
H08.0.1X.6C-17
Operating Manual GTH 175/Pt
Intended use:
The GTH 175/Pt measures the temperature in liquids,
air/gas and soft media.
The measurement is performed with a Pt1000 probe,
permanently connected to the measuring instrument.
• High accuracy and precision are the strengths of the
GTH 175/Pt.
• The device can be used as a reference measuring
instrument for the calibration of other measuring
systems.
GREISINGER electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Tel.: +49 9402 9383-0, Fax: +49 9402 9383-33, eMail: info@greisinger.de

H08.0.1X.6C-17
Dispense exhausted batteries at destined gathering places.
The device must not be disposed in the unsorted municipal waste. Send
the device directly to us (sufficiently stamped), if it should be disposed.
We will dispose the device appropriate and environmentally sound.
General
Read through this document attentively and make yourself familiar to the of the
device before you use it. Keep this document in a ready-to-hand way in order to be
able to look up in the case of doubt
Disposal Notice
Safety Instructions
This device has been designed and tested in accordance to the safety regulations
for electronic devices. However, its trouble-free operation and reliability cannot be
guaranteed unless the standard safety measures and special safety advises given
in this manual will be adhered to when using it.
1. Trouble-free operation and reliability of the device can only be guaranteed if it
is not subjected to any other climatic conditions than those stated under
“Specification”.
Transporting the device from a cold to a warm environment condensation may
result in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
2. Whenever there may be a risk whatsoever involved in running it, the device
has to be
switched off immediately and to be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
there is visible damage to the device
the device is not working as specified
the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.
3. Warning: Do not use these product as safety or emergency stop device, or in
any other application where failure of the product could result in personal injury
or material damage.
Failure to comply with these instructions could result in death or serious injury
and material damage.
4. Do not heat the probe over 200 C

H08.0.1X.6C-17
1. Display of temperature in C
2. Low battery voltage, replace battery
Hint: when battery voltage falls to low, BAT warning cannot be indicated.
check the battery in case of implausible measuring values.
To change battery: Remove battery cover at the backside of the device (push
cover forward), change the battery and close the cover.
The on/off switch is located at the left side of the instrument.
Operating and Maintenance Advice
The battery has to be removed, when storing device above 50°C.
Treat device and probes carefully. Use only in accordance with above specifica-
tion. (do not throw, hit against etc.).
Protect from soiling, especially in the area of the gas inlet at the front plate.
Du not pull at the probe cable!
Tip: We recommend to take out battery if device will not be operated for
some time!
Operating
Display and operating elements
Adjusting
The measuring device will be calibrated before
leaving our works. If you want to adjust the device,
please proceed as follows:
1.Zero point adjusting (0°C): Put ice cubes in a
glass and pour cold water until ice cubes are almost covered. Put sensor into
glass, and stir water. Wait for stable value and turn zero point potentiometer (NP)
by means of a screw driver until display shows "0.00"
2.Scale adjusting: For scale adjustment a known reference temperature is
needed. Expose the probe to the reference and adjust the display using the potentiometer “Scale”
Calibration Service
calibration certificate - DKD certificate - official certifications:
If the measuring instrument is supposed to receive a calibration certificate, it has
to be sent to the manufacturer. (declare test levels, e.g. 0°C; 70°C)

H08.0.1X.6C-17
± 0.1 % of meas. value ± 2 digit (at nom. temperature)
(at range –70.0 ... +199.9 °C – at lower values refer to correction table below)
Pt1000, electrically insulated and mounted in stainless
steel tube of 3mm dia and approx. 100mm length,
temperature-resistant up to 200 C
Probe permanently connected to the device.
Immersion probe: large plastic handle ~ 135 mm
1m ultra flexible silicone cable,
handle and cable temperature-resistant up to 100 C
Insertion probe with tip: large plastic handle ~ 135 mm
1m ultra flexible silicone cable,
handle and cable temperature-resistant up to 100 C
Core temperature probe: small Teflon handle ~ 75 mm
1m Teflon cable,
handle and cable temperature-resistant up to 250 C
approx. 13 mm high, 3½-digit LCD
switch for on/off, 2 adjusting potentiometer
-30 to +45 °C, 0 to 80 %RH (non condensing)
9V-battery type 6F22 (included)
"BAT" displayed automatically in case of low battery
Impact-resistant ABS-housing, front IP65
~ 106 x 67 x 30 mm (H x W x D), without probe
~ 190 g incl. battery and probe
The device corresponds to the essential protection ratings
established in the Regulations of the Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (2004/108/EG).
Additional error: <1%
temperature display
-200 -201.9
-195 -196.7
-190 -191.6
-185 -186.4
-180 -181.2
-175 -176.1
temperature display
-170 . -171.0
-165 -165.9
-160 -160.8
-155 -155.7
-150 -150.6
-145 -145.5
temperature display
-140 -140.4
-135 -135.4
-130 -130.3
-125 -125.3
-120 -120.2
-115 -115.2
temperature display
-110 -110.1
-105 -105.1
-100 -100.1
-90 -90.0
-80 -80.0
-70 -70.0
Specification
Correction table for low temperatures