GW Instek APS-7000 Series, APS-7100, APS-7050, APS-7300, APS-7200 User Manual

Page 1
Programmable AC Power Source
APS-7000 Series
USER MANUAL
ISO-9001 CERTIFIED MANUFACTURER
Page 2
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Page 3
Table of Contents
Table of Contents
SAFETY INSTRUCTIONS ................................................... 4
GETTING STARTED .......................................................... 8
APS-7000 Series Overview ...................... 9
Appearance .......................................... 14
OPERATION .................................................................. 26
Set Up .................................................. 28
Basic Operation ................................... 60
Advanced Settings ............................... 82
Miscellaneous ...................................... 88
Trigger ................................................. 95
Preset Settings ................................... 100
Arbitrary Waveform Function ............. 104
Test Mode Function ........................... 112
REAR PANEL SIGNAL OUTPUTS .................................... 149
COMMUNICATION INTERFACE .................................... 153
Interface Configuration ...................... 154
FAQ .............................................................................. 175
APPENDIX .................................................................... 177
Firmware Update ............................... 177
APS-7000 Default Settings ................. 179
APS-7000 Specifications ..................... 181
APS-7000 Dimensions ........................ 187
Declaration of Conform ity .................. 191
INDEX .......................................................................... 192
3
Page 4
APS-7000 Series User Manual
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the APS-7000 or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
4
Page 5
SAFETY INSTRUCTIONS
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
CAUTION
Do not place any heavy object on the APS-7000. Avoid severe impact or rough handling that
leads to damaging the APS-7000.
Do not discharge static electricity to the APS-
7000.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not disassemble the APS-7000 unless you are
qualified.
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The APS-7000 doesn’t fall under category II, III or IV.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
0 is for measurements performed on circuits not directly
connected to Mains.
Power Supply
WARNING
AC Input voltage range:
115/230 Vac ± 15% (APS-7050, APS-7100)
230 Vac ± 15% (APS-7200, APS-7300)
Frequency: 50/60Hz To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.
Safety Guidelines
5
Page 6
APS-7000 Series User Manual
Cleaning the APS­7000
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 20%~ 80%, no condensation Altitude: < 2000m Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The APS-7000 falls under degree
2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -10°C to 70°C
Relative Humidity: 80%, no condensation
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
6
Page 7
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the instrument in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
7
Page 8
APS-7000 Series User Manual
APS-7000 Series Overview ......................................... 9
Series lineup ........................................................................................ 9
Operating Area ................................................................................... 9
Main Features ................................................................................... 11
Accessories ........................................................................................ 12
Appearance .............................................................. 14
Front Panel ........................................................................................ 14
Rear Panel .......................................................................................... 19
Status Bar Icons ................................................................................ 24
GETTING STARTED
This chapter describes the power source in a nutshell, including its main features and front / rear panel introduction.
8
Page 9
GETTING STARTED
Model name
Max. Output Current
Power Rating
Output Voltage
APS-7050
4.2A/2.1A
500VA
0~310.0 Vrms
APS-7100
8.4A/4.2A
1000VA
0~310.0 Vrms
APS-7200
16.8A/8.4A
2000VA
0~310.0 Vrms
APS-7300
25.2A/16.8A
3000VA
0~310.0 Vrms
Voltage (V)
Current (A)
0
1
2
0
3
4
5
100
200
300
400 500
600
120 155 240
310 480
600
4.2
3.23
2.1
1.62
1.05
0.83
APS-7050 Output Operating Area
155V Range 310V Range 600V Range
(APS-003 Option)
APS-7000 Series Overview
Series lineup
The APS-7000 series consists of 4 models, the APS-7050, APS-7100, APS-7200 and APS-7300, differing only in capacity. Note that throughout the user manual, the term “APS-7000” refers to any of the models, unless stated otherwise.
Operating Area
9
Page 10
APS-7000 Series User Manual
Voltage (V)
Current (A)
0
2
4
0
6
8
10
100
200
300
400 500
600
120 155 240
310 480
600
8.4
6.45
4.2
3.23
2.1
1.67
155V Range 310V Range 600V Range
APS-7100 Output Operating Area
(APS-003 Option )
Voltage (V)
Current (A)
0
5
10
0
15
20
100
200
300
400 500
600
120 155 240
310 480
600
16.8
12.9
8.4
6.45
4.2
3.33
155V Range 310V Range 600V Range
APS-7200 Output Operating Area
(APS-003 Option )
10
Page 11
Voltage (V)
Current (A)
0
5
10
0
15
20
25
100
200
300
400 500
600
120 155 240
310 480
600
25.2
19.4
12.6
9.7
6.3
5.0
155V Range 310V Range 600V Range
APS-7300 Output Operating Area
(APS-003 Option )
Main Features
Performance
Low output ripple and noise Excellent and feature-rich measurement
capacity
Standard maximum output voltage is 310Vrms Maximum output voltage and frequency of
600Vrms(APS-003 Option)/1000Hz(APS-004 Option)
GETTING STARTED
11
Page 12
APS-7000 Series User Manual
Features
OCP, OPP and OTP protection Variable voltage, frequency and current limiter Sequence and simulation function Large 4.3 inch TFT panel Globally adjustable power inlet not restricted by
the power supply environment
USB interface is equipped as standard with the
ability to save and recall setup files.
Only 88mm (2U) case height (APS-7050 and
APS-7100 models only).
Interface
Standard:
Ethernet port USB host USB CDC (APS-7200 and APS-7300 models
only)
Optional:
GPIB RS-232 / USB CDC (APS-7050 and APS-7100
models only)
RS-232 (APS-7200 and APS-7300 models only)
Standard Accessories
Part number
Description
CD ROM
User manual, programming manual
82GW1SAFE0M*1
Safety guide
Region dependent
Type I Power cord (APS-7050)
Region dependent
Type II Power cord (APS-7100)
Accessories
12
Page 13
GETTING STARTED
Region dependent
Type III Power cord (APS-7200, APS-7300)
62PS-7K0SC701 x1 5302-01613001 x1
Mains terminal cover set (APS-7050)
62PS-7K0SC401 x1 5302-01613001 x2
Mains terminal cover set (APS-7100)
GTL-123
Test leads: 1x red, 1x black
Optional Capacity
Part number
Description
APS-003
Output Voltage Capacity: 0 ~ 600Vrms
APS-004
Output Frequency Capacity: 45 ~ 1000Hz
Optional Accessories
Part number
Description
GRA-423
APS-7050 and APS-7100 rack mount kit
GRA-429
APS-7200 rack mount kit
GRA-430
APS-7300 rack mount kit
APS-001
GPIB interface card
APS-002
RS-232 / USB CDC interface card (APS-7050 and APS-7100 only)
APS-007
RS-232 interface card (APS­7200 and APS-7300 only)
GPW-004
Power Cord 8mm2/3C, 3m
Max Length, 105°C,
RNYBS8-6*3P, RNYB8­8*3P
Download
Name
Description
gw_aps.inf
USB driver
13
Page 14
APS-7000 Series User Manual
APS-7050
AC Power Source
Display
Programmable
F 1
F 2
F 3
F 4
Output
: Long Push
Menu Test Preset
V
Local
Off Phase
On Phase
0
1
4
7
2
5
8
3
6
9
V-Limit F-Limit IPK-Limit
F I rms
Cancel
Shift
Range
Enter
Lock
ALM CLR
IPK CLR
Unlock
Surge / Dip
Ramp
TriggerARB
USB A port
Power switch
Front voltage output socket
Menu, Test, Preset keys
LCD display
Function keys
Display mode key
Arrow keys, Scroll wheel
Voltage, Frequency, Current setting and limit
Keypad
Range, Shift, Cancel, Enter, Lock keys
Output key
Caster
Adjuster
foot
Caster
Appearance
Front Panel
APS-7050, APS-7100
APS-7200 APS-7300
14
Page 15
GETTING STARTED
Item
Description
Power Switch
Turns on the mains power.
Front Voltage Output Socket
Neutral
GND
Line
Output voltage terminal using a regional universal plug. There is a Euro and a Universal regional plug.
CAUTION
Maximum allowable output voltage and current are 250Vrms and 15Arms.
For voltages exceeding 250Vrms or current over 15Arms, please use the rear output terminal.
Universal and EURO region plugs socket must be factory installed.
USB A Port
The USB port is used for data transfers and upgrading software.
LCD Screen
Displays the measured values or menu system.
Display Mode Select Key
Display
Selects between Standard mode and Simple mode.
15
Page 16
APS-7000 Series User Manual
Function Keys
F 1
Assigned to the functions displayed on the right-hand side of the screen.
Menu Key
Menu
Enters the Main menu or goes back to one of the display modes.
Test Key
Test
Puts the instrument into the Sequence, Simulation and Program Control mode.
Preset Key
Preset
Puts the instrument into Preset mode.
Arrow Keys
The arrow keys are used to select the digit power of a value that is being edited.
V
V
V-Limit
Used for setting the output voltage.
V-Limit
(Shift + V)
Used for setting the output voltage limit value.
F
F
F-Limit
Used for setting the output frequency.
F-Limit
(Shift + F)
Used for setting the output frequency limit value.
I rms
I rms
IPK-Limit
Used for setting the maximum output current.
IPK-Limit
(Shift + I rms)
Used to set the peak output current limit value.
Range Key
Range
Switches between the 155V, 310V and 600V ranges (the 600V range is an option).
Scroll Wheel
Used to navigate menu items or for incrementing/decrementing values one step at a time.
16
Page 17
GETTING STARTED
Lock Key
Lock
Unlock
Locks the number pad to prevent accidentally changing panel settings.
Unlock Key
(Long press)
Disables the key lock.
Enter Key
Enter
Confirms selections / settings
Cancel Key
Cancel
Clears entries that are made in the number entry dialog when a value is edited using the arrow keys or the scroll wheel.
The Cancel key can also be used to cancel function setting menus or dialogs.
Shift Key
Shift
Turns on the shift state, which enables shortcut operations.
Output Key
Output
Turns the output on or off.
Number Pad
0
142
5
8
3
6
9
Lock
Unlock
7
On Phase Surge / Dip IPK CLR
Off Phase Ramp ALM CLR
ARB Trigger
Local
Used to enter values.
Local Mode
(Shift + 0)
Switches operation back to local mode from remote mode.
ARB Mode
(Shift + 1)
Sets the ARB function.
Trigger Mode
(Shift + 2)
Sets the JI port trigger behavior on the rear panel.
Off Phase
(Shift + 4)
Sets the off phase for the output voltage.
RAMP
(Shift + 5)
Quick settings for Ramp control.
ALM CLR
(Shift + 6)
Clears alarms.
On Phase
(Shift + 7)
Sets the on phase for the output voltage.
17
Page 18
APS-7000 Series User Manual
Surge/Dip
(Shift + 8)
Quick settings for Surge/Dip control.
IPK CLR
(Shift + 9)
Clears peak current hold.
18
Page 19
Rear Panel
SYNC
Signal Output LAN
OUTPUT INPUT
L N N L
600Vac MAX. 115 / 230V 15%
50 / 60Hz
3.6kVA MAX.
Out On / Off
COM
J1
Trigger Out
Trigger In
115 / 230V 15%
50 / 60Hz
1.8kVA MAX.
SYNC
SER. NO. LB
L N
600Vac MAX.
Signal Output
Trigger Out
Trigger In Out On / Off
COM
LAN
J1
OUTPUT
Fan
GPIB
Line voltage input
GPIB
RS232
RS-232/
USB Device
Sync
Signal Output Rear panel voltage
output terminal
Remote control Ethernet port
Line voltage input
APS-7050
APS-7100
Opt. int. port
GETTING STARTED
19
Page 20
APS-7000 Series User Manual
Remote control
Ethernet port
Line voltage input
APS-7200
GPIB
GPIB
RS232
RS-232
Fan
Sync
Signal Output
Rear panel voltage output and sensing terminals
Optional interface port
Circuit breaker (current)
USB port
20
Page 21
Remote control
Ethernet port
Line voltage input
APS-7300
GPIB
GPIB
RS232
RS-232
Fan
Sync
Signal Output
Rear panel voltage
output and sensing terminals
Optional interface
port
Circuit
breaker (current)
USB port
GETTING STARTED
21
Page 22
APS-7000 Series User Manual
Line Voltage Input
APS-7050
115 / 230V 15%
50 / 60Hz
1.8kVA MAX.
Voltage Input: 115/230±15% VAC; Line frequency: 50Hz/60 Hz (Automatically switchable)
APS-7100
OUTPUT INPUT
L N N L
Voltage Input: 115/230±15% VAC ; Line frequency: 50Hz/60 Hz (Automatically switchable)
APS-7200 & 7300
N L
Voltage Input: 230±15% VAC ; Line frequency: 50Hz/60 Hz
Rear Voltage Output Socket
Output voltage terminal.
APS-7050
L N
OUTPUT
APS-7100
OUTPUT INPUT
L N N L
APS-7200 & 7300
L N
Sync Output Socket
SYNC
BNC socket. This socket will output a signal of approximately 10V when the output is on.
Signal Output
Signal output
APS-7050 APS-7100
APS-7200 APS-7300
Connector for monitoring PASS, FAIL and PROCESSING output signals when using the Program mode.
Remote Control
Trigger Out
Trigger In
Out On / Off
COM
APS-7200 and APS-7300
Trigger Out
Trigger In Out On / Off
COM
APS-7050 and APS-7100
Connector for controlling the TRIGGER IN, TRIGGER OUT and OUT ON/OFF states.
22
Page 23
GETTING STARTED
Ethernet Port
LAN
The Ethernet port is used for remote control and digital monitoring from a PC.
Optional Interface Slot
Optional GPIB communication, RS-232/USB B communication and RS-232 communication.
Circuit breaker (APS-7200, APS-7300 only)
I / O
Main power circuit (current) breaker
Rating : 40A (APS-7200) 63A (APS-7300) Note: Check the status of the power
breaker before power-on the APS-7200 or APS-7300.
FAN
Temperature controlled fan.
Remote sense (APS-7200, APS-7300 only)
Compensation of the load wire drop.
23
Page 24
APS-7000 Series User Manual
Status bar
ARB
/
Indicates if the output is ON or OFF.
Indicates the output power as a percentage of full scale.
Indicates that the Surge/Dip function is active.
Indicates that the Ramp function is active.
The alarm icon will appear on the status bar when one of the protection functions are tripped. Applies to Over Power, Over Irms, Over Ipeak and Over Temperature protection.
Indicates that the APS-7000 is in remote mode.
Indicates that a USB drive is detected in the front panel host port.
Indicates that the LAN interface is activated.
Status Bar Icons
24
Page 25
GETTING STARTED
Indicates that the panel lock is active.
ARB
Indicates that the ARB function is active.
25
Page 26
APS-7000 Series User Manual
Set Up ..................................................................... 28
Line Voltage Removal and Installation for APS-7100 .............. 28
Filter Installation .............................................................................. 31
Power Up ........................................................................................... 33
Output Terminals ............................................................................. 34
Wire gauge considerations .............................................................. 39
Installing the Optional Hardware Modules ................................. 41
Installing Optional Software Modules .......................................... 43
Using the Rack Mount Kit ............................................................. 48
How to Use the Instrument ........................................................... 50
Reset to Default Settings ................................................................ 54
View System Version and Serial Number .................................... 55
LCD Configuration.......................................................................... 56
USB Driver Installation .................................................................. 57
Basic Operation ....................................................... 60
Setting the Voltage Range ............................................................... 60
Setting the Voltage Limit ................................................................ 61
Setting the Output Voltage............................................................. 62
Setting the Frequency Limit ........................................................... 64
Setting the Output Frequency........................................................ 65
Setting the Peak Current Limit ...................................................... 66
Setting the Current RMS Level ...................................................... 70
Setting the On/Off Phase .............................................................. 73
Alarm Clear ....................................................................................... 74
Display Modes .................................................................................. 75
Panel Lock ......................................................................................... 77
Turning the Output On .................................................................. 78
Using the remote sense (APS-7200 and APS-7300 only) ......... 79
Local Sense ........................................................................... 79
Remote Sense ...................................................................... 80
Advanced Settings ................................................... 82
Surge/Dip Control .......................................................................... 82
Ramp Control ................................................................................... 85
Miscellaneous.......................................................... 88
OPERATION
26
Page 27
OPERATION
T Ipeak, hold..................................................................................... 88
Power ON Output ........................................................................... 90
Buzzer ................................................................................................ 91
SCPI Emulation ............................................................................... 92
Remote Sense (APS-7200 and APS-7300 only) .......................... 93
Trigger ..................................................................... 95
Trigger Control Settings ................................................................. 95
Preset Settings .......................................................100
Save Preset Settings to Local Memory ....................................... 100
Load Preset Settings to Local Memory ...................................... 101
Manage Preset Settings ................................................................. 102
Arbitrary Waveform Function ..................................104
ARB Mode Overview .................................................................... 105
Selecting an ARB Waveform ....................................................... 109
Test Mode Function ................................................112
Simulate Mode Overview ............................................................. 114
Simulate Settings ............................................................................ 118
Save a Simulation to Local Memory ........................................... 121
Recall a Simulation from Local Memory ................................... 121
Manage Simulation Settings ......................................................... 122
Running a Simulation .................................................................... 124
Sequence Mode Overview ............................................................ 126
Sequence Settings ........................................................................... 130
Save a Sequence to Local Memory ............................................. 134
Recall a Sequence from Local Memory...................................... 134
Manage Sequence Settings ............................................................ 135
Running a Sequence ...................................................................... 137
Program Mode Overview ............................................................. 139
Save a Program to Local Memory............................................... 146
Recall a Program from Local Memory ....................................... 146
Manage Program Settings ............................................................. 147
27
Page 28
APS-7000 Series User Manual
Background
The APS-7100 is equipped with an input power terminal that can accept 115V/230V ± 15%. To connect or replace the power cord (GW Instek part number 4300-31000101), use the procedure below:
Warning
The following procedure should only be attempted by competent persons.
Ensure the AC power cord is not connected to power.
Removal
1. Turn off the power
switch.
1
APS-7100
2. Unscrew the power cord protective sheath on
the rear output socket.
3. Remove the 2 screws holding the power cord
cover and remove.
Set Up
Line Voltage Removal and Installation for APS-7100
28
Page 29
OPERATION
APS-7100
3
2
4. Remove the AC
power cord wires.
4
CAUTION
The power line inputs for the APS-7100 are on the outer cluster of terminals.
The terminals to the inner of the panel are the rear panel outputs.
Installation
1. Connect the AC power cord wires to the AC
input terminals.
White/Blue Neutral (N) Green/Green-yellowGND ( ) Black/Brown Line (L)
29
Page 30
APS-7000 Series User Manual
APS-7100
OUTPUT INPUT
L N N L
115 / 230V 15%
Line
Neutral
Ground
2. Re-install the power cord cover.
3. Screw the power cord sheath back onto the
cover.
APS-7100
2
3
30
Page 31
Background
The APS-7200/7300 has a filter (GW Instek part number, 57RG-30B01801) that must first be inserted under the control panel before operation.
Steps
1. Pull outward as indicated
in the arrow to detach the snap.
See below for details
2. Remove the cover
3. Remove the screws
4. Move the plastic frame in
the direction indicated by the arrow
OPERATION
Filter Installation
31
Page 32
APS-7000 Series User Manual
5. Remove the plastic frame.
6. Replace the filter with a
new one.
7. The unit is now ready to power up.
Note
Please clean regularly to avoid damaging the internal components of the machine
Warning
The following procedure should only be attempted by competent persons.
Ensure the AC power cord is not connected to power.
32
Page 33
Power Up
Steps
1. Socket type (APS-7050):
Connect the power cord to the rear panel socket.
Input Power Terminal (APS-7100, 7200 & 7300): Connect the power cord to the input power terminals.
APS-7200 & 7300, see page 22,
APS-7100, see page 28.
2. Press the POWER key. The splash screen will
appear momentarily before the continuous mode screen appears with the settings loaded.
CAUTION
The power supply takes around 10 seconds to fully turn on and shutdown.
Do not turn the power on and off quickly.
OPERATION
33
Page 34
APS-7000 Series User Manual
Background
The output terminals can be output from either the front panel or from the rear panel. The outputs are limited to 4.2A/2.1A (APS-7050),
8.4A/4.2A (APS-7100), 16.8A/8.4A (APS-7200) or 25.2A/12.6A (APS-7300).
Supported plugs
Multi-region terminal Socket
Supported Standards
IEC, North America, Japan.
EURO CEE type universal plug
WARNING
Dangerous voltages. Ensure that the power to the instrument is disabled before handling the power supply output terminals. Failing to do so may lead to electric shock.
CAUTION
For the front panel output, the maximum output voltage is 250VAC and current is 10A.
Front Panel Output Connection
1. The front panel has a multi-region power
socket depending on the socket type.
2. Insert the plug from the DUT into the socket.
(APS-7050 or APS­7100 shown)
EURO CEE socket
IEC North America, Japan
Output Terminals
34
Page 35
OPERATION
WARNING
Dangerous voltages. Ensure the output is off before unplugging the plug from the front panel socket.
3. Turn the power on. The AC power supply is
now ready to power the DUT.
Rear Panel Output Connection
The rear panel output is used to supply higher power DUTs. The rear panel output connection is similar to the universal rear panel line input connection on the APS-7100, APS-7200 or APS-
7300.
1. Disconnect the unit from the mains power
socket and turn the power switch off.
2. Unscrew the power cord protective sheath
(APS-7050 and APS-7100).
3. Remove the 2 screws holding the power cord
cover and remove (APS-7050 and APS-7100).
APS-7050
3
2
35
Page 36
APS-7000 Series User Manual
APS-7100
3
2
CAUTION
For the APS-7100, there is a single bank for the input and output terminals. Ensure the correct terminals are connected. The APS-7050, APS-7200 and APS-7300 have a dedicated bank of output terminals on the rear panel.
Installation
4. Connect the output AC power cord wires to
the AC output terminals.
Black Neutral (N) Green GND ( ) Red Line (L)
APS-7100
OUTPUT INPUT
L N N L
115 / 230V 15%
Line
Neutral
Ground
Ground
Neutral
Line
APS-7100 shown. The input terminals are already connected and shows which output terminals are to be connected.
36
Page 37
OPERATION
APS-7050, APS­7200 and APS­7300
Line
Neutral
Ground
L N
OUTPUT
APS-7050 shown.
5. Re-install the power cord cover (APS-7050
and APS-7100).
6. Screw the power cord sheath back onto the
cover (APS-7050 and APS-7100).
APS-7050
5
6
37
Page 38
APS-7000 Series User Manual
APS-7100
5
6
7. Turn the power on. The AC power supply is
now ready to power the DUT.
Note
Grounded Neutral Output: APS-7000 allows for a grounded return on the neutral output. It is suit for the medical industry that required between ground with neutral is 0V essentially. And possible to mitigate ground loops that is ideal for reduce ground noise and isolate sensitive equipment from the effects of ground loops.
WARNING
Because the neutral has been referenced to the chassis ground, be careful electric shock by yourself.
38
Page 39
OPERATION
Background
Before connecting the output terminals to a load, the wire gauge of the cables should be considered.
It is essential that the current capacity of the load cables is adequate. The rating of the cables must equal or exceed the maximum current rated output of the instrument.
Recommended wire gauge
Wire Gauge
Nominal Cross Section
Maximum Current
20
0.5 9
18
0.75
11 18
1
13 16
1.5
18 14
2.5
24 12
4
34 10
6
45
8 10
64
6 16
88 4
25
120 2
32
145 1
50
190 00
70
240 000
95
290 0000
120
340
The maximum temperature rise can only be 60 degrees above the ambient temperature. The ambient temperature must be less than 30 degrees.
To minimize noise pickup or radiation, the load wires and remote sense wires should be twisted-pairs of the shortest possible length. Shielding of the sense leads may be necessary in high noise environments. Where shielding is
Wire gauge considerations
39
Page 40
APS-7000 Series User Manual
used, connect the shield to the chassis via the rear panel ground screw. Even if noise is not a concern, the load and remote sense wires should be twisted-pairs to reduce coupling, which might impact the stability of the power supply. The sense leads should be separated from the power leads.
40
Page 41
OPERATION
Background
There are a number of optional modules that can be installed for remote control.
Optional Modules
APS-001
GPIB Interface card
APS-002
RS-232/USB CDC interface card
APS-007
RS-232 interface card
WARNING
Dangerous voltages. Ensure that the power to the instrument is disabled before handling the power supply output terminals. Failing to do so may lead to electric shock.
CAUTION
Ensure the power is off before installing any of the optional modules.
Installation
1. Turn off the power
switch.
1
2. Unscrew the two screws holding the options
panel plate.
2
APS-7100 shown
Installing the Optional Hardware Modules
41
Page 42
APS-7000 Series User Manual
3. Slide the module PCB onto the rails on the
inside of the module slot.
4. Secure the module with the screws that were
removed from step 2.
4
3
APS-7100 shown
5. The module will be recognized upon startup.
42
Page 43
OPERATION
Background
The APS-003 and APS-004 are optional software modules that upgrade the voltage limit and frequency limit to 600Vrms and 1000Hz, respectively. Like the firmware, the software modules can be upgraded using the USB A port on the front panel. See your local distributor or GW Instek to purchase these options.
WARNING
The APS-003 and APS-004 software module license keys have been updated and there are now two license key file formats in use:
- The old license keys were tied to the serial
number of the APS-7000 unit. The license key files used XXXXXX.lis formatting, where XXXXXX was the serial number of the unit the license key is intended for.
- The new license keys are no longer tied to
serial numbers. These new license key files use XXXXXX.lic formatting (APS003.lic & APS004.lic). The new license keys can be exported and transferred to different units.
The old and new license keys are to be used with the following equipments and restrictions:
- Old license key (*.lis): can only be used with
the APS-7050 and the APS-7100 with a firmware version below V1.08.
- New license key (*.lic): APS-7050 and APS-
7100 with a V1.08 firmware version or above; any APS-7200; any APS-7300.
Note
To upgrade APS-003 or APS-004, it is recommended to
Installing Optional Software Modules
43
Page 44
APS-7000 Series User Manual
upgrade the software version to the latest one. For software, please contact GW or local distributors.
OLD LICENCE KEY INSTALLATION (XXXXXX.LIS format)
Note
This installation guide only applies to APS-7050 and APS-7100 equipments with a firmware version below V1.08.
These options require your serial number. The older license keys are not longer available for purchase.
The serial number on the APS unit must match the XXXXXXX.lis filename, or the upgrade will fail.
See page 55 for instructions on how to view your serial number
Steps
1. Insert the USB flash drive into the USB port on
front panel of the APS-7000.
The USB drive should include the
XXXXXX.lis file in a directory named
“gw”(USB\gw:).
2. Press the Menu key. The Menu
setting will appear on the display.
Menu
3. Use the scroll wheel to go to item 11, Special
Function and press Enter.
4. Key in the password when prompted and then
press Enter.
The password is “5004”.
5. Go to Item 5, Add New Module and press Enter.
44
Page 45
OPERATION
6. If the upgrade is successful, “Vlimit Enabled”
or “Flimit Enabled” will appear on the screen.
Vlimit option successfully upgraded
Invalid License
7. If the upgrade is not successful, “Invalid
License” will be displayed. Check to make
sure the serial number and the XXXXXX.lis filename match.
NEW LICENCE KEY INSTALLATION (XXXXXX.LIC format)
Note
This installation guide only applies to firmware versions 1.08 or above on APS-7050, APS-7100, APS-7200 and APS-7300 equipments. Do not attempt this installation procedure with older firmware.
The license keys (APS003.lic & APS004.lic), can be exported and transferred to another APS-7000 power supply of the same model at anytime. A license key can only be used with one machine at a time.
Steps
1. Insert the USB flash drive into the USB port on
front panel of the APS-7000.
The USB drive should include the APS003.lic
or APS004.lic file in a directory named
“gw”(USB\gw:).
2. Press the Menu key. The Menu
setting will appear on the display.
Menu
45
Page 46
APS-7000 Series User Manual
3. Use the scroll wheel to go to item 11, Special
Function and press Enter.
4. Key in the password when prompted and then
press Enter.
The password is “5004”.
5. Go to Item 5, Add New Module and press Enter.
6. If the upgrade is successful, “Vlimit Enabled”
or “Flimit Enabled” will appear on the screen.
7. Reset the power to the unit when prompted.
Invalid License
8. If the upgrade is not successful, “Invalid
License” will be displayed.
NEW LICENCE KEY EXPORT (XXXXXX.LIC format)
Note
This guide will EXPORT the license key back to the USB flash drive* so that the license key can be installed on a different unit.
*The same USB flash drive that originally contained the *.lic license key must be used when exporting the license key.
Steps
1. Insert the USB flash drive into the USB port on
front panel of the APS-7000.
46
Page 47
OPERATION
2. Press the Menu key. The Menu
setting will appear on the display.
Menu
3. Use the scroll wheel to go to item 11, Special
Function and press Enter.
4. Key in the password when prompted and then
press Enter.
The password is “5004”.
5. Press F1 Export APS-003 or F2 Export APS-004.
6. Reset the power to the unit when prompted.
7. The APS003.lic or APS004.lic will now be in
the GW directory of the USB flash drive.
Warning
You can only export one license key at a time.
You cannot have two license keys on the same USB drive.
You cannot copy a license key to another USB drive. Trying to do so will cause the file to be lost or to be mismatched.
47
Page 48
APS-7000 Series User Manual
Background
The APS-7050, APS-7100, APS-7200 and APS7300 have optional Rack Mount kits:
Model
Rack Mount kit part number
APS-7050 & 7100
GRA-423
APS-7200
GRA-429
APS-7300
GRA-430
The APS-7050 and APS-7100 are designed to fit into a 2U rack height. The APS-7200 is designed to fit into a 7U rack height. The APS­7300 is designed to fit into a 9U rack height. Please see your distributor for further rack mount details.
Rack mount diagram (APS­7050 or 7100)
Using the Rack Mount Kit
48
Page 49
OPERATION
Rack mount diagram (APS-
7200)
Rack mount diagram (APS-
7300)
CAUTION
Ensure adequate ventilation is provided when using the rack mount. Ensure at a gap of at least 50mm is given for the side air intakes. Failure to do so may cause the instrument to overheat.
49
Page 50
APS-7000 Series User Manual
Background
The APS-7000 AC power supplies generally use the scroll wheel, arrow keys and Enter keys to edit numerical values or to select menu options.
Menu navigation is performed using the menu keys and function keys on the front panel.
The following section will explain some of these concepts in detail.
Selecting Menu Items
1. Turn the scroll wheel to select
parameters in menus and lists. The selected parameter will be highlighted in orange. The scroll wheel is also used to increment/decrement setting values.
2. Press the Enter key to edit the
parameter or to enter the selected menu.
Enter
How to Use the Instrument
50
Page 51
OPERATION
Example
Selected parameter
The following is an example of the menu list that appears when the Menu key is pressed.
Using the Keypad to edit parameter values
When editing a value the keypad can be used to directly enter the desired value.
1. Type the value of the
parameter using the keypad.
0
1
4
2
5
8
3
6
9
Lock
Unlock
7
On Phase Surge / Dip IPK CLR
Off Phase Ramp ALM CLR
ARB Trigger
Local
2. Press the Enter key to confirm the
edit.
Enter
Example
Parameter
51
Page 52
APS-7000 Series User Manual
Using the Arrow Keys and Scroll wheel to edit parameter values
Use the arrow keys to select a digit power and then use the scroll wheel to edit the value by that power.
1. Use the arrow keys to move the
cursor to the digit of the desired power.
2. Turn the scroll wheel to edit the
value by the resolution of the selected digit.
Cursor
3. Repeat the steps above for all the relevant
digits.
4. Press the Enter key to confirm the
edit.
Enter
Note
By default the cursor starts at the lowest power digit.
52
Page 53
OPERATION
Using the onscreen keyboard
The onscreen keyboard is only used in the Program Mode. The screenshot below is an example of the onscreen keyboard.
On screen keyboard
Entered characters
Using the Function Keys
The function keys are quick settings keys, the function of which depends on the current menu or operation.
1. Press the function key that corresponds to the
setting directly to its left.
2. The setting or parameter is immediately
executed.
Function keys
APS-7050
AC Power Source
Display
Programmable
Output
: Long Push
Menu Test Preset
V
Local
Off Phase
On Phase
0
1
4
7
2
5
8
3
6
9
V-Limit F-Limit IPK-Limit
F I rms
Cancel
Shift
Range
Enter
Lock
ALM CLR
IPK CLR
Unlock
Surge / Dip
Ramp
TriggerARB
F 1
F2
F3
F4
Corresponding quick
settings
3. Repeat the steps above for all the relevant
digits.
53
Page 54
APS-7000 Series User Manual
Background
The default settings can be restored from the Menu key settings. See page 179 for the default factory settings.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2. Use the scroll wheel to go to item 10, Default
Setting.
3. Press Enter x2 to restore the default settings.
Default settings
Reset to Default Settings
54
Page 55
OPERATION
Background
The Menu>System Information setting displays the serial number and version number.
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. The system information should now be listed
on the display.
If not, use the scroll wheel to go to item 1, System Information.
System
Information
View System Version and Serial Number
55
Page 56
APS-7000 Series User Manual
Background
The LCD Configuration setting sets the brightness, contrast and saturation level of the LCD display.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2. Use the scroll wheel to go to item 9, LCD
Configuration and press Enter.
3. Set the brightness, contrast and saturation.
Contrast(%)
1 ~ 100% (Default=50%)
Brightness(%)
1 ~ 100% (Default=50%)
Saturation(%)
1 ~ 100% (Default=50%)
Exit
4. Press Exit[F4] to exit from the
Ramp Control settings.
Default Settings
5. Press Default[F3] to set all the LCD settings to
50%.
LCD settings
Default
settings
LCD Configuration
56
Page 57
OPERATION
Background
If the USB Type B interface is to be used for remote control, the USB driver needs to be installed.
Note
The USB driver, GW_APS.inf, is located on the CD Rom that accompanied this user manual. Alternatively the driver can be downloaded from the GW Instek website.
For information on the USB interface, see page
154.
Steps
1. Connect the rear panel USB -B port on the
APS-7000 to the PC using a USB Type A to B cable.
2. Go the Windows Device Manager.
For Windows 7:
Start > Control Panel > Hardware and Sound > Device Manager
USB Driver Installation
57
Page 58
APS-7000 Series User Manual
3. The APS-7000 will be located under Other
Devices in the hardware tree. Right-click the APS-7XXX and choose Update Driver Software.
4. From the hardware wizard choose Browse my
computer driver software.
58
Page 59
OPERATION
5. Set the file path to the location of the USB
driver, click Next and finish the driver installation.
6. In APS-7000 will now be located in the Ports
node of the hardware tree in the Windows Device Manager if the driver installation was successful.
59
Page 60
APS-7000 Series User Manual
Background
The Range setting determines the general outlet voltage range. The ranges available correspond to common mains output voltage standards.
Steps
1. Press Range to access the Range
menu.
Range
Basic Operation
This section describes the basic operations required to operate the power supply.
Setting the Voltage Range → from page 60 Setting the Voltage Limit → from page 61 Setting the Output Voltage → from page 62 Setting the Frequency Limit page 64 Setting the Output Frequency → page 65 Setting the Peak Current Limit → from page 66 Setting the Current RMS Level → from page 68 Setting the On/Off Phase → from page 73 Clearing the Alarm → from page 74 Setting the Display mode → from page 75 Panel lock → from page 77 Turning the Output on/off → from page 78 Using the remote sense (APS-7200, APS-7300) → from page 79
Before operating the power supply, please see the Getting Started chapter, page 8.
Setting the Voltage Range
60
Page 61
OPERATION
2. Set the voltage range with the scroll wheel or
with the F1 ~ F4 soft-keys.
Range
AUTO, 600V(option), 310V, 155V
Soft-keys
AUTO, 600V(option), 310V, 155V
3. Press Enter to confirm the Range setting.
Range setting
Soft-key
settings
F1 F2 F3 F4
Note
If the range is changed from 155V to 600V, the Irms and IPK values will automatically be changed to a lower value. If the range is changed from 600V to 155V, the Irms and IPK values remain the same.
If the voltage range is changed when the output is on, the output will be automatically turned off.
Background
Setting the voltage limit allows the output voltage to be set to any level within the voltage limit (V Limit) range.
Steps
1. Press Shift + V to access the Volt
Limit menu.
Shift
+
V
V-Limit
Setting the Voltage Limit
61
Page 62
APS-7000 Series User Manual
2. Set the voltage limit with the scroll wheel or
with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the limit to the maximum and minimum, respectively.
Range
10% of full range ~ full range
Soft-keys
MAX, MIN
3. Press Enter to confirm the voltage limit setting.
Volt Limit
Min/Max
settings F3 F4
Note
Each voltage range (155V, 310V, 600V) has an independent voltage limit. Before change volt limit setting, if ACV setting value is bigger than desire volt limit value, so that the volt limit value can't be change.
Background
Before setting the power supply voltage level, set the voltage range and voltage limit.
Steps
1. Press the V key. The ACV
parameter will be editable.
V
V-Limit
2. Set the voltage with the scroll wheel/keypad
or with the F1 ~ F4 soft-keys.
Setting the Output Voltage
The voltage setting sets the voltage level of the power supply.
62
Page 63
OPERATION
Range
0 volts ~ full range
Soft-keys
DEF1, DEF2, MAX, MIN
3. Press Enter to confirm the voltage setting.
Preset Settings
The DEF1 and DEF2 preset settings are user defined settings. By default they are set to 0.00 volts. The MAX and MIN soft-keys set the voltage to the maximum and minimum, respectively.
4. Press the V key and set the desired voltage
with the scroll wheel/keypad.
Range
0 volts ~ full scale of voltage range
5. Press and hold the DEF1 or DEF2 soft-key
until “Saved to DEF1/2” is displayed. This will save the voltage setting to the DEF1 or DEF2 soft-key.
Note
Trying to set the voltage outside of the voltage limit/range will result in a voltage setting error being displayed on the screen.
The voltage level can be set when the output is on.
Example
Voltage setting
Preset voltage
settings
F3 F4
F1 F2
63
Page 64
APS-7000 Series User Manual
Background
Setting the frequency limit allows the frequency output to be set to any level within the limit range.
Steps
1. Press Shift + F to access the Freq
Limit menu.
Shift
+
F
F-Limit
2. Set the frequency limit with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the frequency limit to the maximum and minimum, respectively.
Range
45.00 ~ 500.0Hz (1000Hz option)
Soft-keys
MAX, MIN
3. Press Enter to confirm the limit setting.
Example
Freq Limit
Min/Max
settings F3 F4
Note
Before change freq limit setting, if FREQ setting value is bigger than desire freq limit value, so that the freq limit value can't be change.
Setting the Frequency Limit
64
Page 65
OPERATION
Background
Before setting the frequency, set the frequency limit.
Steps
1. Press the F key. The FREQ
parameter will be editable.
F
F-Limit
2. Set the frequency with the scroll
wheel/keypad or with the F1 ~ F4 soft-keys.
Range
45.00 ~ 500.0Hz (1000Hz option)
Soft-keys
DEF1, DEF2, MAX, MIN
3. Press Enter to confirm the frequency setting.
Preset Settings
The DEF1 and DEF2 preset settings are user defined settings. By default they are set to
50.00Hz and 60.00Hz, respectively. The MAX and MIN soft-keys set the frequency to the maximum and minimum, respectively.
4. Press the F key and set the desired frequency
with the scroll wheel/keypad.
Range
45.00 ~ 500.0Hz (1000Hz option)
5. Press and hold the DEF1 or DEF2 soft-key
until “Saved to DEF1/2” is displayed. This
will save the frequency setting to the DEF1 or DEF2 soft-key.
Setting the Output Frequency
The frequency setting sets the frequency of the output.
65
Page 66
APS-7000 Series User Manual
Example
Frequency setting
Preset frequency
settings
F3 F4
F1 F2
Note
Trying to set the frequency outside of the frequency limit will result in a frequency setting error being displayed on the screen.
The frequency can be set when the output is on.
Background
Setting the peak current limit sets a limit on the current that can be sourced by the power supply. Once the output current over the setting, the output will set to off.
Note
When the peak current limit is tripped, an alarm will sound. Press Shift + 9 to clear the Ipk alarm.
Shift+6 can also clear the Ipk alarm. See page 74 for details.
Steps
1. Press Shift + I rms to access the
Ipeak menu.
Shift
+
I rms
IPK-Limit
Setting the Peak Current Limit
66
Page 67
OPERATION
2. Set the peak current with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the peak current to the maximum and minimum, respectively.
Range
10% ~ 100% peak current value. The peak current value depends on the selected voltage range.
Soft-keys
MAX, MIN
3. Press Enter to confirm the peak current setting.
Example
Ipeak
Min/Max
settings F3 F4
Delay Time Settings
The Delay Time setting essentially defines how long the measurement of the peak current must be sustained for before it is recognized. By default the delay time is turned off.
Load on
Ipk Limit reached
I Pk Limit Delay
I
Ipk Limit acknowledged,
protection triggered
T
67
Page 68
APS-7000 Series User Manual
1. Press Shift + I rms and then press DELAY[F2].
2. Set the desired delay time with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the delay time to the maximum and minimum, respectively.
Range
0(off) ~ 10 seconds
Soft-keys
MAX, MIN
3. Press Enter to confirm the delay time setting.
Example
Delay Time
Min/Max
settings F3 F4
IPK Measurement Range Settings
The IPK Range settings allow you to manually set the peak current measurement range. By default this setting is set to AUTO.
1. Press Shift + I rms and then press IPK Range[F1].
2. Set the desired range with the scroll wheel.
Range
AUTO, 0.28A, 1.4A, 14A, 70A (APS-7050, APS-7100)
AUTO, 14A, 140A (APS-7200, APS-7300)
3. Press Enter to confirm the IPK Range setting.
68
Page 69
OPERATION
Example
APS-7050 & 7100
APS-7200 & 7300
IPK Range settings
IPK Range settings
Note
When using manual IPK range setting, Irms measure range as below
Model
RMS Current Measurement
APS-7050 APS-7100
2.00 ~ 70.00mA (0.28A Range)
2.0 ~ 350.0mA (1.4A Range)
0.020 ~ 3.500A (14A Range)
0.02 ~ 17.50A (70A Range)
APS-7200 APS-7300
0.100 ~ 3.500A (14A Range)
0.10 ~ 35.00A (140A Range)
69
Page 70
APS-7000 Series User Manual
Background
The I rms setting sets the root mean square current.
Setting the RMS current sets a limit on the current that can be sourced by the power supply. Once the output current over the setting, the output will set to off.
Steps
1. Press I rms to access the I rms
menu.
I rms
Ipk-Limit
2. Set the I rms level with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the I rms level to the maximum and minimum, respectively.
Range
0.00 ~ full scale A (dependant on the voltage range)
Soft-keys
MAX, MIN
3. Press Enter to confirm the current setting.
Example
Irms
Min/Max
settings F3 F4
Setting the Current RMS Level
70
Page 71
OPERATION
Note
Setting the I rms level to 0.00 will disable OCP and the OPP protection function is activated. The OPP is 105% of rating.
Setting the I rms to 0A is dangerous.
I rms Delay Time Settings
The Delay Time setting defines how long the I rms measurement must be sustained for before it is recognized. By default the I rms delay time is turned off.
Load on
Irms Limit
reached
Irms Limit Delay
I
Irms Limit
acknowledged,
protection triggered
T
Irms Limt
4. Press I rms and then press DELAY[F2].
5. Set the desired delay time with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the delay time to the maximum and minimum, respectively.
Range
0(off) ~ 10 seconds
Soft-keys
MAX, MIN
6. Press Enter to confirm the delay time setting.
71
Page 72
APS-7000 Series User Manual
Example
Delay Time
Min/Max
settings F3 F4
OC Fold Settings
The over current fold back settings allow the APS-7000 to work as either a constant voltage source or a constant current source.
While the unit is sourcing less current than the Irms current limit, the APS-7000 will act as a constant voltage source. In this mode, the voltage level will remain constant while the current level may vary. This is the normal operating mode.
When the current level reaches the Irms limit, the APS-7000 will act as a constant current source. In this mode the current is constant and the voltage level varies. When the current subsides below the Irms limit again, the unit will again act as a constant voltage source. When OC Fold is turned off, the APS-7000 will act as a current limiting power source when the Irms limit has been reached.
Vrms
setting
ACV
I
Irms limit
Irms
72
Page 73
OPERATION
Note
OC-FOLD can only be active when the I rms level is greater than 0.
7. Press I rms and then press OC-FOLD[F1]
toggle the OC-Fold function on or off.
OC-Fold setting
Background
The on phase setting sets the starting phase of the voltage output. The off phase setting sets the ending phase of the voltage output.
Steps
1. Press Shift + 7 or Shift + 4
the On Phase or Off Phase, respectively.
Shift
+
7
On Phase
or
4
Off Phase
2. Set the On Phase or Off Phase setting with the
scroll wheel/keypad or with the F3 ~ F4 soft­keys. The MAX and MIN soft-keys set the phase to the maximum and minimum, respectively.
Range
0 ~ 359º
Soft-keys
MAX, MIN
Setting the On/Off Phase
73
Page 74
3. Press Enter to confirm the phase setting.
Example
On Phase
Min/Max
settings
F3 F4
Off Phase
Alarm Clear
Background
The ALM CLR (Alarm Clear) function will clear any Over Power, Over Irms, Over Ipeak, Over Temperature, Output Short alarms and Remote Sense Error.
Applicable Alarms
OVER POWER, OVER IRMS, OVER IPEAK, OVER TEMPERATURE, OUTPUT SHORT, REMOTE SENSE ERROR.
Steps
1. Press Shift + 6 to clear any alarms.
Shift
+
6
ALM CLR
Example
Alarm message
ALM indicator
APS-7000 Series User Manual
74
Page 75
OPERATION
Steps
1. Press the Display key.
2. The display mode will toggle each
time the key is pressed.
Display
Standard Mode
Settings Measurements
Configurable
measurements
Hold
measurement
Simple Mode
Measurement Items
Hold
measurement
Configuring the Standard Mode Measurements
1. Press the Meas Item1, Item2 or
Item3 soft-key.
Display Modes
The APS-7000 power supply has two display modes. The standard display mode shows the power supply setup on the left and the 3 configurable measurements on the right. The simple display mode shows all measurement items available on the APS-7000.
75
Page 76
APS-7000 Series User Manual
2. Use the scroll wheel to select a measurement
item and press Enter to confirm.
Sets first measurement item to Voltage
Hold Measurement
The Hold function will “hold” the current measurements on the display. Measurements
won’t be updated on the display until the
function is released.
Press HOLD[F4] to toggle hold on or off.
When APS-7000 rear terminal has detected reverser current, the display would change as follow:
Reverse Current Mode
76
Page 77
OPERATION
Activate the panel lock
Press the Lock key to active the panel lock. “Keys locked” appears on the display.
A lock icon will appear in the top corner when the panel keys are locked.
Lock
Unlock
Disable the panel lock
Hold the Lock key for ~3 seconds to disable the panel lock. “Keys unlocked” will appear on the display and the padlock icon will disappear.
Lock
Unlock
(hold)
Example
Message Lock icon
Panel Lock
The panel lock feature prevents settings from being changed accidentally. When activated, all keys and knobs except the Lock/Unlock key and the Output key (if active) will be disabled.
If the instrument is remotely controlled via the USB/LAN/RS­232/GPIB interface, the panel lock is automatically enabled. See page 153 for remote control details.
77
Page 78
APS-7000 Series User Manual
Warning
Both of these outputs are electrically linked. Only one DUT should be connected to any one of the outputs at a time. Using both outputs at the same time is not supported. Using the front and rear outputs are the same time could cause dangerous operating conditions. See page 34 for details about using the output terminals or sockets.
Turn Output On
Press the Output key. The Output key will light up and ON will be displayed in the status bar to indicate that the output is on.
Output
Turn Output Off
Press the Output key. The Output key light will go out and OFF will be displayed in the status bar to indicate that the output is off.
Output
Turning the Output On
When the output is turned on, the DUT can be connected to either the rear panel output or the front panel output.
78
Page 79
OPERATION
Warning
Ensure the output is off before handling the remote sense connectors.
Use sense cables with a voltage rating exceeding the isolation voltage of the power supply.
Never connect sensing cables when the output is on. Electric shock or damage to the power supply could result.
Remote sense connectors overview
The remote sense connectors are located at the rear panel of the APS-7200 and APS-7300.
Local sense operation
When using local sense, the sensing terminals are not used. No compensation of any possible voltage drop seen on the load cables is performed. Local sense is only recommended when the voltage drop is of no consequence. By default, the power supply is configured for local sense.
1. Check that the remote sense setting is disabled
(page 93).
Using the remote sense (APS-7200 and APS-7300 only)
The APS-7200 and APS-7300 can be operated using local or remote voltage sense. By default, the power supply is configured for local sense.
Local Sense
79
Page 80
Remote Sense
Remote sense operation
Remote sense is used to compensate for the voltage drop seen across load cables due to resistance inherent in the load cables. The remote sense function can compensate a maximum of 10% of the output voltage.
Warning
Ensure the output is off before handling the remote sense connectors.
Use sense cables with a voltage rating exceeding the isolation voltage of the power supply.
Never connect sensing cables when the output is on. Electric shock or damage to the power supply could result.
1. Configure the remote sense setting to ON
(page 93).
2. Connect the Neutral terminal of the remote
sense terminal block to the Neutral terminal of the load.
3. Connect the Live terminal of the remote sense
terminal block to the Live terminal of the load.
APS-7000 Series User Manual
80
Page 81
OPERATION
Connection example
OUTPUT terminal block
Sensing terminal block
Load cables
Sensing cables
Sensing points
Load
APS-7200 / 7300
81
Page 82
APS-7000 Series User Manual
Parameter Settings
Mode
Auto: When the output is on, this mode will automatically generate a trigger at 0º to keep the surge or dip event repeatedly on site.
Manual: When the output is on, this mode will wait for the TRIG[F4] soft­key to be pressed before starting the surge or dip event on site.
OFF: Disables surge/dip control.
Note
The magnitude of the surge/dip part of the resultant waveform depends on the surge/dip ACV setting level.
Advanced Settings
Surge/Dip Control → from page 82 Ramp Control → from page 85
Surge/Dip Control
Surge and dip control allows the power supply to source artificial surges or dips in voltage to a DUT. The surge/dip control feature provides a fast method to generate a surge/dip voltage event on a nominal voltage.
There have 4 parameters for configuring this feature: Mode selection (Mode), surge/dip voltage (ACV), the start time (T1) and testing duration (T2) of the surge/dip voltage period. The nominal voltage and frequency settings are based on the Basic Operation section.
82
Page 83
OPERATION
Triggering Example:
Site selection: The surge/dip site is selected as shown below.
Trigger (0º for Auto)
Dip/Surge site
T2T1
ACV
100mS
ACV
Sets the ACV surge/dip level from the 0V level.
T1
Sets the T1 time.
Range: 0~22ms(Auto Mode)
Range: 0~99ms(Manual Mode)
T2
Sets the width of the surge/dip.
Range: 0~22ms(Auto Mode)
Range: 0~99ms(Manual Mode)
100mS
Fixed 100mS delay after triggering.
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 2, Surge/Dip
Control and press Enter. Alternatively, use the
short-cut key shift+8.
3. Go to the Mode setting using the scroll wheel
and press Enter. Select the desired mode and press Enter again to confirm.
The Manual mode will allow you to manually trigger the surge/dip site. The Automatic setting will automatically trigger the surge/dip site.
Mode
Manual, Auto, OFF
83
Page 84
APS-7000 Series User Manual
4. Set the remaining parameters.
Note: these parameters are not visible when MODE is set to OFF.
Remaining parameters
ACV, T1, T2
Exit
5. Press Exit[F4] to exit from the
Surge/Dip Control settings.
6. After exiting the menu, the
surge/dip control icon will appear in the status bar.
Triggering the Manual Surge/Dip Site
For the manual mode, the surge/dip site is determined by a manual trigger.
1. Configure the nominal voltage
and frequency settings. See the Basic Operation chapter for details.
Page: 62, 64
2. Turn the output on. The nominal
settings above will be output.
Page 78
3. Press Shift to arm the trigger.
Press TRIG[F4] to generate the trigger manually via the front panel.
Shift
Or, Alternatively, pulse the Trigger In pin on the J1 connector high to generate the trigger.
OR Page 150
Note
The TRIG soft-key is only available when the surge/dip control is set to
Manual
.
84
Page 85
OPERATION
Parameter Settings
Time
The Time setting allows you to set the ramp time as ms/Vrms.
Tup = ramp up time/1Vrms Tdn = ramp down time/1Vrms
1Vrms 1Vrms
Tup Tdn
Voltage
The Voltage setting allows you to set the ramp time as Vrms/ms.
Vup = voltage up/1ms Vdn = voltage down/1ms
Vup Vdn
1ms 1ms
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 3, Ramp
Control and press Enter. Alternatively, use the
short-cut key shift+5.
Ramp Control
The Ramp Control function controls how fast the voltage level ramps up and down. This function allows you to ramp the voltage as a unit of time or as a unit of voltage.
85
Page 86
APS-7000 Series User Manual
3. Go to the Mode setting using the scroll wheel
and press Enter. Select either Voltage or Time and press Enter again to confirm.
Mode
Voltage, Time
4. For Voltage mode, set Vup and Vdn.
Vup
0.01 ~ 99.99Vrms
Vdn
0.01 ~ 99.99Vrms
5. For Time mode, set Tup and Tdn.
Tup
0.1 ~ 999.9ms
Tdn
0.1 ~ 999.9ms
Exit
6. Press Exit[F4] to exit from the
Ramp Control settings.
7. Upon exiting the ramp icon will
appear in the status bar.
86
Page 87
OPERATION
Example
Settings: Mode=Time, Tup=1 msec, VAC=100V, Freq=50Hz, Ramp output=on.
Ramp up Tup=1msec
Example:
Settings: Mode=Voltage, Vdn=2Vrms, VAC=100V, Freq=50Hz, Ramp output=off.
Ramp down Vdn =2Vrms
87
Page 88
APS-7000 Series User Manual
Output on
Ipeak Hold time
I
Begin the peak current
measurement
T
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the T Ipeak, hold(msec) setting using the
scroll wheel and press Enter. Set the time and press Enter again to confirm.
T Ipeak
1 ~ 60,000 ms
Miscellaneous
The Miscellaneous menu contains miscellaneous parameter settings.
T Ipeak, hold → from page 88 Power ON Output → from page 90 Buzzer → from page 91 SCPI Emulation → from page 92 Remote Sense → from page 93 (for APS-7200 and APS-7300 only).
T Ipeak, hold
The T Ipeak, hold function sets the hold time for the peak current measurement. After the output is turned on, the APS-7000 will delay starting the peak current measurement by this hold time.
88
Page 89
OPERATION
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
89
Page 90
APS-7000 Series User Manual
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Power ON Output setting using the
scroll wheel and press Enter. Select a setting and press Enter to confirm.
ON
Set Output ON.
OFF
Set Output OFF.
SEQ
Execute the sequence that was loaded before the unit was last turned off.
SIM
Execute the simulation that was loaded before the unit was last turned off.
PROG
Execute the program that was loaded before the unit was last turned off.
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Power ON Output
The Power ON Output setting allows you to have the output turn on automatically after startup. The settings that are loaded are the last settings that were present in the standard mode before the unit was turned off last.
90
Page 91
OPERATION
Example
Settings
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Buzzer setting using the scroll wheel
and press Enter. Turn the setting on or off and press Enter again to confirm.
Buzzer
ON, OFF
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
Buzzer
The Buzzer setting turns the buzzer sound on or off for key presses and alarms.
91
Page 92
APS-7000 Series User Manual
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the SCPI Emulation setting using the
scroll wheel and press Enter. Choose the SCPI emulation mode and press Enter again to confirm.
SCPI Emulation
GW, EXTECH, N/A
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
SCPI Emulation
The SCPI Emulation setting supports both GW Instek as well as EXTECH SCPI remote commands. When set to EXTECH, the SCPI Emulation setting allows the APS-7000 to be a drop-in replacement for the EXTECH 6700 series by emulating the remote commands of the EXTECH units.
92
Page 93
OPERATION
Note
The remote sense function can compensate a maximum of 10% of the output voltage. The maximum output voltage when compensation is used is limited by the rated voltage.
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Remote Sense setting using the scroll
wheel and press Enter. Select the unit and press Enter to confirm.
Remote Sense
ON, OFF
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
Remote Sense (APS-7200 and APS-7300 only)
The remote sense function detects the output voltage at the sensing input terminal. This function compensates for voltage drops across the load cables when the load is connected to the APS-7200 or APS­7300 over a long distance.
93
Page 94
APS-7000 Series User Manual
Display
When the remote sense function is on, the displayed voltage value is the voltage measured at
the sense terminal and the symbol “SV” is
displayed next to the voltage reading on the standard and simple mode display.
Display
Warning
Before connecting the remote sense cables, turn off the output and peripherals. Please see page 79 for more information on the remote sense cabling instructions.
If the remote sense wires are loose or falling, the display would show a message as below.
94
Page 95
OPERATION
Pin Assignment (APS-7050 and APS-7100)
Trigger Out
Trigger In Out On / Off
COM
J1
Pin Assignment (APS-7200 and APS-7300)
Trigger Out
Trigger In
Out On / Off
COM
J1
Note
The following trigger control settings are not applicable when in the Simulation test mode or the Sequence test mode.
Steps
1. Press Shift + 2 to access the
Trigger Control menu.
Shift
+
2
Trigger
Trigger
The trigger settings are used to turn the output on or off or to set the behavior of the Trigger In and Trigger Out pins of the J1 Connector. The trigger can also be used in conjunction with the Sequence mode, Simulation mode and Surge/Dip mode. See page 150 for further details.
Trigger Control Settings
The trigger control menu sets both the trigger in and the trigger out behavior of the J1 connector.
95
Page 96
APS-7000 Series User Manual
2. The trigger control settings appear. The
settings are divided into Input Pin and Output Pin.
Trigger input
settings
Exit
Trigger output
settings
Trigger Input Pin
1. The Input Pin settings configure what action is
taken when the Trigger In pin on the J1 connector is high.
2. Go to the Action setting using the scroll wheel
and press Enter. Choose what will happen when the Trigger In pin is pulsed high(+5V) and then press Enter to confirm.
None
No action is taken. Remote trigger commands are accepted.
Output
Turns the output on or off when triggered.
Setting
Sets a user-defined voltage and frequency setting when triggered.
Preset
Loads a preset setting when triggered.
SurgeDip
Triggers the surge/dip control.
3. Go to the Width setting and set the minimum
pulse width to recognize the trigger input pulse.
0 Minimum pulse width (5ms)
1~ 60ms
1~ 60ms pulse width.
96
Page 97
OPERATION
4. If Output was chosen, set output status.
Status
ON, OFF
5. If Setting was chosen, Set Vset and Fset.
Vset
0 ~ full scale voltage (dependant on the chosen range; 600V optional)
Fset
45.00 ~ 500.0 Hz (1000Hz optional)
6. If Preset was chosen, choose which preset
setting should be loaded when the Trigger Input is set to high(+5V).
Memory
0 ~ 9
Trigger Output Pin
7. Go to the Source setting to choose what
function or test mode will be able to the trigger the output pin to pulse high.
None
No source is able to use the trigger output. Remote trigger commands are accepted.
Output
The trigger output pin is activated when the output is turned on or off.
Setting
When the settings change the trigger output pin is activated.
Preset
When a preset is loaded or saved, the trigger output pin is activated.
All
The trigger output pin is activated when any of the settings, presets or the output are changed.
8. Go to Width to set how long the trigger output
pin is pulsed high for.
Width
1 ~ 60ms
97
Page 98
APS-7000 Series User Manual
0ms: Stays high/low.
9. If Width, above, was set 0ms, you can choose
the output pin polarity with the Level parameter. This setting will pull the output pin high or low when source condition is true. This setting is not available when Width is set to 1~60ms.
Level
HI (≥+2V), LO(≤+0.8V)
Exit
10. Press Exit[F4] to exit from the
Trigger Control settings.
98
Page 99
OPERATION
[1] Remote Control Commands
The following remote control commands are applicable for the trigger input or output. See the programming manual for usage details.
*TRG :INITiate[:IMMediate]:NAME :INITiate[:IMMediate][:TRANsient] :MEMory:SAV :MEMory:RCL :MEMory:TRIGgered :MEMory:TRIGgered? :OUTPut[:STATe]:TRIGgered :OUTPut[:STATe]:TRIGgered? :SYSTem:CONFigure:TRIGger:INPut:MODE :SYSTem:CONFigure:TRIGger:INPut:MODE? :SYSTem:CONFigure:TRIGger:INPut:SOURce :SYSTem:CONFigure:TRIGger:INPut:SOURce? :SYSTem:CONFigure:TRIGger:OUTPut:MODE :SYSTem:CONFigure:TRIGger:OUTPut:MODE? :SYSTem:CONFigure:TRIGger:OUTPut:SOURce :TRIGger:OUTPut:SOURce :TRIGger:OUTPut:SOURce? :TRIGger:OUTPut[:IMMediate] :TRIGger:MEMory:SOURce :TRIGger:MEMory:SOURce? :TRIGger:MEMory[:IMMediate] :TRIGger[:TRANsient]:SOURce :TRIGger[:TRANsient]:SOURce? :TRIGger[:TRANsient][:IMMediate] [:SOURce]:FREQuency:TRIGgered [:SOURce]:FREQuency:TRIGgered? [:SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] [:SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
99
Page 100
APS-7000 Series User Manual
Steps
1. Press Preset and then hold
a number key (0~9) to save the present settings to the corresponding memory number.
Preset
+
0
~
9
(hold) Presets
M0 ~ M9
2. Press the Preset key again to exit from the
preset mode.
Example
For example, pressing Preset & holding 1 will save the present settings to memory slot 1 (saved to M1).
Note
The preset key will become green when active. A beep will be heard (Buzzer set to ON) and a message will displayed when the settings are saved.
Preset Settings
Save Presets to Local Memory → from page 100 Recall Presets to Local Memory → from page 101 Manage Preset Settings → from page 102
Save Preset Settings to Local Memory
Up to 10 preset settings can be saved to internal memory.
100
Loading...