Gossen MetraWatt LINAX 4000L Operating Instructions Manual

14081B
2/8.18
Operating Instructions
LINAX 4000L
Continuous-Line Recorder
Contents
Applications and brief description
The LINAX 4000L is a microprocessor-controlled continuous-line recorder with 1 to 4 line channels, – 1 to 3 line channels and a printer channel. The measuring channels are electrically isolated from each other and are floating. The recorder is connected to transducers and is served to measure proces-related signals.
Reference symbols in the text
<Key> Designation of the keys in the display and control
panel
The information "right", "left" or "top", "bottom" – unless other­wise stated – is on the understanding that the viewer looks at the front.
Important information for your safety!
It must absolutely be read and followed!
A condition of correct and safe operation of the continuous­line recorder LINAX 4000L is that it is transported and stored in a suitable manner, competently installed and started as well as correctly operated and carefully serviced.
Only those persons must work on the recorder who are famil­iar with installation, startup, operation and servicing of com­parable equipment and who have the qualification required for their work.
The contents of these operating instructions and the safety notes affixed to the unit are to be observed.
The regulations, standards and directives mentioned in these operating instructions are for the Federal Republic of Ger­many. When using the recorder in other countries, relevant national rules must be followed.
The recorder is constructed and tested according to DIN EN 61010-1 "Safety requirements for electronic measur­ing instruments", it left the factory in safe and proper condi­tion. To maintain this condition and to ensure safe operation, the safety notes in these operating instructions with the head­ing "Caution" must be followed. Otherwise, persons could be endangered and the unit itself as well as other equipment and facilities could be damaged.
If the information contained in these operating instructions should not be sufficient in certain cases, our product support (see page 12) will be glad to provide further information.
2 GMC-I Messtechnik GmbH
1 Installation and startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Selecting the mounting site . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4.1 Connecting the input signals . . . . . . . . . . . . . . . . . 4
1.4.2 Connecting the power supply . . . . . . . . . . . . . . . . 4
1.5 Chart loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5.1 Recording table for roll chart . . . . . . . . . . . . . . . . . 5
1.5.2 Recording table for fanfold chart . . . . . . . . . . . . . . 5
1.6 Installing the fiber pen insert. . . . . . . . . . . . . . . . . . . . . . 6
1.7 Switching the recorder on . . . . . . . . . . . . . . . . . . . . . . . 6
1.8 Positioning the recording chart . . . . . . . . . . . . . . . . . . . 6
2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1 Removing the recording chart . . . . . . . . . . . . . . . . . . . . 6
2.2 Removing the rec. chart from the take-up roll . . . . . . . . 6
2.3 Changing the chart speed . . . . . . . . . . . . . . . . . . . . . . . 7
2.4 Select response time . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Reconfiguration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1 Changing measuring ranges . . . . . . . . . . . . . . . . . . . . . 8
3.1.1 Hardware matching by means of jumpers . . . . . . . 8
3.2 Set zero and calibrate the upper range limit . . . . . . . . . . 8
3.3 Replacing scales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 Replacing the label for the measuring points . . . . . . . . . 9
4 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1 Fuse replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
6 Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
7 Returns and Environmentally Sound Disposal . . . . . . . . . . . 12
8 Repair and Replacement Parts Service
Calibration Center* and
Rental Instrument Service. . . . . . . . . . . . . . . . . . . . . . . . . . 12
9 Product Support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
GMC-I Messtechnik GmbH 3
1 Installation and startup
1.1 Scope of delivery
(see Figure 1)
The continuous-line recorder LINAX 4000L comes with: – 1 operating instructions – 2 fasteners Be – 1 fiber pen insert Fe per measuring channel – 1 pack of fanfold chart Fp or 1 roll chart Sr – Depending upon the order, the respective number of
screw-plug terminals Sk and reading ruler(s).
Figure 1 Scope of delivery of the LINAX 4000L
1.2 Selecting the mounting site
Figure 2 Dimensional drawing LINAX 4000L (dimensions in mm)
Position of use Inclination to the side –30°... 0 ... + 30°
Inclination to the rear 20°
Inclination to the front 20° Ambient temp. 0 ... 50 °C Relative humidity 75 % annual average,
max. 85 %.
Prevent dewing!
1.3 Installation
(see Figure 2 and Figure 3)
Installation in switchboards
1. Insert the recorder into the switchboard from the front.
2. At the sides of the case, slide the fasteners Be into the guide grooves (see Figure 3). Note The fasteners Be are suited for side-by-side mounting in hori­zontal or vertical direction.
3. After aligning, equally tighten the fasteners Be.
Figure 3 Mounting the fasteners
Installation in grid frames
1. Fasten 4 each centering angle bracket (Ordering number A416A) on the grid frame.
2. Slide the fasteners Be into the guide grooves at the sides of the case (see Figure 3).
3. After alignment, equally tighten the fasteners Be.
Be Fastener Fe Fiber pen inserts Fp Fanfold chart Sk Screw-plug terminals Sr Chart roll
Fe
Sk
Be
Sr
Fp
De
Panel cutout
Be Fastener
4 GMC-I Messtechnik GmbH
1.4 Connection
(see Figure 4)
Caution
The connection between the protective conductor connection and a pro­tective conductor must be made prior to all other connections.
The unit can become dangerous when the protective conductor inside or outside the unit is interrupted or when the protective conductor connec­tion is undone.
Figure 4 Rear panel and wiring diagrams
1.4.1 Connecting the input signals
• Fasten the signal leads in the screw-plug terminals, maximum cross section 2 × 1mm2.
The recorder must only be operated in installed condition.
A power line connection switch of sufficient switching capacity, which permits all-pole disconnection of the recorder from the power line, must be provided within reach of the mounting site. It must not annul the pro­tective effect of the protective conductor.
1.4.2 Connecting the power supply
• Fasten the power supply lines, max. cross section 1 × 4mm2 or 2 × 1.5 mm2, in the screw terminals. The cross section of the protective conductor must at least correspond to the cross section of the line power cable.
Stromsignale
Spannungssignale
Signal inputs
Voltage signals
Current signals