Roto-Shake Genie® Operating Instructions
The Roto-Shake Genie is designed to provide rotating, rocking, and orbital
shaking in one compact, durable and user-friendly unit. A magnetic stainless
steel platform easily converts from rocking to rotating, and securely holds a
non-skid metal rocker tray, various spring-clip mounting plates and bag
mounting strips. An optional orbital shaker attachment (3-D or Horizontal)
transforms the unit into an orbital shaker. Rotating and rocking speeds can
be easily adjusted with the fingertip speed control knob.
APPLICATIONS
The stainless steel platform (200 x 340mm / 8 x 13.5in) is magnetized to
securely hold various sample types and sizes. The following attachments are
supplied with the Roto-Shake Genie:
1 non-skid metal rocker tray (305 x 305mm / 12 x 12in)
4 magnetic stainless steel bag mounting strips
1 clip plate for 12 each 10-13mm tubes
1 clip plate for 6 each 15-17mm tubes
1 clip plate for 3 each 28-30mm tubes
Additional attachments are available (listed under "Roto-Shake Genie
Accessories") to provide maximum application versatility and can be
purchased through most laboratory equipment distributors or visit www.
scientificindustries.com.
The Roto-Shake Genie is ideal for mixing or washing of any sample type in
life science, chemical, clinical/hospital, food, cosmetic, or environmental laboratories and can withstand temperatures from 0°C to 38°C to allow operation
in cold rooms and incubators.
ASSEMBLY INSTRUCTIONS
We recommend that you retain your original packaging for 90 days in case
you need to return the product for any reason to your distributor or Scientific
Industries. (see "Magnetic Platform Assembly Instructions" provided with the
platform for additional detail)
1. Carefully unpack the magnetic platform (Scientific Industries No.
0A-1100-030). The platform is shipped detached from the housing and
must first be assembled (see fig. 1). Unscrew the lock shaft (D) from
the right side of the housing (E) and remove it from the hole.
2. Affix the platform (B) into the housing with the hex hole (J) facing left,
and the lock release (A) facing right, as shown. Rotate the platform to
line up the hex shaft (I) with the hex hole and slide it onto the hex shaft.
Figure 1
3. Depress and hold the lock release. Slide the lock shaft through the
hole in the housing into the platform. Carefully screw the lock shaft fully
into place - do not force it. Hand-tighten only. Release the lock release.
4. To remove the magnetic platform assembly (B), press and hold the lock
release (A). Unscrew the lock shaft (D) from the right side of the
housing (E) and remove it while supporting the right end of the
platform. Release the lock release. Slide the platform to the right and
rotate it to line up with the hex shaft (I). Slide the platform off of the hex
shaft and remove it. The platform may be immersed for cleaning or
decontamination. Do not autoclave.
CAUTION! The Roto-Shake Genie® platform is magnetic. Do not
place materials on the platform that will be damaged by strong
magnets.
OPERATING INSTRUCTIONS
1. Rotating Mode
a. Press lock release (A).
b. Shift the magnetic platform (B) to the right.
c. While shifting the platform right, release the
lock release, and line up the hex shaft (I) to
engage the lock shaft (D) properly.
d. The platform will lock in place when it
reaches the correct position.