Georg Fischer 2630 Data Sheet

Page 1
Signet 2630 Amperometric Chlorine Electrode
Description
The Signet 2630 Amperometric Chlorine electrode is designed to measure free chlorine in drinking water and wastewater treatment applications. The electrodes are available with a measurement range of 0 to 5 ppm. This electrode requires the Signet 2650 Amperometric Electronics module to communicate with the Signet 8630-3 Chlorine Transmitter.
Utilizing smart-sensor technology, this electrode has a unique embedded memory chip and can communicate a wide variety of information to the Signet 2650 electronics and Signet 8630-3 Transmitter.
System Overview
Panel Mount
Signet Transmitter
(sold separately)
3-8630-3P
Displayed information includes electrode type, factory calibration data, service time, chlorine range, high and low pH (with optional Signet pH electrode), temperature values and more.
Signet’s patented DryLoc® connector provides quick assembly and a secure connection. Gold-plated contacts and an O-ring seal ensure a waterproof and reliable interconnect to the Signet 2650 Amperometric Electronics. The Signet 2630 Amperometric Chlorine Electrode has an integrated temperature element for automatic temperature compensation.
Features
• Embedded memory chip accessible via the Signet 8630 transmitter
• Quick assembly with Signet’s patented DryLoc
®
• Integrated temperature element for automatic temperature compensation
• Separate drive electronics (Signet
2650), for easy electrode replacement without running new cable
Applications
Residual Chlorine Monitoring:
Water Distribution• Ground Water • Surface Water• HVAC Applications• (cooling water) Boiler Feed Water• Gray Water • Dechlorination Food and Beverage• RO Membrane • Protection Swimming Pools• Aquariums• Water Parks
Signet 2650 Chlorine Amperometric Electronics
(sold separately)
Signet 3-2630-2 Chlorine Electrode
Signet 3610 Fitting
(sold separately)
(Sold Separately)
LEAD FREE
Pb
RoHS
COMPLIANT
Page 2
Dimensions
3-2630-2
26.4 mm
1.04 in.)
72.6 mm (2.86 in.)
25.15 mm (0.99 in.)
Model 263X Ordering Notes
1) The sensor must have a stable and constant flow of water past its membrane for accurate free chlorine measurement. Set the flow rate through the flow cell from 0.75 -
2.25 l/min.
(0.2 - 0.6 gpm).
107.2 mm (4.23 in.)
Specifications
General
Polarization Source: Signet 2650 Amperometric Electronics Compatibility: 3-3610-1 3-3610-2 3-4630.392 Mounting: Signet DryLoc connection Materials: CPVC Free Chlorine: Membrane Material: PVDF O-ring Material: FPM Working Electrode: Gold Counter Reference Electrode: Silver halide
Wetted Material
PVC, PVDF, FPM
Performance
Electrode: Repeatability: +/- 0.08 ppm (mg/l) or 3% of selected range whichever is less Slope: 10 to 60 nA/ppm (mg/l)
Response Time, T90: < 2 minutes
System: (including electronics and instrument)
A• ccuracy: < ±3% of electrode signal
after calibration Resolution: ± 0.5% of electrode range• Sensor Conditioning: New, first start-up:
4 hours maximum before calibration
Subsequent start-ups: 2 hours maximum• Temperature Element: PT1000, Class B
Performance (continued)
Operational Ranges and Limits:
Free Chlorine Range: 0 to 5 ppm (mg/l) Free Chlorine pH Operating Range:
5.5 to 9.0 pH Operating Temperature: 0 °C to 45 °C (32 °F to 113 °F)
Maximum Operating Pressure
Membrane:
0.48 bar @ 25 °C (7 psi @ 77 °F)
Flow Velocity Across Membrane Surface: Minimum: 15 cm/s (0.49 ft/s) Maximum: 30 cm/s (0.98 ft/s) Interferences: ClO Chemical Compatibility:
, ozone, bromine
2
< 50% ethanol/water, < 50% glycerol/water
Environmental
Storage Temperature (dry):
-10 ºC to 60 ºC (-4 ºF to 140 ºF) Operating Temperature: 0 ºC to 45 ºC (32 ºF to 113 ºF) Relative Humidity: 0 to 95% indoor/outdoor non-condensing to rated ambient
Shipping Weight
0.14 kg 0.30 lb
Standards and Approvals
CE• Manufactured under ISO 9001 for • Quality RoHs compliant• China RoHS
Application Tips
The sensors should • not be used in water containing surfactants, oils, organic chlorine or stabilizers such as cyanuric acid.
Ordering Information
Mfr. Part No. Code Description
3-2630-2 159 001 662 Free Chlorine electrode, 0 to 5 ppm (mg/l)
Accessories and Replacement Parts
Mfr. Part No. Code Description
3-2630.391 159 001 674 Free Chlorine electrolyte, 30 ml 3-2630.392 159 001 675 Free Chlorine replacement membrane (1) 3-2630.396 159 001 676 Free Chlorine replacement kit - (2) electrolyte and
(2) membranes
3-0700.390 198 864 403 pH Buffer Kit (1 each 4, 7, 10 pH buffer in powder form,
makes 50 mL) 3822-7004 159 001 581 pH 4.01 buffer solution, 1 pint (473 ml) 3822-7007 159 001 582 pH 7.00 buffer solution, 1 pint (473 ml) 3822-7010 159 001 583 pH 10.00 buffer solution, 1 pint (473 ml) 3-2700.395 159 001 605 Calibration kit: included 3 polypropylene cups, box used
as cup stand, 1 pint pH 4.01, 1 pint pH 7.00
3-2630.099 Rev B (10/10)
© Georg Fischer Signet LLC 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 • www.gfsignet.com • e-mail: signet.ps@georgfischer.com Specifications subject to change without notice. All rights reserved. All corporate names and trademarks stated herein are the property of their respective companies.
Loading...