Greisinger GRMU 2000 MP Operating Manual

T02.0.1X.6C-02 page 1 of 3
Operating Manual for Redox-Measuring transducer
GRMU 2000 MP
Specification:
Measuring range: -2000 to 2000mV or refer to type plate Accuracy: ±0,2% Output signal: refer to type plate Connection: 4 - 20 mA (2-wire)
Voltage (3- resp. 4-wire)
Electric isolation: input electrically isolated Auxiliary energy:
Reverse voltage protection: 50V permanent
Perm. impedance
Permissible load (at 0-...V): RL(Ohm) > 3000 Ohm Electrode: any Redox-Electrode of our program is suitable.
Input resistance: 10 Electrode socket: Cinch-socket (standard), optionally available with BNC- and DIN-socket Display: none
Optional display: approx. 10 mm high, 3½-digit LCD-display Nominal temperature: 25°C Operating temperature: 0 to 50°C Relative humidity: 0 to 95 % r.F. (non-condensing) Storage temperature: -20 to 85°C Housing: ABS (IP65 - with the exception of electrode) Dimensions: 82 x 80 x 55 mm (without elbow-type plug and sensor sockets) Mounting: With fixing holes for wall mounting (in housing - accessible after cover has been
Mounting distance: 50 x 70mm, max. shaft diameter of mounting screws is 4 mm. Electric connection: elbow-type plug conforming to DIN 43650 (IP65),
EMC: The
(supply voltage) Uv = 12 - 30 V DC (4-20mA)
Uv = 18 - 30 V DC (0-10V) or refer to type plate
(at 4-20mA): R
Example: for Uv = 18V: RA < (18V - 12V) / 0.02A => RA < 300 Ohm
(electrode not included in scope of supply!)
removed).
max. wire cross section: 1.5 mm², wire/cable diamter from 4.5 to 7 mm
of the Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (89/336/EWG). In accordance with EN50081-1 and EN50082-1 Additional error: <1%
(at nominal temperature)
(Ohm) < ( (Uv - 12V) / 0.02A )
A
12
Ohm
GRMU2000MP corresponds to the essential protection ratings established in the Regulations
Disposal notes
This device must not be disposed as ‘residual waste’. To dispose this device, please send it directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly.
GREISINGER electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Fon: +49 9402/8500, Fax: +49 9402/1829, eMail: info@greisinger.de
T02.0.1X.6C-02 page 2 of 3
Assignment of elbow-type plug:
2-wire connection (4-20mA) 3-wire connection (voltage) 4- wire connection (voltage)
1
Uv
2
A
1 = supply voltage +Uv 1 = signal + 1 = signal + 2 = GND / signal 3 = supply voltage +Uv 2 = signal -
3
1
V
= supply voltage -Uv 3 = supply voltage +Uv signal -
Uv
3
1
V
2
= supply voltage -Uv
Uv
General installation instructions:
To mount the connection cable (2-, 3-, or 4-wire depending on type of device) the angle plug screw has to be loosened and the coupling insert has to be removed by means of a screw driver at the position indicated (arrow). Pull out connection cable through PG glanding and connect to the loose coupling insert as described in the wiring diagram. Replace loose coupling insert onto the pins at the transmitter housing and turn cover cap with PG glanding in the direction desired till it snaps on (4 different starting positions at 90° intervals). Re-tighten the screw at the angle plug. Before connecting to supply voltage, ensure that a working electrode is connected.
Safety instructions:
This device has been designed and tested in accordance with the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other
climatic conditions than those stated under "Specification". If the device is transported from a cold to a warm environment condensation may cause in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
2. General instructions and safety regulations for electric, light and heavy current plants, including domestic safety
regulations (e.g. VDE), have to be observed.
3. If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully. Internal
connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages impairing or destroying the device or another device connected.
4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked
accordingly to avoid re-starting. Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time. In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning:
Do not use these product as safety or emergency stop devices, or in any other application where failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.