Gossen MetraWatt DIN VDE 0701, DIN VDE 0702 Operating Instructions Manual

Page 1
Operating Instructions
MINITESTMASTERPROBASE
Testers per DIN VDE 0701 and 0702
3-349-358-15
4/7.07
Page 2
2 GMC-I Gossen-Metrawatt GmbH
6
4
5
1
9
8
2
3
7
1
0
23
18171615
22212019
9
8
1
2
3
4
5236
11
Page 3
GMC-I Gossen-Metrawatt GmbH 3
1 Mains power cable 2 Fuse (for protecting the probe (3)) 3 Probe connector socket for measuring
protective conductor current (PE) or contact current (I
C
)
4 Rotary switch for selecting the desired
measuring function 5USB port * 6 Send key * for data transmission or
storage 7 LEDs for indicating adherence to, or
violation of limit values 8 Earthing contact mains outlet 9 Earthing contact test socket
10 Probe connector socket for insulation
resistance measurement
(only necessary for Safety Class II)
11 LCD * 12 POWER ON key * 13 RCCB test/POWER OFF key * 14 Fuse for 16 A mains outlets
and mains switch
15 CEE test socket 1P+N+PE 16 A-6h 16 CEE test socket 3P+N+PE 16 A-6h 17 CEE test socket 1P+N+PE 32 A-6h 18 CEE mains socket 3P+N+PE 32 A-6h 19 CEE mains socket 1P+N+PE 16 A-6h 20 CEE mains socket 3P+N+PE 16 A-6h 21 CEE mains socket 1P+N+PE 32 A-6h 22 CEE mains socket 3P+N+PE 32 A-6h 23 Barcode scanner connection
(9-pin, subminiature plug) **
24 Probe cable
*not for MINITESTBASE ** MINITESTMASTER and MINITEST 3PMASTER only
* as from series launched in March 2007
Features of MINITEST... series
MINITEST 3PMASTER
MINITESTMASTER
MINITESTPRO
MINITESTBASE
Connection types
Tests on monophase DUTs
✘✘✘✘
Tests on 3-phase DUTs via additional test sockets CEE 16A / CEE 32A
–––
Fusing devices
Fuse for probe connection
✘✘✘✘
RCCB in mains plug
✘✘
Miniature circuit breaker
–––
Report functions
Illuminated two-line LCD *
✘✘✘
Memory for 2,000 tests (10 measured values per test)
✘✘––
Key for transmission of measured values
✘✘✘
Key for storing measured values
✘✘––
Data interface (USB port)
✘✘✘
Barcode scanner connection (9 pin, subminiature plug)
✘✘––
Standard equipment accessories of MINITEST... series
MINITEST 3PMASTER
MINITESTMASTER
MINITESTPRO
MINITESTBASE
Probe cable with test probe
✘✘✘✘
Adapter for earthing contact plug to CEE coupling 3P+N+PE 32 A-6h
–––
Adapter for plug 1P+N+PE 16 A to CEE coupling 3P+N+PE 32 A-6h
–––
Adapter for plug 3P+N+PE 16 A
to
CEE coupling 3P+N+PE 32 A-6h
–––
Adapter for plug 1P+N+PE 32 A
to
CEE coupling 3P+N+PE 32 A-6h
–––
USB connector cable
✘✘✘
PC analysis software programs on CD-ROM
✘✘✘
Operating instructions
✘✘✘✘
Page 4
4 GMC-I Gossen-Metrawatt GmbH
Table of Contents
1 Applications ...............................................4
2
Safety Features and Precautions .......................4
3 Initial Start-Up ............................................6
4 Test Sequence ...........................................6
5
Descriptions of Individual Measurements .........8
5.1 Measurements at the Test Outlet(s) ........................8
5.2 Measurements at the Mains Outlet(s) .....................9
6 Report Functions ......................................10
6.1 Transmission Mode (Transmission only while
Program MINITEST is Active) ...............................10
6.2
Permanent Transmission (Transmission only while
Program MINITEST is Active) ................................... 10
6.3
Memory Mode (Read-Out only with Program
MINITESTMASTER
) ................................ 10
6.3.1 Activate Memory Menu .......................................10
6.3.2 Selecting Memory Function .................................10
6.3.3 Exiting the Memory Menu ....................................11
6.3.4 Reading out Stored Measured Values at the PC ....11
6.3.5 Barcode Scanner Function ...................................11
6.3.6 Switching Off Safely ............................................11
7 PC Analysis Software Programs ..............11
7.1 Program „MINITEST“ ........................................11
7.2 Program „MiniTEST Master“ ............................11
8 Technical Data .........................................12
9 Maintenance ............................................14
9.1 Housing .............................................................14
9.2 Replacing the Fuse .............................................14
9.3 Calibration Intervals ............................................14
9.4
Return and Environmentally Sound Disposal .................. 14
10 Accessories ..............................................15
10.1 Standard Equipment ...........................................15
10.2 Order Information for Available
Accessories ........................................................15
11 Repair and Replacement Parts Service
DKD Calibration Laboratory and Rental
Instrument Service ..........................................16
12 Product Support .......................................16
1 Applications
The test instrument is used to test the elec­trical safety of monophase and 3-phase electrical devices. The test instrument allows for the measurement of protective conductor resistance, insulation resistance and differ­ential current and/or contact current in accordance with DIN VDE 0404.
2 Safety Features and Precautions
The test instrument has been manufactured and tested in accordance with the following safety regulations:
IEC 61010-1 / DIN EN 61010-1 / VDE 0411-1, DIN VDE 0413 parts 2 and 4.
When used for its intended purpose, the safety of the user, the test instrument and the device under test (electrical equipment or electrical medical device) is assured.
Read the operating instructions carefully and completely before placing your test instrument into service. Follow all instructions contained therein. Make sure that the operating instructions are available to all users of the instrument.
Tests may only be performed by a qualified electrician, or under the supervision and direction of a qualified electrician. The user must be instructed by a qualified electrician in the execution and evaluation of tests.
Safety Precautions for MINITEST 3P⏐M
ASTER
The instrument is intended for connection with a 3-phase system up to 32 A.
MINITEST 3PMASTER can also be connected to all customary CEE and earthing contact outlets via the adapters included in the stan­dard equipment.
Attention!
!
The instrument may only be con­nected to the mains via the adapters included in the standard equipment.
Attention!
!
All mains sockets and/or test sock­ets are connected in parallel: Only one DUT at a time may therefore be connected to a socket.
Page 5
GMC-I Gossen-Metrawatt GmbH 5
Observe the following safety precautions:
MINITEST⏐M
ASTER⏐PRO⏐BASE:
The instrument may only be connected to electrical systems with a maximum of 230 V which are protected with a fuse or circuit breaker with a maximum rating of 16 A.
Measurements within electrical systems
are prohibited.
Be prepared for the occurrence of
unexpected voltages at devices under test (for example, capacitors may be dangerously charged).
Make certain that the measurement
cables are in flawless condition, e.g. no damage to insulation, no cracks in cables or plugs etc.
Insulation Resistance Measurement R
ISO
Testing is conducted with up to 500 V. Current limiting is utilized (I < 3.5 mA), but if the terminals (L or N) are touched, electrical shock may occur which could result in consequential accidents.
Differential Current Measurement I
D
/
Contact Current Measurement I
C
It is absolutely essential to assure that the device under test is operated with line voltage during performance of the differential current and/or contact current measurement. Exposed conductive parts may conduct dangerous contact voltage during testing, and may not under any circumstances be touched (MINITESTM
ASTER⏐PRO:
mains power is disconnected if leakage current exceeds approx. 15 mA).
The function test may only be performed
after the device under test has successfully passed the safety test while connected to the test socket!
The measuring and test instrument may not be used:
if it demonstrates visible damage
with damaged connector cables and measuring cables
if it no longer functions properly
after extraordinary stresses due to transport
MINITEST 3PMASTER: with other adapters than those included in the standard equipment.
In such cases, the instrument must be removed from operation and secured against unintentional use.
Meanings of Symbols on the Instrument
The symbols on the instrument have the following meanings:
Warning concerning a point of danger (attention: observe documentation!)
Indicates EC conformity
The device may not be disposed of with the trash. Further informa­tion regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term WEEE.
!
Page 6
6 GMC-I Gossen-Metrawatt GmbH
3 Initial Start-Up
Connection
Connect the test instrument’s mains plug to the mains power outlet.
MINITEST 3P⏐M
ASTER:
The instrument can be connected to all cus­tomary CEE sockets by means of the mains connection adapter (standard equipment).
Switching the MINITEST 3P⏐M
ASTER On:
The instrument is switched on via the 3-pole miniature circuit breaker (m.c.b.).
Switching the MINITESTMASTERPROBASE On
The test instrument can be switched on with the green power-on switch located on the integrated residual current protective device. It can be switched back off again with the blue test key.
Switching the MINITEST⏐B
ASE On
The test instrument is switched on and off with the help of the mains plug.
4 Test Sequence
Testing for electrical safety always begins with a visual inspection. The measurements are then conducted in the order in which they appear next to the selector switch on the test instrument, from top to bottom.
For each type of measurement, the test instrument evaluates the measured values and indicates whether or not limit values in accordance with DIN VDE 0701 or 0702 have been adhered to by means of an LED array.
The number and type of required measure­ments are determined according to the clas­sification of the device under test:
Indication Meaning
Green LED
The measured value is better than the strictest limit value.
Red LED
The measured value is worse than the most lenient limit value.
Yellow LED
The limit value is in-between the strictest and the most lenient limit values. Whether or not the DUT passes testing depends upon the device’s classification. Evaluation should be conducted by a qualified electrician.
Page 7
GMC-I Gossen-Metrawatt GmbH 7
* MINITEST 3PMASTER only
Device Under Test
Connect DUT to the test socket. Connect DUT to mains outlet.
– Switch the DUT on.
Protective conductor resistance
Probe socket PE/ I
C
Protective conductor resistance
Permanent, probe socket PE/ I
C
Insulation resistance
No probe required
Insulation resistance
Probe socket R
ISO
Differential current
No probe required
Contact current
Probe socket PE/ I
C
Safety class I
Earth contact plug
Without exposed parts
✘✘
Test with the mains plug poled in both directions.
With exposed parts, not connected with PE
Test with the mains plug poled in both directions.
Test with the mains plug poled in both directions.
CEE plug *
Without ex­posed parts
✘✘✘
With exposed parts, not connected with PE
✘✘
Permanently connected DUT
Safety class II
With European earth contact plug
Test with the mains plug poled in both directions.
Use of mains outlet number depending on the mains terminal (all test sockets as well as mains outlets No. 8 and 19 can always be used regardless of which mains adapters are used)
Mains outlet number
20 21 22
Mains connection with mains plug 3P+N+PE 32 A
✘✘✘
Connection with earthing contact plug adapter to CEE coupling 3P+N+PE 32 A-6h
max. 16 A
Connection with adapter for plug 1P+N+PE 16 A to CEE coupling 3P+N+PE 32 A-6h
max. 16 A
Connection with adapter for plug 3P+N+PE 16 A to CEE coupling 3P+N+PE 32 A-6h
max. 16 A✘max. 16 A
Connection with adapter for plug 1P+N+PE 32 A to CEE coupling 3P+N+PE 32 A-6h
Page 8
8 GMC-I Gossen-Metrawatt GmbH
5 Descriptions of Individual Measurements
5.1 Measurements at the Test Outlet(s)
Protective Conductor Resistance, R
PE
SOCKET
The probe must be plugged into the
PROBE PE/I
C
socket for the measurement of protective conductor resistance. Measure­ment is performed between the protective conductor terminal at the test outlet and the
PROBE PE/I
C
socket.
Protective Conductor Resistance, R
PE
FIX
This measurement is used for testing the protective conductor at permanently con­nected devices. The test instrument must be connected to the same supply circuit as the device under test during measurement.
Attention!
!
Good test results may be obtained with parallel ground connections, although the protective conductor is interrupted.
The probe must be plugged into the
PROBE PE/IC socket for the measurement of
protective conductor resistance. Measure­ment is performed between the protective conductor terminal at the mains connection and the PROBE PE/I
C
socket.
Testing is performed up through a resistance value of approximately 1.3 Ω with automatic polarity reversal.
The earthing contact resistance and the test current are indicated on the LCD (not for
MINITESTBASE).
Checking the Zero Point and the Fuse
The zero point and the internal fuse can be checked by inserting the plug attached to the test probe cable into the PROBE R
ISO
socket and the tip of the test probe into the
PROBE PE/IC socket with the selector switch in the R
PE
SOCKET or RPE FIX setting. The dis-
played value should not exceed 100 mΩ. Note: If a current of 0 mA is indicated, the
fuse is usually defective. Test and replace it, if required.
Attention!
!
It is absolutely essential to insert the tip of the test probe into the
PROBE PE/IC socket for this test!
Attention!
!
Use only the probe cable which is included with the test instrument. If a different probe cable is used: – The measured value may be
distorted.
– The test instrument may be
damaged.
– Malfunctioning may occur.
Insulation Resistance, R
ISO
Various test types are provided for the mea­surement of insulation resistance.
The test type is selected automatically when the test probe is plugged into the probe connector socket (10).
The original test probe cable must be used in order to take advantage of this function. Test type switching cannot be assured if other test probe cables are used.
For safety class I DUTs without exposed conductive parts: Test insulation between L/N and PE at the test outlet.
The test probe cable may not be plugged into the PROBE R
ISO
socket for this test!
For safety class II DUTs, or for safety class I DUTs with exposed conductive parts which are not connected to the protective conductor: Test between L/N at the test outlet and the PROBE R
ISO
socket.
The test probe must be plugged into the PROBE R
ISO
socket for this test!
The insulation resistance and the test volt­age are indicated on the LCD (not for MINIT- ESTBASE).
Page 9
GMC-I Gossen-Metrawatt GmbH 9
Discharging the Device Under Test
The device under test is automatically discharged when the insulation test is switched off (turn switch to R
PE
or ID).
The LEDs light up rapidly, one after the other from left to right, during discharging.
Furthermore, residual voltage appears at the display (not for MINITESTB
ASE).
Leave the DUT connected to the test instrument during the discharging sequence.
5.2 Measurements at the Mains Outlet(s)
Differential current and/or contact current measurements are performed at the mains outlet(s). The DUT must be unplugged, and plugged back in again. In the case of mains
plugs which are not polarity protected, testing must be conducted with the mains plug poled in both directions.
Attention!
!
The mains outlet(s) always con­duct(s) voltage as soon as the test in­strument has been switched on.
Note!
Make sure that the device under test is switched off before plugging it into the mains outlet.
The device under test must be switched on during testing.
The device under test must be switched back off before it is unplugged from the mains outlet.
Differential Current Measurement, I
D
Total current from all phase conductors is measured during differential current mea­surement. For safety class I devices, this corresponds to the amount of current which is conducted by the protective conductor.
Differential current is displayed in mA at the LCD (not for MINITESTBASE).
Evaluating the Measured Values
Measured values of less than 0.5 mA are indicated by means of a green LED.
Measured values of greater than 3.5 mA are indicated by means of a red LED. These devices may be dangerous, however, for
example for 3-phase current devices, there may be higher admissible limit values of e.g. 1 mA per kW up to 10 mA, depending on the manufac­turer’s specifications. These cases are all indi­cated with the red LED. Evaluation must be
conducted by a qualified electrician in this case.
Measured values of between 0.5 and 3.5 mA are indicated by means of a yellow LED. An evaluation of electrical safety should be con­ducted by a qualified electrician in this case.
Contact Current Measurement, I
C
When measuring contact current, current is measured which is conducted via the probe at the PROBE PE/I
C
socket to the protective
conductor terminal at the device under test.
Evaluating the Measured Values
Measured values of less than 0.25 mA are indicated by means of a green LED. There is no danger in touching these parts.
Measured values of greater than 0.5 mA are indicated by means of a red LED. Devices which demonstrate these values are dangerous because persons who touch them are startled, which may result in consequential accidents.
These devices may not be placed back into service.
Measured values of between 0.25 and
0.5 mA are indicated by means of a yellow LED. Current within this range is not dangerous, but the threshold of perception is approximately 0.1 mA depending upon sensitivity, which causes mild tingling. This is unpleasant for some people, for which reason devices of this sort should be examined.
Self-Test
A device self-test is conducted when the selector switch is turned to the LED Test position.
Testing is indicated by the LEDs which light up one after the other from left to right.
Page 10
10 GMC-I Gossen-Metrawatt GmbH
6 Report Functions
Depending on the selected operating mode, the measured values and the result of each test can be transmitted to a PC via USB port for onward processing or stored to the inter­nal device memory. The transmission or storage process of measured values is indi­cated by all LEDs lighting briefly up from right to left.
Selection of Operating Mode
The corresponding PC analysis software programs have to be installed and started for the three different operating modes in order to receive data, see table above.
Press key „S“ (6) when switching on the instrument to set the requested operating mode. The following operating modes are shown in subsequent order:
„Transmission mode“, siehe chapter 6.1
„Permanent transmission“, see chapter 6.2
„Memory mode“, see chapter 6.3 The operating mode which is indicated when
releasing key „S“ (6) is selected. This setting remains active even after switching off the instrument.
6.1 Transmission Mode (Transmission only while Program MINITEST is Active)
To transmit the current measured value, key „S“ must be pressed in operating mode „Transmission mode“ .
6.2 Permanent Transmission (Transmission only while Program MINITEST is Active)
If operating mode „Permanent transmission“ is activated, each measured value is trans­mitted via the USB port.
6.3
Memory Mode (Read-Out only with Program MINI
TEST⏐MASTER
)
In the test instruments MINITESTMASTER or MINITEST 3PMASTER a maximum of 2,047
DUTs with 10 measured values per DUT can be stored. If an attempt is made to store more than 10 measured values for one DUT, the additional values are ignored and the red LED lights up briefly.
6.3.1 Activate Memory Menu
After selecting memory mode (see above) you have to set the rotary switch to position
Gerä-
teselbsttest/Speichermenü (
Self-test/Memory
Menu)
. Activate the memory menu by pressing key „S“ (6). The following is shown on the LCD: ** MEMORY MENU ** SELECTION -> SWITCH. You may now select the requested memory function via the rotary switch.
6.3.2 Selecting Memory Function
Key NUM – > Menu DUT NUM –
The number of the current memory location is reduced by one each time key „S“ (6) is pressed. The lowest possible value is „1“. The letter „D“ in front of the number signifies that data are already available for this mem­ory location.
Key NUM + > Menu DUT NUM+
The number of the current memory location is increased by one each time key „S“ (6) is pressed. The highest possible value is „2,047“, i.e. a maximum of 2,047 DUTs can be tested. The letter „D“ in front of the num­ber signifies that data are already available for this memory location.
Key DATA > Menu MEASURED VALUE NO.
In this switch position all measured values of a DUT are shown in a consecutive, numeri­cal sequence in the order of their recording.
Operating mode selected at the test instrument
MINITEST 3PMASTER
MINITESTMASTER
MINITESTPRO
PC analysis software depend­ing on the operating mode
Transmission mode
(Transmission of individual measured values)
✘✘✘
MINITEST
Permanent transmission
(Continuous transmission of measured values)
✘✘✘
MINITEST
Memory mode
(Internal measured value memory)
✘✘
—MiniTEST
Master
Page 11
GMC-I Gossen-Metrawatt GmbH 11
Key CLR > Menu DUT DEL ?
Pressing key „S“ (6) deletes the current memory location, i.e. all data for the selected DUT. As a preventive measure against acci­dental deletion you have to keep key „S“ (6) pressed for about half a second before the deletion process is started. The red LED lights up during the deletion.
Key CLR ALL > Menu DELETE MEMORY ?
Pressing key „S“ (6) deletes the entire device memory. As a preventive measure against accidental deletion you have to keep key „S“ (6) pressed for about 1 to 2 seconds before the deletion process is started. The red LED lights up during the deletion.
6.3.3 Exiting the Memory Menu
Set the rotary switch to position Geräteselb­sttest/Speichermenü (Self-test/Memory Menu) and press key „S“ (6) to exit the mem­ory menu.
In operating mode „Memory mode“ the last measured value indicated after a measure­ment is shown when pressing key „S“ (6). The storage operation - as well as transmi­sion of the values via USB port - is indicated by a flickering of the LEDs. If more than 10 values are stored for one DUT, the red LED lights up additionally for approx. half a sec­ond, thus signalling that it was not possible to store the measured values.
6.3.4 Reading out Stored Measured Values at the PC
Prerequisite: The test instrument may not be set to Geräteselbsttest/Speichermenü (Memory Menu or Self-test).
The stored measured values can be trans­mitted from the test instrument to a PC via the analysis software and the USB port where they can be evaluated and docu­mented.
6.3.5 Barcode Scanner Function
Connect the barcode scanner to jack 23. Only barcode scanners by GMC-I Gossen-
Metrawatt GmbH may be used, e. g. type B3261.
Barcode scanners with USB port are not suitable.
In function „NUM+“ and „NUM-“ a text com­prising a maximum of 24 characters which has been scanned with the barcode scanner is recorded as a description of the DUT. The text is shown in the second line of the LCD and stored.
6.3.6 Switching Off Safely
You have to exit the memory menu before switching off the test instrument or discon­necting it from the mains. Set the rotary switch (4) to position Geräteselbsttest/Spei­chermenü (Self-test/Memory Menu) and press key „S“ (6) for approx. half a second.
7 PC Analysis Software Programs
Two PC software programs for the adoption and evaluation of data with the following fea­tures are available on the attached CD ROM:
7.1 Program „MINITEST“
Automatic adoption of measured values: Start test, display measurement results, save results
Customer data management
Other functions: – Selection between
VDE 0701 and 0702 – Establishing the test cycle – Automatic adoption of the contractor’s
data in the test report
Transmit measurement results to a word processing program
7.2 Program „MiniTEST Master“
Read-out of device memory
Presentation of measurement data in tabular form
Exporting the table to Excel
Preparation of test reports
Page 12
12 GMC-I Gossen-Metrawatt GmbH
8 Technical Data
*With automatic polarity reversal
Intrinsic Error and Measuring Error
Influencing Quantities and Influence Error
Measured Quantity Measuring Range Reso-
lution
U
no-load
R
i
I
K
I
N
Protective conductor resistance 0 1.30 Ω
1.0 99.9 Ω
10 mΩ 100 m
Ω
< 5 V < 5 V
> 200 mA *
Insulation resistance 0 9.99 MΩ 10 kΩ 520 V approx.
50 kΩ
< 3.5 mA > 1 mA
Contact current measurement (verification of absence of voltage)
0 9.99 mA 10 μA1 kΩ
Differential current
MINITEST⏐M
ASTER⏐PRO⏐BASE
0.1 . . . 9.99 mA~
10 μA
Differential current
MINITEST 3PMASTER
0.2 . . . 9.99 mA~ 10 μA
Measured Quantity Intrinsic Error Measuring Error
Protective conductor resistance ± (5% rdg. + 4 d) ± (10% rdg. + 6 d) Insulation resistance ± (7% rdg. + 2 d) ± (10% rdg. + 5 d)
Contact current measurement (verification of absence of voltage)
± (5% rdg. + 4 d) ± (10% rdg. + 5 d)
Differential current
MINITEST⏐M
ASTER⏐PRO⏐BASE
± (5% rdg. + 6 d) ± (10% rdg. + 6 d)
Differential current
MINITEST 3P⏐M
ASTER
± (5% rdg. + 10 d) ± (10 % rdg. + 10 d)
Influencing Quantity / Sphere of Influence
Designation per DIN VDE 0404
Influence Error ± % of Measured Value
Change of position E1
Change to test equipment supply voltage
E2 2.5
Temperature fluctuation
E3
Specified influence error valid starting with temperature changes as of 10 K:
0 21° C and 25 40° C 1 for protective conductor resistance
0.5 for all other measuring ranges
Amount of current at DUT E4 2.5
Low frequency magnetic fields E5 2.5
DUT impedance E6 2.5
Capacitance during insulation measurement
E7
2.5
Waveshape of measured current
E8
49 51 Hz 2 with capacitive load
(for equivalent leakage current)
45 60 Hz 1 (for contact current)
2.5 for all other measuring ranges
Page 13
GMC-I Gossen-Metrawatt GmbH 13
Reference Conditions
Ambient temperature
+23° C ±2K
Relative humidity 40 ... 60% Line voltage
MINITEST⏐M
ASTER⏐PRO⏐BASE: 230 V ±1%
MINITEST 3P⏐M
ASTER: 230 V/400 V ±1%
Measured quantity frequency 50 Hz ±0.2%
Measured quantity waveshape Sine (deviation
between RMS and rectified value: ±0.5%)
Ambient Conditions
Operating temp. 0 to + 40° C Storage temp. range –20 to +70° C Humidity Max.75%, no con-
densation allowed
Elevation to 2000 m
Power Supply
Line voltage
MINITEST
MASTERPROBASE
:
230 V 50 Hz
MINITEST 3P
MASTER: 230V/400V50Hz
Throughput rating
MINITEST
MASTERPROBASE
:
max. 3700 VA
MINITEST 3PMASTER: max. 38.4 kVA depending upon load at the mains outlet
Electrical Safety
Safety class I Nominal line voltage 230 V Tes t v ol ta ge
Mains + PE (mains) to test outlet, probe socket PE/IC or R
ISO
: 1.5 kV∼ Mains to PE (mains): 3 kV
Measuring category 300 V CAT II Contamination degree 2 Fuse FF0,315H1000V
or FF0,315H500V or FF0,315H250V
MINITEST 3P⏐M
ASTER
only:
3 x C16A
Residual current protective device (
MINI
TEST
MASTERPRO
)
30 mA with undervoltage trigger and inhibiting of automatic restart
Display and Indicating Devices
LCD (not for MINITESTB
ASE)
Dot matrix display, two lines of 20 characters each
LEDs
9 LEDs for indicating compliance with or violation of limit values: 1 red, 7 yellow and 1 green
Mechanical Design
Dimensions/weight
MINITEST
MASTERPROBASE:
W x H x D: 200 mm x 1mm x 77 mm (without integrated outlets, grommets and rotary switch)
Weight approx. 1.5 kg
MINITEST 3P⏐M
ASTER
W x H x D: 350 mm x 160 mm x 125 mm (without surface-type outlets, grommets, cir­cuit breaker and rotary switch) (overall dimensions without cables)
Weight approx. 3.3 kg
Protection housing IP 44,
terminals IP 20
Table excerpt regarding significance of the IP code
s
Electromagnetic Compatibility (EMC)
Generic Standard EN 61326-1: 1997
EN 61326: 1997/A1: 1998
IP XY
(1
st
digit X)
Protection
against foreign
object entry
IP XY
(2nd
digit Y)
Protection
against the
penetration of
water
2 12.5 mm dia.
0 not protected
4 1.0 mm dia.
4 splashing water
Page 14
14 GMC-I Gossen-Metrawatt GmbH
9 Maintenance
9.1 Housing
No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives or solvents.
9.2 Replacing the Fuse
Instructions on checking the zero point and fuse are given in chapter 5.1.
Attention!
!
Disconnect the instrument from the measuring circuit before removing the fuse for replacement!
Eliminate the cause of the overload after the fuse has blown before putting the instrument back into service!
The fuse holder is situated between the mains power cable and the probe connector socket. The characteristic value of the fuse is indicated on the front plate or in chapter 8.
Attention!
!
Please make absolutely sure that only the specified fuse is inserted! If fuses with other blowing character­istics, other current ratings or other breaking capacities are used, the op­erator is placed in danger, and pro­tective diodes, resistors and other components may be damaged.
The use of mended fuses or short-circuiting the fuse holder is prohibited.
The defective fuse can be disposed of with the trash.
9.3 Calibration Intervals
According to DIN VDE 0702, measuring instruments used for periodic testing must be tested and calibrated in accordance with the manufacturer’s specifications on a regular basis. Depending upon usage, the manufacturer recommends an interval of 1 to 3 years for this test instrument.
Note: These tests can be performed on-site with the SECU-cal 10 calibration adapter acces­sory.
9.4 Return and Environmentally Sound Disposal
The test instrument is a category 9 product (monitoring and control instrument) in accor­dance with ElektroG (German electrical and electronic device law). This device is not subject to the RoHS directive.
We identify our electrical and elec­tronic devices (as of August 2005) in accordance with WEEE 2002/ 96/EC and ElektroG with the sym­bol shown to the right per DIN EN 50419.
These devices may not be disposed of with the trash.
Please contact our service department regarding the return of old devices (see chapter 11).
Page 15
GMC-I Gossen-Metrawatt GmbH 15
10 Accessories
10.1 Standard Equipment
Please refer to page 3 for the standard equipment accessories.
CD ROM (not for MINITEST⏐B
ASE)
System requirements
MS WINDOWS XP SP2 USB port 1.1 or 2.0.
10.2 Order Information for Available Accessories
The accessories available for your measur­ing instrument are regularly examined for compliance with the currently applicable safety standards and are extended to in­clude new application fields, if necessary. The suitable, up-to-date accessories for your measuring instrument are shown on our website www.gossenmetrawatt.com with the associated photo, order number, de­scription and, depending on the scope of the accessories, datasheet and operating in­structions.
Description Type Article
number
Probe for measuring protective conductor resistance, e.g. at rotating devices under test
Brush probe Z745G
Calibration adapter for test instruments per DIN VDE 0701/0702/0751 (max. 200 mA
)
SECU­cal 10 Z715A
Test adapter for electrical devices and extension cables with CEE plug-and­socket devices
VL2 E Z745W
Barcode scanner
B3261
GTZ 3261 000 R0001
Barcode and label printer including software
Z721D Z721D
Label set for printer Z721D (quantity x width: 3x24/1x18/1x9mm, each 8 m long)
Z722D Z722D
Label set for printer Z721D (5 ribbons, each 18 mm wide and 8 m long)
Z722E Z722E
PC software for the generation of reports and lists as supplement to MS-Word/EX­CEL (language version German/English/ French/Finnish/Polish (Exception: EXCEL only in German/English)
PC.doc­WORD ™/EX­CEL Z714A
Case for MINITESTMASTER, MINITESTPRO and MINITESTBASE
Case Z740B
Universal carrying pouch for
MINITESTMASTER, MINITESTPRO
and MINITESTBASE
F2000 Z700D
Universal carrying pouch big for MINITEST 3PMASTER
F2010 Z700F
Page 16
Edited in Germany • Subject to change without notice • A pdf version is available on the Internet
GMC-I Gossen-Metrawatt GmbH Thomas-Mann-Str. 16-20 90471 Nürnberg •
Germany
Phone +49 911 8602-111 Fax +49 911 8602-777 E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.com
11 Repair and Replacement Parts Service
DKD Calibration Laboratory* and Rental Instrument Service
If required please contact:
GMC-I Gossen-Metrawatt GmbH Service Center Thomas-Mann-Str. 20 90471 Nürnberg · Germany Phone: +49 911 8602-0 Fax: +49 911 8602-253 E-mail
service@gossenmetrawatt.com
This address is only valid in Germany. Please contact our representatives or sub­sidiaries for service in other countries.
*
Calibration Laboratory for Measured Electrical Quantities: DKD – K – 19701, accredited per DIN EN ISO/IEC 17025: 2005
Accredited quantities: direct voltage, direct current value, direct current resistance, alternating voltage, alternating current value, AC active power, AC apparent power, DC power, capacitance, frequency and temper­ature
Competent Partner
GMC-I Gossen-Metrawatt GmbH is certified in accordance with DIN EN ISO 9001:2000.
Our calibration laboratory is accredited per DIN EN ISO/IEC 17025 by the Physikalisch­Technischen Bundesanstalt (German Federal Institute of Physics and Metrology) and the Deutscher Kalibrierdienst (German Calibration Service) under registration num­ber DKD–K–19701.
We offer a complete range of expertise in the field of metrology: from test reports and fac-
tory calibration certificates, right on up to DKD calibration certificates.
Our spectrum of offerings is rounded out with free test equipment management.
Our service department includes an on-site DKD calibration bench. If errors are discovered during calibration, our specialized personnel are capable of completing repairs using original replacement parts.
As a full service calibration lab, we can cali­brate instruments from other manufacturers as well.
12 Product Support
If required please contact:
GMC-I Gossen-Metrawatt GmbH
Product Support Hotline
Phone: +49 911 8602-112 Fax: +49 911 8602-709 E-mail
support@gossenmetrawatt.com
Loading...