Georg Fischer 2819-2823, 2850 User Manual

CSA is cross sectional area.
*
www.gfsignet.com
Technical Reference Section: Conductivity/ Resistivity
Principle of operation
AC
+
Half Cells
Voltage
+
z
+
+
+
+
+
+
+
+
+
+
+
+
+
+
1cm
1cm
+
+
2
-1
Graphs
221
Most conductivity electrodes consist of two measuring half-cells. The geometry of the half-cells can be tai­lored to provide highly accurate mea­surements over a specific conduc­tivity range. Cell constants
help to
describe electrode geometry for the purpose of selecting the appropriate
x
electrode for a given application. A cell constant is defined as the length
y
between the two half-cells divided by the area of the cells.
Conductivity Cell Constant = =
Length
CSA*
z
xy
As an example, When x = y = z = 1cm the cell constant becomes = 1cm
Solutions of very low conductivity (high resistivity) such as ultra-pure water are best measured with half-cells that are very close together (i.e., cell constant =
-1
0.01cm
). Highly conductive solutions should be measured with half-cells that
are farther apart and that have relatively little cross sectional area between
-1
them (i.e., cell constant = 20.0cm
Temperature Compensation
The conductivity of a solution is highly dependent upon temperature. Therefore, conductivity measure­ments are almost always converted to an equivalent conductivity at the common reference temperature of 25°C (77°F). This is accomplished by means of temperature compensa­tion algorithms in the instruments, which require temperature as well as conductivity measurement input. To simplify and facilitate this require­ment all Signet conductivity elec­trodes contain high-quality tempera­ture sensing elements intelligently positioned for quick and accurate response.
).
siderably with the ionic composition of the solution and can range from less than 1% to more than 3% per °C. This is true of regional ground water sources as well as for other solutions such as brackish water, acids and bases. Signet instruments allow the entry of custom linear compensation coefficients for these applications. See the instruction manual of any Signet conductivity instrument for details. The conductivity or resistivity of pure water is not a linear function with respect to temperature. In fact, the latest Signet conductivity instru­ments utilize a sophisticated polyno­mial to compensate for the peculiar
effects. For seamless measurement Temperature effects on conductivity are more or less linear for normal water-based solutions, hovering around 2% per °C. However, the actual linear relationship varies con-
accuracy all current Signet conduc-
tivity instruments switch automati-
cally between linear and pure-water
compensation as certain measure-
ment thresholds are crossed. Temperature Compensation Exception
One exception to the requirement for temperature compensation has been estab­lished by USP (United States Pharmacopeia), which prescribes limits of accept­ability for ultra-pure water quality based upon non-compensated measurements. This methodology is used to eliminate measurement variances that may result from differences in the pure-water temperature compensation algorithms used by different manufacturers of conductivity measurement equipment. A more thor­ough treatment of the USP standard and instrument functionality can be found in the instruction manuals of the following Signet conductivity instruments: Model 8900 Multi-Channel, Multi-Parameter Controller (Appendix D), model 8860 Dual Channel Conductivity/Resistivity Controller.
Choosing
Products
Instrument
Multi-Pa-
rameter
Flow
pH/ORP
Conductivity/
Resistivity
Temperature,
Products
Other
Installation
Wiring &
Reference
Technical
Temperature/
Pressure
of Terms
Glossary
Reference
Part No.
Index
p
p
Relay Information
The two most common methods of controlling a process are “on/off” and “pro­portional” control. In on/off control, relay setpoints are defined as either high or low limits on the process variable. When the measurement value reaches a limit the relay is energized, typically for the purpose of opening a valve or start­ing a pump to introduce a chemical reagent to the process. This should cause the measurement value to change in the direction of the setpoint as shown in these on/off control diagrams:
High limit on/off relay control
= HI setpoint
= Hysteresis
= Relay energized
= Relay de-energized
pH
Notice the relay will not de-energize until the setpoint is exceeded by the hysteresis value. This is a program­mable value and is primarily used to prevent ”relay chatter”, which occurs if a relay is set to energize and de­energize at the same value. Because of hysteresis, and because reagent delivery is fairly constant while the relay is energized, a condition known as “overshoot” is inherent to the on/ off control method. Overshoot refers to the introduction of more chemical reagent than is absolutely necessary for achieving a desired adjustment to the process value, and can be expen­sive over time.
Proportional control is a popular alternative to the on/off control method. This method typically makes use of variable-rate metering pumps to reduce overshoot and improve precision. Establishing a propor­tional control scenario requires the selection of setpoint(s), deviation
Low limit on/off control
pH
= LO setpoint
= Hysteresis
= Relay energized
= Relay de-energized
range(s) and maximum pulse rates. The example shown here illustrates how two relays in ”pulse mode” can be used to proportionally control pH within a desired range, or to a single setpoint. This is called “Dual Propor­tional Control”. Of course, a single relay in proportional pulse mode can be used to establish a high or low limit and will also reduce overshoot.
Metering pumps are idle at and be­tween setpoints. When a setpoint is exceeded, the pump begins delivering reagent at a rate proportional to the difference between the measurement value and the setpoint. The larger the difference, the faster the deliv­ery. The programmed deviation value defines how quickly the maximum pulse rate is reached. Depending on the input requirements of the meter­ing pump, proportional control can also be accomplished with scaleable 4 to 20 mA outputs instead of pulsing relays or open collectors.
Dual proportional pulse relay control
Maximum Pulse Rate
0 pH
Deviation=
5.30 pH
LO
5.30 H
HI
7.50 H
Deviation=
3.50 pH
11.00 pH
222
14 pH
pH
Maximum Pulse Rate
Deviation=
5.20 pH
0
2.10 pH
LO and HI
7.30 pH
Deviation=
4.90 pH
12.20 pH
14 pH
www.gfsignet.com
Loading...
+ 4 hidden pages