Georg Fischer 2630 Data Sheet

Signet 2630 Amperometric Chlorine Electrode
Description
The Signet 2630 Amperometric Chlorine electrode is designed to measure free chlorine in drinking water and wastewater treatment applications. The electrodes are available with a measurement range of 0 to 5 ppm. This electrode requires the Signet 2650 Amperometric Electronics module to communicate with the Signet 8630-3 Chlorine Transmitter.
Utilizing smart-sensor technology, this electrode has a unique embedded memory chip and can communicate a wide variety of information to the Signet 2650 electronics and Signet 8630-3 Transmitter.
System Overview
Panel Mount
Signet Transmitter
(sold separately)
3-8630-3P
Displayed information includes electrode type, factory calibration data, service time, chlorine range, high and low pH (with optional Signet pH electrode), temperature values and more.
Signet’s patented DryLoc® connector provides quick assembly and a secure connection. Gold-plated contacts and an O-ring seal ensure a waterproof and reliable interconnect to the Signet 2650 Amperometric Electronics. The Signet 2630 Amperometric Chlorine Electrode has an integrated temperature element for automatic temperature compensation.
Features
• Embedded memory chip accessible via the Signet 8630 transmitter
• Quick assembly with Signet’s patented DryLoc
®
• Integrated temperature element for automatic temperature compensation
• Separate drive electronics (Signet
2650), for easy electrode replacement without running new cable
Applications
Residual Chlorine Monitoring:
Water Distribution• Ground Water • Surface Water• HVAC Applications• (cooling water) Boiler Feed Water• Gray Water • Dechlorination Food and Beverage• RO Membrane • Protection Swimming Pools• Aquariums• Water Parks
Signet 2650 Chlorine Amperometric Electronics
(sold separately)
Signet 3-2630-2 Chlorine Electrode
Signet 3610 Fitting
(sold separately)
(Sold Separately)
LEAD FREE
Pb
RoHS
COMPLIANT
Dimensions
3-2630-2
26.4 mm
1.04 in.)
72.6 mm (2.86 in.)
25.15 mm (0.99 in.)
Model 263X Ordering Notes
1) The sensor must have a stable and constant flow of water past its membrane for accurate free chlorine measurement. Set the flow rate through the flow cell from 0.75 -
2.25 l/min.
(0.2 - 0.6 gpm).
107.2 mm (4.23 in.)
Specifications
General
Polarization Source: Signet 2650 Amperometric Electronics Compatibility: 3-3610-1 3-3610-2 3-4630.392 Mounting: Signet DryLoc connection Materials: CPVC Free Chlorine: Membrane Material: PVDF O-ring Material: FPM Working Electrode: Gold Counter Reference Electrode: Silver halide
Wetted Material
PVC, PVDF, FPM
Performance
Electrode: Repeatability: +/- 0.08 ppm (mg/l) or 3% of selected range whichever is less Slope: 10 to 60 nA/ppm (mg/l)
Response Time, T90: < 2 minutes
System: (including electronics and instrument)
A• ccuracy: < ±3% of electrode signal
after calibration Resolution: ± 0.5% of electrode range• Sensor Conditioning: New, first start-up:
4 hours maximum before calibration
Subsequent start-ups: 2 hours maximum• Temperature Element: PT1000, Class B
Performance (continued)
Operational Ranges and Limits:
Free Chlorine Range: 0 to 5 ppm (mg/l) Free Chlorine pH Operating Range:
5.5 to 9.0 pH Operating Temperature: 0 °C to 45 °C (32 °F to 113 °F)
Maximum Operating Pressure
Membrane:
0.48 bar @ 25 °C (7 psi @ 77 °F)
Flow Velocity Across Membrane Surface: Minimum: 15 cm/s (0.49 ft/s) Maximum: 30 cm/s (0.98 ft/s) Interferences: ClO Chemical Compatibility:
, ozone, bromine
2
< 50% ethanol/water, < 50% glycerol/water
Environmental
Storage Temperature (dry):
-10 ºC to 60 ºC (-4 ºF to 140 ºF) Operating Temperature: 0 ºC to 45 ºC (32 ºF to 113 ºF) Relative Humidity: 0 to 95% indoor/outdoor non-condensing to rated ambient
Shipping Weight
0.14 kg 0.30 lb
Standards and Approvals
CE• Manufactured under ISO 9001 for • Quality RoHs compliant• China RoHS
Application Tips
The sensors should • not be used in water containing surfactants, oils, organic chlorine or stabilizers such as cyanuric acid.
Ordering Information
Mfr. Part No. Code Description
3-2630-2 159 001 662 Free Chlorine electrode, 0 to 5 ppm (mg/l)
Accessories and Replacement Parts
Mfr. Part No. Code Description
3-2630.391 159 001 674 Free Chlorine electrolyte, 30 ml 3-2630.392 159 001 675 Free Chlorine replacement membrane (1) 3-2630.396 159 001 676 Free Chlorine replacement kit - (2) electrolyte and
(2) membranes
3-0700.390 198 864 403 pH Buffer Kit (1 each 4, 7, 10 pH buffer in powder form,
makes 50 mL) 3822-7004 159 001 581 pH 4.01 buffer solution, 1 pint (473 ml) 3822-7007 159 001 582 pH 7.00 buffer solution, 1 pint (473 ml) 3822-7010 159 001 583 pH 10.00 buffer solution, 1 pint (473 ml) 3-2700.395 159 001 605 Calibration kit: included 3 polypropylene cups, box used
as cup stand, 1 pint pH 4.01, 1 pint pH 7.00
3-2630.099 Rev B (10/10)
© Georg Fischer Signet LLC 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057 • www.gfsignet.com • e-mail: signet.ps@georgfischer.com Specifications subject to change without notice. All rights reserved. All corporate names and trademarks stated herein are the property of their respective companies.
Loading...