Graco 313873H User Manual

Instructions
313873H
VRM
Hydraulic, Plural-Component, Variable-Ratio Proportioner. For pouring and dispensing sealants and adhesives. For professional use only.
Not approved for use in European explosive atmosphere locations.
Important Safety Instructions
Read all warnings and instructions in this manual. Save these instructions.
EN
ti17663a
Contents
Related Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Important Two-Component Material Information . 8
Isocyanate Conditions . . . . . . . . . . . . . . . . . . . . . 8
Material Self-ignition . . . . . . . . . . . . . . . . . . . . . . 8
Keep Components A (Red) and B (Blue) Separate 8
Moisture Sensitivity of Isocyanates . . . . . . . . . . . 8
Changing Materials . . . . . . . . . . . . . . . . . . . . . . . 9
A (Red) and B (Blue) Components . . . . . . . . . . . . . 9
Component Identification . . . . . . . . . . . . . . . . . . . 10
Hydraulic Power Pack . . . . . . . . . . . . . . . . . . . . 13
Motor Control Module (MCM) . . . . . . . . . . . . . . 14
Advanced Display Module (ADM) . . . . . . . . . . . 16
Fluid Control Module (FCM) . . . . . . . . . . . . . . . 19
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Initial Machine Setup . . . . . . . . . . . . . . . . . . . . . 20
Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Priming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
ADM Operation Overview . . . . . . . . . . . . . . . . . 26
Machine Operation Overview . . . . . . . . . . . . . . 28
System Setup and Calibration . . . . . . . . . . . . . . 29
Dispensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Flushing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Short-term . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
End of Shift . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Pressure Relief Procedure . . . . . . . . . . . . . . . . . . 42
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Install Upgrade Tokens . . . . . . . . . . . . . . . . . . . 44
Advanced Display Module (ADM) . . . . . . . . . . . 44
Motor Control Module (MCM) . . . . . . . . . . . . . . 45
Fluid Control Module (FCM) . . . . . . . . . . . . . . . 45
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Light Tower (Optional) . . . . . . . . . . . . . . . . . . . . 47
Common Problems . . . . . . . . . . . . . . . . . . . . . . 47
ADM Troubleshooting . . . . . . . . . . . . . . . . . . . . 49
Motor Control Module . . . . . . . . . . . . . . . . . . . . 50
Fluid Control Module . . . . . . . . . . . . . . . . . . . . . 52
Appendix A - ADM Icons Overview . . . . . . . . . . . 54
Setup Screen Icons . . . . . . . . . . . . . . . . . . . . . . 54
Home Screen Icons . . . . . . . . . . . . . . . . . . . . . 55
Appendix B - ADM Setup Screens Overview . . . 56
Appendix C - ADM Run Screens Overview . . . . . 62
Optional Screens . . . . . . . . . . . . . . . . . . . . . . . . 65
Appendix D - ADM Event and Error Codes Overview
66
Event Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Appendix E - USB Operation . . . . . . . . . . . . . . . . 77
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
USB Options . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Download Log Files . . . . . . . . . . . . . . . . . . . . . . 77
Log Files, Folder Structure . . . . . . . . . . . . . . . . 78
Transfer System Settings . . . . . . . . . . . . . . . . . . 80
Update Custom Language . . . . . . . . . . . . . . . . . 81
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Graco Standard Warranty . . . . . . . . . . . . . . . . . . . 84
2
313873H

Related Manuals

Manuals are available at www.graco.com.
Component manuals in English:
System Manuals
313874 VRM Repair-Parts
Power Distribution Box Manual
3A0239 Power Distribution Boxes Instruc-
tions-Parts
Pumpline Manuals
3A0022
3A0021 Vertical Hydraulic Driver
Feed System Manuals
U-Cup Dura-Flo
Repair-Parts
Lowers
Related Manuals
3A1159 VRM Feed Systems
Val ve Manuals
310550 1/2 in. NPT Fluid Port Ball Seat
Applicator
310551 3/4 in. NPT Fluid Port Ball Seat
Applicator
3A1792 DV Series Dispense Valves
Flow Meter Manual
309834 Helical Gear Fluid Flow Meter
313873H 3

Models

Models
Approximate
Full Load
System
approved
Peak Amps Per Phase*
Voltage (phase)
System Watts†
24F872 60 A 230 (3)
24F391
24F873
55 A 400 (3)
55 A 400 (3)
24,000 66 (30) 1.0 (3.8) 1.63 2000 (14, 138)
* Full load amps with all devices operating at maximum capabilities. Fuse requirements at various flow rates and
mix chamber sizes may be less.
** Values are dependent on installed pump size. Values shown are for largest available pump size.
Flow rate is independent of frequency 50/60 Hz.
‡ If a motor control module is replaced, see the Adjust Motor Control Module Selector Switch section in the
VRM repair manual to set the machine to the proper maximum fluid working pressure.
Max Flow Rate** lb/min (kg/min)
Output per Cycle (A+B)** gal. (liter)
Hydraulic Pressure Ratio**
Maximum Fluid Working Pressure ‡ psi (MPa, bar)
4 313873H

Warnings

Warnings
The following warnings are for the setup, use, grounding, maintenance, and repair of this equipment. The exclama­tion point symbol alerts you to a general warning and the hazard symbol refers to procedure-specific risk. Refer back to these warnings. Additional, product-specific warnings may be found throughout the body of this manual where applicable.
WARNING
ELECTRIC SHOCK HAZARD
This equipment must be grounded. Improper grounding, setup, or usage of the system can cause elec­tric shock.
Turn off and disconnect power at main switch before disconnecting any cables and before servicing equipment.
Connect only to grounded power source.
All electrical wiring must be done by a qualified electrician and comply with all local codes and reg­ulations.
TOXIC FLUID OR FUMES HAZARD
Toxic fluids or fumes can cause serious injury or death if splashed in the eyes or on skin, inhaled, or swallowed.
Read MSDSs to know the specific hazards of the fluids you are using.
Store hazardous fluid in approved containers, and dispose of it according to applicable guidelines.
Always wear chemically impermeable gloves when spraying, dispensing, or cleaning equipment.
PERSONAL PROTECTIVE EQUIPMENT
You must wear appropriate protective equipment when operating, servicing, or when in the operating area of the equipment to help protect you from serious injury, including eye injury, hearing loss, inhala­tion of toxic fumes, and burns. This equipment includes but is not limited to:
Protective eyewear, and hearing protection.
Respirators, protective clothing, and gloves as recommended by the fluid and solvent manufacturer.
SKIN INJECTION HAZARD
High-pressure fluid from dispensing device, hose leaks, or ruptured components will pierce skin. This may look like just a cut, but it is a serious injury that can result in amputation. Get immediate surgical
treatment.
Do not point dispensing device at anyone or at any part of the body.
Do not put your hand over the fluid outlet.
+
Do not stop or deflect leaks with your hand, body, glove, or rag.
Follow the Pressure Relief Procedure when you stop dispensing and before cleaning, checking, or servicing equipment.
Tighten all fluid connections before operating the equipment.
Check hoses and couplings daily. Replace worn or damaged parts immediately.
313873H 5
Warnings
WARNING
FIRE AND EXPLOSION HAZARD
Flammable fumes, such as solvent and paint fumes, in work area can ignite or explode. To help prevent fire and explosion:
Use equipment only in well ventilated area.
Eliminate all ignition sources; such as pilot lights, cigarettes, portable electric lamps, and plastic drop cloths (potential static arc).
Keep work area free of debris, including solvent, rags and gasoline.
Do not plug or unplug power cords, or turn power or light switches on or off when flammable fumes are present.
Ground all equipment in the work area. See Grounding instructions.
Use only grounded hoses.
Hold gun firmly to side of grounded pail when triggering into pail.
If there is static sparking or you feel a shock, stop operation immediately. Do not use equipment until you identify and correct the problem.
Keep a working fire extinguisher in the work area.
EQUIPMENT MISUSE HAZARD
Misuse can cause death or serious injury.
Do not operate the unit when fatigued or under the influence of drugs or alcohol.
Do not exceed the maximum working pressure or temperature rating of the lowest rated system component. See Technica l Data in all equipment manuals.
Use fluids and solvents that are compatible with equipment wetted parts. See Technical Data in all equipment manuals. Read fluid and solvent manufacturer’s warnings. For complete information about your material, request MSDS from distributor or retailer.
Do not leave the work area while equipment is energized or under pressure. Turn off all equipment and follow the Pressure Relief Procedure when equipment is not in use.
Check equipment daily. Repair or replace worn or damaged parts immediately with genuine manu­facturer’s replacement parts only.
Do not alter or modify equipment.
Use equipment only for its intended purpose. Call your distributor for information.
Route hoses and cables away from traffic areas, sharp edges, moving parts, and hot surfaces.
Do not kink or over bend hoses or use hoses to pull equipment.
Keep children and animals away from work area.
Comply with all applicable safety regulations.
6 313873H
Warnings
WARNING
MOVING PARTS HAZARD
Moving parts can pinch, cut or amputate fingers and other body parts.
Keep clear of moving parts.
Do not operate equipment with protective guards or covers removed.
Pressurized equipment can start without warning. Before checking, moving, or servicing equip­ment, follow the Pressure Relief Procedure and disconnect all power sources.
BURN HAZARD
Equipment surfaces and fluid that’s heated can become very hot during operation. To avoid severe burns:
• Do not touch hot fluid or equipment.
313873H 7

Important Two-Component Material Information

Important Two-Component Material Information

Isocyanate Conditions

Spraying or dispensing materials containing isocya­nates creates potentially harmful mists, vapors, and atomized particulates.
Read material manufacturer’s warnings and material MSDS to know specific hazards and precautions related to isocyanates.
Prevent inhalation of isocyanate mists, vapors, and atomized particulates by providing sufficient ventila­tion in the work area. If sufficient ventilation is not available, a supplied-air respirator is required for everyone in the work area.
To prevent contact with isocyanates, appropriate per­sonal protective equipment, including chemically impermeable gloves, boots, aprons, and goggles, is also required for everyone in the work area.

Material Self-ignition

Some materials may become self-igniting if applied too thickly. Read material manufacturer’s warnings and material MSDS.

Moisture Sensitivity of Isocyanates

Isocyanates (ISO) are catalysts used in two component foam and polyurea coatings. ISO will react with moisture (such as humidity) to form small, hard, abrasive crystals, which become suspended in the fluid. Eventually a film will form on the surface and the ISO will begin to gel, increasing in viscosity. If used, this partially cured ISO will reduce performance and the life of all wetted parts.
NOTE: The amount of film formation and rate of crystal­lization varies depending on the blend of ISO, the humidity, and the temperature.
To prevent exposing ISO to moisture:
Always use a sealed container with a desiccant dryer in the vent, or a nitrogen atmosphere. Never store ISO in an open container.
Keep the pump wet cups filled with IsoGuard
®
Select rier between the ISO and the atmosphere.
Use moisture-proof hoses specifically designed for ISO, such as those supplied with your system.
Never use reclaimed solvents, which may contain moisture. Always keep solvent containers closed when not in use.
Never use solvent on one side if it has been contam­inated from the other side.
Always lubricate threaded parts with ISO pump oil or grease when reassembling.
, part 24F516. The lubricant creates a bar-

Keep Components A (Red) and B(Blue) Separate

Cross-contamination can result in cured material in fluid lines which could cause serious injury or dam­age equipment. To prevent cross-contamination of the equipment’s wetted parts, never interchange component A (Red) and component B (Blue) parts.
8 313873H

Changing Materials

When changing materials, flush the equipment mul­tiple times to ensure it is thoroughly clean.
Always clean the fluid inlet strainers after flushing.
Check with your material manufacturer for chemical compatibility.
Most materials use ISO on the A (Red) side, but some use ISO on the B (Blue) side. See the follow­ing section.

A (Red) and B (Blue) Components

A (Red) and B (Blue) Components
IMPORTANT!
Material suppliers can vary in how they refer to plural component materials.
Be aware that when standing in front of the manifold on proportioner:
Component A (Red) is on the left side.
Component B (Blue) is on the right side.
NOTE: On the VRM system, A (Red) is on the left side of the base unit and B (Blue) is on the right side of the base unit. The sides switch when going to the dispense valve stand so that A (Red) is on the right side of the dis­pense valve stand and B (Blue) is on the left side of the dispense valve stand.
A (Red)
B (Blue)
For all machines:
The A (Red) side is intended for ISO, hardeners, and catalysts.
The B (Blue) side is intended for polyols, resins, and bases.
NOTE: For machines with material volume ratios other than 1:1, the higher volume side is typically the B (Blue) side.
B (Blue)
A (Red)
ti17663a
313873H 9

Component Identification

Component Identification
AA Advanced Display Module (see page 16) AB Hydraulic Power Pack AC Vertical Hydraulic Driver AD Pump Lower AE Flow Meter AF Power Distribution Box AG Dispense Stand AH Mixer
AE
AP
AN
AS
AA
AJ Pump Wet Cup AK Main Power Switch AL Air Supply Inlet AM Electrical Enclosure AN Ratio Check Dispense Valves/Ports AP Material Line AR A (Red) Dispense Valve AS B (Blue) Dispense Valve
AC
AB
AF
AR
AM
AH
ti17663a
AG
FIG. 1: Component Identification
10 313873H
AN
Component Identification
AN
AL
AK
AJ
AD
ti17664a
FIG. 2: Component Identification
313873H 11
Component Identification
Circuit Breakers
Most circuit breakers are located inside the power distri­bution box. The main block of circuit breakers in the power distribution box is shown below, with detailed information in the following table. See the power distri­bution box manual for more information.
CB1 CB5 CB10CB6
24C687_313873-5_1c
Size
Ref.
230V/
3 phase
400V/
3 phase
Component
CB1 30A 63A Motor Control Module
CB5 5A 5A Miscellaneous
CB6 5A 5A Miscellaneous
CB10 30A 63A Motor Control Module
12 313873H

Hydraulic Power Pack

Component Identification
F
B
257442_313873-5_1t
Shown with shroud removed
E
C
G
A
D
H
257442_313873-5_2t
FIG. 3
Key:
A 8 Gallon Hydraulic Oil Reservoir (see Technical Data on
page 83 for specifications) B Electric Motor C Dipstick D Hydraulic Housing
313873H 13
E Motor Control Module (see page 14) FFan G Hydraulic Oil Filter H Hydraulic Fluid Drain Port
Component Identification

Motor Control Module (MCM)

NOTICE
If the Motor Control Module is replaced, the selec­tor switch must be set prior to initial startup of the Motor Control Module or damage may occur. See system repair manual for MCM replacement proce­dure.
B
A
The MCM is located in the Hydraulic Power Pack.
When installed, the end of the MCM with the power input connection (12) faces down and the end with the access cover (A) faces up.
The Motor Control Module uses an 8-position selector switch to set the system maximum working pressure.
C
12
FIG. 4: MCM Component Identification
1A
10
1B
3
2
5
r_257396_3b9905_01b
6
8
9
11
7
13
r_257396_3b9905_03b
14 313873H
Ref Description
A Access Cover
B Module Status LEDs
C Warning Label
A (Red) MCM only:
ADM, Power Distribution Box
Component Identification
1A, 1B
B (Blue) MCM only:
FCM, Power Distribution Box
NOTE: 1A and 1B are interchangeable.
2 Three-way Splitter to:
Oil Low Level Sensor, Dispense Valve Solenoid, Footswitch
3 Oil Temperature Sensor
5 Electric Motor Temperature Sensor
6LVDT
7 A (Red) MCM only:
Three-way Splitter to: Hydraulic Directional Valve, Oil Overtemperature Switch, Ratio Check Solenoid Valve
B (Blue) MCM only:
Three-way Splitter to: Hydraulic Directional Valve, Oil Overtemperature Switch (Third connection is not used)
8 Pressure Transducer (for material side
controlled by the MCM)
9 Not used
10 MCM to MCM Analog Connection
11 Motor Position Sensor
12 MCM Power Input Connection
13 Motor Power Connection
313873H 15
Component Identification

Advanced Display Module (ADM)

User Interface
A
B
C
D
E
H
G
FIG. 5: ADM Component Identification - Front
Buttons
Call out Button Function
AADM
Enable/ Disable
B System
Status Indicator LED
C Stop Stop all system processes
Enable/disable ADM
Displays system status
F
TI12362a1
Call out Button Function
D Softkeys Defined by icon next to softkey
E Abort Abort current operation
F Enter Accept change, acknowledge error,
select item, toggle selected item
GRun/
Setup
Toggle between Run and Setup
screens Screens Toggle
HArrow
Keys
Navigate within a screen or to a new
screen
16 313873H
Component Identification
J
S
K
L
M
N
R
F
IG. 6: ADM Component Identification - Rear
P
ti12902a
Key:
J Flat Panel Mount K Model Number Identification Label L USB Interface (see Appendix E - USB Operation
beginning on page 77)
M CAN Cable Connection to MCM
N Module Status LEDs P Accessory Cable Connection R Software Token Access Cover S Battery Access Cover
See ADM Troubleshooting on page 49 for LED status information.
See Maintenance section beginning on page 43 for battery replacement and software update procedures.
313873H 17
Component Identification
ADM Screen Components
See the ADM appendix sections beginning with Appendix A - ADM Icons Overview on page 54 for more informa- tion.
Current date and time Current screen
Previous screen Next screen
Operating Mode
Enter/Exit screen
Settings Detail
F
IG. 7: Main Display Components - Typical Setup Screen
Faults, Status
Previous screen no.
Current screen no.
Next screen no.
Current date and time Current screen
Previous screen Next screen
Operating Mode
Change Operating
Mode
Dispense Details
F
IG. 8: Main Display Components - Home Screen (Shot mode shown)
Faults, Status
Initiate Dispense
Initiate Ratio Check Dispense
18 313873H

Fluid Control Module (FCM)

The fluid control module is located inside the electrical enclosure.
Component Identification
A
C
ti12337a1
D
FIG. 9:
Key:
A Module Connection Screws B Access Cover C Module Status LEDs D CAN Connectors (one is connected to MCM, other is
unused)
ti12336a1
B
313873H 19

Setup

Setup

Initial Machine Setup

Perform this setup procedure to prepare the machine for initial operation.
The machine is not properly grounded until this setup procedure is performed. To prevent risk of electric shock, do not start the machine until this setup proce­dure is completed.
1. Locate the machine.
NOTICE
Be careful not to hit hydraulic power pack drain port while moving machine. Applying significant force to drain port may damage the hydraulic tank.
NOTICE
To prevent machine damage, do not expose system to rain.
a. Bolt machine to original shipping pallet before
lifting.
Electrical Cord Requirements
NOTE: Power cord is not supplied. See the following
table.
Table 1: Power Cord Requirements
Cord Requirements
Model
230V, 3 phase 8 (8.4), 3 wire + ground
400V, 3 phase 6 (13.3), 4 wire + ground †
† Residual Current Device (RCD) must be rated at
300 mA if installed.
Electrical Cord Wires by Model
230V, 3 phase: L1, L2, L3, GND 400V, 3 phase: L1, L2, L3, N, GND
Electrical Requirements
See Models on page 4 for detailed electrical require­ments information.
a. Use 5/32 in. or 4 mm hex allen wrench to con-
nect the power leads to L1, L2, L3, and N as applicable.
b. Connect green wire to ground (GND).
AWG (mm
2
)
b. Locate the machine on a level surface. See
Dimensions on page 83 for space require­ments.
L1L2L3N
2. Connect electrical cord.
GND
Installing this equipment requires access to parts
F
which may cause electric shock or other serious injury if work is not performed properly. Have a qualified electrician connect power and ground to main power switch terminals. All electrical wiring must be done by a qualified electrician and comply with all local codes and regulations.
20 313873H
IG. 10: 400V, 3 phase shown
r_24C686_313998_1a
Setup
Power Line Voltage Surges
Power conversion equipment can be sensitive to voltage fluctuations on incoming power. The Motor Control Mod­ule falls under the category of power conversion equip­ment because energy is stored on a capacitive bus and then modulated to control a brushless motor. Engi­neered design takes this into account and withstands a wide range of conditions, but it is possible for supplied power to occasionally fall outside the tolerable range in industrial plants with high-amperage reactive pulsed loads such as welding equipment. If the tolerable range is exceeded, an overvoltage condition is flagged and the system will shut down in an alarm state to protect itself and alert the user of unstable power. Excessive or repeated overvoltage may permanently damage hard­ware.
The MAX-HOLD feature on a multimeter can be used to determine peak DC voltage on the line. DC is the proper setting, as opposed to AC, because peak voltage is the critical parameter that affects the DC voltage level stored on the capacitive bus in power conversion equip­ment. Reading should not regularly exceed approxi­mately 400VDC to avoid tripping the 420VDC alarm level in the Motor Control Module. If power quality is sus­pect, power conditioning or isolation of the device(s) causing poor power quality is recommended. Consult a qualified electrician if there are any concerns about the available power supply.
The chart below shows the permissible magnitude and duration of temporary over-voltage events:
Maximum Permissible Transient Voltage Surges
* Constructed from ITIC 1996 curve, referenced by IEC 61000-2-4
1400
1200
1200Vac, 1697Vdc
1000
800
600
400
Voltage (Volts RMS)
200
0
0.000001 0.00001 0.0001 0.001 0.01 0.1 1 10
<--1 MW Max Surge Power
480Vac, 679Vdc
<--150 KW Max Surge Power
336Vac, 475Vdc
Time (seconds)
<--50 KW Max Surge Power
288Vac, 407Vdc
<--No Power Limit
264Vac, 373Vdc
3. Ground the system.
This equipment must be grounded.
a. VRM: grounded through power cord. See Con-
nect Electrical Cord, step #2 on page 20.
Power Line Test Steps with Multimeter
i. Set multimeter to “DC voltage”.
ii. Connect multimeter probes to supplied power line.
iii. Press “Min Max” successively to show the peak
positive and negative DC voltages.
iv. Confirm readings do not exceed 400VDC (Motor
Control Module alarm issued at 420VDC).
b. Dispense Valve: follow your local code.
c. Fluid supply containers: follow your local code.
d. Dispensing target/container: follow your local
code.
e. Solvent pails used when flushing: follow your
local code. Use only metal pails, which are con­ductive, placed on a grounded surface. Do not place pail on a nonconductive surface, such as paper or cardboard, which interrupts grounding continuity.
f. To maintain grounding continuity when flushing
or relieving pressure, hold grounded metal pail firmly to a metal part of dispense valve then ini­tiate a dispense.
313873H 21
Setup
4. Install feed system.
NOTE: This step only connects the feed system. Do not
allow fluid to flow into the system in this step.
Supply hoses from feed tank should be 2 in. (51 mm) ID minimum.
Use at least 10 layers of PTFE tape and use pipe dope on the fluid inlet fitting to prevent galling.
NOTICE
a. Close inlet ball valve.
b. Connect and tighten component B (Blue) supply
hose to the 2 in. npt(f) fitting on the component B (Blue) fluid inlet at the base of the pump lower.
c. Repeat previous steps for A (Red) material side.
5. Connect system base to dispense stand.
a. Connect fluid hoses.
7. Check hydraulic fluid levels.
The hydraulic reservoirs are filled at the factory. See Technical Data on page 83 for hydraulic fluid specifica­tions.
8. Fill pump wet cups 2/3 full with IsoGuard Select Fluid.
ti14725a
b. Connect electrical wires.
c. Connect ground wire.
6. Connect 1/2 in. air supply to dis­pense stand air inlet.
NOTICE
To prevent machine damage, air supply must be fil­tered and dried.
ti17664a1
9. Perform Startup, page 24.
10.Perform ADM setup.
NOTE: All ADM setup items are located in the Setup screens. See Appendix B - ADM Setup Screens Over­view beginning on page 56 for more information.
See ADM Operation Overview on page 26 for help with operating the ADM including how to modify settings.
a. Navigate to Advanced #1 screen then set gen-
eral system settings. See page 60.
b. Navigate to Advanced #2 screen then set units
of measure. See page 60.
c. Navigate to Advanced #3 screen then
enable/disable system features. See page 61.
22 313873H
d. Navigate to System #1 screen then define con-
trol mode, dispense mode, and pump informa­tion. See page 58.
e. Navigate to System #2 screen then define dis-
pense valve details and other system settings. See page 58.
f. Navigate to System #3 screen then define
labels, pressure imbalance, and flow meter details. See page 59.
g. Navigate to Shots screen then define shots.
See page 56.
h. Navigate to Sequences screen then define
sequences. See page 57.
i. If desired, navigate to Maintenance screen
then reset counters. See page 59.
Setup
11.Perform Flushing procedure, page 40.
NOTICE
The machine is tested with oil at the factory. Flush out the oil with a compatible solvent before loading the machine with material. See Flushing on page 40.
12.If not already open, open the fluid inlet ball valves to load the system with material.
13.Perform System Setup and Cali­bration procedure, page 29.
313873H 23

Startup

Startup
1. Perform all required maintenance tasks. See Main­tenance on page 43.
2. Check for leaks.
3. Check hydraulic fluid levels.
4. Check pump wet cup fluid levels.
5. Check feed system fluid levels.
6. Turn Main Power Switch to the ON position. The splash screen will be displayed on the ADM until it is finished loading.
7. When the ADM is finished loading, press to
enable the ADM. The System Status Indicator Light
next to will illuminate green.
8. Press repeatedly to select a different operating
mode then press to accept.
NOTE: The Setup screens cannot be accessed when Disabled mode is the active operating mode. Also, cer­tain machine functions and setup changes are disabled when Standby mode is selected.
24 313873H

Priming

Priming
1. Place waste containers below both ratio check dis­pense valves.
NOTE: Both ratio check dispense valves will be open when dispensing in Prime Mode. Only one pump moves but material may drip from the other ratio check dis­pense valve when opened.
2. Press repeatedly to select Prime Mode then
press to accept.
3. Check the dispense settings shown at the bottom of the screen.
7. Continue dispensing until clean, air-free material is
dispensed from both sides then press to stop
dispensing.
8. Press to select the B (Blue) side.
9. Press to begin dispensing B (Blue) material.
10. Continue dispensing until clean, air-free material is
dispensed from both sides then press to stop
dispensing.
4. If desired, change the dispense settings.
a. Press to enter editing mode.
b. Use the left and right arrow keys to select the
item to change.
c. Use the numeric keypad to type the new value.
d. Press to accept the new value.
e. Press to exit editing mode.
5. Press to select the A (Red) side.
6. Press to begin dispensing A (Red) material.
313873H 25

Operation

Operation

ADM Operation Overview

ADM Navigation Diagram
Run Screens
Shot #1
Shot #2
Shot #...
Home
Password
(if enabled)
Setup Screens
Sequences #1
Sequences #2
Sequences #...
Entry
Errors #1
Errors #2
Errors #...
Calibration
Events #1
Events #2
Events #...
System #1
System #2
System #3
Maintenance
Maintenance
Optional:
Diagnostic
Supply
Conditioning #1
Conditioning #2
Conditioning #3
Advanced #1
Advanced #2
Advanced #...
26 313873H
Operation
Navigation Overview
For all ADM screens to be accessible and functional, the ADM must be enabled and an operating mode other than Standby or Disabled must be selected.
NOTE: The optional Diagnostic screen can be enabled from Advanced #3 screen, see page 61.
To navigate between screens use the arrow keys on the ADM keypad. To access the Setup screens,
press . If the Setup screens password is turned on, use the ADM keypad to enter the password then
press . For Setup screens information, see Appen- dix B - ADM Setup Screens Overview on page 56. For Run screens information, see Appendix C - ADM Run
Screens Overview on page 62.
Change ADM Values
To edit information in a screen, such as a shot definition or a system setting like time or date format, follow this general process:
1. Press .
NOTE: Operator mode flow settings are edited using a slightly different process. To edit dispensing settings when in Operator mode, see Dispense in Operator Mode section on page 39.
2. Once in the screen, use the arrow keys to navigate to the desired item.
3. Edit the value:
If the item has a drop-down list to select
from, press to display the dropdown list. Use the up and down arrow keys to highlight the
desired item then press to select the item.
If the item is a numeric value, use the numeric
keypad on the ADM to type the new value then
press to accept the value.
If the item is a select/deselect or enable/dis-
able checkbox option, press to toggle the value.
If a screen has been entered by pressing or if the system is in editing mode then navigating to a different screen will be disabled. As applicable, exit the screen and editing mode to re-enable screen navigation.
If necessary, press to cancel editing.
4. Press .
313873H 27
Operation

Machine Operation Overview

Ramp Up Feature
The purpose of the ramp up feature is to enable dis­pensing at the correct ratio but at a reduced flow rate when materials are too thick to dispense at the correct flow rate. As the fluid warms up due to the friction of moving through the system, the Ramp Up feature will slowly increase the flow rate until the desired flow rate is achieved. After a period of idle time the system will cool down and the fluids will not be as warm while dispens­ing, which may result in the Ramp Up feature being acti­vated.
While the system is dispensing, the ramp up feature monitors the torque supplied to the B (Blue) pump to verify it does not apply more torque than the pump can handle. If it does, it will reduce the flow rate in both pumps to maintain the required dispense ratio. As the system warms up and material thickness decreases, the ramp up feature will increase the flow rate until the desired flow rate is achieved.
NOTE: Because the B (Blue) side is the high volume side, it is closer to its maximum flow rate capacity than the A (Red) side pump.
Learning Mode
When a flow rate or dispense ratio is requested that the machine has not learned, the system will use the pump volume of each pump to estimate the pump velocities needed then Learn mode will be used to adjust them to the correct flow rates. The system will begin dispensing at the estimated pump velocity and each stroke per­formed will be used to gather information and adjust the velocity. After a number of strokes, the system will have sufficient data to accurately dispense at the desired set­point and Learning mode will be exited.
When Learn mode is active, a “Learning New Set Point” advisory is generated and the yellow light on the light tower is illuminated. After Learn mode is complete, the advisory is cleared.
When the ramp up feature reduces the flow rate, a “Sys­tem Dispensing Below Requested Set Point” advisory is generated and the yellow advisory lamp on the light tower is illuminated. If the system is able to achieve the desired flow rate the advisory is cleared.
28 313873H
Operation

System Setup and Calibration

Perform this entire procedure if any of the following con­ditions are met:
The machine is new
One or both materials in the system have changed since last performing this procedure
Flow rate, ratio, or ambient temperature has changed
If the software has been updated, verify all software set­tings in the first section of this procedure are still correct. If any incorrect software setting is found, perform this entire setup and calibration procedure.
Software Settings
1. With the machine on, press to enable the ADM. The LED next to the button should be green.
2. Press repeatedly to select Standby mode then
press to accept.
7. Navigate to the System 2 screen.
8. Select a base purge flow rate. A value of approxi­mately 100 g/s or 100 cc/s is recommended.
9. If installed, check the “Supply Low Level Sensor” option.
10. Verify the correct Dispense Valve type is selected.
11. Navigate to the System 3 screen.
3. Press to enter the Setup screens.
4. Navigate to the System 1 screen.
5. Verify the correct pumps and pump sizes are selected. Most systems use Dura-Flo 580 or 430 pumps.
6. Select volume or weight for the dispense mode. Weight mode is recommended because it is easier to calibrate.
313873H 29
Operation
12. Select the flow meter types installed on your sys­tem. Most systems use “HG6000” flow meters for both sides. If no flow meters are installed, select the “Disable” option as indicated below and proceed to step 14.
16. Press .
17. Enter the specific gravities for the two materials in the system.
13. Set the Ratio “Alarm %” to 0 to turn off ratio alarms and set the Ratio “Deviation %” to any number greater than or equal to 5%.
NOTE: The “Alarm %” can be turned back on after this setup and calibration procedure is completed.
14. Set the “Pressure Imbalance Alarm” to 2000 psi (137.9 bar, 13.8 MPa).
15. Navigate to the main Calibration screen.
NOTE: The specific gravities do not need to be exact but should be close.
18. Press to exit the Setup screens.
Prime the Machine
Refer to Priming section on page 25.
Piston Position Learning
19. Navigate to the Calibration screen.
20. Perform Learn Mode.
NOTE: Learn Mode will teach the system the mechani­cal limits of piston travel. It must be performed whenever the pump line is rebuilt or if any other maintenance is performed that may affect the mechanical tolerances in the pump line. If the machine does not appear to be uti­lizing the full extent of the pump stroke, or if the machine appears to be contacting the end of the hydraulic cylin­der, perform the Learn Mode procedure.
a. From the Calibration screen, press to
access the Learn Mode screen.
30 313873H
Operation
b. Place a waste container below the ratio check
valves. The next steps will cause the machine to dispense material.
c. Press then . The pump will travel to
bottom-most position.
d. After the pump stops moving, press then
press . The pump will travel to the top-most position.
NOTE: During this process, the system learned the mechanical limits of piston travel. If the pump did not reach either piston mechanical travel limit for any rea­son, repeat the procedure.
Flow Rate and Ratio Learning
25. At the main Calibration screen, press to erase
all learned data.
NOTE: This will not affect the weight calibration that was just completed.
26. Press to exit the Setup screens.
21. Press to exit the Setup screens.
22. Press repeatedly to select Operator mode
then press to accept.
23. Press to enter the Setup screens then press
left or right to navigate to the main Calibration screen.
Flow Meter / No Flow Meter Calibration
24. If flow meters are installed, refer to Flow Meter or Flow/Ratio Calibration starting on page 34. If flow meters are not installed, refer to No Flow Meter Machine Calibration starting on page 35.
27. Press repeatedly to select Shot mode then
press to accept.
28. Select a defined shot that will provide a 10 second or longer dispense at the flow rate and ratio intended to be used during normal system opera­tion.
NOTE: Changes to the ambient temperature will affect the maximum flow rate for the system. If the ambient temperature decreases, the flow rate should be decreased.
313873H 31
Operation
29. On the main run screen, verify the ratio check valve
button is not active.
NOTE: This verifies material dispenses through the static mixer.
30. Place bucket under the end of the static mixer.
31. Press to begin dispensing then write down the A (Red) and B (Blue) dispense pressures shown on the ADM.
NOTE: During the dispense an off-ratio deviation may be generated and that is ok.
32. Repeat the previous step until the “System Learning New Setpoint” advisory turns off and the light tower yellow lights turns from yellow to green.
NOTE: Continue base purge until clean material comes out of the end of the mixer.
d. When all mixed material is pushed out of the
mixer, press to stop dispensing.
34. Press repeatedly to select Shot Mode then
press to accept.
35. Press to activate the ratio check valves.
NOTE: Be ready to adjust ratio check valves immedi­ately after performing the following step.
36. With buckets below the ratio check valves,
press to begin dispensing.
37. While dispensing, adjust the ratio check opening adjustment screws until both material line pressures are approximately equal to the pressures recorded in step 31.
NOTE: If adjusting the ratio check opening screws after a ratio check dispense, the pressure difference due to the adjustment will not be shown until the next dispense.
NOTE: After the ratio check dispense pressures are properly adjusted a ratio check dispense can be per­formed. The ratio check dispense should be at least 10 seconds.
33. Base purge the mixer to clear mixer of mixed mate-
38. If the pressures are correctly adjusted prior to com-
rial:
pleting the shot, press to stop dispensing.
NOTE: A base purge will dispense only the B (Blue) material to push all mixed material out of the mixer. Base purge settings are defined on the System #2
39. If the pressures were not correctly adjusted prior to the shot finishing, go to step 36 to repeat.
screen, see page 58.
Ratio Check
a. Press repeatedly to select Standby Mode
NOTE: In the following steps, the weight of the dis­pensed materials is used to calibrate the flow meters.
then press to accept.
This works regardless of whether the selected dispense mode is weight or volume.
b. Press .
40. With active and with pre-weighed buckets
c. Press to begin dispensing.
below the ratio check valves, press to begin dispensing a ratio check dispense.
32 313873H
41. Weigh the two buckets and use the net weight of each dispensed material to calculate the actual ratio of the dispensed material.
42. If the calculated ratio of the weighed materials does not match the ratios displayed on the ADM, go to step 1 to re-calibrate the flow meters.
43. If the calculated ratio of the weighed materials matches the ratio displayed on the ADM, then navi­gate to the System 3 screen and change the ratio alarm percentage to the desired percentage.
44. If at any point in the future the ratio, flow rate, or ambient temperature changes from what was used while performing this procedure, go to step 1.
NOTE: If the ratio or flow rate is changed to a ratio or flow rate that has not been calibrated by performing this procedure, the system will generate a “Learning New Setpoint” advisory. The system usually produces a good dispense ratio during the learning process however the advisory is generated to inform the user of the condition. The system can store calibration data in its memory for up to five different flow rates and ratios.
Operation
NOTE: If the ambient temperature changes significantly from the ambient temperature seen while performing this calibration procedure, the system will need to “learn” the new temperature and the flow rate may need to be decreased. The machine will not automatically be aware of the temperature change but it will try to find the correct dispensing properties to compensate for the new temperature. If the ambient temperature changes signifi­cantly, go to step 1.
313873H 33
Operation
Flow Meter or Flow/Ratio Calibration
1. At the main Calibration screen, press to erase
any previously learned flow meter calibration data.
NOTE: At this point, the “Learning New Set Point” advi­sory will be generated.
2. Select to enter the flow meter calibration
screen.
NOTE: After the weights are entered the K-factor will be shown to the right of the weights. The previous K-factor is shown to the left of the weight entry fields.
9. Repeat steps 5-8 until the new K-factor shown is within 1% of the previous K-factor.
10. Select to exit the flow meter calibration screen.
3. On the flow meter calibration screen, enter the flow rate and ratio that will be used during normal opera­tion.
4. If either K-Factor value is 0, enter 3000 if an HG6000 flow meter is installed in that side.
NOTE: In the following steps, the weight of the dis­pensed materials is used to calibrate the flow meters. This works regardless of whether the selected dispense mode is weight or volume.
5. Weigh two buckets and record the weight of each then place below the ratio check valves.
6. With two buckets in place to catch material dis-
pensed from the ratio check valves, press to begin dispensing.
7. After dispensing for at least 10 seconds, press to stop dispensing.
NOTE: If available, a footswitch can also be used.
8. Weigh both buckets and enter the net weight of each material dispensed in the last two fields pro­vided on the screen.
34 313873H
Operation
No Flow Meter Machine Calibration
It is highly recommended that the user operate the machine in weight mode when flow meters are not installed or have been disabled.
NOTE: In weight mode, the ratio displayed is a weight ratio and should not be considered as volumetric ratio.
1. Press to enter the Setup screens then press
left or right to navigate to Advanced #3 screen. Turn off the “Enable Ratio Check Weight Mode Entry” option.
3. Press to exit the setup screen and verify the
ratio check option is not selected.
4. Place a waste container under the mixer.
Press to begin dispensing and record the aver­age A (Red) and B (Blue) pressures shown on the ADM.
NOTE: The dispense can be aborted early after record-
2. Press and navigate to the shot definition screen. Define a shot which is 10 times larger than
the desired flow rate selected. This will set approxi­mately a 10 second dispense time.
Example: If the dispense rate is 300 grams/second, set the amount to 3000 grams.
ing by pressing .
5. Select Standby mode and perform a base purge.
NOTE: A base purge will dispense only the B (Blue) material to push all mixed material out of the mixer. Base purge settings are defined on the System #2 screen, see page 58.
a. Press .
b. Press to begin dispensing.
c. When all mixed material is pushed out of the
mixer, press to stop dispensing.
NOTE: Approximately 1 to 2 liters of base material will need to be dispensed.
6. Select Shot mode and select the ratio check option
by pressing .
313873H 35
Operation
7. Place waste containers below the ratio check noz­zles and start a ratio check dispense. Adjust the ratio check opening screws until the pressures dis­played are near the values previously recorded when dispensing through the mixer (step 4).
8. Press to enter the Setup screens then press
left or right to navigate to Advanced #3 screen. Turn on the “Enable Ratio Check Weight Mode Entry” option.
NOTE: After entering the net weight of B (Blue) material, the ADM will inform the user how close the respective flow of the pump was to the desire rate.
11. Press to accept the information.
NOTE: The ADM will respond by generating a “System Learning Setpoint” advisory and the corresponding advi­sory light will be illuminated on the machine light tower (if installed).
9. Press to exit the setup screen and verify the
ratio check option is selected.
10. With new waste containers below the ratio check nozzles, start the dispense by pressing the foot-
switch or . At the end of the dispense, enter the A (Red) and B (Blue) material weights into the prompt boxes. Enter the weight of each bucket for both materials.
12. Press to accept the advisory. Repeat steps 10 through 11 until the percentages approach zero and the advisory is removed.
NOTE: The machine will be calibrated for the flow and ratio selected once the advisory is removed.
13. If the user needs to operate at a second flow or ratio, repeat the calibration process for the second desired flow or ratio.
NOTE: The machine will store the necessary control data for both calibration points.
36 313873H
NOTE: It is recommended that the user calibrate the machine at the extreme rates where it will be used. Example: If the machine is to be used at mixed flows between 300 & 500 grams/second (at 3:1 ratio), cali­brate the machine at the two 300 and 500 grams/ sec­ond extremes, then stop the calibration process. The machine will be very close to all requested flows and ratios in between.
14. Turn off the “Enable Ratio Check Weight Mode Entry” option in the Advanced #3 screen. The user can verify the machine calibration by performing a ratio check dispense and weighing the A (Red) and B (Blue) materials.
Operation
313873H 37
Operation

Dispensing

Dispense in Shot Mode
To dispense in Shot mode, at least one shot number must be defined. Shots are defined on the Shots screen, see page 56.
1. Navigate to the Home screen.
2. Press repeatedly to select Shot Mode then
press to accept.
3. If desired, change the selected shot.
a. Press .
b. Use numeric keypad to type the desired shot
number.
c. Press to accept. The shot number defini-
tion details will be shown on the bottom of the screen.
NOTE: Only defined shot numbers can be entered. If an undefined shot number is entered, it will be ignored.
Dispense in Sequence Mode
Sequences can be defined on the Sequences screen. Sequences are defined on the Sequences screen, see page 57.
1. Navigate to the Home screen.
2. Press repeatedly to select Sequence Mode
then press to accept.
3. If desired, changed the selected sequence.
a. Press once.
b. Press the right arrow key on the ADM keypad
once.
c. Use the up and down arrow keys to select a
sequence.
d. Press to accept.
4. If desired, press to skip to the next defined
shot position in the sequence. Repeat as desired. To go to the first defined position in the sequence,
4. Press to begin dispensing the active shot. To
abort the shot at any time, press or . The
shot will continue until the predefined amount has been dispensed.
5. Check the ADM for errors and pop-up notifications
that could indicate a faulty dispense. Press to acknowledge any displayed errors.
press .
5. Press to begin dispensing the active shot.
NOTE: To abort the shot at any time, press or
. If the shot is not aborted, material will continue to
dispense until the predefined amount has been dis­pensed. The next position in the Sequence will automat­ically be selected upon completion of the shot.
6. Check the ADM for errors and pop-up notifications
that could indicate a faulty dispense. Press to acknowledge any displayed errors.
38 313873H
Operation
Dispense in Operator Mode
Operator Mode begins dispensing when is pressed and stops when it is pressed again.
NOTE: If a footswitch is used, press and hold to dis­pense. Release to stop dispensing.
1. Navigate to the Home screen.
2. Press repeatedly to select Operator Mode
then press to accept.
3. Check the dispense settings shown at the bottom of the screen.
7. Check the ADM for errors and pop-up notifications
that could indicate a faulty dispense. Press to acknowledge any displayed errors.
4. If desired, change the dispense settings.
a. Press to enter editing mode.
b. Use the left and right arrow keys to select the
item to change.
c. Use the numeric keypad to type the new value.
d. Press to accept the new value.
e. Press to exit editing mode.
NOTE: The user can disable changing the flow, ratio, or both on the Advanced #3 setup screen.
5. Press to begin dispensing.
6. Press to stop dispensing.
313873H 39

Flushing

Flushing
Flush equipment only in a well-ventilated area. Do not dispense flammable fluids. Do not turn on heaters while flushing with flammable solvents.
Flush out old fluid with new fluid, or flush out old fluid with a compatible solvent before introducing new fluid. All fluid components are compatible with common sol­vents. Use only moisture-free solvents. See Technica l Data on page 83 for list of wetted components to verify compatibility of solvent with wetted materials. See sol­vent manufacturers information for material compatibil­ity. To prevent moisture from reacting with isocyanate, always leave the system dry or filled with a moisture-free plasticizer or oil. Do not use water. See Important Two-Component Material Information on page 8.
Grounding the solvent pails used when flushing: follow your local code. Use only metal pails, which are conduc­tive, placed on a grounded surface. Do not place pail on a nonconductive surface, such as paper or cardboard, which interrupts grounding continuity.
9. To maintain grounding continuity when flushing or relieving pressure, hold a metal part of dispense valve firmly to the side of a grounded metal pail,
then press . Continue to dispense until the sys-
tem is thoroughly flushed then press to stop
dispensing.
10. Close the solvent flush ball valve.
11. Disconnect solvent flush feed system.
12. Open the feed system ball valve.
13. Perform Priming procedure, page 25.
If flushing with a compatible solvent, perform the follow­ing procedure.
1. Perform Shutdown procedure, page 41.
2. Close the feed system ball valve at inlet near the pump lower.
3. Connect solvent flush feed system to unused inlet port near the pump lower.
4. Open solvent flush ball valve.
5. Perform Startup procedure, page 24.
6. Press repeatedly to select Operator Mode
then press to accept.
7. Press to enter editing mode.
8. Navigate to the flow rate value, change the value to
50-75% of the maximum flow rate, then press to accept.
40 313873H

Shutdown

Shutdown

Short-term

1. Place container under mixer.
2. If using a moisture-sensitive material, park pumps.
a. Navigate to the Home screen.
b. Press repeatedly to select Standby Mode
then press to accept.
c. Press to park pump. Material will dis-
pense. when the pumps are in the parked posi­tion, they will stop moving.
3. Allow material to drain completely from the mixer prior to base purge.
4. Perform base purge.
NOTE: A base purge will dispense only the B (Blue) material to push all mixed material out of the mixer. Base purge settings are defined on the System #2 screen, see page 58.
7. Place container under the mixer and allow mixer to drain completely.
NOTICE
Preventing material from draining from the mixer may cause material in the mixer to harden and dam­age the dispense block.
8. Turn Main Power Switch to the OFF position.

End of Shift

1. Perform Short-term Shutdown procedure.
2. Remove, disassemble, and flush mixer.
a. Press .
b. Press to begin dispensing.
c. When all mixed material is pushed out of the
mixer, press to stop dispensing.
NOTE: Approximately 1 to 2 liters of base material will need to be dispensed.
5. Press to park pumps again.
6. Press .
313873H 41

Pressure Relief Procedure

Pressure Relief Procedure
1. Perform Shutdown procedure.
2. Place a waste container below each ratio check valve.
In the following step, any pressure in the lines will be instantly relieved which may lead to material spraying out of the valve and splashing in the bucket. Use appropriate protective wear to prevent contact with materials.
3. Press the red override button on top of the pneu­matic valve nearest the mixer. This will open the ratio check valves and relieve any residual pressure in the fluid lines.
ti17666a
42 313873H

Maintenance

Check all sub-component manuals for maintenance schedule and procedures.
Task Schedule
Replace hydraulic oil and filter See table
Inspect fluid lines for leaks and signs of wear
Check wet cup fluid level, add IsoGuard Select fluid as neces­sary
Check hydraulic fluid level Weekly
Verify operation of tank air drying system to prevent isocyanate crystallization
Verify vent holes on bottom of hydraulic power pack shroud are clear and unobstructed
Daily
Weekly
Weekly
Weekly (more often in dusty environ-
ments)
Maintenance
Change Hydraulic Oil and Filter
Change break-in oil in a new unit after the first 250 hours of operation or within 3 months, whichever comes first. After initial break-in, see the following table for rec­ommended oil and filter change schedule.
Table 2: Frequency of Oil Changes
Ambient Tem perature
0 to 90°F (-17 to 32°C)
90°F and above (32°C and above)
Recommended Frequency
1000 hours or 12 months, whichever comes first
500 hours or 6 months, whichever comes first
Check Hydraulic Fluid Level
Check hydraulic fluid level on dipstick (A). Fluid level must be between indent marks (B) on dipstick. Refill as required with approved hydraulic fluid; see Technical Data on page 83. If fluid is dark in color, change fluid and filter.
Check all fittings and connec­tions, tighten as necessary
Use compressed air to remove dust buildup on control boards, fan, motor (under shield), hydrau­lic oil coolers, and component heat sink fins
As necessary
Monthly
A
S
B
ti7861a1
257442_313873-5_2t
313873H 43
Maintenance

Install Upgrade Tokens

NOTE: The Motor Control Module, Fluid Control Mod-
ule, and Temperature Control Module connection to the system is temporarily disabled during the installation of upgrade tokens.
To install software upgrades:
1. Use correct software token stated in the table. See Graco Control Architecture
manual for instructions.
NOTE: Upgrade all modules in the system to the software version on the token, even if you are replacing only one or two modules. Different soft­ware versions may not be compatible.
All data in the module (System Settings, USB Logs, Recipes, Maintenance Counters) may be reset to factory default settings. Download all settings and user preferences to a USB before the upgrade, for ease of restoring them following the upgrade.
See manuals for locations of specific GCA compo­nents.
Module Programming

Advanced Display Module (ADM)

Replace Battery
A lithium battery maintains the ADM clock when power is not connected.
To replace the battery:
1. Disconnect power to the ADM.
2. Remove rear access panel.
The software version history for each system can be viewed in the technical support section at www.graco.com.
Token Application
16G407 Ratio Monitoring (Flow Meters):
- Fluid Control Module
16G365 VRM:
- Advanced Display Module
- Motor Control Module
- Communication Gateway Module
r_257396_3b9905_04b
F
IG. 11: Remove Access Cover
ti12334a1
ti12903a
3. Remove the old battery and replace with a new CR2032 battery.
4. Replace rear access panel.
Install Upgrade Token
See Install Upgrade Tokens on page 44.
Cleaning
Use any alcohol-based household cleaner, such as glass cleaner, to clean the ADM. Spray on the rag then wipe ADM. Do not directly spray the ADM.
44 313873H

Motor Control Module (MCM)

Keep heat sink fins clean at all times. Clean them using compressed air.
NOTE: Do not use conductive cleaning solvents on the module.
Heat Sink Fins
Maintenance
r_257396_3b9905_02b
FIG. 12: Clean Heat Sink Fins
Install Upgrade Token
See Install Upgrade Tokens on page 44.

Fluid Control Module (FCM)

Install Upgrade and Key Tokens
See Install Upgrade Tokens on page 44.
313873H 45
Maintenance
46 313873H

Troubleshooting

For information about ADM error and event codes see Appendix D - ADM Event and Error Codes Overview, page 66.
Before performing any troubleshooting procedure:
1. Perform Pressure Relief Procedure on page 42.
2. Turn Main Power Switch to the OFF position.
3. Allow equipment to cool.
Try the recommended solutions in the order given for each problem to avoid unnecessary repairs. Also, deter­mine that all circuit breakers, switches, and controls are properly set and wiring is correct before assuming there is a problem.
Troubleshooting

Light Tower (Optional)

Signal Description
Green on only System is powered up and there are
no error conditions present
Yellow on An advisory exists
Red flashing A deviation exists
Red on The system is shut down due to an
alarm occurring.
Errors include advisories, deviations, or alarms, so green will only be on when none of these occur. A yellow light can be on at the same time as red (flashing or solid on) when an advisory exists at the same time as a devi­ation or alarm.

Common Problems

Problem Cause Solution General
Display Module completely dark
No or incorrect amount of material dispensed from either side
Significant material leaking from pump seal
Material dispensed not cor­rect weight
Proportioning System
Proportioning pump does not hold pressure when stalled
No power Verify main power switch is ON Thrown circuit breaker Check machine breakers and reset Loose connection Tighten 5-pin cable on Advanced Display Module Bad display module Replace Advanced Display Module Ball valve closed (if installed) Open tank ball valve Tank empty Add fluid Tank clogged Clean tank Air in material Prime the machine Pump shaft worn and/or shaft seal
worn Specific gravity of one or more of the
two materials has changed since cali­bration
Check valve malfunction Remove check valve; clean or replace as necessary Piston worn or broken Replace piston
Pump piston or intake valve leaking 1. Observe gauges to determine which pump is los-
Remove pump shaft assembly and reinstall, see pump manual for instructions and rebuild kit
Perform calibration procedure
ing pressure.
2. Determine in which direction the pump has stalled by observing which directional valve indi­cator light is on.
3. Repair the valve.
313873H 47
Troubleshooting
Problem Cause Solution
Material imbalance Inadequate flow from pump; cavitation Clean inlet strainer screen
Worn pump inlet valve ball/seat or gasket, repair as necessary
Erratic pump movement Pump cavitation Feed pump pressure is too low, adjust pressure to
within required range
Pump output low Obstructed fluid hose or mixer; fluid
hose ID too small Worn piston valve or intake valve in
displacement pump Inadequate feed pump pressure Check feed pump pressure and adjust to within
Open, clear; use hose with larger ID
See pump manual for appropriate repair procedure
required range
48 313873H

ADM Troubleshooting

B
Troubleshooting
C
A
H
G
FIG. 13: ADM Component Identification - Rear
D
E
F
ti12362a1
ADM System Status LEDs (B) Conditions
Module Status
LED Signal Description
Green on Run mode, System on
Green flashing Setup mode, System on
Yellow on Run Mode, System off
J
S
K
L
M
N
R
ti12902a
P
ADM Module Status LEDs (N) Conditions
Module Status
LED Signal Description
Green on System is powered up
Yellow on Communication in progress
Red solid ADM hardware failure
Red flashing Uploading software
USB Module Status LEDs (L) Conditions
Module Status
LED Signal Description
Green flashing System is powered up
Yellow on Downloading information to USB
Green/Yellow Flashing
313873H 49
ADM is busy, USB cannot transfer information when in this mode
Troubleshooting

Motor Control Module

Diagnostic Information
Table 3: LED Status Signal
Module Status LED Signal Description
Green on System is powered up
Yellow on Internal communication in progress
Red solid MCM hardware failure. Replace MCM
Red flashing fast Uploading software
Red flashing slow Token error, remove token and upload
software token again
LED Signals
FIG. 14: LED Signals
50 313873H
r_257396_3b9905_07b
Acceptable Size and Duration of Power Line Voltage Fluctuations
The Motor Control Module is designed to withstand volt­age fluctuations from the incoming power supply. If the incoming power supply goes outside of the tolerable range, an over-voltage condition is flagged and the sys­tem shuts down in an alarm state. Excessive or repeated over-voltage may permanently damage hard­ware. The chart below shows the permissible magnitude and duration of temporary over-voltage events. Consult a qualified electrician if there are any concerns about the available power supply.
Maximum Permissible Transient Voltage Surges
* Constructed from ITIC 1996 curve, referenced by IEC 61000-2-4
1400
Troubleshooting
1200
1000
800
600
400
Voltage (Volts RMS)
200
0
0.000001 0.00001 0.0001 0.001 0.01 0.1 1 10
1200Vac, 1697Vdc
<--1 MW Max Surge Power
480Vac, 679Vdc
336Vac, 475Vdc
<--150 KW Max Surge Power
<--50 KW Max Surge Power
288Vac, 407Vdc
<--No Power Limit
264Vac, 373Vdc
Time (seconds)
313873H 51
Troubleshooting

Fluid Control Module

Diagnostic Information
Module Status LED Signal Diagnosis
Green on System is powered up
Yellow Internal communication in progress
Red solid FCM hardware failure. Replace FCM.
Red flashing fast Uploading software
Red flashing slow Token error. Remove token and
upload software token again.
FIG. 15:
Module Status LEDs
ti12337a1
52 313873H
Troubleshooting
313873H 53

Appendix A - ADM Icons Overview

Appendix A - ADM Icons Overview

Setup Screen Icons

Icon Description
Enter Screen
Exit Screen
Erase Selected Item
Erase All Items Shown
Change Multiple Values
(see Using the button on
page 56)
Ratio Calculator
(see Using the button on
page 56)
Icon Description
Learn Top-Most Piston Position
Go to Next Calibration Screen
Begin Weight Calibration Shot
Dispense Valve Details
Shot Number
Sequence
Flow
Weight
Volume
Return to Previous/Main Screen
Calibrate Piston Position
On Main Calibration screen:
Calibrate Weight Dispense
On Flow Meter Calibration screen:
Use Dispensed Material Weight to Calibrate Flow Meters
Use Dispensed Material Volume to Calibrate Flow Meters
Calibrate Flow Meters
Learn Bottom-Most Piston Position
Duration
Ratio
Calculated Ratio
Tank/Tank Heater
Primary Heater
Heated Hose
Chiller
54 313873H
Appendix A - ADM Icons Overview

Home Screen Icons

Icon Description
Select Operating Mode
Initiate Dispense
Dispense Disabled
Ratio Check
Park Piston
Park Piston Disabled
Close Dispense Valve
Icon Description
Prime A (Red) Side
Prime B (Blue) Side
Enter Screen
Exit Screen
Erase Selected Item
Erase All Items Shown
Weight
Perform Base Purge
Selected Shot Number
No Shot Number Selected
Selected Sequence and Sequence Position
No Sequence Selected
Skip to Next Shot in Sequence
Abort Sequence
Volume
Duration
Piston Cycles
Edit Operator Mode Flow Properties
313873H 55

Appendix B - ADM Setup Screens Overview

Appendix B - ADM Setup Screens Overview
Setup Screens Navigation Diagram
Shot #1
Shot #2
Shot #...
Sequences #1
Sequences #2
Sequences #...
Calibration
System #1
System #2
System #3
Shots
This screen allows the user to edit shot definitions. The contents of this screen change based on the Dispense Mode. Shots may be defined by flow rate and by time (duration), volume, or weight depending on the Dis­pense Mode selection. See System #1 on page 58 for Dispense Mode options.
Maintenance
Supply
Conditioning #1
Conditioning #2
Conditioning #3
Advanced #1
Advanced #2
Advanced #...
4. Type the new value then press enter. All values below the selected shot will change to the new value.
5. Repeat the previous two steps as desired.
6. Press to deactivate.
Using the button
The button can be used to calculate an x:1 ratio
from a non x:1 ratio. For example, if the desired ratio is
5:2, the button can be used to convert 5:2 to 2.5:1.
1. From the Shots screen, press .
2. Navigate to the column.
Using the button
The button can be used to change multiple shot
definitions at once.
3. Press .
4. Enter the non x:1 ratio in the column. The cal­culated x:1 ratio will automatically be displayed in
the column and in the shot definition.
1. From the Shots screen, press .
5. Press .
2. Press .
3. Navigate to a shot definition value.
56 313873H
Appendix B - ADM Setup Screens Overview
Sequences
This screen allows the user to edit sequence definitions. The contents of this screen change based on the Dis­pense mode.
Dispense detail is shown as volume, time, or weight depending on which Dispense Mode is selected. See
System #1 on page 58 for Dispense Mode options.
NOTE: 5 sequences with 20 positions each are avail-
able across 10 pages.
Calibration - Learn Mode
This screen is for learning the mechanical limits for pis­ton travel.
Calibration - Weight
Calibration - Main
This screen shows calibration data for the system and provides access to the individual calibration screens. See System Setup and Calibration on page 29 for how to use the calibration screens.
The date next to each key represents the last time that function was performed.
The “Cal. Min” and “Cal. Max” values are the system recognized extreme ends of piston travel.
This screen is for calibrating weight.
Calibration - Flow Meters
This screen is for calibrating the flow meters.
313873H 57
Appendix B - ADM Setup Screens Overview
System #1
This screen allows the user to set mechanical system settings and the Dispense Mode setting.
Dispense mode can be set to time, volume, or weight. Dispense mode controls how dispense quantities are measured. Dispensing must be calibrated, see System Setup and Calibration on page 29 for more informa­tion.
Pump sizes and inlet pressures must be entered on this screen. The inlet pressure is the minimum allowable fluid inlet pressure. If the inlet pressure is below this value, dispensing is disabled. If pump sizes and inlet pressures are not entered properly, system performance will be affected.
System #2
This screen allows the user to set the Gel Timer proper­ties and set which items are installed on the machine.
NOTICE
The correct dispense valve option must be selected. Selecting an incorrect dispense valve option will lead to erratic machine performance.
When enabling the Gel Timer, the user must select one of the 100 available shot definitions to use as the Gel Shot. This shot will be dispensed when the Idle Period expires. The Idle Period will begin after a dispense is completed. Any dispense operation in the middle of the timer countdown will reset the Idle Period counter. The system will generate an audible alarm that will begin the user-entered number of seconds before the Idle Period expires.
The hydraulic level sensor and hydraulic RTD for the pump line must be enabled when installed in the sys­tem. If the sensors are not marked as enabled, they will be ignored.
58 313873H
Appendix B - ADM Setup Screens Overview
System #3
This screen allows the user to edit the labels for the A (Red) and B (Blue) sides of the machine. The labels set for the A (Red) and B (Blue) sides of the machine are displayed throughout the screens. Labels are limited to five characters.
The pressure imbalance alarm can also be set. This is the difference between the A (Red) and B (Blue) side material pressures before an alarm occurs.
Maintenance
This screen shows shot number and sequence position counters. Use the Counters dropdown menu to select which set of counters to view.
Supply
The flowmeter types are defined on this screen. The available flow meter types are Disabled, HG6000, or SRZ-100. The ratio deviation value is the allowable per­centage before the machine displays a pop-up notifica­tion. The ratio alarm value is the allowable percentage difference before the machine will stop a dispense.
Keyboard Screen
This screen is used to edit the A (Red) and B (Blue) labels on the ADM. Use arrow keys to select the desired
Do not use the check-boxes on this screen. If a low level sensor is used, check the box for the Supply Low Level Sensor option on the System #2 screen, see page 58.
letter and press to accept the letter. Press to exit the keyboard.
313873H 59
Appendix B - ADM Setup Screens Overview
Conditioning #1
NOTE: The VRM does not use temperature condition-
ing. Do not check any of the boxes on this screen.
Conditioning #2
Advanced #1
This screen allows the user to set the language, date format, current date, time, setup screens password, screen saver delay, and turn on or off silent mode.
Advanced #2
NOTE: The VRM does not use temperature condition-
ing.
This screen allows the user to set the units of measure.
Conditioning #3
NOTE: The VRM does not use Night mode conditioning.
Do not enable night mode.
60 313873H
Appendix B - ADM Setup Screens Overview
Advanced #3
This screen allows the user to control the availability of some key system features.
Disable Dispensing From Display: Check this box to disable dispensing from the ADM. A footswitch, dispense valve trigger, or other external signal will be the only way to initiate a dispense.
Disable Modifying Temp Setpoint: Check this box to disable modifying temperature setpoints from the Run screens. This is only applicable if temperature control items are installed and enabled.
Disable Operator Mode Flow Adjustments: When checked, the user will not be able to change the flow on operator mode Home Run screen.
Disable Operator Mode Ratio Adjustments: When checked, the user will not be able to change the ratio on operator mode Home Run screen.
Temperature Alarm Disables Dispense: When checked, any active alarm from any temperature zone will disable dispensing.
Complete Dispense with Setpoint Error: When this box is checked, the shot will continue dispens­ing even if the system never reaches the desired setpoint.
Enable Shot Aborted Notification: When this box is checked, a pop-up notification will be displayed when a shot is aborted.
Enable Ratio Check Weight Mode Entry: This option is for machines without flow meters. When this box is checked, after any ratio check shot a popup window will appear asking the user to enter the dispensed weights. Press the Cancel button to abort the entry or press the Enter button to record the new values.
Advanced #4
This screen is for enabling the optional Diagnostic screen and for enabling USB log downloading and USB errors. For more information about USB operation, see Appendix E - USB Operation on page 77. For more information about the optional screens, see Diagnostic screen on page 65.
Advanced #5
Numbers shown are for reference only and may be different on your system.
This screen displays software information.
313873H 61

Appendix C - ADM Run Screens Overview

Appendix C - ADM Run Screens Overview
Run Screens Navigation Diagram
Home Events #1
Errors #1
Errors #2
Errors #...
Events #2
Events #...
Home Screens
The Home screen is the first screen that displays when the machine is started and when switching from the Setup screens to the Run screens. The Home screen shows the current fluid pressure at the pump outlets.
The operating mode can be selected from the Home screen. The available operating modes are Operator, Sequence, Shot, Standby, Night, and Disabled.
System Off
Maintenance
Optional:
Diagnostic
Disabled
In Disabled mode, machine operation is disabled and the setup screens cannot be accessed.
Standby
This operating mode is only used at initial startup and
when is pressed.
In Standby mode, the user can park the pumps, perform a base purge, and close the dispense valve. See Shut- down on page 41 for base purge and parking the pumps procedure.
62 313873H
Appendix C - ADM Run Screens Overview
Shot
This mode allows the user to dispense using one of the defined shots. See Shots screen on page 56. Shot defi- nition including flow, ratio, progress, and size is shown at the bottom of the screen. The user can also dispense a ratio check shot.
Sequence
Operator
This mode allows users to dispense without using pre-
defined shot numbers or sequences. Use to set
flow rate and ratio. The user can also dispense a ratio check shot.
Prime
This mode allows the user to dispense using one of the
This mode allows users to prime each pump individually. defined sequences. See Sequences on page 57. The shot definition for the shot used in the current sequence position is shown at the bottom of the screen. Shot defi­nition information shown includes flow, ratio, progress, and size.
313873H 63
Appendix C - ADM Run Screens Overview
Errors Screen
The Errors screens display the 50 most recent system errors. Each error entry includes a description and error code along with a date and time stamp. There are 5 pages displaying 10 errors each.
See the Troubleshooting section beginning on page 47 for a detailed description of all system errors.
Events Screen
Maintenance
This screen displays historical information for each pump including material usage and pump cycles. The
material usage units are shown next to the , , or
icons and depend on the selected Dispense mode. The batch counters are resettable but the total counters are not.
The Events screens display the 200 most recent events. Each event includes a description and event code along with a date and time stamp. There are 20 pages display­ing 10 events each.
See the Troubleshooting section beginning on page 47 for a detailed description of all system events.
64 313873H

Optional Screens

The optional Diagnostic screen can be enabled in the
Advanced #4 screen, see page 61.
Diagnostic
Appendix C - ADM Run Screens Overview
The Diagnostic screen shows status information for vari­ous system components.
313873H 65

Appendix D - ADM Event and Error Codes Overview

Appendix D - ADM Event and Error Codes Overview

Event Codes

Event Code and String Triggers
EAA0-R: Base Purge Base purge performed EAD0-R: Pump(s)
Parked EA10-R: Red Pump
Prime EA20-R: Blue Pump
Prime EA00-R: Disp.
Occurred (Shot #) EAR0-R: Night Mode
Recirc On
EB00-R: Stop Button Pressed
EBR0-R: Night Mode Recirc Off
EBR1-R: Setup val­ues changed
ECA1-R: Red Mate­rial SG Modified
ECB2-R: Blue Mate­rial SG Modified
ECH0-R: Learn Mode Executed
EH00-R: Gel Timer Dispense
EL00-R: System Pow­ered On
EM00-R: System Powere d Off
ENC1-R: Cal. Point 1 Weight Entered
ENC2-R: Cal. Point 2 Weight Entered
ENC4-R: Cal. Point 1 Weight Erased
ENC5-R: Cal. Point 2 Weight Erased
END0-R: Ratio Check Dispense
ENN0-R: Automatic Cal. Performed
One or both pumps were parked
The red pump has been primed
The blue pump has been primed
A dispense has occurred of the given shot number.
While in night mode the system has automatically entered a low recircula­tion mode and attempted to turn on all enabled conditioning zones.
The Red stop button was pressed on the Advanced Display Module.
While in night mode the system has automatically stopped the low recircu­lation mode and turned off all condi­tioning zones.
One or more of the setup values has changed.
The Red materials specific gravity was modified.
The Blue materials specific gravity was modified.
A learn mode calibration was success­fully completed.
The gel timer expired and the system automatically took the gel shot.
The System was powered on.
The System was powered off.
A value for the first point in the three point calibration was entered.
A value for the second point in the three point calibration was entered.
The running average for point one of the three point calibration was erased.
The running average for point two of the three point calibration was erased.
A ratio check shot was dispensed from the ratio check calibration screen.
The system was successfully charac­terized with the Automatic calibration.
Event Code and String Triggers
EQU1-R: Settings Downloaded
EQU2-R: Settings Uploaded
EQU3-R: Language Downloaded
EQU4-R: Language Uploaded
EQU5-R: Logs Down­loaded
ER01-R: Shot Count Reset
ER02-R: Seq. Posi­tion Count Reset
ERA1-R: Red Mate­rial Volume Reset
ERA2-R: Red Mate­rial Weight Reset
ERA3-R: Red Cycle Count Reset
ERB1-R: Blue Mate­rial Volume Reset
ERB2-R: Blue Mate­rial Weight Reset
ERB3-R: Blue Cycle Count Reset
The system settings were successfully transferred from the ADM to a USB drive.
The system settings file was success­fully transferred from the USB drive to the ADM.
The custom language file was suc­cessfully transferred from the ADM to a USB drive.
The custom language file was suc­cessfully transferred from the USB drive to the ADM.
The Error/Event and Shot data logs were successfully transferred from the ADM to a USB drive.
A counter from the shot counters main­tenance page was erased
A counter from the sequence counters maintenance page was erased
The resettable totalizer for the Red material volume was reset to zero.
The resettable totalizer for the Red material weight was reset to zero.
The resettable cycle counter for the Red pump was reset to zero.
The resettable totalizer for the Blue material volume was reset to zero.
The resettable totalizer for the Blue material weight was reset to zero.
The resettable cycle counter for the Blue pump was reset to zero.
66 313873H

Error Codes

Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
The three point calibration data is invalid, system will operate in weight mode but will attempt to volumetrically calculate weight. This will lead to consistent shots which will be offset for the desired dispense amount.
The system will ignore the calibration data gathered and will use information gathered during dispenses
An over current was detected on the output
High current has been detected on a phase and has been shutdown to prevent damage
Too much current is being drawn from the wall
A hardware current fault has occurred causing a system shutdown
Unexpected current to heater/chiller
0500
05A1
A4A6
A4B5
A4A3
A4B1
A4A2
A4B4
A4A7
A4B8
A4H1
A4M1
A4N1
A7A6
A7B5
A7A3
A7B1
A7A2
A7B4
A7A7
A7B8
Invalid Weight Cal. Data
Invalid Auto Cal. Data
Red Blanket Overcurrent Blue Blanket Overcurrent Red Inline Overcurrent Blue Inline Overcurrent Red Hose Overcurrent Blue Hose Overcurrent Red Chiller Overcurrent Blue Chiller Overcurrent
Motor Over Current
Motor Over Current
Motor Over Current
Red Blanket Control Fault Blue Blanket Control Fault Red Inline Control Fault Blue Inline Control Fault Red Hose Control Fault Blue Hose Control Fault Red Chiller Control Fault Blue Chiller Control Fault
Error Type Cause Solution
Deviation Invalid data Re-calibrate the machine
If any messages appeared indicating why the
Deviation Invalid data
Alarm
Alarm
Alarm
Bad heaters Measure resistance of heater
Alarm
Alarm
Alarm
Alarm High voltage
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Shorted Temperature Control Module Bad internal wiring of the motor Short circuit of motor wiring Low voltage from the wall during load Short circuit of motor wiring
Motor rotor has become locked
Shorted Temperature Control Module
calibration failed attempt to fix the problem then re-run the calibration
Measure voltage across the disconnect switch. Voltage should measure between 190 and 264 Vac. If temperature rises for a zone that has been disabled, replace Temperature Control Module
Replace motor
Check wiring to the motor to ensure no bare wires are touching and that no wires are shorted to ground Make sure the supply line is properly sized for the load and is above the minimum voltage requirements Check wiring to the motor to ensure no bare wires are touching and that no wires are shorted to ground Unplug the directional valve (so pressure will not build) and try to move the motor again. If this succeeds then the power pack may need to be replaced. If the motor is still unable to move, the bearings or hydraulic pump have likely failed in the motor and will need to be replaced.
If temperature rises for a zone that has been disabled, replace Temperature Control Module
313873H 67
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
No Red
A8A6
Blanket Current
No Blue
A8B5
Blanket Current No Red Inline
A8A3
Current No Blue Inline
A8B1
A8A2
A8B4
A8B7
A8B8
A9C1
B9C0
CAC1
CAC3
CAC4
CAC5
CAC6
CAC7
CAA6
CAB5
CAA3
CAB1
CAA2
CAB4
CAA7
CAB8
D1A1
Current No Red Hose Current
No Blue Hose Current
No Red Chiller Current
No Blue Chiller Current
Motor Over Current
Small Shot Request
Comm. Error Motor Comm. Error Red Tank Comm. Error Blue Tank Comm. Error Mix Head Comm. Error Mix Head 2 Comm. Error Ratio Monitor Comm. Error Red Blanket Comm. Error Blue Blanket Comm. Error Red Inline Comm. Error Blue Inline Comm. Error Red Hose Comm. Error Blue Hose Comm. Error Red Chiller Comm. Error Blue Chiller Setpoint Not Reached
No current to the conditioning zone
A software error has occurred commanding too much current
The requested dispense amount is below the minimum amount of the system (25% of the combined pump volumes is the minimum)
Communication error
The set point was not reached and the pump was shutdown
Error Type Cause Solution
Alarm
Alarm
Tripped circuit breaker Visually check circuit breaker for a tripped condition
Alarm
Alarm
Alarm
Measure voltage across input terminals on power line
Alarm Low power
Alarm
Alarm Bad heater(s) Measure resistance of heater(s)
Alarm
Deviation
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm Module not programmed Program the module
Alarm Module missing power Check power supply connection
Alarm Module bad Replace module
Deviation
Cable unplugged/loose power
Bad Motor Control Module code
Pumps are defined with the wrong size
Requested shot is below the capabilities of the current pump setup
Loose/broken connection Check connection
Material restriction too high for requested flow
filter. Voltage should measure between 190 and 264 Vac
Check for loose or disconnected wires or plugs
Check for MCM software update, load latest MCM software, if problem persists contact Graco
On the ADM go into the Setup screens to the System screens then make sure that the pump sizes are defined correctly
If the user has to be able to take the shot the system must be fitted with smaller pumps
Reduce flow request
68 313873H
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
D2A1
D3A1
D5A1
D6A1
DDA1
DDB2
DFA1
DSC0
F7D1
L111
L122
L311
L322
L6A1
L6B2
Setpoint Not Reached
Setpoint Exceeded
Invalid Learn Mode Data
Position Sensor Fault
Red Pump Cavitation
Blue Pump Cavitation
Pump Not Parked
Pumps Not Defined
Pump Failed to Stall
Red Low Material Level
Blue Low Material Level
Red High Material Level
Blue High Material Level
Red Auto Refill Timeout
Blue Auto Refill Timeout
The set point was not reached Deviation
The set point was exceeded Deviation
This calibration lets the MCM know where the ends of the pump are. If the data gathered during this process is outside of normal parameters the machine will operate with a greatly reduced stroke.
The linear position sensor is returning data that should not be possible during normal operation
Cavitation was detected on the given pump
The pump failed to reach the park position
The type or size of the Red or Blue material pumps have not been defined
When the pump tried to stall to pressure the pump traveled more than it should in normal operation (only applies to dead-headed system)
Low material level in tanks
High material level in tanks
The tank stand has been filing for a time greater than expected
Error Type Cause Solution
Pump cannot reach the requested pressure Pump cannot reach the requested flow System underwent a change that caused a large drop in restriction (such as new orifices)
No material in pumps
Recalibrate the machine Rerun the learn mode calibration
Loose/bad connection
Deviation
Bad linear position sensor
Loose/bad connection to linear position sensor
Bad linear position
Alarm
Deviation
Deviation Verify that feed pumps are supplying material
Deviation
Alarm
Deviation
Deviation Tanks low on material Fill tanks with material
Deviation Loose/broken connection
Deviation
Deviation
Deviation
Deviation
sensor Linear position sensor
may be loose where attached to pump housing
Insufficient material being supplied or insufficient material pressure on feed system
Debris or packout in the incoming fluid filter
Orifices blocked Clear blockage Hose blocked Clear or replace hose as necessary Dispense valve failed to open
Properly setup the system
Failure of the dispense valve
Material leak
Out of material Fill tanks
Bad level sensor Replace level sensor
Defective fill valve
No material is actually being fed
Loose level sensor connection
Bad level sensor Replace level sensor
Increase restriction in the system
Decrease restriction in the system
Erase learned System Data, found in the setup screens under calibration
Make sure the material lines are open and have proper feed pressure
Check to ensure the pressure transducer is properly installed and all wires are properly connected
Verify pump moves to limits, if problem persists replace linear position sensor
Check to ensure the linear position sensor is properly installed and all wires are properly connected
Replace linear position sensor
Re-tighten the sensor and re-calibrate the machine
Verify that incoming ball valves are open
Inspect filter for debris of filler packout and clean or replace as necessary
Check to make sure the dispense valve is properly configured and connected to the MCM On the ADM go into the setup screens -> System-> then make sure that the pump type and size are set (not --) Ensure the valve has a proper air supply and seals properly. If not, service the valve as necessary. Visually inspect the machine and hoses for sign of leakage. NOTE: This error will display after 2 full piston strokes so the leak will be substantial.
If the tanks appear to have plenty of material check to make sure the level sensor is connected to the proper port and that the cord is not damaged
If the tanks appear to have plenty of material check to make sure the level sensor is connected to the proper port and that the cord is not damaged
Make sure the feed pumps are operating properly
Check for loose or disconnected wires or plugs
313873H 69
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
The volume of oil in the tank is
MBH1 Low Oil Level
Low Motor
MBN1
Performance
Pump Failed
N4A1
to Move
Thermal
P400
Pressure Rise
Red Pressure
P4A1
Shutdown
Blue
P4B2
Pressure Shutdown
below the minimum level needed for the system to properly operate
The motor magnetism has decreased to the point where performance is greatly reduced
The MCM attempted to move the pump but no movement was detected
Pressure has risen to an unsafe level due to thermal expansion of materials. All conditioning zones have automatically been turned off.
The material pump pressure exceeded the maximum operating pressure as defined in the setup screens
Error Type Cause Solution
Low oil level Check oil level and if low add more hydraulic fluid
Check to ensure the hydraulic oil level sensor is
Loose/bad connection
Bad level sensor Replace sensor
Alarm
Leak in hydraulic driver
Leak in the hydraulic reservoir, heat exchanger
Advisory
Deviation
Deviation High pressure
Alarm
Alarm
Prolonged exposure to heat or high voltage
Motor failure
Hydraulic power pack failure
Loose/bad connection to the linear position sensor
Failure of the linear position sensor Motor no longer coupled to hydraulic pump Supply tube from hydraulic pump to manifold is loose or broken Broken motor shaft Replace motor
Over-pressure valve dumping to tank
Dispense valve failed to open Bad dispense valve Replace dispense valve
Restriction in the material lines
Invalid maximum pressure defined
Orifices blocked Clear blockage Hose blocked Clear blockage or replace hose as necessary Dispense valve failed to
open
properly connected to the MCM and that the wire has not been damaged
Inspect hydraulic driver end seals and early leak detection tubing. Replace seals as necessary and replace lost oil. Inspect the hydraulic reservoir fittings and filter for leaks. Repair or replace as necessary and replace lost oil.
If error persists and performance can no longer satisfy the user requirements the motor will need to be replaced
Visually check to ensure the pump is moving, if not ensure the motor is wired properly If motor is moving but pump is not and pressure is not building they hydraulic power pack may need servicing Check to ensure the linear position sensor is properly connected to the MCM and the wiring has not be damaged
Replace the linear position sensor
Reset coupler per specifications and retighten set screws
Retighten or replace supply tube
Verify that no outside forces are stopping the pump from moving, then inspect over-pressure valve for damage or debris
Open the dispense valve manually or open the valves to bleed pressure
Check to make sure the dispense valve is properly configured and connected to the MCM
Check to ensure there is no blockage
Make sure the requested pressure is within the max operating pressure, which can be found on the setup screen System 1
Check to make sure the dispense valve is properly configured and connected to the MCM
70 313873H
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
The pressure difference
P4D0
P6A1
P6B2
T1A6
T1B5
T1A3
T1B1
T1A2
T1B4
T1A7
T1B8
T3H1
T3N1
Pressure Imbalance
Red Pressure Sensor Fault Blue Pressure Sensor Fault Red Tank Low Fluid Temp. Blue Tank Low Fluid Temp.
Red Inline Low Fluid Temp.
Blue Inline Low Fluid Temp. Red Hose Low Fluid Temp.
Blue Hose Low Fluid Temp.
Red Chiller Low Fluid Temp. Blue Chiller Low Fluid Temp.
Oil Temp. Cutback
Motor Temp. Cutback
between the Red and Blue material is greater than the defined amount
The pressure sensor is providing invalid/no pressure readings
Fluid temperature is below the defined low alarm limit
The hydraulic oil temperature is approaching a level where damage is possible so the Motor Control Module is limiting the output to a safe level
Motor temperature is approaching a level where damage is possible so the motor control module is limiting the output to a safe level
Error Type Cause Solution
Dispense line is clogged
Pressure imbalance is defined too low
Orifice blocks closed off
Alarm
Alarm Loose/bad connection
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm Low power
Alarm
Alarm Bad heater(s) Measure resistance of heater(s)
Deviation
Advisory
too much on one or both sides
Debris in the orifice block
Material fillers may have packed out in an orifice
Out of material Fill tanks with material Feed system defective Replace defective item
Bad sensor Replace pressure transducer
No material in pump Fill tanks
Tripped circuit breaker Visually check circuit breaker for a tripped condition
Cable unplugged/loose power
No power to fan Check cord to make sure fan has power Debris in fan or fan grill Clear debris from fan/fan grill
Low air volume from fan
No power to fan Check cord to make sure fan has power Debris in fan or fan grill clear debris from fan/fan grill
Low air volume from fan
Ambient environmental conditions are too hot
Motor/pump coupler may be rubbing on hydraulic pump
Ensure the material flow is equally restricted on both material lines On the ADM go into the setup screens -> System-> and ensure the pressure imbalance value is the maximum acceptable to prevent unnecessary alarms which will abort dispenses Verify that one or both of the orifice blocks dispense when adjusted to the fully open position then adjust accordingly Relieve system pressure then remove the orifice from the orifice block and inspect for debris in the cavity Relieve system pressure and remove the orifice from the orifice block and inspect for pack out. Clean or replace as necessary.
Check to ensure the pressure transducer is properly installed and all wires are properly connected
Measure voltage across input terminals on power line filter. Voltage should measure between 190 and 264 Vac
Check for loose or disconnected wires or plugs
Try to stop fan by lightly pressing on the center with a pencil eraser. If the fan slows down easily it will need to be replaced
Try to stop fan by lightly pressing on the center with a pencil eraser. If the fan slows down easily it will need to be replaced
Move machine to an area below 120°F
Reset coupler per specifications and retighten set screws
313873H 71
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
Blue Tank
T4B5
High Fluid Temp.
Red Inline
T4A3
High Fluid Temp. Blue Inline
T4B1
High Fluid Temp.
T4A2
T4B4
T4A7
T4B8
T4C1
T4H1
T4N1
T6A6
T6B5
T6A3
T6B1
T6A2
T6B4
T6A7
T6B8
Red Hose High Fluid Temp.
Blue Hose High Fluid Temp.
Red Chiller High Fluid Temp. Blue Chiller High Fluid Temp.
Motor Control High Temp.
Oil Temp. Shutdown
Motor Temp. Shutdown
Red Tank RTD Fault
Blue Tank RTD Fault
Red Inline RTD Fault
Blue Inline RTD Fault
Red Hose FTS Fault
Blue Hose FTS Fault
Red Chiller RTD Fault
Blue Chiller RTD Fault
Fluid temperature is above the defined high alarm limit
The temperature the MCM has reached a level where product life will be decreased drastically and has been shutdown for protection
The hydraulic oil is at a temperature where performance is impacted significantly and has resulted in a system shutdown
Motor temperature is too high and system has been shutdown to prevent possible damage
RTD 1 is giving no or invalid data
Error Type Cause Solution
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm Loose connections Tighten connections
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm Bad RTD Replace RTD
Defective RTD Replace RTD
Defective High Power Temperature Control Module
No power to fan Check cord to make sure fan has power Debris in fan or heatsink Clear debris from fan or heatsink
Low air volume from fan
Motor may be damaged Replace motor Debris is packed in the
MCM’s heat sink fins No Power to Fan Check cord to make sure fan has power Debris in fan or fan grill Clear debris from fan/fan grill
Low air volume from fan
No power to fan Check cord to make sure fan has power Debris in fan or fan grill Clear debris from fan/fan grill
Low air volume from fan
Ambient environmental conditions are too hot Motor may be damaged Motor may need to be replaced
Loose or bad connection Check RTD wiring
Replace High Power Temperature Control Module
Try to stop fan by lightly pressing on the center with a pencil eraser. If the fan slows down easily it will need to be replaced
Clear debris from MCM heat sink fins
Try to stop fan by lightly pressing on the center with a pencil eraser. If the fan slows down easily it will need to be replaced
Try to stop fan by lightly pressing on the center with a pencil eraser. If the fan slows down easily it will need to be replaced
Move machine to an area below 120°F
72 313873H
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
Red Blanket
T6C6
RTD Fault Blue Blanket
T6C7
T6C8
T8A6
T8A3
T8B1
T8A2
T8B4
T8A7
T8B8
T9A6
T9B5
T9A3
T9B1
RTD Fault Red Chiller RTD Fault Blue Chiller RTD Fault No Heat Red Tank No Heat Blue Tank No Heat Red Inline
No Heat Blue Inline
No Heat Red Hose
No Heat Blue Hose
No Cooling Red Chiller
No Cooling Blue Chiller
Red Blanket Temp. Cutoff Blue Blanket Temp. Cutoff
Red Inline Temp. Cutoff
Blue Inline Temp. Cutoff
RTD 2 is giving no or invalid data
No temperature rise
No temperature decline
Heater overtemperature cutoff
Error Type Cause Solution
Alarm
Alarm
Alarm
Alarm Bad RTD Replace RTD
Alarm
Alarm
Alarm
Alarm Low Power
Alarm
Alarm Bad heater(s) Measure resistance of heater(s)
Alarm Tripped circuit breaker Visually check circuit breaker for a tripped condition
Alarm
Alarm
Alarm
Alarm
Alarm Loose connections Tighten connections
Loose or bad connection Check RTD wiringT6C5
Tripped circuit breaker Visually check circuit breaker for a tripped conditionT8B5
Measure voltage across input terminals on power line filter. Voltage should measure between 190 and 264 Vac
Cable unplugged/loose power
Defective cooling valve
Chilled water supply off Turn on chilled water supply Loose or bad connection Check RTD wiring
Defective RTD Replace RTD
Defective High Power Temperature Control Module
Check for loose or disconnected wires or plugs
Disconnect the valve and measure the voltage across the wires when the chiller is running to ensure 24V is being delivered to the valve. If so, the cooling valve will likely need replacing.
Replace High Power Temperature Control Module
313873H 73
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
Red Blanket
T9C6
Ctrl Shutdown
Blue Blanket
T9C5
Ctrl Shutdown Red Inline
T9C3
Ctrl Shutdown Blue Inline
T9C1
Ctrl
T9C2
T9C4
T9C7
T9C8
V1H1
V4A6
V4B5
V4A3
V4B1
V4A2
V4B4
V4A7
V4B8
V4H0
W0U0
WBH1
Shutdown Red Hose Ctrl
Shutdown Blue Hose
Ctrl Shutdown Red Chiller Ctrl Shutdown Blue Chiller Ctrl Shutdown
Motor Control Undervoltage
Red Blanket Overvoltage Blue Blanket Overvoltage Red Inline Overvoltage Blue Inline Overvoltage Red Hose Overvoltage Blue Hose Overvoltage Red Chiller Overvoltage Blue Chiller Overvoltage
Motor Control Overvoltage
USB Update Failed
Motor Encoder Fault
PCB over temperature
The voltage to the MCM has dropped to a level where performance is greatly affected
High line voltage
The voltage to the MCM has reached an unsafe level and has been shutdown in an attempt to prevent damage
The ADM tried to upload a system settings file but failed
An error has been detected on the motor position sensor
Error Type Cause Solution
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Overheated Temperature Control Module
Tripped circuit breaker Visually check circuit breaker for a tripped condition Supply lines providing low voltage
Incoming line voltage is too high
Supply lines providing high voltage
System Settings file is corrupt
System Settings file is intended for another system
Failing sensors If error persists the motor will need to be replaced
Loose connection
Turn conditioning zone off. Wait a few minutes. If the condition does not clear or regenerates consistently, replace heater module
Check incoming voltage to ensure it is above the minimum operating voltage
Measure voltage across disconnect switch. Voltage should measure between 190 and 264 Vac.
Check incoming voltage to ensure it is below the maximum operating voltage
Replace the system settings file with a backup or new file
Ensure that the first line in the settings.txt file contains the text GMS. If not replace the file with the proper system update file.
Ensure the d-sub connector to the motor is connected and the wiring is intact
74 313873H
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
The motor has reached a
WKH1
WM06
WM05
WM03
WM01
WM02
WM04
WM07
WM08
WMA6
WMB5
WMC6
WMC5
WMC3
WMC1
WMC2
WMC4
WMC7
WMC8
WMH1
High Motor Speed
Red Tank Con. Fault Blue Tank Con. Fault Red Inline Con. Fault Blue Inline Con. Fault Red Hose Con. Fault Blue Hose Con. Fault Red Chiller Con. Fault Blue Chiller Con. Fault Red Blanket High Temp.
Blue Blanket High Temp.
Red Tank Con. Fault Blue Tank Con. Fault Red Inline Con. Fault Blue Inline Con. Fault Red Hose Con. Fault Blue Hose Con. Fault Red Chiller Con. Fault Blue Chiller Con. Fault Motor Controller Fault
speed that should not be reached in normal operation and was shutdown to prevent possible damage
High current to relay 1
Tank blanket is above the defined high alarm limit
Unexpected current to relay 1
A general fault has occurred within the MCM
Error Type Cause Solution
No power to directional valve
Bad directional valve connection
Directional valve failure The directional valve will need to be replaced Hydraulic power pack
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm Defective RTD Replace RTD
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Alarm
Deviation Internal hardware failure
failure Defective encoder Replace encoder Motor no longer coupled to hydraulic pump Supply tube from hydraulic pump to manifold is loose or
broken Broken motor shaft Replace motor
Broken contactor Replace contactor
Defective High Power Temperature Control Module Loose connections Tighten connections
Shorted module
Make sure the directional valve has power
Make sure the cord to the directional valve is connected to the correct port and the cord is not damaged
The hydraulic power pack will need repair
Reset coupler per specifications and retighten set screws
Retighten or replace supply tube
Replace High Power Temperature Control Module
If temperature is being affected by a zone that has been disabled, replace heat module
Cycle power, if the error persists the MCM will need to be replaced
313873H 75
Appendix D - ADM Event and Error Codes Overview
Error Code Error Name Error Description
The requested controlling value (pressure or flow) is outside the limits of the system
The shot that was entered for the gel timer is not a valid shot. This must be fixed before the gel timer will function properly
WSC0
WSD0
Invalid Setpoint Request
Invalid Gel Timer Definition
Error Type Cause Solution
On the ADM go into the setup screens -> System->
Deviation
Deviation
System incorrectly setup
Shot incorrectly defined
Gel timer shot is below the minimum dispense amount or set for a invalid pressure/flow
The MCM has determined that the gel timer shot will not be able to be executed based parameters entered in the ADM
and ensure that all pages have properly defined values
Redefine shot with control parameters within the limits of the system
Select a different shot or modify existing shot data
If you are certain that the shot is within parameters, try running the Learn Mode routine found in the setup screen Calibration. If the error persists, a gel shot with reduced control parameters is required.
76 313873H

Appendix E - USB Operation

Appendix E - USB Operation
NOTICE
Low-quality USB stick drives may lead to burning out the USB port on the ADM. Use only high-quality USB stick-drives.

Overview

There are 3 main uses for the USB.
Ability to download a log of up to the past 50,000 errors, events, or jobs that can contain over 150,000 snapshots of critical dipsense information.
Download, modify, and upload custom language files
Download and upload system configurations
This data includes most user-selectable and
user-configurable settings.
This data does not include pump counters, error
and event logs, shot and sequence counters

USB Options

The only options for USB on the ADM are in Advanced #4, see page 61.
The second option is the Erase icon which will reset the last download date to a time where all logs can be downloaded. This will allow the user to download all USB log entries, which may take over 2 hours if the log files are full. The ADM does not monitor the USB logs to alert the user when data may be overwritten. In order to minimize download times and the risk of losing data it is recommended that the user download the logs every 2 weeks or more often if the machine is used during more than one full shift a day.

Download Log Files

If the “Enable Downloading of USB Logs” is checked, a USB stick-drive can be used to download the log files.
To download the log files, insert a high-quality USB stick-drive into the USB port in the bottom of the ADM. The ADM will automatically begin downloading the log files as well as the custom language file (DISP­TEXT.TXT) and the system settings file (SET­TINGS.TXT). The status of the download will be shown in the Status bar. If an error occurs during download, remove the USB stick-drive then re-insert.
The first option is a checkbox that enables or disables the downloading of the Error Event and Shot Data log files. The Shot Data log runs during all operating modes.
313873H 77
Appendix E - USB Operation

Log Files, Folder Structure

F
IG. 16: DOWNLOAD, DATAxxxx Folders
Each time a stick-drive is inserted into the ADM USB port, a new folder named DATAxxxx is created. The number at the end of the folder name is incremented each time a stick-drive is inserted and data is transferred through the USB. In each DATAxxxx folder there are three log files. They are formatted as .csv (comma separated value) files and can be opened by most text editors or data processing programs such as Excel.
Example 1-ERROR File
The 1-ERROR file is the Errors and Events log file.
78 313873H
Example 2-JOB File
The 2-JOB file is the Shot Data log file.
Appendix E - USB Operation
Example 3-SYSTEM File
The 3-SYSTEM file is the Software Version log file.
313873H 79
Appendix E - USB Operation

Transfer System Settings

Use the following process to transfer system settings from one machine to another.
1. Insert a high-quality USB stick-drive into the ADM USB port on the system with the settings to be transferred. Once the download is complete the SETTINGS.TXT file will be located in the “DOWN­LOAD” folder.
NOTICE
The user should never attempt to modify the SET­TINGS.TXT file in any way. Graco is not responsible for damages caused by an improperly modified setup file.
2. Plug the USB stick-drive into a computer.
3. Navigate to the DOWNLOAD folder.
4. Copy the SETTINGS.TXT file from the DOWNLOAD folder into the UPLOAD folder.
5. Remove the USB stick-drive from the computer and install it into the ADM USB port for the second machine. The software will automatically begin updating.
NOTE: Before the update begins the ADM automatically shuts down the system, aborting any in-progress dis­pensing. When the software is updating a pop-up box will appear to inform the user of the update and the sys­tem will lock. Once the update is complete the ADM will tell the user to cycle power to apply the updates.
6. When the software is done updating, remove the USB stick-drive.
7. Turn the Main Power Switch to the OFF position then back to the ON position.
8. Plug the USB stick-drive into a computer.
9. Navigate to the UPLOAD folder and remove the SETTINGS.TXT file.
NOTE: Immediately after uploading the settings, remove the SETTINGS.TXT file from the UPLOAD folder to pre­vent accidental loss of data the next time the USB stick-drive is inserted into the ADM USB port. If there is a SETTINGS.TXT file in the UPLOAD folder when the USB stick-drive is inserted into the ADM USB port the software will automatically begin updating.
80 313873H

Update Custom Language

Appendix E - USB Operation
Use the following process to customize the text on the ADM. The language file DISPTEXT.TXT can be modi­fied in Excel but must be saved as a Unicode Text file with the extension .TXT in order for it to properly import.
1. Insert a high-quality USB stick-drive into the ADM USB port on the system with the settings to be transferred. Once the download is complete the DISPTEXT.TXT file will be located in the “DOWN­LOAD” folder.
2. Plug the USB stick-drive into a computer.
3. Navigate to the DOWNLOAD folder.
4. Copy the DISPTEXT.TXT file from the DOWNLOAD folder to your computer.
5. Use any data processing software such as Excel to edit the DISPTEXT.TXT file.
a. In the first column, locate the string to change.
b. In the second column of the same row, enter the
new string.
NOTE: Immediately following uploading the language file, remove the DISPTEXT.TXT file from the UPLOAD folder to prevent accidental loss of data the next time the USB stick-drive is inserted into the ADM USB port. If there is a DISPTEXT.TXT file in the UPLOAD folder when the USB stick-drive is inserted into the ADM USB the software will automatically begin updating.
c. Save the file as a Unicode Text file. The name
must remain “DISPTEXT.TXT”.
6. Copy the edited DISPTEXT.TXT file into the UPLOAD folder.
7. Remove the USB stick-drive from the computer and insert it into the ADM USB port. The software will automatically begin updating.
NOTE: Before the update begins the ADM automatically shuts down the system, aborting any in-progress dis­pensing. When the software is updating a pop-up box will appear to inform the user of the update and the sys­tem will lock. Once the update is complete the ADM will tell the user to cycle power to apply the updates.
8. When the software is done updating, remove the USB stick-drive.
9. Turn the Main Power Switch to the OFF position then back to the ON position.
10. Plug the USB stick-drive into a computer.
11. Navigate to the UPLOAD folder and remove the DISPTEXT.TXT file.
313873H 81
Appendix E - USB Operation
Example SETTINGS.TXT File Example DISPTEXT.TXT File
NOTICE
The user should never attempt to modify the SET­TINGS.TXT file in any way. Graco is not responsible for damages caused by an improperly modified setup file.
82 313873H

Technical Data

Air Inlet Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85-100 psi (0.59-0.69 MPa, 5.9-6.9 bar)
Wetted Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Stainless steel, UHMW
Hydraulic Reservoir Capacity . . . . . . . . . . . . . . . . . . . . 8 gal. (30 liters) each
Recommended Hydraulic Fluid . . . . . . . . . . . . . . . . . . . Citgo A/W Hydraulic Oil, ISO Grade 46

Dimensions

The dimensions of the machine vary by machine layout. The distance from the machine base to the dispense stand can be varied significantly because of the hoses.
Ref in. (mm)
A (Height) 79 (2007) B (Width) 63 (1600) C (Depth) 115 (2921)
Technical Data
B
A
C
313873H 83

Graco Standard Warranty

Graco warrants all equipment referenced in this document which is manufactured by Graco and bearing its name to be free from defects in material and workmanship on the date of sale to the original purchaser for use. With the exception of any special, extended, or limited warranty published by Graco, Graco will, for a period of twelve months from the date of sale, repair or replace any part of the equipment determined by Graco to be defective. This warranty applies only when the equipment is installed, operated and maintained in accordance with Graco’s written recommendations.
This warranty does not cover, and Graco shall not be liable for general wear and tear, or any malfunction, damage or wear caused by faulty installation, misapplication, abrasion, corrosion, inadequate or improper maintenance, negligence, accident, tampering, or substitution of non-Graco component parts. Nor shall Graco be liable for malfunction, damage or wear caused by the incompatibility of Graco equipment with structures, accessories, equipment or materials not supplied by Graco, or the improper design, manufacture, installation, operation or maintenance of structures, accessories, equipment or materials not supplied by Graco.
This warranty is conditioned upon the prepaid return of the equipment claimed to be defective to an authorized Graco distributor for verification of the claimed defect. If the claimed defect is verified, Graco will repair or replace free of charge any defective parts. The equipment will be returned to the original purchaser transportation prepaid. If inspection of the equipment does not disclose any defect in material or workmanship, repairs will be made at a reasonable charge, which charges may include the costs of parts, labor, and transportation.
THIS WARRANTY IS EXCLUSIVE, AND IS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO WARRANTY OF MERCHANTABILITY OR WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE.
Graco’s sole obligation and buyer’s sole remedy for any breach of warranty shall be as set forth above. The buyer agrees that no other remedy (including, but not limited to, incidental or consequential damages for lost profits, lost sales, injury to person or property, or any other incidental or consequential loss) shall be available. Any action for breach of warranty must be brought within two (2) years of the date of sale.
GRACO MAKES NO WARRANTY, AND DISCLAIMS ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, IN CONNECTION WITH ACCESSORIES, EQUIPMENT, MATERIALS OR COMPONENTS SOLD BUT NOT MANUFACTURED BY GRACO. These items sold, but not manufactured by Graco (such as electric motors, switches, hose, etc.), are subject to
the warranty, if any, of their manufacturer. Graco will provide purchaser with reasonable assistance in making any claim for breach of these warranties.
In no event will Graco be liable for indirect, incidental, special or consequential damages resulting from Graco supplying equipment hereunder, or the furnishing, performance, or use of any products or other goods sold hereto, whether due to a breach of contract, breach of warranty, the negligence of Graco, or otherwise.
FOR GRACO CANADA CUSTOMERS
The Parties acknowledge that they have required that the present document, as well as all documents, notices and legal proceedings entered into, given or instituted pursuant hereto or relating directly or indirectly hereto, be drawn up in English. Les parties reconnaissent avoir convenu que la rédaction du présente document sera en Anglais, ainsi que tous documents, avis et procédures judiciaires exécutés, donnés ou intentés, à la suite de ou en rapport, directement ou indirectement, avec les procédures concernées.
Graco Information
For the latest information about Graco products, visit www.graco.com.
TO PLACE AN ORDER, contact your Graco distributor or call to identify the nearest distributor. Phone: 612-623-6921 or Toll Free: 1-800-328-0211 Fax: 612-378-3505
All written and visual data contained in this document reflects the latest product information available at the time of publication.
GRACO INC. AND SUBSIDIARIES • P.O. BOX 1441 • MINNEAPOLIS MN 55440-1441 • USA
Copyright 2010, Graco Inc. All Graco manufacturing locations are registered to ISO 9001.
Graco reserves the right to make changes at any time without notice.
For patent information, see www.graco.com/patents.
Original instructions. This manual contains English. MM 313873
Graco Headquarters: Minneapolis
International Offices: Belgium, China, Japan, Korea
www.graco.com
Revised November 2012
Loading...