Basic Service Manual
Direction Number: 5789130-100 English
Rev.5
ADVANCED SERVICE DOCUMENTATION. PROPERTY OF
GE HEALTHCARE. USE OF THESE MATERIALS LIMITED TO AGENTS AND
EMPLOYEES OF GE HEALTHCARE OR OTHER PARTIES EXPRESSLY
LICENSED BY GE HEALTHCARE. UNLICENSED USE IS STRICTLY
PROHIBITED.
GE Medical Systems (China) Co., Ltd.
No. 19, Changjiang Road
WuXi National Hi-Tech Dev. Zone
214028 Jiangsu, China
TEL: +86 510 85225888; FAX: +86 510 85226688
www.gehealthcare.com
Please verify that you are using the latest revision of this document. Information
pertaining to this document is maintained on MyWorkshop/ePDM (GE electronic Product
Data Management). If you need to know the latest revision, contact your distributor, local
GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at
1 800 682 5327 or 1 262 524 5698.
Versana Premier – Basic Service Manual System view i-1
5789130-100 English Rev.5
Page 4
Important precautions
Translation policy
i-2 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 5
Versana Premier – Basic Service Manual System view i-3
5789130-100 English Rev.5
Page 6
i-4 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 7
Versana Premier – Basic Service Manual System view i-5
5789130-100 English Rev.5
Page 8
i-6 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 9
Versana Premier – Basic Service Manual System view i-7
5789130-100 English Rev.5
Page 10
i-8 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 11
Versana Premier – Basic Service Manual System view i-9
5789130-100 English Rev.5
Page 12
i-10 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 13
Versana Premier – Basic Service Manual System view i-11
5789130-100 English Rev.5
Page 14
i-12 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 15
Versana Premier – Basic Service Manual System view i-13
5789130-100 English Rev.5
Page 16
Damage in transportation
All packages should be closely examined at time of delivery. If
damage is apparent, write “Damage In Shipment” on ALL copies
of the freight or express bill BEFORE delivery is accepted or
“signed for” by a GE representative or hospital receiving agent.
Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14
days after receipt, and the contents and containers held for
inspection by the carrier. A transportation company will not pay a
claim for damage if an inspection is not requested within this 14
day period.
Certified electrical contractor statement - For USA Only
All electrical Installations that are preliminary to positioning of
the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. Other connections
between pieces of electrical equipment, calibrations and testing
shall be performed by qualified GE personnel. In performing all
electrical work on these products, GE will use its own specially
trained field engineers. All of GE’s electrical work on these
products will comply with the requirements of the applicable
electrical codes.
The purchaser of GE equipment shall only utilize qualified
personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed
electricians) to perform electrical servicing on the equipment.
i-14 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 17
Omission and errors
WARNING
If there are any omissions, errors or suggestions for improving
this documentation, contact the GE Global Documentation
Group with specific information listing the system type, manual
title, part number, revision number, page number and
suggestion details.
Mail the
information to:
GE Medical Systems (China) Co., Ltd.
No. 19 Changjiang Road
Wuxi National Hi-Tech Dev. Zone
214028 Jiangsu, China
TEL: +86 510 85225888; FAX: +86 510 85226688
GE employees should use TrackWise to report service
documentation issues.
These issues will then be in the internal problem reporting tool
and communicated to the writer.
Service Safety Considerations
DANGERDANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH,
ARE PRESENT IN THIS EQUIPMENT. USE EXTREME
CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
Use all Personal Protection Equipment (PPE) such as gloves,
safety shoes, safety glasses, and kneeling pad, to reduce the
risk of injury.
For a complete review of all safety requirements, refer to
Chapter 1 in the Service Manual.
Versana Premier – Basic Service Manual System view i-15
5789130-100 English Rev.5
Page 18
Legal notes
Proprietary to GE
The contents of this publication may not be copied or duplicated
in any form, in whole or in part, without prior written permission
of GE.
GE may revise this publication from time to time without written
notice.
Permission to use this Advanced Service Software and related
documentation (herein called the material) by persons other
than GE employees is provided only under an Advanced
Service Package License relating specifically to this Proprietary
Material. This is a different agreement from the one under which
operating and basic service software is licensed. A license to
use operating or basic service software does not extend to or
cover this software or related documentation.
If you are a GE employee or a customer who has entered into
such a license agreement with GE to use this proprietary
software, you are authorized to use this Material according to
the conditions stated in your license agreement.
However, you do not have the permission of GE to alter,
decompose or reverse-assemble the software, and unless you
are a GE employee, you may not copy the Material. The
Material is protected by Copyright and Trade Secret laws; the
violation of which can result in civil damages and criminal
prosecution.
If you are not party to such a license agreement or a GE
Employee, you must exit this Material now.
Trademarks
All products and their name brands are trademarks of their
respective holders.
Versana Premier – Basic Service Manual System view i-23
5789130-100 English Rev.5
Page 26
i-24 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 27
Chapter 1
Introduction
This chapter describes important issues related to
safely servicing the Ultrasound system. The service
provider must read and understand all the information
presented here before installing or servicing the units.
Versana Premier – Basic Service Manual System view 1-1
5789130-100 English Rev.5
Page 28
Introduction
1-1-1 Contents in this chapter
• 1-1 ‘Overview’ on page 1-2
• 1-2 ‘Manual Overview’ on page 1-3
• 1-3 ‘Important conventions’ on page 1-7
• 1-4 ‘Product icons’ on page 1-11
• 1-5 ‘Labels locations’ on page 1-12
• 1-6 ‘Safety considerations’ on page 1-14
• 1-7 ‘Dangerous procedure warnings’ on page 1-22
• 1-8 ‘Lockout/Tagout (LOTO) requirements’ on page 1-23
• 1-9 ‘Returning probes and repair parts’ on page 1-24
• 1-10 ‘EMC, EMI and ESD’ on page 1-25
• 1-11 ‘Customer assistance’ on page 1-27
1-1 Overview
1-2 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 29
This manual provides installation and service information for the
Versana Premier Ultrasound system. It is divided in ten chapters
as shown below.
1-2-1 Contents in this manual
The manual is divided into ten chapters.
In the beginning of the manual, before chapter 1, you will find
the Revision overview, the Important precautions including
Translation policy, Damage in transportation, Certified electrical
contractor statement, Omission & errors, Service safety
considerations and Legal notes, and the Table of Contents
(TOC).
Manual Overview
1-2 Manual Overview
Chapter
number
1.
2.
3.
4.
5.
6.
An Index has been included after Chapter 10.
Table 1-1: Contents in this manual
Chapter titleDescription
IntroductionContains a content summary and warnings.
Site preparationsContains pre-setup requirements for the Versana
Premier.
System SetupContains setup procedure with procedure
checklist for the system.
General Procedures
and Functional
Checks
Components and
Functions (Theory)
Service AdjustmentsContains instructions on how to make any
Contains functional checks that must be
performed as part of the installation, or as
required during servicing and periodic
maintenance.
Contains block diagrams and functional
explanations of the electronics.
available adjustments to the Versana Premier.
Versana Premier – Basic Service Manual System view 1-3
5789130-100 English Rev.5
Page 30
Introduction
Table 1-1: Contents in this manual (Continued)
Chapter
Chapter titleDescription
number
7.
8.
9.
Diagnostics/
Troubleshooting
Replacement
procedures
Provides procedures for running diagnostic or
related routines for the Versana Premier.
Provides disassembly procedures and
reassembly procedures for all changeable FRU.
Renewal PartsContains a complete list of replacement parts for
Versana Premier.
10.
Care & MaintenanceProvides periodic maintenance procedures for
Versana Premier.
N/A
IndexA quick way to the topic you’re looking for.
1-2-2 Typical users of the Proprietary Service Manual
• GEHC Service Personnel (setup, maintenance, etc.)
• GEHC Online Center Personnel
• Licensed Hospital’s Service Providers
1-2-3 Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood
before operating the Versana Premier and also kept near the
unit for quick reference.
NOTE: Probe information displayed on screen does not necessarily
reflect the probes available on your ultrasound system. Please
refer to the probe list for available probes and features.
1-4 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 31
Manual Overview
1-2-4 Versana Premier models covered by this manual
Table 1-2: Versana Premier Model Designations
Model NameSystem SW
Versana Premier VAR1.x.x
Versana Premier VSR1.x.x
NOTE: When not otherwise specified, the contents of this manual
applies to all Versana Premier models.
Versana Premier – Basic Service Manual System view 1-5
5789130-100 English Rev.5
Page 32
Introduction
1-2-5 Smart Manual Function
On each page of this Proprietary Service Manual, there is a
smart manual icon to bring you back to Overview of Versana
Premier. By clicking on the system overview list, you can be
brought to the FRU list directly.
Figure 1-1. Smart Manual Icon
1-6 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 33
1-3 Important conventions
1-3-1 Conventions used in book
Important conventions, used in this document, are described
next.
1-3-1-1 Model designations
This manual covers the Versana Premier Ultrasound systems
listed in:
1-2-4 ‘Versana Premier models covered by this manual’ on
page 1-5.
1-3-1-2 Icons
Important conventions
Pictures, or icons, are used wherever they will reinforce the
printed message. The icons, labels, and conventions used on
the product and in the service information are described in this
chapter.
Versana Premier – Basic Service Manual System view 1-7
5789130-100 English Rev.5
Page 34
Introduction
DANGER
WARNING
CAUTION
1-3-1-3 Safety precaution messages
Various levels of safety precaution messages may be found on
the equipment and in the service information. The different
levels of concern are identified by a flag word that precedes the
precautionary message. Known or potential hazards to
personnel are labeled in one of three ways:
• DANGER
• WARNING
• CAUTION
Danger is used to indicate the presence of a hazard that will
cause severe personal injury or death if the instructions are
ignored.
Warning is used to indicate the presence of a hazard that can
cause severe personal injury and property damage if
instructions are ignored.
Caution is used to indicate the presence of a hazard that will or
can cause minor personal injury and property damage if
instructions are ignored. Equipment damage possible.
NOTE: Notes are used to provide important information about an item
or a procedure.
NOTE: Be sure to read the notes; the information contained in a note
can often save you time or effort.
1-8 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 35
1-3-2 Standard hazard icons
Important information will always be preceded by either the
exclamation point (!) contained within a triangle, or the symbols
for “Danger”, “Warning” or “Caution”, as seen throughout this
chapter. In addition to text, several different graphical icons
(symbols) may be used to make you aware of specific types of
hazards that could possibly cause harm.
Important conventions
Table 1-3: Standard hazard icons
ELECTRICAL
MECHANICAL
RADIATION
LASER
HEAT
PINCH
NOTE: Even if a symbol isn’t used on the product or in this manual, it
may be included for your reference.
Versana Premier – Basic Service Manual System view 1-9
5789130-100 English Rev.5
Page 36
Introduction
1-3-2-1 Standard Icons that indicate that a special procedure is to be used
Some others icons make you aware of specific procedures that
should be followed.
Table 1-4: Standard Icons that indicates that a special procedure is to be used
Avoid Static ElectricityTag and Lock OutWear Eye Protection
Hand ProtectionFoot ProtectionWear Eye Protection
Be sure to read the notes; the information contained in a note
can often save you time or effort.
1-10 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 37
Product icons
1-4 Product icons
It is important to refer to the current revision of the Ultrasound
system’s Basic User Manual for a full list of product labels prior
to servicing the system.
Versana Premier – Basic Service Manual System view 1-11
5789130-100 English Rev.5
Page 38
Introduction
It is important to refer to the current revision of the Ultrasound
system’s Basic User Manual for a full list of product labels prior
to servicing the system.
1-5-1 Rating Plate Location
Versana Premier labels are provided in English.
The labels are at the rear of the system. The label content may
be different for different regions. Please refer to the labels on the
system for the actual content.
1-5 Labels locations
Figure 1-2. Versana Premier Label Location
1-12 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 39
1-5-1 Rating Plate Location(continued)
1. Rating Plate
Table 1-5: Label Icons
Label/IconPurpose/MeaningLocation
Labels locations
Every system has a unique
marking for identification, the
Unique Device Identification
(UDI) Label. The UDI label
consists of a series of
alpha-numeric characters and
barcode which uniquely identify
the Versana Premier system as a
medical device manufactured by
General Electric. Scan or enter
the UDI information into the
patient health record as required
by country-specific laws.
Serial Number.
Rating plate
Rating plate
NOTE: If the new label is needed during the service activities, please
click “Ask an Expert” to submit the case in the support central:
http://supportcentral.ge.com/products/sup_products.asp?prod_
id=44177.
Versana Premier – Basic Service Manual System view 1-13
5789130-100 English Rev.5
Page 40
Introduction
1-6 Safety considerations
1-6-1 Contents in this section
• 1-6-2 ‘Introduction’ on page 1-14
• 1-6-3 ‘Human Safety’ on page 1-14
• 1-6-4 ‘Mechanical safety’ on page 1-17
• 1-6-5 ‘Electrical safety’ on page 1-20
1-6-2 Introduction
The following safety precautions must be observed during all
phases of operation, service and repair of this equipment.
Failure to comply with these precautions or with specific
warnings elsewhere in this manual, violates safety standards of
design, manufacture and intended use of the equipment.
1-6-3 Human Safety
• Operating personnel must not remove the system covers.
• Servicing should be performed by authorized personnel
only.
Only personnel who have participated in a Versana Premier
Training Seminar are authorized to service the equipment.
1-14 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 41
1-6-3 Human Safety(continued)
WARNING
WARNING
WARNING
Because of the limited access to cabinets and equipment in the
field, placing people in awkward positions, GE has limited the
lifting weight for one person in the field to 16 KG (35 LBS).
Anything over 16 KG (35 LBS) requires 2 people.
Have two people available to deliver and unpack the Versana
Premier.
Attempts to move the Ultrasound system considerable
distances or on an incline by one person could result in injury
or damage or both.
Ensure that the Ultrasound system is turned off and unplugged.
Wait for at least 20 seconds for capacitors to discharge as
there are no test points to verify isolation. The amber light on
the OP panel ON/OFF button will turn off.
Safety considerations
Ultrasound system components may be energized. Always
refer to the Ultrasound system's Proprietary Service Manual for
LOTO warnings and cautions.
Versana Premier – Basic Service Manual System view 1-15
5789130-100 English Rev.5
Page 42
Introduction
WARNING
WARNING
WARNING
WARNING
1-6-3 Human Safety(continued)
Risk of electrical shock, Ultrasound system must be turned off
and disconnected from power source. Cord must be controlled
at all times.
Wait for at least 20 seconds for capacitors to discharge as
there are no test points to verify isolation. The amber light on
the OP panel on/off button will turn off.
Ultrasound System components may be energized. Always
refer to the Ultrasound system's Proprietary Service Manual for
LOTO warnings and cautions
Use all Personal Protection Equipment (PPE) such as gloves,
safety shoes, safety glasses, and kneeling pad, to reduce the
risk of injury.
Beware of possible sharp edges on all mechanical parts. If
sharp edges are encountered, the appropriate PPE should be
used to reduce the risk of injury.
Wear all PPE including gloves as indicated in the chemical
MSDS.
1-16 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 43
1-6-4 Mechanical safety
WARNING
WARNING
WARNING
WARNING
CAUTION
CAUTION
While the software install procedure is designed to preserve
data, you should save any patient data, images, system setups
to removable media or hardcopy before doing a software
upgrade.
Ultrasound probes are highly sensitive medical instruments
that can easily be damaged by improper handling. Use care
when handling and protect from damage when not in use. Do NOT use a damaged or defective probe. Failure to follow these
precautions can result in serious injury and equipment
damage.
Never use a probe that has fallen to the floor. Even if it looks
OK, it may be damaged.
Safety considerations
When the Ultrasound system is raised for a repair or moved
along any incline, use extreme caution since it may become
unstable and tip over.
Take extra care when moving the system.
The Versana Premier weighs approximately 70 kg (140 lbs) or
more, depending on installed peripherals, when ready for use.
To avoid possible injury and equipment damage when
transporting from one area of use to another:
• Be sure the pathway is clear.
• Limit movement to a slow careful walk.
• Use two or more persons to move the equipment on
inclines or long distance.
Before you move or transport the Ultrasound system, make
sure to lock the monitor arm firmly and flip down the monitor to
prevent damage to the Ultrasound system.
Versana Premier – Basic Service Manual System view 1-17
5789130-100 English Rev.5
Page 44
Introduction
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
1-6-4 Mechanical safety(continued)
To avoid injury when you move the monitor and the monitor
arm, do not put your finger, hand, or object on the joint of the
monitor or the monitor arm.
Ensure that nobody touches the console arm when moving the
operator panel.
Do not move the Ultrasound system if the Operator Panel is in
unlocked position.
Do not transport Versana Premier in a vehicle without locking
the casters (wheels) and securing it as described in chapter 4.
Use protective glasses during drilling, filing smooth surfaces,
and during all other work where eyes need protection.
Use safety shoes when doing work where there is any chance
of foot injury.
1-18 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 45
1-6-4 Mechanical safety(continued)
CAUTION
CAUTION
Use protective gloves when working with sharp edges or when
directed to wear PPE during a removal/replacement procedure.
Be careful not to pinch any of the cables.
NOTE: Special care should be taken when transporting the Ultrasound
system in a vehicle:
• Before transporting, place the system in its special storage
case.
• Lock the wheels (brake).
• Ensure that the system is firmly secured while inside the
vehicle.
• Secure system with straps or as directed otherwise to
prevent motion during transport.
• Prevent vibration damage by driving cautiously. Avoid
unpaved roads, excessive speeds, and erratic stops or
starts.
Safety considerations
Versana Premier – Basic Service Manual System view 1-19
5789130-100 English Rev.5
Page 46
Introduction
WARNING
WARNING
1-6-5 Electrical safety
1-6-5-1 Safe practices
Follow these guidelines to minimize shock hazards whenever
you are using the Ultrasound system:
• To minimize shock hazard, the equipment chassis must be
• The Ultrasound system is equipped with a three-conductor
• The power outlet used for this equipment should not be
• Both the system power cable and the power connector must
Connecting a Versana Premier to the wrong voltage level will
most likely destroy it.
connected to an electrical ground.
AC power cable. This must be plugged into an approved
electrical outlet with safety ground.
shared with other types of equipment.
meet international electrical standards.
DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU
UNIT LEVEL AT ANY CIRCUMSTANCES.
1-20 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 47
1-6-5-2 Probes
Safety considerations
Follow these guidelines before connecting a probe to the
Ultrasound system:
• Inspect the probe prior to each use for damage or
degradation to the:
• housing
• cable strain relief
• lens
• seal
• connector pins
• locking mechanism
• Do not use a damaged or defective probe.
• Never immerse the probe connector or adapter into any
liquid.
• The system has more than one type of probe port. Use the
appropriate probe port designed for the probe you are
connecting.
1-6-5-3 Peripherals
Refer to the Patient Safety Environment section of the User’s
Manual for peripheral isolation information.
Versana Premier – Basic Service Manual System view 1-21
5789130-100 English Rev.5
Page 48
Introduction
DANGER
WARNING
WARNING
WARNING
WARNING
WARNING
1-7 Dangerous procedure warnings
Warnings, such as the example below, precede potentially
dangerous procedures throughout this manual. Instructions
contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING
DEATH, ARE PRESENT IN THIS EQUIPMENT. USE
EXTREME CAUTION WHEN HANDLING, TESTING
AND ADJUSTING.
If the covers are removed from an operating Versana Premier,
some metal surfaces may be warm enough to pose a potential
heat hazard if touched, even while in shutdown mode.
Explosion Warning
DO NOT operate the equipment in an explosive atmosphere.
Operation of any electrical equipment in such an environment
constitutes a definite safety hazard.
DO NOT substitute parts or modify equipment
Because of the danger of introducing additional hazards, ONLY
install GE approved parts. DO NOT perform any unauthorized
modification of the equipment.
SHUT DOWN FORCEDLY OR PLUG IN/OUT ACDC
INVALID MAY CAUSE THE DAMAGE OF SYSTEM
FILES.
AFTER UNPLUG POWER CORD, WAIT FOR AT LEAST
20 SECONDS FOR CAPACITORS TO DISCHARGE AS
THERE ARE NO TEST POINTS TO VERIFY
ISOLATION.
1-22 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 49
Lockout/Tagout (LOTO) requirements
WARNING
1-8 Lockout/Tagout (LOTO)
requirements
Follow Lockout/Tagout requirements by ensuring you are in total
control of the AC power plug at all times during the service
process.
To apply Lockout/Tagout (LOTO):
1. Plan and prepare for shutdown.
2. Shutdown the equipment.
3. Isolate the equipment.
4. Remove/disconnect the battery, if present.
5. Apply Lockout/Tagout Devices.
6. Control all stored and residual energy.
7. Verify isolation.
All potentially hazardous stored or residual energy is relieved.
Energy Control and Power Lockout for Versana Premier.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.
Versana Premier – Basic Service Manual System view 1-23
5789130-100 English Rev.5
Page 50
Introduction
1-9 Returning probes and repair
parts
Equipment being returned must be clean and free of blood and
other infectious substances. GE policy states that body fluids
must be properly removed from any part or equipment prior to
shipment. GE employees, as well as customers, are responsible
for ensuring that parts/equipment have been properly
decontaminated prior to shipment. Under no circumstance
should a part or equipment with visible body fluids be taken or
shipped from a clinic or site (for example, body coils or an
ultrasound probe).
The purpose of the regulation is to protect employees in the
transportation industry, as well as the people who will receive or
open this package.
NOTE: The US Department of Transportation (DOT) has ruled that
“items that were saturated and/or dripping with human blood
that are now caked with dried blood; or which were used or
intended for use in patient care” are “regulated medical waste”
for transportation purposes and must be transported as a
hazardous material.
NOTE: The USER/SERVICE staff should dispose of all the waste
properly, per federal, state, and local waste disposal regulations.
The Ultrasound system is not meant to be used for long-term
storage of patient data or images. The user is responsible for the
data on the system and a regular backup is highly
recommended.
If the system is sent for repair, please ensure that any patient
information is backed up and erased from the system before
shipping. It is always possible during system failure and repair to
lose patient data. GE is not responsible for the loss of this data.
If PHI (Patient Healthcare Information) data needs to be sent to
GE employees for service purposes, GE will ascertain
agreement from the customer. Patient information shall only be
transferred by approved service processes, tools and devices
restricting access, protecting or encrypting data where required,
and providing traceability in the form of paper or electronic
documents at each stage of the procedure while maintaining
compliance with cross-border restrictions of patient information
transfers.
1-24 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 51
1-10 EMC, EMI and ESD
1-10-1 Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance
of a device within its electromagnetic environment. This
environment consists of the device itself and its surroundings
including other equipment, power sources and persons with
which the device must interface. Inadequate compatibility results
when a susceptible device fails to perform as intended due
interference from its environment or when the device produces
unacceptable levels of emission to its environment. This
interference is often referred to as radio–frequency or
electromagnetic interference (RFI/EMI) and can be radiated
through space or conducted over interconnecting power of
signal cables. In addition to electromagnetic energy, EMC also
includes possible effects from electrical fields, magnetic fields,
electrostatic discharge and disturbances in the electrical power
supply.
EMC, EMI and ESD
1-10-2 CE Compliance
Versana Premier conforms to all applicable conducted and
radiated emission limits and to immunity from electrostatic
discharge, radiated and conducted RF fields, magnetic fields
and power line transient requirements.
For applicable standards, refer to the Safety Chapter of the
Ultrasound system User’s Manual.
NOTE: For CE Compliance, it is critical that all covers, screws,
shielding, gaskets, mesh, clamps, are in good condition,
installed tightly without skew or stress. Proper installation
following all comments noted in this service manual is required
in order to achieve full EMC performance.
Versana Premier – Basic Service Manual System view 1-25
5789130-100 English Rev.5
Page 52
Introduction
WARNING
WARNING
1-10-3 Electrostatic discharge (ESD) prevention
DO NOT touch any boards with integrated circuits prior to
taking the necessary ESD precautions.
Always connect yourself, via an arm-wrist strap, to the advised
ESD connection point located on the rear of the Ultrasound
system (near the power connector).
Follow general guidelines for handling of electrostatic sensitive
equipment.
Risk of electrical shock, Ultrasound system must be turned off.
Avoid all contact with electrical contacts, conductors and
components. Always use non-conductive handles designed for
the removal and replacement of ESD sensitive parts. All parts
that have the potential for storing energy must be discharged or
isolated before making contact.
1-26 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 53
1-11 Customer assistance
1-11-1 Contact information
If this equipment does not work as indicated in this service
manual or in the basic user manual, or if you require additional
assistance, please contact the local distributor or appropriate
support resource, as listed below.
Before you call, identify the following information, and acquire
log (Alt+D) to send to the Customer Care team:
1. System ID / serial number.
2. Software version.
3. Date and time of occurrence.
4. Sequence of events leading to issue.
5. Is the issue repeatable?
6. Imaging mode, probe, preset/application.
7. Media brand, speed, capacity, type.
8. Save secondary image capture, cine loop.
Customer assistance
NOTE: Restart the application before resuming clinical scanning.
NOTE: The serial number can be found at the rear of the system.
Versana Premier – Basic Service Manual System view 1-27
5789130-100 English Rev.5
Page 54
Introduction
1-11-2 Phone numbers for Customer Assistance
Table 1-6: Phone numbers for Customer Assistance
LOCATIONPHONE NUMBER
USA
GE Healthcare - GE Medical Systems
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
CanadaPhone:1-800-668-0732
Latin AmericaService
Europe (OLC-EMEA)
GE Ultraschall Deutschland Gmbh & Co. KG
Beethovenstraße 239
Postfach 11 05 60, D-42655 Solingen
Germany
GE Medical Systems (China) Co., Ltd.
No.19 Changjiang Road
WuXi National Hi-Tech Dev.Zone
214028 Jiangsu
China
+86 510 85225888+86 510 85226688
1-28 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 55
1-11-4 Authorized Representative
Table 1-8: Authorized Representative
AUTHORIZED REPRESENTATIVE
The location of the CE marking is shown in the Safety chapter of the Basic User
Manual.
Authorized EU Representative:
GE Medical Systems SCS
283 rue de la Minière 78530 BUC, France
1-11-5 Factory Site
Table 1-9: Factory Site
Factory Site
Customer assistance
TELEPHONE / FAX
NUMBER
TELEPHONE / FAX
NUMBER
GE Medical Systems (China) Co., Ltd.
No.19, Changjiang Road, Wuxi National Hi-Tech Dev.Zone,
214028 Jiangsu, China
TEL: +86 510-85225888
FAX: +86 510-85226688
Versana Premier – Basic Service Manual System view 1-29
5789130-100 English Rev.5
Page 56
Introduction
1-30 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 57
Chapter 2
Site Preparations
This chapter provides the information required to plan
and prepare for the setup of an Ultrasound system.
Included the descriptions of the facility and electrical
needs to be met by the purchaser of the units.
Versana Premier – Basic Service Manual System view 2-1
5789130-100 English Rev.5
Page 58
Site Preparations
2-1-1 Contents in this chapter
• 2-1 ‘Overview’ on page 2-2
• 2-2 ‘General Ultrasound system requirements’ on page 2-3
• 2-3 ‘Facility needs’ on page 2-12
• 2-4 ‘Environmental Dangers’ on page 2-21
2-1 Overview
2-2 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 59
2-2 General Ultrasound system
CAUTION
2-2-1 Contents in this section
• 2-2-2 ‘Ultrasound system environmental requirements’ on
page 2-3
• 2-2-3 ‘Electrical requirements’ on page 2-6
• 2-2-4 ‘EMI limitations’ on page 2-9
• 2-2-5 ‘Probes environmental requirements’ on page 2-11
• 2-2-6 ‘Time and manpower requirements’ on page 2-11
General Ultrasound system requirements
requirements
2-2-2 Ultrasound system environmental requirements
2-2-2-1 If the Ultrasound system is very cold or hot
When unpacking the Ultrasound system, allow the temperature
of the Ultrasound system to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 2-1: System Acclimation Time Chart
Degree C-4.5-20.534042.54547.5505560
Degree F23.928.432.937.4104108.5113117.5122131140
hours
3
2
1
0 0 1 23 468
Versana Premier – Basic Service Manual System view 2-3
5789130-100 English Rev.5
Page 60
Site Preparations
CAUTION
CAUTION
2-2-2-2 Environmental specifications for Ultrasound system
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
Table 2-2: System Environmental Requirements
Temperature
Humidity
Pressure
Operational (Versana
Premier with probe)
3- 40° C (without
battery)
Storage (Versana
Premier)
-5 - 50° C
23 - 122° F
Transport (Versana
Premier)
-5 - 50° C
23 - 122° F
37.4 - 104° F
10 - 30° C (with battery)
50 - 86° F
30 - 80%
non-condensing
10 - 90%
non-condensing
10 - 90%
non-condensing
700 - 1060hPa700 - 1060hPa700 - 1060hPa
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
The Versana Premier system and probe connector is not
waterproof. Do not expose the device to water or any kind of
liquid.
2-4 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 61
2-2-2-3 Cooling
NOTE: Each person in the room places an additional 300 BTU/h
2-2-2-4 Lighting
General Ultrasound system requirements
The cooling requirement for a console Ultrasound system with
monitor and on board peripherals, is up to 3800 BTU/h. This
figure does not include cooling needed for lights, people, or
other equipment in the room.
demand on the cooling system.
Bright light is needed for Ultrasound system installation, updates
and repairs. However, operator and patient comfort may be
optimized if the room light is subdued and indirect. Therefore a
combination lighting system (dim/bright) is recommended. Keep
in mind that lighting controls and dimmers can be a source of
EMI which could degrade image quality. These controls should
be selected to minimize possible interference.
Versana Premier – Basic Service Manual System view 2-5
5789130-100 English Rev.5
Page 62
Site Preparations
2-2-3 Electrical requirements
2-2-3-1 General requirements
NOTE: GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
2-2-3-2 Electrical requirements for the Ultrasound system
In the table below, the electrical specifications for the Ultrasound
system includes monitor and on board peripherals.
Table 2-3: Electrical Specifications for Versana Premier system
Vol tag e
100-240 VAC±10%Max. 450VA50/ 60HZ
Toleranc ePower ConsumptionFrequency
:
:
2-2-3-3 Site power outlets
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
2-6 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 63
2-2-3-4 Site circuit breaker
CAUTION
It is recommended that the branch circuit breaker for the
Ultrasound system be readily accessible.
POWER OUTAGE MAY OCCURE.
The Versana Premier requires a dedicated single branch
circuit. To avoid circuit overload and possible loss of critical
care equipment, make sure you DO NOT have any other
equipment operating on the same circuit.
2-2-3-5 Site power outlets
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
General Ultrasound system requirements
2-2-3-6 Unit power plug
If the Ultrasound system arrives without a power plug, or with
the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
Versana Premier – Basic Service Manual System view 2-7
5789130-100 English Rev.5
Page 64
Site Preparations
2-2-3-7 Power stability requirement
Table 2-4: Power stability requirement
IEC 61000-4-11
Voltage dips, short
interruptions and
voltage variations
on mains supply
< 5%T (> 95% dip) for
0.5 cycle;
40%T (60% dip) for 5
cycles;
70%T (30 dip) for 25
cycles;
< 5%T (>95% dip) for 5
sec
< 5%T (> 95% dip) for
0.5 cycle;
40%T (60% dip) for 5
cycles;
70%T (30 dip) for 25
cycles;
< 5%T (>95% dip) for 5
sec
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
2-8 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 65
2-2-4 EMI limitations
General Ultrasound system requirements
Ultrasound systems are susceptible to Electromagnetic
Interference (EMI) from radio frequencies, magnetic fields, and
transients in the air or wiring. They also generate EMI. The
Ultrasound system complies with limits as stated on the EMC
label. However there is no guarantee that interference will not
occur in a particular installation.
Possible EMI sources should be identified before the Ultrasound
system is installed.
Electrical and electronic equipment may produce EMI
unintentionally as the result of a defect. Some of these sources
include:
• medical lasers
• scanners
• cauterizing guns
• computers
• monitors
• fans
• gel warmers
• microwave ovens
• light dimmers
• mobile phones
• in-house wireless phones (DECT phones)
• wireless computer keyboard and mouse
• air conditioning system
• High Frequency (HF) surgery equipment
• general AC/DC adapters
• portable phones
The presence of a broadcast station or broadcast van may also
cause interference.
See: 2-2-4-1 ‘EMI prevention/abatement’ on page 2-10 for EMI
prevention tips.
Versana Premier – Basic Service Manual System view 2-9
5789130-100 English Rev.5
Page 66
Site Preparations
2-2-4-1 EMI prevention/abatement
Table 2-5: EMI prevention/abatement
EMI RULEDETAILS
Be aware of Radio
Frequency sources
Ground the
Ultrasound system
Replace all screws,
Radio Frequency
gaskets, covers,
cores
Replace broken
Radio Frequency
gaskets
Do not place labels
where Radio
Frequency gaskets
touch metal
Use GE specified
harnesses and
peripherals
Take care with
cellular phones
Properly route
peripheral cables
• Keep the Ultrasound system at least 5 meters (15 feet) away from other EMI
sources.
• Special shielding may be required to eliminate interference problems caused by
high frequency, high powered radio or video broadcast signals.
Poor grounding is the most likely reason an Ultrasound system will have noisy
images. Check grounding of the power cord and power outlet.
• After you finish repairing or updating the Ultrasound system, replace all covers
and tighten all screws.
• Any cable with an external connection requires a magnet wrap at each end.
• Install all covers. Loose or missing covers or Radio Frequency gaskets allow
radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an Radio Frequency gasket are
broken, replace the gasket. Do not turn on the Ultrasound system until any loose
metallic part is removed.
Where applicable, never place a label where Radio Frequency gaskets meet the
Ultrasound system. Otherwise, the gap created will permit Radio Frequency
leakage. Or, if a label has been found in such a position, move the label.
The interconnect cables are grounded and require ferrite beads and other
shielding. Also, cable length, material, and routing are all important; do not change
from what is specified.
Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Where applicable, do not allow cables to lie across the top of the Card Rack or
hang out of the peripheral bays. Loop the excess length for peripheral cables
inside the peripheral bays. Attach the monitor cables to the frame.
2-10 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 67
General Ultrasound system requirements
CAUTION
CAUTION
CAUTION
2-2-5 Probes environmental requirements
2-2-5-1 Operation, storage and transport temperatures for probes
Probes should be operated, stored, or transported within the
parameters outlined below.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 2-6: Probe Environmental Requirements
OperationalStorageTransport
Temperature10° - 40° C
50° - 104° F
Humidity30- 80%
non-condensing
Pressure700 - 1060hPa700 - 1060hPa700 - 1060hPa
-5° - 50° C
23° - 122° F
10 - 90%
non-condensing
-5° - 50° C
23° - 122° F
10 - 90%
non-condensing
Check the room temperature before you use the probe.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
NOTE: Refer to Table 2-1 on page 2-3 to determine the needed
settlement time.
2-2-6 Time and manpower requirements
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
Versana Premier – Basic Service Manual System view 2-11
5789130-100 English Rev.5
Page 68
Site Preparations
2-3-1 Contents in this section
• 2-3-2 ‘Purchaser responsibilities’ on page 2-13
• 2-3-3 ‘Required facility needs’ on page 2-14
• 2-3-4 ‘Desirable features’ on page 2-15
• 2-3-5 ‘Minimal floor plan suggestion’ on page 2-16
• 2-3-6 ‘Recommended floor plan suggestion’ on page 2-17
• 2-3-7 ‘Suggested floor plan, Ultrasound system, and
EchoPAC PC in same room’ on page 2-18
• 2-3-8 ‘Networking setup requirements’ on page 2-18
2-3 Facility needs
2-12 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 69
2-3-2 Purchaser responsibilities
The work and materials needed to prepare the site is the
responsibility of the purchaser. Delay, confusion, and waste of
manpower can be avoided by completing pre-installation work
before delivery. Purchaser responsibility includes:
• Procuring the materials required
• Completing the preparations before delivery of the
Ultrasound system
• Paying the costs for any alterations and modifications not
specifically provided in the sales contract
NOTE: All electrical installations that are preliminary to the positioning
of the equipment at the site prepared for the equipment must be
performed by licensed electrical contractors. Other connections
between pieces of electrical equipment, calibrations, and testing
must also be performed by qualified personnel. The products
involved (and the accompanying electrical installations) are
highly sophisticated and special engineering competence is
required. All electrical work on these products must comply with
the requirements of applicable electrical codes. The purchaser
of GE equipment must only utilize qualified personnel to perform
electrical servicing on the equipment.
Facility needs
The desire to use a non–listed or customer provided product or
to place an approved product further from the Ultrasound
system than the interface kit allows, presents challenges to the
installation team. To avoid delays during installation, such
variances should be made known to the individuals or group
performing the installation at the earliest possible date
(preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the
Ultrasound system. Carpet is not recommended because it
collects dust and creates static. Potential sources of EMI
(electromagnetic interference) should also be investigated
before delivery. Dirt, static, and EMI can negatively impact
Ultrasound system reliability.
Versana Premier – Basic Service Manual System view 2-13
5789130-100 English Rev.5
Page 70
Site Preparations
2-3-3 Required facility needs
NOTE: GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
:
:
2-14 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 71
2-3-3 Required facility needs(continued)
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound unit is only a conduit.
• Dedicated single branch power outlet of adequate
amperage meeting all local and national codes which is
located less than 2.5 m (8 ft.) from the unit’s proposed
location
• Door opening is at least 76 cm (30 in) wide
• Proposed location for unit is at least 0.5m (1.5 ft.) from the
wall for cooling
• Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
unit to connect cables.
• Power outlets for other medical equipment.
• Power outlets for test equipment within 1 m (3.2 ft.) of
Ultrasound system.
• Clean and protected space to store probes (in their cases or
on a rack)
• Material to safely clean probes (done with a plastic
container, never metal)
Facility needs
For the amperage requirements, see: 2-2-3 ‘Electrical
requirements’ on page 2-6.
2-3-4 Desirable features
• Door is at least 92 cm (3 ft.) wide
• Circuit breaker for dedicated power outlet is easily
accessible
• Sink with hot and cold water
• Receptacle for bio–hazardous waste, like used probe
sheaths
• Emergency oxygen supply
• Storage for linens and equipment
• Nearby waiting room, lavatory, and dressing room
• Dual level lighting (bright and dim)
• Lockable cabinet ordered by GE for its software and
proprietary manuals
Versana Premier – Basic Service Manual System view 2-15
5789130-100 English Rev.5
Page 72
Site Preparations
4
1
3
11
6
13
8
10
5
7
9
12
14
2
2-3-5 Minimal floor plan suggestion
CSI 8x10
Scale:
Each square equals one square
foot (app. 31 x 31 cm)
1. Door – at least 762 mm (30
inches)
2. Film Viewer
3. Counter Top, Sink with hot
and cold water and Supplies
Storage
4. Linen Supply
5. Probes/Supplies
6. Examination Table – 1930 x
610 mm (76 x 24 inches)
7. Footswitch
8. Stool
9. Ultrasound system
10. External Peripherals
11. Dedicated Power Outlet Circuit Breaker protected and
easily accessible
12. Network Interface
13. 457 mm (18 inches) distance
of Ultrasound system from
wall or objects
14. GE Cabinet for Software and
Manuals
Figure 2-1. Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
2-16 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 73
2-3-6 Recommended floor plan suggestion
CSI 14x17
Facility needs
10
11
12
15
9
2
3
4
5
1314
16
8
7
1
17
6
18
Scale: Each square equals one square foot (app. 31 x 31 cm)
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with
hot and cold water
6. Overhead Lights Dimmer Dual Level Lighting (bright
and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance
of Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x
610 mm (76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30
inches)
Figure 2-2. A 14 by 17 foot recommended floor plan
Versana Premier – Basic Service Manual System view 2-17
5789130-100 English Rev.5
Page 74
Site Preparations
2-3-7 Suggested floor plan, Ultrasound system, and EchoPAC PC in
same room
1. EchoPAC PC workstation
parts
2. UPS
3. Ethernet network wall outlet
4. 3x mains power outlets
5. Hot and Cold water
6. Dedicated mains power outlet
7. Ethernet network wall outlet
Figure 2-3. Suggested Room with EchoPAC PC workstation and Ultrasound Scanner
2-3-8 Networking setup requirements
2-3-8-1 Stand alone Ultrasound system (without network connection)
None.
2-3-8-2 Scanner connected to hospital’s network
Supported networks:
100/1000 Mbit Ethernet/DICOM network (option)
2-3-8-3 InSite requirements
InSite requires an Ethernet connection via:
• 100/1000 Mbit Interface
2-18 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 75
2-3-8-4 Purpose of the DICOM network function
DICOM services provide the operator with clinically useful
features for moving images and patient information over a
hospital network.
Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote
printers.
As an added benefit, transferring images in this manner frees up
the on-board monitor and peripherals, enabling viewing to be
done while scanning continues.
With DICOM, images can be archived, stored, and retrieved
faster, easier, and at a lower cost.
Facility needs
Versana Premier – Basic Service Manual System view 2-19
5789130-100 English Rev.5
Page 76
Site Preparations
2-3-8-5 DICOM option setup requirements
To configure the Ultrasound system to work with other network
connections, the site’s network administrator must provide
information to complete the form “Worksheet for DICOM
Network Information”. Ensure that there are no spaces in any
field of the form.
See:
Entries must include:
• A host name, local port number, AE Title, IP address and
Net Mask for the Ultrasound system.
• The IP addresses for the default gateway and other routers
at the site for ROUTING INFORMATION.
• The host name, IP address, port and AE Title for each
device the site wants connected to the Ultrasound system
for DICOM APPLICATION INFORMATION. A field for the
make (manufacturer) and the revision of the device, is also
included. This information may be useful for error solving.
Figure 2-4. Worksheet for DICOM Network Information
2-20 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 77
Environmental Dangers
2-4 Environmental Dangers
Commercial devices such as laser cameras, printers, VCRs and external monitors,
usually exceed allowable leakage current limits and, when plugged into separate AC
outlets, are in violation of patient safety standards. Suitable electrical isolation of such
external AC outlets, or providing the device with extra protective earth, will be required
in order to meet UL60601-1 and IEC60601-1 / IEC60601-1-1 standards for electrical
leakage.
2-4-1 Patient Vicinity UL60601-1 (USA)
2.12.20DV (UL60601-1:2003)
In area in which patients are normally cared for, the patient vicinity is the space with
surfaces likely to be contacted by the patient or attendant who can touch the patient.
This encloses a space within the room 1.83 m (6 ft.) beyond the perimeter of the bed
(examination table, dental chair, treatment booth, and the like) in its intended location,
and extending vertically 2.29 m (7.5 ft.) above the floor.
1. Patient environment
Versana Premier – Basic Service Manual System view 2-21
5789130-100 English Rev.5
Page 78
Site Preparations
2-4-2 Patient Environment IEC60601-1 and ANSI AAMI ES60601-1
Sub Clause 3.79 and figure A.9 (IEC60601-1:2005 and ANSI AAMI
ES60601-1:2005)
Such an area is an environment in which medical diagnosis, monitoring or treatment is
carried out. It is very difficult to attach unique dimensions to the PATIENT
ENVIROMENT.
In practice a distance of 2,5 m (8.2 ft.) above the floor on which the medical personnel
stand and a horizontal distance of 1,5 m (4.9 ft.) have justified themselves as
indicative of the dimensions of the Patient Environment.
The patient environment/vicinity will be depicted as a dashed line in this procedure.
See example below.
1. Patient environment
Figure 2-5.Patient environment
2-22 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 79
Chapter 3
System Setup
This chapter contains information needed to install
Versana Premier system.
Included is a procedure that describes how to receive
and unpack the equipment and how to file a damage or
loss claim.
How to prepare the facility and unit of the actual
installation, and how to check and test the unit, probes,
and external peripherals for electrical safety are also
included in this procedure.
Versana Premier – Basic Service Manual System view 3-1
5789130-100 English Rev.5
Page 80
System Setup
3-1-1 Contents in this chapter
• 3-1 ‘Overview’ on page 3-2
• 3-2 ‘Setup reminders’ on page 3-3
• 3-3 ‘Receiving and unpacking the equipment’ on page 3-6
• 3-5 ‘Preparing for setup’ on page 3-19
• 3-6 ‘Completing the setup’ on page 3-20
• 3-7 ‘System Configuration’ on page 3-25
• 3-8-3 ‘Peripherals Installation Instructions’ on page 3-31
• 3-9 ‘Connectivity setup’ on page 3-67
• 3-11 ‘Paperwork after setup’ on page 3-99
3-1 Overview
3-2 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 81
3-2-1 Average setup time
DANGER
CAUTION
CAUTION
CAUTION
• Unpacking the Versana Premier: 30 minutes
• Set up Versana Premier options: 30 minutes
• DICOM Network Configuration: 30 minutes or more,
depending on the configuration
• Install Insite: 0.5 hour
The Versana Premier installation and functional checkout will
take approximately 1 hour. Versana Premier consoles with
optional equipment may take slightly longer.
3-2-2 Setup warnings
Setup reminders
3-2 Setup reminders
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE
OF OPENING THE AC GROUND LINE (I.E., METER’S
GROUND SWITCH IS OPEN), DON’T TOUCH THE
ULTRASOUND SYSTEM!
To prevent electrical shock, connect the unit to a properly
grounded power outlet. DO NOT use a three to two prong
adapter. This defeats safety grounding.
DONOT wear the ESD wrist strap when you work on live
circuits and more than 30 V peak is present.
DO NOT operate this unit unless all board covers and frame
panels are securely in place. System performance and cooling
require this.
Versana Premier – Basic Service Manual System view 3-3
5789130-100 English Rev.5
Page 82
System Setup
DANGER
CAUTION
3-2-2 Setup warnings(continued)
1. There are no operator serviceable components. To prevent
shock, do not remove any covers or panels. Should
problems or malfunctions occur, unplug the power cord.
Only qualified service personnel should carry out servicing.
NOTE: For information regarding packing labels, refer to LABELS
ON PACKAGE.
2. After being transported, the unit may be very cold or hot. If
this is the case, allow the unit to acclimate before you turn it
on. It requires one hour for each 2.5°C increment it's
temperature is below 3°C or above 40°C.
Equipment damage possibility. Turning the system on without
acclimation after arriving at site may cause the system to be
damaged.
If the Ultrasound system is very cold or hot, do not turn on its power
until it has had a chance to acclimate to its operating environment.
Degree C
Degree F
hours
The following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
Table 3-1: System Acclimation Time Chart
50454035302520151050-5
122113104958677685950412323
420000000246
3-4 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 83
3-2-2 Setup warnings(continued)
CAUTION
CAUTION
Operator Manual(s)
The User Manual(s) should be fully read and understood
before operating the Versana Premier and kept near the
Ultrasound system for quick reference.
Acoustic Output Hazard
Although the ultrasound energy transmitted from the Versana
Premier probe is within AIUM/NEMA standards, avoid
unnecessary exposure. ultrasound energy can produce heat
and mechanical damage.
Setup reminders
Versana Premier – Basic Service Manual System view 3-5
5789130-100 English Rev.5
Page 84
System Setup
CAUTION
CAUTION
3-3 Receiving and unpacking the
3-3-1 Purpose of this section
This section describes how to receive and unpack Versana
Premier.
3-3-2 Contents in this section
• 3-3-3 ‘Warnings for receiving and unpacking’ on page 3-6
• 3-3-4 ‘The Tilt indicator’ on page 3-7
• 3-3-5 ‘Receiving the Versana Premier’ on page 3-8
• 3-3-6 ‘Unpacking the Versana Premier’ on page 3-14
equipment
3-3-3 Warnings for receiving and unpacking
Two people are needed to unpack the Ultrasound system
because of its weight. Attempts to move the Ultrasound system
considerable distances or on an incline by one person could
result in injury or damage or both.
Two people are required whenever a part weighing 16 KG (35
LBS) or more must be lifted.
Remember to use relevant personal protecting equipment
(PPE) during packing and unpacking. Check with your local
EHS representative.
3-6 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 85
Receiving and unpacking the equipment
3-3-4 The Tilt indicator
3-3-4-1 Overview
Improper handling during transportation may harm the
equipment inside the package even if the package itself is
undamaged.
To make it easier to detect if the handling during transportation
has been improper, a set of Tilt indicator has been attached to
the transportation box.
Table 3-2: Tilt Watch
DescriptionIllustration
Tilt Watch
Versana Premier – Basic Service Manual System view 3-7
5789130-100 English Rev.5
Page 86
System Setup
3-3-5 Receiving the Versana Premier
3-3-5-1 Examine all packages
Examine package closely at time of delivery, as described in the
procedure below.
Table 3-3: Examine all packages
StepTaskIllustrations
1. Is damage apparent?
• If YES; continue with the instructions in
3-3-5-4 ‘Damage in transportation’ on
page 3-11.
• If NO; continue with the next step.
2. Is the Tilt Indicator red colored inside the
middle of the indicator?
• If YES: The Tilt Indicator has been
activated. Continue with the instructions in
3-3-5-4 ‘Damage in transportation’ on
page 3-11 before you continue with the next
step.
• If NO: continue with the next step.
3. Continue with the instructions in
3-3-6 ‘Unpacking the Versana Premier’ on
page 3-14.
1 - Red Color
3-8 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 87
3-3-5-2 Position of the Tilt indicator
The Tilt indicator has been attached to the transportation box as
illustrated in the figure below.
Receiving and unpacking the equipment
Figure 3-1. Tilt indicator
1. Tilt Indicator
NOTE: Before cutting the straps, check Tilt Tag
been triggered. If damaged, report it to the carrier. If not, then
cut the straps around the crate.
Versana Premier – Basic Service Manual System view 3-9
5789130-100 English Rev.5
to make sure it has not
Page 88
System Setup
3-3-5-3 If Tilt Indicator has triggered or is missing
The purpose of the tilt indicator label is to alert people handling a
product that it is sensitive to tipping and it must remain upright at
all times. It is basically an active "Up Arrow" that changes color if
the package is tipped89 degrees or more from horizontal.
These labels can be false activated if tipped less than 89
degrees, and shocked or vibrated at the same time. This event
does occur, but is considered uncommon. If a package is
received with an activated tilt indicator label, there is high
degree of certainty it tipped 89 degrees or more from horizontal
during shipment.
An activated tilt indicator label does not indicate if the package
was simply “Tipped” (laid down with no impact shock) or “Tipped
Over” (free fall, with an impact shock). Using both shock
indicator labels and tilt indicator labels will help identify if a Tip
Over impact shock occurred.
Table 3-4: Tilt Indicator has triggered or is missing
StepTask
1.
If the Tilt Indicator is missing:
Note on the shipping papers at the time of
receipt that the Tilt Indicator label is missing.
If the Tilt Indicator has triggered:
Note on the shipping papers at the time of
receipt that the Tilt Indicator label was activated.
2.
Inspect the product for possible concealed
damage.
3-10 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 89
Receiving and unpacking the equipment
3-3-5-4 Damage in transportation
Follow this procedure if damage is apparent:
1. Write “Damage In Shipment” on ALL copies of the freight or
express bill BEFORE delivery is accepted or “signed for “by
a GE representative or hospital receiving agent.
2. Report the damage to the carrier.
• Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in
any event, within 14 days after receipt, and the contents
and containers held for inspection by the carrier.
• A transportation company will not pay a claim for
damage if an inspection is not requested within this 14
day period.
3-3-5-5 Versana Premier transportation box label
The Versana Premier transportation box label is located at the
front of the transportation box.
Figure 3-2.Versana Premier transportation box label 1
Figure 3-3.Versana Premier transportation box label 2
Versana Premier – Basic Service Manual System view 3-11
5789130-100 English Rev.5
Page 90
System Setup
3-3-5-5 Versana Premier transportation box label(continued)
Table 3-5: Package Label Icons
Label/IconPurpose/MeaningLocation
Keep this way up.Package
Fragile.Package
Package
Keep it away from rain.
2 levels at most.Package
3-12 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 91
Receiving and unpacking the equipment
Table 3-5: Package Label Icons
Label/IconPurpose/MeaningLocation
Humidity LimitationPackage
Temperature LimitationPackage
Pressure LimitationPackage
Versana Premier – Basic Service Manual System view 3-13
5789130-100 English Rev.5
Page 92
System Setup
CAUTION
CAUTION
CAUTION
3-3-6 Unpacking the Versana Premier
When a new system arrives, check that any components are not
damaged and are not in short supply. If shipping damage or
shortage occurs, contact the address shown in Chapter 1.
Please carefully unpack the system, and do not dispose the
package of Versana Premier, so that it can be reused for
service.
Do not lift the unit by the Keyboard. Equipment damage may
result.
The crate with the Versana Premier weighs approximately
50kg. Be prepared for a sudden shift of weight as the unit is
removed from its base (pallet).
NOTE: Please check the Versana Premier console is well assembly
after unpacking the system.
Table 3-6: Unpacking the Versana Premier
StepDescriptionCorresponding Graphic
1Tear off the "stop open" mark.
3-14 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 93
Receiving and unpacking the equipment
Table 3-6: Unpacking the Versana Premier
StepDescriptionCorresponding Graphic
2Cut off the two packing straps around the crate.
Note: To avoid injury, hold the strap clasp with
one hand when cutting the strap.
3Remove the top cover.
4Remove the three plastic locks.
Note: Rotate the inside plastic lock
counterclockwise to remove it and then remove
the outside lock.
5Remove the outside shipping box.
Versana Premier – Basic Service Manual System view 3-15
5789130-100 English Rev.5
Page 94
System Setup
Table 3-6: Unpacking the Versana Premier
StepDescriptionCorresponding Graphic
6Remove the 2 foams beside the system.
Note: There is no foam if the system is
transported by air. Ignore this step if there is no
foam.
7Remove the clear plastic (wrapped around the
8Cut the cable ties around the four wheels.
Versana Premier) from the unit.
3-16 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 95
Receiving and unpacking the equipment
Table 3-6: Unpacking the Versana Premier
StepDescriptionCorresponding Graphic
9Remove the barrier bag from the unit.
Note: Remove the PE bag if the system is
transported by air.
10Remove the foams beside the monitor.
11Remove any additional packing material
surrounding the system.
12Unlock the wheels, and then hold the control
panel at the front side to move the system until
the front two wheels on the ground and then
slowly move the whole system on the ground.
3-3-6-1 Moving into Position
Please refer to Basic User Manual on how to move the system.
3-3-7 Packing the Equipment
Please pack Versana Premier in the reverse order of unpacking.
Versana Premier – Basic Service Manual System view 3-17
5789130-100 English Rev.5
Page 96
System Setup
3-4 Packing materials - recycling
information
The packing materials for Versana Premier are recyclable:
• The Transportation Box is made of spruce or similar
material. (“PHYTOSANITARY CERTIFICATE” included in all
shipments to The People's Republic of China.)
• Lever lockings (hinges) are made of zinc plated steel.
• The inner reinforcements are made of Ethafoam
(Polyethylene foam).
• The plastic foil is made of LDPE (Low Density
Polyethylene).
3-18 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 97
3-5 Preparing for setup
3-5-1 Verify customer order
Compare items received by the customer to that which is listed
on the delivery order. Report any items that are missing, back
ordered, or damaged.
3-5-2 Physical inspection
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to ‘Damage in
transportation’ on page i-14 in the beginning of this manual.
Preparing for setup
3-5-3 EMI protection
The Versana Premier has been designed to minimize the effects
of Electro-Magnetic Interference (EMI). Many of the covers,
shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason,
it is imperative that all covers and hardware are installed and
secured before the unit is put into operation.
See 2-2-4 ‘EMI limitations’ on page 2-9 for more information
about EMI protection.
Versana Premier – Basic Service Manual System view 3-19
5789130-100 English Rev.5
Page 98
System Setup
3-6 Completing the setup
3-6-1 Purpose of this section
This section describes how to complete the installation of
Versana Premier.
3-6-2 Contents in this section
• 3-6-3 ‘System specifications’ on page 3-20
• 3-6-4 ‘Electrical specifications’ on page 3-21
• 3-6-6 ‘Connections on the I/O Rear Panel’ on page 3-24
• 3-6-7 ‘Connecting probes’ on page 3-24
• 3-6-8 ‘Powering the system’ on page 3-24
3-6-3 System specifications
3-6-3-1 System requirements verification
• Verify that the site meets the requirements listed in
Chapter 2.
(See: 2-3 ‘Facility needs’ on page 2-12.)
• Verify that the specifications below don’t conflict with any
on-site conditions.
3-6-3-2 Physical dimensions
Table 3-7: Physical dimensions of Versana Premier
HeightWidthDepthUnit
less than 1780 less than 620less than 810mm
less than 70 less than 24.4less than 31.8Inches
3-6-3-3 Console Weight
• Weight: approx. 70 kg (154.3 lbs)
3-20 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Page 99
3-6-4 Electrical specifications
WARNING
Connecting a Versana Premier to the wrong voltage level will
most likely destroy it.
3-6-4-1 Verification of the system’s voltage setting
Verify that the mains voltage specified for the Versana Premier
is available on-site.
Refer to the latest revision of the Basic User Manual for a full list
of product labels prior to serving the system.
3-6-4-2 Electrical specifications for Versana Premier
In the table below, the electrical specifications for Versana
Premier includes monitor and on board peripherals.
Table 3-8: Electrical specifications for Versana Premier
Completing the setup
Model NameVol tag eTolerances
Versana Premier VA
100-240V±10% Max.450VA50/60 Hz
Versana Premier VS
Power
consumptionFrequency
Versana Premier – Basic Service Manual System view 3-21
5789130-100 English Rev.5
Page 100
System Setup
3-6-5 Peripherals/Accessories Connector Panel
Versana Premier peripherals and accessories can be properly
connected using the side connector panel.
Figure 3-4. Rear Connector Panel
1. Power out portDC 12V 1.5A Out (Only for Digital Box in China)
2. Diagnostic LEDDiagnostic LED 1, Diagnostic LED 2, Diagnostic LED 3
3. Dual USB portUSB for use
4. Ethernet10/100MBase-T for LAN connectivity
5. USB portUSB 3.0 Port for use
6. HDMI portFor video port (Resolution 1920x1080)
7. VGAVGA Video Out
8. Audio Line OutAudio Line Out
9. S-VideoS-Video Out
10. Composite OutComposited Video Out
11. Circuit breakerCircuit breaker
12. AC Inlet100-240V
13. GNDGround terminal
3-22 System viewVersana Premier – Basic Service Manual
5789130-100 English Rev.5
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.