Signet 4632 Chlorine Dioxide Analyzer System
Features
• Reagent free measuring
• Complete panel system allows for quick and
easy installation
• Built-in flow regulator maintains constant flow
and pressure to the sensors regardless of inlet
pressure
• Pre-wired panel includes a 100/240 VAC
power supply, two 4 to 20 mA outputs and two
mechanical relays
The Signet 4632 Chlorine Dioxide Analyzer System is an
integrated all-in-one system designed to measure
Chlorine dioxide residual up to 2 ppm/mg/l.
The unique integrated clear flow cell combines sensors,
flow regulator, filter and variable area flow indicator
in one compact unit. An integrated flow regulator with
removable filter accepts inlet pressures of 1 to 8 bar
(15 to 120 psi), while maintaining constant flow and
minimal pressure to the sensors.
Water flows vertically into sensor tip eliminating bubble
entrapment. The flow cell is designed to maintain a
minimum amount of water to ensure sensors stay
submerged, even when the system and flow is
turned off.
The Signet 4630 Chlorine Analyzer System allows quick
setup and easy installation and is supplied with a
100-240 VAC power supply, two 4 to 20 mA outputs and
two dry contact mechanical relays.
MEASURING EQUIPMEN
Applications
Residual Chlorine Monitoring:
• Cooling Towers
• Fruit and Vegetable Washing
• Water Distribution
• Wastewater Odor Control
• Poultry and Meat Processing
• UPW Treatment
• Hospital and Healthcare Facilities
U.S. Patent No: 8,336,375 B2

Specifications
General
Compatible 3-2632-1 Chlorine Dioxide Electrode, 0 to 2 ppm / 3-2650-7 Amperometric
Electronics
Materials
Panel Black Acrylic
Flow Cell Acrylic
Wiring Enclosure Polycarbonate
Wetted Materials
Flow Cell, Spacer Rings Acrylic
Flow Regulator Housing Polycarbonate
Strainer, E-clip, Regulator Spring, Float Stainless Steel
Valves, Vent Polypropylene
Flow Cell O-rings, Diaphragm EPDM, FKM
Chlorine Electrode PVC, PVDF, FPM
pH electrode PPS, Glass, UHMW PE, FPM
Sealing Tape on Valves, Plug and Vent PTFE
Plug Polyethylene
Max. Temperature/Pressure Rating
System Inlet Pressure Rating 1 to 8 bar 15 to 120 psi
Pressure Regulator
Flow Tolerance ± 15% or rated specification above
Flow Rate Limits 30.24 to 45.36 LPH 8 to 12 US gal/h
Storage Temperature 0 °C to 65 °C 32 °F to 149 °F
Operating Temperature 0 °C to 45 °C 32 °F to 113 °F
Electrical
AC Input - Standard Configuration 100 to 240 VAC nominal 50 to 60 Hz, 0.17 A at 100 VAC
DC Input - Optional Configuration 12 to 24 VDC ±10% regulated, 250 mA max.
Environmental
Relative Humidity 0 to 95%
Maximum Altitude 2000 m (6,562 ft)
Enclosure NEMA 4X (with output wire glands sealed)
Shipping Weight
Standards and Approvals
< 0.69 bar (10 psi) variation over all ranges of flow and pressure
10 kg 22 lb
CE, UL, CUL, FCC
China RoHS
Manufactured under ISO 9001 for Quality and ISO 14001 for Environmental
Management and OHSAS 18001 for Occupational Health and Safety
www.gfsignet.com

Ordering Information
Mfr. Part No. Code Description
3-4632-10
3-4632-11
159 001 768 Chlorine Dioxide panel, 0 to 2 ppm/mg/l, no pH sensor
159 001 769 Chlorine Dioxide panel, 0 to 2 ppm/mg/l, with pH sensor
Accessories and Replacement Parts
Mfr. Part No. Code Description
3-2632-1 159 001 767 Chlorine Dioxide electrode, 0 to 2 ppm (mg/L)
3-2650-7 159 001 670 Chlorine - Inline Amperometric Electronics, Digital (S3L), 4.6 m (15 ft) cable
3-2724-00 159 001 545 pH sensor, flat glass , PT1000 temp element, 3/4" MNPT
3-2750-7 159 001 671 pH - Inline Electronics, Digital (S3L) , 4.6 m (15 ft) cable
3-8630-3P 159 001 673 Panel mount chlorine and pH transmitter
3-4630.390 159 001 688 Rebuild kit: O-rings, boots, screws, 1 filter screen
3-4630.391 159 001 689 Pressure regulator with 1 spare filter screen
3-4630.392 159 001 690 Acrylic flow cell complete with all components and connections
3-2632.391 159 310 160 Chlorine Dioxide electrolyte, 30 mL
3-2632.398 159 310 165 Chlorine Dioxide maintenance kit - (2) electrolyte, (2) PTFE membranes, (2) Silicone Bands,
and Polishing Paper
3-2630.394 159 310 164 Free Chlorine and Chlorine Dioxide Replacement PTFE membrane (1)
(sensors sold after Nov 1, 2012)
7300-0024 159 001 693 VDC Power Supply
3-4632.099 Rev A (09/13)
© Georg Fischer Signet LLC
3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 • www.gfsignet.com • e-mail: signet.ps@georgfischer.com
Specifications subject to change without notice. All rights reserved. All corporate names and trademarks stated herein are the property of their respective companies.