GE Solar 9500 User manual

Solar® 9500
Information Monitor
Service Manual
Software Version 3
2001085-061 Revision A
NOTE: Due to continuing product innovation, specifications in this manual are subject to change without
notice.
RAC RAMS, SAM, SOLAR, TRAM-NET, TRAM-RAC, and UNITY NETWORK are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office. All other trademarks contained herein are the property of their respective owners.
© GE Medical Systems Information Technologies, 2003. All rights reserved.
T-2 Solar 9500 Information Monitor Revision A

Contents

1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Manual Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Service Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Equipment Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
The Solar 9500 Information Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Supported Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Processing Unit Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Installing Cable Shroud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Display Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Primary Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Secondary Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Remote Display 1, 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Touchscreen Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Revision A Solar 9500 Information Monitor i
2001085-061
Scrolling vs. Non-scrolling Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Browser Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Processing Unit/Display Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Flat Panel Display Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Display Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Acquisition Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Tram-rac Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
Tram-rac Power Supply Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
Processing Unit/Tram-rac Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
Tram-rac 2 Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
Tram-rac 4A Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Unity Network ID Connectivity Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Dual Tram-rac Housings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
About Tram-net . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Processing Unit/RAMS Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
Processing Unit/Polled Parameter Interconnection . . . . . . . . . . . . . . . . . . . . . . . 3-16
Local Area Network (LAN) Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
About Ethernet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
Twisted Pair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
Network Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-19
Printer Interconnect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Laser Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20
PRN 50 Digital Writer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-22
Connecting to the Unity Network MC Network . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23
Connecting to the Unity Network IX Network . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-24
Keyboard and Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Turning Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Turning Power Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Using the Menu Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27
Alternate Power Down Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27
ii Solar 9500 Information Monitor Revision A
2001085-061
4 Software Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Monitor Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Layout Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Display/Layout Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Unit Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12
Controls and Menu Timeouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-14
Module Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Bed/Unit Name Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Serial Interface Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
Network Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
Unix Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
IP Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-18
Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-20
Audio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-21
Diagnostic Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-21
Locale Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-22
Creating a Custom Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Editing Basics for All Customizable Windows . . . . . . . . . . . . . . . . . . . . . . . . . . .4-23
Display Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-24
An Overview of Configured Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-25
Creating A Bed Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-26
Sample Creation of a Simple Bed Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-29
Print System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-31
Stratification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-32
5 Module Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
The Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Calibrate Non-invasive Blood Pressure (NBP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Manufacturer Recommendation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
Calibrate Mainstream CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Manufacturer Recommendation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9
0% CO2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
10% CO2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Calibration Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Revision A Solar 9500 Information Monitor iii
2001085-061
Calibrate Sidestream CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Manufacturer Recommendation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
0% CO2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
10% CO2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-14
Calibration Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-14
Calibrate Barometric Pressure for Interfaced CO2 . . . . . . . . . . . . . . . . . . . . . . . . 5-15
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Manufacturer Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Manufacturer Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Cleaning Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Exterior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Cleaning the Touch Screen Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Cleaning Inside the Solar 9500 CPU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Battery Pack Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Battery Pack Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Battery Recycling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
NLX Motherboard Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Electrical Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-10
Power Outlet Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-11
Ground (Earth) Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-11
Ground (Earth) Wire Leakage Current Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-13
Enclosure Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-14
Patient (Source) Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15
Patient (Sink) Leakage Current Test
(Mains Voltage on the Applied Part) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-16
Test Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-17
Checkout Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Required Tools/Special Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-18
Initial Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-18
Touch Screen Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-19
Speaker Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-19
iv Solar 9500 Information Monitor Revision A
2001085-061
Tram-rac Housing Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-19
Tram-net Communication Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-20
Unity Network MC Network Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Weblink Access Check (optional feature) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Printer Check (printer optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Unity Network ID Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
M-Port Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-22
Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-22
PM Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-22
Repair Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Fault Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
First Things to Ask . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Main Power and Display Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Fuse Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
AC Line Voltage Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
120 VAC, 50/60 Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9
240 VAC, 50/60 Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9
Troubleshooting Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10
Diagnostic Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-11
Log File Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-12
Log File Timestamps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-12
Diagnosing System Problems with Houston . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-13
Houston Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-13
Overview of the Menu Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14
Navigating Multi-page Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-21
Revision A Solar 9500 Information Monitor v
2001085-061
8 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Processing Unit Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Subassemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Power Supply Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
Video PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Riser Interface PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Riser I/O PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-9
M-Port PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-9
Signal Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Power Supply PCB Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-11
Video PCB Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-14
Riser Interface PCB Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-17
Riser I/O PCB Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-32
M-Port PCB Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-33
9 Field Replaceable Units, Parts Lists and Drawings . . . . 9-1
Ordering Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Field Replaceable Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Disassembly/Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Opening the Unit for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-3
Replace the Power Supply Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4
Replace the NLX Motherboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
Replace Riser I/O PCB Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-7
Replace/Add Video PCB or M-Port PCB Assembly . . . . . . . . . . . . . . . . . . . . . . . .9-8
Replace NLX Riser Interface PCB Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
Replace the Hard Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-11
Upper Level Assembly 901007-003A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-18
Power Supply Assembly 419080-001B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-22
vi Solar 9500 Information Monitor Revision A
2001085-061

1 Introduction

Revision A Solar 9500 Information Monitor 1-1
2001085-061

Manual Informatio n

Revision History

Each page of this manual has the document part number and revision letter at the bottom of the page. The revision letter identifies the document’s update level. The revision history of this document is
summarized below.
Revision Date Comment
A 20 January 2003 Initial release of this manual.

Manual Purpose

This manual supplies technical information for service representatives and technical personnel so they can maintain the equipment to the assembly level. Use it as a guide for maintenance and electrical repairs considered field repaira ble. Where necessary the manual identifies additional sources of relevant information and technical assistance.
Introduction: Manual Information
Revision History

Intended Audience

See the operator’s manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use.
This manual is intended for service representatives and technical personnel who maintain, troubleshoot, or repair this equipment.
1-2 Solar 9500 Information Monitor Revision A
2001085-061
Introduction: Safety Information

Safety Information

Responsibility of the Manuf acturer

GE Medical Systems Information Technologies is responsible for the effects of safety, reliability, and performance only if:
Assembly operations, extensions, readju stments, modifications, or repairs are carried out by persons authorized by GE.
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
The equipment is used in accordance with the instructions for use.

General

This device is intended for use under the direct supervision of a licensed health care practitioner.
This device is not intended for home use. U.S. law restricts this device to be sold by or on the order of a physician. Contact GE Medical Systems Information Technologies for information
before connecting to the equipment any devices that are not recommended in this manual.
Parts and accessories used must meet t he requireme nts of t he appli cable IEC 601 series safety standards, and/or the system configuration must meet the requirements of the IEC 60601-1-1 medical electri cal systems standard.
Periodically, and whenever the integrity of the device is in doubt, test all functions.
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include:
use of the accessory in the PATIENT VICINITY; and
evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard.
If the installation of the equipment, in the USA, will use 240V rather than 120V, the source must be a center-tapped, 240V, single-phase circuit.
Revision A Solar 9500 Information Monitor 1-3
2001085-061
Introduction: Safety Information

Warnings, Cautions, and Notes

The terms danger, warning, and caution are used throughout this manual to point out hazards and to designate a degree or level or seriousness. Familiarize yourself with their definitions and significance.
Hazard is defined as a source of potential injury to a person. DANGER indicates an imminent hazard which, if not avoided, will
result in death or serious injury. WARNING indicates a potential hazard or unsafe practice which, if not
avoided, could result in death or serious injury. CAUTION indicates a potential hazard or unsafe practice which, if not
avoided, could result in minor personal injury or product/property damage.
NOTE provides application tips or other useful information to assure that you get the most from your equipment.
1-4 Solar 9500 Information Monitor Revision A
2001085-061
Introduction: Safety Information
PRESS
4P41

Equipment Symbols Some of the following symbols appear on the equipment.

ATTENTION: Consult accompanying documents before using the equipment.
In Europe, this symbol means dangerous or high voltage. In the United States, this symbol represents the caution notice below:
To reduce the risk of electric shock, do NOT remove cover (or back). Refer servicing to qualified personnel.
Defibrillator-proof type CF equipment; type CF equipment is specifically designed for applications where a conductive connection directly to the heart is established. The paddles indicate the equipment is defibrillator proof.
Defibrillator-proof type BF equipment; type BF equipment is suitable for intentional external and internal application to the patient, excluding direct cardiac application. Type BF equipment is type B equipment with an F-type isolated (floating) part. The paddles indicate the equipment is defibrillator proof.
Type B equipment; type B equipment is suitable for intentional external and internal application to the patient, excluding direct cardiac application.
Equipotentiality
Alternating current (AC)
Power;
Fuse
Press to open.
I = ON; O= OFF
Medical Equipment With respect to electric shock, fire and mechanical hazards only in accordance with UL 2601-1, and CAN/CSA C22.2 NO. 601.1.
Revision A Solar 9500 Information Monitor 1-5
2001085-061

Service Information

Service Requirements

Follow the service requirements listed below.
Introduction: Service Information
Refer equipment servicing to GE Medical Systems Information Technologies authorized service personnel only.
Any unauthorized attempt to repair equipment under warr anty voids that warranty.
It is the user’s responsibility to report the need for service to GE Medical Systems Information Technologies or to one of their authorized agents.
Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required.

Equipment Identification

Every GE Medical Systems Information Technologies device has a unique serial number for identification. A sample of the information found on a serial number label is shown below.
D 0 XX 0005 G XX
Month Manufactured
A = January B = February C = March D = April E = May F = June G = July H = August J = September K = October L = November M = December
Year Manufactured
0 = 2000 1 = 2001 2 = 2002 (and so on)
Product Code
Two-character product descriptor
Product Sequence Number
Manufacturing number (of total units manufactured.)
Division
F = Cardiology G = Monitoring
Device Characteristics
One or 2 letters that further describe the unit, for example: P = prototype not conforming to marketing specification R = refurbished equipment S = special product documented under Specials part numbers U = upgraded unit
1-6 Solar 9500 Information Monitor Revision A
2001085-061

2 Overview

Revision A Solar 9500 Information Monitor 2-1
2001085-061
Overview: The Solar 9500 Information Monitor

The Solar 9500 Information Monitor

The basic Solar 9500 Information Monitor consists of a Solar 9500 processing unit, a compatible touchscreen display, and one or more acquisition devices.
The processing unit is the he art of t he s ystem. Prima ry use r cont rols a re provided by the touchscreen display but a keyboard and mouse may also be used.
001A
Up to two additional displays may be attached directly to the Solar 9500 Information Monitor to display layout configurations separate from the Primary Display. The additional displays are connected via optional video boards installed in the Solar 9500 processing unit.
002A
2-2 Solar 9500 Information Monitor Revision A
2001085-061

Front View

Overview: The Solar 9500 Information Monitor
On the front panel two indicator lights labeled AC and CPU show the operating condition of the unit. A connector panel on the right side of the unit provides all the connections to interface equipment to the system.

Rear View

003A
Located at the rear of the unit are the equipotential lug, power supply inlet and switch, and a mounting bracket.
&$87,21
The Solar 9500 processor must be mounted in the vertical position. Never operat e the unit in the horizontal position.
004A
Revision A Solar 9500 Information Monitor 2-3
2001085-061

Interconnection

Overview: The Solar 9500 Information Monitor
Shown below is the right side interconnect panel and a description of each connectors use.
UNITY NETWORK MC NETWORK
TRAM-NET
PCI BUS SLOTS
M-PORTS
UNITY NETWORK MC NETWORK is the Unity Network
PARALLEL PORT
DISPLAY (NON SCROLLING)
SERIAL PORT - 1
MOUSE
KEYBOARD
SERIAL PORT - 2
NOT USED
UNITY NETWORK IX NETWORK
NOT USED
NOT USED
NOT USED
connection that provides real time data (Parameter waveforms and numerics) between GE devices.
UNITY NETWORK IX NETWORK is the Unity Network connection that provides non-real time data between the Solar 9500 system and other hospital intranet devices.
TRAM-NET is the GE acquisition network connection t hat provides real time data between a Tram-rac’s patient connected acquisition devices (Trams and single parameter modules) and the Solar 9500 Information Monitor.
PCI BUS SLOT is for optiona l add in boar ds (e. g. t he s econd displ ay).
M-PORTS are for Unity Network ID, PRN 50, or PRN 50-M connection.
PARALLEL PORT is for connecting to a PostScript compatible printer.
SERIAL PORTS 1 & 2 are for touch screen, PRN 50, PRN 50-M, RAMS interface connections, or for polled parameter service.
KEYBOARD is a PS / 2 keyboard connection.
MOUSE is a PS / 2 mouse connection.
005B
:$51,1*
Connect devices solely manufactured or app roved by GE directly to the Unity Networks. Contact GE Technical Support before connecting equipment from other manufacturers.
2-4 Solar 9500 Information Monitor Revision A
2001085-061
Overview: Technical Specifications

Technical Specifications

Item Description
Type High definition raster or TFT Active Matrix Color LCD Color depth 16 bit color Resolution 1024 horizontal x 768 vertical Vertical Frequency 70 Hz non-scrolling (erase bar), 60 Hz scrolling Horizontal Frequency 57 kHz (minimum) Video bandwidth 110 MHz (minimum) Linearity 1.5% of vertical height Max
Solar 9500 Features and Options
Display Specifications
Item Description
User interface SAW Touchscreen (ELO touch systems SMART SET compatible) Waveform presentation Non-scrolling (erase bar) or scrolling Number of waveforms 30 waveforms Number of parameters 30 parameter windows Sweep speeds 1, 5, 10, 15, 25, 35, 50 mm/sec Seconds per waveform Max of 12 seconds at 25 mm/sec (based on display configuration) Pressure waveform scales Full, individual or free Software options 12SL ECG analysis, user display customization, browser, polled serial parameter
data, arrhythmia waveform review
Processing Specifications
Item Description
Main Processor Intel 233 MHz Pentium with MMX technology, main memory 128 MB RAM Graphics Controller Cirrus Logic GD-5465, 2MB RAMBUS DRAM Optional Graphics Processor
with Hardware Windowing Unity IX Network
Communication Processor Unity Network MC Network
Communication Processor
Revision A Solar 9500 Information Monitor 2-5
Fujitsu Sparclite MB86831 32bit, 50 MHz 6MB VRAM.
Intel 82558, 10/100 Base-TX
Intel 82596, 10 Base-T
2001085-061
Overview: Technical Specifications
Processing Specifications
Tram-net Communication
Intel 82596
Processor M-Port Network
Intel 82559
Communication Processor M-Port Ethernet Hub/Repeater Intel LXT914
Data I/O Connectors
Item Description
Unity Network IX Network RJ45 - standard Unity Network MC Network RJ45 - standard Tram-net 3-DB9F - standard M-Port 3-RJ45 - standard Main graphics DB15HD - standard Scrolling Graphics DB15HD - 2 optional Parallel port DB25F - standard Serial ports 2-DB9M - standard Keyboard/mouse 2-6 pin DIN, P/S2 style USB 4 pin USB - standard (not used) Audio line out Mini-jack standard (not used) Audio line in Mini-jack standard (not used) Audio Mic out Mini-jack standard (not used)
Environmental Specifications
Item Description
Power Requirements 110 - 120 (± 10%), 50/60-Hz
220-240 (± 10%), 50/60-Hz Power Consumption 200 Watts (includes Solar 9500, Tram-rac and Tram) Thermal dissipation 683 Btu/hr Internal 10 watt-hour UPS
Provides 12 second backup power for clean disk shutdown battery
Cooling Forced air
2-6 Solar 9500 Information Monitor Revision A
2001085-061
Operating Conditions Temperature
Overview: Technical Specifications
Environmental Specifications
10° C to 35° C (50° F to 95° F) Humidity
15% to 85% (noncondensing)
Storage Conditions Temperature
Humidity
–40° C to 70° C (-40° F to 158° F)
15% to 95% (non-condensing)
Physical Specifications (Solar 9500 CPU)
Item Description
Height 332 mm (13.0 in) Width 348 mm (13.7 in) Depth 156 mm (6.1 in) Weight 9 kg (19.8 lbs max. depending on options)
Certification
Item Description
Safety Standards Solar 9500 Processing Unit:
UL 2601-1 Classified
UL Classified for CAN/CSA C22.2 No. 601.1
IEC 60601-1 Certified
CE Marking for the 93/42/EEC Medical Device Directive (Refer to operator’s
manual for CE Marking specifics.)
Revision A Solar 9500 Information Monitor 2-7
2001085-061
Overview: Technical Specifications
Classification
Item Description
Type of protection against
Class I Equipment electrical shock
Degree of protection against
Type B Applies Part electrical shock
Degree of protection against harmful ingress of water
Degree of safety of application in the presence of a flammable
Ordinary Equipment (enclosed equipment without protection against ingress of
water)
Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide. anesthetic mixture with air or with oxygen or nitrous oxide
Method(s) of sterilization or
Not Applicable disinfection recommended by the manufacturer
Mode of operation Continuous operation
2-8 Solar 9500 Information Monitor Revision A
2001085-061

Supported Devices

The following is a list of supported peripheral devi ces compatible with the Solar 9500 via the Unity Network ID connectivity device. Refer to the Unity Network ID connectivity device service manual for interface adapter part numbers.
127(
Overview: Supported Devices
Due to continuous product innovation, this list may no longer be comprehensive. If necessary, call your sales representative for a current list of supported peripheral devices.
Refer to the operator documentation for further information about supported devices.
Supported Anesthesia Machines
Manufacturer Model
Dräger Medical Cato
Cicero EM (9 pin) Cicero EM (25 pin) Cicero PM 8060 (9 pin) Cicero PM 8060 (25 pin) PM 8050 Julian
North American Dräger Medical Narkomed 2B, 2C, 3, 4
Continuous Cardiac Output
Manufacturer Model
Abbott Laboratories Q-Vue
Q2
Baxter Edwards Vigilance
Vigilance European
Gas Analyzers
Manufacturer Model
Datex Capnomac Ultima Datex-Ohmeda 5250 RGM: Resp Gas
Rascal II Anes Gas
Revision A Solar 9500 Information Monitor 2-9
2001085-061
Overview: Supported Devices
Anesthesia Ventilators
Manufacturer Model
Datex-Ohmeda 7800/7810
7900 Aestiva 3000
Ventilators
Manufacturer Model
Bear Medical Bear 1000 Bird 8400ST/6400ST/VIP Dräger Medical Babylog 8000
Evita Evita 2 Evita 2 Dura Evita 4
Hamilton Medical Amadeus
Veolar Galileo
Nellcor Puritan Bennett 7200E/SPE/AE
Adult Star/1500/2000 Infant Star/500/950 840
Siemens Medical SV 300
SV 300A SV 900C/D/E
2-10 Solar 9500 Information Monitor Revision A
2001085-061

3 Installation

Revision A Solar 9500 Information Monitor 3-1
2001085-061
Installation: System Components

System Components

The Solar 9500 Information Monitoring System consists of four basic components:
Solar 9500 processing unit,
A color touch screen Primary Display,
Tram modu le, and
Tram-rac housing (holds Tram and single parameter modules).
Additional, option al components include:
Additional single or dual parameter modules,
RAMS,
One or two additional displays,
Unity Network ID
PostScript Printer and/or PRN 50 or PRN 50-M, and a
Browser server
CIC
Shown below is an example of a Solar 9500 Information Monitoring System.
SOLAR 9500
INFORMATION
MONITORING SYSTEM
TRAM-RAC
HOUSING
WITH TRAM
MODULE
TOUCHSCREEN
DISPLAY
SOLAR 9500
PROCESSOR
LASER PRINTER
(OPTIONAL)
TO/FROM OTHER PARTS
OF THE HOSPITAL
TO/FROM OTHER PARTS
OF THE HOSPITAL
PART of the
UNITY NETWORK
006A
3-2 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Processing Unit Setup

Processing Unit Setup

Check the unit for proper voltage setting before i nstallation. Although the AC mains volt age on Sola r 9500 process ing unit is factor y set for yo ur requireme nts, the power supply could be damaged if S1 is in the wrong position.
&$87,21
Make sure the voltage selection switch is in the correct position before connecting AC mains. The power supply will be damaged if the switch is in the 115 volt position and a high range voltage (195-270VAC) is applied. The supply will not function if in the 230 volt position when the low range voltage (90-135VAC) is applied.
1. Lay the Solar 9500 processing unit on its back.
2. Simultaneously press in on the two release tabs and lift up the front cover.
007A
3. View the power s upply voltag e setti ng thro ugh the window at the top of the unit.
008A
Revision A Solar 9500 Information Monitor 3-3
2001085-061

Mounting

Installation: Processing Unit Setup
4. Replace cover by carefully aligning the top edge of the cover into the hinge teeth on the back cover, then swing the front cover down until the release tabs are fully engaged.
The processing unit M UST be installed in a vertic al position with the supplied bracket before use.
&$87,21
Operating the device in a horizontal position or without the supplied bracket may cause damage to the equipment.
Wall channel
009B
3-4 Solar 9500 Information Monitor Revision A
2001085-061

Installing Cable Shroud

After installing all of the interface cables, use the following instructions to install the cable shroud.
Installation: Processing Unit Setup
010A
1. Remove the front cover as instructed in “Proce ssing Unit Setup” on page 3.
2. Snap the cable shroud, pn 420235-001, onto the side of the Solar 9500 housing.
3. Secure with the three self-tapping screws provided, pn 45177-205.
4. Replace the front cover as instructed in “Processing Unit Setup” on page 3.
&$87,21
Do not force the cover. Ensure that the front cover is aligned properly on its hinges before closing. Forcing the front cover into place with improper alignment will cause damage to the cover.
Revision A Solar 9500 Information Monitor 3-5
2001085-061

Display Descriptions

A single Solar 9500 processing unit can be connected to up to three displays as shown in the block diagram below and described as follows:
Installation: Display Descriptions
011A

Primary Display

Secondary Display

Remote Display 1, 2

This is the first in a series of monitors or may be the sole monitor connected to the Solar 9500 processing unit. The primary display is distinguished from the other monitors by the eight control buttons displayed at the bottom of the screen. Real Time data (parameter waveforms and numerics) and/or Non-Real Time data (a display with trends, alarm history, etc.) are displayed on this screen.
A secondary display has touchscree n capabilitie s and displays Real Time and/or Non-Real Time data on a two or three display system.
A remote display is a view only monitor and is not intended to be used as a touchscreen display. No user interface or applications in display setup are available. Up to two remote displays can be connected to the Solar 9500 processing unit.
3-6 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Display Descriptions

Touchscreen Interface

A touchscreen display is the recommended user interface, although a keyboard and/or mouse may be used. The touchscreen must be compatible with ELO Touchsystems’ Smart Set protocol. You may connect up to two touchscreens to the Serial Ports 1 and 2 of the Solar 9500 processing units. See chapter 4, Software Configuration for information on Serial Port Configuration.

Scrolling vs. Non-scrolling Waveforms

Scrolling refers to the ability to continuously move the wavef orms acros s the display from right to left. Non-scrolling waveforms are static with an “erase bar” moving across the display. As the erase bar sweeps the display, it erases the data in fro nt of the bar a nd writes new dat a in back of the bar.
The base Solar 9500 system only supports non-scrolling waveforms. For scrolling and non-scrolling waveform capability, the PCI bus add-in video board must be ordered. The So la r 9500 can support up to two add-in video boards.

Browser Support

The web browser feature is only supported on the motherboard video connection (non-scrolling) and requires a screen description of Primary, Secondary, or Applications Only. See “Display/Layout Setup” on page 4-4 for details.
Revision A Solar 9500 Information Monitor 3-7
2001085-061
Installation: Processing Unit/Display Interconnection

Processing Unit/Display Interconnection

Each display in a Solar 9500 system must be connected according to its configuration. Illustrated below are simplified interconnect diagrams for the various display configurations.
Two Displays - Primary Non-Scrolling
A
Non-Scrolling
(Motherboard)
Scrolling
BC
VIDEO, MAIN
TOUCHSCREEN
TOUCHSCREEN
(secondary only)
VIDEO A
Two Displays - Primary Scrolling
PRIMARY
DISPLAY
SECONDARY
or
REMOTE
Three Displays - Primary Non-Scrolling
PRIMARY
DISPLAY
Non-Scrolling
(Motherboard)
Scrolling
ABC
Scrolling
VIDEO, MAIN
TOUCHSCREEN
SECONDARY
REMOTE
TOUCHSCREEN (secondary only)
VIDEO A
REMOTE DISPLAY
VIDEO B
Three Displays - Primary Scrolling
or
PRIMARY
DISPLAY
VIDEO A
TOUCHSCREEN
SECONDARY
or
REMOTE
TOUCHSCREEN (secondary only)
VIDEO, MAIN
REMOTE DISPLAY
VIDEO B
012B
ABC
Scrolling
Scrolling
PRIMARY
DISPLAY
VIDEO B
TOUCHSCREEN
SECONDARY
or
REMOTE
TOUCHSCREEN
(secondary only)
VIDEO A
AB
Scrolling
Non-Scrolling
(Motherboard)
C
Scrolling
3-8 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Processing Unit/Display Interconnection

Flat Panel Display Interconnection

To connect a Flat Panel Display, a medical grade power supply is required. Shown below is an illustration.
TOUCH SCREEN CONNECTION 2000727-002

Display Mounting

VIDEO CONNECTION 2002215-00X
FLAT PANEL DISPLAY BOTTOM VIEW
MEDICAL GRADE POWER SUPPLY
013C
Use GE or approved display mounting solutions when installing one or more displays.
&$87,21
When using a non-UL2601/IEC60601 display ALWAYS
use an approved isolation transformer, and
protect the display from possible ingress of liquids.
Revision A Solar 9500 Information Monitor 3-9
2001085-061
Installation: Acquisition Interfaces

Acquisition Interfaces

Local bed acquisition comes from three possible sources:
a Unity Network ID connectivity device
a Tram and single parameter modules in one or two Tram-rac housings via the Tram-net interfa ce, and
the RAMS (Random Access Mass Spectrometer) via an RS-232 interface.
The Tram-net interface is a GE proprietary network used for communications between a bedside monitor and the Tram-rac(s).

Tram-rac Housing

The Tram-rac housing (remote acquisition case) acquires patient data for the Solar 9500 Information Monitor System. The Tram-rac Housing Service Manual has more information. There are two Tra m-rac hous ings available for the monitor:
Tram-rac 2 housing, which holds a single Tram module, and
Tram-rac 4A housing, which can hold a Tram module, a SAM module, and/or additional single parameter modules.
127(
The Solar 9500 is NOT compatible with the RAC 2A housing.
Shown below is a Tram-rac 4A housing with a Tram module and two single parameter modules inserte d .
014A
3-10 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Acquisition Interfaces

Tram-rac Power Supply Connection

If a Tram-rac power supply is used, connect the power cord as shown below.

Processing Unit/Tram-rac Interconnection

015A

Tram-rac 2 Housing

The Solar 9500 processing unit will support up to two Tram-rac housings. These housings can be connected in parallel or series. When connected in series the center Tram-net connector of the Tram-rac connects to the secon d Tram-rac housing. The T ram-rac hous ing fu rthest from the monitor must have a power supply.
Shown below is the connection from the Tram-rac 2 housing to a Solar 9500 processing unit.
127(
To connect the Solar 9500 processing unit to the Tram-rac housing, use cable pn 700520-00x or equivalent. A Tram-net hub assembly is NOT used with the Solar 9500 processing unit
700520-00X
TRAM
NET
016B
Revision A Solar 9500 Information Monitor 3-11
2001085-061

Tram-rac 4A Housing

Installation: Acquisition Interfaces
Shown below is the connection from the Tram-rac 4A housing to a Solar 9500 processing unit.
700520-00X

Unity Network ID Connectivity Device

Shown below is a connection from the Unity Network ID to a Solar 9500 processing unit. The Unity Network ID may be connected to any M-Port.
017B
I 0
104A
3-12 Solar 9500 Information Monitor Revision A
2001085-061

Dual Tram-rac Housings

Shown below are examples of how to connect t wo Tra m-ra c ho usings t o a Solar 9500 information processor.
Installation: Acquisition Interfaces
NOTE
WHEN CONNECTED IN PARALLEL, THE UNPOWERED RAC-4A MAY BE SUBSTITUTED BY A RAC-2
700520-00X
Parallel Connection
700520-00X
Serial Connection
018B
411090-00X
019B
Revision A Solar 9500 Information Monitor 3-13
2001085-061

About Tram-net

Installation: Acquisition Interfaces
The three 9-pin connectors on the processing unit make a Tram-net network available for the acquisition devices.
Tram-net is a small network that offers ample flexibility, a high rate of communication, and relatively inexpensive cabling. Data is transmitted at the rate of 921.6K bits per second. It uses a star topology, sometimes referred to as a rooted tree topology. This means that the wiring of the network can be pictured as a star or a series of stars. The center of each star is called a hub, and at the points of the star are called nodes. There are cables between the nodes and the hubs, but no cables exist between nodes.
020A
127(
Do not confuse the internal hub or node with the Tram-net hub assembly used with other GE modular monitors. The hub described in these paragraphs refe rs to the elect ronic dis trib ution poi nt of dat a.
Data is acquired at a node, and is transmitted through a hub to all the other nodes. Each node has an address so data will be received by the node with the correct destination address. It is impossible for a node to communicate with another node without the data going through a hub somewhere along its journey. The hub controls all of the data ‘traffic’ in the system.
In a Tram-net system, the head hub is contained in the patient monitor, but there will be intermediate hubs in the Tram-rac housing and Tram module as well.
3-14 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Acquisition Interfaces

Processing Unit/RAMS Interconnection

Before connecting the RAMS to the Solar 9500 processing unit you must first configure the serial channel of the processing unit via the Serial Interface Controls in the Service Menu. See chapter 4, Serial Interface Controls for details. The RAMS interfaces with a Solar 9500 via the RAMS RS-232 connection to the Solar processing unit with an interface cable (pn 414803-003) as shown below.
021A
Revision A Solar 9500 Information Monitor 3-15
2001085-061
Installation: Processing Unit/Polled Parameter Interconnection

Processing Unit/Polled Parameter Interc onnection

The optional polled parame ter service allows access to patient parameter data from the Solar 9500 to other data gath ering device s. Only the poll ed parameter service is available. Admit/discharge, time, software version, and waveform services are NOT available via the serial interface.
Before connecting the data gathering device to the Solar 9500 processing unit, you must first configure the serial channel of the processing unit via the Serial Interface Controls in the Service Menu. Refer to chapter 4, Serial Interface Controls for details.
The serial port of the Solar 9500 has a standard PC pinout and functions as a DTE (see EIA-232-E and EIA/TIA-574 standards for further information). The following table shows the pinout of the serial port on the Solar 9500 processing unit.
Pin Number Pin Name Direction To/From the Solar 9500
1 DCD IN 2RXDIN 3TXDOUT 4DTROUT 5GNDGND 6DSRIN 7RTSOUT 8CTSIN 9RIIN
To connect the Solar 9500 processing unit to a standard PC for data gathering, a null modem ca ble with 9-pin female (socket) ends should be used. The serial ports are not isolated on the Solar 9500 processing unit, and, if required, an external isolation mechanism must be used.
3-16 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Local Area Network (LAN) Interfaces

Local Area Network (LAN) Interf aces

The Solar 9500 supports two separate LAN ethernet connections. One connection is dedicated to the Unity Network MC network and one is dedicated to the Unity Network IX network, hereafter referred to as MC network or IX network. Both use a twisted-pair network connection.
The MC (Mission Critical) network is used to connect the Solar 9500 processing unit to other bedsides. This network contains all the waveforms, parameters, alarms and other time-sensitive data. This network is also used to send status information about the printers and for “pushing” layout, unit defaults, and Solar 9500 software between Solar 9500 systems. The MC network is a 10BaseT network.
The IX (Information Exchange) network is used for connection to a Weblink Multiple Access Server (MAS) and possibly a network laser printer. The U IX network is a 10BaseT/100BaseTX Network with auto­sensing between protocols.
Below are the locations of the MC and IX network connector ports on the Solar 9500 processing unit.

About Ethernet

UNITY NETWORK MC NETWORK (Real Time Data)
UNITY NETWORK IX NETWORK (Non-Real Time Data)
022B
See chapter 4, Software Configuration for information on how to setup and configure the networks.
Ethernet is a local area network used as the main link of the Unity Network, a comprehensive information communication system. The Unity Networks offer the high rate of communication of 10 megabits per second. This local area network links all patient monitors, central stations, and other GE equipment th roughout the hospital. Dep ending on the construction of the hospital, thick-net, thin-net, or CAT-5 twisted pair cabling is used. The Solar 9500 is designed to be used with twisted­pair cabling. Consult GE when trying to interface with ei ther thick-net or thin-net cabling.
Revision A Solar 9500 Information Monitor 3-17
2001085-061

Twisted Pair

Installation: Local Area Network (LAN) Interfaces
Twisted pair is the most popular cabling because it is easy to install and flexible to work with. It uses the star topology with a concentrator as the hub of the segment. Each of the network devices is connected directly to the concentrator so longer lengths of cable are required. A maximum of 100 meters or 328 feet is the longes t length of twi ste d pair cable all owed. The number of devices is limited to the amount of connectors at the concentrator.
For example, one segment may connect all the patient monitors and central stations in the ICU (Intensive Care Unit) and another may connect the monitoring system in the CCU (Critical Care Unit). Each segment could be a fully-functio nin g stand- alo ne sys te m if th ey were not connected to each other. However, with a bridge or repeater to connect the ICU (one segment) with the CCU (the other segment), information can pass between any of the nodes (patient monitors and central stations) on either branch similar to a patient transfer from one unit to another.
023A
3-18 Solar 9500 Information Monitor Revision A
2001085-061

Network Terms

Installation: Local Area Network (LAN) Interfaces
Illustrated below is a simplified view of a network and a definition of its terms.
23B
024A
Revision A Solar 9500 Information Monitor 3-19
2001085-061

Printer Interconnect

The following types of printer connections are supported by the Solar 9500:
127(
Installation: Printer Interconnect
A network printer connected to eit her et hernet i nterf ace via a hub or switch
A printer connected to a parallel port A PRN 50 or PRN 50-M connected to an M-Port A PRN 50 or PRN 50-M connected to a serial port
Only one parallel port printer can be used.
Only one PRN 50 or PRN 50-M writer can be used.
A PRN 50 or PRN 50-M writer assigned to an M-Port will not be used by the Solar 9500 if a serial port is configured to be used for a PRN 50 or PRN 50-M writer.
If more than one PRN 50 or PRN 50-M writer is connected to M­Ports, only the one that was connected first will be used. If it is disconnected, then one of t he other a vai lab le wri ters w ill be us ed.

Laser Printer

Parallel port printers and writers are broadcast to the network and can be shared by other Solar 9500s. The parallel port printer or writer name is identified and broadcasted as the bed name of the Solar 9500 it is connected to; for example, Bed 1 WRITER.
:$51,1*6
SHOCK HAZARD. Laser printers are UL 60950/IEC 60950 certified equipment, which may not meet the leakage current requirements of patient care equipment. This equipment must not be located in the patient vicinity unless the medical system standard IEC 60601­1-1 is followed.
Do not connect a laser printer to a multiple portable socket outlet (MPSO) supplying patient care equipment. The use of a MPSO for a system will result in an enclosure leakage current equal to the sum of all the individual earth leak age currents of the s ystem if th ere is an interruption of the MP SO protective earth conductor.
An optional laser printer can be directly connected to the Solar 9500 processing unit at the para llel port or t o either netwo rk used by th e Solar
9500. However, this printer must be PostScript level 2 compatible. See
“Printing” on page 4-12 for detailed information about printer
configuration.
3-20 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Printer Interconnect
All printers are shared across the network whether they are directly connected to the network or indirectly through a Solar 9500 with a parallel port printer. This is accomplished by assigning a Solar 9500 to act as the print server for each printer on the network. A single Solar 9500 can act as a print server for multiple printers.
&$87,216
Multiple Solar 9500s can not act as print servers for a single printer. Using them in this manner will result in unpredictable behavior.
Solar 9500 printers may not be shared with other devices (e.g. CIC).
Shown are two methods of connecting to the parallel port. Connect the printer to the processing unit with cable pn 6119-104 as shown.
PARALLEL PORT
PRINT SERVER
MC NETWORK PORT
PARALLEL PORT
Printer to Parallel Port
MC NETWORK PORT
PRINT SERVER
025A
PARALLEL PORT
MC NETWORK
026A
Printer to Parallel Port, System to the Unity Network MC Network
Revision A Solar 9500 Information Monitor 3-21
2001085-061

PRN 50 Digital Writer

Installation: Printer Interconnect
A PRN 50 or PRN 50-M digital writer can be connected to one of the Solar 9500 M-Ports or serial ports. A writer connected to an M-Port will not be used by the Solar 9500 if a writer is configured to a serial port.
Cable PN 418335-001
Note:
If connecting to a PRN 50 with AutoPort, then adapter PN 2001973-001 is required
PRN 50 or PRN 50-M to M-Port
Refer to “Serial Interface Controls” on page 4-16 for instructions on configuring a port for a PRN 50 or PRN 50-M digital writer.
PN 2006550-001
PRN 50 PRN 50-M
Cable PN 422843-001 Cable PN 418335-001 No adapter Adapter PN 2006550-001
Cable PN 418335-001Adapter
PRN 50-M to Serial Port
105A
101B
3-22 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Printer Interconnect

Connecting to the Unity Network MC Network

When connecting a network printer to the MC network, the IP address of the printer should start with 126.50.x.x. and the netmask should be
255.0.0.0. This assumes that the default MC network address of the
Solar 9500 is unchanged from the factory defaul t. Consult the operating manual of the printer for setting the IP address and netmask.
PRINT SERVER
10 BASE - T
MC NETWORK PORT
MC NETWORK PORT
MC NETWORK
027A
Revision A Solar 9500 Information Monitor 3-23
2001085-061
Installation: Printer Interconnect

Connecting to the Unity Network IX Network

When connecting a network printer to the IX network, the IP address should be a unique, valid node within the subnet of the IX network. The netmask should correspond with the rules determined by the institution for the IX network.
127(
To share a printer between Solar 9500s on the IX network, the Solar 9500s must also be connected to the MC network, since all printer status information passes th rough the MC network.
IX NETWORK
PRINT SERVER
MC NETWORK PORT
IX NETWORK PORT
UNITY NETWORK MC NETWORK
10 BASE-T
IX NETWORK  PORT
MC NETWORK PORT
028A
3-24 Solar 9500 Information Monitor Revision A
2001085-061
Installation: Keyboard and Mouse

Keyboard and Mouse

The optional keyboard and mouse can be used in lieu of the touchscreen interface. They are also used when servicing or troubleshooting a Solar
9500. A port is provided for each device at the interconnect panel and
they are NOT interchangeable. Follow the directions below when connecting these devices:
KEYBOARD
MOUSE
029B
127(
The keyboard and mouse must be connected to the Solar 9500 processor BEFORE power up or it will not function.
The keyboard will not operate unless the mouse is connected.
Revision A Solar 9500 Information Monitor 3-25
2001085-061

Turning Power On

Installation: Turning Power On
The Solar 9500 processing unit is preset at the factory for a speci f ic AC voltage. Before applying power to the monitor, be sure its voltage setting matches your power requirements. Refer to the label on the back of the processor for the voltage and current requirements. Refer also to Power Requirements in chapter 2, Technical Specifications.
When all cables are properly connected and the attached display has been energized, the power switch on the back of the Solar 9500 processor should be pressed to the “1” or ON position.
The power-up process takes a few minutes to complete.
Five seconds after turning power on, you should see the Solar 9500 logo displayed on each of the displays.
Approximately one minute after turning power on, the graphical user interface, waveforms, and parameters will be displayed on the screen and normal processing will be started.
Several seconds af te r t he p rima ry di spl ay i s up, t he se cond and t hird displays will be functional.
About 2.5 minutes after the power-up sequence is complete, all applications will be available and printing will be active.
127(
Do not switch the power off and on between cases. Unit must be powered down before servicing. Refer to “Turning Power Off” on page 3-27.
3-26 Solar 9500 Information Monitor Revision A
2001085-061

Turning Power Off

The preferred method of turning on the Solar 9500 power off is to first shut the system down by usin g the on-screen menu command, then use the on/off mains power switch located on the back of the unit.

Using the Menu Command

To power down the processing unit, select the Setups button on the bottom of the primary display. Next select the Configuration/Service menu option. This will take you to the Configuration menu, where you
can select Shut System Down.... A confirmation screen displays.
Installation: Turning Power Off
&$87,21
Do not move the system while turning power off. Wait 40 seconds after removing power before moving the system.
Select Shut system down to start the orderly shut down process. Wait until the message Caution: Do not move the Solar 9500 monitor for 40 seconds after turning power off appears and complete the shut down by pressing the 0 on the on/off power switch on the back of the processing unit.

Alternate Power Down Method

If necessary, you can directly turn off the Solar 9500 with the on/off switch. However, the system sees this as an AC power failure which causes it to run on the internal battery followed by a start of the automatic shutdown procedure. Since the battery is intended for orderly shutdown during an AC failure condition, we recommend that you shut down the system through t he me nu co mmand. Wi th a new, fully charge d battery, only 13 continuous shutdowns are supported before a recharge is necessary.
&$87,21
Do not move the system while turning power off. Wait 40 seconds after removing power before moving the system.
030A
Revision A Solar 9500 Information Monitor 3-27
2001085-061
For your notes
Installation: Turning Power Off
3-28 Solar 9500 Information Monitor Revision A
2001085-061

4 Software Configuration

Revision A Solar 9500 Information Monitor 4-1
2001085-061
Software Configuration: Monitor Service Menu

Monitor Service Menu

The Service Menu is for use by qualified field engineers and factory service personnel to configure and service the Solar 9500 Information Monitor.
To access the SERVICE MODE menu, execute the following menu sequence, starting from the MAIN menu:
Setups Configuration/Service...
&$87,21
The Service Menu is intended for use only by qualified personnel. Unnecessary tampering with service mode menu options for experimentation purposes may cause a malfunction of the equipment and is not recommende d by the manufacturer.
Service mode...
At the numeric touchscreen keypad, enter the day and month using leading zeros. (July 4 = 0407)
The following Service menu screen displays.
031B
4-2 Solar 9500 Information Monitor Revision A
2001085-061

Layout Controls

Allow Configuration
Software Configuration: Monitor Service Menu
127(
Configuration editing is an optional software package. If your Solar 9500 is not enabled for configuration editing, the Layout Controls will be inactive.
Configuration editing allows you to create, modify, and save screen layouts. Configuration editing uses a number of windows to help you design your screen. These windows are combined and layered to maximize functionality and to permit you to configure the Solar 9500 according to your personal preference.
Configuration editing functions are addressed in greater detail later in this chapter.
Select Allow Configuration and the words Configuration ON display in red at the top of the primary display to show that configuration editing is enabled.
Disable Configuration
Delete Layouts
When you are finished with configuration editing, turn configuration editing off. Configuration editing can be disabled from the service menu by selecting Disable Configuration or from the Setups menu by selecting Turn layout configuration off.
A layout is the base layer or background canvas of the screen layout (including Parameter blocks, buttons and waveform windows). Selecting Delete Layout opens a display that allows you to choose which layouts to remove.
127(
Currently active layouts, layouts listed in the Layout Setup menu, and GE clinical layouts cannot be deleted.
Revision A Solar 9500 Information Monitor 4-3
2001085-061

Display/Layout Setup

Software Configuration: Monitor Service Menu
Selecting Default Layouts... displays the Application-Layout Setup window. If your unit has optional graphics cards, you can program the Default Layouts for additional displays and define the default layout to be used on each display.
Under Layout Setup cho o se the default layouts for the Primary, the Secondary, the Remote 1 and 2, and the Patient View. Layout Setup shown is for a system with 3 displays.
Screen Description Options
032A
Selecting Display Setup... displays the Appl ication Displ ay S etu p screen. For each video output, choose the Screen Description. Each application can be assigned to any of the available video outputs whose screen descriptions are Primary, Secondary or Apps Only. The column labeled Motherboard Main refers to the graphics port on the Solar 9500 motherboard. The column Video Slot A refers to the first add-in video card, the column Video Slot B refers to the second add-in video card. A highlighted green circle indicates which video/screen description the application appears on.
127(
An application can be assigned to only one display. Screen descriptions cannot be duplicated, each must be unique. (e.g. If Primary is selected for one video output, then Primary cannot be selected for another video output.)
The 12SL and Browser applications are optional software packages. These applications can only be assigned if they are enabled. The Browser can only be run on the Motherboard Main video connection (non-scrolling) and requires a screen description of Primary, Secondary, or Applications Only.
You cannot assign any non-real-time applications to a display that doesn’t have a user input device such as a remote display.
4-4 Solar 9500 Information Monitor Revision A
2001085-061
Software Configuration: Monitor Service Menu
Display Setup shown is for a system with 3 displays.
033B
Primary
The primary display is the first in a series of monitors or the sole monitor connected to the Solar 9500 processing unit. It is distinguished from the other monitors by the eight control buttons displayed at the bottom of the screen. Real Time data (parameter waveforms and numerics) and/or Non-Real Time data (a display with trends, alarm history, etc.) display on he screen.
Secondary
A secondary display has touchscree n capabilitie s and displays Real Time and/or Non-Real Time data on a two or three display system.
Remote 1, 2
A remote display is a view only monitor and not intended to be used as a touchscreen display. Only Real Time data displays. Up to two remote displays can be connected to the Solar 9500 processing unit.
127(
You cannot assign Non-Real Time applicat ions to a remote display.
Unused
The display is not configured for use.
Revision A Solar 9500 Information Monitor 4-5
2001085-061
Waveform Options
Software Configuration: Monitor Service Menu
The base Solar 9500 system only supports non-scrolling waveforms using the Motherboard Main connection. For scrolling and non-scrolling waveform capability, the PCI bus add-in video board must be ordered. The Solar 9500 can support up to two add-in video boards. (Slot A and B.)
Scrolling refers to the ability to continuously move the wavef orms acros s the display. Non-scrolling waveforms are static with an “erase bar” moving across the display. As the erase bar sweeps the display, it erases the data in front of the bar and writes new data in back of th e bar.
Turn Waveform Scrolling (erase bar mode) ON and OFF for optional graphics cards only. Turn on Wide Waveform Lines to make the waveforms bolder for distant viewing (e.g., a surgeon’s display).
After completing setup options, select SAVE to save or Cancel to remove the choices. When SAVE is selected the following message appears:
In order to make the display changes active a monitor reset is required. Press SAVE to save the changes and reset the monitor immediately. Press CANCEL to discard the changes and exit.
:$51,1*
DO NOT make changes to the Display Setup while actively monitoring a patient because a rese t of the unit is required and important data may be lost.
4-6 Solar 9500 Information Monitor Revision A
2001085-061

Unit Defaults

Parameters...
Software Configuration: Monitor Service Menu
Selecting the Parameters... option under Unit Defaults displays the Setup Parameter Defaults window. All of the parameter menu options as well
as arrhythmia and parameter alarm limits, alarm levels and alarm volume can be accessed, modified, and saved as unit defaults.
Setup Parameter Defaults
In the Setup Parameter Defaults window, there are buttons for each available parameter. Select the appropriate button to open the corresponding defaults window and view or modify the current default settings for an available parameter. Refer to “Individual Parameter Windows” on the followi ng pages.
&$87,21
Incorrect Parameter Settings. Changing the parameter defaults does NOT affect the current parameter and alarm settings until you select Save, and then Restore on the Setup Parameter Defaults screen.
127(
The difference between run-time changes and parameter default changes is whether the parameter is accessed through the main menu or the service menu. Parameter menus accessed fro m the main menu, display “Control” in the upper-right corner whereas they display “Defaults” via the se rvice menu. For ex ample, “EC G Control” vs. “ECG Defaults.”
034B
Revision A Solar 9500 Information Monitor 4-7
2001085-061
Software Configuration: Monitor Service Menu
127(
When setting the parameter defaults for CO2, only the CO2 waveform scale, CO2 units, and CO2 color can be set. The CO2 modules, the SAM, and the RAMS all have other parameter configuration defaults that cannot be changed. Examples are: O2 waveform scale, O2 units, O2 co lor, N2O compensation (either on/off or a percent), and O2 compensation.
Alarm Defaults
To set alarm defaults, select a parameter from the Setup Parameter Defaults window, then select To Alarm Setup... The Alarm Defaults
window displays (see next page).
035A
4-8 Solar 9500 Information Monitor Revision A
2001085-061
Software Configuration: Monitor Service Menu
A
B
C
D
E
F
G
036A
A Alarm Limits Set alarm limits on the left side of the screen by
selecting the up/down arrows.
B Alarm Level and State Set alarm levels to Crisis, Warning, Advisory or
Message. Set alarm state to On, Off, or Always On.
NOTE: Setting an alarm state to Always On means
that the user does not have the option to change the individual parameter alarm to the Off state.
C Alarm Volume Set alarm volume by selecting the up/down arrows. Set
alarm volume to On, Off, or Always On. NOTE: Setting the alarm volu me to Always On means
the user does not have the option to turn the alarm volume off.
D Defaults Select Restore... to restore alarm defaults to either User
Defaults or GE clinical layouts Defaults.
E Alarm Graphing Set alarm graphing to On or Off to enable or disable
automatic graphing of alarms as they occur.
F Arrhythmia Alarm Levels... Select Arrhythmia Alarm Levels to ope n th e Arrhythmia
Defaults window. All 15 monitored arrhythmia conditions have alarm level choices Crisis, Warning, Advisory or Message.
G System Alarm Levels... Select System Alarm Levels... to open the System Alarm
Defaults window. Set ECG Leads Fail and SpO2 Probe Off Patient to System Warning or System Advisory.
Revision A Solar 9500 Information Monitor 4-9
2001085-061
Software Configuration: Monitor Service Menu
127(
After modifying any defaults remember to return to the Setup Parameter Defaults window and select one of the following options.
037A
Save – After modifying a parameter default settings, select Save to save the changes.
Restore – Select Restore to restore the most recently saved settings. When Restore is selected, a message window opens:
This will change all parameter settings back to the user default settings. Do you wish to continue?
Select Yes to change the settings, or No to close the message window without making changes.
Reset – Select Reset to reset all settings back to the original factory settings. When Reset is selected, a message window opens:
This will change all parameter settings back to the original GE clinical layouts default settings. Do you wish to continue?
Select Yes to change the settings, or No to close the message window without making changes.
Print – Select Print to initiate a printout of the saved user default settings for all of the available parameters.
GE Medical Systems Information Technologies recommends printing default settings after making changes.
Close – Select Close to close the Setup Parameter Defaults window.
4-10 Solar 9500 Information Monitor Revision A
2001085-061
End Case...
Software Configuration: Monitor Service Menu
Select End Case... to set the default behavior when the End Case... button is selected.
Patient Info...
038A
Select Patient Info... to set the default units of measure for patient information.
039A
Revision A Solar 9500 Information Monitor 4-11
2001085-061

Printing

Configure Printers...
Software Configuration: Monitor Service Menu
Patient data on the Solar 9500 can be printed for review to a PostScript compatible printer or writer. The Solar 9500 supports both parallel and network printers. A network printer sha res the netwo rk with other Solar 9500 monitors. (This means it is not connected to any one particular Solar 9500 monitor). However, there must be one Solar 9500 on the network that is configured with the printer’s name and address. If there is more than one printer on the network, each must have its own name and address. The Solar 9500 configured with that name and address then controls and queries that printer on the network for its status.
Refer to “Serial Interface Controls” on page 16 for instructions on configuring a port for a printer or writer.
Select Configure Printers... to open the Net Printer Setup window.
127(
Configure Printer... is only used for setting up network printers. Parallel printers do not ne ed configuring. PRN 50 and PRN 50-M digital writers are configured in “Serial Interface Controls” on page 16.
040B
The following options are available:
New allows you to add another printer to the list. When adding a printer to the network it is necessary to enter the
Name by which the printer will be referred to, the IP Address of the printer, and the Port Number of the printer. The port number may be found in the printer’s network documentation. For example, Hewlett Packard LaserJets use Port # 9100.
Remove allows you to remove a printer from the list.
Undo All allows you to erase all changes made.
4-12 Solar 9500 Information Monitor Revision A
2001085-061
Clear Print Queue...
Software Configuration: Monitor Service Menu
Show Keyboard displays an on-screen key board. This button switches to Hide Keyboard when pressed. Select Hide Keyboard to remove the keyboard from the display.
Ok closes the window and prompts you to reboot. (You may choose not to reboot, but changes are not effective until you do.)
Cancel allows you to cancel the changes and closes the window.
127(
A parallel port printer does not need configuration and is not configured with this window.
To test the network printer(s), choo se Select Printer... from the Setups menu. Select a network printer and print the Software Revision page. It may take up to 5 minutes before the printer name(s) are displayed in the Network Printers window.
Select Clear Print Queue... to remove data waiting to be printed to a printer. The following confirmation screen displays.
041A
Revision A Solar 9500 Information Monitor 4-13
2001085-061
Software Configuration: Monitor Service Menu

Controls and Menu Timeouts

Select Change Timeouts... to set the length of time a pop-up window stays open. The choices are 15 seconds, 30 seconds, 1 minute and No Timeout.
127(
Service related windows do not time out.
042A
4-14 Solar 9500 Information Monitor Revision A
2001085-061

Module Calibration

Software Configuration: Monitor Service Menu
Select Calibrate... in the Module Calibration window to calibrate a module. Calibration menus are provided for CO2 and NBP.
127(
These buttons to display the calibration means are active only if the associated Parameter is active and a module is present.
An on screen keypad is also available for all calibration windows where numeric entry is required. See chapter 5, Module Calibration for further information on module calibration.
043A

Bed/Unit Name Setup

The Bed/Unit Na me S e tu p allows you to enter the bed and unit name. Select Change Name... to open a window with two data entry fields and Show Keyboard, Ok and Cancel buttons. The data fields are Unit Name and Bed Name.
044A
To enter data, select the Show Keyboard button. You can now use the onscreen keyboard to enter the unit name. After entering the uni t name, press the Tab button on the screen keypad to advance the cursor to Bed Name or on to Hide Keyboard. (You can a lso put the cursor in either field by touching the screen.) When complete, select Ok to close the window and save the changes or select Cancel to close the window without saving.
Revision A Solar 9500 Information Monitor 4-15
2001085-061
Software Configuration: Monitor Service Menu

Serial Interface Controls

Select Serial Interfaces... to open the Serial Port Configuration Control window.
045A
In this window you define what device is connected to Serial Port 1 and Serial Port 2. Under Serial Port 1 and Serial Port 2 are two choice lists labeled Device and Display. The Device list includes Off, Touch Screen, RAMS M-200, Polled Parameter, and Writer.
When Touchscreen is selected, the options Display 1, Display 2 or Display 3 (depending on the number of di splays currently available) ap pears in the Display box.
Select Writer when a PRN 50 or PRN 50-M is connected.
127(
If the serial port is configured for a writer, a PRN 50 or PRN 50-M connected to an M-Port will not be recognized by the Solar 9500.
Patient data (Trends, Arrhythmia Review, Alarm History, etc.), Procedures (CO, PA, Wedge, 12SL, etc.) and System information (error logs, etc.) are not supported on a writer.
After making your selections choose OK or Cancel. A confirmation screen appears for you to confirm any changes.
127(
When touchscreen calibration of the display is required, it may take several seconds for the background tasks to complete before the calibration screen appears.
When calibrating the touchscreen, only touch the displayed calibration target. Otherwise, the target area for selecting screen items will be out of position.
Because Polled Parameter is a user-purchased option, it may not be selectable.
Direct Digital Writers (DDW) are NOT supported by the Solar 9500.
4-16 Solar 9500 Information Monitor Revision A
2001085-061

Network Services

Unix Access

Software Configuration: Monitor Service Menu
The Solar 9500 system uses the Unity Network to send Layouts, Parameters and Alarm defaults, and Solar 9500 software to other Solar 9500 beds in the same care unit or other care units on the network.
This window is used to diagnose system problems if the user interface is functional.
Select Run An Xterm... to display a window that allows interaction with the Operating System. When this window opens it displays a system name prompt (i.e., S9500-xxxx%). To close the Run An Xterm window touch the close icon in the upper right corner of the X-term window or, from a keyboard, type the word exit then press Enter.
&$87,21
Use caution when running an Xterm. Critical data may be unintentionally altered or deleted. This window is for system diagnosis by GE service personnel.
Revision A Solar 9500 Information Monitor 4-17
2001085-061

IP Address

Software Configuration: Monitor Service Menu
Select Change Address... in the IP Address window to open two windows, one with the current address, netmask and action buttons, OK and Cancel and the other with a numeric keypad.
&$87,21
The IP address of the Unity Network MC network has been set at the factory to be unique. When changing the IP address, make sure no other equipment on the network has the same address. When connected to the network the Solar 9500 will warn the user if duplicate IP addresses have been assigned.
046B
4-18 Solar 9500 Information Monitor Revision A
2001085-061
Software Configuration: Monitor Service Menu
When Change Address... is selected:
an IP Address window and numeric keypad appear,
the current IP Addresses and netmasks display if they have been entered,
the numeric keypad all ows you to enter an address or netmask,
you can select OK to exit and store,
you can select Cancel to exit without saving, and
if OK is selected, a confirmation box appears so you can confirm any change.
The following network addresses are invalid:
First octet of network address >239
1.0.0.0
1.1.0.0
1.1.1.0
0.1.0.0
0.0.1.0
1.1.1.0
127.0.0.1 (loopback address)
Addresses currently in use on th e network
Browser Server
About Netmasks
The Browser Server IP address must be a legal address on the Unity Network IX network. The example below shows an acceptable and an unacceptable address.
Acceptable Not Acceptable
IX Network 151.1.2.3 151.1.2.3 Browser Server 151.1.2.2 151.1.3.3 Netmask 255.255.255.0 255.255.255.0
The variable netmask capability of the Solar 9500 allows for an advanced network with intelligent routers. Typically for the Unity Network MC network the netmask should be 255.0.0.0. For the Unity Network IX network the netmask should be set to match the rules for subnet segments as defined by the instit uti on’s network topology.
Revision A Solar 9500 Information Monitor 4-19
2001085-061

Time and Date

Software Configuration: Monitor Service Menu
Changing time and date occurs immediately on all display devices.
127(
When a monitor i s conne cted to the MC n etwork , the ti me and date is automatically updated from the network time.
:$51,1*
Loss of patient data history. Changing the time or date settings may result in the loss of patient data history. If one monitor’s time or date i s ch anged, all mon itors o n th e network ‘listen’ and follow suit within 3-5 seconds. Changing the time base of one monitor may cause some loss of patient data history for all the monitors on the network.
When you select Change Time and Date... the Time and Date Control window displays.
047B
You can change the Day, Month, Year, Hour and Minute by selecting a numeric field and using the arrow buttons to increase or decrease the number. You can take one of the following actions:
Select Current Time to reset the time based on the current time of the system clock.
Select OK to display the message, “Press ‘Change’ to change the time. Press ‘Cancel’ to ignore the time change.”
Select Cancel to exit without saving.
4-20 Solar 9500 Information Monitor Revision A
2001085-061

Audio

Software Configuration: Monitor Service Menu
Select Test Audio... to open a window with a slider bar to adjust the test audio volume from 0 to 100% and two buttons, Play Tone and Close.
049A
To test the audio subsystem do the following:
1. Move the slider to desired volume level.
2. Select Play Tone for a tone.
3. Select Stop Tone to turn tone off.

Diagnostic Messages

4. Select Close to end test and close the window.
Under Diagnostic Messages you can display and print all system and Tram Log Files. See chapter 7, Troubleshootin g fo r mor e info r mation.
Revision A Solar 9500 Information Monitor 4-21
2001085-061

Locale Settings

Software Configuration: Monitor Service Menu
Use the Locale Settings option to change languages and country-specific features.
1. Select Locale... The following window appears.
050A
2. Select a language and/or country, then click OK. A dialog box similar to the following appears:
051A
3. Select Yes. The Solar 9500 restarts in the new language.
4-22 Solar 9500 Information Monitor Revision A
2001085-061
Software Configuration: Creating a Custom Layout

Creating a Custom Layout

The Solar 9500 Information Monitor allows you to create and configure custom screen layouts to suit your needs.
With configuration editing enabled, any of the windows on the screen can be resized from any corner. Windows can also be moved around on the screen, by grabbing the edge of the window.
127(
Windows cannot be moved outside of the bed window in which they were created.
If the layout you just configured is some thing you would like to reuse, save the layout before disabling configuration editing. For more information on saving a layout, refer to “Creating a Bed” in this chapter.
When Configuration Editing is enabled, pressing a command button, parameter window or waveform window, will cause a different option menu to be displayed instead of the normal action.

Editing Basics for All Customizable Windows

The following are basic editing principles for working with customizable windows.
:$51,1*6
Do not configure the screen in such a way that parameter and alarm data is obscured. This could delay recognition of critical events when monitoring a patient.
Always turn layout configuration OFF when finished with configuration editing. Failure to turn configuration off may result in accidental configuration changes by the operator and loss of important data.
127(
Configuration editing must be enabled. It may be easier to use a mouse for these operations. For all
operations use the left button only. The right button is not used.
Click and hold the cursor on the b ackground of a win dow to di splay a pop-up menu listing options for that window.
Select the edge o f a window t o s how th e cursor and allow it t o be dragged to an appropriate spot on the display.
Select any corner of a window to show a right angled or “L” shaped cursor and allow the window to be resized.
Revision A Solar 9500 Information Monitor 4-23
2001085-061

Display Features

W W
M W
U P W
Software Configuration: Creating a Custom Layout
All window pop-up option menus contain the Delete key which removes the window from the display.
Waveform Window Options
Delete
052B
053A
Below is a s ample of the Solar 9500 Information Monitor display screen or Bed Window.
aveform indow
nmonitored arameter
indow
essage indow
Bed Window
Parameter Windows
Procedure Timer
Command Button
Waveform Window Button
4-24 Solar 9500 Information Monitor Revision A
Lower Control Buttons
054B
2001085-061
Software Configuration: Creating a Custom Layout

An Overview of Configured Windows

Bed Window
A Bed Window is the top level of configurable windows. It contains other windows which can be saved and retrieved. Only one bed window can be displayed at a given time.
Waveform Window
Waveform Windows contain scrolling or non-scrolling waveforms. You can save and retrieve conten ts of a Wa vefo rm Window layo ut. Wave form Windows can also be configured to print.
Parameter Window
A Parameter Window contains parameter data for any one of the following:
Command Button
Message Window
ECG
BPs 1-8
NBP
Temp 1,6,8
Respiration
SpO2
SvO2
CO2
Gas (Anesthetic Agent)
Cardiac output
BIS
Vent
CCO
A Command Button ties a button to a command operation such as Silence Alarms, NBP Go/Stop, Zero All, View Other Patient, Switch Layout, To Alarm Setup, or Browser.
A Message Window dis plays relevant system messages such as alarm state and print status.
Procedure Timer Window
A Procedure Timer Window functions like a stopwatch to allow specific procedures to be timed.
Revision A Solar 9500 Information Monitor 4-25
2001085-061
Software Configuration: Creating a Custom Layout
Subgroups Window

Creating A Bed Layout

A Subgroups Window allows window layering functionality, with creation, save and retrieval of reusable building blocks.
The top level of the configuration editing menu is Bed Options. This level allows you to create custom bed layouts using basic monitoring building blocks.
To access the Bed Options menu, touch or click anywhere in a blank area between the top header window and the lower control buttons. The Bed Options menu will now open displaying the following options:
Save Bed
Waveform Window
Parameter Window
Command Button
Message Window
Create Sub-group
Retrieve Sub-group
Procedure timer
Background
Delete
127(
If you are using a mouse for Configuration editing, click and hold the left mouse button to scroll do wn t he menu . When y ou rel e as e the l eft button, the highlighted menu item is selected.
4-26 Solar 9500 Information Monitor Revision A
2001085-061
Software Configuration: Creating a Custom Layout
Bed Options
Save Bed – Allows you to name and save the curre ntly di splayed bed for future retrieval.
127(
Bed layout names may contain alphanumeric characters 0 – 9, A – Z, a space, a dash, and an underscore only. If unacceptable characters are entered, a tone will sound.
Waveform Window – Allows you to create and configure a Waveform window.
After you’ve created the Waveform window, touch or click inside it to open the Waveform window options. Selecting Delete removes the waveform window.
For information on selecting waveforms to display in a waveform window, refer to the Solar 9500 Information Monitor Operator’s Manual.
Parameter Window – Allows you to create and configure a parameter window.
After you’ve created the Parameter window, touch or click inside it to open the Parameter window opt ions. Choosing Selections opens the parameter selection window, and choosing Delete removes the parameter window.
Within the Parameter selection window, you can defi ne the paramete r to be displayed in that parameter window.
Command Button – Allows you to create and configure a Command button.
After you’ve created the command button, touch or click inside it to open the Button options. Selecting Assign Function opens the Button setup window. Selecting Background opens the Background color selection window.
Within the Button setup window, select the appropriate function from the list of available options. These options are:
Switch Layout Silence Alarms Zero All NBP Go/Stop Alarm Setup… View Other Patient… Browser... Edit Remote 1 Display Edit Remote 2 Display
After highlighting your choice, select OK to activate your choice and close the window or select Close to close the window without making changes.
Revision A Solar 9500 Information Monitor 4-27
2001085-061
Software Configuration: Creating a Custom Layout
Message Window – Allows you to create and configure a Message window.
After you’ve created the message window, touch or click inside it to open the Message Options window.
To define the message type, select Message Selection. This opens the Message Type Selection window. This window offers choices of Window Type (choose from Alarm State or Scrolling Msgs) or Close. After defining the message type, select Background to define the background color for the message window, or select Delet e to delete this message window from the display. Select Close to close this window.
Create Sub-group – Allows you to create a subgroup, or collection of windows, for configuring your display. Subgroups are usually made of commonly-used sets of parameter windows, waveform windows, command buttons, and mess age windows . Thes e subgr oups c an be sa ved and retrieved for ease of customizing.
After you’ve created a subgroup, touch or click inside it to open the Sub­group options. Sub-group options include the following:
Save Sub-group — allows you to name and save the subgroup for retrieval at a later time.
127(
Bed layout names may contain alphanumeric characters 0 – 9, A – Z, a space, a dash, and an underscore only. If unacceptable characters are entered, a tone will sound.
Command Button — allows you to create and configure a command button within the sub-group.
Waveform Window — allow s you to create and configure a waveform window within the sub- group.
Parameter Window — allows you to create and configure a parameter window within the sub-group.
Message Window — allows you to create and configure a message window within the sub- group.
Procedure timer — allows yo u to creat e a proce dure t imer wi thi n t he sub-group.
Background — opens the Background color selection window, allowing you to set the background color of the subgroup.
Delete — allows you to delete the sub-group.
Retrieve Sub-group – Allows you to retrieve a previously saved subgroup by bringing up the “Select Layout Configuration” window. Choose the desired sub-group from the list and select OK.
Procedure timer – Allows you to open a timer window. After you’ve created the timer, touc h or click inside it to open the Timer optio n s. Select Delete to remove the timer window.
Background – Allows you to open the Background color selection window, where you can set the background color for the bed.
4-28 Solar 9500 Information Monitor Revision A
2001085-061
Software Configuration: Creating a Custom Layout
Delete – Allows you to delete the entire bed window leaving just the background, for creating a new bed.
&$87,21
Use discretion when sel ecting Dele te. All obj ects added t o the bed layout will be removed immediately and without confirmation.
127(
If you accid entall y dele te a bed th at has been prev iousl y sav ed, sele ct Setups on the Lower Control Buttons then select Switch Layout... to bring up the Select Layout Configuration window. Choose the desired bed layout from this list and select OK.

Sample Creation of a Simple Bed Layout

To create a simple bed layout, follow these steps:
Create a Waveform Window
1. Turn on Allow Configuration which is found in the Service menu.
2. Delete the existing screen by finding a portion of the background, invoking the Bed Options pop-up menu and then selecting Delete. If the display is crowded, you may have to first delete other windows in order to get to the bed window’s background.
3. Move the cursor into the middle of the bed window. Click and hold to invoke the Bed Options pop-up menu and select Waveform Window.
4. Move and resize the waveform window by grabbing the edge to move and the corner to resize. Make the window about 1 bed window.
Repeat the last two steps but place this waveform window directly below the first.
2 as high as the
Revision A Solar 9500 Information Monitor 4-29
2001085-061
Software Configuration: Creating a Custom Layout
Create a Parameter Window
Setting up the Waveform Window:
5. In the top waveform window, press on the Configure Window in the
6. Use the arrow in the waveform wi ndow to place the waveforms in t he
Repeat for lower waveform window.
7. Move the cursor into the open area of the bed window, invoke the
8. Move the paramete r win d ow t o the r ig ht of the wa ve f o rm win do w .
9. Press in the middl e of the parameter window to invoke the
10. Select a desired parameter.
Repeat and create several other desired parameter window.
lower left corner of the window. After the menu appears add several waveforms to the window by clicking on the toggle buttons.
desired spot.
pop-up and select Parameter Window.
Parameter Options pop-up menu and select Selections.
Save the Bed
Test the Layout
11. Invoke the pop-up menu over a blank area of the bed window and select Save Bed or choose th e S av e La yout option in the Setups menu.
12. Move the cursor to the text entry field and type TEST.
13. Press OK.
To confirm that the TEST layout was saved, go to the Setups menu and select Switch Layout. When the selection menu appears select MMS Default then press OK. Now repeat, but select TEST instead.
4-30 Solar 9500 Information Monitor Revision A
2001085-061
Software Configuration: Creating a Custom Layout

Print System Settings

Select Print System Settings to print a Solar 9500 System Configuration Record.
127(
GE recommends printing the system configuration for each unit after completion of customer configuration. Retain copies for future reference.
The following information is on the Solar 9500 System Conf iguration Record:
IP Address Settings
Unity Network MC Riser IP Address Unity Network MC Riser IP Netmask Unity Network IX Motherboard IP Address Unity Network IX Motherboard IP Netmask M-Port IP Netmask Browser Motherboard IP Address
Application - Layout Setup
Primary Monitor Layout Secondary Monitor Layout View Only Monitor 1 Layout View Only Monitor 2 Layout Patient View Layout
Serial Interfaces
Serial Port 1 Device Serial Port 1 Display Serial Port 2 Device Serial Port 2 Display
Default Patient Information
Default Patient Units Default Location ID Default Site Number
Application - Display Setup
Motherboard Video Description Video Slot A Description Video Slot B Description Waveform Scrolling Wide Waveform Lines Browser Startup Display Wedge/Cath. Insertion Display Cardiac Output Display ST Display Trends Display Alarm Setup Display Alarm History Display 12 SL Display Ventilator Summary Display
Alarm Graph Setup
ECG Leads BP Site Other Chart Speed Arrhythmia Print Duration Parameter Print Duration Graticules Grid
Stratification
Arrhythmia Review Package 12 SL Package Browser Package Layout Configuration Package Serial Polled Parameters Package
Default Unit Settings
Bed Name Unit Name Language Menu Timeout End Case Defaults
Network Printer Configuration
Entry for each network printer
Printer Selections
Patient Data Printer Procedures Printer Manual Graph Printer Alarm Graph Printer System Printer Remote Graph Printer
Revision A Solar 9500 Information Monitor 4-31
2001085-061

Stratification

Software Configuration: Creating a Custom Layout
Select Stratification to configure software optio ns. Each option is password controlled based on the ethernet address. This message displays:
A separate password is required to access each item. Contact GEMS­IT to obtain the password fo r the monitor. Yo u will need your ethernet address:
xx.xx.xx.xx.xx.xx (w h e r e x = your monitor’s ethernet address)
These options are:
Arrhythmia Review - arrhythmia waveform review
12SL - 12 lead analysis
Browser server access
Layout Configuration - the ability to create and save custo mized layouts
Polled Parameter - serial polled parameter service
106A
4-32 Solar 9500 Information Monitor Revision A
2001085-061

5 Module Calibration

Revision A Solar 9500 Information Monitor 5-1
2001085-061

General

Module Calibration: General
Module or sensor calibrat ions are perf ormed on the CO2, Anes the tic gas , SVO2, and NBP (Tram) modules. Some of the module/sensor calibrations can be performed by the user; others must be performed by service personnel only. The following list describes the module and in which manual the procedure is described.
NBP (Tram) This manual - This chapter. MGA-IR Mainstream CO2 Module This manual - This chapter. MGA-IR Sidestream CO2 Module This manual - This chapter. Capnostat Mainstream Solar 9500 Operator’s Manual Capnostat Dual CO2 Module Solar 9500 Operator’s Manual SAM/SAM 80 Solar 9500 Operator’s Manual
and SAM/SAM 80 Service Manual SVO2 Solar 9500 Operator’s Manual RAMS RAMS Service Manual
5-2 Solar 9500 Information Monitor Revision A
2001085-061

The Service Menu

Module Calibration: The Service Menu
The Service Me nu is for us e by qual ified pers onnel t o troubles hoot, re pair or configure the Solar 9500 Information Processor.
&$87,21
The Service Menu is intended for use only by qualified personnel. Unnecessary tampering with service mode menu options for experimentation purposes may cause a malfunction of the equipment and is not recommende d by the manufacturer.
1. To access the SERVICE MODE menu, execute the following menu sequence, starting from the MAIN menu:
Setups Configuration Service... Service mode...
2. At the numeric touchscreen keypad, enter day and month from monitor screen with leading zeros. (July 4 = 0407)
Enter Password
The following Service menu screen displays.
031B
Revision A Solar 9500 Information Monitor 5-3
2001085-061
Module Calibration: The Service Menu
3. In the Module Calibration section, select Calibrate... The Calibrate... window with buttons NBP..., CO2... and Close display.
060A
127(
These buttons are active only if their Parameters are present and turned on.
5-4 Solar 9500 Information Monitor Revision A
2001085-061
Module Calibration: Calibrate Non-invasive Blood Pressure (NBP)

Calibrate Non-invasive Blood Pressure (NBP)

About the Procedure

The overall accuracy of non-invasive blood pressure (NBP) readings by the monitor depend on the following:
the zero pressure reading, and
the voltage span of the NBP sensor in the monitor.
This procedure provides a method of verifying these items are accurate and also checks the NBP pneumatic circuit plumbing for leaks.

Manufacturer Recommendation

The manufacturer recommends performing this procedure upon initially receiving the monitor, before it is used on a patient, and once each year thereafter. Also, perform the procedure each time the monitor is opened for service or repair, simply to verify the NBP pneumatic circuit plumbing did not develop inadvertent air leaks.

Test Equipment

The following items are required to complete the NBP calibration procedure:
Manometer (Sensym PDM200M or mercury manometer),
NBP tube, pn. 414873-001,
NBP cuff, pn. 9461-301 (any size will work),
Something to wrap the NBP cuff around (PVC pipe or other tube),
NBP bulb (hand pump) to manually over-inflat e the cuff.
The table below lists items for connecting the NBP tube between the manometer and NBP cuff:
Description Part Number Quantity
NBP cuff coupling 400787-001 1 NBP hose coupling 46100-002 1 NBP tee 4745-101 2 Manometer tubing 401582-001 2 ft
Revision A Solar 9500 Information Monitor 5-5
2001085-061
Module Calibration: Calibrate Non-invasive Blood Pressure (NBP)

Calibration Procedure

:$51,1*
When the NBP cuff is used in this procedure, it must be tightly wrapped around a rigid cyli n der or pipe. Do not put the NBP cuff around a human arm during the calibration procedures due to the potential for injury.
1. Enter the Service Menu as described on page 5-3.
2. Select the NBP... button to display the NBP Calibration menu.
061A
NBP Zero Calibration
3. With the NBP cuff and tubing disconnected, perform a zero calibration.
4. Under Calibrate zero select the Start button. The message Zeroing appears in the Calibration messages area. When complete, the message Zero Calibration complete appears.
5. Connect a cuff and manometer to the monitor as shown on the following page.
5-6 Solar 9500 Information Monitor Revision A
2001085-061
Module Calibration: Calibrate Non-invasive Blood Pressure (NBP)
You need an NBP tube.
If you have a Tram 200SL, 250SL, 300A, or 350A module, you need pn 9461-203.
If you have a Tram 400 module with an NBP connector that protrudes from the front panel, you need pn 9461-208.
If you have a Tram 400SL, 450SL, 600SL, 650SL, 800A, 850A or 850SL module, you need pn 414873-001.
NBP Calibration Setup
You need a length of pipe to wrap the NBP cuff around. You can get this from your local hardware store.
You need an NBP cuff. Any size will work . If you don’t have one, order pn 9461-301.
Y ou n eed a coupling t o con nect the NBP tubing to the NBP cuff. Use pn 400787-001.
You need a coupling to connect your manometer adapter to the NBP tube. Use pn 46100-002.
You need an accurate manometer. A digital one, such as a Sensym PDM200M, is nice, but you may use a mercury manometer as well.
You need about two feet of NBP tubing. Order pn 401582-001.
Y o u n eed two tees to co nne ct the manometer hose and the balloon hose to the hose that connects the NBP hos e to the NBP cuff. Use pn 4745-101.
You need a bulb to manually over-inflate the NBP cuff.
062A
Revision A Solar 9500 Information Monitor 5-7
2001085-061
Module Calibration: Calibrate Non-invasive Blood Pressure (NBP)
Set up the Manometer
6. Turn the manometer on and adjust the range switch to the
Start the Gain Calibration Test
7. Under Calibrate gain select the Start button. The message Inflating
8. When the pressure has stabilized, select Enter pressure. Use the on-
1000 mmHg setting.
to cal pressure displays in the Calibration messages window . The NBP module starts pumping up the cuff and the pressure in
mmHg displays under the Cuff pressure text and on the manometer. When the pressure reaches approximately 250 mmHg, the pump shuts off and the pressure drops slowly to about 240 mmHg before stabilizing. If the pressure continues to drop at a rate of 1mmHg or more for every five seconds, there is a leak in the NBP plumbing. If there is a leak in the NBP plumbing, repair it and restart this calibration procedure.
screen keyboard to enter a pressure value that is one mmHg lower than the current manometer reading. The message Gain calibration complete appears in the Calibration messages window.
9. To verify the calibration select Start under Check calibration.
10. The message in the Calibration messages window changes to Calibration check in progress. Verify that the pressure readings (shown as CUFF in the NBP parameter box) on the monitor and manometer are equal (± 1 mmHg) for at least one full minute. The messages window changes to At calibration check pressure.
11. Select Stop under Check calibration and the message Calibration check cancelled appears in the window. The cuff will deflate.
12. As part of the leakage test, use the bulb (hand pump) to increase pressure beyond 300 mmHg. The monitor should deflate the cuff at pressures exceeding 330 mmHg.
13. End the test. Close all open windows.
127(
After approximately 5 minutes, the monitor will deflate the cuff and switch to normal operation.
14. Turn the manometer off and remove the test equipment from the monitor.
5-8 Solar 9500 Information Monitor Revision A
2001085-061
Module Calibration: Calibrate Mainstream CO2

Calibrate Mainstream CO2

About the Procedure

This procedure provides a method of verifying the accuracy of the MGA-IR Mainstream CO2 module.

Manufacturer Recommendation

The manufacturer recommends performing this procedure upon initially receiving the module, before it is used on a patient, and once each year thereafter. Also, perform the procedure each time the module is opened for service or repair.

Test Equipment

The following items are requ ired to successfully complete the CO2 calibration procedure:

Calibration Procedure

Description Part Number Quantity
CO2 Module Calibration Kit 405910-001 1
1. Connect the Calibration Kit to the MGA-IR Mainstream CO2 Module and Sensor as shown below.
MAINSTREAM CO
2
1
CO
2
2
F
E
063A
2. Enter the Service Menu as described on page 5-3.
Revision A Solar 9500 Information Monitor 5-9
2001085-061

0% CO2 Calibration

Module Calibration: Calibrate Mainstream CO2
3. Select the CO2... button to display the CO2 Calibration menu.
064A
4. Under Set Barometric Pressure, enter the current pressure by using the up and down arrows, the slide bar or touching the keypad icon and entering the pressure from the keypad pop-up.
127(
Barometric pressure is used to calculate the percentage of CO2 in the patient’s airway from the mmHg value. This calculation is based on the following equation: %CO2 = mmHg CO2 x 100% / mmHg barometric pressure.
5. Disconnect the calibration adapter from the tubing and move adapter through room air to induce 0% CO2 as shown below.
065A
6. Under Calibrate Module select the Start Calibration. The text on this button changes to Abort Calibration. The Module Status window displays the messages Calibrating and Press button when 0% CO2 is
supplied...
127(
If you select Abort Calibration before 0% or 10% gases are entered, you must restart the calibration procedure.
5-10 Solar 9500 Information Monitor Revision A
2001085-061

10% CO2 Calibration

Module Calibration: Calibrate Mainstream CO2
7. Select Press when gas is supplied and the messages change to Waiting for 10% Gas and Press when 10% CO2 is supplied...
8. Reconnect the calibration adapter to the tubing and apply 10% calibration gas to the airway adapter of the sensor. When CAL gas is applied during calibration, a waveform similar to that displayed below shows the CO2 level.
066A
127(
To view the CO2 waveform, close the Service Menu window and move the CO2 calibration window.
9. Next, select Press when gas is supplied to begin 10% calibration.
A blanking out of both Module Status windows indicates that the calibration was completed.

Calibration Completion

If the calibration failed one of the following messages appear:
127(
One Of The Cal Gases is Wrong
or
Bad 10% Gas Used
The previously stored calibration factors are held if the calibration fails.
Revision A Solar 9500 Information Monitor 5-11
2001085-061
Module Calibration: Calibrate Sidestream CO2

Calibrate Sidestream CO2

About the Procedure

This procedure provides a method of verifying the accuracy of the MGA­IR Sidestream CO2 module.

Manufacturer Recommendation

The manufacturer recommends performing this procedure upon initially receiving the module, before it is used on a patient, and once each year thereafter. Also, perform the procedure each time the module is opened for service or repair.

Test Equipment

The following items are requ ired to successfully complete the CO2 calibration procedure:

Calibration Procedure

Description Part Number Quantity
CO2 Module Calibration Kit 405910-001 1
The calibration procedure is relatively simple and should be performed once per year. The calibration procedure is as follows:
1. Connect the Calibration Kit to the MGA-IR Sidestream CO2 Module and Sensor as shown below.
067A
2. Enter the Service Menu as described on page 5-3.
5-12 Solar 9500 Information Monitor Revision A
2001085-061
Module Calibration: Calibrate Sidestream CO2
3. Select CO2... to display the CO2 Calibration menu.
064A
4. Under Set Barometric Pressure enter the current pressure by using the up and down arrows, the slide bar or touching the keypad icon and entering the pressure from the keypad pop-up.
127(
Barometric pressure is used to calculate the percentage of CO2 in the patient’s airway from the mmHg value. This calculation is based on the following equation: %CO2 = mmHg CO2 x 100% / mmHg barometric pressure.

0% CO2 Calibration

5. Disconnect the calibration adapter from the tubing and move adapter through room air to induce 0% CO2 as shown below.
065A
6. Under Calibrate Module select Start Calibration. The text on this button changes to Abort Calibration. The Module Status window displays the messages Calibrating and Press button when 0% CO2 is
supplied...
127(
If you select Abort Calibration before 0% or 10% gases are entered, you must restart the calibration procedure.
Revision A Solar 9500 Information Monitor 5-13
2001085-061

10% CO2 Calibration

Module Calibration: Calibrate Sidestream CO2
7. Select Press when gas is supplied and the messages change to Waiting for 10% Gas and Press when 10% CO2 is supplied.
8. Reconnect the calibration adapter to the tubing and apply 10% calibration gas to the airway adapter of the sensor. When CAL gas is applied during calibration, a waveform similar to that displayed below will show the CO2 level.
066A
127(
To view the CO2 waveform, close the Service Menu window and move the CO2 calibration window.

Calibration Completion

9. Select Press when gas is supplied to begin 10% gas calibration.
A blanking out of both Module Status windows indicates that the calibration was completed.
If the calibration failed one of the following messages appear:
127(
One Of The Cal Gases is Wrong
or
Bad 10% Gas Used
The previously stored calibration factors are held if the calibration fails.
5-14 Solar 9500 Information Monitor Revision A
2001085-061
Module Calibration: Calibrate Barometric Pressure for Interfaced CO2

Calibrate Barometric Pressure for Inte rfaced CO2

About the Procedure

This procedure sets the baromet ric press ure for an inte rfa ced CO2 usi ng Calibrate... in the Module Calibration window of the Service Menu.
127(
The buttons to display the calibration means are active only if the associated Parameter is active and an interface device is present.

Procedure

1. Enter the Service Menu as described on page 5-3.
2. Select Calibrate... to display the following window.
043A
3. CO2... will be active when an interface device with CO2 parameter is connected to the Solar 9500. Select CO2...
108A
4. Enter the current pressure by using the up and down arrows, the slide bar or touching the keypad icon and en ter ing the pres sure from the keypad pop-up.
127(
Barometric pressure is used to calculate the percentage of CO2 in the patient’s airway from the mmHg value. This calculation is based on the following equation: %CO2 = mmHg CO2 x 100% / mmHg barometric pressure.
Revision A Solar 9500 Information Monitor 5-15
2001085-061
For your notes
Module Calibration: Calibrate Barometric Pressure for Interfaced CO2
5-16 Solar 9500 Information Monitor Revision A
2001085-061
Loading...