Georg Fischer 4150 Specifications

Signet 4150 Turbidimeter
The Signet 4150 Turbidimeter system provides accurate and reliable compliant water quality monitoring for municipal and industrial applications.
Features
• Simple and easy single unit installation with built-in pressure regulator
• Versions compliant with either U.S. EPA 180.1 for North and South America and Asia or ISO 7027 for Europe
• Time saving and e ciencies of cuvette
technology simplifi es calibration
• Spannable 4 to 20 mA output or RS 485 output
• Two adjustable alarm relays
• Bright backlit display
• Easy access for wiring and maintenance
• Ultrasonic cleaning option ensures long and steady on-line measurement
• Inexpensive standards allow for multiple system calibrations
The 4150 measures turbidity via a 90 degree light which refl ects particles as they fl ow through a small volume, low fl ow glass cuvette. Air bubbles are eliminated from the cuvette by adjusting the backpressure valve on the outlet tube. The cuvette is located in a watertight dark chamber for continuously accurate on-line measurement. A replaceable desiccant pack provides a dry-stable environment to ensure reliable measurements.
Simple and fast calibration can be accomplished in under fi ve minutes by placing the in-line glass cuvette from the measuring chamber into the cuvette holder while still in service and the inlet and outlet tubing remains connected. The inexpensive calibration standard allows for dry and multiple system calibrations without mixing chemicals. After calibration, the unit is up and running with simple re-insertion of the glass cuvette back into the measuring chamber.
Additional features include a message indicator when the desiccant needs replacing and as an option, auto/ ultrasonic cleaning of the glass in-line cuvette for longer runs between maintenance.
The 4150 is available in two measuring ranges. The 0 to 100 NTU/FNU version is for low range applications such as drinking water. The 0 to 1000 NTU/FNU range can be used for various applications including raw water and wastewater reclamation.
©
CUS
L
I
D
S
E
T
Applications
• Monitor Filter Performance
• Raw or Filtered Water
• Municipal Water Distribution
• Wastewater Reclamation and Tertiary E uent
• Aquatic Life Support
Specifi cations
General Flow Rate Range 0.1 L/min to 1 L/min (0.026 GPM to 0.26 GPM)
Measurement Range 0 to 100.0 NTU/FNU or 0 to 1000.0 NTU/FNU Accuracy ±2% of reading or ±0.02 NTU/FNU below 40 NTU/FNU whichever is greater
±5% of reading above 40 NTU/FNU
NTU = FNU = FTU
Mounting
Horizontal plane, integral mounting bracket (with standard hole pattern) Use 8 mm (5/16”) OD, 5 mm (3/16”) ID fl exible tubing for the water supply/outlet (customer supplied)
Resolution
up to 0.0001 NTU/FNU (below 10 NTU/FNU)
Display
Two-Line LCD w/backlight
Alarm Relays
120-240 VAC, 2A Form C Relay
Analog Signal w/Field Selectable Range
Active 4-20 mA, 600 Ω or RS485
Wetted Materials
Tubing Vinyl Measuring Cuvette Borosilicate Glass Glass Washer Seal Silicone Pressure Regulator Polypropylene 316 stainless steel acetal Inlet Tube 316 stainless steel
Maximum Inlet Pressure
345 kPa (50 psi) based on tubing connection provided
Pressure regulator rated up to 200 psi
50 PSI limit for tubing connector
Power Supply
100 – 240 VAC, 47 – 63 Hz, 80 VA
Insulation Rating
Double Insulated
Pollution Degree 2
Overvoltage Category II
Altitude
2000 meters (6,561 ft) maximum
Relative Humidity
Maximum 95% RH non-condensing
Enclosure Rating
Power Supply Box NEMA 4X / IP66
Operating Temperature/Pressure
1 °C to 50 °C 34 °F to 122 °F
(5 to 15 psig) 35 to 104 kPa
Environmental Conditions
Not recommended for outdoor use
Shipping Weight
2.5 kg 5.5 lb
Standards and Approvals
CE, FCC RoHS compliant, China RoHS Compliant to U.S. EPA 180.1 for white light Compliant to PN EN ISO 7027 for infrared light ETL Listed UL 61010-1, cETL and CSA C22.2 No. 61010-1