GF Signet 2750-7 User Manual

Signet 2750-7 pH Electronics
*3-2750-7.090*
3-2750-7.090 Rev C 05/11 English
CAUTION!
• Remove electrical power from the device or transmitter before wiring input or output connections.
• Follow instructions carefully to avoid personal injury or damage to the electronics.
1. Description
The Signet 2750-7 pH Electronics amplify the output from the Signet 2724-00 pH Electrode and output a Digital (S3L) signal to the Signet 8630 Chlorine Transmitter. Features:
Signet's patented DryLoc® connector provides quick assembly and a secure connection. Gold-plated contacts and an O-ring seal ensure a waterproof and reliable interconnect to the sensor.
Separate drive electronics from sensor make for easy sensor replacement without running new cable.
DryLoc connectivity allows easy sensor removal for calibration.
2. Installation
Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system.
English
Flow Tee Installation 4630 Flow Cell Installation
Signet 3-2727-00 Flat pH Electrode
The bottom yellow nut is not used when using the 2750-7 pH electronics with the Signet 4630 fl ow cell.
Signet 4630 fl ow cell
Do Not Use Lubricant or Sealing Tape on Threads. Do Not Overtighten.
CHEMICAL COMPATIBILITY WARNING
The retaining nuts of pH and ORP sensors are not designed for prolonged contact with aggressive substances. Strong acids, caustic substances and solvents or their vapor may lead to failure of the retaining nut, ejection of the sensor and loss of the process fl uid with possibly serious consequences, such as damage to equipment and serious personal injury. Retaining nuts that may have been in contact with such substances, e.g. due to leakage or spilling, must be replaced.
3. Wiring Wiring to the Signet 8630 Chlorine Transmitter
Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into terminals.
White
Red
Black
Shield
Signal Ground
Digital (S3L) data
+5 VDC
18 17 16
For calibration and confi guration please refer to the 3-8630 Signet Chlorine Transmitter manual.
4. Specifi cations
General
Compatible Electrode:
Signet 2724-00 Flat pH electrode
Compatible Instrument:
Signet 3-8630-3P Chlorine Transmitter
Mounting:
DryLoc connection
Materials:
®*
Valox
(PBT)
Cable:
4.6 m (15 ft) 3 conductor shielded, 22 AWG
Shipping Weight:
0.64 kg (1.41 lb)
Performance
Electronics Accuracy:
± 0.03 pH @ 25 °C
Resolution:
0.02 pH
Response Time (includes electrode response):
< 6 s for 95% of change
Operational Range:
0.00 to 14.00 pH
Electrical
Input Specifi cations:
Input Impedance: > 10 Temperature Drift: ± 0.002 pH per ºC Input Resolution: 0.02 pH, 0.3 ºC
Output Specifi cations:
Digital (S
3
L): Serial ASCII, TTL level 9600 bps
Maximum Cable Length: 30 m (100 ft)
Environmental
Storage Temperature:
–20 °C to 85 °C (–4 °F to 185 °F)
Operating Temperature (Electronics Only):
0 °C to 85 °C (32 °F to 185 °F)
Relative Humidity:
0 to 95%, non-condensing
Enclosure Requirements:
NEMA 4X/IP65 with electrode connected
Standards and Approvals
CE Manufactured under ISO 9001 and ISO 14001
China RoHS (Go to www.gfsignet.com for details)
11
Ω
* Valox® is a registered trademark of SABIC Innovative Plastics
96 mm (3.78 in.)
23 mm (0.91 in.)
61.0 mm (2.40 in.)
Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090 For the most up-to-date information, please refer to our website at www.gfsignet.com
3-2750-7.090 Rev C 05/11 English © Georg Fischer Signet LLC 2011