The accessories available for your instrument are checked for
compliance with currently valid safety regulations at regular
intervals, and are amended as required for new applications.
Currently up-to-date accessories which are suitable for your measuring instrument are listed at the following web address along
with photo, order number, description and, depending upon the
scope of the respective accessory, data sheet and
operating instructions:
www.gossenmetrawatt.com
We calibrate and recalibrate all instruments supplied by GMC-I
Messtechnik GmbH, as well as other manufacturers, at our service center, for example after one year within the framework of
your test equipment monitoring program, as well as prior to use
etc. and offer you test equipment management free of charge.
See also chapter 9.6 on page 71.
Repair and Replacement Parts Service
Calibration Center* and Rental Instrument Service
This address is only valid in Germany. Please contact our
representatives or subsidiaries for service in other countries.
* DAkkS Calibration Laboratory for Measured Electrical Quantities
D-K-15080-01-01 accredited per DIN EN ISO/IEC 17025
Accredited quantities: direct voltage, direct current value, direct current
resistance, alternating voltage, alternating current value, AC active
power, AC apparent power, DC power, capacitance, frequency,
temperature
4GMC-I Messtechnik GmbH
Competent Partner
GMC-I Messtechnik GmbH is certified in accordance with
DIN EN ISO 9001.
Our DAkkS calibration laboratory is accredited by the Deutsche
Akkreditierungsstelle GmbH (National accreditation body for the
Federal Republic of Germany) in accordance with DIN EN ISO/IEC
17025 under registration number D-K-15080-01-01.
We offer a complete range of expertise in the field of metrology:
from test reports and factory calibration certificates, right on up to
DAkkS calibration certificates.
Our spectrum of offerings is rounded out with free test equipment management.
As a full service calibration lab, we can calibrate instruments from
other manufacturers as well.
Contact Persons
GMC-I Messtechnik GmbH5
Table of Contents
ContentsPageContentsPage
1Safety Features and Precautions ....................................... 8
1.1Use for Intended Purpose ................................................................10
1.2Meanings of Danger Symbols ..........................................................10
1.3Meanings of Acoustic Warning Signals ............................................10
You have selected an instrument which provides you with a high
level of safety.
This instrument fulfills the requirements of the applicable European guidelines and national regulations. This is confirmed by
means of the CE mark. A corresponding declaration of conformity
can be
requested from GMC-I Messtechnik GmbH.
The TRMS digital multimeter has been manufactured and tested
in accordance with the following safety regulations:
IEC 61010–1:2010/DIN EN 61 010–1:2011/VDE 0411–1:2011.
When used for its intended purpose (see page 10), safety of the
operator, as well as that of the instrument, is assured. Their safety
is however not guaranteed, if the instrument is used improperly or
handled carelessly.
In order to maintain flawless technical safety conditions, and to assure
safe use, it is imperative that you read the operating instructions
thoroughly and carefully before placing your instrument into service, and
that you follow all instructions contained therein.
The multimeter is equipped with an automatic socket blocking
mechanism for your safety, and in order to safeguard your
instrument. This mechanism is linked to the rotary switch and only
allows access to those jacks which are actually required for the
selected function. It also prevents the user from turning the rotary
switch to impermissible functions after the measurement cables
have already been plugged in.
Measuring Categories and their Significance per IEC 61010-1
CATDefinition
Measurements in electrical circuits which are not directly connected to the
I
mains: e.g. electrical systems in motor vehicles and aircraft, batteries etc.
Measurements in electrical circuits which are electrically connected to the
II
low-voltage mains: via plug, e.g. in household, office and laboratory applications
Measurements in building installations: stationary consumers, distributor
III
terminals, devices connected permanently to the distributor
The measuring category and the maximum rated voltage which
are printed on the device apply to your measuring instrument, e.g.
600 V CAT II I or 1000 V CAT II.
For the application of measuring cables see page 72.
Observe the following safety precautions:
• The multimeter may not be used in potentially explosive
atmospheres.
• The multimeter may only be operated by persons who are
capable of recognizing contact hazards and taking the
appropriate safety precautions. Contact hazards according to
the standards exist anywhere, where voltages of greater than
33 V RMS or 70 V DC may occur. Avoid working alone when
taking measurements which involve contact hazards. Be
certain that a second person is present.
• Maximum allowable voltage between the voltage measuring
sockets or all connector sockets and ground is 1000 V for
measuring category II, and 600 V for measuring category I I I.
• Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
8GMC-I Messtechnik GmbH
Safety Warnings
• Make certain that the measurement cables are in flawless
condition, e.g. no damage to insulation, no interruptions in
cables or plugs etc.
• No measurements may be made with this instrument in
electrical circuits with corona discharge (high-voltage).
• Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be present.
• Measurements under moist ambient conditions are not
permitted.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8; “Technical Data” in the table entitled
“Measuring Functions and Measuring Ranges” in the
“Overload Capacity” column.
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents and voltages are
otherwise not indicated, and the instrument may be damaged.
• The instrument may not be operated if the fuse cover or the
battery compartment lid has been removed, or if its housing is
open.
• The input for the current measuring range is equipped with a
fuse link.
Maximum permissible voltage for the measuring circuit (= rated
voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only (see page 65)! The fuse must have a
breaking capacity of at least 30 kA.
Opening of Equipment / Repair
The equipment may be opened only by authorized service personnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid.
Even original spare parts may be installed only by authorized service personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
Repair and Parts Replacement by Authorized Service Personnel
When the instrument is opened, voltage conducting parts may be
exposed. The instrument must be disconnected from the measuring circuit before the performance of repairs or the replacement of
parts. If repair of a live open instrument is required, it may only be
carried out by trained personnel who are familiar with the dangers
involved.
Defects and Extraordinary Strains
If it may be assumed that the instrument can no longer be
operated safely, it must be removed from service and secured
against unintentional use.
Safe operation can no longer be relied upon:
• If the device demonstrates visible damage,
• If the instrument no longer functions, or if malfunctioning
occurs,
• After long periods of storage under unfavorable conditions,
e.g. humidity, dust or extreme temperature (see “Ambient Conditions” on page 64).
GMC-I Messtechnik GmbH9
Safety Warnings
!
1.1 Use for Intended Purpose
• The respective multimeter is a portable device which can be
held in the hand during the performance of measurements.
• Only those types of measurements described in chapter 5 may
be performed with the measuring instrument.
• The measuring instrument, including measurement cables and
plug-on test probes, may only be utilized within the specified
measuring category (see page 65 and the table on page 8
regarding significance).
• Overload limits may not be exceeded. See technical data on
page 60 for overload values and overload limits.
• Measurements may only be performed under the specified
ambient conditions. See page 64 regarding operating
temperature range and relative humidity.
• The measuring instrument may only be used in accordance
with the specified degree of protection (IP code) (see page 66).
1.2 Meanings of Danger Symbols
Warning concerning a point of danger
(attention: observe documentation!)
Warning concerning dangerous voltage at the measurement
input: U > 15 V AC or U > 25 V DC
1.3 Meanings of Acoustic Warning Signals
Voltage warning: > 1000 V (intermittent acoustic signal)
Current warning: > 11 A (continuous acoustic signal)
1 Display (LCD) (see page 13 for significance of symbols)
2
MAN / AUTO
Increase parameter values
shift key for manual/automatic measuring range selection
“Operating Mode” menu:
Selection of individual menu entries against
the direction of flow
3
ON / OFF | LIGHT
key for switching device and display illumination on and off
4 FUNC | ENTER multifunction key
“Operating Mode” menu: Acknowledge entry (ENTER)
ISO ON / OFF Insulation resistance measurement
U
Key for switching insulation resistance measurement on and off
5 Increase measuring range or move decimal point to the right (MAN function)
6
Rotary switch
for measuring functions (see page 14 for significance of symbols)
7 DAkkS calibration mark
8 Connector socket for ground / connected to ground
9 Connector socket for current measurement with automatic blocking
10 Connector socket for voltage, resistance, temperature, diode
capacitance measurement
with automatic blocking
11 DATA / MIN / MAX
Key for freezing, comparing, deleting the measured value, and for Min/Max function
Decrease values
“Operating Mode” menu:
12
MEASURE | SETUP
Select individual menu entries in flow direction
Key for switching between measuring and menu functions
13 ZERO | ESC
12GMC-I Messtechnik GmbH
Key for zero balancing
“Operating Mode” menu: Exit menu level and return to a higher level, exit
parameters entry function without saving
METRAHIT COIL
only – Function COIL:
UCOIL ON / OFFInterturn short-circuit measurement as long as
key is being pressed, see chapter 5.8.
14 Decrease measuring range or move decimal point to the left (MAN function)
15 Power pack connector jack
16 Infrared interface
and
Symbols used in the Digital Display
Battery full
Battery OK
Battery weak
Battery (almost) dead, U < 1.8 V
Data transmission to /
from the multimeter is active
IR interface in standby mode
(ready to receive starting commands)
1 Battery level indicator
2 ON: continuous operation (automatic shutdown deactivated)
3 MAN: manual measuring range selection active
4 Digital display with decimal point and polarity display
5 max/min: Min/Max value storage
6 DATA: display memory, “freeze measured value”
7 STORE: memory mode active
8 ISO: insulation resistance measurement active / selected test voltage
9 1: x current clip factor (transformation ratio)
10 IR: infrared interface display
11 Scale for analog display
12 Pointer for analog display (bar graph – pointer)
triangle appears: indicates overranging
13 Diode measurement selected
14 Selected type of current
15 TRMS measurement
16 Pt100(0): selected platinum resistance thermometer with
automatic recognition of Pt100/Pt1000
17 Type K: temperature measurement with type K (NiCr-Ni) thermocouple
18 sec (seconds): unit of time
19 REL: relative measurement with reference to offset
20 Unit of measure
21 ZERO: zero balancing active
22 Duty Cycle Measurement
(this function only available with customer-specific variant)
23 Warning regarding dangerous voltage: U > 15 V AC or U > 25 V DC
24 Continuity test with acoustic signal active
25 h (hours): unit of time
Position of the power pack adapter socket,
see also chapter 3.1.
Fuse for current measuring ranges (see chapter 9.3)
The device may not be disposed of with the trash.
Further information regarding the WEEE mark can be
accessed on the Internet at
www.gossenmetrawatt.com under the search term
WEEE (see also chapter 9.5).
Calibration mark (blue seal):
CAT II / III Measuring category II or I II device, see also
See also “Recalibration” on page 71.
“Measuring Categories and their Significance per IEC
61010-1” on page 8.
Continuous, doubled or reinforced insulation
Indicates EC conformity
▲
IR
▼
GMC-I Messtechnik GmbH15
Position of the infrared interface, window on the top of the
instrument
Initial Start-Up – Setup
Attention!
!
Note
3Initial Start-Up
3.1 Batteries
Be certain to refer to chapter 9.2 regarding correct battery
installation!
Momentary battery voltage can be queried in the “Info” menu (see
chapter 6.3).
Disconnect the instrument from the measuring circuit before opening the battery compartment lid in order to replace the batteries.
Operation with Power Pack (not included, see chapter 10.3)
Installed batteries are disconnected electronically if the NA X-TRA
power pack is used, and need not be removed from the
instrument. If rechargeable batteries are used, they must be
recharged externally.
If the external power supply is switched off, the device is switched
to battery operation without interruption.
3.2 Activation
Switching the Instrument On Manually
➭ Press the ON / OFF | LIGHT key until the display appears.
Power-up is acknowledged with a brief acoustic signal. As
long as the key is held depressed, all of the segments at the
liquid crystal display (LCD) are illuminated.
The LCD is depicted on page 13.
The instrument is ready for use as soon as the key is released.
Display Illumination
After the instrument has been switched on, background illumination can be activated by briefly pressing the ON / OFF | LIGHT key. Illumination is switched back off by once again pressing the same
key, or automatically after approximately 1 minute.
Switching the Instrument On with a PC
The multimeter is switched on after transmission of a data block
from the PC, assuming the “irStb” parameter has been set to “ir on” (see chapter 6.4).
However, we recommend using the power saving mode: “ir off”.
Electrical discharge and high frequency interference may
cause incorrect displays to appear, and may disable the
measuring sequence.
Disconnect the device from the measuring circuit. Switch the
instrument off and back on again in order to reset. If the
problem persists, briefly dislodge the battery from the
connector contacts (see also chapter 9.2).
3.3 Setting the Operating Parameters
Setting Time and Date
See “t iME” and “dAtE” parameters in chapter 6.4.
Display Modes for the Digital Display
Selection can be made from two different display modes
(see “0.diSP” parameter in chapter 6.4).
16GMC-I Messtechnik GmbH
ON / OFF
LIGHT
FUNC
ENTER
ON / OFF
LIGHT
Auto-Range
88.8.8.8
Illumination On
800.00
ON/ OFF
LIGHT
Illumination Off
800.00
ON/ OFF
LIGHT
0ff
ON/ OFF
LIGHT
long
(1 s)
3.4 Switching the Instrument Off
Switching the Instrument Off Manually
➭ Press the ON / OFF | LIGHT key until 0FF appears at the display.
Shutdown is acknowledged with a brief acoustic signal.
Automatic Shutdown
The instrument is switched off automatically if the measured value
remains unchanged for a long period of time (maximum measured
value fluctuation of approximately 0.8% of the measuring range
per minute or 1 C or 1 F per minute), and if none of the keys or
the rotary switch have been activated before a selected period of
time in minutes has elapsed (see “APoFF” parameter on page 53).
Shutdown is acknowledged with a brief acoustic signal.
Exceptions include:
Transmission and memory mode operation, continuous operation
and whenever a dangerous voltage is applied to the input
(U > 15 V AC or U > 25 V DC).
Disabling Automatic Shutdown
The instrument can be set to continuous operation.
➭ Simultaneously press the
The “Continuous On” function is indicated at by means of the
ON display to the right of the battery symbol.
The “Continuous On (on)” setting can be cancelled by changing the respective parameter or by switching the instrument off manually.
In this case the parameter will be reset to 10 min (see “APoFF” on
page 56).
and keys to this end.
Initial Start-Up – Setup
GMC-I Messtechnik GmbH17
Control Functions
4Control Functions
4.1 Selecting Measuring Functions and Measuring Ranges
4.1.1 Automatic Range Selection
The multimeter is equipped with auto-ranging for all measuring
functions, except for temperature measurement, and diode and
continuity testing. Auto-ranging is active as soon as the
instrument is switched on. The instrument automatically selects
the measuring range which allows for highest possible resolution
of the applied quantity. When the instrument is switched to
frequency measurement, the previously selected voltage
measuring range remains active.
AUTO-Range Function
The multimeter is switched automatically to the next higher range
at (3099 d + 1 d 0310 d), and to the next lower range at
(280 d - 1 d 2799 d).
With high resolution the multimeter is switched automatically to
the next higher range at (30999 d + 1 d 03100 d), and to the
next lower range at (2800 d - 1 d 27999 d).
4.1.2 Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges can be
selected manually in accordance with the following table by
pressing the MAN / AUTO button.
The desired measuring range can then be selected with the
or
key.
The instrument is automatically returned to range selection when
the MAN / AUTO key is pressed, the rotary switch is activated or the
instrument is switched off and back on again.
Overview: Auto-Ranging and Manual Range Selection
FunctionDisplay
MAN / AUTO
Range switching sequence for:
V:300 mV*
Hz: 300 Hz
:
300
or
MAN / AUTO
* Via manual measuring range selection only
A:
300A 3 mA 30 mA 300 mA 3A 10A
A : See chapter 5.9.3 und chapter 5.9.4
F:30 nF 300 nF 3F 30F 300F
M
:
300 k
@ISO
Return to automatic measuring range selection—
Manual mode active:
utilized measuring range is fixed
3 V 30 V 300 V 1000 V
3 kHz 30 kHz 300 kHz (Hz(U))
3 k
30 k
300 k
3 M
30 M
300 M
3 M
3000 M
30 M
MAN
MAN
The multimeter is held in the selected measuring range. If the
range limit is exceeded, 0L appears at the display. You should then
switch to the next higher measuring range with the help of the
key.
18GMC-I Messtechnik GmbH
Control Functions
Note
4.1.3 Quick Measurements
Measurements performed using a suitable fixed measuring range
are executed more quickly than those which utilize automatic
range selection. Quick measurement is made possible with the
following two functions:
• Manual measuring range selection, i.e. selection of the measuring
range with the best resolution (see chapter 4.1.2)
or
•With the DATA function (see chapter 4.4). In this way, the
appropriate measuring range is selected automatically after the
first measurement and the second measurement is executed
more quickly.
The selected measuring range remains active for the subsequent
series of measurements with these two functions.
4.2 Zero Offset / Relative Measurements
Zero balancing or a reference value for relative measurements can
be stored to memory depending upon deviation from the zero
point:
Deviation from zero
– with short-circuited measurement cables for V, , A
– with open input for capacitance unit of measure: F
0 to 200 digitsZERO REL
> 200 to 1500 digitsREL
Display
The relevant reference or correction value is deducted individually
for the respective measuring function as an offset from all future
measurements and remains in memory until deleted, or until the
multimeter is switched off.
Zero balancing and reference value adjustment can be used for
auto-ranging, as well as for manual measuring range selection.
Note: Zero balancing is not possibIe in switch position
M@UISO
.
Zero Balancing
➭ Plug the measuring cables into the instrument and connect the
free ends to each other, except for capacitance measurement
in which case the ends of the cables are not connected to
each other.
➭ Briefly press the ZERO | ESC key.
The instrument acknowledges zero balancing with an acoustic
signal, and the “ZERO REL” symbol appears at the LCD. The
value measured at the moment the key is pressed serves as a
reference value.
➭ Zero balancing can be cleared by once again pressing the
ZERO | ESC key.
As a result of TRMS measurement, the multimeter displays a
residual value of 1 to 10/35 digits with short-circuited
measurement cables as the zero point for V AC / I AC or
V(AC+DC) / I (AC+DC) measurements (non-linearity of the
TRMS converter). This has no influence on specified
accuracy above 1% of the measuring range (or 3% in the
mV, V(AC+DC) ranges).
Setting the Reference Value
➭ Plug the measuring cables into the instrument and measure a
reference value (max. 1500 digits).
GMC-I Messtechnik GmbH19
Control Functions
➭ Briefly press the ZERO | ESC key.
The instrument acknowledges storage of the reference value
with an acoustic signal, and the “ZERO REL” or the “REL”
symbol appears at the LCD. The value measured at the
moment the key is pressed serves as a reference value.
➭ The reference value can be cleared by once again pressing the
ZERO | ESC key.
Notes Regarding Relative Measurement
• Relative measurement effects the digital display only. The
analog display continues to read out the original measured value.
• In the case of relative measurement, F or AC quantities may
also appear as negative values.
4.3 Display (LCD)
4.3.1 Digital Display
4.3.2 Analog Display
Measured Value, Polarity
The analog display demonstrates the dynamic performance of a
moving-coil mechanism. This display is especially advantageous
for observing measured value fluctuation, and for balancing
procedures.
Display mode Pointer: the current measured value is tracked in
real-time.
The analog scale displays a negative range of 5 scale divisions for
the measurement of zero-frequency quantities, allowing for
precise observation of measured value fluctuation around zero. If
the measured value exceeds the negative range of 5 scale
divisions, polarity is reversed at the analog display.
Scaling of the analog scale is automatic. This is very helpful for
manual measuring range selection.
Measured Value, Unit of Measure, Type of Current, Polarity
The measured value with decimal and plus or minus sign appears
at the digital display. The selected unit of measure and current
type are displayed as well. A minus sign appears to the left of the
value during the measurement of zero-frequency quantities, if the
Overranging
Overranging in the positive range is displayed by means of the
right triangle symbol.
plus pole of the measured quantity is applied to the “” input.
The “0.d iSP” parameter can be used to determine whether lead-
ing zeros will be appear or be suppressed at the measured value
display (see chapter 6.4).
Refresh Rate
In the bar graph and pointer modes, the analog display is
refreshed 40 times per second.
Overranging
If the upper range limit of 3100 digits is exceeded “0L” (overload)
appears at the display.
Exceptions: “0L” appears at the display as of 1000.0 V in the case
of voltage measurement in the 1000 V range, as of 5.100 V for
diode testing, and as of 11.00 A in the 10 A range.
20GMC-I Messtechnik GmbH
Note
V, A, Hz
t [s]
100%
10%
Activated
Reactivated
Save
Save
3000 digits
of Measuring Range
F, %
Control Functions
4.4 Measured Value Storage: DATA (auto-hold / compare)
An individual measured value can be automatically “frozen” with
the DATA function (auto-hold). This is useful, for example, when
contacting the measuring points with the test probes requires
your full attention. After the measuring signal has been applied
and the measured value has settled in in accordance with the
“condition” listed in the table below, the measured value is frozen
at the digital display and an acoustic signal is generated. The test
probes can now be removed from the measuring points, and the
measured value can be read from the digital display. If the
measuring signal falls below the value specified in the table, the
function is reactivated for storage of the next value.
Measured Value Comparison (DATA Compare)
If the currently frozen value deviates from the first saved value by
less than 100 digits, the acoustic signal is generated twice. If
deviation is greater than 100 digits, only one brief acoustic signal
is generated.
The DATA function has no effect on the analog display, at
which the current measured value continues to appear.
DATA
Function
activatebriefblinksonce
save
(stabilized
measured
value)
reactivate
1)
Press
DATA /
Min/Max
However, when the digital display is “frozen”, the decimal
point is fixed as well (fixed measuring range, symbol: MAN).
The selected measuring range should not be manually
changed as long as the DATA function is active.
The DATA function is deactivated by pressing and holding the
DATA/MIN/MAX key (approx. 1 second), when the measuring
function is changed or when the instrument is switched off and
back on again.
GMC-I Messtechnik GmbH21
change to
Min/Max
exitlong
1)
Reactivation results from falling short of specified measured value limits.
2)
Two acoustic signals are generated the first time a measured value is saved as
a reference value. For subsequent data hold, two acoustic signals are only
generated if the currently frozen value deviates from the
than 100 digits.
Key: MV = measured value, R = measuring range
briefSee table in chapter 4.4.2
Condition
Measuring
Function
V, A, F, Hz, %> 10%
V, A, F, Hz, %< 10%
Measuring
Signal
of R
0L
of R
= 0L
Response from Instrument
Display
MV
Digital
is displayed
stored
MV
is cleared is cleared
first
saved value by less
DATA
static
blinks
Acou-
stic
once
twice
twice
2)
Control Functions
Example
The voltage measuring range is set manually to 30 V.
The first measured value is 5 V and is stored to memory because
it is greater than 10% of the measuring range (= 3 V), and is thus
reliable above the background noise level. As soon as the
measured values drops to less than 10% of the measuring range,
i.e. amounts to less than 3 V which corresponds to removal of the
test probes from the measuring point, the instrument is ready to
store a new value.
4.4.1 Function DATA during insulation measurement *
This DATA function differs from the standard DATA-Compare
function.
In the interference voltage measurement function V(ac+dc)1M
the special DATA function for insulation measurement is activated
by pressing the DATA key – DATA blinks in the display. During inter-
ference voltage measurement, no DATA values are determined or
stored yet. By briefly pressing the FUNC | ENTER key, the insulation
measurement is activated and the test voltage is applied.An intermittent acoustic signal indicates to the user that the test voltage is
applied (accompanied by a blinking of the ISO symbol in the display). After applying the test probes, a check is being made as to
whether a valid measured value is present.As soon as a stable
measured value appears in the display, the DATA value is stored
(frozen in the display). A long acoustic signal indicates that the
measurement is completed, the test voltage is switched off again,
DATA no longer blinks and the user can read and note down the
measured value. Byrepeatedlypressing the FUNC | ENTER key, the
test voltage is applied again and DATA is activated (DATA blinks in
the display).
If no DATA value is determined or if OL is shown in the display, the
test voltage is automatically switched off after approx. 10 seconds. In order to deactivate the DATA function, the FUNC | ENTER
key must be pressed and held.
* This function is especially available for variant METRAHIT ISO AERO,
for other variants as from firmware version 1.18.
22GMC-I Messtechnik GmbH
4.4.2 Saving Minimum and Maximum Values – MIN/MAX Function
Note
Minimum and maximum measured values applied to the
measuring instrument’s input after the Min/Max function has been
activated can be “frozen” at the display. The most important use
of this function is the determination of minimum and maximum
values during long-term measured value observation.
The Min/Max function can be activated in all measuring functions.
The Min/Max function has no effect on the analog display, at
which the current measured value continues to appear.
Apply the measured quantity to the instrument and set the
measuring range with the MAN / AUTO key before activating the
Min/Max function.
The Min/Max function is deactivated by pressing and holding the
DATA/MIN/MAX key (approx. 1 second), when the measuring
function is changed or when the instrument is switched off and
back on again.
As opposed to the DATA function, the Min/Max function can
also be used for temperature measurement.
Function
Min/Max
activate
and save
save and
display
return to 1
stoplongare deleted
Press
DATA /
Min/Max
1
2 x briefare saved
brief
2
brief
3
brief
Min. and Max.
Measured Values
storage continues
in background,
new min. and max.
values are displayed.
same as 1,
stored values are not
deleted
Control Functions
Response from Instrument
Display
Measured
Valu e
Digital
current
measured
value
saved min.
value
saved max.
value
same as 1
current
measured
value
Max.
Min.
max
and
min
min.
max.
same
as 1
is
deleted
Acous-
Signal
tic
twice
once
once
once
twice
GMC-I Messtechnik GmbH23
Control Functions
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
4.5 Measurement Data Recording
The multimeter is capable of recording measurement data using
an adjustable sampling rate for long periods of time in the form of
measurement series. Data are stored to a battery-backed memory module, and are retained even after the multimeter is switched
off. The system acquires measured values relative to real-time.
Stored measured values can subsequently be read out with the
help of METRAwin 10 software. The only prerequisite is a PC which
is connected by means of an interface cable to the
USB X-TRA bidirectional interface adapter, which is plugged onto
a multimeter. See also chapter 7, “Interface Operation”.
Memory Parameters Overview
ParameterPage: Header
CLEAr
EMpty
0CCUP
rAtE
StArt
StoP
25: Clear Memory
25: Clear Memory – appears after CLEAr
25: Querying Memory Occupancy
55: rAtE – set the sampling rate
24: Starting Recording via Menu Functions
25: Ending Recording
The STORE Menu Function
➭ First set the sampling rate for memory mode operation (see
chapter 6.4 the rAtE parameter) , and then start memory mode
operation.
➭ First select the desired measuring function and an appropriate
measuring range.
➭ Check the battery charge level before starting long-term
measurement recordings (see chapter 6.3).
Connect the NA X-TRA power pack if required.
Starting Recording via Menu Functions
➭ Switch to the “SETUP ” mode by pressing MEASURE | SETUP, and
select the “StorE ” menu.
... StorE 000.0 % StArt
1nFo
➭ Memory mode operation is started by activating FUNC | ENTER.
STORE appears underneath the analog display and indicates
that the memory mode has been activated. “StoP ” appears at
the digital display.
➭ Press MEASURE | SETUP in order to return to the measuring
function.
24GMC-I Messtechnik GmbH
Control Functions
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
During Recording
STORE is displayed underneath the analog display during memory
mode operation, and
StoP
000.3%
memory occupancy
can be monitored:
The following message appears as soon as memory is full:
“100.0%”.
In order to be able to observe measured values during recording,
switch to the measuring function by pressing MEASURE | SETUP.
The display is returned to the memory menu after once again
pressing MEASURE | SETUP.
A new memory block is created when another measuring function
is selected with the rotary switch or the FUNC | ENTER key. Data
storage then continues automatically.
Ending Recording
➭ “StoP” appears at the display after pressing MEASURE | SETUP.
StoP start
➭ Acknowledge the “StoP ” display by pressing FUNC | ENTER.
STORE is cleared from the display indicating that recording has
been ended.
➭ Press MEASURE | SETUP in order to return to the measuring
function.
➭ Memory mode operation can also be exited by switching the
multimeter off.
Querying Memory Occupancy
Memory occupancy can be queried during recording with the
help of the “1nFo ” menu (see also chapter 6.3).
Memory occupancy range: 000.1% to 099.9%.
1nFo
bAtt:
...
0CCUP
%:
017.4
%
Memory occupancy can be queried before recording is started via
the “StorE ” menu.
... StorE017.4 % StArt
1nFo
Clear Memory
This function deletes all measured values from memory!
This function cannot be executed during memory mode
operation.
... StorE017.4 % StArt
1nFo
CLEArempty
GMC-I Messtechnik GmbH25
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
@UISO
, A/Hz
5Measurements
5.1 Voltage Measurement
Notes Regarding Voltage Measurement
• The multimeter may only be operated with installed batteries.
Dangerous voltages are otherwise not indicated, and the instrument
may be damaged.
• The multimeter may only be operated by persons who are capable of recognizing contact hazards and taking the appropriate
safety precautions. Contact hazards exist anywhere, where
voltages of greater than 33 V RMS may occur.
The test probes may only be only gripped up to the finger
Do not touch the metallic test probes under any
guard.
circumstances.
• Avoid working alone when taking measurements which involve
contact hazards. Be certain that a second person is present.
• Maximum allowable voltage between terminals 9 or 10 and
ground (8) is 1000 V for measuring category I I, and 600 V for
measuring category III.
• Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
• No measurements may be made with this instrument in
electrical circuits with corona discharge (high-voltage).
• Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be present.
• Be aware of the fact that dangerous voltage spikes are not displayed
during measurement with the low-pass filter.
We recommend measuring voltage without the low-pass filter first, in
order to be able to detect any dangerous voltages.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8, “Technical Data”, in the table entitled
“Measuring Functions and Measuring Ranges” in the “Overload Capacity” column.
•
Rotary selector switch position “V1M
/ M
@UISO
” is
available for the detection of interference voltage during
insulation resistance measurement.
Use switch position V, V or V in order to perform
precise voltage measurements.
26GMC-I Messtechnik GmbH
5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC)
Note
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Tem p
MISO
– (+)
+ (–)
V= : 100 V1000 V
020.00
V
DC
AC
TRMS
020.00
V
DC
V
V
V
V
> 15 V AC or > 25 V DC:
> 1000 V:
Warnings regarding dangerous voltage:
Max. 1000 V, 3 kHz
V: 10 mV1000 V
Hz: 1Hz 300 kHz
P
max
= 3 x 106 V x Hz
Measuring Ranges:
230.0
FUNC
ENTER
MEASURE
SETUP
... SET ... CLIP
FUNC
ENTER
FUNC
ENTER
0ff
5 ranges: 300 mV, 3 V, 30 V,
300 V and 1000 V
➭ Set the CL iP parameter to 0FF in the current clip setup menu.
Otherwise all measured values are displayed in amperes,
corrected by the amount resulting from the selected transformation ratio for an interconnected clip-on current sensor.
... SETrAtE ... CLIP
1nFo
1 / 10/100/1000 / 0ff
➭ In accordance with the voltage to be measured, turn the rotary
switch to V or V .
➭ Connect the measurement cables as shown.
The “” connector jack should be grounded.
An intermittent acoustic signal warns the operator if the
measured value exceeds the upper range limit in the 1000 V
range.
Make sure that a current measuring range (“A”) has not been
activated when the multimeter is connected for voltage
measurement! If the fuse’s blowing limits are exceeded as a
result of operator error, both the operator and the instrument are
in danger!
With the rotary switch in the V position, the multimeter is always in
the 1 V measuring range immediately after it is switched on. As
soon a the MAN / AUTO key is pressed, and assuming the
measured value is less than 280 mV, the multimeter is switched to
the mV measuring range.
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
GMC-I Messtechnik GmbH27
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
Attention!
!
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
@UISO
, A/Hz
5.1.2 Alternating Voltage and Frequency Measurement V AC and Hz
with Selectable Low-Pass Filter
➭ Set the CL iP parameter to 0FF in the current clip setup menu.
Otherwise all measured values are displayed in amperes,
corrected by the amount resulting from the selected transformation ratio for an interconnected clip-on current sensor.
1nFo ... SETrAtE ... CLIP
1 / 10/100/1000 / 0ff
➭ In accordance with the voltage or frequency to be measured,
turn the rotary switch to V~ or Hz/%.
➭ Connect the measurement cables as shown.
The “” connector jack should be grounded.
Frequency Measurement
➭ Connect the measured quantity in the same way as for voltage
measurement.
➭ Manually select the measuring range for the voltage amplitude.
When the instrument is switched to frequency measurement,
the previously selected voltage measuring range remains active.
➭ You can switch back and forth between frequency
measurement with and without low-pass filter.
Repeatedly press the FUNC | ENTER multifunction key, until the
Hz or Hz/Fil unit of measure appears at the display.
Lowest measurable frequencies and maximum allowable
voltages are included in chapter 8, “Technical Data”.
Measurement with Low-Pass Filter
Voltage Measurement
Be aware of the fact that dangerous voltage spikes are not
An intermittent acoustic signal warns the operator if the
measured value exceeds the upper range limit in the 1000 V
range.
Make sure that a current measuring range (“A”) has not been
activated, when the multimeter is connected for voltage
measurement! If the fuse’s blowing limits are exceeded as a
result of operator error, both the operator and the instrument are
in danger!
➭ You can switch back and forth between voltage measurement
with and without low-pass filter.
➭ Repeatedly press the FUNC | ENTER multifunction key, until the V
displayed during this type of measurement (see also “Voltage Comparator”. We recommend measuring voltage
without the low-pass filter first, in order to be able to detect
any dangerous voltages.
A 1 kHz/-3 dB low-pass filter can be activated if required, in order
to filter out capacitively induced high frequency pulses of greater
than 1 kHz, for example when performing measurements at
cables, i.e. undesired voltages of greater than 1 kHz can be
suppressed.
“Fil” appears at the display in order to indicate the respectively
activated low-pass filter. The multimeter is automatically switched
to manual measuring range selection.
or V/Fil unit of measure appears at the display.
28GMC-I Messtechnik GmbH
Specified measuring accuracy is not reached with signals of
COM mA A
V
Tem p
MISO
Max. 1000 V, 3 kHz
V~: 10 mV 1000 V
Hz: 1Hz 300 kHz
~
P
max
= 3 x 106 V x Hz
Measuring Ranges:
> 15 V AC or > 25 V DC:
> 1000 V:
Warnings regarding dangerous voltage:
230.0
5 ranges: 300 mV, 3 V, 30 V,
300 V and 1000 V
0230.6
V
AC
TRMS
0050.3
Hz
AC
V~
Hz / %
V~
Hz
0229.9
V
AC
TRMS
Fil
V~ with Filter
1 kHz
V~
0050.0
AC
Fil
Hz
Hz with Filter
!
!
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
1 kHz
Hz
MEASURE
SETUP
... SET ... CLIP
FUNC
ENTER
FUNC
ENTER
0ff
%
%
Hz: vpm 0FF
RPM: vpm 0FF
The following function is
only available with customer-specific variants.
greater than 500 Hz when the filter is active.
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
Voltage Comparator for Displaying Dangerous Voltage
The input signal or measuring signal is checked by a voltage
comparator for dangerous spikes, because these do not appear
at the display when the low-pass filter is used.
At voltages of greater than 15 V AC or 25 V DC, a danger symbol
appears at the display: .
Duty Cycle Measurement
(Function not available for METRAHIT ISO)
➭ Set the rotary switch to V~.
➭ Repeatedly press the FUNC | ENTER multifunction key until %
appears at the display.
➭ Connect the measurement cables as shown.
Make sure that a current measuring range (“A”) has not been
activated, when the multimeter is connected for frequency or
duty cycle measurement!
The ratio of pulse duration to pulse period is measured with
periodic square-wave signals and displayed as a percentage.
@UISO
, A/Hz
The applied frequency must remain constant during duty
cycle measurement.
30GMC-I Messtechnik GmbH
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
RPM
Revolutions
min
----------------------------
Pulses
Revolution
-------------------------§
x
60s
s
--------
=
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
@UISO
, A/Hz
RPM Measurement (function only available with customer-specific variant)
Revolutions per minute (also known as rotational frequency) are
measured by acquiring pulses. As a prerequisite for this measurement, the number of measurable pulses per revolution must first
be set in the RPM setup menu (vpm0FF, see below).
➭ Set the rotary switch to V~.
➭ Repeatedly press the FUNC | ENTER multifunction key until RPM
is briefly displayed. The measured value then appears in RPM,
for example “244.3 r”.
Measured value RPM = revolutions per minute.
Parameter vpm = pulses per revolution.
Pulses per Revolution Setup Menu
... SETrAtE ... vpm
1nFo
0ff/1 : 0.1 ... 1000
5.2 Resistance Measurement,
➭ Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
➭ Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Refer to section 5.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to “”.
➭ Connect the device under test as shown.
Use short or shielded measurement cables in the case of
high-impedance resistance.
Improving Accuracy by means of Zero Balancing
Cable resistance and contact resistance can be eliminated in all
measuring ranges by means of zero balancing (see section 4.2).
Temperature measurement is performed with a Pt100 or Pt1000
resistance thermometer, or a type K thermocouple (accessory,
not included), which is connected to the voltage input.
Selecting the Unit of Measure for Temperature
1nFo
... tEMP°C / °F
(°C = default setting)
5.3.1 Measurement with Resistance Thermometers
➭ Set the rotary switch to “Temp
RTD
”.
Press the FUNC | ENTER key in order to change to the other measuring function if required.
The sensor type, i.e. Pt100 or Pt1000, is detected automatically
and displayed.
There are two different ways to compensate for cable resistance:
Automatic Compensation
➭ Acknowledge by pressing the ZERO | ESC key.
“Short leads” appears at the display.
If you prefer to enter cable resistance directly, you can skip the
following entry prompt.
➭ Short circuit the measuring instrument’s connector cables.
“000.0” appears at the display. After pressing the FUNC | ENTER
key, automatic compensation of cable resistance is
activated for all subsequent measurements. The short-circuit
can now be eliminated, and the device is ready for use.
Enter Cable Resistance
➭ Press the ZERO | ESC key once again in the automatic
compensation menu.
➭ Enter the known resistance of the connector cables with the
scroll keys:
Select the digit to be changed with the
keys, and change
the respectively selected digit with the keys. The default
value is 0.43 . Values can be selected within a range of 0 to
50 .
➭ Upon pressing the FUNC | ENTER key, the selected value is acti-
vated and the display is returned to the measuring function.
Cable resistance remains in memory even after the instrument
has been switched off.
GMC-I Messtechnik GmbH33
Measurements: V/Hz, , , Temp (°C/°F), , M
COM mA A
V
Tem p
MISO
0025.6
C
Pt1000
RTD
0000.0
Short Leads
r000.5
Enter Cable Resistance
Automatic
RTD
Compensation
Measuring Ranges
RTD
Pt100
–200.0+850.0 C
RTD
Pt1000
–150.0+850.0 C
FUNC
ENTER
ZERO
ESC
ZERO
ESC
FUNC
ENTER
Te mp
RTD
Tem p
TC
@UISO
, A/Hz
34GMC-I Messtechnik GmbH
Note
Note
COM mA A
V
Tem p
MISO
Te mp
RTD
0023.3
C
Tem p
TC
Typ e K
TC
TE (TC)
0256.0
C
Pt1000
RTD
Measuring Range
TE (TC)
K (NiCr-Ni)
–250.0 +1372.0 C
FUNC
ENTER
5.3.2 Measurement with Thermocouples, Temp TC
RTD
”.
➭ Set the rotary switch to “Temp
The last selected temperature measurement or the last
selected temperature sensor, i.e. type K or Pt100/Pt1000,
remains in memory and is accordingly displayed. Press the
FUNC | ENTER key in order to change to the other
measuring function if required.
➭ The reference temperature is measured at the internal
reference junction (see parameter “1tEMP” in Seite 54 regarding
querying).
The internal reference temperature (temperature of the
internal reference junction) is measured by a temperature
sensor inside of the instrument. This may be somewhat
above room temperature as a result of internal heat-up, or
moving from warmer to colder surroundings or vice versa.
➭ Connect the sensor to the two accessible jacks. The
instrument displays the measured temperature using the
selected unit of measure.
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
GMC-I Messtechnik GmbH35
Measurements: V/Hz, , , Temp (°C/°F), , M
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Tem p
MISO
0V !
!
0.443
V
DC
R < 1, 10, 20, 30, 40, 90
FUNC
ENTER
R: 300
Measuring Range:
000.8
(0.1 to 310 )
5.4 Continuity Test
➭ Disconnect supply power from the electrical circuit of the de-
vice to be measured, and discharge all high-voltage capacitors.
➭ Make sure that the device under test is voltage-free.
Depending upon the selected limit value, the multimeter
generates a continuous acoustic signal in the case of continuity or
short-circuiting, i.e. at a value of less than the selected limit value.
“0L” appears at the display in the case of an open connection.
The limit value can be adjusted in the “SET ” menu (see also sec-
tion 6.4):
@UISO
, A/Hz
1nFo ... SETrAtE ... bEEP
1, 10, 20, 30, 40, 90
(10 = default setting)
36GMC-I Messtechnik GmbH
5.5 Diode Testing with Constant Current of 1 mA
Note
COM mA A
V
Tem p
MISO
Forward DirectionReverse Direction
0V !
0.L
to 5.100 V
Measuring Range:
+
–
0.443
V
!
000.8
DC
FUNC
ENTER
➭ Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
➭ Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Refer to section 5.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to “”.
➭ Press the FUNC | ENTER key and the diode symbol appears at
the display.
➭ Connect the device under test as shown.
Conducting Direction and Short-Circuit
The instrument displays forward voltage in volts (display: 4
places). As long as voltage drop does not exceed the maximum
display value of 5.1 V, several series connected components or
reference diodes with small reference voltages, as well as
Z-diodes and LEDs, can be tested.
Reverse Direction and Interruption
The measuring instrument indicates overload 0L.
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
Resistors and semiconductor paths connected in parallel to
the diode distort measurement results!
Disconnect supply power from the electrical circuit of the device to
be measured, and discharge all high-voltage capacitors.
➭ Make sure that the device under test is voltage-free.
Capacitors must always be discharged before measurement is
performed.
Interference voltages distort measurement results!
Refer to section 5.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to “”.
➭ Connect the (discharged!) device under test to the sockets
with the measurement cables as shown.
The “–” pole of polarized capacitors must be connected to the
“” jack.
Resistors and semiconductor paths connected in parallel to
the capacitor distort measurement results!
@UISO
, A/Hz
38GMC-I Messtechnik GmbH
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
Note
Note
COM mA A
V
Tem p
MISO
00.00
Short circuit the measurement cables.
R
x
0V !
!
M
@UISO
V
Interference Voltage Detection
000.0
V
DC
AC
TRMS
1M
ISO 500 V
@UISO
, A/Hz
5.7 Insulation Resistance Measurement – M
5.7.1 Preparing for Measurement
Testing the Measurement Cables
The test probes at the ends of the measurement cables
should be short circuited before performing insulation
resistance measurements with the selector switch in the
or the position, in order to make sure that a value close
to 0 is displayed at the instrument. Incorrect connection or
a broken measurement cable can be detected in this way.
@UISO
Insulation resistance may only be measured at voltage-free
devices. The measurement cables may not come into contact with one another during high-resistance insulation
measurements.
M
➭ Set the rotary switch to “
➭ Connect the measurement cables to the two accessible jacks.
➭ Interference voltage measurement (V AC+DC TRMS) is
conducted in this switch position.
The FUNC | ENTER key for U
order to start the insulation resistance measurement if the
device under test is voltage-free.
ISO ON / OFF may only be pressed in
@UISO
”.
GMC-I Messtechnik GmbH39
The M
tion resistance measurement and interturn short-circuit
detection (METRAHIT COIL). However, if interference voltage is
inadvertently applied with the switch in this position, it
appears at the display. If an interference voltage of 50 V is
present, insulation resistance measurement is disabled. The
LCD continues to display the interference voltage value. If a
voltage of greater than 1000 V is present, an acoustic signal
is generated as well.
switch position may only be used for insula-
@UISO
Measurements: V/Hz, , , Temp (°C/°F), , M
Caution: High-Voltage!
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MAN / AUTO
DATA/MIN/MAX
... 500 V10 V ...
M
@UISO
V
1M
Autorange
002.00
M
M
ISO
FUNC
ENTER
UISO ON/ OFF
000.0
V
Activate
Test Vol t a g e
DC
AC
TRMS
500 V
ISO 500 V
M
@UISO
V
1M
@UISO
, A/Hz
The selected test voltage appears at the display during testing.
Do not touch the conductive ends of the test probes
instrument has been activated for the measurement of insulation
resistance.
You may otherwise be exposed to a current of
2.5 mA (limited in the measuring instrument), and although
this is not life endangering, the resulting electrical shock is
quite discernible. If, on the other hand, measurement is
being performed on a capacitive device under test, for
when the
5.7.2 Performing the Insulation Measurement
➭ Starting the insulation resistance measurement:
Press and hold the U
ISO ON / OFF key until the display settles in,
in order to measure insulation resistance.
Insulation resistance measurement is ended by releasing the
key.
example a cable, it may be charged with up to approximately 1200 V. Touching the device under test after measure-
ment has been performed is life endangering in this case!
Selecting the Test Voltage (U
= 50 to 1000 V*)
ISO
The desired test voltage can be selected in the “SET” menu (see
also section 6.4):
1nFo ... SETrAtE ... v1so
... 500, 1000 V*
* Selectable test voltages and the default value depend upon the
respective device variant.
Alternatively, the requested test voltage can be selected via the
cursor keys
V
1M
. Prerequisite: the multimeter is in switch position
and in the AUTO range function.
40GMC-I Messtechnik GmbH
Auto-ranging is active during insulation resistance measurement.
For the automatic freezing of valid measured values, a DATA function specially geared to insulation measurement can be activated,
see section 4.4.1.
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
Attention!
!
Note
Note
+
–
0V !
Discharge
025.0
UISO
V
DC
AC
TRMS
ISO 500 V
@UISO
, A/Hz
Automatic Recognition of Interference Voltage
5.7.3 Ending the Measurement and Discharging
during Insulation Resistance Measurement
If the instrument detects interference voltage of greater than
15 V AC or > 25 V DC during insulation testing (condition: U
, e.g. Riq < 100 k at 100 V, see Seite 62, footnote 1), Error
U
ISO
interference
is briefly displayed at the LCD. The instrument is then
automatically switched to voltage measurement, and the currently
measured voltage value is displayed.
➭ Briefly press the U
ISO ON / OFF key.
After measurement has been completed, any remaining residual
A dead zone results in erroneous measurements for
automatic interference voltage detection during insulation
resistance measurement. The dead zone lies within a range
of 80% to 120% of the selected test voltage (in the case of
an interference voltage whose value is equal to that of measuring voltage, the two voltages neutralize each other).
voltage is displayed which may result from cable capacitance.
The instrument’s internal 1 M resistor causes rapid discharging.
However, contact to the device under test must be maintained.
The falling voltage value can be observed directly at the LCD.
Do not disconnect the device under test until the voltage value has
dropped to < 25 V!
Manual switching to insulation resistance measurement is
disabled for as long as voltage is applied to the test terminals.
If interference voltage is no longer present, the M
@ISO
measure-
ment can be started by once again pressing the UISO ON / OFF key.
The instrument’s batteries are rapidly depleted during insulation resistance measurement. Deactivate insulation resistance measurement between measurements for this reason.
Use only alkaline manganese batteries in accordance with
IEC 6 LR61.
If Error appears at the display, the cable (the device under
test) is most likely capacitively charged to a significant extent. Remedy: Short circuit the cable (the device under test).
Repeat the measurement.
Rotary selector switch position “V1M
/ M
able for the detection of interference voltage during insulation resistance measurement.
Special case indication of low insulation resistances up to short circuit
First „Error“ is displayed, and then briefly „Short“, followed by the
current resistance measured value.
GMC-I Messtechnik GmbH41
Use switch position V, V or V in order to perform
precise voltage measurements.
@UISO
” is avail-
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
COM mA A
V
Tem p
MISO
0V !
!
M
@UISO
V
1M
COIL
V
U
W
Example
COIL
METRAHITCOIL
COIL
COIL
TEST ADAPTER
Star Connection
M
Connect COIL TEST ADAPTER
and Measuring Cables
UISO
1200 V !
> 500 A !
(Pulse)
M
@UISO
V
Interference Voltage Detection
000.0
V
DC
AC
TRMS
1M
1000V
COIL
@UISO
, A/Hz
5.8 Interturn Short-circuit Measurement – Function
This function can only be performed with the METRAHIT COIL device variant and only in combination with the COIL TEST ADAPTER.
With this measurement, a partially periodic time value is determined in proportion to the inductivity of the respective winding for
each motor coil or coil combination via a cyclical discharge at high
voltage. A comparison of the measurement results allows for examining the symmetry of the motor coils and the detection of a
potential interturn short-circuit as a result.
5.8.1 Preparing the Measurement
COIL/M
@UISO
Interturn short-circuit measurements may only performed on
voltage-free windings.
➭ Connect the
COIL TEST ADAPTER
with the two enabled jacks of
the multimeter via the contact-protected plugs at the adapter
connection cables: the red cable with the COIL jack and the
black cable with the COM jack.
COIL/M
➭ Set the rotary switch at „
@UISO
“.
➭ Connect the two measurement cables with the two jacks of
the
COIL TEST ADAPTER (marked with a motor symbol).
➭ Establish contact between the inductive DUT and the (alligator)
clips applied to the test probes. In the case of 3-phase current
motors, for example, contact the winding connections U–V, V–
W und U–W successively.
➭ An interference voltage measurement V AC+DC TRMS is con-
ducted in this switch position.
Only if the DUT is voltage-free, are you allowed to press key UON / OFF for interturn short-circuit measurement (and hold it until
the measured value is stable).
42GMC-I Messtechnik GmbH
COIL
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
Caution: High-Voltage!
Attention!
!
@UISO
, A/Hz
The switch position
COIL/M
may only be used for
@UISO
interturn short-circuit measurement. However, any interference voltage accidentally applied is indicated in this switch
position.
If an interference voltage of more than 50 V is present in the
equipment, interturn short-circuit measurement is disabled.
The LCD panel continues to display the interference voltage
and the „Caution: High-Voltage“ symbol. If a voltage of more
than 1000 V is applied, an additional acoustic signal is generated.
Do not touch the conductive ends of the test probes as long as
the U
COIL ON / OFF key is being pressed.
A voltage of up to 1,200 Volt is present at the two enabled output jacks of the multimeter as well as, in the form of pulsed
voltage, at the two output jacks of the
COIL TEST ADAPTER
(marked with a motor symbol).
Multimeter: A 2.5 mA current (limited in the mesuring instrument) may flow through your body which, although it does
not reach any life-threatening values, is perceptible nevertheless as a slight electric shock.
COIL TEST ADAPTER
: Pulsed current of more than 500 A may
be present at the output jacks.
The DUT may charge up: Wait after each mesurement until the
voltage at the DUT has decreased (interference voltage indication). If not, touching the DUT may pose a danger to life!
Test voltage (U
= 1000 V)
ISO
The test voltage of the interturn short-circuit measurement has
been fixed at 1000 V and cannot be changed. The display shows
the momentary test voltage during measurement.
➭ Contact the winding in question (e.g. U–V) to check for ab-
sence of voltage.
Do not measure without support, but rather with a self-retaining contact, for example alligator clips. Poor contact
may lead to flying sparks, and slipping off the DUT may be
life-threatening!
➭ Activating interturn short-circuit measurement:
Press and hold the U
COIL ON / OFF key for measurement until the
display of the time value has stabilized. In the display, a test
voltage of 1000 V blinks.
➭ Read off the numeric value (indicated in µsec) and note down
the value together with the contacted winding.
➭
Release the
U
COIL
ON / OFF
key to stop the measurement.
GMC-I Messtechnik GmbH43
Measurements: V/Hz, , , Temp (°C/°F), , M
0120.0
ZERO
ESC
U
COIL ON
/ OFF
000.0
V
Apply Test Voltage
DC
AC
TRMS
M
@UISO
V
1M
COIL
Note Winding & Measured Value
μ
1000Vsec
1000V
035.0
μ
035.0
μ
035.0
μ
1
2
3
U-V
V-W
U-W
035.0
μ
4
. . .
1000Vsec
1000Vsec
REL
1000Vsec
REL
1000Vsec
REL
➭ Allow the winding to be discharged via the multimeter before
removing the contact, see Kap. 5.8.3.
➭ Connect the next winding (e.g. V–W) and repeat the measuring
procedure as described above.
@UISO
, A/Hz
Signalling of Measuring Progress
For better guidance, the first
three test steps are marked
accordingly.
On starting the first measurement, a (horizontal) segment
left of the numeric field is
shown.
On starting the second measurement, two segments are
shown.
On starting the third measurement, three segments are
shown. From the fourth measurement onwards, the three
segments are shown intermittently.
44GMC-I Messtechnik GmbH
Measurements: V/Hz, , , Temp (°C/°F), , M
Note
+
–
0V !
Discharge
025.0
U
ISO
V
DC
AC
TRMS
1000V
@UISO
, A/Hz
Automatic Evaluation of the Measuring Results
With the start of the second measurement, the automatic evaluation of the measuring results begins. The analog display* shows
the maximum deviation in percent from all preceding measurements of the U
measurement series. This allows for a direct
COIL
comparison between the first and the subsequent measurements. By establishing a relation between the third measurement
and the results of the preceding measurements, you automatically
receive a concluding comparison of the 3 measurements.
Alternatively, you can measure at all 3 motor coils (star or delta
connected), note down the values and compare them with each
other for conclusion.
The admissible asymmetry depends on the motor type:
A motor with short-circuit rotor will not display any major asymmetry (typically 1%), whereas a permanently excited motor may
display a certain degree of asymmetry, depending on the rotor
position. At a deviation of more than 10 %, however, there is definitely a fault in the DUT (e.g. interturn short-circuit).
If a measured value is equal to 0, there is a short-circuit. If there is
no discharge, the measured coil is open.
In order to start a new measurement series, briefly press the
MEA-
SURE | SETUP key twice or switch to another function (via rotary
switch operation).
5.8.3 Ending the Measurement and Discharging
➭ Release the UCOIL ON / OFF key.
After measurement has been completed, any remaining residual
voltage is displayed which may result from cable capacitance.
The instrument’s internal 1 M resistor causes rapid
discharging.
The contact with the motor coils must be maintained.
The falling voltage value can be observed directly at the LCD.
Do not disconnect the device under test until the voltage value has
dropped to < 25 V!
Rotary selector switch position “V1M
/ M
@UISO
” is
available for the detection of interference voltage prior to
interturn short-circuit measurement.
Use switch position V, V or V in order to perform
precise voltage measurements.
* the deviation is presented as a bargraph in function COIL
(default setting), additionally REL is shown.
GMC-I Messtechnik GmbH45
Measurements: V/Hz, , , Temp (°C/°F), , M
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
@UISO
, A/Hz
5.9 Current Measurement
Notes Regarding Current Measurement
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents are otherwise not
indicated, and the instrument may be damaged.
• Set up the measuring circuit in a mechanically secure fashion,
and secure it against inadvertent breaks. Select conductor
cross-sections and lay out connections such that they do not
overheat.
• An continuous acoustic signal warns of current greater than
11 A.
• The input for the current measuring range is equipped with a
fuse link. Maximum permissible voltage for the measuring
circuit (= rated voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only! The fuse must have a breaking capacity of at least 30 kA.
• If the fuse for the active current measuring range blows, “FUSE”
appears at the digital display, and an acoustic signal is
generated at the same time.
• If a fuse should blow, eliminate the cause of overload before
placing the instrument back into service!
• Fuse replacement is described in section 9.3.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8, “Technical Data” in the table entitled
“Measuring Functions and Measuring Ranges” in the “Overload Capacity” column.
Direct Current Measurement – CLIP Parameter = OFF
➭ Set the CL iP parameter to 0FF in the current clip setup menu.
Otherwise all displayed measured values are corrected by the
amount resulting from the selected transformation ratio for an
interconnected clip-on current sensor.
... SETrAtE ... CLIP
1nFo
1 / 10/100/1000 / 0ff
Switch
A0/2 A DC
A1A DC AC TRMS
A~0/2 A~ AC TRMSAC (A) clip: clip-on current transformer
Hz (A)1Hz ~ AC Hz (A) clip: clip-on current transformer
FUNC
DisplayAdditional clip function (via CLIP 1:1/10/
100/1000 in the SET menu)
46GMC-I Messtechnik GmbH
5.9.1 Direct and Pulsating Current Measurement, Direct Connection,
COM mA A
V
Tem p
MISO
~
R
x
~
R
x
1
3
2
!
A
A
003.50
mA
DC
003.50
mA
DC
AC
TRMS
A
A
A: 0.1 A . . . 11 A
A: 10 A . . . 11 A
6 ranges: 300 A, 3 mA, 30 mA, 300 mA, 3 A and 10 A
88.8.8.8
!
I > 11 A
Current
Measuring Ranges:
Current measurement
may only be performed
with installed batteries!
FUNC
ENTER
MEASURE
SETUP
... SET ... CLIP
FUNC
ENTER
FUNC
ENTER
0ff
A DC and A (DC+AC)
➭ First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
➭ In accordance with the current to be measured, set the rotary
switch to A or A
➭ Select the type of current appropriate for the measured
quantity by briefly pressing the FUNC | ENTER multifunction key.
Each time the key is pressed, the instrument is switched back
and forth between A DC and A (DC + AC)
cated by means of an acoustic signal. The current type is indicated at the LCD by means of the DC or the (DC+AC)
symbol.
➭ Safely connect the measuring instrument (without contact
resistance) in series to the power consumer (2) as shown.
➭ Switch supply power to the measuring circuit back on (3).
➭ Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
➭ Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
➭ Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
.
, which is indi-
TRMS
TRMS
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
GMC-I Messtechnik GmbH47
Measurements: V/Hz, , , Temp (°C/°F), , M
COM mA A
V
Tem p
MISO
~
R
x
~
R
x
1
3
2
A~
Hz
A~
Hz
003.50
mA
AC
TRMS
0050.1
Hz
AC
88.8.8.8
!
!
I > 11 A
Current measurement
may only be performed
with installed batteries!
Current
FUNC
ENTER
Measuring Ranges:
Hz: 1 Hz . . . 31 kHz
A ~: 10 A . . . 11 A
6 ranges: 300 A, 3 mA, 30 mA, 300 mA, 3 A and 10 A
3 ranges: 300 Hz, 3 kHz and 30 kHz
MEASURE
SETUP
... SET ... CLIP
FUNC
ENTER
FUNC
ENTER
0ff
5.9.2 Alternating Current and Frequency Measurement,
Direct Connection, A AC and Hz
➭ First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
➭ In accordance with the current or frequency to be measured,
turn the rotary switch to A~ or Hz.
➭ Select the desired measured quantity by briefly pressing the
FUNC | ENTER multifunction key. Each time the key is pressed,
AC
TRMS
switching is acknowledged with an acoustic signal.
➭ Safely connect the measuring instrument (without contact
resistance) in series to the power consumer as shown.
➭ Switch supply power to the measuring circuit back on (3).
➭ Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
➭ Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
➭ Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
and Hz are alternately selected, and
@UISO
, A/Hz
48GMC-I Messtechnik GmbH
5.9.3 Direct and Pulsating Current Measurement
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Tem p
MISO
~
R
x
Current
0035.0
A
1:1000
A
DC
0035.0
A
1:1000
A
DC
AC
TRMS
FUNC
ENTER
R
i
Ri ~ 9 M
V
V
MEASURE
SETUP
... SET ... CLIP
FUNC
ENTER
FUNC
ENTER
0ff
Measuring Ranges:
mA / A: see table
with Clip-On Current Sensor A DC and A (DC+AC)
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(V input), all current displays appear with the correct value in
accordance with the selected transformation ratio. The only
prerequisite is that the current sensor is equipped with at least
one of the below listed transformation ratios, and that the ratio
has been previously selected in the following menu (Clip0FF).
Current Clip Setup Menu
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
... SETrAtE ... CLIP
1nFo
0ff/1 / 10/100/1000
Tra n sf. R a t io
CL IP
1:1
1 mV / 1 mA
1:10
1m V / 10 mA
1:100
1m V / 100 mA
1:1000
1 mV/1 A
Maximum allowable operating voltage is equal to the current
Measuring RangesClip Type
300 mV3 V30 V
300.0 mA3.000 A30.00 AWZ12C
3.000 A30.00 A300.0 A
30.00 A300.0 A3,000 kA
300.0 A3,000 kA30.00 kA
transformer’s nominal voltage.
additional error resulting from the clip-on current sensor must also be
taken into consideration.
(default value: Clip = 0ff = voltage display)
GMC-I Messtechnik GmbH49
When reading the measured value,
WZ12B,
Z201A/B
Z202A/B
METRAFLEX
Z202A/B, Z203A/B,
WZ12C
METRAFLEX
Measurements: V/Hz, , , Temp (°C/°F), , M
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Tem p
MISO
0035.0
A
AC
TRMS
A~
Hz
1:1000
0050.1
Hz
AC
1:1000
FUNC
ENTER
V~
Hz
~
R
x
Current
Ri ~ 9 M
R
i
Measuring Ranges:
Hz: 1 Hz . . . 31 kHz
MEASURE
SETUP
... SET ... CLIP
FUNC
ENTER
FUNC
ENTER
0ff
mA / A: see table
3 ranges: 300 Hz, 3 kHz and 30 kHz
@UISO
, A/Hz
5.9.4 Alternating Current Measurement with Clip-On Current Sensor,
A AC and Hz
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(V input), all current displays appear with the correct value in
accordance with the selected transformation ratio. The only
prerequisite is that the current sensor is equipped with at least
one of the below listed transformation ratios, and that the ratio
has been previously selected in the following menu (Clip0FF).
Current Clip Setup Menu
1nFo
0ff/1 / 10/100/1000
Tra n sf. R a t io
CL IP
1 mV / 1 mA
1m V / 10 mA
1m V / 100 mA
Maximum allowable operating voltage is equal to the current
1:1
1:10
1:100
1:1000
1 mV/1 A
transformer’s nominal voltage.
additional error resulting from the clip-on current sensor must also be
taken into consideration.
(default value: Clip = 0ff = voltage display)
50GMC-I Messtechnik GmbH
... SET rAtE ... CLIP
WZ12B,
Z201A/B
Z202A/B
METRAFLEX
Z202A/B, Z203A/B,
WZ12C
METRAFLEX
Measuring RangesClip Type
300 mV3 V30 V
300.0 mA3.000 A30.00 AWZ12C
3.000 A30.00 A300.0 A
30.00 A300.0 A3,000 kA
300.0 A3,000 kA30.00 kA
When reading the measured value,
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Tem p
MISO
~
R
x
Current
R
i
Ri 50
A~
Hz
A~
Hz
003.50
A
AC
TRMS
0050.1
Hz
AC
FUNC
ENTER
1:1000
1:1000
MEASURE
SETUP
... SET ... CLIP
FUNC
ENTER
FUNC
ENTER
0ff
Measuring Ranges:
Hz: 1 Hz . . . 31 kHz
mA / A: see table
3 ranges: 300 Hz, 3 kHz and 30 kHz
5.9.5 Alternating Current Measurement with A AC and Hz Clip-On
Current Transformer
Current/Current Transformer Output
When a clip-on current transformer is connected to the
multimeter (mA/A input), all current displays appear with the
correct value in accordance with the selected transformation
ratio. The only prerequisite is that the current transformer is
equipped with at least one of the below listed transformation
ratios, and that the ratio has been previously selected in the
following menu (Clip0FF).
Current Clip Setup Menu
... SET rAtE ... CLIP
1nFo
1 / 10/100/1000 / 0ff
Tra n sf. R a t ios
CL IP
1:1
1mA / 1 mA
1:10
1 mA / 10 mA
1:100
1 mA / 100 mA
1:1000
1 mA / 1 A
DMM Measuring RangesClip Types
30 mA300 mA3 A
30.00 mA300.0 mA3.000 A
300 mA3.000 A30.00 A
3.000 A30.00 A300.0 A
30.00 A300.0 A3000.0 A
WZ12A, WZ12D,
WZ11A, Z3511,
Z3512, Z3514
Measurements: V/Hz, , , Temp (°C/°F), , M
@UISO
, A/Hz
GMC-I Messtechnik GmbH51
Device and Measuring Parameters
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
6Device and Measuring Parameters
The instrument’s “SETUP ” mode (menu mode) makes it possible to
set operating and measuring parameters, query information and
activate the interface.
➭ The menu mode is accessed by pressing the
MEASURE | SETUP key, assuming that the instrument is switched
on and set to “Measure” (measuring mode operation).
“1nFo ” appears at the display.
➭ The main menus, i.e. the “SET ” and “tEMP ” menus, as well as
the “SEnd” and “StorE ” menus, are accessed, and the display is
returned to “InFo ”, by repeatedly activating the
keys
(in any direction).
➭ After selecting the desired main menu, sub-menus are
accessed by pressing the FUNC | ENTER key.
➭ The desired parameter is selected by repeatedly pressing the
keys.
➭ In order to check or change a parameter, acknowledge it with
the FUNC | ENTER key.
➭ The
keys can be used to position the cursor at the entry
position.
The desired value is selected with the help of the
➭ Changes can only be accepted with the FUNC | ENTER key.
➭ You can return to the sub-menu without making any changes
by pressing the ZERO | ESC key, and to the main menu by
pressing the ZERO | ESC key once again etc.
➭ You can switch to the measuring mode from any menu level by
pressing the FUNC | ENTER key.
keys.
After repeatedly pressing the MEASURE | SETUP key (without first
turning the multimeter off), you can return to the last selected
menu or parameter from the measuring mode.
Example: Setting Time
... setrAte ... t iME
1nFo
10:24 (hh:mm)
Setting hours and minutes:
Advance to desired entry position.
Change the setting, the entry position blinks.
Press and hold the key
The new time setting is activated after
acknowledgement.
to change the setting rapidly.
52GMC-I Messtechnik GmbH
6.1 Paths to the Various Parameters 6.2 List of All Parameters
bAtt
:
uers ion
:
1temp
dAtE
:
t iME
:
0CCvP
:
1nFo
tEMPSEnd SET StorE
StArt
StoP
000. 0
CLEAr
StArt
StoP
Sub-Menus / Parameters
rAtE
resol *
0
.d iSP
a
.d iSP
vpm *
CL iP
APoFF
v_1so
bEEP
irStb
Addr
dAtE
t iME
C
F
Main Menus
QuerySetSetSetSet
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
01nfo
MEASURE
SETUP
020.00
V
DC
AC
TRMS
Selection
FUNC
ENTER
Selection
FUNC
ENTER
Selection
FUNC
ENTER
Selection
FUNC
ENTER
* Function only available with customer-specific variant
GMC-I Messtechnik GmbH53
Device and Measuring Parameters
Parameter
0.d iSP 55: 0.diSP – show/hide leading zeros
a.disp
Addr
APoFF
bAtt
bEEP
CLEAr
CL IP
dAtE
EMpty
1nfo
irStb
itemp
0CCvP
rAtE
resol
SEnd
SET
StArt
StoP
store
tEMP
t iME
v1so
vpm
uErS ion
Page: Header
55: A.diSP – Select display mode for the analog display
59: Configuring Interface Parameters
56: APoFF – specified time for automatic shutdown and continuous ON
54: bAtt – query battery voltage
56: bEEP – set limit value for continuity testing
24: Measurement Data Recording
49: Direct and Pulsating Current Measurement with Clip-On Current Sensor
A DC and A (DC+AC)
50: Alternating Current Measurement with Clip-On Current Sensor, A AC and Hz
54: dAtE – query date, 57: dAtE – enter date
24: Measurement Data Recording
54: Querying Parameters – InFo Menu (as moving letters)
59: Configuring Interface Parameters
54: ItEMP – query reference temperature
24: Measurement Data Recording
55: rAtE – set the sampling rate
55: rESoL – high resolution for V DC and W (customer-specific function)
58: Activating the Interface
55: Entering Parameters – SETUP Menu
24: Measurement Data Recording
33: Temperature Measurement: Temp RTD and Temp TC
54: tiME – query time, 57: tiME – set time
56: U_ISo – select test voltage
56: RPM – revolutions per minute (customer-specific function)
54: vErSion – query firmware version
Device and Measuring Parameters
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
6.3 Querying Parameters – InFo Menu (as moving letters)
bAtt – query battery voltage
:
1nFobAtt
2.75 V.
vErSion – query firmware version
1nFobAtt:
uErSion
:
1.00
ItEMP – query reference temperature
The temperature of the internal reference junction is measured with a
temperature sensor in close proximity to the input sockets.
1nFo bAtt: ...
1tEMP: 24
°C
dAtE – query date
1nFobAtt:
D = day, M = month, Y = year
...
dAtE: 31.12.
05 (DD.MM.YY)
Date and time must be reentered after replacing the batteries.
tiME – query time
1nFo bAtt
t iME: 13:46:
56
:
...
(hh:mm:ss)
h = hours, m = minutes, s = seconds
Date and time must be reentered after replacing the batteries.
OCCUP – query memory occupancy
:
...
0CCvP:
000.0
1nFobAtt
%
54GMC-I Messtechnik GmbH
Device and Measuring Parameters
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
6.4 Entering Parameters – SETUP Menu
rAtE – set the sampling rate
The sampling rate specifies the time interval after which the
respective measured value is transmitted to the interface, or to
measured value memory.
Any one of the following sampling rates can be selected:
[mm:ss.t]: 00:00,1, 00:00,2, 00:00.5, 00:01,0, 00:02,0, 00:05.0
[h:mm:ss.t] (h=hours, m=minutes, s=seconds
rESoL – high resolution for V DC and (customer-specific function)
Switching back and forth between 3¾ and 4¾ places is possible
for direct voltage and resistance measurement.
... SETrAtE ... resol
1nFo
0.diSP – show/hide leading zeros
This parameter determines whether or not leading zeros will
appear in the measured value display.
1nFo
...
SET rAtE ... 0.diSP
0000.0 : with leading zeros (default value)
0.0 : leading zeros suppressed
A.diSP – Select display mode for the analog display
One of two different display modes can be selected for the
analog display:
• bArG: bar graph
• Point: pointer
... SETrAtE ... A.diSP
1nFo
bArG / Point
CLIP – set current clip factor
3000 / / 30000
(3000 = default setting)
GMC-I Messtechnik GmbH55
See chapter 5.9.3, chapter 5.9.4 and chapter 5.9.5.
Device and Measuring Parameters
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
APoFF – specified time for automatic shutdown and continuous ON
The instrument is switched off automatically if the measured value
remains unchanged for a long period of time and if none of the
keys or the rotary switch have been activated before the specified
time “APoFF” (entered in minutes) has elapsed.
If the on setting is selected, the multimeter is set for long-term
measurement and
ON appears in the display to the right of the
battery symbol. In this case, the multimeter can only be switched
off manually. The “on” setting can only be cancelled by changing
the respective parameter, and not by switching the instrument off.
... SETrAtE ... APoFF
1nFo
10 ... 59 min on
(10 minutes = default setting)
U_ISo – select test voltage
The desired test voltage for insulation resistance measurement
can be selected here:
... SETrAtE ... v1so
1nFo
... 500, 1000 V*
RPM – revolutions per minute (customer-specific function)
See section 5.1.2 regarding settings.
bEEP – set limit value for continuity testing
... SETrAtE ... bEEP
1nFo
1, 10, 20, 30, 40, 90
(10 = default setting)
irStb – status of the infrared receiver in the stand-by mode
See chapter 7.2 on page 59 regarding settings.
Addr – set device address
See chapter 7.2 on page 59.
* Selectable test voltages and the default setting depend upon the
respective customer-specific variant.
56GMC-I Messtechnik GmbH
Device and Measuring Parameters
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
ZERO
ESC
ON / OFF
LIGHT
dAtE – enter date
Entering the current date makes it possible to acquire measured
values in real-time.
1nFo
... SETrAtE ... dAtE
31.12 (DD: day . MM: month)
2005 (YYYY: year)
Date and time must be reentered after replacing the batteries.
tiME – set time
Entering the correct time makes it possible to acquire measured
values in real-time.
... SETrAtE ... t iME
1nFo
10:24 (hh:mm)
Date and time must be reentered after replacing the batteries.
6.5 Default Settings
Previously entered changes can be undone, and the default
settings can be reactivated. This may be advisable under the
following circumstances:
• after the occurrence of software or hardware errors,
• if you are under the impression that the multimeter does not
work correctly.
➭ Disconnect the device from the measuring circuit.
➭ Remove the batteries temporarily (see also chapter 9.2).
➭ Simultaneously press and hold the and
keys, and connect the battery at the same time.
A sequence of acoustic signals consisting of two immediately
consecutive tones acknowledges successful resetting.
GMC-I Messtechnik GmbH57
Interface Operation
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
7Interface Operation
The multimeter is equipped with an infrared interface for the transmission of measurement data to a PC. Measured data are optically transferred through the instrument housing by means of infrared light to an interface adapter (accessory), which is attached
to the multimeter. The adapter’s USB interface allows for the
establishment of a connection to the PC via an interface cable.
Beyond this, commands and parameters can be transmitted from
the PC to the multimeter as well. The following functions can be
executed:
• Configuration and read-out of measuring parameters,
• Measuring function and measuring range selection,
• Start measurement,
• Read out stored measured values.
7.1 Activating the Interface
The interface is automatically activated for receiving operation
(multimeter receives data from the PC) as soon as the interface is
addressed by the PC, assuming that the “irStb” parameter has
been set to “ir on” (see chapter 7.2), or the instrument is already
switched on (the first command wakes up the multimeter, but
does not yet execute any further commands).
The “continuous transmission” operating mode is selected
manually as described below. In this operating mode, the
instrument continuously transmits measurement data to the PC
via the interface adapter, which can then be displayed with the
help of a terminal program.
Starting Continuous Transmission Operation with Menu Functions
... SEndStArtsend
1nFo
The symbol blinks at the display in order to indicate interface
operation.
Stopping Continuous Transmission Operation with Menu Functions
1nFo
... SEndsend
stop send
The symbol is cleared from the display.
Automatic Activation and Deactivation of Transmission Mode Operation
If the sampling rate is 10 seconds or longer, the display is
switched off automatically between samples in order to prolong
battery service life. The only exception is when the multimeter is
set to continuous operation.
As soon as an event occurs, the display is automatically switched
back on.
58GMC-I Messtechnik GmbH
Interface Operation
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
7.2 Configuring Interface Parameters
irStb – status of the infrared receiver in the stand-by mode
There are two possible switching statuses for the infrared
interface when the multimeter is switched off:
iron: IR appears at the display and the infrared interface is
active, i.e. signals such as making commands can be
received, and power is consumed even though the
multimeter is switched off.
ir oFF:IR does not appear at the display and the infrared
interface is switched off, signals cannot be received.
... SETrAtE ... irStb
1nFo
ir on / ir oFF
(irstb = ir oFF = default setting)
Addr – Address
If several multimeters are connected to the PC via an interface
adapter, a separate address can be assigned to each instrument.
Address number 1 should be selected for the first instrument, 2
should be assigned to the second and so forth.
Key: R = meas. range, d = digit(s), rdg. = measured value (reading)
Intrinsic Uncertainty
under Reference Conditions
(... % rdg. + ... d)
4)
1 + 6
with ZERO function active
4)
4)
4)
4)
(... % rdg. + ... d)
8)
0.1 + 2
(... % R + ... d)
0.2% R + 8 d1000 V
0.2% R/kHz + 8 d
0.2% R + 8 d
0.2% R/kHz + 8 d
(... % rdg. + ... d)
0.5 %+ 15
0.5 %+ 15
1% + 5 K
9)
suring range; limitation: up to 30% of the range at up to 100 kHz in the mV measuring range, 30% of the range in the 3 A measuring range. The voltage
measuring ranges with max. 30 kHz apply in the A measuring range.
Plus sensor deviation
With ZERO function active
Residual value of 1 to 10 d with short circuited terminal tips,
exception: mV / A range of 1 to 35 d at zero point due to the TRMS converter
10 minute cool-down period
Required signal range 30% to 100% of the voltage measuring range
Overload
Capacity
ValueTime
1000 V
DC
AC
max. 10 s
RMS
Sine
Hz (V) 6):
6)
Hz(A )
:
max. 10 s
1000 V
7)
Hz (A):
DC
AC
RMS
Sine
6)
1000 V
DC/AC
max. 10 sPt 1000
RMS
Sine
2)
Cont.
Technical Data
Insulation Resistance Measurement
Measuring RangeResolution
0,3 V ... 1000 V
5 ... 310.0 k 0.1 k
0.280 ... 3.100 M1 k
02.80 ... 31.00 M10 k
028.0 ... 310.0 M100 k
0280 ... 3100 M1 M
1)
During insulation measurement (M
U
limits:
interference
Ri < 100 k @ Uiso 100 V, Ri < 250 k@ Uiso 250 V, Ri < 500 k@ Uiso 500 V,
Ri < 1000 k @ Uiso 1000 V
2)
Interference voltage measurement TRMS (V AC + DC) with 1 M input resistance,
Bandwidth 15 Hz ... 500 Hz, measuring error 3% + 30 Digit
and in the 3 V and 30 V range, all values < 0.29 V are suppressed
2)
> 10....20 V and U
1)
Test voltage
U
ISO
Ri=1M
500/1000 V
):
If
„Error“
U
ISO
50/100/250/500 V
50/100/250/500/1000 V
50/100/250/500/1000 V
50/100/250/500/1000 V
Interference voltage measurement TRMS (V AC + DC) with an input resistance of
1 M, frequency response width 15 Hz ... 500 Hz, accuracy of 3% + 30 digits;
in the 3 V and 30 V range, all values below 0,29 V are suppressed
Nominal Voltage
U
ISO
Ri=1M
1000 V
Intrinsic Uncertainty at
Reference Conditions
3 + 30 > 100 digits
10 + 5 digits
Short circuit measurement in the inductivity range:
10 H to 50 mH @ 100 Hz
Internal Clock
Time formatTT.MM.JJJJ hh:mm:ss
(% rdg. + d)
Resolution0.1 s
Measuring
Function
U
interference
M
@UISO
M
@UISO
/
Open-
Nom.
Circuit
Voltage
U
Voltage
N
U
— —— —
50, 100,
250,
500 V
@UISO
1000 Vmax.
M
62GMC-I Messtechnik GmbH
o
max.
1.1x U
1.1x U
Nom.
ShortCircuit
Current I
k
Acoustic
Signal
for
Current I
N
U>1000V 1000 V
1.0 mA < 1.2mAU>1000V 1000 V
Iso
0.5 mA < 1.2mAU>1000V 1000 V
Iso
Overload Capacity
ValueTime
dauernd
10 s
10 s
Accuracy1 minute per month
Temperature influence
50 ppm/K
Influencing Quantities and Influence Error
Influencing
Quantity
Temperature
1)
With zero balancing
Sphere of
Influence
0 C ... +21 C
and
+25 C... +40 C
Measured Quantity /
Measuring Range
V0.2 + 5
V0.4 + 5
300 ... 3 M0.5 + 5
30 M1 + 5
mA/A 0.5 + 5
mA/A 0.8 + 5
30 nF ... 300 F1 + 5
Hz0.2 + 5
C/F (Pt100/Pt1000)0.5 + 5
Influence Error
1)
(...% rdg. + ... d) / 10 K
Technical Data
Influ-
encing
Quan-
Frequency
2)
Power limiting: frequency x voltage, max. 3 x 106 V x Hz
3)
The accuracy specification is valid as of a display value of 10% and up
tity
Measured
Quantity /
Measuring
Sphere of Influence
Intrinsic Uncertainty
... % rdg. + ... d)
Range
> 15 Hz ... 45 Hz2 + 5 > 300 Digit
...
> 65 Hz ... 2 kHz2 + 5 > 300 Digit
300 V
1000 V> 65 Hz ... 5 kHz3 + 5 > 60 Digit
> 2 kHz ... 10 kHz3 + 5 > 300 Digit
> 15 Hz ... 45 Hz
10 A
> 65 Hz ... 10 kHz
> 15 Hz ... 45 Hz
10 A
> 65 Hz ... 10 kHz
> 65 Hz ... 10 kHz3 + 5 > 300 Digit
2)
> 65 Hz ... 10 kHz3 + 30 > 300 Digit
V
2)
A
A
+DC
A
A
AC
AC
AC
AC
AC
300 mV
300 A ...
300 A ...
300 mV /
3 V / 30 V
30 mA...3 A
3 + 10 > 300 Digit
3 + 30 > 300 Digit
to 100% of the measuring range for both measuring modes with the
TRMS converter in the A AC and A (AC+DC) ranges.
Influencing
Quantity
Crest Factor CF
5)
Except for sinusoidal waveshape
Influencing
Quantity
Relative
Humidity
Battery voltage1.8 to 3.6 Vditto
Sphere of
Influence
1 ... 3
> 3 ... 5 3% rdg.
Sphere of
Influence
75%,
3 days,
instrument off
Measured Quantity /
Measuring Range
V, A
Influence Error
1% rdg.
Measured QuantityInfluence Error
V A, , F, Hz, C1 x intrinsic uncertainty
Included in intrinsic uncer-
tainty
3)
5)
Influencing
Quantity
Common Mode
Interference
Voltage
Series Mode
Interference
Voltage
Sphere of Influence
Interference quantity max. 1000 VV> 120 dB
Interference quantity max. 1000 V
50 Hz ... 60 Hz, sine
Interference quantity: V ,
respective nominal value of the
measuring range,
max. 1000 V , 50 Hz ... 60 Hz sine
Interference quantity max. 1000 V
Measured Qty. /
Measuring Range
3 V , 30 V> 80 dB
300 V> 70 dB
1000 V> 60 dB
V> 50dB
V> 110 dB
Damping
GMC-I Messtechnik GmbH63
Technical Data
Response Time (after manual range selection)
Measured Quantity /
Measuring Range
V, V
A, A
300 ... 3 M2s
30 M, M
@UISO
Continuity< 50 ms
C (Pt 100)Max. 3 s
30 nF ... 300 FMax. 5s
>10 Hz1.5 s
Response Time
Digital Display
1.5 s
Max. 5 s
1.5 s
Measured Quantity
Jump Function
From 0 to 80%
of upper range limit value
From to 50%
of upper range limit value
From 0 to 50%
of upper range limit value
Reference Conditions
Ambient temperature
Relative humidity40% to 75%
Meas. Qty. Freq.45 Hz ...65 Hz
Meas. Qty. Waveshape
Battery voltage3 V 0.1 V
+23 C 2K
Sine
Ambient Conditions
Accuracy range0 C ... +40 C
Operating temp. range
Storage temp. range 25 C ... + 70 C (without batteries)
Relative humidity40 to 75%, no condensation allowed
Elevationto 2000 m
Deployment
10 C ... +50 C
Indoors, except within specified ambient
conditions
Display
LCD panel (65 mm x 36 mm) with analog and digital display including
unit of measure, type of current and various special functions
Background illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Analog
DisplayLCD scale with pointer
ScalingLinear
Polarity displayWith automatic switching
Overflow displayWith the symbol
Measuring rate40 measurements per second and display
Digital
Display/Char. Height 7-segment characters / 15 mm
Number of places3¾ places, 3100 steps, convertible to
Overflow display“OL” is displayed for 3100 digits
Polarity display“–” (minus sign) is displayed
Measuring rate10 and 40 measurements per second with
Refresh Rate2 times per sec., every 500 ms
:
5 ... 0 ... 30 with 35 scale divisions
for, 0 ... 30 with 30 scale divisions in all
other ranges
refresh
4¾ places* in measuring function V DC and
if plus pole is connected to “”
the Min/Max function except for the
capacitance, frequency and duty cycle
measuring functions
* Function only available with customer-specific variant
64GMC-I Messtechnik GmbH
Technical Data
Power Supply
Battery2 ea. 1.5 V mignon cell (2 ea. size AA),
Service lifeWith alkaline manganese batteries: approx.
Battery testBattery capacity display with battery
Power OFF function
Power pack socket
Measuring
Function
V200
V150
M
1)
Times 0.7 for interface operation
Nominal
Voltage U
1000 V2 M200
alkaline manganese per IEC LR6
200 hours (without M
@UISO
measurement)
symbol in 4 segments: .
Querying of momentary battery voltage via
menu function.
The multimeter is switched off automatically:
– If battery voltage drops to below approx.
1.8 V
– If none of the keys or the rotary switch are
activated for an adjustable duration of 10
to 59 minutes, and the multimeter is not in
the continuous operation mode
If the power pack has been plugged into
the instrument, the installed batteries are
disconnected automatically.
Rechargeable batteries can only be
recharged externally.
1)
1)
Number of Possible
Measurements with
Nominal Current per
VDE 0413
Resistance
of the
N
100 V1 M50
100 V100 k3000
500 V500 k600
DUT
Service life
in Hours
Electrical Safety
Safety classI I per DIN EN 61 010-1:2011/
Measuring category CAT IICAT II I
Nominal voltage1000 V 600 V
Fouling factor2
Test voltage5.2 kV~ per DIN EN 61010-1:2011/
VDE 0411-1:2011
VDE 0411-1:2011
Fuses
Fuse linkFF 10 A/1000 V AC/DC;
10 mm x 38 mm; switching capacity: 30 kA
at 1000 V AC/DC;
protects the current measurement input in
the 300 A through 10 A ranges
Electromagnetic Compatibility (EMC)
Interference emission
Interference immunity
EN 61326-1:2013, class B
EN 61326-1:2013
EN 61326-2-1:2013
GMC-I Messtechnik GmbH65
Technical Data
Data Interface
Typ e
Data transmissionSerial, bidirectional (not IrDa compatible)
ProtocolDevice specific
Baud Rate38,400 baud
Functions– Select/query measuring functions
The USB X-TRA plug-in interface adapter (see accessories) is
used for adaptation to the PC’s USB port.
Optical via infrared light through the housing
and parameters
Query momentary measurement data
–
Internal Measured Value Storage
Memory capacity4 MBit / 540 kB for approx. 15,000
measured values with indication of date
and time
Mechanical Design
HousingImpact resistant plastic (ABS)
Dimensions200 x 87 x 45 mm
(without protective rubber cover)
WeightApprox. 0.35 kg with batteries
ProtectionHousing: IP 54
(pressure equalization by means of the housing)
Table Excerpt Regarding Significance of the IP Code
IP XY
st
(1
Protection against
digit X)
foreign object entry
4 1.0 mm dia.4splashing water
5dust protected5water jets
IP XY
(2nd digit Y)
Protection against the
penetration of water
66GMC-I Messtechnik GmbH
Technical Data
GMC-I Messtechnik GmbH67
Maintenance – Recalibration
Note
Note
Note
MEASURE
SETUP
FUNC
ENTER
9Maintenance and Recalibration
Disconnect the instrument from the measuring circuit before
opening the battery compartment lid or fuse cover when
replacing batteries or fuses!
9.1 Displays – Error Messages
MessageFunctionMeaning
FUSE
0L
ur
error
Current measurement
In all operating modes
MeasurementIndicates overflow
M
measurement
ISO
M
measurement
ISO
Blown fuse
Battery voltage has fallen below 1.8 V
Measured value of less than 10% of the measuring range
Interference voltage detected
9.2 Batteries
Removing the Batteries During Periods of Non-Use
The integrated quartz movement draws power from the
batteries even when the instrument is switched off. It is
advisable to remove the batteries during long periods of
non-use for this reason (e.g. vacation). This prevents
excessive depletion of the battery, which may result in
damage under unfavorable conditions.
Battery Replacement
Stored measurement data are lost when the batteries are
replaced. In order to prevent data loss, it is advisable to
backup your data to a PC with the help of METRAwin 10 software before replacing the batteries.
The selected operating parameters remain in memory,
although date and time must be reentered.
Battery
The current battery charge level can be queried in the “1nfo”
menu:
:
1nFobAtt
2.75 V.
Make sure that no battery leakage has occurred before initial
start-up, as well as after long periods of storage. Continue to
inspect the batteries for leakage at short, regular intervals.
If battery leakage has occurred, carefully and completely clean the
electrolyte from the instrument with a damp cloth, and replace the
battery before using the instrument.
If the “” symbol appears at the display, the batteries should
be replaced as soon as possible. You can continue working with
the instrument, but reduced measuring accuracy may result.
The instrument requires two 1.5 V batteries in accordance with
IEC LR 6, or two equivalent rechargeable NiCd batteries.
68GMC-I Messtechnik GmbH
Maintenance – Recalibration
Attention!
!
Attention!
!
Blown Fuse
fuse
A
DC
Replacing the Batteries
Disconnect the instrument from the measuring circuit before
opening the battery compartment lid in order to replace the
batteries.
➭ Set the instrument face down onto the working surface.
➭ Turn the slotted screw on the lid with the battery symbols
counterclockwise.
➭ Lift off the lid and remove the batteries from the battery
compartment.
➭ Insert two new 1.5 V mignon batteries into the battery
compartment, making sure that the plus and minus poles
match up with the provided polarity symbols.
➭ When replacing the battery compartment lid, insert the side
with the guide hooks first.
Tighten the screw by turning it clockwise.
➭ Please dispose of depleted batteries in accordance with
environmental protection regulations!
9.3 Fuses
Testing the Fuse
The fuse is tested automatically:
• When the instrument is switched on with the rotary switch in
the A position
• When the instrument is already on and the rotary switch is turned to
the A position
• In the active current measuring range when voltage is applied
GMC-I Messtechnik GmbH69
If the fuse is blown or has not been inserted, “FuSE” appears at
the digital display. The fuse interrupts the current measuring
ranges. All other measuring ranges remain functional.
Replacing the Fuse
If a fuse should blow, eliminate the cause of overload before
placing the instrument back into service!
Disconnect the instrument from the measuring circuit before
opening the fuse cover in order to replace the fuse!
➭ Set the instrument face down onto the working surface.
➭ Turn the slotted screw on the cover with the fuse symbol
counterclockwise.
➭ Lift off the cover and pry the fuse out using the flat side of the
fuse cover.
➭ Insert a new fuse. Make sure that the fuse is centered, i.e.
between the tabs at the sides.
➭ When replacing the fuse cover, insert the side with the guide
hooks first. Tighten the screw by turning it clockwise.
➭ Dispose of the blown fuse with the trash.
Maintenance – Recalibration
Attention!
!
Note
Pb Cd Hg
Use specified fuses only!
If fuses with other blowing characteristics, other current
ratings or other breaking capacities are used, the operator
is placed in danger, and protective diodes, resistors and
other components may be damaged.
The use of repaired fuses or short-circuiting the fuse holder
is prohibited.
Testing the Fuse with the Instrument Switched On
After inserting the fuse with the instrument switched on, the
instrument must be switched off briefly and then switched
back on, or briefly switched to a non current measuring
range and then back to the A measuring range.
If contact is poor or the fuse is blown, FUSE appears at the
display.
9.4 Housing Maintenance
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth for cleaning. Avoid
the use of cleansers, abrasives or solvents.
9.5 Return and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German electrical and
electronic device law). This device is subject to the RoHS directive. Furthermore, we make reference to the fact that the current
status in this regard can be accessed on the Internet at
www.gossenmetrawatt.com by entering the search term WEEE.
We identify our electrical and electronic devices in accordance with WEEE 2012/19/EU and ElektroG with
the symbol shown at the right per DIN EN 50419.
These devices may not be disposed of with the trash.
Please contact our service department regarding the return of old
devices (see page 4).
If you use batteries or rechargeable batteries in your instrument or
accessories which no longer function properly, they must be duly
disposed of in compliance with the applicable national regulations.
Batteries or rechargeable batteries may contain harmful substances or heavy metal such as lead (PB), cadmium (CD) or mercury (Hg).
They symbol shown to the right indicates that batteries or
rechargeable batteries may not be disposed of with the
trash, but must be delivered to collection points specially
provided for this purpose.
70GMC-I Messtechnik GmbH
Maintenance – Recalibration
9.6 Recalibration
The respective measuring task and the stress to which your measuring instrument is subjected affect the ageing of the components and may result in deviations from the guaranteed accuracy.
If high measuring accuracy is required and the instrument is frequently used in field applications, combined with transport stress
and great temperature fluctuations, we recommend a relatively
short calibration interval of 1 year. If your measuring instrument is
mainly used in the laboratory and indoors without being exposed
to any major climatic or mechanical stress, a calibration interval of
2-3 years is usually sufficient.
During recalibration* in an accredited calibration laboratory
(DIN EN ISO/IEC 17025) the deviations of your instrument in rela-
tion to traceable standards are measured and documented. The
deviations determined in the process are used for correction of
the readings during subsequent application.
We are pleased to perform DAkkS or factory calibrations for you in
our calibration laboratory. Please visit our website at
www.gossenmetrawatt.com ( Company DAkkS Calibration
Center or FAQs Calibration questions and answers)).
By having your measuring instrument calibrated regularly, you fulfill the requirements of a quality management system per DIN EN
ISO 9001.
* Verification of specifications or adjustment services are not part of the
calibration. For products from our factory, however, any necessary adjustment is frequently performed and the observance of the relevant
specification is confirmed.
9.7 Manufacturer’s Guarantee
All METRAHIT digital multimeters and calibration instruments are
guaranteed for a period of 3 years after shipment. The manufacturer’s guarantee covers materials and workmanship. Damages
resulting from use for any other than the intended purpose or
operating errors, as well as any and all consequential damages,
are excluded.
The calibration certificate confirms that the product conformed to
the specified technical data at the time of calibration. We guarantee the observance of the specified technical data within the
admissible tolerance limits for a period of 12 months from delivery.
GMC-I Messtechnik GmbH71
Accessories
Attention!
!
10Accessories
10.1 General
The extensive accessories available for our measuring
instruments are checked for compliance with currently valid safety
regulations at regular intervals, and are expanded as required for
new applications. Currently up-to-date accessories which are
suitable for your measuring instrument are listed at the following
web address along with photo, order number, description and,
depending upon the scope of the respective accessory, data
sheet and operating instructions: www.gossenmetrawatt.de
(Products Measuring Technology – Portable Multimeters
METRAHIT ... Accessories).
10.2 Technical Data for Measurement Cables
(scope of delivery: KS17-2 safety cable set)
Electrical Safety
Maximum Rated Voltage
Measuring Category
600 V1000 V
CAT IVCAT I I I CAT I I
Maximum Rated Current 1 A1 A16 A
with safety cap applied
without safety cap applied
—
——
Please observe the maximum values of the electrical safety of the instrument.
Ambient Conditions (EN 61010-031)
Tem pe ra tu re–2 0 C ... + 50 C
Relative humidity50 to 80%
Contamination level
2
1000 V
Application KS17-2
In conformity with standard DIN EN 61010-031, measurements in an environment according to measuring
category III may only be performed with the safety cap
applied to the test probe of the measurement cable.
For establishing contact in 4 mm jacks you have to remove the
safety cap by levering out the snap lock of the safety cap with another sharp object (e.g. the second test probe).
10.3 Power Pack NA X-TRA (not included)
Use only power packs from GMC-I Messtechnik GmbH only in
combination with your instrument. This assures operator safety by
means of an extremely well insulated cable, and safe electrical
isolation (nominal secondary ratings: 5 V / 600 mA).
Installed batteries are disconnected electronically if the power
pack is used, and need not be removed from the instrument.
72GMC-I Messtechnik GmbH
10.4 Interface Accessories (not included)
USB X-TRA Bidirectional Interface Adapter
This adapter makes it possible to connect cable multimeters, as
well as STARLINE multimeters which are equipped with a serial IR
interface, to the USB port at a PC. The adapter allows for data
transmission between the multimeter and the PC.
METRAwin 10 PC Analysis Software
METRAwin 10 PC software is a multilingual, measurement data log-
ging program* for recording, visualizing, evaluating and
documenting measured values from METRAHIT multimeters.
Details on the system requirements are given in the installation
instructions of METRAwin 10/METRAwin 45.
Accessories
* runs under any IBM compatible Windows operating system