• Durable DryLoc®
connector with gold
plated contacts
• Special design allows
for installation at any
angle, even inverted
• Quick temperature
response
Flat Glass
Description
The Signet 2774 - 2777 pH and ORP
Electrodes feature a unique foul-
®
proof DryLoc
connector with goldplated contacts designed specifically
for use with the Signet 2750 and 2760
preamplifiers, sensor electronics,
and connectors. These dependable
and highly responsive electrodes feature a PTFE double reference junction and a large reference chamber
that combine to extend the servicelife. Embedded positioning of the
temperature element in the pH/ORP
sensing tip allows the temperature
response to be quick and accurate.
The electrodes are offered with either
flat or bulb style sensing elements.
The flat versions allow sediment and
particles to sweep past the measurement surface, minimizing risks of
abrasion, breakage and coating. The
bulb versions can be used for general-purpose applications. Due to the
specially designed chambers which
keep electrolyte in place, all versions
can be installed at any angle, even
inverted.
Submersible Installation
Other Instruments
Customer supplied
Flow
Signet 2760
(sold separately)
Panel, Pipe, Tank, Wall Mount
678
RELAY 2
10
RELAY 1
pH
12
2
14
0
ENTER
+GF+
SIGNET
Signet or Other Instrument
5700 8750 8900
Pipe extension
or conduit with
3/4 in. FNPT
threads
(customer supplied)
Signet 2750 or
Signet 2760
(sold separately)
• Easy sensor replace-
®
ment using DryLoc
electrode connector
• High temperature
versions available
• Mounts into standard
3/4 inch threads
• Compatible with all
Signet pH/ORP instruments
Applications
• Water Treatment &
Water Quality
Monitoring
• Demineralizer, Regeneration & Rinse
• Cooling Tower and
Boiler Protection
• Aquatic Animal Life
Support Systems
• Pool and Spa Control
• Neutralization
Systems
Signet 2774-2777
DryLoc pH/ORP Electrodes
Fittings - Customer supplied
92
Signet 2774-2777
DryLoc pH/ORP
Electrodes
www.gfsignet.com
www.gfsignet.com
Page 2
DimensionsSpecifications
38.6mm
(1.52 in.)
35.5mm
(1.4 in.)
29.2mm
(1.15 in.)
128.3mm
(5.05 in.)
22.8mm
(0.9 in.)
22.1mm
(0.87 in.)
General
Compatibility:
Signet Models 2750 and 2760
Operating Range:
• 2774/2776: 0 to 14 pH
• 2775/2777: +/-2000 mV (ORP)
Pipe Size Range: 1 inch and up.
• In-line/vertical mounting: Use the
electrode’s 3/4 in. threads to install
into pipe fitting. Electrode can be
mounted at any angle.
• Submersible mounting: Use
threads on Model 2750 or 2760;
requires 3/4 in. NPT or ISO7-R/ 3/4
in. male threaded extension.
Temperature Sensor:
• pH: 3KΩ or PT1000 RTD
• ORP: none
Temperature response time:
τ
< 1 min.
95%
Shipping Weight: 0.25 kg (0.55 lbs.)
Standards and Approvals
• CE
• Manufactured under ISO 9001:2000
for Quality and ISO 14001:2004 for
Environmental Management
Choosing
Products
Instrument
Multi-Pa-
rameter
Flow
pH/ORP
Conductivity/
Resistivity
Pressure, Level
Temperature,
Max. Temperature/Pressure rating:
Operating Temperature:
0°C to 85°C (32°F to 185°F)
Maximum Operating Pressure:
6.9 bar (100 psi)
Storage Temperature:
> 0°C (32°F)
Higher temperature and pressure sensors are available.
Model 2774-2777 Compatibility Selection Chart for Electronics and Instruments
Compatible Signet Instruments
Sensor Temp.
27743 KΩ2760-1, -2,
2774-1 PT 1000
277510 KΩ2760-1, -2,
2775-1 None---- 2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
27763 KΩ2760-1, -2,
2776-1 PT 1000
Compatible Electrodes
277710 KΩ2760-1, -2,
2777-1 None---- 2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
*Must have internal preamplifier in instrument
**Must use sensor with compatible temperature element; contact technical service for questions
Element
570087508900PLC
-11, or -21
-11, or -21
-11, or -21
-11, or -21
2760-1, -2,
-11, or -21
--
2760-1, -2,
-11, or -21
2760-1, -2,
-11, or -21
--
2760-1, -2,
-11, or -21
--
any 2750 w/
digital output
any 2750 w/
digital output
--
any 2750 w/
digital output
any 2750 w/
digital output
any 2750 w/4 to
20 mA output
any 2750 w/4 to
20 mA output
any 2750 w/4 to
20 mA output
any 2750 w/4 to
20 mA output
Other Mfr’s Instruments with
internal preamplifiers*
2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
2760-3, -4, -31, or -41 2760-1, -2, -11, or -21
Other Mfr’s Instruments requiring
external preamplified signals**
Products
Other
Installation
Wiring &
Reference
Technical
Temperature/
Pressure
Graphs
of Terms
Glossary
Reference
Part No.
www.gfsignet.com
www.gfsignet.com
Index
93
Page 3
Electrode Key Features
and Benefits:
• Ryton™ body for chemical compatibility with most harsh chemicals.
Also able to withstand high temperatures.
• Porous PTFE reference junctions are
highly chemically resistant; resists
fouling and dirt buildup.
• First reference chamber with KNO
protects Ag/AgCl wire for a prolonged
sensor life.
• Capillary TC (temperature sensor)
embedded in tip of pH/ORP electrode
for quicker temperature response
than most other electrodes on the
market.
®
• DryLoc
connector with corrosion
resistant gold pins for quick and easy
sensor removal.
3
Reference
Ag/AgCl
Wire
Double
Porus
PTFE
Reference
Junctions
pH/ORP
Electrode
Electrode Cut-Away View
DryLoc
Connector
Ryton
Body
Reference
Electrolyte
Chamber
st
Reference
1
Chamber with
KNO
Capillary
TC
Model 2774-2777
Ordering Notes:
1) pH and ORP sensors
®
require connection to
model 2750 or 2760.
2) To replace electrodes,
simply twist from preamplifier to replace
with new electrode.
™
3) Conduit and mounting
brackets for submersible installation must
always be used (customer supplied).
4) All of these sensors
can be installed upside-down.
3
5) Special order options
may have longer delivery time. Consult your
local George Fischer
sales representative
for lead times.
2721 Remote Preamplifier
The 2721 remote preamplifier should
be used with special order sensors
that are built with cables (Signet
Models 277X-HT, 277X-1-HT, or other
Signet sensors ordered with cables).
It can also be used for applications
where another manufacturer’s sensor is used with a Signet 5700 or 8750
instrument.
43 mm
(1.7 in.)
305 mm
(12.2 in.)
30 mm
(1.2 in.)
Female
BNC
connector
Application Tips:
• Use the flat glass
electrodes when
a self-cleaning
feature is desired;
especially useful in
applications with
abrasive chemicals.
• Use bulb protected
electrodes for
general purpose
applications
• ORP electrodes are
generally used for
chemical reaction
monitoring, not
control.
• Ensure that sensor materials are
chemically compatible with the
process liquid.
94
www.gfsignet.com
www.gfsignet.com
Page 4
Ordering Information
Electrode Part Number- Choose either a pH or ORP Electrode
pH Electrodes
3-2774Flat pH surface electrode
3-2776Bulb pH electrode with bulb protection
Temperature Element - Choose one
-3K Ohm RTD for pH for connection to 8750 or 5700 instruments when used with the 2760 preamplifier.
-1PT1000 RTD for pH for connection to the 8900 when used with the 2750 sensor electronics
Special Order Options for pH Electrodes - Options -HT and -C can only be used with the 3-2721 preamplifier.
(These options cannot be used with the 2750 or 2760)
-HT
For high temperature and high pressure applications, up to 110°C (230°F) @ 150 psig; DryLoc
tor is removed and replaced with a 4.6m (15 ft.) cable.
-CRemove DryLoc
-ISO ISO 7/1 R - 3/4 Threaded electrodes are available.
ORP Electrodes
3-2775Flat ORP surface electrode
3-2777Bulb ORP electrode with bulb protection
Temperature Element - Choose one
-10 K ID resistor for ORP electrodes for connection to the 8750 and 5700 when used with the 2760 preamplifier
or the 8900 when used with the 2750 sensor electronics
-1No T.C. for ORP electrodes for use with other suppliers instruments when used with the 2760 connector
Special Order Options for ORP Electrodes - Options -HT and -C can only be used with the 3-2721 preamplifier
(These options cannot be used with the 2750 or 2760
-HT
For high temperature and high pressure applications, up to 110°C (230°F) @ 150 psig; DryLoc
tor is removed and replaced with a 4.6m (15 ft.) cable.
-CRemove DryLoc
-ISO ISO 7/1 R - 3/4 Threaded electrodes are available.
3-2775Example Part Number
®
connector and add 4.6m (15 ft.) cable. Other cable lengths are available
®
connector and add 4.6m (15 ft.) cable. Other cable lengths are available
®
connec-
®
connec-
Choosing
Products
Instrument
Multi-Pa-
rameter
Flow
pH/ORP
Conductivity/
Resistivity
Pressure, Level
Temperature,
Preamplifiers & Sensor Electronics - Choose one (required for initial installation of any pH or ORP electrode)
The Sensor Electronics below can be used with the Signet 8900 or as a blind transmitter with 4 to 20 mA output for connection to a PLC, PC, chart recorder, and other devices requiring 4 to 20 mA input.
3-2750-1 In-line Sensor electronics w/junction box
3-2750-2 In-line Sensor electronics w/junction box and EasyCal
3-2750-3 Submersible Sensor electronics w/4.6m (15 ft) cable and 3/4 in. NPT threads
3-2750-4 Submersible Sensor electronics w/4.6m (15 ft) cable and 3/4 in. ISO threads
The Preamplifiers below can be used with the Signet 5700 or 8750 Instruments
3-2760pH/ORP Preamplifier (for use with the 8750 or 5700 instrument) or Connector (for use with other manufacturer’s
3-2760-21Example Part Number
instruments) with 4.6m (15ft.) cable
Preamplifier or Connector Style; Choose one
-1Preamplifier with 3/4 inch NPT threads
-2Preamplifier with 3/4 inch ISO threads
-3Connector with 3/4 inch NPT threads
-4Connector with 3/4 inch ISO threads
Mounting Configurations; Choose one
-Submersible mounting - gray body
1In-line mounting - yellow body
Mfr. Part No. CodeMfr. Part No. CodeMfr. Part No. CodeMfr. Part No. Code