GE Protimeter Timbermaster Quick Start Manual

Page 1
GE GE Protimeter
Sensing Timbermaster
Introduction
The Protimeter Timbermaster is a conductivity moisture meter designed for use in wood. Moisture measure­ments can be taken using the integral pin electrodes, or using the meter in combination with moisture probes or hammer electrodes. When used with the temperature probe, the moisture measurements are automatically cor­rected with respect to temperature. This feature is particularly relevant for users testing wood that is signifi­cantly above or below 20°C (70°F). The Timbermaster is switched ON by
pressing momentarily and switched OFF by pressing and holding
for 2 seconds or more. The instru­ment will switch OFF automatically after 5 minutes unless the default set­ting is changed (see section 3).
1 Using the Protimeter Timbermaster without the Temperature Probe
The Timbermaster is calibrated for wood at 20°C (70°F). In general, tim­ber that is hotter than 20°C will give higher readings and timber colder than 20°C will give lower readings. An approximate manual correction of
0.5% moisture content per 5°C may be subtracted from timber that is above 20°C. For timber that is below 20°C, a manual correction of 0.5% moisture content per 5°C may be added to the measured value.
1
®
Page 2
1.1 Readings with Integral Electrode Pins
Remove the cap to expose the needle electrodes and switch the instrument
ON by pressing . Select the appro- priate wood calibration scale (A, B,C,E,F,G, H or J) by referring to the enclosed Protimeter wood calibration tables and pressing . Push the pins into the surface of the wood and observe the reading.
1.2 Readings with Moisture Probe or Hammer Electrode
Connect the moisture probe or (optional) Hammer Electrode to the
3.5 mm socket on the right hand side of the Timbermaster and switch ON
by pressing . Select the appropri­ate wood calibration scale (A, B, C, E, F, G, H or J) by referring to the enclosed Protimeter wood calibration tables and pressing
. Drive the
moisture probe pins or hammer elec­trode needles into the wood and observe the reading.
2 Using the Protimeter Timbermaster with the Temperature Probe
If the timber being measured is signif­icantly above or below 20°C (70°F) then the T imber master shou ld be used in conjunction with the Temperature Probe. When this probe is connected, the Timbermaster automatically cor­rects the measured moisture value with respect to temperature.
2.1 Automatically Temperature Corrected (ATC) Readings
Switch the Timbermaster ON and select the appropriate wood calibra­tion scale as detailed in sections 1.1 or 1.2. Using either a Hammer Elec­trode, or a hammer and nail of nomi­nal 2 mm diameter, make a hole in the wood to be tested. Remove the Ham­mer Electrode or nail and push the Temperature Probe into the hole until the tip is at the required depth. Con­nect the Temperature Probe to the Timbermaster via the 2.5 mm socket.
2
Page 3
To obtain the automatically tempera­ture corrected (ATC) moisture value, take moisture readings as detailed in sections 1.1or 1.2 while the Tempera­ture Probe is positioned in the wood and connected to the Timbermaster . If the temperature of the wood is assumed to be equal to the ambient air temperature, ATC moisture values can be obtained by holding the con­nected temperature probe in air. Switch between temperature and moisture displays by pressing
.
3Setup Mode
The setup mode is entered by press­ing and
simul taneously. This
action displays the following infor­mation about the instrument in the following sequence:
1. Firmware version number, for example 1.08.
2. Product part number, for example bLd5601.
3. Firmware date in yy-mm-dd for­mat, for example (00-05-28).
The user then has the option of chang­ing the default setting for the tempera­ture display (°C or °F) and the automatic switch OFF time (disable automatic switch OFF or set from 1 to 9 minutes) by changing the codes as detailed in the table below. The first code digit is
changed by pressing and the second digit by pressing
. Confirm the new
settings by pressing .
Code Description
0=0 No action 0=1 Resets all user settings to the
defaults (°C, 5 minutes) 1=0 Selects °C for temperature display 1=1 Selects °F for temperature display 2=0 Disables automatic switch OFF 2=1 Set automatic switch OFF to 1 min. 2=2 Set automatic switch OFF to 2 min.
2=.. ....and so on to ...
2=9 Set automatic switch OFF to 9 min.
3
Page 4
4 Calibration Check
The calibration of the Timbermaster is checked by holding the electrode needles across the exposed wires of the ‘calcheck’ device (supplied) or across the terminals of the Protimeter Checkbox (optional accessory). When checking the calibration, the A scale should be selected and the tem­perature probe must be disconnected. Correctly calibrated Timbermaster will register a (%MC) value in the range of 17.8 to 18.3.
5 Care and Maintenance
When the Timbermaster is not in use, keep it in its pouch together with its accessories. Store the kit in a stable, dust-free environment out of direct sunlight. Remove the batteries from the instrument if it is to be stored for periods of more than four weeks, or when the low battery power symbol appears on the display. Check the condition of accessories used with the Timbermaster instrument on a regular basis and replace them if they become worn or damaged.
The information contained in this leaflet is given in good faith. As the method of use of the instrument (and its accesso­ries) and the interpretation of the read­ings are beyond the control of the manufacturers, they cannot accept responsibility for any loss, consequen­tial or otherwise, resulting from its use.
USA
1100 Technology Park Drive Billerica, MA 01821-4111 www.gesensing.com/protimeterproducts
Ireland
Shannon Industrial Estate Shannon, County Clare www.gesensing.com/protimeterproducts
Protimeter Timbermaster is a registered trademark of GE
®
User Instructions
4
INS5601, Rev. A
January 2006
Loading...