Signet 4150 Stilling Chamber with automatic air release valve
*3-4150.096*
3-4150.096 Rev C 09/13 English
Description
The Signet 4150 Stilling Chamber is used to remove air bubbles
from the fl ow stream before it passes through the measuring
cell of the Turbidimeter. Air bubbles will cause unstable or
incorrect readings if not eliminated from the fl ow stream.
The stilling chamber is a cylinder approximately 5 cm x 66 cm
(2 in. x 26 in.). Water enters the chamber via the outlet fi tting
near the top of the cylinder (see fi gure 1). As it passes through
the chamber, the fl ow velocity is reduced by the larger volume
of the chamber. This allows time for any entrained air to rise to
the top of the cylinder, where the automatic air vent releases it
into the environment.
For proper operation the stilling chamber must be mounted
lower than the sensor. For optimum results there should be at
least 76 cm (30 in.) as measured from midpoint of the air vent
to the inlet at the sensor.
The air vent has a maximum working pressure of 6 bar (90 psi)
and maximum temperature range to 106 °C (225 °F).
Inlet with
aerated water
Outlet with
clear water
MINIMUM 76 cm (30 in.)
66 cm
(26 in.)
24 cm
(9½ in.)
Pressure
Regulator
DESSICANT COMPARTMENT
OPEN ONLY FOR MAINTENANCE
3-4150 Turbidimeter
5/16 in. O.D. x 3/16 in. I.D. Tubing
Operating Instructions
Braukman Automatic Air Release Valve
1. Before fi lling the chamber, make sure the red vent cap is securely tight
and that the release valve is in the closed position (see fi gure 2) by
turning valve body clockwise. HAND-TIGHTEN ONLY to avoid damaging
the valve.
2. With the valve in the closed position, fl ush the system as required to
remove dirt, debris and contaminates.
3. Fill the system with the appropriate solution.
4. Start operation by turning the air release valve body counter-clockwise to
the open position (see fi gure 2).
5. For proper operation, make sure that the red vent cap is tightened all the
way to the stop position.
HAND-TIGHTEN ONLY.
Fig 1
CLOSED
OPENED
Figure 2 Shown in
Open Position
Cleaning Instructions
1. Turn the valve body clockwise to the closed position, isolating the
valve from the system. See fi gure 2.
2. Remove the fl oat assembly by unscrewing the top of the body
and lifting the black cover assembly up. See fi gure 3.
3. Carefully clean the seat area of any dirt or debris.
4. Carefully clean any dirt or debris from inside the vent chamber.
5. Reinstall the clean fl oat assembly, making sure that the O-ring is not
damaged and is seated properly. See fi gure 3.
6. Reinstall the black cover assembly on the air release valve body.
HAND-TIGHTEN ONLY.
7. Operate by turning the air release valve body counter-clockwise to the
open position.
HAND-TIGHTEN ONLY. See fi gure 3.
8. For proper operation, make sure that the red vent cap is tightened all
the way to the stop. HAND-TIGHTEN ONLY.
Ordering Information
Part Number Code Description
4150-0003 159 001 587 Stilling/Bubble Chamber
BLACKCOV ER
ASSEMBLY
O-RING
SEAT AREAMAKE SURE
AREA IS FREE
OF DIRTAN D
DEBRIS
Figure 3
RED VENT CAP
WITH LEAKAGE
GUARD.
VENTLEVER
SPRING
FLOAT
CONSTRUCT ED
FOR EASY IN
LINE SERVICE .
SHUTOFFIS
BUILT-IN NO
DRAINDOWN
REQUIRED.
3-4150-1 159 001 596 Turbidimeter, White Light, 0 to 1000 NTU/FNU
3-4150-2 159 001 597 Turbidimeter, Infrared, 0 to 1000 NTU/FNU
3-4150-3 159 001 598 Turbidimeter, White Light, 0 to 100 NTU /FNU With AutoClean
3-4150-4 159 001 599 Turbidimeter, Infrared, 0 to 100 NTU/FNU With AutoClean
3-4150-5 159 001 600 Turbidimeter, White Light, 0 to 1000 NTU/FNU With AutoClean
3-4150-6 159 001 601 Turbidimeter, Infrared, 0 to 1000 NTU/FNU With AutoClean
4150-0001 159 001 593 Pressure Regulator
4150-0004 159 001 589 Replacement cuvette with ultrasonic transducer
4150-0005 159 001 595
Tubing Kit: 1-shutoff clamp, 1-backpressure valve,
2-connecting tubing with fi ttings, 1-drain vent
4150-0007 159 001 602 Replacement cuvette set (3 glass cuvettes)
3-4150.380 159 001 588 Replacement Desiccant
3-4150.381 159 001 613 **Replacement Desiccant cap with gasket (Special Order Only)
3-4150.382 159 001 650 Turbidity Lamp Replacement kit, White
3-4150.386 159 001 652 O-ring kit for cuvette
3-4150-24V 159 001 723 **24 Volt Power Supply (Special Order Only)
3822-4000 159 001 592 *Formazin Stock Solution, 4000 NTU/FNU, 500 ml
3822-4001 159 001 585 *Calibration Kit, Turbidity, 100, 10 & 0.02 NTU/FNU
3822-4002 159 001 591 *Formazin Stock Kit
3822-4003 159 001 586 *Calibration Kit, Turbidity, 1000, 10 & 0.02 NTU/FNU
* Material Safety Data Sheets (MSDS) are available online at www.gfsignet.com
** signet-specialproduct@georgfi scher.com.
Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090
For the most up-to-date information, please refer to our website at www.gfsignet.com
3-4150.096 Rev C 09/13 English © Georg Fischer Signet LLC 2013