ground bond testing
only.
RS232/IN
RS232/IN
These two RS232 interfaces are the
communication links between the
master (safety tester) to slave
(scanner box). They are also used to
connect multiple scanner boxes in a
daisy chain.
In addition the RS232/IN port is
also used to perform firmware
updates.
RS232/OUT
RS232/OUT
Line voltage input
T 2A 250V
Line voltage input: 100-240VAC,
50/60Hz.
Line voltage fuse
Line voltage fuse:
T 2A/250V
GND
GND
Connect the GND (ground)
terminal to the earth ground to
ensure operational safety.
17
GSB-01/GSB-02 Series User Manual
Background
The GPT-9000 is a high voltage instrument that
outputs dangerous voltages. The following
section describes precautions and procedures
that must be followed to ensure a safe work
environment when GSB-01/GSB-02 is
connected to the GPT-9000 series safety testers.
WARNING
The GPT-9000 generates voltages in excess of
5kVAC or 6kVDC. Follow all safety precautions,
warnings and directions given in the following
section when using the instrument with the
scanner boxes.
1. Only technically qualified personnel should be
allowed to operate the safety tester and scanner
box(es).
2. The operating workplace must be fully isolated,
especially when the instrument is in operation.
The instruments should be clearly labeled with
appropriate warning signage.
3. The operator should not wear any conductive
materials, jewelry, badges, or other items, such
as wrist watches.
4. The operator should wear insulation gloves for
high voltage protection.
5. Ensure the earth ground of the line voltage is
properly grounded.
Safety Considerations
Workplace Precautions
18
GETTING STARTED
6. Ensure any devices that are adversely affected
by magnetic fields are not placed near the tester
and scanner box(es).
19
GSB-01/GSB-02 Series User Manual
Background
The GPT-9000 is a high voltage instrument that
outputs dangerous voltages. The following
section describes precautions and procedures
that must be followed to ensure that the tester
along with any conned scanner boxes are
operated in a safe manner.
WARNING
The GPT-9000 generates voltages of up to 5kVAC
or 6kVDC. Follow all safety precautions, warnings
and directions given in the following section when
using the instrument.
1. Never touch the safety tester, lead wires,
terminals, probes, connected scanner box(es) or
any other connected equipment when the tester
is testing.
2. Do not turn the safety tester on and off quickly
or repeatedly. When turning the power off,
please allow a few moments before turning the
power back on. This will allow the protection
circuits to properly initialize.
Do not turn the power off when a test is
running, unless in an emergency.
3. Only use those test leads supplied with the
instrument. Leads with inappropriate gauges
can be dangerous to both the operator and the
instrument.
For GB testing, never use the Sense leads on the
SOURCE terminals.
4. Do not short the HIGH VOLTAGE terminal
with ground. Doing so could charge the chassis
to dangerously high voltages.
Operating Precautions
20
GETTING STARTED
5. Ensure the earth ground of the line voltage is
properly grounded.
6. Only connect the test leads to the HIGH
VOLTAGE/SOURCE H/SENSE H terminals
before the start of a test. Keep the test leads
disconnected at all other times.
7. Always press the STOP button when pausing
testing.
8. Do not leave the safety tester unattended.
Always turn the power off when leaving the
testing area.
9. When remotely controlling the safety tester,
ensure adequate safety measures are in place to
prevent:
Inadvertent output of the test voltage.
Accidental contact with the instrument during
testing. Ensure that the instrument and DUT are
fully isolated when the instrument is remotely
controlled.
10. Ensure an adequate discharge time for the
DUT.
When DCW or IR tests are performed, the DUT,
test leads and probes become highly charged.
The GPT-9000 has discharge circuitry to
discharge the DUT after each test. The time
required for a DUT to discharge depends on the
DUT and test voltage.
Never disconnect the safety tester before a
discharge is completed.
21
GSB-01/GSB-02 Series User Manual
Background
The GSB-01/GSB-02 are used with high voltage
devices and as such, daily safety checks should
be made to ensure safe operation.
1. Ensure all test leads are not broken and are free
from defects such as cracks or splitting.
2. Ensure the scanner box(es) are always
connected to an earth ground.
3. Test the operation of the safety tester + scanner
box(es) with a low voltage/current output:
Ensure the safety tester generates a FAIL
judgment when the HIGH VOLTAGE and
RETURN terminals are shorted (using the
lowest voltage/current as the testing
parameters).
WARNING
Do not use high voltages/currents when the HIGH
VOLTAGE and RETURN terminals are shorted. It
may result in damage to the instrument.
Basic Safety Checks
22
OPERATION
Menu Tree Additions ....................................................... 24
Test Lead Connection ...................................................... 25
Connecting GSB-01 Units ....................................................................... 25
Create an Automatic Test ....................................................................... 51
Running an Automatic Test ................................................................... 54
Automatic Test Results ........................................................................... 58
OPERATION
23
GSB-01/GSB-02 Series User Manual
status
SET
T
PASS/FAIL result
status
OTS
P
status
YAER
D
status
EIV
W
status
IDE
T
Press
START
Press
STOP
Press
MANU/
AUTO
Scan for
connected
scanner boxes
Press
STOP
Press
STOP
Press
EDIT/
SAVE
Press
STOP
2
PAGE View
(AUTO mode
only)
1
Press
UTILITY
Press
ESC
Switch to AUTO
mode
MANU UTILITY
menu (MANU
mode only)
1
Press
UTILITY
1 Press EDIT/SAVE to save settings, or ESC to cancel and return to the previous screen.
2 Press the STOP key twice for a FAIL result.
3 When in MANU mode, selecting MANU number 000 will enter the special manual mode.
4 The Sweep mode function is only accessible in the special manual mode.
Common Utility
Settings
1
Switch to MANU
mode
Special manual
mode
Press
PAGE
Hold
MANU/
AUTO
MANU
no.
# 000
3
SCAN Configuration
(MANU mode only)
1
Press
ESC
Save the SCAN
configuration
Press
EDIT/
SAVE
PAGE View
(AUTO mode
only)
1
Press
PAGE
CA NS
Press
Press
ESC
Press
PAGE
Menu Tree Additions
When the scanner boxes are added to the GPT-9000 safety tester the
scan utility and the scanner configuration menu become available.
These additional menu functions are highlighted in the menu tree
below.
24
OPERATION
Background
The following will describe how to connect
scanner boxes to a GPT-9000 safety tester. Up to
4 scanner boxes can be connected. When wiring
the scanner boxes to the safety tester or to each
other, only the HV wiring leads (GHT-108)
should be used.
When connected, all the high voltage and all
the return terminals will have been connected
in a daisy chain manner.
WARNING
Ensure the safety tester is off when connecting the
scanner boxes to the safety tester.
Front Panel
1. Connect the high voltage terminal on the safety
tester to the high voltage terminal on the 1st
scanner box, as shown below.
2. Connect the return terminal on the safety tester
to the return terminal on the 1st scanner box.
3. Connect the return terminal on the 2
nd
scanner
box to the return terminal on the 3rd scanner
box.
Test Lead Connection
This section describes how to connect the GPT-9000 to a number of
scanner boxes. It is recommended that only models of the same
type are connected together.
ENSURE THE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACINGTHE FUSE
GPIB
RS232 RS232/IN
HV2 HV1
Return Return
RS232/OUT
RS232/IN
Safety Tester
Scanner box #1
HV1HV2
RETURN
CH1
CH2
CH3
CH4CH6
CH5CH7
CH8
GND
ENSURE THE POWERISREMOVED FROM
THE INSTRUMENT BEFORE REPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T2A250V
RS232/IN
RS232/OUT
SER. NO. LB
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
RS232/OUT
RS232/IN
Scanner box #2
Scanner box #3
HV1HV2
RETURN
CH1
CH2
CH3
CH4CH6
CH5CH7
CH8
GND
ENSURE THE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T 2A 250V
RS232/IN
RS232/OUT
SER. NO. LB
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
HV2 HV1
Return Return
Scanner box #4
RS232/OUT
RS232/IN
28
OPERATION
Background
The following will describe how to connect
scanner boxes to a GPT-9000 series safety tester
with ground bond test support. Up to 4 scanner
boxes can be connected. In the examples below
only 3 scanner boxes are connected.
When wiring the scanner boxes to the safety
tester or to each other, only the GB wiring leads
(GHT-109) should be used.
Front Panel
1. Connect the SOURCE H and SOURCE L
terminals on the safety tester to the same
terminals on the 1st scanner box.
2. Connect the SENSE H and SENSE L terminals
on the safety tester to the same terminals on the
1st scanner box.
Connecting GSB-02
29
GSB-01/GSB-02 Series User Manual
POWER
STARTSTOP
REMOTE
MANU/AUTO EDIT/SAVE UTILITY
ESCPAGE
GPT-9804
Tester
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
Rx
SENSE H
SOURCE H
SENSE L
SOURCE L
RETURN
AC / DC Withstanding Voltage /
Insulation Resistance / Ground Bond
PASSFAILREADY TEST
SOURCE H SOURCE L
SENSE H SENSE L
RETURN
HI-POT
IR
GSB-02
Multiplex Scanner Box
GB Rx
POWER
HI
1CHANNEL
PASS
FAIL
LO2LO
3
HI
4LO5LO6
H I7H I
8
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
LOLOLO LO
HIHIHIHI
SOURCE H SOURCE L
SENSE H SENSE L
RETURN
HI-POT
IR
GSB-02
Multiplex Scanner Box
GB Rx
POWER
HI
1CHANNEL
PASS
FAIL
LO2LO
3
HI
4LO5LO6
H I7H I
8
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
LOLOLO LO
HIHIHIHI
SOURCE H SOURCE L
SENSE H SENSE L
RETURN
HI-POT
IR
GSB-02
MultiplexScanner Box
GB Rx
POWER
HI
1CHANNEL
PASS
FAIL
LO2LO
3
HI
4LO5LO6
H I7HI
8
HIGHVOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
LOLOLO LO
HIHIHIHI
SOURCE L
Safety Tester
Scanner box #1
Scanner box #1
Scanner box #1
SOURCE H
SENSE L
SENSE H
Rear Panel
1. Connect the SOURCE H and SOURCE L
terminals on the 1st scanner box in series to the
same terminals on the 2nd and 3rd scanner box.
2. Connect the SENSE H and SENSE L terminals
on the 1st scanner box in series to the same
terminals on the 2nd and 3rd scanner box.
3. Connect the RS232 ports from the safety tester
to each scanner box in a daisy chain using the
RS-232C cables.
ENSURE THE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACINGTHE FUSE
GPIB
RS232 RS232/IN
RS232/OUT
RS232/IN
RS232/OUT
RS232/IN
Safety Tester
Scanner box #1
Scanner box #2
Scanner box #3
SOURCE L
SOURCE H
SENSE L
SENSE H
SOURCE L
SOURCE H
SENSE L
SENSE H
31
GSB-01/GSB-02 Series User Manual
Background
The terminals on the rear panel of the scanner
boxes are divided into two sections, input and
output.
The input section is used (as shown previously)
to daisy chain scanner boxes together.
The output section is used to connect up to 8
DUTs. The GSB-01 is only used for DCW, ACW
and IR testing. The GSB-02 (shown) replaces
two high voltage terminals with a pair of
SENSE H/SOURCE H terminals for GB testing.
For all 8 channels (GSB-01) or channel 1~6(GSB-
02) the terminals can be assigned as either HV
or LO (High voltage or return) terminals on the
GPT safety tester. This allows the scanner boxes
to be quite flexible to your testing needs.
NOTE
The GB outputs (channel 7 and 8) can only be
configured as HI terminals or be disabled.
Input and Output
Areas on the Rear
Panel
HV1HV2
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
CH3
CH4CH6
CH5
CH7
GND
ENSURE THE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T 2A 250V
RS232/IN
RS232/OUT
CH8
CH1
CH2
GB Rx
SENSE H
SOURCE H
SENSE L
SOURCE L
HI-POT
IR
RETURN
SER. NO. LB
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
InputOutput
Examples
Below are some possible DUT test setups as
examples. Note that you are not limited to
those shown below.
DUT Connection
32
OPERATION
Example 1: GB connection with common LO terminals. GTL-116B and
GTL-116R test leads are to be used.
GTL-116RGTL-116B
HV1HV2
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
CH3
CH4CH6
CH5
CH7
GND
ENSURETHE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T 2A 250V
RS232/IN
RS232/OUT
CH8
CH1
CH2
GB Rx
SENSE H
SOURCE H
SENSE L
SOURCE L
HI-POT
IR
RETURN
SER.NO. LB
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
GB Output connection
DUT
DUT
Example 2: IR, ACW, DCW connection with separately configured HI &
LO terminals. GHT-116R and GHT-116B test leads are to be used.
GHT-116BGHT-116R
HV1HV2
RETURN
CH1
CH2
CH3
CH4CH6
CH5CH7
CH8
GND
ENSURE THE POWER ISREMOVED FROM
THE INSTRUMENT BEFOREREPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T 2A 250V
RS232/IN
RS232/OUT
SER. NO. LB
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
CH1 ~ CH4 configured as HI outputs
CH5 ~ CH8 configured as LO terminals
DUT
DUT
DUT
DUT
33
GSB-01/GSB-02 Series User Manual
Example 3: IR, ACW, DCW connections with common return terminal.
These tests are best performed sequentially. GHT-116R and GHT-116B
test leads are to be used.
GHT-116R
GHT-116B
HV1HV2
RETURN
CH1
CH2
CH3
CH4CH6
CH5CH7
CH8
GND
ENSURE THE POWER ISREMOVED FROM
THE INSTRUMENT BEFOREREPLACING THE FUSE
100 240VAC
50/60Hz 50VA MAX.
T 2A 250V
RS232/IN
RS232/OUT
SER. NO. LB
OUTPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
INPUT
HIGH VOLTAGE
MAX.
6.0 kVDC
5.0 kVAC
CH1 to CH8 configured as HI
DUT
DUT
DUT
DUT
DUT
DUT
DUT
DUT
34
OPERATION
Background
The power up sequence as well as the
connections between each scanner box and the
safety tester are critical to operate the scanners
correctly.
Steps
1. Make sure the safety tester and all scanner
boxes are turned off.
2. Connect each scanner box to a master safety
tester in a daisy chain, as shown on page 25.
Make sure the RS232 cables are connected to
the correct ports.
Make sure the Input and Output cables are
connected properly.
3. Turn on the scanner boxes from the
power switches.
Each power LED will be lit red.
POWER
4. Turn the GPT-9000 power on. It is
essential that the GPT-9000 is turned
on after the scanner boxers.
POWER
Startup Screen
S y s e m t
H a r w a rd
e C
h
C h e c i g . .n.
F i r w a rm
e C h
SYST E S ME L F T E S T
k
e c i g . .n.k
e c i g . .n.k
Start Up Procedure
Startup
35
GSB-01/GSB-02 Series User Manual
Background
The scanner connection can be checked in the
Common Utility menu.
Steps
1. Ensure the tester is in VIEW status.
I R G B
E
A CWDC W
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
W
MM A UN N A_0M A NU = * * * - 0 0
m A
2.Press the UTILITY key.
UTILITY
3. Press the SCAN soft-key to view
any connected scanner boxes. Any
scanner boxes will be displayed in
order from 1 to 4.
C A NS
Example
S C A NN E B ORX 2 : - - - ---
S C A NN E B ORX 1 : G B - 0S1
S C A NN E B ORX 3 : - - - --S C A NN E B ORX 4 : - - - ---
C A NSL C DB U Z Z N T EIR C T R L
MOCMO U NT I L I T Y
In the example above, only one scanner box
(model: GSB-01) is connected.
Note
If the any connected scanners are not properly
connected they will not appear in the Scan Utility.
Connection Check
36
OPERATION
Note
Setting the test settings, test mode and general
operation will not be discussed in this manual.
Please see the GPT-9000 User Manual for more
operation details.
Creating a Test Setup
This section describes how to create, edit and run tests using the
scanner box interfaces. In principal we will be showing you how to
configure the output terminals on the scanner box rear panel. For
the GSB-01, each terminal can be configured as either an HV output
(hereafter referred to as HI) or as a return terminal (hereafter
referred to as LO). For the GSB-02, channels 1 to 6 can be
configured to HI or LO terminals as well. However for channels 7
and 8, the GSB-02 can also configure these channels into pairs of
Source/Sense HI.
Thus each terminal on the scanner boxes must be configured based
on the desired test setup, as referred to in the previous chapter on
page 25.
Before operating the safety tester please read the safety precautions
as outlined in the Safety Considerations chapter on page 18.
37
GSB-01/GSB-02 Series User Manual
Background
The flowchart below shows the basic workflow
for creating tests for the scanner boxes.
Configure the
test settings for
the Selected
Manu test
Select Manu Test
Number
Configure the
outputs of each
scanner for the
Selected test
For Auto Tests,
repeat the above
steps for each
Manu Test that
will be included in
the Auto Test
Save the scanner
configuration
Background
To edit any of the manual test settings, the
tester must be in EDIT status.
Any settings or parameters that are edited only
apply to the currently selected MANU number.
Scanner Box Test Creation Workflow
Select a Manual Test
38
OPERATION
Steps
1. If the tester is in AUTO mode,
press and hold the MANU/AUTO
key for three seconds to switch to
MANU mode.
The tester can only switch between
AUTO and MANU mode when in
the VIEW status.
MANU/AUTO
2. Use the scroll wheel to set the
MANU number.
I R G B A CWD CW
E
FR E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 mAIHES = 0T
0 0 . 1 S0AM P =R
W2M A NU = * * * - 0 0
MANU number
m A
3. Press the EDIT/SAVE key when in
VIEW status to enter the EDIT
status. This will enter the EDIT
status for the chosen test number
only.
EDIT/SAVE
I R G B A C WD CWI / LHOI M ETR
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0MM A UN N A_
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
T2M A N U = * * * - 0 0
m A
4. The Status changes from VIEW to EDIT.
Note
Pressing the EDIT/SAVE key again will save the
settings for the current test and return back to
VIEW status.
39
GSB-01/GSB-02 Series User Manual
Background
After a MANU number has been chosen and
the tester is in EDIT status, the settings for the
current manual test can be configured.
Steps
1. To choose the test function, press the ACW,
DCW, IR or GB soft-keys.
A CWD C WI R
G B
2. The test function soft-key is highlighted.
3. Press the UP / DOWN arrow keys
to bring the cursor to a function
setting.
4. Use the scroll wheel to set the
value of the function setting.
Example
I R G B A CWD CWI / LHOI M ETR
E
F RE Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 mAIHES = 0T
0 0 . 1 S0AM P =R
T2M A NU = * * * - 0 0
cursor
m A
5. Repeat steps 3 and 4 for the remaining settings.
6. Press the EDIT/SAVE key to save
and save the manual test and go
back to VIEW status.
EDIT/SAVE
Note
For detailed instructions on how to set the various
function settings, please see the user manual.
Configure the Manual Test Settings
40
OPERATION
Background
The scanner box output is configured
separately for each manual test. This allows you
to have one manual test to test multiple DUTs
at the same time from a number of scanner
boxes.
For automatic tests, each manual test can be
seen as configuring the output of one step of
the automatic test.
This section will assume that you are only
configuring a single manual test. For automatic
tests, repeat the instructions below for each
manual test that is added to the automatic test.
Note
For the GSB-02, the GB outputs can only be set to
HI or disabled.
Note
The following settings will only apply to the current
manual test.
Overview
c h 2 c h c h3
Cursor
S E T T I C H NA E LN 2M AN U = * * * - 0 0
c h 1 4 c
G S B - 0 1 :
G S B - 0 2 :
G S B - 0 1 :
G S B - X X :
Selected MANU test
file number
Connected scanner
boxes from 1~4
Channel settings
c h 7 c hc 6 h5h
N G
8
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
L
H
X
X
E N DS
N I TI
InitializeShow
configuration
Steps
1. Press the EDIT/SAVE key when in
VIEW status to enter the EDIT
status. This will enter the EDIT
status for the currently selected
manual test.
EDIT/SAVE
Configuring the Scanner Box Outputs
41
GSB-01/GSB-02 Series User Manual
2. Press the PAGE key to bring up the
Scanner Configuration Page view
for the currently selected manual
test.
PAGE
3. Press the UP/DOWN and
LEFT/RIGHT arrows keys to move
the cursor to the desired channel
and scanner box.
4. Use the scroll wheel to set the
selected channel on the selected
scanner box as H or L or X.
H
Sets the channel as a HV terminal (or
Source H/Sense H terminals for GSB-02)
L Sets the channel as a return terminal
(For the GSB-02, the GB terminals can only
be configured to H or X(disabled))
X Disables the channel
Example
c h 2 c h c h3
Cursor
S E T T I C H NA E LN 2M AN U = * * * - 0 0
c h 1 4 c
G S B - 0 1 :
G S B - 0 2 :
G S B - 0 1 :
G S B - X X :
Connected scanner
boxes
Not connected
Channel settings
c h 7 c hc 6 h5h
N G
8
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
H
H
X
X
L
H
X
X
E N DS
N I TI
In the example above scanner box#1 has ch1~7
configured as H and ch8 configured as L.
Scanner box#2 has ch1~8 configured as H.
5. Press the EDIT/SAVE key to save
the scanner output settings for the
current manual test.
EDIT/SAVE
42
OPERATION
6. A test can now be started, see page on 44 to get
started.
Initialize
7. Pressing the INIT key will initialize
all the channels to “X”, disabling
all the channels.
N I TI
Send
8. Pressing SEND will output the
channel settings onto the
connected scanner boxes.
E N DS
Note
If a number of terminals are turned on at the same
time, then the voltage/current that is set on the
safety tester must be divided by the number of
channels that are turned on for the test. For
example, if channels 1 to 3 are turned on for an
ACW test, then the test current must be divided by
3 to determine the result from each channel
(assuming identical DUTs).
43
GSB-01/GSB-02 Series User Manual
Background
A test can be run when the tester is in READY
status.
Note
The tester cannot start to run a test under the
following conditions:
A protection setting has been tripped; when a
protection setting has been tripped the
corresponding error message is displayed on
the screen.
The INTERLOCK function is ON and the
Interlock key is not inserted in the signal I/O
port.
The STOP signal has been received remotely.
If Double Action is ON, ensure the START
button is pressed immediately after the STOP
button (<0.5s).
Note
When a test is running the voltage output cannot
be changed, unless the test is under the special
manual mode.
Steps
1. Ensure a MANU test is selected and the tester is
in VIEW status.
I R G B
m A
A CWD CW
E
FR E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0AM P =R
W
VIEW status
MM A UN N A_2M A NU = * * * - 0 0
Selected manual test
Running a Manual Test
44
OPERATION
2. Press the STOP button to put the
tester into the READY status.
STOP
I R G B
m A
A CWD CW
E
FR E Q = 0 H z 6
0
100
k V
AER
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0AM P =R
D
MM A UN N A_2M A NU = * * * - 0 0
READY status
Y
3. The READY indicator will be lit
blue when in the READY status.
READY
4. Press the START button when the
tester is in the READY status. The
manual test starts automatically
and the tester goes into the TEST
status.
START
5. The TEST indicator will be lit
orange when in the TEST status.
TEST
I R G B A CWD CW
E
FR E Q = 0 H z 6
0
100
k V
SET
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0AM P =R
MM A UN N A_2M A NU = * * * - 0 0
TEST status
T
m A
00 00
45
GSB-01/GSB-02 Series User Manual
6. The HI or LO indicators of the channels that are
turned on for the manual test will be shown on
the respective scanner boxes.
Multiplex Scanner Box
GSB-01
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL2345678
PASS
FAIL
1
Ch8 onCh1 on
7. The test will start by showing the remaining
ramp up time, followed by the remaining test
time. The test will continue unit the test is
finished or the test is stopped.
I R G B A CWD CW
E
F R E Q = 0 H z 6
0
100
k V
SET
MT I E = 0 0 3 . 2 SR
0 0 m AER F =# 0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0AM P =R
MMA UN N A_2M A N U = * * * - 0 0
T
remaining TIMER time
00 37
m A
remaining RAMP time
Stop the Test
8. To stop the test at any time when it
is running, press the STOP button.
The test will stop immediately.
When the STOP button is pressed,
a judgment is not made on the test.
All panel keys except the STOP
button are locked when the tester
is in STOP status.
STOP
I R G B A CWDC W
E
F R E Q = 0 H z 6
0
100
k V
OTS
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
MM A UN N A_2M A NU = * * * - 0 0
STOP status
P
m A
00 00
46
OPERATION
9. To put the tester back into READY
status, press the STOP button
again.
STOP
Exit TEST Status
To exit testing, press the
MANU/AUTO key when the
tester is in the READY status. The
tester will revert to the VIEW
status for the current test.
MANU/AUTO
I R G B
m A
A CWDC W
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
W
MM A UN N A_2M A NU = * * * - 0 0
Note
Do not touch any terminals, test leads or any other
connections when the test is on.
47
GSB-01/GSB-02 Series User Manual
Background
If the test is allowed to run to completion (the
test is not stopped or a protection setting is not
tripped) then the tester will judge the test(s) as
either PASS or FAIL.
The pass/fail result of the test(s) is shown on
the safety tester display and the scanner box(s).
Please note that when multiple DUTs are tested
simultaneously and a fail result is produced, it
indicates that at least one DUT failed the test. It
doesn’t indicate which of the DUTs passed or
failed. Each DUT would have to be tested
separately to determine that.
Note
The test will be judged PASS when:
The HI SET and LO SET limits have not been
tripped during the test time.
The test will be judged FAIL when:
Either the HI SET or LO SET limit has been
tripped during the test time.
A protection setting has been tripped during
the test time.
Test may not finish if the scanner boxes are not
properly connected.
MANU Test Results
48
OPERATION
PASS Judgment
1. When the test is judged as PASS,
PASS will be displayed, the buzzer
will sound, the PASS indicator will
be lit green and the channel
indicator will be lit green.
PASS
Safety tester
example
I R G B A C WD CW
E
F R E Q = 0 H z 6
0
100
k V
SAP
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
MM A UN N A_2M A NU = * * * - 0 0
S
m A
00 37
Scanner display
example
Multiplex Scanner Box
GSB-01
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL2345678
PASS
FAIL
1
Pass
2. The PASS judgment will be held on both the
safety tester and scanner displays until the
STOP or START button is pressed.
Pressing the STOP button will
return the tester to the READY
status.
STOP
Pressing the START button will
restart the test.
START
Note
The buzzer will only sound if the Pass Sound is set
to ON.
The START button is disabled when the buzzer is
beeping.
49
GSB-01/GSB-02 Series User Manual
FAIL Judgment
3. When the test is judged as FAIL,
FAIL will be displayed on the
safety tester, the FAIL indicator
will be lit red, the channel
indicator on the scanner will be lit
red and the buzzer will sound.
As soon as a test is judged FAIL,
power is cut from the terminals.
FAIL
Safety tester
example
I R G B A C WD CW
F R E Q = 0 H z 6
0
100
k V
IAF
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
2M A UN0 0_2M A NU = * * * - 0 0
L
m A
01 37
Scanner display
example
Multiplex Scanner Box
GSB-01
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL2345678
PASS
FAIL
1
Fail
4. The FAIL judgment will be held on
the safety tester display and the
scanner(s) until the STOP button is
pressed. Pressing the STOP button
twice will return the tester to the
READY status.
STOP
5. The READY indicator will be lit
blue in the READY status.
READY
I R G B
m A
A CWD CW
E
F R E Q = 0 H z 6
0
100
k V
AER
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
D
MM A UN N A_2M A NU = * * * - 0 0
READY status
Y
Note
The buzzer will only sound if Fail Sound is set to
ON.
50
OPERATION
Background
It is assumed that you already know how to
create an automatic test. This section will only
give the essential instructions for creating an
automatic test. For more comprehensive details,
please see the GTP-9000 user manual.
Steps
1. If the tester is in MANU mode,
press and hold the MANU/AUTO
key for three seconds. This will put
the tester into Auto mode.
The tester can only switch between
AUTO and MANU mode when in
the VIEW status.
MANU/AUTO
Automatic Tests
This section describes how to create and run automatic tests for the
GSB-01/GSB-02 scanner boxes.
An automatic test comprises of up 16 MANU tests that are run
sequentially. In order to run an automatic test using the GSB01/GSB-02 scanner boxes, the scanner outputs for each manual test must first be configured, and then each manual test in the automatic
test can be stepped through. In other words, the channel outputs
are directed by the manual tests that comprise the automatic test.
It is assumed that you have read and understood how to create an
automatic test on the GPT-9000 safety tester. If you are not sure
how to create and run an automatic test on the safety tester, please
see the GPT-9000 user manual.
Before operating the GPT-9000 please read the safety precautions as
outlined in the Set Up chapter on page 18.
Create an Automatic Test
51
GSB-01/GSB-02 Series User Manual
I R G B A CWD CW
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
M A NU = * * * - 0 0 1
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
VIEW status
MM A UN N A_ E
m A
2. Use the scroll wheel to choose the
AUTO number.
AUTO #
001~100
I R G B A CWD CW
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
AUTO number
m A
3. Press the EDIT/SAVE key when in
VIEW status to enter the EDIT
status. This will enter the EDIT
status for the chosen AUTO
number.
EDIT/SAVE
I R G B
m A
A CWDC WA D D
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0MA U OT N A_
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
T1A U T O = 0 0 1 - 0 0
4. Press the DOWN arrow keys to
bring the cursor to the MANU
number, if it is not already.
I R G B
cursor
MANU number
m A
A CWD CWA D D
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0MM A UN N A_
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
T1A U T O = 0 0 1 - 0 0
52
OPERATION
5. Use the scroll wheel to choose a
MANU number to add to the
automatic test.
MANU number
001~100
6. Press the ADD soft-key to add the
selected manual test to the
automatic test as another step.
A D D
7. Repeat steps 5 and 6 for any other tests that you
wish to add to the automatic test.
8. A maximum of 16 steps can be added. FULL
will be displayed when 16 steps are added.
Note
The test order can be edited in the Page View
menu after the AUTO test is saved. See the user
manual for details.
9. When in EDIT status, press the
EDIT/SAVE key to save the
automatic test. After the test is
saved the tester will revert back to
VIEW status.
EDIT/SAVE
I R G B
m A
A CWDC WA D D
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0MA U OT N A_
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
W1A U T O = 0 0 1 - 0 0
Note
Pressing the EDIT/SAVE key again will return the
tester back to EDIT status for the selected AUTO
test.
53
GSB-01/GSB-02 Series User Manual
Background
Running an automatic test with the GSB-01/02
scanner boxes is the same as running an
automatic test without them.
Note
The tester cannot start to run an AUTO test
under the following conditions:
Any protection modes have been tripped.
The INTERLOCK function is ON and the
Interlock key is not inserted in the signal I/O
port.
The STOP signal has been received remotely.
If Double Action is ON, ensure the START
button is pressed immediately after the STOP
button (<0.5s).
Warning
Do not touch any terminals, test leads or the DUT
when a test is running.
Steps
1. Ensure the tester is in VIEW status.
I R G B A CWD CW
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
m A
2. Press the STOP button to put the
tester into the READY status.
STOP
Running an Automatic Test
54
OPERATION
I R G B A CWD CW
E
F R E Q = 0 H z 6
0
100
k V
AER
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
Y
MA U OT N A_0A U T O = 0 0 1 - 1 0
D
m A
3. The READY indicator will be lit
blue when in the READY status.
READY
4. Press the START button when the
tester is in the READY status. The
AUTO test starts automatically and
the display changes to TEST status.
START
5. The TEST indicator will be lit
orange when in the TEST status.
TEST
6. The HI or LO indicators of the channels that are
turned on for the current step of the test will be
shown on the respective scanner boxes.
Multiplex Scanner Box
GSB-01
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL2345678
PASS
FAIL
1
Ch8 onCh1 on
7. Each test will start by showing the remaining
ramp up time, followed by the remaining test
time. Each manual test of the automatic test will
be tested in sequence until the last test has
finished or the test is stopped.
55
GSB-01/GSB-02 Series User Manual
I R G B A CWD CW
F R E Q = 0 H z 6
0
100
k V
SET
MT I E = 0 0 3 . 2 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
T
remaining TIMER time
00 37
m A
remaining RAMP time
E MA U OTN A_ 1A U T O = 0 0 1 - 0 0
PASS/FAIL HOLD
1. If Pass Hold is set to ON or Fail Mode is set to
HOLD for a manual test, then the tester will
“hold” the testing after a Pass/Fail result for
that particular test.
I R G B A CWD CW
F R E Q = 0 H z 6
0
100
k V
LOH
MT I E = 0 0 3 . 2 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
D
00 37
m A
E MA U OTN A_ 1A U T O = 0 0 1 - 0 0
HOLD status
2. The PASS or FAIL indicator will
also be lit on the safety tester and
the corresponding channel
indicator will be lit red or green on
the scanner box(es). The buzzer
will NOT sound.
PASS
FAIL
Multiplex Scanner Box
GSB-01
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL2345678
PASS
FAIL
1
Channel indicator
3. To continue to the next test after
HOLD is displayed on-screen,
press the START button.
START
56
OPERATION
4. To stop the test when HOLD is
displayed on-screen, press the
STOP button.
STOP
Note
When in HOLD status, only the START and STOP
buttons can be pressed, all other keys are disabled.
Stop a Running
Test
1. To stop the AUTO test at any time
when it is running, press the STOP
button. The AUTO test will stop
immediately. When the STOP
button is pressed, a judgment is
not made on the current test and
any remaining tests are aborted.
All panel keys except the STOP
and START buttons are locked
when the tester has been stopped.
All the results up until when the
AUTO test was stopped are shown
on-screen.
STOP
E MA U OT N A_ *A U T O = 0 0 1 - * *
# 0 1 : F A I# 0 2 : P A S# 0 3 : S T O# 0 4 : - - # 0 5 : - - -# 0 6 : - - -# 0 7 : - - -# 0 8 : - - # 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
L
-
S
-
P
-
-
-
Example of an automatic test that has been
stopped. Dashes (-) indicate aborted test steps.
1. To put the tester back into READY
status, press the STOP button
again.
STOP
57
GSB-01/GSB-02 Series User Manual
Exit Testing
To exit testing, press the
MANU/AUTO key when the
tester is in the READY status. The
tester will revert to the VIEW
status for the current automatic
test.
MANU/AUTO
I R G B A C WD CW
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
m A
Background
If all the test steps are allowed to run to
completion (the AUTO test is not stopped or a
protection setting is not tripped) then the tester
will judge each step as either PASS or FAIL.
This is shown as a table after the automatic test
has finished running. If the test has been
stopped, then any remaining tests will not be
run and thus the AUTO test will not finish
running.
Overview
P A S
P A SSS
P A S
S K ISP
E MA U OTN A_ 1A U T O = 0 0 1 - 0 0
# 0 1 :# 0 2 :# 0 3 :# 0 4 : P A S
# 0 5 :# 0 6 :# 0 7 :# 0 8 : S T O
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
S
P
F A I
F A ILL
FAIL judgmentPASS judgment
skipped stepstep stopped
Note
The PASS/FAIL judgment for an automatic test
as a whole depends on the results of all the
steps (manual tests) that compose the automatic
test:
Each step must be passed for a PASS judgment
Automatic Test Results
58
OPERATION
(excluding skipped tests).
A FAIL result for a single step will result in
FAIL for the whole automatic test.
A STOP. No step can be stopped for a
PASS/FAIL judgment to be made. In other
words, if a test is stopped, it is judged as neither
PASS nor FAIL.
No step can contain an ERROR or ILOCK
message.
E R R
O
P A S
S
E MA U OTN A_ 1A U T O = 0 0 1 - 0 0
# 0 1 :# 0 2 :# 0 3 :
I L O
C
ILOCK messageERROR message
RK
ERROR: Indicates that V, I or R is not correct.
This usually occurs if the testing leads are not
properly connected.
ILOCK: Indicates that the interlock key is
disconnected (if configured to be used).
PASS Judgment
When all the tests have been
judged as PASS, the PASS
indicator will be lit green, each
channel used in the test will be lit
green and the buzzer will sound.
PASS
Multiplex Scanner Box
GSB-01
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL2345678
PASS
FAIL
1
Green channel indicators,
indicating a pass
Example
P A S
P A SSS
P A S
P A SSS
E MA U OT N A_ *A U T O = 0 0 1 - * *
# 0 1 :# 0 2 :# 0 3 :# 0 4 : P A S
# 0 5 :# 0 6 :# 0 7 :# 0 8 : P A S
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
S
S
P A S
P A SSS
59
GSB-01/GSB-02 Series User Manual
Note
The Pass Sound setting must to set to ON for the
buzzer to sound.
FAIL Judgment
When any of the tests have been
judged as FAIL, the FAIL indicator
will be lit red, the channel
indicator for the failed channel(s)
will be lit red and the buzzer will
sound.
FAIL
Multiplex Scanner Box
GSB-01
H ILOH ILOH ILOH ILOH ILOH ILOH ILOH I
LO
CHANNEL2345678
PASS
FAIL
1
Red channel indicators, indicating a fail
F A I
P A SLS
P A S
P A SSS
E MA U OT N A_ *A U T O = 0 0 1 - * *
# 0 1 :# 0 2 :# 0 3 :# 0 4 : P A S
# 0 5 :# 0 6 :# 0 7 :# 0 8 : P A S
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
S
S
F A I
P A SLS
Note
The Fail Sound setting must to set to ON for the
buzzer to sound.
60
OPERATION
View Results
1. When the PASS or FAIL overview
table is shown on the screen, turn
the scroll wheel right to scroll
through each test step.
I R G B A CWD C W
F R EQ = 0 H z 6
0
100
k V
SAP
MT I E = 0 0 3 . 2 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 0 S0AM P =R
S
00 37
m A
E MM A UNN A_ 3S T E P : 0 2 - 0 0
PASS/FAIL result
step number
MANU number for
current step
2. Turn the scroll wheel left to return
back to the overview table.
Return to Ready
Status
1. The PASS/FAIL results will be
held on the screen until the STOP
button is pressed.
2. To put the tester back into READY
status, press the STOP button
(twice for a fail result).
STOP
3. The READY indicator will be lit
blue in the READY status.
READY
I R G B A CWD CW
E
F R E Q = 0 H z 6
0
100
k V
AER
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
Y
READY status
MA U OT N A_0A U T O = 0 0 1 - 1 0
D
m A
61
GSB-01/GSB-02 Series User Manual
Exit Testing
To exit testing, press the
MANU/AUTO key when the
tester is in the READY status. The
tester will revert to the VIEW
status.
MANU/AUTO
I R G B A C WD CW
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES = 0T
0 0 . 1 S0A MP =R
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
m A
62
OPERATION
Description
The interface settings choose the remote
interface configuration. USB, RS232 and GPIB
(optional) can be selected. With the scanner
box(es) installed, remote control is only
possible via USB. The RS-232 interfaces are
used to link the scanner boxes to the safety
tester.
Steps
1. Ensure the tester is in VIEW status. Save the
current test if necessary.
I R G B
E
A CWD CW
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =#0 .0
1 . 0 0 mAIHES= 0T
0 0 . 1 S0AM P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
VIEW status
2.Press the UTILITY key.
UTILITY
3. Press the INTER soft-key to bring
up the Interface Common Utility
menu.
N T EIR
I n t
L C DB U Z Z N T EIR C T R L
MOCMO U NT I L I T Y
B u d :a 1 1 5 2 0 0
R S 2 3 2f a c :e
e r
Common Utility Settings
The Common Utility settings are system-wide settings that apply to
both MANU tests and AUTO tests. The scanner boxes introduce
one new interface setting.
Scanner Box Interface Settings
63
GSB-01/GSB-02 Series User Manual
4. Use the scroll wheel to select
SCANNER BOX.
5. Press EDIT/SAVE to save the
settings and exit to VIEW status.
EDIT/SAVE
Note
The baud rate setting is fixed to 115200 baud.
Note
When Interface is set to SCANNER, the remote
interface is automatically set to USB.
Note
The ESC key can be pressed at any time to cancel
and exit back to VIEW status.
Clear the configuration and results from the
scanner box.
Note
Only applicable in the VIEW or READY status.
Syntax
GSB:CLR
GSB:SCAN
Query
Description
Scans for any connected scanner boxes and returns
the model type and order.
Note
Only applicable in the VIEW or READY status.
Query Syntax
GSB:SCAN?
REMOTE CONTROL
Scanner Commands
65
GSB-01/GSB-02 Series User Manual
Return parameter
<string>
Returns:
The connection status and order of
each scanner in the following format:
<1st>,<2nd>,<3rd>,<4th>
Where 01=GSB-01, 02=GSB-02,
XX=not connected.
Query Example
GSB:SCAN?
>01,02,XX,XX
Number #1 scanner box =GSB-01,#2=GSB-02, #3=not
connected,#4= not connected.
GSB<X>:CHANnel
Query
Description
Queries setting state of each scanner box for a
specified MANU number.
Note
Only applicable in the VIEW or READY status.
Query Syntax
GSB<X>:CHANnel?
Parameter
<X>
1 ~ 100 (MANU number)
Return parameter
<string>
Returns the status of each scanner
box in a table as follows:
Where the first line is the channel
index and lines 2~5 are connection
settings for the scanner boxes in
order from 1st to 4th.
H=Set to H/HV terminal.
L=Set to Return terminal.
X=Off
Indicates that channel 1 and 4 have passed for the
GSB-01 and that channel 2 and 6 have failed for the
GSB-02.
GSB<X>:HI
Set
Query
Description
Sets or queries the HI settings for each scanner box
for a selected MANU test number.
Syntax
Query Syntax
GSB<X>:HI <string>
GSB<X>:HI?
Parameter
<X>
1~100 (MANU number)
Parameter/Return
parameter
<string>
8 character Hex code of a binary number
(16 digits), prepended with “S”.
HI = binary 1, other = binary 0.
First 2 Hex/8 binary digits = 1st scanner
box, second 2 Hex/8 binary digits = 2nd
scanner box, third 2 Hex/8 binary digits
= 3rd scanner box, 4th 2 Hex/8 binary
digits = 4th scanner box.
String format:
Hex code
Binary
Ch1 ~ CH8
1st scanner
Ch1 ~ CH8
2nd scanner
Ch1 ~ CH8
3rd scanner
Ch1 ~ CH8
4th scanner
XXXXXXXX:XXXXXXXX:XXXXXXXX:XXXXXXXX
XX:XX:XX:XX
Example1
GSB1: HI S80000000
Sets Channel 1 of scanner 1 to HI. All other channels
are off. (8000000 Hex =
10000000000000000000000000000000 binary)
68
REMOTE CONTROL
Example2
GSB1: HI SFF000000
Sets Channel 1~8 of scanner 1 to HI. All other
channels are off. (FF00000 Hex =
11111111000000000000000000000000 binary)
Note
Only applicable in the VIEW or READY status.
If a channel is set to binary 0, it can still be set to LOW
using the GSB<X>:LOW command.
GSB<X>:LOW
Set
Query
Description
Sets or queries the LOW (return) settings for each
scanner box for a selected MANU test number.
Syntax
Query Syntax
GSB<X>:LOW <string>
GSB<X>:LOW?
Parameter
<X>
1~100 (MANU number)
Parameter/Return
parameter
<string>
8 character Hex code of a binary number
(16 digits), prepended with “S”.
LOW = binary 1, other = binary 0.
First 2 Hex/8 binary digits = 1st scanner
box, second 2 Hex/8 binary digits = 2nd
scanner box, third 2 Hex/8 binary digits
= 3rd scanner box, 4th 2 Hex/8 binary
digits = 4th scanner box.
String format:
Hex code
Binary
Ch1 ~ CH8
1st scanner
Ch1 ~ CH8
2nd scanner
Ch1 ~ CH8
3rd scanner
Ch1 ~ CH8
4th scanner
XXXXXXXX:XXXXXXXX:XXXXXXXX:XXXXXXXX
XX:XX:XX:XX
Example1
GSB1: LOW S80000000
Sets Channel 1 of scanner 1 to LOW. All other
channels are not set. (8000000 Hex =
10000000000000000000000000000000 binary)
69
GSB-01/GSB-02 Series User Manual
Example2
GSB1: LOW SFF000000
Sets Channel 1~8 of scanner 1 to LOW. All other
channels are not set. (FF00000 Hex =
11111111000000000000000000000000 binary)
Note
Only applicable in the VIEW or READY status.
If a channel is set to binary 0, it can still be set to HI
using the GSB<X>:HI command.
70
FAQ
•
I get a scanner error when I run a test.
•
The tester will stop testing midway through an automatic test.
•
When I press the START button the tester will not start testing?
FAQ
I get a scanner error when I run a test.
A scanner error indicates that the HV and return terminals haven’t
been properly connected between the safety tester and the
scanner(s).
The tester will stop testing midway through an automatic
test.
You may have PASS HOLD turned on or FAIL MODE set to HOLD.
Press the Start button to continue to the next test. See the user
manual for setting details.
When I press the START button the tester will not start
testing?
The tester must first be in the READY status before a test can be
started. Ensure the tester displays READY before pressing the
START button.
If “Double Action” is enabled, the START button must be pressed
0.5 seconds after the STOP button is pressed, otherwise the tester
will not start testing.
71
GSB-01/GSB-02 Series User Manual
If “Interlock” is enabled, the interlock key must be inserted into the
signal I/O port on the rear before a test can be started.
Lastly, ensure that the Start Ctrl setting is correctly configured in
the Common Utility menu. For example, to enable the START
button to start a test, ensure that the Start Ctrl setting is set to
FRONT PANEL.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com / marketing@goodwill.com.tw.
72
APPENDIX
Steps
1. Turn the instrument
off.
POWER
2. Take off the power cord and remove the fuse
socket using a minus driver.
Replace the fuse in the fuse holder.
Rating
100V~240V
T2A 250V
APPENDIX
Fuse Replacement
73
GSB-01/GSB-02 Series User Manual
Error Messages
Description
SCANNER ERR
Indicates that there is a problem with the
scanner connection.
Error Messages
Test Errors
The following error messages or messages may appear on the GPT
screen when configuring or running tests.
74
APPENDIX
Scanner
Model
GSB-01
GSB-02
High Voltage Rating
5kV AC/6kV DC
5kV AC/6kV DC
High Current Rating
40A AC
Number of HV channels
8 6 Number of GB channels
None
2
System Interface
RS-232
RS-232
Power Source
AC 100-240V ±10%, 50/60Hz
Other
Operation Environment
Indoor use, altitude up to 2000m
Ambient temperature 0°C to 40°C
Relative humidity 70%
Installation category II
Pollution degree 2
Storage Temperature &
Humidity
-10°C to 70°C, 85% RH
Accessories
GB wiring leads: GHT-109 x1 (only for GSB-02)
HV wiring leads: GHT-108 x1
Power cord x1
Quick start guide x1
User manual x1 (CD)
RS232C cable: GTL-235
HV test leads: GHT-116R x8(GSB-01), x6(GSB-02)
Return test leads: GHT-116B x1
GB H test leads: GTL-116R x2(GSB-02)
GB L test leads: GTL-116B x1(GSB-02)
Dimensions
GSB-01: 330(W) x 101(H) x 399(D)
GSB-02: 330(W) x 101(H) x 413(D)
Weight
Approx. 5.5kgs
Note
When using the scanner box outputs, IR tests
support a maximum of 2000MΩ.
Specifications
The specifications apply when the GSB-01/02 is powered on for at
least 30 minutes at 15˚C~35˚C.
Specifications
75
GSB-01/GSB-02 Series User Manual
Multiplex Scanner Box
POWER
H I
1CHANNEL
PASS
FAIL
LO2LO
3
H I
4LO5LO6
H I7H I
8
LOLOLO LO
H IH IH IH I
322.0
330.0
88.0
100.6
349.0
399.2
RETURN
GSB-01
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
Dimensions
GSB-01
76
GSB-02
SOURCE H SOURCE L
SENSE H SENSE L
RETURN
HI-POT
IR
GSB-02
Multiplex Scanner Box
GB Rx
POWER
H I
1CHANNEL
PASS
FAIL
LO2LO
3
H I
4LO5LO6
H I7H I
8
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
LOLOLO LO
H IH IH IH I
322.0
330.0
88.0
100.6
349.0
412.6
APPENDIX
77
GSB-01/GSB-02 Series User Manual
Type of Product: Multiplex Scanner Box
Model Number: GSB-01, GSB-02
◎EMC
EN 61326-1
EN 61326-2-1
Electrical equipment for measurement, control and
laboratory use –– EMC requirements (2013)
Current Harmonics
EN 61000-3-2: 2006 +A1:2009
+A2:2009
Radiated Immunity
EN 61000-4-3: 2006 +A2:2010
Voltage Fluctuations
EN 61000-3-3: 2008
Electrical Fast Transients
EN 61000-4-4: 2012
-------------------------
Surge Immunity
EN 61000-4-5: 2006
-------------------------
Conducted Susceptibility
EN 61000-4-6: 2009
-------------------------
Power Frequency Magnetic Field
EN 61000-4-8: 2010
-------------------------
Voltage Dip/ Interruption
EN 61000-4-11: 2004
Low Voltage Equipment Directive 2006/95/EC
Safety Requirements
EN 61010-1: 2010
EN 61010-2-030: 2010
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declare that the below mentioned product
are herewith confirmed to comply with the requirements set out in the
Council Directive on the Approximation of the Law of Member States
relating to Electromagnetic Compatibility (2004/108/EC) and Low Voltage
Directive (2006/95/EC).
For the evaluation regarding the Electromagnetic Compatibility and Low
Voltage Directive, the following standards were applied: