Georg Fischer 2850 Operating Manual

Signet 2850 Conductivity/Resistivity Sensor Electronics (PVDF)
3-2850.090-2 Rev. C 03/15
Operating Instructions
English
3-2850-51, 3-2850-52
with ¾-in. adapter
3-2850-61, 3-2850-62, 3-2850-63
with Universal adapter
3-2850-51-XXV(D), 3-2850-52-XXV(D)
PVDF Integral System
Description
• Signet 2850 Conductivity/Resistivity Sensor Electronics provides either a two-wire 4 to 20 mA output or Digital (S3L) format.
• The 4 to 20 mA output models provide eight ranges for each electrode cell constant, plus the ability to invert each range.
• The EasyCal feature allows the devices to automatically recognize standard conductivity test solution values for simple fi eld calibration.
• The Conductivity Sensor provided with integral systems will have its custom cell constant information programmed into the electronics at the factory to provide a 2% sensor accuracy. See page 6 for details.
Table of Contents
Description ............................................................................................ 1
Dimensions ........................................................................................... 1
Warranty Information ............................................................................ 2
Product Registration ............................................................................. 2
Safety Information................................................................................. 2
Specifi cations........................................................................................ 2
Conductivity Sensor Dimensions .......................................................... 3
Operating Range Chart......................................................................... 3
In-Line Installation................................................................................. 4
Tank Installation .................................................................................... 4
4 to 20 mA Wiring ................................................................................. 5
Digital (S3L) Wiring ................................................................................ 5
Dual Input Wiring (Dual Digital (S3L) Output) ........................................ 5
Cell Constant Selection ........................................................................ 6
Range Selection for 4 to 20 mA Output ................................................ 7
Calibration............................................................................................. 8
EasyCal ................................................................................................ 8
Dual Input Calibration ........................................................................... 8
Maintenance ......................................................................................... 9
Troubleshooting .................................................................................... 9
Electronic Certifi cation ........................................................................ 10
Ordering Information ........................................................................... 12
English
Deutsch
Français
Español
Italiano
Dimensions
2850-51, -52 Integral mount 2850-61, -62, -63 Universal mount
95 mm
3.74 in.
85 mm
3.34 in.
95 mm
3.74 in.
82 mm
3.24 in.
Warranty Information
Specifi cations
Refer to your local Georg Fischer Sales offi ce for the most current warranty statement.
All warranty and non-warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales offi ce or distributor. Product returned without a Service Form may not be warranty replaced or repaired.
Signet products with limited shelf-life (e.g. pH, ORP, chlorine electrodes, calibration solutions; e.g. pH buffers, turbidity standards or other solutions) are warranted out of box but not warranted against any damage, due to process or application failures (e.g. high temperature, mishandling (e.g. broken glass,
chemical poisoning, dry-out) or
damaged membrane, freezing
and/or extreme temperatures).
Product Registration
Thank you for purchasing the Signet line of Georg Fischer measurement products.
If you would like to register your product(s), you can now register online in one of the following ways:
• Visit our website www.gfsignet.com. Under Service and Support click on
Product Registration Form
• If this is a pdf manual (digital copy),
click here
Safety Information
1. Depressurize and vent system prior to installation or removal.
2. Confi rm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure specifi cations.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents, care should be taken and appropriate eye, face, hand, body, and/or respiratory protection should be used.
Caution / Warning / Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death
Electrostatic Discharge (ESD) / Electrocution Danger
Alerts user to risk of potential damage to product by ESD, and/or risk of potential of injury or death via electrocution.
Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during installation and service of Signet products.
Pressurized System Warning
Sensor may be under pressure, take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury.
Note / Technical Notes
Highlights additional information or detailed procedure.
General
Compatible Electrodes ............. Any Signet Conductivity/
Resisitivity electrode
(2818-2823, 2839-2842)
Temperature Compensation .....PT-1000 RTD (2% per °C)
Pure Water Compensation ....... Auto-switching when using
0.01 cell and raw conductivity
value < 0.5 S
EasyCal
Automatic recognition of the following conductivity values:
• 146.93 S, 1408.8 S, 12856 S (@ 25 °C) (Test solutions per ASTM D1125-95)
• 10 S, 100 S, 200 S, 500 S, 1000 S, 5000 S, 10,000 S,
50,000 S, 100,000 S (@ 25 °C)
Power
• 12 to 24 VDC (4 to 20 mA output)
• 5 VDC ±5% regulated, 3.0 mA max. for Digital (S
3
L) output
(Reverse polarity and short circuit protected)
Digital (S
3
L) Output ................. Serial ASCII, TTL level 9600 bps
Accuracy
• Conductivity ..........................±2% of reading
• Temperature ..........................±0.5 °C
Resolution
• Conductivity ..........................0.1% of reading
• Temperature ..........................< 0.2 °C
Update Rate
• Single Input models ..............< 600 ms
• Dual Input models .................< 1.2 s
Current Output
Field-selectable ranges. Factory set span:
• 0.01 cell
(2818, 2819, 2839)
.. 4 to 20 mA = 0 to 100 S
• 0.10 cell (2820, 2840) ........... 4 to 20 mA = 0 to 1000 S
• 1.0 cell (2821, 2841) ............. 4 to 20 mA = 0 to 10,000 S
• 10.0 cell (2822, 2842) ........... 4 to 20 mA = 0 to 200,000 S
• 20.0 cell (2823) ..................... 4 to 20 mA = 0 to 400,000 S
Max. Loop Resistance .............. 50 @ 12 VDC
325 @ 18 VDC
600 @ 24 VDC
Accuracy ................................... ±2% of output span
Resolution.................................7 A
Update Rate .............................<600 ms
Error Indication ......................... 22 mA
3
Available data via Digital (S
L) Output
• Raw conductivity
• Calibrated conductivity
• Calibrated temperature-compensated conductivity
• Temp. Error Indication ........... Open input and out-of-range
diagnostics for temperature or
internal electronic error.
Environmental
Material ..................................... PBT
Enclosure Rating ...................... NEMA 4X/IP65
Operating Temperature .............-10 to 85 °C (14 to 185 °F)
Storage Temperature ................-20 to 85 °C (-4 to 185 °F)
Relative Humidity......................0 to 95%, non-condensing
Shipping Weight .......................0.75 kg (1.75 lb.)
Standards and Approvals
• CE, RoHS Compliant
• Manufactured under ISO 9001 for Quality,
ISO 14001 for Environmental Management and OHSAS 18001 for Occupational Health and Safety.
China RoHS (Go to www.gfsignet.com for details) This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received,
including interference that may cause undesired operation.
2
2850 Conductivity/Resistivity Sensor Electronics
Conductivity Sensor Dimensions
73.7 mm
2.90 in.
32.39 mm (1.28 in.)
3-2839
3-2818
59.3 mm
2.33 in.
73.66 mm (2.9 in.)
41.91 mm (1.65 in.)
12.7 mm (0.5 in.)
35.8 mm
1.41 in.
3-2840
4.6 m (15 ft) cable (std)
152 mm (6 in.)
12.7 mm (0.5 in.)
3-2819, 3-2820, 3-2821
Operating Range Chart
107 mm (4.2 in.)
49.5 mm (1.95 in.)
21.5 mm
0.85 in.
41.9 mm
1.65 in.
¾ in. NPT
2820-x
20.3 mm (0.8 in.)
2821-x
7.6 mm (0.3 in.)
3-2841, 3-2842
26.8 mm
1.13 in.
T1 = 50 mm (2.0 in.) T2 = 64 mm (2.5 in.)
152 mm (6.0 in.)
126 mm (5.0 in.)
3-2819, 3-2820, 3-2821
Sanitary
13 mm (0.5 in.)
203 mm
8.0 in.
3-2823
89 mm
3.5 in.
(0.01 Cell)
0.010 μS
0
UPW @ 25°C
(0.10 Cell)
2818/2819/2839
0
USP
10 μS
.
2821/2841
(1.0 Cell)
2820/2840
1 μS
0
0.5 μS 20 μS
Bottled
(20.0 Cell)
(10.0 Cell)
100 μS
50 μS
.
Water
2823
2822/2842
1,000 μS 10,000 μS
Cooling Tower Waste Waters
Deionization Regen. Chemicals
Alkali Cleaners, Acids/Bases
50,000 μS
200,000 μS
400,000 μS 200 μS 100,000 μS 0.055 μS
2850 Conductivity/Resistivity Sensor Electronics
3
In-Line Installation Tank Installation
While the 2850-5X and 2850-6X electronics cannot be submerged, either model will accommodate tank installation.
Select any electrode with a 5 m (15 ft) cable. The cable may be cut to length, but it CANNOT BE EXTENDED.
Most of the Conductivity/Resistivity electrodes used with the 2850 can be installed into standard 3/4 in. (NPT or ISO 7-R3/4) fi ttings.
The 2850-5X models are designed for integral systems, where the electronics are attached directly to the sensor.
The preferred installation for in-line applications directs fl ow straight into the electrode. This confi guration reduces the probability of entrapped air bubbles, and provides the best continuous sampling of the fl uid content.
The universal adapter included with 2850-6X models can be attached to the top of a tank or mounted to a surface near the tank.
Adapt the 2850-5X models to tank installations by attaching a clamp to the conduit.
If the electrode is mounted vertically in a tee, do not recess the electrode orifi ces inside the tee. Mounting upside down may help prevent air entrapment.
An oversized tee or fl ow cell may be helpful for inline installations.
At least 4 threads (ANSI B1.20.1) must be engaged to meet the pressure rating as published in the specifi cations.
Use the 2850-6X models for in-line installations where the electronics must be separated from the electrode.
Conductivity electrodes with sanitary fl ange fi ttings must be installed using this option.