GE Sievers 900, Sievers 5310 C Quick Start Manual

Sievers 900 and 5310 C On-Line TOC Analyzers
Quick Start Guide
IMPORTANT
For more detailed instructions, you can download manuals for the Sievers 900 On-Line TOC Analyzer, the Sievers 5310 C On-Line TOC Analyzer, or the Sievers Inorganic Carbon Remover (ICR) from our Web site. Go to www.geinstruments.com, click Library, then Manuals, and make your selection.
Sievers 900 and 5310 C On-Line TOC Analyzer Installation Instructions (continued)
Linear range (900) Linear range (5310 C)
Precision
Accuracy
Analysis time
0.03 ppb - 50 ppm TOC 4 ppb - 50 ppm TOC
<1% of RSD
±2% or ±0.5 ppb, whichever is greater
4 minutes
Sample ow rate
(nominal)
Power requirements
Fuses
Analysis mode: 0.5 mL/min Fast Flush (between samples): 1.0 mL/min
100-240 ± 10% VAC, 50 watts, 50/60 Hz
No user-replaceable fuses; built-in circuit breaker
Sample temperature
Ambient temperature
Sample pressure (in iOS System)
Maximum relative humidity
1 ºC to 95 ºC**
10 ºC to 40 ºC
Up to 250 psig**
Up to 95%, noncondensing
Maximum altitude 3,000 m (9,843 ft)
Installation/Overvoltage category
Serial (RS-232), USB, parallel printer port, 4-20 mA output, two alarms, binary output
II
Calibration stability Typically stable for 12 months
Chemical reagents (prepackaged)
300 mL acid; 300 mL or 150 mL oxidizer
Safety certicates CE, ETL listed. Conforms to UL Std. 61010-1
Certied to CSA C22.2 No. 61010-1
Pollution degree
2
Display
Color, touch-sensitive LCD
Weight
16.9 kg (37.2 Ibs)
* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
** If the sample temperature and pressure are above 60 ºC and 100 psi, the 900 and 5310 C On-Line Analyzer models with the stainless steel iOS is required.
Outputs
Step 6 C
Install the Analog outputs and alarms (continued).
Step 9
Flush the Analyzer
Before placing the Analyzer into normal operation, you should
perform a reagent ush to remove any gas bubbles that may
have formed in the reagent lines; you should then leave the Analyzer powered on for 12 hours to allow the DI water reservoir to properly complete rinse down action.
Size
62.4 cm height x 45.2 cm width x 26.4 cm depth (24.6 in x 17.8 in x 10.4 in)
Required sample line
ow rate
30 mL - 300 mL/min (for On-Line Mode)
Normal operating environment
Intended for indoor use only; avoid sunlight
IP Rating
IP 45
Step 7
Connect the inlet system and waste outlet.
1. Connect the 1/4” Teon tubing with the capsule or inline lter to the sample inlet on the iOS System. Tighten 1/4 turn past nger-tight with a 9/16” open-end wrench. Do not over-
tighten the nut.*
* For raw water and heavy particulates applications, additional
pre-ltration is required. Please contact GE Analytical Instruments
if you need recommendations.
2. Connect the 3/4” OD waste line tubing to the waste outlet on
the iOS System by sliding the tubing over the barb tting.
Table 3: Serial and 4-20 mA Inputs and Outputs (TB3)
12
4-20 mA ( - output)
11
4-20 mA ( + output)
10
Not used
9
Not used
8
Reserved
7
Reserved
6
Reserved
5
Not used
4
Serial (In)
3
Serial (Out)
Pin Number (from left)
Output
2
Serial (Ground)
1
Not used
Step 8
Congure basic Analyzer settings.
Before using the Analyzer, you will need to congure various basic
settings. Some of these settings will not need to be changed again, unless yo u move the Analyzer or recongure the operational environ­ment. Refer to the 900 or 5310 C On-Line Operation and Maintenance Manual for complete details.
Step 8 B
Enable password protection.
Step 8 C
Set the clock and time zone.
Step 8 D
Name the Analyzer location (optional).
Step 8 E
Set reagent expiration dates.
Step 8 F
Set up the data history.
Step 8 G
Set up the printer (pptional).
Step 8 H
Export or print constants.
Step 8 A
Power the Analyzer on.
Step 8 I
Set up data I/O.
©2016. General Electric Company. All Rights Reserved.
DQS 96003-01 Rev A
Americas: T 800 255 6964, 303 444 2009 | geai@ge.com Europe/Middle East/Africa: T 44 (0) 161 864 6800 | geai.europe@ge.com
Asia Pacic: geai.asia@ge.com www.geinstruments.com
For more detailed instructions, you can download operation and maintenance manuals for the Sievers 900 On-Line Analyzer, the Sievers 5310 C On-Line TOC Analyzer, and the 900 Inorganic Carbon Remover (ICR) from our Web site. Go to www.geinstruments.com, click Library, then Manuals, and make your selection.
Where to Find Operation and Maintenance Manuals
Sievers 900 and 5310 C Series On-Line TOC Analyzer Installation Instructions
To avoid exposure to the chemical reagents, wear
acid-resistant gloves, protective clothing, and safety goggles or a face shield when changing the reagent supplies.
Reagent containers are for single-use only – DO NOT REFILL. Relling or reusing reagent containers
will void all Analyzer and parts warranties and nullify any performance claims.
Operation of the Analyzer without the in-line lter
on the sample inlet line will damage the Analyzer and void the warranty. To avoid damaging the
Analyzer, install the lter and replace the lter
element as needed.
See the separate “Warnings” document for additional important information.
Step 1
Unpack and inspect the Analyzer.
Step 2
Complete the identication records.
Step 3
Select a location for the Analyzer. Refer to the System
Specications chart.
WARNINGS
To avoid false TOC readings and possible damage to
the Analyzer, always make sure the sample is owing
through the iOS System and the DI water reservoir is
lled before starting analysis.
Figure 1 - Reagent cartridges
NOTE: Hazardous reagents (ammonium persulfate
and phosphoric acid) are used in the Analyzer. Before installing the reagents, please read the Material Safety Data Sheets (MSDS) contained in a pouch on the top of the reagent shipping box for proper handling precautions and spill or leak procedures.
Step 5
Fill the DI water reservoir.
1. Remove the rubber inlet cover on the DI water reservoir.
2. Slide the bottle nozzle into the inlet hole, and gently squeeze the
bottle. Be sure to ll the reservoir until the water reaches the ll
line, just below the inlet.
Step 4
Install the reagent cartridges. (See Figure 1.)
The oxidizer and acid reagents are shipped from GE Analytical
Instruments in specic packaging for safety in transit. Carefully
read the attached MSDS sheets prior to opening the packaging. Within the packaging are two reagent cartridges. Each reagent
cartridge is identied with labels that indicate which cartridge
contains ammonium persulfate and which cartridge contains phosphoric acid.
1. Locate the acid supply line, indicated by the Acid label. Attach
the PEEK nut to the acid reagent container and tighten nger-
tight.
2. Slide the valve on the acid container to the open position by pushing the green button all the way in. Refer to the labeling on the container for proper positioning.
Step 6
Install accessories and cables.
If you will be using a 900 ICR unit with your conguration, make
sure it is installed. Follow the complete set of operation instructions in the ICR Operation and Maintenance Manual. (See downloading instructions on the back page of this Quick Start Guide.)
Step 6 A
Install the 900 Inorganic Carbon Remover (ICR).
Installation of the Sievers 900/5310 C On-Line TOC Analyzer requires an external source of AC power connected to the enclosure using a water­tight conduit connector. The electrical connection should be performed by a qualied electrician. An external switch or circuit breaker is recom­mended to facilitate maintenance and servicing of the Analyzer. It should be installed near the Analyzer and be clearly marked as the disconnecting device for the Analyzer. (See Figure 2.)
NOTE: Before installing any wiring inside the Analyzer, put on a grounding strap for ESD protection.
Step 6 B
Connect to a power supply.
The Analyzer has three terminal blocks, arranged horizontally in the electrical enclosure inside the Analyzer; remove the enclosure cover by loosening the front screws. Consult Tables 1–3 for a list of functions on each terminal block. The output
and alarm connections should be installed by a qualied
electrician. Refer to the 900 or 5310 C On-Line Operations and Maintenance Manual for complete details.
Step 6 C
Install the analog outputs and alarms.
3. Slide the acid container into the reagent enclosure, with the valve pointing down and toward the front of the reagent enclosure.
4. Locate the oxidizer supply line, indicated by the Oxidizer label. Attach the PEEK nut to the oxidizer container and
tighten nger-tight.
5. Slide the valve on the oxidizer container to the open position by pushing the green button all the way in. Refer to the labeling on the container for proper positioning.
6. Slide the oxidizer container into the reagent enclosure, with the valve pointing down and toward the front of the reagent enclosure.
Table 1: Inputs and Outputs (TB2)
10 Ground, for binary input
9 24 V (+ output, for binary input)
8 Start/Stop - (for binary input)
7 Start/Stop + (for binary input)
6 Alarm 2 (NO*) (output)
5 Alarm 2 (NC*) (output)
4 Alarm 2 (Common) (output)
3 Alarm 1 (NO*) (output)
2
Alarm 1 (NC*) (output)
1
Alarm 1 (Common) (output)
* NC = normally closed *NO = normally open
Pin Number (from left)
Input/Output
Table 2: Inputs and Outputs (TB1)
8 900 Lab: Reserved
5310 C Lab: Stream 1 - (input) 0
7 900 Lab: Reserved
5310 C Lab: Stream 1 + (input)
6 End of Batch (NO*) (output)
5 End of Batch (NC*) (output)
4 End of Batch (Common) (output)
3 Reserved
2
Reserved
1 Reserved
* NC = normally closed *NO = normally open
Pin Number (from left)
Input/Output
Figure 2 - Wiring AC Power Conduit (Arrows indicate connection points.)
Loading...