Georg Fischer 2850 Operating Manual

Page 1
Signet 2850 Conductivity/Resistivity Sensor Electronics (PVDF)
3-2850.090-2 Rev. C 03/15
Operating Instructions
English
3-2850-51, 3-2850-52
with ¾-in. adapter
3-2850-61, 3-2850-62, 3-2850-63
with Universal adapter
3-2850-51-XXV(D), 3-2850-52-XXV(D)
PVDF Integral System
Description
• Signet 2850 Conductivity/Resistivity Sensor Electronics provides either a two-wire 4 to 20 mA output or Digital (S3L) format.
• The 4 to 20 mA output models provide eight ranges for each electrode cell constant, plus the ability to invert each range.
• The EasyCal feature allows the devices to automatically recognize standard conductivity test solution values for simple fi eld calibration.
• The Conductivity Sensor provided with integral systems will have its custom cell constant information programmed into the electronics at the factory to provide a 2% sensor accuracy. See page 6 for details.
Table of Contents
Description ............................................................................................ 1
Dimensions ........................................................................................... 1
Warranty Information ............................................................................ 2
Product Registration ............................................................................. 2
Safety Information................................................................................. 2
Specifi cations........................................................................................ 2
Conductivity Sensor Dimensions .......................................................... 3
Operating Range Chart......................................................................... 3
In-Line Installation................................................................................. 4
Tank Installation .................................................................................... 4
4 to 20 mA Wiring ................................................................................. 5
Digital (S3L) Wiring ................................................................................ 5
Dual Input Wiring (Dual Digital (S3L) Output) ........................................ 5
Cell Constant Selection ........................................................................ 6
Range Selection for 4 to 20 mA Output ................................................ 7
Calibration............................................................................................. 8
EasyCal ................................................................................................ 8
Dual Input Calibration ........................................................................... 8
Maintenance ......................................................................................... 9
Troubleshooting .................................................................................... 9
Electronic Certifi cation ........................................................................ 10
Ordering Information ........................................................................... 12
English
Deutsch
Français
Español
Italiano
Dimensions
2850-51, -52 Integral mount 2850-61, -62, -63 Universal mount
95 mm
3.74 in.
85 mm
3.34 in.
95 mm
3.74 in.
82 mm
3.24 in.
Page 2
Warranty Information
Specifi cations
Refer to your local Georg Fischer Sales offi ce for the most current warranty statement.
All warranty and non-warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales offi ce or distributor. Product returned without a Service Form may not be warranty replaced or repaired.
Signet products with limited shelf-life (e.g. pH, ORP, chlorine electrodes, calibration solutions; e.g. pH buffers, turbidity standards or other solutions) are warranted out of box but not warranted against any damage, due to process or application failures (e.g. high temperature, mishandling (e.g. broken glass,
chemical poisoning, dry-out) or
damaged membrane, freezing
and/or extreme temperatures).
Product Registration
Thank you for purchasing the Signet line of Georg Fischer measurement products.
If you would like to register your product(s), you can now register online in one of the following ways:
• Visit our website www.gfsignet.com. Under Service and Support click on
Product Registration Form
• If this is a pdf manual (digital copy),
click here
Safety Information
1. Depressurize and vent system prior to installation or removal.
2. Confi rm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifi cations.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents, care should be taken and appropriate eye, face, hand, body, and/or respiratory protection should be used.
Caution / Warning / Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death
Electrostatic Discharge (ESD) / Electrocution Danger
Alerts user to risk of potential damage to product by ESD, and/or risk of potential of injury or death via electrocution.
Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during installation and service of Signet products.
Pressurized System Warning
Sensor may be under pressure, take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury.
Note / Technical Notes
Highlights additional information or detailed procedure.
General
Compatible Electrodes ............. Any Signet Conductivity/
Resisitivity electrode
(2818-2823, 2839-2842)
Temperature Compensation .....PT-1000 RTD (2% per °C)
Pure Water Compensation ....... Auto-switching when using
0.01 cell and raw conductivity
value < 0.5 S
EasyCal
Automatic recognition of the following conductivity values:
• 146.93 S, 1408.8 S, 12856 S (@ 25 °C) (Test solutions per ASTM D1125-95)
• 10 S, 100 S, 200 S, 500 S, 1000 S, 5000 S, 10,000 S,
50,000 S, 100,000 S (@ 25 °C)
Power
• 12 to 24 VDC (4 to 20 mA output)
• 5 VDC ±5% regulated, 3.0 mA max. for Digital (S
3
L) output
(Reverse polarity and short circuit protected)
Digital (S
3
L) Output ................. Serial ASCII, TTL level 9600 bps
Accuracy
• Conductivity ..........................±2% of reading
• Temperature ..........................±0.5 °C
Resolution
• Conductivity ..........................0.1% of reading
• Temperature ..........................< 0.2 °C
Update Rate
• Single Input models ..............< 600 ms
• Dual Input models .................< 1.2 s
Current Output
Field-selectable ranges. Factory set span:
• 0.01 cell
(2818, 2819, 2839)
.. 4 to 20 mA = 0 to 100 S
• 0.10 cell (2820, 2840) ........... 4 to 20 mA = 0 to 1000 S
• 1.0 cell (2821, 2841) ............. 4 to 20 mA = 0 to 10,000 S
• 10.0 cell (2822, 2842) ........... 4 to 20 mA = 0 to 200,000 S
• 20.0 cell (2823) ..................... 4 to 20 mA = 0 to 400,000 S
Max. Loop Resistance .............. 50 @ 12 VDC
325 @ 18 VDC
600 @ 24 VDC
Accuracy ................................... ±2% of output span
Resolution.................................7 A
Update Rate .............................<600 ms
Error Indication ......................... 22 mA
3
Available data via Digital (S
L) Output
• Raw conductivity
• Calibrated conductivity
• Calibrated temperature-compensated conductivity
• Temp. Error Indication ........... Open input and out-of-range
diagnostics for temperature or
internal electronic error.
Environmental
Material ..................................... PBT
Enclosure Rating ...................... NEMA 4X/IP65
Operating Temperature .............-10 to 85 °C (14 to 185 °F)
Storage Temperature ................-20 to 85 °C (-4 to 185 °F)
Relative Humidity......................0 to 95%, non-condensing
Shipping Weight .......................0.75 kg (1.75 lb.)
Standards and Approvals
• CE, RoHS Compliant
• Manufactured under ISO 9001 for Quality,
ISO 14001 for Environmental Management and OHSAS 18001 for Occupational Health and Safety.
China RoHS (Go to www.gfsignet.com for details) This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received,
including interference that may cause undesired operation.
2
2850 Conductivity/Resistivity Sensor Electronics
Page 3
Conductivity Sensor Dimensions
73.7 mm
2.90 in.
32.39 mm (1.28 in.)
3-2839
3-2818
59.3 mm
2.33 in.
73.66 mm (2.9 in.)
41.91 mm (1.65 in.)
12.7 mm (0.5 in.)
35.8 mm
1.41 in.
3-2840
4.6 m (15 ft) cable (std)
152 mm (6 in.)
12.7 mm (0.5 in.)
3-2819, 3-2820, 3-2821
Operating Range Chart
107 mm (4.2 in.)
49.5 mm (1.95 in.)
21.5 mm
0.85 in.
41.9 mm
1.65 in.
¾ in. NPT
2820-x
20.3 mm (0.8 in.)
2821-x
7.6 mm (0.3 in.)
3-2841, 3-2842
26.8 mm
1.13 in.
T1 = 50 mm (2.0 in.) T2 = 64 mm (2.5 in.)
152 mm (6.0 in.)
126 mm (5.0 in.)
3-2819, 3-2820, 3-2821
Sanitary
13 mm (0.5 in.)
203 mm
8.0 in.
3-2823
89 mm
3.5 in.
(0.01 Cell)
0.010 μS
0
UPW @ 25°C
(0.10 Cell)
2818/2819/2839
0
USP
10 μS
.
2821/2841
(1.0 Cell)
2820/2840
1 μS
0
0.5 μS 20 μS
Bottled
(20.0 Cell)
(10.0 Cell)
100 μS
50 μS
.
Water
2823
2822/2842
1,000 μS 10,000 μS
Cooling Tower Waste Waters
Deionization Regen. Chemicals
Alkali Cleaners, Acids/Bases
50,000 μS
200,000 μS
400,000 μS 200 μS 100,000 μS 0.055 μS
2850 Conductivity/Resistivity Sensor Electronics
3
Page 4
In-Line Installation Tank Installation
While the 2850-5X and 2850-6X electronics cannot be submerged, either model will accommodate tank installation.
Select any electrode with a 5 m (15 ft) cable. The cable may be cut to length, but it CANNOT BE EXTENDED.
Most of the Conductivity/Resistivity electrodes used with the 2850 can be installed into standard 3/4 in. (NPT or ISO 7-R3/4) fi ttings.
The 2850-5X models are designed for integral systems, where the electronics are attached directly to the sensor.
The preferred installation for in-line applications directs fl ow straight into the electrode. This confi guration reduces the probability of entrapped air bubbles, and provides the best continuous sampling of the fl uid content.
The universal adapter included with 2850-6X models can be attached to the top of a tank or mounted to a surface near the tank.
Adapt the 2850-5X models to tank installations by attaching a clamp to the conduit.
If the electrode is mounted vertically in a tee, do not recess the electrode orifi ces inside the tee. Mounting upside down may help prevent air entrapment.
An oversized tee or fl ow cell may be helpful for inline installations.
At least 4 threads (ANSI B1.20.1) must be engaged to meet the pressure rating as published in the specifi cations.
Use the 2850-6X models for in-line installations where the electronics must be separated from the electrode.
Conductivity electrodes with sanitary fl ange fi ttings must be installed using this option.
4
2850 Conductivity/Resistivity Sensor Electronics
Page 5
4 to 20 mA Wiring
Maximum length of 4 to 20 mA loop is 300 meters (1000 ft)
4
CTS
3 2
ON
1
3
V-
2 1
V+
SW1
4
CTS
3 2
ON
1
D3
to
9 VDC
4 20
4-20 mA Loop Output
26 VDC
Sensr Gnd (SHIELD)
Temp. IN (WHITE)
Signal RTN (BLK)
Signal IN (RED)
Digital (S3L) Wiring
The Digital (S Signet 8900 Multi-Parameter Controller and the Signet 9900 Transmitter.
Refer to the wiring sections of the 8900 or 9900 manual to determine cable length limitations.
IMPORTANT: Provide Earth Ground as
Note: The 2819-2823 and 2839-2842 sensors can
3
L) output is compatible with the
indicated by symbol.
connect directly to the 9900 via the 9900's optional Direct Conductivity/Resistivity
I/O Module 3-8900.401-X
+5VDC (Black)
Frequency
Input
Freq. Input (Red)
1
GND (Shield)
+5VDC (Black)
Frequency
Input 2
Freq. Input 2 (Red)
Input
Input
S3L
OR S3L
2
1
3
S L (Red)
GND (White/Shield)
+5VDC (Black)
3
S L (Red)
GND (White/Shield)
Analog Output 1
(if applicable)
Analog Output 2
(if applicable)
9900 S3L Input
3-8900.621C
1
2
3
4
5
6
7
8
9
10
11
+
12
-
13
+
14
-
SHLD
DATA
GND
V+
Module 3-9900.394 (159 001 699).
White Red Black
Dual Digital (S3L) Input Wiring
Dual Input is available on Digital (S3L) output model 3-2850-63 only, for use with the Signet 8900 Multi-Parameter Controller.
Sensors may be the same or different cell constants.
If the Dual Input model is used with a single sensor, connect the sensor to Input 1 and set all of the Input 2 switches to OFF (OPEN).
I/O Module 3-8900.401-X
+5VDC (Black)
Frequency
Input
Freq. Input (Red)
1
GND (Shield)
+5VDC (Black)
Frequency
Input 2
Freq. Input 2 (Red)
OR S3L
Input
2
S3L
Input
1
3
S L (Red)
GND (White/Shield)
+5VDC (Black)
3
S L (Red)
GND (White/Shield)
Analog Output 1
(if applicable)
Analog Output 2
(if applicable)
3-8900.621C
1
2
3
4
5
6
7
8
9
10
11
+
12
-
13
+
14
-
Dual Input applications usually include one sensor with a smaller cell constant and one sensor with a larger cell constant. Place the larger cell constant on Input 1.
This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN).
Gnd
3
Data
2 1
V+
3
Gnd
2
Data
1
V+
SW1
SW1
CTS
ON
4 3 2
1
D3
3-9900.395
H COMM Module
CTS
ON
CTS
ON
4 3
Input 2
2
1
4 3
Input 1
2
1
D3
DC Power
Loop Voltage
Sensr Gnd (SHIELD)
Temp. IN (WHITE)
Signal RTN (BLK)
Signal IN (RED)
Sensor 2
Sensor 1
IMPORTANT: Provide Earth Ground as indicated by
symbol.
2850 Conductivity/Resistivity Sensor Electronics
5
Page 6
Cell Constant Selection
S 1
2
3
4
ON
CTS
S 1
1
2
3ONCTS
Single Input 2850 Electronics (2850-51, 2850-52, 2850-61, 2850-62)
• Single input models use only SW3.
• Use SW3 #1-3 to select the cell constant for the rst sensor.
• Use SW3 #4 to disable the PT1000 Temp Compensation function in the 2850 (as required for USP applications).
• Recommended:
D3
W
ON
1
2
SW3
Input 1
Make all switch settings before supplying power. Switch changes made after supplying power will take 15 to 20 seconds before becoming effective.
Integral 2850 Systems (Single Input)
• The 2850-51-XX and 2850-52-XX are shipped from the factory with the custom cell constant and temperature offset programmed into the single-input electronics.
• SW3 will be preset to "Custom cell constant".
• No additional modifi cations are required.
Replacing Integral System Electrode or Electronics
• When installing a new conductivity electrode or replacing System Electronics, the SW3 switches must be confi gured to select the appropriate cell constant.
• See Single Input section above.
• Leaving SW3 in the custom cell constant confi guration will effect accuracy.
• Use calibration standards to calibrate (see page 8).
Dual Input 3-2850-63 Electronics
• Switch banks SW3 and SW4 are used to select the cell constant of the electrodes.
• Use SW3 #1-3 to select the cell constant for the rst sensor.
• Use SW3 #4 to disable the PT1000 Temp Compensation function in the 2850 (as required for USP applications). This disables the function for BOTH INPUTS.
• Use SW4 #1-3 to select the cell constant for the second sensor.
W
D3
ON
1
2
SW3
Input 1 Input 2
• Set SW4 #4 to OPEN to disable Input 2.
• Recommended: Make all switch settings before supplying power. Switch changes made after supplying power will take 15 to 20 seconds before becoming effective.
NOTE: Dual Input applications usually include one sensor with a smaller cell constant and one
sensor with a larger cell constant. Place the larger cell constant on Input 1. This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN).
2
3
1
CTS
4
3
2
3
1
CTS
CTS
ON
4
4
1
3
3
2
SW4
NOTE:
To work correctly with the 9900, the 2850 must be set for the custom cell constant or the actual electrode cell constant and the 9900 set for a 1.0 cell constant.
SW3 or SW4
Custom Cell Constant
ON
1
0.01/cm
ON
1
0.1/cm
ON
1
1.0/cm
ON
1
10.0/cm
ON
1
20.0/cm
ON
1
Examples
CTS
Closed Open
4
3
2
-1
CTS
Closed Open
4
3
2
-1
CTS
Closed Open
4
3
2
-1
CTS
Closed Open
4
3
2
-1
CTS
Closed Open
4
3
2
-1
CTS
Closed Open
4
3
2
1.0 cell constant,
SW3 = Cell Constant and
Temperature Compensation
#1 #2 #3 #4 Cell Constant
C C C O Custom C O O O 0.01 cm O C O O 0.1 cm C C O O 1.0 cm O O C O 10.0 cm C O C O 20.0 cm
-1
-1
-1
-1
-1
SW4 = Cell Constant and Input 2
#1 #2 #3 #4 Cell Constant
C C C O Custom C O O O 0.01 cm O C O O 0.1 cm C C O O 1.0 cm O O C O 10.0 cm C O C O 20.0 cm
-1
-1
-1
-1
-1
temp comp Active
ON
1
CTS
4
3
2
1.0 cell constant,
temp comp Inactive
ON
CTS
Closed Open
Closed
#4: (O) Open = Temp Comp ACTIVE (C) Closed = Temp Comp INACTIVE
6
2850 Conductivity/Resistivity Sensor Electronics
#4: (O) Open = Input 2 INACTIVE (C) Closed = Input 2 ACTIVE
Open
1
4
3
2
Page 7
Range Selection for 4 to 20 mA Output
• The Range selection switch bank (SW2) provides eight range selections for each cell constant.
• Each range can be inverted, making a total of 16 range options.
• Select a range from the table below and set SW2 as indicated.
Example (refer to shaded selections of chart):
• The electrode installed is the 3-2840-1V, with a 0.10 cell constant.
• The required output is 4 to 20 mA = 0 to 100 S.
• Set SW2 switch bank to C-C-O-O.
• To invert range (4 to 20 mA = 100 to 0 S), close switch #4.
CTS
ON
2
SW2
4
3
SW1
D3
1 3 2
CTS
ON
4
1
3
2
SW2
SW3
CTS
ON
4
1
3
2
1
SW2 Switch Setting 0.01 Cell
0.10 Cell
1.0 Cell 10.0 Cell 20.0 Cell
#1 #2 #3 #4 Resistivity Ranges in BOLD
CCCO10 to 20 MΩ 0 to 2 S 0 to 20 S 0 to 200 S 0 to 400 S
CCCC20 to 10 MΩ 2 to 0 S 20 to 0 S 200 to 0 S 400 to 0 S
OCCO 2 to 10 MΩ 0 to 5 S 0 to 50 S 0 to 500 S 0 to 1000 S
OCCC 10 to 2 MΩ 5 to 0 S 50 to 0 S 500 to 0 S 1 000 to 0 S
COCO 0 to 2 MΩ 0 to 10 S 0 to 100 S 0 to 1 000 S 0 to 2 000 S
COCC 2 to 0 MΩ 10 to 0 S 100 to 0 S 1 000 to 0 S 2 000 to 0 S
O O C O 0 to 1 S 0 to 50 S 0 to 500 S 0 to 5 000 S 0 to 10 000 S
O O C C 1 to 0 S 50 to 0 S 500 to 0 S 5 000 to 0 S 10 000 to 0 S
CCOO
0 to 5 S
0 to 100 S
0 to 1 000 S 0 to 10 000 S 0 to 20 000 S
C C O C 5 to 0 S 100 to 0 S 1 000 to 0 S 10 000 to 0 S 20 000 to 0 S
O C O O 0 to 10 S 0 to 200 S 0 to 2 000 S 0 to 50 000 S 0 to 100 000 S
OCOC 10 to 0 S 200 to 0 S 2 000 to 0 S 50 000 to 0 S 100 000 to 0 S
C O O O 0 to 50 S 0 to 500 S 0 to 5 000 S 0 to 100 000 S 0 to 200 000 S
C O O C 50 to 0 S 500 to 0 S 5 000 to 0 S 100 000 to 0 S 200 000 to 0 S
OOOO 0 to 100 S 0 to 1 000 S 0 to 10 000 S 0 to 200 000 S 0 to 400 000 S
O O O C 100 to 0 S 1000 to 0 S 10 000 to 0 S 200 000 to 0 S 400 000 to 0 S
Switch #4 inverts the loop output rangeNOTE:
(O) Open = 4 to 20 mA (C) Closed = 20 to 4 mA
2850 Conductivity/Resistivity Sensor Electronics
7
Page 8
Calibration
• All 2850 Conductivity/Resistivity Sensor Electronics are factory calibrated for maximum out-of-the-box accuracy.
• The EasyCal feature allows calibration to be performed at the sensor.
The 2839-2842 Conductivity sensors come with a certifi cate of calibration. This certifi cate provides information regarding the actual cell constant and temperature offset as tested and verifi ed according to NIST standards. This information has been programmed into the 2850 electronics as its default factory value. If required, the sensor can be recertifi ed by contacting the GF Signet repair department.
a. SW2: 4 to 20 mA output range selection switch bank b. Power and Output terminal block: 4 to 20 mA or
Digital (S
3
L) Outputs
c. SW3 and SW4: Electrode Cell Constant selection
switch banks d. D3: LED indicator e. SW1: EasyCal Push-button f. Input 1 & Input 2: Conductivity Electrode connections
a
b
c
3 2 1
SW2
CTS
ON
4 3 2
1
4
CTS
SW4
3
Input 2
2
ON
1
4
CTS
SW3
3 2
Input 1
ON
1
f
NOTE:
When the electrode is changed from its original factory provided electrode or the application requires a cell constant
d
SW1
e
D3
change, reset SW3 or SW4 switches (see page 6). Then perform a wet calibration with calibration standards.
EasyCal
EasyCal is a single-point calibration system. During this procedure, if the measured value is within ±10% of any of the test
values listed below, the 2850 will automatically recognize the test value and calibrate the output to that value.
NOTE: The fi rst step (Reset) is recommended each time an electrode is replaced, but is NOT necessary upon initial installation or
periodic calibration.
1. Reset the 2850 Sensor to factory calibration: Set all switches on SW3 (and SW4 for Dual Input) to OPEN. Apply power, wait
at least 3 seconds, then press and hold SW1 until the LED (D3) comes on steady then goes off again (approx. 10 seconds).
When the LED goes off, release SW1; reset is complete. Reset SW3 and SW4 to the proper settings.
2. Place the electrode/sensor assembly into the conductivity test solution appropriate to your operating range. Shake the
electrode to dislodge any air bubbles visible on the surface of the electrode.
• 146.93 S, 1408.8 S, 12856 S (per ASTM D1125-95)
• 10 S, 100 S, 200 S, 500 S, 1000 S, 5000 S, 10,000 S, 50,000 S, 100,000 S.
Allow at least 2 minutes for the electrode response to stabilize.
3. Press and hold SW1 for approximately 8 to 10 seconds. During this time, the LED will come on steady then go back off.
(If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section).
Calibration is complete. Return the system to service.
Dual Input Calibration
Dual Input 3-2850-63 Electronics
• The 2850-63 can only be calibrated if both inputs are used, or if Input 2 is disabled.
• To disable Input 2: Set SW4, #4 to OPEN.
• Place each sensor into a test solution that represents the operating range of that sensor. When SW1 is pressed, the 2850 will
calibrate both inputs simultaneously.
• Dual Input applications usually include one sensor with a smaller cell constant and one sensor with a larger cell constant.
Place the larger cell constant on Input 1. This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN).
8
2850 Conductivity/Resistivity Sensor Electronics
Page 9
Maintenance
The 2850 requires no periodic maintenance.
• Coatings on the electrode may cause slow response or drift.
• Clean metallic surfaces with a mild detergent and a non-abrasive brush or cotton swab.
Troubleshooting
LED and Output Condition Possible Causes Suggested Solutions
Current Out: LED off, current output is 22 mA
(S3L) out: "Check Sensor"
During EasyCal, the LED blinks rapidly for 4 seconds.
After completing calibration procedure, the output values are inaccurate.
The output of the 2850 indicates a value that is off by a factor of 10 (e.g. 10x, 100x the correct value)
• SW3 and SW4 set to all OPEN (factory setting).
• No electrode installed.
• SW3 (and SW4) not set correctly.
• System not ready.
• Measured value of the test solution is outside the ±10% tolerance.
• The electronics or the electrode is defective.
• Insuffi cient time allowed for electrode stabilization during calibration procedure.
• Test solutions are contaminated.
• EasyCal performed with temp comp Inactive, then temp comp Active for measurement. (or vice versa)
• The cell constant selection on the 2850 (SW3 or SW4) does not match the actual input.
• Normal for new system.
• Install electrode.
• Set SW3 and SW4 correctly.
• Retry (recycle power).
• Use fresh test solutions and restart the calibration.
• Replace the electronics or the electrode.
• Recalibrate and wait at least 2 minutes after placing electrode in solution before pressing SW1.
• Use fresh test solutions and restart the calibration.
• Match the 2850 switch settings to agree with the sensor being used.
2850 Conductivity/Resistivity Sensor Electronics
9
Page 10
Electronic Certifi cation
Signet offers conductivity simulators in fi ve different values. These tools enable the user to validate the performance of the electronics independently of the electrode. This requirement is defi ned by ASTM D 1125-95 (Standard Test Methods for Electrical Conductivity and Resistivity of Water) which is commonly used for USP 24 applications.
Cond. - Resist.
Simulator
@ 25°C
For use with Signet
2850 Conductivity
NIST Traceable
3-2850.101-1 (159 001 392) 1.0 S ±0.1%
3-2850.101-2 (159 001 393) 2.5 S ±0.1%
3-2850.101-3 (159 001 394) 10.0 S ±0.1%
3-2850.101-4 (159 001 395) 18.2 M ±0.1%
3-2850.101-5 (159 001 396) 10.0 M ±0.1%
Before using the Simulator:
Before using the simulators, disconnect all electrodes and reset the 2850 Electronics to factory calibration:
1. Set all switches on SW3 (and SW4 for Dual Input) to OPEN. Turn the power on, wait at least 3 seconds
2. Press and hold SW1 until the LED (D3) comes on steady then goes off again (approx. 10 seconds).
3. When the LED goes off, release SW1. Reset is complete.
To validate Input 1:
1. Turn the power off.
2. Set SW3 to accept a 0.01 cell constant. Leave SW4 in the all-OPEN position.
3. Install the simulator into the 2850 at the Input 1 conductivity electrode terminals.
4. Connect the 2850 output terminals to the appropriate display instrument. Turn the power on.
• The display should indicate a conductivity value within ±2% of the simulator value.
• If the output is x10 or x100 of the correct value, recheck step 2. The 2850 is not set to accept a 0.01 cell constant.
• If the correct value is displayed, the electronics are calibrated to ASTM D 1125 specifi cations. Remove the simulator and reset SW3 to the correct cell constant before reinstalling the electrode.
• If the correct value is not displayed, contact the factory.
To validate Input 2:
1. Turn the power off.
2. Set SW4 to accept a 0.01 cell constant.
3. Install the simulator into the 2850 at the Input 2 conductivity electrode terminals.
4. Set SW3 switches to the all-OPEN position.
5. Turn the power on and repeat step 4.
Remember to reset all switches to their correct position before resuming normal operation.
10
2850 Conductivity/Resistivity Sensor Electronics
Page 11
Notes
2850 Conductivity/Resistivity Sensor Electronics
11
Page 12
Ordering Information
Signet 2850 Conductivity/Resistivity Sensor Electronics
*All versions include EasyCal
Mfr. Part No. Code Description
3-2850-51 159 001 398 2850 Sensor Electronics with Digital (S
3
L) Output and ¾ inch adapter
3-2850-52 159 001 399 2850 Sensor Electronics with 4 to 20 mA Output and ¾ inch adapter
3-2850-61 159 001 400 2850 Sensor Electronics with Digital (S
3
L) Output and Universal adapter
3-2850-62 159 001 401 2850 Sensor Electronics with 4 to 20 mA Output and Universal adapter
3-2850-63 159 001 402 2850 Sensor Electronics with Dual Inputs, Dual Digital (S3L) Outputs and Universal adapter
(for use with 8900 only)
3
3-2850-51-39V 159 001 818 Integral 2850 system, Digital (S
L) output, 0.01 cell, PVDF NPT threads
3-2850-51-40V 159 001 819 Integral 2850 system, Digital (S3L) output, 0.1 cell, PVDF NPT threads
3-2850-51-41V 159 001 820 Integral 2850 system, Digital (S3L) output, 1.0 cell, PVDF NPT threads
3-2850-51-42V 159 001 821 Integral 2850 system, Digital (S3L) output, 10.0 cell, PVDF NPT threads
3-2850-51-39VD 159 001 822 Integral 2850 system, Digital (S
3
L) output, 0.01 cell, PVDF ISO threads
3-2850-51-40VD 159 001 823 Integral 2850 system, Digital (S3L) output, 0.1 cell, PVDF ISO threads
3-2850-51-41VD 159 001 824 Integral 2850 system, Digital (S3L) output, 1.0 cell, PVDF ISO threads
3-2850-51-42VD 159 001 825 Integral 2850 system, Digital (S3L) output, 10.0 cell, PVDF ISO threads
3-2850-52-39V 159 001 826 Integral 2850 system, 4 to 20 mA output, 0.01 cell, PVDF NPT threads
3-2850-52-40V 159 001 827 Integral 2850 system, 4 to 20 mA output, 0.1 cell, PVDF NPT threads
3-2850-52-41V 159 001 828 Integral 2850 system, 4 to 20 mA output, 1.0 cell, PVDF NPT threads
3-2850-52-42V 159 001 829 Integral 2850 system, 4 to 20 mA output, 10.0 cell, PVDF NPT threads
3-2850-52-39VD 159 001 830 Integral 2850 system, 4 to 20 mA output, 0.01 cell, PVDF ISO threads
3-2850-52-40VD 159 001 831 Integral 2850 system, 4 to 20 mA output, 0.1 cell, PVDF ISO threads
3-2850-52-41VD 159 001 832 Integral 2850 system, 4 to 20 mA output, 1.0 cell, PVDF ISO threads
3-2850-52-42VD 159 001 833 Integral 2850 system, 4 to 20 mA output, 10.0 cell, PVDF ISO threads
Parts and Accessories
3-9000.392-1 159 000 839 Liquid-tight connector kit, 1 set,
1
/2 in. NPT
3-9000.392-2 159 000 841 Liquid-tight connector kit, 1 set, PG 13.5
3-2850.101-1 159 001 392 Plug-in NIST-traceable recertifi cation tool, 1.0 S
3-2850.101-2 159 001 393 Plug-in NIST-traceable recertifi cation tool, 2.5 S
3-2850.101-3 159 001 394 Plug-in NIST-traceable recertifi cation tool, 10.0 S
3-2850.101-4 159 001 395 Plug-in NIST-traceable recertifi cation tool, 18.2 M
3-2850.101-5 159 001 396 Plug-in NIST-traceable recertifi cation tool, 10.0 M
Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090 For the most up-to-date information, please refer to our website at www.gfsignet.com
3-2850.090-2 Rev. C 03/15 English © Georg Fischer Signet LLC 2015