Signet 2650 DryLoc® Amperometric Electronics
English
3-2650.090
3-2650.090 Rev. F 05/11 English
CAUTION!
• Remove power before wiring input or output connections.
• Follow instructions carefully to avoid personal injury or damage
to the electronics.
1. Description
The Signet 2650 Amperometric Electronics provide the polarization voltage and signal conditioning required by all Signet
Amperometric Sensors. Sophisticated circuitry makes system setup and verifi cation easy.
Features:
• Passes data stored in sensor memory, such as factory calibration data, service time, range and more.
• Signet's patented DryLoc® connector provides quick assembly and a secure connection. Gold-plated contacts and an O-ring seal
ensure a waterproof and reliable interconnect to the sensor.
• Separate drive electronics from sensor make for easy sensor replacement without running new cable.
• DryLoc connectivity allows easy sensor removal for calibration.
2. Installation - Tee Fitting
4630 Flow Cell Installation
Insert sensor fully into fi tting. Align
1
tab on sensor with notch on fi tting.
Lubricate O-rings with a viscous,
non-petroleum based lubricant
(grease) compatible with the
3
2
system.
Install the yellow nut on to the fi tting
2
and hand tighten. Do Not Use Tools.
The bottom yellow nut is
not used when using the 2650
electronics with the Signet 4630
fl ow cell.
Signet 4630
fl ow cell
1
Align the 2650 Electronics with
3
the pins on the sensor. Engage
it by gently turning the lock ring
¼-turn counter-clockwise.
Do Not Use Lubricant or Sealing Tape on Threads.
Do Not Overtighten.
CHEMICAL COMPATIBILITY WARNING
The retaining nuts of Amperometric sensors are not designed for prolonged contact with aggressive substances. Strong
acids, caustic substances and solvents or their vapor may lead to failure of the retaining nut, ejection of the sensor and
loss of the process fl uid with possibly serious consequences, such as damage to equipment and serious personal injury.
Retaining nuts that may have been in contact with such substances, e.g. due to leakage or spilling, must be replaced.
3. Wiring
Chlorine Digital (S
• Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into terminals.
• Refer to the wiring diagram above to connect the 2650 cable to the terminals on the
8630 Chlorine Transmitter.
• Maximum cable length is 30 m (100 ft).
• Cable should have noise immunity for the sensor signal of up to 30 m (100 ft).
• Cable capacitance must be less than 30 pF/foot.
3
L) Wiring
White
Red
Black
Shield
Signal Ground
Digital (S3L) data
+5 VDC
15
14
13
Gnd
I/O
V
+
4. Technical Data Signet 2650 DryLoc Amperometric Electronics
General
Compatible Sensors:
All Signet Amperometric DryLoc Sensors
Compatible Instruments:
Signet 3-8630-3P Chlorine Transmitter
Mounting:
DryLoc connection
Materials:
®
Valox
(PBT)
Cable:
4.6 m (15 ft) 3 conductor shielded, 22 AWG
Shipping Weight:
0.64 kg (1.14 lb)
Performance
Electronics Accuracy:
< 5 nA or 1% of reading, whichever is greater
@ 25 °C over full input range
Temperature:
± 1.0 °C (PT1000) over full operation range
(when calibrated at ambient temperature)
System Response:
500 ms (update rate)
Operational Range:
± 450 mV
Resolution:
0.1 nA
* Valox® is a registered trademark of SABIC Innovative Plastics
Electrical
Input Specifi cations:
Sensor: Raw signal
Temperature: PT1000 RTD
Output Specifi cations:
Digital (S
3
L):
Serial ASCII, TTL level 9600 bps
Maximum Cable Length:
30 m (100 ft)
Power Supply Input:
Digital (S
3
L) mode:
5 to 6.5 V ± 10% , 3 mA max
Environmental
Storage Temperature:
–20 °C to 85 °C (–4 °F to 185 °F)
Operating Temperature:
0 °C to 85 °C (32 °F to 185 °F). Electronics only
Relative Humidity:
0 to 95%, non-condensing (no electrode connected)
Enclosure Requirements:
NEMA 4X/IP65 with electrode connected
Standards and Approvals
CE
Manufactured under ISO 9001 and ISO 14001
China RoHS (Go to www.gfsignet.com for details)
Dimensions
35.0 mm
(1.38 in.)
101 mm
(4.0 in.)
61.0 mm
(2.40 in.)
Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090
For the most up-to-date information, please refer to our website at www.gfsignet.com
3-2650.090 Rev. F 05/11 English © Georg Fischer Signet LLC 2011