UV sensor is with extremely
precise structure. Once don't
use it , be sure to store it in
the dry environment. For
example, put the whole sensor
including Desiccant ( Drier )
into to the Plastic bag and seal
the bag as tightly as possible
( refer the following figure ).
Take the sensor out of the bag only when use it.
Comply to above method will extend the life of UV
sensor. Otherwise, the gain of the UV SENSOR may
be decreased and shorten the calibration period. It is
also necessary to replace the Desiccant ( Drier )
periodically.
Maximum indication 1999.
Measurement
ranges & Lo range : 1999 uW/cm^2 x 1 uW/cm^2
resolutionHi range : 19990 uW/cm^2 x 10 uW/cm^2
UV sensor Band pass 290 nm to 390 nm.
spectrum
Accuracy± ( 4 % FS + 2 dgt )FS : full scale
Sensor structureThe exclusive UV photo sensor with
Buttons. SwitchPower button, Zero button, Range switch
Sample TimeApprox. 1 sec.
Over Range
indication
Weight 251 g / 0.55 LB (including battery)
Operating Temp.0 to 50 .℃
OperatingLess than 85% R.H.
Humidity
Power SupplyDC 9V battery, 006P , MN 1604 (PP3)
Power Approx. DC 6 mA.
Consumption
SizeMain instrument : 131 x 70 x 25 mm.
AccessoriesInstruction manual........... 1 PC.
IncludedUV sensor probe.............. 1 PC.
Optional Acc.Soft carrying case, CA-52A, CA-03.
2 ranges : Hi & Lo range
* 1000 uW/cm^2 = 1 mW/cm^2
* Calibration is executed under the UVA light &
and compare with the standard UVA light meter.
* Spec. tested under the environment RF Field
Strength less than 3 V/M & frequency less than the
30 MHz only.
the cosine correction filter.
or equivalent.
Sensor probe head : 45 mm dia x 32 mm.
Sensor probe handle : 125 x 24 mm dia.
2
4. FRONT PANEL DESCRIPTION
4-1 Display
4-2 Power Button
4-3 Range Switch (Hi/Lo Switch)
4-4 Zero Button
4-5 Stand
4-6 Battery Compartment/Cover
4-7 UV Probe Handle
4-8 UV Sensor
3
5. MEASURING PROCEDURE
Under the " Lo range " if the " Display show " ", it
1) Push the " Power Button ( 4-2, Fig. 1 ) continuously until
the " Display " ( 4-1, Fig. 1 ) on then release the " Power
Button : will power ON the meter.
* After power ON if push the " Power Button ( 4-2, Fig. 1 )
once a while again will power OFF the meter.
2) Select the max. range on the " Range Switch "
( 4-3, Fig. 1 ).
Lo range : 1999 uW/cm^2 x 1 uW/cm^2
Hi range : 19990 uW/cm^2 x 10 uW/cm^2
* 1000 uW/cm^2 = 1 mW/cm^2
3) Hold the " Probe Handle " ( 4-7, Fig. 1 ) & let " UV Sensor"
( 4-8, Fig. 1 ) face to measuring UV light source , then the
Display ( 4-1, Fig. 1 ) will show values of UV light on the
display reading.
6. MEASURING CONSIDERATION
1) As the " Lo range " is designed & to measure the UV light
values more than 2000 uW/cm^2 . If the measured UV light
values more than 1999 uW/cm^2, it should select the " Range
Switch " to the " Hi range " to get the exact measuring value.
*
means the measurement value already over range, then
should select the " Range Switch " ( 4-3. Fig. 1 ) to the
" Hi " position ( 19990 uW/cm^2 range ).
2)Zero Button ( 4-4, Fig. 1 ) can be operated under the
Lo range ( 1999 uW/cm^2 range ) and the display
value 100 uW/cm^2 only.≦
4
2) Storage for the " UV sensor "
When LCD display shows " " in the left corner,
UV sensor is with extremely precise structure.
Once don't use it , be sure to store it in the dry environment.
For example, put the whole sensor including Desiccant
( Drier ) into to the Plastic bag and seal the bag as tightly
as possible. Take the sensor out of the bag only when use
it. Comply to above method will extend the life of UV
sensor. Otherwise, the gain of the UV SENSOR may be
decreased and shorten the calibration period. It is also
necessary to replace the Desiccant ( Drier ) periodically.
7. BATTERY REPLACEMENT
1)
It is necessary to replace the battery. However, in-spec
measurement may still be made for several hours after
low battery appears.
2) Slide the " Battery Cover " ( 4-6, Fig. 1 ) away from the
instrument and remove the battery.
3) Replace with 9V battery and reinstate the cover.
4) Make sure the battery cover is secured after changing the
battery.
5
0806-UV340