GE Dinamap ProCare User manual

Page 1
Page 2
Page 3
DINAMAP
®
ProCare Monitor
Operation Manual
Page 4
Page 5
Contents
Introduction
About the DINAMAP® ProCare Monitor ............................................................7
Indications ...................................................................................................................7
Warnings ......................................................................................................................8
Cautions .......................................................................................................................9
Product Compliance ...............................................................................................10
Symbols.......................................................................................................................11
Getting Started
Unpacking the Monitor and Accessories ...........................................................15
Setting up BP Connections ....................................................................................15
Setting up SpO2 Connections ..............................................................................16
Setting up Temperature Connections ...................................................................6
Setting up the Printer (Installing the Paper) .......................................................17
Power Sources ..........................................................................................................17
Powering the Monitor: Battery and DC Supply ................................................17
Configuration Mode Settings .................................................................................19
Setting the Date and Time .....................................................................................20
Inflation Pressure Default Settings (Refer to the BP Section for options) ....21
Alarm Default Settings (Refer to Alarms Section for options).........................21
SPO2 Configuration Settings .................................................................................22
(Refer to the Masimo® section for options) ................................22
Procedures .............................................................................................20
Procedures .............................................................................................21
Procedure for units with Nellcor® Technology
(Refer to the Nellcor® section for options) .................................22
Procedure for units with Masimo® Technology
Product Overview
Buttons .......................................................................................................................25
Front Panel ................................................................................................................27
Rear Panel .................................................................................................................29
Left-Side Panel ..........................................................................................................30
Right-Side Panel ........................................................................................................30
Windows ....................................................................................................................31
Indicators ...................................................................................................................31
Operating Modes ....................................................................................................31
Parameter Modes ....................................................................................................32
User Modes ...............................................................................................................32
Sounds ........................................................................................................................33
Power Sources ..........................................................................................................34
Specifications ............................................................................................................35
Entering Configuration Mode ............................................................31
Mechanical ............................................................................................35
Power Requirements ...........................................................................35
Environmental .......................................................................................36
Printer
Description ................................................................................................................41
Page 6
Installing the Paper ..................................................................................................41
Print Button ...............................................................................................................41
Printer Alarms ...........................................................................................................42
Storage .......................................................................................................................42
Alarms
Cautions .....................................................................................................................42
Alarm Codes .............................................................................................................45
Adjusting Alarm Limits ............................................................................................45
Alarms Button .......................................................................................45
Adjusting the Alarm Volume .................................................................................46
Silencing and Acknowledging an Alarm .............................................................46
Silence Button .......................................................................................46
Alarm Sounds ...........................................................................................................46
Alarm Detection and Priorities ..............................................................................47
Limit Alarms ...........................................................................................47
Parameter Status Alarms .....................................................................48
Printer Alarms ........................................................................................48
Memory Alarms ....................................................................................48
Battery Alarms........................................................................................48
Specifications ............................................................................................................50
Alarm Message Code Table ..................................................................................50
System Failure Alarms .........................................................................49
History
Description ................................................................................................................55
Buttons Associated with History ...........................................................................55
Windows Associated with History .......................................................................56
Indicators Associated with History ......................................................................56
Erasing Stored History .........................................................................55
BP
Description ................................................................................................................59
General Warnings ....................................................................................................60
General Cautions .....................................................................................................61
Buttons Associated with BP ...................................................................................62
Windows Associated with BP ...............................................................................63
Indicators Associated with BP ...............................................................................63
Parameter Modes ....................................................................................................63
BP Modes of Operation .........................................................................................64
User Settings .............................................................................................................66
Menu Settings ...........................................................................................................67
Sounds Associated with BP ...................................................................................67
General Notes .......................................................................................61
Manual BP Determinations ................................................................64
Auto Cycle Determinations ................................................................64
STAT BP Determinations ....................................................................65
Mode Settings .......................................................................................66
Limit Settings .........................................................................................66
Neonate Settings .....................................................................66
Page 7
Contents
Procedures ................................................................................................................67
Alarms ........................................................................................................................70
Specifications ............................................................................................................71
Cuff Connections .................................................................................69
Quick-Disconnect ....................................................................69
Air Hose ....................................................................................69
Neonate Air Hose ...................................................................70
Critikon US Patents ..............................................................................70
Factory Default Settings ......................................................................71
SpO2
NELLCOR® OxiMax® SpO2 ................................................................................75
Description ................................................................................................................75
General Warnings ....................................................................................................76
General Cautions .....................................................................................................76
General Notes ..........................................................................................................77
Configuration Settings Associated with SpO2 ..................................................77
Buttons Associated with SpO2 .............................................................................78
Windows Associated with SpO2 ..........................................................................78
Indicators Associated with SpO2 .........................................................................78
Parameter Modes ....................................................................................................78
User Settings .............................................................................................................79
Limit settings ..........................................................................................79
Menu Settings ...........................................................................................................79
Sounds Associated with SpO2 ..............................................................................79
Procedures ................................................................................................................79
Warnings ....................................................................................................................80
Cautions .....................................................................................................................80
Alarms ........................................................................................................................81
Specifications ............................................................................................................85
Masimo Set® SpO2 ................................................................................................89
Description ................................................................................................................89
General Cautions .....................................................................................................90
General Notes ..........................................................................................................92
Configuration Settings Associated with SpO2 ..................................................92
Buttons Associated with SpO2 .............................................................................93
Windows Associated with SpO2 ..........................................................................93
Indicators Associated with SpO2 .........................................................................93
Patient safety .........................................................................................80
Patient safety .........................................................................................80
Troubleshooting ....................................................................................82
Measurement Range ...........................................................................85
Accuracy and Motion Tolerance ......................................................85
Saturation ..................................................................................85
Pulse Rate................................................................................. 85
NELLCOR® Sensor Accuracy ............................................................85
OxiMax® ...................................................................................85
OxiCliq® ...................................................................................85
Reusable Sensor Models .......................................................86
Factory Default Settings ......................................................................86
NELLCOR® Patents .............................................................................87
Page 8
Parameter Modes ....................................................................................................93
User Settings .............................................................................................................94
Limit settings ..........................................................................................94
Menu Settings ...........................................................................................................94
Sounds Associated with SpO2 ..............................................................................94
Procedures ................................................................................................................94
Warnings ....................................................................................................................95
Cautions .....................................................................................................................95
Alarms ........................................................................................................................96
Specifications ......................................................................................................... 100
Patient safety .........................................................................................95
Patient safety .........................................................................................95
Troubleshooting ....................................................................................97
Measurement Range ........................................................................ 100
Accuracy and Motion Tolerance ................................................... 100
Saturation ............................................................................... 100
Pulse Rate .............................................................................. 100
Masimo® Sensor Accuracy ................................................ 101
LNOP ......................................................................................101
Resolution .............................................................................. 101
Low Perfusion Performance ............................................... 101
Interfering Substances ......................................................................101
Factory Default Settings ................................................................... 102
Masimo Patents ................................................................................. 102
TURBOTEMP®
Description ............................................................................................................. 105
General Warning .................................................................................................. 106
General Cautions ..................................................................................................106
Configuration Settings Associated with Temperature ...................................107
Buttons Associated with Temperature ............................................................. 107
Windows Associated with Temperature .......................................................... 107
Indicators Associated with Temperature ......................................................... 107
Parameter Modes ................................................................................................. 109
User Settings ..........................................................................................................109
Menu Settings ........................................................................................................ 109
Sounds Associated with TURBO
Procedures for Oral Predictive Mode Determinations .................................109
Procedures for Rectal Predictive Mode Determinations ............................. 110
Procedures for Monitor Mode Determinations
(Axillary Determinations) .................................................................................. 111
Alarms ..................................................................................................................... 113
Specifications ......................................................................................................... 114
Predictive Mode ................................................................................105
Monitor Mode ................................................................................... 106
Measurement in Progress Indicators ............................................. 107
Predictive Mode ...................................................................107
Measurement NOT in Progress Indicators .................................. 108
TEMP® .................................................... 109
Factory Default Settings ................................................................... 114
IVAC® Patents ...................................................................................114
Page 9
Contents
Pulse Rate
Description ............................................................................................................. 117
General Notes ....................................................................................................... 117
Buttons Associated with Pulse Rate .................................................................. 117
Windows Associated with Pulse Rate ..............................................................118
Indicators Associated with Pulse Rate .............................................................. 118
Parameter Modes ................................................................................................. 118
User Settings ..........................................................................................................118
Menu Settings ........................................................................................................ 118
Sounds Associated with Pulse Rate ..................................................................118
Alarms ..................................................................................................................... 119
Specifications ......................................................................................................... 120
Appendix A: Principles of Noninvasive Blood Pressure Determination
Principles of Noninvasive Blood Pressure Determination ........................... 123
Limit settings ....................................................................................... 118
SpO2 Associated Alarms (If SpO2 is the source) .......... 119
BP Associated Alarms (If BP is the source) ..................... 119
Factory Default Settings ................................................................... 120
Systolic Search ................................................................................... 124
Reference Used to Determine NIBP Accuracy .......................... 125
DINAMAP® Monitors With Intra-Arterial Reference
(DINAMAP® Classic Technology) .................................125
DINAMAP® Monitors With Auscultatory Reference
(DINAMAP® Auscultatory Technology) ....................... 125
Appendix B: Maintenance
Maintenance .......................................................................................................... 129
Cleaning the Monitor ....................................................................... 129
Materials .................................................................................129
Cuff Cleaning and Disinfection ...................................................... 129
General ................................................................................... 129
Materials .................................................................................130
Procedure ...............................................................................130
Temperature Devices .......................................................................130
SpO2 Sensors ..................................................................................... 131
Storage and Battery Care ....................................................................................131
General Caution .................................................................................................... 133
Fuses ........................................................................................................................ 133
Calibration .............................................................................................................. 133
Leak Testing ...........................................................................................................134
Disposal of Product Waste ................................................................................. 134
Replacing the Battery ....................................................................... 132
Batteries ...............................................................................................134
Patient Applied Parts ........................................................................ 134
Packaging Material ............................................................................ 135
Monitor ................................................................................................135
Page 10
Appendix C: Connection Details
Connection Details ............................................................................................... 139
Host Port Connector (rear panel) .................................................. 139
Appendix D: Warranty, Service, and Spare Parts
Warranty, Service, and Spare Parts ...................................................................143
Warranty ................................................................................................................. 143
Assistance and Parts ............................................................................................. 143
Repairs ..................................................................................................................... 144
Packing Instructions ............................................................................................. 144
Service Manuals .................................................................................................... 144
Appendix E: Reorder Codes
Reorder Codes ......................................................................................................147
Page 11
DINAMAP
®
ProCare Monitor
Operation Manual
1
Page 12
DINAMAP® ProCare Monitor
2
Page 13
DINAMAP® ProCare Monitor
Operation Manual
This manual is for DINAMAP ProCare Monitors models 100, 200, 300, and 400, with or without printers.
ProCare 100: BP, Pulse
ProCare 200: BP, Pulse, and Temp
ProCare 300: BP, Pulse, and SpO
ProCare 400: BP, Pulse, Temp, and SpO
The model of the Monitor determines which parameters are in your monitor. Please refer to applicable sections.
Reissues and Updates
Changes occurring between issues are addressed through Change Information Sheets, Addendums, and replacement pages. If a Change Information Sheet does not accompany this manual, it is correct as printed.
Errors and Omissions
If errors or omissions are found in this manual, please notify: GE Medical Systems Information Technologies Technical Publications 4502 Woodland Corporate Boulevard Tampa, FL 33614 1-800-558-7044
2
2
Part No. 2009360-001 C
The content of this document including all figures and drawings is proprietary information of GE Medical Systems Information Technologies, provided solely for purposes of operation, maintenance or repair, and dissemination for other purposes or copying thereof is prohibited without prior written consent by GE Medical Systems Information Technologies, Tampa, Florida.
Illustrations may show design models; production units may incorporate changes.
3
Page 14
Hierarchy of Warnings and Cautions
A general warning is a statement that alerts the user to the possibility of injury, death, or other serious adverse reactions associated with the misuse of the device. A warning relates to steps in a procedure.
A general caution is a statement that alerts the user to the possibility of a problem with the device associated with its use or misuse. Such problems include device malfunction, device failure, damage to the device or damage to other property. A caution relates to steps in a procedure.
© Copyright 2002, 2004. GE Medical Systems Information Technologies. All rights reserved.
World Headquarters GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, WI 53223 USA Tel: + 1 414 355 5000
1 800 558 5120 (US only)
Fax: + 1 414 355 3790
European Representative GE Medical Systems Information Technologies GmbH Munzinger Straße 3-5 D-79111 Freiburg Germany Tel: + 49 761 45 43 - 0 Fax: + 49 761 45 43 - 233
Asia Headquarters GE Medical Systems Information Technologies Asia; GE (China) Co., Ltd. 24th Floor, Shanghai MAXDO Center, 8 Xing Yi Road, Hong Qiao Development Zone Shanghai 200336, P.R. China Tel: + 86 21 5257 4650 Fax: + 86 21 5208 2008
4
Page 15
Introduction
Page 16
6
Page 17
Introduction
About the DINAMAP
The ProCare Monitor provides a small, portable, easy-to-use monitoring alternative for sub-acute hospital and non­hospital settings. The battery-operated Monitor offers noninvasive determination of systolic blood pressure, diastolic blood pressure, mean arterial pressure, pulse rate, oxygen saturation, and temperature. Monitors are available with or without integrated printers. ProCare Monitors are intended for use in various markets, from the physician’s office to sub-acute triage and medical/surgical units.
ProCare 100: BP, Pulse
ProCare 200: BP, Pulse, and Temp
ProCare 300: BP, Pulse, and SpO
ProCare 400: BP, Pulse, Temp, and SpO
The model of the Monitor determines which parameters are in your monitor. Please refer to applicable sections.
Using the ProCare Monitor, a clinician can view, print, and recall clinical data that is derived from each parameter. The Monitor is also capable of alerting the clinician to changes in the patient’s condition or when it is unable to effectively monitor the patient’s condition. All of the main operations of the ProCare Monitor are easy-to-use and only a button-touch away. Please review the factory default settings and, where applicable, enter settings appropriate for your use.
®
ProCare Monitor
2
2
Indications
The ProCare Monitor is intended to monitor one patient at a time in a clinical setting.
Contraindications
This device is not designed, sold, or intended for use except as indicated.
Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
7
Page 18
Warnings
• Do not use the ProCare Monitor in the presence of magnetic resonance imaging (MRI) devices. There have been reports of sensors causing patient burns when operating in an MRI environment.
• Do not use the Monitor in the presence of flammable anesthetics.
• To help prevent unintended current return paths with the use of high frequency (HF) surgical equipment, ensure that the HF surgical neutral electrode is properly connected.
• To avoid personal injury, do not perform any servicing unless qualified to do so.
• WARNING: These Monitors should not be used on patients who are connected to cardiopulmonary bypass machines.
• If powering the Monitor from an external power adapter or converter, use only GE Medical Systems Information Technologies-approved power adapters and converters.
• The Monitor does not include any user-replaceable fuses. Refer servicing to qualified service personnel.
• To reduce the risk of electric shock, do not remove the cover or the back. Refer servicing to a qualified service person.
• If the accuracy of any determination reading is questionable, first check the patient’s vital signs by alternate means and then check the ProCare Monitor for proper functioning.
• Use of portable phones or other radio frequency (RF) emitting equipment near the system may cause unexpected or adverse operation.
• The equipment or system should not be used adjacent to, or stacked with, other equipment. If adjacent or stacked use is necessary, the equipment or system should be tested to verify normal operation in the configuration in which it is being used.
• The use of accessories, transducers and cables other than those specified may result in increased emissions or decreased immunity performance of the equipment or system.
8
Page 19
Introduction
Cautions
• Do not use replacement batteries other than the type supplied with the Monitor. Replacement batteries are available from GE Medical Systems - Accessories and
Supplies.
•The ProCare Monitor is designed to conform to Electromagnetic Compatibility (EMC) standard IEC 601-1-2, 1993 and will operate accurately in conjunction with other medical equipment which also meets this requirement. To avoid interference problems affecting the Monitor, do not use the Monitor in the presence of equipment which does not conform to these specifications.
• Place the ProCare Monitor on a rigid, secure surface. Monitor must only be used with mounting hardware, poles, and stands recommended by GE Medical Systems Information Technologies.
• The weight of the accessory basket contents should not exceed 5 lb (2.7kg).
• Arrange the external AC/DC power converter, air hoses, and all cables carefully so they do not constitute a hazard.
• Verify calibration of BP parameter (temp and pulse oximeter do not require calibration). Ensure that the display is functioning properly before operating the ProCare Monitor.
• Do not immerse the Monitor in water. If the Monitor is splashed with water or becomes wet, wipe it immediately with a dry cloth.
• Do not gas sterilize or autoclave.
•The ProCare Monitor, when used with GE Medical Systems Information Technologies-approved applied parts and accessories, is protected against defibrillator damage.
Note
• The electromagnetic compatibility profile of the ProCare Monitor may change if accessories other than those specified for use with the ProCare Monitor are used.
9
Page 20
Product Compliance
The DINAMAP® ProCare Monitor is classified in the following categories for compliance with IEC 601-1:
• Internally powered or Class II when powered from external supply
•Transportable
• For continuous operation
• Not suitable for use in the presence of flammable anesthetics
• Not for use in the presence of an oxygen-enriched atmosphere (oxygen tent)
• Type BF applied parts
• IPX1, degree of protection against ingress of water
• Sterilization/Disinfection, see Appendix B
• Software is developed in accordance with IEC 601-1-4.
• This equipment is suitable for connection to public mains via power adaptors as defined in CISPR 11.
•The SpO
parameter conforms to EN 865:1997 with
2
the exception of Clauses 36, 48, sub-clause
51.108.1and to ISO 9919.
• Defibrillation protected. When used with the recommended accessories, the Monitor is protected against the effects of defibrillator discharge. If monitoring is disrupted by the defibrillation, the Monitor will recover.
DINAMAP® PROCARE MONITOR CLASSIFIED WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL AND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH CAN/CSA C22.2 NO.
601.1. ALSO EVALUATED TO IEC-601-2-30.
This product conforms with the essential requirements of the Medical Device Directive. Accessories without the CE mark are not guaranteed to meet the Essential
0086
Requirements of the Medical Device Directive.
10
Page 21
Symbols
The following symbols are associated with the ProCare Monitor. Note: The model of the Monitor determines which symbols appear on it.
Attention, consult accompanying documents
Silence
Alarms
+ / - Increase / decrease adjustable settings
Menu
Inflate/Stop
Cycle
History
Print
On/Off
Battery Power
External communications port connector
Charging
Defibrillator-proof type BF equipment
External DC power input
Class II equipment according to IEC 60536
11
Page 22
Packaging label depicting the transportation and storage atmospheric pressure range of 500 to 1060 hPa.
IPX1
The DINAMAP® ProCare Monitor is protected against vertically falling drops of water and conforms with the IEC 529 standard at level of IPX1. Vertically falling drops shall have no harmful effects to the Monitor.
12
Page 23
Getting Started
Page 24
14
Page 25
Getting Started
Unpacking the Monitor and Accessories
Before attempting to use the ProCare Monitor, take a few minutes to become acquainted with the Monitor and its accessories. Unpack the items carefully. This is also a good time to check for any damage or accessory shortage. If there is a problem or shortage, contact GE Medical Systems Information Technologies.
It is recommended that all the packaging be retained, in case the Monitor must be returned for service in the future.
Setting up BP Connections
1. Connect the end of the air hose that has quick-release clips to the BP connector on the front of the Monitor. Make sure that the hose is not kinked or compressed. Note: To disconnect the hose from the Monitor, squeeze the quick-release clips together and pull the plug from the BP connector.
2. Select appropriate cuff size. Measure patient’s limb and select appropriately sized cuff according to size marked on cuff or cuff packaging. When cuff sizes overlap for a specified circumference, choose the larger size cuff. Precaution: Accuracy depends on use of proper size cuff.
3. Inspect cuff for damage. Replace cuff when aging, tearing, or weak closure is apparent. Do not inflate cuff when unwrapped. Precaution: Do not use cuff if damaged.
4. Connect the cuff to the air hose. Refer to the BP section for complete cuff connection instructions. Warning: It is mandatory that the appropriate hose and cuff combination be used. Any attempt to modify the hose will inhibit the Monitor from switching between the neonatal and adult measurement modes. Note: Care should be taken in reconnecting the cuff to a hose, ensuring that threads of the cuff and hose are in alignment and no cross-threading occurs.
15
Page 26
5. Refer to the BP section of this manual for complete instructions on taking an accurate BP determination.
Note: Use only CRITIKON
®
Blood Pressure Cuffs. The size, shape, and bladder characteristics can affect the performance of the instrument. Inaccurate readings may occur unless CRITIKON
®
Blood Pressure Cuffs are used.
Refer to Appendix E for reorder codes.
Setting up SpO2 Connections
1. Plug the appropriate SpO2 sensor into the SpO2 sensor extension cable.
2. Then plug the SpO sensor connector on the Monitor.
sensor extension cable into the SpO2
2
3. Refer to the applicable SpO complete instructions on monitoring SpO
section of this manual for
2
.
2
Setting up Temperature Connections
1. Connect the temperature probe cable to the temperature probe connector on the Monitor.
2. Insert the temperature probe into the probe holster at the side of the Monitor.
3. Refer to the
TURBO
section of this manual for complete instructions on taking a temperature reading.
TEMP®
16
Page 27
Getting Started
Setting up the Printer (Installing the Paper)
1. With the Monitor powered on, turn it so that the side with the printer is facing you.
2. While grasping the side of the Monitor, lift the printer door open by placing your thumb in the indented area and pulling. The printer door will pop open.
3.Place the roll of paper into the compartment so that the end of the paper comes off the right-side of the roll (paper is wound around the roll clockwise). Push the roll all the way to the back of the printer cavity, making sure the paper extends out of the printer cavity at least two inches.
4.Firmly press the door to close it.
5.Refer to the Printer section of this manual for complete instructions.
Power Sources
The ProCare Monitor is designed to operate from an internal lead­acid battery (see “Specifications” in Product Overview section).
Notes
• The ProCare Monitor is not designed to operate without an internal battery.
Be sure to unplug the Monitor before transport.
Powering the Monitor: Battery and DC Supply
Prior to each use, inspect the power supply cord to ensure proper connection and condition.
Before the ProCare Monitor is used for the first time, the battery should be charged in the Monitor for at least 8 hours.
17
Page 28
With external DC power connected, the green CHARGING indicator will light to indicate that the battery is charging.
When the Monitor is operating on battery power and the BATTERY LOW alarm is not active, the BATTERY indicator is backlit green. When the Monitor is operating on battery power and the BATTERY LOW alarm is active, the BATTERY indicator flashes green, the LOW indicator flashes amber, and the medium priority alarm sounds until it is acknowledged. Press the Silence button to acknowledge this alarm. Once it is acknowledged, the indicators continue to flash, but the audible alarm is silenced for 10 minutes. Once the BATTERY LOW condition becomes active, the Monitor should be connected to a DC power supply to recharge the battery (refer to “Power Requirements” in Product Overview section). If the Monitor continues to be used without charging the battery, the Monitor eventually enters a fail-safe mode (E13: BATTERY TOO LOW TO OPERATE). Refer to the Alarms Section for information of an E13 error code.
Battery charging will take place as long as the Monitor remains connected to an external DC power source.
Notes
• To prolong the life of the battery, keep the Monitor connected to a DC power supply whenever possible. NEVER allow the battery to become completely discharged. A fully charged battery will
power the Monitor for approximately 5 hours. To ensure
full charge cycles, replace only with a recommended battery. If the Monitor is to be stored for some time, first charge the battery and then remove it and store it separately from the Monitor.
18
Page 29
Getting Started
Turning the Monitor On and Off
To turn the ProCare Monitor on, push the power On/Off button.
As the Monitor powers up, it will run a short self-test routine, which will flash all the indicator lights and then beep the warning speaker.
To turn the Monitor off, push the power On/Off button again. This will terminate any measurements that may be in progress and automatically deflate the cuff. Note: Pressing and holding the On/Off button for 15 seconds will reset the CPU processor. After resetting the CPU processor, check all the unit configuration settings.
Configuration Mode Settings
Monitor settings such as HIGH/LOW alarm settings changed in the Clinical Mode will not be retained after the monitor is powered off. To retain alarm and parameter settings, the changes must be done in the configuration mode. Date/Time settings are also entered in the configuration mode.
To enter the configuration mode: with the Monitor off, press and hold the Menu button at the same time as pressing and holding the On/Off button for 3 seconds. The Monitor enters the configuration mode.
As the monitor turns on in the configuration mode, a brief display appears showing the software revision and the BP technology of the Monitor. These displays appear only during the first part of the power up sequence and are not selectable and cannot be changed.
Display
Major software revision Systolic Minor software revision Diastolic Type of BP technology min
19
Window
Page 30
The Menu selections appear in the following order. Refer the each manual section for settings options. Note: Menu selections for a Masimo unit have three SpO settings. Refer to the Masimo SpO
section for options.
2
2
Setting:
Window LED Display
• Inflate
Pressure: Diastolic XXX (numeric)
• SpO
• SpO2 Sat: SpO
• SpO2* Sensitivity: SpO
Mode: SpO
2
2
2
2
• Temp: °C or °F
• Year: Systolic
• Month: MAP/Cuff
• Day: Diastolic
• Hour: min
• Minute: min
• Mode: Systolic
*Masimo units only
Setting the Date and Time
To set the date and time on the ProCare Monitor, you must access the configuration mode. Press MENU to skip the default settings that do not require changes. The following list shows the windows in which the date/time settings appear.
Setting:
• Year: Systolic
• Month: MAP/Cuff
• Day: Diastolic
• Hour: min
• Minute: min
Window LED Display
Procedures
1. Press the Menu button to move from one setting to another. Use the +/- buttons to increment or decrement the setting.
20
Page 31
Getting Started
Note: For the date and time to be saved, you must
advance the menu through the minute setting.
2. To exit the configuration mode, press the ON/OFF button.
3. To continue with other changes, press the Menu button. CFG will appear in the Systolic window. To change parameter settings, press the Menu button and select the parameter function. To change alarm settings, press the Alarms button.
Inflation Pressure Default Settings (Refer to the BP Section for options)
Procedures
1. Enter the configuration mode: with the Monitor off, press and hold the Menu button at the same time as pressing and holding the On/Off button for 3 seconds, or press Menu until the Inflate Pressure is lit on the display and the pressure is displayed in the Diastolic window.
2. Use the +/- buttons to increment or decrement the inflate pressure default setting. Increments are 5mmhg from 100mmhg to 250mmhg.
3. To exit the configuration mode, turn the unit off. To continue with additional configuration settings, press Menu.
Alarm Default Settings (Refer to Alarms Section for options)
Procedures
1. Enter the configuration mode: with the Monitor off, press and hold the Menu button at the same time as pressing and holding the On/Off button for 3 seconds. After the unit enters the configuration mode, press Alarms. At any point in the configuration mode menu, Alarms default may be selected.
2. To set or change the default setting, press the Alarms button to select alarm setting. Use the +/- buttons to increment or decrement the individual settings. Note: For the Alarms default setting to be saved, you must advance the menu through the SPO2 settings.
3. To exit the configuration mode, turn the unit off. To continue with additional configuration settings, press Menu.
21
Page 32
SPO2 Configuration Settings
Procedure for units with Nellcor® Technology
(Refer to the Nellcor® section for options)
1. Enter the configuration mode: with the Monitor off, press and hold the Menu button at the same time as pressing and holding the On/Off button for 3 seconds.
2. Press the menu button until Rate window.
3. Use the +/- buttons to select the option.
4. Press the Menu button once. SAt appears in the Pulse Rate window.
5. Use the +/- buttons to select the option.
6. To exit the configuration mode, turn the unit off. To continue with additional configuration settings, press
Menu.
Procedure for units with Masimo® Technology (Refer to the Masimo® section for options)
1. Enter the configuration mode: with the Monitor off, press and hold the Menu button at the same time as pressing and holding the On/Off button for 3 seconds.
2. Press the Menu button until Rate window.
3. Use the +/- buttons to select the option.
4. Press the Menu button once. SAt appears in the Pulse Rate window.
5. Use the +/- buttons to select the option.
6. Press the Menu button once. SEn appears in the Pulse Rate window.
7. Use the +/- buttons to select the option.
8. To exit the configuration mode, turn the unit off. To continue with additional configuration settings, press
Menu.
n0d appears in the Pulse
n0d appears in the Pulse
22
Page 33
Product Overview
Page 34
24
Page 35
Product Overview
Buttons
1 Silence button: Press to mute audible alarms. Any alarm
active that is acknowledgeable is also removed whenever this key is pressed. When pressed after alarm sounds (silence active), the silence icon (bell) lights to indicate that audible alarms have been silenced for 2 minutes
2 Alarms button: Press to view or adjust parameter alarm
settings
3 +/- button (Plus/Minus): Press the + button to increase an
adjustable setting and the - button to decrease an adjustable setting. This button is active only when a user­setting mode (limit or menu) is active
4 Menu button: Press to access menu settings that can be
adjusted while in clinical mode (i.e., ALARM VOLUME,
PULSE VOLUME, INFLATE PRESSURE; refer to Operating Modes in this section for a description of clinical mode)
5 SpO
6 BP connector: BP cuff hose attaches here 7 Inflate/Stop button: Press to start a manual BP
8 Temperature probe holster: Temperature probe is stored
9 Cycle button: Press to start Auto Cycle or STAT mode
sensor connector: SpO2 sensor extension cable
2
attaches here
determination or stop any BP determination
here
25
Page 36
10 Temperature probe cover storage: Box of probe covers is
stored here
11 History button: Press to activate the history mode. When
activated, it displays the most recent entries stored. Press and hold the button for 2 seconds to clear all entries stored
12 Print button: Press to print currently displayed values or all
stored entries when in history mode
13 On/Off button: Controls on/off state of monitor; push for
power on and push again for power off
14 Temperature probe connector: Temperature probe cable
attaches here
26
Page 37
Product Overview
Front Panel
15 Silence icon: when Silence button is pressed after alarm
sounds (silence active), silence icon (bell) lights to indicate that audible alarms have been silenced for 2 minutes
16 Systolic window: 3-digit red LED indicates measured
systolic BP in mmHg
17 Diastolic window: 3-digit red LED indicates measured
diastolic BP in mmHg
18 Alarm volume indicator: lights to indicate you are making
a change to the alarm volume
19 Pulse volume: lights to indicate you are making a change
to the pulse volume
20 Inflate pressure: lights to indicate you are making a
change to the inflation pressure
21 Pulse Rate window: 3-digit yellow LED shows pulse rate in
beats per minute
22 SpO
23 SpO
pulse indicator: Red LED bar flashes to indicate that
2
real-time pulse rate measurements are being derived from SpO
signals
2
window: 3-digit red LED indicates oxygen saturation
2
in %
27
Page 38
24 MAP/Cuff window: 3-digit red LED indicates measured
MAP in mmHg and shows instantaneous cuff pressure during BP determination
25 Min window: Displays the BP mode if manual or STAT is
the cycle time when in Auto Cycle mode
26 Battery power indicator: Green LED indicates the Monitor
is operating on battery power
27 Low battery power indicator: Yellow LED indicates LOW
charge status of internal battery
28 Charging indicator: Green LED indicates presence of
external power source and battery charging
29 Temperature window: 4-digit red LED indicates measured
temperature
28
Page 39
Product Overview
Rear Panel
30 Data interface connector: Host communications port
(15 pin D-type RS-232 serial port) for use only with equipment conforming to IEC 601-1, configured to comply with IEC 601-1-1
31 Printer compartment
29
Page 40
Left-Side Panel
32 Printer compartment
Right-Side Panel
33 External DC power socket: To be used with approved GE
Medical Systems Information Technologies AC-DC power converter ONLY
30
Page 41
Product Overview
Windows
Each derived vital sign has an associated window for displaying the value. For each window, the vital sign’s name and unit of measure are labeled above and to the right of it, respectively. An additional window--the min window--is available for displaying the BP mode or chosen AUTO CYCLE selection.
Indicators
Indicators are text messages and icons that are positioned on the front of the Monitor and can be either backlit red or green. For each vital sign that has user-adjustable limits, two indicators (HIGH, LOW) appear to the right of its window.
Operating Modes
The ProCare Monitor can operate in one of four modes: clinical, configuration, advanced configuration, and service.
Clinical mode is the Monitor’s normal operating mode. While this mode is active, alarm limits and a few other commonly used settings are adjustable. All parameters are available for monitoring in this mode.
Configuration and advanced configuration modes display the software revision and allows you to configure defaults for some settings that are available in clinical mode, as well as some less commonly used settings that are only adjustable in these modes. No parameters are operable in these modes, therefore, patient monitoring should be suspended.
Entering Configuration Mode
1. With the Monitor off, press and hold the Menu button at the same time as pressing the On/Off button for 3 seconds.
2. The Monitor automatically switches on in configuration mode.
Refer to the Service Manual for instructions for use concerning advanced configuration and service mode.
31
Page 42
Parameter Modes
The ProCare Monitor has three parameter modes: offline, ready, or operate.
A parameter is in offline mode when its vital sign(s) are not checked against user-set limits, alarm conditions detected by the parameter do not generate alarms, and no vital sign(s) data is displayed.
A parameter is in ready mode when its vital signs are not checked against user-set limits, alarm conditions detected by the parameter do not generate alarms, and no vital sign(s) data is displayed. However, information regarding the connection of a sensor is communicated in this mode.
A parameter is in operate mode when the appropriate vital signs are being checked against user-set limits and alarming conditions detected by the parameter generate alarms. Vital sign(s) data from the parameter is displayed.
When the Monitor is turned on and no sensor is connected, the parameter remains in offline mode. Upon detection of a sensor, the parameter auto-switches to ready mode. Upon detection of valid patient data, the parameter auto-switches to operate mode.
User Modes
The ProCare Monitor has four user modes that are available during clinical operating mode: Menu, Cycle, Limit
Adjustment, and History.
The Menu mode allows you to access and change the three settings associated with the following indicators: ALARM VOLUME, PULSE VOLUME, AND INFLATE PRESSURE. To activate this mode, press the Menu button. With each press of the Menu button, the indicator appears in flashing green with the associated value appearing in red in the Diastolic window. As the Menu button is pressed, the indicators appear in the following order: ALARM VOLUME, PULSE VOLUME, AND INFLATE PRESSURE. To change the associated value, simply use the +/- button to increment or decrement, respectively. After 15 seconds of not pressing the
32
Page 43
Product Overview
Menu button, the Menu mode is automatically exited. Otherwise, you can exit the Menu mode by pressing the Menu button one more time after viewing the oldest entry.
Upon exiting Menu mode, the main monitoring screen is displayed. Alarm and pulse volume settings are retained after power-off. INFLATE PRESSURE is reset to its configured default after power-off.
The Cycle mode allows you to access Auto Cycle and STAT modes. Refer to the BP section for more information.
The History mode allows you to access the stored patient data. Refer to the History section for more information.
The Limit Adjustment mode allows you to change alarm limit settings that are used while monitoring a patient. All limit alarm settings return to their default settings after power-off. The range and increment/decrement steps for each derived vital sign that has adjustable limits are described in each parameter section. The step size specified (which cannot be adjusted) tells how much the limit value will change per increment/decrement key press and also dictates how close together a pair of limits can be.
Limit-adjustable vital signs are displayed in the following order: Systolic (HIGH, LOW), Diastolic (HIGH, LOW), Pulse Rate (HIGH, LOW), and SpO Note: The Temperature and MAP (mean arterial pressure) vital signs are not checked against alarm limits.
(HIGH, LOW).
2
Sounds
The Monitor generates sounds based upon user interaction, parameter events, parameter and system alarms, and low battery alarms.
User interaction sounds include positive key tone and negative key tone. The positive key tone sounds one high­pitched tone when you press a button and the Monitor is able to perform the intended function. The negative key tone sounds three low-pitched tones when you press a button and the Monitor is unable to perform the intended function.
33
Page 44
Parameter-specific sounds are related to the functioning of each parameter: NIBP, Temperature, and SpO
. Refer to the
2
individual parameter sections for definitions and sounds.
Alarm-specific sounds include medium and high priority. The medium-priority alarm sounds three high-pitched tones. The high-priority alarm sounds three high-pitched tones followed by two high-pitched tones. When alarms of both priorities are active, only the high-priority alarm sound is audible.
A sound is generated when AC power is lost while patient monitoring is active.
Power Sources
The ProCare Monitor is designed to operate from an internal lead-acid battery. For replacement rechargeable batteries, please refer to the Service section of this manual.
34
Page 45
Product Overview
Specifications
Mechanical
Dimensions Height: 9.7 in (24.7 cm)
Width: 8.6 in (21.9 cm) without Temperature
10.0 in (25.4 cm) with Temp Depth: 5.3 in (13.5 cm)
Weight, Including Battery 5.68 lb (2.58 kg)
Mountings Self-supporting on rubber feet
Portability Carried by handle
Classification Information Mode of operation:
continuous Degree of protection against harmful ingress of water: Drip­proof IPX1
Power Requirements
Power Converter US: P/N: 2009460-001
Protection against electrical
shock: Class II AC input: 120 VAC/60 Hz 24W DC output voltage; 12VDC at 1A The AC mains power adapter contains a nonresettable and nonreplaceable fuse.
Power Converter UK & EUR: P/N UK: 2008538-001
P/N EUR: 2009539-001
Protection against electrical
shock: Class II AC input: 230-240 VAC/50 Hz 92mA DC output voltage; 12VDC at 1A
35
Page 46
The AC mains power adapter contains a nonresettable and nonreplaceable fuse.
Monitor Protection against electrical
shock: Internally powered or
Class II when powered from
specified external power
supply.
DC input voltage: 12 VDC,
supplied from a source
conforming to IEC 601-1.
Fuses: The Monitor contains
four fuses. The fuses are auto-
resettable and mounted within
the Monitor. The fuses protect
the low voltage DC input, the
battery, the remote alarm
output, and the +5 V output
on the host port connector.
Battery: 6 volt, 3.3 amp-hours
protected by internal auto-
resetting fuse and thermal
protection.
Minimum operation time:
5 hrs (5 min cycle with adult
cuff at 25 °C, SpO
active at
2
60 bpm, Temp in monitor
mode) from full charge.
Time
for full recharge: Approx. 5 hrs
from full discharge when the
Monitor is switched off and
approx. 8 hrs when the Monitor
is switched on.
Environmental
Operating Temperature + 5 °C to + 40 °C
(+ 41 °F to + 104 °F)
Operating Atmospheric Pressure 700 hPa to 1060 hPa
36
Page 47
Product Overview
Storage Temperature – 20 °C to + 50 °C
(– 4 °F to + 122 °F)
Storage/Transportation Atmospheric Pressure 500 hPa to 1060 hPa
Humidity Range 5% to 95% noncondensing
Radio Frequency Complies with IEC
Publication 601-1-2 (April
1993) Medical Electrical Equipment, Electromagnetic Compatibility Requirements and Tests and CISPR 11 (Group 1, Class B) for radiated and conducted emissions.
The DINAMAP® ProCare Monitor is protected against vertically falling drops of
IPX1
water and conforms with the IEC 529 standard at level of IPX1. No harmful effects will come of vertically falling drops of water making contact with the Monitor.
37
Page 48
38
Page 49
Printer
Page 50
40
Page 51
Printer
Description
The ProCare Monitor comes with an optional text-only printer. Each time a printout is started, the following information is printed: GE Medical Systems header, monitor name; model number; current software revision; a place for patient name and hand-written comments; a place for the column labels, parameter labels, or unit of measure; date; and vital signs data, if available. On some models, a pencil icon
appears instead of the name and comments labels.
Installing the Paper
Refer to the Getting Started section for instructions.
Print Button
The Print button initiates two types of printouts: currently displayed values and History.
By pressing the Print button while the main monitoring screen is viewable, a printout of the currently displayed values is generated. Since the currently displayed values may have been derived at different times (e.g., the BP values are 5 minutes old, while Temp was just completed), the Monitor prints the value along with a time stamp of when that value was derived. Values are printed in order of the most recent to the oldest.
By pressing the Print button when in History mode, all entries currently stored in the history are printed in order of the most recent to the oldest. Note: If the Print button was pressed during the first 10 seconds of SpO and pulse rate readings.
The availability of the printer is determined at the time the printout is started. When the printer is unavailable, the Print button makes a negative key sound when you press it. The printer is unavailable if it is out of paper, too hot, or if the battery is too low, or if the Monitor is in any of the following modes: Cycle, Alarm limit adjustment, Menu, config, or service.
If no print is visible on the paper, check that the paper roll has been installed in the correct position (refer to diagram).
monitoring, dashes will appear for SpO2
2
41
Page 52
To tear off the printout, use a slight sideways action to pull the paper sharply up across the edge of the door.
Printer Alarms
When any of the alarm conditions occur, the alarm code flashes in the min window and a high-priority alarm sounds. When a printer alarm condition is active, you can acknowledge the alarm by pressing the Silence button.
E10: PRINTER NO PAPER This alarm is generated when the Print button is pressed and the printer detects that there is no paper.
E11: PRINTER TOO HOT This alarm is generated when a printout is in progress and the printer becomes too hot to print.
E12: BATTERY TOO LOW TO PRINT This alarm is generated when a printout is in progress and the battery becomes too low to print.
Storage
Store thermal paper in a cool, dry place. The printed strip (thermal paper recording) should not be
• exposed to direct sunlight,
• exposed to temperatures over 100 °F/38 °C or relative humidity over 80%,
• placed in contact with adhesives, adhesive tapes, or plasticizers such as those found in all PVC page protectors.
Note: When in doubt about long-term storage conditions, store a photocopy of the thermal paper recording.
Cautions
• The paper is thermally activated; therefore, do not store it in a hot place as discoloration may result.
• Use only replacement paper rolls (P/N 770137 for single rolls, 089100 for box of 10) from GE Medical Systems - Accessories and Supplies.
42
Page 53
Alarms
Page 54
44
Page 55
Alarms
The ProCare Monitor provides visible and audible indications of patient-, and system-related alarm conditions. An alarm can generate an audible indication, visual indication, alarm message code, and electronic record in the history.
Alarm Codes
All alarm indications are accompanied by an audible signal unless the Silence button is selected. A microprocessor system failure will generate a high-pitched audible alarm regardless of the setting of the Silence button.
There are three categories of alarms: patient-related (limit and parameter status alarms), system-related (printer, battery, and memory alarms), and system failures.
Adjusting Alarm Limits
Alarms Button
Pressing the Alarms button activates the limit adjustment user mode. When the limit adjustment mode is initially activated, it first displays the systolic high limit value with the HIGH indicator flashing in red. With each subsequent press of the Alarms button, the next limit in the sequence is displayed for adjustment. Press the + button to increase the displayed limit value and the - button to decrease the displayed limit value.
To exit limit adjustment mode after setting alarms, you must progress through all limits until you reach the last one. Then, after reaching the last limit, immediately press the Alarms button. The main monitoring screen appears indicating that you have exited Alarms mode. If monitoring is active, the current vital signs appear after you have completed your alarm setting changes. Or, you can let the mode time-out by not touching the Alarms button until the main monitoring screen appears indicating that you have exited Alarms mode. The Monitor returns to all default limit settings after power­off; these defaults are adjustable via configuration mode.
Adjustable vital signs limits are displayed in the following order: Systolic (HIGH, LOW), Diastolic (HIGH, LOW), Pulse Rate (HIGH, LOW), and SpO parameter section for ranges.
(HIGH, LOW). Refer to each
2
45
Page 56
Note: The Temperature and MAP vital signs are not checked against alarm limits.
Adjusting the Alarm Volume
You can adjust the alarm volume by pressing the Menu button. ALARM VOLUME will flash in green with the value in the Diastolic window. You can set the ALARM VOLUME range from 1 to 10 (10 being the loudest). Changing the volume applies to all alarms.
The positive key tone that sounds when you press the +/­buttons relates directly to the user-set alarm volume. Refer to User Modes.
Silencing and Acknowledging an Alarm
Silence Button
To silence a patient-related alarm (limit and parameter status alarms) at anytime, press the Silence button. The silence icon (a bell) lights to indicate that audible alarms have been silenced for 2 minutes.
Some patient- and system-related alarms are acknowledgeable. (Refer to the Alarm Message Codes table at the back of this section.) For acknowledgeable alarms that are active when the Silence button is pressed, any associated audible or visual indication is removed and any associated alarm message code is no longer displayed.
The silence icon has three states:
• Solid red: silence icon is active.
• Blinking red: alarm silence is not active and at least one alarm condition is active.
• Off: alarm silence is not active and no alarm condition is active.
Alarm Sounds
The Monitor produces two different alarm sounds based upon the priority of the alarm: medium and high priority. The medium-priority alarm sounds three high-pitched tones. The high-priority alarm sounds three high-pitched tones followed
46
Page 57
Alarms
by two high-pitched tones. When alarms of both priorities are active, only the high-priority alarm sound is audible.
Alarm Detection and Priorities
Limit Alarms
The Monitor checks each derived vital sign (except MAP and Temperature) against user-set limits. A high-limit alarm is generated when that value is greater than its high limit. A low-limit alarm is generated when that value is less than its low limit. All limit alarms are considered high priority.
Parameter Range
Systolic High 35 - 290 200
Systolic Low 30 - 285 80
Diastolic High 15 - 220 120
Diastolic Low 10 - 215 30
Pulse Rate High: NELLCOR 35 - 250 150
Pulse Rate Low: NELLCOR 30 - 245 50
Pulse Rate High: MASIMO 35 - 235 150
Pulse Rate Low: MASIMO 30 - 230 50
High 21 - 100 100
SpO
2
SpO
Low 20 - 99 90
2
*To change Alarm Default settings, refer to Config Mode in
Getting Started section.
Factory
Default*
When a limit violation occurs, the following happens:
• The derived vital sign that is out of limits and the associated HIGH or LOW indicator flash.
• The high priority alarm sound becomes audible unless silence is active.
47
Page 58
Parameter Status Alarms
The Monitor generates parameter status alarms when unusual patient or sensor conditions are detected. All parameter status alarms are considered high priority alarms. Refer to the Alarm Message Codes table in this section.
When a parameter status alarm occurs, the following happens:
• Its code flashes in the associated window.
• The high priority alarm sound becomes audible unless alarm silence is active.
Printer Alarms
Refer to Printer section.
Memory Alarms
Memory alarms occur when battery-backed RAM is corrupted. When it occurs, all settings are reset to their factory defaults. When detected, the alarm message code appears in the Systolic window and the medium-priority alarm sounds.
Battery Alarms
Two types of battery alarms exist: a BATTERY LOW alarm indicating the battery needs recharging and an E13 BATTERY TOO LOW TO OPERATE indicating that the battery is so low monitoring has stopped. For information on the BATTERY LOW alarm, refer to “Powering the Monitor” in the Getting Started section.
When the E13 BATTERY TOO LOW TO OPERATE is detected, the Monitor ceases all monitoring, blanks the display except for the Low Battery indicators, and displays E13 in the Systolic window. A medium priority alarm sounds and continues to sound for 10 minutes, unless the unit is powered off.
After the first 10 minutes, if the Monitor was idle (not monitoring) when the condition was detected, then it automatically powers off.
48
Page 59
Alarms
Otherwise, if the unit had been monitoring when the condition was detected, the alarm changes to the critical priority alarm level. The Monitor continues to display the Low Battery indicators and error code E13 as before. This continues for 10 minutes, unless the unit is powered off. At the end of this period, if the Monitor has not be shut off, it then automatically powers off.
System Failure Alarms
System failures are caused by a system software or hardware failure. When a system failure occurs, the error code is displayed in the Systolic window, and the Monitor enters a fail-safe mode in which a constant alarm tone sounds for up to 10 minutes before the tone is automatically switched off. Refer to the Alarm Message Codes table in this section.
49
Page 60
Specifications
Default Settings Alarm Volume 5
50
Page 61
Alarms
Alarm
Message
Code
E89 BP NO DETERMINATION Yes
E85 BP LEVEL TIMEOUT Yes
E84 BP TOTAL TIMEOUT Yes
E83 BP INFLATION TIMEOUT Yes
E82 EXCESS AIR IN CUFF Yes
E80 BP OVERPRESSURE Yes
E21 SPO2 REPLACE SENSOR Yes
E22
E23 SPO2 SENSOR OFF FINGER Yes
E25 SPO2 NO SIGNAL Yes
Alarm Detected
SPO2 REPOSITION SENSOR
(Masimo ONLY)
Acknowledgeable by
pressing Silence?*
Yes
E61 TEMP PROBE BROKEN No
E63
E66 TEMP PROBE TOO HOT No
E10 PRINTER NO PAPER Yes
E11 PRINTER TOO HOT Yes
E12 BATTERY TOO LOW TO PRINT Yes
E13 BATTERY TOO LOW TO OPERATE No
TEMP DISCONNECTED OR
WRONG TYPE PROBE
51
Yes
Page 62
E00 MEMORY LOST Yes
Alarm
Message
Code
930
940
950
951
952
970
971
Alarm Detected
SYSTEM FAILURE**
SpO2 no status from module for 30 ±10s,
fatal error reported by module
SYSTEM FAILURE**
TEMP data samples less than 45 in 5s interval
SYSTEM FAILURE**
NIBP pump on during idle or overcurrent
detected
SYSTEM FAILURE**
NIBP valve stuck closed during cuff typing
SYSTEM FAILURE**
NIBP PT2 higher than 150 for greater than
15s while idle
SYSTEM FAILURE**
Time base failure
SYSTEM FAILURE**
RAM test failure
Acknowledgeable by
pressing Silence?*
No
No
No
No
No
No
No
972
SYSTEM FAILURE**
ROM checksum failure
No
SYSTEM FAILURE**
973
Secondary SPI communication error during
initialization
No
SYSTEM FAILURE**
974
Calibration data invalid on initialization or
unit never calibrated
No
SYSTEM FAILURE**
975
*Acknowledging an alarm by pressing the Silence button, cancels the alarm.
** Refer to the ProCare Service Manual or call Technical Support for definitions and instructions.
Calibration data storage exhausted while
saving data in cal mode
No
52
Page 63
History
Page 64
54
Page 65
History
Description
The History mode allows you to access the stored patient data. The ProCare Monitor can hold up to 40 stored entries in history. When full, the oldest entry is removed so the most recent entry can be stored. Additionally, entries are automatically removed when they become older than 24 hours.
The age of each entry is maintained and displayed in the min window with a minus sign (-) in front of it when other data stored for that entry is displayed. For entries that are greater than 59 minutes old, the age is displayed as HH:MM (hour:min). For entries that are less than or equal to 59 minutes old, the age is displayed in total minutes.
When viewing entries in History that are out of limits, the corresponding HIGH or LOW indicator appears in red.
An entry is stored in history at the completion of a BP determination and at the completion of a successful predictive temperature measurement. At the end of a BP determination, systolic, diastolic, MAP, pulse rate, and SpO values are stored. At the end of a temperature determination, only the temperature value is stored.
2
Buttons Associated with History
To activate the History mode, press the History button. The HISTORY indicator flashes green while this mode is active.
With each press of the History button, the patient data stored with the next oldest entry is displayed. Entries are displayed from the most recent to the oldest. For example, the most recent entry could have an age of -0 minutes and the oldest entry could have an age of -23:59.
After 15 seconds of not pressing the History button, the History mode is automatically exited. Otherwise, you can exit the History mode by pressing the History button one more time after viewing the oldest entry. Upon exiting
History mode, the main monitoring screen is displayed.
Erasing Stored History
To erase stored patient data, press and hold the History button for a minimum of 2 seconds. All entries that were
55
Page 66
stored in history, as well as any patient data displayed on the Monitor that relates to the previous determination or the previous temperature measurement is erased.
Windows Associated with History
Each window on the Monitor can be active during History mode. When the History button is pressed the patient data stored for each entry is displayed in the applicable windows. Patient data is displayed from most recent to oldest, indicated by the age in the min window.
Indicators Associated with History
The History indicator is used to show the state of the history mode. When History mode is active, the History indicator flashes green.
56
Page 67
BP
Page 68
58
Page 69
BP
Description
The BP parameter in the ProCare Monitor is available with two types of BP technologies: one calibrated to intra-arterial pressure and one calibrated to the auscultatory method (specific technologies are available in select markets). Please refer to the front of your ProCare Monitor to see which BP technology your monitor contains. The monitors containing auscultatory technology will have an auscultatory label above the model number on the front of the Monitor. Then, refer to the Appendix A: Principles of Noninvasive Blood Pressure Determination for specific information regarding both technologies. All user interface options, instructions for use, and alarms will be the same for both technologies. The BP parameter is included in all models. Blood pressure is monitored noninvasively in the ProCare Monitor by oscillometric method. Note: For neonatal populations, the reference is always the intra-arterial pressure monitoring method.
The ProCare Monitor has three BP modes: 1. Manual, 2. Auto Cycle, and 3. STAT. The mode is selected by the user. The actual BP determination is automated and, once it is complete, the values for systolic pressure, diastolic pressure, mean arterial pressure, and pulse rate are shown in their respective windows.
Before each BP determination, the Monitor performs a test to ensure that the cuff pressure is below a specified level. The determination is delayed until this condition is met. The Monitor senses the type of hose being used and automatically uses adult/pediatric monitoring parameters or neonatal monitoring parameters, as appropriate.
Audible and visible alarms occur when any of the values for systolic pressure, diastolic pressure, or pulse rate are outside their selected high or low limits. Note: The Monitor will continue to operate until the battery is completely depleted in order to obtain the full use of the battery. However, if the battery reaches its “empty” point during a BP determination, it will simply stop in the middle of the determination.
59
Page 70
Instructions for cleaning and disinfecting BP cuffs are in Appendix B.
General Warnings
• Connect cuffs and inflation systems only to systems designed for non-invasive blood pressure monitoring. Devices with luers and locking luer connectors may be inadvertently connected to intravascular fluid systems that may allow air to be pumped into a blood vessel.
•The ProCare Monitor will not measure blood pressure effectively on patients who are experiencing seizures or tremors.
• Arrhythmias will increase the time required by the ProCare Monitor to determine a blood pressure and may extend the time beyond the capabilities of the Monitor.
• In Manual mode, the ProCare Monitor displays the results of the last blood pressure determination for 30 minutes or until another determination is completed. If a patient’s condition changes between one determination and the next, the Monitor will not detect the change or indicate an alarm condition.
• Devices that exert pressure on tissue have been associated with purpura, skin avulsion, compartmental syndrome, ischemia and/or neuropathy. To minimize these potential problems, especially when monitoring at frequent intervals or over extended periods of time, make sure the cuff is applied appropriately and examine the cuff site and the limb distal to the cuff regularly for signs of impeded blood flow.
• Do not apply external pressure against cuff while monitoring. Doing so may cause inaccurate blood pressure values.
• Use care when placing cuff on extremity used to monitor other patient parameters.
•The ProCare Monitor is designed for use only with CRITIKON
®
dual-tube cuffs.
60
Page 71
BP
• Use only accessories approved for use with DINAMAP® Monitors. Failure to use recommended accessories may result in inaccurate readings.
• Blood pressure cuffs should be removed from the patient when the Monitor is powered off. If the extremity remains cuffed under these conditions or if the interval between blood pressure determinations is prolonged, the patient’s limb should be observed frequently and the cuff placement site should be rotated as needed.
General Cautions
• Accuracy of BP measurement depends on using a cuff of the proper size. It is essential to measure the circumference of the limb and to select the proper size cuff. In addition, the air hoses are color-coded according to patient population. The gray 12- or 24-foot hose (3.66 m or 7.3 m) is required on patients who require cuff sizes from infant through thigh cuffs. The light blue 12-foot hose (3.66 m) is required for the neonatal cuff sizes #1 through #5.
• If it becomes necessary to move the cuff to another limb, make sure the appropriate size cuff is used.
• The pulse rate derived from a BP determination may differ from the heart rate derived from an EKG waveform because the ProCare Monitor measures actual peripheral pulses, not electrical signals or contractions from the heart. Differences may occur because electrical signals at the heart occasionally fail to produce a peripheral pulse or the patient may have poor peripheral perfusion. Also, if a patient’s beat-to­beat pulse amplitude varies significantly (e.g., because of pulsus alternans, atrial fibrillation, or the use of a rapid-cycling artificial ventilator), blood pressure and pulse rate readings can be erratic, and an alternate measuring method should be used for confirmation.
General Notes
• A patient’s vital signs may vary dramatically during the use of cardiovascular agents such as those that raise or
61
Page 72
lower blood pressure or those that increase or decrease heart rate.
• Several conditions may cause the BP parameter to calculate and display only the mean arterial pressure (MAP) without systolic and diastolic readings. These conditions include very low systolic and amplitude fluctuations, so an accurate calculation for these values can’t be made (e.g., patient in shock); too small of a difference between systolic and MAP calculations in relationship to the difference between diastolic and MAP; or a leak has occurred in the ProCare Monitor. If only the MAP value is displayed, an alarm message code is displayed in the systolic window, while the diastolic window remains blank.
• This equipment is suitable for use in the presence of electrosurgery.
Buttons Associated with BP
The buttons associated with BP are Inflate/Stop and Cycle.
The Inflate/Stop button starts and stops BP determinations. When a determination is in progress, pressing this button stops the determination. While in STAT mode, pressing this button cancels STAT mode, as well a determination if in progress. When in Auto Cycle mode, pressing this button starts a determination or cancels a determination if in progress; it does not change the mode. Note: While an E80: BP OVERPRESSURE alarm is active, all presses of this button are ignored.
The Cycle button initiates the Cycle mode, which is where you can choose STAT or an auto cycle time. Cycle mode selections appear in the following order: Stat, 1, 2, 3, 4, 5, 10, 15, 20, 30, 60, 90, 120 (minutes), and - - (two dashes). Choose Stat to start STAT mode. Choose 1-120 to select the desired cycle time and start Auto Cycle mode. When you reach the desired setting, do not press the Cycle button again. After 2 seconds the cycle mode is deactivated. Choose the two dashes to cancel Auto Cycle mode. Note: While an E80: BP OVERPRESSURE alarm is active, all presses of this button are ignored.
62
Page 73
BP
Windows Associated with BP
The windows associated with BP are Systolic, Diastolic, MAP/Cuff, Pulse Rate, and min. The Systolic, Diastolic, MAP/Cuff, and Pulse Rate (if SpO
are automatically cleared when a new BP determination is started. In manual mode, the displayed information is also cleared when it becomes older than 30 minutes.
The Systolic and Diastolic windows display values after a determination has completed successfully. While in STAT mode, the Systolic window flashes the early systolic value if it is available.
The MAP/Cuff window displays the derived mean arterial pressure (MAP) following the completion of a successful determination. During any type of BP determination, the pressure inside the cuff appears in this window.
The min window displays the BP mode of operation and the age of the previous BP determination. When both types of information are present, they flash alternately in this window. When in Manual mode, two dashes (- -) are displayed. When in Auto Cycle mode, the chosen Cycle time is displayed (e.g.,
15) When in STAT mode, Stat is displayed. When displayed, the age of the previous BP determination is preceded by a minus sign (e.g., - 5).
is not active) windows
2
Indicators Associated with BP
The indicators associated with BP are Systolic HIGH and LOW, Diastolic HIGH and LOW, AUTO CYCLE and HISTORY.
The AUTO CYCLE indicator appears solid green when Auto mode is on. It flashes green when changes are being made to the current BP mode (e.g., Cycle mode is active). The HISTORY indicator appears solid green when the age of the previous BP determination is displayed in the min window.
Parameter Modes
The BP parameter is in operate mode upon power-up. It does not have a ready mode.
63
Page 74
BP Modes of Operation
The ProCare Monitor has three BP modes: 1. Manual, 2. Auto Cycle, and 3. STAT. The mode is selected by the user. BP determinations are automated and, upon completion, the values for systolic pressure, diastolic pressure, mean arterial pressure, and pulse rate are shown in their respective windows.
Manual BP Determinations
Manual mode allows you to take one blood pressure determination. Manual mode is always the BP mode of operation upon power-up. A normal, uninterrupted Manual determination takes about 40 seconds. Following a determination, the cuff pressure must drop below 5 mmHg (neonate) or 15 mmHg (adult) before another determination can be started.
Manual BP determinations are started by pressing the Inflate/Stop button. To stop a Manual BP determination press the Inflate/Stop button. The values displayed in the Systolic, Diastolic, MAP, and Pulse Rate (if SpO active) windows are cleared after 30 minutes have lapsed.
Auto Cycle Determinations
Auto Cycle mode allows you to take multiple determinations at user-defined intervals. In the Auto Cycle mode, the pressure must be below 5 mmHg (neonate) or 15 mmHg (adult) for at least 30 seconds before the next auto determination can be started.
is not
2
Auto Cycle mode is started by selecting the Cycle button. When in Auto Cycle mode, the AUTO CYCLE indicator appears solid green. Manual determinations can be taken while in Auto Cycle mode without affecting when the next auto determination is to start. You can also change the time interval while in Auto Cycle mode.
Once the Cycle button is pressed, the first Auto Cycle determination is started, and the time between determinations appears in the min window. You can change the time interval between Auto Cycle determinations by continuing to press the Cycle button until you reach the
64
Page 75
BP
desired time interval. Time interval selections appear in the following order: Stat, 1, 2, 3, 4, 5, 10, 15, 20, 30, 60, 90, 120 (minutes), and - - (two dashes). When you reach the desired time interval, do not press the Cycle button again; after 2 seconds, the chosen time interval is retained and remains in the min window.
Pressing the Cycle button when in Auto Cycle mode activates Cycle mode again with two dashes (- -) appearing in the min window. If you press the Cycle button immediately after the first press, the next time interval appears in the min window. If you do not press the Cycle button immediately after the first press, Cycle mode is deactivated. Press the Inflate/Stop button to stop the determination in progress without canceling the Auto Cycle mode. Choose the two dashes (- -) to cancel Auto Cycle mode.
If the first Auto Cycle determination results in a limit alarm, another repeat determination is taken to verify the alarm. This occurs only once.
Whenever an Auto Cycle determination results in a E89: BP NO DETERMINATION alarm, up to nine more repeat determinations are attempted in order to achieve valid values. If at any time during this repeat cycle, the BP NO DETERMINATION alarm is acknowledged by pressing the Silence button or the Inflate/Stop button, additional determinations are not attempted. If the repeat cycle completes all nine repeat determinations without reaching a valid value, the Monitor returns to normal Auto Cycle mode. However, an Auto Cycle mode determination must complete successfully before a repeat cycle will follow a future Auto Cycle mode determination that results in an E89: BP NO DETERMINATION alarm.
STAT BP Determinations
STAT mode allows you to take as many determinations as possible within a 5-minute time period. The Monitor will begin another determination once the pressure is below 5 mmHg for 8 seconds (neonates) or 15 mmHg for 4 seconds (adults), unless the 5-minute period has ended or STAT mode has been canceled.
65
Page 76
Note: BP and BP-derived pulse rate alarm limits are disabled while in Stat mode.
STAT BP determinations are started by selecting the Cycle button. Once the Cycle button is pressed, choose Stat. The Monitor automatically begins a 5-minute period of STAT determinations. Note: If the Monitor was previously in Auto mode, the first STAT BP determination begins after 2 seconds.
After the first STAT determination, an early systolic value flashes in the associated window. If STAT mode is started when a determination is already in progress, that determination becomes the first in the series of STAT determinations. At the end of STAT mode, the BP mode prior to entering STAT mode is resumed. To cancel STAT mode, press the Inflate/Stop button.
User Settings
Mode Settings
There is one mode setting associated with this parameter: Cycle. The Cycle mode is started by pressing the Cycle button. While the Cycle mode is active, Cycle selections are displayed in the min window. Cycle selections appear in the following order: Stat, 1, 2, 3, 4, 5, 10, 15, 20, 30, 60, 90,
120, - - .
Limit Settings
There are two limit settings associated with this parameter: HIGH and LOW. Both limit settings are available for Systolic and Diastolic windows. The range for Systolic HIGH is 35 to 290 mmHg; the range for Systolic LOW is 30 to 285 mmHg. The range for Diastolic HIGH is 15 to 220 mmHg; the range for Diastolic LOW is 10 to 215 mmHg. The settings appear in increments of 5 mmHg.
Neonate Settings
The Monitor determines if a neonate BP cuff is being used and then checks the systolic limit and adjusts it to an appropriate limit for neonates.
66
Page 77
BP
Menu Settings
The INFLATE PRESSURE menu setting is associated with the BP parameter. This option lets you adjust the target pressure that the Monitor initially pumps to for the next determination. The settings appear in increments of 5 mmHg. The range available is 100 mmHg to 250 mmHg. If the Monitor senses a neonate BP cuff in use, it checks the systolic limit and adjusts it to an appropriate limit for neonates.
When you change the target pressure, any values from the previous determinations are cleared from their associated windows. Note: Default inflation pressure can be changed in Config Mode. Refer to the Getting Started section.
Manual determinations use the value chosen for INFLATE PRESSURE as their target inflation pressure unless a determination is started within 2 minutes of the completion of a previous determination. For a determination that is started within 2 minutes of the previous determination, the target inflation pressure is adaptive (calculates the target inflation pressure based on the previous determination). For all Auto Cycle and STAT determinations (with the exception of the first determination of the series), the target inflation pressure is always adaptive. If recent determination results are available, the target pressure is derived from the last systolic value.
Sounds Associated with BP
There is one tone associated with this parameter: a single tone. The tone sounds at the completion of any BP determination.
Procedures
1. Connect the end of the air hose which has quick-release clips to the BP connector on the front of the Monitor. Make sure that the hose is not kinked or compressed. Note: To disconnect the hose from the Monitor, squeeze the quick-release clips together and pull the plug from the BP connector.
67
Page 78
2. Select the appropriate blood pressure measurement site. Because normative values are generally based on this site and as a matter of convenience, the upper arm is preferred. When upper arm size or shape, the patient’s clinical condition, or other factors prohibit use of the upper arm, the clinician must plan patient care accordingly, taking into account the patient’s cardiovascular status and the effect of an alternative site on blood pressure values, proper cuff size, and comfort. The figure shows the recommended sites for placing cuffs.
Warning: Do not place the cuff on a limb being used for intravenous infusion or any area where circulation is compromised or has the potential to be compromised.
3. If patient is standing, sitting, or inclined, ensure that cuffed limb is supported to maintain cuff at level of patient’s heart. If cuff is not at heart level, the difference in systolic and diastolic values due to hydrostatic effect must be considered. Add 1.80 mmHg to values for every inch (2.54 cm) above heart level. Subtract 1.80 mmHg from values for every inch (2.54 cm) below heart level.
4. Select appropriate cuff size. Measure patient’s limb and select appropriately sized cuff according to size marked on cuff or cuff packaging. When cuff sizes overlap for a specified circumference, choose the larger size cuff.
Precaution: Accuracy depends on use of proper size cuff. Caution: Do not use an infant cuff with an auscultatory reference DINAMAP
®
ProCare Monitor. The neonatal #5 cuff and neonatal hose may be used on patients with an arm circumference of 8 - 15 cm. Note: Use only CRITIKON
®
Blood Pressure Cuffs. The size, shape, and bladder characteristics can affect the performance of the instrument. Inaccurate readings may occur unless CRITIKON
®
Blood Pressure Cuffs are used.
68
Page 79
BP
5. Inspect cuff for damage. Replace cuff when aging, tearing, or weak closure is apparent. Do not inflate cuff when unwrapped.
Precaution: Do not use cuff if structural integrity is suspect.
6. Connect the cuff to the air hose.
7. Inspect patient’s limb prior to application.
Precaution: Do not apply cuff to areas where skin is not intact or tissue is injured.
8. Palpate artery and place cuff so that patient’s artery is aligned with cuff arrow marked “artery.”
9. Squeeze all air from cuff and confirm that the connection is secure and unoccluded and that tubing is not kinked.
10.Wrap cuff snugly around the patient’s limb. Cuff index line must fall within the range markings. Ensure that hook and loop closures are properly engaged so that pressure is evenly distributed throughout cuff. If upper arm is used, place cuff as far proximally as possible.
11.Proper cuff wrapping should be snug, but should still allow space for a finger between patient and cuff. Cuff should not be so tight as to prevent venous return between determinations.
Warning: Using a cuff that is too tight will cause venous congestion and discoloration of the limb, but using a cuff that is too loose may result in no readings and/or inaccurate readings.
12.Proceed with monitoring in the Manual, Auto Cycle, or STAT mode.
What to do When Taking BPs on Different Patients
To ensure the previous patient’s BP will not be used for adaptive target inflation pressure when taking a BP on a new patient, you can 1.) clear the history by holding the history key for more than 2 seconds, or 2.) if in manual mode, wait for more than 2 minutes since the last determination was taken on the previous patient.
In manual mode, the Monitor will not use the displayed BP values for adaptive target inflation pressure if it has been more than 2 minutes since the end of the previous determination. In manual mode, the BP values are displayed for a maximum of 30 minutes. In auto mode, the displayed BP values are used for adaptive target inflation pressure independant of the age of the displayed values.
69
Page 80
Cuff Connections
Quick-Disconnect Air Hose
Push and twist the male slip hose connector into the cuff connector to ensure a tight fit.
Neonate Air Hose
Push and twist the male slip hose connector into the cuff connector to ensure a tight fit.
Alarms
Alarm Acknowledgeable
by pressing Silence?
BP SYSTOLIC HIGH yes BP SYSTOLIC LOW yes BP DIASTOLIC HIGH yes BP DIASTOLIC LOW yes E89: BP NO DETERMINATION yes E85: BP LEVEL TIMEOUT yes E84: BP TOTAL TIMEOUT yes E83: BP PUMP TIMEOUT yes E80: BP OVERPRESSURE yes E82: EXCESS AIR IN CUFF yes
The alarm codes appear in the Systolic window. When an alarm is active, it can be acknowledged by starting a new determination, with the exception of the E80: BP OVERPRESSURE alarm. BP OVERPRESSURE can only be acknowledged by pressing the Silence button. During STAT mode determinations, systolic, diastolic and BP-derived pulse rate values are not checked against their limits.
Critikon US Patents
4,360,029; 4,501,280; 4,546,775; 4,638,810; 5,052,397; 4,349,034; 4,543,962; 4,627,440; 4,754,761; 5,170,795
European Patents
EP122123, EP205805, EP207807
70
Page 81
BP
Specifications
Cuff Pressure Range 0 to 290 mmHg (adult/ped) (Normal operating range) 0 to 140 mmHg (neonate)
Blood Pressure Accuracy Meets or exceeds ANSI/AAMI
standard SP-10 (mean error ≤5 mmHg, standard deviation ≤8 mmHg)
Maximum Determination 120 s (adult/ped) Time 85 s (neonate)
Overpressure Cutoff 300 to 330 mmHg (adult/ped)
150 to 165 mmHg (neonate)
Pulse Rate Range
Pulse Rate Accuracy ± 3.5%
Factory Default Settings
Systolic Limits (mmHg) HIGH: 200 LOW: 80 Diastolic Limits (mmHg) HIGH: 120 LOW: 30 Inflation Pressure* 160 mmHg (adult/ped)
Cycle button 15
*Default inflation pressure can be changed in Config Mode. Refer to the Getting Started section.
All DINAMAP® Monitor regulatory and accuracy studies have been performed using CRITIKON Cuffs. Use only CRITIKON size, shape, and bladder characteristics can affect the performance of the instrument. Inaccurate readings may occur unless CRITIKON
30 to 200 beats/min (adult/ped) 30 to 220 beats/min (neonate)
110 mmHg (neonate)
®
®
Blood Pressure Cuffs. The
®
Blood Pressure Cuffs are used.
Blood Pressure
71
Page 82
72
Page 83
SpO
2
Page 84
74
Page 85
NELLCOR® SpO
NELLCOR* OxiMax® SpO2
Description
The SpO2 parameter in the ProCare Monitor is available in two different leading technologies: NELLCOR MASIMO SET see which SpO SpO
technology logo will be on the front fascia of the
2
®
. Please refer to the front of your Monitor to
technology your monitor contains. The
2
Monitor. This section refers to NELLCOR
®
and
®
SpO2 technology.
2
The SpO begin SpO
parameter is included in Models 300 and 400. To
2
monitoring, plug the sensor into the NELLCOR®
2
DOC-10 sensor extension cable, then plug the cable into the Monitor. Simply place the SpO
sensor on the patient’s
2
finger; monitoring begins automatically. Functional oxygen saturation (SpO continuously monitored in the ProCare Monitor using pulse oximetry technology from NELLCOR
) of arterial blood is noninvasively and
2
®
. Functional SpO2 is the ratio of oxygenated hemoglobin to hemoglobin that is capable of transporting oxygen. This ratio, expressed as a percentage, is shown in the SpO
window, which is
2
continually updated.
Pulse rate derived from SpO
appears in the Pulse Rate
2
window. A tone sounds at a rate corresponding to the pulse rate and at a pitch corresponding to the SpO
saturation
2
level. The pitch is highest at 100% oxygen saturation, and it continuously decreases as the saturation level falls. The Monitor displays a pulse amplitude bar. The pulse amplitude bar graph is proportional to the arterial blood flow.
Audible and visible alarms occur when SpO
levels are
2
outside the alarm limits. When a parameter status alarm occurs, an alarm message code appears in the SpO
2
window. Note: Limit alarms, printing, and trending are not available for the first 10 seconds of SpO
monitoring.
2
*NELLCOR® is a trademark of Nellcor Puritan Bennett
75
Page 86
General Warnings
• Do not place SpO2 sensor on patient during magnetic resonance imaging (MRI). Adverse reactions include potential burns to patients as a result of contact with attachments heated by the MRI radio frequency pulse, potential degradation of the magnetic resonance image, and potential reduced accuracy of SpO measurements. Always remove oximetry devices and attachments from the MRI environment before scanning a patient.
• The use of cardio-green and other intravascular dyes at certain concentrations may affect the accuracy of the SpO
measurement.
2
•The SpO
function is calibrated to read functional
2
arterial oxygen saturation. Significant levels of dysfunctional hemoglobins such as carboxyhemoglobin or methemoglobin may affect the accuracy of the SpO measurement.
• Pulse oximetry readings and pulse signals can be affected by certain environmental conditions, sensor application errors, and certain patient conditions. See the appropriate sections of this manual for specific safety information.
• The use of accessories, transducers, and cables other than those specified may result in increased emission and/or decreased immunity and inaccurate readings of the Monitor.
2
2
General Cautions
• As with any clip-on sensor, pressure is exerted. The clinician should be cautious in using a clip-on sensor on patients with compromised circulation (e.g., because of peripheral vascular disease or vasoconstricting medications).
• Do not perform any testing or maintenance on a sensor while it is being used to monitor a patient.
• Bright light sources (e.g., infrared heat lamps, bilirubin lights, direct sunlight, operating room lights) may interfere with the performance of the SpO
76
function.
2
Page 87
NELLCOR® SpO
To prevent such interference, cover the sensor with opaque material.
General Notes
• A patient’s vital signs may vary dramatically during the use of cardiovascular agents such as those that raise or lower blood pressure or those that increase or decrease pulse rate.
•The ProCare Monitor that is labeled with NELLCOR Technology is compatible only with NELLCOR OxiMax extension cable.
• Software development, software validation, and Risk and Hazard Analysis has been performed to a registered quality system.
• User or patient-applied parts are latex-free.
®
sensors and NELLCOR® DOC-10 sensor
®
®
2
Configuration Settings Associated with SpO
There are two configuration settings associated with this parameter: Response Mode (n0d) and SatSeconds™ (SAt).
Response Mode (n0d) allows the user to specify the averaging technique to optimize measurements in the presence of various patient movement. Choose Mode 1 (Normal Response) when patients are active as in exercise protocols. Choose Mode 2 (Fast Response; default setting) for the general patient population. The configuration of this setting affects whether the pulse rate derived from SpO displayed in the Pulse Rate window while SpO mode. If Mode 1 is chosen, the pulse rate is not displayed. If Mode 2 is chosen, the pulse rate is displayed.
With traditional alarm management, upper and lower alarm limits are set for monitoring SpO soon as an alarm limit is violated by as little as one percentage point, an audible alarm immediately sounds. When the % SpO sounds each time the limit is violated. To prevent these nuisance alarms, the ProCare Monitor uses the SatSeconds technique. The SatSeconds
fluctuates near an alarm limit, the alarm
2
technique (SAt) limit-controls
. During monitoring, as
2
2
is in operate
2
2
is
77
Page 88
the time that the % SpO2 level may fall outside the alarm before an audible alarm sounds. Choose either 0, 10, 25, 50, or 100 seconds. If 0 is chosen this limit hold-off feature is disabled.
The SatSeconds
“Safety Net” is for patients with saturation
levels having frequent excursions below the limit, but not staying below the limit long enough for the SatSeconds
time setting to be reached. When 3 or more limit violations occur within 60 seconds, an alarm sounds even if the
SatSeconds
Buttons Associated with SpO
time setting has not been reached.
2
There are no buttons associated with this parameter.
Windows Associated with SpO
2
There is one window associated with this parameter: SpO2. While in offline mode, nothing is displayed in the SpO
2
window. When in ready mode and the parameter senses that a sensor is connected, a single dash (-) appears in this window. When it is in operate mode and the parameter is reporting a valid data, the derived SpO
value appears in this
2
window. The values are displayed in %. Note: If SpO
is the source for Pulse Rate, the Pulse Rate
2
window is associated with this parameter.
Indicators Associated with SpO
2
There is one indicator associated with this parameter: the pulse amplitude indicator bar. The red LED bar flashes to indicate that pulse rate measurements are being derived from SpO
signals and the height of the bar is proportional to
2
the arterial blood flow.
Parameter Modes
The SpO2 parameter is in offline mode upon power-up. The parameter automatically switches from offline to ready mode when a sensor is connected to the Monitor. The parameter automatically switches from ready to operate mode when pulses are detected.
If an SpO alarm is acknowledged, the parameter automatically
SENSOR OFF FINGER or SPO2 LOST PULSE
2
78
Page 89
NELLCOR® SpO
switches from operate to ready mode. If a SpO2 SENSOR DISCONNECTED alarm is acknowledged, the parameter automatically switches from operate to offline mode. Note: Once either of these alarms have been acknowledged, SpO
monitoring is not resumed until a valid pulse is
2
detected again.
User Settings
Limit settings
There are two limit settings associated with this parameter: HIGH and LOW. The range for HIGH is 21 to 100%. The range for LOW is 20 to 99%. The settings appear in increments of 1%.
Menu Settings
The PULSE VOLUME menu setting is associated with the SpO
parameter. This option lets you adjust the volume of
2
the tone that sounds after each detected heartbeat. It can be adjusted from 0 - 10 (10 being the loudest). If you set the volume to zero, no tone will sound.
2
Sounds Associated with SpO
An audible tone is provided by the Monitor for each pulse detected by the NELLCOR the audible tone is directly related to the calculated saturation value. As the saturation value increases, the pitch frequency increases continuously with each 1% increase in SpO
. As the saturation value decreases, the pitch frequency
2
continuously decreases. This audible tone is silenced while an alarm sounds or the PULSE VOLUME is set to 0. Refer to “Menu Settings” in this section.
2
®
SpO2 parameter. The pitch of
Procedures
1. Select a sensor that is appropriate for the patient and the clinical situation.
• Use only NELLCOR NELLCOR another pulse oximetry cable will have an adverse effect on performance. Do not attach any cable that is intended for computer use to the sensor port. Do not connect any device other than NELLCOR sensor to the sensor connector.
®
DOC-10 pulse oximetry cables. Use of
®
sensors PURPLE and WHITE and
®
-approved
79
Page 90
Warning: Do not use a damaged sensor or one with exposed electrical contacts. Do not use sensor, cables, or connectors that appear damaged.
®
Note: Use only NELLCOR
OxiMax® compatible sensors, which are available from GE Medical Systems ­Accessories and Supplies.
2. Following the directions for use supplied with the sensor, apply the sensor to the patient.
Warnings
Patient safety:
• If the sensor is are not applied properly, the patient’s skin could be injured or the ability of the ProCare Monitor to measure oxygen saturation could be compromised. For example, a clip-on sensor should never be taped shut. Taping the sensor could damage the patient’s skin or impair the venous return, thus causing venous pulsation and inaccurate measurement of oxygen saturation.
• Excessive pressure from the sensor may cause necrosis of the skin.
• For additional warnings and information, refer to the NELLCOR
®
sensor’s directions for use.
Monitor performance:
•When an SpO pressure cuff, the SpO cuff is inflated. If SpO
sensor is on a limb that has a blood
2
data will not be valid when the
2
readings are required during
2
the entire blood pressure determination, attach the SpO
sensor to the limb opposite the one with the
2
blood pressure cuff.
• Remove nail polish and artificial nails. Placing a sensor on a polished or an artificial nail may affect accuracy.
Cautions
Patient safety:
• Do not place any clip-on sensor in a patient’s mouth or on a patient’s nose or toe.
• Do not place a clip-on finger sensor on a patient’s thumb or across a child’s foot or hand.
• The sensor disconnect error message and associated alarm indicate that the sensor is either disconnected or
80
Page 91
NELLCOR® SpO
the wiring is faulty. The user should check the sensor connection and, if necessary, replace the sensor, DOC­10 cable, or both.
• Observe the sensor site frequently to assure adequate distal circulation.
least every 2 hours and rotated at least every 4 hours.
Monitor performance:
• Placing a sensor distal to an arterial line may interfere with adequate arterial pulsation and compromise the measurement of SpO
• Place the sensor so that the LEDs and the photodiode are opposite each other.
3. Plug the SpO cable. Then plug the SpO SpO
sensor connector.
4. Proceed with monitoring. SpO
2
continuously and can run simultaneously with other measurements.
Alarms
Alarm Acknowledgeable
SPO2 HIGH no SPO2 LOW no E20: SPO2 SENSOR DISCONNECTED yes E21: SPO2 REPLACE SENSOR yes E23: SPO2 SENSOR OFF FINGER yes E25: SPO2 LOST PULSE no
Sensor sites should be checked at
.
2
sensor into the SpO2 sensor extension
2
sensor extension cable into the
2
determinations run
2
by pressing Silence?
2
81
Page 92
Troubleshooting
This section discusses potential difficulties and suggestions for resolving them. If the difficulty persists, contact a qualified service person or your local representative. The service manual, which is for use by qualified service personnel provides additional troubleshooting information.
PROBLEM: The pulse amplitude bar indicates a pulse, but no oxygen saturation or pulse rate values appear on the screen.
CAUSE:
• Excessive patient motion may be making it impossible for the SpO
function to find a pulse pattern.
2
• The sensor may be damaged.
• The patient’s perfusion may be too low to allow the
function to measure saturation and pulse rate.
SpO
2
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor is applied securely and properly, and
SpO
2
replace it if necessary; move the sensor to a new site; or use a disposable adhesive sensor that may tolerate more motion.
• Replace the sensor.
PROBLEM: The SpO
value or the pulse rate changes
2
rapidly; the pulse amplitude bar is erratic. CAUSE:
• Excessive patient motion may be making it impossible for the SpO
function to find a pulse pattern.
2
• An electrosurgical unit (ESU) may be interfering with performance.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the sensor is applied securely and properly, and replace it if necessary; move the sensor to a new site; use a sensor that tolerates more motion.
If an ESU is interfering:
• Move the SpO
cable as far from the ESU as possible.
2
82
Page 93
NELLCOR® SpO
• Plug the Monitor and the ESU into different AC circuits.
• Move the ESU ground pad as close to the surgical site as possible.
• The sensor may need to be replaced with a new sensor.
• If the patient weighs less than 3 kg or more than 40 kg, apply an OxiMax OXI-A/N, OXI-P/I), or OxiCliq oxygen transducer to an appropriate site. These sensors have Faraday shields which provide added protection from high electronic noise and ambient light.
PROBLEM: The oxygen saturation measurement does not correlate with the value calculated from a blood gas determination.
CAUSE:
•The SpO for the effects of pH; temperature; CO hemoglobin; or 2,3-DPG.
• Accuracy can be affected by incorrect sensor application or use; intravascular dyes; bright light; excessive patient movement; venous pulsations; electrosurgical interference; and placement of a sensor on an extremity that has a blood pressure cuff, arterial catheter, or intravascular line.
SOLUTION:
• Check that calculations have been corrected appropriately for the relevant variable. In general, calculated saturation values are not as reliable as direct laboratory hemoximeter measurements.
• If there is excessive light, cover the sensor with opaque material.
• Circulation distal to the sensor site should be checked routinely. Refer to the directions for use supplied with the sensor for requirements on moving the sensor to another site to ensure adhesion, skin integrity, and correct optical alignment. If skin integrity changes, move the sensor to another site.
• Try to keep the patient still, or change the sensor site to one with less motion.
calculation may not have correctly adjusted
2
®
, reusable sensors (except DS-100,
; fetal
2
2
83
Page 94
• Observe all instructions, warnings, and cautions in this manual and in the directions for use of the sensor.
PROBLEM: A valid SpO
signal was present but has
2
disappeared.
CAUSE:
• A BP determination on the same limb is in progress.
SOLUTION:
Check the patient.
• An alarm message code appears on the screen, and the audible alarm will sound immediately.
• Move the sensor to the arm that is not connected to a blood pressure cuff.
PROBLEM: A REPLACE SENSOR error code has been detected. CAUSE:
• The sensor or cable may be the wrong type or defective, the cabling may be improperly connected.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the proper sensor/cable is applied securely and properly, and replace it if necessary.
• Disconnect and reconnect the sensor.
PROBLEM: A SENSOR DISCONNECTED error code has been detected, but the sensor is still connected. CAUSE:
• The sensor is not completely connected. The sensor extension cable or sensor wiring is faulty.
SOLUTION:
Check the patient.
• Check the sensor connection to the sensor extension cable and the sensor extension cable connection to the monitor. Then, if needed, replace the sensor or the sensor extension cable.
84
Page 95
NELLCOR® SpO
Specifications
Measurement Range
SpO
2
Pulse Rate 20 to 250 beats/min
Accuracy and Motion Tolerance
Saturation
Without Motion - Adults* 70 to 100% ±2 digits Without Motion - Neonate* 70 to 100% ±3 digits With Motion - Adults/Neo** 70 to 100% ±3 digits Low Perfusion 70 to 100% ±2 digits
Pulse Rate
Without Motion 20 to 250 beats/min ±3 digits With Motion normal physiologic range
Low Perfusion 20 to 250 beats/min ±3 digits
1 to 100%
0 to 69% unspecified
55 to 125 beats/min ±5 digits
2
*Adult specifications are shown for OxiMax® MAX-A and MAX-N sensors. Neonate specifications are shown for OxiMax will vary by sensor type. **Applicability:
NELLCOR® Sensor Accuracy Note: All NELLCOR
NELLCOR
OxiMax® MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N sensors.
®
®
DOC-10 cable; the SCP-10 cable is not
sensors must be used with the
®
MAX-N. Saturation accuracy
compatible with the ProCare Monitor.
Sensor Model
SpO2 Range 70% - 100%
OxiMax
®
MAX-A*, MAX-AL* ±2 digits
MAX-N* MAX-N*
(Adult) ±2 digits
(Neonate) ±3 digits MAX-P* ±2 digits MAX-I* ±2 digits
MAX-R*
OxiCliq
®
±3.5 digits
OxiCliq A ±2.5 digits OxiCliq P ±2.5 digits
85
Page 96
OxiCliq N† (Adult) ±2.5 digits OxiCliq N
(Neonate) ±3.5 digits
OxiCliq I ±2.5 digits
Reusable Sensor Models
D-YS (Infant to Adult) ±3 digits D-YS (Neonate) ±4 digits D-YS with D-YSE ±3.5 digits D-YS with D-YSPD ±3.5 digits DS-100A ±3 digits OXI-A/N (Adult) ±3 digits OXI-A/N (Neonate) ±4 digits OXI-P/I ±3 digits
* The accuracy specification under motion conditions is ±3. For a definition of motion, contact NELLCOR representative.
The MAX-N and the OxiCliq N were tested on patients >40 kg.
The accuracy specification has been determined between saturations of
80%-100%.
®
Technical Services or your local
Neonatal Sensor Accuracy
When sensors are used on neonatal subjects as recommended, the specified accuracy range is increased by ±1 digit, to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood. For example, MAX-N accuracy on neonates is ±3 digits, rather than ±2 digits.
Sensor Light Source
Wavelength Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation Infrared: 22.5 mW (max)
Red: 30 mW (max)
Factory Default Settings
SpO
(%) HIGH Alarm Limit: 100
2
(%) LOW Alarm Limit: 90
SpO
2
Response Mode 2 (for Mode 2: Fast Response) SatSeconds
0
86
Page 97
NELLCOR® SpO
NELLCOR® Patents
US Patent No. 4,621,643; 4,653,498; 4,700,708; 4,770,179; 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,421,329; 5,485,847; 5,533,507; 5,577,500; 5,803,910; 5,853,364; 5,865,736; 6,083,172; Re. 35,122 and foreign equivalents.
2
87
Page 98
88
Page 99
MASIMO SET® SpO
Masimo Set* SpO Description
2
®
2
The SpO2 parameter in the ProCare Monitor is available in two different leading technologies: NELLCOR MASIMO SET Monitor to see which SpO contains. The SpO
®
. Please refer to the front of your ProCare
technology your monitor
technology logo will be on the front
2
2
fascia. This section refers to MASIMO SET
®
®
SpO2
and
technology.
The SpO Functional oxygen saturation (SpO
parameter is included in Models 300 and 400.
2
) of arterial blood is
2
noninvasively and continuously monitored in the ProCare Monitor using pulse oximetry technology from MASIMO
®
SET
. Functional SpO2 is the ratio of oxygenated hemoglobin to hemoglobin that is capable of transporting oxygen. This ratio, expressed as a percentage, is shown in the SpO
2
window, which is continually updated.
Pulse rate when associated with SpO
appears in the Pulse
2
Rate window. A tone sounds at a rate corresponding to the pulse rate and at a pitch corresponding to the SpO
2
saturation level. The pitch is highest at 100% oxygen saturation, and it becomes lower as the saturation level falls. The Monitor displays a pulse amplitude bar. The pulse amplitude bar graph is proportional to the arterial blood flow.
Audible and visible alarms occur when SpO
levels are
2
outside the alarm limits. When a parameter status alarm message code occurs, an alarm message appears in the SpO
window.
2
Note: Limit alarms, printing, and trending are not available for the first 10 seconds of SpO
*MASIMO SET® is a trademark of Masimo Corporation. Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to the device.
monitoring.
2
89
Page 100
General Cautions
• If any measurement does not seem reasonable, first check the patient’s vital signs by alternate means and the check the pulse oximeter for proper functioning.
• Inaccurate measurements may be caused by:
- Incorrect sensor application or use
- Significant levels of dysfunctional hemoglobins (e.g., carboxyhemoglobin or methemoglobin)
- Intravascular dyes such as indocyanine green or methylene blue
- Exposure to excessive illumination, such as surgical
lamps (especially ones with xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight (exposure to excessive illumination can be corrected by covering the sensor with a dark or opaque material)
- excessive patient movement
- venous pulsations
- placement of sensor on an extremity with a blood
pressure cuff, arterial catheter, or intravascular line
• The pulse oximeter can be used during defibrillation, but the readings may be inaccurate for a short time.
• loss of pulse signal can occur in any of the following situation:
- the sensor is too tight
- there is excessive illumination from light sources such as surgical lamp, a bilirubin lamp, or sunlight
- a blood pressure cuff is inflated on the same extremity as the one with a SP0
- the patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia
- there is arterial occlusion proximal to the sensor
- the patient is in cardiac arrest or is in shock
sensor attached
2
90
Loading...