GE ML_CL_INW SERVICE MANUAL_SM_2021423-503_H A7 Anesthesia System Service Manual

Page 1
GE Healthcare
Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual
Version 6.5.6
Instructions
2021423-503 Revision H
Page 2
Listed below are GE Medical Systems Information Technologies trademarks used in this document.
MAC-LAB
®
, CARDIOLAB®, CENTRICITY®, TRAM®, TRAMNET®, and TRAM-RAC® are trademarks of GE Medical Systems Information Technologies, a division of General Electric Company going to market as GE Healthcare.
All other trademarks contained herein are the property of their respective owners.
© GE Medical Systems Information Technologies, 2009 - 2014. All rights reserved.
T-2 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
25 August 2014
Page 3
CE Marking Information
Compliance
The Mac-Lab/CardioLab/ComboLab system bears CE mark CE-0459 indicating its conformity with the provisions of the Council Directive 93/ 42/EEC concerning medical devices and fulfills the essential requirements of Annex I of this directive.
For devices manufactured in the United States, the CE mark is applied under the authority of Notified Body GMED (0459).
The country of manufacture and appropriate Notified Body can be found on the equipment labeling.
The product complies with the requirements of standard IEC 60601-1-2.
The safety and effectiveness of this device has been verified against previously distributed devices. Although all standards applicable to presently marketed devices may not be appropriate for prior devices (that is, electromagnetic compatibility standards), this device will not impair the safe and effective use of those previously distributed devices. Refer to the user’s information.
CE Marking Information
Recommendations
Refer to the Mac-Lab/CardioLab Service Manual for Compliance Information.
Users should be aware of known RF sources, such as radio or TV stations and hand-held or mobile two-way radios, and consider them when installing a medical device or system.
Be aware that adding accessories or components, or modifying the medical device or system may degrade the EMI performance. Consult with qualified personnel regarding changes to the system configuration.
The Mac-Lab/CardioLab System, if operating under the conditions defined in the EMC Standard IEC 60601-1-2 (Radiated Immunity 3 V/m) and at field strengths above 3 volts per meter may cause waveform distortions at radio frequency (RF) electromagnetic interference (EMI).
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING SERVICE MANUAL.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual CE-1
Page 4
CE Marking Information
The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers and cables sold by the manufacturer of the Mac-Lab/CardioLab System as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the Mac-Lab/CardioLab System.
The Mac-Lab/CardioLab System should not be used adjacent to or stacked with other equipment and that if adjacent to or stacked use is necessary, the Mac-Lab/CardioLab System should be observed to verify normal operation in the configuration in which it will be used.
Review the AAMI Committee Technical Information Report (TIR) 18, “Guidance on Electromagnetic Compatibility of Medical Devices for Clinical/Biomedical Engineers”. This guidance document provides a means to evaluate and manage the EMI environment in the hospital.
The following actions can be taken to reduce the risk of medical device EMI and achieve EMC:
Assess the EMC environment of the healthcare facility (for example,
identify radio transmitters in around the facility) and identify areas where critical medical devices are used (for example, ER, ICU, CCU, NICU).
Increase the distance between sources of EMI and susceptible
devices.
Remove the devices that are highly susceptible to EMI.
Lower power transmitted from electrical and electronic equipment
(EMI sources) under hospital control (that is, paging systems).
Label devices susceptible to EMI.
Educate healthcare facility staff (nurses and doctors) to be aware of,
and to recognize, potential EMI related problems.
Exception
ECG Parameter
IEC 60601-1-2 clause 36.202.1 - Immunity: Radiated Immunity –
The level of compliance is less than 1V/m
If operating under the conditions defined in EMC Standard EN 60601-1­2 (Radiated Immunity 3V/m), field strengths less than 1V/m may cause waveform distortions and erroneous numeric data at various electromagnetic interference (EMI) frequencies.
CE-2 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 5
Recommendations
CE Marking Information
Review the AAMI EMC Committee technical information report (TIR-18) titled Guidance on electromagnetic compatibility of medical devices for clinical / biomedical engineers - Part 1: Radiated radio-frequency electromagnetic energy. This TIR provides a means to evaluate and manage the EMI environment in the hospital.
The following actions can be taken:
managing (increasing) distance between sources of EMI and
susceptible devices
managing (removing) devices that are highly susceptible to EMI
lower power from internal EMI sources under hospital control (i.e.
paging systems)
labeling devices susceptible to EMI
educating staff (nurses and doctors) to be aware of, and to recognize,
potential EMI related problems
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual CE-3
Page 6
CE Marking Information
CE-4 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 7
Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Manual Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Product References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Illustrations and Names . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Chapter Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Safety Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
ITE Conformance Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-12
Types of Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Mac-Lab System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
CardioLab EP System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
ComboLab System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
CardioLab II Plus Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
Security Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Non-Validated Software Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-15
Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-15
Validated Third Party Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Service Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Service Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-19
Product Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Required Tools and Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-22
Power Input Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-22
Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23
Electromagnetic Compatibility and Interference . . . . . . . . . . . . . . . . . . . . . . . . . .1-23
Exception . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-23
EU Battery Directive 2006/66/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-24
Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-25
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-25
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-26
Compliance Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-29
Mac-Lab/CardioLab System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-29
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual i
Page 8
2 Equipment Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Typical System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Control Room Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Procedure Room Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
IT Control Room Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Network Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
System Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Acquisition Workstation System Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
GE Client Workstation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
Mobile Workstand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-36
Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-37
TRAMNet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-37
CardioLab II Plus Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-39
CARTO™ System (Patient Interface Unit) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-45
Analog Output Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-47
Network Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-48
Cabling and Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-53
HP z600 Computer I/O Backplane Connections . . . . . . . . . . . . . . . . . . . . . . . . .2-53
HP xw8400 Computer I/O Backplane Connections . . . . . . . . . . . . . . . . . . . . . . .2-54
Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-55
HP z600 Video Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-56
HP xw8400 Video Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-57
Communications Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-58
TRAM/Amplifier Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-59
Amplifier/Stimulator Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-60
Audio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-64
CARTO™XP System Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-65
CARTO™ 3 System Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-66
Review Workstations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-69
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-72
Operating System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-72
Database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-72
Application Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-72
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Mac-Lab/CardioLab Service Performance Testing . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
FRU Checkout Procedure Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Tools Recommended . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Visual Inspection/Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Electrical Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Functional Checkout Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Mac-Lab SCOP (Standard Checkout Procedure) . . . . . . . . . . . . . . . . . . . . . . . . .3-8
CardioLab SCOP (Standard Check Out Procedure) . . . . . . . . . . . . . . . . . . . . . .3-10
TRAM Module Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
ii Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 9
CardioImage/Image Capture Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
IEB Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Printer Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
CARTO Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Audible Indicators Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
RF Ablation Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Stimulator Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
DASH / Solar Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Remote Operators Terminal Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Analog Out Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Centricity Cardiology INW Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
Remote Connection Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16
Planned Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Mobile Workstand Preventative Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
Adjusting Tilt and Tilt Tension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
Adjusting Swivel or Pivot Tension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Clean Fiber Optic Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Initial Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Matrox Parhelia Fan Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Replace Main Video Board with Matrox M9140 Quad Video Board . . . . . . . . . . . 4-5
Prevent Failure of the Matrox Odyssey Driver Service . . . . . . . . . . . . . . . . . . . .4-12
CardioLab Amplifier Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Noise on Intracardiac Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Real-Time Testing During a Case. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
CardioImage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Normal Operation and Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-22
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Switching Ablation Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Windows XP/2003 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-30
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
GEMS_TaskMonitor Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Networking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-33
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual iii
Page 10
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Centricity Cardiology INW Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Replication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-37
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-37
Cleanup Database Corruption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
INNOVA CENTRAL/Touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Check Connection Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-39
Determine IP Addresses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-39
CARTO XP/CARTO 3 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-40
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-40
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-40
EtCO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
EtCO2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-43
Capnostat Module Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
Capnostat Module Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-56
Calibrate the CapnoFlex LF Module (Sensor + Pump) . . . . . . . . . . . . . . . . . . . .4-58
Mac-Lab Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-59
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-61
CVIS to Mac-Lab/CardioLab Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-65
Problem Scenarios . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-67
Test DICOM Connectivity (CVIS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-68
Test DICOM Connectivity (X-Ray) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-70
HL7 Export to CVIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-71
InSite ExC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-72
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-72
GEMS Log Collector Task. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-72
List of Log Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-75
List of SNMP Faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-75
Application Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-77
Prerequisite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-77
Option Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-78
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-78
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-78
Event Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-79
Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-80
Application Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-80
Install/Update Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-80
Install Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-81
Upgrade Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-82
iv Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 11
Maintenance Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-83
Replication Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-84
Device Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-85
Log File Locations for Mac-Lab/CardioLab System . . . . . . . . . . . . . . . . . . . . . . . 4-86
5 Recovery Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Acquisition System Disaster Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
High-Level Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Log On to the Acquisition System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Save Acquisition System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
INW Networked Acquisition Systems Only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Restore System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Set Default Baseline Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31
Install/Enable Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
Configure the Acquisition System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
Configure the Acquisition System on the INW Network . . . . . . . . . . . . . . . . . . . 5-40
Initialize Applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-48
Confirm File Paths and Network Storage Settings . . . . . . . . . . . . . . . . . . . . . . . . 5-51
Additional Installation for Purchased Options . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-51
Reconfigure Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-52
Reconfigure Mapped Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-52
Configure Data Export . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-52
Reload CVIS Client . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-54
Update Operator Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-54
Review System Disaster Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-55
High-Level Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-55
Log On to the Review System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-56
Save Review System Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-56
Gather Information from INW Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-60
Configure New Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-61
Configure the Review System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-63
Configure the Review System on the INW Network . . . . . . . . . . . . . . . . . . . . . . 5-75
Initialize Applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-81
Confirm Windows XP is Activated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-84
Confirm File Paths and Network Storage Settings . . . . . . . . . . . . . . . . . . . . . . . . 5-84
Additional Installation for Purchased Options . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-84
Reconfigure Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-84
Reconfigure Mapped Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-85
Configure Data Export . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-85
Reload CVIS Client . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-85
Update Operator Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-86
Perform Necessary Software Updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-86
System Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-87
Review Workstation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-87
INW Server Disaster Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-88
High-Level Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-88
Record Server Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-90
Reimaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-91
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual v
Page 12
G4/G5/G6 Server Setup and Data Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-98
Reconfigure Centricity Cardiology INW Network . . . . . . . . . . . . . . . . . . . . . . . 5-119
Post Configuration Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-128
Cardiovascular Information System (CVIS) Configuration . . . . . . . . . . . . . . . . 5-148
Perform On Demand Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-155
Verify Backups Ran Successfully . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-156
Update Operator Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-157
System Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-158
Verify Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-158
6 Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
HP z600 Computer with ADA Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
HP z600 Computer with DVD-RAM Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
HP xw8400 Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
GE Client Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
DASH Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Solar Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Centricity Cardiology INW G6 Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Centricity Cardiology INW G5 Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Centricity Cardiology INW G4 Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Integrated Electronics Box (IEB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Uninterrupted Power Source (UPS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Monitors and Printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
TRAMNet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
MicroPace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
RMOT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
Speakers and Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
vi Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 13
A Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Electrical and IT Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Hospital Staff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Medical Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Miscellaneous Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-18
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual vii
Page 14
viii Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 15

1 Introduction

2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-1
Page 16
Introduction

Manual Information

Purpose

This manual contains the instructions necessary to operate and service the Mac-Lab/CardioLab system safely in accordance with its function and intended use. These instructions include but are not limited to:
Describing fundamental controls and indicators.
Operating procedures.
Connecting and disconnecting detachable parts and accessories.
Troubleshooting and service procedures.

Intended Audience

This manual is intended for the person who maintains and troubleshoots this equipment.

Revision History

Revision Date Comment
A 6 November 2009 Initial release of the manual.
B 21 April 2010 Updated with serviceability
improvements.
C 1 August 2011 Updated to address new hardware.
D 8 November 2011 Added power cords.
E 28 February 2012 Updated Hard Drive, Memory and RMOT
parts.
F 7 May 2013 Updated to include installation and
troubleshooting procedures for the Matrox Quad video board.
G 18 November
2013
H 24 March 2014 Updated to include information for EIZO
Updated to include information for new speakers.
monitors.
1-2 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 17

Conventions

Introduction
The following conventions are used throughout this manual.
Bold Indicates keys on the keyboard, text to be entered, or hardware items
such as buttons or switches on the equipment.
Italics Indicates software terms that identify menu items, buttons, or
options in various windows.
[Key1] + [Key2] Indicates a keyboard operation. A (+) sign between the names of two
keys indicates that you must press and hold the first key while pressing the second key once.
For example, “Press Ctrl + Esc” means to press and hold down the
Ctrl key while pressing the Esc key.
<Space> Indicates you must press the spacebar on the keyboard when typing a
precise text string with one or more spaces. The point where the spacebar must be pressed is indicated as: <Space>. The purpose of the < > brackets is to ensure you press the spacebar when required.

Product References

Related Documents

Enter Indicates you must press the “Enter” or “Return” key on the keyboard.
Do not type “enter”.
The name of the product described in this manual is Mac-Lab/CardioLab/ Centricity Cardiology INW System.
Part Number Document
2021423-015 Mac-Lab Operator’s Manual
2021423-016 CardioLab Operator’s Manual
2021423-017 Centricity Cardiology INW Operator’s Manual
2021423-021 Mac-Lab/CardioLab/Centricity Cardiology INW Pre-Installation Manual
2021423-022 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual
(v6.5.4 and earlier)
2027332-103 Mac-Lab/CardioLab Security Guide
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-3
Page 18
Introduction

Illustrations and Names

All illustrations in this manual are provided as examples only. They may not necessarily reflect your system’s configuration or the data on your system.
In this manual, all names appearing in examples and illustrations are fictitious. The use of any real person’s name is purely coincidental.

Chapter Contents

Chapter 1, Introduction: This chapter describes the Service
Chapter 2, Equipment Overview: This chapter describes the
Chapter 3, Maintenance: This chapter contains preventive
Chapter 4, Troubleshooting: This chapter contains
Chapter 5, Recovery Procedures: This chapter details procedures
Chapter 6, Parts List: This chapter contains a list of Field
Appendix A, Abbreviations: This appendix lists all abbreviations
manual and chapter contents. Also included are general safety requirements, service requirements, equipment symbols, and serial number identification information.
equipment and its technical characteristics, connector locations, and preparation for use.
maintenance information and cleaning guidelines.
troubleshooting instructions and functional test directions.
to restore a workstation or server in the field after failure.
Replaceable parts.
used with the system.
1-4 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 19

Safety Information

Safety Messages

Definitions
Introduction
DANGER
— Safety messages which indicate an imminently hazardous situation which, if not avoided, WILL result in death or serious injury.
WARNING
— Safety messages which indicate a potentially hazardous situation which, if not avoided, COULD result in death or serious injury.
Applicable Messages
CAUTION
— Safety messages which indicate a potentially hazardous situation which, if not avoided may result in minor or moderate injury, loss of data or system failure.
NOTE
Messages which provide additional user information.
DANGER
EXPLOSION HAZARD — Using this equipment in the presences of flammable gasses (including anesthetics and oxygen) may cause an explosion or fire.
Always use this equipment in a well ventilated area away from the presence of potentially flammable gasses
WARNING
SYSTEM STABILITY — All systems should be installed and maintained according to installation instructions and manufacturers specifications. Any alterations may affect the stability of the system.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-5
Page 20
Introduction
WARNING
PATIENT MONITORING — The system is intended to be used as a recording system for catheterization, electrophysiology and related specialty laboratories. A defibrillator or ECG monitor should be attached for patients in need of uninterrupted ECG display. An additional means to display SPO
patients in need of uninterrupted SPO
should be available for
2
display. A
2
temporary pacemaker needs to be available for patients in need of uninterrupted delivery of pacing.
WARNING
SUPERVISED USE — This device is intended for use under the direct supervision of a licensed health care practitioner.
WARNING
ACCESSORIES — All system components sold by GE Healthcare for use with this system are tested to meet IEC 60601-1 when used with the system. All other components used with the system must meet IEC 60601-1.
WARNING
IEC — All equipment not complying with IEC 60601-1 should be placed outside the patient environment.
WARNING
SHOCK HAZARD — DO NOT allow hydraulic fluid from any device to contact any conductive surface. Hydraulic fluid is electrically conductive.
An electrical shock hazard exists between chassis ground and isolated (floating) ground when power is applied. Unplug the unit from the power source before proceeding. To reduce the risk of electric shock or damage to your equipment, do not disable the power cord grounding feature. This equipment is designed for connection to a earth grounded power outlet. The grounding plug is an important safety feature.
1-6 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 21
Introduction
WARNING
SHOCK HAZARD — Keep the conductive parts of lead electrodes and associated parts away from other conducting parts, including earth.
WARNING
SHOCK HAZARD — Total system earth leakage current must not exceed 500 microamperes.
Voltages between the three power outlet connections must be accurate BEFORE measuring power outlet ground-to­neutral loop resistance.
Electrostatic discharge (ESD) can damage electronic components. Be sure you are properly grounded when using or servicing the system.
WARNING
SHOCK HAZARD — To reduce the risk of electric shock or damage to the equipment, do not disable the power cord grounding feature. This equipment is designed for connection to a grounded (earthed) power outlet.
WARNING
SHOCK HAZARD — Use appropriate Lock Out/Tag Out procedures whenever an energy hazard exists, such as disconnecting equipment to perform maintenance.
WARNING
SHOCK HAZARD — Electrical shock hazard between chassis ground and isolated (floating) ground when power is applied. Unplug the unit from the power source before proceeding.
To reduce the risk of electric shock or damage to your equipment, do not disable the power cord grounding feature. This equipment is designed for connection to a grounded (earthed) power outlet. The grounding plug is an important safety feature.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-7
Page 22
Introduction
WARNING
SHOCK HAZARD — High voltages exist in the unit. Use insulated tools. Remove jewelry and use only one hand when possible.
Always disconnect power from the equipment before attempting service.
WARNING
SHOCK HAZARD — Only connect IEC 60601-1 certified medical equipment. If fluid of any kind should leak into the system, discontinue use of the equipment and contact GE Healthcare immediately.
WARNING
SHOCK HAZARD — To reduce the risk of electric shock, DO NOT remove cover or back. This equipment contains no user serviceable parts. Refer servicing to qualified service personnel.
WARNING
SHOCK HAZARD — Proper connections to the patient are critical for the proper operation of the CardioLab II Plus Amplifier and patient safety. Use extreme caution when making patient connections. Do NOT connect non­isolated equipment to the amplifier. Proper connections to the stimulator are critical for the proper operation of the CardioLab II Plus Amplifier and patient safety. Use extreme caution when making stimulator connections. The stimulator must be a IEC 60601-1 certified product Ablation operations should NOT be executed when a stimulator unit is connected.
WARNING
SHOCK HAZARD — Do not allow any foreign materials such as fluids to come in contact with the CardioLab II Plus Amplifier. If fluid of any kind should leak into the CardioLab II Plus Amplifier, discontinue using the amplifier and contact GE Healthcare. DO NOT allow hydraulic fluid from any device to contact any conductive surface. Hydraulic fluid is electrically conductive.
1-8 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 23
Introduction
WARNING
SHOCK HAZARD — Use appropriate Lock Out/Tag Out procedures before performing any service on potentially energized equipment.
WARNING
SHOCK HAZARD — Remote speakers must be powered by a 60601-1 certified power supply.
WARNING
CONDUCTION HAZARD — Do not leave conductive parts of lead electrodes open. This condition could leave the patient or user more susceptible to shock.
WARNING
GROUNDING — Electrostatic discharge (ESD) can damage electronic components. Be sure you are properly grounded before opening any equipment.
WARNING
POWER SUPPLY — The supply line must be fused and capable of delivering 16A.
Route the AC power cable away from moving parts.
Connect only to a properly earth grounded outlet.
WARNING
DEFIBRILLATOR PRECAUTIONS — DO NOT contact the amplifier or patient during defibrillation.
Defibrillation equipment is required in the event that the patient needs to be cardioverted or defibrillated.
WARNING
INTERNAL DEFIBRILLATION — Prior to internal defibrillation, disconnect the intercardiac catheters from the system to prevent damage to the amplifier.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-9
Page 24
Introduction
WARNING
INTERPRETATION HAZARD — A qualified physician must overread all computer-generated measurements and tracings. Computerized interpretation is only significant when used in conjunction with clinical findings.
Signal quality should be checked only by a licensed health care practitioner.
All computer assisted measurement functions must be verified by a licensed health care practitioner.
WARNING
ANNUAL EQUIPMENT MAINTENANCE — The system should be checked yearly to ensure compliance with medical device standards.
WARNING
PINCH POINT HAZARD — Keep hands, hair, jewelry and loose clothing away from moving parts. Otherwise, serious injury could result.
WARNING
REPLACEMENT PARTS — Replace only with the same type and rating of fuse.
WARNING
SOFTWARE — DO NOT add unauthorized software to the system. Doing so may make the software unstable.
WARNING
REMOVABLE MEDIA — Exit the software application properly before turning off the system. Failure to do so may result in loss of ability to store the data properly.
Follow recommended instructions when exiting the application or turning off the system to ensure data integrity.
1-10 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 25
Introduction
CAUTION
ELECTROSTATIC SHOCK — Working on electronic devices while not properly grounded my cause damage to the equipment.
Always use a wrist grounding strap and grounding pad when working on electronic equipment.
CAUTION
SALE OF DEVICE — Federal (USA) law restricts this device to sale by or on the order of a physician.
CAUTION
EQUIPMENT CABLING DAMAGE — Route optical cables thorough conduit in the ceiling or floor to avoid damage to the cables or cable connectors.
CAUTION
EQUIPMENT USE —
DO NOT reuse single use devices.
CAUTION
EMC — No known electromagnetic or other interference between the system exists. However, precautions should be taken to avoid the use of cellular telephones, pagers, or other transmitters.
CAUTION
TRANSDUCER — All air must be removed from the transducer or associated tubing before calibration for accuracy.
Pressure channels should be calibrated to a known pressure value if the pressure value is in question.
CAUTION
REMOTE MONITOR — Always power the system through separate isolation transformers which are independently connected to a wall outlet.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-11
Page 26
Introduction

ITE Conformance Statement

The Mac-Lab/CardioLab/Centricity Cardiology INW Server software
applications are intended to be installed and run on an applicable ITE compliant hardware platforms (minimum Hardware Essential Performance be ITE compliant) which may not meet leakage and/or EMC requirements of patient care area equipment. This means but is not limited to the following (unless the medical electrical standard [EN60601-1-1 and EN60601-1-2] is followed):
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware does not directly deliver energy to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware does not indirectly control energy delivered to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware is not directly connected to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware is not installed or used in the Patient vicinity.
Patient Vicinity summarized as follows:
A space, within a location intended for the examination and
treatment of patients, extending 6 ft (1.83 m) beyond the normal location of the bed, chair, table, treadmill, or other device that supports the patient during examination and treatment and extending vertically to 8 ft 2.4 in. (2.5 m) above the floor.
Areas where health care staff members monitor patients and
perform charting and administrative tasks are not considered to be Patient Vicinity areas.
1-12 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 27

Types of Systems

Mac-Lab System

Introduction
The Mac-Lab System is intended for use in a catheterization and related cardiovascular specialty laboratories under the direct supervision of a licensed healthcare practitioner. It is intended to monitor, calculate and/ or record cardiovascular data from adult and pediatric patients undergoing cardiac catheterization procedures. The data may be manually entered or acquired via interfaced devices. Data includes: ECG, heart rate, pulse oximetry (SpO
temperature, valve gradients and areas, cardiac output, hemodynamic measurements, invasive and noninvasive blood pressure and procedural information and optional intracardiac electrocardiogram (IECG). Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, and generate reports on the data. Additionally, the system may acquire, amplify, display and record data received from other interfaced medical devices typically used during these procedures, such as imaging devices. The Mac-Lab System does not control the delivery of energy, administer drugs, perform any life­supporting or life-sustaining functions, or analyze data acquired during the procedure. The Mac-Lab System does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities.
), respiration rate, EtCO2,
2

CardioLab EP System

The CardioLab EP System is intended for use in an electrophysiological laboratory and related specialty laboratories under the direct supervision of a licensed healthcare practitioner. It is intended to monitor, calculate and/or record electrophysiological data from adult and pediatric patients undergoing electrophysiological studies. Data includes: ECG, pressure, and intracardiac electrocardiogram (IECG) waveforms, heart rate, pulse oximetry (SpO
temperature, invasive and noninvasive blood pressure, and procedural information. Physiological parameters such as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from the signal data, displayed and recorded. The system allows the user to monitor the acquisition of data, review the data, and generate reports on the data. Additionally, the system may acquire, amplify, display and record data received from other interfaced medical devices typically used during these procedures, such as imaging devices and RF generators. The CardioLab EP System does not control the delivery of energy, administer drugs, perform any life-supporting or life-sustaining functions, or analyze data acquired during the procedure. The CardioLab EP System does not transmit alarms or arrhythmias and does not have arrhythmia detection capabilities.
), respiration rate, EtCO2,
2
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-13
Page 28
Introduction

ComboLab System

The ComboLab System is the combination of the both the Mac-Lab and CardioLab EP systems. The ComboLab System is intended for use in either a catheterization laboratory or electrophysiological laboratory and related speciality laboratories under the direct supervision of a licensed healthcare practitioner. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab EP System, although only one may be used at a time.

CardioLab II Plus Amplifier

The CardioLab II Plus Amplifier is intended for use in an electrophysiological catheter laboratory or catheterization laboratory where ECG, direct cardiac, intracardiac and pressure signals need to be recorded from a patient. This device is intended for use under the direct supervision of a licensed health care practitioner.
1-14 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 29

Security Information

Validated configurations consist of product scenario implementations that are tested against GE product failure and sample cases run to simulate a real environment while conducting testing. GE provides service and support based on validated configurations only and bears responsibility for GE products only.

Non-Validated Software Applications

This is to detail the policy for non-validated software applications applied to the following products:
Mac-Lab Hemodynamic Systems
CardioLab Electrophysiology Systems
ComboLab Hemodynamic and Electrophysiology Systems
Centricity Cardiology INW Server
Introduction

Policy

Our policy is that we do not recommend installations of any other third party software applications (that is, not mentioned above) on the products above as they could affect performance, warranty and serviceability. Examples of such might include other Anti-Virus programs, print servers, interface engines, anti-spyware, anti-adware, viruses, malware, and so on. GE does not support the use or installation of these non-validated applications or tools. If such programs are installed the customer bears the risk of the product not functioning properly and any risks that are introduced for patient safety. GE will not guarantee support for an environment that is different than a validated implementation. In cases where this is found and technical assistance is requested, GE will provide support, by first eliminating the third party software on the product and bringing it back to a known and validated configuration. If it is found that the third party software is the cause of product failure, any costs associated with this will be the responsibility of the hospital.
At GE Healthcare we understand the uniqueness of our customers and pride ourselves in the personalized solutions we provide to help you in your quest to care for patients. We hope this section will help you to better understand our policy on non-validated software applications and we hope you can work with us as we proceed in further developing our products for your future benefit.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-15
Page 30
Introduction

Validated Third Party Software

Currently validated and allowed third party software such as Enterprise Back-up, Archival, Anti-virus and other applications to run on the same computer are:
Enterprise Backup Software
IBM Tivoli Storage Manager v6.1
Symantec Backup Exec for Windows Server v12.5
CA ARCserv Backup for Windows r12.5
EMC Networker v7.5 + SP1
Validated Anti-Virus Software
Mac-Lab/CardioLab/ComboLab IT (Version 6.5.6)
The Mac-Lab/CardioLab 6.5.6 system has been validated to run with the software listed below. Except where noted, the software below supports English, French, German, Italian, Spanish, Swedish, Chinese, and Japanese:
Supported Anti-Virus Software Supported Anti-Virus Software Version
CA eTrust Antivirus r8.1
McAfee® VirusScan® (with McAfee ePolicy Orchestrator v4.0, Patch 5)
Norton Antivirus™ 2009
Symantec EndPoint Protection 11.0.5 (English, French, German, Italian,
Trend Micro OfficeScan Client/Server Edition
The Antivirus Server software can be installed at a hospital domain level and the Centricity Cardiology INW server can be a client for the Antivirus software.
8.7i, Patch 2
Spanish, Chinese)
11.0.4 MR4 + MP2 (Japanese)
10.0
1-16 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 31
Previously Validated Anti-Virus Software
The Mac-Lab/CardioLab system has previously been validated to run with the software listed below. It is recommended that systems run with one of the currently validated Anti-Virus software packages to avoid any unforeseen issues. Please refer to “Validated Anti-Virus Software” on page 1-16.
Supported Anti-Virus Software Supported Anti-Virus Software Version
Introduction
CA eTrust™ AntiVirus Enterprise (for a single Acquisition, Review, or Server)
McAfee® VirusScan® (with McAfee ePolicy Orchestrator v4.0, Patch 3)
McAfee Active VirusScan SMB Edition (for Acquisition, Review, or Server).
McAfee ProtectionPilot™ 1.0.1, 1.5
McAfee ViruScan Enterprise 8.5i
Norton AntiVirus Edition (for Acquisition and Review) (Norton AntiVirus software must be the boxed copy. Do not download the software from internet.)
Norton Antivirus™ (Norton AntiVirus software must be the boxed copy. Do not download the software from internet.)
Symantec EndPoint Protection 11.0.4
Symantec AntiVirus (for Acquisition, Review, or Server)
Symantec AntiVirus Corporate Edition (for Acquisition, Review, or Server)
r8.1
8.7i
8.0
2005
2007
9.0
10.2
Trend Micro OfficeScan Client/Server Edition
Trend Micro OfficeScan Client/Server Edition (for Acquisition, Review, or Server)
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-17
8.0 + SP1
8.0 (English)
7.3 (non-English and non-Swedish)
Page 32
Introduction
3rd Party: PedCath
Language PedCath Version
English v7.6.7
French v7.5.5
Swedish/German (see note) v7.6.5
Italian v7.5.5
Spanish v7.5.5
Japanese v7.4.3J
Chinese (see note) v7.4.3J
NOTE
Chinese systems will use the Japanese version of PedCath. Swedish and German systems will use the English version of PedCath.
1-18 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 33

Service Information

Service Requirements

Follow the service requirements listed below:
Equipment servicing should be performed by GE Healthcare
Any unauthorized attempt to repair equipment under warranty voids
It is the user’s responsibility to report the need for service to GE
Failure on the part of the responsible individual, hospital, or
Regular maintenance, irrespective of usage, is essential to ensure
Introduction
authorized service personnel only.
that warranty.
Healthcare or one of their authorized agents.
institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
that the system is always functional when required.

Product Identification

Every system has a unique serial number for identification. The serial number appears on the product labels which are located on the top of the CPU, the back of the Integrated Electronics Box (IEB), the right side of the RAC 4A, the back of the TRAM and the back of the Amplifier.
2
1
Identification Tag
Item Name Description
1 Manufacture Date Date of manufacture in YYYY-MM format.
2 Part Number of Product GE Healthcare part number for product.
3 Product Code Description Identifier for type of product.
4 Serial Number Unique serial number for identification.
Encoded as barcode below displayed number.
3
4
5
5 Manufacturer Information Name and address of device
manufacturer.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-19
Page 34
Introduction
XXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXX
XXXXXXXXX
XX
XXXXXXXXXXX
XXXXXXXXXXXXX
A
B
C
D
E
F
G
H
I
AA5 XX XXXX XXX X
Old Identification Tags
Systems manufactured earlier may have one of the two following styles of identification labels instead.
Identification Tag
Item Name
A Name of Device
B Manufacturer
C Serial Number
D 3-digit Product Code
E Year Manufactured
F Fiscal Week
G Product Sequence Number
H Manufacturing Site
I Miscellaneous
1-20 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 35
Identification Tag
XXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXX
XXXXXXXXX
XX
XXXXXXXXXXX
XXXXXXXXX
A
B
C
D E
F
G
H
A X XX XXXX XX
Item Name
Introduction
A Name of Device
B Manufacturer
C Serial Number
D Facility Code
E Product Sequence Number
F Product Code
G Year Manufactured
H Month Manufactured
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-21
Page 36
Introduction

Required Tools and Supplies

The following is a list of tools needed to service the Mac-Lab/CardioLab system:
Adjustable wrench with a 1-inch capacity
Phillips screwdriver
Needle nose pliers
Pliers
Socket wrench (3/16 inch to 1 inch size bits)
Phillips head screwdriver (small and large sizes)
Flat head screwdriver (small and large sizes)
Cutting pliers
Power outlet ground tester
Grounding strap
Simulator with ECG, pressure, cardiac output, and temperature
outputs
Cables to connect simulator to ECG inputs of amplifier and TRAM
Cables to connect the simulator to the catheter input modules (2 mm
shielded)
Adapter cable to connect the simulator to the amplifier and TRAM
pressure inputs
Cables to connect the simulator to the TRAM, cardiac output, and
temperature inputs

Power Input Requirements

If the equipment is installed in the USA and uses 240V rather than 120V, the source must be a center-tapped, 240V single phase circuit (120V, 20A; 240V, 10A).
1-22 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 37

Safety Information

Follow all of the instructions and warnings included in this manual and save these instructions for later use. However, these instructions in no way supersede current medical practices regarding patient care and safety.

Electromagnetic Compatibility and Interference

This catheterization system meets the requirements of the European Medical Device Directive, and by reference, the requirements for electromagnetic compatibility specified by EN 60601-1-2. This equipment is not likely to interfere with other equipment, except with certain types of rate adaptive pacemakers that rely on bioelectric impedance measurement (BIM) technology. Also, interference with the operation of this equipment is not likely to result from proximity to general equipment.
This equipment meets or exceeds Class A immunity levels. However, avoid the use of strong intentional radio transmitters in the immediate vicinity, including cellular telephones. The use of strong unintentional radiators, such as electrosurgical equipment, may cause some level of interference. GE Healthcare offers optional ECG patient cables with electrosurgical noise filtering, which reduces the level of interference on ECG traces.
Introduction

Exception

The Mac-Lab/CardioLab system and CardioLab Plus II Amplifier meets the requirements of IEC 60601-1-2 Medical Electrical Equipment, Part 1: General Requirements for Safety, 2. Collateral Standard: Electromagnetic compatibility - Requirements and tests.
If using electrosurgical or RF ablation equipment in conjunction with this equipment, select an ESU filtered cable. For more information about management of the equipment in the clinical environment (from an electromagnetic compatibility perspective) consult the Association for the Advancement of Medical Instrumentation's technical information report
Guidance on Electromagnetic Compatibility of Medical Devices for Clinical/Biomedical Engineers.
ECG Parameter
IEC 60601-1-2 clause 36.202.1 - Immunity: Radiated Immunity –
The level of compliance is less than 1V/m
If operating under the conditions defined in EMC Standard EN 60601-1­2 (Radiated Immunity 3V/m), field strengths less than 1V/m may cause waveform distortions and erroneous numeric data at various electromagnetic interference (EMI) frequencies.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-23
Page 38
Introduction
Recommendations:
Review the AAMI EMC Committee technical information report (TIR-18) titled Guidance on electromagnetic compatibility of medical devices for clinical / biomedical engineers - Part 1: Radiated radio-frequency electromagnetic energy. This TIR provides a means to evaluate and manage the EMI environment in the hospital.
The following actions can be taken:
managing (increasing) distance between sources of EMI and
susceptible devices
managing (removing) devices that are highly susceptible to EMI
lower power from internal EMI sources under hospital control (i.e.
paging systems)
labeling devices susceptible to EMI
educating staff (nurses and doctors) to be aware of, and to recognize,
potential EMI related problems

EU Battery Directive 2006/66/EC

The Mac-Lab/CardioLab product conforms to the EU Battery Directive 2006/66/EC.
The Uninterruptable Power Supply (UPS) supplied with the IEB of the Mac-Lab/CardioLab contains lead/acid storage batteries. These batteries are a permanent part of the UPS assembly and are not intended to be removed or changed. Attempting to remove or change these batteries may result in an unsafe condition.
The separate collection symbol is affixed to a battery, or its packaging, to advise you that the battery must be recycled or disposed of in accordance with local or country laws. The letters below the separate collection symbol indicate whether certain elements (Pb=Lead, Cd=Cadmium, Hg=Mercury) are contained in the battery. To minimize potential effects on the environment and human health, it is important that all marked batteries that you remove from the product are properly recycled or disposed. Information on the potential effects on the environment and human health of the substances used in batteries is available at the urls below:
http://www.gehealthcare.com/euen/weee-recycling/index.html
http://www.epa.gov/mercury/about.htm
http://www.cdc.gov/niosh/npg/npgd0383.html
http://www.epa.gov/lead/
http://www.cdc.gov/niosh/npg/npgd0368.html
http://www.cdc.gov/niosh/npg/npgd0087.html
http://www.epa.gov/waste/hazard/wastemin/minimize/factshts/ cadmium.pdf
1-24 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 39

Responsibility of the Manufacturer

GE Healthcare is responsible for the effects of safety, reliability, and performance only if:
Installation operations, extensions, readjustments, modifications, or
repairs are performed by persons authorized by GE Healthcare.
The equipment is used in accordance with the instructions provided.
The user attaches only accessory equipment recommended for
interaction with the amplifier.

General

This system is intended for use under the direct supervision of a licensed health care practitioner.
The Mac-Lab/CardioLab system, including the CardioLab Plus II Amplifier, maintains leakage currents below medical limits with an isolation transformer which is incorporated into the design of the integrated electronics box (IEB). The system power is supplied by a dedicated power supply on an isolated circuit for proper behavior of the system.
Introduction
To ensure patient safety, use only parts and accessories manufactured or recommended by GE Healthcare. Parts and accessories used must meet the requirements for the applicable EN 60601 series safety standards, and the system configuration must meet the requirements of the EN 60601-1 medical electrical systems standard. Contact GE Healthcare before connecting any device to this system that is not recommended for use in this manual.
If the equipment is installed in the USA and uses 240V rather than 120V, the source must be a center-tapped, 240V single-phase circuit.
To ensure patient safety, use only parts and accessories manufactured or recommended by GE Healthcare.
This equipment should only be connected as specified by GE Healthcare. Installation of the equipment should only be performed by qualified GE Healthcare service representatives.
Contact GE Healthcare for information before connecting any device to this equipment that is not recommended in this manual. Only cables and accessory equipment purchased from GE Healthcare should be used in conjunction with the system.
It is important that all of the instructions in this manual be followed. However, these instructions in no way supersede current medical practices regarding patient care and safety.
Refer servicing of the equipment to GE Healthcare authorized service personnel. Any attempt to repair equipment under warranty will void that warranty.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-25
Page 40
Introduction

Equipment Symbols

Equipment Symbols Descriptions
Failure on the part of responsible individuals or institutions to implement a satisfactory equipment maintenance schedule may lead to undue equipment failure and possible health hazards.
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice includes:
Use of the accessory in the Patient Vicinity.
Evidence that the safety certification of the Accessory is performed in
accordance with the appropriate EN 60601-1 and EN 60601-1 harmonized national standard.
The CardioLab II Plus Amplifier is a component/accessory of a Class II medical device.
CE Marking
Underwriters Laboratories, Inc.
Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, mechanical and other specified hazards, only in accordance with UL 2601-1, CAN/CSA C22.2 No.
601.1, IEC 60601-1 and IEC 60601-1-1.
Power Ratings Chart:
V~ = V = Voltage, ~ = Alternating Current
I = Current
Fuse =
Equipotential Ground Point
A nonessential electrical connection which may be used to connect the chassis to other equipment to achieve the same potential.
1-26 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 41
Equipment Symbols Descriptions
Type CF Equipment that is Defibrillator Proof
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and internal application to the patient excluding direct cardiac application. The “paddles” outside the box indicate that the part is defibrillator proof.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a higher degree of protection against electric shock than that provided by Type B applied parts.
Type CF Equipment
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and internal application to the patient excluding direct cardiac application.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a higher degree of protection against electric shock than that provided by Type B applied parts.
Type BF Equipment that is Defibrillator Proof
Introduction
Type BF Applied Part that is defibrillator proof: Isolated (floating) applied part suitable for intentional external and internal application to the patient excluding direct cardiac application. “Paddles outside the box indicate the applied part is defibrillator proof.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a higher degree of protection against electric shock than that provided by Type B applied parts.
Atmospheric Pressure Limits
Temperature Limits
Humidity Limits
Attention
Pay attention to the documents delivered with the equipment.
Input Symbol
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-27
Page 42
Introduction
Equipment Symbols Descriptions
General Warning
Manufacturer
Date of Manufacture
Consult Instructions for Use
WARNING: SHOCK HAZARD
Remove power cord from the mains source by grasping the plug. Do not pull on the cable. This may cause cable damage, which could result in loss of protective earthing or exposure to mains voltage.
1-28 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 43
Introduction

Compliance Information

Mac-Lab/CardioLab System

Declaration - Electromagnetic Emissions
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Emissions Tests Compliance Electromagnetic environment guidance
RF emissions CISPR 11 Group 1 The Mac-Lab / CardioLab System uses RF
energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class A The Mac-Lab / CardioLab System is suitable
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/Flicker emissions IEC 61000-3-3
Complies
for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
IEC 60601
Test Level
± 6 kV contact ± 8 kV air
± 2kV for power supply lines (signal lines exempt from testing, length is less than 3 m)
± 1kV differential mode ± 2kV common mode
Compliance
Level
± 6 kV contact ±8 kV air
± 2kV for power supply lines
± 1kV differential mode ±2kV common mode
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-29
Page 44
Introduction
Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity Test
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
IEC 60601 Test Level
<5% 100-120Vac <5% 200-240Vac (>95% dip) for
0.5 cycle
Compliance
Level
<5% 100-120Vac <5% 200-240Vac (>95% dip) for
0.5 cycle
Electromagnetic Environment - Guidance
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Mac-Lab / CardioLab System requires continued operation during power mains interruptions, it is recommended that the Mac-Lab / CardioLab System be powered from an Uninterruptable
power supply or a battery. 40% 100-120Vac 40% 200-240Vac (>60% dip) for 5 cycles
70% 100-120Vac 70% 200-240Vac (>30% dip) for 25 cycles
<5% 100-120Vac <5% 200-240Vac (>95% dip) for 5 seconds
40% 100-120Vac 40% 200-240Vac (>60% dip) for 5 cycles
70% 100-120Vac 70% 200-240Vac (>30% dip) for 25 cycles
<5% 100-120Vac <5% 200-240Vac (>95% dip) for 5 seconds
3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a
typical location typical commercial or hospital environment.
1-30 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 45
Introduction
Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity Test
Conducted RF emissions IEC 61000-4-6
Radiated RF emissions IEC 61000-4-3
IEC 60601 Test Level
3 Vrms 105kHz to 80 MHz with 2 Hz modulation
3V/m80 MHz to
2.5 GHz
Compliance
Level
Portable and mobile RF communications equipment should be used no
closer to any part of the Mac-Lab / CardioLab System, including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance (d)
3 Vrms d = 1.2 P
3 V/m d = 1.2 P
d = 2.3 P
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters(m).Field Strengths from fixed RF
transmitters, as determined by a electromagnetic site survey
less than the compliance level in each frequency range
occur in the vicinity of equipment marked with the following symbol:
Electromagnetic Environment - Guidance
2
. Interference may
1
, should be
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflected from structures, objects and people.
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Mac-Lab / CardioLab System is used exceeds the applicable RF compliance level above the Mac-Lab / CardioLab System should be observed to verify normal operation. If abnormal perfor­mance is observed, additional measures may be necessary, such as re-orienting or relocating the Mac-Lab / CardioLab System.
2
Over the frequency range 150 kHz to 80 MHz. Field strengths should be less than 3 V/m
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 1-31
Page 46
Introduction
d 1.2 P=
d 1.2 P=
d 2.3 P=
Recommended Separation Distances Between
Portable and Mobile RF Communications Equipment and the Mac-Lab/CardioLab system
The Mac-Lab/CardioLab system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Mac-Lab/CardioLab system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Mac-Lab/CardioLab system as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output
Power of Transmitter
Separation Distance According to Frequency of Transmitter
m
W
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE
At 80 MHz and 800 MHz, separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflected from structures, objects and people.
1-32 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 47

2 Equipment Overview

2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-1
Page 48
Equipment Overview

Typical System Configuration

NOTE
For configuration, component and connection information for HP xw8200 and HP xw8000 computers, refer to the Mac-Lab/CardioLab/ Centricity Cardiology INW Service Manual v6.5 (PN 2021423-022).
This system may consist of either Mac-Lab (hemodynamic), CardioLab (electrophysiologic), or ComboLab (both systems).
The system may be standalone or connected to a network. Network installations include a Centricity Cardiology INW server. In this manual, INW refers to Centricity Cardiology INW. It may be connected to an optional CVIS server.
The system may also include a CardioLab II Plus Amplifier, also referred to as CLab II Plus.
The following block diagram shows a typical setup.
Dashed lines indicate CardioLab/ComboLab only
Stimulator
Keyboard
Workstation
Control Room
Speakers
Mouse
Real-Time
Monitor
HP
CPU
UPS
Printer
Review
Monitor
IEB
Image
Monitor
3" Floor Conduit
3" Floor Conduit
3" Ceiling Conduit
Amplifier
CIM
Monitor Boom
Catheter
Personality
Module
RF
Generator
Remote
Real-Time
Monitor
TRAM
RAC4A
Remote
Review
Monitor
Bed
Procedure Room
INW
Server
IT Control Room
Network
Switch
CVIS
To Hospital Network
2-2 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 49

Control Room Components

Acquisition Computer
USB Mouse
Keyboard
Barcode scanner
Integrated Electronics Box (IEB): The IEB provides power to
local system components, signal pass through video and communications.
Uninterruptible Power Supply (UPS): The UPS provides
emergency power to the HP computer.
Monitors: Two monitors (Real-Time and Review) are standard. A
third Image monitor is optional.
Local speakers: Provide output for ECG QRS tone and optional
audible indicators.
Laser Printer: Prints images or reports.
Cardiac Stimulator: Provides direct cardiac stimulation to assist
the physician during an EP case.
Equipment Overview

Procedure Room Components

Monitors: One remote monitor is standard (and is software
switchable between Real-Time and Review windows). A second monitor is optional.
Remote speakers: Provide output for ECG QRS tone and optional
audible indicators in the procedure room.
TRAMNet: The TRAMNet gathers patient data for the Mac-Lab
system. The TRAMNet consists of a Remote Acquisition Case (RAC­4A) and a Transport Remote Acquisition Module (TRAM).
CardioLab II Plus Amplifier: The Amplifier gathers data for both
the ComboLab and CardioLab systems.
CARTO™ System (Patient Interface Unit): Provides 3D
electroanatomical cardiac maps to the CardioLab.
Analog Output Box: Distributes Heart rate and blood pressure
signals from the TRAMNet and Amplifier to other equipment.
Review Workstation: Client installed in physician office or outside
lab. Provides post-procedure analysis capabilities such as generating reports.
Nurses Workstation: Client used during the study. Provides the
operator with the ability to participate in an active study and perform actions such as document administered medications and supplies consumed.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-3
Page 50
Equipment Overview
DMS Express Server
(Interface Server)
INW Server
Review
Workstation
Mac-Lab CardioLab
Hospital
Enterprise
OR

IT Control Room Equipment

Network Switch
Centricity Cardiology INW Server
CVIS Server

Network Connections

2-4 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 51

System Hardware

WARNING
SHOCK HAZARD — Use appropriate Lock Out/Tag Out procedures before performing any service on potentially energized equipment.

Acquisition Workstation System Components

HP Computer
The HP computer is used for either Acquisition systems or Review workstations and has the following components:
Primary Hard Drive with two partitions:
C: Application partition
D: Data partition
Optional Magneto-Optical drive (Drive E:).
NOTE
The optional Magneto-Optical drive is only available on HP xw8400 computers.
Equipment Overview
DVD-RAM drive (Drive F:).
NOTE
The drive used to store application studies is determined by the file paths set in under System Settings in the main application menu.
Secondary Hard Drive (Drive G:).
Four standard expansion cards (plus optional cards):
HP xw8400: Matrox Parhelia video board provides video out for
the Real-Time (Video 1) and Review (Video 2) monitors.
HP z600: Matrox M9140 video board provides video out for the
Real-Time (Video 1), Review (Video 2), and optional (Video 3) monitors.
TRAMNet to Ethernet Adapter board for communication with
the TRAM in the RAC 4A.
PCI Communication board to provides Amplifier
Communication, Analog and Digital I/O, and Switched video (video out from the IEB).
Dual NIC Adapter board provides Network connection.
Optional on HP xw8400: Adaptec Ultra320 SCSI PCIe board for
supporting MO drive.
Optional on HP xw8400: Matrox Odyssey XA board used to
capture images from an X-Ray or other imaging system.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-5
Page 52
Equipment Overview
1
2
2
4
3
5
Optional on HP xw8400: ATI FIREGL V5200 256MB PCI-E or
ATI FIREGL V3350 256MB PCI-E Video board. Provides video out for optional (video 3) monitor.
Optional on HP z600: Matrox Odyssey EA board used to capture
images from an X-Ray or other imaging system.
External Components and Connections for the HP z600
Label Part Number Description
1 2028599-002 DVD-RAM Drive
2 N/A Power Switch
3 N/A LEDs, USB, Firewire, and Audio connections
4 N/A Backplane I/O panel
5 N/A Board Slots
6 N/A AC Power
Not Shown 2026539-001 HP USB 2-Button Optical Mouse with Wheel
Not Shown Various Keyboard (dependent on language)
2-6 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 53
Equipment Overview
External Components and Connections for the HP xw8400
Label Part Number Description
1 2028599-002 DVD-RAM Drive
2 N/A Power Switch, LEDs, USB, Firewire, and Audio
connections
3 N/A AC Power
4 N/A Backplane I/O panel
5 N/A Board Slots
Not Shown 2003340-001 Magneto-Optical Drive (optional)
Not Shown 2026539-001 HP USB 2-Button Optical Mouse with Wheel
Not Shown Various Keyboard (dependent on language)
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-7
Page 54
Equipment Overview
Disassembly
WARNING
GROUNDING HAZARD — Electrostatic discharge (ESD) can damage electronic components. Be sure you are properly grounded before opening any equipment.
Remove Case Side Panel
1. Lift the tab near the top.
2. Swing the side panel out.
Remove Front Panel - HP xw8400 only
1. Lift tabs indicated.
2. Swing the front panel out.
2-8 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 55
Equipment Overview
1
2
1
2
2
Remove Hard Drive - HP xw8400 only
1. Press tabs as indicated.
2. Slide drive out.
Remove Hard Drive - HP z600 only
1. Press tab to release handle.
2. Pull drive out.
Remove the AGP/PCI Card Support Assembly - HP xw8400 only:
1. Squeeze in the tabs, where indicated. Gently push in, then pull out.
2. Lift off the clamp.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-9
Page 56
Equipment Overview
Remove PCI Boards
1. Open the PCI retainer by pressing down on the green tabs and moving the retainer
1
away.
2. Matrox M9140 only: Carefully hold the board release tab toward the top of the case.
3. Gently pull the board out of the computer case.
Remove DVD-RAM Drive
1 2
1. Remove screw, if present.
2. Lift tab indicated.
3. Slide drive out.
Remove Power Supply - HP z600 only
1. Pull green handle to release power supply.
2. Slide power supply out.
2-10 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 57
Equipment Overview
1
2
DVD-RAM Drive
The DVD-RAM drive can also read/write to CDs. The recommended disks have a capacity of approximately 4.7 GB per side.
NOTE
The drive used to store application studies is determined by the file paths set in under System Settings in the main application menu.
12 3
1: DVD-RAM Drive Jumpers
2: IDE Connector
3: Power Connector
SATA-to-IDE Converter - HP z600 Only
The DVD-RAM drive is connected using a SATA-to-IDE converter.
Label Description
1 SATA Connection
2 4-pin Power Connection
Not Shown IDE Connection (on back of device)
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-11
Page 58
Equipment Overview
Optical Drive
The optical drive is an optional upgrade to HP Computers.
NOTE
The optional Magneto-Optical drive is not available on HP z600 computers.
NOTE
Drive use is determined by the file paths set in under System Settings in the main application menu.
1: Optical Drive Jumpers
2: SCSI Adapter (PN 2012522-002)
NOTE
Two sets of DIP switches are located on the top optical drive. These are factory preset and must not be changed.
2-12 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 59
HP z600 Installed Boards
1
2
3
4
5
6
HP z600 Custom Dual NIC Adapter Board
Equipment Overview
Part Number: 2047766-001
Location: Slot 1
Function: Network connection card, ablation connection card.
I/O: Two network connection jacks, one serial port (for connection to
ablation devices)
Uses internal serial connection from motherboard:
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-13
Page 60
Equipment Overview
1
2
3
4
5
6
HP z600 Matrox Quad Board M9140 LP Video Board
Part Number: 2049434-001
Location: Slot 2
Function: Video out for Monitor 1, Monitor 2 and Monitor 3 (optional)
I/O: One proprietary connector. Provides four DVI/VGA output ports
through a custom connector.
2-14 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 61
Equipment Overview
HP z600 Matrox Odyssey eA Board (Optional)
1
2
3
4
5
6
Part Number: 2049435-001
Location: Slot 3
Function: Image capture for X-Ray or other imaging systems
I/O: Two DVI connectors
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-15
Page 62
Equipment Overview
1
2
3
4
5
6
HP z600 TRAMNet to Ethernet Adapter board
Part Number: 2009937-003
Location: Slot 5
Function: Provides communication with the TRAM in the RAC 4A,
and converts signals from the TRAM to the ethernet.
I/O: Network Interface Port and 9-Pin D connector
2-16 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 63
Equipment Overview
1
2
3
4
5
6
HP z600 PCI Communication Board
Part Number: 2014935-002
Location: Slot 6
Functions:
Communicates with the CardioLab II Plus Amplifier via the fiber
optic connectors
Analog and Digital communication with the IEB via the D-
Connector
Provides the video switching signal for the Switched Video
output on the IEB via the D-Connector
I/O: D-Connector to IEB, Fiber Optic Jacks to the Amplifier
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-17
Page 64
Equipment Overview
HP xw8400 Installed Boards
HP xw8400 PCI Communication Board
Part Number: 2014935-002
Location: Slot 1
Functions:
Communicates with the CardioLab II Plus Amplifier via the fiber
optic connectors
Analog and Digital communication with the IEB via the D-
Connector
Provides the video switching signal for the Switched Video
output on the IEB via the D-Connector
I/O: D-Connector to IEB, Fiber Optic Jacks to the Amplifier
2-18 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 65
Equipment Overview
HP xw8400 ATI FIREGL V5200 256MB PCI-E Video Board (Optional)
Part Number: 2026189-002
Location: Slot 2
Function: Video out for optional (video 3) monitor
I/O: Two DVI connectors (only the left one is used)
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-19
Page 66
Equipment Overview
HP xw8400 Dual NIC Adapter Board
Part Number: 2017795-002
Location: Slot 3
Function: Network connection card
I/O: Two network connection jacks
2-20 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 67
Equipment Overview
HP xw8400 PCIX 2 SCSI Card
NOTE
This board is optional with the MO upgrade kit (available only for 2034493-002 – 2034493-004 Acquisition configurations).
Part Number: 2038632-001
Location: Slot 4
Function: Provides SCSI cable connection internally to an optional
MO Drive.
I/O: Two SCSI connectors (only internal SCSI is used)
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-21
Page 68
Equipment Overview
HP xw8400 Matrox Parhelia PCI 256 MB Video Card
Part Number: 2026188-001
Location: Slot 5 (PCI slot)
Function: Provides video out for the Real-Time (Video 1) and Review
(Video 2) monitors. This board supports resolutions of 1280x1024 for Mac-Lab, 1600x1200 for CardioLab.
I/O: Two DVI connectors
Included adaptors: DVI to HD15 Video Adapter (2021697-001).
2-22 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 69
HP xw8400 Matrox Odyssey XA Board (Optional)
Equipment Overview
Part Number: 2030688-001
Location: Slot 6
Function: Image capture for X-Ray or other imaging systems
I/O: Two DVI connectors
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-23
Page 70
Equipment Overview
HP xw8400 TRAMNet to Ethernet Adapter board
Part Number: 2009937-002
Location: Slot 7
Function: Provides communication with the TRAM in the RAC 4A,
and converts signals from the TRAM to the ethernet.
I/O: Network Interface Port and 9-Pin D connector
2-24 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 71
Integrated Electronics Box (IEB)
The IEB provides power to local system components and signal pass through video and communications. Three versions of the IEB are available to handle different line voltages: 100, 120, and 240 VAC.
The IEB consists of the following components:
Video Multiplexer (VID MUX) board to distribute video signals from
the video cards in the HP computer to the monitors.
I/O board to provide signal pass through for other system
components.
Isolation transformer:
Conforms to IEC 60601-1
2.0 kVA medical grade high-isolation transformation
Provides either 120 VAC (for 100 and 120 VAC systems) or 240
Equipment Overview
VAC. The primary and secondary windings are wired in parallel for 120 VAC and in series for 240 VAC.
CAUTION
CONNECTIONS — If the transformer must be disconnected for any reason, ensure the connections are correct.
UPS (either 100, 120, or 240 VAC) and Power Distribution panel
provide conditioned power to local system components.
NOTE
The UPS is external to the IEB and provides uninterrupted power to the HP computer. The monitors and other peripherials will not remain on if power is lost.
Depending on which IEB is installed, a modem and modem power supply for dial-up communication may be present. The power supply accepts any line voltage and converts it into 9 VDC.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-25
Page 72
Equipment Overview
External Components and Connections
Label Part Number Description
1 2003681-003 VID MUX board
2 2003680-001 I/O Board
3 N/A Power Connections. The bottom right connector is for the HP
computer and the four vertical connectors on the right and left are for peripheral equipment.
4 N/A Thermal Overcurrent Circuit Breaker. This breaker protects all
internal components and external devices connected to the IEB.
5 N/A Main Power Switch. This switch controls power to all internal
components and external devices except the HP computer.
Both the Main Power switch on the front of the IEB and the Thermal Overload circuit breaker on the back must be on for the system to power up.
2-26 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 73
Video Multiplexer (VID/MUX) Board
Label Description
Equipment Overview
1 Switched video out (Video 1 or 2). This is used when the facility has only
one remote monitor. To switch monitors:
Mac-Lab: Go to the main menu and select Windows > Physician Video
CardioLab: Use the dropdown box just under the main menu.
2 Video 1 (Real-time) out to local monitor (bottom) and remote monitor (top)
3 Video 2 (Review) out to local monitor (bottom) and remote (top)
4 Video 3 (CardioImage) out to local monitor (bottom) and remote (top)
5(L to R) Video 1, Video 2, Video 3 in from HP Computer Matrox AGP Video
board (Vid 1 and Vid 2) and the Matrox PCI Video board (Vid 3)
The VID/MUX board amplifies (±5V 1.25A from the internal IEB DC power supply. See “Internal Components and Connections” on page 2-31) and distributes the video signals from the HP Computer (See“HP xw8400
Video Connections” on page 2-57 or “HP z600 Video Connections” on
page 2-56).
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-27
Page 74
Equipment Overview
I/O Board
Label Description
1 Analog I/0 Communications. Connects to the HP Computer PCI
Comms board.
2 Analog in from another system (Fluoro, and so on).
3 Connection to the Analog Output Box.
4 Analog out (Stimulator, and so on).
5 I/O TRAM (includes 16.5VDC power).
6 Output to HP Computer TRAMNet to Ethernet to Adapter board.
Refer to the following cabling diagrams:
“Communications Connections” on page 2-58.
“TRAM/Amplifier Connections” on page 2-59.
2-28 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 75
Equipment Overview
Power Connections
Label Description
1 and 5 AC out to local equipment (6A @ 120 VAC max.; 3A @ 240 VAC max.)
2 Thermal Overcurrent circuit breaker. Protects all internal components
and external devices connected to the IEB.
3 AC main power in (15A @ 120 VAC max; 7A, @ 240 VAC max.)
4 AC power out to UPS (2A @ 120 VAC max; 1A, @ 240 VAC max.)
Refer to the cabling diagram, “Power Connections” on page 2-55.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-29
Page 76
Equipment Overview
Disassembly
Open Case
1. Remove 7 side-panel screws.
2. Lift the side panel out.
Remove Connection Boards
1. Open the case.
2. Remove the board screws.
3. Disconnect interior cables.
WARNING
GROUNDING HAZARD — Electrostatic discharge (ESD) can damage electronic components. Be sure you are properly grounded before opening any equipment.
2-30 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 77
Internal Components and Connections
Equipment Overview
Label Part Number Description
1 2003681-003
2003680-001
2 2013535-001 Power Supply: Provides ±5 VDC 1.25A for the Vid/
VID/MUX Board Assembly
I/O DB Connect Board Assembly
MUX board and +16.5 VDC for the RAC-4A.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-31
Page 78
Equipment Overview
Uninterruptible Power Supply (UPS)
The UPS is designed to maintain power to the computer during a power outage.
NOTE
The UPS is external to the IEB and provides uninterrupted power to the HP computer. The monitors and other peripherials will not remain on if power is lost.
Part Number Description
2034077-001 UPS Powercom, 120V, 50/60 Hz, 1500 VA, BNT
2034078-001 UPS Powercom, 230V, 50/60 Hz, 1500 VA, BNT
2034079-001 UPS Powercom, 100V, 50/60 Hz, 1500 VA, BNT
2-32 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 79

GE Client Workstation

Equipment Overview
There are several configuration options for use of the Review workstations in a networked system environment.
Nursing Workstation software allows the user to join a Mac-Lab or
CardioLab case in progress to chart and to document within the case log.
Remote Review Workstation software allows the user to review a
Mac-Lab or CardioLab case in full disclosure in order to compose or complete reports or coronary diagrams or to review patient waveforms and measurements.
Pre/Post Procedure Workstation software allows the user to start
or continue a Mac-Lab or CardioLab case in order to chart and to document within the case log.
NOTE
Automatic vitals logging is also available with the optional DASH and Solar monitor interface.
NOTE
When used in a patient care area, the hardware must meet IEC60601-1. This can be achieved through the use of a medical grade computer, separate isolation transformer or other means as determined by appropriate hospital personal.
The GE Client Review computer has the following components:
Primary Hard Drive with two partitions:
C: Application partition
D: Data partition
DVD-RAM drive (Drive E:).
NOTE
The drive used to store application studies is determined by the file paths set in under System Settings in the main application menu.
Secondary Hard Drive (Drive G:).
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-33
Page 80
Equipment Overview
1
2
3
4
5
External Components and Connections for the GE Client Workstation
Label Part Number Description
1 N/A Power and Reset Switches
2 N/A DVD Drive
3 N/A USB and Audio connections (behind cover)
4 N/A AC Power
5 Various Keyboard (dependent on language)
6 N/A Com Port for DASH/Solar connection
7 N/A Video Input
8 N/A USB 2-Button Optical Mouse with Wheel
9 N/A RJ-45 Port to Network
2-34 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 81
Disassembly
1
1
Equipment Overview
WARNING
GROUNDING HAZARD — Electrostatic discharge (ESD) can damage electronic components. Be sure you are properly grounded before opening any equipment.
Remove Case Side Panel
1. Loosen thumbscrews.
2. Pull side panel off.
Remove Hard Drive
1. Remove screws for hard
1
1
drive.
2. Slide drive out and disconnect cables.
Remove DVD Drive
1. Disconnect cables from DVD drive.
2. Remove DVD drive cage
2
screws.
3. Remove DVD drive cage from case.
4. Remove screws for DVD drive.
5. Remove DVD drive from cage.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-35
Page 82
Equipment Overview

Mobile Workstand

Refer to the instructions included with the mobile workstand for details on how to place the GE hardware on an assembled GE mobile workstand (PN 2008203-001).
6
1
2
5
Label Part Number Description
1 N/A DASH Monitor
2 N/A Mouse or Barcode Scanner location
3 N/A Keyboard location
23
4
4 2017588-001, 2021331-001, 2026192-001,
2037934-001 or 2048966-001
5 2004195-001 Isolation Transformer 120/220V 50/60HZ 600VA
6 2079070-002 LCD Monitor
2-36 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Review Workstation Computer
Page 83

Monitors

TRAMNet

Equipment Overview
The standard workstation comes with two monitors and may have a third optional monitor. The monitors include:
A Real-Time monitor to display current patient information. It is
usually located on the left.
A Review monitor to display past patient information.
An optional Image monitor (used with the CardioImage option)
displays live or frozen video images.
A single remote monitor is standard and receives input from the switched video out (Video 1 or 2) jack on the IEB (see “External
Components and Connections” on page 2-26).
To switch video for the Mac-Lab go to the main menu and select
Windows > Physician Video.
To switch video for the CardioLab use the drop-down box just under
the main menu.
Label Description
1 TRAM Module
2 To Analog Out Box
3 To IEB TRAM Connector
The TRAMNet consists of a Remote Acquisition Case (RAC-4A) and one or more TRAM modules. The TRAMNet is located in the procedure room near the procedure table.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-37
Page 84
Equipment Overview
The TRAM module receives patient inputs and then transmits signal data to the IEB and other equipment through the Analog Output Box (refer to the cabling diagrams in “TRAM/Amplifier Connections” on page 2-59.
TRAM Module Inputs:
Electrocardiogram (ECG) - green connector (5 or 10 leads)
Thermal Dilution Cardiac Output (TDCO) - brown connector
Invasive Blood Pressure (IBP) - 3 red connectors
Non-Invasive Blood Pressure (NIBP) - black connector
Pulse Oximetry (SpO
) - blue connector
2
2-38 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 85
Outputs: Analog output connector.
The following TRAM modules are supported by the Mac-Lab system:
200SL
400/450SL (Ohmeda, Nellcor)
451/451N/451M (Ohmeda, Nellcor, Massimo)
600/650SL (Ohmeda, Nellcor)
The following software versions are supported by the Mac-Lab system:
7A, 7B, 7C, 7D, 7E, 7F, 7H
8A, 8B, 8C, 8D
9A, 9B
10A, 10B
11A, 11B, 11C

CardioLab II Plus Amplifier

The CardioLab II Plus Amplifier is designed to amplify various types of cardiac signals that are acquired from a patient and transmit them back to the CardioLab system. These signals include 12 Lead ECG, intracardiac and pressure signals.
Equipment Overview
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-39
Page 86
Equipment Overview
Front View
Front View 64/32 Channels- CardioLab II Plus Amplifier
Name Description
Pressure Inputs Four pressure inputs will each accept a pressure transducer that is calibrated to 5uV/V/mmHg. See the Accessory
Parts List for GE Healthcare Systems Information Technologies recommended pressure transducers.
ECG Input The ECG input connector will accept a 10 lead ECG input cable.
Stimulator Input The stimulator input will accept a stimulator input cable. Use the 3 ft, 15 ft or 30 ft Touchproof Stimulator Cables
(part numbers, 301-00204-03, 301-00204-15, 301-00204-30, respectively) with the CardioLab II Plus Amplifier.
Catheter Inputs Up to three catheter inputs will each accept a 32 pole catheter input module. Use the CardioLab II Plus Catheter
Input Modules (part number 301-00202-08) with the CardioLab II Plus Amplifier.
Auxiliary Inputs The auxiliary input will accept the Auxiliary Input Cable (part number 301-00205-08).
Power Indicator The power indicator is on as long as the CardioLab II Plus Amplifier is turned on.
Communication Indicators
Three communication indicators show the status of the communication between the acquisition computer and the CardioLab II Plus Amplifier.
When the XMT indicator is on, the CardioLab II Plus Amplifier is transmitting data to the acquisition computer.
The XMT indicator should remain on as long as the power indicator is on.
When the RCV CMD indicator is on, the CardioLab II Plus Amplifier is receiving data from the acquisition
computer. The RCV CMD indicator will turn on as long as an instruction is being received by the amplifier.
When the RCV ERR indictor is on, the CardioLab II Plus Amplifier may not be receiving data from the
acquisition computer in a valid format or there may be a problem with the fiber optic cable.
2-40 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 87
Front View 128/96 Channels- CardioLab II Plus Amplifier
Equipment Overview
Name Description
Pressure Inputs Four pressure inputs will each accept a pressure transducer that is calibrated to 5uV/V/mmHg. See the Accessory
Parts List for GE Healthcare Systems Information Technologies recommended pressure transducers.
ECG Input The ECG input connector will accept a 10 lead ECG input cable.
Stimulator Input The stimulator input will accept a stimulator input cable. Use the 3 ft, 15 ft or 30 ft Touchproof Stimulator Cables
(part numbers, 301-00204-03, 301-00204-15, 301-00204-30, respectively) with the CardioLab II Plus Amplifier.
Catheter Inputs Up to seven catheter inputs will each accept a 32 pole catheter input module. Use the CardioLab II Plus Catheter
Input Modules (part number 301-00202-08) with the CardioLab II Plus Amplifier.
Auxiliary Inputs The auxiliary input will accept the Auxiliary Input Cable (part number 301-00205-08).
Power Indicator The power indicator is on as long as the CardioLab II Plus Amplifier is turned on.
Communication Indicators
Three communication indicators show the status of the communication between the acquisition computer and the CardioLab II Plus Amplifier.
When the XMT indicator is on, the CardioLab II Plus Amplifier is transmitting data to the acquisition computer.
The XMT indicator should remain on as long as the power indicator is on.
When the RCV CMD indicator is on, the CardioLab II Plus Amplifier is receiving data from the acquisition
computer. The RCV CMD indicator will turn on as long as an instruction is being received by the amplifier.
When the RCV ERR indictor is on, the CardioLab II Plus Amplifier may not be receiving data from the
acquisition computer in a valid format or there may be a problem with the fiber optic cable.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-41
Page 88
Equipment Overview
Back View
Back View - CardioLab II Plus Amplifier
Name Description
FDDI Connection to Computer
Power Input The GE-supplied power cord connects to the Power Input.
Equipotentiality Terminal
This connector will accept a FDDI type fiber optic cable. The fiber optic cable allows for two-way communication between the amplifier and the CardioLab system.
The Equipotentiality Terminal is connected to earth ground.
2-42 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 89
CardioLab II Plus Catheter Input Module
Equipment Overview
The equivalent to the Stimulus Inputs 1-4 on the Stimulator Input Cable are labelled on the CardioLab II Plus Amplifier as Stimulus Inputs A1, A2, B1 and B2.
The four stimulus inputs on the Stimulator Input Cable connect to the stimulator.
Each Catheter Input Module is broken down into two groups of 16 poles each. For example, Block A is broken into A (1-16) and A (17-32).
Each pacing channel, requires one positive and one negative pole. In order to maximize the number of pacing channels, only one positive and one negative designation should be made for each group.
For detailed instructions on how to designate pacing channels and enable pacing for particular sites, refer to the on-line help in the software application.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-43
Page 90
Equipment Overview
Accessory Parts List
GE Healthcare Systems Information Technologies recommends the following accessory parts be used with the CardioLab II Plus Amplifier:
Accessory Parts List
Accessory Name Part Number
Abbott Transpac IV Disposable Transducer with 4 ft extension Cable*
Abbott Transpac Reusable Cable* Abbott #42661-04-42
3M Disposable Electrode for ECG Monitoring*
* This accessory is not stocked by GE Healthcare Systems Information Technologies.
Abbott #42582-08
3M Red Dot #2256
2-44 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 91

CARTO™ System (Patient Interface Unit)

1
Label Description
1 CARTO XP PIU Output to Amplifier
Equipment Overview
CAUTION
EXCESSIVE SIGNAL NOISE — Using a CARTO XP PIU earlier than version 10 with a crossover cable may cause excessive signal noise.
If this excessive signal noise occurs, then discontinue using the crossover cable until the PIU is replaced with a later version. (The new PIU must be purchased from Biosense Webster).
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-45
Label Description
1 CARTO 3 PIU Output to Amplifier
Page 92
Equipment Overview
NOTE
For specific information related to the operation, troubleshooting, or repair of the CARTO System, refer to the OEM documentation. For information about the CardioLab/CARTO System interface, refer to the CardioLab Operator manual.
The CARTO System acquires and displays electroanatomical 3D images of heart chambers. This information is displayed on the CARTO System and can be exported into the CardioLab application.
Two connections are required: one from the CARTO System Patient Interface Unit to the CardioLab II Plus Amplifier and another from the CARTO System computer to the HP computer. Refer to the cabling diagram in “CARTO™XP System Connections” on page 2-65 or
“CARTO™ 3 System Connections” on page 2-66 for specific information.
To quickly view the status of the interface, check the CARTO System icon at the bottom-right of the Real-Time window (a heart against a grid). The color changes, depending on the system state:
Gray: Disconnected
Blue: Standby
Green: Passive Interface
Yellow: Active Interface
2-46 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 93

Analog Output Box

Equipment Overview
Label Description
1 Input from TRAMNet
2 Input from Amplifier through the IEB
The analog output box distributes heart rate and blood pressure
signals to multiple outputs (refer to the connection diagram in
“TRAM/Amplifier Connections” on page 2-59. This may be done to
synch auxiliary equipment or display the information on a scope or other device.
Patient information comes from either the TRAMNet or the Amplifier (through the IEB) and is output through the output jacks on the top of the box:
Channels 2-4 and 13-16 are switchable between the TRAMNet (Mac-
Lab) and CardioLab II Plus Amplifier (CardioLab).
The remaining channels are used exclusively by CardioLab.
Channels 1-12 may be distributed to up to 4 other devices using the
selector switch.
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-47
Page 94
Equipment Overview
(
4

Network Components

Servers
Centricity Cardiology INW Server provides
Network repository for patient studies
Acquisition system and Review workstation access to patient
studies
CVIS Server provides
Central list management for staff, supplies, and so on.
Patient scheduling
Patient information
Inventory Management
Statistical Reports
Centricity Cardiology INW G6 Server
The G6 server has the following specifications:
Intel Xeon E5504, Quad-Core, 2.00 GHz, 4MB L3 Cache, 80W
Two tiers:
Low: Three 146GB Drives
High: Six 300GB Drives
Separate C: and D: drives (see the following chart):
The C: drive is RAID 1
The D: drive is RAID 5
DDiisskk1
6GB
11446
RRAAIIDD11++0
(CC––DDrriivvee))
2
1
DDiisskk2
6GB
11446
0
DDiisskk3 33000
Parity
0GB
3
DDiissk 33000
Parity
k
0GB
4
DDiisskk5
0GB
33000
Parity
5
DDiisskk6 33000
Parity
0GB
6
DDiisskk7
0GB
33000
Parity
7
DDiisskk8
0GB
33000
Parity
8
2-48 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 95
Equipment Overview
12
3
Label Description
1 CD/DVD Drive
2 Tape Drive
3 Power Switch, USB Connectors, External Monitor Connection
4 Hard Drives
Picture Name Description
4
Part Number: 2050659-001
HP Network Interface Card
HP Tape Drive Controller
Function: Network Interface Card
Location: Slot 4, 5
Part Number: 2050328-001
Function: Tape Drive Controller Card
Location: Slot 4, 5
Centricity Cardiology INW G5 Server
The G5 server has the following specifications:
Intel Dual-Core Xeon 5120/1.86 GHz (1066 MHz) with 4MB Level 2
cache (2MB per core)
Three tiers:
Low: Three 72 GB drives
Mid: Three 146 GB drives
High: Eight 146 GB drives
Separate C: and D: drives (see the following chart):
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-49
Page 96
Equipment Overview
The C: drive is RAID 1 mirrored
The D: drive is RAID 5 mirrored
1 2
Label Description
1 Hard Drives
2 CD/DVD Drive
3 Floppy Disk Drive
4 Tape Drive
Picture Name Description
3
4
Part Number: 2036800-038
HP SMART ARRAY P/400
2-50 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Function: Hard drive RAID controller
Location: Slot 4, 5
Page 97
Centricity Cardiology INW G4 Server
1
23
4
The G4 server has the following specifications:
3.2 GHz Xeon processor with 1 GB RAM
Three tiers:
Low: Three 72 GB drives
Mid: Three 146 GB drives
High: Four 300 GB drives
Separate C: and D: drives (see the following chart):
The C: drive is RAID 1 mirrored
The D: drive is RAID 5 mirrored
Equipment Overview
Label Description
1 Hard Drives
2 AIT Drive
3 CD/DVD Drive
4 Floppy Drive (behind cover)
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-51
Page 98
Equipment Overview
Picture Name Description
HP PCI GIGABIT NIC
HP SMART ARRAY 641
HP BBWC ENABLE - MEM ­128 MB DDR
Part Number: 2026881-001
Function: Provides reliable network connection
NOTE
On G4 servers, connect the ethernet cable to the NIC card. Do not use the motherboard network port.
Location: Slot 5
Part Number: 2016993-001
Function: Hard drive RAID controller
Location: Slot 2
Part Number: 2026817-001
Function: Improves server performance and decreases
data loss potential if power is lost
Location: Connected to the SMART ARRAY card in Slot 2
2-52 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Page 99

Cabling and Connections

z600 Computer
I/O Backplane
1
2
3
4
6
5
WARNING
SHOCK HAZARD — Use appropriate Lock Out/Tag Out procedures before performing any service on potentially energized equipment.

HP z600 Computer I/O Backplane Connections

Equipment Overview
Label Part Number Description
1 N/A PS2 Connection for Keyboard
2 2020280-001
(Note: This number is
for the cable only)
3 HP 237241-001 USB Port for Mouse (Must use this port)
4 N/A RJ-45 Port to Network
5 N/A To Speakers
6 2025965-002 USB Port for Printer (cable to use depends on the printer)
2021423-503H Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2-53
PS2 connection for Barcode Scanner Y-Cable (other side connects to Keyboard)
Page 100
Equipment Overview

HP xw8400 Computer I/O Backplane Connections

Label Part Number Description
1 N/A PS2 Connection for Keyboard
2 2020280-001
(Note: This number is
for the cable only)
3 6119-101
2031697-001
4 HP 237241-001 USB Port for Mouse (Must use this port)
5 N/A To Speakers
6N/AHP xw8400 only: To Ablation Interface
7 N/A RJ-45 Port to Network
8 2025965-002 USB Port for Printer (cable to use depends on the printer)
PS2 connection for Barcode Scanner Y-Cable (other side connects to Keyboard)
HP xw8400 only: To Local Printer (cable to use depends on the printer).
2-54 Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual 2021423-503H
Loading...