HFR: “A” and “B” Side Flow Meter (One for each
side)
Part No.Description
24J319S3000 Flow Meter Kit
24J320G3000 Flow Meter Kit
24J321G3000HR Flow Meter Kit
24J322HG6000 Flow Meter Kit
24J323HG6000HR Flow Meter Kit
HFR for NVH: Flow Meter Kits
Part No.Description
24T18224:1 and 16:1 NVH Cart
Flow Meter Kit
24T1831:1 NVH Cart Flow Meter Kit
24T20024:1 and 16:1 NVH Modular
Flow Meter Kit
24T2011:1 NVH Modular Flow Meter Kit
HFR: Flow Meter Calibration Kit
(Per Applicator)
Part No.Description
24J324L-Head Flow Meter Calibration Kit
24J325S-Head Flow Meter Calibration Kit
24J326P2 Flow Meter Calibration Kit
24J357GX-16 Flow Meter Calibration Kit
24F227EP/Fusion Flow Meter Calibration Kit
255247MD2 1:1 Flow Meter Calibration Kit
255245MD2 10:1 Flow Meter Calibration Kit
23A1657F
Related Manuals
Related Manuals
Component manuals in English. Manuals are available
at www.graco.com.
Manual No.Description
Systems
313997HFR Operation
313998HFR Repair-Parts
3A2797HFR for NVH Foam - Cart,
Setup-Operation
3A1961HFR for NVH Foam - Modular,
Setup-Operation
Flow Meters
3087783000 Series Flow Meters
3098346000 Series Flow Meters
Dispense Valves
312753L-Head Operation-Maintenance
312752S-Head Operation-Maintenance
313536GX-16 Operation
313872EP Gun, Instructions-Parts
Overview
On HFR systems, the flow meters are used only as a
method of monitoring system performance, which will
help with system maintenance. The ratio displayed on
the ADM is a running average of the data collected.
NOTE: The flow meters will not correct any ratio errors
caused by system performance.
When the ratio monitoring system is active, the ratio will
be displayed on the screen below in the form of ratio:1
(e.g. 24.03:1). The display will show “--:--” when the ratio
monitoring system is not active. The recommended minimum value for an alarm is 3%.
NOTE: For circulation or NVH systems, the ratio monitoring system only becomes active when the machine is
in high pressure circulation.
313213Probler P2, Instructions-Parts
309550Fusion Gun Instructions-Parts
312185MD2 Valve Instructions-Parts
313380GX-7 DI Auto-Robotic Spray Gun,
Operation-Maintenance
Accessories
3A0395HFR and VRM Tank Feed Systems
Instructions-Parts
3A0861Ratio Check Assembly for Fusion/EP
Guns
3A1657F3
Installation
Installation
Flow Meter Electronics Kit
24J318
1. Perform Shutdown procedure. See HFR or NVH
operation manual for detailed instructions.
2. Use two 111800 hex head screws (supplied) to
mount electronics assembly 24J318 to the bottom of
the electrical enclosure. If there are transformers
installed, remove two of the hex head screws from
the transformers and use those two holes to mount
the electronics assembly and discard the supplied
hex head screws.
NOTE: For NVH machines, mount electronics assembly
to the bottom of the “B” side electrical enclosure.
3. Connect CAN cable to CAN connection labeled “1”
on the fluid control module.
ti17011a
4. Connect opposite end of the CAN cable (111) to any
free CAN port on any other fluid control module or
temperature control module in the system. If none
are available, use CAN cables (111 and 113) to connect to CAN splitter inside the power distribution
box. See references on page 20.
ti17009a
ti17010a
43A1657F
Installation
HFR: 3000 Series Flow Meter
1. Perform HFR Pressure Relief Procedure. See
HFR operation manual for detailed instructions.
2. Perform dispense valve Pressure Relief Proce-dure. See dispense valve manual, page 3, for
detailed instructions.
3. Verify all air, hydraulic, and material pressures have
been relieved before continuing.
4. Perform HFR Shutdown procedure. See HFR oper-
ation manual for detailed instructions.
5. If the flow meter is for the A (Red) side, connect inlet
fitting (2) and outlet fitting (4) to the flow meter. See
F
IG. 1.
If the flow meter is for the B (Blue) side, connect
inlet fitting (3) and outlet fitting (5) to the flow meter.
See F
IG. 1.
NOTE: The connections for each side of the machine
are different sizes to prevent attaching the incorrect
hoses. The A (Red) side has the smaller connections.
8. Route the flow meter data cable as shown in F
IG. 2.
ti17013a
FIG. 2: Flow Meter Data Cable Routing
9. Connect the other end of the flow meter data cable
to the fluid control module installed in the electrical
enclosure.
If the data cable is for the A (Red) side flow meter,
connect to port 1 on the FCM. See F
IG. 2.
If the data cable is for the B (Blue) side flow meter,
connect to port 2 on the FCM. See F
IG. 2.
A
2 or 3
B
4 or 5
ti17012a
FIG. 1: Flow Meter Installed on Fluid Manifold --
A (Red) side shown
6. Connect the flow meter assembly to the fluid
manifold (A) on the HFR system as shown in F
IG. 1.
7. Connect the 3-pin connector (B) on the flow meter
data cable to the flow meter. See F
IG. 1.
ti17014a
FIG. 3: Flow Meter Data Cable Connections
3A1657F5
Installation
HFR: 6000 Series Flow Meter
1. Perform HFR Pressure Relief Procedure. See
HFR operation manual for detailed instructions.
2. Perform dispense valve Pressure Relief Proce-dure. See dispense valve manual, page 3, for
detailed instructions.
3. Verify all air, hydraulic, and material pressures have
been relieved before continuing.
4. Perform HFR Shutdown procedure. See HFR oper-
ation manual for detailed instructions.
5. Remove all fittings from the outlet ports on the front
of the fluid manifold. See F
each fitting is from to make sure they are installed
on the correct side later.
IG. 4. Note which side
8. Connect flow meter and fittings to swivel fitting
already connected to fluid manifold. See F
IG. 5.
304
301
303
302
ti17016a
FIG. 5: Connect Fittings to Flow Meter
9. Connect the flow meter assembly to the fluid manifold as shown. See F
IG. 6.
10. Re-install the fittings that were previously removed.
See F
IG. 6.
NOTICE
Make sure to install the fittings in the correct material lines. Failure to do so will result in material
cross-contamination of the fittings and material
hoses.
ti17015a
FIG. 4: Remove Fluid Manifold Fittings
6. Connect swivel fitting (302) to fluid manifold. See
F
IG. 5.
7. Connect fittings (303 and 304) to flow meter (301).
See F
IG. 5.
11. Connect the 4-pin connector of the flow meter data
cable to the flow meter. See F
IG. 6.
ti17017a
FIG. 6: Flow Meter Installation
63A1657F
12. Route the flow meter data cable as shown in FIG. 7.
ti17013a
FIG. 7: Flow Meter Data Cable Routing
13. Connect the other end to the fluid control module
installed in the electrical enclosure.
If the data cable is for the A (Red) side flow meter,
connect to port 1 on the FCM. See F
IG. 2.
If the data cable is for the B (Blue) side flow meter,
connect to port 2 on the FCM. See F
IG. 8.
Installation
ti17014a
FIG. 8: Flow Meter Data Cable Connections
3A1657F7
Installation
NVH Flow Meters
1. Perform HFR for NVH Pressure Relief Procedure.
See HFR for NVH operation manual for detailed
instructions.
2. Perform dispense valve Pressure Relief Proce-dure. See dispense valve manual, page 3, for
detailed instructions.
3. Verify all air, hydraulic, and material pressures have
been relieved before continuing.
4. Perform HFR for NVH Shutdown procedure. See
HFR operation manual for detailed instructions.
5. Remove all fittings from the inlet and outlet ports on
the front of the fluid manifold. See F
IG. 9.
6. If applicable, connect the A (Red) and B (Blue)
return fittings. See F
IG. 10.
FIG. 10: Connect Return Fittings
7. Connect A (Red) flow meter and fittings to fluid
manifold. See F
IG. 11.
FIG. 9: Remove Fluid Manifold Fittings
83A1657F
FIG. 11: A (Red) Flow Meter Installation
Installation
8. Connect B (Blue) flow meter and fittings to fluid
manifold. See F
IG. 12.
1:1 Ratio
24:1 or 16:1
Ratio
12. Connect the other end to the fluid control module
installed in the electrical enclosure.
If the data cable is for the A (Red) side flow meter,
connect to port 1 on the FCM. See F
IG. 14.
If the data cable is for the B (Blue) side flow meter,
connect to port 2 on the FCM. See F
IG. 14.
ti17014a
FIG. 14: Flow Meter Data Cable Connections
13. Install token into the FCM and turn the system
power on. Once the software installation is complete, remove the token.
FIG. 12: B (Blue) Flow Meter Installation
9. Connect the supply and return hoses from the applicator to the corresponding fittings.
10. Connect the 4-pin connector of the flow meter data
cable to the flow meter.
11. Route the flow meter data cable as shown in
F
IG. 13.
NOTE: For NVH - Modular systems, route the flow
meter data cable as shown in F
IG. 7, page 7.
FIG. 13: Flow Meter Data Cable Routing
ti17219a
FIG. 15: Token Installation
3A1657F9
Loading...
+ 19 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.