APPLICATIONS
The stainless steel platform (200 x 340mm / 8 x 13.5in) is magnetized to securely hold various sample types and sizes. The following attachments are supplied with the Roto-Shake Genie:
1 non-skid metal rocker tray (305 x 305mm / 12 x 12in) 4 magnetic stainless steel bag mounting strips 1 clip plate for 12 each 10-13mm tubes 1 clip plate for 6 each 15-17mm tubes 1 clip plate for 3 each 28-30mm tubes
Additional attachments are available (listed under "Roto-Shake Genie Accessories") to provide maximum application versatility and can be purchased through most laboratory equipment distributors or visit www. scientificindustries.com.
The Roto-Shake Genie is ideal for mixing or washing of any sample type in life science, chemical, clinical/hospital, food, cosmetic, or environmental labo­ratories and can withstand temperatures from 0°C to 38°C to allow operation in cold rooms and incubators.
ASSEMBLY INSTRUCTIONS
We recommend that you retain your original packaging for 90 days in case you need to return the product for any reason to your distributor or Scientific
Industries. (see "Magnetic Platform Assembly Instructions" provided with the platform for additional detail)
1. Carefully unpack the magnetic platform (Scientific Industries No. 0A-1100-030). The platform is shipped detached from the housing and must first be assembled (see fig. 1). Unscrew the lock shaft (D) from the right side of the housing (E) and remove it from the hole.
2. Affix the platform (B) into the housing with the hex hole (J) facing left, and the lock release (A) facing right, as shown. Rotate the platform to line up the hex shaft (I) with the hex hole and slide it onto the hex shaft.
Figure 1
3. Depress and hold the lock release. Slide the lock shaft through the hole in the housing into the platform. Carefully screw the lock shaft fully into place - do not force it. Hand-tighten only. Release the lock release.
4. To remove the magnetic platform assembly (B), press and hold the lock release (A). Unscrew the lock shaft (D) from the right side of the housing (E) and remove it while supporting the right end of the platform. Release the lock release. Slide the platform to the right and rotate it to line up with the hex shaft (I). Slide the platform off of the hex shaft and remove it. The platform may be immersed for cleaning or decontamination. Do not autoclave.
CAUTION! The Roto-Shake Genie® platform is magnetic. Do not place materials on the platform that will be damaged by strong magnets.
OPERATING INSTRUCTIONS
1. Rotating Mode
a. Press lock release (A). b. Shift the magnetic platform (B) to the right. c. While shifting the platform right, release the
lock release, and line up the hex shaft (I) to engage the lock shaft (D) properly.
d. The platform will lock in place when it
reaches the correct position.