To compensate for thermocouple error, you must first
set the offset value. To do so (recommended each time
the thermocouple probe is changed):
1. Press and hold the key.
2. Use the and keys, respectively, to increase
or decrease the offset value. The allowable adjustment
range is ±9°F (±5°C).
3. Press the key to save the setting and exit offset
setup mode.
To adjust the offset of the secondary channel in the
DT80-2, press the key and repeat steps 2 and 3.
The instrument will automatically save the offset value
upon shutdown, and default to it upon the next startup.
SPECIFICATIONS
• Display: dual 5-digit LCD
• Temperature units: Degrees Celsius or Fahrenheit
• Display resolution: 0.1
• Measurement range: -328° to 1922°F (-200° to
1050°C) using a J-type thermocouple; -328° to 2498°F
(-200° to 1370°C) using a K-type thermocouple
• Measurement accuracy: ±(0.05% of reading + 1.4°F)
• Supplied Probe(s) Range: 50º to 500ºF (-45º to 260ºC)
• Operating temperature range: 40° to 104°F
(5° to 40°C), below 80% RH
• Storage temperature range: 14° to 140°F
(-10° to 60°C), below 70% RH