Greisinger GTMU - MP User guide

WEEE-Reg.-Nr.: DE93889386
T20.0.4X.6C-03
Transmitter for temperature
from version V3.2 Operating Manual GTMU - MP
Content
1 INTENDED USE 2 2 GENERAL ADVICE 2 3 SAFETY INSTRUCTIONS 2 4 DISPOSAL NOTES 2 5 ASSIGNMENT OF ELBOW-TYPE PLUG (DEPENDING ON TYPE) 3 6 GENERAL INSTALLATION INSTRUCTIONS: 3 7 DESIGN TYPES, DIMENSION 3 8 DISPLAY FUNCTIONS 4
8.1 MEASURING DISPLAY 4
8.2 MIN/MAX VALUE MEMORY 4
9 ERROR AND SYSTEM MESSAGES 4 10 CONFIGURATION OF THE DEVICE 5 11 NOTES TO THE CALIBRATION SERVICES 6 12 SPECIFICATION 6
GREISINGER electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
+49 (0) 9402 / 9383-0+49 (0) 9402 / 9383-33 info@greisinger.de
T20.0.4X.6C-03 Operating Manual GTMU - MP page 2 of 6
This device must not be disposed as “residual waste”. To dispose this device, please send it directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly.
_____________________________________________________ _____________________________________________________________________________
1 Intended use
The device measures the temperature of air or non corrosive / non ionizing gases. The safety instructions (see chapter 3) have to be observed.
The device must not be used for purposes and under conditions for that the device had not been designed. The device must carefully dealt with and has to be used according to the specifications (do not throw, knock, etc.). It has to be protected against dirt. Do not expose the sensor to aggressive gases (like ammonia) for longer time. Avoid condensation, as after drying there may remain residues, which may affect the precision negatively. In dusty environment additional protection has to be applied (special protection caps).
2 General advice
Read through this document attentively and make yourself familiar to the operation of the device before you use it. Keep this document in a ready-to-hand way in order to be able to look up in the case of doubt.
3 Safety instructions
This device has been designed and tested in accordance to the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using it.
1. Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic conditions than those stated under “Specification”. Transporting the device from a cold to a warm environment condensation may result in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
2. General instructions and safety regulations for electric, light and heavy current plants, including domestic safety regulations (e.g. VDE), have to be observed.
3. If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully. Internal connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages impairing or destroying the device or another device connected.
4. Whenever there may be a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning: Do not use this product as safety or emergency stop device or in any other application where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage.
4 Disposal notes
T20.0.4X.6C-03 Operating Manual GTMU - MP page 3 of 6
_____________________________________________________ _____________________________________________________________________________
5 Assignment of elbow-type plug (depending on type)
4-20mA (2-wire connection) voltage (3-wire connection) voltage (4-wire connection)
1 = supply voltage +Uv 1 = signal + 1 = signal + 2 = GND / signal 3 = supply voltage +Uv 2 = signal - (4) = supply voltage -Uv 3 = supply voltage +Uv signal - (4) = supply voltage –Uv
The type “current-” or “voltage output” is set by works and cannot be changed.
6 General installation instructions:
To mount the connection cable (2-, 3-, or 4-wire depending on type of device) the elbow-type plug screw has to be loosened and the coupling insert has to be removed by means of a screw driver at the position indicated (arrow). Pull out connection cable through PG gland and connect to the loose coupling insert as described in the wiring diagram. Replace loose coupling insert onto the pins at the transducer housing and turn cover cap with PG gland in the direction desired till it snaps on (4 different starting positions at 90° intervals). Re-tighten the screw at the angle plug.
7 Design types, dimension
design-type 1 design-type 2 design-type 3 design-type 4
T20.0.4X.6C-03 Operating Manual GTMU - MP page 4 of 6
Display
Description
Possible fault cause
Remedy
Err.1
Measuring range exceeded
Wrong signal
Temperatures above the measuring range are not allowed.
Err.2
Measuring value below measuring range
Wrong signal
Temperatures below the measuring range are not allowed.
Err.7
System fault
Error in device
Disconnect from supply and reconnect. If error remains: return to manufacturer
Err.9
Sensor error
Sensor or cable defective
Check sensors, cable and connections, damages visible?
Er.11
Calculation not possible
Calculation variable missing or invalid
Check temperature
8.8.8.8
Segment test
The transducer performs a display test for 2 seconds after power up. After that it will change to the display of the measuring.
SET
1 2 3
°C
_____________________________________________________ _____________________________________________________________________________
8 Display Functions
8.1 Measuring display
During normal operation the temperature in [°C] or [°F] is displayed.
8.2 Min/Max Value Memory
watch Min values (Lo): press shortly once display changes between „Lo„ and Min values watch Max values (Hi): press shortly once display changes between „Hi„ and Max values restore current values: press or once again current values are displayed clear Min-values: press for 2 seconds Min values are cleared. Display shows shortly „CLr„. clear Max-values: press for 2 seconds Max values are cleared. Display shows shortly „CLr„.
After 10 seconds the currently measured values will be displayed again.
9 Error and system messages
T20.0.4X.6C-03 Operating Manual GTMU - MP page 5 of 6
Parameter
value
information
SET
and
UNIT
Unit of temperature displays factory setting: °C
°C
Temperatures in °Celsius
°F
Temperatures in °Fahrenheit
VT
Display at zero output of humidity display value (output scaling)
-xxx.0 … +xxx.0
Entry of display value at which the temperature output should deliver 4 mA (or 0V) (factory setting: 0.0)
VT
Display at maximum output for humidity display value (output scaling)
-xxx.0 … +xxx.0
Entry of display value at which the temperature output should deliver 20 mA (or 1 V / 10 V) (factory setting: 100.0)
Offset correction of humidity measuring *)
oFF
deactivated (factory setting)
-5.0 … +5.0
Selectable from -5.0 to +5.0 % rel. humidity
Scale correction of humidity measuring *)
oFF
deactivated (factory setting)
-15.00 … +15.00
Selectable from -15.00 to +15.00 % scale correction
ATTENTION!
If the jumper is removed from the shown contacts, the configuration is inaccessible, values are protected.
Never connect other contacts!
_____________________________________________________ _____________________________________________________________________________
10 Configuration of the device
Follow these instructions to configure the functions of the device:
Press SET until the first parameter UNIT appears in the display If a parameter should be changed, press or ,
The device changed to the setting – edit with or Confirm the value with SET Jump to the next parameter with SET.
Pressing SET again stores the settings, the instruments restarts (segment test)
Please note: If there is no key pressed within the menu mode within 2 minutes, the
configuration will be cancelled, the entered settings are lost!
*) if higher values are needed, please check sensor, if necessary return to manufacturer for inspection.
Calculation: corrected value = (measured value – Offset) * (1+Scale/100)
T20.0.4X.6C-03 Operating Manual GTMU - MP page 6 of 6
Measuring range
-50.0 ... 400.0 °C or -58.0 ... 752.0 °F
Accuracy Display Add. Output signal
(at nom. temperature 25°C) ±0.4% of meas. value. ±0.2°C ± 0.2% FS
Media
Non corrosive gasses
Sensor
Pt1000
Meas. frequency
1 per second
Adjusting
Digital offset and scale adjustment for humidity and temperature
Min-/Max-value memory
Min and max measured values are stored
Output signals
refer to type plate,
Scaling
both freely scalable by entering display values for 4mA (resp. 0V) and 20mA (resp. 1V/10V) output
Connection
For option AV01, AV10
4 - 20 mA (2-wire) 0 - 1 (10) V (3-wire)
Auxiliary energy: (supply voltage, each output)
4 - 20 mA: Uv = 12 - 30 V DC 0-1V: Uv = 12 - 30 V DC, max. 10mA 0-10V: Uv = 18 - 30 V DC, max. 10mA
Perm. impedance
4 - 20 mA: RA(Ohm) < ( (Uv - 12V) / 0,02A )
example: for Uv = 18V: RA < (18V - 12V) / 0.02A => RA < 300 Ohm
Permissible load
0 - 1 (10) V: RL(Ohm) > 3000 Ohm
Display
approx. 10 mm high, 4-digit LCD-display
Operating elements
3 keys
Ambient conditions Nom. temperature Operating temperature
Relative humidity Storage temperature
25°C Electronics: -25 ... 70 °C
Please consider that the allowable range of the housing and electronics must not be exceeded, especially when using the probe at higher temperatures!
Electronics: 0 ... 95 %RH (not condensing)
-25 ... 70 °C
Housing
ABS (IP65, except sensor head)
Dimensions
82 x 80 x 55 mm (without elbow-type plug and sensor tube) for Option Kabel: Sensor head Ø14mm*68mm, 1m teflon cable, high humidity sensor
Mounting
Holes for wall mounting (in housing - accessible after cover has been removed).
Mounting distance
50 x 70 mm, max. shaft diameter of mounting screws is 4 mm
Electrical connection
Elbow-type plug conforming to DIN 43650 (IP65), max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm
EMC
The device corresponds to the essential protection ratings established in the Regulations of the Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EG). In accordance with EN 61326-1 : 2006, additional errors: <1 % FS. When connecting long leads adequate measures against voltage surges have to be taken.
_____________________________________________________ _____________________________________________________________________________
11 Notes to the calibration services
Calibration certificates - DKD-certificates - other certificates: If device should be certificated for its accuracy, it is the best solution to return it with the referring sensors to the manufacturer. (please state desired test values, e.g. 25°C)