Rev. C2016/06/01Removed breast and testes from indication of use
Rev. D2016/08/18Included IP33, Ophthalmic and Aorta presets, IEC
(YYYY/MM/DD)
Included Power button under device labels
Included revised rating label
Included sections in Privacy and Security chapter
statement
Included the revised rating label, battery label and
external charger label
Included the scope for all the standards in Table i-1
60601-11 and 12 standards, Tricefy App
REASON FOR CHANGE
Please verify that you are using the latest revision of this document. Information
pertaining to this document is maintained on ePDM (GE electronic Product Data
Management). If you need to know the latest revision, contact your distributor, local GE
Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at
1 800 682 5327 or 1 262 524 5698.
Vscan Extend – User Manuali-1
5721203-100Rev. D
This page intentionally left blank.
i-2Vscan Extend – User Manual
5721203-100
Rev. D
Regulatory Requirements
Conformance Standards
The GE product families are tested to meet all applicable
requirements in relevant EU Directives and European/
International standards. Any changes to accessories, peripheral
units or any other part of the system must be approved by the
manufacturer: GE Medical Systems. Ignoring this advice may
compromise the regulatory approvals obtaine d for the product.
This product complies with the regulatory requirement of the
following:
Table i-1: Regulatory Requirements
Standard/DirectiveScope
93/42/EECMedical Devices Directive (MDD)
2007/47/EC (MDD amendment)
Directive 2011/65/EU RoHS
2002/96/EC WEEE
The CE label affixed to the product testifies compliance to
the Directive. The location of the CE marking is shown in the
Safety chapter of this manual.
EN55011Industrial, scientific and medical equipment -
Radio-frequency disturbance characteristics - Limits and
methods of measurement
IEC* 60601-1
CAN/CSA-C22.2 No 601.1
IEC* 60601-2-37Medical electrical equipment - Part 2-37. Particular
IEC* 60601-1-2Medical Electrical Equipment - part 1-2. Collateral standard:
IEC* 60601-1-4Medical Electrical Equipment - part 1-4. Collateral standard:
IEC* 60601-1-6Medical Electrical Equipment - part 1-6. Collateral standard:
NEMA/AIUM UD-3Standard for real-time display of thermal and mechanical
Medical Electrical Equipment, Part 1; General Requirements
for Safety
requirements for the safety of ultrasonic medical diagnostic
and monitoring equipment
Electromagnetic compatibility - Requirements and tests.
Programmable electrical medical systems
Usability.
acoustic output indices on diagnostic ultrasound equipment.
Vscan Extend – User Manuali-3
5721203-100Rev. D
Table i-1: Regulatory Requirements (Continued)
Standard/DirectiveScope
ISO10993-1Biological evaluation of medical devices
EN 300 328Electromagnetic compatibility and Radio spectrum Matters
ISO 14971Medical devices - Application of risk management to medical
IEC* 62304Medical device software - Software life-cycle processes
IEC* 62366Medical devices - Application of usability engineering to
IEC* 60601-1-11Requirements for medical electrical equipment and medical
IEC* 60601-1-12Requirements for medical electrical equipment and medical
* including national deviations
(ERM); Wideband transmission systems
devices
medical devices
electrical systems used in the home healthcare environment
electrical systems intended for use in the emergency
medical services environment
Certifications
• GE Vingmed Ultrasound is ISO 13485 certified.
i-4Vscan Extend – User Manual
5721203-100
Rev. D
Classifications
Class II Equipment
The following classifications are in accordance with the IEC/
EN 60601-1.
Type and degree of protection against electric shock:
• Vscan Extend has an internal battery which allows the
operation during AC power absence.
• The AC adapter is Class II.
• Vscan Extend has type BFApplied Part.
Vscan Extend main unit is rated IP33:
• 3: Protected against solid foreign objects of 2,5 mm Ø and
greater.
• 3: Protected against spraying water.
Vscan Extend probe (immersible portion) is IPX7.
EQUIPMENT in which protection against electric shock does not
rely on BASIC INSULATION only, but in which additional safety
precautions such as DOUBLE INSULATION or REINFORCED
INSULATION are provided, there being no provision for
protective earthing or reliance upon installation conditions.
Type BF Applied part
TYPE BF APPLIED PART providing a specified degree of
protection against electric shock, with particular regard to
allowable LEAKAGE CURRENT.
Table i-2: Leakage Current
Normal modeSingle fault condition
Patient leakage current<100 microA<500 microA
Original Documentation
• The original document was written in English.
Vscan Extend – User Manuali-5
5721203-100Rev. D
Country Specific Approval
• USA AND TERRITORIES
DANGER
• JAPAN
DANGER
• CHINA
DANGER
Importer Information
Turkey
The following optional feature IS NOT available in the USA
and its territories:
– Ophthalmalic
The following optional feature IS NOT available in JAPAN:
– Ophthalmalic
The following optional feature IS NOT available in CHINA:
– Ophthalmalic
Contents:
‘Overview’ on page 1-2
‘Warnings’ on page 1-7
‘Contact Information’ on page 1-10
Vscan Extend – User Manual1-1
5721203-100Rev. D
Introduction
Attention
Overview
This manual covers the following two configurations of the
Vscan Extend. Refer to the relevant section based on the
configuration purchased.
• Vscan Extend
deep scanning (holding the phased array transducer G3S)
• Vscan Extend
deep and shallow scanning (holding the phased array
transducer G3S and the linear array transducer G8L)
Vscan Extend is a trademark of General Electric Company.
TM
configured with a sector probe allowing
TM
configured with a Dual Probe allowing
General description
Vscan Extend is a pocket-sized, battery powered general
purpose diagnostic ultrasound system. The system consists of a
handheld unit with a 5 inch touch screen display and a
permanently attached probe.
The battery can be charged either in the system or alone. The
system is capable of transferring images wirelessly to a DICOM
server or via Windows Share. Data can also be exported to a
standard PC using a wired USB export.
Capabilities also include access to GE Marketplace, which
allows the user to download Bladder Volume, Lung Protocol,
Barcode Reader and Interface to Case Exchange Software
Apps.
1-2Vscan Extend – User Manual
5721203-100
Rev. D
Principles of operation
Medical ultrasound images are created by computer and digital
memory from the transmission and reception of mechanical
high-frequency waves applied through a probe. The mechanical
ultrasound waves spread through the body, producing an echo
where density changes occur. The echoes return to the probe
where they are converted back into electrical signals.
These echo signals are amplified and processed by several
analog and digital circuits having filters with many frequency and
time response options, transforming the high-frequency
electrical signals into a series of digital image signals which are
stored in memory. Once in memory, the image can be displayed
in real-time on the image monitor.
A probe is an accurate, solid-state device, providing multiple
image formats. The digital design and use of solid-state
components provides highly stable and consistent imaging
performance with minimal required maintenance.
Safety
Overview
Intended use
Read and understand all instructions in the User's Manual
before attempting to use the ultrasound unit. Keep the manual
with the equipment at all time. Periodically review the
procedures for operation and safety precautions.
Vscan Extend is a general purpose diagnostic ultrasound
imaging system for use by qualified and trained healthcare
professionals enabling visualization and measurement of
anatomical structures and fluid.
Vscan Extend – User Manual1-3
5721203-100Rev. D
Introduction
Indications for use (for all countries except USA, China and Japan)
Vscan Extend is a general purpose diagnostic ultrasound
imaging system for use by qualified and trained healthcare
professionals enabling visualization and measurement of
anatomical structures and fluid. It's pocket-sized portability and
simplified user interface enables integration into examination
and training sessions indoors and in other environments
described in the user manual. The information can be used for
basic/focused assessments and adjunctively wi th othe r m ed ica l
data for clinical diagnosis purposes during routine, periodic
monitoring, and triage.
With the phased array transducer on the sector probe, the
specific clinical applications and exam types include: Cardiac;
Abdominal; Renal; OB/GYN; Urology; Fetal, Evaluation of
Presence of Fluid; Imaging Guidance for Needle/Catheter
Placement (e.g. paracentesis, pericardiocentesis, thoracentesis,
amniocentesis); Peripheral Vascular Imaging (e.g. arteries and
veins); Thoracic/Lung (e.g. pleural motion/sliding, line artifacts);
Adult Cephalic; and Pediatrics.
With the addition of the linear array transducer on the single
dual headed probe solution, the specific clinical applications and
exam types are expanded to include: Peripheral vascular
imaging (e.g. lower extremity, carotid); Procedure Guidance for
Arterial or Venous Vessels (e.g. central lines, upper extremity);
Small Organs (e.g. thyroid); Musculoskeletal (Long Bone; Hip,
shoulder, elbow and Knee Joints); Evaluation of Presence of
Fluid; Thoracic/Lung (e.g. pleural motion/sliding, line artifacts);
Ophthalmic*; and Pediatrics.
WARNING
*Ophthalmic use is provided as an option. Ophthalmic
scanning MUST only be used with the linear functionality of th e
Dual Probe.
If the Vscan Extend purchased does NOT have the Ophthalmic
preset option, DO NOT use for ophthalmic use or any use
causing the acoustic beam to pass through the eye.
1-4Vscan Extend – User Manual
5721203-100
Rev. D
Indications for use (for USA, China and Japan)
Vscan Extend is a general purpose diagnostic ultrasound
imaging system for use by qualified and trained healthcare
professionals enabling visualization and measurement of
anatomical structures and fluid. It's pocket-sized portability and
simplified user interface enables integration into examination
and training sessions indoors and in other environments
described in the user manual. The information can be used for
basic/focused assessments and adjunctively wi th othe r m ed ica l
data for clinical diagnosis purposes during routine, periodic
monitoring, and triage.
With the phased array transducer on the sector probe, the
specific clinical applications and exam types include: Cardiac;
Abdominal; Renal; OB/GYN; Urology; Fetal, Evaluation of
Presence of Fluid; Imaging Guidance for Needle/Catheter
Placement (e.g. paracentesis, pericardiocentesis, thoracentesis,
amniocentesis); Peripheral Vascular Imaging (e.g. arteries and
veins); Thoracic/Lung (e.g. pleural motion/sliding, line artifacts);
Adult Cephalic; and Pediatrics.
With the addition of the linear array transducer on the single
dual headed probe solution, the specific clinical applications and
exam types are expanded to include: Peripheral vascular
imaging (e.g. lower extremity, carotid); Procedure Guidance for
Arterial or Venous Vessels (e.g. central lines, upper extremity);
Small Organs (e.g. thyroid); Musculoskeletal (Long Bone; Hip,
shoulder, elbow and Knee Joints); Evaluation of Presence of
Fluid; Thoracic/Lung (e.g. pleural motion/sliding, line artifacts);
and Pediatrics.
Overview
Contraindication for use (for USA, China and Japan)
The Vscan Extend ultrasound device is not intended for
ophthalmic use or any use causing the acoustic beam to pass
through the eye.
Vscan Extend – User Manual1-5
5721203-100Rev. D
Introduction
Intended users
Prescription Device
Operator profile
Vscan Extend is intended to be used by qualified and trained
healthcare professionals that are legally authorized by law in the
country, state or other local municipality in which he pr actice s to
use the device. The list of the potential users includes but is not
limited to (based on title/geographical location): primary care
physicians, point-of-care users, sonographers, medical
healthcare technicians, nurses, midwives, paramedics, nurse
practitioner, physician assistants, medical students.
The users may or may not be working under supervision or
authority of a physician.
For USA only:
CAUTION: Federal law restricts this device to sale by or on the
order of a physician.
CAUTION
Qualified and trained healthcare professionals with at least a
basic level of general ultrasound training that includes limited
image acquisition techniques and interpretation (i.e. po sition the
probe correctly on the patient and determine at least normal vs.
abnormal anatomy views during scanning).
The operator must read and understand the user manual.
Contact GE sales representative fo r product training assistance
and visit the Vscan web portal for reference materials.
1-6Vscan Extend – User Manual
5721203-100
Rev. D
Important Safety Considerations
To prevent damage of the equipment or injury to your self or
others, read the following safety warnings before using Vscan
Extend.
Warnings
Warnings
WARNING
• Vscan Extend is a precision instrument. Handle Vscan
Extend and its accessories with care. Do not subject Vscan
Extend to mechanical shock or impact.
• Do not attempt to disassemble or alter any part of the unit
including the probe, the battery, the AC/DC adapter and
accessories. Disassembly or modification may result in
electrical shock.
• Stop using the unit if it emits smoke or noxious fumes.
Failure to do so may result in electrical shock or fire.
• Stop using the unit if the casing is damaged, including the
probe. Failure to do so may result in electrical shock.
• Do not use the device if the gorilla glass is broken.
• Do not use the AC/DC adapter if showing visible damages.
• Use only the designated power accessories (battery and
charger). Failure to do so may result in electrical shock or
fire.
• Do not place the battery near a heat source or expose it to
direct flame. Such exposure may lead to corrosive liquid
leakage, electrical shock or fire.
• To reduce risk for electrical shock, do not plug or unplug
the AC/DC adapter from mains socket with wet hands.
• Avoid dropping or subjecting the unit, including the probe,
the battery and accessories to severe impacts. This could
result in electrical shock, corrosive liquid leakage and
injury.
• Keep good hand contact with Vscan Extend during
scanning to avoid heating up of the unit and termination of
scan due to built-in temperature limits.
• At times, user may be required to enter PIN to save patient
data. It is extremely important to remember this PIN in
order to avoid loss of patient data in case of entering wrong
PIN multiple times or if the user forgets the PIN.
Vscan Extend – User Manual1-7
5721203-100Rev. D
Introduction
Important Safety Considerations (continued)
WARNING
WARNING
• Do not short-circuit the battery terminal with metallic
objects. This may result in overheating and burns.
• Do not store or carry a battery loosely with metallic
devices.
• Disconnect the battery charger when not in use to avoid
fire hazard.
• Keep the charger dry. Failure to observe this precaution
may result in fire and electric shock
• Keep this unit out of reach of children. Strangulation
resulting from baby or child entanglement in probe cable
may occur.
Before charging or using a battery it is important that you read
and understand the battery safety and environment information.
• Do not damage the rechargeable battery. A damaged
battery can cause an explosion or fire, and can re su lt in
personal injury and/or property damage. To prevent injury
or damaged do not use or charge the battery if it appears to
be damaged. Signs of damage include, but are not limited
to, discoloration, warping, and leaking batt er y fluid . Do no t
expose the battery to fire, high temperature, or direct
sunlight. Do not immerse or expose the battery to water.
Do not use or store the battery inside a vehicle during hot
weather. Do not drop or puncture the battery. Do not open
the battery or short-circuit its contacts.
WARNING
Avoid contact with the rechargeable battery if it appears to be
leaking. Battery fluid is corrosive, and contact with it can result
in personal injury and/or property damage.
To prevent injury or damage:
• If the battery leaks, avoid contact with the battery fluid. If
any liquid from the battery should come in contact with the
eye, immediately wash the eye with plenty of water and
seek medical advice as soon as possible. Do not rub your
eyes!
• If battery fluid gets onto your skin or clothing, immediately
use clean water to wash off the battery fluid.
1-8Vscan Extend – User Manual
5721203-100
Rev. D
Important Safety Considerations (continued)
Warnings
WARNING
CAUTION
Charge and use the rechargeable battery only in strict
accordance with the instructions. Charging or using the battery
in unauthorized equipment can cause an explosion or fire, and
can result in personal injury and/or equipment damage.
To prevent injury or damage:
• Do not charge or use the battery if it appears to be
damaged or leaking.
• Charge the battery only in a Vscan Extend device or in the
Vscan Extend battery charger. Be sure to follow all
instructions that are provided with the battery charger.
• Discontinue charging a battery that gives off extreme heat
or a burning odor.
• Use the battery only in the Vscan Extend.
• Use the battery only for its intended use and according to
the instructions in the product documentation.
• Ensure to backup data regularly and to erase the data
before sending the Vscan Extend for service.
• To access or add patient information, a device pin is
required. Ensure not to lose the pin.
CAUTION
Vscan Extend – User Manual1-9
5721203-100Rev. D
Vscan Extend cannot be sterilized. Apply a sterile sheath if the
device is going to be used in an application where a higher
degree of cleanliness is required. The sheath should cover the
probe and any part of the probe cable that might come in
contact with the area requiring higher cleanliness.
Introduction
Contacting GE Ultrasound
For additional information or assistance, please contact your
local distributor or the appropriate support resource listed on the
following pages:
TEL: (1) 800-437-1171
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
Please contact your local Applications or Sales Representative.
For service contact your local Service Representative.
1-10Vscan Extend – User Manual
5721203-100
Rev. D
Accessories Catalog Requests
To request the latest GE Accessories catalog or equipment
brochures in the United St ates, call the Response Center
TEL: (1) 800-643-6439
In other locations, contact your local Applications, Sales or
Service Representative.
Placing an order
To place an order, order supplies or ask an accessory-related
question in the United States, call the GE Access Center
TEL: (1) 800-472-3666
In other locations, contact your local Applications, Sales or
Service Representative.
Global ultrasound support center phone numbers
For countries not listed in the tables below, please contact the
local distributor.
Contact Information
When contacting Support you will have to provide your system
ID. If the system ID is unknown, please give the Temporary
System ID “Vscan Extend” to be properly routed for support.
Americas
Table 1-1: Americas
RegionTelephone
United States
Canada800-668-0732
Mexico0800 9043400
Puerto Rico0800 4371171
Brazil0800 122345
Argentina0800 3331984
Chile0800 367000
1
For USA only: when contacting GE CARES you will have to provide your system ID. If system ID is
unknown, please give the Temporary System ID “Vscan Extend” to be properly routed for support.
Table 1-2: Europe, Middle East and Africa (Continued)
RegionTelephone
Saudi Arabia800 1243002
South Africa800 111 671
Spain902 400 246
Sweden0201201436
Switzerland0800556958
TurkeyContact local distributor
UAE800 3646
UK0845 8503392
Ukraine+38 044 391 37 56 (57)
Contact Information
Vscan Extend – User Manual1-13
5721203-100Rev. D
Introduction
Asia and Pacific
Table 1-3: Asia and Pacific
RegionTelephone
China8008108188
Hong Kong(852) 21006288
Taiwan0800-021-770
India(91) 1800-425-8025
Singapore(65) 62773444
Australia1-800-659-465
New Zealand0800 65 94 65
Japan0120-055-919
Korea(82) 2-1544-6119
Bangladesh(880) 29135488
Sri Lanka(94) 114891178
BhutanContact GE India
MaldivesContact GE India
NepalContact local distributor
Malaysia1800 88 3911
Thailand(66) 26248400
VietnamContact local distributor
PhilippinesContact local distributor
IndonesiaContact local distributor
LaosContact local distributor
Brunei DarussalamContact local distributor
CambodiaContact local distributor
Contents:
‘Introduction’ on page 2-2
‘Owner responsibility’ on page 2-3
‘Important safety considerations’ on page 2-5
‘Maximum probe temperature’ on page 2-24
‘Device labels and symbols’ on page 2-25
Vscan Extend – User Manual2-1
5721203-100Rev. D
Safety
Overview
Introduction
This chapter describes the important safety measures which
should be taken before operating the Vscan Extend. Procedures
for simple care and maintenance of the Vscan Extend are also
described.
Various levels of safety precautions may be foun d on the
equipment, and different levels of severity are identified by one
of the following icons that precede precautionary statements in
the text.
The following icons are used to indicate pr ec au tio ns :
DANGER
WARNING
CAUTION
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions will cause:
• Severe or fatal personal injury
• Substantial property damage.
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions may cause:
• Severe personal injury
• Substantial property damage.
Indicates that a potential hazard may exist which through
inappropriate conditions or actions will or can cause:
• Minor injury
• Property damage.
2-2Vscan Extend – User Manual
5721203-100
Rev. D
Overview
Owner responsibility
Owner responsibility
It is the responsibility of the owner to ensure that anyone
operating Vscan Extend reads and understands this section of
the manual. However, there is no representation that the act of
reading this manual renders the reader qualified to operate,
inspect, test, align, calibrate, troubleshoot, repair or modify the
system. The owner should make certain that only properly
trained, fully-qualified service personnel undert ake maintenance
of the equipment. There are no user serviceable parts in the
system or accessories. If servicing is required, contact GE. See
‘Contact Information’ on page 1-10 for more information.
The owner of Vscan Extend should ensure that only properly
trained, fully qualified personnel are authorized to operate the
system. Before authorizing anyone to operate the system, it
should be verified that the person has read, and fully
understands, the operating instructions contained in this
manual. It is advisable to maintain a list of authorized operators.
Should the system fail to operate correctly, or if Vscan Extend
does not respond to the commands described in this manual,
the operator should contact the nearest field GE Ultrasound
Service Office.
For information about specific requirements and regulations
applicable to the use of electronic medical equipment, consult
the local, state and federal agencies.
Vscan Extend – User Manual2-3
5721203-100Rev. D
Safety
Overview (continued)
The owner of Vscan Extend MUST be aware of the data
protection policies while configuring the DICOM and Windows
Share servers. GE is not responsible for data sharing. Bef ore
configuring the servers, the owner of the Vscan Extend MUST
abide by the applicable privacy acts specific to their region or
country.
CAUTION
CAUTION
For USA only:
Federal law restricts this device to sale by or on the order of a
physician.
Vscan Extend should be used in compliance with law. Some
jurisdictions restrict certain uses, such as gender
determination.
Notice against user modification
Never modify this product, including system components,
cables, and so on. User modification may cause safety hazards
and degradation in system performance. All modification must
be done by a GE qualified person.
Software upgrade following GE recommendations can be done
by the user.
2-4Vscan Extend – User Manual
5721203-100
Rev. D
Loading...
+ 221 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.