• Operating temperature: -10 °C to 70 °C (14 °F to 158 °F)
• Storage temperature: -15 °C to 80 °C (5 °F to 176 °F)
• Relative humidity: 0 to 95%, non-condensing
• Maximum altitude: 2000 m (6562 ft)
• Insulation category: II
• Pollution degree: 2
• Rating: NEMA 4X/IP65 front
Standards and Approvals:
• CE, UL listed
• Immunity: EN50082-2
• Emissions: EN55011 Class B
• Manufactured under ISO 9001 for Quality, ISO 14001
for Environmental Management and OHSAS 18001 for
Occupational Health and Safety.
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules. Operation
is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received, including
interference that may cause undesired operation.
China RoHS (Go to www.gfsignet.com for details)
2. Installation
• ProcessPro transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary
hardware to install the transmitter. This manual includes complete panel mounting instructions.
• Field mounting requires one of two separate mounting kits. The 3-8052 integral kit joins the sensor and instrument together into a
single package. The 3-8050 Universal kit enables the transmitter to be installed virtually anywhere.
• Detailed instructions for integral mounting or other fi eld installation options are included with the 3-8052 Integral kit or the 3-8050
Universal kit.
2.1 Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel
cutout, an adhesive template is provided as an installation guide. Recommended clearance on all sides
between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT
RELEASE.
Press quick-clips outward and remove.
gasket
latch
System Pwr
Loop -
2
System Pwr
Loop +
1
Panel Mount
Installation Detail
panel
Output -
4
Output +
3
terminals
Sensr Gnd
(SHIELD)
7
Sensr IN
(RED)
6
Sensr V+
(BLACK)
5
quick-clips
mounting
bracket
Page 2
10
9
8
7
6
5
4
3
2
1
16
15
14
13
12
11
Output 2-
Output 2+
Output 1-
Output 1+
Loop 2-
Loop 2+
System Pwr
Loop -
System Pwr
Loop +
AUX
Power -
AUX
Power +
Snsr 2 Gnd
(WHITE)
Snsr 2 IN
(RED)
Snsr 2 V+
(BLACK)
Snsr 1 Gnd
(WHITE)
Snsr 1 IN
(RED)
Snsr 1 V+
(BLACK)
3. Electrical Connections
1
15
S
D
2
_
+
Internal open-collector
output circuit
Outputs
Isolation
Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.
Wiring Procedure
1. Remove 13 mm - 16 mm (0.5 in. - 0.625 in.) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.
Wiring Tips:
• Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
• If your system uses a single sensor, it must be located within 122 m (400 ft.) of the transmitter.
• If your system uses two sensors, the total length of cable connected to the transmitter is limited to 122 m (400 ft.)
• The 3 conductors from a dual-sensor system can be tied together and then a single set of wires continued on to the transmitter.
• For best performance, ground the sensor SHIELD wire to a local earth ground at a point near the sensor.
(Experiment by connecting the sensor shield wire to different local ground points to identify best signal quality.)
Terminals 3-4: Primary Loop Power
12-24 VDC ±10% system power and current loop connections.
SINGLE SENSOR system MUST use terminals 3-4.
Terminals 5-6: 12-24 VDC ± 10%
Power and current loop connections for second sensor.
• Maximum loop impedance:
50 Ω max. at 12 V
325 Ω max. at 18 V
600 Ω max. at 24V
Terminals 11-16: Dual digital sensor inputs
See next page for more information.
2
1
Terminals 1-2: Auxiliary power is used
only if two sensors are connected to
the transmitter.
2
8350-3 Temperature Transmitter
Terminals 7-10: Open-Collector Outputs
Two transistor outputs, programmable (see CALIBRATE menu) as:
• High or Low setpoint with adjustable hysteresis
• Proportional pulse output
• May be disabled (Off) if not used.
Page 3
16
15
14
13
12
11
Signet 2350
Temperature
Sensors
Terminals
White (Snsr 2 Gnd)
Red (Snsr 2 IN)
Black (Snsr 2 V+)
White (Snsr 1 Gnd)
Red (Snsr 1 IN)
Black (Snsr 1 V+)
3.1 Sensor Input Connections
+
+
-
+
-
+
+
+
-
+
+
+
+
+
-
+
6
5
4
3
2
1
6
5
4
3
2
1
6
5
4
3
2
1
6
5
4
3
2
1
6
5
4
3
2
1
*
Example: Two transmitters connected to PLC/Recorder with separate power supply
Loop 2 Loop 2 +
System Power Loop 1 System Power Loop 1 +
AUX Power AUX Power +
Loop 2 -
Loop 2 +
System Power Loop 1 -
System Power Loop 1 +
AUX Power -
AUX Power +
PLC or Recorder
Channel 3
4-20 mA in
Channel 4
4-20 mA in
Transmitter 1
Terminals
Transmitter 2
Terminals
Channel 1
4-20 mA in
Channel 2
4-20 mA in
Power
Supply
12-24 VDC
* *
*
* *
*
*
*
Stand-alone application, no current loop used
Transmitter
Terminals
Power
Supply
12-24 VDC
Connection to a PLC with built-in power supply
Internal
PLC
Connection
Connection to a PLC/Recorder, separate supply
PLC or Recorder
Loop 2 -
Loop 2 +
System Power Loop 1 -
System Power Loop 1 +
AUX Power -
AUX Power +
Transmitter
Terminals
Loop 2 -
Loop 2 +
System Power Loop 1 -
System Power Loop 1 +
AUX Power -
AUX Power +
Transmitter
Terminals
Loop 2 -
Loop 2 +
System Power Loop 1 -
System Power Loop 1 +
AUX Power -
AUX Power +
Power
Supply
12-24 VDC
PLC Terminals
Channel 1
4-20 mA
Power
Supply
12-24 VDC
Channel 2
4-20 mA
Channel 2
4-20 mA
Channel 1
4-20 mA
NC
NC
AUX power required only where two sensors are used
AUX power required only where two sensors are used
AUX power required only where two sensors are used
AUX power required only where two sensors are used
A
B
A
B
C
7HPSÛ&
7HPSÛ&
ENTER
Signet Temperature
Transmitter
7HPSÛ&
7HPSÛ&
ENTER
Signet Temperature
Transmitter
Transmitter
A + B 122 m (400 ft.)
J-box
Transmitter
A + B + C 122 m (400 ft.)
7HPSÛ&
7HPSÛ&
ENTER
Signet Temperature
Transmitter
Transmitter
122 m (400 ft.)
3.2 System Power/Loop Connections
8350-3 Temperature Transmitter
3
Page 4
Temp1: 25.0 ˚C
Temp2: 25.0 ˚C
ENTER
Signet Temperature
Transmitter
3.3 Open Collector Output
Hysteresis
Time
Low Setpoint
Process
Relay energized
Relay de-energized
Hysteresis
Time
High Setpoint
Process
10
5
Open Collector Output Rate:
0 to 100 Pulses/min.
endpoint
Pulse rate
0 pulses
100 pulses
Starting point
The Open Collector output can be used as a switch that responds
when the process value moves above or below a setpoint, or it
can be used to generate a pulsing signal with a rate proportional
to the process value.
Delta Temperature (Differential Mode)
If two sensors are connected to the transmitter, the DELTA
(Sensor 1 - Sensor 2) can be assigned to output. To enable
differential mode, the selected output "Source" selection must be
set to "DT".
• Low:
Output energizes when
process variable is less
than the setpoint. The
output de-energizes
when the process
variable moves above
the setpoint plus the
hysteresis value.
• High:
Output energizes
when process variable
is greater than the
setpoint. The output de-energizes when the process variable
moves below the setpoint plus the hysteresis value.
• Proportional Pulsing
The Open Collector output will generate a 100 mS pulse at the rate defi ned by settings in the
CALIBRATE menu (see page 6)
In the example:
• The output will be 0 pulses/min. at temperatures less than 5º.
• The output will be 50 pulses/min. at 7.5º.
• The output will be 100 pulses/min. at temperatures above 10º.
Sensor 1 must be equal to ID 1. If it is not, go to the Calibrate
menu and switch the sensor ID.
In the example below, Sensor 1 is measuring 80º while sensor 2
is measuring 20º. The delta is 60º ( Sensor 1 - Sensor 2 = 60º)
To track the delta temperature using a 4-20 mA output:
Loop1 Source: DT
Loop1 Range: 0-100
Loop1: = 13.60 mA
Sensor 1 (80)
Sensor 2 (20)
To set a delta temperature
setpoint using an Open
Collector output:
Output1 Source: DT
Output1 Mode: Low
Output1 Setpoint: 70
Output1 = ON
ENTER
The starting point, endpoint and maximum pulse rate are select able in the menu.
4. Menu Functions
• During normal operation, ProcessPro displays the VIEW menu.
• When editing CALIBRATE or OPTIONS menus, ProcessPro returns to the VIEW menu if no activity occurs for
10 minutes.
• To select a VIEW display, press the ▲ or ▼ keys. The selections will scroll in a continuous loop.
• Changing the display selection does not interrupt system operations.
• No key code is necessary to change display selection.
• Output settings cannot be edited from the VIEW menu.
View Menu
DisplayDescription
Monitor Temperature 1 and Temperature 2 simultaneously.
This is the permanent view display.
Monitor the delta temperature (Channel 1 rate - channel 2 rate = Delta
Temperature). This is a permanent display.
Monitor the 4 to 20 mA output for Loop 1 and 2.
Monitor date for scheduled maintenance or date of last calibration.
(See description in Calibrate Menu.)
4
Temp1: 28.7ºC
Temp2: 19.5ºC
DeltaTemperature:
9.2ºC
The displays below are temporary. After 10 minutes the display returns to the permanent display.
Loop 1 Output:
12.00 mA
Loop 2 Output:
7.65 mA
Last CAL:
02-10-09
8350-3 Temperature Transmitter
Page 5
Output1 Setpnt:
2
0.00 °C
Output1 Setpnt:
1
0.00 °C
ENTER
Output1 Setpnt:
1
0
.00
°
C
Output1 Setpnt:
1
9
.00 °C
Output1 Setpnt:
Saving
Output1 Setpnt:
19.00 °C >
20.00 °C >
Output1 Setpnt:
>
Switch ID:
ProcessPro Editing Procedure:
CALIBRATE: ----
Enter Key Code
CALIBRATE: *---
Enter Key Code
CALIBRATE: **--
Enter Key Code
CALIBRATE: ***-
Enter Key Code
>
Switch ID:
Step 1. Press and hold ENTER key:
• 2 seconds to select the CALIBRATE menu
• 5 seconds to select the OPTIONS menu.
Step 2. The Key Code is ▲-▲-▲-▼ keys in sequence.
• After entering the Key Code, the display will show the fi rst item in the selected menu.
Step 3. Scroll menu with ▲ or ▼ arrow keys.
Step 4. Press ► key to select menu item to be edited.
• The fi rst display element will begin fl ashing.
Step 5. Press ▲ or ▼ keys to edit the fl ashing element.
• ► key advances the fl ashing element.
Step 6. Press ENTER key to save the new setting and return to Step 3.
Notes on Step 1:
• The View Menu is normally displayed.
• The CALIBRATE and OPTIONS menus require a KEY CODE.
Press &
hold for
access:
VIEW
ENTER
Notes on Step 2:
If no key is pressed for 5 minutes while display is showing "Enter
Key Code", the display will return to the VIEW menu.
2s5s
CALIBRATE
OPTIONS
Notes on Steps 3 and 4:
• Refer to pages 6 and 7 for complete listing of menu items and their use.
• From the Step 3 display, pressing the ▲ and ▼ keys simultaneously will return
the display to the VIEW menu.
• If no key is pressed for 10 minutes, display will also return to the VIEW menu.
Step 3: Finished Editing?
Press the ▲ and ▼ keys simultaneously after saving
the last setting to return to normal operation.
First item in CALIBRATE menu
Step 3
Step 4
Notes on Steps 5 and 6:
• All output functions remain active during editing.
• Only the fl ashing element can be edited.
• ► key advances the fl ashing element in a continuous loop.
• Edited value is effective immediately after pressing ENTER key.
• If no key is pressed for 10 minutes unit will restore the last saved value and return to step 3.
• Step 6 (pressing ENTER key) always returns you to Step 3.
• Repeat steps 3-6 until all editing is completed.
8350-3 Temperature Transmitter
Step 5: Made an Error?
Press the ▲ and ▼ keys simultaneously while
any element is fl ashing. This will recall the last
saved value of the item being edited and return
you to Step 3.
Step 5
Step 6
5
Page 6
Calibrate Menu
Display
(Factory settings shown)
Switch ID
>
Temp Units:
°C >
Set:
Temp1
Set:
Temp2
Loop 1 Source
Temp 1
Loop 1 Range: °C
0.0 → 100.0 >
Output1 Source
Temp1 >
Description
Reverses the identifi cation of Sensor inputs: Temp1 becomes identifi ed as Temp2, temp2
becomes identifi ed as Temp1.
Use this function to make corrections if the sensors are in the wrong order for a DELTA
TEMPERATURE measurement.
Select Temperature units: °C or °F.
Provides a maximum 20 °C offset to match 8350 to external reference.
Enter “-999” to restore the original Factory calibration value.
Changes to this setting will become effective when the display exits the Calibrate menu.
Select the INPUT SIGNAL represented by this 4-20 mA output:
Temp 1, Temp 2, or Delta Temperature (Temp 1- Temp 2.)
Check the 2350 Sensor instructions for the range capability of your sensor.
Be sure to modify this setting if you change the Temperature Units.
Select the INPUT SIGNAL represented by this Open Collector Output:
Temp 1, Temp 2, or Delta Temperature (Temp 1- Temp 2.)
Output1 Mode:
Low >
Output1 Setpnt:
25.0 °C >
Output1 Hys:
°C >
Output 1 Range:
25.0-45.0 >
Output 1 PlsRate:
120 Pulses/Min >
Last CAL:
2-10-09
Select the desired mode of operation for the Open Collector output. Options available are Off,
Low, High or Pulse. The signal may be disabled (Off) if not used.
In Low or High Mode, the Open Collector output will be activated when the temperature
reaches this value. Be sure to modify this setting if you change the Temperature Units.
The Open Collector output will be de-energized at Setpoint ± Hysteresis, depending on High
or Low Setpoint selection.
If the output is in PULSE mode, set the start and end point of the range and also set the
maximum pulse rate. (The maximum PULSE rate setting is 400 pulses per minute.)
The combined Output Range and Pulse rate settings shown here indicate:
"Start pulsing when the process value is 25° and increase the pulse rate up to the maximum
of 120 pulses per minute when the process value reaches 45°"
Use this “note pad” to record important dates, such as annual recertifi cation or scheduled
maintenance.
Menu items will repeat for Loop 2 and Output 2.
6
8350-3 Temperature Transmitter
Page 7
Options Menu
Display
(Factory settings shown)
Contrast:
3 >
Averaging
Off >
Output1 Active
Low >
Output2 Active
Low >
Loop1 Adjust
4.00 mA >
Loop1 Adjust
20.00 mA >
Test Loop 1:
>
Test Loop 2:
>
Description
Adjust the LCD contrast for best viewing. A setting of 1 is lowest contrast, 5 is highest. Select
lower contrast if the display is in warmer ambient surroundings.
OFF provides the quickest output response to changes in temperature. Longer averaging
period produces more stable display and output response. Select LOW, HIGH or OFF.
Active HIGH: This setting is used to turn a device (pump, valve) ON at the setpoint.
Active LOW: This setting is used to turn a device OFF at the setpoint.
Adjust the minimum and maximum current output. Use this setting to match the system output
to any external device. The display value represents the precise current output.
Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 mA
Press ▲ or ▼ keys to manually order any output current value from 3.6 mA to 21.00 mA to test
current loop output.
Test Output 1:
>
Test Output 2:
Press ▲ or ▼ keys to manually toggle the state of open collector output.
The value entered in Set Temperature
fi eld is greater than 20 °C (36 °F)
deviation from sensor input. (The 8350
allows a maximum of 20 °C offset.).
Value in SET TEMPERATURE fi eld
is -999.
Hysteresis value too large.
Defective transmitter.
Value entered in Output PlsRate fi eld is
greater than 400.
Memory fault occurred.
• Correct sensor wiring.
• Recycle power with all sensors connected.
• Replace defective sensor.
• Confi rm calibration values.
• Enter -999 in Set Temperature fi eld.
• Replace defective sensor.
Entering "-999" in this fi eld will remove all user
calibration input and restore the factory values.
Change the hysteresis value
Replace transmitter
Select a maximum pulse rate of 400 pulses per
minute or less.
• Press any key to reload factory presets.
• Reprogram all setpoints.
• If this message appears again, replace the
8350.
Ordering Information
Mfr. Part No. Code Description
3-8350-1 159 000 192 Temperature transmitter Field mount
3-8350-1P 159 000 193 Temperature transmitter Panel mount
3-8350-2 159 000 194 Temperature transmitter Field mount with relays
3-8350-2P 159 000 195 Temperature transmitter Panel mount with relays
3-8350-3 159 000 196 Temperature transmitter Field mount with dual input/output
3-8350-3P 159 000 197 Temperature transmitter Panel mount with dual input/output
Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • or call (in the U.S.): (800) 854-4090
For the most up-to-date information, please refer to our website at www.gfsignet.com