GE MOA 280i Operation And Maintenance Manual

Sievers* Nitric Oxide Analyzer NOA 280i
Operation and Maintenance Manual
6060 Spine Road
Boulder, CO 80301 USA
phone 800.255.6964 • 303.444.2009
fax 303.444.9543 DLM 14291 Rev A
www.geinstruments.com Printed in USA ©2006
IDENTIFICATION RECORDS
Record the following numbers as they are listed on the identification labels located on the back panel of the NOA and the front of the vacuum pump.
Analyzer serial no.
Pump serial no.
Warranty Start Date
Date of receipt/Installation
*Trademark of General Electric Company; may be registered in one or more countries.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
ii
PRINTING HISTORY
The information contained in this document may be revised without notice. Sievers Instruments makes no warranty of any kind with regard to this material. Sievers Instruments shall not be liable for errors contained herein or for incidental, or consequential damages in connection with the furnishing, performance or use of this material.
No part of this document may be photocopied or reproduced without the prior written consent of Sievers Instruments, Inc.
Initial Printing December 2000 DLM 14290-01 Revision A March 2001 DLM 14291 Revision A May 2006
Printed in USA
GE Analytical Instruments ©2006 DLM 14291 Rev. A
iii
Standard Limited Warranty
GE Analytical Instruments warrants its products (Sievers®, GE Analytical™ and Leakwise™) against defects in materials and workmanship. GE Analytical Instruments will, at its option, repair or replace instrument components that prove to be defective with new or remanufactured components (i.e., equivalent to new). The warranty set forth is exclusive and no other warranty, whether written or oral, is expressed or implied.
Warranty Term
The GE Analytical Instruments warranty term is thirteen (13) months ex-works, or twelve (12) months from installation or start up by GE Analytical Instruments certified service personnel. In no event shall the standard limited warranty coverage extend beyond thirteen (13) months from original shipment date.
Warranty Service Warranty Service is provided to customers through telephone support (1-800-255-
6964), Monday - Friday, from 8:00 a.m. to 5:00 p.m. (Mountain Time), excluding all company and legal holidays. Telephone support is provided for troubleshooting and determination of parts to be shipped from GE Analytical Instruments to the customer in order to return the product to operation. If telephone support is not effective, the product may be returned to GE Analytical Instruments for repair or replacement. In some instances, suitable instruments may be available for short duration loan or lease. GE Analytical Instruments warrants that any labor services provided shall conform to the reasonable standards of technical competency and performance effective at the time of delivery. All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative, or designate. GE Analytical Instruments warrants these services for 30 days after the authorization and will correct any qualifying deficiency in labor provided that the labor service deficiency is exactly related to the originating
GE Analytical Instruments ©2006 DLM 14291 Rev. A
iv
event. No other remedy, other than the provision of labor services, may be applicable. Repair components (parts and materials), but not consumables, provided in the course of a repair, or purchased individually, are warranted for 90 days ex-works for materials and workmanship. In no event will the incorporation of a warranted repair component into an instrument extend the whole instrument’s warranty beyond its original term. Consumables (e.g., dilution standards, verification solutions, and UV lamps, etc.) are warranted to the extent of their stated shelf life, provided these items are maintained within the stated environmental limitations. Warranty claims for consumables and verification standards are limited to the replacement of the defective items, prorated from the time of claim to the expiration of shelf life.
Shipping
A Repair Authorization Number (RA) must be obtained from the Technical Support Group before any product can be returned to the factory. GE Analytical Instruments will pay freight charges, exclusive of any taxes and duties, for replacement or repaired products shipped to the customer site. Customers shall pay freight charges, including all taxes and duties, for all products returning to GE Analytical Instruments. Any product returned to the factory without an RA number will be returned to the customer.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
v
GE Analytical Instruments ©2006 DLM 14291 Rev. A
vi
TABLE OF CONTENTS
1. INTRODUCTION...................................................................................................................................... 1-1
SAMPLE INLET SYSTEMS.....................................................................................................................................1-5
SAMPLE FLOW CONTROL DEVICE ........................................................................................................................ 1-6
OZONE FLOW CONTROL MODULE.......................................................................................................................1-6
OZONE GENERATOR..........................................................................................................................................1-6
CHEMILUMINESCENT REACTION CHAMBER AND OPTICAL FILTER ............................................................................ 1-6
PHOTOMULTIPLIER TUBE AND COOLED HOUSING ................................................................................................. 1-7
VACUUM PUMP AND OZONE DESTRUCTION TRAP................................................................................................. 1-7
ELECTRONICS.................................................................................................................................................... 1-7
ANALOG, PRINTER AND RS-232 OUTPUTS ......................................................................................................... 1-8
EXHALATION PRESSURE TRANSDUCER .................................................................................................................1-9
THERMAL MASS FLOW METER ............................................................................................................................ 1-9
2. SPECIFICATIONS..................................................................................................................................... 2-1
3. MENUS AND CONTROL OVERVIEW ..................................................................................................... 3-1
MAIN MENU...................................................................................................................................................... 3-1
STATUS SCREEN................................................................................................................................................ 3-1
ANALYSIS ......................................................................................................................................................... 3-2
MEASUREMENT MENU ........................................................................................................................................ 3-3
MAIN MENU OPTIONS........................................................................................................................................ 3-6
Control..................................................................................................................................................... 3-6
Calibration..............................................................................................................................................3-7
Messages..................................................................................................................................................3-7
Maintenance............................................................................................................................................ 3-8
TIME-OUT FUNCTION ........................................................................................................................................3-8
4. INSTALLATION........................................................................................................................................ 4-1
LOCATION ........................................................................................................................................................ 4-1
POWER REQUIREMENTS .....................................................................................................................................4-1
ENVIRONMENTAL CONSIDERATIONS.....................................................................................................................4-2
TOOLS AND ADDITIONAL SUPPLIES...................................................................................................................... 4-2
Tools......................................................................................................................................................... 4-2
Gases........................................................................................................................................................4-2
Data Collection....................................................................................................................................... 4-2
VACUUM PUMP SETUP.......................................................................................................................................4-4
GE Analytical Instruments ©2006 DLM 14291 Rev. A
vii
Step 1 – Add Oil to the pump...............................................................................................................
4-4
Step 2 - Install Pump Inlet Fitting ...................................................................................................... 4-4
Step 3- Install the Chemical Trap Mounting Bracket.......................................................................4-5
Step 4 - Install the Pump Outlet Fitting............................................................................................ 4-6
Step 5 - Install the Chemical Trap and Vacuum Hoses....................................................................4-6
Step 6 - Connect Power Cord to Vacuum Pump and Turn On Pump Power Switch.................... 4-9
CONNECTIONS TO NOA.....................................................................................................................................4-9
Vacuum Pump Power Cord and Vacuum Hose................................................................................... 4-9
Vacuum Test............................................................................................................................................ 4-9
Gas for Ozone Generator.................................................................................................................... 4-11
Frit Restrictor....................................................................................................................................... 4-12
Computer, Printer and Analog Signal Connections........................................................................ 4-12
Setting the Clock.................................................................................................................................. 4-13
CONFIGURATION MENU OPTIONS...................................................................................................................... 4-14
Com Port................................................................................................................................................ 4-14
Pressure Units.......................................................................................................................................4-15
SETTING THE CONSUMABLES INSTALLATION DATA..............................................................................................4-16
START-UP....................................................................................................................................................... 4-16
5. INSTALLATION AND SETUP: GAS-PHASE MEASUREMENTS............................................................5-1
INSTALLATION OF GAS SAMPLING PACKAGE .........................................................................................................5-1
INSTALLATION OF THERMAL MASS FLOWMETER....................................................................................................5-2
NOA SETUP FOR GAS-PHASE MEASUREMENTS..................................................................................................... 5-4
Exhalation Mode..................................................................................................................................... 5-4
Nitric Oxide Mode..................................................................................................................................5-7
6. CALIBRATION.......................................................................................................................................... 6-1
ZERO GAS CALIBRATION ....................................................................................................................................6-1
Calibration with Zero Air Filter..........................................................................................................6-2
Calibrating with Zero Air Cylinder.....................................................................................................6-2
Zero Gas Calibration Warnings............................................................................................................ 6-4
NO CALIBRATION GAS....................................................................................................................................... 6-4
Calibration Gas Warnings.....................................................................................................................6-6
Calculation of Gas Concentration....................................................................................................... 6-8
INDEPENDENT CALIBRATION OF PPB AND PPM RANGES..........................................................................................6-8
ACCURACY OF PPB LEVEL MEASUREMENTS USING PPM LEVEL CALIBRATION ............................................................. 6-9
FLOW/RESPONSE CHARACTERISTICS OF NOA 280I ...........................................................................................6-10
CALIBRATION AT LOWER FLOW RATES .............................................................................................................. 6-11
GE Analytical Instruments ©2006 DLM 14291 Rev. A
viii
7.
ON-LINE EXHALED NITRIC OXIDE .......................................................................................................7-1
ASSEMBLY OF THE ACCURATE NO BREATH KIT.................................................................................................... 7-2
CONNECTION OF THERMAL MASS FLOWMETER..................................................................................................... 7-3
CONNECTION OF GAS SAMPLING AND PRESSURE TUBING ......................................................................................7-3
INSPIRATORY GAS CONNECTIONS........................................................................................................................ 7-4
Inspiratory Gas Filter............................................................................................................................ 7-4
NOA SETUP..................................................................................................................................................... 7-4
PERFORMING THE MANEUVER............................................................................................................................. 7-5
SELECTION OF NO PLATEAU..............................................................................................................................7-6
FLOW/PRESSURE CHARACTERISTICS OF ACCURATE NO RESTRICTORS ................................................................... 7-8
MODELS OF NITRIC OXIDE PRODUCTION IN THE AIRWAYS .....................................................................................7-8
CLEANING THE ACCURATE NO BREATH KIT AND FLOWMETER...............................................................................7-9
Disassemble the Valve........................................................................................................................... 7-9
Prewash the Components...................................................................................................................7-10
Sterilization.......................................................................................................................................... 7-10
Rinsing.................................................................................................................................................... 7-10
Drying..................................................................................................................................................... 7-10
CHECKING THE INSPIRATORY GAS FILTER..........................................................................................................7-11
8. OFF-LINE EXHALED NITRIC OXIDE (BAG SAMPLING) ...................................................................... 8-1
ASSEMBLY OF VITAL CAPACITY BAG COLLECTION KIT........................................................................................... 8-1
ASSEMBLY OF DEADSPACE DISCARD BAG COLLECTION KIT .................................................................................... 8-2
CLEANING THE BAGS .........................................................................................................................................8-5
COLLECTING THE SAMPLES – VITAL CAPACITY BAG KIT........................................................................................8-6
Connecting the bag to the filler.........................................................................................................8-7
Instructing the Subject and Collecting the Samples ....................................................................... 8-7
Disconnecting the bag from the filler and sealing the bag...........................................................8-8
COLLECTING THE SAMPLES – DEADSPACE DISCARD BAG KIT.................................................................................. 8-9
Connecting the bag to the filler.........................................................................................................8-9
Instructing the Subject and Collecting the Samples ..................................................................... 8-10
Disconnecting the bag from the filler and sealing the bag.........................................................8-11
ANALYZING THE SAMPLES ................................................................................................................................ 8-11
Analysis Setup....................................................................................................................................... 8-11
NOA SETUP................................................................................................................................................... 8-11
CLEANING THE BAG KITS ................................................................................................................................. 8-13
Vital Capacity Bag Kit......................................................................................................................... 8-13
Deadspace Discard Bag Kit ................................................................................................................. 8-14
GE Analytical Instruments ©2006 DLM 14291 Rev. A
ix
F
LOW/PRESSURE CHARACTERISTICS OF BAG KITS.............................................................................................. 8-14
STABILITY OF NO IN MYLAR BAGS....................................................................................................................8-15
OFF-LINE VERSUS ON-LINE EXHALED NO MEASUREMENTS..................................................................................8-17
TESTING BAGS FOR PINHOLE LEAKS.................................................................................................................. 8-17
9. BREATH-BY-BREATH AND CHAMBER SAMPLING FOR EXHALED NITRIC OXIDE..........................9-1
BREATH-BY-BREATH MEASUREMENTS.................................................................................................................. 9-1
Spontaneously Breathing Subjects......................................................................................................9-1
Ventilated Subjects...............................................................................................................................9-2
NOA SETUP.....................................................................................................................................................9-3
NO/Pressure Offset................................................................................................................................9-3
Humidified Circuits................................................................................................................................9-3
CHAMBER SAMPLING.......................................................................................................................................... 9-4
NOA SETUP.....................................................................................................................................................9-6
10. NASAL NITRIC OXIDE......................................................................................................................10-1
RECOMMENDED SETUP.....................................................................................................................................10-1
PERFORMING THE MANEUVER...........................................................................................................................10-2
NOA SETUP...................................................................................................................................................10-3
11. INSTALLATION AND SETUP: LIQUID MEASUREMENTS..............................................................11-1
SUPPLIES........................................................................................................................................................ 11-1
Gases......................................................................................................................................................11-1
Reagents................................................................................................................................................ 11-1
Lab Equipment...................................................................................................................................... 11-1
SETUP OF PURGE VESSEL.................................................................................................................................11-2
Connections of tubing to glassware.................................................................................................. 11-3
Procedure for Tightening Swagelok Fittings.................................................................................................11-4
DILUTION OF ANTI-FOAMING AGENT.................................................................................................................11-7
NOA SETUP FOR LIQUID MEASUREMENTS.......................................................................................................... 11-8
DEPROTEINIZATION PROCEDURES....................................................................................................................11-10
Cold ethanol precipitation............................................................................................................... 11-11
Zinc Sulfate/Sodium Hydroxide precipitation.............................................................................. 11-11
12. MEASUREMENT OF NITRIC OXIDE AND NITRITE IN LIQUID SAMPLES....................................12-1
APPARATUS FOR NITRITE REDUCTION ............................................................................................................... 12-1
PREPARATION OF THE NITRITE REDUCING AGENT ..............................................................................................12-2
ADJUSTMENT OF PURGE GAS FLOW RATE .........................................................................................................12-3
GE Analytical Instruments ©2006 DLM 14291 Rev. A
x
A
DJUSTMENT OF LIQUID LEVEL......................................................................................................................... 12-3
Preparation of Stock Solution............................................................................................................ 12-5
Preparation of Dilute Standards....................................................................................................... 12-7
WATER BLANKS .............................................................................................................................................. 12-8
INJECTION TECHNIQUE .................................................................................................................................... 12-9
PREPARATION OF CALIBRATION CURVE .............................................................................................................12-9
LINEAR RANGE AND OFF-SCALE PEAKS.............................................................................................................. 12-9
REPEATABILITY ............................................................................................................................................. 12-10
NITRITE CONTAMINATION .............................................................................................................................. 12-10
SAMPLE ANALYSIS ......................................................................................................................................... 12-12
BACKGROUND NITRITE................................................................................................................................... 12-12
REPLACING THE REDUCING AGENT AND OPENING THE PURGE VESSEL ............................................................... 12-13
SEPTUM REPLACEMENT.................................................................................................................................. 12-14
CLEANING THE PURGE VESSEL........................................................................................................................ 12-14
CLEANING OF THE IFD FILTER ....................................................................................................................... 12-16
LONG-TERM MAINTENANCE OF THE PURGE VESSEL AND BUBBLER........................................................................ 12-16
13. MEASUREMENT OF NITRATE, NITRITE AND NITRIC OXIDE IN LIQUID SAMPLES................. 13-1
APPARATUS FOR NITRATE REDUCTION .............................................................................................................. 13-1
Preparation of the Nitrate Reducing Agent.................................................................................... 13-2
Preparation of 1M NaOH..................................................................................................................... 13-2
Startup Procedures for Nitrate Reduction...................................................................................... 13-3
ADJUSTMENT OF PURGE GAS FLOW RATE.........................................................................................................13-4
LEAK CHECK FOR PURGE VESSEL...................................................................................................................... 13-5
ADJUSTMENT OF LIQUID LEVEL......................................................................................................................... 13-5
Preparation of Stock Solution............................................................................................................ 13-6
Preparation of Dilute Standards....................................................................................................... 13-7
WATER BLANKS .............................................................................................................................................. 13-9
INJECTION TECHNIQUE .................................................................................................................................... 13-9
PREPARATION OF CALIBRATION CURVE ...........................................................................................................13-10
ANALYSIS OF SAMPLES AND STANDARDS...........................................................................................................13-10
Serum and Plasma Samples.............................................................................................................. 13-10
NITRATE CONTAMINATION ............................................................................................................................. 13-11
REPLACING THE REDUCING AGENT AND OPENING THE PURGE VESSEL ............................................................... 13-11
OPENING THE GAS BUBBLER .......................................................................................................................... 13-12
SEPTUM REPLACEMENT.................................................................................................................................. 13-13
CLEANING THE PURGE VESSEL........................................................................................................................ 13-13
GE Analytical Instruments ©2006 DLM 14291 Rev. A
xi
C
LEANING THE GAS BUBBLER.........................................................................................................................13-15
CLEANING THE BUBBLER TUBING.................................................................................................................... 13-16
CLEANING OF THE IFD FILTER .......................................................................................................................13-16
Long-term maintenance of the purge vessel and bubbler..........................................................13-17
14. OTHER LIQUID MEASUREMENT TECHNIQUES............................................................................14-1
MEASUREMENT OF NITROSOTHIOLS ...................................................................................................................14-1
Cu(I)/Cysteine Reagent.......................................................................................................................14-1
Preparation of Reducing Agent.......................................................................................................................14-1
Preparation of Nitrosothiols Standards..........................................................................................................14-2
Copper(I)/Iodide/Iodine Reagent......................................................................................................14-2
Preparation of the Reducing Agent................................................................................................................ 14-2
Preparation of S-Nitroso-Albumin...................................................................................................................14-3
Treatment of Plasma Samples ........................................................................................................................14-3
MEASUREMENT OF IRON-BOUND NO ................................................................................................................. 14-3
HEADSPACE MEASUREMENT OF NITRIC OXIDE .................................................................................................... 14-4
Apparatus for Headspace Analysis.................................................................................................... 14-5
Sample Collection................................................................................................................................14-5
Preparation of Standards for Headspace......................................................................................... 14-6
DYNAMIC HEADSPACE ANALYSIS........................................................................................................................ 14-7
15. MAINTENANCE.................................................................................................................................15-1
CHANGING THE VACUUM PUMP OIL ..................................................................................................................15-1
CHANGING THE HOPCALITE TRAP.....................................................................................................................15-4
CLEANING THE CHEMILUMINESCENCE REACTION CELL......................................................................................... 15-5
VACUUM TEST .............................................................................................................................................. 15-10
RESET THE CELL CLEANING TIMER..................................................................................................................15-10
LIGHT LEAK TEST.......................................................................................................................................... 15-10
COOLER MAINTENANCE..................................................................................................................................15-11
TESTING AND CLEANING THE FLOW RESTRICTOR FRIT...................................................................................... 15-13
GAS SAMPLING PARTICLE FILTER.................................................................................................................... 15-14
SECURITY .....................................................................................................................................................15-14
16. TROUBLESHOOTING.......................................................................................................................16-1
ERRORS..........................................................................................................................................................16-1
POSSIBLE ERRORS AND REMEDIES...................................................................................................................... 16-2
E 01 – Setup Data Corrupted, Check Before Running .................................................................... 16-2
E 02 – Cell Pressure was Above the Limit........................................................................................ 16-2
GE Analytical Instruments ©2006 DLM 14291 Rev. A
xii
E 03 – Ozone Supply Pressure was Below the Limit.......................................................................
16-2
E 05 – Cooler Temperature Above the Limit................................................................................... 16-3
WARNINGS ..................................................................................................................................................... 16-3
W 09–Pump Oil Needs to be Replaced..............................................................................................16-4
W 10–Pump Oil Needs to be Replaced Soon..................................................................................... 16-4
W 11–Hopcalite Needs to be Replaced.............................................................................................16-4
W 12–Hopcalite Needs to be Replaced Soon.................................................................................... 16-4
W 13–Reaction Cell Needs to be Cleaned........................................................................................ 16-4
W 14–Reaction Cell Needs to be Cleaned Soon............................................................................... 16-4
W 15–Cooler Needs to be Serviced.................................................................................................... 16-5
W 16–Cooler Needs to be Serviced Soon..........................................................................................16-5
CLEARING THE ERROR AND WARNING STACKS ................................................................................................... 16-6
START-UP TESTS............................................................................................................................................. 16-6
Vacuum Pump....................................................................................................................................... 16-7
Cooler Temp......................................................................................................................................... 16-7
Supply Pressure....................................................................................................................................16-7
PMT Signal............................................................................................................................................. 16-7
TROUBLESHOOTING THE NOA.......................................................................................................................... 16-8
No Power to NOA.................................................................................................................................. 16-9
No Display............................................................................................................................................ 16-10
CELL PRESSURE TOO HIGH OR TOO LOW......................................................................................................... 16-11
GAS SAMPLING PROBLEMS.............................................................................................................................. 16-11
HIGH BACKGROUND NO AFTER CALIBRATION..................................................................................................16-12
LIQUID MEASUREMENTS PROBLEMS ................................................................................................................. 16-12
Low Sensitivity.................................................................................................................................................16-12
Leaks in Purge System....................................................................................................................................16-13
Low Conversion for Nitrate............................................................................................................................16-13
Poor Repeatability............................................................................................................................. 16-13
Syringe Problems.............................................................................................................................................16-14
Foaming of theVCl3 Reagent..........................................................................................................................16-14
Injection Technique........................................................................................................................................16-14
Contamination .................................................................................................................................................16-14
High background Signal and Rising Baselines................................................................................ 16-15
Ghost Peaks......................................................................................................................................... 16-15
GE Analytical Instruments ©2006 DLM 14291 Rev. A
xiii
GE Analytical Instruments ©2006 DLM 14291 Rev. A
xiv
1. INTRODUCTION
The Model 280i Nitric Oxide Analyzer (NOA™) from Sievers Instruments is a high­sensitivity detector for measuring nitric oxide based on a gas-phase chemiluminescent reaction between nitric oxide and ozone:
NO + O3 -> NO
*
NO
-> NO2+ hν
2
*
2
+ O2
Emission from electronically excited nitrogen dioxide is in the red and near­infrared region of the spectrum, and is detected by a thermoelectrically cooled, red-sensitive photomultiplier tube. The detection limit of the NOA for measurement of gas-phase NO is ~0.5 part per billion by volume. The detection limit for measurement of NO and its reaction products in liquid samples is ~ 1 picomole.
In biological systems, nitric oxide is produced from the enzymatic oxidation of arginine. Three isoforms of the enzyme nitric oxide synthase (NOS) have been identified in many cell types: endothelial NOS, neuronal NOS, and inducible NOS. The biological functions of NO include action as a vasodilator, neurotransmitter, cytotoxic agent, inhibit of platelet aggregation, and activator of smooth muscle proliferation. Nitric oxide is also present in exhaled breath and may be a useful marker of airway inflammation.
-
In solution, nitric oxide reacts with molecular oxygen to form nitrite (NO with oxyhemoglobin and superoxide anion (O
-
) to form nitrate (NO- ) . NO also
2
), and
2
reacts with thiols to form S-nitroso compounds, amines to form nitrosamines, and metals to form metal-nitrosyl complexes. In the gas phase, NO reacts with high concentrations of oxygen to form nitrogen dioxide.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-1
UNITED STATES REGULATORY REQUIREMENTS
CAUTION – INVESTIGATIONAL DEVICE:
Limited by United States Law to Investigational use.
EXHALED BREATH AND LIQUID APPLICATIONS:
For Research use Only. Not for Use in Diagnostic Procedures.
SAFETY WARNINGS
WARNING:
High voltage is present in the instrument when power cord is connected. To avoid
potentially dangerous shock, disconnect the power cord before removing the
cover.
WARNING:
This is a safety Class I product provided with a protective earthing ground
incorporated into the power cord. The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor, inside or outside the instrument is likely to make the instrument
dangerous. Intentional interruption is prohibited.
WARNING:
This symbol indicates that to comply with European Union Directive 2002/96/EC for
waste electrical and electronic equipment (WEEE), the Analyzer should be disposed of
separately from standard waste.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-2
ENGLISH
Any operation requiring access to the inside of the equipment, could result in
injury. To avoid potentially dangerous shock, disconnect from power supply before
For continued protection against fire hazard replace fuse with same type and
This symbol, on the instrument indicates that the user should refer to the
This is a safety Class I product. It must be wired to a mains supply with a protective
earthing ground incorporated into the power cord. Any interruption of the
protective conductor, inside or outside the equipment, is likely to make the
instrument dangerous. Intentional interruption is prohibited.
If this instrument is used in a manner not specified by Sievers Instruments Inc. USA,
the protection provided by the instrument may be impaired.
Disposal of RAM Card Lithium batteries must follow the local environmental
Chaque opération à l’intérieur de l’appareil, peut causer du préjudice. Afin
d’éviter un shock qui pourrait être dangereux, disconnectez l’appareil du
Afin de protéger l’appareil continuellement contre l’incendie, échangez le
Le symbol, indique que l’utilisateur doit consulter le manuel
Ceci est un produit de Classe de sécurité I. L’instrument doit être branché
sur l’alimentation secteur par un fil de secteur prévu d’une prise de masse.
Chaque interruption du conducteur protégeant, à l’intérieur ou á l’extérieur
de l’appareil peut rendre l’instrument dangereux. Interruption
Si l’instrument n’est pas utilisé suivant les instructions de Sievers
Instruments, Inc., USA, les dispositions de sécurité de l’appareil ne sont plus
Les batteries RAM Card Lithium doivent être déposés suivant les régulations
manual for operating instructions.
fusible par un fusible du même type et valeur.
WARNING:
opening the equipment.
WARNING:
rating.
WARNING:
WARNING:
WARNING:
WARNING:
regulations.
FRANÇAIS
ATTENTION:
réseau avant de l’ouvrir.
intentionnelle est interdite.
d’environnement locales.
ATTENTION:
ATTENTION:
d’instructions.
ATTENTION:
ATTENTION:
valables.
ATTENTION:
ESPAÑOL
ATENCION:
Cualquier operación que requiera acceso al interior del equipo, puede causar
una lesión. Para evitar peligros potenciales, desconectarlo de la alimentación a
Para protección contínua contra el peligro de fuego, sustituir el fusible por uno
Este símbolo, en el instrumento indica que el usuario debería referirse al
Esto es un producto con clase I de seguridad. Debe conectarse a una red que
disponga de tierra protectora en el cable de red. Cualquier interrupción del
conductor protector, dentro o fuera del equipo, puede ser peligroso. Se prohibe
Si este instrumento se usa de una forma no especificada por Sievers
Instruments, Inc., USA, puede desactivarse la protección suministrada por el
Las pilas de litio de la RAM Card deshechadas deben seguir las regulaciones
red antes de abrir el equipo.
del mismo tipo y características.
manual para instrucciones de funcionamiento.
la interrupción intencionada.
medioambientales locales.!
ATENCION:
ATENCION:
ATENCION:
ATENCION:
instrumento.
ATENCION:
DEUTSCH
Vor dem Öffnen des Gerätes Netzstecker ziehen!
Für kontinuierlichen Schutz gegen Brandgefahr dürfen bei
Sicherungswechsel nur Sicherungen der gleichen Stärke verwendet
Dieses, auf dem Gerät weist darauf hin, dab der Anwender zuerst
das entsprechende Kapitel in der Bedienungsanleitung lesen sollte.
Dies ist ein Gerät der Sicherheitsklasse I und darf nur mit einem
Netzkabel mit Schutzleiter betrieben werden. Jede Unterbrechung des
Schutzleiters auβerhalb oder innerhalb des Gerätes kann das Gerät
elektrisch gefährlich machen. Absichtliches Unterbrechen des
Schutzleiters ist ausdrücklich verboten.
Wenn das Gerät nicht wie durch die Firma Sievers Instruments, Inc.,
USA, vorgeschrieben und im Handbuch beschrieben betrieben wird,
können die im Gerät eingebauten Schutzvorrichtungen beeinträchtigt
Die Entsorgung der Lithium-Batterie in der RAM-Karte darf nur nach den
geltenden Umweltschutzregeln erfolgen.
WARNHINWEIS:
WARNHINWEIS:
werden!
WARNHINWEIS:
WARNHINWEIS:
WARNHINWEIS:
werden.
WARNHINWEIS:
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-3
ITALIANO
ATTENZIONE:
Qualsiasi intervento debba essere effettuato sullo strumento può
essere potenzialmente pericoloso a causa della corrente elettrica.
Il cavo di alimentazione deve essere staccato dallo strumento prima
Per la protezione da rischi da incendio in seguito a corto circuito,
sostituire I fusibili di protezione con quelli dello stesso tipo e
Il simbolo, sullo strumento avverte l’utilizzatore di consultare il
Manuale di Istruzioni alla sezione specifica.
Questo strumento è conforme alle specifiche per I prodotti in Classe I
- Il cavo di alimentazione dalla rete deve essere munito di “terra”.
Qualsiasi interruzione del cavo di terra all’interno ed all’esterno dello
strumento potrebbe risultare pericolòsa. Son o proibite interruzioni
Se questo strumento viene utilizz ato in maniera non conforme alle
specifiche di Sievers Instruments, Inc. USA, le protezioni di cui esso è
Le batterie al Litio sulla RAM CARD, quando sono esaurite, devono
essere gettate secondo le regolamentazioni vigenti localmente.
della sua apertura.
ATTENZIONE:
caratteristiche.
ATTENZIONE:
ATTENZIONE:
intenzionali.
ATTENZIONE:
dotato potrebbero essere alterate.
ATTENZIONE:
DUTCH
Iedere handeling binnenin het toestel kan beschadiging
veroorzaken. Om iedere mogelijk gevaarlijke shock te vermijden
moet de aansluiting met het net verbroken worden, vóór het
Voor een continue bescherming tegen brandgevaar, vervang de
zekering door een zekering van hetzelfde type en waarde.
Het symbool, geeft aan dat de gebruiker de instructies in de
handleiding moet raadplegen.
Dit is een produkt van veiligheidsklasse I. Het toestel moet
aangesloten worden op het net via een geaard netsnoer. Bij
onderbreking van de beschermende geleider, aan de binnenzijde
of aan de buitenzijde van het toestel, kan gebruik het toestel
gevaarlijk maken. Opzettelijke onderbreking is verboden.
Indien het toestel niet gebruikt wordt volgens de richtlijnen van
Sievers Instruments, Inc., USA gelden de veiligheidsvoorzieningen
RAM kaart Lithium batterijen dienen volgens de lokale
afvalwetgeving verwijderd te worden.
OPGELET:
openen van het toestel.
OPGELET:
OPGELET:
OPGELET:
OPGELET:
niet meer.
OPGELET:
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-4
The Model 280i NOA is used for measurement of NO in exhaled breath and measurement of nitrite, nitrate/nitrite and nitrosothiols in biological fluids, cell culture media, and other liquid samples. A schematic of the 280i NOA is shown in Figure 1-1 and consists of the following major components:
Sampling Inlet Systems
Sample Flow Control Device
Ozone Flow Control Module
Ozone Generator
Chemiluminescence Reaction Chamber and Optical Filter
Photomultiplier Tube and Cooled Housing
Vacuum Pump and Chemical Trap
Front Panel Display
Four Button Keypad
PMT Amplifier
Power Supply
Analog to Digital Converter
Microprocessor and Output Electronics
Analog, Printer, and RS-232 Outputs
Exhalation Pressure Transducer
Optional Thermal Mass Flowmeter

Sample Inlet Systems

The NOA 280i has a complete range of sample inlet systems for measurement of NO and its reaction products including:
Gas Sampling Kit for measurement of gas­phase NO. The kit
®
includes a Nafion
drier,
0.45 µm particle filter, PVC sampling lines with Luer® adapters and a calibration tee.
Figure 1-1: Schematic of Model 280i Nitric
Oxide Analyzer
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-5
Purge vessel for the measurement of NO, nitrite, nitrate and other reaction products in liquid samples. The purge vessel can also be used for headspace analysis with a gas-tight syringe.
Accurate NO™ Exhaled Breath Kit for on-line measurement of exhaled NO using elevated pressure to close the soft pallet and constant low exhalation flow (30 – 250 mL/s BTPS)
Bag Collection and Sampling Kit for off-line measurement of exhaled NO using collection in Mylar® bags.

Sample Flow Control Device

The vacuum pump continuously draws gas into the analyzer at a constant flow rate. A porous metal frit restrictor sealed in a 1/8" adapter is connected to a Swagelok® bulkhead union at the rear of the NOA. The standard restrictor provides a flow rate of ~200 mL/min and restrictors for other flow rates are available from Sievers Instruments.

Ozone Flow Control Module

The connection for the gas supply for the ozone generator (oxygen or 95% O2/ 5% CO2) is made using Teflon tubing and a Swagelok bulkhead connector. The gas must be a regulated supply from an external cylinder, lecture bottle or house oxygen. The flow rate of gas into the ozone generator (~30 mL/min) is controlled using a regulator and small diameter tubing restrictors. The regulator and bulkhead connectors are located on the back of the NOA. The regulator is adjusted to 6 psi pressure, which is measured by a pressure transducer, and monitored on the front panel display.

Ozone Generator

An electrostatic ozone generator and high voltage transformer are used to generate ozone at a concentration of ~2% by volume from oxygen. This large excess of ozone is sufficient for measurement of NO up to 500 ppm.

Chemiluminescent Reaction Chamber and Optical Filter

GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-6
Nitric oxide and ozone are mixed in a small volume (~20 mL) reaction cell. This small volume permits measurement of low concentrations of NO at low flow rates and produces sharp peaks for analysis of liquid samples. For maximum sensitivity, the reaction cell is operated at low pressure (typically 4-7 torr). A few other chemicals such as sulfur-containing compounds, undergo a chemiluminescent reaction with ozone but emit light at shorter wavelengths. To minimize interference from these species, an optical filter that transmits only red wavelengths (>600 nm) is installed between the reaction cell and the photomultiplier tube.

Photomultiplier Tube and Cooled Housing

The light from the chemiluminescent reaction of NO with O3 is measured using a red-sensitive photomultiplier tube. For maximum sensitivity, the PMT is cooled to ­12 °C using a thermoelectric cooled housing. The cooler is operated continuously whenever the main power switch is on. The temperature of the cooled housing is measured using a K-type thermocouple and monitored on the front panel display.

Vacuum Pump and Ozone Destruction Trap.

A vacuum pump is used to draw the sample into the NOA and maintain the reaction cell at low pressure. The exhaust from the reaction cell exits the analyzer at the rear of the instrument using a metal tube connected to Tygon tubing. Ozone in the exhaust is removed using a chemical trap containing Hopcalite™. This material reacts with ozone, removing it from the exhaust before the gas reaches the vacuum pump. Since the Hopcalite is consumed, the chemical trap must be periodically replaced. The oil used in the vacuum pump is a synthetic motor oil (Mobil 1™ weight 10W-30) which provides better protection than conventional pump oil. For long pump lifetime, the oil must be changed at regular intervals. The exhaust from the pump contains some oil mist, which can be removed using a charcoal trap. The microprocessor keeps track of the trap and oil lifetimes, and notifies the user when it is time to replace the traps or change the oil.

Electronics

There are 7 circuit boards in the NOA:
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-7
PMT amplifier
Analog to digital converter (ADC) board
Microprocessor and Outputs board
Power supply board
24 V power supply
Front panel display
Keypad.
The PMT amplifier processes the signal from the PMT. To provide both high sensitivity and wide dynamic range, the amplifier has two gain ranges. High Gain is used for measurements requiring high sensitivity (liquid samples and exhaled breath). The amplifier in high gain has a linear response up to ~1 ppm of NO gas or ~400 picomoles of NO
-
or NO
2
-
for liquid samples. The low gain decreases the
3
sensitivity of the amplifier, permitting measurement of up to ~500 ppm of NO gas or ~200 nanomoles of NO
-
or NO
2
-
for liquid samples. The analog output signal (mV
3
only) is obtained from the amplifier. A switch on the amplifier sets the full-scale voltage. When the switch is in the down position the output range is 0–1V. When the switch is in th eup position, the output range is 0–10V.
The amplifier is also connected to the ADC board which also monitors three pressure transducers (cell, supply and exhalation), the cooler thermocouple and the thermal mass flowmeter. The microprocessor and firmware calculate gas concentration, monitor the performance of the analyzer, keep track of maintenance items, and control the output of data. The power supply board and 24V power supply provides the high voltage for the PMT, power to the PMT cooler and the DC power for the electronics. The front panel display and keypad are used for display of the data and the operation of the analyzer.

Analog, Printer and RS-232 Outputs

In addition to displaying data on the front panel, data can also be: sent to a recorder or integrator using the analog output, sent to a computer using the RS­232 output, and printed using the parallel printer port. The analog output can be set to 0-1 V or 0-10 V full-scale. The printer output shows the minimum, maximum and average value for a selected print interval ranging from 5 seconds to 10 minutes. The RS-232 output provides data at sampling rates from 32 samples per second to 6 samples per minute.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-8

Exhalation Pressure Transducer

An exhalation pressure transducer is present in the analyzer for use with the Accurate NO breath kit and for detection of exhalations during breath-by-breath measurements.

Thermal Mass Flow Meter

On-line measurement of exhalation flow rates can be performed using a optional flow meter. The signal from the flow meter is included in the RS-232 output.
This completes the introduction to the NOA and its components:
Chapter 2 lists the specifications of the analyzer.
Chapter 3 has an overview of the firmware and controls.
Chapter 4 contains the basic installation procedures.
Chapter 5 contains the installation and setup for gas-phase measurements.
Chapter 6 describes how to calibrate the NOA for gas-phase measurements.
Chapter 7 describes on-line measurement of NO in exhaled breath.
Chapter 8 describes off-line measurement of exhaled NO.
Chapter 9 describes breath-by-breath and chamber sampling for exhaled NO
measurements.
Chapter 10 describes measurement of nasal nitric oxide.
Chapter 11 contains the installation and setup liquid measurement.
Chapter 12 describes the setup and measurement of nitrite in liquid
samples.
Chapter 13 describes the setup and measurement of nitrate in liquid
samples
Chapter 14 describes the measurement of nitrosothiols and other reaction
products in liquid samples.
Chapter 15 describes maintenance of the NOA.
Chapter 16 lists troubleshooting procedures and error and warning
messages.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
1-9

2. SPECIFICATIONS

Sensitivity
.................................................... Gas < 1 ppb
.................................................... Liquid ~1 picomole
Range
.................................................... Gas < 1 – 500,000 ppb
.................................................... Liquids nanomolar to millimolar
Response Time
.................................................... Electronics 67 msecs to 90% full
scale
.................................................... Lagtime 1 second
Repeatability
.................................................... Gas ± 5%
.................................................... Liquid ± 5 10%
Sample Size
.................................................... Gas 10 300 mL/min
.................................................... Liquid 0.001 – 5 mL
Display ........................................... Back-lit LCD screen
.................................................... ppb/ppm or mV
Outputs
.................................................... Analog 0 1V, 0 – 10 V
.................................................... Digital RS-232 (9600-38.4K baud)
.................................................... Printer parallel port
Data Sampling Rate............................ 0.002 - 32 samples/second
GE Analytical Instruments ©2006 DLM 14291 Rev. A
2-1
Power Requirements................................120 V, 60 Hz (6A)
100 V, 50 or 60 Hz (7A) 230 V, 50 Hz (3A) NOA
.........................................................Height 16 in. (41 cm)
.........................................................Width 6.2 in. (16 cm)
.........................................................Length 20 in. (51
cm) ....................................................
.........................................................Weight 35 lbs. (16
kg)
Vacuum Pump with installed trap
.........................................................Height 14.5 in. (37 cm)
.........................................................Width 7.5 in. (19 cm)
.........................................................Length 19 in. (48
cm)
.........................................................Weight 47 lbs. (21.5
kg)
Operating Environment
Ambient Temperature ....................................32°F to 86°F (0°C to 30°C)
Relative Humidity.......................................... 0% to 90%
Fuse Requirements
Main Fuse ...................................100 VAC model: T, 10A, 250V
120 VAC model: T, 5A, 250V
230 VAC model: T, 5A, 250V
Ozone Fuse..................................100VAC model: T, 200mA, 250V
120 VAC model: T, 200mA, 250V 230 VAC model: T, 100mA, 250V
GE Analytical Instruments ©2006 DLM 14291 Rev. A
2-2

3. MENUS AND CONTROL OVERVIEW

Operation of the NOA 280i is performed using the four front panel buttons (UP Arrow, DOWN Arrow, ENTER, and CLEAR) to run the menu-based firmware. Use the UP or DOWN Arrow buttons to scroll through the menu options and select values. The ENTER Button is used to select menu options and to save setpoints. The CLEAR Button is used to exit menus and clear entries. The CLEAR Button is also used to display the Status Menu from the Main Menu, select the Main Menu from the Measurement Menu or to return to the Measurement Menu from the Main Menu. A cursor is used for selection of the menu options and the location
M A I N M E N U
A N A L Y S I S C A L I B R A T I O N
C O N T R O L M E S S A G E S
0 4 J U L 0 0 0 8 : 1 7 : 3 1
of the cursor is indicated by a highlighted menu option.

Main Menu

There are five options in the Main Menu: Analysis, Control,, Calibration,, Messages, and Maintenance,. A title field is located at the top of the menu, and the date and time are displayed in the message line at the bottom of the display. Select an option by using the Arrow buttons to highlight the desired
M A I N T E N A N C E
Figure 3-1: Main Menu
option, and press the ENTER button.

Status Screen

Press the CLEAR button from the Main Menu to display the Status Screen. This screen shows the current values for the PMT cooler temperature, reaction cell
GE Analytical Instruments ©2006 DLM 14291 Rev. A
3-1
ff)
S T A T U S C O O L E R T E M P – 1 2 . 0 ° C C E L L P 7 6 0 . 0 T o r r
S U P P L Y P 0 . 0 P S I G P M T S I G N A L 0
E N T E R / C L E A R c o n t i n u e
Figure 3-2: Status Screen (PMT O
pressure, oxygen supply to the ozone generator, and the PMT signal, (counts at analog to digital converter). Before starting the NOA, the status screen should be checked to confirm that the cooler temperature, cell and supply pressures are within the specifications required for the start-up tests.
Pressing either the ENTER or CLEAR button will return to the Main Menu.

Analysis

The Analysis option is used to start and stop the NOA. The NOA has three modes
.
of operation: Start, Stand-by and Stop. From the Main Menu, pressing the ENTER button with the Analysis option highlighted will display the Analysis Menu.
Selecting the Start option will:
Switch to the Startup Screen.
Check the cell pressure to see if it is above 300 torr, then turn on the
pump.
A N A L Y S I S
S T A R T
S T A N D - B Y
Figure 3-3: Analysis Menu with NOA in Stop Mode.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
3-2
V A C U U M P U M P P A S S
C O O L E R T E M P P A S S C E L L P R E S S P A S S
S U P P L Y P R E S S T E S T I N G P M T S I G N A L
E N T = S T A T U S C L R = S T O P
Figure 3-4: Start-up Screen.
Check the PMT cooler temperature to see if it is –12 ± 2 °C.
Wait until the reaction cell pressure is < 100 torr.
Check the ozone supply pressure is >4 psig (6 psig recommended).
Turn on the PMT and record an ozone-off baseline signal.
Turn on the ozone generator, and wait for an increase in the PMT signal
due to the background chemiluminescence from ozone.
Display the PMT signal in the Measurement Display.
If any of the above conditions are not met, the start-up screen will FAILED for that test. Pressing ENTER from the Start-up Screen will display the Status Screen to aid in troubleshooting failed tests. Pressing CLEAR will return to the Analysis Menu.

Measurement Menu

Once the start-up testing is completed, the Measurement Menu is displayed.
Figure 3-5:Measurement Menu (Nitric Oxide Mode)
GE Analytical Instruments ©2006 DLM 14291 Rev. A
3-3
D A T A O U T P U T
O U T P U T E N A B L E D
E N T E R t o g g l e o u t p u t
Figure 3-6: Data Output Menu.
This menu shows that NOA’s mode (Nitric Oxide or Exhalation), the PMT amplifier’s setting (HI or LO sensitivity) and the signal from the PMT (mV or gas concentration). Two shortcuts are available from the Measurement Menu; DATA and WARN. When DATA is highlighted, pressing the ENTER button moves to the Data Output Menu. The data output (Com port and printer) are enabled at start-up, but can be disabled to pause data collection by pressing the ENTER button. The outputs are re-enabled by pressing the ENTER button. Pressing CLEAR returns to the Measurement Menu.
The firmware keeps track of usage and when maintenance is required, a WARN shortcut is displayed in the Measurement Menu. The UP or DOWN Arrow buttons can be used to scroll between DATA and WARN. With the WARN shortcut highlighted, pressing the ENTER button will display the Warning Menu with a list of the current warnings (see Chapter 14 for information on maintenance and warnings).
In the Nitric Oxide mode the current values for the cooler temperature, cell pressure and ozone supply pressure are displayed at the bottom of the Measurement Menu. In the Exhalation mode, a bar graph of the exhalation pressure is displayed.
From the Measurement Menu, pressing the CLEAR button will return to the Main Menu and from the Main Menu, pressing CLEAR will return to the Measurement Menu.
GE Analytical Instruments ©2006 DLM 14291 Rev. A
3-4
Loading...
+ 181 hidden pages