ULTRAFLO U3000 / U4000 Ultrasonic Flowsensor
Features
• Large, easy to read graphic display
• Easy to install
• Clamp-on sensors for dry servicing
• Simple to follow programming menu
• Simple ‘Quick Start’ set up procedure
• Analog, pulse and alarm outputs
• Reynolds number correction
The Ultraflo brings simplicity to the non-invasive measurement of liquid flow, offering the user quick and accurate flow measurement with its easy to follow menu and simple set up. Dry servicing, providing minimum downtime and maximum availability, even in a continuously running system. Compact, rugged and reliable, the Ultraflo has been designed to provide sustained performance in industrial environments.
Applications
• HVAC & Energy System Audits
• Pump Verification
• Process Control
• Chemical Addition
• Hydraulic Systems
• Fire Systems
• Leak Detection
• Boiler Testing
Specifications
General
DSP Measurement Technique Transit time Flow Velocity Range 0.1 m/s - 20 m/s Accuracy
Repeatability ±0.5% of measured value or ±0.02 m/s whichever is the greater Response Time Selectable Flow Units Velocity m/sec, ft/sec.
Selectable Total Volume Units litres, m³, gallons, US gallons, barrels Total Volume 12 Digits Menu Languages EN, DE, FR, RU, SWE, IT, SP, P, NO, DEN
Environmental
Operating Temperature -20 °C to 50 °C -4 °F to 122 °F Storage Temperature -25 °C to 75 °C -13 °F to 167 °F Pipe Wall Temperature -20 °C to 135 °C -4 °F to 275 °F Operating Humidity
Applicable Pipe Types
Pipe Materials PVDF-SYGEF, PP-PROGEF, PE-ELGEF, PB-INSTAFLEX, ABS, PVC-U/PVC-C, Mild Steel, Ductile
Pipe Dimension (OD) 13 mm to 2000 mm 0.5 in. to 78 in. Pipe Wall Thickness 1 mm to 75 mm 0.04 in. to 3 in. Pipe Lining Applicable pipe linings include Rubber, Glass, Concrete, Epoxy, Steel Pipe Lining Thickness 0 mm to 25 mm 0 in. to 1 in.
Electrical
Supply Voltage 12 - 24 V AC or DC; 86 - 264 V AC (47Hz to 63Hz) Power Consumption Max. 10.5 W
Outputs
Analog Output Range 4 to 20 mA, 0 to 20 mA, 0 to 16 mA
Pulse Output Type Digital MOSFET relay
Alarm Outputs Type 2 x MOSFET relays
USB Interface (U4000 only) Protocol Supports full speed (12Mbits/sec) data connection
RS-232 Interface (U4000 only) Protocol "Serial RS-232 communication including XON/XOFF handshaking"
Pipe ID >75 mm ±0.5% to ±3% of flow reading for flow rate >0.2 m/s Pipe ID 13 mm - 75 mm
< 500 ms depending on pipe diameter.
Volume l/s, l/min, l/h, gal/min, gal/h, USgals/min, USgals/h, Barrel/h,
Max. 90% relative humidity @ 50 °C (122 °F)
Iron, Stainless Steel 316, Copper
Resolution 0.1% of full scale Load Max. 620 Ω Isolation 1500 V Opto-isolated Alarm Current Adjustable between 0–26 mA
Pulse Repetition 1 to 250 pps, user programmable Pulse Width 2 to 500 ms, user programmable Voltage Max. 48 V Current Max. 500 mA Isolation 1500 V opto isolated
Voltage Max. 48 V Current Max. 500 mA Isolation 1500 V opto isolated Alarm Function High / Low flow rate, flow volume or signal error
Software USB driver software is provided with the package Connector Mini USB
Terminal Block GND, RxD, TxD
±3% of flow reading for flow rate >0.2 m/s
Barrel/day, m³/s, m³/min, m³/h.
www.gfsignet.com