GE DASH 3000 V5, DASH 4000 V5 Quick Reference Manual

BEDSIDE MONITORING
Dash 3000/4000 V5
TAB TRIM
COVER & NON-TAB TRIM
Quick Reference Guide
© 2004 General Electric Company PN: 0304-CS-QRGD Rev. A 2018408-010
COVER & NON-TAB TRIM
TAB TRIM
BEDSIDE MONITORING
Dash 3000/4000 V5
TAB TRIM
COVER & NON-TAB TRIM
Quick Reference Guide
COVER & NON-TAB TRIM
TAB TRIM
Components
Display
Date
Time
Unit Name
Bed Number
Patient Name
Parameter
Window
More Menus
Waveforms
Time / Date
ECG
Parameter
Window
Unit Name Bed Number
TAB TRIM
COVER & NON-TAB TRIM
Patient Name
Waveforms
More Menus
Basic Use/
Admit/Discharge
Parameter
Windows
COVER & NON-TAB TRIM
TAB TRIM
Basic Use/
Admit/Discharge
The Basics
Direct Action Keys
Operations
Trim Knob: The Trim Knob
is found on the monitor or remote control and is used for highlighting and selecting menu options.
Turn the Trim Knob to
highlight desired menu options.
Once you have highlighted
the menu option, press to select.
TAB TRIM
COVER & NON-TAB TRIM
POWER:
Tur ns
Display
On or Off
GRAPH GO/STOP:
Starts and Stops
Manual Strips
and Print Window.
NBP
GO/STOP:
Starts and Stops NBP.
ZERO ALL:
Zeros
Invasive
Pressures.
SILENCE ALARM:
Silences and Pauses Alarms.
Also Serves as a Quick Admit Key
TRIM
KNOB:
Turn and
Push.
COVER & NON-TAB TRIM
TAB TRIM
Patient Cable Connectors
All patient cable connectors
are located on the side of the monitor.
Definition of Terms
Hardwire: This means that
the ECG signal is being acquired from a cable attached to the patient and connected to a bedside monitor.
Telemetry: This means that
the ECG signal is being acquired from a telemetry transmitter/transceiver. This patient is able to ambulate without being limited by a cable.
TTX: Refers to a telemetry
transmitter/transceiver.
RED:
Invasive
Pressures
BLUE:
SPO
TAB TRIM
COVER & NON-TAB TRIM
BROWN:
2
CO / Temp
YELLOW:
CO
2
BLACK:
NBP
GREEN:
ECG
COVER & NON-TAB TRIM
TAB TRIM
Silence Alarm
Pressing the Silence Alarm key
once will silence an active alarm for one minute. The message SILENCED will appear on the display. Any new alarm at an equal or greater priority will sound.
NOTE: If Crisis Alarm Breakthrough is set in defaults, all crisis alarms will break through Alarm Silence and Alarm Pause.
Pressing the Silence Alarm key
twice if an alarm is sounding will start an ALARM PAUSE. The length of pause will vary depending on the monitor’s mode. The message ALARM PAUSE will appear on the display.
NOTE: Alarm Pause Lengths:
— Adult ICU Mode: 5 minutes.
— Neonatal ICU Mode: 3 minutes.
— Operating Room Mode: 5 minutes. 15 minutes,
Alarm Paused (permanent pause).
Alarms will reactivate if the
Silence Alarm key is pressed again.
An Alarm Pause will
immediately be activated if the Silence Alarm key is pushed in the absence of an alarm.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Monitor Applications
The bedside monitor can be used in four different ways depending on hospital need. Differences between each application will be apparent in the monitors admit menu.
Standard
This application uses a monitor mounted in a room. It has only hardwire capability and does not accommodate telemetry.
To Admit a Patient
Select MORE MENUS.
Select ADMIT MENU.
Select ADMIT PATIENT.
NOTE: The Admit key, on the front of the monitor, can also be used to admit the monitor.
To Discharge a Patient
Remove all ECG leads from
the patient.
Select MORE MENUS.
Select ADMIT MENU.
Select DISCHARGE
PATIENT.
Turn the Trim Knob to move
the cursor in front of Discharge and press to select.
A message DISCHARGED
and ALL ALARMS OFF will appear on the display when the monitor is in a discharged mode.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
How to Enter Demographic
Information:
Select MORE MENUS.
Select ADMIT MENU.
Select CHANGE ADMIT
INFO. An information window with menu options is displayed.
Rotate the Trim Knob
control to move the pointer (>); repeat the press, turn, press process to enter characters or make selections.
Select RETURN after all
information is entered.
Select desired option:
SAVE CHANGES or DO NOT SAVE CHANGES.
TAB TRIM
COVER & NON-TAB TRIM
Recall Defaults:
This menu allows the
clinician to recall previously named monitor defaults while monitoring an admitted patient.
Up to five sets of defaults
can be programmed.
COVER & NON-TAB TRIM
TAB TRIM
Rover
This application allows the monitor to be moved or “roved” to the patients bedside. It has only hardwire capability and does not accommodate telemetry.
To Admit a Patient
Connect the AC Power source.
Connect the Network cable.
NOTE: This step is not required when using wireless Dash monitors. It is required for all other Dash monitors.
Push the Power button to
activate the display.
Select MORE MENUS.
Select ADMIT MENU.
Select SET UNIT NAME.
From the information window,
move cursor in front of the desired unit.
Select BED SET NUMBER.
From the information window,
move cursor in front of the desired bed number.
Select ADMIT PATIENT.
NOTE: If the Unit Name or Bed Number windows do not appear, check that the network cable is connected.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Rover
To Discharge a Patient
Remove all ECG leads from
the patient.
Select MORE MENUS.
Select ADMIT MENU.
Select DISCHARGE
PATIENT.
Turn the Trim Knob to move
the cursor in front of Discharge and press to select.
A message DISCHARGED
and ALL ALARMS OFF will appear on the display when the monitor is in a discharged mode.
NOTE: It is recommended to leave the network cable plugged in and the Dash display on for two minutes following discharge (if applicable).
Push the Power button to turn
the display off.
Store the monitor with AC
power cord plugged in and display off.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Combo
This application uses a monitor mounted in a room, but the ECG data can be acquired from either a hardwire cable from the monitor or a telemetry transmitter/transceiver.
To Admit a Patient to
Hardwire
Select MORE MENUS.
Select ADMIT MENU.
Select ADMIT PATIENT.
To Change the ECG
Source from Hardwire to Telemetry
Select MORE MENUS.
Select ADMIT MENU.
Select ECG SOURCE.
Turn the Trim Knob to move
the cursor in front of the desired telemetry transmitter/ transceiver number or Monitor (Discharge Telem) for hardwire capability from the information window and press to select.
TAB TRIM
COVER & NON-TAB TRIM
NOTE: If the Telemetry transmitter/transceiver is being used for the ECG signal, the TTX number will appear in the ECG parameter box.
COVER & NON-TAB TRIM
TAB TRIM
Combo
To Discharge a Patient:
Remove all ECG leads from
the patient.
Select MORE MENUS.
Select ADMIT MENU.
Select DISCHARGE
PATIENT.
Turn the Trim Knob to move
the cursor in front of the desired discharge option and press to select.
Return: Exit to Main Menu.Monitor: Discharges only
the bedside monitor.
Telemetry: Discharges
patient from telemetry.
Both: Discharges both the
monitor and telemetry.
A message DISCHARGED
and ALL ALARMS OFF will appear on the display when the monitor is in a discharged mode.
NOTE: When discharging ONLY the bedside monitor, all stored vital sign data will be deleted. The only data which remain available will be: HR, ST, PVC, and Alarm Histories.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Rover-Combo
This application combines the mobility feature of Rover monitoring with the telemetry capabilities of Combo monitoring.
To Admit a Patient
Connect the AC Power source.
Connect the Network cable.
NOTE: This step is not required when using wireless Dash monitors. It is required for all other Dash monitors.
Push the Power button to
activate the display.
Select MORE MENUS.
Select ADMIT MENU.
Select SET UNIT NAME.
From the information window,
move cursor in front of the desired unit.
Select SET BED NUMBER.
From the information window,
move cursor in front of the desired bed number.
Select ECG SOURCE.
From the information window,
move cursor in front of the desired transmitter/ transceiver or monitor (hardwire) and press to select.
NOTE: If the Unit Name, Bed Number or ECG Source windows do not appear, check that the network cable is connected.
Select ADMIT PATIENT.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Rover-Combo
To Discharge a Patient:
Remove all ECG leads
from the patient.
Select MORE MENUS.
Select ADMIT MENU.
Select DISCHARGE
PATIENT.
Turn the Trim Knob to move
the cursor in front of the desired discharge option and press to select:
Return: Exit to Main
Menu.
Monitor: Discharges only
the bedside monitor.
Telemetry: Discharges
patient from telemetry.
Both: Discharges both the
monitor and telemetry.
NOTE: When discharging ONLY the bedside monitor, all stored vital sign data will be deleted. The only data which remain available will be: HR, ST, PVC, and Alarm Histories.
A message DISCHARGED
and ALL ALARMS OFF will appear on the display when the monitor is in a discharged mode.
Push the Display On/Off
button to turn the display off.
NOTE: It is recommended to leave the network cable plugged in and the Dash display on for two minutes following
discharge.
Store the monitor with the AC
power cord plugged in and the display off.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
ECG
Skin Preparation and Lead
Placement:
The quality of the signal received from the electrodes is a direct result of skin prep and lead placement.
Clip or shave hair from
application sites.
Gently rub the area with a
gauze pad to remove dead skin cells.
Cleanse site with alcohol or
mild soap and water.
Dry skin completely.
Apply electrodes according
to manufacturers recommendations.
Pace Detection
Indicators
PVC Count
(Full Arrhythmia Only)
ST Measurement
Point
QRS
Indicator
TAB TRIM
COVER & NON-TAB TRIM
Heart Rate
Heart Rate
Alarm Limits
ST Analysis Data
ECG/ Pace
COVER & NON-TAB TRIM
TAB TRIM
ECG/ Pace
Display Lead: Top or first lead
displayed on the bedside and graphed with alarms and manual prints.
To Change the Display Lead:Select ECG.Select DISPLAY LEAD II.A popup menu opens.Move the cursor in front of
the desired display lead and press to select.
Select MAIN MENU to
exit.
Size: Changes the size of all
ECG waveforms displayed and graphed. 1X is the standard size.
NOTE: At least a 0.5 millivolt QRS complex at standard size is needed for beat detection.
To Adjust ECG Size:Select ECG.Select ECG SIZE.Turn the Trim Knob to
highlight desired selection and press to select.
Select MAIN MENU to
exit.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Detect Pace: Turns pacemaker
detection On/Off.
Two different options of
pacemaker modes are available. The clinician must judge which mode is best for each patient. Pace 2 is the recommended pacemaker detection mode. Pace 1 should be used if Pace 2 does not adequately detect pacemaker spikes.
To Activate or Deactivate the
Pacemaker Mode:
Select ECG.Select DETECT PACE.Turn the Trim Knob in
front of the desired pacemaker mode and press to select.
Select MAIN MENU to
exit.
NOTE: A P appears in the ECG parameter window when pace detection is enabled. An *” will appear if a paced beat is
detected.
Pace 2 Mode: Analyzes
waveforms with the added capability of minimizing the chance of counting severe residual pacemaker energy as a QRS complex. Pace 2 may not adequately detect all QRS morphologies. Arrhythmia calls such as Asystole or Pause may be made with heart rate identified as less than actual.
NOTE: Pacemaker patients should be kept under close observation.
TAB TRIM
COVER & NON-TAB TRIM
Pace 1 Mode: Analyzes the
presence of pacer spikes, assesses the waveform for residual pacemaker energy and determines the presence of an R-wave following the pacemaker spike. If an event occurs during the first few milliseconds following the spike, it will be counted as a paced spike.
COVER & NON-TAB TRIM
TAB TRIM
Guidelines for Successful
Pacemaker Monitoring:
Multi-vector pace detection is
determined by simultaneous analysis of the displayed ECG Lead and the Vector Lead.
Adequate pacemaker
detection is directly dependant on the quality of the ECG waveform. Proper skin preparation and electrode placement are essential.
Ensure that the pace detection
mode is activated.
All detected pacemaker spikes
will appear upright, uniform and white on the display screen.
If the monitor is not
adequately detecting pacemaker spikes as evidenced by heart rate double counting, pacemaker spikes not detected, alarms for low heart rate or asystole it is recommended that you change the electrode placement to a recommended configuration.
After changing electrode
placement, always RELEARN the ECG waveform. If adjusting the electrode placement does not resolve the detection issue, change the pace detection mode to Pace 1 and RELEARN.
TAB TRIM
COVER & NON-TAB TRIM
Recommended Alternative Lead Placements
COVER & NON-TAB TRIM
TAB TRIM
Limits:
To Change the ECG Alarm
Limit:
Select ECG.Select ECG LIMITS.Select desired alarm limit.Turn the Trim Knob to the
desired alarm limit and press to select.
Select MAIN MENU to
exit.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
View All ECG: Allows six leads
of ECG to be viewed on the display.
To View All ECG:Select ECG.Select VIEW ALL ECG.Six waveforms will be
displayed.
– Press GRAPH GO/STOP to
print displayed leads.
Press the Trim Knob to
remove displayed leads.
Select MAIN MENU to
exit.
Arrhythmia: Arrhythmia
processing can be manually changed to one of the following conditions:
Full: Expands detection to all
arrhythmia conditions defined by the software level.
Lethal: Arrhythmia
processing is limited to Asystole, VFib/VTac and VTach.
Off: Disables all arrhythmia
alarms. Parameter alarms remain active.
To Change the Arrhythmia
Processing Mode:
Select ECG.Select ARRHYTHMIA.Turn the Trim Knob to
move the cursor in front of the desired arrhythmia processing mode and press to select.
Select MAIN MENU to
exit.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Relearn: During ECG
monitoring, it may be necessary to relearn the ECG waveform if a change in the patients normal ECG pattern has occurred, or the electrode placement has changed.
A change in the ECG pattern
could result in:
Incorrect arrhythmia calls.Loss of ST measurement.Inaccurate heart rate
detection.
To Relearn the ECG
Waveform:
Select the ECG parameter
window.
Select RELEARN.The ECG parameter
window will replace the HR with an X and the message Learning will appear above the display lead.
Select MAIN MENU to
exit.
TAB TRIM
COVER & NON-TAB TRIM
Turning the ECG Parameter
Off or On: See monitor setup for details.
COVER & NON-TAB TRIM
TAB TRIM
Single Lead vs. Multi-Lead:
This option examines ECG leads utilized for arrhythmia analysis.
Multi-Lead Analysis examines
ECG leads I, II, III and V to help eliminate false alarms and improve the ability of the system to:
Detect beats that occur
isoelectric to a single chest lead.
Discriminate artifact that
appears in one lead compared to other lead vectors.
Provides a Smart Lead
Fail feature where the failed lead is identified and, if available, another lead is provided for display.
Continue arrhythmia
processing after a lead change.
Single Lead Analysis uses
only the top displayed lead to process heart rate and arrhythmia information.
Single lead analysis can be
beneficial when troubleshooting heart rate and arrhythmia alarms.
To Change to Single Lead
Analysis:
Select the ECG Parameter
Window.
Select LD ANALYSIS:
MULTI-LEAD.
Move the cursor in front of
Single Lead and press to select.
Select MAIN MENU to
exit.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
12 Lead ECG Analysis
NOTE: For the most accurate serial comparison, use the same electrode configuration used on previous analysis.
To Obtain a 12-Lead on a
Patient:
Place patient in a supine
position.
Correctly identify and apply
all 10 electrodes.
Select the ECG Parameter
Window.
Select 12 LEAD ECG
ANALYSIS.
NOTE: Accurate demographics must be entered to ensure proper analysis.
Select ADMIT INFO
Select CHANGE ADMIT
INFO.
Electrode Placement
Enter Information:
First Name: Completely
entered.
Last Name: Completely
entered.
Patient ID: Facility Specific,
(i.e., Medical Record Number, Social Security Number, etc.)
Sex
Front
Leads
V1 Fourth intercostal space at the right
V2 Fourth intercostal space at the left
V3 Midway, between locations V2 and
V4 Mid-clavicular line in the fifth
V5 Anterior axillary line on the same
V6 Mid-axillary line on the same
LA Left deltoid or left wrist.
RA Right deltoid or right wrist.
LL Left thigh or left ankle.
LA Right thigh or right ankle
12 Lead Electrode Placement
sternal border; right chest.
sternal border; left chest.
V4; left chest.
intercostal space; left chest.
horizontal level as V4; left chest.
horizontal level as V4 & V5; left chest.
Birth DateAge: Correct age impacts
the analysis.
HeightWeightLocation ID: Identifies
origin of 12-lead to MUSE for transmission. (Set in defaults. For rover modes, set in admit menu).
Site Number: Identifies
hospital to MUSE for transmission. (Set in defaults).
NOTE: Depending on facility policy, other fields may be required; i.e., Technician ID, Optional Field, Order Number.
(continued on next card)
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
ST
12 Lead ECG
TAB TRIM
ST
12 Lead ECG
(continued from previous card)
Select 12 LEAD ECG NOW.
Message appears:
PERFORMING ANALYSIS. Please wait… After 30
seconds, an unconfirmed 12-lead ECG Analysis window will appear.
Transmission and/or printing
options include the following:
Transmit–Print: Send
12 lead for storage to MUSE and print copy to laser printer.
Transmit: Send
12 lead for storage to MUSE. No printed copy on unit.
Print: Print copy to laser
printer on unit. No 12 lead stored at MUSE.
Delete: Erase the analysis
without storing or printing.
Return: Exit to 12-lead
menu.
TAB TRIM
COVER & NON-TAB TRIM
12 Lead is completed. The last 12 lead analysis window is available for review under REVIEW 12 LEAD
ECG TO TRANSMIT OR PRINT.
COVER & NON-TAB TRIM
TAB TRIM
ST Analysis
ST Menu Options:
ST information with trends
and complexes may appear in a waveform position.
ST by lead data may appear
in the ECG Parameter Window.
The lead with the most
deviation from the isoelectric line appears in the ECG window. The ST value is updated regularly and changes to the alarming lead when limits are exceeded.
NOTE: ST options may vary depending on monitor default configuration.
NOTE: The accuracy of the ST Analysis is dependent on the placement of the electrodes.
Pace Detection
Indicators
PVC Count
(Full Arrhythmia Only)
ST Measurement
Point
QRS
Indicator
TAB TRIM
COVER & NON-TAB TRIM
Heart Rate
Heart Rate
Alarm Limits
ST Analysis Data
ST Trends and Complexes
The monitor has an ST
display which consists of three 30-minute ST trends OR three ECG complexes.
A reference complex for each
of the selected leads is displayed for visual reference purposes when ST is on.
The current complex is
superimposed (green) over the reference complex (gray).
Data is outside of
the displayed
scale
The reference
complex is in back
of the current
complex
COVER & NON-TAB TRIM
TAB TRIM
Turn ST ON/OFF: Turns ST
analysis off and returns to the main menu.
To Turn ST Analysis
Program Off:
Select the ECG parameter
window.
Select ST ANALYSIS.Select TURN ST OFF.Display automatically
returns to the Main Menu.
To Turn ST Analysis
Program On:
Select the ECG parameter
window.
Select ST ANALYSIS.ST complexes and numerics
automatically appear on the display.
Select MAIN MENU to
exit.
Store New References:
Displays a new set of reference complexes on the screen to use as a visual reference.
To store new references:
Select the ECG parameter
window.
Select ST ANALYSIS.
Select STORE NEW
REFERENCES.
Select YES from the popup
menu. The reference complexes
shown on the display will now reflect the current complexes.
Select MAIN MENU to
exit.
NOTE: Selecting STORE NEW REFERENCE does not affect actual ST processing, it is for visual reference only.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
ST Limits: Displays a menu
and information window to set and adjust ST deviation limits.
To Adjust All ST Limits:Select the ECG parameter
window.
Select ST ANALYSIS.Select ADJUST ALL
LIMITS.
Turn the Trim Knob to the
+ / – desired limit and press to select.
All limits are automatically
adjusted in the information window.
The monitor adjusts the low
and high limits around the current value of the lead(s).
Select MAIN MENU to
exit.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Identify V Lead: Identifies the
V Lead being used for ECG and ST analysis.
To Change the V Lead:Select the ECG parameter
window.
Select ST ANALYSIS.Select IDENTIFY V LEAD.Move the cursor in front of
the desired V Lead and press to select.
NOTE: With a 5-leadwire cable, the V Lead is used in ST Analysis and arrhythmia analysis.
Changing this label changes the label on the V-lead trend and complex.
NOTE: With a 10-leadwire cable, the V Lead is used for arrhythmia analysis only. Changing this label DOES NOT change the label on the ST display. Use the ST display menu to change the label.
Select MAIN MENU to
exit.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
ST Display: Changes the leads
for the trends and complexes display.
To Change the Displayed ST
Leads:
Select the ECG parameter
window.
– Select ST ANALYSIS. – Select ST DISPLAY. – Turn the Trim Knob to
deselect one of the three selections before selecting another.
Select MAIN MENU to
exit.
NOTE: A maximum of three choices can be selected for the ST Display.
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
TAB TRIM
COVER & NON-TAB TRIM
COVER & NON-TAB TRIM
TAB TRIM
Loading...
+ 75 hidden pages