The Roto-Shake Genie is designed to provide rotating, rocking, and orbital
shaking in one compact, durable and user-friendly unit. A magnetic stainless
steel platform easily converts from rocking to rotating, and securely holds a
non-skid metal rocker tray, various spring-clip mounting plates and bag
mounting strips. An optional orbital shaker attachment (3-D or Horizontal)
transforms the unit into an orbital shaker. Rotating and rocking speeds can
be easily adjusted with the fingertip speed control knob.
APPLICATIONS
The stainless steel platform (200 x 340mm / 8 x 13.5in) is magnetized to
securely hold various sample types and sizes. The following attachments are
supplied with the Roto-Shake Genie:
1 non-skid metal rocker tray (305 x 305mm / 12 x 12in)
4 magnetic stainless steel bag mounting strips
1 clip plate for 12 each 10-13mm tubes
1 clip plate for 6 each 15-17mm tubes
1 clip plate for 3 each 28-30mm tubes
Additional attachments are available (listed under "Roto-Shake Genie
Accessories") to provide maximum application versatility and can be
purchased through most laboratory equipment distributors or visit www.
scientificindustries.com.
The Roto-Shake Genie is ideal for mixing or washing of any sample type in
life science, chemical, clinical/hospital, food, cosmetic, or environmental laboratories and can withstand temperatures from 0°C to 38°C to allow operation
in cold rooms and incubators.
ASSEMBLY INSTRUCTIONS
We recommend that you retain your original packaging for 90 days in case
you need to return the product for any reason to your distributor or Scientific
Industries.(see "Magnetic Platform Assembly Instructions" provided with the
platform for additional detail)
1. Carefully unpack the magnetic platform (Scientific Industries No.
0A-1100-030). The platform is shipped detached from the housing and
must first be assembled (see fig. 1). Unscrew the lock shaft (D) from
the right side of the housing (E) and remove it from the hole.
2. Affix the platform (B) into the housing with the hex hole (J) facing left,
and the lock release (A) facing right, as shown. Rotate the platform to
line up the hex shaft (I) with the hex hole and slide it onto the hex shaft.
Page 3
Figure 1
3. Depress and hold the lock release. Slide the lock shaft through the
hole in the housing into the platform. Carefully screw the lock shaft fully
into place - do not force it. Hand-tighten only. Release the lock release.
4. To remove the magnetic platform assembly (B), press and hold the lock
release (A). Unscrew the lock shaft (D) from the right side of the
housing (E) and remove it while supporting the right end of the
platform. Release the lock release. Slide the platform to the right and
rotate it to line up with the hex shaft (I). Slide the platform off of the hex
shaft and remove it. The platform may be immersed for cleaning or
decontamination. Do not autoclave.
CAUTION! The Roto-Shake Genie® platform is magnetic. Do not
place materials on the platform that will be damaged by strong
magnets.
OPERATING INSTRUCTIONS
1. Rotating Mode
a. Press lock release (A).
b. Shift the magnetic platform (B) to the right.
c. While shifting the platform right, release the
lock release, and line up the hex shaft (I) to
engage the lock shaft (D) properly.
d. The platform will lock in place when it
reaches the correct position.
Page 4
2. Rocking Mode
a. Press the lock release (A).
b. Shift the magnetic platform (B) to the left.
c. While shifting the platform left, release the
lock release and rotate the platform to line
up the pin (H) with the pin slot (K) in the
end of the platform. Shift the platform onto
the pin.
d. The platform will lock in place when it
reaches the correct position.
3. Orbital Mode Requires optional orbital shaker attachment (3-D
Attachment, SI-1150 or Horizontal Attachment, SI-1175)
a. Remove magnetic platform (B) and all accessories from the Roto-
Shake Genie
®
.
b. Place orbital shaker attachment on the Roto-Shake Genie base,
aligning the hex shaft (I) with the hex on the pulley of the
orbital shaker attachment. Ensure that the base feet are in direct
contact with the bottom of the Roto-Shake Genie.
To Load Samples:
1. Both sides of the versatile magnetic platform can be loaded
a. To rock open vessels (rocking): Select rocking mode. Center
the metal rocker tray on the magnetic
platform. Place samples anywhere on the
tray. (NOTE: Metal rocker tray can only
be used in rocking mode)
b. To shake/mix open vessels (orbital
shaking): Insert optional orbital shaker
attachment. Center sample containers on
the platform.
c. To rotate or rock bags: Select mode of
operation and secure bags in the desired
position using 2 - 4 magnetic stainless steel
mounting strips. Additional mounting strips
and tray expansion kit are available for
higher sample quantities (refer to "RotoShake Genie Accessories").
d. To rotate or rock tubes: Select mode of
operation and load samples into the
appropriate clip plate. Three tube clip plate sizes
Page 5
are available (one clip plate of each size is supplied with the Roto-
Shake Genie). Additional clip plates are available for holding
additional tubes (refer to "Roto-Shake Genie Accessories")
Size
Large
Medium
Small
NOTE: Clip plates may be aligned at various angles to produce different
motions within the tubes. Before operation, rotate the platform to ensure
that the tubes will clear the Roto-Shake Genie housing.
2. Turn the speed control knob to the lowest setting and turn the power
switch on. The platform will begin to rotate, rock or orbital shake.
3. Select the speed by turning the speed control knob until the desired
speed is attained. If using the Roto-Shake Genie in a cold room, allow
the unit to acclimate to the room temperature before setting the exact
speed.
CAUTION! Failure to follow operating instructions can compromise
user's safety.
CAUTION! Ensure that the power is off before shifting operating
modes.
CAUTION! Ensure that the power is off before removing the mag netic platform assembly.
Diameter of Tube
28 - 30mm
15 - 17mm
10 - 13mm
Maximum Volume of Tube
50ml
15ml
1.5 - 2.0ml
CARE AND HANDLING
1. Roto-Shake Genie® and Accessories:
The Roto-Shake Genie should be given the care normally required for any
electrical appliance. Avoid wetting or unnecessary exposure to fumes. The
surface may be cleaned with most cleaning agents. It cannot and should not
be autoclaved. The magnetic platform assembly is stainless steel and can be
cleaned with most cleaning agents. It cannot and should not be autoclaved.
For thorough cleaning and decontamination, the platform should be removed
from the Roto-Shake Genie. The clip plates and mounting strips are made of
magnetic stainless steel and can be cleaned with most cleaning agents or
autoclaved. The metal rocker tray cannot and should not be autoclaved.
Clean up any spills promptly to prevent damage. Replacement parts are
available through most laboratory equipment distributors or directly from
Scientific Industries.
CAUTION! Ensure that the power is off before cleaning the Roto Shake Genie.
CAUTION! Do not immerse.
Page 6
2. Fuses:
The electronics are protected by a quick blow 5 x 20mm Fuse, 0.5 amp for
all models.
To replace the fuse, locate the fuse holder on the back of the housing. Insert
a screwdriver into the slot in the cap and push in gently while turning ¼ turn
counterclockwise. Allow the cap to come off with the fuse. Remove the fuse
from the cap and insert the proper replacement fuse. Push the fuse and cap
back into the fuse holder and rotate the cap clockwise until it locks in.
CAUTION! Replace fuse with the correct type and value only.
CAUTION! Ensure that the power is off before replacing the fuse.
SPECIFICATIONS
Dimensions (HxWxD): 267 x 508 x 252mm
(10.5 x 20.0 x 9.9in)
Depth with Rocker Tray: 305mm (12in)
Weight: 11.8 Kg (26 lb)
Speed: Rocking - 10-70 cycles per minute
Rocking Angle: ±10 degrees from horizontal
Environmental: The Roto-Shake Genie
"Installation Category 2".
0°C (32°F) to 38°C (100°F), 95% maximum
relative humidity.
Maximum Load:4.5 Kg (10 lb)
®
is classified as
Standard Accessories:
1 clip plate for 12 each 10-13mm tubes
1 clip plate for 6 each 15-17mm tubes
1 clip plate for 3 each 28-30mm tubes
4 magnetic stainless steel bag mounting strips
1 non-skid metal rocker tray (305 x 305mm / 12 x 12in)
Additional accessories are available for increased capacity.
Page 7
Roto-Shake Genie® Accessories
Fig.Cat. No.Description
1SI-11231 Clip Plate for 12 each 10-13mm tubes
2SI-11241 Clip Plate for 6 each 15-17mm tubes
3SI-11251 Clip Plate for 3 each 28-30mm tubes
4SI-11206 Clip Plates for 12 each 10-13mm tubes
5SI-11216 Clip Plates for 6 each 15-17mm tubes
6SI-11226 Clip Plates for 3 each 28-30mm tubes
7SI-1126Bag Mounting Strips (pack of 16)
8SI-1127Expansion Kit - two stainless steel trays, 24 bag mounting strips
9SI-11302 Clip Plates for Hybridization Tubes
10SI-11342 Universal Clip Plates
11SI-11351 Magnetic Covered 100 Microtube Box
12SI-11361 Clip Plate for 100ml Volumetric Flasks
130A-1100-070Non-Skid Tray
14SI-11503-D Orbital Shaker Attachment
15SI-1175Horizontal Orbital Shaker Attachment
16SI-1170Stackable Tray
17SI-1137Dual Port Mixing Tube (*)
Page 8
ROTO-SHAKE GENIE
PARTS ASSEMBLY LIST
To order parts for the Roto-Shake Genie:
Contact your local distributor or visit
www.scientificindustries.com. Please specify
Part No., quantity and electric voltage.
®
Indicator No. Part No.DescriptionIndicator No. Part No.Description
10K-1100-902 Power Supply Board Kit80K-1100-900 Hex Shaft Kit
2AESP0001100, 120V DPST Switch-Red9KBP0002Speed Control Knob