This manual is for DINAMAP® PRO Monitor Models 100V2,
200V2, 300V2, and 400V2, all with printers.
• PRO 100V2: BP and Pulse
• PRO 200V2: BP, Pulse, and Temp
• PRO 300V2: BP, Pulse, and SpO
• PRO 400V2: BP, Pulse, Temp, and SpO
The model of the Monitor determines which menu option
buttons appear on the LCD. Please refer to applicable
sections.
The Model Number 100-400V2 is generic in nature and
reflects the range of product codes available. Your product
may be labeled with a specific product code such as
DINAMAP
version of the product's design.
®
PRO Model 410. "V2" refers to the second
Reissues and Updates
Changes occurring between issues are addressed through
Change Information Sheets, Addendums, and replacement
pages. If a Change Information Sheet does not accompany
this manual, it is correct as printed.
2
2
Errors and Omissions
If errors or omissions are found in this manual, please notify:
GE Medical Systems Information Technologies
4502 Woodland Corporate Boulevard
Tampa, FL 33614
1-877-274-8456
Part No. 2009802-001B
The content of this document including all figures and
drawings is proprietary information of GE Medical Systems
Information Technologies, provided solely for purposes of
operation, maintenance or repair, and dissemination for
3
other purposes or copying thereof is prohibited without prior
written consent by GE Medical Systems Information Technologies, Tampa, Florida.
Illustrations may show design models; production units may
incorporate changes.
Hierarchy of Warnings and Cautions
A general warning is a statement that alerts the user to the
possibility of injury, death, or other serious adverse reactions
associated with the misuse of the device. A warning relates
to steps in a procedure.
A general caution is a statement that alerts the user to the
possibility of a problem with the device associated with its
use or misuse. Such problems include device malfunction,
device failure, damage to the device or damage to other
property. A caution relates to steps in a procedure.
Disposal of Product Waste .................................................................................. 103
Appendix G .........................................................................105
Host Port Connector (rear panel)....................................................................... 105
6
Introduction
About the DINAMAP® PRO Monitor
DINAMAP® PRO Monitors provide noninvasive
determination of systolic blood pressure, diastolic blood
pressure, mean arterial pressure, pulse rate, temperature, and
oxygen saturation. These portable AC- and DC-operated
monitors are primarily intended for use in hospital acute care
settings such as outpatient surgery, accident and emergency,
labor and delivery, GI/endoscopy, and medical/surgical
units.
The PRO Monitor comes in four different models: PRO
100V2, 200V2, 300V2, and 400V2, all with printers.
• PRO 100V2: BP and Pulse
• PRO 200V2: BP, Pulse, and Temp
• PRO 300V2: BP, Pulse, and SpO
• PRO 400V2: BP, Pulse, Temp, and SpO
All of the main operations of the PRO Monitor are easy to
use. Please review the factory default settings and, where
applicable, enter settings appropriate for your use. The
“Using the Monitor” section of this manual explains how to
use the Monitor in its most simple form, while the “Using the
Menu System” section explains how to customize
measurements by using the menu system.
2
2
Indications
The PRO Monitor is intended to monitor one patient at the
bedside.
Contraindications
This device is not designed, sold, or intended for use except
as indicated.
Federal law (U.S.A.) restricts this device to sale by or on the
order of a clinician.
Warnings
• Do not use the PRO Monitor in the presence of
magnetic resonance imaging (MRI) devices. There have
7
been reports of sensors causing patient burns when
operating in an MRI environment.
• Do not use the Monitor in the presence of flammable
anesthetics.
• To help prevent unintended current return paths with
the use of high frequency (HF) surgical equipment,
ensure that the HF surgical neutral electrode is
properly connected.
• To avoid personal injury, do not perform any servicing
unless qualified to do so.
• WARNING: These Monitors should not be used on
patients who are connected to cardiopulmonary bypass
machines.
• If powering the Monitor from an external power
adapter or converter, use only power adapters and
converters approved by GE Medical Systems
Information Technologies.
• The Monitor does not include any user-replaceable
fuses. Refer servicing to qualified service personnel.
• To reduce the risk of electric shock, do not remove the
cover or the back. Refer servicing to a qualified service
person.
• If the accuracy of any determination reading is
questionable, first check the patient’s vital signs by
alternate means and then check the PRO Monitor for
proper functioning.
Cautions
• Do not use replacement batteries other than the type
supplied with the Monitor. Replacement batteries are
available from GE Medical Systems Information Technologies. See Appendix D.
• The PRO Monitor is designed to conform to
Electromagnetic Compatibility (EMC) standard IEC
601-1-2, 1993 and will operate accurately in
conjunction with other medical equipment which also
meets this requirement. To avoid interference problems
affecting the Monitor, do not use the Monitor in the
8
presence of equipment which does not conform to
these specifications.
• Place the PRO Monitor on a rigid, secure surface.
Monitor must only be used with mounting hardware,
poles, and stands recommended by GE Medical
Systems Information Technologies. See Appendix D.
• The weight of the accessory basket contents should not
exceed 6.6 lb (3 kg).
• Arrange the power cord, air hoses, and all cables
carefully so they do not constitute a hazard.
• Verify calibration of BP parameter (temp and pulse
oximeter do not require calibration). Ensure that the
display is functioning properly before operating the
PRO Monitor.
• Do not immerse the Monitor in water. If the Monitor is
splashed with water or becomes wet, wipe it
immediately with a dry cloth.
• Do not gas sterilize or autoclave.
• The PRO Monitor, when used with applied parts and
accessories approved by GE Medical Systems
Information Technologies, is protected against
defibrillator damage.
Notes
• Waveforms may be distorted and readings inaccurate
when electrosurgical cautery equipment is used while
monitoring with the PRO Monitor.
• The electromagnetic compatibility profile of the PRO
Monitor may change if accessories other than those
specified for use with the PRO Monitor are used.
• Trend data are retained in the PRO Monitor when it is
turned off, except when the default is overridden by
selecting the Trend button under the Service menu.
9
Product Compliance
The DINAMAP® PRO Monitor is classified in the following
categories for compliance with IEC 601-1:
• Class l, internally powered
•Transportable
• For continuous operation
• Not suitable for use in the presence of flammable
anesthetics
• Not for use in the presence of an oxygen-enriched
atmosphere (oxygen tent)
• Type BF applied parts
• IPX1, degree of protection against ingress of water
• Sterilization/Disinfection, see Appendix F
®
DINAMAP
CLASSIFIED WITH RESPECT TO ELECTRIC SHOCK, FIRE
AND MECHANICAL AND OTHER SPECIFIED HAZARDS
ONLY IN ACCORDANCE WITH CAN/CSA C22.2 NO.
601.1. ALSO EVALUATED TO IEC-601-2-30.
PRO MONITOR
0086
This product conforms with the essential requirements
of the Medical Device Directive. Accessories without
the CE mark are not guaranteed to meet the Essential
Requirements of the Medical Device Directive.
10
Symbols
The following symbols are associated with the PRO Monitor.
Note: The type of model determines which symbols appear
on the Monitor.
Attention, consult accompanying documents
Defibrillator-proof type BF equipment
Power ON/OFF
SpO
2
START/STOP BP
AUTO BP
Battery Power
MAP
mmHg
MAP (Mean Arterial Pressure)
Temperature
Beats Per Minute
BPM
SILENCE
External Communications Port Connector
Artifact
External AC or DC Power Indicator
External DC Power Input
External AC Power Input
SpO
Motion Artifact / Low Perfusion (US)
2
SpO
Motion Artifact / Low Perfusion (Europe)
2
11
Packaging label depicting the
transportation and storage
atmospheric pressure range
of 500 to 1060 hPa.
IPX1
The DINAMAP
against vertically falling drops of water and
conforms to IEC-529 standard at level of
IPX1. Vertically falling drops of water shall
have no harmful effects to the Monitor.
®
PRO Monitor is protected
12
Getting Started
Unpacking the Monitor and Accessories
Before attempting to use the PRO Monitor, take a few
minutes to become acquainted with the Monitor and its
accessories. Unpack the items carefully, and check them
against the contents checklist enclosed in one of the
accessory boxes. This is also a good time to check for any
damage or shortage. If there is a problem or shortage,
contact GE Medical Systems Information Technologies.
It is recommended that all the packaging be retained, in case
the Monitor must be returned for service in the future.
Power Sources
The PRO Monitor is designed to operate from either an
internal lead-acid battery, AC mains or an IEC 601-1
compliant DC power source (see Appendix A). For
replacement rechargeable batteries, please refer to the
Service section of this manual.
The Monitor contains five fuses. Two AC line input fuses are
mounted internally and are replaceable only by qualified
service personnel. The remaining three fuses are autoresetable and mounted within the Monitor. These fuses
protect the low voltage DC input, the battery, and the +5 V
output on the host port connector.
Powering the Monitor
Before the PRO Monitor is used for the first time, the battery
should be charged in the Monitor for at least 8 hours.
Refer to the illustration of the rear panel connections.
Looking at the rear of the PRO Monitor, remove the battery
compartment cover. Insert the rechargeable battery into the
compartment so that the battery terminals fit into the power
clips at the bottom of the compartment. Then replace the
cover. Insert the power cord plug into the mains external
power socket (2) and plug into an AC outlet.
Refer to the illustration of the front panel controls and
indicators. With mains or external DC power connected, the
green external power indicator LED (14) will light to indicate
that external power is being applied and that the battery is
13
charging. If the battery is not inserted, the external power
indicator LED will flash (short flash approx. every 4 sec).
When the Monitor is running on battery power, a battery
icon appears in LCD area 3 (toggling with the time indicator)
indicating the charge status.
During battery-only operation, the yellow battery power
indicator LED (17) will light. When the battery becomes
discharged beyond the low battery threshold, the indicator
will begin to flash, and the Monitor will sound warning beeps
every 30 seconds. At this point, the Monitor should be
connected to an AC outlet to recharge the battery. If the
Monitor continues to be used without charging the battery,
the message WARNING: THE BATTERY IS TOO LOW FOR MONITOR TO FUNCTION. TURN MONITOR OFF
appears. The Monitor shuts down all functions until it is
turned off and the battery is recharged or replaced. To run
the Monitor on AC power, it must be powered off and then
on again.
Battery charging will take place as long as the Monitor
remains connected to an external AC power source. A
battery that is fully discharged can be fully recharged in 1
hour 50 minutes when the Monitor is switched off or 8 hours
if the Monitor is switched on.
Notes
• To prolong the life of the battery, keep the Monitor
connected to an AC outlet whenever possible. NEVER
allow the battery to become completely discharged. A
fully charged battery will power the Monitor for
approximately 2 hours and should survive between 200
and 500 charge/discharge cycles. When it is necessary
to replace the battery, refer to the “Compatibility Table
and Reorder Codes” listed in Appendix D. To ensure full
charge cycles, replace only with a recommended
battery. If the Monitor is to be stored for some time, first
charge the battery and then remove it and store it
separately from the Monitor.
• For continued safety, use only a power cord of listed
type SJT, three-conductor, min. No. 18 AWG, terminated
14
Getting Started
in a medical/hospital grade attachment plug, provided
with the following cord tag: “Hospital Grade Plug."
Grounding integrity can only be maintained when
equipment is connected to an equivalent receptacle
marked "Hospital Grade."
• Where the integrity of the external earth conductor in
the installation or its arrangement is in doubt, the
Monitor must be operated from its internal battery.
General Caution
• Do not touch either the pin of the DC input connector
(3) or the terminals within the battery compartment (1)
and the patient at the same time.
15
Rear Panel Connections
1 Battery compartment cover: Retains and protects internal
battery.
2 Mains input: Used to connect to AC power supply.
3 External power socket: To be used with approved GE
Medical Systems Information Technologies AC-DC power
converter ONLY.
4 Inactive temperature cable storage: Inactive temperature
probe cable attaches here (Models 200V2 and 400V2).
5 Pole clamp: Used to clamp monitor to pole or stand.
6 Data interface connector: Host communications port
(15 way D-type RS-232 serial port) for use only with
equipment conforming to IEC 601-1, configured to
comply with IEC 601-1-1.
16
Getting Started
Front Panel Controls and Indicators
7 Systolic pressure display: 3-digit red LED indicates
measured systolic BP in mmHg.
8 Active temperature probe holster: Temperature probe
that is being used stored here (Models 200V2 and
400V2).
9 Diastolic pressure display: 3-digit red LED indicates
measured diastolic BP in mmHg.
10 Temperature probe cover storage: Box of probe covers
stored here (Models 200V2 and 400V2).
11 Inactive temperature probe holster: Extra temperature
probe can be stored here (Models 200V2 and 400V2).
12 Temperature display: 4-digit red LED indicates measured
temperature (Models 200V2 and 400V2).
13 °C °F display: Indicates whether temperature is being
displayed in degrees Celsius or Fahrenheit (Models
200V2 and 400V2).
17
14 External power indicator: Green LED indicates external
power status and battery charging status of monitor.
15 Temperature probe connector: Temperature probe cable
attaches here (Models 200V2 and 400V2).
16 ON/OFF switch: Controls on/off state of monitor; push
for power on and push again for power off.
17 Battery power indicator: Yellow LED indicates operation
and charge status of internal battery.
18 SpO
sensor connector: SpO2 sensor extension cable
2
attaches here (Models 300V2 and 400V2).
19 Mean arterial pressure display: 3-digit red LED indicates
measured MAP in mmHg and shows instantaneous cuff
pressure during BP determination.
20 SpO
pulse indicator: Yellow LED in heart symbol flashes
2
to indicate that real-time pulse rate measurements are
being derived from SpO
signals (Models 300V2 and
2
400V2).
21 Rotor: Used to highlight and select items in LCD menus; if
monitor is off, pressing rotor will switch monitor on.
22 Pulse BPM display: 3-digit yellow LED shows pulse rate in
beats per minute.
23 SpO
display: 3-digit red LED indicates oxygen saturation
2
in % (Models 300V2 and 400V2).
24 SpO2 motion/artifact indicator LED: For NELLCOR, LED
Illuminates when motion artifact is detected (Models
300V2 and 400V2). For Masimo, LED illuminates when
low perfusion or low signal quality is detected. (Models
300V2 and 400V2).
25 LCD (liquid crystal display): Displays all alarms, user
interface messages, and configuration options.
26 Alarm silence switch: Alternately mutes and enables
audible alarms; when pushed once after alarm sounds
(silence on), switch lights to indicate that audible alarms
have been silenced for 2 minutes.
27 AUTO BP key: Press to start Auto BP mode.
28 Light sensor: Automatically measures ambient light to set
LED display intensity.
29 START/STOP BP key: Press to start or stop a BP, Auto,
Stat, or Vitals determination.
30 Cuff connector: BP cuff hose attaches here.
18
Getting Started
Switching the Monitor On and Off
To switch the DINAMAP PRO Monitor on, press and hold
the power ON/OFF switch (16) for at least 10 seconds or
press the rotor (21).
As the Monitor powers up, it will run a short self-test routine,
which will flash all the indicator lights and then beep the
warning speaker. After a few seconds the system will be
ready for operation, as indicated by the appearance of the
main menu on the LCD (25).
To switch the Monitor off, push the power ON/OFF switch
(16) again. This will terminate any measurements that may be
in progress and automatically deflate the cuff.
When the Monitor is operating on the internal battery only,
battery life is enhanced by the use of the sleep mode.
However, the PRO Monitor will not enter sleep mode if an
alarm is active. If no controls are used and no determinations
are being made, the Monitor will enter sleep mode after a
time which can be preset by the operator. All LED displays
will be blanked except for a dash in the far-left systolic
position, and any existing readings will be transferred to the
LCD, which displays the message “Sleep Mode Active.”
Moving the rotor or pressing a key will “wake up” the
Monitor.
Liquid Crystal Display (LCD)
Menu Area
This area displays the name of the menu that has option
buttons available for selection. Normal text in the menu area
19
appears dark on a light background, while the text of
selected buttons appears light on a dark background.
Note: Some menus have six option buttons. In these cases,
there is no space available to display the menu title.
Area 2
This area displays data from one of three different sources.
• Source 1: SpO
plethysmograph (Models 300V2 and
2
400V2)
• Source 2: Last three BP readings
• Source 3: Error and warning messages
Note: Refer to “Display Button” in the “Using the Menu
System” section for instructions on setting Area 2.
Area 3
This area displays the time, the time lapsed since the last BP
determination, the battery icon (if operating on battery
power, the time and battery icon toggle), and the BP and
Printer modes.
Using the Printer
Installing the Paper
Turn the PRO Monitor so that the side is facing you. While
grasping the side of the Monitor, firmly press the notched
indentations on the printer door to open it. The printer door
will pop open. With the Monitor powered on, place the roll
of paper into the compartment so that the end of the paper
comes off the top, and thread it between the two printer
plates. As the paper touches the plates, the paper will begin
to auto-feed itself into the printer. Feeding the end of the
paper strip through the slot in the door, firmly press the
notched indentation on the side of the printer door to close
it. Use the paper release lever to clear a paper jam or
manually feed the paper.
Note: Make sure that the roll of paper is tightly wound.
20
Getting Started
Any time the paper is loaded, the printer automatically prints
a test strip with the DINAMAP
visible on the paper, check that the paper roll has been
installed in the correct position (refer to diagram). To tear off
the printout, use a slight sideways action to pull the paper
sharply up across the serrated edge of the door.
Printer Alarms
If the Monitor is switched on with no paper installed or with
the printer door open, the message “No Paper” will appear
next to “PRNT” in Area 3 of the LCD. When new paper is
installed and the printer door is closed, the message will
change to “Manual” for Manual print or “Auto” for Auto
print, depending on the status before the paper change.
If the paper runs out during a print request or if an attempt is
made to print when no paper is installed, the message
“Printer - No Paper” will appear in Area 2 of the LCD and an
audible alarm will sound. In addition, the message “No
Paper” will appear next to “PRNT” in Area 3 of the LCD. To
clear the alarm, press the rotor. The message in Area 3 of the
LCD will remain until new paper is installed and the printer
door is closed. (See “Using the Menu System.”)
Installing new paper will cause the DINAMAP PRO header
to be printed, thereby confirming that the paper is installed
correctly and that the printer is operational. The message
next to “PRNT” in Area 3 of the LCD will change to “Auto”
or “Manual” to identify the operating mode of the printer.
After power-off, the operating mode of the printer returns
to the previous user-selected setting (Auto or Manual)
unless specified otherwise in the Print button under the
Service Button.
®
PRO name on it. If no print is
21
Storage
Store thermal paper in a cool, dry place. The printed strip
(thermal paper recording) should not be
• exposed to direct sunlight,
• exposed to temperatures over 100 °F/38 °C or relative
humidity over 80%, or
• placed in contact with adhesives, adhesive tapes, or
plasticizers such as those found in all PVC page
protectors.
Note: When in doubt about long-term storage conditions,
store a photocopy of the thermal paper recording.
Cautions
• The paper is thermally activated; therefore, do not store
it in a hot place as discoloration may result.
• Use only replacement paper rolls (58 mm) from GE
Medical Systems Information Technologies.
22
Using the Monitor
Noninvasive Blood Pressure Determination
Description
The BP parameter is included in Models 100V2, 200V2,
300V2, and 400V2. Blood pressure is monitored
noninvasively in the PRO Monitor by the oscillometric
method, which measures the amplitude of the pressure
oscillations within the blood pressure cuff. Further
information about the oscillometric method is in Appendix C.
The PRO Monitor has four BP modes: 1. Manual, 2. Auto,
3. Stat, and 4. Vitals (UK: All Obs). The mode, which is
selected by the user, is shown on the LCD (25). The BP
measurements are automatic, and once the cycle is
complete the LED displays (7, 9, 19, 22) show systolic
pressure, diastolic pressure, mean arterial pressure, and pulse
rate.
1. Manual BP determinations are started by pressing the
START/STOP BP key (29). In the Manual mode, the blood
pressure is determined one time.
2. Auto BP determinations are started by selecting the
AUTO BP key (27) or the Auto button under the Set BP
(UK: BP Mode) button in the Main menu.
When Auto mode is selected, a number at the right of the
Auto button indicates the time interval between each
reading. To change the time interval, choose the box
around the number and turn the rotor until the desired
interval is reached. The interval can be set between 1 and
120 minutes (1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90, and
120 minutes). Press the rotor to confirm the setting.
3. Stat determinations are started by selecting the Stat
button under the Set BP button (UK: BP Mode) in the
Main menu. In the Stat mode, the blood pressure is
determined as many times as possible in 5 minutes.
4. Vitals (UK: All Obs) determinations are started by
selecting the Vitals (UK: All Obs) button in the Main
menu. (Refer to the “Using the Menu System” section.)
Selection of this button initiates a BP determination while
23
allowing SpO2 and predictive temperature determinations
to be monitored and recorded (depending on Monitor
model). In the Vitals (UK: All Obs) mode, the blood
pressure is determined one time.
Before each BP determination, the Monitor performs a test
to ensure that the cuff pressure is below a specified level.
The determination is delayed until this condition is met.
During the delay, the BP values are displayed as zero.
The Monitor senses the type of hose being used and
automatically uses adult/pediatric monitoring parameters or
neonatal monitoring parameters, as appropriate.
Audible and visible alarms occur when a value for systolic
pressure, diastolic pressure, mean arterial pressure, or pulse
rate is outside the selected high or low limit.
Instructions for cleaning and disinfecting BP cuffs are in
Appendix F.
General Warnings
• The PRO Monitor will not measure blood pressure
effectively on patients who are experiencing seizures or
tremors.
• Arrhythmias will increase the time required by the PRO
Monitor to determine a blood pressure and may extend
the time beyond the capabilities of the Monitor.
• In Manual mode, the PRO Monitor displays the results
of the last blood pressure determination for a duration
of time set by the user or until another determination is
completed. If a patient’s condition changes between
one determination and the next, the Monitor will not
detect the change or indicate an alarm condition.
• Devices that exert pressure on tissue have been
associated with purpura, skin avulsion, compartmental
syndrome, ischemia and/or neuropathy. To minimize
these potential problems, especially when monitoring
at frequent intervals or over extended periods of time,
make sure the cuff is applied appropriately and
24
Using the Monitor
examine the cuff site and the limb distal to the cuff
regularly for signs of impeded blood flow.
• Do not apply external pressure against cuff while
monitoring. Doing so may cause inaccurate blood
pressure values.
• Use care when placing cuff on extremity used to
monitor other patient parameters.
• The PRO Monitor is designed for use only with dualtube cuffs.
• Use only accessories recommended by GE Medical
Systems Information Technologies. Failure to use
recommended accessories may result in inaccurate
readings. See Appendix D.
• Blood pressure cuffs should be removed from the
patient when the Monitor is powered off. If the
extremity remains cuffed under these conditions or if
the interval between blood pressure determinations is
prolonged, the patient’s limb should be observed
frequently and the cuff placement site should be
rotated as needed.
General Cautions
• Accuracy of BP measurement depends on using a cuff
of the proper size. It is essential to measure the
circumference of the limb and to select the proper size
cuff. The air hoses are color-coded according to size of
the patient. The gray 12- or 24-foot hose (3.66 m or 7.3
m) is required on patients who require cuff sizes from
infant through thigh cuffs. The teal (blue-green) 12-foot
hose (3.66 m) is required for the neonatal cuff sizes #1
through #5.
• If it becomes necessary to move the cuff to another
limb, make sure the appropriate size cuff is used.
• The pulse rate derived from a BP determination may
differ from the heart rate derived from an EKG
waveform because the PRO Monitor measures actual
peripheral pulses, not electrical signals or contractions
from the heart. Differences may occur because
electrical signals at the heart occasionally fail to
25
produce a peripheral pulse or the patient may have
poor peripheral perfusion. Also, if a patient’s beat-tobeat pulse amplitude varies significantly (e.g., because
of pulsus alternans, atrial fibrillation, or the use of a
rapid-cycling artificial ventilator), blood pressure and
pulse rate readings can be erratic, and an alternate
measuring method should be used for confirmation.
General Notes
• A patient’s vital signs may vary dramatically during the
use of cardiovascular agents such as those that raise or
lower blood pressure or those that increase or decrease
heart rate.
• Because treatment protocols based on the patient’s
blood pressure may rely on specific values and differing
measurement methods, such as auscultatory, clinicians
should note a possible variance from values obtained
with the PRO Monitor in planning patient care
management. The PRO Monitor values are based on the
oscillometric method of noninvasive blood pressure
measurement and correspond to comparisons with intraaortic values within ANSI /AAMI Standards for accuracy
(a mean difference of ± 5 mmHg, and a standard
deviation of ± 8 mmHg).
• Several conditions may cause the BP parameter to
calculate and display only the mean arterial pressure
(MAP) without a systolic and diastolic reading. These
conditions include very low systolic and amplitude
fluctuations, so an accurate calculation for these values
can’t be made (e.g., patient in shock); too small of a
difference between systolic and MAP calculations in
relationship to the difference between diastolic and
MAP; or a leak has occurred in the PRO Monitor
(1. Check all BP connections 2. Monitor may need
calibration and leak testing). If only the MAP value is
displayed, the systolic and diastolic will display dashes
(---) and an alarm message “N99-BP FAILED” will be
displayed.
26
Using the Monitor
Procedures
1. Connect the end of the air hose which has quick-release
clips to the cuff connector (30) on the front of the
Monitor. Make sure that the hose is not kinked or
compressed.
Note: To disconnect the hose from the Monitor, squeeze
the quick-release clips together and pull the plug from the
cuff connector (30).
2. Select the appropriate blood pressure measurement site.
Because normative values are generally based on this site
and as a matter of convenience, the upper arm is
preferred. When upper arm size or shape, the patient’s
clinical condition, or other factors prohibit use of the
upper arm, the clinician must plan patient care
accordingly, taking into account the patient’s
cardiovascular status and the effect of an alternative site
on blood pressure values, proper cuff size, and comfort.
The figure shows the recommended sites for placing cuffs.
Warning: Do not place the cuff on a limb being used for
intravenous infusion or any area where circulation is
compromised or has the potential to be compromised.
3. If patient is standing, sitting, or inclined, ensure that cuffed
limb is supported to maintain cuff at level of patient’s
heart. If cuff is not at heart level, the difference in systolic
and diastolic values due to hydrostatic effect must be
considered. Add 1.80 mmHg to values for every inch
(2.54 cm) above heart level. Subtract 1.80 mmHg from
values for every inch (2.54 cm) below heart level.
4. Select appropriate cuff size. Measure patient’s limb and
select appropriately sized cuff according to size marked
on cuff or cuff packaging. When cuff sizes overlap for a
specified circumference, choose the larger size cuff.
Precaution: Accuracy depends on use of proper size cuff.
27
5. Inspect cuff for damage. Replace cuff when aging,
tearing, or weak closure is apparent. Do not inflate cuff
when unwrapped.
Precaution: Do not use cuff if structural integrity is
suspect.
6. Connect the cuff to the air hose. Thread the cuff
connectors onto the hose connectors until finger tight. Do
not overtighten.
Warning: It is mandatory that the appropriate hose and
cuff combination be used. Any attempt to modify the
hose will inhibit the Monitor from switching between
the neonatal and adult measurement modes.
Note: In normal use, each cuff will have its own hose, so it
will not usually be necessary to disconnect them. If it is
necessary to do so, carefully unscrew the cuff from the
hose. Care should be taken in reconnecting the cuff to a
hose, ensuring that threads of the cuff and hose are in
alignment and no cross-threading occurs.
7. Inspect patient’s limb prior to application.
Precaution: Do not apply cuff to areas where skin is not
intact or tissue is injured.
8. Palpate artery and place cuff so that patient’s artery is
aligned with cuff arrow marked “artery.”
9. Squeeze all air from cuff and confirm that connection is
secure and unoccluded and that tubing is not kinked.
10.Wrap cuff snugly around the patient’s limb. Cuff index
line must fall within the range markings. Ensure that hook
and loop closures are properly engaged so that pressure
is evenly distributed throughout cuff. If upper arm is used,
place cuff as far proximally as possible.
11.Proper cuff wrapping should be snug, but should still
allow space for a finger between patient and cuff. Cuff
should not be so tight as to prevent venous return
between determinations.
Warning: Using a cuff that is too tight will cause venous
congestion and discoloration of the limb, but using a
cuff that is too loose may result in no readings and/or
inaccurate readings.
12.Proceed with monitoring in the Manual, Auto, Stat, or
Vitals (UK: All Obs) mode.
28
Using the Monitor
Manual Mode
To start a determination, press the START/STOP BP key (29).
A normal, uninterrupted Manual cycle takes about 40
seconds. The cuff pressure must drop below 5 mmHg
(neonate) or 15 mmHg (adult) before another determination
can be started. BP information will be displayed for a
duration of time set by the user. The factory default setting is
2 minutes (refer to “Retain Values” in the Using the Menu System section for information on how to change this
setting). The user can set the display duration in 2, 5, 10, 15,
20, 30, 45, 60, 90, 120 minute increments. If another
determination is initiated during the user-set timeframe, the
display will change. This applies to Manual and Vitals (UK: All
Obs) modes. After power-off, the operating mode returns to
the default setting of Manual. The default setting of Manual
can be overridden to return to the previous user-selected
setting (Auto or Manual) by selecting Set BP (UK: BP Mode)
under the Service menu.
Note: The START/STOP BP key is an on-off switch; pressing it
will stop any BP determination (Manual, Auto, Stat, or Vitals)
that is in progress.
Auto Mode
Auto BP determinations are started by selecting the AUTO
BP key (27) or the Auto button under the Set BP button (UK:
BP Mode) in the Main menu.
Selecting the AUTO BP key (27) brings up the Set BP menu
(UK: BP Mode) and automatically starts an Auto BP
determination as long as the Monitor is in Manual BP mode.
If the Monitor is already in Auto BP mode, selection of the
AUTO BP key (27) brings up the Set BP menu (UK: BP
Mode) without starting a new determination until the preset
time interval has expired. Pressing the START/STOP BP key
during a series of Auto BP determinations will cancel the
determination in progress.
29
When Auto mode is selected, a number at the right of the
Auto button indicates the time interval between each
reading. To change the time interval, choose the box around
the number and turn the rotor until the desired interval is
reached. The interval can be set between 1 and 120 minutes
(1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90, and 120 minutes).
Press the rotor to confirm the setting. After power-off, the
operating mode returns to the default setting of Manual. The
default setting of Manual can be overridden to return to the
previous user-selected setting (Auto or Manual) by selecting
Set BP (UK: BP Mode) under the Service menu.
In the Auto mode, the pressure must be below 5 mmHg
(neonate) or 15 mmHg (adult) for at least 30 seconds before
another determination can be started. BP information will be
displayed on the LED until the next determination is started.
This applies to Auto mode only.
Note: To cancel an Auto BP determination, select the
Manual button in the Set BP menu (UK: BP Mode).
Stat Mode
Multiple BP readings can be taken at any time by selecting
the Stat button under the Set BP button (UK: BP Mode) in
the Main menu. Stat mode can also be accessed by pressing
the AUTO BP key (27) and then selecting the Stat button
when the Set BP menu (UK: BP Mode) appears.
If a Manual determination is not in progress, a 5-minute
series of determinations will start. If a Manual determination
is in progress, that determination will become the first in the
series. A normal, uninterrupted Stat sequence will give the
first set of systolic, diastolic, and mean arterial pressure
values and pulse rate within 15 to 20 seconds. Selecting the
Stat button during a series of Stat determinations will cancel
the determination in progress and the rest of the series. BP
information will be displayed on the LED until the
determination has been canceled or completed. This applies
to Stat mode only.
The series begins with cuff inflation to a pressure above the
previous systolic pressure or, if no previous systolic value is
stored, to approximately 150 mmHg for adult/pediatrics. The
30
Using the Monitor
initial target pressure selection for neonates is 110 mmHg.
Artifact rejection is relaxed in the Stat mode for adult/
pediatric patients to allow for accelerated determinations. If
a BP or Stat reading has been made previously, the first new
systolic value will flash on the LED display (7) within a few
seconds and will continue to flash until the end of the
determination. At that point a short tone will sound and the
updated systolic, diastolic, and mean arterial pressures and
pulse rate will appear on their LED displays (7, 9, 19, 22). The
Monitor will begin another determination once the pressure
is below 5 mmHg for 8 seconds (neonates) or 15 mmHg for
4 seconds (adults), unless the 5-minute period has ended or
the determination has been canceled.
Note: Alarm limits are disabled while in Stat mode.
TURBO
Description
The Temperature parameter is included in Models 200V2
and 400V2. The PRO Monitor uses IVAC* TURBO
technology and can be used with both oral and rectal
temperature probes. The TURBO
consists of an electronic thermometer that uses a heatsensing device known as a thermistor to sense temperature.
The thermistor is part of the electrical circuit and is located at
the tip of the probe. To obtain temperatures, the probe tip
measures the rate of change in temperature when the
thermistor comes into contact with surrounding tissue. A
final temperature is calculated based on this rate of change.
During a temperature determination, the temperature display
(12) provides a progress meter and probe ready indicator. In
the far-left position, a single horizontal line indicates the
probe is ready to start a determination after removal from the
probe holster. In the far-right position of the temperature
display, a “chase sequence” around the outside space
indicates a predictive temperature determination is in
progress. During monitor mode, the temperature readings
flash constantly.
*IVAC is a trademark of Alaris Medical Systems
TEMP
TEMP™ parameter
TEMP™
31
Temperature is shown on the temperature display in degrees
Celsius or Fahrenheit, and the unit of measure is indicated by
the °C °F display (13). The default, which is Celsius, can be
changed in the Clinician Menu (please refer to the “Using the
Menu System” section of this manual).
Temperature determinations can be either predictive or
monitoring. The user can make this selection from the Main
Menu using the More... and then Temp buttons (please refer
to the “Using the Menu System” section of this manual).
Predictive Mode
In predictive mode, a final temperature is displayed with an
audible single tone. Upon initiation of a measurement, the
old temperature measurement, if present, is cleared. A
predictive measurement is initiated when the probe is
removed from the probe holster. When a predictive mode
measurement is in progress, it is terminated when one of the
following occurs:
• A final value is determined.
• The probe is inserted into the probe holster.
• The temperature measurement mode is automatically
switched to monitor mode because a predictive result
was unable to be computed.
• A Temperature alarm is issued.
A predictive temperature measurement value is
automatically cleared when it is older than the display time
set by the user and the probe has been returned to the probe
holster. When the value is older than 5 minutes and the
probe has not been returned to the probe holster, that value
is clear.
Monitor Mode
Monitor mode is most commonly used for axillary
temperature determinations. In monitor mode, the display is
updated continually as the patient’s temperature rises or falls.
Monitor mode is automatically initiated when a predictive
mode measurement terminates after approximately 10
seconds of not being able to successfully compute a result.
32
Using the Monitor
When a monitor mode measurement is in progress, it is
terminated when the probe is inserted into the probe holster.
Note: These temperature readings are not stored in trends
and not reported via host comms.
General Warning
• The performance of the Monitor may be degraded if it
is operated outside of the environmental conditions
specified in Appendix A.
General Cautions
• Be careful not to overextend the coiled cord of the
temperature probe. Overextension can damage the
probe coil connector interfaces.
• Accurate oral temperatures (blue) can only be obtained
by placing the probe under the tongue in the right or
left sublingual pocket. Temperatures in other locations
in the mouth can vary by more than 2 °F or 1 °C.
• Accurate rectal temperatures can only be obtained by
using the red temperature probe. Red and blue
temperature probes are not interchangeable.
• Do not allow the tip of the predictive temperature
probe to come into contact with a heat source (e.g.,
hands or fingers) prior to taking a temperature
determination. If this occurs, allow 5 seconds for the
probe tip to cool before proceeding.
• Electromagnetic Compatibility: Operating the
thermometer near equipment which radiates highenergy electromagnetic and radio frequencies
(electrosurgical/cauterizing equipment, portable
radios, cellular telephones, etc.) may cause false alarm
conditions. If this happens, reposition the thermometer
away from the source of interference and perform a
new measurement.
• Use only IVAC
the TURBO
and thermal characteristics of the probe covers can
affect the performance of the instrument. Inaccurate
®
Probes and P850A probe covers with
TEMP™ Thermometer. The size, shape,
33
readings may occur unless IVAC® probe covers are
used.
• If a patient’s temperature is below 96.0 °F, the unit will
automatically switch from the normal mode into the
monitor mode within 60 seconds. It will continue to
monitor the patient’s temperature until the probe is
removed from the patient and returned to the storage
well.
Procedures for Oral Predictive Mode Determinations
For oral temperature measurement, use the blue oral probe.
1. Connect the temperature probe cable to the temperature
probe connector.
2. Remove the temperature probe from the probe holster.
An audible single-tone sounds. Place a protective
temperature probe cover on the probe.
3. Have the patient open his/her mouth slightly. Holding the
probe loosely, insert the probe tip into the sublingual
pocket where the richest blood is located.
4. Hold the probe during the entire temperature
measurement process, and keep the probe tip in contact
with the tissue at all times. Do not allow the patient to
reposition the probe.
5. The determination begins automatically. Hold the
temperature probe steady until the determination is
complete. This takes approximately 10 seconds, during
which time a “chase sequence” of arrows in the right-side
of the Temperature window appears to indicate progress.
6. When the determination is complete, an audible tone
sounds and the temperature appears on the display.
34
Using the Monitor
7. Remove the probe.
Discard the
disposable cover by
holding the probe as
you would a syringe
and pressing the
button on the probe
handle. Place the
probe in the probe
holster. An audible single-tone sounds. Once you place
the probe in the probe holster, the temperature values will
be cleared in the user selected timeframe.
Procedures for Rectal Predictive Mode Determinations
To measure rectal temperature use the red rectal probe.
1. Connect the temperature probe cable to the temperature
probe connector.
2. Remove the temperature probe from the probe holster.
An audible single-tone sounds. Place a protective
temperature probe cover on the probe. Failure to firmly
install the probe cover may result in the probe cover
becoming loose or disengaged during use. Be careful not
to press the probe ejection button where the cord exits
the probe as this might loosen or eject the probe cover.
3. Touch the tissue about a half inch (1.3 cm) above the
sphincter muscle and carefully insert the probe, using
current hospital technique for penetration. (The use of a
lubricant is optional.)
4. The determination begins automatically. To ensure
continuous tissue contact and maximize patient comfort,
hold the probe in position until the determination is
complete. This takes approximately 10 seconds, during
which time a “chase sequence” of arrows in the right-side
of the Temperature window appears to indicate progress.
5. When the determination is complete, an audible tone
sounds and the temperature appears on the display.
6. Remove the probe. Discard the disposable cover by
holding the probe as you would a syringe and pressing
the button on the probe handle. Place the probe in the
probe holster. An audible single-tone sounds. Once you
place the probe in the probe holster, the temperature
values will be cleared in the user selected timeframe.
35
Procedures for Monitor Mode Determinations (Axillary
Determinations)
1. Connect the temperature probe cable to the temperature
probe connector.
2. Remove the temperature
probe from the probe
holster. An audible singletone sounds. Place a
protective temperature
probe cover on the probe
and insert the probe in the
patient's axilla, making sure
the tip of the probe is in
contact with the skin and
positioned as close as possible to the axillary artery with
the patient's arm held close to his/her side.
5. The determination begins automatically. The dashes in the
vital signs area are replaced with a flashing temperature
value as soon as the probe warms up. Leave the probe in
place for the same length of time as required by standard
hospital procedure for taking an axillary temperature. The
Monitor does not beep to indicate a final reading.
6. Remove the probe. Discard the disposable cover by
holding the probe as you would a syringe and pressing
the button on the probe handle. Place the probe in the
probe holster. An audible single-tone sounds. Once you
place the probe in the probe holster, the temperature
values will be cleared in the user-selected timeframe.
Notes
• If there is a long delay from the time the probe is
removed from the probe holster until it is inserted into
the patient’s mouth, it is possible that the instrument will
not display a final temperature. If this occurs, insert the
probe into the probe holster, remove it again, and start a
new measurement.
• If an alarm is actively sounding, an audible tone will not
sound.
• If the probe is not inserted/placed within 30 seconds of
probe removal, the Monitor may issue an alarm.
36
Using the Monitor
• If the probe tip temperature is 92.0° F or higher (33.3° C)
when taken out of the probe holster, the thermometer
will not be able to perform a predictive measurement.
Instead, the thermometer will automatically go into
monitor mode. The temperature reading will then flash.
A correct final temperature reading may require 3
minutes or longer. The Monitor will not beep at final
temperature. It will continue to monitor the patient’s
temperature until the probe is removed from the patient
and returned to the probe holster.
Masimo Set* SpO
2
®
Description
The SpO
different leading technologies: NELLCOR
MASIMO SET
Monitor to see which SpO
contains. The SpO
fascia. This section refers to MASIMO SET
parameter in the PRO Monitor is available in two
2
®
. Please refer to the front of your PRO
technology your monitor
technology logo will be on the front
2
2
®
OxiMax™and
®
SpO2
technology.
The SpO
400V2. To begin SpO
parameter is included in Models 300V2 and
2
monitoring, simply place the SpO2
2
sensor on the patient’s finger; monitoring begins
automatically. Functional oxygen saturation (SpO
) of arterial
2
blood is noninvasively and continuously monitored in the
PRO Monitor using pulse oximetry technology from
MASIMO SET
®
. Functional SpO2 is the ratio of oxygenated
hemoglobin to hemoglobin that is capable of transporting
oxygen. This ratio, expressed as a percentage, is shown in the
SpO
window, which is continually updated.
2
*Masimo SET® is a trademark of Masimo Corporation. Possession or
purchase of this device does not convey any express or implied license to
use the device with replacement parts which would, alone, or in
combination with this device, fall within the scope of one or more of the
patents relating to the device.
37
Pulse rate derived from SpO2 appears in the Pulse Rate
window, and the SpO
pulse indicator flashes
2
synchronization with the real-time pulse rate measurements
that are derived from the SpO
signal. A tone sounds at a
2
rate corresponding to the pulse rate and at a pitch
corresponding to the SpO
saturation level. The pitch is
2
highest at 100% oxygen saturation, and it becomes lower as
the saturation level falls. The Monitor displays a pulse bar
graph. The pulse bar graph is proportional to the signal
quality. The artifact indicator LED (24) illuminates when low
perfusion or poor signal quality is detected.
Audible and visible alarms occur when SpO
levels are
2
outside the alarm limits. When a limit alarm occurs, an alarm
message appears in SpO
window.
2
Note: Limit alarms, printing, and trending are not available
for the first 10 seconds of SpO
monitoring.
2
If you select the Alarms button, the Alarms menu appears.
This menu is used to adjust the violation limits for BP and
SpO
. Refer to “Alarms Button” in the “Using the Menu
2
System” section.
If you select the Suspend button, the SpO
alarm is
2
suspended for 2 minutes and then the PRO Monitor returns
to normal SpO
that SpO
appear in the SpO
counting down. Selecting Cancel will cancel the SpO
suspension and return to monitoring SpO
monitoring. A message informing the user
2
is suspending appears in Area 2 and dashes
2
LED while the SpO2 alarm suspend is
2
.
2
alarm
2
38
Using the Monitor
If the Monitor is unable to detect a pulse for 10 seconds
during normal SpO
alternating patient values with dashes. The Monitor returns to
normal SpO
good pulse determinations are made.
2
General Cautions
• If any measurement does not seem reasonable, first
check the patient’s vital signs by alternate means and
the check the pulse oximeter for proper functioning.
• Inaccurate measurements may be caused by:
- Incorrect sensor application or use
- Significant levels of dysfunctional hemoglobins (e.g.
carboxyhemoglobin or methemoglobin)
- Intravascular dyes such as indocyanine green or
methylene blue.
- Exposure to excessive illumination, such as surgical
lamps (especially ones with xenon light source),
bilirubin lamps, fluorescent lights, infrared heating
lamps, or direct sunlight (exposure to excessive
illumination can be corrected by covering the sensor
with a dark or opaque material)
- excessive patient movement
- venous pulsations
- placement of sensor on an extremity with a blood
pressure cuff, arterial catheter, or intravascular line
• The pulse oximeter can be used during defibrillation,
but the readings may be inaccurate for a short time.
• loss of pulse signal can occur in any of the following
situation:
- the sensor is too tight
monitoring, the values in the LED flash,
2
reporting of values when several consecutive
39
- there is excessive illumination from light sources such
as surgical lamp, a bilirubin lamp, or sunlight
- a blood pressure cuff is inflated on the same extremity
as the one with a SP0
sensor attached.
2
- the patient has hypotension, severe vasoconstriction,
severe anemia, or hypothermia
- there is arterial occlusion proximal to the sensor
- the patient is in cardiac arrest or is in shock
Sensors
®
• Before use, carefully read the Masimo
sensor
directions for use.
®
• Use only Masimo
oximetry sensors for SP0
2
measurements. Other oxygen transducers (sensors)
may cause improper SpO
• Tissue damage can be caused by incorrect application
or use of an Masimo
performance.
2
®
sensor, for example by wrapping
the sensor too tightly. Inspect the sensor site as
directed in the sensor Directions for Use to ensure skin
integrity and correct positioning and adhesion of the
sensor.
• Do not use damaged Masimo
Masimo
®
sensor with exposed optical components. Do
®
sensors. Do not use an
not immerse the sensor in water, solvents, or cleaning
solutions (the sensors and connectors are not
waterproof). Do not sterilize by irradiation, steam, or
ethylene oxide. See the cleaning instructions in the
directions for use for reusable Masimo
®
sensors.
• Do not use damaged patient cables. Do not immerse
the patient cables in water, solvents, or cleaning
solutions (the patient cable connectors are not
waterproof). Do not sterilize by irradiation, steam, or
ethylene oxide. See the cleaning instructions in the
directions for us e for reusable Masimo
• Do not use the SpO
function during magnetic
2
®
patient cables.
resonance imaging (MRI). Adverse reactions include
potential burns to patients as a result of contact with
attachments heated by the MRI radio frequency pulse,
potential degradation of the magnetic resonance
image, and potential reduced accuracy of SpO
2
40
Using the Monitor
measurements. Always remove oximetry devices and
attachments from the MRI environment before
scanning a patient.
• The use of cardio-green and other intravascular dyes at
certain concentrations may affect the accuracy of the
SpO
measurement.
2
•The SpO
arterial oxygen saturation. Significant levels of
dysfunctional hemoglobins such as carboxyhemoglobin
or methemoglobin may affect the accuracy of the SpO
measurement.
• The use of accessories, transducers, and cables other
than those specified may result in increased emission
and/or decreased immunity and inaccurate readings of
the Monitor.
• As with any clip-on sensor, pressure is exerted. The
clinician should be cautious in using a clip-on sensor on
patients with compromised circulation (e.g., because of
peripheral vascular disease or vasoconstricting
medications).
• Do not perform any testing or maintenance on a sensor
while it is being used to monitor a patient.
• Bright light sources (e.g., infrared heat lamps, bilirubin
lights, direct sunlight, operating room lights) may
interfere with the performance of the SpO
To prevent such interference, cover the sensor with
opaque material.
function is calibrated to read functional
2
function.
2
2
General Notes
• A patient’s vital signs may vary dramatically during the
use of cardiovascular agents such as those that raise or
lower blood pressure or those that increase or decrease
pulse rate.
• The PRO Monitor that is labeled with MASIMO SET
Technology is compatible only with MASIMO SET
®
sensors.
• Software development, software validation, and Risk and
Hazard Analysis has been performed to a registered
quality system.
41
®
• Use or patient-applied parts are latex-free.
Procedures
1. Select a sensor that is appropriate for the patient and the
clinical situation.
Warning: Do not use a damaged sensor or one with
exposed electrical contacts.
Note: Use only MASIMO
from Masimo
®
Corporation and GE Medical Systems -
®
sensors, which are available
Accessories and Supplies.
2. Following the directions for use supplied with the sensor,
apply the sensor to the patient.
Warnings
Patient safety:
• If you fail to apply the sensor properly, the patient’s
skin could be injured or the ability of the PRO Monitor
to measure oxygen saturation could be compromised.
For example, a clip-on sensor should never be taped
shut. Taping the sensor could damage the patient’s skin
or impair the venous return, thus causing venous
pulsation and inaccurate measurement of oxygen
saturation.
• Excessive pressure from the sensor may cause necrosis
of the skin.
Monitor performance:
•When an SpO
pressure cuff, the SpO
cuff is inflated. If SpO
sensor is on a limb that has a blood
2
data will not be valid when the
2
readings are required during
2
the entire blood pressure determination, attach the
SpO
sensor to the limb opposite the one with the
2
blood pressure cuff.
• Remove nail polish and artificial nails. Placing a sensor
on a polished or an artificial nail may affect accuracy.
Cautions
Patient safety:
• Do not place any clip-on sensor in a patient’s mouth or
on a patient’s nose or toe.
• Do not place a clip-on finger sensor on a patient’s
thumb or across a child’s foot or hand.
42
Using the Monitor
• Observe the sensor site to assure adequate distal
circulation. Sensor sites should be checked at least
every 2 hours and rotated at least every 4 hours.
Monitor performance:
• Placing a sensor distal to an arterial line may interfere
with adequate arterial pulsation and compromise the
measurement of SpO
• Place the sensor so that the LEDs and the photodiode
are opposite each other.
3. Plug the SpO
cable. Then plug the SpO
SpO
sensor connector.
4. Proceed with monitoring. SpO
2
continuously and can run simultaneously with other
measurements.
sensor into the SpO2 sensor extension
2
.
2
sensor extension cable into the
2
determinations run
2
Troubleshooting Masimo SET
This section discusses potential difficulties and suggestions
for resolving them. If the difficulty persists, contact a qualified
service person or your local representative.
The service manual, which is for use by qualified service
personnel provides additional troubleshooting information.
PROBLEM: The pulse amplitude bar indicates a pulse, but no
oxygen saturation or pulse rate values appear on the screen.
CAUSE:
• Excessive patient motion may be making it impossible
for the SpO2 function to find a pulse pattern.
• The sensor may be damaged.
• The patient’s perfusion may be too low to allow the
SpO2 function to measure saturation and pulse rate.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
SpO2 sensor is applied securely and properly, and
replace it if necessary; move the sensor to a new site; or
®
SpO2 Parameter
43
use a disposable adhesive sensor that may tolerate more
motion.
• Replace the sensor.
PROBLEM: The SpO
value or the pulse rate changes
2
rapidly; the pulse amplitude bar is erratic.
CAUSE:
• Excessive patient motion may be making it impossible
for the SpO
function to find a pulse pattern.
2
• An electrosurgical unit (ESU) may be interfering with
performance.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; use a sensor
that tolerates more motion.
If an ESU is interfering:
• Move the SpO
cable as far from the ESU as possible.
2
• Plug the Monitor and the ESU into different AC circuits.
• Move the ESU ground pad as close to the surgical site as
possible.
• The sensor may be damp or may need to be replaced
with a new sensor.
PROBLEM: The oxygen saturation measurement does not
correlate with the value calculated from a blood gas
determination.
CAUSE:
• The SpO2 calculation may not have correctly adjusted
for the effects of pH; temperature; CO2; fetal
hemoglobin; or 2,3-DPG.
• Accuracy can be affected by incorrect sensor application
or use; intravascular dyes; bright light; excessive patient
movement; venous pulsations; electrosurgical
interference; and placement of a sensor on an extremity
that has a blood pressure cuff, arterial catheter, or
intravascular line.
44
Using the Monitor
SOLUTION:
• Check that calculations have been corrected
appropriately for the relevant variable. In general,
calculated saturation values are not as reliable as direct
laboratory hemoximeter measurements.
• If there is excessive light, cover the sensor with opaque
material.
• Circulation distal to the sensor site should be checked
routinely. The site must be inspected every 8 hours to
ensure adhesion, skin integrity, and correct optical
alignment. If skin integrity changes, move the sensor to
another site.
• Try to keep the patient still, or change the sensor site to
one with less motion.
• Observe all instructions, warnings, and cautions in this
manual and in the directions for use of the sensor.
PROBLEM: A valid SpO
disappeared.
CAUSE:
• A BP determination on the same limb is in progress.
SOLUTION:
• An alarm message codes will appear on the screen, and
the audible alarm will sound immediately.
PROBLEM: An SpO2 PLACEMENT error has occurred.
CAUSE:
• Weak or “noisy” signal.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; or replace the
sensor.
• Change sensor type.
• Consider increasing perfusion using heat.
signal was present but has
2
45
• If there is excessive light, cover the sensor with opaque
material.
PROBLEM: A sensor error indicating a bad sensor has
occurred.
CAUSE:
• The sensor or cable may be defective, or the cabling
may be improperly connected.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor/cable is applied securely and properly, and
replace it if necessary.
• Disconnect and reconnect the sensor.
®
NELLCOR
OxiMAX™ SpO
2
Description
The SpO
different leading technologies: NELLCOR
MASIMO SET
see which SpO
SpO
Monitor. This section refers to NELLCOR SpO
Functional oxygen saturation (SpO
parameter in the PRO Monitor is available in two
2
®
. Please refer to the front of your Monitor to
technology your monitor contains. The
technology logo will be on the front fascia of the
2
2
®
OxiMAX and
technology.
2
) of arterial blood is
2
noninvasively and continuously monitored in the PRO
Monitor using pulse oximetry technology from NELLCOR
Functional SpO
is the ratio of oxygenated hemoglobin to
2
hemoglobin that is capable of transporting oxygen.
Pulse rate derived from SpO
display (22), and the SpO
appears in the Pulse BPM
2
pulse indicator (20) flashes
2
synchronization with the real-time pulse rate measurements
*NELLCOR® is a trademark of Nellcor Puritan Bennett
46
®
.
Using the Monitor
that are derived from the SpO2 signal. A tone sounds at a
rate corresponding to the SpO
highest at 100% oxygen saturation and becomes a little
lower with each 1% decrease in saturation level. The
Monitor can display a pulse amplitude bar and a
plethysmographic waveform on the LCD (25). The pulse
amplitude bar graph is proportional to the arterial blood
flow. The artifact indicator LED (24) lights continuously when
the Monitor detects motion sufficient enough to affect
readings.
saturation level. The pitch is
2
Audible and visible alarms occur when SpO
levels are
2
outside the alarm limits. When a limit alarm occurs, a
message appears in Area 2 of the LCD display.
If you select the Alarms button, the Alarms menu appears.
This menu is used to adjust the violation limits for BP and
SpO
. Refer to “Alarms Button” in the “Using the Menu
2
System” section.
If you select the Suspend button, the SpO
alarm is
2
suspended for 2 minutes and then the PRO Monitor returns
to normal SpO
that SpO
appear in the SpO
counting down. Selecting Cancel will cancel the SpO
suspension and return to monitoring SpO
monitoring. A message informing the user
2
is suspending appears in Area 2 and dashes
2
LED while the SpO2 alarm suspend is
2
.
2
alarm
2
If the Monitor is unable to detect a pulse for 10 seconds
during normal SpO
monitoring, the values in the LED flash,
2
alternating patient values with dashes. The Monitor returns to
47
normal SpO2 reporting of values when several consecutive
good pulse determinations are made.
General Warnings
• Do not use the SpO2 function during magnetic
resonance imaging (MRI). Adverse reactions include
potential burns to patients as a result of contact with
attachments heated by the MRI radio frequency pulse,
potential degradation of the magnetic resonance
image, and potential reduced accuracy of SpO
measurements. Always remove oximetry devices and
attachments from the MRI environment before
scanning a patient.
• The use of cardio-green and other intravascular dyes at
certain concentrations may affect the accuracy of the
SpO
measurement.
2
•The SpO
function is calibrated to read functional
2
arterial oxygen saturation. Significant levels of
dysfunctional hemoglobins such as carboxyhemoglobin
or methemoglobin may affect the accuracy of the SpO
measurement.
• Pulse oximetry readings and pulse signals can be
affected by certain environmental conditions, sensor
application errors, and certain patient conditions. See
the appropriate sections of this manual for specific
safety information.
• The use of accessories, transducers, and cables other
than those specified may result in increased emission
and/or decreased immunity and inaccurate readings of
the Monitor.
2
2
General Cautions
• As with any clip-on sensor, pressure is exerted. The
clinician should be cautious in using a clip-on sensor on
patients with compromised circulation (e.g., because of
peripheral vascular disease or vasoconstricting
medications).
48
Using the Monitor
• Do not perform any testing or maintenance on a sensor
while it is being used to monitor a patient.
• Bright light sources (e.g., infrared heat lamps, bilirubin
lights, direct sunlight, operating room lights) may
interfere with the performance of the SpO
To prevent such interference, cover the sensor with
opaque material.
General Notes
• A patient’s vital signs may vary dramatically during the
use of cardiovascular agents such as those that raise or
lower blood pressure or those that increase or decrease
heart rate.
• When the front of the PRO Monitor is labeled with the
Nellcor logo, use only NELLCOR sensors for SPO
measurements.
• Software development, software validation, and Risk and
Hazard Analysis has been performed to a registered
quality system.
Procedures
1. Select a sensor that is appropriate for the patient and the
clinical situation.
Warning: Do not use a damaged sensor or one with
exposed electrical contacts. Do not use sensors, cables,
or connectors that appear damaged.
Note: Use only NELLCOR
are available from: GE Medical Systems - Accessories and
Supplies.
2. Following the directions for use supplied with the sensor,
apply the sensor to the patient.
®
OxiMAX™ sensors, which
function.
2
2
Warnings
Patient safety:
• If you fail to apply the sensor properly, the patient’s
skin could be injured or the ability of the PRO Monitor
to measure oxygen saturation could be compromised.
For example, a clip-on sensor should never be taped
shut. Taping the sensor could damage the patient’s skin
or impair the venous return, thus causing venous
49
pulsation and inaccurate measurement of oxygen
saturation.
• Excessive pressure from the sensor may cause necrosis
of the skin.
• For additional warnings and information, refer to the
NELLCOR
®
sensor’s directions for use.
Monitor performance:
•When an SpO
pressure cuff, the SpO
cuff is inflated. If SpO
sensor is on a limb that has a blood
2
data will not be valid when the
2
readings are required during
2
the entire blood pressure determination, attach the
SpO
sensor to the limb opposite the one with the
2
blood pressure cuff.
• Remove nail polish and artificial nails. Placing a sensor
on a polished or an artificial nail may affect accuracy.
Cautions
Patient safety:
• Do not place any clip-on sensor in a patient’s mouth or
on a patient’s nose or toe.
• Do not place a clip-on finger sensor on a patient’s
thumb or across a child’s foot or hand.
• The sensor disconnect error message and associated
alarm indicate that the sensor is either disconnected or
the wiring is faulty. The user should check the sensor
connection and, if necessary, replace the sensor, cable,
or both.
• Observe the sensor site to assure adequate distal
circulation. Sensor sites should be checked at least
every 2 hours and rotated at least every 4 hours.
Monitor performance:
• Placing a sensor distal to an arterial line may interfere
with adequate arterial pulsation and compromise the
measurement of SpO
• Place the sensor so that the LEDs and the photodiode
are opposite each other.
.
2
50
Using the Monitor
3. Plug the SpO2 sensor into the SpO2 sensor extension
cable. Then plug the SpO
SpO
sensor connector (18).
2
4. Proceed with monitoring. SpO
continuously and can run simultaneously with other
measurements.
Troubleshooting the NELLCOR SpO
This section discusses potential difficulties and suggestions
for resolving them. If the difficulty persists, contact a qualified
service person or your local representative.
The service manual, which is for use by qualified service
personnel provides additional troubleshooting information.
PROBLEM: The pulse amplitude bar indicates a pulse, but no
oxygen saturation or pulse rate values appear on the screen.
CAUSE:
• Excessive patient motion may be making it impossible
for the SpO
function to find a pulse pattern.
2
• The sensor may be damaged.
• The patient’s perfusion may be too low to allow the
function to measure saturation and pulse rate.
SpO
2
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor is applied securely and properly, and
SpO
2
replace it if necessary; move the sensor to a new site; or
use a disposable adhesive sensor that may tolerate more
motion.
• Replace the sensor.
sensor extension cable into the
2
determinations run
2
Parameter
2
PROBLEM: The SpO
value or the pulse rate changes
2
rapidly; the pulse amplitude bar is erratic.
CAUSE:
• Excessive patient motion may be making it impossible
for the SpO
function to find a pulse pattern.
2
• An electrosurgical unit (ESU) may be interfering with
performance.
51
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; use a sensor
that tolerates more motion.
If an ESU is interfering:
• Move the SpO
cable as far from the ESU as possible.
2
• Plug the Monitor and the ESU into different AC circuits.
• Move the ESU ground pad as close to the surgical site as
possible.
• The sensor may be damp or may need to be replaced
with a new sensor.
• If the patient weighs less than 3 kg or more than 40 kg,
apply an OXIMAX reusable sensor (except DS-100, OXIA/N, OXI-P/I) or OxiCliq oxygen transducer to an
appropriate site. These sensors have Faraday shields
which provide added protection from high electronic
noise and ambient light.
PROBLEM: The oxygen saturation measurement does not
correlate with the value calculated from a blood gas
determination.
CAUSE:
•The SpO
for the effects of pH; temperature; CO
calculation may not have correctly adjusted
2
; fetal
2
hemoglobin; or 2,3-DPG.
• Accuracy can be affected by incorrect sensor application
or use; intravascular dyes; bright light; excessive patient
movement; venous pulsations; electrosurgical
interference; and placement of a sensor on an extremity
that has a blood pressure cuff, arterial catheter, or
intravascular line.
SOLUTION:
• Check that calculations have been corrected
appropriately for the relevant variable. In general,
calculated saturation values are not as reliable as direct
laboratory hemoximeter measurements.
52
Using the Monitor
• If there is excessive light, cover the sensor with opaque
material.
• Circulation distal to the sensor site should be checked
routinely. The site must be inspected every 8 hours to
ensure adhesion, skin integrity, and correct optical
alignment. If skin integrity changes, move the sensor to
another site.
• Try to keep the patient still, or change the sensor site to
one with less motion.
• Observe all instructions, warnings, and cautions in this
manual and in the directions for use of the sensor.
PROBLEM: A valid SpO
disappeared.
CAUSE:
• A BP determination on the same limb is in progress.
SOLUTION:
• An alarm message (No signal) will appear on the screen,
and the audible alarm will sound immediately.
PROBLEM: A bad signal error has occurred.
CAUSE:
• Weak or “noisy” signal.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; or replace the
sensor.
• Change sensor type.
• Consider increasing perfusion using heat.
• If there is excessive light, cover the sensor with opaque
material.
signal was present but has
2
PROBLEM: A sensor error indicating a bad sensor has
occurred.
CAUSE:
53
• The sensor or cable may be defective, or the cabling
may be improperly connected.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor/cable is applied securely and properly, and
replace it if necessary.
• Disconnect and reconnect the sensor.
54
Using the Menu System
Introduction
The PRO Monitor is equipped with a liquid crystal display
(25) and a rotor (21). Used together, these allow the
operator to view and edit most of the Monitor’s parameters
and functions. When the Monitor is in use, a number of
option buttons appear on the liquid crystal display (LCD).
The model of the Monitor determines which menu option
buttons appear on the LCD. The number of buttons and the
specific options depend on the menu level. The rotor
provides the means of choosing menu options and changing
monitor settings.
Liquid Crystal Display
The LCD is divided into three areas, each of which has a
distinct function.
Menu Area
This area displays the menu buttons that are available for
selection. Normal text in the menu area appears dark on a
light background, while the text of selected buttons appears
light on a dark background.
Area 2
This area displays BP and SpO
messages. The Display mode menu is used to select the data
to be displayed.
Area 3
This area displays the time, the time lapsed since the last
Auto BP determination (if in Auto BP mode), the battery icon
(if operating on battery power, the time and battery icon
toggle), and the BP and printer modes.
Note: In cold ambient temperatures (below 50 °F / 10 °C),
updates on the LCD can be delayed by approximately 1
second. This delay on the LCD does not affect the
performance of the Monitor.
data and error and warning
2
55
Normal Mode Menu
56
Using the Menu System
Clinician Mode Menu
57
Rotor
Rotating the rotor causes option buttons to be highlighted
(light text on a dark background). Turning the rotor produces
a click. Turning it clockwise moves the highlighting clockwise
over the available buttons, while turning it counterclockwise
reverses the direction of the highlighting. Pressing the rotor
selects the highlighted button and produces an audible tone.
Some menus (e.g., Alarms) contain values that can be
changed by the operator. After the value is highlighted, the
user selects it by pressing the rotor. Turning the rotor
clockwise will cause the value to increase, and turning the
rotor counterclockwise will cause the value to decrease.
Pressing the rotor again will confirm the changed value.
Menu Tree
The menu tree on the previous page shows all possible
choices available within the menu structure, from the top
level downward.
Main Menu
This menu is the top level menu. It is displayed when the
Monitor is first switched on and after the rotor has been
inactive for 2 minutes, unless the Monitor is in sleep mode
(Pwr Sav).
58
Using the Menu System
Vitals Button (UK: All Obs)
Selection of this button initiates a BP determination while
allowing SpO
be monitored and recorded (depending on Monitor model).
When the Vitals determination is complete, a single “warble”
sounds and all patient data are displayed on the LEDs and
held for 2 minutes. The LCD shows*:
After this 2 minute static display, BP and temp values are
displayed for the period of time pre-selected by the user for
manual values (refer to “Retain Values” in the Using the Menu System section for information on how to change this
setting). After the 2 minute static display, the SpO
parameter returns to active monitoring and dynamically
displays the SpO
and predictive temperature determinations to
2
Note: If the printer is in “Auto”
mode, the Print button does
not appear as an option.
2
values.
2
Notes
• If the Monitor is performing a Vitals determination, the
Vitals button cannot be selected.
• If a BP determination is in progress, the Vitals button
cannot be selected.
• A Vitals determination is canceled if the BP
determination is canceled.
• A Vitals determination can be canceled by pressing
either the AUTO BP or START/STOP key.
• During the 2 minute freeze period, SpO
monitoring and
2
alarms are suspended.
Clear
Selection of this button halts measurements and returns the
user to the Main menu.
*The model of the Monitor determines which menu option buttons appear on the LCD.
59
Note: If the SpO2 plethysmograph is displayed on the LCD,
the waveform pauses for 2 minutes or until the Clear button
is selected. SpO
values are also retained the same manner
2
as the BP and Temperature values.
Print
Selection of this button causes the current data to be printed.
Notes
• The Print button appears only when Print is set to
Manual mode.
• If the printer is in Auto print mode, the data will be
printed automatically.
More... Button
Selection of this button displays the More... menu. The
More... menu has six options (depending on model of
Monitor), most of which have submenus. For this reason,
instructions for the More... button are in a separate section.
Set BP Button (UK: BP Mode)
Selection of this button displays the Auto, Stat, and Manual
BP menu.
Auto
Selection of this option starts an Auto BP determination.
When Auto Mode is selected, a number at the right of the
Auto button indicates the time interval between each
reading. To change the time interval, choose the box around
the number and turn the rotor until the desired interval is
reached. The interval can be set between 1 and 120 minutes
(1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90, and 120 minutes).
Press the rotor to confirm the setting. After power-off, the
60
Using the Menu System
operating mode returns to the default setting of Manual. The
default setting of Manual can be overridden to return to the
previous user-selected setting (Auto or Manual) by selecting
Set BP under the Service menu. To cancel an Auto BP
determination, select the START/STOP key. To cancel an
Auto BP mode, select the Manual button in the Set BP (UK:
BP Mode) menu.
Manual
Selection of this option starts a Manual BP determination.
After power-off, the operating mode returns to the default
setting of Manual. The default setting of Manual can be
overridden to return to the previous user-selected setting
(Auto or Manual) by selecting Set BP under the Service menu.
Tgt Pressure
Selection of this option allows the user to set the BP target
inflation pressure. The initial target pressure can be set
between 100 and 250 mmHg in 5 mmHg increments. The
factory default is 150 mmHg for adults and 110 for neonates.
(This is indicated by “AUTO” at the end of adjustable range.)
When using a neonate blood pressure cuff, if the target
pressure is set to greater than 140 mmHg under the Set BP
or Clinical menu, the Monitor automatically defaults to a
target pressure of 110 mmHg. If the target pressure is set
between 100 and 140 mmHg, then that setting is the target
pressure that will be used. When the target pressure is
changed, the next determination will use the new target
inflation value if no systolic is available. Initial target pressure
is restored to the factory default setting after power-off. The
target pressure can be adjusted permanently in the Clinician
menu of the Service mode (refer to “Press” in the “Using the
Menu System” section).
Stat
Selection of this option allows the user to start Stat
determinations. When Stat is selected, blood pressure is
determined as many times as possible in 5 minutes.
Note: Alarm limits are disabled while in Stat mode.
Main
Selection of this button returns the user to the Main menu.
61
Alarms Button
Selection of this button displays the Alarms menu. This menu
is used to adjust the violation limits for BP, Pulse Rate, and
SpO
. The values and ranges for these parameters are not
2
stored when the Monitor is turned off. The user may edit the
limits, but they are restored to the default values each time
the Monitor is switched on. To permanently change the
alarm limits, refer to “Alarms” under “Service Button” in the
“Using the Menu System” section.
Alarm Limits Table
ParameterRangeDefault
Systolic High35 - 245180
Systolic Low30 - 24030
Diastolic High15 - 195130
Diastolic Low10 - 19015
MAP High20 - 215140
MAP Low15 - 21050
BPM High
(Masimo)
BPM Low (Masimo)30 - 23040
BPM High (Nellcor)35 - 250150
BPM Low (Nellcor)30 - 24550
SpO2 High21 - 100Off
SpO2 Low20 - 9990
35 - 235160
Volume
Selection of this button displays the alarm volume submenu.
The volume range is from 1 to 10, with 10 being the loudest.
The alarm volume is stored when the Monitor is turned off
and restored to the user’s preference each time the Monitor
is switched on. Selection of the Check button allows the
current volume setting to be heard. Selection of the Main
button returns the user to the Main menu.
62
Using the Menu System
Auto
Selection of this button updates the alarm limits relative to
the current parameter values. Pressing this button cancels
any limit violation alarm that is invalided by this automatic
limit change. The alarm limits are updated as follows:
Notes
• In no case will the updated alarm limits be set beyond
the valid limits in the Alarm Limits table.
• If no values are available, the limits will remain
unchanged.
Main
Selection of this button returns the user to the Main menu.
Trend Button
Selection of this button displays the Trend mode menu.
63
Display
Selection of this button allows the operator to view the trend
data.
Note: If the trend data have been lost (e.g., if the clock
settings have been changed), the message “Trend Empty”
will appear instead of the Newer, Older, and Print page
buttons.
Newer and Older.
forward and backward through the recorded data. If no
information is available, these buttons will not appear.
Print page.
information to be printed. If no information is available, this
button will not appear.
Main.
menu.
Clear
Selection of this button produces an advisory that the trend
will be lost. Choosing Yes will erase the trend memory.
Choosing No will retain the trend memory. This button
disappears from the menu while printing and when Trend is
empty.
Print All
Selection of this button prints all the historical data available.
When selected, this button temporarily changes to Cancel
until the history has completed printing. Once printing is
complete, the Cancel button returns back to the Print All
button. This button disappears from the menu when Trend is
empty.
Selection of this button causes the displayed
Selection of this button returns the user to the Main
These buttons may be used to move
64
Using the Menu System
Main
Selection of this button returns the user to the Main menu.
Print Button
Selection of this button displays the Print menu.
Auto/Man
Pressing this button toggles between Automatic and Manual
Printing modes. The current mode is displayed on Area 3 of
the LCD. The Automatic mode prints the readings after each
determination. The Manual mode, which is the factory
default mode, requires the user to press the Now button to
print the readings.
Now
Selection of this button causes the last readings of the
available parameters to be printed. If no readings are
available, the message “No reading” is printed for that
parameter. An error message appears if there is no paper in
the printer.
History
Selection of this button causes the entire contents of the
trend memory to be printed. When selected, this button
temporarily changes to Cancel until the history has
completed printing.
Main
Selection of this button returns the user to the Main menu.
65
More... Menu
This menu is used to set the various operating modes of the
Monitor.
SpO
Button (Models 300V2 and 400V2)
2
Selection of this button displays the SpO
which is used to set the SpO
pulse tone volume and SpO2
2
Average (Masimo ONLY).
mode menu,
2
Average
For PRO Monitors with Masimo SET technology, the Average
value is used in the calculation of SpO
values. Average can
2
be set to 4, 6, 8, 10, 12 ,14, or 16 seconds.
Volume
The pulse tone volume can be set in the range of Off to 9.
The value Off should be selected if no pulse tone is desired.
The volume setting is stored when the Monitor is turned off
and is restored to the user’s preference each time the
Monitor is switched on.
Info
With Nellcor SpO
technology installed, the info button
2
displays the current settings for Response Mode and
SatSeconds™.
66
Using the Menu System
Response Mode (Nellcor Only)
Response mode allows the user to specify the averaging
technique to optimize measurements in the presence of
various patient movement. Choose Mode 1 (normal
Response) when patients are active as in exercise protocols.
Choose Mode 2 (Fast response; default setting) for the
general patient population.
SatSeconds™ (Nellcor Only)
The SatSeconds™ technique limit-controls the time that the %
SpO2 level may fall outside the alarm before an audible
alarm sounds. Choose either 0, 10, 25, 50, or 100 seconds.
If 0 is chosen this feature is disabled.
The SatSeconds™ “Safety Net” is for patients with saturation
levels having frequent excursions below the limit, but not
staying below the limit long enough for the SatSeconsd time
setting to be reached. When 3 or more limit violations occur
within 60 seconds, an alarm sounds even if the the
SatSeconds time setting has not been reached.
With Masimo SET technology installed, the Info button
displays the current settings for the Sensitivity and FastSAT
features. Sensitivity can be set to Normal (the factory
default) or Max. FastSAT can be set to On or Off (the factory
default).
Main
Selection of this button returns the user to the Main menu.
Config Button
Selection of this button displays the Config mode menu,
which allows the Power Save mode and time to be adjusted.
67
Pwr Sav (Sleep Mode)
Selection of this button allows the operator to specify the
time, in minutes, that elapses before the Monitor goes into
“sleep” mode (LEDs blanked and LCD displaying values from
LEDs). Sleep mode is available only if the Monitor is
operating from its battery. Sleep mode conserves power
while the Monitor is not in use. Once the Monitor is in Sleep
mode, the user can return it to normal operation by touching
any button or the rotor.
The Monitor enters Sleep mode only if the following are true:
• No alarm is active
•SpO
is not actively reporting patient statistics
2
• The keys and rotor have not been used for the preset
time
• The Monitor is running from its battery
• No determinations are in progress
• The Monitor has been running from the battery for the
entire preset time
The Monitor awakens from Sleep mode if any of the
following occur:
• The rotor is turned or pressed
• Any of the keys are pressed
• An alarm condition is issued
• The battery supply level becomes discharged to a critical
level
• A mains or suitable DC supply is connected
• An Auto BP or Temp determination starts
•An SpO
signal is detected
2
• A BP determination is started through the host comm
68
Using the Menu System
Time
Selection of this button allows the operator to change the
internal time and date of the Monitor. The clock, which is
maintained by an internal battery after power down, uses 24hour format. The date is in the British format of dd/mmm/
yyyy; however, to avoid confusion the month number has
been substituted with a three-letter abbreviation. Leap years
are calculated automatically.
Accept
user that the trend will be lost when the clock settings are
changed. Choosing Yes will cause the Monitor to accept the
new clock settings and erase the trend memory. Choosing
No will cause the Monitor to retain the existing clock settings
and the trend memory. Either choice returns the user to the
Main menu.
Main
menu.
Rotor
Selection of this button displays a panel for setting the
volume of the beep that sounds when the rotor is turned.
The range of adjustment is Off (default) to 9, and the setting
is retained when the Monitor is turned off.
. Selection of this button produces an advisory to the
. Selection of this button returns the user to the Main
69
Main
Selection of this button returns the user to the Main menu.
Temp
Selection of this button displays the temperature submenu,
which allows the user to choose the temperature label.
When C (Celsius) is selected, the °C indicator lights. When F
(Fahrenheit) is selected, the °F indicator lights.
C or F. Selection of this button toggles the temperature
display between Celsius and Fahrenheit.
Predictive. Selection of this button toggles between the two
modes of temperature measurement. TURBO
TEMP
predictive measurement intiates when the probe is removed
from the holster and terminates when either a value is
determined, the probe is re-holstered, or after a 2-minute
time span when the software defaults to “Monitor” mode.
Monitor. Monitor mode is most commonly used for axillary
temperature determinations. In monitor mode, the display is
updated continually as the patient’s temperature rises or falls.
Display Button
Selection of this button displays the Display mode menu.
This menu is used to specify whether Area 2 of the LCD will
display SpO
or BP data. If neither SpO2 nor 3 NIBP is
2
selected, Area 2 of the LCD will remain blank except for the
pulse amplitude bar (if SpO
data are available) and any
2
error or warning messages that may appear. The Display
mode setting is maintained when the Monitor is switched off
and on.
70
Using the Menu System
SpO2 Pleth
When this option is checked and SpO
the plethysmograph waveform and the pulse amplitude bar
will be displayed.
3 NIBP
When this option is checked, the last 3 NIBP readings will be
displayed. If SpO
will also be displayed.
Main
Selection of this button returns the user to the Main menu.
Service Button
Selection of this button displays a keypad that allows the
clinician to access some parts of the Service mode menu. To
access the Clinician menu, use the rotor to select the
numbers 1, 2, 3, 4 sequentially.
Notes
•SpO
• Service modes that affect the calibration or alignment of
is automatically disabled when entering Service
2
mode.
the instrument are not available to the user. These
modes are described in the Service Manual.
data is available, the pulse amplitude bar
2
data are available,
2
71
Clinician Menu
Press
. Selection of this button displays a panel for setting the
default BP target inflation pressure. Adjusting the default
target pressure will automatically update the current inflation
target pressure and will be used for the next reading. The
range of adjustment is 100 mmHg to 180 mmHg, and the
setting is retained when the Monitor is turned off.
The initial target pressure can be set between 100 and 180
mmHg in 5 mmHg increments. The factory default is 150
mmHg for adults and 110 for neonates. This is indicated by
the “AUTO” label at the end of the adjustable range. When
the target pressure is changed, the next determination will
use the new target inflation value if no systolic is available.
When adjusted under the Clinician menu of the Service
mode, the target pressure is adjusted permanently.
Retain Values. The Retain Values setting indicates the
amount of time that the manual BP and Temp measurement
values stay on screen. This is also the setting for the amount
of time that Spot Mode values are displayed after a freeze
period.
OK. Selection of this button returns the user to the Clinician
menu.
Info
. Selection of this button causes the most recent
calibration date of the NIBP system to be displayed.
Selection of OK returns the user to the Clinician menu.
72
Using the Menu System
OK. Selection of this button returns the user to the Clinician
menu.
More...
which allows the user to permanently change default mode
settings.
Trend. Selection of this button displays the message:
Automatically clear trend on power-up? Selection of Yes
overrides the default setting by clearing all trends on powerup and returns the Monitor to the More... menu. Selection of
No retains the default setting by saving all trends after poweroff and returns the Monitor to the More... menu. Selection of
Cancel returns the user to the More... menu.
Print. Selection of this button displays the message: Restore
Print mode on power-up? Selection of Yes restores the Print
mode to the default setting (previous user-selected mode)
after power-off and returns the Monitor to the More... menu.
Selection of No restores the Print mode to the Manual mode
after power-off and returns the Monitor to the More... menu.
Selection of Cancel returns the Monitor to the More... menu.
. Selection of this button displays the More... menu,
SpO
. Selection of this button displays the message: Enter
2
SpO
Configuration Mode? Selection of No returns the
2
Monitor to the More... menu. Selection of Yes brings up the
SpO
menu.
2
73
For monitors with Nellcor SpO2 technology, the SpO2
configuration menu includes Response Mode and
SatSeconds™. Response Mode can be set to “1” or “2”. The
factory default is “2”.
SatSeconds™ technique limit-controls the time that the %
SpO
level may fall outside the alarm before an audible
2
alarm sounds. Choose either 0, 10, 25, 50, or 100 seconds.
If 0 is chosen this feature is disabled
For monitors with Masimo SpO
technology, sensitivity can
2
be set to Normal (the factory default) or Max. FastSAT can
be set to On or Off.
Set BP. Selection of this button displays the message: Restore
BP mode on power-up? Selection of Yes restores the BP
mode to previous user-selected mode after power-off and
returns the Monitor to the More... menu. Selection of No
restores the BP mode to the default setting of Manual after
power-off and returns the Monitor to the More... menu.
Selection of Cancel returns the Monitor to the More... menu.
Alarms. Selection of this button displays the message: Enter
alarm configuration mode? Selection of No returns the
Monitor to the More... menu. Selection of Yes brings up the
Alarms menu. Selection of Reset changes all alarm limits
back to the factory defaults and returns the Monitor to the
More... menu. Selection of Save permanently saves the userselected alarm limits and returns the Monitor to the More...
menu. Selection of Cancel returns the Monitor to the More...
menu.
Main. Selection of this button returns the user to the Main
menu.
74
Using the Menu System
Silence. Selection of this button will cause all alarms except
the FAILSAFE alarm to be muted. A confirmation menu will
appear in Area 2 of the LED. Selection of either Yes or No
returns the user to the Clinician mode menu. If silence is
confirmed, the Alarm Silence button (26) illuminates and
alarms are permanently muted. If silence is not confirmed,
the alarm will be audible.
Caution: Alarms will be muted until either the Monitor is
switched off and on again or the Alarm Silence button (26)
is pressed.
Main
. Selection of this button returns the user to the Main
menu.
Error and Warning Messages
The error panel appears in Area 2 of the LCD and indicates
the error and its code, if it has one. In this example, a limit
violation alarm (which has no error code) has occurred. A list
of alarm error messages and their codes is in Appendix B.
Alarm conditions are addressed in two ways: the Alarms
button and OK button.
Alarms Button
Selection of this button takes the user to the Alarms menu,
where the alarm limits can be adjusted. This button is
available only when a parameter alarm limit has been
violated.
75
OK Button
Selection of this button acknowledges the error. The Monitor
clears the identified error and then returns the user to the
Main menu.
76
Appendix A
Technical Specifications
BP
Cuff Pressure Range0 to 290 mmHg (adult)
(Normal operating range)0 to 140 mmHg (neonate)
± 0.2° F (when tested in a
calibrated liquid bath; meets
ASTM E1112, Table 1, in range
specified)
Predictive mode accuracy± 1.0° F
± 0.6° C
Determination timeapproximately 10 seconds
Use only IVAC probes and probe covers. The size, shape, and
thermal characteristics of the probe covers can affect the
performance of the instrument. Inaccurate readings or retention
problems may occur unless IVAC probes and probe covers are used.
Refer to Appendix D for reorder codes.
IVAC® Patents
U.S. D300,728; D300,909
NELLCOR SpO
2
Measurement Range
SpO
2
1 to 100%
Pulse Rate20 to 250 beats/min
Perfusion Range0.03 to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adults*70 to 100% ±2 digits
Without Motion - Neonate* 70 to 100% ±3 digits
With Motion - Adults/Neo** 70 to 100% ±3 digits
Low Perfusion70 to 100% ±2 digits
78
Appendix A
Pulse Rate
Without Motion20 to 250 beats/min ±3 digits
With Motionnormal physiologic range
55 to 125 beats/min ±5 digits
Low Perfusion20 to 250 beats/min ±3 digits
*Adult specifications are shown for OXIMAX MAX-A and MAX-N sensors.
Neonate specifications are shown for OXIMAX MAX-N. Saturation accuracy
will vary by sensor type.
**Applicability: OXIMAX MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N
sensors.
Default Settings
SpO
(%)HIGH: 100
2
(%)LOW: 90
SpO
2
Response Mode2 (for Mode 2: Fast Response)
Sat Seconds0
Audible IndicatorPitch changes continuously
with saturation; volume from 0
(off) to 9
WaveformsPulse plethysmograph
waveform on LCD gain
compensated
Sensor Connect/Monitor detect attachment
Disconnect From Patientor disconnection of sensor
from patient within 15 s
Pulse DetectionMonitor will detect pulse or
enter no signal state within 15
s of being attached to patient
Loss of PulseMonitor will detect loss of
pulse from patient and enter
no signal state within 10 s
Pulse Rate25 to 240 beats/min
Perfusion Range0.02 to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adult/Ped*70 to 100% ±2 digits
Without Motion - Neonate*70 to 100% ±3 digits
With Motion - Adult/Ped/Neo**† 70 to 100% ±3 digits
Low Perfusion‡70 to 100% ±2 digits
0 to 69% unspecified
Pulse Rate
Without Motion25 to 240 beats/min ±3 digits
With Motionnormal physiologic range
25 to 240 beats/min ±5 digits
*The Masimo SET® SpO2 parameter with LNOP-Adt sensors has been
validated for no motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies in the range of 70-100% SpO
a laboratory co-oximeter and ECG monitor. This variation equals plus or
minus one standard deviation. Plus or minus one standard deviation
encompasses 68% of the population.
**The Masimo SET
validated for motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies while performing rubbing and tapping
motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non repetitive
motion before 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia
studies in the range of 70-100% SpO2 against a laboratory co-oximeter and
ECG monitor. This variation equals plus or minus one standard deviation.
Plus or minus one standard deviation encompasses 68% of the population.
†The Masimo SET
validated for neonatal motion accuracy in human blood studies on neonates
while moving the neonate’s foot at 2 to 4 cm against a laboratory cooximeter and ECG monitor. This validation equals plus or minus, one
®
SpO2 parameter with LNOP-Adt sensors has been
®
SpO2 parameter with LNOP-Neo Pt sensors has been
against
2
80
Appendix A
standard deviation. Plus or minus one standard deviation encompasses 68%
of the population.
‡The Masimo SET
accuracy in bench top testing against a Biotek Index 2 simulator and
Masimo’s simulator with signal strengths of greater than 0.02% and a %
transmission of greater than 5% for saturations ranging from 70 to 100%.
This variation equals plus or minus, one standard deviation. Plus or minus
one standard deviation encompasses 68% of the population.
SpO2 parameter has been validated for low perfusion
SpO2 Range 70% - 100%
)1%
2
Low Perfusion Performance
>0.02% Pulse AmplitudeSaturation (% SpO
) ±2 digits
2
and % Transmission >5%Pulse Rate ±3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The
level of increase is approximately equal to the amount of
carboxyhemoglobin present. Dyes, or any substance
containing dyes, that change usual arterial pigmentation may
cause erroneous readings.
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644;
5,769,785; 6,002,952; 6,036,642; 6,067,462; 6,206,830;
6,157,850, and international equivalents.
Mechanical
DimensionsHeight: 9.8 in (25.0 cm)
Width: 9.8 in (24.8 cm)
Depth: 6.9 in (17.5 cm)
Weight, Including Battery7.8 lb (3.5 kg)
MountingsSelf-supporting on rubber feet
or pole mountable
PortabilityCarried by recessed handle or
pole mounted
Classification InformationMode of operation:
continuous
Degree of protection against
harmful ingress of water: Dripproof IPX1
Power Requirements
AC input voltage: 100-240
VAC, 50 / 60 Hz (nominal)
90 ~ 253 VAC, 47 ~ 63 Hz
(range), 50VA.
Protection against electrical
shock: Class 1
DC input voltage: 24 VDC
(nominal), 12-30 VDC, 36VA,
82
Appendix A
supplied from a source
conforming to IEC 601-1.
AC input is protected by two
internal fuses, replaceable by
qualified service personnel
only. DC input line is
protected by an internal autoresetting fuse.
Battery: 12 volt, 2.3 amphours protected by internal
auto-resetting fuse.
Minimum operation time: 2
hrs (5 min cycle with adult cuff
at 25 °C with power save
mode enabled) from full
charge.
Time for full recharge: 1 hr 50
min from full discharge when
the Monitor is switched off
and 8 hrs when the Monitor is
switched on.
Environmental
Operating Temperature+ 5 °C to + 40 °C
(+ 41 °F to + 104 °F)
Operating Atmospheric
Pressure700 hPa to 1060 hPa
Storage Temperature– 20 °C to + 50 °C
(– 4 °F to + 122 °F)
Storage/Transportation
Atmospheric Pressure500 hPa to 1060 hPa
Humidity Range0% to 95% noncondensing
Radio FrequencyComplies with IEC
Publication 601-1-2 (April
1993) Medical Electrical
Equipment, Electromagnetic
83
IPX1
Compatibility Requirements
and Tests and CISPR 11
(Group 1, Class A) for radiated
and conducted emissions.
The DINAMAP® PRO Monitor is
protected against vertically falling drops
of water and conforms to IEC-529
standard at level of IPX1. Vertically falling
drops of water shall have no harmful.
84
Appendix B
Alarm Codes
All alarm indications are accompanied by an audible signal
unless Alarm Silence is selected.
A microprocessor system failure will generate a high-pitched
audible alarm regardless of the setting of the Alarm Silence
switch.
There are three categories of alarms: patient alarms, system
alarms, and failsafe alarm.
Patient Alarms
Patient alarms include those alarms issued when the patient’s
systolic pressure, diastolic pressure, pulse rate, or oxygen
saturation is outside the set limits. Whenever one of these
conditions occurs, the associated display (SYSTOLIC, MAP,
DIASTOLIC, PULSE, or SpO
reading and an audible alarm will be issued.
Pressing the Alarm Silence switch (causing the integral LED
to be lit) silences the audible alarm for 2 minutes, but the
alarm display reading and SILENCE LED indicator will
continue to flash at the same rate.
) will flash the most recent
2
System Alarms
System alarms alert the operator to certain abnormal
conditions or internal system failures. Pressing the rotor
cancels the alarm information box which is displayed on the
LCD. Codes for different procedural and system alarms are
on the next page.
Failsafe Alarm
The failsafe alarm, which is the most powerful alarm of the
PRO Monitor, indicates a serious failure of the Monitor. This
alarm occurs immediately upon any failure of a self-test and
indicates system failure. When the failsafe alarm occurs, the
Monitor disables all features to ensure patient safety.
85
Hierarchy of Alarms
Alarms in the DINAMAP® PRO Monitor are in three priority
levels. They are:
Alarm
Failsafe1
Patient and system2 (High priority alarm)
Low battery3
The Priority 1 alarm (i.e., Failsafe) will override any other
alarm. Priority 2 alarms will override only the low battery
alarm. The low battery alarm will not override any other
alarm.
Priority Level
86
Probable Cause
Effect of
Clear via
SelectKnob
Switch
Effect of
Alarm Silence
Unable to make an NIBP
determination due to
insufficient signal
Clear
2 minutes
silence
Appendix B
minutes. Motion artifact
silence
Inflation time > 40 seconds
Clear
2 minutes
One cuff pressure for > 1
Clear
2 minutes
Determination time > 2
minute. Motion artifact
Clear
2 minutes
silence
or air leak detected
silence
ClearOverpressure detected
2 minutes
silence
and Volume
Audible Tone
Procedural and Error Alarm Codes
LCD Description
LED
Alarm
Display
Code
High priority
N99 -
No
High priority
alarm. Volume
adjustable
N55 -
NIBP FAILED
No
change
N99
87
High priority
alarm. Volume
adjustable
N44 -
TIMEOUT: PRESS
No
change
N55
High priority
alarm. Volume
adjustable
N33 -
TIMEOUT: TOTAL
No
change
N44
High priority
adjustable
alarm. Volume
N00 -
No
N00
alarm. Volume
TIMEOUT: INFLT
change
N33
adjustable
OVER PRESSURE
change
signal.
2
Probable Cause
Effect of
Clear via
SelectKnob
Silence Switch
Effect of Alarm
and Volume
Audible Tone
Procedural and Error Alarm Codes (cont.)
High priority
No or very low SpO
Clear
2 minutes
alarm. Volume
Check or reposition sensor
silence
adjustable
sensor not connected.
2
No sensor code detected.
SpO
Clear
2 minutes
silence
High priority
alarm. Volume
Sensor failure
adjustable
High priority
signal weak or noisy.
2
SpO
2 minutes
Sensor failure
Clear
silence
alarm. Volume
adjustable
High priority
Sensor off patient
ClearSpO
2 minutes
alarm. Volume
2
silence
adjustable
High priority
Temperature probe not
Clear
2 minutes
alarm. Volume
connected or inoperable
silence
adjustable
Predictive temperature
2 minutes
High priority
determination > 60 sec or
attempting axillary temp
Clear
silence
alarm. Volume
adjustable
Predictive temperature exceeds
upper range
Clear
2 minutes
silence
High priority
alarm. Volume
adjustable
ClearBroken temperature probe
2 minutes
High priority
alarm. Volume
silence
adjustable
NO SIGNAL
LCD Description
P55 -
SpO
LED
Display
No
change
Code
Alarm
P55
SENSOR
2
P00 -
No
P00
2
SpO
change
PLACEMENT?
2
SpO
No
No
(Masimo Only)
change
Code
SENSOR OFF?
2
SpO
Values
zeroed
No
Code
E33 -
No
E33
TEMP: FAIL
change
E00 -
No
E00
TEMP: FAIL
change
BlankHIGH TEMP
No
Code
TEMP PROBE
BROKEN
No
change
No
Code
88
Probable Cause
Effect of
Clear via
SelectKnob
Switch
Effect of
Alarm Silence
Replace or recharge battery.
From onset of alarm. 5 NIBP
measurements available.
Beep rate increases linearly
as battery discharges
No effect
2 minutes
silence
Appendix B
Replace or recharge battery.
NIBP measurement disabled
Paper ran out or printer
door open
NIBP algorithm returned
value outside specified
accuracy range
No effectNo effect
Clear
2 minutes
silence
Clear
2 minutes
silence
No effectNo effectInternal system fault
Procedural and Error Alarm Codes
and Volume
Audible Tone
LCD Description
LED
Display
Code
Alarm
3 beeps every
10 seconds,
adjustable
LOW BATTERY,
Flashing
No
change
No
Code
volume
battery icon
Steady tone,
LOW BATTERY -
No
89
maximum
volume
SYSTEM
DISABLED
Blank
Code
High priority
alarm. Volume
adjustable
High priority
PRINTER -
NO PAPER
NIBP RANGE
No
change
Values
No
No
Code
alarm. Volume
adjustable
Steady tone,
ERROR
Error code,
posted
Code
Other:
maximum
Blank
N, P,
volume
description
E, I, S
90
Appendix C
Principles of Noninvasive Blood Pressure
Determination
The oscillometric method of determining NIBP is
accomplished by a sensitive transducer which measures cuff
pressure and minute pressure oscillations within the cuff. A
single determination (in normal mode) is initiated before
taking repeated determinations in auto or stat mode. As a
determination is taken, the algorithm stores the pattern of the
patient's oscillation size as a function of the pressure steps. In
auto and stat mode, as few as six pressure steps may be
necessary to complete the determination process. When
employing fewer pressure steps, the system uses the stored
information from the previous blood pressure determination
to decide the best pressure steps to take. The algorithm
measures the consistency of pulse size to tell if the
oscillations taken at a step are good and if more steps are
needed.
The first determination initially pumps up to a target pressure
of about 150 mmHg for adult patients depending on initial
target pressure preset. After inflating the cuff, the NIBP
parameter begins to deflate, the oscillations versus cuff
pressure is measured, and, finally, systolic pressure, mean
pressure, and diastolic pressure are calculated and the screen
is updated.
During an NIBP determination, the parameter deflates the
cuff one step each time it detects two pulsations of relatively
equal amplitude. The time between deflation steps depends
on the frequency of these matched pulses (pulse rate of the
patient). However, if the monitor is unable to find any pulse
within several seconds, it will deflate to the next step. The
process of finding two matched pulses at each step provides
artifact rejection due to patient movement and greatly
enhances the accuracy of the monitor. The figure shows a full
determination sequence for an adult patient. In stat mode,
the artifact rejection technique of two matching pulses at
each step is disabled and some steps may have only one
pulse.
91
Full NIBP Determination Sequence for Adult
At each step the microprocessor stores cuff pressure, the
matched pulse amplitude, and the time between successive
pulses. The stepped deflation and matched pulse detection
continues until diastolic pressure is determined or total cuff
pressure falls below 8 mmHg. The parameter then deflates
the cuff (to zero detected pressure), analyzes the stored data,
and updates the screen.
The operating cycle is composed of four parts: inflation time,
deflation time, evaluation time, and wait time. Wait time,
which varies from mode to mode, is affected by the cycle
time (auto mode) or operator intervention (manual mode).
The figure shows the basic operating cycle for an NIBP
determination.
92
Appendix C
NIBP - Auto Mode
Systolic Search
If systolic pressure is not found, the NIBP parameter can
search at cuff pressures higher than the initial target pressure.
The parameter will inflate the cuff above the initial target
pressure to get better data in the systolic region. The
maximum pressure allowed in systolic search is limited by
the normal range for cuff pressures. In any operating mode, if
a patient's systolic pressure exceeds the inflation pressure,
the parameter will begin normal deflation sequence, detect
the absence of a systolic value, stop deflation, reinflate to a
higher (than initial) inflation pressure (290 mmHg maximum),
and resume the normal deflation sequence.
Do not use the auscultatory method to verify the accuracy of
the NIBP parameter. Auscultatory method (using cuff and
stethoscope) calculates the mean pressure value from
audible sounds at systolic and diastolic, but the NIBP
parameter method detects all three values.
Reverting and Accelerated Determination
When the determination data does not agree with the
previous determination, there may be a reversion to a
determination with more steps. The algorithm will try to
93
make an accelerated determination of blood pressure if it has
been 16 minutes or less since the last determination.
94
Appendix D
Reorder Codes
PROD
PRODUCT
DINAMAP PRO Monitor Operations Manual-English 2009802-001
DINAMAP PRO Monitor Service Manual2009829-001
Battery, 12V Lead Acid2010422-001
Printer Paper (box of 10)089100 E9050KP
DINAMAP Rolling Stand003215E9050JB
Nurse Call487208CR
NIBP:
Air Hose 12 ft Adult/Pediatric, Screw Connector107365E9050LH
Air Hose 24 ft Adult/Pediatric, Screw Connector107366E9050LJ
Air Hose 12 ft Neonatal, Quick Disconnect107368E9050LK
Air Hose 12 ft Adult/Pediatric, Quick Disconnect 88847E9050KN
CUFF Assortment Packs:
CLASSIC-CUF
®
Assortment Pack2692E2692J
Includes: 1 each: Infant, Child, Small Adult, Adult,
Large Adult, Thigh Cuff
CLASSIC-CUF
Warning: There are no user serviceable parts inside the
DINAMAP
personnel.
Warranty
All repairs on products under warranty must be performed or
approved by GE Medical Systems Information Technologies.
Unauthorized repairs will void the warranty.
Products not covered by warranty should be repaired only
by qualified electronics service personnel.
Assistance and Parts
If the product malfunctions or if assistance, service, or spare
parts are required, contact GE Medical Systems Information Technologies Technical Support.
Before contacting GE Medical Systems Information Technologies it is helpful if you can duplicate the problem
and check and confirm the operation of all accessories to
ensure that they are not the cause of the problem.
®
PRO Monitor. Refer all servicing to qualified
When calling, please have the following information at hand:
• product name and model number and complete
description of the problem
• the serial number of your Monitor
• your name and address
• a purchase order number if out-of-warranty repairs or
spare parts are required
• your GE Medical Systems Information Technologies
account number, if applicable
• the 6-digit part number for spare or replacement parts
97
Repairs
If your product requires warranty, extended warranty, or
non-warranty repair service, call GE Medical Systems
Information Technologies and a representative will assist you.
Estimates for non-warranty repairs are provided at no charge;
however, the product must be sent to GE Medical Systems
Information Technologies for an estimate.
To facilitate prompt service in cases where the product has
external chassis or case damage, please advise the
representative when you call.
The representative will record all necessary information and
will provide a Return Authorization Number. Prior to
returning any product for repair, a Return Authorization
Number must be obtained.
Packing Instructions
If you have to return goods for service, follow these
recommended packing instructions:
• Remove all hoses, cables, sensors, power cords, and
ancillary products from the Monitor before packing.
• Wherever possible use the original shipping carton and
packing materials.
• Observe the environmental conditions detailed in
Appendix A.
It is recommended that all returned goods be insured. Claims
for loss or damage to the product must be initiated by the
sender.
Service Manuals
Service Manuals containing calibration and repair
information can be ordered from GE Medical Systems
Information Technologies. These manuals also include full
schematic diagrams, assembly drawings, and spare parts lists.
Refer to Appendix D for the reorder number of the Service
Manual.
98
Appendix F
Maintenance
Cleaning the Monitor
The Monitor and accessories are to be kept clean and used
according to the instructions provided here and in the
Service Manual.
The exterior of the Monitor may be wiped clean with a soft
cloth slightly dampened with mild detergents. The Monitor
and accessories should be inspected once yearly for wear
and damage.
•Do not immerse unit.
•Do not clean with isopropyl alcohol or other solvents.
•Do not immerse hoses.
Cuff Cleaning and Disinfection
General
The cuff must be thoroughly cleaned with the specified
detergent before reuse. The additional use of household
bleach as described below provides at least intermediatelevel disinfection.
• Apply cuff hose plugs before cleaning.
• The following cleansing procedure was repeated 20
times on DURA-CUF
SOFT-CUF
performance of the cuff.
• While this procedure is adequate for cleaning/
disinfection, it may not remove all stains.
•Do not immerse hoses.
•Do not immerse cuffs without prior application of cuff
hose caps.
®
Blood Pressure Cuffs without affecting the
®
Blood Pressure Cuffs and once on
*Trademark
Materials
• Enzymatic detergent such as ENZOL* enzymatic
detergent (US) or Cidezyme* enzymatic detergent (UK)
• Distilled water
• 10% solution of household bleach (5.25% sodium
hypochlorite) in distilled water
• Soft cloths and soft-bristled brushes
99
•Spray bottles
Procedure
1. Prepare the enzymatic detergent according to the
manufacturer’s instructions and the 10% bleach solution,
in separate spray bottles.
2. Spray the detergent liberally on device. If the material is
dried on, allow the cuff to sit for 1 minute. For soil on the
soft part of the closure or the cuff itself, wipe the material
off with a soft cloth. For persistent contamination on the
soft part of the closure, use a soft-bristled brush to loosen
particles. Rinse with copious amounts of distilled water.
Repeat until no visible contamination remains. For soil on
the hook part of the closure, use a soft-bristled brush to
remove the material, and rinse with copious amounts of
distilled water. Repeat until no visible contamination
remains.
3. Spray the 10% bleach solution on the affected area until
the area is saturated. Allow the cuff to sit for 5 minutes.
4. Wipe away any excess solution and rinse the cuff again
with distilled water. Allow 2 hours for drying.
The user has the responsibility to validate any deviations
from the recommended method of cleaning and disinfection.
For additional information on infection control procedures,
contact GE Medical Systems Information Technologies
Technical Support.
Temperature Devices
Do not immerse predictive temperature probes. The probe
may be cleaned with an alcohol solution. Use a cloth or
sponge—just damp, not wet—and avoid getting any liquid
into the interior of the probe.
SpO
Sensors
2
Adhesive sensors are sterile and for single use only. Reusable
sensors should be cleaned before reuse with a 70% alcohol
solution. Do not immerse the sensor completely in water,
solvents, or cleaning solutions (because the connector is not
waterproof). Do not sterilize the sensor by irradiation, steam,
or ethylene oxide. If disposable sensors or their packaging
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.