The information in this manual only applies to Mac-Lab/CardioLab/ Centricity CardiologyINW Service
Manual software version 6.5. It does not apply to earlier software versions. Due to continuing product
innovation, specifications in this manual are subject to change without notice.
CARTO is a trademark of Biosense Webster, Inc.
Listed below are GE Medical Systems Information Technologies trademarks used in this document.
MAC-LAB®, CARDIOLAB®, CENTRICITY®, TRAM®, TRAMNET®, and TRAM-RAC® are trademarks of
GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
All other trademarks contained herein are the property of their respective owners.
J11 September 2013Updated to include installation and
troubleshooting procedures for the
Matrox Quad video board.
1-2Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 9
Purpose
Intended Audience
Introduction
This manual contains the instructions necessary to operate and service
the Mac-Lab/CardioLab system safely in accordance with its function
and intended use. These instructions include but are not limited to:
Describing fundamental controls and indicators.
Operating procedures.
Connecting and disconnecting detachable parts and accessories.
Troubleshooting and service procedures.
This manual is intended for the person who maintains and troubleshoots
this equipment.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual1-3
1-4Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 11
Statement of Indications for Use
The Mac-Lab/CardioLab/Centricity Cardiology INW Server software
applications are intended to be installed and run on an applicable ITE
compliant hardware platforms (minimum Hardware Essential
Performance be ITE compliant) which may not meet leakage and/or EMC
requirements of patient care area equipment. This means but is not
limited to the following (unless the medical electrical standard
[EN60601-1-1 and EN60601-1-2] is followed)
The Mac-Lab/CardioLab/Centricity Cardiology INW Server hardware
does not directly deliver energy to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server hardware
does not indirectly control energy delivered to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server hardware is
not directly connected to the Patient.
The Mac-Lab/CardioLab/Centricity Cardiology INW Server hardware is
not installed or used in the Patient vicinity.
Introduction
Mac-Lab System
Patient Vicinity summarized as follows:
A space, within a location intended for the examination and treatment of
patients, extending 6 ft (1.83 m) beyond the normal location of the bed,
chair, table, treadmill, or other device that supports the patient during
examination and treatment and extending vertically to 8 ft 2.4 in. (2.5 m)
above the floor.
Areas where healthcare staff members monitor patients and perform
charting and administrative tasks are not considered to be Patient
Vicinity areas.
The Mac-Lab system is intended for use under the direct supervision of a
licensed healthcare practitioner to monitor, calculate, or record
cardiovascular data from patients as they undergo cardiac
catheterization. Cardiovascular data may be manually entered or
acquired via interfaced GE Medical Systems Information Technologies
TRAM modules, MUSE cardiovascular system, and other interfaced
information systems. Data includes: ECG waveforms, heart rate, pulse
oximetry (SpO2), respiration rate, valve gradients and areas, cardiac
output, hemodynamic measurements, invasive and non-invasive blood
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual1-5
Page 12
Introduction
CardioLab System
pressure, procedural information, and optional intracardiac
electrocardiogram (IECG).
This is applicable to pediatric and adult patients requiring cardiac/
circulatory system catheterization. This is intended for use in
catheterization and related cardiovascular specialty laboratories.
The intended use of the CardioLab system is to acquire, filter, digitize,
amplify, display, and record electrical signals obtained during
electrophysiological studies and related procedures conducted in an
electrophysiological laboratory. Signal types acquired include ECG
signals, direct cardiac signals, and pressure recordings. Physiological
parameters such as diastolic, systolic, and mean blood pressure, heart
rate, and cycle length may be derived from the signal data, displayed and
recorded. The system allows the user to monitor the acquisition of data,
review the data, store the data, perform elementary caliper-type
measurements of the data, and generate reports on the data.
Additionally, the system may acquire, amplify, display, and record data
received from other medical devices typically used during these
procedures, such as imaging devices and RF generators.
The CardioLab EP system does not control the delivery of energy,
administer drugs, perform any life-supporting or life-sustaining
functions, or analyze physiological data or other data acquired during an
EP procedure. It does not transmit alarms or arrhythmias and does not
have arrhythmia detection capabilities.
1-6Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 13
Chapter Contents
Introduction
1 Introduction: This chapter describes the Service manual and
chapter contents. Also included are general safety requirements,
service requirements, equipment symbols, and serial number
identification information.
2 Equipment Overview: This chapter describes the equipment and its
technical characteristics, connector locations, and preparation for
use.
3 Maintenance: This chapter contains a preventive maintenance
schedule and cleaning guidelines.
4 Troubleshooting: This chapter contains troubleshooting
instructions and functional test directions.
5 Recovery: This chapter details procedures to restore a workstation
or server in the field after failure.
6 Parts List: This chapter contains a list of Field Replaceable parts.
A Abbreviations: This appendix lists all abbreviations used with the
system.
Security Information
Validated configurations consist of product scenario implementations
that are test against GE product failure and sample cases run to
simulate a real environment while conducting testing. GE provides
service and support based on validated configurations only and bears
responsibility for GE products only.
Non-Validated Software Applications
This is to detail the policy for non-validated software applications
applied to Invasive Cardiology monitoring products:
GE Mac-Lab Hemodynamic Systems
GE CardioLab Electrophysiology Systems
GE ComboLab Hemo and EP Systems
GE Centricity INW Server
Please remember that these products are a Class II device under the
FDA regulations and should be treated as secondary monitoring medical
devices.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual1-7
Page 14
Introduction
Policy
Our policy is that we do not recommend installations of any other third
party software applications (i.e. not mentioned above) on the products
above as they could affect performance, warranty and serviceability.
Examples of such might include other Anti-Virus programs, print
servers, interface engines, anti-spyware, anti-adware, viruses, malware,
etc. GE does not support the use or installation of these non-validated
applications or tools. If such programs are installed the customer bears
the risk of the product not functioning properly and any risks that are
introduced for patient safety. GE will not guarantee support for an
environment that is different than a validated implementation. In cases
where this is found and technical assistance is requested, GE will
provide support, by first eliminating the third party software on the
product and bringing it back to a known and validated configuration. If it
is found that the third party software is the cause of product failure, any
costs associated with this will be the responsibility of the hospital.
At GE Healthcare we understand the uniqueness of our customers and
pride ourselves in the personalized solutions we provide to help you in
your quest to care for patients. We hope this section will help you to
better understand our policy on non-validated software applications and
we hope you can work with us as we proceed in further developing our
products for your future benefit.
Validated Third Party Software
Currently validated and allowed third party software such as Enterprise
Back-up, Archival, Anti-virus and other applications to run on the same
computer are:
1-8Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Symantecv. 8.0v. 8.1v. 10.2
McAfeev. 4.5.1v. 7.xv. 8.5i with
5.xIT (v. 6.0 & v. 6.1)IT (Version 6.5)
Protection Pilot 1.5
Norton Professionalv. 8.0v. 8.12007
Trend Micro--Corporate v. 8.0
(English)
v.7.3 (Except
English and
Swedish)
eTrust--EZ Antivirus 2005
v. 7.1
CA Antivirus--r8.1
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual1-9
Page 16
Introduction
Archival: Storage Area Network (SAN) (equal to a mapped network
drive)
3rd Party: PedCath
LanguagePedCath Version
Englishv. 7.6.5
Frenchv. 7.5.5
Italianv. 7.5.5
Spanishv. 7.5.5
Japanesev. 7.4.3J
NOTE
Chinese systems will use the Japanese version of PedCath. Swedish/
German systems will use the English version of PedCath.
1-10Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 17
Service Information
Service Requirements
Follow the service requirements listed below:
Equipment servicing should be performed by GE Medical Systems
Any unauthorized attempt to repair equipment under warranty voids
It is the user’s responsibility to report the need for service to GE
Failure on the part of the responsible individual, hospital, or
Regular maintenance, irrespective of usage, is essential to ensure
Introduction
Information Technologies authorized service personnel only.
that warranty.
Medical Systems Information Technologies or one of their authorized
agents.
institution using this equipment to implement a satisfactory
maintenance schedule may cause undue equipment failure and
possible health hazards.
that the system is always functional when required.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-11
Page 18
Introduction
XXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXX
XXXXXXXXX
XX
XXXXXXXXXXX
XXXXXXXXXXXXX
AA5 XX XX XXXX X X
A
B
C
D
E
F
G
H
I
Equipment Identification
Every system has a unique serial number for identification. The serial
number appears on the product labels which are located on the top of the
CPU, front and back of the Integrated Electronics Box (IEB), right side of
the RAC 4A, back of the TRAM and back of the Amplifier.
Identification Tags
Identification Tags
ItemName
AName of Device
BManufacturer
CSerial Number
D3 digit product code
EYear Manufactured
FFiscal Week
GProduct Sequence Number
HManufacturing Site
IMisc.
1-12Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 19
Old Identification Tags
[NAME OF DEVICE]
GE MEDICAL SYSTEMS
MILWAUKEE
WI
MADE IN USA
A2PX1808F
A 2 PX 1808 XX
A
B
C
D
E
F
G
H
Introduction
Identification Tag
ItemNameDescription
AName of DeviceCathlab Computer, CLABII Plus Amplifier or Cath Lab IEB Box
BManufacturerGE Medical Systems Information Technologies
CSerial NumberUnique Identifier
DFacility CodeF or G = Milwaukee, P = Bangalore, S = Border Ops (Juarez/El Paso)
EProduct Sequence
Manufacturing number of total units manufactured
Number
FProduct CodeTwo character product descriptor
GYear Manufactured9 = 1999, 0 = 2000 (and so on)
HMonth ManufacturedA = January, B = February (and so on)
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-13
Page 20
Introduction
Required Tools and Supplies
The following is a list of tools needed to service the Mac-Lab/CardioLab
system:
Adjustable wrench with a 1-inch capacity
Phillips screwdriver
Needle nose pliers
Pliers
Socket wrench (3/16 inch to 1 inch size bits)
Phillips head screwdriver (small and large sizes)
Flat head screwdriver (small and large sizes)
Cutting pliers
Power outlet ground tester
Grounding strap
Signal generator with ECG, pressure, cardiac output, and
temperature outputs.
Cables to connect simulator to ECG inputs of amplifier and TRAM
Cables to connect the simulator to the catheter input modules (2 mm
shielded)
Adapter cable to connect the simulator to the amplifier and TRAM
pressure inputs
Cables to connect the simulator to the TRAM, cardiac output, and
temperature inputs
Power Input Requirements
If the equipment is installed in the USA and uses 240V rather than
120V, the source must be a center-tapped, 240V single phase circuit
(120V, 20A; 240V, 10A).
1-14Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 21
Safety Information
Follow all of the instructions and warnings marked included in this
manual and save these instructions for later use. However, these
instructions in no way supersede current medical practices regarding
patient care and safety.
Electromagnetic Compatibility and Interference
This catheterization system meets the requirements of the European
Medical Device Directive, and by reference, the requirements for
electromagnetic compatibility specified by EN 60601-1-2. This equipment
is not likely to interfere with other equipment, except with certain types
of rate adaptive pacemakers that rely on bioelectric impedance
measurement (BIM) technology. Also, interference with the operation of
this equipment is not likely to result from proximity to general
equipment.
This equipment meets or exceeds Class A immunity levels. However,
avoid the use of strong intentional radio transmitters in the immediate
vicinity, including cellular telephones. The use of strong unintentional
radiators, such as electrosurgical equipment, may cause some level of
interference. GE Medical Systems Information Technologies offers
optional ECG patient cables with electrosurgical noise filtering, which
reduces the level of interference on ECG traces.
Introduction
If using electrosurgical or RF ablation equipment in conjunction with
this equipment, select an ESU filtered cable. For more information about
management of the equipment in the clinical environment (from an
electromagnetic compatibility perspective) consult the Association for the
Advancement of Medical Instrumentation's technical information report
Guidance on Electromagnetic Compatibility of Medical Devices for
Clinical/Biomedical Engineers.
Responsibility of the Manufacturer
GE Medical Systems Information Technologies is responsible for the
effects of safety, reliability, and performance only if:
Installation operations, extensions, readjustments, modifications, or
repairs are performed by persons authorized by GE Medical Systems
Information Technologies.
The equipment is used in accordance with the instructions provided.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-15
Page 22
Introduction
General
The user attaches only accessory equipment recommended for
interaction with the amplifier.
This system is intended for use under the direct supervision of a licensed
health care practitioner.
The Mac-Lab/CardioLab system maintains leakage currents below
medical limits with an isolation transformer which is incorporated into
the design of the integrated electronics box (IEB). The system power is
supplied by a dedicated power supply on an isolated circuit for proper
behavior of the system.
To ensure patient safety, use only parts and accessories manufactured or
recommended by GE Medical Systems Information Technologies. Parts
and accessories used must meet the requirements for the applicable EN
60601 series safety standards, and the system configuration must meet
the requirements of the EN 60601-1 medical electrical systems standard.
Contact GE Medical Systems Information Technologies before
connecting any device to this system that is not recommended for use in
this manual.
If the equipment is installed in the USA and uses 240V rather than
120V, the source must be a center-tapped, 240V single-phase circuit.
To ensure patient safety, use only parts and accessories manufactured or
recommended by GE Medical Systems Information Technologies.
This equipment should only be connected as specified by GE Medical
Systems Information Technologies. Installation of the equipment should
only be performed by qualified GE Medical Systems Information
Technologies service representatives. Connection of the system to the
equipment is not recommended in this manual.
Contact GE Medical Systems Information Technologies for information
before connecting any device to this equipment that is not recommended
in this manual. Only cables and accessory equipment purchased form GE
Medical Systems Information Technologies should be used in conjunction
with the system.
It is important that all of the instructions in this manual be followed.
However, these instructions in no way supersede current medical
practices regarding patient care and safety.
Refer servicing of the equipment to GE Medical Systems Information
Technologies authorized service personnel. Any attempt to repair
equipment under warranty will void that warranty.
1-16Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 23
Introduction
Failure on the part of responsible individuals or institutions to
implement a satisfactory equipment maintenance schedule may lead to
undue equipment failure and possible health hazards.
The use of ACCESSORY equipment not complying with the equivalent
safety requirements of this equipment may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice
includes:
Use of the accessory in the Patient Vicinity.
Evidence that the safety certification of the Accessory is performed in
accordance with the appropriate EN 60601-1 and EN 60601-1
harmonized national standard.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-17
Page 24
Introduction
Equipment Symbols
Equipment SymbolsDescriptions
CE Marking
Underwriters Laboratories, Inc.
Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, mechanical
and other specified hazards, only in accordance with UL 2601-1, CAN/CSA C22.2 No.
A nonessential electrical connection which may be used to connect the chassis to other
equipment to achieve the same potential.
Type CF Equipment that is Defibrillator Proof
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and
internal application to the patient excluding direct cardiac application. The “paddles” outside
the box indicate that the part is defibrillator proof.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the
specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a
higher degree of protection against electric shock than that provided by Type B applied
parts.
1-18Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 25
Equipment SymbolsDescriptions
Type CF Equipment
Type CF Applied Part: Isolated (floating) applied part suitable for intentional external and
internal application to the patient excluding direct cardiac application.
(Medical Standard Definition:) F-Type applied part (floating/isolated) complying with the
specified requirements of IEC 60601-1/UL 2601-1/CSA 601.1 Medical Standards provide a
higher degree of protection against electric shock than that provided by Type B applied
parts.
Attention
Pay attention to the documents delivered with the equipment.
Input Symbol
Introduction
Compliance Information
Mac-Lab/CardioLab System
Table 1. Declaration - Electromagnetic Emissions
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Mac-Lab / CardioLab System should assure that it is used in such an environment.
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11Class AThe Mac-Lab / CardioLab System is suitable
Harmonic emissions IEC 61000-3-2Class A
Voltage fluctuations/Flicker emissions IEC
61000-3-3
Complies
for use in all establishments other than
domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-19
Page 26
Introduction
Table 2. Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment -
Guidance
Electrostatic discharge (ESD)
± 6 kV contact ± 8 kV air± 6 kV contact ±8 kV airFloors should be wood, concrete or
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
± 2kV for power supply
lines ± 1 kV for input/
output lines
Surge IEC 61000-4-5± 1kV differential mode
± 2kV common mode
Voltage dips, short
interruptions and voltage
variations on power supply
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for 0,5 cycle
input lines IEC 61000-4-11
40% 100-120Vac
40% 200-240Vac
(>60% dip) for 5 cycles
70% 100-120Vac
70% 200-240Vac
(>30% dip) for 25 cycles
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for 5
seconds
± 2kV for power supply lines ±1 kV
for input/output lines
± 1kV differential mode ±2kV
common mode
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for 0,5 cycle
40% 100-120Vac
40% 200-240Vac
(>60% dip) for 5 cycles
70% 100-120Vac
70% 200-240Vac
(>30% dip) for 25 cycles
<5% 100-120Vac
<5% 200-240Vac
(>95% dip) for 5 seconds
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Mac-Lab / CardioLab System
requires continued operation
during power mains interruptions, it
is recommended that the Mac-Lab
/ CardioLab System be powered
from an Uninterruptable power
supply or a battery.
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
3 A/m3 A/mPower frequency magnetic fields
should be at levels characteristic of
a typical location typical
commercial or hospital
environment.
1-20Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 27
Introduction
Table 3. Declaration - Electromagnetic Immunity
The Mac-Lab / CardioLab System is intended for use in the electromagnetic environment specified below. The customer or the user of
the Mac-Lab / CardioLab System should assure that it is used in such an environment.
Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment -
Guidance
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-21
Page 28
Introduction
Table 3. Declaration - Electromagnetic Immunity
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Mac-Lab / CardioLab System,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance (d)
Conducted RF emissions
IEC 61000-4-6
Radiated RF emissions
IEC 61000-4-3
3 Vrms
105kHz to 80 MHz
3V/m80 MHz to
2,5 GHz
3 Vrms
d = 1.17 P
d = 1.17 P
3 V/m
d = 2.33 P
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters(m).Field
Strengths from fixed RF
transmitters, as determined by a
electromagnetic site survey a,
should be less than the compliance
level in each frequency range b.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
1-22Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 29
Introduction
Table 3. Declaration - Electromagnetic Immunity
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflected from
structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Mac-Lab / CardioLab System is used exceeds the applicable RF compliance level above the Mac-Lab / CardioLab
System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Mac-Lab / CardioLab System.
b Over the frequency range 150 kHz to 80 MHz. Field strengths should be less than 3 V/m.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-23
Page 30
Introduction
Centricity Cardiology INW
File Server EMC Immunity Performance
The Centricity Cardiology INW information system file server was
subjected to:
Electrical Fast Transient (EFT) bursts applied to the AC power
mains, per CISPR 24GemsFT bursts on the AC power mains may
temporarily reduce network performance/bandwidth. Operator
intervention is not required and normal performance returns upon
cessation of the EFT event.
Users should be aware of known RF sources, such as radio or TV
stations and hand-held or mobile two-way radios, and consider them
when installing a device or system.
File Server Accessory EMC Immunity Performance
The Sony CRT monitor was subjected to:
Fast Transient Surges (FTS) of 2 KV applied to the AC power lines
common mode (line to earth), per CISPR24. FTSs of 2 KV or greater may
cause the Sony CRT monitor to experience unrecoverable hardware
failure.
NOTE
The Sony CRT monitor meets/passes the IEC61000-4-5 ITE
requirement of 1 KV.
The UPS was subjected to:
Fast Transient Surges (FTS) of 2 KV applied to the AC power lines
common mode (line to earth), per CISPR24. FITSs of 2 KV or greater
may cause the UPS to go into bypass mode and require a manual reset
for normal operation.
1-24Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 31
Conventions
Safety Messages
Introduction
DANGER
— Safety messages which indicate an imminently
hazardous situation which, if not avoided, WILL result in
death or serious injury.
WARNING
— Safety messages which indicate a potentially
hazardous situation which, if not avoided, COULD result
in death or serious injury.
CAUTION
— Safety messages which indicate a potentially
hazardous situation which, if not avoided may result in
minor or moderate injury, loss of data or system failure.
Definitions
NOTE
Messages which provide additional user information.
Items shown in Black text are keys on the keyboard, text to be
entered, or hardware items such as buttons or switches on the
equipment.
Items shown in Italicized text are software terms which identify
menu items, buttons, or options in various windows.
To perform an operation which appears with a plus (+)sign between
the names of two keys, you press and hold the first key while
pressing the second key once. This is called a keystroke combination.
For example, Press Ctrl+Esc means to press and hold down the Ctrl
key while pressing the Esc key.
When instructions are given for typing a precise text string with one
or more spaces, the point where the spacebar must be pressed is
indicated as: <Space>. The purpose of the < > brackets is to ensure
you press the spacebar when required.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-25
Page 32
Introduction
Enter means to press the Enter or Return key on the keyboard. Do
Hazard Statements
Dangers
Warnings
not type enter.
DANGER
EXPLOSION/FIRE HAZARD — Using this equipment in
the presences of flammable gasses (including anesthetics
and oxygen) may cause an explosion or fire.
Always use this equipment in a well ventilated area away
from the presence of potentially flammable gasses
WARNING
PATIENT SHOCK HAZARD — Only connect UL/IEC
60601-1 certified medical equipment.
WARNING
EQUIPMENT DAMAGE — DO NOT allow hydraulic
fluid from any device to contact any conductive surface.
Hydraulic fluid is electrically conductive.
If fluid of any kind should leak into the system,
discontinue use of the equipment and contact GE Medical
Systems Information Technologies immediately.
WARNING
MISINTERPRETATION OF DATA — The system may
report erroneous heart rate values when pacemaker
spikes or arrhythmias are present. Reported values
should be verified by a qualified physician.
1-26Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 33
Introduction
WARNING
PATIENT SHOCK/EQUIPMENT DAMAGE — Keep the
conductive parts of lead electrodes and associated parts
away from other conducting parts, including earth.
WARNING
PATIENT SHOCK/EQUIPMENT DAMAGE — DO NOT
contact the amplifier or patient during defibrillation.
Standby defibrillation equipment is required in the event
that the patient needs to be cardioverted or defibrillated.
Disconnect leadwires from patient before defibrillating.
Otherwise, serious injury, death, or damage to the
equipment could result.
WARNING
PATIENT SAFETY — All equipment not complying with
EN 60601-1 should be placed outside the patient
environment.
WARNING
POWER SUPPLY SAFETY — Total system earth
leakage current must not exceed 500 microamperes.
Voltages between the three power outlet connections must
be accurate BEFORE measuring power outlet ground-toneutral loop resistance.
Electrostatic discharge (ESD) can damage electronic
components. Be sure you are properly grounded when
using or servicing the system.
WARNING
POWER SUPPLY/SHOCK HAZARD — To reduce the
risk of electric shock or damage to the equipment, do not
disable the power cord grounding feature. This equipment
is designed for connection to a grounded (earthed) power
outlet.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-27
Page 34
Introduction
WARNING
MISINTERPRETATION OF DATA — A qualified
physician must overread all computer-generated
measurements and tracings. Computerized interpretation
is only significant when used in conjunction with clinical
findings.
Signal quality should be checked only by a licensed health
care practitioner.
All computer assisted measurement functions must be
verified by a licensed health care practitioner.
WARNING
LEAKAGE CURRENT SAFETY — The system should be
tested yearly for adequate grounding on patient leakage
current.
Keep leakage current within acceptable limits when
connecting auxiliary equipment to the system.
When more than two electrical devices are interconnected,
the summation of leakage currents must be checked
yearly.
WARNING
LOCK OUT/TAG OUT — Use appropriate Lock Out/Tag
Out procedures whenever an energy hazard exists, such
as disconnecting equipment to perform maintenance.
WARNING
MAINTENANCE/SERVICE — High voltages exist in the
unit. Use insulated tools. Remove jewelry and use only
one hand when possible.
Always disconnect power from the equipment before
attempting service.
1-28Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 35
Introduction
WARNING
PERSONAL/PATIENT SAFETY — Keep hands, hair,
jewelry and loose clothing away from moving parts.
Otherwise, serious injury could result.
WARNING
REPLACEMENT PARTS — Replace only with the same
type and rating of fuse.
WARNING
POWER SUPPLY — The supply line must be fused and
capable of delivering 16A.
Route the AC power cable away from moving parts.
Connect only to a properly earth grounded outlet.
WARNING
PATIENT SAFETY — This device is intended for use
under the direct supervision of a licensed health care
practitioner.
WARNING
SHOCK HAZARD — Electrical shock hazard between
chassis ground and isolated (floating) ground when power
is applied. Unplug the unit from the power source before
proceeding.
To reduce the risk of electric shock or damage to your
equipment, do not disable the power cord grounding
feature. This equipment is designed for connection to a
grounded (earthed) power outlet. The grounding plug is
an important safety feature.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-29
Page 36
Introduction
Cautions
CAUTION
— Federal (USA) law restricts this device to sale by or on
the order of a physician.
CAUTION
INACCURACY OF DATA — DO NOT display any
channels which are based on unused catheter input
modules. Otherwise, erroneous values may occur.
ECG signal quality should only be checked by qualified
medical personnel.
CAUTION
EQUIPMENT CABLING DAMAGE — Route optical
cables thorough conduit in the ceiling or floor to avoid
damage to the cables of cable connectors.
CAUTION
EQUIPMENT USE —
CAUTION
EMC — No known electromagnetic or other interference
between the system exists. However, precautions should
be taken to avoid the use of cellular telephones, pagers, or
other transmitters.
CAUTION
EQUIPMENT DAMAGE — Exit the software application
properly before turning off the system. Failure to do so
may damage the optical disk.
Remove the optical disk before moving the system.
Failure to do so may result in damage to the drive or the
disk.
DO NOT reuse single use devices.
1-30Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 37
Introduction
CAUTION
TRANSDUCER — Remove air that is trapped in the
transducer or associated tubing by flushing the system
according to Cath lab procedures.
CAUTION
REMOTE MONITOR — Always power the system
through separate isolation transformers which are
independently connected to a wall outlet. You can power
the laser printer through the IEB/Workstation power
supply. Otherwise, the printer may draw enough current
away from the system and cause it to reboot.
CAUTION
SOFTWARE — DO NOT add unauthorized software to
the Mac-Lab/CardioLab system. Doing so may make the
Mac-Lab/CardioLab software unstable.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual1-31
Page 38
Introduction
1-32Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 39
2Equipment Overview
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual2-1
Page 40
Equipment Overview
IEB
Amplifier
CIM
TRAM
RAC4A
Bed
RF
Generator
Catheter
Personality
Module
Mouse
Printer
Speakers
Stimulator
Real-Time
Monitor
Review
Monitor
Remote
Real-Time
Monitor
3" Ceiling Conduit
3" Floor Conduit
3" Floor Conduit
Remote
Review
Monitor
Image
Monitor
INW
Server
CVIS
Network
Switch
To Hospital Network
Keyboard
Workstation
Dashed lines indicate CardioLab/ComboLab only
Procedure Room
Control Room
IT Control Room
Monitor
Boom
HP
CPU
UPS
Typical System Configuration
This system is used to assess the cardiac condition of patients and may
consist of either Mac-Lab (hemodynamic), CardioLab
(electrophysiologic), or ComboLab (both systems).
The system may be standalone or connected to a network. Network
installations include an INW and an optional CVIS server.
The block diagram below shows a typical setup.
2-2Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
001
Page 41
System Hardware
Network Components
Equipment Overview
WARNING
— LOCK OUT/TAG OUT
Use appropriate Lock Out/Tag Out procedures before
performing any service on potentially energized
equipment.
INW Server provides
Network repository for patient studies
Acquisition and review workstation access to patient studies
CVIS Server provides
Central list management for staff, supplies, etc.
Patient scheduling
Patient information
Inventory Management
Statistical Reports
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual2-3
Part Number: 2026817-001
Function: Improves server performance and decreases
data loss potential if power is lost
Location: Connected to the SMART ARRAY card in Slot 2
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual2-7
Page 46
Equipment Overview
INW G3 Server
The G3 server has the following specifications:
2.8 GHz Xeon processor with 512 MB RAM
Three tiers:
Low: One 72 GB drives
Mid: Three 146 GB drives (RAID 5 mirrored)
High: Six 146 GB drives (RAID 5 mirrored)
401
2-8Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 47
Acquisition Workstation Configuration
Control Room Components
Integrated Electronics Box (IEB): The IEB provides power to
local system components, signal pass through video and
communications.
Uninterruptible Power Supply (UPS): The UPS provides
emergency power to the HP computer.
HP Acquisition computer
USB Mouse
Keyboard
Monitors: Two monitors (Real-Time and Review) are standard. A
third Image monitor is optional.
Local speakers: Provide output for ECG QRS tone and optional
audible indicators.
A laser printer: Prints images or reports.
Barcode scanner
Equipment Overview
Remote Components
Monitors: One remote monitor is standard (and is software
switchable between Real-Time and Review windows). A second
monitor is optional.
Remote speakers: Provide output for ECG QRS tone and optional
audible indicators in the procedure room.
Other Related System Equipment
TRAMnet: The TRAMnet gathers patient data for the Mac-Lab
system. The TRAMnet consists of a Remote Acquisition Case (RAC4A) and a Transport Remote Acquisition Module (TRAM).
CLab II Plus Amplifier: The Amplifier gathers data for both the
ComboLab and CardioLab systems.
CARTO™ XP System (Patient Interface Unit): Provides 3D
electroanatomical cardiac maps to the CardioLab.
Analog Output Box: Distributes HR and BP signals from the
TRAMnet and Amplifier to other equipment.
Review or Nurses workstation
Video Printer
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog y INW Service Manual2-9
Page 48
Equipment Overview
Acquisition Workstation System Components
Integrated Electronics Box (IEB)
The IEB provides power to local system components, signal pass through
video and communications. A modem for dial-up communications may be
present, depending on which IEB is installed. Three versions of the IEB
are available to handle different line voltages: 100, 120, and 240 VAC.
The IEB consists of the following components:
Video Multiplexer (VID MUX) board to distribute video signals from
the video cards in the HP computer to the monitors.
I/O board to provide signal pass through for other system
components.
Isolation transformer:
Conforms to IEC 60601-1
2.0 kVA medical grade high-isolation transformation
Provides either 120 VAC (for 100 and 120 VAC systems) or 240
VAC. The primary and secondary windings are wired in parallel
for 120 VAC and in series for 240 VAC.
CAUTION
— If the transformer must be disconnected for any
reason, ensure the connections are correct.
Power Distribution panel provides conditioned power to local system
components.
Depending on which IEB is installed, a modem and modem power
supply for dial-up communication may be present. The power supply
accepts any line voltage and converts it into 9 VDC.
2-10Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 49
External Components and Connections
5
Equipment Overview
190b
LabelPart NumberDescription
12003681-003VID MUX board
22003680-001I/O Board
3N/APower Connections. The bottom right connector is for the HP
computer and the four vertical connectors on the right and left
are for peripheral equipment.
4N/AThermal Overcurrent Circuit Breaker. This breaker protects all
internal components and external devices connected to the
IEB.
5N/AMain Power Switch. This switch controls power to all internal
components and external devices except the HP computer.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-11
Page 50
Equipment Overview
Both the Main Power switch on the front of the IEB and the Thermal
Overload circuit breaker on the back must be on for the system to power
up.
Video Multiplexer (VID MUX) Board
LabelDescription
1Switched video out (Video 1 or 2). This is used when the facility has only
one remote monitor. To switch monitors:
Mac-Lab: Go to the main menu and select Windows > Physician
Video
CardioLab: Use the dropdown box just under the main menu.
2Video 1 (Real-time) out to local monitor (bottom) and remote monitor (top)
3Video 2 (Review) out to local monitor (bottom) and remote (top)
4Video 3 (CardioImage) out to local monitor (bottom) and remote (top)
5(L to R) Video 1, Video 2, Video 3 in from HP Computer Matrox AGP Video
board (Vid 1 and Vid 2) and the Matrox PCI Video board (Vid 3)
197c
2-12Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 51
Equipment Overview
The VID MUX board amplifies (±5V 1.25A from the internal IEB DC
power supply. See “Internal Components and Connections” on page 2-17)
and distributes the video signals from the HP Computer (See“HP8400
Video Connections” on page 2-61, “HP8200 Video Connections” on
page 2-62 or “HP8000 Video Connections” on page 2-63).
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-13
Page 52
Equipment Overview
I/O Board
LabelDescription
1Analog I/0 Communications. Connects to the HP Computer PCI
Comms board.
2Analog in from another system (Fluoro, etc.).
3Connection to the Analog Output Box.
4Analog out (Stimulator, etc.).
5I/O TRAM (includes 16.5VDC power).
6Output to HP Computer TRAMnet to Ethernet to Adapter board.
Refer to the following cabling diagrams:
“HP8400 Communications Connections” on page 2-65
“HP8200 Communications Connections” on page 2-66
“HP8000 Communications Connections” on page 2-67
“HP8400 TRAM/Amplifier Connections” on page 2-68
“HP8200 TRAM/Amplifier Connections” on page 2-69
196c
2-14Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 53
Equipment Overview
“HP8000 TRAM/Amplifier Connections” on page 2-70
Power Connections
LabelDescription
198c
1 & 5AC out to local equipment (6A @ 120 VAC max.; 3A @ 240 VAC max.)
2Thermal Overcurrent circuit breaker. Protects all internal components
and external devices connected to the IEB.
3AC main power in (15A @ 120 VAC max; 7A, @ 240 VAC max.)
4AC power out to UPS (2A @ 120 VAC max; 1A, @ 240 VAC max.)
See the cabling diagram, “Power Connections” on page 2-60.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-15
Page 54
Equipment Overview
Disassembly
Open Case
191
b
1. Remove 7 side-panel screws.
2. Lift the side panel out.
Remove Connection Boards
193
a
1. Open the case.
2. Remove the board screws.
3. Disconnect interior cables.
WARNING
— Electrostatic discharge (ESD) can damage electronic
components. Be sure you are properly grounded before
opening any equipment.
2-16Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 55
Internal Components and Connections
1
23
Equipment Overview
195c
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-17
Page 56
Equipment Overview
LabelPart NumberDescription
12003681-003
22013535-001Power Supply: Provides ±5 VDC 1.25A for the Vid
32003263-001Terminal Strip
Uninterruptible Power Supply (UPS)
The UPS is designed to maintain power to the computer during a power
outage.
NOTE
The UPS provides uninterrupted power to the HP computer. The
monitors and other peripherials will not remain on if power is lost.
2003680-001
Video and
I/O boards (rear view)
MUX board and +16.5 VDC for the RAC-4A.
199
2-18Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 57
Equipment Overview
Part NumberDescription
2034077-001UPS Powercom, 120V, 50/60 Hz, 1500
VA, BNT
2034078-001UPS Powercom, 230V, 50/60 Hz, 1500
VA, BNT
2034079-001UPS Powercom, 100V, 50/60 Hz, 1500
VA, BNT
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-19
Page 58
Equipment Overview
HP Computer
The HP computer is used for either Acquisition systems or Review
workstations and has the following components:
Floppy drive (Drive A:) - On 8000/8200 only.
Primary Hard Drive with two partitions
C: Application partition
D: Data partition
Magneto-Optical drive (Drive E:)- Optional on 8200/8400.
DVD-RAM drive or CD-RW drive (Drive F:).
NOTE
The drive used to store application studies is determined by the
file paths set in under System Settings in the main application
menu.
Secondary Hard Drive (Drive G:).
Four standard expansion cards (plus two optional cards):
Video board provides video out for the Real-Time (Video 1) and
Review (Video 2) monitors. (Matrox Parhelia PCI for 8400 and
8200, Matrox AGP Parhelia for 8000)
TRAMnet to Ethernet Adapter board for communication with the
TRAM in the RAC 4A
PCI Communication board to provides Amplifier
Communication, Analog and Digital I/O, and Switched video
(video out from the IEB).
Dual NIC Adapter board provides Network connection.
(optional on 8400): Adaptec Ultra320 SCSI PCIe board for
supporting MO drive.
(optional on 8200 and 8400): Matrox Odyssey XA board used to
capture images from an X-Ray or other imaging system.
(optional on 8200 and 8000): Matrox Odyssey XPro board used to
capture images from an X-Ray or other imaging system.
(optional on 8400): ATI FIREGL V5200 256MB PCI-E Video
board. Provides video out for optional (video 3) monitor.
(optional on 8200): ATI FIREGL V5100 128MB PCI-E Video
board. Provides video out for optional (video 3) monitor
(optional on 8000): Matrox PCI Parhelia Video board provides
video out for the optional (video 3) monitor.
2-20Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 59
Equipment Overview
External Components and Connections for the HP8400
LabelPart NumberDescription
12028599-002DVD-RAM Drive
2N/APower Switch, LEDs, USB, Firewire, and Audio
connections
3N/AAC Power
4N/ABackplane I/O panel
5N/ABoard Slots
Not Shown2003340-001Magneto-Optical Drive (optional)
Not Shown2026539-001HP USB 2-Button Optical Mouse with Wheel
Not ShownVariousKeyboard (dependent on language)
306
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-21
Page 60
Equipment Overview
4
5
6
1
2
3
External Components and Connections for the HP8200
connections
4N/AAC Power
5N/ABackplane I/O panel
6N/ABoard Slots
Not Shown2003340-001Magneto-Optical Drive (optional)
Not Shown2026539-001HP USB 2-Button Optical Mouse with Wheel
Not ShownVariousKeyboard (dependent on language)
2-22Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 61
Equipment Overview
External Components and Connections for the HP8000
Not Shown2026539-001HP USB 2-Button Optical Mouse with Wheel
Not ShownVariousKeyboard (dependent on language)
090b
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-23
Page 62
Equipment Overview
Disassembly
WARNING
— Electrostatic discharge (ESD) can damage electronic
components. Be sure you are properly grounded before
opening any equipment.
Remove Case Side Panel
1.Lift the tab near the top.
2.Swing the side panel out.
Remove Front Panel
1.Lift tabs indicated.
2.Swing the front panel out.
2-24Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 63
Equipment Overview
1
2
2
Remove Hard Drive
1.Press tabs as indicated.
2.Slide drive out.
Remove Floppy Drive
1.Remove screw.
2.Lift tab indicated.
3.Slide drive out.
Remove Clamps
1.Lift the clamp fingers.
2.Squeeze down the tabs,
where indicated. Gently
push in, then pull out.
3.Lift off the clamp.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-25
Page 64
Equipment Overview
Remove the ATI FIREGL
V5100/V5200 128MB/256MB
PCI-E Video Board (HP8200/
8400 only)
1.Lift up the lock arm as
shown.
2.Squeeze down the tabs,
where indicated. Gently
push in, then pull out.
3.Lift off the clamp.
Remove DVD-RAM Drive
1.Remove screw.
2.Lift tab indicated.
3.Slide drive out.
2-26Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 65
Equipment Overview
DVD-RAM Drive
The DVD-RAM drive can also read/write to CDs. The recommended disks
have a capacity of approximately 4.7 GB per side.
NOTE
The drive used to store application studies is determined by the file
paths set in under System Settings in the main application menu.
092b
1: DVD-RAM Drive Jumpers
2: IDE Connector
Optical Drive
The optical drive is optional upgrade to HP Computers.
NOTE
Drive use is determined by the file paths set in under System
Settings in the main application menu.
1: Optical Drive Jumpers
2: SCSI Adapter (p/n 2012522-002)
NOTE
Two sets of DIP switches are located on the top optical drive.
These are factory preset and must not be changed.
093b
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-27
Page 66
Equipment Overview
HP8400 Installed Boards
8400 PCI Communication Board
309
Part Number: 2014935-002
Location: Slot 1
Functions:
Communicates with the C-Lab II Plus Amplifier via the fiber
optic connectors
Analog and Digital communication with the IEB via the D-
Connector
Provides the video switching signal for the Switched Video
output on the IEB via the D-Connector
I/O: D-Connector to IEB, Fiber Optic Jacks to the Amplifier
2-28Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 67
Equipment Overview
8400 ATI FIREGL V5200 256MB PCI-E Video Board (Optional)
Part Number: 2026189-002
Location: Slot 2
Function: Video out for optional (video 3) monitor
I/O: Two DVI connectors (only the left one is used)
304
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-29
Page 68
Equipment Overview
8400 Dual NIC Adapter Board
Part Number: 2017795-002
Location: Slot 3
Function: Network connection card
I/O: Two network connection jacks
307
2-30Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 69
8400 PCIX 2 SCSI Card
NOTE
This board is optional with the MO upgrade kit.
Equipment Overview
Part Number: 2038632-001
Location: Slot 4
Function: Provides SCSI cable connection internally to an optional
MO Drive.
I/O: Two SCSI connectors (only internal SCSI is used)
310
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-31
Page 70
Equipment Overview
8400 MATROX PARHELIA PCI 256 MB Video Card
305
Part Number: 2026188-001
Location: Slot 5 (PCI slot)
Function: Provides video out for the Real-Time (Video 1) and Review
(Video 2) monitors. This board supports resolutions of 1280x1024 for
Mac-Lab, 1600x1200 for CardioLab.
I/O: Two DVI connectors
Included adaptors: DVI to HD15 Video Adapter (2021697-001).
2-32Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 71
8400Matrox Odyssey XA Board (Optional)
Equipment Overview
Part Number: 2030688-001
Location: Slot 6
Function: Image capture for X-Ray or other imaging systems
I/O: Two DVI connectors
311
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-33
Page 72
Equipment Overview
8400 TRAMnet to Ethernet Adapter board
Part Number: 2009937-002
Location: Slot 7
Function: Provides communication with the TRAM in the RAC 4A,
and converts signals from the TRAM to the ethernet.
I/O: Network Interface Port and 9-Pin D connector
312
2-34Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 73
HP8200 Installed Boards
1
7
6
5
4
3
2
NOTE
Slot 4 is not used.
8200 Dual NIC Adapter Board
Equipment Overview
Part Number: 2017795-001
Location: Slot 1
Function: Network connection card
I/O: Two network connection jacks
201a
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-35
Page 74
Equipment Overview
1
7
6
5
4
3
2
8200 ATI FIREGL V5100 128MB PCI-E Video Board (Optional)
Part Number: 2026189-001
Location: Slot 2
Function: Video out for optional (video 3) monitor
I/O: Two DVI connectors (only the left one is used)
202a
2-36Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 75
Equipment Overview
1
7
6
5
4
3
2
8200 PCI Communication Board
Part Number: 2014935-002
Location: Slot 3
Functions:
Communicates with the C-Lab II Plus Amplifier via the fiber
optic connectors.
Analog and Digital communication with the IEB via the D-
Connector.
Provides the video switching signal for the Switched Video
output on the IEB via the D-Connector.
I/O: D-Connector to IEB, Fiber Optic Jacks to the Amplifier
203a
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-37
Page 76
Equipment Overview
1
7
6
5
4
3
2
8200 MATROX PARHELIA PCI 256 MB Video Card
205a
Part Number: 2026188-001
Location: Slot 5 (PCI slot)
Function: Provides video out for the Real-Time (Video 1) and Review
(Video 2) monitors. This board supports resolutions of 1280x1024 for
Mac-Lab, 1600x1200 for CardioLab.
I/O: Two DVI connectors
Included adaptor: DVI to HD15 Video Adapter (2021697-001).
2-38Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 77
8200 Matrox Odyssey XA Board (Optional)
1
7
6
5
4
3
2
Equipment Overview
Part Number: 2030688-001 (replaces initially released part number
2017708-001, Matrox Odyssey XPro board, which is obsolete)
Location: Slot 6
Function: Image capture for X-Ray or other imaging systems
I/O: Two DVI connectors
206a
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-39
Page 78
Equipment Overview
8200 TRAMnet to Ethernet Adapter board
Part Number: 2009937-002
Location: Slot 7
Function: Provides communication with the TRAM in the RAC 4A,
and converts signals from the TRAM to the ethernet.
I/O: Network Interface Port and 9-Pin D connector
207a
2-40Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 79
HP8000 Installed Boards
NOTE
Slot 1 is not used.
8000 Matrox AGP Parhelia Video board
Equipment Overview
102a
Part Number: 2017707-001
Location: Slot 2 (AGP slot)
Function: Provides video out for the Real-Time (Video 1) and Review
(Video 2) monitors. This board supports resolutions of 1280x1024 for
Mac-Lab, 1600x1200 for CardioLab.
I/O: Two DVI connectors
Included adaptor: DVI to HD15 Video Adapter (2021697-001).
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-41
Page 80
Equipment Overview
8000 TRAMnet to Ethernet Adapter board
Part Number: 2009937-002
Location: Slot 3
Function: Provides communication with the TRAM in the RAC 4A,
and converts signals from the TRAM to the ethernet.
I/O: Network Interface Port and 9-Pin D connector
103a
2-42Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 81
8000 PCI Communication board
Equipment Overview
Part Number: 2014935-002
Location: Slot 4
Functions:
Communicates with the C-Lab II Plus Amplifier via the fiber
optic connectors.
Analog and Digital communication with the IEB via the D-
Connector.
Provides the video switching signal for the Switched Video
output on the IEB via the D-Connector.
I/O: D-Connector to IEB, Fiber Optic Jacks to the Amplifier
104a
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-43
Page 82
Equipment Overview
8000 Dual NIC Adapter Board
Part Number: 2017795-001
Location: Slot 5
Function: Network connection card
I/O: Two network connection jacks
105a
2-44Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 83
8000 Matrox Odyssey XPro Board (Optional)
Equipment Overview
Part Number: 2017708-001
Location: Slot 6
Function: Image capture for X-Ray or other imaging systems
I/O: Two DVI connectors
106a
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-45
Page 84
Equipment Overview
8000 Matrox PCI Parhelia Video Board (Optional)
Part Number: 2017709-001
Location: Slot 7
Function: Video out for optional (video 3) monitor
I/O: Two DVI connectors (only the left one is used)
107a
2-46Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 85
Monitors
Equipment Overview
120b
LabelDescription
1DVI Jacks. Note: the video cable may be plugged into either jack--the
monitor automatically switches to the correct jack.
2USB and Firewire ports. (Not Used)
3AC power connection
The standard workstation comes with two monitors and may have a
third optional monitor. The monitors include:
A Real-Time monitor to display current patient information. It is
usually located on the left.
A Review monitor to display past patient information.
An optional Image monitor (used with the CardioImage option)
displays live or frozen video images.
A single remote monitor is standard and receives input from the
switched video out (Video 1 or 2) jack on the IEB (see “External
Components and Connections” on page 2-11).
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-47
Page 86
Equipment Overview
To switch video for the Mac-Lab go to the main menu and select
Windows > Physician Video.
To switch video for the CardioLab use the dropdown box just under
the main menu.
2-48Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 87
Other Related Equipment
TRAMnet
Equipment Overview
029b
LabelDescription
1TRAM Module
2To Analog Out Connector
3To IEB TRAM Connector
The TRAMnet consists of a Remote Acquisition Case (RAC-4A) and one
or more TRAM modules. The TRAMnet is located in the procedure room
near the procedure table.
The TRAMnet receives power (+16.4 VDC) from the IEB DC power
supply via the serial cable.
The TRAM module receives patient inputs and then transmits signal
data to the IEB and other equipment through the Analog Output Box
(refer to the cabling diagrams on page“HP8400 TRAM/Amplifier
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-49
Page 88
Equipment Overview
Connections” on page 2-68, “HP8200 TRAM/Amplifier Connections” on
page 2-69 and “HP8000 TRAM/Amplifier Connections” on page 2-70).
TRAM Module Inputs
Electrocardiogram (ECG) - green connector (5 or 10 leads)
Thermal Dilution Cardiac Output (TDCO) - brown connector
Invasive Blood Pressure (IBP) - 3 red connectors
Non-Invasive Blood Pressure (NIBP) - black connector
Pulse Oximetry (SpO2) - blue connector
Outputs: Analog output connector.
The following TRAM modules are supported by the Mac-Lab system:
200SL
400/450SL (Ohmeda, Nellcor)
451/451N/451M (Ohmeda, Nellcor, Massimo)
600/650SL (Ohmeda, Nellcor)
The following software versions are supported by the Mac-Lab system:
7A, 7B, 7C, 7D, 7E, 7F, 7H
8A, 8B, 8C, 8D
9A, 9B
10A, 10B
11A, 11B, 11C
2-50Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 89
CARTO™ XP System (Patient Interface Unit)
LabelDescription
Equipment Overview
160c
1Output to Amplifier
CAUTION
EXCESSIVE SIGNAL NOISE — Using a PIU earlier
than version 10 with a crossover cable may cause
excessive signal noise.
If this excessive signal noise occurs, then discontinue
using the crossover cable until the PIU is replaced with a
later version. (The new PIU must be purchased from
Biosense Webster).
NOTE
For specific information related to the operation, troubleshooting, or
repair of the CARTO™ XP System, refer to the OEM documentation.
For information about the CardioLab/CARTO™ XP System interface,
refer to the CardioLab Operator manual.
The CARTO™ XP System acquires and displays electroanatomical 3D
images of heart chambers. This information is displayed on the
CARTO™ XP System and can be exported into the CardioLab
application.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-51
Page 90
Equipment Overview
Two connections are required: one from the CARTO™ XP System Patient
Interface Unit to the Clab II Plus Amplifier and another from the
CARTO™ XP System computer to the HP computer. Refer to the cabling
diagram on page“HP8400 CARTO™XP System Connections” on page 273, “HP8200 CARTO™XP System Connections” on page 2-74 and
“HP8000 CARTO™ XP System Connections” on page 2-75 for specific
information.
To quickly view the status of the interface, check the CARTO™ XP
System icon at the bottom-right of the Real-Time window (a heart
against a grid). The color changes, depending on the system state:
Gray: Disconnected
Blue: Standby
Green: Passive Interface
Yellow: Active Interface
2-52Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 91
Analog Output Box
Equipment Overview
050b
LabelDescription
1Input from TRAMnet
2Input from Amplifier through the IEB
The analog output box distributes HR and BP signals to multiple outputs
(refer to the connection diagram on page “HP8400 TRAM/Amplifier
Connections” on page 2-68, “HP8200 TRAM/Amplifier Connections” on
page 2-69 and “HP8000 TRAM/Amplifier Connections” on page 2-70).
This may be done to synch auxiliary equipment or display the
information on a scope or other device.
Patient information comes from either the TRAMnet or the Amplifier
(through the IEB) and is output through the output jacks on the top of
the box:
Channels 2-4 and 13-16 are switchable between the TRAMnet (Mac-
Lab) and Clab 2 Plus Amplifier (CardioLab).
The remaining channels are used exclusively by CardioLab.
Channels 1-12 may be distributed to up to 4 other devices using the
selector switch.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-53
Page 92
Equipment Overview
Cabling and Connections
WARNING
LOCK OUT/TAG OUT — Use appropriate Lock Out/Tag
Out procedures before performing any service on
potentially energized equipment.
2-54Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 93
Amplifier/Stimulator Connections
IEB
HP
Computer
MicroPace
Stimulator
MicroPace
CPU
Amplifier
RF
Generator
1
9
6
2
4
5
Catheter
Connection
Box
External
ECG
Auxiliary
Reference
3
7
8
Catheter
Input
Module
Personality
Module
Typical Setup (with MicroPace)
Equipment Overview
004
LabelPart NumberDescription
12003411-001BNC Cable (50’). Connects back of MicroPace (ECG 1
Input) to back of IEB (Analog Out 1).
2N/ALarge Cable with D-Connectors. Connects back of
MicroPace (Auxiliary Port) to side of Catheter
Connection Box.
3N/ASmall Cable. Connects front of MicroPace (Green
Connection Port or in emergencies Red Connection
Port) to side of Catheter Connection Box.
4301-00204-03Red or Black leads connect to front of Catheter
Connection Box to Amplifier Stimulator Input.
52003434-001Fiber Optic Cable (75’)
62003587-001DB44 Connect Cable
72003408-00250’ Serial Cable
8301-00202-08Catheter Input Module
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-55
Page 94
Equipment Overview
IEB
HP
Computer
Bloom
Patient
Isolation
Box
Amplifier
RF
Generator
BatteryBattery
1
6
2
4
5
Catheter
Connection
Box
External
ECG
Auxiliary
Reference
3
7
8
Catheter
Input
Module
Personality
Module
Typical Setup (with Bloom)
003
LabelPart NumberDescription
12003410-001BNC Cable (30’). Connects Bloom and IEB.
2N/ALarge Cable with Square Connectors. Connects
back of Bloom to back of Patient Isolation Box.
3N/ASmall Cable. Connects front of Patient Isolation Box to
side of Catheter Connection Box.
4301-00204-03Red or Black leads connect to front of Catheter
Connection Box to Amplifier Stimulator Input.
52003434-001Fiber Optic Cable (75’)
62003587-001DB44 Connect Cable
72003408-00250’ Serial Cable
8301-00202-08Catheter Input Module
2-56Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 95
Equipment Setup
LabelPart NumberDescription
C and C1N/ABloom (This is a typical example.Specific equipment may vary by location).
Equipment Overview
002a
Connects to the IEB (not shown) via the External Input BNC jack on the back (C1)
and the Patient Isolation Box (D) via a large cable with square connectors. (See
next section for detail.)
DN/APatient Isolation Box Rear connection to Bloom (C1) via large cable with square
connectors. Front connection to Catheter Connection Box (E). Front panel
provides access to internal batteries.
EN/ACatheter Connection Box Side connection to the Patient Isolation Box (D) front
connector. Red and Black leads connect from the front to the Amplifier (F) Stim
Input.
G301-00202-08Catheter Input Module Catheter leads (not shown) plug into the block end.
Connector on other end plugs into the Amplifier (4) Cardiac Input (Labeled A, B,
C, etc.)
1N/ABloom connection to IEB (BNC cable not shown).
2N/ABloom connection to Patient Isolation Box.
3N/APatient Isolation box to Catheter Input Module cable.
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-57
Page 96
Equipment Overview
12
LabelPart NumberDescription
4301-00204-03Catheter Connection Box to Amplifier cable.
5301-00038-00External ECG Connections. Colored leads connect to patient. Connector on
6301-00205-08Auxiliary Reference Cable. Plugs into the Amplifier (4) Auxiliary Input.
Bloom (Connection Detail)
other end connects into the Amplifier(4) ECG Input.
ConnectorDescription
1External Input. BNC connection to the IEB (cable not shown).
2J20. Connection to Patient Isolation Box (large square
connector)
001
2-58Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 97
HP/IEB Connections
3
4
1
2
HP8000 Computer
I/O Backplane
5
6
3
6
5
1
2
4
7
7
HP8200/8400 Computer
I/O Backplane
HP Computer I/O Backplane Connections
Equipment Overview
025
LabelPart NumberDescription
1N/APS2 Connection for Y-cable to keyboard (if
connecting with PS2 version Barcode Scanner)
22020280-001
(Note: This number is
PS2 connection for Y-Cable to PS2 version Barcode
Scanner
for the cable only)
36119-101
To Local Printer (cable to use depends on printer)
2031697-001
4N/AUSB Ports for Mouse (Must use right-most port), and
for USB version Barcode Scanner
5N/ATo Speakers
6N/ATo Ablation Interface
7N/ATo Network
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-59
Page 98
Equipment Overview
33
2
1
HP Computer
IEB
4
UPS
Power Connections
LabelPart NumberDescriptionLength
1700775-001
700774-001
2405535-013
80274-006
2003927-001
32003927-001
403689-010
120 VAC main power in
250 VAC main power in
AC power to HP CPU (100V)
AC power to HP CPU (120V)
AC power to HP CPU (230V)
AC to Local Equipment--Monitors, Printers, etc.
AC to Remote Monitors
8'
8'
8'
100'
42003927-001AC power to UPS8'
018
2-60Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
Page 99
HP8400 Video Connections
7
6
4
5
8
3
1
9b
9a
2
HP8400
Computer
IEB
Equipment Overview
LabelPart NumberDescriptionLength
12003928-001Real-time display (Video 1) in to IEB 5'
22003928-001Review display (Video 2) in to IEB5'
32003928-001Cardio Image display (Video 3) in to IEB5'
42003928-001Out to Local Cardio Image Monitor (Video 3)5'
52003928-001Out to Local Review Monitor (Video 2)5'
62003928-001Out to local Real-time monitor (Video 1)5'
72003442-001Out to remote monitor (switched Video 1 or Video 2) 100'
82003442-001Out to optional remote monitors (Video 1, 2, or 3)100'
421
2021423-022JMac-Lab/CardioLab/ Centricity Cardiolog yINW Service Manual2-61
Page 100
Equipment Overview
7
6
4
5
8
3
1
9b
9a
2
HP8200
Computer
IEB
LabelPart NumberDescriptionLength
9a2018060-001Image capture. DVI to 5-BNC female connector cable
9b2003411-001BNC Cable to optional imaging system50'
HP8200 Video Connections
plugs into either DVI plug. Only one BNC connection is
used to connect to the image input from another system
(9b)
419
LabelPart NumberDescriptionLength
12003928-001Real-time display (Video 1) in to IEB 5'
22003928-001Review display (Video 2) in to IEB5'
32003928-001Cardio Image display (Video 3) in to IEB5'
2-62Mac-Lab/CardioLab/ Centricity CardiologyINW Service Manual2021423-022J
42003928-001Out to Local Cardio Image Monitor (Video 3)5'
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.