The Signet 3719 pH/ORP Wet-Tap allows installation and removal of pH or ORP electrodes without the need for process shutdown during routine electrode maintenance and calibration. Process isolation is achieved with a double O-ring seal on a unique and compact retraction assembly; no separate valve is required. A cam-activated automatic locking mechanism, SafeLoc™, and the short stroke design help to assure operator safety.
English
Operating Instructions
Description
WARNING!
English
Deutsch
Français
Español
Table of Contents
Warranty Information ......................................................... 2
Product Registration .......................................................... 2
Safety Information .............................................................2
Dimensions ........................................................................ 2
Specifi cations .................................................................... 3
System Overview............................................................... 3
Installation ......................................................................3-5
Location, Orientation and Required Clearance ............ 3-4
For pipe sizes 2.5 to 12 in. .......................................... 4
For pipe sizes less than 2.5 in. .................................... 4
Wet-Tap into Fitting...................................................... 5
Electrode Installation ................................................... 5
Electrode Removal ............................................................ 6
Electrode Maintenance ...................................................... 6
Troubleshooting ................................................................. 7
Replacing O-rings .............................................................. 8
Ordering Information........................................................ 12
Page 2
i
.
Warranty Information
Dimensions
Refer to your local Georg Fischer Sales offi ce for the most current warranty statement.
All warranty and non-warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales offi ce or distributor. Product returned without a Service Form may not be warranty replaced or repaired.
Signet products with limited shelf-life (e.g. pH, ORP, chlorine electrodes, calibration solutions; e.g. pH buffers, turbidity standards or other solutions) are warranted out of box but not warranted against any damage, due to process or application failures (e.g. high temperature, mishandling (e.g. broken glass,
chemical poisoning, dry-out) or
damaged membrane, freezing
and/or extreme temperatures).
Product Registration
Thank you for purchasing the Signet line of Georg Fischer measurement products.
If you would like to register your product(s), you can now register online in one of the following ways:
• Visit our website www.gfsignet.com. Under Service and Support click on
Product Registration Form
• If this is a pdf manual (digital copy),
click here
3719-1 Wet-Tap
11/2 in. outlet for 21/2 to 4 in. pipe
76 mm (3.0 in.)
49 mm (1.94 in.)
25 mm
(1.00 in.)
39 mm (1.52 in.)
Pipe thread
1
1
/2 in. NPT
or
ISO 7-R1
20 mm (0.8 in.)
1
/
2
3719-2 Wet-Tap
2 in. outlet for 6 to 12 in. pipe
130 mm (5.13 in.)
Pipe thread
2 in. NPT
or
ISO 7-R2
75 mm (2.94 in.)
45 mm (1.75 in.)
76 mm (3.0 in.)
39 mm (1.52 in.)
41 mm
20 mm (0.8 in.)
34 mm
(1.32
(1.62 in
Safety Information
Do not exceed maximum temperature/pressure specifi cations.
Wear safety goggles or face shield during installation/service.
Do not attempt to disassemble the retraction housing and the electrode piston.
Do not alter product construction.
Failure to follow safety instructions may result in severe personal injury.
Retract electrode before any pipe cleaning operation.
Caution / Warning / Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death
Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during installation and service of Signet products.
Pressurized System Warning
Sensor may be under pressure, take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury.
Note / Technical Notes
Highlights additional information or detailed procedure.
35 mm
(1.4 in.)
2756-WTP(-1) pH electrode, plastic
2757-WTP ORP electrode, plastic
218 mm
(8.57 in.)
CAUTION!
When using clamp-on saddle fi ttings (customer-supplied), the system temperature and pressure is limited by the saddle specifi cations. The process temperature is limited to 40 °C (104 °F) maximum when using saddles.
2 Signet 3719 pH/ORP Wet-Tap
Page 3
Specifi cations System Overview
General
Compatible sensors (See 3-2764.090 for more information): 2756 -WTP DryLoc plastic pH electrode 2756-WTP-1 DryLoc plastic pH electrode 2757-WTP DryLoc plastic ORP electrode
d
e
Shipping Weight
Wet-Tap assembly ............ 1.2 kg (2.7 lbs.)
Electrode ........................... 0.13 kg (0.3 lb)
Process Connection ............. 3-3719-11: 1½ in. NPT
3-3719-21: 2 in. NPT 3-3719-12: ISO 7/1-R1½ 3-3719-22: ISO 7/1-R2
Wetted materials
Retraction Housing ............... CPVC
O-rings .................................. FKM
Other Materials
Locking Shroud..................... PVC
Hardware .............................. 316 stainless steel
Storage Temperature
Wet-Tap assembly ............ -15 to 120 °C (5 to 248 °F)
Standards and Approvals
Manufactured under ISO 9001 and 14001
25
China RoHS (Go to www.gfsignet.com for details)
c
a
b
a) 3719 pH/ORP Wet-Tap b) Customer-supplied Low-Profi le Clamp-on Saddle Fitting
(ASTM sizes 2½ to 12 in.)
c) DryLoc
(DryLoc d) 3-2751-1 or 2760 DryLoc
e) 3-2751-2 electronics
NOTE: Output signal options (See specifi c manual for details): 2751:
- Digital (S
- 4 to 20 mA 2760:
- Analog mV Output- Digital (S
®
ordering information on page 8.)
pH or ORP Electrode (6 versions available; see
®
refers to the electrode connector style)
®
pH/ORP Sensor Electronics/
Preamplifi er
3
L)
3
L)
All components (items a–d) are sold separately.
Installation
Notes on Location, Orientation and Required Clearance
The 3719-1X is designed for use in pipes up to 4 in.
The 3719-2X is designed for use in pipes from 6 to 12 in.
Select a location that will provide suffi cient clearance to remove and insert the electrode.
The 3719 can be mounted in any orientation, including horizontal and inverted. Avoid the 12 o'clock position. In the presence of sediment, avoid the 6 o'clock position. See next page for dimensional information.
If inverted, use caution when removing the sensor.
Residual fl uid may be
present in the retraction housing. Keep electrode connector clean and dry at all times.
Any angle OK
3Signet 3719 pH/ORP Wet-Tap
Page 4
Installation continued Installation continued
Provide 20 inches (minimum) clearance from the top of the pipe for electrode removal..
218 mm (8.57 in.)
115 mm (4.5 in.)
250 mm (9.9 in.)
OR
89.4 mm (3.5 in.)
45 mm (1.75 in.)
Installation in Pipe Sizes less than 2.5 Inches
It is possible to install the 3719 into pipe sizes below
2.5 inches by creating a “fl ow cell” with standard piping components.
One simple solution, using a tee fi tting and reducer bushings, is shown in the example below.
Many similar confi gurations are conceivable in a wide variety of materials, but be very careful to verify dimensional compatibility.
Select an appropriate installation orientation to avoid the entrapment of air inside the fl ow cell.
Contact your local Georg Fischer Sales and Support offi ce for assistance.
3-3719-1X pH/ORP Wet-Tap
1½ in. Threads
2 in /d50 mm Tee (S x S x S)
Flush Style Reducer Bushing (Spg x FT) 2 in. x 1½ in.
Flush Style Reducer Bushings (Spg x S) 2 in./d50 x ½ in./12 mm 2 in./d50 x ¾ in./20 mm 2 in./d50 x 1 in./25 mm 2 in./d50 x 1¼ in./30 mm 2 in./d50 x 1½ in./40 mm
Installation in Pipe Sizes 2.5 to 12 Inches
For reliable in-line measurements of pH and ORP, it is imperative to position the electrode tip into the process stream.
Because of its compact “short stroke” design, the 3719 requires low-profi le fi ttings to ensure proper positioning in pipe sizes DN65 to DN300 (2.5 to 12 inches).
It is strongly recommended
to use low-profi le fi ttings (customer-supplied).
Choose the 3719 version (-1X or
-2X) appropriate to the size of the branch connection of the required tting:
Use -1X for sizes DN65 to DN100
(2.5 to 4 in), and -2X for sizes DN150 to DN300 (6 to 12 in.)
3-3719-1X pH/ORP Wet-Tap (Fully Inserted)
4 Signet 3719 pH/ORP Wet-Tap
Page 5
Installation continued
Installation (Wet-Tap Into Fitting)
The 3719 is packaged with no electrode installed, and with the electrode piston in the fully inserted position.
Examine the female threads at the top of the electrode piston. Do not install electrode if threads are damaged.
Examine the two O-rings at the lower end of the assembly. Do not install if O-rings are missing or there are any signs of damage.
The internal O-rings (4 each) have been lubricated during assembly with Synthetic grease (PTFE).
The pipe can be pressurized after completion of step 4.
DO NOT ATTEMPT TO REMOVE THE RETRACTION HOUSING FROM A PRESSURIZED PIPING SYSTEM!
12
Grasp the retraction housing below the locking shroud; turn the locking shroud 1/4-turn clockwise.
Locking
Shroud
Retraction
Housing
Lubricated
O-rings
Pull the locking shroud straight up until both O-rings are fully seated inside the retraction housing.
The electrode piston is locked in position by SS locking pins.
DO NOT tamper with the locking pins! If the piston is depressed with no electrode installed, the pipe contents are exposed.
34
Turn the shroud 1/4-turn counterclockwise and lift it completely away from the electrode piston.
Electrode
Piston
• Thread the Wet-Tap into the pipe fi tting.
• Smaller pipes may require bracing to support the Wet-Tap weight and the longitudinal forces required for operation.
• Use an appropriate thread sealant to prevent leaks.
• The piping system can now be safely pressurized.
• Inspect the installation for leaks.
Electrode Installation
Before installation, lubricate O-rings with a non­petroleum based, viscous lubricant (grease) compatible with the system.
1. Remove the safety plug from top of electrode piston. Slide electrode straight down into electrode piston. Thread electrode into place until connector shoulder is fl ush with top of electrode piston. Hand tighten only.
WARNING:
Do not fl ex the electrode when inserting into the Wet-Tap assembly. The electrode can be permanently damaged if it is fl exed during installation.
2. Place the Locking Shroud over electrode; turn ¼-turn clockwise to unlock the piston, then press down fi rmly on the locking shroud to lower the electrode piston into the pipe.
3. Turn the shroud 1/4-turn counterclockwise to lock the piston.
4. Install the matching electronics assembly or preamplifi er onto the electrode connector.
Lubricate O-Ring
1
DO NOT ATTEMPT TO REMOVE THE RETRACTION HOUSING FROM A PRESSURIZED PIPING SYSTEM!
Lubricate O-Ring
2751 2760 or 2751
2
OR
3
4
5Signet 3719 pH/ORP Wet-Tap
Page 6
Electrode Removal
The electrode in any pH or ORP system requires periodic service, calibration or replacement. When removing the electrode from the Wet-Tap assembly, it is very important to exercise caution and follow the instructions carefully.
1. Remove the preamplifi er from the top of
the Wet-Tap assembly.
2. Turn the locking shroud
to unlock the piston.
3. Pull up on the locking shroud to retract the
electrode piston into the pipe.
1
/4-turn clockwise
1
DANGER!
The piston should retract easily from a pressurized pipe. If the piston offers resistance, there is a danger that the piston is coated with deposits from the process.
STOP! DO NOT FORCE THE PISTON UP!
It may damage the O-rings or break off the piston tip.
Return the locking shroud to the LOCKED position and follow the steps in the box below.
2
3 4
3
5
5
4. Turn the locking shroud ¼-turn
counterclockwise and lift up to remove it from the Wet-Tap assembly.
5. Remove the electrode by turning it counterclockwise.
For safety, keep your body clear of the top of the Wet-Tap
assembly while removing the electrode.
If the piston does not retract easily, or if any fl uid is observed leaking from the electrode threads, the pipe must be drained before the 3719 can be safely removed.
• Stop the fl ow and depressurize the pipe.
Remove the entire 3719 assembly from the pipe.
Remove the electrode from the assembly
Clean any scaling and debris found on or in the piston tip and surrounding area.
DANGER!
If any fl uid is observed leaking from the electrode threads, STOP! DO NOT REMOVE THE ELECTRODE. There is a danger that the piston tip has been damaged.
Tighten the electrode back down to reseal the assembly and follow the steps in the box below.
6 Signet 3719 pH/ORP Wet-Tap
Page 7
Replacing O-rings
Tools required:
• Table vise
Side cutter to cut O-rings
1 3/8" wrench
• Small screwdriver
• 3/16 Allen wrench
• Scissors
Replacement O-ring kit for 3719 Wet-Tap (3-3719.392 code 159 310 304)
Lubricant
Remove shroud
1
Remove nose piece
4
Press and hold release pin, push
2
down to extend Wet-Tap piston
Release pin
Remove old O-rings (use the screwdriver to
5
carefully lift o-ring from groove and cut with scissors). Inspect O-ring grooves for any damage or nicks that may compromise the integrity of the liquid seal. Clean the Wet-Tap piston and nose piece.
NOTE: Clean with a soft brush and a chemical compatible solvent to remove any debris
Carefully remove the Stainless Steel
3
set screw
Open the lubrication pouch and lightly
6
lubricate the new O-rings. Carefully roll O-rings over the tip of the piston and into each O-ring groove.
Lubricant
Locate the countersink hole on the
7
nosepiece. Verify that it is clean and free of debris. Locate the threaded hole on the main body. Verify that threads are free of debris and are not damaged. If threads are damaged, the Wet-Tap should be replaced.
Assemble the nose piece back onto the Wet-Tap main body.
Stainless Steel set screw hole
countersink hole
Replace the Stainless Steel set screw. Do not over
8
tighten. When properly installed, the set screw should be flush with the Wet-Tap body.
watch a video of these instructions.
Click here
Replace the shroud.
9
The Wet-Tap is ready for reinstallation.
7Signet 3719 pH/ORP Wet-Tap
Page 8
Ordering Information
Mfr. Part No. Code Description
3-3719-11 159 000 804 pH/ORP Wet-Tap, 1
3-3719-21 159 000 805 pH/ORP Wet-Tap, 2 in. NPT
3-3719-12 159 000 806 pH/ORP Wet-Tap, ISO 7/1-R 1.5
3-3719-22 159 000 807 Wet-Tap Assembly, ISO 7/1-R 2
Parts and Accessories
3-2756-WTP 159 001 390 Electrode, pH, DryLoc
3-2756-WTP-1 159 001 384 Electrode, pH, DryLoc®, plastic bulb, 3 KΩ, Wet-Tap
3-2757-WTP 159 001 391 Electrode, ORP, DryLoc
3-2751-1 159 001 804 DryLoc pH/ORP Sensor Electronics, In-line
3-2751-2 159 001 805 DryLoc pH/ORP Sensor Electronics, In-line with EasyCal
3-2760-11 159 001 367 In-line Preamplifi er with ¾ in. NPT threads and 4.6 m (15 ft) cable
3-2760-21 159 001 368 In-line Preamplifi er with ¾ in. ISO threads and 4.6 m (15 ft) cable
3-3719.390 159 000 855 3719 Locking Shroud
3-3719.392 159 310 304 O-ring kit
1
/2 in. NPT
®
, plastic bulb, Pt1000, Wet-Tap
®
, plastic bulb, 10 KΩ ID, Wet-Tap
3-3719.390 Locking shroud
3-2756-WTP DryLoc pH electrode 3-2756-WTP-1 DryLoc pH electrode 3-2757-WTP DryLoc ORP electrode
3-3719.392
O-ring Kit
Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090 For the most up-to-date information, please refer to our website at www.gfsignet.com