Observe this documentation, in particular all included safety information, in order to protect yourself and others from injury, and to
prevent damage to the instrument.
• Carefully and completely read and adhere to these operating
instructions, as well as the instrument’s condensed operating
instructions.
The respective documents can be found at
http://www.gossenmetrawatt.com. Retain these documents
for future reference.
• Use only the specified accessories with the instrument!
• Carefully and completely read and adhere to the product documentation for optional accessories. Retain these documents
for future reference.
• Tests may only be performed by a qualified electrician, or
under the supervision and direction of a qualified electrician.
The user must be instructed by a qualified electrician concerning performance and evaluation of the tests.
• Wear suitable and appropriate personal protective equipment
(PPE) whenever working with the instrument.
• Implement adequate measures for protection against electrostatic discharge (ESD).
• Observe and comply with all applicable DIN, VDE, EN, IEC and
ANSI guidelines. Inform yourself concerning the operating
company’s specifications and guidelines, as well as those of
the respective battery manufacturers.
• Batteries are electrochemical components with very high
short-circuit currents. Avoid short-circuits which endanger
yourself as a user, the entire system or operating personnel.
• If the instrument doesn’t function flawlessly, remove it from
operation and secure it against inadvertent use.
• The instrument may only be used as long as it’s in good working order.
Inspect the housing before use. Pay particular attention to any
possible cracks and the insulation around the sockets.
Damaged components must be replaced immediately.
• Accessories and cables may only be used as long as they’re
fully intact.
Inspect all cables and accessories before use. Pay particular
attention to damage, broken insulation or kinked cables.
Damaged components must be replaced immediately.
• Do not use the instrument after long periods of storage under
unfavorable conditions (e.g. humidity, dust or extreme temperature).
• Do not use the instrument after extraordinary stressing due to
transport.
• Do not use the instrument in potentially explosive atmospheres.
• Use the instrument only within the specified ambient conditions.
• Use the instrument only in accordance with the specified protection class (IP code).
• The instrument must not be exposed to direct sunlight.
• The instrument and the included accessories may only be
used for the measurements described in this manual.
• Only use accessories and the charger which are included with
the battery tester or listed as optional accessories.
• Only use the instrument and the accessories within the specified measuring category.
• The measurement cables and all external cabling must not
exceed an overall length of 1 meter.
• Plugging in the measurement cables and the power pack must
not necessitate any undue force.
• Connect the ground wire before connecting the voltage conducting cable, and disconnect the voltage conducting cable
before disconnecting the ground wire.
• Check measurement cable continuity.
• Avoid short circuits due to incorrectly connected measurement
cables.
• Never touch the conductive ends of the alligator clips, the test
probes or the Kelvin probes.
• Ensure that the alligator clips, test probes and Kelvin probes
make good contact.
• Do not move or remove the Kelvin probes during a measurement.
Unwanted sparking may otherwise occur due to test current.
• Do not remove or move the alligator clips, test probes or Kelvin
probes until measurement has been completed. Completion of
measurement is indicated by means of an acoustic acknowledgment.
• Avoid simultaneously contacting the battery and the frame or
hardware, which may be grounded.
• The battery tester is equipped with a transponder module.
Determine whether or not use of the 125 kHz frequency is permissible in your country.
• The battery tester is equipped with a Bluetooth
Determine whether or not use of the implemented frequency
range of 2.402 to 2.480 GHz is permissible in your country.
• The input at the P+ measurement jack is equipped with a fuse
link. The instrument may only be used as long as this fuse is in
flawless condition.
A defective fuse may only be replaced by our repair service.
• The instrument only works with inserted batteries.
• The battery charger may only be connected to electrical systems (TN or TT) with a maximum of 230/400 V which comply
with applicable safety regulations (e.g. IEC 60346, VDE 0100)
and are protected with a fuse or circuit breaker with a maximum rating of 16 A.
• Do not use the instrument while the internal batteries are being
charged.
• Do not use the instrument if the battery compartment cover
has been removed.
• Always create a backup copy of your measurement data.
®
module.
1
METRACELL BT PROGMC-I Messtechnik GmbH
Attention!
!
Note
Block
Battery
System
2Important Information
Please read this important information!
2.1Intended Use /
Use for Intended Purpose
Periodic testing and well-organized maintenance are necessary in
order to assure the availability of stationary battery systems.
The METRACELL BT PRO is a multifunctional instrument for testing batteries and battery blocks (referred to below as “devices
under test” or “DUTs”). It can be used to determine the current status of a DUT and pinpoint concealed defects. The battery tester is
used primarily for testing DUTs in stationary battery systems.
The battery tester stores measured values internally. All recorded
measured values can be viewed on site. Individual acquired values
can be conveniently re-measured if necessary. Management of
devices under test, as well as evaluation and backup of the measured values, is handled by the included “BT PRO Manager” software. The measured values can also be transferred to an external
computer as a CSV file.
The battery tester is a portable instrument which can be held in the
hand while performing measurements.
Safety of the operator, as well as that of the instrument, is only
assured when it’s used for its intended purpose.
2.2Use for Other than Intended Purpose
Using the instrument for any purposes other than those described
in the condensed operating instructions or these instrument operating instructions is contrary to use for intended purpose.
2.3Liability and Guarantee
GMC-I Messtechnik GmbH assumes no liability for property damage, personal injury or consequential damage resulting from
improper or incorrect use of the product, in particular due to failure
to observe the product documentation. Furthermore, all guarantee
claims are rendered null and void in such cases.
Nor does GMC-I Messtechnik GmbH assume any liability for data
loss.
2.4Opening the Instrument / Repairs
The instrument may only be opened by authorized, trained personnel in order to ensure flawless, safe operation and to assure
that the guarantee isn’t rendered null and void. Even original
replacement parts may only be installed by authorized, trained
personnel.
Unauthorized modifications to the instrument are prohibited.
If it can be ascertained that the battery tester has been opened by
unauthorized personnel, no guarantee claims can be honored by
the manufacturer with regard to personal safety, measuring accuracy, compliance with applicable safety measures or any consequential damages.
3Documentation
3.1Identifiers
The following identifiers are used in this documentation:
IdentifierMeaning
Safety information that must be complied with
Warning
Important information which must be
Important
PrerequisiteA condition etc. which must be fulfilled
1. Procedural stepSteps of a procedure which must be
ResultResult of a procedural step
• Enumeration
– Enumeration
Figure 1: Caption
Table 1:Description of the content of a table
Footnote
3.2Definition of Terms
A block is the smallest unit. A group of two or more blocks is
called a battery. For example, this can involve several blocks connected in series (e.g. a traction battery) or a group of batteries
consisting of two or more blocks in a UPS system.
Measurements can be conducted for individual blocks or batteries. From an organizational standpoint, a device under test is
always set up in the battery tester as a battery which includes one
or more blocks which will be measured. And thus if you only want
to measure one block, set up a battery with just one block.
taken into consideration and complied
with
before a given action can be taken
completed in the specified order
Bullet lists
Description of the content of a figure
Comment
2
Figure 2: Example – Blocks and Battery within a System
GMC-I Messtechnik GmbHMETRACELL BT PRO
2
11
3
4
5
6
7
10
9
1
8
4Getting Started
1. Read and adhere to the product documentation. In particular
observe all safety information in the documentation, on the
instrument and on the packaging.
“The Instrument” 3
“Operation”
“Menu and Functions”
3. Install BT PRO Manager software and familiarize yourself with
the program
4. Create a battery database 20.
5. Complete all required preparation in order to be able to perform measurements
6. Perform measurements
7. Evaluate your measurements “Viewing, Editing, Transferring
and Evaluating Measured Values”
7
18.
1
2
2
10
18.
30.
50.
5The Instrument
5.1Scope of Delivery
Please check for completeness.
1 METRACELL BT PRO
1 Rubber holster
4 1.2 V AA batteries
1Power pack
1 Set of multimeter test probes (KS29)
2 Alligator clips (KY95-3)
1 Set of Kelvin probes for 4-wire measurement (Z227D)
1 Carrying strap
1 Carrying case
1BTPRO Manager (battery tester management software)
1 Test report / factory calibration certificate
1 Condensed operating instructions
5.2Optional Accessories
Some measurements necessitate optional accessories:
– AC/DC current clamp sensor
– CP1800 (Z204A) for measurements up to 1250 ADC or
– CP330 (Z202B) for measurements up to 300 A
– METRATHERM IR BASE temperature sensor (Z680A)
– Spring-loaded contact pins as replacement parts for the Kelvin
probes (Z227F)
Please contact our product support team for further information
on accessories
53.
DC
Values recorded by and stored at the DMA 35 portable density
meter from Anton Paar GmbH can be transferred to the battery
tester. A description of how to use the density meter with the
METRACELL BT PRO is included in these operating instructions
(
30). Please contact Anton Paar GmbH for any further infor-
mation concerning the DMA 35 density meter.
5.3Instrument Overview
Front
Figure 3: Front Panel
3
METRACELL BT PROGMC-I Messtechnik GmbH
1
2
3
Attention!
!
!
Max.
6VDC
800 m
Symbols on the instrument and the included accessories:
No.Function
1Rubber holster (removable)
2Infrared interface
3Charging socket (charging cable connection)
4Measurement input S−
5Measurement input P−/COM
6Measurement input S+
7Measurement input P+
8Carrying strap eyelets
9Function keys
10Display
11RFID reader
Back
IR
Infrared interface location
Warning concerning a point of danger
(attention, observe documentation!)
Double insulation (protection category II)
Charging socket location
( “Instrument Overview”
Attention!
Only rechargeable NiMH batteries may be inserted when the charger is connected.
Fuse
Static magnetic field
European conformity marking
3)
Figure 4: Back of Instrument
No.Function
1Retaining magnet
2Retaining clip
3Battery compartment cover (underneath the holster)
The device may not be disposed of with household
trash “Returns and Environmentally Sound Disposal” 54.
5.4Relevant Standards
The battery tester has been manufactured and tested in accordance with the following safety regulations:
IEC 61 010-1
EN 61 010-1
VDE 0411-1
EN 60 529
VDE 0470, part 1
DIN EN 61 326-1
VDE 0843-20-1
Safety requirements for electrical equipment
for measurement, control and laboratory use
–
General requirements
Test instruments and test procedures
Degrees of protection provided by enclosures
(IP code)
Electrical equipment for measurement, control and laboratory use – EMC requirements –
Part 1: General requirements
4
GMC-I Messtechnik GmbHMETRACELL BT PRO
Note
Attention!
!
<Date>
<Time>
BTpro
<Version>
5.5Measurement Inputs
All measurement inputs are referenced to ground and reference
potential, P–/COM input. The measurement inputs are not electrically isolated from each other.
The measuring ranges depend on the respective measurement
inputs.
Measurement
input
S−Input for measuring DC voltage.
P−/COM Reference potential (ground potential) of all measure-
S+Measurement input for direct and alternating voltage
P+Current conducting cable to plus pole during resis-
Function
Measuring range: ±2450.00 mVDCResolution: 0.01 mV
Input impedance: > 10 M
Sensing lead to minus pole during resistance mea-
surement.
ment inputs.
Current conducting cable to minus pole during resistance measurement.
measurements.
Measuring
ranges:±24.5000 V
±600.000 V
Input impedance: 1.6 M
±300.000 V
Sensing lead to plus pole during resistance measurement.
tance measurement.
Max. 24 V
Maximum test voltage must not exceed
24 V DC at input P+.
The instrument is damaged if this value is
exceed.
600 V CAT III refers to measurement inputs S+, S−
and P−/COM.
DC
Resolution: 0.1 mV
DC
Resolution: 1 mV
DC
Resolution: 10 mV
AC
5.6LCD Display
All measured values, functions and settings can be observed at
the instrument’s display. As a rule, the display at the LCD panel
appears in two lines. Three and four-line displays are also used in
some cases, for example with measured values.
After switching the battery tester on, the initial display appears
including the instrument’s designation and the version, as well as
date and time.
Figure 5: Figure: Initial Display
After pressing any key while the initial display is shown, the battery
tester is switched to the main menu.
All other displays and information which appear at the instrument
depend on the previously selected function. The currently selected
function is always identified by a preceding arrow ->.
5.7Function Keys / Navigation
KeyFunction
“Function” key
– Switch on and off ( 7)
– Press briefly: Return to next higher menu level or
cancel measurement
– Press and hold: End (
measurement
Scroll down key “▼”
Selection of menu items or digits in downward direction.
Scroll up key “▲”
Selection of menu items or digits in upward direction.
The display is surrounded by an RFID reader. If individual batteries
are equipped with an RFID tag (transponder), the tag can be
scanned before measurement, thus making it possible to subsequently assign the measurement data to the respective DUT automatically
125 kHz Technology
The RFID reader can only read RFID tags which use a frequency of
125 kHz. This technology offers several advantages:
• As one of the first RFID technologies, it’s widely used and easy
to obtain.
• It uses robust tags which are relatively insensitive to interfer-
ence from metals and liquids, which is crucial in battery envi-
ronments.
• Universal use (license-free ISM band)
• Minimal energy consumption (low-frequency long-wave band –
LF passive tags)
• Reading range: approx. 10 cm
• Small data transfer volumes with fast reading process (e.g.
only object ID as battery identification)
Reading RFID Tags
In order to read an RFID tag, hold the battery tester display 1 to
2 cm above the RFID tag. RF appears at display.
32.
5.9Attachment Options
The battery tester provides acoustic feedback for all important
procedural steps, e.g. readiness to record measured values. Consequently, there’s no need to continuously observe the display.
In order to ensure freedom of movement, comfort and quick work
for the operator, the battery tester can be secured by means of
various retaining devices for the entire duration of measurement:
• Carrying strap:
The operator can use the carrying strap to suspend the battery
tester around his neck. The condensed operating instructions
include a description of how to attach the carrying strap.
• Retaining clip:
The operator can attach the battery tester to his belt with the
help of the retaining clip.
• Retaining magnet:
The battery tester can be attached to magnetic surfaces such
as battery cabinet doors with the retaining magnet.
5
METRACELL BT PROGMC-I Messtechnik GmbH
Note
Note
5.10 Technical Data
Power SupplyRechargeable NiMH batteries, 4 ea. 1.2 V type AA (recommended: Ansmann maxE 2500 mAh)
Operating temperature: +5 ... +40 °C
Storage temperature:-20 ... +60 °C
Relative humidity:Max. 75%, no condensation allowed
Elevation: Max. 2000 m
Measuring category:600 V CAT III
Pollution degree:2
Protection category:II per IEC 61 010-1/EN 61010-1/VDE 0411-1
Fuse link: 1 ea. SIBA 600 V/10 A FF
Test Voltage:Test voltage at measuring connection P+ may not exceed 24 V
Interference emission: EN 61326-1:2013, class A
Interference immunity: EN 61 326-1:2013
EN 61326-2-1:2013
Protection: Housing: IP40
per DIN VDE 0470, part 1/EN 60 529
(protection against ingress of solid foreign objects:
1.0 mm dia., protection against ingress of
water: not protected)
Housing:Approx. 9.6 × 15.4 × 3.3 cm (W × H × D)
Weight: Approx. 0.45 kg (without rubber holster)
Display:LCD, monochrome, luminous
IrDA:Connection for DMA 35 density meter, version 3
RFID:Connection for RFID tag
®
Bluetooth
: Connection for PC, headset and DMA 35 density meter, version 4
DC
.
Technical data for the AC/DC current clamp sensors, the temperature sensor and the DMA 35 density meter can be found in the
respective product documentation.
5.11 Characteristic Values
Measuring Function
Multimeter/
Connector
Measured QuantityV
Display Range
Measuring Range
-2450.00 …
+2450.00 mV
-2450.00 …
+2450.00 mV
DC
Multimeter/
Float/
Discharge/Charge
V
DC
-24.5000 …
+24.5000 V
-24.5000 …
+24.5000 V
Resolution0.01 mV0.1 mV1 mV10 mV0.01 m0.1 °C
Input Impedance /
Test Cu r r e n t
Intrinsic Uncertainty
S+
S−•••
P+•
Measuring
connec-
tions
P−/
COM
1)
Only applicable if the temperature sensor is connected to the battery tester.
2)
Within a frequency range of 45 to 500 Hz.
>10M1.6 M1.6 M1.6 Ml
±(0.05%
rdg. + 10 d)
±(0.05%
rdg. + 10 d)
• •••
• • ••••
Multimeter/
Interval U/
Internal U+I
V
DC
-600.000 …
+600.000 V
-600.000 …
+600.000 V
±(0.05%
rdg.+ 50 d)
MultimeterResistanceTemperature
V
AC
-300.000 …
+300.000 V
-300.000 …
+300.000 V
±(2.0%
rdg. + 10 d)
2)
Rel + R
ct
00.00 …
1000.00 m
°C
-230.0 …
+230.0 °C
00.10 …
1000.00 m
approx. 2 A> 10 M
p
±(3.0%
rdg. + 8 d)
1)
The characteristic values of the AC/DC current clamp sensors and the DMA 35 density meter, as well as additional characteristic
values for the temperature sensor, can be found in the respective product documentation.
6
GMC-I Messtechnik GmbHMETRACELL BT PRO
Attention!
!
Attention!
!
Note
Attention!
!
Attention!
!
Note
6Operation
6.1Power Supply (rechargeable batteries)
Battery Operation
The battery tester is operated with rechargeable NiMH batteries.
Pre-charged batteries are shipped loose. You have already
inserted the batteries during initial startup (
ating instructions).
Operating time is approximately 10 hours when the batteries are
fully charged. A warning is generated when battery voltage drops
to below 4.8 V. The battery tester can be operated for approximately 1 more hour after this warning is generated.
Leaking batteries can damage the device.
Check the batteries at short, regular intervals and after long
periods of storage.
Charging Process
It takes about 3½ hours to charge the batteries.
Damage to the batteries or the instrument.
When charging, use only:
- The included rechargeable batteries or the
batteries specified in the technical data
- The included charger
Always connect the battery charger to the
battery tester before connecting the charger
to the electrical outlet.
condensed oper-
The battery tester must be switched off.
All measurement cables and accessories must be removed
from the battery tester.
1. Remove the holster from the battery tester: Press against the
bottom of the holster at the front and push it back and off of
the battery tester. As soon as the lower part of the battery tester is exposed, pull the instrument completely out of the holster.
2. Place the battery tester face down on a stable surface.
3. Loosen and remove the screw from the battery compartment
cover.
4. Unlock and remove the battery compartment cover.
5. Remove any batteries from the battery compartment.
6. Insert 4 approved rechargeable batteries into the battery compartment, making sure that the plus and minus poles match up
with the provided polarity symbols.
Only the included batteries or those specified in the technical data ( 6) may be inserted and used.
7. Place the battery compartment cover onto the battery compartment and press it until it audibly clicks into place.
8. Secure the battery compartment cover with the screw.
9. Pull the holster back over the battery tester: Observe the cutouts in the holster for the front panel and the connections.
Push the battery tester into the holster at the top first as far as
it will go. Then press on the front at the bottom until the battery
tester is fully reinserted into the holster.
The batteries have now been inserted. The battery tester can
be used.
Do not switch the battery tester on during charging. The
battery tester may not be operated during charging and is
switched off automatically after 3 seconds.
The battery tester is switched off.
4 NiMH batteries are inserted
(see section on “Replacing the Batteries”
on inserting batteries).
All measurement cables and accessories must be removed
from the battery tester.
1. Insert the charger’s connector plug into the battery tester’s
charging socket “Instrument Overview”
2. Connect the charger to the 230 V mains outlet.
The batteries in the battery tester are charged.
3. Wait until the charging process has been completed.
It takes about 3½ hours to fully charge the batteries.
4. Unplug the charger from the electrical outlet.
5. Remove the connector plug from the battery tester’s charging
socket.
The batteries are charged. The battery tester can now be
used.
Replacing the Batteries
You can replace the batteries if necessary.
7 for information
3.
6.2Switching the Instrument On/Off
Switching the Instrument On
1. Press and hold the Function key for approximately 2 seconds.
The initial display appears with the instrument’s designation
and version, as well as date and time ( “LCD Display”
2. Press any key.
The main menu appears at the display. The battery tester can
now be used.
Switching the Battery Tester Off
Normally, the battery tester is switched off manually. If the battery
tester has not been used for more than 10 minutes, it’s switched
off automatically.
1. Select POWER OFF in the main menu with the help of the ▼
key.
2. Press the OK key.
The display goes blank. The battery tester has been switched
off.
The battery tester can also be switched off by pressing the
Function key. The key must be pressed and held for approximately 10 seconds to this end.
However, it’s preferable to switch the instrument off using
the SWITCH OFF function.
5).
Risk of injury due to contact with dangerous voltages.
Only use the battery tester when the battery compartment
cover is inserted and firmly screwed into place.
Required tools: Philips screwdriver
7
METRACELL BT PROGMC-I Messtechnik GmbH
Note
Note
Note
Note
6.3System Settings
You can enter basic settings to the battery tester.
Setting Date and Time
Switch back and forth between standard time and daylight
savings time manually. The battery tester is not equipped
with an automatic time change function.
Although the battery tester can be paired with several
devices via Bluetooth
eral devices via Bluetooth® is not possible.
®
Bluetooth
needed.
connections can be terminated when they’re no longer
®
, simultaneous operation with sev-
1. Switch the battery tester on 7.
2. Press the OK key.
The main menu appears.
3. Select SETTINGS from the main menu
using the ▼▲ keys.
4. Press the OK key.
The SETTINGS menu is displayed.
5. Select SYSTEM using the ▼▲ keys.
6. Press the OK key.
The SYSTEM menu is displayed.
7. Select SET TIME&DATE using the ▼▲ keys.
8. Press the OK key.
The SET TIME&DATE menu is displayed. The cursor is positioned
at the first digit of the date display (dd.mm.yy: dd = day, mm =
month and yy = year).
9. Use the ▼▲ keys to enter the first digit of the two-digit display
for the current day.
10. Press the OK key to move to the second digit.
11. Repeat steps 9. and 10. until the date has been completely
entered.
After entering and confirming the last digit for the year, the cursor jumps to the first digit of the time display (ss:mm:hh: ss =
second, mm = minute and hh = hour)
12. Set the digits for time in the same way as described for setting
the date.
After entering the last digit for current time, your settings are
saved. The SYSTEM menu is displayed.
Date and time have been set.
Language Selection
Device language can be set to either English or German.
1. Switch the battery tester on
2. Press the OK key.
The main menu appears.
3. Select SETTINGS from the main menu
using the ▼▲ keys.
4. Press the OK key.
The SETTINGS menu is displayed.
5. Select SYSTEM using the ▼▲ keys.
6. Press the OK key.
The SYSTEM menu is displayed.
7. Select SEL. LANGUAGE using the ▼▲ keys.
8. Press the OK key.
The SEL. LANGUAGE menu is displayed.
9. Select the desired language using the ▼▲ keys.
10. Press the OK key.
The device language is changed.
7.
6.4Establishing and Terminating Wireless Connections
The battery tester can be wirelessly connected to various devices:
• To the DMA 35 density meter from Anton Paar GmbH
– DMA 35 version 3 via infrared “DMA35 Measurement
(IrDA)”
49
– DMA 35 version 4 via Bluetooth
ment (BT)”
• To a PC in order to view, evaluate and save measurement data
“Saving Measured Values to a PC (data backup)”
• To a headset for improved audibility of signals generated by the
instrument “Establishing and Testing a Bluetooth® Connection with a Headset”
49
8
®
“DMA35 Measure-
50
In case of connection problems, it may be helpful to terminate (
references to specific sections above).
Establishing and Testing a Bluetooth
Acoustic signals are used to confirm various actions before, during
and after a measuring procedure, as well as certain instrument
reactions.
The instrument can be paired with a headset, making it possible to
better hear these signals in noisy work environments. The connection can be checked in advance with a test signal.
Read and adhere to the documentation included with your
headset.
The headset supports Bluetooth
1. Switch the battery tester on
2. Press the OK key.
The main menu appears.
3. Select SETTINGS from the main menu
using the ▼▲ keys.
4. Press the OK key.
The SETTINGS menu is displayed.
5. Select BT-SETTINGS using the ▼▲ keys.
The BT-SETTINGS menu is displayed.
6. Select LINK HEADSET using the ▼▲ keys.
7. Press the OK key.
8. The PUT HEADSET IN PAIRING MODE -> prompt appears.
9. Set the headset to pairing mode. Read the documentation
included with your headset to find out how.
10. Position the headset close to the battery tester.
11. Acknowledge the PUT HEADSET IN PAIRING MODE -> prompt by
pressing the OK key.
BT PAIRING: SEARCH HEADSET! appears at the display.
12. The headset and the battery tester are connected to each
other. Pairing progress is displayed as a percentage. As soon
as the devices are successfully connected, SUCCESSFUL!
appears at the display.
13. Put on the headset.
14. Press the OK key.
The BT-SETTINGS menu is displayed.
15. Press the Function key.
The main menu appears.
16. Select SETTINGS from the main menu using the ▼▲ keys.
17. Press the OK key.
The SETTINGS menu is displayed.
18. Select SYSTEM using the ▼▲ keys.
19. Press the OK key.
The SYSTEM menu is displayed.
20. Select TEST HEADSET using the
21. Press the OK key.
A test signal is played back at the headset.
9) and reestablish Bluetooth
®
Connection with a Headset
®
profile “HSP”.
7.
▼▲ keys.
®
connections (see
8
GMC-I Messtechnik GmbHMETRACELL BT PRO
Terminating the Bluetooth® Connection
All paired devices are disconnected when the Bluetooth
tion is terminated. It’s not possible to disconnect individual
devices.
1. Switch the battery tester on 7.
2. Press the OK key.
The main menu appears.
3. Select SETTINGS from the main menu
using the ▼▲ keys.
4. Press the OK key.
The SETTINGS menu is displayed.
5. Select BT-SETTINGS using the ▼▲ keys.
The BT-SETTINGS menu is displayed.
6. Select REM. PAIRINGS using the ▼▲ keys.
7. Press the OK key.
The DO YOU REALLY WANT TO DELETE? prompt appears.
8. Press the OK key.
DELETING DATA! PLEASE WAIT appears briefly at the display.
REM. PAIRINGS appears at the display.
Bluetooth
minated.
®
connection to all paired devices has now been ter-
®
connec-
6.5Testing the RFID Reader
The RFID reader can be tested ( 5).
A compatible RFID tag (125 kHz) is available.
1. Switch the battery tester on
2. Press the OK key.
The main menu appears.
3. Select SETTINGS from the main menu
using the ▼▲ keys.
4. Press the OK key.
The SETTINGS menu is displayed.
5. Select SYSTEM using the ▼▲ keys.
6. Press the OK key.
The SYSTEM menu is displayed.
7. Select TEST TRANSP. using the ▼ key.
8. Press the OK key.
WAITING FOR RFID – – – – –> ( ) <– – – – – appears at the display.
Furthermore, a repetitive acoustic signal indicates readiness to
read.
9. Hold the battery tester’s RFID reader roughly 1 to 2 cm above
the RFID tag.
The RFID tag is read. The RFID tag’s number appears at the
display.
7.
Refer to section on “Performing Measurements”
assign RFID tags to DUTs and read them out during measurement.
Menu is displayed in selected language,
return to
LANGUAGE
SEL.
display
WAITING FOR
RFID
<RFID
Number>
Brief tone signals selection,
display remains in
menu
List with
all detected
BT devices is
displayed.
BT PAIRING:
SEARCH
HEADSET
DELETING
DATA!
PLEASE WAIT!
NO BT
DEVICES
FOUND
Detected BT
devices are
listed
11
METRACELL BT PROGMC-I Messtechnik GmbH
=
=
OK
ON
=
A
=OK=
=
ON
=
B
7.2Display Menu
General Key
-> MEASUREMENT
DATA
DELETING DATA
Display menu/function,
the arrow indicates the selected function.
Feedback or information after action
PLEASE WAIT
BLOCK 12.2500V
0004/4
Display measured value (example)
Display measured value evaluation
AN:0001 BN:0010
Descriptive textExplanation of measured value, parameter
and other displays
Scroll within menu levels and functions.
Start function,
submenu selection
Switch on,
termination of an action currently being executed
Move up one view in the menu.
The initial display appears after switching the battery tester on
(
7). The main menu is then displayed after pressing any key.
If you’re already in a submenu, press the FUNCTION key repeatedly
until the display is returned to the main menu.
This is the central menu from which all other hierarchically subordinate menus and functions can be accessed. All main menu functions can be selected with the help of the ▼▲ keys.
Table 10:List of Functions Available in the Main Menu
FunctionDescription
MEASUREMENTAvailable measuring functions
DATA MENUData management functions
12
15
SETTINGSParameter settings for system, measure-
ments and data transfer
16
INFOInformation concerning firmware version, bat-
tery charge level and memory occupancy
17
POWER OFFSwitch off the battery tester
7.
Move down one view in the menu.
Jump to submenu level (press briefly).
Jump to submenu level (press and hold).
Jump back to higher menu level (initial position).
Remains at current position.
No action required.
Further action and displays depend on
systems to be linked or downstream systems.
0Currently selected digit (display with black
background).
XPlaceholder for one digit.
*Parameter not for measuring functions
INTERVAL U and INTERVAL U+I
**Parameters only for measuring functions
INTERVAL U and INTERVAL U+I
***Parameters only for measuring functions
INTERVAL U and INTERVAL U+I
Set/select special key assignments for
parameters
Cursor for selecting the digit to be
changed (inverted cursor), storage takes
place after selection of the last digit, then
return to higher menu level
Increase digit
Decrease digit
Cancel entry/selection of digits without
saving, return to higher menu level
RFID tags must be held roughly 1 to 2 cm above the display
7.3Measurement Menu
Select MEASUREMENT in the main menu and then press the OK
key in order to open the menu.
Table 11:List of Functions Available in the MEASUREMENT Menu
MeasurementDescription
MULTIMETERDC and AC voltage measurements without
FLOATPeriodic measurement of block voltages.
DISCHARGEMultiple measurement of block voltages at
CHARGEMultiple measurement of block voltages at
RESISTANCEPeriodic measurement of the internal resis-
TEMPERATUREMeasurement of block temperature with an IR
CONNECTORMeasurement of voltage drop to determine
INTERVAL UMeasurement of the voltage of a battery at any
INTERVAL U + IMeasurement of the voltage and current of a
DMA 35 (IrDA)Measurement of acid density and electrolyte
DMA 35 (BT)Measurement of acid density and electrolyte
QUITExit the MEASUREMENT menu and return to the
storing measured values
This measurement is used for quarterly recording of float voltage, for example in UPS systems.
short intervals during discharging (capacitance
test)
short intervals during charging (capacitance
test)
tance of the blocks
temperature sensor
connector loss between blocks
desired time intervals (voltage curve)
battery at any desired time intervals (voltage
and current curves)
Example: Recording of discharge current
during discharging.
temperature within a block.
Measurements are performed with the DMA 35
density meter (version 3) from Anton Paar
GmbH. Data are transferred from the sensor to
the battery tester via infrared (Infrared Data Association – IrDA).
temperature within a block.
Measurements are performed with the DMA 35
density meter (version 4) from Anton Paar
GmbH. Measured values are transferred from
the sensor to the battery tester via Bluetooth
®
main menu.
12
GMC-I Messtechnik GmbHMETRACELL BT PRO
Note
Temperature
The required parameters for the device under test must be
defined before starting a measurement. Use the battery
database to this end ( “Performing Measurements”
30
and “Transferring the Battery Database to the Battery
Tester”
23).
Manual entries should only be made in exceptional cases.
Multimeter
Select the MEASUREMENT MULTIMETER function sequence via
the main menu and press the OK key in order to open the menu.
Table 12:Possible Measuring Ranges for MULTIMETER Measure-
ment
Measuring
Range
2.45 V
DC
24.5 V
DC
600 V
DC
300 V
AC
Description
(S–)Measuring range selection for direct voltage mea-
surement via measuring socket S–
(S+)Measuring range selection for direct voltage mea-
surement via measuring socket S+
(S+)Measuring range selection for direct voltage mea-
surement via measuring socket S+
(S+)Measuring range selection for alternating voltage
measurement via measuring socket S+
QUITReturn to MEASUREMENT menu
Float, Discharge, Charge and Resistance
In order to open the menu, select one of the following function
sequences from the main menu:
MEASUREMENT FLOAT or MEASUREMENT
DISCHARGE or MEASUREMENT CHARGE or MEASUREMENT
RESISTANCE, and then press the OK key.
Table 13:Parameters and Functions for the FLOAT, DISCHARGE,
CHARGE and RESISTANCE Measurements
Function/
Setting
Description
STARTMeasurement is started after selecting or en-
tering the battery parameters.
SELECT.BATTThe menu for selecting and loading a stored
battery database is accessed via this function.
If a database exists, the stored data records
appear.
AREA: XXXXManual assignment of a unique four-digit
number for the location of the battery system
BATTERY: XXXXManual assignment of a unique four-digit
number as a designation of the battery in the
battery system
BLOCKS:XXXXManual assignment a unique four-digit num-
ber identifying the number of blocks within the
battery
MIN: XX.XXX VManual measuring range setting: lower volt-
age limit value
MAX: XX.XXX VManual measuring range setting: upper volt-
age limit value
(max. +24.50 V)
QUITReturn to MEASUREMENT menu
Select the MEASUREMENT TEMPERATURE function sequence
via the main menu and press the OK key in order to open the menu.
Table 14:Parameters and Functions for the TEMPERATURE Mea-
surement
Function/
Setting
Description
STARTMeasurement is started after selecting or en-
tering the battery parameters.
SELECT.BATTThe menu for selecting and loading a stored
battery database is accessed via this function.
If a database exists, the stored data records
appear.
AREA: XXXXManual assignment of a unique four-digit
number for the location of the battery system
BATTERY: XXXXManual assignment of a unique four-digit
number as a designation of the battery in the
battery system
BLOCKS:XXXXManual assignment a unique four-digit num-
ber identifying the number of blocks within the
battery
MIN: XXXX.X CManual measuring range setting: lower tem-
perature limit value
(max. –230 °C)
MAX: XXXX.X CManual measuring range setting: upper tem-
perature limit value
(max. +230 °C)
QUITReturn to MEASUREMENT menu
Start / Battery Parameters Menu for Connector Measurement
Select the MEASUREMENT CONNECTOR function sequence via
the main menu and press the OK key in order to open the menu.
Table 15:Parameters and Functions for the CONNECTOR Mea-
surement
Function/
Setting
Description
STARTMeasurement is started after selecting or en-
tering the battery parameters.
SELECT.BATTThe menu for selecting and loading a stored
battery database is accessed via this function.
If a database exists, the stored data records
appear.
AREA: XXXXManual assignment of a unique four-digit
number for the location of the battery system
BATTERY: XXXXManual assignment of a unique four-digit
number as a designation of the battery in the
battery system
BLOCKS:XXXXManual assignment a unique four-digit num-
ber identifying the number of blocks within the
battery
MIN: XXXX.X mVManual measuring range setting: lower volt-
age limit value
MAX: XXXX.X mVManual measuring range setting: upper volt-
age limit value
(max. +2400.0 mV)
QUITReturn to MEASUREMENT menu
13
METRACELL BT PROGMC-I Messtechnik GmbH
Note
Note
Note
Note
Start / Battery Parameters Menu for Interval U Measurement
Select the MEASUREMENT INTERVAL U function sequence via
the main menu and press the OK key in order to open the menu.
Table 16:Parameters and Functions for the INTERVAL U Mea-
surement
Function/
Setting
STARTMeasurement is started after selecting or en-
SELECT.BATTThe menu for selecting and loading a stored
AREA: XXXXManual assignment of a unique four-digit
BATTERY: XXXXManual assignment of a unique four-digit
INT [s]: XXXXManually adjustable time interval (in seconds)
MIN: XXX.XX VManual measuring range setting: lower volt-
MAX: XXX.XX VMeasuring range setting: upper voltage limit
QUITReturn to MEASUREMENT menu
Interval U + I
Select the MEASUREMENT INTERVAL U+I function sequence via
the main menu and press the OK key in order to open the menu.
Table 17:Parameters and Functions for INTERVAL U+I Measure-
Function/
Setting
STARTMeasurement is started after selecting or en-
SELECT.BATTThe menu for selecting and loading a stored
AREA: XXXXManual assignment of a unique four-digit
BATTERY: XXXXManual assignment of a unique four-digit
INT [s]: XXXXManually adjustable time interval (in seconds)
Description
tering the battery parameters.
battery database is accessed via this function.
If a database exists, the stored data records
appear.
number for the location of the battery system
number as a designation of the battery in the
battery system
for recording the measured value.
(min. 1 second)
age limit value
Same as discharge cut-off voltage. If
the limit value is fallen short of, the
instrument generates an acoustic signal at regular intervals. This indicates
that the discharging process has been
completed and measurement can be
ended.
value
Same as charge cut-off voltage. If the
limit value is exceeded, the instrument
generates an acoustic signal at regular
intervals. This indicates that the
charging process has been completed
and measurement can be ended.
(max. +600 V)
ments
Description
tering the battery parameters.
battery database is accessed via this function.
If a database exists, the stored data records
appear.
number for the location of the battery system
number as a designation of the battery in the
battery system
for recording the measured value.
(min. 1 second)
Function/
Setting
MIN: XXX.XX VManual measuring range setting: lower volt-
MAX: XXX.XX VManual measuring range setting: upper volt-
CLAMP A: XXXXManual measuring range setting: Current
CLAMP mV: XXXXManual measuring range setting: Voltage
QUITReturn to MEASUREMENT menu
DMA35 (IrDA and Bluetooth
Select the MEASUREMENT DMA35 (IrDA) or DMA35 (BT) function sequence via the main menu and then press the OK key in order
to open the menu.
Table 18:Parameters and Functions for the DMA35 (IrDA) and
Function/
Setting
STARTMeasurement is started after selecting or en-
SELECT.BATTThe menu for selecting and loading a stored
AREA: XXXXManual assignment of a unique four-digit
BATTERY: XXXXManual assignment of a unique four-digit
BLOCKS:XXXXManual assignment a unique four-digit num-
MIN: X.XXXXManual measuring range setting: lower limit
MAX: X.XXXXManual measuring range setting: upper limit
QUITReturn to MEASUREMENT menu
Description
age limit value
Same as discharge cut-off voltage. If
the limit value is fallen short of, the
instrument generates an acoustic signal at regular intervals. This indicates
that the discharging process has been
completed and measurement can be
ended.
age limit value
Same as charge cut-off voltage. If the
limit value is exceeded, the instrument
generates an acoustic signal at regular
intervals. This indicates that the
charging process has been completed
and measurement can be ended.
(max. +600 V)
value of the transformation ratio of the current
clamp
value of the transformation ratio of the current
clamp
®
)
DMA35 (BT) Measurements
Description
tering the battery parameters.
battery database is accessed via this function.
If a database exists, the stored data records
appear.
number for the location of the battery system
number as a designation of the battery in the
battery system
ber identifying the number of blocks within the
battery
value for acid density
value for acid density
14
GMC-I Messtechnik GmbHMETRACELL BT PRO
Note
7.4Data Menu
Select DATA MENU in the main menu and then press the OK key in
order to open the menu.
Table 19:Functions in the DATA Menu
FunctionDescription
SHOW DATADisplay of stored measured values
EXPORT (CSV)Transfer of stored measured values to the PC
SHOW BATT-DEFDisplay of the battery database
CLEAR DATAComplete, irretrievable deletion of all stored
CLEAR BAT-DEFComplete, irretrievable deletion of a battery
QUITReturn to the main menu
Measured Values
Select the DATA SHOW DATA function sequence from the main
menu and press the OK key in order to open the MEASUREMENTS
menu.
Table 20:Information on Stored MEASUREMENTS
DesignationDescription
Measuring Function
[XXXXX]Meaning varies depending on measurement:
DD.MM.YYDate of measurement
HH:MM:SSTime of measurement
AN:XXXXLocation number assigned to the battery sys-
BN:XXXXBattery number assigned to the battery
as CSV file
measured values
database
Designation of the measuring function
– Number of blocks over which measure-
ments have been performed
– Number of measurements (e.g. for
INTERVAL U)
tem
DesignationDescription
AN:XXXXLocation number assigned to the battery sys-
tem
BN:XXXXBattery number assigned to the battery
Table 23:Measured RESISTANCE Values
DesignationDescription
BLOCKMeasurement over one block
XXXX/XBattery number / number of blocks
XX.XXXX VMeasured voltage value
R(el) =Measured value of the internal electrical resis-
tance of a block
R(ct) =Measured value of the internal electrochemical
resistance of a block
Table 24:Measured TEMPERATURE Values
DesignationDescription
BLOCKMeasurement over one block
XXXX/XBattery number / number of blocks
XXXX.X CMeasured temperature value
AN:XXXXLocation number assigned to the battery sys-
tem
BN:XXXXBattery number assigned to the battery
Table 25:Measured CONNECTOR Values
DesignationDescription
BLOCKMeasurement over one block
XXXX/XBattery number / number of blocks
XXXX.XX mVMeasured voltage value
AN:XXXXLocation number assigned to the battery sys-
tem
BN:XXXXBattery number assigned to the battery
Navigate through all of the measurements in the MEASUREMENT VALUES menu using the ▼▲ keys. After selecting a given mea-
surement and pressing the OK key, measured values and measured value details are displayed.
If an RFID tag has been assigned to a system and the tag is
read before measurement, the RFID tag’s number is displayed instead of the AN and BN numbers in the corresponding measured values / measured value details.
Table 21:Measured FLOAT Values
DesignationDescription
BLOCKMeasurement over one block
XXXX/XBattery number / number of blocks
X.XXXX VMeasured voltage value
AN:XXXXLocation number assigned to the battery sys-
tem
BN:XXXXBattery number assigned to the battery
Table 22:Measured DISCHARGE and CHARGE Value s
DesignationDescription
BLOCKMeasurement over one block
XXXX/XBattery number / number of blocks
X.XXXX VMeasured voltage value
Table 26:Measured INTERVAL U Values
DesignationDescription
COUNT:Number of measurements performed over a
freely defined time interval
X.XXX VDisplay of overall battery voltage which was
measured during a freely defined time interval
TIME.Time interval during which the battery’s overall
voltage was measured
Table 27:Measured INTERVAL U + I Values
DesignationDescription
COUNT:Number of measurements performed
X.XXX VOverall battery voltage measured during the
defined time interval
X.XXXX ABattery current measured during the defined
time interval
TIME.Time interval during which the battery’s overall
voltage and battery current were measured
Table 28:Measured DMA (IrDA) and DMA35 (BT) Values
DesignationDescription
BLOCKMeasurement over one block
XXXX/XBattery number / number of blocks
X.XXX kg/lAcid density
X.XXX CTe mpe ra tu re
15
METRACELL BT PROGMC-I Messtechnik GmbH
DesignationDescription
AN:XXXXLocation number assigned to the battery sys-
tem
BN:XXXXBattery number assigned to the battery
Menu for the Display of Measured Value Evaluations
In addition to individual measurements, measured value evaluations can also be displayed. Select a measurement in the SHOW DATA menu using the ▼▲ keys. Then press and hold the OK key
until the menu with the measured value evaluations is displayed.
Complete information is available in the section on “Viewing, Editing, Transferring and Evaluating Measured Values”
50.
Table 29:Measured Value Evaluations
DesignationDescription
SUM: [XXXX]Sum of all measured values measured over
the blocks of a battery. The number of blocks
over which measurement has been performed
is shown in square brackets (e.g. for FLOAT or
RESISTANCE).
NUMBER [XXXX]Number of all measured values that were
measured during a multiple measurement over
a block. The number of measurements is
shown in square brackets (e.g. for TEMPERA-
).
TURE
AVER.:Mean value of the measured values measured
either over all blocks of a battery or over one
block in the case of a multiple measurement.
MIN: XXXXSmallest measured value of the measurement
with indication of the number of the block at
which the measured value was measured.
MAX: XXXXLargest measured value of the measurement
with indication of the number of the block at
which the measured value was measured.
7.5The SETTINGS Menu
Select SETTINGS in the main menu and then press the OK key in
order to open the menu.
Table 30:SETTINGS Menu Functions
FunctionDescription
SYSTEMSystem settings
MEASUREMENTSSettings for measured value recording
®
BT-SETTINGSSettings for Bluetooth
CSV EXPORTSettings for the transmission of data and mea-
sured values as a CSV file
SERVICE MENUThis function is for service only ( “Service
and Support”
53). Access is restricted by
means of a password.
QUITExit the SETTINGS menu and return to the main
menu.
SYSTEM
Select the SETTINGS SYSTEM function sequence from the main
menu and press the OK key in order to open the SYSTEM menu.
Table 31:Functions in the SYSTEM Menu
FunctionDescription
SET TIME&DATESettings for device date and time
SEL. LANGUAGEUser language selection
TEST HEADSETPlayback of a test signal at a headset in order
to check the Bluetooth
QUITReturn to SETTINGS menu
connection
®
function
MEASUREMENTS
Select the SETTINGS MEASUREMENTS function sequence from
the main menu and press the OK key in order to open the MEASURE-MENTS menu.
Table 32:List of Functions Available in the MEASUREMENTS Menu
FunctionDescription
POLARITY ONSwitch polarity on: the negative or positive
measured value is displayed.
POLARITY OFFSwitch polarity off: the absolute value of the
measured value is displayed.
QUITReturn to SETTINGS menu
BT-SETTINGS
Select the SETTINGS BT SETTING function sequence from the
main menu and press the OK key in order to open the BT SETTINGS
menu.
Table 33:Functions in the BT SETTINGS Menu.
FunctionDescription
LINK COMPUTERFunction for establishing a Bluetooth® con-
nection between the battery tester and a PC
®
LINK HEADSETFunction for establishing a Bluetooth
connection between the battery tester and a
headset
®
LINK DMA35 V4Function for establishing a Bluetooth
connection between the battery tester and a version 4 DMA 35 density meter from Anton Paar
GmbH
®
REM. PAIRINGSTermination of the Bluetooth
connection to
all paired devices
(device connections cannot be terminated individually)
QUITReturn to SETTINGS menu
CSV EXPORT
Select the SETTINGS CSV EXPORT function sequence from the
main menu and press the OK key in order to open the CSV EXPORT
menu.
Table 34:Functions in the CSV EXPORT Menu
FunctionDescription
EXPORT TARGETSelection of the destination (e.g. PC) for the
transmission of measured values as a CSV file
(a Bluetooth
®
connection must first be estab-
lished between the battery tester and the PC)
FILENA. FIXEDFunction for defining a uniform filename when
transferring measured values to a PC as a
CSV file
The file is named “BTPRO_Export.csv”.
FILEN. UNIQ.Function for generating different filenames
each time measured values are transferred to
a PC as a CSV file The file is named “BTPRO_Export_<SystemID>_<Date>_<Time>.csv”.
QUITReturn to SETTINGS menu
16
Loading...
+ 40 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.