Georg Fischer 2839-1V, 2840-1V, 2841-1V, 2842-1V Specifications

Signet 2839-1V(D) to 2842-1V(D) PVDF Conductivity Electrodes
Features
• ± 2% accuracy - Custom calibration certificate provided
• Dual-threaded
• Compact electrode length for easy in-line installation in small pipe sizes
• Triple orifice flow-through design
reduces clogging and bubble entrapment
• 316 SS electrodes with injection molded PVDF process connections and insulators
• Meets USP requirements
2839-1V 2840-1V
2841-1V 2842-1V
The Signet 2839-1V(D) to 2842-1V(D) Conductivity/ Resistivity Electrodes are available in four cell constants from 0.01 to 10.0 cm-1, and are suitable for a wide variety of applications from high purity water quality monitoring to weak acids and bases. 316 SS electrode surface finishes are controlled in a precision bead blasting operation to ensure measurement accuracy and repeatability.
The PVDF insulator and process connections are injection over-molded to minimize variance between electrodes. Double threaded connections in either ¾ in. NPT or ISO 7/1-R 3/4 enable quick and easy installation in submersible or in-line configurations. Transmitter integral mounting kit and junction boxes are available as accessories.
A Certificate of Calibration is included with all 28391-V(D) to 2842-1V(D) Conductivity/Resistivity Electrodes. The electrodes are calibrated to meet ± 2% accuracy. Electrodes can be shipped back to the GF Signet factory for recertification.
Applications
• Water Treatment & Water Quality Monitoring
• Reverse Osmosis
• Deionization
• Cooling Tower and Boiler Protection
• Distillation
• Desalination
• Demineralizer
• Semiconductor
• Aquatic Animal Life Support Systems
Specifications
General
Operating Range
2839 0.055 to 100 μS 0.02 to 50 ppm 2840 1 to 1,000 μS 0.5 to 500 ppm 2841 10 to 10,000 μS 5 to 5,000 ppm 2842 100 to 200,000 μS 50 to 100,000 ppm
Cell Constant Accuracy ± 2%. When the information provided on the certificate of calibration is entered into the
transmitter/controller. ± 5% when entered as a standard cell constant
Dual-Threaded Process Connection
Cable Length (use for the 2839, 40 ,41 and 42)
Temperature Element PT1000 Temp. Response,
Temperature Accuracy ±0.5 °C ±0.9 °F
Wetted Materials
Internal O-ring (2841 and 2842) FPM Insulator Material PVDF Electrode Material 316 SS Threaded Process Connection PVDF
Max. Temperature/Pressure Rating
Storage Temperature -20 °C to 131 °C -4 °F to 268 °F
Shipping Weight
2839 0.34 kg 0.74 lb 2840, 2841, 2842 0.30 kg 0.66 lb
Standards and Approvals
standard 4.6 m (15 ft) maximum 30 m (100 ft) all sensors when used with the 9900
0.01 cells 4.6 m (15 ft) used with 8850, 8860, and 2850*
0.01 cell 5 sec.
0.10 cell 10 sec.
1.0 cell 20 sec.
10.0 cell 30 sec.
-1V versions: ¾ in. NPT
-1VD versions: ISO 7/1-R 3/4
131 °C @ 2.76 bar 268 °F @ 40 psi
RoHS compliant China RoHS Manufactured under ISO 9001 for Quality and ISO 14001 for Environmental Management
and OHSAS 18001 for Occupational Health and Safety
18.2 MΩ to 10 KΩ 1 MΩ to 1 KΩ
* 2850 cable length 4.6 m (15 ft) maximum for all cells.
See Temperature and Pressure graphs for more information.