• Remove power to unit before wiring input and output connections.
• Follow instructions carefully to avoid personal injury.
• This product should only be used for the purposes and in the manner described in this manual.
Conductivity/Resistivity
Relay 1
Installation
1. Punch out panel and de-burr edges. Recommended clearance on all sides between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. Connect wires to terminals.
5. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE. Press
quick-clips outward and remove.
6. If cleaning is necessary, wipe the front of the unit with a damp cloth.
7. The live contacts on the back of this unit must be covered to avoid accidental shock hazard.
gasket on
front side
96mm (3.8 in.)
96 mm
(3.8 in.)
102 mm
(4.0 in.)
Optional
Rear Cover
56 mm
(2.2 in.)
Panel Cutout
92 x 92 mm
(+ 0.8, - 0 mm)
3.6 x 3.6 in.
(+0.031, -0 in.)
panel
of panel
RELAY 1
COM
N/O
N/C
- OPEN COLL 4
OPTION
OUTPUT
RELAY 3,4
quick-clip
terminals
mounting
RELAY 4
RELAY 3
RELAY 2
COM
N/O
POWER
N/OCOM N/C
L
N/C
COM
N/O
N
N/C
OPEN COLL 3
-
SGNL 2
TEMP 2
ISO GND
SHLD
SGNL 1
MP 1
E
T
OPEN
HLD
ISO GND
COLL 3,4
S
bracket
C1 23.45 µS
C2 8.72 µS
Signet
Transmitter
Relay 2
ENTER
Specifi cations
General
Compatible electrodes: Signet 2818 to 2823 (Standard or
Certifi ed Cond/Resist Sensors)
Enclosure:
• Rating: NEMA 4X/IP65 front (NEMA 4X Rear cover available)
• Case: PBT
• Panel case gasket: Neoprene
• Window: Polyurethane-coated polycarbonate
• Keypad: Sealed 4-key silicone rubber
• Weight: 8860-AC: 0.58 kg (1.28 lb) 8860: 0.55 kg (1.2 lb)
Display (Alphanumeric 2 x 16 LCD) :
• Contrast: User selected, 5 levels
• Update rate: 1.5 s
Sensor input range:
• Conductivity: 0.01 μS/cm to 400 000 μS/cm
• Resistivity: 10 k•cm to 100 M•cm
• TDS: 0.023 to 200 000 PPM nominal (adjustable μS/PPM)
• Temperature:
-25 °C to 120 °C (-13 °F to 248 °F) PT 1000; 25 ºC = 1096
Accuracy:
• Conductivity/Resistivity: ± 2% of reading
• Temperature: ± 0.5 ºC (0 ºC to 100 ºC)
Electrical
Power requirements:
3-8860-AC: 100 to 240 VAC ± 10% , 50 to 60 Hz, 20VA or
11 to 24 VDC ±10%, regulated, 0.5 A max
3-8860: 11 to 24 VDC ±10%, regulated, 0.5 A max
Three 4 to 20 mA Outputs:
• Passive, isolated, fully adjustable and reversible
4 to 20 mA outputs are independently source selectable
• Max loop impedance:
150 @ 12 V, 450 @ 18 V, 750 @ 24 V
• 50 mA sink or source, 30 VDC max. pull-up voltage
• Programmable for:
• High or Low setpoint with adjustable hysteresis
• Pulse operation (max. rate: 400 pulses/min)
• USP standards
• Time delay: 0 s to 6400 s.
Relay outputs (up to 4 SPDT relays available)
• Maximum resistive load: 5 A @ 250 VAC, 5 A @ 30 VDC
• Isolation between open contacts: 500 V minimum
• Programmable for:
• High or Low setpoint with adjustable hysteresis
• Pulse operation (max. rate: 400 pulses/min)
• USP standards
• Time delay: 0 s to 6400 s
Environmental
Ambient operating temperature: -10 °C to 55 °C (14 °F to 131 °F)
•
• Storage temperature: -15 °C to 80 °C (5 °F to 176 °F)
• Relative humidity: 0 to 95%, non-condensing
• Maximum altitude: 2000 m (6562 ft)
• Insulation category: II
• Pollution degree: 2
Standards and Approvals
CE
• EN61326: Immunity & Emissions for Control Equipment
• EN61010: Safety
UL listed
Manufactured under ISO 9001 for Quality, ISO 14001 for
Environmental Management and OHSAS 18001 for Occupational
Health and Safety.
China RoHS (Go to www.gfsignet.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and, (2)
This device must accept any interference received, including
interference that may cause undesired operation.
Page 2
Electrical Connections
Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.
This product must be provided with a means to disconnect all current carrying conductors connected to the main AC line, or,
as part of the building installation.
Wiring Procedure
1. Remove 9 mm to 12 mm (0.35 in. to 0.47 in.) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
Wiring Tips:
• Do not route sensor cable in conduit containing AC power wiring - electrical noise may interfere
with sensor signal.
• Routing sensor cabling in grounded metal conduit may prevent moisture damage, electrical
noise, and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Do not insert two wires into a single terminal. If necessary, splice the wires together before
inserting into the terminal.
RELAY 1
RELAY 2
RELAY 3 RELAY 4
Electrical shock hazard exists!
Never connect live AC lines to the
instrument.
Always connect a ground wire to
the ground terminal when using
AC power.
Caution:
N/C N/O COM N/C N/O COM N/C N/O COM N/C N/O COM
- OPEN COLL 4
- OPEN COLL 3
+ OPEN COLL 4
+ OPEN COLL 3
- LOOP 1
- LOOP 2
+ LOOP 1
- LOOP 3
+ LOOP 2
+ LOOP 3
POWER
L
N
-
+
LISTED
U
L
E171559
¨
OUTPUT OPTION
OPEN
COLL 3,4
RELAY 3,4
SHLD
ISO GND
SGNL 1
TEMP 1
SHLD
ISO GND
Output Option Switch
• In OPEN COLLECTOR position, relays 3 and 4 are open
collector outputs as shown.
• In RELAY 3, 4 position, relays 3 and 4 are dry contact relays
identical to relays 1 and 2.
• The menu references in the 8860 display will not change.
Outputs 3 and 4 will be identifi ed as "Relay" regardless of
switch setting.
LR92369
SGNL 2
TEMP 2
L
100 - 240V
N
50-60Hz, 20VA
-
11 - 24V
0.5 A
+
~
3-8860.610 D
4 to 20 mA loop Outputs
The current loops in the 8860 are passive circuits.
12-24VDC must be provided from an external source.
A single loop is illustrated for clarity.
RANGE defines where the pulsing starts
and where it reaches the maximum rate.
Pulse rate
µS -----
>
0 pulses
100 pulses
The Pulse Rate determines what the
maximum rate will be. The 8860 maximum
capability is 400 pulses per minute.
• The 8860 will accept two independent sensor input signals.
The two sensors may be of different cell constant values.
• Do not route sensor cable in any conduit containing AC power
wiring - electrical noise may interfere with the signal.
• Cable shield MUST be maintained through cable splice.
• Use three conductor shielded cable for cable extensions up
to 30 m (100 ft) max. for measurements below 10 M (above
0.10 μS).
• Maximum cable length for resistivity measurements above
10 Mis 25 ft, and solution temperatures must be between20 °C and 100 °C.
0.010μS
100 MΩ
UPW @ 25°C
0.054μS
18.3 MΩ
Bottled
USP
Rinse
Water
3-2820, 3-2840 ( 0.10 Cell)
3-2819, 3-2839
1μS
1MΩ
10μS
100KΩ
Water
3-2822, 3-2842 (10.0 Cell)
3-2821, 3-2841 (1.0 Cell)
(0.01 Cell)
100μS
10KΩ
3-2823
1,000μS10,000μS
200μS
Cooling Tower
Waste Waters
Brackish Sea Water
(Salinity)
50,000μS
Deionization Regen. Chemicals
Alkali Cleaners, Acids/Bases
(20.0 Cell)
200,000μS
400,000μS
100,000μS
SHLD
ISO GND
TEMP 1
SGNL 1
RED
WHITE
BLACK
SILVER (SHLD)
Relay and Open Collector Functions
The 8860 is equipped with 4 SPDT relays that can be confi gured
for High alarm, Low alarm, Proportional pulse or USP operation.
• Low: Output triggers when process variable is less than
setpoint.
• High: Output triggers when process variable is higher than
setpoint.
• Off: Disables output pulse.
SILVER (SHLD)
BLACK
WHITE
RED
SHLD
ISO GND
SGNL 2
TEMP 2
• Pulse Operation
The output emits a 100 mS pulse at a rate defi ned by the
Source, Pulse Range and Max PlsRate settings, and by the
process condition. The maximum pulse rate is 400 pulses per
minute.
Example: As the process falls below 10 μS the output will
start pulsing in relation to the process value, the max pulse
endpoint and the programmed pulses/min. Pulse rate
will increase as the process approaches the programmed
endpoint.
• During normal operation, the ProcessPro displays the VIEW menu.
• When using the CALIBRATE or OPTIONS menus, the ProcessPro will return to the VIEW
menu if no activity occurs for 10 minutes.
• To select the item you want displayed, press the UP or DOWN arrow keys. The items
will scroll in a continuous loop. Changing the display selection does not interrupt system
operations.
• No keycode is necessary to change display selection.
• Output settings cannot be edited from the VIEW menu.
View Menu
DisplayDescription
C1 1000.00 μS/cm
C2 30.00 μS/cm
Monitor C1 Conductivity and C2 Conductivity simultaneously. This is the
permanent display when Channel 2 is ON.
C1 1000.00 μS/cm
25.0 ºC
Monitor Channel 1 Conductivity and Channel 1 Temperature. This is the
permanent display when Channel 2 is OFF.
The VIEW displays below are temporary. The permanent display will return after 10 minutes
C2 30.00 μS/cm
25.0 ºC
Monitor Channel 2 Conductivity and Channel 2 Temperature. This view is
available only when Channel 2 is ON.
Reject C1 → C2
97.00
Monitor Percent Reject, Difference, or Ratio
(Channel 1 to Channel 2 or Channel 2 to Channel 1)
Loop 1 12.03 mA
Loop 2 5.69 mA
Monitor Loop 1 and Loop 2 current output simultaneously.
Loop 3 13.7 mA
Monitor Loop 3 and status of Relays 3 and 4 (Open Collector 3 and 4).
R3 ON R4 PLS
Last Cal
Monitor date for scheduled maintenance or date of last calibration.
Step 2. The Key Code is UP-UP-UP-DOWN keys in sequence.
• After entering the Key Code, the display will show the fi rst item in the selected menu.
Step 3. Scroll menu with UP or DOWN arrow keys.
Step 4. Press RIGHT ARROW key to select menu item to be edited.
• The fi rst display element will begin fl ashing.
Step 5. Press UP or DOWN keys to edit the fl ashing element.
• RIGHT ARROW key advances the fl ashing element.
Step 6. Press ENTER key to save the new setting and return to Step 3.
Notes on Step 1:
• The View Menu is normally displayed.
• The CALIBRATE and OPTIONS menus require a KEY CODE.
Press &
hold for
access:
VIEW
ENTER
Notes on Step 2:
If no key is pressed for 5 minutes while display is showing "Enter
Key Code", the display will return to the VIEW menu.
2s5s
CALIBRATE
OPTIONS
Notes on Steps 3 and 4:
• Refer to pages 6 and 7 for complete listing of menu items and their use.
• From the Step 3 display, pressing the UP and DOWN keys simultaneously will
return the display to the VIEW menu.
• If no key is pressed for 10 minutes, display will also return to the VIEW menu.
Step 3: Finished Editing?
Press the UP and DOWN keys simultaneously after
saving the last setting to return to normal operation.
First item in CALIBRATE menu
Step 3
Step 4
Notes on Steps 5 and 6:
• All output functions remain active during editing.
• Only the fl ashing element can be edited.
• RIGHT ARROW key advances the fl ashing element in a continuous loop.
• Edited value is effective immediately after pressing ENTER key.
• If no key is pressed for 10 minutes unit will restore the last saved value and return to step 3.
• Step 6 (pressing ENTER key) always returns you to Step 3.
• Repeat steps 3-6 until all editing is completed.
Step 5: Made an Error?
Press the UP and DOWN keys simultaneously
while any element is fl ashing. This will recall the
last saved value of the item being edited and
return you to Step 3.
Select CUSTOM only if you are connecting a certifi ed conductivity sensor. Select
STANDARD for all non-certifi ed sensors.
Chan 1 Cell:
1.0 >
Cell: Custom
1.0000 >
Chan 1 Set:
Temperature >
Chan 1 Set:
Conductivity >
Chan 1 Units:
uS/cm >
Chan 1 TDS:
2.0000 uS/PPM >
Function:
Reject C1 → C2 >
Loop 1 Source:
Chan 1 Cond >
For STANDARD sensors: Select the nominal cell constant: 0.01, 0.1, 1.0, 10.0 or 20.0.
For CUSTOM sensors: Enter the precise cell constant from the certifi cate provided with your
sensor, or from the information label on the sensor.
Adjust the temperature of the system based on an accurate external reference.
This single-point wet calibration procedure requires a test solution of known value.
Enter all zeroes here to restore TEMPERATURE and CONDUCTIVITY to factory calibration.
Select the units of measure: μS/cm, mS/cm, kΩ•cm, MΩ•cm, PPB, PPM
If the Units selection is PPM or PPB, set the ratio of μS to Total Dissolved Solids. The factory
preset value is 2 μS per 1 PPM of TDS. (Always μS/PPM, even if units is PPB).
See page 10 for additional information.
Select a functional relationship between C1 and C 2: • Ratio is (C1:C2) or (C2:C1)
• Percent Reject is 100%(1-C2/C1) or 100%(1-C1/C2) • Difference is (C1-C2) or (C2-C1)
Select the measured value or calculated FUNCTION you want Loop 1 to represent:
Chan 1 Cond, Chan 2 Cond, Chan 1 Temp, Chan 2 Temp, or Function
Loop 1 Range: uS
0.0000 → 100.000 >
Relay 1 Mode:
Low >
Relay 1 Source:
Chan 1 Cond >
Relay 1 Setpnt:
10.0000 uS >
Relay 1 Hys:
0.5000 uS >
Relay 1 Delay:
10.0 secs >
Relay 1 Rng: uS
10.0000 → 40.0000 >
Relay 1 PlsRate:
120 Pulses/Min >
Set the minimum (4 mA =) and maximum (20 mA =) range for Loop 1.
Make sure that the values are consistent with the units of the source.
Select operating mode for Relay 1: OFF, LOW, HIGH, USP or PULSE.
For USP mode: • Relay 1 SOURCE must be Cond 1 or Cond 2
• Temp Comp (Options menu) must be set to None
Select the INPUT SIGNAL (or FUNCTION) monitored by Relay 1:
• Cond 1 • Cond 2 • Temp 1 • Temp 2 • Function
Set Relay 1 activation point. The maximum value acceptable is 999999.
USP setpoints are high alarms, where the setpoint is a percentage below the USP limit.
Relay 1 will be deactivated at setpoint ± Hysteresis (depending on High or Low selection).
When the relay Mode is USP (defi ned as a HIGH alarm), Hysteresis is displayed in μS.
Set up to 6400 seconds delay time for relay response.
Relay 1 will be activated only if the source value exceeds the setpoint for this time period.
If Relay 1 is PULSE mode, set the start and end point of the conductivity range and also set
the maximum pulse rate. (The maximum PULSE rate setting is 400 pulses per minute.)
The combined Relay 1 Range and Pulse rate settings shown here indicate:
"Start pulsing when the conductivity value is 10 μS and increase the pulse rate up to the
maximum of 120 pulses per minute when the conductivity value reaches 40 μS"
Last CAL:
04-20-08 >
• All changes in this menu become effective when saved, except the "Set Cond" and "Set Temp" settings.
• All outputs affected by a change in the "Set Cond" and "Set Temp" settings are frozen until you exit the Calibrate menu.
6
Use this "notepad" to record important dates, such as annual recertifi cation or scheduled
maintenance.
USP (United States Pharmacopoeia) has defi ned a set of conduc-
tivity values (limits) to be used for pharmaceutical water monitoring. This standard requires non-temperature compensated
conductivity measurement be used to warn if the con ductance
approaches the USP limit. The limits vary according to the temperature of the sample.
The 8860 has the USP limits stored in memory. It will automatically determine the proper USP limit based on the measured
temperature.
Using the USP function
In the 8860, USP setpoints are defi ned as a percentage below
the USP limit, so a USP alarm is always a HIGH alarm. The 8860
can be set to warn you if the conductivity approaches within a set
percentage of the USP limit.
The following settings and conditions are required for a USP
relay function:
1. In the CALIBRATE menu:
• RELAY MODE must be set to USP.
• RELAY SOURCE must be Chan 1 or Chan 2 Cond.
• SOURCE UNITS must be set to μS.
2. In the OPTIONS menu:
• The TC Mode of the USP channel must be set to None.
(Service tip: If a relay is constantly on when it should be off,
check the relay settings listed above.)
Example:
• The USP setpoint is 40%.
• The water temperature is 19 ºC, so the USP limit is 1.0 μS.
• The relay will be activated when the conductivity value reaches
0.6 μS, or 40% below the 1.0 USP limit.
• If the water temperature drifts to more than 20 ºC, the 8860 will
automatically adjust the USP limit to 1.1.
• The relay will now be activated when the conductivity value
reaches 40% below 1.1 μS, or 0.66 μS.
Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In
general, select lower contrast if the display is in warmer ambient surroundings.
Select ºC or ºF.
Turn CH 2 OFF if not in use. This will remove all menu functions related to CH 2.
Select 50 Hz or 60 Hz electrical noise suppression, according to the AC power used in your
area. Select the proper setting for all applications, whether AC or DC powered.
Set the method for temperature compensation to NONE, LINEAR or PURE WATER. You
must select NONE for USP systems. Select LINEAR for applications where the water is
less than 5 MΩ (or greater than 0.2 μS). Select PURE WATER for applications where the
water is greater than 5 MΩ (or less than 0.2 μS).
For LINEAR or PURE WATER temperature compensation, select a % per ºC slope.
Maximum slope setting is 9.99 % per ºC. If Temp Comp setting is NONE, this item will not
be displayed.
Averaging:
OFF provides the most instantaneous response to process changes. Select LOW (4 sec)
or HIGH (8 sec) averaging if your process experiences frequent or extreme fl uctuations.
Off >
Chan 1 Decimal:
Set the decimal to the best resolution for your application. The display will automatically
scale up to this restriction. Select *****., ****.*, ***.** **.*** or *.****
*.**** >
Loop 1 Adjust:
4.00 mA >
Loop 1 Adjust:
20.00 mA >
Relay 3 Active:
High >
Test Loop 1:
Adjust the minimum and maximum current output. The display value represents the precise
current output. Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 mA
Use this setting to match the system output to any external device.
These settings repeat for Loop 2 and Loop 3.
Select active HIGH or active LOW operation for relay 3
Recommended: Use active LOW if OUTPUT OPTION switch is set for OPEN
COLLECTOR operation.
Active HIGH: Power is applied to relay coil when process value reaches SETPOINT.
Active LOW: Power is removed from relay coil when process value reaches SETPOINT.
Press UP and DOWN keys to manually change Loop 1output current.
Limits are 3.6 mA to 21.00 mA. Hold UP or DOWN keys to scroll the output value.
>
Test Relay 1:
Press UP and DOWN keys to manually toggle the relay state.
>
Notes:
Channel settings will repeat when Channel 2 is enabled.
Loop settings will repeat for Loop 2 and Loop 3.
Check setting for related values when making changes (for example, if temp is set for ºC with a temp alarm at 25 ºC, and you change
the temp to ºF, be sure to change the alarm setpoint to 77 ºF.)
Relay settings will repeat for Relays 2, 3 and 4 (except "Relay Active" selection; applies to Relay 3 and 4 only.)
The 8860 has been electronically calibrated at the factory.
• Procedure A verifi es the accuracy and linearity of the 8860 by simulating temperature and conductivity values with precision (±0.1%) fi xed resistors.
• Procedure B is a wet calibration. This procedure uses the sensor input and NIST traceable test solutions.
A) Accuracy Verifi cation with Precision Resistors (Electronic Calibration):
1. Simulate the Temperature
The temperature input to the 8860 is a PT-1000 thermistor, where 1000 Ohms () is equal to 0 ºC and a change of 3.84 equals a
• Connect a resistor (1000 to 1096 ) between "Temp" and "Iso. Gnd" terminals.
• Set Temp: Adjust the temperature to exact value based on the measured resistance. (See Editing Procedure, Calibrate menu).
• To verify the 8860 temperature linearity, connect a second resistor value to the terminals.
• If the 8860 does not display the correct value, service is required. ( ± 0.5 ºC)
2. Simulate the Conductivity
You may calculate the exact Resistance needed to simulate a specifi c conductivity value , or you may calculate the exact Conductivity
based on a resistor value:
Resistance = Cell constant e.g. 0.1 Cell = 5,000 or 5 k conductivity (Siemens*) 0.000020 (Siemens*)
Conductivity = Sensor cell e.g. 0.1 Cell = 0.000001 Siemens*
Simulation resistance () 100,000 () or 1μS/cm
-6
(*1 μS = 1 X 10
Siemens or 0.000001 Siemens)
• Connect the conductivity resistor between the "Sgnl 1 " and "Iso Gnd" (or Sgnl 2 and Iso Gnd) terminals.
• Set Cond: Adjust the conductivity value based on the resistor value. (See Editing Procedure and Calibrate menu).
• Verify the linearity of the 8860 by connecting a second Conductivity resistor of a different value.
• If the 8860 does not display the correct value (± 2% of reading), service is required.
B) Wet Calibration with NIST Traceable Solutions:
When using NIST traceable standards, review the temperature information provided with the test solution. Prevent contamination of the
test solution. The sensor must be at the temperature specifi ed on the test solution label.
• Remove the sensor from the system. Rinse the sensor in a small amount of test solution.
• Place the sensor into the test solution.
• Place a reference thermometer into the same solution.
• Allow suffi cient time for the temperature to stabilize.
• Set Temp: Adjust the temperature value based on the reference thermometer. (See Editing Procedure.)
• Set Cond: Adjust the conductivity value based on the test solution value. (See Editing Procedure.)
• Verify the linearity of the 8860 by placing the sensor into a second test solution of a different value.
• If the 8860 does not display the correct value (Temperature ± 0.5 ºC, Conductivity ± 2% of reading), service is required.
Conductivity measurement is highly dependent on temperature.
The basic rule is that higher temperatures result in greater
conductance (less resistance).
Temperature effects are expressed as the percentage of
conductivity change (in μS) per °C. The conductivity value is
generally referenced to 25 ºC. The 8860 has three temperature
compensation options:
None
USP standards for pharmaceutical waters require that the
measurement be made without temperature compensation. USP
limits are discussed on page 7.
Pure Water (Standard Compensation)
This selection is used for measurements of very clean water,
less than 0.2 μS. Temperature effects are not linear in this
range, so the temperature coeffi cient is not easily determined.
This selection is recommended for all Resistivity applications
measuring from 5 M to 18 M. This selection conforms to
ASTM standard D1125 and D5391.
Linear
This selection allows you to calculate a custom temperature
compensation value for Conductivity measurements in the range
of 0.2 μS and greater (Resistivity applications measuring less
than 5 M). The procedure is outlined in the section on the right.
Calculating a Linear Temperature Coeffi cient
1. Set TC Mode to NONE (see OPTIONS menu, page 8).
2. Heat a sample solution close to the maximum process
temperature. Place sensor in the sample solution and
allow several minutes for stabilization. Record the 8860
temperature and conductivity values in the spaces provided:
Displayed temperature: T1 = _______ °C
Displayed conductivity: C1 = _______ μS
3. Cool the sample solution close to the minimum process
temperature. Place sensor in the sample solution allowing
several minutes for stabilization. Record displayed
temperature and conductivity values in the spaces provided:
Displayed temperature: T2 = _______ °C
Displayed conductivity: C2 = _______ μS
(A 10% change in conductivity between steps 2 and 3 is
recommended.)
4. Substitute recorded readings (steps 2 and 3) into the following
formula:
TC Slope = 100 x (C1 - C2)
(C2 x (T1 - 25)) - (C1 x (T2 - 25))
TDS Factor
Some industries need to display a conductivity value as Total
Dissolved Solids (TDS), measured in units of parts per million
(PPM) or parts per billion (PPB).
• 1 PPM is equivalent to 1 mg per liter.
• 1 PPB is equivalent to 1 μg per liter.
• The 8860 calculates PPM or PPB by dividing the μS value
by a TDS Factor that you defi ne. TDS factors can vary
widely, ranging from 1.50 μS to 2.50 μS per PPM. Methods
for establishing a TDS factor are beyond the scope of this
manual.
• The 8860 will accept TDS factor values from 0.01 μS to
99999.9 μS per PPM.
(factory preset = 2.00 μS per PPM)
NOTE: The 8860 TDS factor must be set in PPM.
Example: A sample solution has a conductivity of 205 μS @
48 °C. After cooling the solution, the conductivity was measured
at 150 μS @ 23 °C. (C1 = 205, T1 = 48, C2 = 150, T2 = 23)
The TC is calculated as follows:
TC Slope = 100 x (205 - 150) =5500 = 1.42%/°C
(150 x (48 - 25)) - (205 x (23 - 25)) 3860
Display is over range. This may be a
normal condition if your process operates
at/near the limits of the sensor range.
The menu item being set is a percentage
value and must be less than 100.
The menu item being set cannot be zero
or a negative value.
The Pulse Rate for Relay and Open
Collector outputs cannot be greater than
400 pulses per minute.
A value of "0" is being set into
".. Set: Conductivity" menu item.
The calibration offset entered is
beyond the allowable tolerances of the
instrument.
Check sensor for correct range.
Check Decimal setting in OPTIONS menu.
Check Calibrate menu settings for incompatible
SOURCE and RANGE values.
Select a value from 0 to 100.
Select a value greater than zero.
Select a pulse rate less than 400
This will remove any user calibration from the
"..Set Conductivity" and "..Set Temperature"
items in the Calibrate menu.
Check calibration procedure for accuracy.
Check sensor for proper operation.
Check any cable extensions for poor splices or
termination.
Technical Note
If a Current Loop is locked at 3.6 mA, the problem is related to the temperature circuit:
This occurs only if the 8860 detects a resistance from the temperature sensor that is less than 250 Ω or greater than 2800 Ω.
• Check the sensor wiring for open/short or poor connections on white (TEMP IN) and black (ISO GND)wires.
• The PT1000 temperature device in the sensor is defective.
3-8860 159 000 677 Conductivity/Resistivity Controller, DC
3-8860-AC 159 000 678 Conductivity/Resistivity Controller, AC
Accessories
Mfr. Part No Code Description
3-8050.395 159 000 186 NEMA 4X Rear Cover
3-9000.392 159 000 368 Liquid-tight Connector Kit for rear cover (includes 3 connectors)
3-8050.392 159 000 640 Model 200 Retrofi t Adapter
3-5000.399 198 840 224 5x5 inch Adapter Plate for Signet retrofi t
3-0000.596 159 000 641 Heavy Duty Wall Mount Bracket
3-5000.598 198 840 225 Surface Mount Bracket
3-8050.396 159 000 617 RC Filter Kit (for relay use)
7300-7524 159 000 687 Power Supply, 7.5W 24 V
7300-1524 159 000 688 Power Supply, 15W 24 V
7300-3024 159 000 689 Power Supply, 30W 24 V
7300-5024 159 000 690 Power Supply, 50W 24 V
7300-1024 159 000 691 Power Supply, 100W 24 V
Conductivity/Resistivity Electrodes
Mfr. Part No. Code Description
3-2818-1 159 001 718 Conductivity/Resistivity (CR) Cell, 0.01, SS
3-2819-1 198 844 010 CR Cell, 0.01, SS
3-2820-1 198 844 000 Conductivity Cell, 0.1, SS
3-2821-1 198 844 001 Conductivity Cell, 1.0, SS
3-2822-1 198 844 002 Conductivity Cell, 10, SS
3-2823-1 198 844 003 Conductivity Cell, 20, SS
3-2819-S1 159 000 085 CR Sanitary, 0.01, SS, 1 to 1 1/2 in.
3-2819-S1C 159 000 087 CR Sanitary, 0.01, SS, 1 to 1 1/2 in., Certifi ed
3-2819-S2 159 000 086 CR Sanitary, 0.01, SS, 2 in.
3-2819-S2C 159 000 088 CR Sanitary, 0.01, SS, 2 in., Certifi ed
3-2819-T1 159 000 081 CR Sanitary, 0.01, Titanium, 1 to 1 1/2 in.
3-2819-T1C 159 000 083 CR Sanitary, 0.01, Titanium, 1 to 1 1/2 in., Certifi ed
3-2819-T2 159 000 082 CR Sanitary, 0.01, Titanium, 2 in.
3-2819-T2C 159 000 084 CR Sanitary, 0.01, Titanium, 2 in., Certifi ed
3-2820-S1 159 000 089 CR Sanitary, 0.1, SS, 1 to 1 1/2 in.
3-2820-S1C 159 000 091 CR Sanitary, 0.1, SS, 1 to 1 1/2 in., Certifi ed
3-2820-S2 159 000 090 CR Sanitary, 0.1, SS, 2 in.
3-2820-S2C 159 000 092 CR Sanitary, 0.1, SS, 2 in., Certifi ed
3-2820-T1 159 000 624 CR Sanitary, 0.1, Titanium, 1 to 1 1/2 in.
3-2820-T2 159 000 625 CR Sanitary, 0.1, Titanium, 2 in.
3-2821-S1 159 000 093 CR Sanitary, 1.0, SS, 1 to 1 1/2 in.
3-2821-S1C 159 000 095 CR Sanitary, 1.0, SS, 1 to 1 1/2 in., Certifi ed
3-2821-S2 159 000 094 CR Sanitary, 1.0, SS, 2 in.
3-2821-S2C 159 000 096 CR Sanitary, 1.0, SS, 2 in., Certifi ed
3-2821-T1 159 000 626 CR Sanitary, 1.0, Titanium, 1 to 1 1/2 in.
3-2821-T2 159 000 627 CR Sanitary, 1.0, Titanium, 2 in.
3-2839-1 159 000 921 Conductivity Cell, 0.01, 15 ft cable, NPT
3-2839-1D 159 000 923 Conductivity Cell, 0.01, 15 ft cable, ISO
3-2840-1 159 000 786 Conductivity Cell, 0.1, 15 ft cable, NPT
3-2840-1D 159 000 788 Conductivity Cell, 0.1, 15 ft cable, ISO
3-2841-1 159 000 790 Conductivity Cell, 1.0, 15 ft cable, NPT
3-2841-1D 159 000 792 Conductivity Cell, 1.0, 15 ft cable, ISO
3-2842-1 159 000 794 Conductivity Cell, 10.0, 15 ft cable, NPT
3-2842-1D 159 000 796 Conductivity Cell, 10.0, 15 ft cable, ISO
NOTE: Alternate wetted materials and lengths are available through special order.
Cable length extensions to 30 m (100 ft) are available through special order.
Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • or call (in the U.S.): (800) 854-4090
For the most up-to-date information, please refer to our website at www.gfsignet.com