Manual system for proportional mixing of plural component coatings.
For professional use only.
Approved for use in explosive atmospheres (except the EasyKey and 3KS Power Supply
Module).
Important Safety Instructions
Read all warnings and instructions in this
manual. Save these instructions.
See page 4 for model information, including maximum working
pressure. Equipment approval labels are on page 3. Some
components shown are not included with all systems.
313881ProMix 3KS Kit Installation
313883ProMix 3KS Kit Repair-Parts
312775ProMix 2KS Manual System Installation
312776ProMix 2KS Manual System Operation
312777ProMix 2KS Manual System Repair-Parts
312781Fluid Mix Manifold
312782Dispense Valve
312783Color Change Valve Stacks
312787Color Change Module Kit
312784Gun Flush Box Kits
310745Gun Air Shutoff Kit
312786Dump Valve and Third Purge Valve Kits
312785Network Communication Kits
308778G3000/G3000HR Flow Meter
313599Coriolis Flow Meter
313290Floor Stand Kit
313542Beacon Kit
313386Basic Web Interface/Advanced Web Inter-
face
40679915V256 Automatic System Upgrade Kit
40680015V825 Discrete I/O Board Kit
Equipment Approvals
Equipment approvals appear on the following labels
which are attached to the Fluid Station and Power Supply Module. See F
Power Supply Module and Fluid Station Label
ProMix® 3KS
Electronic Proportioner
FM08ATEX0074
II 2 G
Ex ia IIA T3
MAX AIR WPR
.77100
MPabarPSI
MAX FLUID WPR
MPabarPSI
MAX TEMP 50°C (122°F)
PART NO.SERIES SERIAL MFG. YR.
Intrinsically safe equipment
for Class I, Div 1, Group D, T3
CUS
Ta = -20°C to 50°C
Install per 289833
IG. 1 on page 4 for label locations.
ATEX Certificate is listed here
Intrinsically Safe (IS) System. Install
per IS Control Drawing No. 258682.
EasyKey Interface IS Associated
Apparatus for use in non hazardous
CUS
Intrinsically safe
equipment for Class I,
Div 1, Group D, T3
Ta = -20°C to 50°C
location, with IS Connection to Smart
Fluid Plate IS
Apparatus for use in:
Class I, Division 1, Group D T3 C
Hazardous Locations
Read Instruction Manual
Warning: Substitution of components
may impair intrinsic safety.
PART NO. SERIES
MFG. YR.
Fluid Station Label
®
ProMix 3KS
FM08ATEX0073
II 2 G
Ex ia IIA T3
FLUID PANEL
MAX AIR WPR
.77
MPabarPSI
SERIAL
GRACO INC.
P.O. Box 1441
Minneapolis, MN
55440 U.S.A.
TI14376a
100
GRACO INC.
P.O. Box 1441
Minneapolis, MN
55440 U.S.A.
TI14374a
ATEX Certificate is listed here
Power Supply Module Label
ProMix 3KS
PART NO.
GRACO INC.
P.O. Box 1441
Minneapolis, MN
55440 U.S.A.
®
SERIES NO. MFG. YR.
Intrinsically safe connections
for Class I, Div 1, Group D
CUS
Ta = -20°C to 50°C
Install per 258682
Um: 250 V
POWER REQUIREMENTS
VOLTS
AMPS
85-250 ~
2 AMPS MAX
50/60 Hz
II (2) G
[Ex ia] IIA
FM08ATEX0072
TI14375a
ATEX Certificate is listed here
313882C3
Page 4
System Configuration and Part Numbers
System Configuration and Part Numbers
Configurator Key
The configured part number for your equipment is printed on the equipment identification
labels. See F
each of the following categories, depending on the configuration of your system.
IG. 1 for location of the identification labels. The part number includes digits from
3K
System
Component C
Fluid MeterComponent C Change
TK0 = No Meter
1 = G3000
2 = G3000HR
3 = 1/8 in. Coriolis
4 = Solvent Meter
Label Location
on Fluid Station
Maximum Fluid
Working Pressure
is listed here
0 = No Valves (single component C)
1 = Two Valves (low pressure)
2 = Four Valves (low pressure)
3= Two Valves (high pressure)
4= Four Valves (high pressure)
ProMix® 3KS
Electronic Proportioner
FM08ATEX0074
II 2 G
Ex ia IIA T3
CUS
Intrinsically safe
equipment for Class I,
Div 1, Group D, T3
Ta = -20°C to 50°C
MAX AIR WPR
.77100
MPabarPSI
MAX FLUID WPR
MPabarPSI
MAX TEMP 50°C (122°F)
Intrinsically Safe (IS) System. Install
per IS Control Drawing No. 258682.
EasyKey Interface IS Associated
Apparatus for use in non hazardous
location, with IS Connection to Smart
Fluid Plate IS
Apparatus for use in:
Class I, Division 1, Group D T3 C
Hazardous Locations
Read Instruction Manual
Warning: Substitution of components
may impair intrinsic safety.
PART NO. SERIES
MFG. YR.
Not
Designated
00
SERIAL
GRACO INC.
P.O. Box 1441
Minneapolis, MN
55440 U.S.A.
Not
Designated
Label Location on
Power Supply Module
TI14370aTI14361a
Configured Part Number
TI14376a
FIG. 1: Identification Label
4313882C
Page 5
System Configuration and Part Numbers
Hazardous Location Approval
Models using a G3000, G3000HR, or intrinsically safe Coriolis meter for A, B, and C meters are approved for
installation in a Hazardous Location - Class I, Div I, Group D, T3 or Zone I Group IIA T3.
Maximum Working Pressure
Maximum working pressure rating is dependent on the fluid component options selected. The pressure rating is
based on the rating of the lowest rated fluid component. Refer to the component pressure ratings below.
Example: Model TK1400 has a maximum working pressure of 3000 psi (21 MPa, 210 bar).
Check the identification label on the EasyKey, power supply module, or fluid station for the system maximum working pressure. See F
ProMix Fluid Components Maximum Working Pressure
Base System (no meters [option 0], no color/component C change [option 0]) . . . . .3000 psi (21.0 MPa, 210 bar)
ProMix 3KS Power Supply Module
Wall Mount Fluid Station, 50 cc Integrator and Static Mixer
IS Power Cable, red color coded, 50 ft (15.25 m)
CAN Communication Cable, green color coded, 10 ft (3.05 m)
Meter and Solenoid Cable, 10 ft (3.05 m)
Third Component Network Cable, yellow color coded, 6 ft (1.83 m)
C Side Dump Valve, if color valve(s) selected
313882C5
Page 6
System Configuration and Part Numbers
6313882C
Page 7
Accessories
Accessory
Gun Flush Box Gun Insert Selection
15V354 Third Purge Valve Kit
15V536 Solvent Flow Switch Kit
15V213 Power Cable, 100 ft (30.5 m)
15G710 Fiber Optic Cable, 100 ft (30.5 m)
15V034 10 cc Integrator Kit
15V033 25 cc Integrator Kit
15V021 50 cc Integrator Kit
24B618 100 cc Integrator Kit
15W034 Strobe Light Alarm Indicator Kit
15V337 Advanced Web Interface
15V256 Automatic Mode Upgrade Kit
Accessories
313882C7
Page 8
Warnings
Warnings
The following warnings are for the setup, use, grounding, maintenance, and repair of this equipment. The exclamation point symbol alerts you to a general warning and the hazard symbol refers to procedure-specific risk. Refer back
to these warnings. Additional, product-specific warnings may be found throughout the body of this manual where
applicable.
WARNING
FIRE AND EXPLOSION HAZARD
Flammable fumes, such as solvent and paint fumes, in work area can ignite or explode. To help prevent
fire and explosion:
•Use equipment only in well ventilated area.
•Eliminate all ignition sources; such as pilot lights, cigarettes, portable electric lamps, and plastic drop
cloths (potential static arc).
•Keep work area free of debris, including solvent, rags and gasoline.
•Do not plug or unplug power cords, or turn power or light switches on or off when flammable fumes
are present.
•Ground all equipment in the work area. See Grounding instructions.
•Use only grounded hoses.
•Hold gun firmly to side of grounded pail when triggering into pail.
•If there is static sparking or you feel a shock, stop operation immediately. Do not use equipment
until you identify and correct the problem.
•Keep a working fire extinguisher in the work area.
ELECTRIC SHOCK HAZARD
This equipment must be grounded. Improper grounding, setup, or usage of the system can cause
electric shock.
•Turn off and disconnect power at main switch before disconnecting any cables and before servicing
equipment.
•Connect only to grounded power source.
•All electrical wiring must be done by a qualified electrician and comply with all local codes and
regulations.
INTRINSIC SAFETY
Only models with a G3000, G250, G3000HR, G250HR, or intrinsically safe Coriolis meter for A, B, and C
meters are approved for installation in a Hazardous Location - Class I, Div I, Group D, T3 or Zone I
Group IIA T3. To help prevent fire and explosion:
•Do not install equipment approved only for a non-hazardous location in a hazardous area. See the
ID label for the intrinsic safety rating of your model.
•Do not substitute or modify system components as this may impair intrinsic safety.
SKIN INJECTION HAZARD
High-pressure fluid from gun, hose leaks, or ruptured components will pierce skin. This may look like just
a cut, but it is a serious injury that can result in amputation. Get immediate surgical treatment.
•Tighten all fluid connections before operating the equipment.
•Do not point gun at anyone or at any part of the body.
•Do not put your hand over the spray tip.
•Do not stop or deflect leaks with your hand, body, glove, or rag.
•Follow Pressure Relief Procedure in this manual, when you stop spraying and before cleaning,
checking, or servicing equipment.
8313882C
Page 9
Warnings
WARNING
EQUIPMENT MISUSE HAZARD
Misuse can cause death or serious injury.
•Do not operate the unit when fatigued or under the influence of drugs or alcohol.
•Do not exceed the maximum working pressure or temperature rating of the lowest rated system
component. See Technical Data in all equipment manuals.
•Use fluids and solvents that are compatible with equipment wetted parts. See Technical Data in all
equipment manuals. Read fluid and solvent manufacturer’s warnings. For complete information
about your material, request MSDS forms from distributor or retailer.
•Check equipment daily. Repair or replace worn or damaged parts immediately with genuine manufacturer’s replacement parts only.
•Do not alter or modify equipment.
•Use equipment only for its intended purpose. Call your distributor for information.
•Route hoses and cables away from traffic areas, sharp edges, moving parts, and hot surfaces.
•Do not kink or over bend hoses or use hoses to pull equipment.
•Keep children and animals away from work area.
•Comply with all applicable safety regulations.
TOXIC FLUID OR FUMES HAZARD
Toxic fluids or fumes can cause serious injury or death if splashed in the eyes or on skin, inhaled, or
swallowed.
•Read MSDS’s to know the specific hazards of the fluids you are using.
•Store hazardous fluid in approved containers, and dispose of it according to applicable guidelines.
•Always wear impervious gloves when spraying or cleaning equipment.
PERSONAL PROTECTIVE EQUIPMENT
You must wear appropriate protective equipment when operating, servicing, or when in the operating
area of the equipment to help protect you from serious injury, including eye injury, inhalation of toxic
fumes, burns, and hearing loss. This equipment includes but is not limited to:
•Protective eyewear
•Clothing and respirator as recommended by the fluid and solvent manufacturer
•Gloves
•Hearing protection
313882C9
Page 10
Glossary of Terms
Glossary of Terms
Air Chop - the process of mixing air and solvent
together during the flush cycle to help clean the lines
and reduce solvent usage.
Analog - relating to, or being a device in which data are
represented by continuously variable, measurable,
physical quantities, such as length, width, voltage, or
pressure.
Closed Loop Flow Control - refers to the process
when the flow rate is adjusted automatically to maintain
a constant flow.
Coriolis Meter - a non-intrusive flow meter often used in
low flow applications or with light viscosity, shear sensitive, or acid catalyzed materials. This meter uses vibration to measure flow.
Digital Input and Output - a description of data which
is transmitted as a sequence of discrete symbols, most
commonly this means binary data represented using
electronic or electromagnetic signals.
Discrete I/O - refers to data that constitutes a separate
entity and has direct communication to another control.
Dose Size - the amount of resin (A), catalyst (B), and
reducer (C) that is dispensed into an integrator.
Dose Time Alarm - the amount of time that is allowed
for a dose to occur before an alarm occurs.
Ethernet - a method for directly connecting a computer
to a network or equipment in the same physical location.
Fiber Optic Communication - the use of light to transmit communication signals.
Fill Time - the amount of time that is required to load
mix material to the applicator.
Flow Control Resolution - a settable value that allows
the flow control system to maximize its performance.
The value is based on maximum desired flow rates.
Flow Rate Analog Signal - the type of communication
signal that can be used on the ProControl module.
Flow Rate Tolerance - the settable percent of acceptable variance that the system will allow before a flow
rate warning occurs.
Flow Set Point - a predefined flow rate target.
Grand Total - a non-resettable value that shows the
total amount of material dispensed through the system.
Gun Trigger Input Signal - used to manage ratio assurance dose times and flow control processes.
Intrinsically Safe (IS) - refers to the ability to locate certain components in a hazardous location.
Idle - if the gun is not triggered for 2 minutes the system
enters Idle mode. Trigger the gun to resume operation.
Job Total - a resettable value that shows the amount of
material dispensed through the system for one job. A job
is complete when a color change or complete system
flush occurs.
K-factor - a value that refers to the amount of material
that passes through a meter. The assigned value refers
to an amount of material per pulse.
Ki - refers to the degree fluid flow over shoots its set
point.
Kp - refers to the speed in which the fluid flow reaches
its set point.
10313882C
Page 11
Glossary of Terms
Manual Mode - when the proportioning or flow control
system is controlling the inputs without any input from
an outside control.
Mix - when cross-linking of the resin (A), catalyst (B),
and reducer (C) occurs.
Mix Input Signal- refers to system mode status where
system begins a dose sequence each time the mix signal is made “High”.
Modbus/TCP - a type of communication protocol used
to communicate Digital I/O signals over an ethernet.
Network Station - a means to identify a particular individual proportioning or flow control system.
Overdose Alarm - when either the resin (A), catalyst
(B), or reducer (C) component dispenses too much
material and the system cannot compensate for the
additional material.
Potlife Time - the amount of time before a material
becomes unsprayable.
Potlife Volume - the amount of material that is required
to move through the mix manifold, hose and applicator
before the potlife timer is reset.
Color/Catalyst Dump - refers to the time required to
flush the lines from the color or catalyst change module
to the mix manifold during a color or catalyst change.
Purge - when all mixed material is flushed from the system.
Purge Time - the amount of time required to flush all
mixed material from the system.
Ratio Tolerance - the settable percent of acceptable
variance that the system will allow before a ratio alarm
occurs.
Sequential Color Change - the process when a color
change is initiated and the system automatically flushes
the old color and loads a new color.
Sequential Dosing - Components A, B, and C dispense
sequentially in the necessary volumes to attain the mix
ratio.
Solvent Fill - the time required to fill the mixed material
line with solvent.
Standby - refers to the status of the system.
Third Purge Valve - refers to the use of three purge
valves used to flush some waterborne materials. The
valves are used to flush with water, air and solvent.
V/P - refers to the voltage to pressure device in the flow
control module.
Color/Catalyst Fill - refers to the time required to fill the
lines from the color or catalyst change module to the mix
manifold.
313882C11
Page 12
Overview
Overview
Usage
The Graco ProMix 3KS Kit adapts a ProMix 2KS system to be an electronic 3-component paint proportioner. It can
blend most 3-component solvent and waterborne epoxy, polyurethane, and acid-catalyzed paints. It is not for use with
“quick-setting” paints (those with a potlife of less than 15 minutes).
•Can proportion at ratios from 0.1:1 to 50:1 in 0.1
increments with the wall mount fluid station.
•Has user selectable ratio assurance and can maintain up to +/-1% accuracy, depending on materials
and operating conditions.
•Models are available to operate air spray or
air-assisted systems with a capacity of up to 3800
cc/min.
•Color change options are available for low pressure
(300 psi [2.1 MPa, 21 bar]) air spray and high pressure (3000 psi [21 MPa, 210 bar]) systems with up
to 25 color change valves and up to 4 catalyst and 4
component C change valves.
NOTE: Optional accessories are available for in field
installation to achieve 25 colors or for component C
change.
Component Identification and Definition
See FIG. 2 and Table 1 for the system components. Components marked with a star (★) are available as part of the
ProMix 3KS Kit. All other components are part of the ProMix 2KS System or are available as accessories.
EK
PSM★
CCM
PS◆PS★◆
BCV
FO
ACV
2KS
MA
2KM
◆ See the ProMix 3KS Repair-Parts manual
for optional cable lengths.
IG. 2. ProMix 3KS System, shown with G3000 Meters and Color/Catalyst/Component C Change
F
12313882C
6 ft (1.83 m) fluid
hose (supplied)
MB
3KS★
CCV★
MC★
3KM★
BC
TI14543a
Page 13
Table 1: Component Descriptions
ComponentDescription
Overview
EasyKey (EK)
★ 3KS Power
Supply Module
(PSM)
Booth Control (BC)
2KS Fluid Station
(2KS)
★ 3KS Fluid Station
(3KS)
2K Fluid Manifold
(2KM)
★ 3K Fluid Manifold
(3KM)
Used to set up, display, operate, and monitor the system. The EasyKey accepts 85-250
VAC, 50/60 Hz line power and converts that power to acceptable low voltage and optical
signals used by other system components.
Accepts 85-250 VAC, 50/60 Hz line power and converts that power to acceptable low voltage signals used by other system components.
Used by the operator for daily painting functions including: choosing recipes, initiating job
complete, reading/clearing alarms, and placing the system in Standby, Mix, or Purge
mode. It is typically mounted inside the booth or near the painter.
Includes air control solenoids, flow switches, and mountings for the A, B, and solvent flow
meters and the 2K fluid manifold assembly. Its control board manages all proportioning
functions.
Includes air control solenoids, flow switches, and mountings for the C and solvent flow
meters and the 3K fluid manifold assembly. Its control board manages all proportioning
functions.
•Pneumatically Operated Dose Valves for component A and B
•Purge Valves for solvent and air purge
•Sampling Valves for calibrating the flow meters and performing ratio checks
•Shutoff Valves for component A and B to close their fluid passages to the mix mani-
fold, to allow for accurate calibration and ratio checks
•Mix Manifold, which includes the fluid integrator and static mixer.
➜ Fluid Integrator is the chamber where component A and B align at the
selected ratio and begin to mix.
➜ Static Mixer has 24 elements to uniformly blend the materials downstream
of the fluid integrator.
•Pneumatically Operated Dose Valve for component C
•Purge Valve for solvent
•Sampling Valve for calibrating the flow meter and performing ratio check
•Shutoff Valve for component C to close the fluid passage to the mix manifold, to
allow for accurate calibration and ratio check
•Mix Manifold, which includes the fluid integrator and static mixer.
➜ Fluid Integrator is the chamber where component C aligns with blended
components A/B at the selected ratio and begins to mix.
➜ Static Mixer has 24 elements to uniformly blend the materials downstream
of the fluid integrator.
313882C13
Page 14
Overview
Table 1: Component Descriptions
ComponentDescription
Flow Meters (MA,
MB, ★MC, MS)
Color Change
Valves (ACV) and
Color Change
Module (CCM)
Catalyst Change
Valves (BCV)
★ Component C
Change Valves
(CCV)
Four optional flow meters are available from Graco:
•G3000 is a general purpose gear meter typically used in flow ranges of 75-3800
cc/min. (0.02–1.0 gal/min.), pressures up to 4000 psi (28 MPa, 276 bar), and viscosities of 20–3000 centipoise. The K-factor is approximately 0.119 cc/pulse.
•G3000HR is a high resolution version of the G3000 meter. It is typically used in flow
ranges of 38–1900 cc/min. (0.01–0.5 gal/min.), pressures up to 4000 psi (28 MPa,
276 bar). and viscosities of 20–3000 centipoise. The K-factor is approximately 0.061
cc/pulse.
•S3000 is a gear meter used for solvents in flow ranges of 38-1900 cc/min. (0.01–0.50
gal/min.), pressures up to 3000 psi (21 MPa, 210 bar), and viscosities of 20–50 centipoise. The K-factor is approximately 0.021 cc/pulse.
•Coriolis is a specialty meter capable of a wide range of flow rates and viscosities.
This meter is available with 1/8 in. or 3/8 in. diameter fluid passages. For detailed
information on the Coriolis meter, see manual 313599.
The K-factor is user-settable; at lower flow rates use a lower K-factor.
➜ 1/8 in. fluid passages: set K-factor to .020 or .061.
➜ 3/8 in. fluid passages: set K-factor to .061 or 0.119.
An optional component. It is available as a color change valve stack for either low or high
pressure with up to 25 color change valves. Each stack includes one additional valve for
solvent to clean the fluid line between color changes.
An optional component. It is available as a catalyst change valve stack for either low or
high pressure with up to 4 catalyst change valves. Each stack includes one additional
valve for solvent to clean the fluid line between catalyst changes.
An optional component. It is available as a component C change valve stack for either low
or high pressure with up to 4 component C change valves. Each stack includes one additional valve for solvent to clean the fluid line between component C changes.
Dual Fiber Optic
Used to communicate between the EasyKey and Wall Mount Fluid Station.
Cable (FO)
★ Fluid Station
Used to provide power to the Wall Mount Fluid Station.
Power Supply Cable
(PS)
Applicator
Not shown. See ProMix 2KS manuals for details.
Handling: use Air
Flow Switch (AFS)
or Gun Flush Box
(GFB)
14313882C
Page 15
Overview
MA
RVA
APV
FI
DVC
AT
DVA
SVASVB
FI
DVB
SM
Key: ProMix 2KS Fluid Station
MAComponent A Meter
DVA Component A Dose Valve
RVA Component A Sampling
Val ve
SVA Component A Shutoff
Val ve
MBComponent B Meter
DVB Component B Dose Valve
RVB Component B Sampling
Val ve
MB
MS
RVB
TI12556a
SPV
SVB Component B Shutoff
Val ve
MSSolvent Meter
SPV Solvent Purge Valve
APV Air Purge Valve
SMStatic Mixer
FIFluid Integrator
ATAir Purge Valve Air Supply
Tu b e
Key: ProMix 3KS Fluid Station
MC
CPV
RVC
1
SVC
MCComponent C Meter
DVC Component C Dose Valve
RVC Component C Sampling Valve
SVC Component C Shutoff Valve
CPV Component C Purge Valve
SMStatic Mixer
FIFluid Integrator
1
3KS fluid inlet. Connect fluid supply line from
2KS fluid manifold outlet here.
2
Connect fluid supply line to gun.
SM
2
TI14382a
FIG. 3. ProMix 2KS and ProMix 3KS Wall Mount Fluid Stations
313882C15
Page 16
Booth Control
Booth Control
Used by the operator for daily painting functions including: changing recipes, signalling job complete, reading/clearing alarms, and placing the system in Standby,
Mix, or Purge mode. It is typically mounted inside the
booth or near the painter.
Table 2: Booth Control Key and Indicator Functions
(see F
IG. 4)
Key/Indicator Definition and Function
Display•Displays recipe number in Run
mode.
•If an alarm occurs, displays the
alarm code (E1 to E31) and red
Alarm indicator blinks.
•Recipe number displays after
alarm is reset.
Recipe
Indicator
•Green LED stays lit while a recipe
is in use.
•LED shuts off when Up or
Down keys are pressed or if an
alarm occurs.
•LED blinks while a new recipe is
loading and turns solid after loading is complete.
•LED blinks when purging.
•Select a new recipe by pressing
Table 2: Booth Control Key and Indicator Functions
IG. 4)
(see F
Key/Indicator Definition and Function
Down KeyScrolls recipe numbers down.
Mix Mode Key •Starts Mix mode*.
•Green LED remains lit while in Mix
mode or in Idle mode.
*After doing a purge, press key
once. The system will do a mixed
material fill sequence, then go to
Standby. Press again to start Mix
mode.
Standby Mode
Key
•Starts Standby mode.
•Green LED remains lit while in
Standby mode.
Purge Mode
Key
•Starts Purge mode.
•Green LED remains lit while in
Purge mode.
Up
or Down keys, then
pressing Enter .
Alarm Reset
Key and
Indicator
•Red LED blinks when an alarm
occurs.
•Press key to reset alarm. LED
shuts off after alarm is reset.
Job Complete
Key and
Indicator
•Signals that job is complete, and
resets A, B, and C totalizers.
•Green LED blinks once after key is
pressed.
Enter KeyEnters selected recipe and starts color
change sequence.
Up KeyScrolls recipe numbers up.
Alarm Indicator (red)
Recipe Indicator (green)
F
IG. 4. Booth Control (see Table 2)
Display
TI11614A
16313882C
Page 17
EasyKey Display and Keyboard
EasyKey Display and Keyboard
LCD Display
F
IG. 5. EasyKey Display and Keypad
Keypad
TI11630A
Navigation KeysAlarm Reset Key
Display
Shows graphical and text information related to setup
and spray operations. Back light will turn off after 10
minutes without any key press. Press any key to turn
back on.
Keypad
Used to input numerical data, enter setup screens, scroll
through screens, and select setup values.
In addition to the numbered keys on the EasyKey keypad, which are used to enter values in setup, there are
keys to navigate within a screen and between screens,
and to save entered values. See Table 3.
Table 3: EasyKey Keypad Functions (see FIG. 5)
KeyFunction
Setup: press to enter or exit Setup mode.
Enter: if cursor is in menu box, press Enter
key to view menu. Press Enter to save a
value either keyed in from the numerical
keypad or selected from a menu.
Up Arrow: move to previous field or menu
item, or to previous screen within a group.
Down Arrow: move to next field or menu
item, or to next screen within a group.
Left Arrow: move to previous screen group.
Right Arrow: move to next screen group.
Alarm Reset: resets alarms.
313882C17
Page 18
EasyKey and 3KS Power Supply Module Connection Ports
EasyKey and 3KS Power Supply Module Connection
Ports
EasyKey
AC Power
Audible AlarmFiber Optic Strain
Relief Port
Main Power
Access Port
Switch
Ground Screw
Ground Screw
I/S Power
to 2KS
Panel
I/S Power
to 3KS
Panel
Graco Web
Interface
Discrete I/O Cable
Connector Ports
TI12638aTI12657a
3KS Power Supply Module
TI14366aTI14372a
AC Power
Switch
Main Power
Access Port
FIG. 6. Connections and AC Power Switches
18313882C
Page 19
EasyKey and 3KS Power Supply Module Connection Ports
EasyKey AC Power Switch
Turns system AC power on or off.
3KS Power Supply Module AC
Power Switch
Turns 3KS AC power on or off.
EasyKey I/S Power
Power circuit to 2KS Fluid Station.
3KS Power Supply Module I/S
Power
Power circuit to 3KS Fluid Station.
EasyKey Potlife Exceeded
Audible Alarm
Alerts the user when a Potlife Exceeded alarm occurs.
EasyKey Graco Web Interface
Port
Used to communicate from a PC to:
➜ Upgrade software
➜ View software version
➜ Download
You can access data on an office or industrial network
through the internet with the proper configuration. See
manual 313386 for more information.
Clear by pressing the Alarm Reset
key.
313882C19
Page 20
Run Mode Screens
Run Mode Screens
NOTE: See FIG. 9 for a map of the Run screens.
Detailed screen descriptions follow.
Splash Screen
At power up, the Graco logo and software revision will
display for approximately 5 seconds, followed by the
Status Screen (see page 22).
F
IG. 7. Splash Screen
The Splash screen will also momentarily display “Establishing Communication.” If this display remains for more
than one minute, check that the fluid station circuit board
is powered up (LED is on) and that the fiber optic cable
is properly connected (see ProMix 2KS Installation manual).
NOTE: If the software version of the fluid plate does not
match the version of the EasyKey, the EasyKey will
update the fluid plate, and the fluid plate programming
screen will appear until the update is completed.
F
IG. 8. Fluid Plate Programming Screen
20313882C
Page 21
Run Mode Screens
Press the Setup key to
enter Setup mode.
TI12783a
FIG. 9. Run Screens Map
313882C21
Page 22
Run Mode Screens
Status Screen
•Use the Up or Down keys to scroll through the
Run screens.
•Press the Setup key to enter the Setup screens
from the Status screen.
•The other keys have no function in this Status
screen.
1
2
3
4
5
6
7
10
11
9
8
6
Potlife Timer: shows remaining potlife time in min-
utes. Two times are shown if there are two guns.
7
Status Bar: shows current alarm or operation mode
(standby, mix, purge, recipe change, or the current
alarm).
8
Current Flow Rate: in cc/min.
9
Animation: when the gun is triggered, the gun
appears to spray and the component A, B, or C
hose lights up, showing which component dose
valve is open.
10
Current Date and Time
11
Screen Number and Screen Arrows: displays the
current screen number and the total number of
screens in a group. The Up and Down arrows on the
right edge of the screen indicate the scroll feature.
The total number of screens in some groups may
vary depending on system configuration selections.
F
IG. 10. Status Screen
Key to F
1
2
3
4
5
IG. 10:
Active Recipe: shows the active recipe.
NOTE: At power up the system defaults to Recipe
61, which is not a valid recipe number.
Stage 1 Target Ratio (A:B): for the active recipe.
The ratio can be from 0.0:1–50.0:1, in 0.1 increments.
Stage 1 Actual Ratio (A:B): in hundredths, calculated after each dose of A and B.
Stage 2 Target Ratio (A+B):C: for the active recipe.
The ratio can be from 0.0:1–50.0:1, in 0.1 increments.
Stage 2 Actual Ratio (A+B):C: in hundredths, calculated after each dose of C.
22313882C
Page 23
Run Mode Screens
Totals Screen
IG. 11. Totals Screen
F
This screen shows the job totals, grand totals, and job
number. Use the tabs to reset job totals (Job Complete),
reset solvent totals (Rst Solvent), or go to Level Control Screen, page 24.
Solvent Totals and the Rst Solvent tab only appear if
“Meter” is selected under Solvent Monitor in Configure
Screen 5 on page 31.
Reset Solvent Screen
FIG. 13. Reset Solvent Total Screen
The screen will ask if you want to reset solvent total.
Select Yes or No.
Alarms Screens
NOTE: Grand totals are not resettable.
Reset Total Screen
F
IG. 12. Reset Total Screen
If job is reset, job number will increment by one for
default.
F
IG. 14. Alarms Screen
Two screens show the last 10 alarms. Use the Up or
Down keys to scroll between the two screens.
See Table 9 on page 77 for a list of alarm codes.
313882C23
Page 24
Run Mode Screens
Level Control Screen
F
IG. 15. Level Control Screen
This screen shows the current volume for each fluid.
Adjust the actual volumes on this screen, or use the tab
to go to Usage (Totals Screen, page 23).
See F
IG. 16. If the tank volume reaches the low-level
threshold, the EasyKey screen will display the Tank
Level Low alarm and prompt the user to do one of the
following:
1. Refill tank volume to clear the alarm.
2. Resume mixing by selecting “Spray 25% of Remainder.” If this selection is chosen, a second alarm will
occur after 25% of the remaining volume is mixed.
Refill tank volume to clear the alarm.
IG. 16. Tank Level Low Screen (Tank A Shown)
F
24313882C
Page 25
Setup Mode
Setup Mode
Press the Setup key to enter Setup mode.
Press the Setup key to
enter Setup mode.
NOTE: See F
IG. 17 for a map of the Setup screens.
Detailed screen descriptions follow.
This screen appears only if a
password is activated.
To access Advanced Setup Screens, page
34 and Recipe Setup Screens, page 38.
To access System Configuration Screens,
page 28.
Press the Setup key to exit
Setup mode and return to the Status
FIG. 17. Setup Screens Map
screen.
This screen appears momentarily
if a password is activated.
TI12784a
313882C25
Page 26
Setup Mode
Password Screen
If a password has been activated (see Configure
Screen 1, page 29), the Password screen will appear. You must enter the password to access the Set Up
Home Screen. Entering the wrong password returns the display to the Status Screen.
NOTE: If you forget the password, you can reset the
password (to 0), using the ProMix Web Interface (see
manual 313386).
F
IG. 18. Password Screen
NOTE: If a password is activated, Setup Locked dis-
plays momentarily after exiting Setup mode and returning to the Status Screen. A lock symbol appears
Set Up Home Screen
FIG. 20. Set Up Home Screen
This screen displays when you enter Setup mode. From
it you can go to Recipe and Advanced Setup Screens
(pages 34-41) or System Configuration Screens
(pages 28-31). Press the Enter key to go to the
selected screen set.
The screen also displays software versions and internet
addresses of various components. The values shown in
F
IG. 20 are only examples and may vary on your screen.
See Table 4 for further information.
on the Status Screen.
IG. 19. Setup Locked Screen
F
26313882C
Page 27
Setup Mode
Table 4: Component Software Versions
Display (may vary
from examples
Component
shown)Description
EK (EasyKey)2.00.012EasyKey software version.
FP (Fluid Plate)2.00.012Fluid Plate software version.
BC (Booth Control)-.-Booth Control not installed, not detected, or not oper-
ational.
1.XXBooth Control software version 1.00 or 1.01.
2.XXBooth Control software version 2.XX.
C1/C2 (Color Change
Modules 1 and 2)
-.-Color Change Module 1/2 not installed, not detected,
or not operational.
1.XXColor Change Module software version 1.00 or 1.01.
2.XXColor Change Module software version 2.XX.
AK (Autokey)2K-ManualNo AutoKey installed or detected. System operates in
2K Manual Mode only
2K-Auto2K AutoKey detected. System can operate in 2K Man-
ual, Semi-automatic, or Automatic Mode.
3K-Auto3K AutoKey detected. System can operate in 3K Man-
ual, Semi-automatic, or Automatic Mode.
XP (XPORT)V6.6.0.2Example of XPORT network module software version.
Other versions are acceptable.
MC (Micro Controller)1042.0198Example of fluid plate micro controller version. Other
versions are acceptable.
IP (Internet Address)192.168.178.5Example of the address EasyKey is set to for basic
and advanced web interface reporting.
MAC (MAC address)00204AAD1810Example of internet MAC address. Each EasyKey will
have a different value in this format.
313882C27
Page 28
Setup Mode
System Configuration Screens
NOTE: See FIG. 21 for a map of the System Configura-
tion Screens. Detailed screen descriptions follow.
NOTE: Each screen displays the current screen number
and the total number of screens in the group.
TI12785a
F
IG. 21. System Configuration and Option Screens Map
28313882C
Page 29
Setup Mode
Configure Screen 1
F
IG. 22. Configure Screen 1
Language
Defines the language of the screen text. Select English
(default), Spanish, French, German, Italian, Dutch, Japanese (Kanji), Korean, and Chinese (Simplified).
Password
The password is only used to enter Setup mode. The
default is 0, which means no password is required to
enter Setup. If a password is desired, enter a number
from 1 to 9999.
Configure Screen 2
FIG. 23. Configure Screen 2
Month
Enter current month.
Day
Enter current day.
Year
Enter current year (four digits).
Time
NOTE: Be sure to write down the password and keep it
in a secure location.
Display Units
Select the desired display units:
•cc/liter (default)
•cc/gallon
Enter current time in hours (24 hour clock) and minutes.
Seconds are not adjustable.
Date Format
Select mm/dd/yy or dd/mm/yy.
313882C29
Page 30
Setup Mode
Configure Screen 3
F
IG. 24. Configure Screen 3
Run Mode
Indicates that this is a Manual system.
Dump Valve A
This field only appears if the system includes an optional
dump valve A. If dump valve A is included, set to On.
Dump Valve B
This field only appears if the catalyst change option is
detected from the cc board, meaning that dump valve B
is present. On is the only setting.
Dump Valve C
Dose Size
Select the total dose size (cc) from the pulldown menu:
100, 50, 25, or 10.
NOTE: Selection DD is for dynamic dosing, which is not
used with ProMix 3KS.
FIG. 26. Dose Size Menu
Example:
To calculate the dose size of each component, you must
first calculate the A+B:C dose sizes. For example, if the
total dose size is 50 cc and the A+B:C ratio is 1:1, the
dose size of A+B is 25 cc and the dose size of C is also
25 cc. Then use the ratio of A:B (2:1 in this example) to
calculate the dose size of A as 16.67 cc (2/3 of 25) and
the dose size of B as 8.33 cc (1/3 of 25).
This field only appears if the component C change
option is detected from the cc board, meaning that dump
valve C is present. On is the only setting.
3rd Flush Valve
Off is default. If optional 3rd flush valve is used, set to
On.
Configure Screen 4
IG. 25. Configure Screen 4
F
Dose Time Alarm
Enter the dose time (1 to 99 seconds). This is the
amount of time allowed for a dose to occur before a
dose time alarm occurs.
Number of Guns
Enter the number of spray guns (1 or 2).
Gun Flush Box
Enter the number of gun flush boxes (Off, 1, or 2).
30313882C
Page 31
Configure Screen 5
F
IG. 27. Configure Screen 5
Special Outputs
Select special outputs (0-4). Each output has two different start times and durations.
Setup Mode
Solvent Monitor
Select solvent monitor (Off, Flow Switch, or Meter).
Web Browser IP
The default web browser IP address prefix is
192.168.178.__ Assign a unique number for each
EasyKey in your system (1-99) and enter it here.
Control Network ID
Used for the Graco Gateway network system. See
Graco Gateway manual 312785 for further information.
313882C31
Page 32
Setup Mode
Option Screens
NOTE: See FIG. 21 on page 28 for a map of the Option
Screens. Detailed screen descriptions follow.
NOTE: Each screen displays the current screen number
and the total number of screens in the group.
Option Screen 1
F
IG. 28. Option Screen 1
Flush Volume Check
This field only appears if Solvent Monitor is set to
“Meter” in Configure Screen 5, page 31.
If set to “On”, Minimum Flush Volume will appear in Rec-
ipe Setup Screen 2, page 39.
Stage Fill Volume
This field refers to the amount of material that is
required to fill the mixed material line from the A/B fluid
panel to the C fluid panel, before adding component C.
This value must be the same as the selected dose size.
Default is 50 cc.
Verification Screen
FIG. 29. Verification Screen
Verification
This screen appears if Flush and Fill Input or K-Factor
Input are changed from “Recipe” to “Global” in Option
Screen 1.
Flush and Fill Input
If set to “Global”, Color/Catalyst Purge and Color/Catalyst Fill are added to Advanced Setup Screen 1, page
35. Advanced Setup Screen 2, 3, and 5 are added.
See pages 35-37.
If set to “Recipe”, Color/Catalyst Purge and Color/Catalyst Fill are added to Recipe Setup Screen 2, page 39.
Recipe Setup Screen 3, 4, and 7 are added. See
pages 40-41.
K-Factor Input
If set to “Global,” Advanced Setup Screen 4, page 36 is
added.
If set to “Recipe,” Recipe Setup Screen 5, page 41, is
added.
Minimum Material Fill Volume
Enter 0-9999 cc.
32313882C
Page 33
Option Screen 2
F
IG. 30. Option Screen 2
Auto Dump
If the auto dump feature is being used, set to “On”. Once
the auto dump is enabled, the gun flush box is enabled
and the potlife alarm is active for 2 minutes, the system
will automatically flush out the old material.
Setup Mode
Flow Rate Monitor
If set to “On,” Recipe Setup Screen 6 on page 41 is
added, enabling setting of high and low flow limits.
If set to “Off,” flow rate monitoring is disabled and Rec-ipe Setup Screen 6 on page 41 will not appear.
313882C33
Page 34
Setup Mode
Advanced Setup Screens
NOTE: See FIG. 31 for a map of the Advanced Setup
Screens. Detailed screen descriptions follow.
Advanced Setup screens 2, 3,
4, and 5 appear depending on
selections made in Option
screens 1 and 2.
TI12786a
F
IG. 31. Advanced Setup Screens Map
34313882C
Page 35
Setup Mode
NOTE: Each screen displays the current screen number
and the total number of screens in the group. The total
number of screens in a group and the fields displayed
on each screen may vary depending on selections
made in the System Configuration Screens and
Option Screens.
Advanced Setup Screen 1
F
IG. 32. Advanced Setup Screen 1
Gun 1/Gun2 Potlife Volume
Color/Catalyst Fill
This field only appears if the system includes a color
change module and Flush and Fill Input is set to
“Global” in Option Screen 1, page 32. Enter the fill time
(0 to 99 seconds). It refers to the time required to fill the
lines from the color/catalyst/component C module to the
dose valve or dump valve.
Advanced Setup Screen 2
FIG. 33. Advanced Setup Screen 2
Enter the potlife volume (1 to 1999 cc) for each gun.
This is the amount of material required to move through
the mix manifold, hose and applicator/gun before the
potlife timer is reset.
Use the following information to determine approximate
pot life volume (PLV) in cc:
Hose ID (inches)Volume (cc/foot)*
3/165.43
1/49.648
3/821.71
Integrator manifold and mixer volume = 75 cc
Spray Gun Volume = 20 cc
(Hose Volume* x Feet of Hose) + 75 + 20 = PLV
Color/Catalyst Purge
This field only appears if the system includes a color
change module and Flush and Fill Input is set to
“Global” in Option Screen 1, page 32. Enter the purge
time (0 to 99 seconds). It refers to the amount of time
required to flush the lines from the color/catalyst/component C module to the dose valve or dump valve.
This screen appears only if Flush and Fill Input is set to
“Global” in Option Screen 1, page 32.
First Purge Source
Select “Air,” “Solvent,” or “3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure
Screen 3 on page 30).
Chop Type
Select “Air/Solvent” or “Air/3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure Screen 3 on page 30). This refers to the process of mixing air and solvent (or air and 3rd flush fluid) together
during the flush cycle, to help clean the lines and reduce
solvent usage.
Final Purge Source
Select “Air,” “Solvent,” or “3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure
Screen 3 on page 30).
Air Chop Time
Enter the air chop time (0.0 to 99.9 seconds).
Solvent Chop Time/3rd Flush Valve Chop Time
Enter the solvent or 3rd flush valve chop time (0.0 to
99.9 seconds).
313882C35
Page 36
Setup Mode
Advanced Setup Screen 3
F
IG. 34. Advanced Setup Screen 3
This screen appears only if Flush and Fill Input is set to
“Global” in Option Screen 1, page 32.
If Number of Guns is set to “2” in Configure Screen 4,
page 30, a Gun 2 column will appear in this screen.
First Purge Time (Stage 1)
Enter the first purge time (0 to 999 seconds) for components A and B.
Advanced Setup Screen 4
FIG. 35. Advanced Setup Screen 4
This screen appears only if K-Factor Input is set to
“Global” in Option Screen 1, page 32.
K-factor A Meter
Enter the k-factor (cc/pulse) for flow meter A. This is the
amount of material that passes through the flow meter
per pulse (electrical pulse signal).
K-factor B Meter
C First Purge Time (Stage 2)
Enter the component C first purge time (0 to 999 seconds).
Total Chop Time (Stage 1)
Enter the total chop time (0 to 999 seconds) for components A and B.
Final Purge Time (Stage 1)
Enter the final purge time (0 to 999 seconds) for components A and B.
C Final Purge Time (Stage 2)
Enter the component C final purge time (0 to 999 seconds).
Mixed Fill Time
Enter the mixed fill time (0 to 999 seconds). It refers to
the amount of material that is required to fill from the
dose valves to the applicator/gun.
Enter the k-factor (cc/pulse) for flow meter B.
K-factor C Meter
Enter the k-factor (cc/pulse) for flow meter C.
K-factor Solvent Meter
This field only appears if Solvent Monitor in Configure
Screen 5, page 31, is set to “Meter.” Enter the k-factor
(cc/pulse) for the solvent flow meter.
36313882C
Page 37
Advanced Setup Screen 5
F
IG. 36. Advanced Setup Screen 5
This screen appears only if Flush and Fill Input is set to
“Global” in Option Screen 1, page 32 and Special Outputs is set to 1, 2, 3, or 4 in Configure Screen 5, page
31. The I/O board has four programmable outputs.
Setup Mode
313882C37
Page 38
Setup Mode
Recipe Setup Screens
NOTE: See FIG. 37 for a map of the Recipe screens. Detailed screen descriptions follow.
Recipe 0 Screens
Recipe screens 3, 4, 5, 6, and
7 appear depending on
selections made in Option
screens 1 and 2
TI12787a
F
IG. 37: Recipe Screens Map
38313882C
Page 39
Setup Mode
NOTE: Each screen displays the current screen number
and the total number of screens in the group. The total
number of screens in a group and the fields displayed
on each screen may vary depending on selections
made in the System Configuration Screens and
Option Screens.
Recipe Setup Screen 1
F
IG. 38. Recipe Setup Screen 1
Ratio
Enter the mix ratio of component A over component B
(0.0:1 to 50:1) and the ratio of A+B over C (0.0:1 to
50:1).
Ratio Tolerance
Enter the ratio tolerance (1 to 99%). This refers to the
percent of acceptable variance that the system will allow
before a ratio alarm occurs.
Recipe Setup Screen 2
FIG. 39. Recipe Setup Screen 2
Minimum Flush Volume
This field only appears if Flush Volume Check is set to
“On” in Option Screen 1 on page 32. Enter the minimum flush volume (0 to 999 cc). Entering 0 disables this
function.
Potlife Time
Enter the potlife time (0 to 999 minutes). Entering 0 disables this function.
Color/Catalyst Purge
This field only appears if the system includes a color
change module and Flush and Fill Input is set to “Recipe” in Option Screen 1, page 32. Enter the purge time
(0 to 99 seconds). It refers to the amount of time
required to flush the lines from the color, catalyst, or
component C module to the dose valve or dump valve.
Component A (Color) Valve (if present)
This field only appears if the system includes a color
change module. Enter the component A valve number
(1 to 25).
Component B (Catalyst) Valve (if present)
This field only appears if the system includes a color
change module. Enter the component B valve number
(1 to 4).
Component C (Reducer) Valve (if present)
This field only appears if the system includes a color
change module. Enter the component C valve number
(1 to 4).
313882C39
Color/Catalyst Fill
This field only appears if the system includes a color
change module and Flush and Fill Input is set to “Recipe” in Option Screen 1, page 32. Enter the fill time (0
to 99 seconds). It refers to the time required to fill the
lines from the color, catalyst, or component C module to
the dose valve or dump valve.
Page 40
Setup Mode
Recipe Setup Screen 3
F
IG. 40. Recipe Setup Screen 3
This screen appears only if Flush and Fill Input is set to
“Recipe” in Option Screen 1, page 32.
First Purge Source
Select “Air,” “Solvent,” or “3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure
Screen 3 on page 30).
Chop Type
Recipe Setup Screen 4
FIG. 41. Recipe Setup Screen 4
This screen appears only if Flush and Fill Input is set to
“Recipe” in Option Screen 1, page 32.
If Number of Guns is set to “2” in Configure Screen 4,
page 30, a Gun 2 column will appear in this screen.
First Purge Time (Stage 1)
Enter the first purge time (0 to 999 seconds) for components A and B.
Select “Air/Solvent” or “Air/3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure Screen 3 on page 30). This refers to the process of mixing air and solvent (or air and 3rd flush fluid) together
during the flush cycle, to help clean the lines and reduce
solvent usage.
Final Purge Source
Select “Air,” “Solvent,” or “3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure
Screen 3 on page 30.)
Air Chop Time
Enter the air chop time (0.0 to 99.9 seconds).
Solvent Chop Time/3rd Flush Valve Chop Time
Enter the solvent or 3rd flush valve chop time (0.0 to
99.9 seconds).
C First Purge Time (Stage 2)
Enter the component C first purge time (0 to 999 seconds).
Total Chop Time (Stage 1)
Enter the total chop time (0 to 999 seconds) for components A and B.
Final Purge Time (Stage 1)
Enter the final purge time (0 to 999 seconds) for components A and B.
C Final Purge Time (Stage 2)
Enter the component C final purge time (0 to 999 seconds).
Mixed Material Fill Time
Enter the mixed material fill time (0 to 999 seconds). It
refers to the amount of material that is required to fill
from the dose valves to the applicator/gun.
40313882C
Page 41
Setup Mode
Recipe Setup Screen 5
F
IG. 42. Recipe Setup Screen 5
This screen appears only if K-Factor Input is set to “Recipe” in Option Screen 1, page 32.
K-factor A Meter
Enter the k-factor (cc/pulse) for flow meter A. This is the
amount of material that passes through the flow meter
per pulse (electrical pulse signal).
Recipe Setup Screen 6
FIG. 43. Recipe Setup Screen 6
This screen appears only if Flow Rate Monitor is set to
“On” in Option Screen 2 on page 33.
Flow Rate Monitor
Select the desired flow rate monitoring (Off, Warning, or
Alarm).
K-factor B Meter
Enter the k-factor (cc/pulse) for flow meter B.
K-factor C Meter
Enter the k-factor (cc/pulse) for flow meter C.
K-factor Solvent Meter
This field only appears if Solvent Monitor in Configure
Screen 5, page 31, is set to “Meter.” Enter the k-factor
(cc/pulse) for the solvent flow meter.
Low Flow Limit
Enter the low flow rate limit (1 to 3999 cc/min).
High Flow Limit
Enter the high flow rate limit (1 to 3999 cc/min).
Recipe Setup Screen 7
F
IG. 44. Recipe Screen 7
This screen appears only if Flush and Fill Input is set to
“Recipe” in Option Screen 1, page 32 and Special Outputs is set to 1, 2, 3, or 4 in Configure Screen 5, page
31. The I/O board has four programmable outputs.
313882C41
Page 42
Setup Mode
Recipe 0 Screens
NOTE: See FIG. 37 on page 38 for a map of the Recipe
0 screens. Detailed screen descriptions follow.
Recipe 0 is typically used:
•in multiple color systems to purge out material lines
without loading a new color
•at the end of a shift to prevent hardening of catalyzed material.
NOTE: Each screen displays the current screen number
and the total number of screens in the group. The total
number of screens in a group and the fields displayed
on each screen may vary depending on selections
made in the System Configuration Screens and
Option Screens.
Recipe 0 Screen 1
Air Chop Time
Enter the air chop time (0.0 to 99.9 seconds).
Solvent Chop Time/3rd Flush Valve Chop Time
Enter the solvent or 3rd flush valve chop time (0.0 to
99.9 seconds).
Recipe 0 Screen 2
FIG. 46. Recipe 0 Screen 2
F
IG. 45. Recipe 0 Screen 1
First Purge Source
Select “Air,” “Solvent,” or “3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure
Screen 3 on page 30).
Chop Type
Select “Air/Solvent” or “Air/3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure Screen 3 on page 30). This refers to the process of mixing air and solvent (or air and 3rd flush fluid) together
during the flush cycle, to help clean the lines and reduce
solvent usage.
Final Purge Source
Select “Air,” “Solvent,” or “3rd Flush Valve” (available
only if 3rd Flush Valve is set to “On” in Configure Screen 3 on page 30.)
If Number of Guns is set to “2” in Configure Screen 4,
page 30, a Gun 2 column will appear in this screen.
Color/Catalyst Purge Time
This field only appears if the system includes a color
change module. Enter the purge time (0 to 999 seconds).
First Purge Time (Stage 1)
Enter the first purge time (0 to 999 seconds) for components A and B.
C First Purge Time (Stage 2)
Enter the component C first purge time (0 to 999 seconds).
Total Chop Time
Enter the total chop time (0 to 999 seconds) for components A and B.
Final Purge Time (Stage 1)
Enter the final purge time (0 to 999 seconds) for components A and B.
C Final Purge Time (Stage 2)
Enter the component C final purge time (0 to 999 seconds).
42313882C
Page 43
Setup Mode
Recipe 0 Screen 3
F
IG. 47. Recipe 0 Screen 3
This screen only appears if Solvent Monitor is set to
“Meter” in Configure Screen 5, page 31 and Flush Volume Check is set to “On” in Option Screen 1, page 32
or 3rd Flush Valve is set to “On” in Configure Screen 3
on page 30.
Minimum Flush Volume
This field only appears if Flush Volume Check is set to
“On” in Option Screen 1 on page 32. Enter the minimum flush volume (0 to 999 cc).
Recipe 0 Screen 4
FIG. 48. Recipe 0 Screen 4
This screen appears only if Flush and Fill Input is set to
“Recipe” in Option Screen 1, page 32 and Special Outputs is set to 1, 2, 3, or 4 in Configure Screen 5, page
31. The I/O board has four programmable outputs.
Exiting Fill Source
This field only appears if 3rd Flush Valve is set to “On” in
Configure Screen 3 on page 30. Select “Off,” “Air,” “Solvent,” or “3rd Valve.”
Exiting Fill Time
This field only appears if Exiting Fill Source is set to
“Air,” “Solvent,” or “3rd Valve.” Enter the time in seconds.
313882C43
Page 44
Setup Mode
Calibration Screen
F
IG. 49. Calibration Screen
Use this screen to calibrate a meter. Set to “Meter A,”
“Meter B,” “Meter C,” or “Solvent Meter” (available if Solvent Monitor in Configure Screen 5, page 31, is set to
“Meter”).
•Start - start calibration
•Abort - stop calibration
•Purge - purge sampling valves after calibration
See Meter Calibration, page 62, for when and how to
calibrate a meter.
44313882C
Page 45
System Operation
System Operation
Operation Modes
Mix
System mixes and dispenses material.
Standby
Stops the system.
Purge
Purges the system, using air and solvent.
Sequential Dosing
Components A, B, and C dispense sequentially in the
necessary volumes to attain the mix ratio.
Recipe (Color) Change
The process when the system automatically flushes out
the old color and loads a new color. See pages 64-76.
General Operating Cycle,
Sequential Dosing
1. The spray gun operator enters and loads the
desired recipe. The color change LED blinks while
recipe is loading, then turns solid when complete.
2. The operator presses the Mix key to begin
operation.
3. The controller sends signals to activate the solenoid
valves. The solenoid valves activate Dose Valves A,
B, and C. Fluid flow begins when the gun is triggered.
Stage 1 (see FIG. 50, ProMix 2KS Detail)
4. Components A and B are introduced into the 2KS
fluid integrator (FI) one at a time as follows.
a. Dose Valve B (DVB) opens, and fluid flows into
the integrator.
b. Flow Meter B (MB) monitors the fluid volume
dispensed and sends electrical pulses to the
ProMix 2KS controller. The controller monitors
these pulses and signals.
c. When the target volume dispenses, Dose Valve
B closes.
NOTE: The dispense volume of component A and B
is based on the mix ratio and dose size set by the
user and calculated by the ProMix 2KS controller.
d. Dose Valve A (DVA) opens, and fluid flows into
the 2KS integrator and is aligned proportionately with component B.
e. Flow Meter A (MA) monitors the fluid volume
dispensed and sends electrical pulses to the
ProMix 2KS controller.
f.When the target volume is dispensed, Dose
Valve A closes.
313882C45
Page 46
System Operation
5. Components A and B are pre-mixed in the 2KS integrator, then uniformly blended in the 2KS static
mixer (SM) before flowing through the hose to the
3KS Fluid Manifold inlet.
Stage 2 (see FIG. 50, ProMix 3KS Detail)
6. Dose Valve C (DVC) opens, and fluid flows into the
3KS integrator and is aligned proportionately with
components A+B (mixed in Stage 1).
7. Flow Meter C (MC) monitors the fluid volume dispensed and sends electrical pulses to the ProMix
3KS controller.
8. When the target volume is dispensed, Dose Valve C
closes.
9. Components A+B and C are pre-mixed in the 3KS
integrator, then uniformly blended in the 3KS static
mixer (SM).
Table 5: Sequential Dosing Operation
Dose 1Dose 2
NOTE: To control output from the 3KS static mixer
to the gun, install an optional fluid pressure regulator.
10. Components A+B and C are alternately fed into the
2KS integrator as long as the gun is triggered.
11. If the gun is not triggered for two minutes, the system switches to Idle mode, which closes off the mix
manifold dose valves.
12. When the gun is triggered again, the ProMix 3KS
continues the process where it left off.
NOTE: Operation can be stopped at any time by
pressing the Standby key or shutting off the
main power switch.
Ratio = 2.0:1
A = 2
B = 1
C = 1
Stage 1 (A:B)
Stage 2
(A+B):CStage 1 (A:B)
Stage 2
(A+B):C
46313882C
Page 47
System Operation
MA
RVA
APV
FI
AT
DVC
DVA
SVASVB
FI
DVB
SM
Key: ProMix 2KS Fluid Station
MAComponent A Meter
DVA Component A Dose Valve
RVA Component A Sampling
Val ve
SVA Component A Shutoff
Val ve
MBComponent B Meter
DVB Component B Dose Valve
RVB Component B Sampling
Val ve
MB
MS
RVB
TI12556a
SPV
SVB Component B Shutoff
Val ve
MSSolvent Meter
SPV Solvent Purge Valve
APV Air Purge Valve
SMStatic Mixer
FIFluid Integrator
ATAir Purge Valve Air Supply
Tu b e
Key: ProMix 3KS Fluid Station
MC
CPV
RVC
1
SVC
MCComponent C Meter
DVC Component C Dose Valve
RVC Component C Sampling Valve
SVC Component C Shutoff Valve
CPV Component C Purge Valve
SMStatic Mixer
FIFluid Integrator
1
3KS fluid inlet. Connect fluid supply line from
2KS fluid manifold outlet here.
2
Connect fluid supply line to gun.
SM
2
TI14382a
FIG. 50. ProMix 2KS and ProMix 3KS Wall Mount Fluid Stations
313882C47
Page 48
System Operation
Mix Manifold Valve Settings
To open dose or purge valves, turn hex nut (E) counterclockwise. To close, turn clockwise. See Table 6 and
F
IG. 51.
E
TI11581a
F
IG. 51. Valve Adjustment
Table 6: Mix Manifold Valve Settings
ValveSettingFunction
Dose
(F
IG. 51)
Hex nut (E) 1-1/4
turns out from fully
Limits maximum fluid flow rate into integrator and
minimizes valve response time.
closed
Purge
IG. 51)
(F
Hex nut (E) 1-1/4
turns out from fully
Limits maximum fluid flow rate into integrator and
minimizes valve response time.
closed
Shutoff
(SVA and
SVB, F
IG. 60)
Sampling
(RVA, RVB,
and RVC,
F
IG. 60)
Fully open during
Run/Mix operation
Fully closed during
Run/Mix operation
Closes component A, B, and C ports to integrator
during ratio check or meter calibration. Open ports
during Run/Mix operation.
Open to dispense component A, B, and C while calibrating meters. Do not open sampling valves unless
fluid shutoff valves are closed.
48313882C
Page 49
System Operation
Air Flow Switch (AFS) Function
Air or Air-assisted Guns
The air flow switch (AFS) detects air flow to the gun and
signals the ProMix controller when the gun is triggered.
The AFS functions with the flow meters to ensure that
system components are functioning correctly.
For example, if a flow meter fails or clogs, pure resin or
catalyst could spray indefinitely if the ProMix does not
detect the condition and intervene, which is why the
AFS is so important.
If the ProMix detects through the AFS signal that the
gun is triggered, yet there is no fluid flow through the
meter, a Dose Time Alarm (E-7 or E-8) occurs after 40
seconds and the system shuts down.
Operating Without Air Flow Switch
It is not recommended to run without an air flow switch.
If a switch fails, replace it as soon as possible.
Airless Gun
It is not recommended to use an airless gun with the
ProMix 3KS. Two issues can arise from operating without an air flow switch:
•Without a gun trigger/air flow switch input the ProMix 3KS does not know it is spraying and will not
generate a Dose Time Alarm (E-7 or E-8). This
means there is no way to detect a failed meter. You
could spray pure resin or catalyst for 2 minutes with-
out knowing.
•Since the ProMix 3KS does not know it is spraying
because there is no gun trigger/air flow switch input,
it will go into System Idle (E-15) every 2 minutes
when in Mix mode.
System Idle Warning (E-15)
F
IG. 52: Air Flow Switches
TI13350a
AFS
This warning occurs if the ProMix is set to Mix ,
and 2 minutes have elapsed since the system received
a flow meter pulse.
In applications using the AFS, triggering the gun clears
the warning and you can start spraying again.
Without the AFS, triggering the gun does not clear the
alarm. To start spraying again, you must press Standby
, then Mix , then trigger the gun.
313882C49
Page 50
System Operation
Start Up
1. Go through the Pre-Operation Checklist in Table 7.
Table 7: Pre-Operation Checklist
✓Checklist
System grounded
Verify all grounding connections were made. See
the ProMix 3KS Installation manual.
All connections tight and correct
Verify all electrical, fluid, air, and system connections are tight and installed according to the
Installation manual.
Fluid supply containers filled
Check component A, B, and C and solvent supply
containers.
Mix manifold valves set
Check that mix manifold valves are set correctly.
Start with the settings recommended in Mix Mani-fold Valve Settings, page 48, then adjust as
needed.
Fluid supply valves open and pressure set
Component A, B, and C fluid supply pressures
should be equal unless one component is more
viscous and requires a higher pressure setting.
Solenoid pressure set
➜ Graco logo, software revision, and “Establishing
Communication” will display, followed by Status
screen. See page 20.
➜ At power up the system defaults to Recipe 61,
which is not a valid recipe number. Initiate a
color change to Recipe 0 or a valid recipe number (1-60).
➜ In bottom left corner, the system status displays,
which can be Standby, Mix, Purge, or an alarm
notification.
FIG. 54. Status Screen
3. Make sure that the Booth Control is working. The
active recipe number should display and the
75-100 psi inlet air supply (0.5-0.7 MPa, 5.2-7 bar)
2. Turn the AC Power Switch on the EasyKey and on
the Power Supply Module ON (I = ON, 0 = OFF).
I = ON
TI12656aTI14372a
F
IG. 53. Power Switch
Standby LED
should be lit.
4. If this is the first time starting up the system, purge it
as instructed in Purging Fluid Supply System,
page 59. The equipment was tested with lightweight
oil, which should be flushed out to avoid contaminating your material.
5. Make sure that the Booth Control is in Standby
mode.
50313882C
Page 51
System Operation
6. Adjust component A, B, and C fluid supplies as needed for your application. Use
lowest pressure possible.
7. Do not exceed the maximum rated working pressure shown on the system identification
label or the lowest rated component in the system.
8. Open the fluid supply valves to the
system.
9. Adjust the air pressure. Most applications require about 80 psi (552
kPa, 5.5 bar) air pressure to operate properly. Do
not use less than 75 psi (517 kPa, 5.2 bar).
10. If using a gun flush box, place the gun into the box
and close the lid. Press the Purge key on the
Booth Control. The purge sequence automatically
starts.
If the gun flush box is not used, trigger the gun into a grounded metal
pail until the purge sequence is complete.
When done purging, the Booth Control automatically switches to Standby mode.
11. Adjust the flow rate.
The fluid flow rate shown on the EasyKey Status
screen is for either component A, B, and C, depending on which dose valve is open. The fluid supply
lines on the screen highlight to show which dose
valve is open.
Watch the fluid flow rate displayed on the Status
screen while the gun is fully open. Verify that the
flow rate of components A, B, and C are within 10%
of each other.
If the fluid flow rate is too low: increase air pressure to component A, B, and C fluid supplies or
increase the regulated fluid pressure.
If the fluid flow rate is too high: reduce the air
pressure, close the fluid manifold dose valves further, or adjust the fluid pressure regulator.
NOTE: Pressure adjustments of each component will
vary with fluid viscosity. Start with the same fluid pressure for component A, B, and C, then adjust as needed.
NOTICE
Do not use the first 4-5 oz. (120-150 cc) of material as
it may not be thoroughly mixed due to alarms while
priming the system.
12. Turn on atomizing air to the gun. Check the spray
pattern as instructed in your spray gun manual.
NOTICE
Do not allow a fluid supply tank to run empty. It is possible for air flow in the supply line to turn gear meters
in the same manner as fluid. This can lead to the proportioning of fluid and air that meets the ratio and tolerance settings of the equipment. This can further
result in spraying uncatalyzed or poorly catalyzed
material.
F
IG. 55. Status Screen Flow Rate Display
313882C51
Page 52
System Operation
Shutdown
Overnight Shutdown
1. Leave the power on.
2. Run Recipe 0 to purge solvent through meters and
gun.
Service Shutdown
1. Follow Pressure Relief Procedure on page 52.
2. Close main air shutoff valve on air supply line and
on ProMix.
3. Shut OFF the Power Switch on the EasyKey and on
the Power Supply Module (0 position). F
4. If servicing the EasyKey or Power Supply Module,
also shut off power at main circuit breaker.
0 = OFF
IG. 56.
Pressure Relief Procedure
NOTE: The following procedures relieve all fluid and air
pressure in the ProMix 3KS system. Use the procedure
appropriate for your system configuration.
Relieve pressure when you stop spraying, before
changing spray tips, and before cleaning, checking, or
servicing equipment.
Single Color Systems
1. While in Mix mode (gun triggered), shut off the A, B,
and C fluid supply pumps/pressure pots. Close all
fluid shutoff valves at the pump outlets.
2. With the gun triggered, push the manual override on
the A, B, and C dose valve solenoids to relieve pressure. See F
NOTE: If a Dose Time alarm (E-7, E-8) occurs, clear
the alarm.
IG. 57.
TI12657aTI14372a
F
IG. 56. Power Switch
3. Do a complete system purge, following the instructions under Purging Using Recipe 0, page 59.
4. Shut off the fluid supply to the solvent purge valve
(SPV) and the air supply to the air purge valve
(APV), F
5. With the gun triggered, push the manual override on
the A, B, and C purge valve solenoids to relieve air
and solvent pressure. See F
vent pressure is reduced to 0.
NOTE: If a Purge Volume alarm (E-11) occurs, clear
the alarm.
IG. 60.
IG. 57. Verify that sol-
52313882C
Page 53
System Operation
Systems with Color Change and without
Dump Valves
NOTE: This procedure relieves pressure through the
sampling valve.
1. Complete all steps under Single Color Systems,
page 52.
2. Close the A side shutoff valve (SVA), F
the A side sampling valve (RVA).
3. Direct the A side sampling tube into a waste container.
4. See F
IG. 59. Open the color change module. Using
the solenoid identification labels as a guide, press
and hold the override button on each color solenoid
until flow from the sampling valve stops.
5. Press and hold the solvent solenoid override until
clean solvent comes from the sampling valve, then
release.
6. Shutoff the solvent supply to the color change stack
solvent valve.
IG. 60. Open
6. See F
IG. 59. Using the solenoid identification labels
as a guide, press and hold the override button on
each catalyst solenoid until flow from dump valve B
stops.
7. Press and hold the dump valve C solenoid override,
F
IG. 57.
8. See F
IG. 59. Open the color change module. Using
the solenoid identification labels as a guide, press
and hold the override button on each color solenoid
until flow from dump valve C stops.
9. Press and hold the dump valve A solenoid override,
F
IG. 57.
10. Press and hold the A side (color) solvent solenoid
override until clean solvent comes from the dump
valve, then release.
11. Press and hold the dump valve B solenoid override,
F
IG. 57.
12. Press and hold the B side (catalyst) solvent solenoid
override until clean solvent comes from the dump
valve, then release.
7. Press and hold the solvent solenoid override until
solvent flow from the sampling valve stops.
8. Open the A side shutoff valve (SVA), F
IG. 60. Close
the A side sampling valve (RVA).
Systems with Color/Catalyst/Component C
Change and Dump Valves
NOTE: This procedure relieves pressure through the
dump valves.
1. Complete all steps under Single Color Systems,
page 52.
2. Shut off all color/catalyst/component C supplies to
the valve stacks.
3. Press and hold the dump valve A solenoid override,
F
IG. 57.
4. See F
IG. 59. Open the color change module. Using
the solenoid identification labels as a guide, press
and hold the override button on each color solenoid
until flow from dump valve A stops.
13. Press and hold the dump valve C solenoid override,
F
IG. 57.
14. Press and hold the C side solvent solenoid override
until clean solvent comes from the dump valve, then
release.
15. Shutoff the solvent supply to the color/catalyst/component C change stack solvent valves.
16. Press and hold the A, B, and C solvent solenoid
overrides and dump valve overrides until solvent
flow from the dump valves stops.
5. Press and hold the dump valve B solenoid override,
F
IG. 57.
313882C53
Page 54
System Operation
Valve Overrides
AFS #1
AFS #2
SOLV SW
GFB 1 PS
GFB 2 PS
Not Used
Flow Solv
Flow C
Flow A
Flow B
Dose A
Dose B
Dose C
J9
1
J1
J13
+
1
G
s
J12
+
G
s
1
+
G
s
J3
+
G
s
1
+
+
+
Dump C
Air
1
Solvent
+++
Purge C
Dump A
3rd Flush
++++++
J14J15J8
Power
Fiber
Optic
CAN
CAN
Dump B
GFB 2
GFB 1
11
Dump Valve A
Dump Valve B
Optional Solenoid
Dose Valve A
Solenoid
Dose Valve B
Solenoid
Purge Valve A
Solenoid
Purge Valve B
Solenoid
GFB 1
GFB 2
3rd Purge Valve
Solenoid
Locations
TI12652a
FIG. 57. Component A and B Solenoids
54313882C
Page 55
J8
(Communication)
J3
(Power Input)
(Color Change
Module)
System Operation
Valve
Overrides
VDC
J2
J1
(Booth Control)
FIG. 58. Component C Solenoids
Dose Valve C
Solenoid
Purge Valve C
Solenoid
Dump Valve C
location (optional)
TI14704a
313882C55
Page 56
System Operation
Solenoid
Identification
Label
Module #1
ColorCatalyst
Solenoid
Identification
Label
Color
TI12826a
Solenoid
Identification
Label
Solvent Solenoid
Overrides
Module #2
ColorCatalyst
Solenoid
Identification
Label
Color
Mount Component C
Valves Here
Solvent Solenoid
Overrides
TI12826a
FIG. 59: Color Change Solenoids
56313882C
Page 57
MA
RVA
APV
DVA
FI
System Operation
DVB
MB
MS
RVB
TI12556a
FI
AT
DVC
SVASVB
SM
Key: ProMix 2KS Fluid Station
MAComponent A Meter
DVA Component A Dose Valve
RVA Component A Sampling
Val ve
SVA Component A Shutoff
Val ve
MBComponent B Meter
DVB Component B Dose Valve
RVB Component B Sampling
Val ve
Key: ProMix 3KS Fluid Station
MC
CPV
MCComponent C Meter
DVC Component C Dose Valve
RVC Component C Sampling Valve
SVC Component C Shutoff Valve
CPV Component C Purge Valve
SMStatic Mixer
FIFluid Integrator
SPV
SVB Component B Shutoff
Val ve
MSSolvent Meter
SPV Solvent Purge Valve
APV Air Purge Valve
SMStatic Mixer
FIFluid Integrator
ATAir Purge Valve Air Supply
Tu b e
1
3KS fluid inlet. Connect fluid supply line from
2KS fluid manifold outlet here.
RVC
1
SVC
2
Connect fluid supply line to gun.
SM
2
TI14382a
FIG. 60. ProMix 2KS and ProMix 3KS Wall Mount Fluid Stations
313882C57
Page 58
System Operation
Purging
Read Warnings, page 8. Follow the Grounding
instructions in your system Installation manual.
To avoid splashing fluid in the eyes, wear eye protection.
If using an electrostatic gun shut off the electrostatics
before flushing the gun.
2. Set the solvent supply pressure regulator at a pressure high enough to completely purge the system in
a reasonable amount of time but low enough to
avoid splashing or an injection injury. Generally, a
setting of 100 psi (0.7 MPa, 7 bar) is sufficient.
There are 4 purging procedures in this manual:
•Purging Mixed Material (below)
•Purging Using Recipe 0 (page 59)
•Purging Fluid Supply System (page 59)
•Purging Sampling Valves and Tubes (page 61)
Use the criteria listed in each procedure to determine
which procedure to use.
Purging Mixed Material
There are times when you only want to purge the fluid
manifold, such as:
•end of potlife
•breaks in spraying that exceed the potlife
•overnight shutdown
•before servicing the fluid manifold assembly, hose or
gun.
In the first stage fluid manifold (2KS), solvent purges the
component B (catalyst, right) side of the mix manifold
and the inner tube of the integrator. Air purges the component A (resin, left) side and the outer tube of the integrator. This air/solvent chop purges the hose connecting
the 2KS static mixer to the 3KS fluid manifold.
3. If using a gun flush box, place the gun into the box
and close the lid. Press the Purge key on the
Booth Control. The purge sequence automatically
starts.
If the gun flush box is not used, trigger the gun into a grounded metal
pail until the purge sequence is complete.
When done purging, the Booth Control automatically switches to Standby mode.
NOTE: After doing a purge, press the Mix
key once. The system will do a mixed material fill
sequence, then go to Standby. Press again to start
Mix mode.
4. If the system is not completely clean, repeat step 3.
NOTE: If necessary, adjust purge sequence so only
one cycle is required.
In the second stage fluid manifold (3KS), solvent purges
the component C (right) side of the mix manifold and the
inner tube of the integrator. The air/solvent chop from
the 2KS purges the left side of the 3KS fluid manifold
and the outer tube of the 3KS integrator.
1. Press the Standby
key on the Booth Control.
Trigger the gun to relieve pressure. Engage trigger
lock.
5. If spray tip was removed, reinstall it.
6. Adjust the solvent supply regulator back to its normal operating pressure.
Trigger the gun to relieve pressure. If you are using a
high pressure gun, engage the trigger lock. Remove
spray tip and clean tip separately.
58313882C
Page 59
System Operation
Purging Using Recipe 0
Recipe 0 is typically used:
•in multiple color systems to purge out material lines
without loading a new color
•at the end of a shift to prevent hardening of catalyzed material.
To setup Recipe 0, go to Advanced Setup. Select the
Recipe tab and change the Recipe to 0. The Recipe 0
Setup Screen appears. Set the chop times from 0-999
seconds in increments of 1 second.
1. Press the Standby
Trigger the gun to relieve pressure.
If you are using a high pressure gun, engage the trigger lock. Remove spray tip and clean tip separately.
key on the Booth Control.
Purging Fluid Supply System
Follow this procedure before:
•the first time material is loaded into equipment*
•servicing
•shutting down equipment for an extended period of
time
•putting equipment into storage
* Some steps are not necessary for initial flushing,
as no material has been loaded into the system
yet.
1. Press the Standby
Trigger the gun to relieve pressure.
If you are using a high pressure gun, engage the trigger lock. Remove spray tip and clean tip separately.
key on the Booth Control.
If using an electrostatic gun shut off the electrostatics
before flushing the gun.
2. If using a gun flush box, place the gun into the box
and close the lid.
3. Select Recipe 0 and press Enter .
4. If a gun flush box is not used, trigger
the gun into a grounded metal pail
until the purge sequence is complete.
5. The color change LED blinks while Recipe 0 runs
and turns solid after purge sequence is complete.
6. If the system is not completely clean, you can
repeat Recipe 0 by pressing Enter
.
If using an electrostatic gun, shut off the electrostatics
before flushing the gun.
2. Attach solvent supply lines as follows:
•Single color/single catalyst/single compo-nent C systems: disconnect the component A,
B, and C fluid supplies at the flow meter inlets,
and connect regulated solvent supply lines.
•Multiple color/multiple catalyst/multiple component C systems: connect the solvent
supply lines to the designated solvent valve on
the appropriate color, catalyst, or component C
valve stack. Do not connect a solvent supply to
any flow meter.
313882C59
Page 60
System Operation
3. Adjust the solvent fluid supply pressure. Use the
lowest possible pressure to avoid splashing.
4. Remove the Fluid Station cover to access the solenoid valves. See F
IG. 57.
5. Purge as follows:
•Single color/single catalyst/single compo-
nent C systems: Purge component A side.
Press the manual override on the Dose Valve A
solenoid valve and trigger the gun into a
grounded metal pail.
Purge component B side. Press the manual
override on the Dose Valve B solenoid valve
and trigger the gun into a grounded metal pail
until clean solvent flows from the gun.
Purge component C side. Press the manual
override on the Dose Valve C solenoid valve
and trigger the gun into a grounded metal pail
until clean solvent flows from the gun.
Repeat to thoroughly clean the fluid integrator.
•Multiple color/multiple catalyst/multiple
component C systems: Select Recipe 0 and
press Enter to purge the complete system.
The color change LED blinks while Recipe 0
runs and turns solid after purge sequence is
complete.
Repeat to thoroughly clean the fluid integrator.
6. Reinstall the Fluid Station cover.
7. Shut off the solvent fluid supply.
8. Disconnect the solvent supply lines and reconnect
the component A, B, and C fluid supplies.
9. See page 50 for Start Up procedure.
60313882C
Page 61
System Operation
Purging Sampling Valves and Tubes
Follow this procedure after meter calibration.
1. Press the Standby
2. See F
IG. 60. Close both fluid shutoff valves and
sampling valves.
3. Route the sampling tubes into a grounded waste
container.
4. On a single color system, attach a solvent supply
line to Flow Meter A inlet.
5. On the EasyKey, press the Setup key and
access the Advanced Setup screens.
6. Press the Right Arrow key to select the Calibra-
tion screen. Press the Down Arrow
select Purge from the menu. Press the Enter
key
.
key on the Booth Control.
key and
8. Close sampling valves.
NOTE: Select Abort on Calibration screen to cancel
current calibration and close dose or purge valves.
9. Fully open both fluid shutoff valves.
10. On a single color system, reconnect component A
fluid supply line to flow meter A.
NOTE: After calibration it is necessary to clean out contaminated mix material. Do a manual purge and resume
the recipe just tested, or do Recipe 0 then go on to the
next recipe.
Dose A, solvent purge valve (B side), component C
purge valve, and color change solvent valves (if
used) will open.
F
IG. 61. Calibration Screen
7. To avoid splashing, slowly open the sampling valves
and dispense solvent until the valves and tubes are
clean.
NOTE: When performing a calibration purge, the
solvent valve(s) close automatically after 2 minutes
or when Abort is selected on the screen.
313882C61
Page 62
Meter Calibration
Meter Calibration
To avoid splashing fluid in the eyes, wear eye protection. The fluid shutoff valves and ratio check valves are
retained by mechanical stops that prevent accidental
removal of the valve stem while the manifold is pressurized. If you cannot turn the valve stems manually,
relieve the system pressure, then disassemble and
clean the valve to remove the resistance.
4. Place the beakers (minimum size - 250 cc) in holders. Put the sampling tubes into the beakers.
NOTE: If tubes need replacing, use 5/32 in. or 4 mm
OD tubing.
5. On the EasyKey, press the Setup key to access
setup screens.
6. Press the Right Arrow key to select the Calibra-
Calibrate the meter:
•The first time the system is operated.
•Whenever new materials are used in the system,
especially if the materials have viscosities that differ
significantly.
•At least once per month as part of regular maintenance.
•Whenever a flow meter is serviced or replaced.
NOTE:
•K-Factors on the Calibration Screen are
updated automatically after the calibration procedure is completed.
•K-Factor values on the screen are viewable
only. If needed, you can manually edit the
K-Factors in Advanced Setup Screen 4 (page
36) or Recipe Setup Screen 5 (page 41).
•All values on this screen are in cc, independent
of the units set in Configure Screen 1.
tion Screen. Press
either Dose Valve A, B, C, or Solvent. Press the
Down Arrow
Start only one at a time.
7. Dispense component A, B, C, or Solvent into beaker.
the Enter key to select
key and select Start from the menu.
•The controller will use the active recipe
K-factors for meter calibration. The active recipe
must be recipe 1 to recipe 60. Recipes 0 and 61
do not have K-factor values.
1. Before calibrating meter A, B, or C, prime the system with material. For a color/catalyst/component C
change system, make sure the color/catalyst/component C valve is open.
2. Shut off all spray or dispense devices connected to
the ProMix.
3. Close all fluid shutoff valves and sampling valves.
62313882C
a. To avoid splashing, slowly open sampling
valves.
b. For more accurate calibration, adjust the valve
to dispense at a flow rate similar to your production spray flow rate.
c. Dispense a minimum of 250 cc; make sure
enough material is dispensed to accurately read
the volume with your beaker. The volumes do
not have to be equal or at any particular ratio.
d. Close sampling valve tightly.
Page 63
Meter Calibration
8. The volume that the ProMix measured displays on
the EasyKey.
9. Compare the amounts on the EasyKey to the
amount in the beakers.
NOTE: For maximum accuracy, use a gravimetric
(mass) method to determine the actual volumes dispensed.
10. If the screen and actual volumes are different, enter
the actual dispensed volume in cc for A, B, C, or
Solvent Volume field, and press the Enter key
If the value was substantially different, repeat the
calibration process.
NOTE: If the screen and actual volume is the same
or if for any reason you want to cancel the calibration procedure, scroll to Abort on the Calibration
Screen menu and press the Enter key
.
11. After the volume for A, B, C, or Solvent is entered,
the ProMix 3KS controller calculates the new flow
meter K-Factor and shows it on the Calibration
Screen.
NOTE: K-Factor values on the screen are viewable
only. If needed, you can manually edit the K-Factors
in Advanced Setup Screen 4 (page 36) or Recipe Setup Screen 5 (page 41).
12. Always purge sampling valves after calibrating
meters. Use one of the following methods.
.
•Follow the Purging Sampling Valves and
Tubes procedure, page 61.
•Place the sampling valve fluid tubes into a com-
patible cleaning fluid (TSL or solvent) or cap
them.
NOTE: If fluid hardens in sampling tubes, replace
them with 5/32 in. or 4 mm OD tubing.
13. Make sure all sampling valves are closed and all
fluid shutoff valves are fully open.
14. Before you begin production, clear the system of
solvent and prime it with material.
a. Go to Mix mode.
b. Trigger the gun into a grounded metal pail until
mixed material flows from the gun nozzle.
c. To begin operation, see Start Up, page 50.
313882C63
Page 64
Color Change
Color Change
Color Change Procedures
Multiple Color Systems
1. Shut off air to the gun.
2. Place the gun in the gun flush box if used, and close
the lid.
3. Switch to Standby
4. Use the scroll keys,
color. Press Enter
sequence.
5. If a gun flush box is not used, trigger
the gun into a grounded metal pail
until the color change sequence is
complete.
6. When the color change indicator light stops flashing
on the Booth Control, the color change sequence is
complete.
NOTE: The color change timer does not start until
the gun is triggered and fluid flow is detected. If no
flow is detected within 2 minutes, the color change
operation aborts. The Booth Control enters Standby
mode at the previous color.
7. When you are ready to spray, remove the gun from
the gun flush box if used, and close its door.
NOTE: The gun flush box door must be closed for
the atomizing air valve to open.
mode at the Booth Control.
or, to select the new
to begin the color change
Color Change Sequences
FIG. 62 through FIG. 71 illustrate various color change
sequences. See Table 8 to determine which figure to reference, based on the recipe change and system configuration. The time sequences are detailed in the following
paragraphs.
NOTE: See Setup Mode on page 25 to select purge
sources and set desired purge, chop, and fill times.
NOTES:
•The system uses old recipe data for the purge cycle.
However, it opens the new color/catalyst valve
based on the new recipe data.
•The system uses the new recipe data for the fill
cycle.
•For the one gun flush box (GFB) option, the spray
gun must be inserted in the GFB during the entire
color change cycle (purge and fill). The GFB trigger
output will be on during the recipe change cycle.
•For the two gun flush box (GFB) option, both spray
guns must be inserted in the GFBs during the entire
color change cycle (purge and fill). The system will
turn each GFB trigger output on and off based on
the preset time for each gun.
•For Special Outputs options, the system will turn
each output on and off based on the preset times.
Each Special Output has two different start times
and durations.
•For systems without dump valves, the First Purge
begins after the Color/Catalyst/Component C
Change steps are completed.
•Dump Valve B is required for a Catalyst Change
8. Press the Mix key to start spraying.
Single Color Systems
1. Follow procedure for Purging Fluid Supply System, page 59.
2. Load the new color. See Start Up, page 50.
3. Press the Mix key to start spraying.
64313882C
system.
•Dump Valve C is required for a Component C
Change system.
•When going from Recipe X to Recipe 0, only the
purge cycle data from Recipe 0 is used.
•When going from Recipe 0 to Recipe X, only the fill
cycle data from Recipe X is used.
Page 65
Color Change
Color Purge/Dump
•This sequence flushes out the color with solvent,
from the color valve to the Dump A valve.
•The color change solvent valve and the Dump A
valve open during the Purge Time.
•The color change solvent valve closes when the
Purge Time expires.
Color Fill
•This sequence fills the line with the new color all the
way to the Dump A valve.
•The new color valve and the Dump A valve open
during the Fill Time.
•The new color valve and the Dump A valve close
when the Fill Time expires.
Catalyst Purge/Dump
•This sequence flushes out the catalyst with solvent,
from the catalyst valve to the Dump B valve.
Component C Fill
•This sequence fills the line with the new component
C all the way to the Dump C valve.
•The new component C valve and the Dump C valve
open during the Fill Time.
•The new component C valve and the Dump C valve
close when the Fill Time expires.
First Purge
Select the First Purge Source (air, solvent, or 3rd valve)
and First Purge Time. For most applications, air is
selected.
The system purges the old material from the dose
valves to the gun, using only the selected purge media
(usually air). The selected purge valve opens during the
First Purge Time and closes when the time expires.
First Purge C
Select the First Purge C Source (solvent or water) and
First Purge C Time. For most applications, solvent is
selected.
•The catalyst change solvent valve and the Dump B
valve open during the Purge Time.
•The catalyst change solvent valve closes when the
Purge Time expires.
Catalyst Fill
•This sequence fills the line with the new catalyst all
the way to the Dump B valve.
•The new catalyst valve and the Dump B valve open
during the Fill Time.
•The new catalyst valve and the Dump B valve close
when the Fill Time expires.
Component C Purge/Dump
•This sequence flushes out component C with solvent, from the component C valve to the Dump C
valve.
•The component C change solvent valve and the
Dump C valve open during the Purge Time.
The system purges the old material from the dose
valves to the gun, using only the selected purge media.
The purge C valve opens during the First Purge C Time
and closes when the time expires.
Chop Cycle
Select the Chop Type (air/solvent or air/3rd valve) and
Chop Times.
The air purge valve opens only during the air chop cycle,
and the solvent (or 3rd valve) opens only during the solvent chop cycle. The number of chop cycles is determined by dividing the Total Chop Time by the sum of the
Air and Solvent Chop Times.
•The component C change solvent valve closes
when the Purge Time expires.
313882C65
Page 66
Color Change
Final Purge
Select the Final Purge Source (air, solvent, or 3rd valve)
and Final Purge Time. For most applications, solvent is
selected.
The system fills the line with solvent from the dose
valves to the gun, using only the selected purge media
(usually solvent). The selected purge valve opens during
the Final Purge Time and closes when the time expires.
Final Purge C
Select the Final Purge C Source (solvent or water) and
Final Purge C Time. For most applications, solvent is
selected.
The system fills the line with solvent from the dose
valves to the gun, using only the selected purge media.
The purge C valve opens during the Final Purge C Time
and closes when the time expires.
Table 8: ProMix 3KS Color Change Chart Reference
Starting
Recipe
Ending
Recipe
Change
Fill
This sequence fills the line from the dose valves to the
gun, and is also referred to as the mixed material fill.
The system begins mixing components A, B, and C until
the Fill Time expires.
Purge Active
The system turns on the Purge/Recipe Change Purge
Active output during these steps.
Fill Active
The system turns on the Recipe Change Fill Active output during this step.
0 or 61XYesNoFIG. 69
0 or 610n/aYesF
0 or 610n/aNoF
IG. 66
IG. 67
66313882C
Page 67
Color Change
Recipe Change
X -> Y
A1 to A2, B1 to B2, C1 to C2
3rd Flush Valve
Dump AIf any component isn't changing, that time segment is skipped.
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)For first pur ge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
Two Second Initial Purge Ti me
First Purge Time
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Mixed Fill Time
Valve
Time Sequence
FIG. 62: ProMix 3KS Color Change Chart 1
313882C67
Page 68
Color Change
Recipe Change
X -> Y
A1 to A2, B1 to B2, C1 to C2
No 3rd Flush Valve
Dump AIf any component isn't changing , that time segment is sk ipped.
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)For first purge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
First Purge Time
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Mixed Fill Time
Valve
Time Sequence
FIG. 63: ProMix 3KS Color Change Chart 2
68313882C
Page 69
Recipe Change
X -> Y
No components change
3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)For first pur ge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
Two Second Initial Purge Ti me
First Purge Time
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Mixed Fill Time
Valve
Time Sequence
Color Change
FIG. 64: ProMix 3KS Color Change Chart 3
313882C69
Page 70
Color Change
Recipe Change
X -> Y
No components change
No 3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)For first purge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
First Purge Time
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Mixed Fill Time
Valve
Time Sequence
FIG. 65: ProMix 3KS Color Change Chart 4
70313882C
Page 71
Recipe Change
y
X, 0, 61 -> 0
No more A, B, C
3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
B Purge Time
C Purge Time
Purge A (Air)For first pur ge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)3rd flush valve is only used if enabled in EasyKe
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
Two Second Initial Purge Ti meTwo second initial purge onlty happens if
First Purge Time3rd flush valve is enabled
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Valve
Time Sequence
Color Change
FIG. 66: ProMix 3KS Color Change Chart 5
313882C71
Page 72
Color Change
Recipe Change
X, 0, 61 -> 0
No more A, B, C
No 3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
B Purge Time
C Purge Time
Purge A (Air)For first purge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
First Purge Time
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Valve
Time Sequence
FIG. 67: ProMix 3KS Color Change Chart 6
72313882C
Page 73
Recipe Change
0,61 -> X
New A, B, C
3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)Only one of these three valves is used depending on the "Exiting F ill Source"
Purge B (Solvent)selected in EasyKey.
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
Exiting Fill Time
Mixed Fill Time
Valve
FIG. 68: ProMix 3KS Color Change Chart 7
Color Change
Time Sequence
313882C73
Page 74
Color Change
Recipe Change
0,61 -> X
New A, B, C
No 3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)
Purge B (Solvent)
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
Exiting Fill Time
Mixed Fill Time
Valve
Time Sequence
FIG. 69: ProMix 3KS Color Change Chart 8
74313882C
Page 75
Recipe Change
X -> Y
External Color Change
3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)For first purge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
Two Second Initial Purge Time
First Purge Time
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Wait for External CC Input
Mixed Fill Time
Valve
Time Sequence
FIG. 70: ProMix 3KS Color Change Chart 9
Color Change
313882C75
Page 76
Color Change
Recipe Change
X -> Y
External Color Change
No 3rd Flush Valve
Dump A
Dump B
Dump C
Solvent A
Solvent B
Solvent C
Component A
Component B
Component C
A Purge Time
A Fill Time
B Purge Time
B Fill Time
C Purge Time
C Fill Time
Purge A (Air)For first purge and final purge,
Purge B (Solvent)only one of the purge valves are used
Purge D (3rd/Water)
Purge C (Solvent/Water)
Dose A
Dose B
Dose C
No Gun Flush Boxes:
Gun Trigger(s) by Operator
1 Gun Fluxh Box:
Gun Flush Box Output 1
Gun Flush Box Output 2
2 Gun Flush Boxes:
Gun Flush Box Output 1
Gun Flush Box Output 2
First Purge Time
First Purge C Time
Chop Time
Final Purge Time
Final Purge C Time
Wait for External CC Input
Mixed Fill Time
Valve
Time Sequence
FIG. 71: ProMix 3KS Color Change Chart 10
76313882C
Page 77
Alarms and Warnings
Alarms and Warnings
NOTICE
Do not use the fluid in the line that was dispensed off
ratio as it may not cure properly.
System Alarms
System alarms alert you of a problem and help prevent
off-ratio spraying. If an alarm occurs, operation stops
and the following occurs:
•A red LED illuminates steadily or blinks on the Booth
Control.
•Booth Control displays an alarm E-Code, E-1 to
E-31. See F
IG. 72.
•Alarm sounds.
•Status bar on the EasyKey Display shows the alarm
E-Code with a description (see Table 9).
Alarm Indicator (red)
Recipe Indicator (green)
IG. 72. Booth Control
F
Display
TI11614A
System Warnings
Table 9 lists the System Warning Codes. Warnings do
not stop operation or sound an alarm. They are saved in
the date/time stamped log, which can be viewed on a
PC, using the ProMix 2KS Web Interface (see manual
313386).
To Reset Alarm and Restart
NOTE: When an alarm occurs be sure to determine the
E-Code before resetting it. See Table 9. If you forget
which E-Code occurred, use the Alarms Screens (page
23) to view the last 10 alarms, with date and time
stamps.
To reset alarms, see Table 10. Many alarms can be
cleared by simply pressing the Alarm Reset
Table 9: System Alarm/Warning Codes
Code DescriptionDetails
E-1Communication Error AlarmPage 78
E-2Potlife AlarmPage 78
E-3Ratio High AlarmPage 79
E-4Ratio Low AlarmPage 80
E-5Overdose A/B Dose Too Short AlarmPage 81
E-6Overdose B/A Dose Too Short AlarmPage 81
E-7Dose Time A AlarmPage 82
E-8Dose Time B AlarmPage 82
E-9Mix in Setup AlarmPage 83
E-10Remote Stop AlarmPage 83
E-11Purge Volume AlarmPage 83
E-12CAN Network Communication Error
Alarm
E-13High Flow AlarmPage 85
E-14Low Flow AlarmPage 85
E-15System Idle WarningPage 85
E-16Setup Change WarningPage 85
E-17Power On WarningPage 85
E-18Defaults Loaded WarningPage 85
E-19I/O AlarmPage 86
E-20Purge Initiate AlarmPage 87
E-21Material Fill AlarmPage 87
E-22Tank A Low AlarmPage 88
E-23Tank B Low AlarmPage 88
E-24Tank S Low AlarmPage 88
E-25Auto Dump Complete AlarmPage 88
E-26Color/Catalyst Purge AlarmPage 88
E-27Color/Catalyst Fill AlarmPage 88
E-29Tank C Low AlarmPage 88
E-30Overdose C AlarmPage 81
E-31Dose Time C AlarmPage 82
key.
Page 84
313882C77
Page 78
Alarm Troubleshooting
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-1: COMM_ERROR
CauseSolution
No power to the EasyKey.Connect power to EasyKey.
No power to Fluid Station. The intrinsically safe power
cable between the EasyKey and Fluid Station is not connected.
No power to Fluid Station. The fluid control board fuse is
blown.
The fiber optic cable between the EasyKey and Fluid
Station is not connected.
The fiber optic cable is cut or bent.Verify that the cable has not been cut or bent at a radius
Dirty fiber optic cable ends.Disconnect fiber optic cable ends and clean with a
A communication cable or connector failed.Replace cable.
E-2: POTLIFE_ALARM
CauseSolution
The potlife time has been exceeded for the mixed material.
NOTICE
To prevent mixed material from curing in the equipment, do not shut off power. Follow one of the solutions at right.
Verify that the cable is correctly connected. See Installation manual.
Verify condition of fuse and replace if necessary. See
Repair-Parts manual.
Verify that the cable is correctly connected. See Installation manual.
smaller than 1.6 in. (40 mm).
lint-free cloth.
Press the Alarm Reset
alarm. Purge the system with solvent, fresh mixed mate-
rial, or a new color:
•Solvent Purge - See Purging Mixed Material on
page 58. The system purges until the preset purge
time is complete.
key to stop the audible
•New Mixed Material Purge - Go to Mix mode and
spray the required volume to restart the timer.
•Color Change - Perform a color change, page 64.
78313882C
Page 79
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-3: RATIO_HIGH_ALARM
Sequential Dosing System
The mix ratio is higher than the set tolerance on the previous dose cycle.
CauseSolution
There is too little restriction in the system.•Check that the system is fully loaded with material.
•Check that the supply pump’s cycle rate is set properly.
•Check that the spray tip/nozzle is properly sized for
the flow and application, and that it is not worn.
•Check that the fluid regulator is set properly.
If the alarm occurs during start up, after purging, the flow
rate was probably too high.
If the alarm occurred after you were spraying for some
time, the pressures from the fluid supplies could be
unbalanced.
Restrict gun needle travel to slow down the initial fluid
delivery rate until fluid hoses are loaded with material.
Adjust component A, B, and C fluid supply regulator
pressures until they are about equal. If the pressures are already about equal, verify that component A, B, and C
dose valves are operating properly.
Slow actuation of the component A, B, or C valves. This
can be caused by:
Manually operate the Dispense A, B, and C solenoid
valves as instructed in the ProMix 3KS Repair-Parts
manual to check operation.
•Air pressure to the valve actuators is too low.•Increase air pressure. Air pressure must be 75-120
psi (0.52-0.84 MPa, 5.2-8.4 bar); 120 psi is recommended.
•Something is restricting the solenoid or tubing and
interrupting valve actuation air.
•There may be dirt or moisture in the air supply. Filter
appropriately.
•A dose valve is turned in too far.•Refer to Table 6: Mix Manifold Valve Settings,
page 50, for adjustment guidelines.
•Fluid pressure is high and air pressure is low.•Adjust air and fluid pressure. See recommended air
pressure above.
313882C79
Page 80
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-4: RATIO_LOW_ALARM
Sequential Dosing System
The mix ratio is lower than the set tolerance on the previous dose cycle.
CauseSolution
There is too much restriction in the system.•Check that the system is fully loaded with material.
•Check that the supply pump’s cycle rate is set properly.
•Check that the spray tip/nozzle is properly sized for
the flow and application, and that it is not clogged.
•Check that the fluid regulator is set properly.
If the alarm occurs during start up, after purging, the flow
rate was probably too high.
If the alarm occurred after you were spraying for some
time, the pressures from the fluid supplies could be
unbalanced.
Restrict gun needle travel to slow down the initial fluid
delivery rate until fluid hoses are loaded with material.
Adjust component A, B, and C fluid supply regulator
pressures until they are about equal. If the pressures are already about equal, verify that component A, B, and C
dose valves are operating properly.
Slow actuation of the component A, B, or C valves. This
can be caused by:
Manually operate the Dispense A, B, and C solenoid
valves as instructed in the ProMix 3KS Repair-Parts
manual to check operation.
•Air pressure to the valve actuators is too low.•Increase air pressure. Air pressure must be 75-120
psi (0.52-0.84 MPa, 5.2-8.4 bar); 120 psi is recommended.
•Something is restricting the solenoid or tubing and
interrupting valve actuation air.
•There may be dirt or moisture in the air supply. Filter
appropriately.
•A dose valve is turned in too far.•Refer to Table 6: Mix Manifold Valve Settings,
page 50, for adjustment guidelines.
•Fluid pressure is high and air pressure is low.•Adjust air and fluid pressure. See recommended air
pressure above.
80313882C
Page 81
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-5: OVERDOSE_A/B_DOSE_TOO_SHORT_ALARM,
E-6: OVERDOSE_B/A_DOSE_TOO_SHORT_ALARM, and
E-30: OVERDOSE_C_ALARM
E-5: the A dose overshoots and, when combined with B, is too large for the mix chamber capacity.
E-6: the B dose overshoots and forces an A side dose that, when combined with B, is too large for the mix chamber
capacity.
E-30: the C dose overshoots and, when combined with A+B, is too large for the mix chamber capacity.
CauseSolution
Valve seal or needle/seat are leaking. Check F
IG. 11
Repair the valve (see valve manual 312782).
Totals Scree n on page 23. If A, B, or C are dosing simultaneously (sequential dosing only), there is a leak.
Sampling valve is leaking.Tighten or replace valve.
Flow meter fluctuations caused by pressure pulsations.Check for pressure pulsations:
1. Close all the manifold valves.
2. Turn on the circulating pumps and all the booth
equipment (such as fans and conveyors).
3. Check if the ProMix 3KS is reading any fluid flow.
4. If the ProMix 3KS shows there is fluid flow and there
are no leaks from the gun or any other seals or fittings, the flow meters are probably being affected by
pressure pulsations.
5. Close the fluid shutoff valve between the fluid supply
system and the flow meter. The flow indication
should stop.
6. If necessary, install pressure regulators or a surge
tank on the fluid inlets to the ProMix 3KS to reduce
the fluid supply pressure. Contact your Graco distributor for information.
Slow actuation of component A, B, or C valves.See E-3: RATIO_HIGH_ALARM and E-4:
RATIO_LOW_ALARM, pages 79-80.
Running a high mix ratio and a high flow rate.It may be necessary to restrict the flow rate through the
component B or C dose valve by adjusting its hex nut (E).
See page 48.
313882C81
Page 82
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-7: DOSE_TIME_A_ALARM, E-8: DOSE_TIME_B_ALARM, and E-31: DOSE_TIME_C_ALARM
E-7: gun trigger input is active (AFS or Integration) and no A meter pulses are detected during the dose time
selected.
E-8: gun trigger input is active (AFS or Integration) and no B meter pulses are detected during the dose time
selected.
E-31: gun trigger input is active (AFS or Integration) and no C meter pulses are detected during the dose time
selected.
CauseSolution
System is in Mix mode and gun is only partially
Fully trigger the gun.
triggered, allowing air but no fluid to pass through gun.
Fluid flow rate is too low.Increase flow rate.
Dose time setting is too short for the current flow rate.Increase the dose time setting.
Flow meter or cable failed or flow meter clogged.To check meter sensor operation, remove meter cap to
expose sensor. Pass a ferrous metal tool in front of the
sensor.
TI12792a
If there is a meter or cable failure, you will see a large difference between the amount of fluid dispensed and the
flow meter volume displayed by the EasyKey. Clean or
repair meter as necessary. Also see meter manual
308778.
Follow Meter Calibration procedure, page 62.
Slow actuation of component A, B, or C valves.See E-3: RATIO_HIGH_ALARM and E-4:
RATIO_LOW_ALARM, pages 79-80.
The supply pump is not turned on.Turn on the supply pump.
There is an air leak downstream from the air flow switch. Check the air lines for leaks and repair.
The air flow switch is stuck open.Clean or replace air flow switch.
System is in Mix mode with 0 volume entered for Min
Material Fill Volume (see Option Screen 1, page 32),
Verify condition of fuse and replace if necessary. See
Repair-Parts manual.
and Fuse F1 is blown.
82313882C
Page 83
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-9: MIX_IN_SETUP_ALARM
CauseSolution
Attempt to operate system while in Setup mode.System must be in Standby to change current recipe,
and cannot be operated.
E-10: REMOTE_STOP_ALARM
CauseSolution
Automation has requested that the system abort all oper-
Abort operations. Troubleshoot automation system.
ations.
E-11: PURGE_VOLUME_ALARM
CauseSolution
ProMix 3KS solvent flow switch is not activated while
purging.
Verify that the gun is not shut off and that the solvent flow
switch is activated while purge is taking place.
Minimum flush volume is set too high.Increase solvent supply or decrease minimum volume
setting.
No meter pulses during Color/Catalyst Dump.Color change solvent supply not set up or functional.
Check Color Change setup.
313882C83
Page 84
Alarm Troubleshooting
E-12: CAN_COMM_ERROR_ALARM
CauseSolution
Table 10. Alarm Troubleshooting
Communication between the Color Change Module and
the Fluid Station is interrupted.
•Verify that all cables are connected securely and that
the Color Change and Booth Control power LEDs
turn on. If the power LED does not turn on, the problem is probably caused by a bad connection. The nut
on the connector must make at least 5 complete
turns to ensure a good connection. If the power LED
still does not light, the cable or board is bad.
•Check the color change board DIP switch settings.
See the ProMix 3KS installation manual.
•Check the fluid plate board DIP switch setting. An
incorrect setting will not cause E-12 alarms, but a
correct setting will help prevent E-12 caused by electrical noise. See the ProMix 3KS installation manual.
•Check EasyKey software version (displayed at power
up for all versions and when the lock key is pressed
for version 2.02.000 and above). If older than
1.06.002, upgrade. Be sure to save settings through
BWI or AWI before upgrading, as they will be erased.
•The sticker on the color change board shows the
software part number and version, for example
15T270 1.01. If the version is older than 1.01,
replace the board.
Communication between the Color Change Module and
the Fluid Station is interrupted. The fluid control board
fuse is blown.
Communication between the Booth Control and the Fluid
Station is interrupted.
•If all software versions and DIP switch settings are
correct and you still have E-12 alarms, then the system has a bad connection, bad cable, or bad circuit
board. Use a multimeter on the CAN connectors to
test whether there is a good connection between
systems. If there is, you have a bad circuit board. If
there is not, you have a bad connector, connection,
or cable.
Verify condition of fuse and replace if necessary. See
Repair-Parts manual.
Verify that the cable is correctly connected.
84313882C
Page 85
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-13: HIGH_FLOW_ALARM or E-14: LOW_FLOW_ALARM (may also be set as Warnings)
CauseSolution
Fluid system is producing too much or too little flow.Troubleshoot fluid system for restrictions, leaks,
exhausted fluid supply, incorrect settings, etc. Increase
or decrease flow rate, as required.
E-15: SYSTEM_IDLE_WARNING
CauseSolution
Mix input is high, but the gun has not been triggered for 2
If not painting, clear alarm and resume operation.
minutes.
If painting, shut down and inspect fluid meter and air flow
switch.
E-16: SETUP_CHANGE_WARNING
CauseSolution
The system setup parameters have been changed.No action required. See Event Log available through
advanced web interface.
E-17: POWER_ON_WARNING
CauseSolution
The power to the system has been cycled.No action required. See Event Log available through
advanced web interface.
E-18: DEFAULTS_LOADED_WARNING
CauseSolution
The factory defaults have been installed on the system.No action required. See Event Log available through
advanced web interface.
313882C85
Page 86
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-19: I/O ALARM
CauseSolution
The Mix and Purge digital inputs are on at the same time. Ensure that only one input is on at a time. At least 1 sec
delay is required when switching from Mix to Purge or
vice versa.
NOTE: The I/O alarm incorporates several sub-alarms relating to internal data issues, as detailed below. These
alarms may not apply to all software versions.
Fluid Plate Reboot (FP Reboot): Occurs if the system
detects a fluid plate control board reboot or power cycle
not triggered from the EasyKey. The system reverts to
Recipe 61, and mixed material may be in the lines.
Autokey Lost: Occurs if the Autokey is lost or changed
after having been detected. (A short term loss of the
Autokey will not be registered.) Some system functions
may become unavailable. For example, an automatic
system will not respond to PLC or robot control.
Illegal Source: Occurs if a recipe outside of the range
1-60 is detected as the source data for global recipe data
copies. This is possible if an invalid configuration file is
sent to the EasyKey.
2K/3K Error: Occurs if the recipe data is incompatible
with the current Autokey setting (2K or 3K). This is possible if the Autokey is changed or an invalid configuration
file is sent to the EasyKey.
Init Error: Occurs if the recipe data codes specifying the
type of machine they were made on are not what is
expected. For example, a 3KS machine receives a configuration file originally made on a 2KS machine.
Config Error: Occurs if a configuration file sent to the
EasyKey specifies a different hardware setup than what
exists. For example, the configuration file specifies 2
color change boards but only 1 is present.
Flush the system or perform a color change. If possible,
identify the origin of the reboot or power cycle.
Reinstall the Autokey, or verify that the Autokey is set
properly.
Verify that the source data is from a valid recipe (1-60).
Verify that the Autokey is set properly or that the configuration file is valid.
Verify that the configuration file is valid.
Verify that the configuration file specifications and the
hardware conform.
Range Error: Occurs if a valve used in a recipe is not
present in the current hardware setup. For example, a
Verify that the recipe specifications and the hardware
conform.
recipe calls for valve 30 but the system has only 12
valves.
Level Control (LC) Error: Occurs if level control data is
Verify that the Autokey is set properly.
received by the EasyKey, and the current Autokey setting
(2K or 3K) has changed since the level control data was
originally initialized.
Level Control (LC) Range Error: Occurs if level control
Set level control data correctly.
data includes a valve range exceeding the capability of
the machine.
Modbus (MB) Overflow: Occurs if the Modbus connec-
Verify the Modbus protocol to the EasyKey.
tion to a PLC experiences data overflow.
86313882C
Page 87
E-20: PURGE_INITIATE_ALARM
CauseSolution
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
System detects atomizing air to the gun when purge is
Shut off gun air.
selected.
For systems with a gun flush box, gun is not in the box
when purge is selected.
For systems with auto dump on, gun is not in the box
when auto dump is initiated.
Place gun in gun flush box. Verify that gun flush box is
operating properly.
Place gun in gun flush box. Verify that gun flush box is
operating properly.
For systems with a gun flush box, Fuse F2 is blown.Verify condition of fuse and replace if necessary. See
Repair-Parts manual.
E-21: MATERIAL_FILL_ALARM
CauseSolution
For systems with minimum mixed material fill volume
Check for restrictions or leaks in the fluid supply system.
entered, the system detects that fill volume is not
achieved during mixed material fill time.
Check if the fill volume is properly configured:
•Adjust fill volume.
•Adjust fill time.
For systems without color change and with minimum
mixed material fill volume entered, Fuse F1 is blown.
Verify condition of fuse and replace if necessary. See
Repair-Parts manual.
313882C87
Page 88
Alarm Troubleshooting
Table 10. Alarm Troubleshooting
E-22: TANK_A_LOW_ALARM, E-23: TANK_B_LOW_ALARM, E-24: TANK_S_LOW_ALARM,
or E-29: TANK_C_LOW_ALARM
CauseSolution
The tank volume reaches the low-level threshold. The EasyKey screen will display the alarm and prompt
the user to do one of the following:
•Refill tank volume to clear the alarm.
•Resume mixing by selecting “Spray 25% of remaining volume.” If this selection is chosen, a second
alarm will occur after 25% of the remaining volume is
mixed. Refill tank volume to clear the alarm.
E-25: AUTO_DUMP_COMPLETE_ALARM
CauseSolution
A potlife alarm is active for more than 2 minutes, the gun
Be sure to spray all mixed material before potlife expires.
flush box is enabled and gun is in the gun flush box, and
an auto dump flush sequence is complete.
E-26: COLOR/CATALYST_PURGE_ALARM
CauseSolution
System detects no meter pulses, or a disruption in meter
Check that meter cable is connected.
pulses lasting longer than 1 second throughout the
Color/Catalyst purge time duration.
Clean or repair meter.
E-27: COLOR/CATALYST_FILL_ALARM
CauseSolution
System detects no meter pulses, or system must detect
Check that meter cable is connected.
at least 10cc of material from each side throughout the
Color/Catalyst fill time duration.
Gun, dump valve, or correct color/catalyst valve not
Clean or repair meter.
Open the valve.
open.
Exhausted fluid supply.Check fluid level and refill if necessary.
Fuse F1, F2, or both are blown.Verify condition of fuses and replace if necessary. See
Repair-Parts manual.
E-29: TANK_C_LOW_ALARM (see above)
E-30: OVERDOSE_C_ALARM (see page 81)
E-31: DOSE_TIME_C_ALARM (see page 82)
88313882C
Page 89
Alarm Troubleshooting
313882C89
Page 90
Schematic Diagrams
Schematic Diagrams
System Pneumatic Schematic
AIR INPUT
PURGE AIR
3/8 AIR FILTER
MANUAL DRAIN
5 MICRON
WALL MOUNT ONLY
05
AIR INPUT
CONTROL AIR
TO MANIFOLD 1/4 TUBE
AIR EXHAUST MUFFLER
4-WAY SOLENOID
4-WAY SOLENOID
4-WAY SOLENOID
4-WAY SOLENOID
4-WAY SOLENOID
3-WAY SOLENOID
3-WAY SOLENOID
12 VDC
12 VDC
12 VDC
12 VDC
12 VDC
AIR EXHAUST MUFFLER
12 VDC
12 VDC
COLOR
CHANGE
CONTROL
DOSE A
OPEN
OPEN
OPEN
OPEN
OPEN
VALV E
DOSE B
VALV E
PURGE A
VALV E
PURGE B
VALV E
PURGE C
VALV E
(OPTIONAL)
DUMP A
VALV E
(OPTIONAL)
DUMP B
VALV E
(OPTIONAL)
FLUSH AIR TO FLUID INLET 1/4 TUBE
MAC
36 SERIES SOLENOID VALVES
A
B
A
B
A
B
A
B
A
B
A
A
CLOSE
5/32 TUBE
CLOSE
5/32 TUBE
CLOSE
5/32 TUBE
CLOSE
5/32 TUBE
CLOSE
5/32 TUBE
5/32 TUBE
OPEN
5/32 TUBE
OPEN
COLOR
VALV E
STACKS
COLOR 1
COLOR 2
COLOR 3
COLOR 4
COLOR 5
COLOR 6
COLOR 7
COLOR 8
COLOR SOLVENT
COLOR 9
COLOR 10
COLOR 11
COLOR 12
CATALYST 1
CATALYST 2
CATALYST 3
CATALYST 4
CATALYST SOLVENT
3-WAY SOLENOID
3-WAY SOLENOID
MANIFOLD
12 VDC
12 VDC
GFB 1
VALV E
A
A
5/32 TUBE
5/32 TUBE
OPEN
OPEN
(OPTIONAL)
GFB 2
VALV E
(OPTIONAL)
90313882C
Page 91
System Pneumatic Schematic
Schematic Diagrams
TO MANIFOLD 1/4 TUBE
AIR EXHAUST MUFFLER
4-WAY SOLENOID
4-WAY SOLENOID
3-WAY SOLENOID
12 VDC
12 VDC
12 VDC
COLOR
CHANGE
CONTROL
DOSE C
OPEN
5/32 TUBE
OPEN
OPEN
VALV E
PURGE C
VALV E
DUMP C
VALV E
(OPTIONAL)
MAC
36 SERIES SOLENOID VALVES
A
B
A
B
A
CLOSE
5/32 TUBE
CLOSE
5/32 TUBE
COLOR
VALV E
STACKS
3RD COMP. 1
3RD COMP. 2
3RD COMP. 3
3RD COMP. 4
3RD COMP. SOLVENT
COLOR 13 (OPTIONAL)
COLOR 14 (OPTIONAL)
COLOR 15 (OPTIONAL)
COLOR 16 (OPTIONAL)
COLOR 17 (OPTIONAL)
COLOR 18 (OPTIONAL)
COLOR 19 (OPTIONAL)
COLOR 20 (OPTIONAL)
COLOR 21 (OPTIONAL)
COLOR 22 (OPTIONAL)
COLOR 23 (OPTIONAL)
COLOR 24 (OPTIONAL)
COLOR 25 (OPTIONAL)
MANIFOLD
313882C91
Page 92
Schematic Diagrams
System Electrical Schematic
NOTE: The electrical schematic illustrates all possible wiring expansions in a ProMix 3KS system. Some compo-
nents shown are not included with all systems.
Non-Hazardous Area
85-250
VAC
L1NGND
L1
N
GND
PROMIX 2KS EASYKEY INTERFACE
TERMINAL
BLOCK
L1NGND
L1
N
GND
LINE
FILTER
L1
N
85-250 VAC
L1
1
POWER
+
DC OK
N
2
SUPPLY
-
+
+24 VDC
COMMON
GND LUG
123
-
COMMON
POWER
ROCKER
SWITCH
121A1B2A
OPEN OPEN
UNUSED
UNUSED
UNUSED
UNUSED
12345
J1
POWER HARNESS
2B
UNUSED
BARRIER
BOARD
HARNESS
+12VDC I/S (RED)
123
COM (BLACK)
J5
SHIELD
+24VDC
OPEN
123
J4
MIX INPUT
PURGE INPUT
JOB COMPLETE INPUT
EXTERNAL CLR CHG READY
RESET ALARM INPUT
DIGITAL INPUT COMMON
DIGITAL INPUT COMMON
RECIPE BIT 0 INPUT
12345
6
1234567
J2
RJ45
ALARM
+-
COMMON
+
1234
DISPLAY
BOARD
1234567891011J412345678910111213141516171819
RJ45
3'
POWER DIST.
TERMINAL
BLOCKS
-
RJ45
J9
RJ45
J6
RECIPE BIT 1 INPUT
RECIPE BIT 2 INPUT
RECIPE BIT 3 INPUT
RECIPE BIT 4 INPUT
RECIPE BIT 5 INPUT
RECIPE CHANGE INPUT
8
J3
REMOTE
I/O
INTEGRATION
+
+
+
-
-
-
J2
DIGITAL OUTPUT COMMON/POWER
PURGE/RECIPE CHG ACTIVE OUTPUT
MIX ACTIVE OUTPUT
MIX READY OUTPUT
FILL ACTIVE
1234567
J4
BOARD
SHIELD
+
-
DIGITAL OUTPUT COMMON/POWER
SPECIAL OUTPUT #1
SPECIAL OUTPUT #2
FLOW CAL. ACTIVE
FLOW RATE ALARM OUTPUT
DIGITAL OUTPUT COMMON/POWER
12345
8
123456789
FLOW CONTROL CAL. (BLK)
GUN TRIGGER (WHT )
DIGITAL IN COMMON (RED)
REMOTE STOP (GRN)
123456789
SPECIAL OUTPUT #3
SPECIAL OUTPUT #4
DIGITAL OUTPUT COMMON/POWER
6
J5
10
ALARM RESET (BRN)
ALARM OUTPUT (BLU)
DIGITAL OUTPUT COMMON (ORG)
POT LIFE (YEL)
FLOW RATE ANALOG IN (PUR)
FLOW RATE ANALOG COMMON (GRAY)
10
J5
(+24) YEL
I/O HARNESSES
20
BEACON
(COM) GRAY
ORG
RED
BRN
TERMINAL
BLOCKS
12345
RS485 INTEGRATION A (WHT/BLU)
RS485 INTEGRATION B (BLU/WHT)
RS485 INTEGRATION GROUND (SHIELD)
12345
J10
6
RS485 NETWORK A (WHT/ORG)
RS485 NETWORK B (ORG/WHT)
RS485 NETWORK GROUND (SHIELD)
6
CABLE
WEB SERVER
RJ45
BULKHEAD
3'
(25'-200' OPTIONS)
RJ45
MODULE
RJ45
3'
FO IN (BLK)
FO OUT (BLU)
RJ45
J8
J7
RJ45
P1
MEMBRANE
SWITCH
WITH
RIBBON
CABLE
DISPLAY
92313882C
Page 93
Schematic Diagrams
System Electrical Schematic
NOTE: The electrical schematic illustrates all possible wiring expansions in a ProMix 3KS system. Some compo-
nents shown are not included with all systems.
Non-Hazardous Area
TO FO IN - J6
(BLK)
TO FO OUT - J4
(BLU)
TO J3
3RD COMPONENT
FLUID PANEL
HAZARDOUS
AREA
SMART FLUID PANEL
TO J10
CONTROL BOX
NON-HAZARDOUS
AREA
85-250
VAC
L1NGND
50' STD.
123
TERMINAL
BLOCK
GND LUG
POWER
ROCKER
123
121A1B2A
OPEN OPEN
NL1L1 NGND
12345
SWITCH
J1
2B
HARNESS
BARRIER
BOARD
123
85-250 VAC
L1
N
1
2
POWER
SUPPLY
-
-
+
+
COMMON
COMMON
+24 VDC
DC OK
PROMIX 3KS BARRIER/PS ENCLOSURE
(10'-100' OPTIONS)
+12VDC I/S (RED)
COM (BLACK)
SHIELD
+24VDC
OPEN
COMMON
123
123
J5
J4
POWER
HARNESS
313882C93
Page 94
Schematic Diagrams
System Electrical Schematic
NOTE: The electrical schematic illustrates all possible wiring expansions in a ProMix 3KS system. Some compo-
nents shown are not included with all systems.
Hazardous Area
TECNO
V/P
FLUID
PRESS.
SENS.
SIG (RED)
COM (BLK)
+ PRESSURE (GRN)
COM (RED)
EX+ (WHT)
- PRESSURE (BLK)
SHIELD (BARE)
J4
J2
12345
1
2
FLOW
CONTROL
BOARD
12345
J1
10' STD.
V/P ANALOG OUT (WHT)
PRESS. (GRN)
+12 V (RED)
GND (BLK)
CHASSIS (BARE)
I.S. METERS
(25'-100' OPTIONS) - GREEN COLOR CODED
(10')/
(40')
12345
J5
SMART FLUID PANEL
CONTROL BOX
BLUE
CABLE
HAZARDOUS
AREA
6
(50' STD.)/
DUMP B
MANIFOLD
BLACK
RED
65432
MH2
J10
+12VDC I/S
COM
SHIELD
123
(100' OPTION)
GFB #1
BLACK
RED
J8
12 VDC
3-WAY SOLENOID
GFB #2
BLACK
RED
1
DUMP A 3RD FLUSH
BLACK
RED
BLUE
BLACK
BLACK
RED
65432
1
J15
J6
J4
FO IN
FO OUT
(BLU)
CABLE
12 VDC
4-WAY SOLENOID
PURGE B
BLACK
65432
(BLK)
RED
PURGE A
BLACK
RED
YELLOW
J14
BLACK
1
12 VDC
DOSE B
GREEN
BLACK
BLACK
65432
J9
4-WAY SOLENOID
DOSE A
SIG
RED
BLACK
RED
1
12345678910
AIR FLOW SWITCH 1
50' STD.
(10'-100' OPTIONS) - RED COLOR CODED
SIG (WHITE)
SHIELD/GRN
PWR (RED)
J3
FLOW METER B
COM (BLACK)
6
SIG (WHITE)
FLOW METER SOLVENT
SHIELD/GRN
PWR (RED)
COM (BLACK)
SIG (WHITE)
12345
J12
SHIELD/GRN
6
GROUND
TERMINAL
UNUSED
UNUSED
UNUSED
12345
J13
UNUSED
UNUSED
GRD (BLK)
UNUSED
32541
6
+12VDC (RED)
SHIELD (BARE)
CAN H (WHT)
CAN L (BLU)
J11
UNUSED
UNUSED
UNUSED
32541
J7
UNUSED
UNUSED
FLOW METER A
3X CABLE
AIR FLOW SWITCH 2
SOLVENT FLOW SWITCH
GFB 2 PRESSURE SWITCH
GFB 1 PRESSURE SWITCH
PWR (RED)
COM (BLACK)
COM
SIG
COM
SIG
COM
SIG
COM
SIG
COM
12345
J1
SMART
FLUID
PANEL
CONTROL
BOARD
NON-HAZARDOUS
AREA
PROMIX 2KS
EASYKEY
INTERFACE
TO FO OUT (BLU) - J8
PROMIX 3KS
PS & BARRIER
INTERFACE
TO FO IN (BLK) - J7
94313882C
Page 95
Schematic Diagrams
System Electrical Schematic
NOTE: The electrical schematic illustrates all possible wiring expansions in a ProMix 3KS system. Some compo-
15 amp maximum circuit breaker required
8 to 14 AWG power supply wire gauge
below 70 dBA
below 85 dBA
perfluoroelastomer; PTFE
1%, user selectable
*Dependent on flow rate, dose size, and meter resolution.
See individual component manuals for additional technical data.
313882C97
Page 98
Graco Standard Warranty
Graco warrants all equipment referenced in this document which is manufactured by Graco and bearing its name to be free from defects in
material and workmanship on the date of sale to the original purchaser for use. With the exception of any special, extended, or limited warranty
published by Graco, Graco will, for a period of twelve months from the date of sale, repair or replace any part of the equipment determined by
Graco to be defective. This warranty applies only when the equipment is installed, operated and maintained in accordance with Graco’s written
recommendations.
This warranty does not cover, and Graco shall not be liable for general wear and tear, or any malfunction, damage or wear caused by faulty
installation, misapplication, abrasion, corrosion, inadequate or improper maintenance, negligence, accident, tampering, or substitution of
non-Graco component parts. Nor shall Graco be liable for malfunction, damage or wear caused by the incompatibility of Graco equipment with
structures, accessories, equipment or materials not supplied by Graco, or the improper design, manufacture, installation, operation or
maintenance of structures, accessories, equipment or materials not supplied by Graco.
This warranty is conditioned upon the prepaid return of the equipment claimed to be defective to an authorized Graco distributor for verification of
the claimed defect. If the claimed defect is verified, Graco will repair or replace free of charge any defective parts. The equipment will be returned
to the original purchaser transportation prepaid. If inspection of the equipment does not disclose any defect in material or workmanship, repairs will
be made at a reasonable charge, which charges may include the costs of parts, labor, and transportation.
THIS WARRANTY IS EXCLUSIVE, AND IS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED
TO WARRANTY OF MERCHANTABILITY OR WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE.
Graco’s sole obligation and buyer’s sole remedy for any breach of warranty shall be as set forth above. The buyer agrees that no other remedy
(including, but not limited to, incidental or consequential damages for lost profits, lost sales, injury to person or property, or any other incidental or
consequential loss) shall be available. Any action for breach of warranty must be brought within two (2) years of the date of sale.
GRACO MAKES NO WARRANTY, AND DISCLAIMS ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE, IN CONNECTION WITH ACCESSORIES, EQUIPMENT, MATERIALS OR COMPONENTS SOLD BUT NOT
MANUFACTURED BY GRACO. These items sold, but not manufactured by Graco (such as electric motors, switches, hose, etc.), are subject to
the warranty, if any, of their manufacturer. Graco will provide purchaser with reasonable assistance in making any claim for breach of these
warranties.
In no event will Graco be liable for indirect, incidental, special or consequential damages resulting from Graco supplying equipment hereunder, or
the furnishing, performance, or use of any products or other goods sold hereto, whether due to a breach of contract, breach of warranty, the
negligence of Graco, or otherwise.
FOR GRACO CANADA CUSTOMERS
The Parties acknowledge that they have required that the present document, as well as all documents, notices and legal proceedings entered into,
given or instituted pursuant hereto or relating directly or indirectly hereto, be drawn up in English. Les parties reconnaissent avoir convenu que la
rédaction du présente document sera en Anglais, ainsi que tous documents, avis et procédures judiciaires exécutés, donnés ou intentés, à la suite
de ou en rapport, directement ou indirectement, avec les procédures concernées.
Graco Information
For the latest information about Graco products, visit www.graco.com.
TO PLACE AN ORDER, contact your Graco distributor or call to identify the nearest distributor.
Phone: 612-623-6921 or Toll Free: 1-800-328-0211 Fax: 612-378-3505
All written and visual data contained in this document reflects the latest product information available at the time of publication.
Graco reserves the right to make changes at any time without notice.
Original instructions. This manual contains English. MM 313882
Graco Headquarters: Minneapolis
International Offices: Belgium, China, Japan, Korea
GRACO INC. P.O. BOX 1441 MINNEAPOLIS, MN 55440-1441
Copyright 2010, Graco Inc. is registered to ISO 9001
www.graco.com
Revised June 2012
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.