Mac-Lab/CardioLab/Centricity
Cardiology INW Service Manual
Version 6.5.6
Instructions
2021423-503 Revision H
Page 2
The information in this manual only applies to Mac-Lab/CardioLab/Centricity Cardiology INW Service
Manual software version 6.5.6. It does not apply to earlier software versions. Due to continuing product
innovation, specifications in this manual are subject to change without notice.
Listed below are GE Medical Systems Information Technologies trademarks used in this document.
MAC-LAB
®
, CARDIOLAB®, CENTRICITY®, TRAM®, TRAMNET®, and TRAM-RAC® are trademarks of GE
Medical Systems Information Technologies, a division of General Electric Company going to market as GE
Healthcare.
All other trademarks contained herein are the property of their respective owners.
T-2Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
25 August 2014
Page 3
CE Marking Information
Compliance
The Mac-Lab/CardioLab/ComboLab system bears CE mark CE-0459
indicating its conformity with the provisions of the Council Directive 93/
42/EEC concerning medical devices and fulfills the essential
requirements of Annex I of this directive.
For devices manufactured in the United States, the CE mark is applied
under the authority of Notified Body GMED (0459).
The country of manufacture and appropriate Notified Body can be found
on the equipment labeling.
The product complies with the requirements of standard IEC 60601-1-2.
The safety and effectiveness of this device has been verified against
previously distributed devices. Although all standards applicable to
presently marketed devices may not be appropriate for prior devices
(that is, electromagnetic compatibility standards), this device will not
impair the safe and effective use of those previously distributed devices.
Refer to the user’s information.
CE Marking Information
Recommendations
Refer to the Mac-Lab/CardioLab Service Manual for Compliance
Information.
Users should be aware of known RF sources, such as radio or TV stations
and hand-held or mobile two-way radios, and consider them when
installing a medical device or system.
Be aware that adding accessories or components, or modifying the
medical device or system may degrade the EMI performance. Consult
with qualified personnel regarding changes to the system configuration.
The Mac-Lab/CardioLab System, if operating under the conditions
defined in the EMC Standard IEC 60601-1-2 (Radiated Immunity 3 V/m)
and at field strengths above 3 volts per meter may cause waveform
distortions at radio frequency (RF) electromagnetic interference (EMI).
MEDICAL ELECTRICAL EQUIPMENT needs special precautions
regarding EMC and needs to be installed and put into service according
to the EMC information provided in the ACCOMPANYING SERVICE
MANUAL.
Portable and mobile RF communications equipment can affect
MEDICAL ELECTRICAL EQUIPMENT.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service ManualCE-1
Page 4
CE Marking Information
The use of ACCESSORIES, transducers and cables other than those
specified, with the exception of transducers and cables sold by the
manufacturer of the Mac-Lab/CardioLab System as replacement parts
for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the Mac-Lab/CardioLab System.
The Mac-Lab/CardioLab System should not be used adjacent to or
stacked with other equipment and that if adjacent to or stacked use is
necessary, the Mac-Lab/CardioLab System should be observed to verify
normal operation in the configuration in which it will be used.
Review the AAMI Committee Technical Information Report (TIR) 18,
“Guidance on Electromagnetic Compatibility of Medical Devices for
Clinical/Biomedical Engineers”. This guidance document provides a
means to evaluate and manage the EMI environment in the hospital.
The following actions can be taken to reduce the risk of medical device
EMI and achieve EMC:
Assess the EMC environment of the healthcare facility (for example,
identify radio transmitters in around the facility) and identify areas
where critical medical devices are used (for example, ER, ICU, CCU,
NICU).
Increase the distance between sources of EMI and susceptible
devices.
Remove the devices that are highly susceptible to EMI.
Lower power transmitted from electrical and electronic equipment
(EMI sources) under hospital control (that is, paging systems).
Label devices susceptible to EMI.
Educate healthcare facility staff (nurses and doctors) to be aware of,
If operating under the conditions defined in EMC Standard EN 60601-12 (Radiated Immunity 3V/m), field strengths less than 1V/m may cause
waveform distortions and erroneous numeric data at various
electromagnetic interference (EMI) frequencies.
CE-2Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 5
Recommendations
CE Marking Information
Review the AAMI EMC Committee technical information report (TIR-18)
titled Guidance on electromagnetic compatibility of medical devices for
clinical / biomedical engineers - Part 1: Radiated radio-frequency
electromagnetic energy. This TIR provides a means to evaluate and
manage the EMI environment in the hospital.
The following actions can be taken:
managing (increasing) distance between sources of EMI and
susceptible devices
managing (removing) devices that are highly susceptible to EMI
lower power from internal EMI sources under hospital control (i.e.
paging systems)
labeling devices susceptible to EMI
educating staff (nurses and doctors) to be aware of, and to recognize,
potential EMI related problems
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service ManualCE-3
Page 6
CE Marking Information
CE-4Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manualvii
Page 14
viiiMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 15
1Introduction
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-1
Page 16
Introduction
Manual Information
Purpose
This manual contains the instructions necessary to operate and service
the Mac-Lab/CardioLab system safely in accordance with its function
and intended use. These instructions include but are not limited to:
Describing fundamental controls and indicators.
Operating procedures.
Connecting and disconnecting detachable parts and accessories.
Troubleshooting and service procedures.
Intended Audience
This manual is intended for the person who maintains and troubleshoots
this equipment.
Revision History
RevisionDateComment
A6 November 2009Initial release of the manual.
B21 April 2010Updated with serviceability
improvements.
C1 August 2011Updated to address new hardware.
D8 November 2011Added power cords.
E28 February 2012Updated Hard Drive, Memory and RMOT
parts.
F7 May 2013Updated to include installation and
troubleshooting procedures for the
Matrox Quad video board.
G18 November
2013
H24 March 2014Updated to include information for EIZO
Updated to include information for new
speakers.
monitors.
1-2Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 17
Conventions
Introduction
The following conventions are used throughout this manual.
BoldIndicates keys on the keyboard, text to be entered, or hardware items
such as buttons or switches on the equipment.
ItalicsIndicates software terms that identify menu items, buttons, or
options in various windows.
[Key1] + [Key2]Indicates a keyboard operation. A (+) sign between the names of two
keys indicates that you must press and hold the first key while pressing
the second key once.
For example, “Press Ctrl + Esc” means to press and hold down the
Ctrl key while pressing the Esc key.
<Space>Indicates you must press the spacebar on the keyboard when typing a
precise text string with one or more spaces. The point where the
spacebar must be pressed is indicated as: <Space>. The purpose of
the < > brackets is to ensure you press the spacebar when required.
Product References
Related Documents
EnterIndicates you must press the “Enter” or “Return” key on the keyboard.
Do not type “enter”.
The name of the product described in this manual is Mac-Lab/CardioLab/
Centricity Cardiology INW System.
2021423-022Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual
(v6.5.4 and earlier)
2027332-103Mac-Lab/CardioLab Security Guide
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-3
Page 18
Introduction
Illustrations and Names
All illustrations in this manual are provided as examples only. They may
not necessarily reflect your system’s configuration or the data on your
system.
In this manual, all names appearing in examples and illustrations are
fictitious. The use of any real person’s name is purely coincidental.
Chapter Contents
Chapter 1, Introduction: This chapter describes the Service
Chapter 2, Equipment Overview: This chapter describes the
Chapter 3, Maintenance: This chapter contains preventive
Chapter 4, Troubleshooting: This chapter contains
Chapter 5, Recovery Procedures: This chapter details procedures
Chapter 6, Parts List: This chapter contains a list of Field
Appendix A, Abbreviations: This appendix lists all abbreviations
manual and chapter contents. Also included are general safety
requirements, service requirements, equipment symbols, and serial
number identification information.
equipment and its technical characteristics, connector locations, and
preparation for use.
maintenance information and cleaning guidelines.
troubleshooting instructions and functional test directions.
to restore a workstation or server in the field after failure.
Replaceable parts.
used with the system.
1-4Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 19
Safety Information
Safety Messages
Definitions
Introduction
DANGER
— Safety messages which indicate an imminently
hazardous situation which, if not avoided, WILL result in
death or serious injury.
WARNING
— Safety messages which indicate a potentially
hazardous situation which, if not avoided, COULD result
in death or serious injury.
Applicable Messages
CAUTION
— Safety messages which indicate a potentially
hazardous situation which, if not avoided may result in
minor or moderate injury, loss of data or system failure.
NOTE
Messages which provide additional user information.
DANGER
EXPLOSION HAZARD — Using this equipment in the
presences of flammable gasses (including anesthetics and
oxygen) may cause an explosion or fire.
Always use this equipment in a well ventilated area away
from the presence of potentially flammable gasses
WARNING
SYSTEM STABILITY — All systems should be installed
and maintained according to installation instructions and
manufacturers specifications. Any alterations may affect
the stability of the system.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-5
Page 20
Introduction
WARNING
PATIENT MONITORING — The system is intended to
be used as a recording system for catheterization,
electrophysiology and related specialty laboratories. A
defibrillator or ECG monitor should be attached for
patients in need of uninterrupted ECG display. An
additional means to display SPO
patients in need of uninterrupted SPO
should be available for
2
display. A
2
temporary pacemaker needs to be available for patients in
need of uninterrupted delivery of pacing.
WARNING
SUPERVISED USE — This device is intended for use
under the direct supervision of a licensed health care
practitioner.
WARNING
ACCESSORIES — All system components sold by GE
Healthcare for use with this system are tested to meet
IEC 60601-1 when used with the system. All other
components used with the system must meet IEC
60601-1.
WARNING
IEC — All equipment not complying with IEC 60601-1
should be placed outside the patient environment.
WARNING
SHOCK HAZARD — DO NOT allow hydraulic fluid from
any device to contact any conductive surface. Hydraulic
fluid is electrically conductive.
An electrical shock hazard exists between chassis ground
and isolated (floating) ground when power is applied.
Unplug the unit from the power source before proceeding.
To reduce the risk of electric shock or damage to your
equipment, do not disable the power cord grounding
feature. This equipment is designed for connection to a
earth grounded power outlet. The grounding plug is an
important safety feature.
1-6Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 21
Introduction
WARNING
SHOCK HAZARD — Keep the conductive parts of lead
electrodes and associated parts away from other
conducting parts, including earth.
WARNING
SHOCK HAZARD — Total system earth leakage current
must not exceed 500 microamperes.
Voltages between the three power outlet connections must
be accurate BEFORE measuring power outlet ground-toneutral loop resistance.
Electrostatic discharge (ESD) can damage electronic
components. Be sure you are properly grounded when
using or servicing the system.
WARNING
SHOCK HAZARD — To reduce the risk of electric shock
or damage to the equipment, do not disable the power cord
grounding feature. This equipment is designed for
connection to a grounded (earthed) power outlet.
WARNING
SHOCK HAZARD — Use appropriate Lock Out/Tag Out
procedures whenever an energy hazard exists, such as
disconnecting equipment to perform maintenance.
WARNING
SHOCK HAZARD — Electrical shock hazard between
chassis ground and isolated (floating) ground when power
is applied. Unplug the unit from the power source before
proceeding.
To reduce the risk of electric shock or damage to your
equipment, do not disable the power cord grounding
feature. This equipment is designed for connection to a
grounded (earthed) power outlet. The grounding plug is
an important safety feature.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-7
Page 22
Introduction
WARNING
SHOCK HAZARD — High voltages exist in the unit. Use
insulated tools. Remove jewelry and use only one hand
when possible.
Always disconnect power from the equipment before
attempting service.
WARNING
SHOCK HAZARD — Only connect IEC 60601-1 certified
medical equipment. If fluid of any kind should leak into
the system, discontinue use of the equipment and contact
GE Healthcare immediately.
WARNING
SHOCK HAZARD — To reduce the risk of electric shock,
DO NOT remove cover or back. This equipment contains
no user serviceable parts. Refer servicing to qualified
service personnel.
WARNING
SHOCK HAZARD — Proper connections to the patient
are critical for the proper operation of the CardioLab II
Plus Amplifier and patient safety. Use extreme caution
when making patient connections. Do NOT connect nonisolated equipment to the amplifier. Proper connections to
the stimulator are critical for the proper operation of the
CardioLab II Plus Amplifier and patient safety. Use
extreme caution when making stimulator connections.
The stimulator must be a IEC 60601-1 certified product
Ablation operations should NOT be executed when a
stimulator unit is connected.
WARNING
SHOCK HAZARD — Do not allow any foreign materials
such as fluids to come in contact with the CardioLab II
Plus Amplifier. If fluid of any kind should leak into the
CardioLab II Plus Amplifier, discontinue using the
amplifier and contact GE Healthcare. DO NOT allow
hydraulic fluid from any device to contact any conductive
surface. Hydraulic fluid is electrically conductive.
1-8Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 23
Introduction
WARNING
SHOCK HAZARD — Use appropriate Lock Out/Tag Out
procedures before performing any service on potentially
energized equipment.
WARNING
SHOCK HAZARD — Remote speakers must be powered
by a 60601-1 certified power supply.
WARNING
CONDUCTION HAZARD — Do not leave conductive
parts of lead electrodes open. This condition could leave
the patient or user more susceptible to shock.
WARNING
GROUNDING — Electrostatic discharge (ESD) can
damage electronic components. Be sure you are properly
grounded before opening any equipment.
WARNING
POWER SUPPLY — The supply line must be fused and
capable of delivering 16A.
Route the AC power cable away from moving parts.
Connect only to a properly earth grounded outlet.
WARNING
DEFIBRILLATOR PRECAUTIONS — DO NOT contact
the amplifier or patient during defibrillation.
Defibrillation equipment is required in the event that the
patient needs to be cardioverted or defibrillated.
WARNING
INTERNAL DEFIBRILLATION — Prior to internal
defibrillation, disconnect the intercardiac catheters from
the system to prevent damage to the amplifier.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-9
Page 24
Introduction
WARNING
INTERPRETATION HAZARD — A qualified physician
must overread all computer-generated measurements and
tracings. Computerized interpretation is only significant
when used in conjunction with clinical findings.
Signal quality should be checked only by a licensed health
care practitioner.
All computer assisted measurement functions must be
verified by a licensed health care practitioner.
WARNING
ANNUAL EQUIPMENT MAINTENANCE — The
system should be checked yearly to ensure compliance
with medical device standards.
WARNING
PINCH POINT HAZARD — Keep hands, hair, jewelry
and loose clothing away from moving parts. Otherwise,
serious injury could result.
WARNING
REPLACEMENT PARTS — Replace only with the same
type and rating of fuse.
WARNING
SOFTWARE — DO NOT add unauthorized software to
the system. Doing so may make the software unstable.
WARNING
REMOVABLE MEDIA — Exit the software application
properly before turning off the system. Failure to do so
may result in loss of ability to store the data properly.
Follow recommended instructions when exiting the
application or turning off the system to ensure data
integrity.
1-10Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 25
Introduction
CAUTION
ELECTROSTATIC SHOCK — Working on electronic
devices while not properly grounded my cause damage to
the equipment.
Always use a wrist grounding strap and grounding pad
when working on electronic equipment.
CAUTION
SALE OF DEVICE — Federal (USA) law restricts this
device to sale by or on the order of a physician.
CAUTION
EQUIPMENT CABLING DAMAGE — Route optical
cables thorough conduit in the ceiling or floor to avoid
damage to the cables or cable connectors.
CAUTION
EQUIPMENT USE —
DO NOT reuse single use devices.
CAUTION
EMC — No known electromagnetic or other interference
between the system exists. However, precautions should
be taken to avoid the use of cellular telephones, pagers, or
other transmitters.
CAUTION
TRANSDUCER — All air must be removed from the
transducer or associated tubing before calibration for
accuracy.
Pressure channels should be calibrated to a known
pressure value if the pressure value is in question.
CAUTION
REMOTE MONITOR — Always power the system
through separate isolation transformers which are
independently connected to a wall outlet.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-11
Page 26
Introduction
ITE Conformance Statement
The Mac-Lab/CardioLab/Centricity Cardiology INW Server software
applications are intended to be installed and run on an applicable
ITE compliant hardware platforms (minimum Hardware Essential
Performance be ITE compliant) which may not meet leakage and/or
EMC requirements of patient care area equipment. This means but
is not limited to the following (unless the medical electrical standard
[EN60601-1-1 and EN60601-1-2] is followed):
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware does not directly deliver energy to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware does not indirectly control energy delivered to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware is not directly connected to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server
hardware is not installed or used in the Patient vicinity.
Patient Vicinity summarized as follows:
A space, within a location intended for the examination and
treatment of patients, extending 6 ft (1.83 m) beyond the normal
location of the bed, chair, table, treadmill, or other device that
supports the patient during examination and treatment and
extending vertically to 8 ft 2.4 in. (2.5 m) above the floor.
Areas where health care staff members monitor patients and
perform charting and administrative tasks are not considered to
be Patient Vicinity areas.
1-12Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 27
Types of Systems
Mac-Lab System
Introduction
The Mac-Lab System is intended for use in a catheterization and related
cardiovascular specialty laboratories under the direct supervision of a
licensed healthcare practitioner. It is intended to monitor, calculate and/
or record cardiovascular data from adult and pediatric patients
undergoing cardiac catheterization procedures. The data may be
manually entered or acquired via interfaced devices. Data includes:
ECG, heart rate, pulse oximetry (SpO
temperature, valve gradients and areas, cardiac output, hemodynamic
measurements, invasive and noninvasive blood pressure and procedural
information and optional intracardiac electrocardiogram (IECG).
Physiological parameters such as diastolic, systolic, and mean blood
pressure, heart rate, and cycle length may be derived from the signal
data, displayed and recorded. The system allows the user to monitor the
acquisition of data, review the data, and generate reports on the data.
Additionally, the system may acquire, amplify, display and record data
received from other interfaced medical devices typically used during
these procedures, such as imaging devices. The Mac-Lab System does not
control the delivery of energy, administer drugs, perform any lifesupporting or life-sustaining functions, or analyze data acquired during
the procedure. The Mac-Lab System does not transmit alarms or
arrhythmias and does not have arrhythmia detection capabilities.
), respiration rate, EtCO2,
2
CardioLab EP System
The CardioLab EP System is intended for use in an electrophysiological
laboratory and related specialty laboratories under the direct
supervision of a licensed healthcare practitioner. It is intended to
monitor, calculate and/or record electrophysiological data from adult and
pediatric patients undergoing electrophysiological studies. Data
includes: ECG, pressure, and intracardiac electrocardiogram (IECG)
waveforms, heart rate, pulse oximetry (SpO
temperature, invasive and noninvasive blood pressure, and procedural
information. Physiological parameters such as diastolic, systolic, and
mean blood pressure, heart rate, and cycle length may be derived from
the signal data, displayed and recorded. The system allows the user to
monitor the acquisition of data, review the data, and generate reports on
the data. Additionally, the system may acquire, amplify, display and
record data received from other interfaced medical devices typically used
during these procedures, such as imaging devices and RF generators.
The CardioLab EP System does not control the delivery of energy,
administer drugs, perform any life-supporting or life-sustaining
functions, or analyze data acquired during the procedure. The CardioLab
EP System does not transmit alarms or arrhythmias and does not have
arrhythmia detection capabilities.
), respiration rate, EtCO2,
2
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-13
Page 28
Introduction
ComboLab System
The ComboLab System is the combination of the both the Mac-Lab and
CardioLab EP systems. The ComboLab System is intended for use in
either a catheterization laboratory or electrophysiological laboratory and
related speciality laboratories under the direct supervision of a licensed
healthcare practitioner. The ComboLab System allows the user to run
either the Mac-Lab System or the CardioLab EP System, although only
one may be used at a time.
CardioLab II Plus Amplifier
The CardioLab II Plus Amplifier is intended for use in an
electrophysiological catheter laboratory or catheterization laboratory
where ECG, direct cardiac, intracardiac and pressure signals need to be
recorded from a patient. This device is intended for use under the direct
supervision of a licensed health care practitioner.
1-14Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 29
Security Information
Validated configurations consist of product scenario implementations
that are tested against GE product failure and sample cases run to
simulate a real environment while conducting testing. GE provides
service and support based on validated configurations only and bears
responsibility for GE products only.
Non-Validated Software Applications
This is to detail the policy for non-validated software applications
applied to the following products:
Mac-Lab Hemodynamic Systems
CardioLab Electrophysiology Systems
ComboLab Hemodynamic and Electrophysiology Systems
Centricity Cardiology INW Server
Introduction
Policy
Our policy is that we do not recommend installations of any other third
party software applications (that is, not mentioned above) on the
products above as they could affect performance, warranty and
serviceability. Examples of such might include other Anti-Virus
programs, print servers, interface engines, anti-spyware, anti-adware,
viruses, malware, and so on. GE does not support the use or installation
of these non-validated applications or tools. If such programs are
installed the customer bears the risk of the product not functioning
properly and any risks that are introduced for patient safety. GE will not
guarantee support for an environment that is different than a validated
implementation. In cases where this is found and technical assistance is
requested, GE will provide support, by first eliminating the third party
software on the product and bringing it back to a known and validated
configuration. If it is found that the third party software is the cause of
product failure, any costs associated with this will be the responsibility of
the hospital.
At GE Healthcare we understand the uniqueness of our customers and
pride ourselves in the personalized solutions we provide to help you in
your quest to care for patients. We hope this section will help you to
better understand our policy on non-validated software applications and
we hope you can work with us as we proceed in further developing our
products for your future benefit.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-15
Page 30
Introduction
Validated Third Party Software
Currently validated and allowed third party software such as Enterprise
Back-up, Archival, Anti-virus and other applications to run on the same
computer are:
Enterprise Backup Software
IBM Tivoli Storage Managerv6.1
Symantec Backup Exec for Windows Serverv12.5
CA ARCserv Backup for Windowsr12.5
EMC Networkerv7.5 + SP1
Validated Anti-Virus Software
Mac-Lab/CardioLab/ComboLabIT (Version 6.5.6)
The Mac-Lab/CardioLab 6.5.6 system has been validated to run with the
software listed below. Except where noted, the software below supports
English, French, German, Italian, Spanish, Swedish, Chinese, and
Japanese:
Supported Anti-Virus SoftwareSupported Anti-Virus Software Version
The Antivirus Server software can be installed at a hospital domain level
and the Centricity Cardiology INW server can be a client for the
Antivirus software.
8.7i, Patch 2
Spanish, Chinese)
11.0.4 MR4 + MP2 (Japanese)
10.0
1-16Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 31
Previously Validated Anti-Virus Software
The Mac-Lab/CardioLab system has previously been validated to run
with the software listed below. It is recommended that systems run with
one of the currently validated Anti-Virus software packages to avoid any
unforeseen issues. Please refer to “Validated Anti-Virus Software” on
page 1-16.
Supported Anti-Virus SoftwareSupported Anti-Virus Software Version
Introduction
CA eTrust™ AntiVirus Enterprise (for a
single Acquisition, Review, or Server)
McAfee Active VirusScan SMB Edition (for
Acquisition, Review, or Server).
McAfee ProtectionPilot™1.0.1, 1.5
McAfee ViruScan Enterprise8.5i
Norton AntiVirus Edition (for Acquisition
and Review) (Norton AntiVirus software
must be the boxed copy. Do not download
the software from internet.)
Norton Antivirus™ (Norton AntiVirus
software must be the boxed copy. Do not
download the software from internet.)
Symantec EndPoint Protection11.0.4
Symantec AntiVirus (for Acquisition,
Review, or Server)
Symantec AntiVirus Corporate Edition (for
Acquisition, Review, or Server)
r8.1
8.7i
8.0
2005
2007
9.0
10.2
Trend Micro OfficeScan Client/Server
Edition
Trend Micro OfficeScan Client/Server
Edition (for Acquisition, Review, or Server)
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-17
8.0 + SP1
8.0 (English)
7.3 (non-English and non-Swedish)
Page 32
Introduction
3rd Party: PedCath
LanguagePedCath Version
Englishv7.6.7
Frenchv7.5.5
Swedish/German (see note)v7.6.5
Italianv7.5.5
Spanishv7.5.5
Japanesev7.4.3J
Chinese (see note)v7.4.3J
NOTE
Chinese systems will use the Japanese version of PedCath. Swedish
and German systems will use the English version of PedCath.
1-18Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 33
Service Information
Service Requirements
Follow the service requirements listed below:
Equipment servicing should be performed by GE Healthcare
Any unauthorized attempt to repair equipment under warranty voids
It is the user’s responsibility to report the need for service to GE
Failure on the part of the responsible individual, hospital, or
Regular maintenance, irrespective of usage, is essential to ensure
Introduction
authorized service personnel only.
that warranty.
Healthcare or one of their authorized agents.
institution using this equipment to implement a satisfactory
maintenance schedule may cause undue equipment failure and
possible health hazards.
that the system is always functional when required.
Product Identification
Every system has a unique serial number for identification. The serial
number appears on the product labels which are located on the top of the
CPU, the back of the Integrated Electronics Box (IEB), the right side of
the RAC 4A, the back of the TRAM and the back of the Amplifier.
2
1
Identification Tag
ItemNameDescription
1Manufacture DateDate of manufacture in YYYY-MM format.
2Part Number of ProductGE Healthcare part number for product.
3Product Code DescriptionIdentifier for type of product.
4Serial NumberUnique serial number for identification.
Encoded as barcode below displayed
number.
3
4
5
5Manufacturer InformationName and address of device
manufacturer.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-19
Page 34
Introduction
XXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXX
XXXXXXXXX
XX
XXXXXXXXXXX
XXXXXXXXXXXXX
A
B
C
D
E
F
G
H
I
AA5 XXXXXX XXXX
Old Identification Tags
Systems manufactured earlier may have one of the two following styles
of identification labels instead.
Identification Tag
ItemName
AName of Device
BManufacturer
CSerial Number
D3-digit Product Code
EYear Manufactured
FFiscal Week
GProduct Sequence Number
HManufacturing Site
IMiscellaneous
1-20Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 35
Identification Tag
XXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXX
XXXXXXXXX
XX
XXXXXXXXXXX
XXXXXXXXX
A
B
C
D
E
F
G
H
A X XX XXXX XX
ItemName
Introduction
AName of Device
BManufacturer
CSerial Number
DFacility Code
EProduct Sequence Number
FProduct Code
GYear Manufactured
HMonth Manufactured
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-21
Page 36
Introduction
Required Tools and Supplies
The following is a list of tools needed to service the Mac-Lab/CardioLab
system:
Adjustable wrench with a 1-inch capacity
Phillips screwdriver
Needle nose pliers
Pliers
Socket wrench (3/16 inch to 1 inch size bits)
Phillips head screwdriver (small and large sizes)
Flat head screwdriver (small and large sizes)
Cutting pliers
Power outlet ground tester
Grounding strap
Simulator with ECG, pressure, cardiac output, and temperature
outputs
Cables to connect simulator to ECG inputs of amplifier and TRAM
Cables to connect the simulator to the catheter input modules (2 mm
shielded)
Adapter cable to connect the simulator to the amplifier and TRAM
pressure inputs
Cables to connect the simulator to the TRAM, cardiac output, and
temperature inputs
Power Input Requirements
If the equipment is installed in the USA and uses 240V rather than
120V, the source must be a center-tapped, 240V single phase circuit
(120V, 20A; 240V, 10A).
1-22Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 37
Safety Information
Follow all of the instructions and warnings included in this manual and
save these instructions for later use. However, these instructions in no
way supersede current medical practices regarding patient care and
safety.
Electromagnetic Compatibility and Interference
This catheterization system meets the requirements of the European
Medical Device Directive, and by reference, the requirements for
electromagnetic compatibility specified by EN 60601-1-2. This equipment
is not likely to interfere with other equipment, except with certain types
of rate adaptive pacemakers that rely on bioelectric impedance
measurement (BIM) technology. Also, interference with the operation of
this equipment is not likely to result from proximity to general
equipment.
This equipment meets or exceeds Class A immunity levels. However,
avoid the use of strong intentional radio transmitters in the immediate
vicinity, including cellular telephones. The use of strong unintentional
radiators, such as electrosurgical equipment, may cause some level of
interference. GE Healthcare offers optional ECG patient cables with
electrosurgical noise filtering, which reduces the level of interference on
ECG traces.
Introduction
Exception
The Mac-Lab/CardioLab system and CardioLab Plus II Amplifier meets
the requirements of IEC 60601-1-2 Medical Electrical Equipment, Part 1:
General Requirements for Safety, 2. Collateral Standard:
Electromagnetic compatibility - Requirements and tests.
If using electrosurgical or RF ablation equipment in conjunction with
this equipment, select an ESU filtered cable. For more information about
management of the equipment in the clinical environment (from an
electromagnetic compatibility perspective) consult the Association for the
Advancement of Medical Instrumentation's technical information report
Guidance on Electromagnetic Compatibility of Medical Devices for
Clinical/Biomedical Engineers.
If operating under the conditions defined in EMC Standard EN 60601-12 (Radiated Immunity 3V/m), field strengths less than 1V/m may cause
waveform distortions and erroneous numeric data at various
electromagnetic interference (EMI) frequencies.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-23
Page 38
Introduction
Recommendations:
Review the AAMI EMC Committee technical information report (TIR-18)
titled Guidance on electromagnetic compatibility of medical devices for
clinical / biomedical engineers - Part 1: Radiated radio-frequency
electromagnetic energy. This TIR provides a means to evaluate and
manage the EMI environment in the hospital.
The following actions can be taken:
managing (increasing) distance between sources of EMI and
susceptible devices
managing (removing) devices that are highly susceptible to EMI
lower power from internal EMI sources under hospital control (i.e.
paging systems)
labeling devices susceptible to EMI
educating staff (nurses and doctors) to be aware of, and to recognize,
potential EMI related problems
EU Battery Directive 2006/66/EC
The Mac-Lab/CardioLab product conforms to the EU Battery Directive
2006/66/EC.
The Uninterruptable Power Supply (UPS) supplied with the IEB of the
Mac-Lab/CardioLab contains lead/acid storage batteries. These batteries
are a permanent part of the UPS assembly and are not intended to be
removed or changed. Attempting to remove or change these batteries
may result in an unsafe condition.
The separate collection symbol is affixed to a battery, or its packaging,
to advise you that the battery must be recycled or disposed of in
accordance with local or country laws. The letters below the separate
collection symbol indicate whether certain elements (Pb=Lead,
Cd=Cadmium, Hg=Mercury) are contained in the battery. To minimize
potential effects on the environment and human health, it is important
that all marked batteries that you remove from the product are
properly recycled or disposed. Information on the potential effects on
the environment and human health of the substances used in batteries
is available at the urls below:
1-24Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 39
Responsibility of the Manufacturer
GE Healthcare is responsible for the effects of safety, reliability, and
performance only if:
Installation operations, extensions, readjustments, modifications, or
repairs are performed by persons authorized by GE Healthcare.
The equipment is used in accordance with the instructions provided.
The user attaches only accessory equipment recommended for
interaction with the amplifier.
General
This system is intended for use under the direct supervision of a licensed
health care practitioner.
The Mac-Lab/CardioLab system, including the CardioLab Plus II
Amplifier, maintains leakage currents below medical limits with an
isolation transformer which is incorporated into the design of the
integrated electronics box (IEB). The system power is supplied by a
dedicated power supply on an isolated circuit for proper behavior of the
system.
Introduction
To ensure patient safety, use only parts and accessories manufactured or
recommended by GE Healthcare. Parts and accessories used must meet
the requirements for the applicable EN 60601 series safety standards,
and the system configuration must meet the requirements of the EN
60601-1 medical electrical systems standard. Contact GE Healthcare
before connecting any device to this system that is not recommended for
use in this manual.
If the equipment is installed in the USA and uses 240V rather than
120V, the source must be a center-tapped, 240V single-phase circuit.
To ensure patient safety, use only parts and accessories manufactured or
recommended by GE Healthcare.
This equipment should only be connected as specified by GE Healthcare.
Installation of the equipment should only be performed by qualified GE
Healthcare service representatives.
Contact GE Healthcare for information before connecting any device to
this equipment that is not recommended in this manual. Only cables and
accessory equipment purchased from GE Healthcare should be used in
conjunction with the system.
It is important that all of the instructions in this manual be followed.
However, these instructions in no way supersede current medical
practices regarding patient care and safety.
Refer servicing of the equipment to GE Healthcare authorized service
personnel. Any attempt to repair equipment under warranty will void
that warranty.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-25
Page 40
Introduction
Equipment Symbols
Equipment SymbolsDescriptions
Failure on the part of responsible individuals or institutions to
implement a satisfactory equipment maintenance schedule may lead to
undue equipment failure and possible health hazards.
The use of ACCESSORY equipment not complying with the equivalent
safety requirements of this equipment may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice
includes:
Use of the accessory in the Patient Vicinity.
Evidence that the safety certification of the Accessory is performed in
accordance with the appropriate EN 60601-1 and EN 60601-1
harmonized national standard.
The CardioLab II Plus Amplifier is a component/accessory of a Class II
medical device.
CE Marking
Underwriters Laboratories, Inc.
Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, mechanical
and other specified hazards, only in accordance with UL 2601-1, CAN/CSA C22.2 No.
601.1, IEC 60601-1 and IEC 60601-1-1.
Power Ratings Chart:
V~ = V = Voltage, ~ = Alternating Current
I = Current
Fuse =
Equipotential Ground Point
A nonessential electrical connection which may be used to connect the chassis to other
equipment to achieve the same potential.
1-26Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 41
Equipment SymbolsDescriptions
Type CF Equipment that is Defibrillator Proof
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and
internal application to the patient excluding direct cardiac application. The “paddles” outside
the box indicate that the part is defibrillator proof.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the
specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a
higher degree of protection against electric shock than that provided by Type B applied
parts.
Type CF Equipment
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and
internal application to the patient excluding direct cardiac application.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the
specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a
higher degree of protection against electric shock than that provided by Type B applied
parts.
Type BF Equipment that is Defibrillator Proof
Introduction
Type BF Applied Part that is defibrillator proof: Isolated (floating) applied part suitable for
intentional external and internal application to the patient excluding direct cardiac
application. “Paddles outside the box indicate the applied part is defibrillator proof.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the
specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a
higher degree of protection against electric shock than that provided by Type B applied
parts.
Atmospheric Pressure Limits
Temperature Limits
Humidity Limits
Attention
Pay attention to the documents delivered with the equipment.
Input Symbol
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-27
Page 42
Introduction
Equipment SymbolsDescriptions
General Warning
Manufacturer
Date of Manufacture
Consult Instructions for Use
WARNING: SHOCK HAZARD
Remove power cord from the mains source by grasping the plug. Do not pull on the cable.
This may cause cable damage, which could result in loss of protective earthing or exposure
to mains voltage.
1-28Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 43
Introduction
Compliance Information
Mac-Lab/CardioLab System
Declaration - Electromagnetic Emissions
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Mac-Lab / CardioLab System should assure that it is used in such an environment.
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11Class AThe Mac-Lab / CardioLab System is suitable
Harmonic emissions IEC 61000-3-2Class A
Voltage fluctuations/Flicker emissions IEC
61000-3-3
Complies
for use in all establishments other than
domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601
Test Level
± 6 kV contact
± 8 kV air
± 2kV for power
supply lines
(signal lines
exempt from
testing, length is
less than 3 m)
± 1kV differential
mode ± 2kV
common mode
Compliance
Level
± 6 kV contact
±8 kV air
± 2kV for power
supply lines
± 1kV differential
mode ±2kV
common mode
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-29
Page 44
Introduction
Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity Test
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601
Test Level
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for
0.5 cycle
Compliance
Level
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for
0.5 cycle
Electromagnetic Environment - Guidance
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Mac-Lab / CardioLab System requires
continued operation during power mains interruptions, it is recommended
that the Mac-Lab / CardioLab System be powered from an Uninterruptable
power supply or a battery.
40% 100-120Vac
40% 200-240Vac
(>60% dip) for 5
cycles
70% 100-120Vac
70% 200-240Vac
(>30% dip) for 25
cycles
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for 5
seconds
40% 100-120Vac
40% 200-240Vac
(>60% dip) for 5
cycles
70% 100-120Vac
70% 200-240Vac
(>30% dip) for 25
cycles
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for 5
seconds
3 A/m3 A/mPower frequency magnetic fields should be at levels characteristic of a
typical location typical commercial or hospital environment.
1-30Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 45
Introduction
Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity Test
Conducted RF
emissions
IEC 61000-4-6
Radiated RF
emissions
IEC 61000-4-3
IEC 60601
Test Level
3 Vrms
105kHz to 80
MHz with 2 Hz
modulation
3V/m80 MHz to
2.5 GHz
Compliance
Level
Portable and mobile RF communications equipment should be used no
closer to any part of the Mac-Lab / CardioLab System, including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance (d)
3 Vrmsd = 1.2 P
3 V/md = 1.2 P
d = 2.3 P
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters(m).Field Strengths from fixed RF
transmitters, as determined by a electromagnetic site survey
less than the compliance level in each frequency range
occur in the vicinity of equipment marked with the following symbol:
Electromagnetic Environment - Guidance
2
. Interference may
1
, should be
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflected from
structures, objects and people.
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the Mac-Lab / CardioLab System is used
exceeds the applicable RF compliance level above the Mac-Lab / CardioLab System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Mac-Lab / CardioLab System.
2
Over the frequency range 150 kHz to 80 MHz. Field strengths should be less than 3 V/m
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual1-31
Page 46
Introduction
d1.2 P=
d1.2 P=
d2.3 P=
Recommended Separation Distances Between
Portable and Mobile RF Communications Equipment and the Mac-Lab/CardioLab system
The Mac-Lab/CardioLab system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Mac-Lab/CardioLab system can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the Mac-Lab/CardioLab system as
recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output
Power of Transmitter
Separation Distance According to Frequency of Transmitter
m
W
150 kHz to 80 MHz80 MHz to 800 MHz800 MHz to 2.5 GHz
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE
At 80 MHz and 800 MHz, separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflected from
structures, objects and people.
1-32Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 47
2Equipment Overview
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-1
Page 48
Equipment Overview
Typical System Configuration
NOTE
For configuration, component and connection information for HP
xw8200 and HP xw8000 computers, refer to the Mac-Lab/CardioLab/
Centricity Cardiology INW Service Manual v6.5 (PN 2021423-022).
This system may consist of either Mac-Lab (hemodynamic), CardioLab
(electrophysiologic), or ComboLab (both systems).
The system may be standalone or connected to a network. Network
installations include a Centricity Cardiology INW server. In this manual,
INW refers to Centricity Cardiology INW. It may be connected to an
optional CVIS server.
The system may also include a CardioLab II Plus Amplifier, also referred
to as CLab II Plus.
The following block diagram shows a typical setup.
Dashed lines indicate CardioLab/ComboLab only
Stimulator
Keyboard
Workstation
Control Room
Speakers
Mouse
Real-Time
Monitor
HP
CPU
UPS
Printer
Review
Monitor
IEB
Image
Monitor
3" Floor Conduit
3" Floor Conduit
3" Ceiling Conduit
Amplifier
CIM
Monitor
Boom
Catheter
Personality
Module
RF
Generator
Remote
Real-Time
Monitor
TRAM
RAC4A
Remote
Review
Monitor
Bed
Procedure Room
INW
Server
IT Control Room
Network
Switch
CVIS
To Hospital Network
2-2Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 49
Control Room Components
Acquisition Computer
USB Mouse
Keyboard
Barcode scanner
Integrated Electronics Box (IEB): The IEB provides power to
local system components, signal pass through video and
communications.
Uninterruptible Power Supply (UPS): The UPS provides
emergency power to the HP computer.
Monitors: Two monitors (Real-Time and Review) are standard. A
third Image monitor is optional.
Local speakers: Provide output for ECG QRS tone and optional
audible indicators.
Laser Printer: Prints images or reports.
Cardiac Stimulator: Provides direct cardiac stimulation to assist
the physician during an EP case.
Equipment Overview
Procedure Room Components
Monitors: One remote monitor is standard (and is software
switchable between Real-Time and Review windows). A second
monitor is optional.
Remote speakers: Provide output for ECG QRS tone and optional
audible indicators in the procedure room.
TRAMNet: The TRAMNet gathers patient data for the Mac-Lab
system. The TRAMNet consists of a Remote Acquisition Case (RAC4A) and a Transport Remote Acquisition Module (TRAM).
CardioLab II Plus Amplifier: The Amplifier gathers data for both
the ComboLab and CardioLab systems.
CARTO™ System (Patient Interface Unit): Provides 3D
electroanatomical cardiac maps to the CardioLab.
Analog Output Box: Distributes Heart rate and blood pressure
signals from the TRAMNet and Amplifier to other equipment.
Review Workstation: Client installed in physician office or outside
lab. Provides post-procedure analysis capabilities such as generating
reports.
Nurses Workstation: Client used during the study. Provides the
operator with the ability to participate in an active study and
perform actions such as document administered medications and
supplies consumed.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-3
Page 50
Equipment Overview
DMS Express Server
(Interface Server)
INW Server
Review
Workstation
Mac-LabCardioLab
Hospital
Enterprise
OR
IT Control Room Equipment
Network Switch
Centricity Cardiology INW Server
CVIS Server
Network Connections
2-4Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 51
System Hardware
WARNING
SHOCK HAZARD — Use appropriate Lock Out/Tag Out
procedures before performing any service on potentially
energized equipment.
Acquisition Workstation System Components
HP Computer
The HP computer is used for either Acquisition systems or Review
workstations and has the following components:
Primary Hard Drive with two partitions:
C: Application partition
D: Data partition
Optional Magneto-Optical drive (Drive E:).
NOTE
The optional Magneto-Optical drive is only available on HP
xw8400 computers.
Equipment Overview
DVD-RAM drive (Drive F:).
NOTE
The drive used to store application studies is determined by the
file paths set in under System Settings in the main application
menu.
Secondary Hard Drive (Drive G:).
Four standard expansion cards (plus optional cards):
HP xw8400: Matrox Parhelia video board provides video out for
the Real-Time (Video 1) and Review (Video 2) monitors.
HP z600: Matrox M9140 video board provides video out for the
I/O: Two network connection jacks, one serial port (for connection to
ablation devices)
Uses internal serial connection from motherboard:
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-13
Page 60
Equipment Overview
1
2
3
4
5
6
HP z600 Matrox Quad Board M9140 LP Video Board
Part Number: 2049434-001
Location: Slot 2
Function: Video out for Monitor 1, Monitor 2 and Monitor 3 (optional)
I/O: One proprietary connector. Provides four DVI/VGA output ports
through a custom connector.
2-14Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 61
Equipment Overview
HP z600 Matrox Odyssey eA Board (Optional)
1
2
3
4
5
6
Part Number: 2049435-001
Location: Slot 3
Function: Image capture for X-Ray or other imaging systems
I/O: Two DVI connectors
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-15
Page 62
Equipment Overview
1
2
3
4
5
6
HP z600 TRAMNet to Ethernet Adapter board
Part Number: 2009937-003
Location: Slot 5
Function: Provides communication with the TRAM in the RAC 4A,
and converts signals from the TRAM to the ethernet.
I/O: Network Interface Port and 9-Pin D connector
2-16Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 63
Equipment Overview
1
2
3
4
5
6
HP z600 PCI Communication Board
Part Number: 2014935-002
Location: Slot 6
Functions:
Communicates with the CardioLab II Plus Amplifier via the fiber
optic connectors
Analog and Digital communication with the IEB via the D-
Connector
Provides the video switching signal for the Switched Video
output on the IEB via the D-Connector
I/O: D-Connector to IEB, Fiber Optic Jacks to the Amplifier
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-17
Page 64
Equipment Overview
HP xw8400 Installed Boards
HP xw8400 PCI Communication Board
Part Number: 2014935-002
Location: Slot 1
Functions:
Communicates with the CardioLab II Plus Amplifier via the fiber
optic connectors
Analog and Digital communication with the IEB via the D-
Connector
Provides the video switching signal for the Switched Video
output on the IEB via the D-Connector
I/O: D-Connector to IEB, Fiber Optic Jacks to the Amplifier
2-18Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 65
Equipment Overview
HP xw8400 ATI FIREGL V5200 256MB PCI-E Video Board (Optional)
Part Number: 2026189-002
Location: Slot 2
Function: Video out for optional (video 3) monitor
I/O: Two DVI connectors (only the left one is used)
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-19
Page 66
Equipment Overview
HP xw8400 Dual NIC Adapter Board
Part Number: 2017795-002
Location: Slot 3
Function: Network connection card
I/O: Two network connection jacks
2-20Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 67
Equipment Overview
HP xw8400 PCIX 2 SCSI Card
NOTE
This board is optional with the MO upgrade kit (available only for
2034493-002 – 2034493-004 Acquisition configurations).
Part Number: 2038632-001
Location: Slot 4
Function: Provides SCSI cable connection internally to an optional
MO Drive.
I/O: Two SCSI connectors (only internal SCSI is used)
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-21
Page 68
Equipment Overview
HP xw8400 Matrox Parhelia PCI 256 MB Video Card
Part Number: 2026188-001
Location: Slot 5 (PCI slot)
Function: Provides video out for the Real-Time (Video 1) and Review
(Video 2) monitors. This board supports resolutions of 1280x1024 for
Mac-Lab, 1600x1200 for CardioLab.
I/O: Two DVI connectors
Included adaptors: DVI to HD15 Video Adapter (2021697-001).
2-22Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 69
HP xw8400 Matrox Odyssey XA Board (Optional)
Equipment Overview
Part Number: 2030688-001
Location: Slot 6
Function: Image capture for X-Ray or other imaging systems
I/O: Two DVI connectors
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-23
Page 70
Equipment Overview
HP xw8400 TRAMNet to Ethernet Adapter board
Part Number: 2009937-002
Location: Slot 7
Function: Provides communication with the TRAM in the RAC 4A,
and converts signals from the TRAM to the ethernet.
I/O: Network Interface Port and 9-Pin D connector
2-24Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 71
Integrated Electronics Box (IEB)
The IEB provides power to local system components and signal pass
through video and communications. Three versions of the IEB are
available to handle different line voltages: 100, 120, and 240 VAC.
The IEB consists of the following components:
Video Multiplexer (VID MUX) board to distribute video signals from
the video cards in the HP computer to the monitors.
I/O board to provide signal pass through for other system
components.
Isolation transformer:
Conforms to IEC 60601-1
2.0 kVA medical grade high-isolation transformation
Provides either 120 VAC (for 100 and 120 VAC systems) or 240
Equipment Overview
VAC. The primary and secondary windings are wired in parallel
for 120 VAC and in series for 240 VAC.
CAUTION
CONNECTIONS — If the transformer must be
disconnected for any reason, ensure the connections are
correct.
UPS (either 100, 120, or 240 VAC) and Power Distribution panel
provide conditioned power to local system components.
NOTE
The UPS is external to the IEB and provides uninterrupted
power to the HP computer. The monitors and other peripherials
will not remain on if power is lost.
Depending on which IEB is installed, a modem and modem
power supply for dial-up communication may be present. The
power supply accepts any line voltage and converts it into 9 VDC.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-25
Page 72
Equipment Overview
External Components and Connections
LabelPart NumberDescription
12003681-003VID MUX board
22003680-001I/O Board
3N/APower Connections. The bottom right connector is for the HP
computer and the four vertical connectors on the right and left
are for peripheral equipment.
4N/AThermal Overcurrent Circuit Breaker. This breaker protects all
internal components and external devices connected to the
IEB.
5N/AMain Power Switch. This switch controls power to all internal
components and external devices except the HP computer.
Both the Main Power switch on the front of the IEB and the Thermal
Overload circuit breaker on the back must be on for the system to power
up.
2-26Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 73
Video Multiplexer (VID/MUX) Board
LabelDescription
Equipment Overview
1Switched video out (Video 1 or 2). This is used when the facility has only
one remote monitor. To switch monitors:
Mac-Lab: Go to the main menu and select Windows > Physician Video
CardioLab: Use the dropdown box just under the main menu.
2Video 1 (Real-time) out to local monitor (bottom) and remote monitor (top)
3Video 2 (Review) out to local monitor (bottom) and remote (top)
4Video 3 (CardioImage) out to local monitor (bottom) and remote (top)
5(L to R) Video 1, Video 2, Video 3 in from HP Computer Matrox AGP Video
board (Vid 1 and Vid 2) and the Matrox PCI Video board (Vid 3)
The VID/MUX board amplifies (±5V 1.25A from the internal IEB DC
power supply. See “Internal Components and Connections” on page 2-31)
and distributes the video signals from the HP Computer (See“HP xw8400
Video Connections” on page 2-57 or “HP z600 Video Connections” on
page 2-56).
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-27
Page 74
Equipment Overview
I/O Board
LabelDescription
1Analog I/0 Communications. Connects to the HP Computer PCI
Comms board.
2Analog in from another system (Fluoro, and so on).
3Connection to the Analog Output Box.
4Analog out (Stimulator, and so on).
5I/O TRAM (includes 16.5VDC power).
6Output to HP Computer TRAMNet to Ethernet to Adapter board.
Refer to the following cabling diagrams:
“Communications Connections” on page 2-58.
“TRAM/Amplifier Connections” on page 2-59.
2-28Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 75
Equipment Overview
Power Connections
LabelDescription
1 and 5AC out to local equipment (6A @ 120 VAC max.; 3A @ 240 VAC max.)
2Thermal Overcurrent circuit breaker. Protects all internal components
and external devices connected to the IEB.
3AC main power in (15A @ 120 VAC max; 7A, @ 240 VAC max.)
4AC power out to UPS (2A @ 120 VAC max; 1A, @ 240 VAC max.)
Refer to the cabling diagram, “Power Connections” on page 2-55.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-29
Page 76
Equipment Overview
Disassembly
Open Case
1. Remove 7 side-panel screws.
2.Lift the side panel out.
Remove Connection Boards
1. Open the case.
2. Remove the board screws.
3. Disconnect interior cables.
WARNING
GROUNDING HAZARD — Electrostatic discharge
(ESD) can damage electronic components. Be sure you are
properly grounded before opening any equipment.
2-30Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 77
Internal Components and Connections
Equipment Overview
LabelPart NumberDescription
12003681-003
2003680-001
22013535-001Power Supply: Provides ±5 VDC 1.25A for the Vid/
VID/MUX Board Assembly
I/O DB Connect Board Assembly
MUX board and +16.5 VDC for the RAC-4A.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-31
Page 78
Equipment Overview
Uninterruptible Power Supply (UPS)
The UPS is designed to maintain power to the computer during a power
outage.
NOTE
The UPS is external to the IEB and provides uninterrupted power to
the HP computer. The monitors and other peripherials will not
remain on if power is lost.
Part NumberDescription
2034077-001UPS Powercom, 120V, 50/60 Hz, 1500 VA, BNT
2034078-001UPS Powercom, 230V, 50/60 Hz, 1500 VA, BNT
2034079-001UPS Powercom, 100V, 50/60 Hz, 1500 VA, BNT
2-32Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 79
GE Client Workstation
Equipment Overview
There are several configuration options for use of the Review
workstations in a networked system environment.
Nursing Workstation software allows the user to join a Mac-Lab or
CardioLab case in progress to chart and to document within the case
log.
Remote Review Workstation software allows the user to review a
Mac-Lab or CardioLab case in full disclosure in order to compose or
complete reports or coronary diagrams or to review patient
waveforms and measurements.
Pre/Post Procedure Workstation software allows the user to start
or continue a Mac-Lab or CardioLab case in order to chart and to
document within the case log.
NOTE
Automatic vitals logging is also available with the optional
DASH and Solar monitor interface.
NOTE
When used in a patient care area, the hardware must meet
IEC60601-1. This can be achieved through the use of a medical
grade computer, separate isolation transformer or other means
as determined by appropriate hospital personal.
The GE Client Review computer has the following components:
Primary Hard Drive with two partitions:
C: Application partition
D: Data partition
DVD-RAM drive (Drive E:).
NOTE
The drive used to store application studies is determined by the
file paths set in under System Settings in the main application
menu.
Secondary Hard Drive (Drive G:).
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-33
Page 80
Equipment Overview
1
2
3
4
5
External Components and Connections for the GE Client Workstation
LabelPart NumberDescription
1N/APower and Reset Switches
2N/ADVD Drive
3N/AUSB and Audio connections (behind cover)
4N/AAC Power
5VariousKeyboard (dependent on language)
6N/ACom Port for DASH/Solar connection
7N/AVideo Input
8N/AUSB 2-Button Optical Mouse with Wheel
9N/ARJ-45 Port to Network
2-34Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 81
Disassembly
1
1
Equipment Overview
WARNING
GROUNDING HAZARD — Electrostatic discharge
(ESD) can damage electronic components. Be sure you are
properly grounded before opening any equipment.
Remove Case Side Panel
1.Loosen thumbscrews.
2.Pull side panel off.
Remove Hard Drive
1.Remove screws for hard
1
1
drive.
2.Slide drive out and
disconnect cables.
Remove DVD Drive
1.Disconnect cables from
DVD drive.
2.Remove DVD drive cage
2
screws.
3.Remove DVD drive cage
from case.
4.Remove screws for DVD
drive.
5.Remove DVD drive from
cage.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-35
Page 82
Equipment Overview
Mobile Workstand
Refer to the instructions included with the mobile workstand for details
on how to place the GE hardware on an assembled GE mobile workstand
(PN 2008203-001).
2-36Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Review Workstation Computer
Page 83
Monitors
TRAMNet
Equipment Overview
The standard workstation comes with two monitors and may have a
third optional monitor. The monitors include:
A Real-Time monitor to display current patient information. It is
usually located on the left.
A Review monitor to display past patient information.
An optional Image monitor (used with the CardioImage option)
displays live or frozen video images.
A single remote monitor is standard and receives input from the
switched video out (Video 1 or 2) jack on the IEB (see “External
Components and Connections” on page 2-26).
To switch video for the Mac-Lab go to the main menu and select
Windows > Physician Video.
To switch video for the CardioLab use the drop-down box just under
the main menu.
LabelDescription
1TRAM Module
2To Analog Out Box
3To IEB TRAM Connector
The TRAMNet consists of a Remote Acquisition Case (RAC-4A) and one
or more TRAM modules. The TRAMNet is located in the procedure room
near the procedure table.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-37
Page 84
Equipment Overview
The TRAM module receives patient inputs and then transmits signal
data to the IEB and other equipment through the Analog Output Box
(refer to the cabling diagrams in “TRAM/Amplifier Connections” on
page 2-59.
TRAM Module Inputs:
Electrocardiogram (ECG) - green connector (5 or 10 leads)
Thermal Dilution Cardiac Output (TDCO) - brown connector
Invasive Blood Pressure (IBP) - 3 red connectors
Non-Invasive Blood Pressure (NIBP) - black connector
Pulse Oximetry (SpO
) - blue connector
2
2-38Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 85
Outputs: Analog output connector.
The following TRAM modules are supported by the Mac-Lab system:
200SL
400/450SL (Ohmeda, Nellcor)
451/451N/451M (Ohmeda, Nellcor, Massimo)
600/650SL (Ohmeda, Nellcor)
The following software versions are supported by the Mac-Lab system:
7A, 7B, 7C, 7D, 7E, 7F, 7H
8A, 8B, 8C, 8D
9A, 9B
10A, 10B
11A, 11B, 11C
CardioLab II Plus Amplifier
The CardioLab II Plus Amplifier is designed to amplify various types of
cardiac signals that are acquired from a patient and transmit them back
to the CardioLab system. These signals include 12 Lead ECG,
intracardiac and pressure signals.
Equipment Overview
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-39
Page 86
Equipment Overview
Front View
Front View 64/32 Channels- CardioLab II Plus Amplifier
NameDescription
Pressure InputsFour pressure inputs will each accept a pressure transducer that is calibrated to 5uV/V/mmHg. See the Accessory
Parts List for GE Healthcare Systems Information Technologies recommended pressure transducers.
ECG InputThe ECG input connector will accept a 10 lead ECG input cable.
Stimulator InputThe stimulator input will accept a stimulator input cable. Use the 3 ft, 15 ft or 30 ft Touchproof Stimulator Cables
(part numbers, 301-00204-03, 301-00204-15, 301-00204-30, respectively) with the CardioLab II Plus Amplifier.
Catheter InputsUp to three catheter inputs will each accept a 32 pole catheter input module. Use the CardioLab II Plus Catheter
Input Modules (part number 301-00202-08) with the CardioLab II Plus Amplifier.
Auxiliary InputsThe auxiliary input will accept the Auxiliary Input Cable (part number 301-00205-08).
Power IndicatorThe power indicator is on as long as the CardioLab II Plus Amplifier is turned on.
Communication
Indicators
Three communication indicators show the status of the communication between the acquisition computer and the
CardioLab II Plus Amplifier.
When the XMT indicator is on, the CardioLab II Plus Amplifier is transmitting data to the acquisition computer.
The XMT indicator should remain on as long as the power indicator is on.
When the RCV CMD indicator is on, the CardioLab II Plus Amplifier is receiving data from the acquisition
computer. The RCV CMD indicator will turn on as long as an instruction is being received by the amplifier.
When the RCV ERR indictor is on, the CardioLab II Plus Amplifier may not be receiving data from the
acquisition computer in a valid format or there may be a problem with the fiber optic cable.
2-40Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 87
Front View 128/96 Channels- CardioLab II Plus Amplifier
Equipment Overview
NameDescription
Pressure InputsFour pressure inputs will each accept a pressure transducer that is calibrated to 5uV/V/mmHg. See the Accessory
Parts List for GE Healthcare Systems Information Technologies recommended pressure transducers.
ECG InputThe ECG input connector will accept a 10 lead ECG input cable.
Stimulator InputThe stimulator input will accept a stimulator input cable. Use the 3 ft, 15 ft or 30 ft Touchproof Stimulator Cables
(part numbers, 301-00204-03, 301-00204-15, 301-00204-30, respectively) with the CardioLab II Plus Amplifier.
Catheter InputsUp to seven catheter inputs will each accept a 32 pole catheter input module. Use the CardioLab II Plus Catheter
Input Modules (part number 301-00202-08) with the CardioLab II Plus Amplifier.
Auxiliary InputsThe auxiliary input will accept the Auxiliary Input Cable (part number 301-00205-08).
Power IndicatorThe power indicator is on as long as the CardioLab II Plus Amplifier is turned on.
Communication
Indicators
Three communication indicators show the status of the communication between the acquisition computer and the
CardioLab II Plus Amplifier.
When the XMT indicator is on, the CardioLab II Plus Amplifier is transmitting data to the acquisition computer.
The XMT indicator should remain on as long as the power indicator is on.
When the RCV CMD indicator is on, the CardioLab II Plus Amplifier is receiving data from the acquisition
computer. The RCV CMD indicator will turn on as long as an instruction is being received by the amplifier.
When the RCV ERR indictor is on, the CardioLab II Plus Amplifier may not be receiving data from the
acquisition computer in a valid format or there may be a problem with the fiber optic cable.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-41
Page 88
Equipment Overview
Back View
Back View - CardioLab II Plus Amplifier
NameDescription
FDDI Connection
to Computer
Power InputThe GE-supplied power cord connects to the Power Input.
Equipotentiality
Terminal
This connector will accept a FDDI type fiber optic cable. The fiber optic cable allows for two-way communication
between the amplifier and the CardioLab system.
The Equipotentiality Terminal is connected to earth ground.
2-42Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 89
CardioLab II Plus Catheter Input Module
Equipment Overview
The equivalent to the Stimulus Inputs 1-4 on the Stimulator Input Cable
are labelled on the CardioLab II Plus Amplifier as Stimulus Inputs A1,
A2, B1 and B2.
The four stimulus inputs on the Stimulator Input Cable connect to the
stimulator.
Each Catheter Input Module is broken down into two groups of 16 poles
each. For example, Block A is broken into A (1-16) and A (17-32).
Each pacing channel, requires one positive and one negative pole. In
order to maximize the number of pacing channels, only one positive and
one negative designation should be made for each group.
For detailed instructions on how to designate pacing channels and enable
pacing for particular sites, refer to the on-line help in the software
application.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-43
Page 90
Equipment Overview
Accessory Parts List
GE Healthcare Systems Information Technologies recommends the
following accessory parts be used with the CardioLab II Plus Amplifier:
Accessory Parts List
Accessory NamePart Number
Abbott Transpac IV Disposable Transducer
with 4 ft extension Cable*
* This accessory is not stocked by GE Healthcare Systems Information Technologies.
Abbott #42582-08
3M Red Dot #2256
2-44Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 91
CARTO™ System (Patient Interface Unit)
1
LabelDescription
1CARTO XP PIU Output to Amplifier
Equipment Overview
CAUTION
EXCESSIVE SIGNAL NOISE — Using a CARTO XP
PIU earlier than version 10 with a crossover cable may
cause excessive signal noise.
If this excessive signal noise occurs, then discontinue
using the crossover cable until the PIU is replaced with a
later version. (The new PIU must be purchased from
Biosense Webster).
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-45
LabelDescription
1CARTO 3 PIU Output to Amplifier
Page 92
Equipment Overview
NOTE
For specific information related to the operation, troubleshooting, or
repair of the CARTO System, refer to the OEM documentation. For
information about the CardioLab/CARTO System interface, refer to
the CardioLab Operator manual.
The CARTO System acquires and displays electroanatomical 3D images
of heart chambers. This information is displayed on the CARTO System
and can be exported into the CardioLab application.
Two connections are required: one from the CARTO System Patient
Interface Unit to the CardioLab II Plus Amplifier and another from the
CARTO System computer to the HP computer. Refer to the cabling
diagram in “CARTO™XP System Connections” on page 2-65 or
“CARTO™ 3 System Connections” on page 2-66 for specific information.
To quickly view the status of the interface, check the CARTO System
icon at the bottom-right of the Real-Time window (a heart against a
grid). The color changes, depending on the system state:
Gray: Disconnected
Blue: Standby
Green: Passive Interface
Yellow: Active Interface
2-46Mac-Lab/CardioLab/Centricity Cardiology INW Service Manual2021423-503H
Page 93
Analog Output Box
Equipment Overview
LabelDescription
1Input from TRAMNet
2Input from Amplifier through the IEB
The analog output box distributes heart rate and blood pressure
signals to multiple outputs (refer to the connection diagram in
“TRAM/Amplifier Connections” on page 2-59. This may be done to
synch auxiliary equipment or display the information on a scope or
other device.
Patient information comes from either the TRAMNet or the Amplifier
(through the IEB) and is output through the output jacks on the top of
the box:
Channels 2-4 and 13-16 are switchable between the TRAMNet (Mac-
Lab) and CardioLab II Plus Amplifier (CardioLab).
The remaining channels are used exclusively by CardioLab.
Channels 1-12 may be distributed to up to 4 other devices using the
selector switch.
2021423-503HMac-Lab/CardioLab/Centricity Cardiology INW Service Manual2-47
Page 94
Equipment Overview
(
4
Network Components
Servers
Centricity Cardiology INW Server provides
Network repository for patient studies
Acquisition system and Review workstation access to patient
studies
CVIS Server provides
Central list management for staff, supplies, and so on.