Georg Fischer 5500 Operating Manual

Page 1
Signet 5500 Flow Monitor
d
r
*3-5500.090-1*
3-5500.090-1 Rev. K 3/06 English
Contents
1. Power Connections
2. Compatible Sensor Wiring
3. Sensor Pulse Output Connections
4. Auxiliary Pulse Output Connections
5. Totalizer Reset Connections
6. 4 - 20 mA Current Output Connections
7. Relay Connections
Remove terminal blocks for easy wiring
1.
2
+
1
Gnd
+
Gnd
Total
AUX
Std. Sensor
U
L
LISTED 77CJ
Freq. IN
Freq. IN
Sen. Pwr.
Open Collector Sensor
12-24 V
10 W
+
Iso. Gnd
output
PLS
output
-
Flow
reset
Patent No. D376,328
CAUTION!
• Remove power to unit before wiring input and output connections.
• Follow instructions carefully to avoid personal injury.
8. Output Functions
9. Menu Functions
10. Parts and Accessories
11. Specifi cations
12. Quick Reference Menu Parameters
13. Troubleshooting
14. Maintenance
2.
3.
Std. Sensor
U
L
LISTED 77CJ
Freq. IN
Freq. IN
Open Collector Sensor
2
+
1
Gnd
+
Gnd
Total
AUX
PLS
Iso. Gnd
Sen. Pwr.
output
output
12-24 V
10 W
+
-
Flow
reset
Patent No. D376,328
1. Power Connections
2. Compatible Sensor Wiring
CAUTION! Never connect 115 VAC or 230 VAC to rear power
terminals. High voltage AC will damage instrument and void warranty.
5500
12-24 V
10 W
Terminals
External power supply
-
+
12 - 24 VDC
OR
12 - 24 VAC
Std. Sensor
= Double Insulate
= DC or AC powe
+
-
Technical Notes:
• To reduce the possibility of noise interference, isolate AC power lines from signal lines.
• Maximum 4-20 mA loop impedance (sec. 6) is affected by the supply voltage.
Freq. IN
Freq. IN
Iso. Gnd
Sen. Pwr.
Open Collector Sensor
Signet Sensors:
Red
Black
Shld.
Red
Black
Shld.
5500 Terminals
515 525 2517
2000 2507 2536 2540 2550 Vortex*
Page 2
t
3. Sensor Pulse Output Connections (sec. 8.1)
3.1 Signet instrument with internal open-collector excitation voltage and pull-up resistor.
Technical Notes:
• Use 2-conductor shielded twisted-pair cable for output lines up to 30 m (100 ft.) max.
• Pull-up resistor NOT REQUIRED when connected to
+
100Ω
PLS
output
Gnd
5500 Terminals
Freq. IN* Red
Shld
Iso. Gnd
Red
Shld
Signet
Instrument
*(Open-collector sensor terminals)
Signet instruments.
3.2 Pulse output connection options for instruments without internal open-collector excitation voltage and pull-up resistor shown below.
NPN sink input
+
28 VDC
max.
-
Gnd
Input
Other
instrument
PNP source input
100Ω
+
PLS
output
Gnd
5500 Terminals
10 kΩ
Input
Gnd
-
+
External power supply
Other
instrument
28 VDC
max.
100Ω
+
PLS
output
10 kΩ
Gnd
5500 Terminals
4. Auxiliary Pulse Output Connections
4.1 Pulse mode (sec. 8.2) - External device control
• The auxiliary output can be confi gured for either TTL, CMOS, or open-collector compatible inputs using the proper pull-up resistor and an external power supply.
• TTL inputs: An external 5 VDC power supply and 10 kΩ pull­up resistor is recommended.
• CMOS inputs:
An external 3 to 15 VDC power supply and 10 kΩ pull-up
resistor is recommended.
• 12 VDC Open-collector inputs:
An external 12 VDC power supply and 10 kΩ pull-up resistor is
recommended.
• 24 VDC Open-collector inputs:
An external 24 VDC power supply and 10 kΩ pull-up resistor is
recommended.
4.2 Totalizer pulse mode (sec. 8.2)
• External counter or accumulator inputs without internal open­collector excitation voltage and pull-up resistor.
NPN sink input
Counter
100Ω
+
AUX
output
Gnd
5500 Terminals
10 kΩ
Input
Gnd
-
+
External power supply
or
Accumulator
28 VDC
max.
External power supply
10 Ω
+
-
3 - 28
VDC
Signal output
Signal ground
100Ω
+
AUX
output
Gnd
5500 Terminals
Technical Notes:
• Open-collector transistor, optically isolated, 5 mA max. sink, 28 VDC max. pull-up voltage.
• Programmable pulse width, 0.10 - 999. seconds (sec. 9.2 M)
• Selectable Active high or active low pulse logic (sec. 9.3 J).
• Use 2-conductor shielded twisted pair cable for AUX output lines to 30 m (100 ft) max.
External power supply
+
28 VDC
max.
Counter
or
Accumulator
PNP source inpu
100Ω
+
AUX
output
Gnd
5500 Terminals
10 kΩ
-
Gnd
Input
2
Signet 5500 Flow Monitor
Page 3
5. Totalizer Reset Connections (sec. 8.3)
Momentary pushbutton switch
2
1
To ta l reset
5500 Terminals
6. 4-20 mA Current Output Connections (sec. 8.4)
Technical Notes:
• Use 2-conductor shielded twisted pair cable for reset lines up to 30 m (100 ft) max. External reset switch overrides software security feature (section 9.3 I).
• To reduce the possibility of AC noise interference, isolate AC power lines from reset lines.
Fuse**
1/8A
+
*
4
20
-
-
+
4-20
mA
7. Relay Connections
Two internal relay contact sets (COM, NO, and NC) may be used for external device control. Front panel LED annunciators indicate the activation status of each relay. Each relay can control up to two devices simultaneously, as shown. Relay operation modes include Low alarm, High alarm, Pulse, and Totalizer (sec. 8.5).
Common device connections include:
Pulse mode - metering pump control
Pulse mode - solenoid valve control
Low or High mode - warning lamps
Low or High mode - bells or sirens
Low or High mode - external heavy-duty relay
5500 Terminals
mA
Technical Notes: **1/8A fuse recommended (customer supplied) * 4-20 mA output is internally powered (non-isolated), maximum loop impedance 350 Ω with a 12 V instrument supply voltage, 950 Ω with a 24 V instrument supply voltage.
To isolate output and prevent ground loop problems:
1. Use monitor device with isolated inputs, or
2. Use separate DC supply for 5500 and monitor device, or
3. Power 5500 with12 - 24 VAC step down transformer
Relay
Energized
NO
5500
Terminals
NO
C
NC
+
External AC/DC
Relay 1Relay 2
NO
power
Relay
De-energized
C
NO
NC
C
NC
Device
+
A
-
-
-
Device
B
+
C
NC
Wiring Example Right Device A IS powered when relay 2 is de-energized (front panel LED "off"). Power is discontinued when the relay 2 setpoint is reached (front panel LED "on"). Device B IS NOT powered when the relay 2 is de-energized. Power is applied after the relay 2 setpoint is reached.
Signet 5500 Flow Monitor
Technical Notes:
• Maximum relay contact ratings: 5 A @ 30 VDC, 5 A @ 125 VAC, or 3 A @ 250 VAC
• An external heavy-duty relay must be used for devices with surge currents or operating currents that exceed the above specifi cations.
3
Page 4
8. Output Functions
This section gives a detailed description of all 5500 output functions.
8.1 Sensor pulse output (sec. 3): This output emits an open-collector frequency equal to the sensor input. The output connects directly to a second Signet instrument for expanded process control with a single sensor input (sec. 3.1). The output is an open-collector type that requires an external pull up resistor and power supply for other instrument inputs (sec. 3.2).
8.2 Auxiliary pulse output (sec. 4): The auxiliary output can be confi gured as Low alarm, High alarm, Pulse or Totalizer output. Output logic can be selected for either active low (factory default) or active high operation (sec. 9.3 J). The auxiliary output is an open-collector type that requires an external pull-up resistor and power supply to operate. Refer to diagrams A - C below for operation details.
A. LOW alarm mode
The output is active (relay energized) when the flow drops below the setpoint, and is inactive (relay de-energized) when the flow rises above the setpoint plus hysteresis (sec. 9.2 F-G, 9.2 J-K).
• Auxiliary output: ideal for external device inputs (sec. 4)
Flow
• Relay outputs: ideal for external warning lamps, sirens, etc. (sec. 7)
Low setpoint=
Adjustable hysteresis=
Output active (relay energized)=
Output inactive (relay de-energized)=
B. HIGH alarm mode
The output is active (relay energized) when the flow rises above the setpoint and is inactive (relay de-engergized) when the flow falls below the setpoint plus hysteresis (sec. 9.2 F-G, 9.2 J-K).
• Auxiliary output: ideal for external device inputs (sec. 4)
• Relay outputs: ideal for external warning lamps, sirens, etc. (sec. 7)
High setpoint=
Adjustable hysteresis=
Output active (relay energized)=
Output inactive (relay de-energized)=
Flow
C. PULSE mode The output is active (relay energized) for the programmed pulse width each time the programmed pulse volume is measured by the flow sensor (sec 9.2 H-I, 9.2 L-M).
• Auxiliary output: ideal for external device control (sec. 4)
• Relay outputs: ideal for metering pump or solenoid valve control (sec. 7)
Flow
Sensor
FLOW
volume 2volume 3volume 4volume 5volume 6
volume
volume 1
= Output active (relay energized)
= Output inactive (relay de-energized)
= Programmed pulse volume (e.g. 12.5 Gal.)
= Programmed pulse width (relay energize time)
Totalizer function The output emits an adjustable signal pulse width for each whole engineeing unit measured by the flow sensor (also displayed on totalizer). To enable the totalizer function, the operator must set the relay or auxiliary pulse volume setpoint to 1.0000 (sec. 9.2 H, 9.2 L).
• Totalizer function: ideal for external counter or accumulator use (sec 4.2)
adjustable
pulse width
VIEW menu
totalizer display
=
1234
8.3 Totalizer reset (sec. 5): This feature allows the operator to reset the totalizer with a dry switch contact closure at any time. The external reset switch overrides software security feature (sec. 5).
8.4 4-20 mA output (sec. 6): The 4-20 mA output is internally powered by the instrument. 4 mA and 20 mA setpoints are fully adjustable within the programmed fl ow range (sec. 11 B). The output can also be reverse spanned.
8.5 Relay 1 and 2 outputs (sec 7): Relay outputs 1 and 2 can be independently confi gured as Low alarm, High alarm, Pulse or Totalizer output. Refer to diagrams A - C above for operation details.
4
Signet 5500 Flow Monitor
Page 5
9. Menu Functions
To access either CALIBRATE or OPTIONS menus, press and hold the ENTER key as illustrated below:
Press & hold ENTER key for menu access:
ENTER
VIEW
(sec. 9.1)
2
seconds
5
seconds
CALIBRATE
(sec. 9.2)
OPTIONS
(sec. 9.3)
9.1 VIEW Menu
Menus:
• VIEW menu (sec. 9.1): The VIEW menu is displayed during standard operation. The operator can navigate freely through the menu by pressing either UP or DOWN arrow keys. The VIEW menu also offers a resettable totalizer feature that allows the operator to reset the displayed total at the press of a button.
• CALIBRATE Menu (sec. 9.2): The CALIBRATE menu contains all critical display setup and output parameters. A simple secu­rity code feature prevents unauthorized tampering. The opera­tor is required to enter a simple access code for menu access. The same code also unlocks OPTIONS menus.
• OPTIONS Menu (sec. 9.3): The OPTIONS menu contains setup and display features that are seldom accessed for minor display or output adjustments.
1
H↓
A.
B.
C.
D.
E.
F.
G.
H.
Choose: Change: Save:
Flow & resettable total
Permanent total
Loop output
Range
Relay 1
Relay 2
Aux output
Last calibration
2
3
Menu Displays A - H: (Factory default displays shown in menu column 1)
A. Flowrate and resettable total display: can be reset
OPTIONS = OPTIONS =
(sec. 9.3I) (sec. 9.3I)
by the front panel keypad or by remote reset switch (sec. 5). A simple access code feature prevents
A.
A.
accidental total reset, when enabled (sec. 9.3 I). External reset switch inputs override security
To reset total press:
ENTER
Press keys in sequence,
(display shows: )
feature, when enabled.
B. Permanent total display: records volumetric fl ow
total for the life of the product and is non-resettable.
C. Loop output display: shows the current output level
B.
To exit without changes:
OR
quick press
B.
C.
To reset total press:
ENTER
at the rear terminals.
D. Flow Range display: shows the programmed min
and max meter range (sec. 9.2 B)
E. Relay 1 display: this display shows the pro-
grammed operation mode and setpoint for relay 1 (sec. 9.2 F-I)
F. Relay 2 display: this display shows the programmed
operation mode and setpoint for relay 2 (sec. 9.2 F-I).
G. Aux output display: this display shows the programmed operation mode and setpoint for the
auxiliary output (sec. 9.2 J-M).
H. Last calibration: this display shows a user defi ned
setup date for maintenance records. This feature is not an internal timer or calender.
A↑
Signet 5500 Flow Monitor
5
Page 6
9.2 CALIBRATE Menu
Press keys in sequence
will appear during code entry.
1
N↓
A.
B.
C.
D.
E.
F.
G.
H.
Choose:
Flow units
Flow range
Flow K-Factor
Total units
Total K-Factor
Relay Low or High Mode Selected
Relay
Relay
Relay Pulse Mode Selected, step F
Relay
Change:
Unit + timebase: h,m,s,d
Meter Min→Max
K-Factor
- Display label only -
K-Factor
Mode & setpoint
Hysteresis setting
Mode & setpoint
to enter menu,
Menu Settings A - N: (Factory default displays shown in menu column 1)
A. Sets fl ow units label (gpm) and timebase (gpm).
2
3
Save:
Flow units label: Aa - Zz, 0 - 9, / Timebase options: h=hours, m=minutes,
s=seconds, d=days (timebase entry affects fl ow rate display and 4 to 20 mA output)
B. Sets Min→Max meter and dial range, 00.000 to
99999. (does not affect 4 to 20 mA output)
C. Sets fl ow K-Factor: 0.0001 to 99999. (see technical notes below)
D. Sets 8-digit totalizer units label: Aa - Zz, 0 - 9, / (does not affect totalizer display or outputs)
E. Sets totalizer K-Factor: 0.0001 to 99999. (see techical notes below)
Menu items F - I repeat for relay 2 setup. F. Sets relay operation mode Low or High, and fl ow
rate setpoint, 0.0000 - 99999. (sec. 8.5)
G. Sets relay hysteresis, 0.0000 - 99999. (sec. 8.5) (Set to zero to disable hysteresis feature)
ENTER
"SAVING"
briefly
displays
H. Sets relay pulse volume setpoint, 0.0001 - 99999.
(sec. 8.5). Set to 1.0000 to enable totalizer
function (sec. 8.5).
I. Sets relay pulse width, 0.10 - 999. seconds (sec. 8.5)
I.
J.
K.
L.
M.
N.
Last calibration
To return to VIEW menu:
6
Relay
Aux output
Aux output
Aux output
Aux output
Pulse width
Aux Output Low or High Mode Selected
Mode & setpoint
Hysteresis
Aux Output Pulse mode selected, step J
Mode & setpoint
Pulse width
Setup date
A↑
To restore original value:
quick
press
quick press
J. Sets auxiliary output mode Low or High, and fl ow
rate setpoint, 0.0000 - 99999. (sec. 8.2).
K. Sets auxiliary output hysteresis, 0.0000 - 99999. (Set to zero to disable hysteresis feature)
L. Sets auxiliary output pulse volume setpoint,
0.0001 - 99999. (sec. 8.2). Set to 1.0000 (factory default) to enable totalizer function (sec. 8.2).
M. Sets auxiliary output pulse width, 0.10 - 999. seconds
N. Sets user defi ned setup date for maintenance
records. This feature is not an internal timer or calender
Technical Notes: The fl ow and total K-Factors are independent of each other. The K-Factor settings represent the number of pulses generated by the Signet fl ow sensor for each engineering unit of fl uid measured (published in fl ow sensor manual).
Signet 5500 Flow Monitor
Page 7
9.3 OPTIONS Menu
Press keys in sequence
will appear during code entry.
1
J↓
A.
B.
C.
D.
E.
F.
G.
H.
I.
Choose:
Display contrast Low -- -- -- -- -- High
Flow display Decimal
Total display Total decimal
Display averaging
Loop output
Loop output
Loop output
Loop output
Totalalizer reset
4 mA setpoint
20 mA setpoint
4 mA adjust
20 mA adjust
VIEW menu=
to enter menu,
Menu Settings A - J: (Factory default displays shown in menu column 1)
A. Selects LCD display contrast: 5 levels
2
Change:
3
Save:
B. Selects fl ow display decimal:
*****
. to
** . ***
C. Selects totalizer display decimal:
. to
********
****** . **
D. Selects LCD display averaging: Off = 0 seconds, Low= 4 seconds, High= 8 seconds (also effects 4 - 20 mA output)
E. Sets 4 mA output setpoint. 4 mA and 20 mA setpoints are reversible.
4s 8s0s
ENTER
F. Sets 20 mA output setpoint. 20 mA and 4 mA
setpoints are reversible.
"SAVING"
briefly
displays
G. Sets 4 mA current output: 3.0 to 5.0 mA (overrides 4.00 mA factory calibration)
H. Sets 20 mA current output: 19 to 21 mA (overrides 20.00 mA factory calibration)
I. Selects totalizer reset options: Lock "Off" disables
the VIEW menu access code feature; Lock "On" enables the feature (sec. 9.1 A). When enabled, the operator must enter an access code for totalizer reset privileges. The totalizer will be reset in either case if a contact closure is detected at rear reset
VIEW menu=
terminals (sec. 5).
J.
Auxiliary out logic
To return to VIEW menu:
quick press
Active Low or High
A↑
To restore original value:
quick press
J. Selects Auxiliary output logic:
• Active Low logic (factory default): Output changes from high to low when active, see diagram below.
• Active high logic: Output changes from low to high when active, see diagram below.
Active Low Logic (factory default)
Inactive (Off)
Active (On)
Active High Logic
Active (On)
Inactive (Off)
Signet 5500 Flow Monitor
7
Page 8
10. Parts and Accessories
Splashproof rear cover #3-5000.395 (code 198 840 227)
5 x 5 inch adapter plate for Signet retrofi t #3-5000.399 (code 198 840 224)
• Power supply, 115 VAC - 24 VAC, #3-5000.075
• Front snap-on bezel, #3-5000.525 (code 198 840 226)
• Assorted fl ow unit/multiplier decals, #3-5500.611 (code 198 840 230)
• 5500 instruction sheet #3-5500.090-1 (code 198 869 914)
11. Specifi cations
General Sensor compatibility: All Signet fl ow sensors Accuracy: ±0.5% of reading Input: Optically isolated
Enclosure:
• Rating: NEMA 4X/IP65 front
• Dimensions: 1/4 DIN, 96 x 96 x 88 mm (3.8 x 3.8 x 3.5 in.)
• Case materials: ABS plastic
• Keypad material: Sealed 4-key silicone rubber
• Weight: Approximately 500 g (18 oz.)
Optional surface mount bracket #3-5000.598 (code 198 840 225)
Electrical Power:
• 12 to 24 VDC or 12 to 24 VAC, unregulated, 50-60 Hz, 10 W max.
Relay contacts (2 sets):
• Mechanical SPDT contacts
• Maximum voltage rating: 5 A @ 30 VDC, 5 A @ 125 VAC, or 3 A @ 250 VAC, (power factor = 1.0)
• Hysteresis: User adjustable
Display:
• Type: Microprocessor controlled air-core meter movement and backlit alphanumeric 2 x 16 LCD
• Update rate: Flow = 1s, totalizer = <200 ms
• Contrast: User selected, 5 levels
• Relay annunciators: 2 LEDs
Totalizers:
• 8-digit resettable with security option
• 8-digit non-resettable
Environmental Operating temp.: -10 to 55°C (14 to 131°F), 50°C (122°F) max. with optional rear cover Storage temp.: -15 to 80°C (5 to 176°F) Relative humidity: 0 to 95%, non-condensing Altitude: 4000 m max. Pollution degree: 2
Agency Approvals
• CE, UL listed
• Manufactured under ISO 9001
Current output:
• 4 to 20 mA, non-isolated, internally powered, fully adjustable and reversible
• Update rate: < 200 ms
• Max loop impedance: 350 Ω max with a 12 V instrument supply voltage, 950 Ω max with a 24 V instrument supply voltage
• Accuracy: ±0.1% of max range
Sensor (PLS) Auxiliary (AUX) pulse outputs:
• Open-collector transistor, optically isolated, 5 mA max. sink, 28 VDC max. pull-up voltage, programmable (AUX) pulse width
100Ω
+
5500 rear PLS/AUX
output terminals
GND
Internal PLS/AUX circuits
Totalizer reset:
• Front keypad or external contact closure, 30 m (100 ft) max cable length
• External contact closure overrides security feature (sec. 5)
Noise immunity: EN50082-2 Noise emissions: EN55011 Safety: EN61010-1
8
Signet 5500 Flow Monitor
Page 9
Dimensions
96 mm (3.8 in.)
Panel Gasket
Optional Splashproof Rear Cover
Front
View
Rear View
RELAY 1
Freq. IN
Std. Sensor
LISTED
U
L
®
77CJ
Signet Model #3-5500
3
2
1
0
Signet Monitor
88 mm (3.5 in.)
+
Gnd
PLS
Freq. IN
Iso. Gnd
Sen. Pwr.
output
Open Collector Sensor
12-24 V
10 W
+
-
4
GPM
Gnd
+
AUX
output
NO
5
6
RELAY 2
7
8
Mounting Panel
Mounting Clamp (2 ea.)
9
10
96 mm (3.8 in.)
91 mm (3.6 in.)
88 mm (3.5 in.)
SIDE VIEW
ENTE
76 mm (3 in.)
96 mm (3.8 in.)
88 mm (3.5 in.)
2
+
1
-
4-20
Tot a l reset
Relay 1Relay 2
C
NC
NO
mA
C
NC
Flow
88 mm (3.5 in.)
Patent No. D376,328
92 x 92 mm
(3.62 x 3.62 in.)
12. Quick Reference Menu Parameters
VIEW Menu Setup Parameters (sec. 9.1)
Menu
Parameters
A.
• Resettable total
B.
C.
D.
E.
F.
Display
Description
• Flow rate and units
Permanent
total
Current loop
output
Min → Max meter
and dial range
• Relay 1 mode
• Relay 1 setpoint
• Relay 2 mode
• Relay 2 setpoint
Panel cutout
Range
00.000 - 99999. GPM
000000.00 - 99999999.
000000.00 - 99999999.
3 - 21 mA
00.000 - 99999.
Low or High, 0.0000 - 99999.
Pulse, 0.0001 - 99999.
Low or High, 0.0000 - 99999.
Pulse, 0.0001 - 99999.
Factory
Default
0000.0
0000000.0
0000000.0
n/a
0.000 - 100.00
Low
10.000 GPM
High
90.000 GPM
G.
H.
Signet 5500 Flow Monitor
• Aux output mode
• Aux output setpoint
Last calibration
date
Low or High, 0.0000 - 99999.
Pulse, 0.0001 - 99999.
00 - 00 - 00 to
39 - 39 - 99
Pulse
1.0000 Gallons
01 - 01 - 98
9
Page 10
CALIBRATE Menu Setup Parameters (sec. 9.2)
A.
B.
C.
D.
E.
F.
G.
H.
Menu
Parameters
Display
Description
flow unit= digits 1-3
flow timebase= digits 4
Range
Aa - Zz, 0 - 9, /, _;
Mm, Ss, Hh, Dd
Factory
Default
_GPM
_ = blank display digit option
Min → max
range
Flow K-Factor
Total units
Total K-Factor
• Relay 1 mode
• Relay 1 setpoint
00.000 - 99999.
0.0001 - 99999.
Aa - Zz, 0 - 9, /, _
(8-digits max.)
0.0001 - 99999.
Low or High
0.0000 - 99999.
00.000 - 100.00 GPM
60.000
_Gallons
60.000
Low
10.000 GPM
_ = blank display digit option
Relay mode and setpoint
• Relay 1 Hysteresis
• Hysteresis volume
• Relay 1 mode
• Relay 1 pulse volume
I.
Relay 1 pulse width
0.0000 - 99999.
Pulse
0.0001 - 99999.
0.10 - 999. seconds
2.0000 GPM
10.000
Gallons
10.0
seconds
displays repeat for relay 2 setup
J.
K.
L.
M.
N.
• Aux output mode
• Aux output setpoint
Aux output hysteresis
volume
• Aux output mode
• Aux output setpoint
Aux output pulse
width
Last calibration
date
OPTIONS Menu Setup Parameters (sec. 9.3)
Menu
Parameters
A.
Display
Description
Display contrast
Low or High
0.0000 - 99999.
0.0000 - 99999.
Pulse
0.0001 - 99999.
0.10 - 999. seconds
00 - 00 - 00 to
39 - 39 - 99
Range
0 to 5
Low
1.0000
0.0000
1.000
Gallons
0.10
seconds
01 - 01 - 98
Factory
Default
3
10
B.
C.
D.
E.
Flow decimal
Totalizer decimal
Display averaging
4 mA setpoint
**.*** to *****.
******.** to ********.
Off= 0 sec., Low= 4sec.,
High= 8 sec.
Unlimited
****.*
*******.*
Off
0.0000 GPM
Signet 5500 Flow Monitor
Page 11
OPTIONS Menu Setup Parameters Continued...
F.
G.
H.
J.
Menu
Parameters
I.
Auxiliary (AUX) output
Display
Description
20 mA setpoint
4 mA adjust
20 mA adjust
VIEW menu total
reset access code
pulse logic
Range
Unlimited
3.0 to 5.0 mA
19 to 21 mA
On or Off
Active Low or
Active High
Factory
Default
100.00 GPM
4.00 mA
20.00 mA
Off
Low
13. Troubleshooting
Display Problem Solution
-- -- -- -- Display timebase too large Change fl ow timebase (S=Seconds, M=Minutes, H=Hours, D=Days) in CALIBRATE menu to a smaller value (e.g. GPD to GPM)
Check settings for Relay 1
Check settings for Relay 2
Check settings for AUX output
Must be between
0.0001 - 99999
Must be between
0.10 - 999
Flow K-Factor can not be zero
Total K-Factor can not be zero
SETUP READ ERROR Press any Key
Relay 1 pulse width too large for frequency input or pulse volume too small
Relay 2 pulse width too large for frequency input or pulse volume too small
Auxiliary output pulse width too large for frequency input or pulse volume too small
Relay 1, Relay 2, or Auxiliary output (AUX) pulse volume setpoint must be within 0.0001 - 99999.
Relay 1, Relay 2, or Auxiliary output (AUX) pulse width must be within
0.10 - 999 seconds Flow display K-Factor cannot be
zero Totalizer K-Factor cannot be
zero Power fault occurred while saving
setup menu entry
Reduce Relay 1 pulse width setting (sec. 9.2 I) or increase pulse volume setting (sec. 9.2 H)
Reduce Relay 2 pulse width setting (sec. 9.2 I) or increase pulse volume setting (sec. 9.2 H)
Reduce auxiliary output pulse width setting (sec. 9.2 M) or increase pulse volume setting (sec. 9.2 L).
Enter volumetric setpoint within 0.0001 - 99999. engineering units (sec. 9.2 H or 9.2 L)
Enter pulse width within 0.10 - 999 seconds (sec. 9.2 I or 9.2 M)
Enter K-Factor greater than zero (sec. 9.2 C)
Enter K-Factor greater than zero (sec. 9.2 E)
Press any key to reload factory defaults then reprogram fl ow system setup parameters. Note: totalizer displays remain undistrubed and will resume after instrument reset.
14. Maintenance
Clean the instrument case and front panel with a soft cloth and mild liquid soap solution.
Signet 5500 Flow Monitor
11
Page 12
George Fischer Signet, Inc. 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090
3-5500.090-1 Rev. K 3/06 English © George Fischer Signet, Inc. 2001 Printed in U.S.A. on recycled paper
Loading...