This device measures the oxygen concentration in gas mixtures and air. The actual
measurement takes place at the opening of the sensor. Due to the design of the
sensor, the device has to be calibrated at regular intervals (at fresh air =20.95%
oxygen) to get accurate measuring values. If the sensor is used up, this will be
detected at calibration and the sensor element has to be replaced before the next
measurement.
2 General note
Read this document carefully and get used to the handling of the device before you
use it. Keep this paper ready to hand in order to look it up if a question turns up.
H59.0.21.6C-10 page 3 of 8
4 Safety instructions
This device has been designed and tested in accordance to the safety regulations
for electronic devices. However, its trouble-free operation and reliability cannot be
guaranteed unless standard safety measures and special safety advises given in
this manual will be adhered to when using it.
1. Trouble-free operation and reliability of the device can only be guaranteed if it
is not subjected to any other climatic conditions than those stated under
“Specification”.
2. Transporting the device from a cold to a warm environment condensation may
result in a failure of the function. In such a case make sure the device
temperature has adjusted to the ambient temp. before trying a new start-up.
3. The circuitry has to be designed most carefully if the device should be
connected to other devices. Internal connection in third party devices (e.g.
connection GND and earth) may result in not-permissible voltages impairing or
destroying the device or another device connected.
4. Whenever there may be a risk whatsoever involved in running it, the device
has to be switched off immediately and to be marked accordingly to avoid restarting. Operator safety may be a risk if:
- there is visible damage to the device or the device is not working as specified.
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or
maintenance.
5. Warning: Do not use this product as safety or emergency stop device or in
other applications where failure of the product could result in personal injury or
material damage. Failure to comply with these instructions could result in death
or serious injury and material damage.
6. This device only serves as supervision by the monitoring of essential or other
for the customer important systems.
It must not be used instead of compulsory approval monitoring devices and it is
not designed for that purpose. If this device is used for the monitoring of such
systems on its own, the manufacturer will not assume liability for damages
whatsoever.
7. Caution, acid! The sensor contains KOH. This can cause severe chemical
burns. If leaking, avoid contact!
If there was contact:
- to skin: Flush contacted area with large amounts of water
for several minutes.
- to clothing: remove contaminated clothing.
- to eyes: Flush with large amounts of water for several
minutes, obtain medical treatment.
After swallowing:
- give large volumes of water. DO NOT induce vomiting!
- Obtain medical treatment.
H59.0.21.6C-10 page 4 of 8
1. display of the oxygen concentration in %
2. BAT-warning: battery is used up and has to be
replaced (measuring is valid)
On/Off switch
Press shortly: change between display of actual value-,
:Min-value display,
:Max- value display
Press long: delete min/max values
Press shortly: short display of the sensor rating (calculated after the last
calibration) (i.e. “100.P“ = 100%)
Press long: start calibration (see also calibration)
5 Operating and maintenance advice
The battery has to be taken out, when the device is stored above 50°C.
Handle the device and the sensor with care and only use it according its
specification. The connectors have to be saved from humidity and dirt.
Unplug the sensor not by pulling the cable, but the plug.
Tip: It is recommended to take the battery out, when the device is stored for a
longer period of time.
6 Operation
Display- and Operation elements
First start
If not already done: plug the sensor to the device.
Switch the device on with the
Start the calibration by pressing the
displayed. After the calibration the device is ready for measuring.
key.
key for 2 seconds, until is
H59.0.21.6C-10 page 5 of 8
7 Oxygen measuring
The oxygen measurement depends on air pressure. The current pressure can be
entered during configuration.
If you measure at atmosphere and the calibration took place at the same
conditions, the pressure value has not to be adjusted necessarily, because the
same pressure is used for calibration and measuring.
If the measurement is carried out at other pressure conditions than the
calibration: The actual pressure has to be entered in the configuration for
measurement and calibration.
The sensor‟s temperature dependency is compensated internally. However for
best results try to keep the same temperature both during calibration and
measuring.
Temperature differences between sensor and gas may cause additional errors!
Therefore wait an appropriate period of time until the sensor has adjusted its
temperature to the gas that is measured. A suitable flow of the gas around the
sensor increases the adjustment significantly.
Try to avoid warming of the sensor by touching it during measurement or
calibration.
When measuring bottled gas, consider that the gas coming out is cooler than the
ambient because of the pressure loss! Strong air flows can produce an over
pressure at the sensor – possible source of measuring/calibration errors!
The nominal life time can be shortened by:
- Wrong storage / operation temperature
- Permanent use with dry gases (compressed gas).
It helps to put the sensor to normal-humid ambient air in measuring breaks
(“flush” system with fresh air).
The optimum operation position is: with the sensor inlet pointing downwards
8 Calibration
In order to compensate for ageing of the sensor and air pressure fluctuations due
to weather changes, the sensor has to be calibrated at regular intervals. We
recommend to calibrate it at least once a week or, for optimum measuring results,
directly before starting the measuring process. Adjust the absolute pressure at the
configuration point “P.Ab” before carrying out a calibration.
Calibration: The sensor will be calibrated to the atmospheric oxygen concentration
of 20.95%. The sensor has to be subjected to air (make sure that rooms are
thoroughly aired).
Press the key for 2 sec till appears.
The calibration will be automatically completed as soon as the measuring values
for oxygen are stable (takes a few seconds). After the calibration the device will
show the rating of the sensor state for a short time.
If an error message (CE.3, CE.4, CE.6) is displayed, the sensor signal is
invalid. Press any key to restart, the previous calibration will be restored.
H59.0.21.6C-10 page 6 of 8
11 Errors and system messages
no
display
The battery is used up and has to be exchanged, or the device is
defective
The battery is used up, for a short period further measurements are
possible
The battery is used up and has to be replaced, measurements are no
more possible
Calibration error: sensor voltage is too low (sensor used up or wrong
chosen air pressure)
Calibration error: sensor voltage is too high (wrong chosen air
pressure or sensor defective (membrane crack, etc.)
Calibration error: signal not stable
Measuring error: range has been exceeded
Measuring error: measuring value has fallen below permitted range
System error: the device is defective or considerably out of the
allowed ambient temperature range
12 Display when switching-on the device
Start and display test
If switch-off delay “P.oF” (power off) is active, this is signalled at the
turn-on procedure.
9 Sensor rating
The rating of the sensor state is calculated and saved after each successful
calibration.
Watch sensor rating: Press shortly the
time i.e.
The rating is displayed in 10 percent steps: 100% = optimal sensor status. Lower
values are indicating that the sensor life time will soon be reached. (50% doesn‟t
mean 50% of the life time, but 50% of the reference signal!). But also a erroneous
pressure entry may cause low sensor ratings.
= 100%.
key, the display shows for a short
10 Replace sensor element
Unplug the sensor, unscrew the white protection housing and take the sensor
element. Take the rubber disk off and put it on the new sensor. Put the new sensor
element into the housing, screw it down and plug it.
Then start the calibration.
H59.0.21.6C-10 page 7 of 8
Altitude
Abs. air pressure
Altitude
Abs. air pressure
0 m
1013 mbar
800 m
921 mbar
100 m
1001 mbar
1000 m
899 mbar
200 m
989 mbar
1200 m
877 mbar
300 m
978 mbar
1400 m
856 mbar
400 m
966 mbar
1600 m
835 mbar
500 m
954 mbar
1800 m
815 mbar
600 m
943 mbar
2000 m
795 mbar
13 Configuration of the device
1. Switch the device off. Press the
until is displayed (about 3s).
Switch-off delay (factory setting: 20):
The switch-off delay is shown in minutes. If no key is pressed, the instrument
automatically switches-off after the entered period of time.
2. Press
3. Enter your desired switch-off delay using the
values are:
off:The switch-off delay is deactivated (permanent operation)
1...120: The switch-off delay in minutes.
4. Confirm the value via the
Enter the absolute air pressure (or pressure of the gas that is measured). Its
value depends on your altitude (see the table below) and the present weather.
5. Press the key, the current value appears.
6. Enter the current air pressure using the
Selectable values are: 500 .. 1999 mbar
7. Confirm the value via the
will reboot afterwards.
key, the currently selected value is displayed.
Air pressure (factory setting: 980 mbar)
key during the switch-on of the device,
and keys. Selectable
key. appears in the display.
and keys.
key. The values will be saved and the device
If no key is pressed within 60 seconds during the configuration, the set-up
will be aborted.
Eventually made changes won’t be stored!
Air pressure depending on the altitude above sea level
when activated, the device switches automatically off, if it is
not operated for longer time (selectable 1..120min).
Housing:
impact-resistant ABS, transparent panel, front side IP65
Dimensions:
approx. 106 x 67 x 30 mm (L x W x D) without sensor cable
Weight:
approx. 185g incl. battery and sensor
EMC:
The device corresponds to the essential protection ratings
established in the Regulations of the Council for the
Approximation of Legislation for the member countries
regarding electromagnetic compatibility (2004/108/EG).
Device meets EN 61326-1:2006, Additional fault: <1%