Signet 2630 Amperometric Chlorine Electrode
Description
The Signet 2630 Amperometric Chlorine
electrode is designed to measure free
chlorine in drinking water and wastewater
treatment applications. The electrodes
are available with a measurement range
of 0 to 5 ppm. This electrode requires the
Signet 2650 Amperometric Electronics
module to communicate with the Signet
8630-3 Chlorine Transmitter.
Utilizing smart-sensor technology,
this electrode has a unique embedded
memory chip and can communicate a wide
variety of information to the Signet 2650
electronics and Signet 8630-3 Transmitter.
System Overview
Panel Mount
Signet Transmitter
(sold separately)
3-8630-3P
Displayed information includes electrode
type, factory calibration data, service time,
chlorine range, high and low pH (with
optional Signet pH electrode), temperature
values and more.
Signet’s patented DryLoc® connector
provides quick assembly and a secure
connection. Gold-plated contacts and
an O-ring seal ensure a waterproof and
reliable interconnect to the Signet 2650
Amperometric Electronics.
The Signet 2630 Amperometric Chlorine
Electrode has an integrated temperature
element for automatic temperature
compensation.
Features
• Embedded memory chip
accessible via the Signet
8630 transmitter
• Quick assembly with
Signet’s patented
DryLoc
®
connector
• Integrated temperature
element for automatic
temperature
compensation
• Separate drive
electronics (Signet
2650), for easy electrode
replacement without
running new cable
Applications
Residual Chlorine
Monitoring:
Water Distribution•
Ground Water •
Surface Water•
HVAC Applications•
(cooling water)
Boiler Feed Water•
Gray Water •
Dechlorination
Food and Beverage•
RO Membrane •
Protection
Swimming Pools•
Aquariums•
Water Parks•
Signet 2650 Chlorine
Amperometric Electronics
(sold separately)
Signet 3-2630-2 Chlorine
Electrode
Signet 3610 Fitting
(sold separately)
(Sold Separately)
LEAD FREE
Pb
RoHS
COMPLIANT
Dimensions
3-2630-2
26.4 mm
1.04 in.)
72.6 mm
(2.86 in.)
25.15 mm
(0.99 in.)
Model 263X
Ordering Notes
1) The sensor must have
a stable and constant
flow of water past
its membrane for
accurate free chlorine
measurement. Set the
flow rate through the
flow cell from 0.75 -
2.25 l/min.
(0.2 - 0.6 gpm).
107.2 mm
(4.23 in.)
Specifications
General
Polarization Source:
Signet 2650 Amperometric Electronics
Compatibility:
3-3610-1
3-3610-2
3-4630.392
Mounting: Signet DryLoc connection
Materials: CPVC
Free Chlorine:
Membrane Material: PVDF
O-ring Material: FPM
Working Electrode: Gold
Counter Reference Electrode:
Silver halide
Wetted Material
PVC, PVDF, FPM•
Performance
Electrode:
Repeatability:
+/- 0.08 ppm (mg/l) or 3% of
selected range whichever is less
Slope: 10 to 60 nA/ppm (mg/l)
Response Time, T90: < 2 minutes
System: (including electronics and instrument)
A• ccuracy: < ±3% of electrode signal
after calibration
Resolution: ± 0.5% of electrode range•
Sensor Conditioning:•
New, first start-up:
4 hours maximum before calibration
Subsequent start-ups: 2 hours maximum•
Temperature Element: • PT1000, Class B
Performance (continued)
Operational Ranges and Limits:
Free Chlorine Range:•
0 to 5 ppm (mg/l)
Free Chlorine pH Operating Range:•
5.5 to 9.0 pH
Operating Temperature:
0 °C to 45 °C (32 °F to 113 °F)
Maximum Operating Pressure
Membrane:•
0.48 bar @ 25 °C (7 psi @ 77 °F)
Flow Velocity Across Membrane Surface:
Minimum: 15 cm/s (0.49 ft/s)
Maximum: 30 cm/s (0.98 ft/s)
Interferences:
ClO
Chemical Compatibility:
, ozone, bromine
2
< 50% ethanol/water,
< 50% glycerol/water
Environmental
Storage Temperature (dry):
-10 ºC to 60 ºC (-4 ºF to 140 ºF)
Operating Temperature:
0 ºC to 45 ºC (32 ºF to 113 ºF)
Relative Humidity:
0 to 95% indoor/outdoor
non-condensing to rated ambient
Shipping Weight
0.14 kg 0.30 lb
Standards and Approvals
CE•
Manufactured under ISO 9001 for •
Quality
RoHs compliant•
China RoHS•
Application Tips
The sensors should •
not be used in
water containing
surfactants, oils,
organic chlorine or
stabilizers such as
cyanuric acid.
Ordering Information
Mfr. Part No. Code Description
3-2630-2 159 001 662 Free Chlorine electrode, 0 to 5 ppm (mg/l)
Accessories and Replacement Parts
Mfr. Part No. Code Description
3-2630.391 159 001 674 Free Chlorine electrolyte, 30 ml
3-2630.392 159 001 675 Free Chlorine replacement membrane (1)
3-2630.396 159 001 676 Free Chlorine replacement kit - (2) electrolyte and
(2) membranes
3-0700.390 198 864 403 pH Buffer Kit (1 each 4, 7, 10 pH buffer in powder form,
makes 50 mL)
3822-7004 159 001 581 pH 4.01 buffer solution, 1 pint (473 ml)
3822-7007 159 001 582 pH 7.00 buffer solution, 1 pint (473 ml)
3822-7010 159 001 583 pH 10.00 buffer solution, 1 pint (473 ml)
3-2700.395 159 001 605 Calibration kit: included 3 polypropylene cups, box used
as cup stand, 1 pint pH 4.01, 1 pint pH 7.00
3-2630.099 Rev B (10/10)
© Georg Fischer Signet LLC
3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 • www.gfsignet.com • e-mail: signet.ps@georgfischer.com
Specifications subject to change without notice. All rights reserved. All corporate names and trademarks stated herein are the property of their respective companies.