Series 2-5000 Minihelic®II
Differential Pressure Gage
Specifications: Installation & Operating Instructions
ø2-19/32
[65.88]
1-7/16 [36.51]
9/32 [7.14]
Series 2-5000 Minihelic®II Differential
Pressure Gages have clean design, small size,
low cost and sufficient accuracy for all but the most
demanding applications. With housing molded from
mineral- and glass-filled nylon and a lens molded
from polycarbonate, this gage will withstand rough
use and exposure, as well as high total pressure up
to 30 psig [2.067 bar]. Over-pressure is accommodated by a blow-out membrane molded in conjunction with the diaphragm.
INSTALLATION
1. Select a location free from excessive vibration
and where ambient temperature will be between 20
to 120°F (-6.7 to 49°C). Sensing lines may be any
length necessary without affecting accuracy.
However, long runs of tubing will dampen readings
slightly and cause a minor increase in response
time. If pulsing pressure or vibration cause excessive pointer oscillation, please contact factory for
ways to provide additional damping.
2. This gage is calibrated and zeroed in the vertical
position at the factory. If the gage is used in any
other position, it must be re-zeroed each time the
position is changed. Gages with ranges under 5˝
w.c.(1.24 kPa), or the equivalent, should be used
only in the vertical position unless special calibration
was specified when ordering.
2-3/8 [60.33]
Bulletin A-36-A
LOW PRESSURE PORT
ø7/16 [11.11] HOLE IN
PANEL FOR SURFACE MOUNT
(2) ø5/32 HOLES
IN PANEL FOR SURFACE
ø3-1/16
[77.79]
1-[25.40]
2.330 [59.18]
SURFACE MOUNT
HOLES
SPECIFICATIONS
Service: Air and compatible gases.
Wetted Materials: Consult factory.
Housing: Glass filled nylon; polycarbonate lens.
Accuracy: ±5% of full scale at 70°F (21.1°C).
Pressure Limits: 30 psig (2.067 bar) continuous
to either pressure connection.
Temperature Limits: 20 to 120°F (-6.67 to
48.9°C).
Size: 2-1/16˝ (52.39 mm) diameter dial face.
Mounting Orientation: Diaphragm in vertical
position. Consult factory for other position orientations.
Process Connections: Barbed, for 3/16˝ I.D.
tubing (standard); 1/8˝ male NPT (optional).
Weight: 6 oz (170.1g).
CAUTION: FOR USE ONLY WITH AIR OR
COMPATIBLE GASES.
MOUNT
HIGH PRESSURE PORT
ø7/16 [11.11] HOLE IN
PANEL FOR SURFACE MOUNT
Figure A
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, IN 46361, U.S.A. Fax: 219/872-9057
Phone: 219/879-8000 www.dwyer-inst.com
e-mail: info@dwyer-inst.com
Bulletin A-36-A
#4 MOUNTING
#6 MOUNTING
SCREW
SCREW
MOUNTING
LUG
PANEL MOUNTED
PANEL MOUNTED INSTALLATION
3. To surface-mount the gage in a 3/16 (4.78
mm) max panel thickness, drill two 5/32˝ (3.97
mm) holes on a horizontal line, 2.330˝ (59.18
mm) apart for mounting screws. Next, drill two
7/16˝ (11.11 mm) holes 1.000˝ (25.4 mm) apart
on a vertical line for pressure connections.
Secure the gage with two #4 mounting screws.
Be careful not to block the slotted hole near the
right-hand mounting hole, this provides a path
for pressure relief in the event of over-pressurization.
4. To panel-mount the gage, cut a 2-5/8˝ (66.68
mm) diameter hole. Install mounting lugs to the
back of the gage with two #4 mounting screws,
thread two #6 mounting screws through mounting lugs and tighten.
®
5. Before installation, the Minihelic
Pressure Gage may need to be zeroed before
placing into operation. If zeroing is required, you
must first remove the locking clip located near
the bottom face of the gage (refer to Fig A). Use
a small blade screwdriver or the allen wrench
supplied with the Minihelic
Pressure Gage to remove the clip. The clip is
attached with a hinge design so it will not get
lost after removal. After removing the clip, firmly
hold the case with one hand and rotate the front
cover in the counterclockwise direction with the
palm of the other hand. The Minihelic
Differential Pressure Gage utilizes a bayonet
locking design. The cover will rotate approximately 9 degrees then stop and can be
removed. Caution: When removing the cover
be careful to slowly pull the cover straight off the
gage housing. Damage can occur to the scale if
the cover is pulled off on an angle. If difficult to
loosen cover, place a small sheet of rubber
between the cover and the palm of the hand.
The zero adjust screw is located below the
scale at the “zero” label near the base of the
pointer. Use the hex allen wrench supplied and
II Differential
®
II Differential
®
II
#4 MOUNTING
SCREW
3/16 (4.78) MAX
PANEL THICKNESS
SURFACE MOUNTED
6. To measure positive pressure, connect
tubing to port marked “HI” and vent “LO” port
to atmosphere. For negative pressure (vacuum), connect to port marked “LO” and vent
“HI” port to atmosphere. For differential pressure, connect higher pressure to port marked
“HI” and lower to “LO” port. If gage is supplied with 1/8˝ NPT connections, be careful
not to over-tighten fittings to avoid damage to
the gage.
CALIBRATION CHECK
Select a second gage or manometer of
known accuracy and in an appropriate range.
Use short lengths of rubber or vinyl tubing to
connect the high-pressure side of the gage
and the test gage to two legs of a tee. Very
slowly, apply pressure through the third leg.
Allow enough time for pressure to equalize
throughout the system and for fluid to drain. if
a manometer is being used. Compare readings. If the gage being tested exceeds rated
accuracy, it should be returned to the factory
for recalibration.
MAINTENANCE
No lubrication or periodic servicing is
required. Keep case exterior and cover clean.
Occasionally, disconnect pressure lines to
vent both sides of the gage to atmosphere
and re-zero per installation paragraph 5.
Upon final installation of the Series 2-5000
®
Minihelic
II Differential Pressure Gage, no
routine maintenance is required. A periodic
check of the system calibration is recommended. The Series 2-5000 is not field serviceable and should be returned if repair is
needed (field repair should not be attempted
and may void warranty). Be sure to include a
brief description of the problem plus any relevant application notes. Contact customer
service to receive a return goods authorization number before shipping.
adjust until the pointer is on zero. This must be
done with both pressure connections vented to
atmosphere and the gage oriented in the final
mounting position. Replace cover.
©Copyright 2007 Dwyer Instruments, Inc Printed in U.S.A. 2/07 FR# 14-440434-10 Rev.4
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, IN 46361, U.S.A. Fax: 219/872-9057
Phone: 219/879-8000 www.dwyer-inst.com
e-mail: info@dwyer-inst.com