GE Aestiva 5 Operation Manual

Page 1
GE Healthcare
Aestiva/5
Operation Manual - Part 2
Software Revision 4.X
Page 2
User Responsibility
w
CAUTION
This Product will perform in conformity with the description thereof contained in this Operation manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Customer Service Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the prior written approval of Datex-Ohmeda. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda.
U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A., check local laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code, the year of manufacture, and a sequential unit number for identification.
AAA F 12345
This alpha character indicates the year of product manufacture and when the serial number was assigned; “D” = 2000, “E” = 2001, “F” = 2002, etc. “I” and “O” are not used.
Aestiva, S/5, PSVPro, Tec 5, Tec 6
Datex-Ohmeda Inc.
Other brand names or product names used in this manual are trademarks or registered trademarks of their respective holders.
and
Tec 7
are registered trademarks of
Page 3
1/Introduction
Table of Contents
2/Cleaning and Sterilization
How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . 1-3
Maintenance summary and schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Operator maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Datex-Ohmeda approved service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Patient path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Scavenging path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Clean and sterilize . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
To wash (by hand or machine) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Autoclave . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Special requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Assemble . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Disassemble the patient path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
i
3/Setup and Connections
Canister disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Disassemble the scavenging path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
How to clean and disinfect the flow sensors . . . . . . . . . . . . . . . . . . . . . . . .2-13
How to clean and sterilize the optional CO
bypass assembly . . . . . . . . . 2-15
2
Breathing system setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Canister setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Pneumatic and electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
How to install gas cylinders (high-pressure leak test) . . . . . . . . . . . . . . . .3-11
Cylinder yokes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
DIN connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
How to install the gooseneck lamp (12 V) . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
How to attach equipment to the top shelves . . . . . . . . . . . . . . . . . . . . . . .3-15
How to install equipment on the foldout shelf . . . . . . . . . . . . . . . . . . . . . .3-17
Installation notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
1006-0939-000
i
Page 4
Aestiva
4/User Maintenance
ii
Repair policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Manifold maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Expiratory valve maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Receiver maintenance (active gas scavenging only) . . . . . . . . . . . . . . . . . .4-8
Flow sensor maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Breathing circuit maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Bellows maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Bellows tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
5/Alarms and Troubleshooting
6/Illustrated Parts
sensor calibration - 21% O
O
2
sensor calibration - 100% O
O
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
2
Flow sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
How to prevent water build-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
About alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Alphabetical list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
Breathing system problems (no alarm) . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Electrical problems (power failure, etc.) . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Pneumatic problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
Breathing system parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Top level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Main manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
Test tools and system parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
7/Theory of Operation and Specifications
Ventilator Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
ii
1006-0939-000
Page 5
Table of Contents
Breathing system schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Electrical block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Pneumatic circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Suction regulators (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-12
flowmeter (optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-12
O
2
Breathing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Breathing system specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
Pneumatic specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-17
Flow specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-18
iii
Index
Warranty
Electrical power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
Power cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
Electrical fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-20
Battery information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-20
Electro-magnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-20
Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-21
Environmental requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-22
Ventilation operating specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
Aestiva ventilator accuracy data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-25
1006-0939-000
iii
Page 6
Aestiva
iv
iv
1006-0939-000
Page 7
1 Introduction
1-1
In this section
1006-0939-000
How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Maintenance summary and schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1-1
Page 8
Aestiva
How to use this manual
This is part two of the Aestiva operation and maintenance manual. It tells you how to:
• Remove and clean parts
OO
22
22
mm
sssseeeennnnssssoooorr
ss
rr
••••SSSSeeeetttt uuuupppp tttthhhheeee ssssyyyysssstttteeeemm
••••IIIIddddeeeennnnttttiiiiffffyyyy aaaannnndddd rrrreeeeppppllllaaaacccceeee wwwwoooorrrrnnnn oooorrrr ddddaaaammmmaaaaggggeeeedddd ppppaaaarrrrttttss
••••CCCCaaaalllliiiibbbbrrrraaaatttteeee tttthhhheeee OO
The second half of this section is a maintenance schedule.
The last section, troubleshooting, tells you what causes each alarm and what you can do about it.
Use this manual together with Part 1, which includes the operating instructions and preoperative checkout.
w
WARNING If an alarm occurs, safeguard the patient first, before
troubleshooting or repair procedures.
1-2
1006-0939-000
Page 9
N
N
Symbols used in the manual or on the equipment
w
Warnings and
occur if you do not follow all instructions in this manual.
Warnings tell about a condition that can cause injury to the operator or the patient.
Cautions tell about a condition that can cause damage to the equipment. Read and follow all warnings and cautions.
Other symbols replace words on the equipment or in Datex-Ohmeda manuals. No one device or manual uses all of the symbols. These symbols include:
Cautions tell you about dangerous conditions that can
w
1 Introduction
1-3
l
O o q p œ †
~
x
On (power)
Not autoclavable
Í
Off (power)
Type B equipment
m
Standby
Type BF equipment
µ
Standby or preparatory state for part of the equipment
“ON” only for part of the equipment
H
Type CF equipment
Caution, ISO 7000-0434
w
“OFF” only for part of the equipment
wW
Direct current
Alternating current Dangerous Voltage
Protective earth ground
Attention, refer to product instructions, IEC 601-1
This way up
Earth ground
y
r
1006-0939-000
Electrical input Electrical output
Pneumatic inlet Pneumatic outlet
Frame or chassis ground
REF
Stock Number
1-3
Page 10
Aestiva
å Y
t
T
+
-
Alarm silence button
Equipotential Systems with this mark agree with
Variability Read top of float
Variability in steps Vacuum inlet
Plus, positive polarity Suction bottle outlet
Minus, negative polarity
SN
O2+
Serial Number
the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their Operation and Maintenance Manuals. The xxxx is the certification number of the Notified Body used by Datex­Ohmeda’s Quality Systems.
O2 Flush button
P
N
ˆ
z
Z
Lamp, lighting, illumination Cylinder
Movement in one direction Isolation transformer
Movement in two directions Linkage system
Lock Risk of Explosion
Unlock Low pressure leak test
1-4
1006-0939-000
Page 11
1 Introduction
134°C
R u q t
Autoclavable
Bag position/ manual ventilation
Mechanical ventilation
r
Open drain (remove liquid)
Close drain
U
Inspiratory flow
Expiratory flow
Q
O2 sensor connection End case
The primary regulator is set to pressure less than 345 kPa.
European Union Representative
The primary regulator is set to pressure less than 414 kPa.
1-5
1006-0939-000 1-5
Page 12
Aestiva
Maintenance summary and schedule
These schedules show the minimum frequency. You will have to service the equipment more frequently if you use it:
• In unusual conditions (dirty gas supplies, high temperature, high humidity, etc.).
• More frequently than normal.
Operator maintenance
Examine all components and do the maintenance procedures more frequently if necessary.
Minimum Frequency
Maintenance
Daily • Clean the external surfaces.
• 21% O
calibration (circuit O2 sensor).
2
• Flow sensor calibration
Two weeks • Drain the vaporizers and discard the agent. This is not
necessary for Tec 6 vaporizers.
Monthly • 100% O
calibration (cIrcuit O2 sensor).
2
• Put Krytox (or a lubricant approved for use with 100% ) on all tee handle threads.
O
2
During cleaning and setup
• Inspect the parts for damage. Replace or repair as
necessary
Annually • Replace the external o-rings on the vaporizer ports.
As necessary • Install new cylinder gaskets on cylinder yokes.
• Replace the absorbent in the absorber.
• Drain the breathing system.
• Drain the breathing circuit module.
1
• Drain the overflow trap on the optional suction
regulator.
• Replace the circuit O
• Replace the disposable flow sensors (plastic)
sensor.
2
2
• Replace the autoclavable flow sensors (metal)3.
• Replace the receiver filter (active gas scavenging only).
1. This is included in the preoperative test procedure.
2. Under typical use the sensor meets specifications for 3 months
3. Under typical use the sensor meets specifications for 1 year.
1-6 1006-0939-000
Page 13
Datex-Ohmeda approved
service
1 Introduction
Minimum Frequency Maintenance
6 months Have an approved service person do the service tests
and scheduled service maintenance.
1-7
1006-0939-000 1-7
Page 14
Aestiva
1-8 1006-0939-000
Page 15
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
2 Cleaning and Sterilization
2-1
Obey applicable safety precautions:
• During the cleaning of the liquid collecting tray, avoid skin or eye contact with the contents of the absorber. In the event of skin or eye contact, immediately rinse the affected area with water and seek medical assistance.
• Read the material data sheet for each cleaning agent.
• Read the operation and maintenance manual for all sterilization equipment.
• Wear gloves and safety glasses. A damaged O
sensor can leak and cause
2
burns (contains potassium hydroxide).
• Do not breathe the fumes.
CCAAAAUUUUTTTTIIIIOOOONNNN
w CC
To prevent damage:
• Refer to the manufacturer’s data if you have questions about a cleaning agent.
• Do not use organic, halogenated, or petroleum based solvents, anesthetic agents, glass cleaners, acetone, or other harsh cleaning agents.
• Do not use abrasive cleaning agents (such as steel wool, silver polish or cleanser).
• Keep all electronic parts away from liquids.
• Do not permit liquid to go into the equipment housings.
• Do not soak synthetic rubber parts for more than 15 minutes. Swelling or faster aging can occur.
• Only autoclave parts that are marked 134°C.
In this section Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Clean and sterilize . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Disassemble the patient path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Canister disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Disassemble the scavenging path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
How to clean and disinfect the flow sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
How to clean and sterilize the optional CO
1006-0939-000 2-1
bypass assembly . . . . . . . . . . . . . . . . . 2-15
2
Page 16
Aestiva
Summary
2-2
Patient path
The parts in Figure 2-1 send exhaled gas back to the patient. They may require more frequent cleaning/sterilization than parts in Figure 2-2. Refer to your hospital’s infection control policy.
Autoclave (134°C) or wash (mild detergent pH <10.5)
AB.23.003
*
If plastic, refer to cleaning/disinfection procedure. If metal, autoclave (134°C)
Wipe with a damp cloth.
*
Figure 2-1 • These parts return
exhaled
gas to the patient
AB.23.079
2-2 1006-0939-000
Page 17
Scavenging path
AB.23.003
2 Cleaning and Sterilization
Autoclave (134°C) or wash (mild detergent pH <10.5)
2-3
Figure 2-2 • These parts do not send gas back to the patient
AB.23.080
1006-0939-000 2-3
Page 18
Aestiva
Clean and sterilize
2-4
To wash (by hand or
machine)
The Disassembly part of this section tells you how to remove parts for cleaning.
Patient path
AB.23.085
Scavenging path
AB.23.083
Use a mild detergent (pH <10.5). Then, rinse and dry completely. All parts except the O
User maintenance tells you how to disassemble parts and clean inside them if necessary.
2-4 1006-0939-000
sensor, and flow sensors can be washed.
2
Page 19
Autoclave
Metal Only
2 Cleaning and Sterilization
Patient path
2-5
AB.23.087
Upside down
AB.23.086
Scavenging path
AB.23.083
Autoclave at 134°C. Inspect the parts for deterioration. The User Maintenance section tells you how to do this.
1006-0939-000 2-5
Page 20
2-6
Aestiva
Special requirements
*
AB.23.088
* Hang the bellows upside down
(extended) to dry. If not, the
convolutions can stick together.
•To clean the circuit O2 sensor, wipe it with a damp cloth. Do not put the sensor in liquid.
• To clean/disinfect metal/or plastic flow sensors, use the flow sensor cleaning procedure. Do not get the connectors wet.
• Disassemble the bellows before you wash it. If not, it will take a very long time to dry. Hang the bellows upside down to dry.
• Assemble the bellows before you autoclave. Autoclave the bellows upside down.
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
CCAAAAUUUUTTTTIIIIOOOONNNN
w CC
Assemble
Do not use talc, zinc stearate, calcium carbonate, corn starch or equivalent materials to prevent tackiness. These materials can go into the patient’s lungs and airways and cause irritation or injury.
Do not put the circuit O2 sensor or flow sensor connector in liquid.
w Do not autoclave the Circuit O
sensor or the plastic flow sensors.
2
w Do not clean the interior surfaces of the flow sensors. Use a damp
on external surfaces only.
cloth
The Setup section tells you how to assemble the breathing system.
Inspect all parts for deterioration. Replace them if necessary.
The Preoperative tests in Part 1 (the first half of this manual), tell you how to test the system for correct operation.
2-6 1006-0939-000
Page 21
Disassemble the patient path
2 Cleaning and Sterilization
Step 1
Open the access panel.
Step 2
Push up on the latch located under the flow sensor module.
Step 3
Remove the flow sensor module. You will feel some resistance. Continue to pull.
2-7
AA.96p.009
AA.96p.010
AA.96p.011
Step 4
Disconnect the cable and remove the O
sensor.
2
AA.96p013, 014
1006-0939-000 2-7
Page 22
2-8
Aestiva
Step 5
Completely loosen the thumbscrew on the breathing circuit module.
AA.96p016
Step 6
Remove the module. You will feel some resistance. Continue to pull.
Step 7
Push the metal button to the first stop. Pull out the bag arm.
Step 8
Push the metal button to the second stop and open the top panel.
AA.96p017
AA.96p018
AA.96p020
2-8 1006-0939-000
Page 23
Step 9
Push the release button and remove the bellows assembly (pull up).
Step 10
Remove the main manifold.
2 Cleaning and Sterilization
2-9
AA.96p021, 23
AA.96p024
Autoclave assemblies marked 134°C. Refer to the Section “Clean and sterilize” for complete instructions. To assemble the circuit refer to the Setup section.
1006-0939-000 2-9
Page 24
Aestiva
Canister disassembly
2-10
w CCCCAAAAUUUUTTTTIIIIOOOONNNNTo prevent damage, pull the release handle forward. Then turn the handle
Step 1
Pull the release forward.
AA.96p026
wZ
Step 2
Turn the release clockwise.
AA.96p028
Step 3
Pull down and out on the canisters.
AA.96p029
2-10 1006-0939-000
Page 25
Step 4
Push in the buttons on each side of the top dish.
Step 5
Lift up the drain dish.
Remove the top dish and drain dish.
2 Cleaning and Sterilization
2-11
AA.96p053
AA.96p054
Continue with the next section or go to "Clean and sterilize". To assemble, refer to the Setup section.
1006-0939-000 2-11
Page 26
Aestiva
Disassemble the scavenging path
These parts send exhaled gas to the disposal system, not the patient.
2-12
Exhalation valve
Complete the basic disassembly procedure.
Push in the latches and pull out the exhalation valve.
AB.23.089
Scavenging reservoir
Loosen the knob and remove the rear cover.
Pull the reservoir up and to the rear.
Twist and pull down to remove the tube.
AB.23.090
2-12 1006-0939-000
Page 27
How to clean and disinfect the flow sensors
wwwwCCCCAAAAUUUUTTTTIIIIOOOONNNNDo not autoclave plastic flow sensors.
wwwwDo not use high-pressure gas, or brushes to clean the flow sensors.
wwwwDo not use cleaning solvents that are not approved for use with
Polycarbonates (e.g. CIDEX Plus).
CIDEX sterilization Both Datex-Ohmeda and the manufacturer of CIDEX (Johnson & Johnson)
have tested this procedure.
• CIDEX must be 14 day mixture, with activator vial REF REORDER # 2245
• One liter of this solution cleans four (4) flow sensors
2 Cleaning and Sterilization
2-13
Procedure (Figure 2-3) 1. Remove the flow sensor module from the absorber. Refer to “Disassemble
the patient path” in this section.
2. Remove the flow sensors from the module.
• Push in the latch.
• Pull off the cover.
• Remove the flow sensors.
3. Submerge the flow sensor and tubes in activated CIDEX solution. Keep the connector dry.
4. Keep the solution in the tubes for the sterilization period.
5. Submerge the flow sensor and tubes in distilled water. Again, do not get the connector wet.
6. Rinse as indicated in CIDEX instructions.
7. Do steps 5 and 6 again to remove all CIDEX.
8. COMPLETELY dry the flow sensor and the tubes before you use the sensor.
Use a dry syringe, or connect vacuum or pressure to remove all liquid from the sensor (sensor, tubes, and connector):
•Minimum time: 1 min
•Maximum vacuum: 30 in Hg
•Maximum flow: 10 L/min flow
•Maximum pressure: 345 kPa (50 psi).
1006-0939-000 2-13
Page 28
2-14
Aestiva
Step 2
Steps 3 and 4
Steps 5, 6 and 7
Step 8
Dry for > 1 min with these precautions:
• Maximum vacuum 30 in Hg
• Maximum flow 10 L/min
• Maximum pressure 345 kPa (50 psi)
Figure 2-3 • Steps 2-8
2-14 1006-0939-000
Page 29
2 Cleaning and Sterilization
HHHHoooowwww ttttoooo cccclllleeeeaaaannnn aaaannnndddd sssstttteeeerrrriiiilllliiiizzzzeeee tttthhhheeee ooooppppttttiiiioooonnnnaaaallll CCCCOO
These instructions assume that the system has the optional CO2 bypass
feature installed.
11
SSSStttteeeepppp 11
Remove the absorber canisters. (Also described in “Canister disassembly” in this section.)
a. Pull the canister release handle forward.
w CCCCAAAAUUUUTTTTIIIIOOOONN
b. Turn the release handle clockwise.
c. Lift out the canisters.
Note: The canisters should drop under their own weight when released. If they do not, clean and lubricate the pins in the drain dish and sockets in the base.
22
SSSStttteeeepppp 22
Release the CO2 bypass assembly by
pushing both buttons (one on each side of the assembly).
NN
To prevent damage, pull the canister release handle forward before turning the
handle.
OO
bbbbyyyyppppaaaassssssss aaaasssssssseeeemmmmbbbbllllyy
22
22
yy
2-15
33
SSSStttteeeepppp 33
Pull down on the CO2 bypass assembly
(a) to remove it and lift the drain dish (b) up.
a
b
1006-0939-000 2-15
Page 30
2-16
Aestiva
44
SSSStttteeeepppp 44
Clean and sterilize the bypass assembly:
a. Immerse the assembly into a
solution of mild detergent and water.
b. Agitate the assembly in the
solution while repeatedly actuating and releasing the plate.
c. Rinse the assembly by
immersing it in clean water and agitating it again with the plate actuated and released.
d. Autoclave the assembly at
134°C.
e. Cool about 40 minutes at room
temperature.
a
Note: Optionally, the bypass assembly can be divided into two smaller sections for easier cleaning by releasing the top clamp (a) and pulling the hose from the assembly. When reassembling, reattach the hose to the bottom part of the assembly. Secure the hose with the clamp and pinch the clamp one notch past finger-tight with pliers.
55
SSSStttteeeepppp 55
Reinstall the bypass assembly:
a. Reinstall the drain dish.
b. Insert the bypass assembly.
c. Push back and adjust to locate
the access holes for the top ports.
d. Push the buttons on each side of
the assembly and push the assembly up; both buttons should snap into place.
e. Pull down on the assembly to
make sure it is locked into place.
c
b
a
d
2-16 1006-0939-000
Page 31
66
SSSStttteeeepppp 66
Reinstall the canisters:
2 Cleaning and Sterilization
a. Ensure seals and rim are free of soda lime dust.
b. Place the canisters in the drain dish.
c. Turn the canister release handle counterclockwise to its locked position while gently applying pressure on the
canisters to hold them against the canister guides.
77
SSSStttteeeepppp 77
Test the Bag circuit for leaks according to the Breathing System tests in the Appendix section of Part 1 of the Operation manual.
• Do the test in both the Absorber mode and the Bypass mode.
• Ensure the alarm message “No CO in section 3 of Part 1 of the Operation manual.
88
SSSStttteeeepppp 88
Before you use the system, complete the Preoperative Test procedure. Refer to the ‘Appendix - Preoperative Tests’ section of the Operation Manual Part 1, System Controls, Operation and Checkout.
Absorption” activates when in Bypass mode. See “CO2 Bypass mode operation”
2
2-17
1006-0939-000 2-17
Page 32
2-18
Aestiva
2-18 1006-0939-000
Page 33
3 Setup and Connections
IIIImmmmppppoooorrrrttttaaaannnnttttDatex-Ohmeda strongly recommends that you use O2 monitoring with this equipment.
Refer to local standards for mandatory monitoring.
IIIImmmmppppoooorrrrttttaaaannnnttttEuropean Standard EN 740 requires anesthetic agent monitoring when anesthetic
vaporizers are in use.
IIIImmmmppppoooorrrrttttaaaannnnttttEuropean Standard EN 740 requires CO2 monitoring during ventilation.
ww
WWWWAAAARRRRNNNNIIIINNNNGGGGSSSSAlways make sure that the pipeline supply hoses and the breathing circuit
ww
components are not toxic and will not:
• Cause an allergic reaction in the patient.
• React with the anesthetic gases or agent to produce dangerous by-products.
ww
ww
To prevent incorrect values or equipment malfunction, use only Datex-Ohmeda cables, hoses and tubing.
ww
ww
This system operates correctly at the electrical interference levels of IEC 601-1-2. Higher levels can cause nuisance alarms that may stop mechanical ventilation.
ww
ww
To help prevent false alarms from devices with high-intensity electrical fields:
• Keep the electrosurgical leads away from the breathing system and the
flow and oxygen sensors.
• Do not put the electrosurgical leads on any part of the anesthesia system.
3-1
ww
ww
To protect the patient when electrosurgical equipment is used:
• Monitor the correct operation of all life support and monitoring equipment.
• Keep backup manual ventilation available in case the electrosurgical
equipment prevents safe use of the ventilator.
• Do not use conductive masks or hoses.
In this section Breathing system setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Canister setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Pneumatic and electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
How to install gas cylinders (high-pressure leak test) . . . . . . . . . . . . . . . . . . . . . . .3-11
How to attach equipment to the top shelves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
How to install equipment on the foldout shelf . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
Installation notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
1006-0939-000 3-1
Page 34
Aestiva
Breathing system setup
Step 1
To assemble the patient circuit, start at step 5.
3-2
Install the main manifold.
AA.96p024
Step 2
Install the Bellows.
•Make sure the release button aligns correctly.
AA.96p022
Step 3
Gently close the control panel.
AA.96p012
Step 4
Install the bag arm.
Install a rebreathing bag on the bag arm.
AA.96p019
3-2 1006-0939-000
Page 35
Step 5
Identify the breathing circuit module (Circle or Mapleson).
3 Setup and Connections
Look at the tabs on the rear or the label on the front.
Step 6
Install the breathing circuit module. You will feel some resistance. Continue to push.
Step 7
Completely tighten the thumbscrew.
AB.23.211
Circle
Mapleson
3-3
AA.96p017
AA.96p016
Step 8
Install the O2 sensor.
• Completely tighten the sensor.
• Connect the cable.
AA.96p013, 014
1006-0939-000 3-3
Page 36
3-4
Aestiva
Step 9
Install the flow sensor module.
• You will hear a “click” when it locks into position.
• Refer to Note 1 or the User Maintenance section.
Step 10
Close the front panel.
AA.96p011
AA.96p009
Step 11
Connect the patient circuit.
Circle Module
AB.23.104
Mapleson D / Bain Module
To-fro
AB.23.152
3-4 1006-0939-000
Page 37
Note 1
Flow sensor assembly
3 Setup and Connections
3-5
AA.96p039
AA.96p038
1006-0939-000 3-5
Page 38
Aestiva
Canister setup
Each canister holds 1.35 kg of loose absorbent or 1.13 kg of commercially packaged (pre-pack) absorbent. Datex-Ohmeda recommends Sodium or Bar­ium hydroxide based absorbent.
3-6
WWWWAAAARRRRNNNNIIIINNNNGGGG
w
Step 1
Pull the release forward.
Change absorbent often to prevent the build up of non-metabolic gases when the system is not in use.
Inspect absorbent at the end of a case. During non-use, absorbent can go back to the original appearance. Refer to the absorbent labeling for more information about color changes.
If the absorbent completely dries out, it may give off CO (carbon monoxide) the next time you use it. For safety, replace the absorbent.
AB.23.061
Step 2
Turn the release clockwise.
AB.23.061
3-6 1006-0939-000
Page 39
Step 3
Push in the buttons and install the top dish and drain dish.
Pull down to make sure the top dish is locked in place.
Step 4
Fill the canisters with absorbent and install them. Push down on the drain dish if necessary.
3 Setup and Connections
AA.96p054
3-7
AA.96p029
Step 5
Turn the release counter-clockwise.
Step 6
Push in the release.
AA.96p028
AA.96p026
1006-0939-000 3-7
Page 40
Aestiva
Pneumatic and electrical connections
WWAAAARRRRNNNNIIIINNNNGGGG
wWW
Equipment connected to the electrical outlets can increase the leakage current. Regularly test the leakage current.
3-8
Mains inlet
CCAAAAUUUUTTTTIIIIOOOONNNN
wCC
Use only medical grade gas supplies. Other types of gas supplies may contain water, oil, or other contaminants.
The Aestiva gas supplies also supply these devices through internal connec­tions:
• The venturi suction regulator (optional)
• The external O
• Ventilator drive gas
• Pneumatic outlets
flowmeter (optional)
2
AA.96.116
Outlets
Labels show outlet voltage ratings and circuit breaker amp ratings.
AA.96.116
Tec 6 power
AA.96.095
3-8 1006-0939-000
Page 41
Pipeline inlets
Label identifies the gas.
Pneumatic outlet
Labels identify the gas.
3 Setup and Connections
AA.96.116
3-9
Note: The outlet pressures range from pipeline pressure to 550-690 kPa when cylinder supplies are used.
Scavenging
Adapters may be necessary. Refer to the Illustrated Parts section.
AA.96.116
AA.96.137
1006-0939-000 3-9
Page 42
3-10
Aestiva
Suction regulator
Optional item.
Note: Venturi regulators use the system Air or O
Standard regulators must be connected to an external vacuum supply.
supply.
2
External vacuum (non-venturi)
Filter
Patient
External O2 flowmeter
Optional item.
Serial port
Refer to the communications appendix for command and data formats.
Collection Bottle
AA.96.172
AA.96.045
AA.96.116
3-10 1006-0939-000
Page 43
How to install gas cylinders (high-pressure leak test)
3 Setup and Connections
Cylinder yokes
1. Find the cylinder wrench.
AA.96.123
2. Close the cylinder valve on the cylinder to be replaced.
AA.96.122
3. Fully loosen the tee handle.
AA.96.124
4. Open the cylinder yoke.
3-11
AA.96.125
5. Remove the used cylinder and the used gasket.
AA.96.126
6. Remove the cap from the cylinder valve on the new cylinder.
7. Point the cylinder valve away from all flammable items and items that can be damaged by a release of pressure.
8. Quickly open and close the cylinder valve. This removes dirt from the
cylinder outlet.
1006-0939-000 3-11
Page 44
3-12
Aestiva
w CC
CCAAAAUUUUTTTTIIIIOOOONNNN
No gasket or more than one gasket can cause a leak.
9. Install a new gasket.
10.Align the cylinder post with the index pins.
11.Close the yoke gate and tighten the tee handle.
12.Install a cylinder plug and gasket in all empty cylinder yokes.
13.Do a high-pressure leak test:
• Disconnect pipeline supplies.
• Disconnect all equipment from the pneumatic outlet and turn Off the auxiliary flowmeter.
• Set the system switch to Standby.
• Open the cylinder.
• Record the cylinder pressure.
• Close the cylinder.
• If the cylinder pressure decreases more than 690 kPa in one minute, there is a leak.
To repair a leak:
Install a new cylinder gasket and tighten the tee handle.
Do this step again. If the leak continues, do not use the system.
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGGGGDo not leave gas cylinder valves open if the pipeline supply is in use. Cylinder supplies
could be depleted, leaving an insufficient reserve supply in case of pipeline failure.
DIN connections
1. Close the cylinder valve on the cylinder to be replaced.
AA.96.127
2. Loosen the adapter and remove the cylinder.
AA.96.129
3. Remove the cap from the cylinder valve on the new cylinder.
4. Point the cylinder outlet away from all items that can be damaged by a release of high-pressure gas.
3-12 1006-0939-000
Page 45
3 Setup and Connections
5. Open and immediately close the cylinder valve to remove dirt or contaminants.
6. Do a high pressure leak test:
• Disconnect pipeline supplies.
• Disconnect all equipment from the pneumatic outlet and turn Off the auxiliary flowmeter.
• Set the system switch to Standby.
• Open the cylinder.
• Record the cylinder pressure.
• Close the cylinder.
• If the cylinder pressure decreases more than 690 kPa in one minute, there is a leak.
To repair a leak:
Tighten the cylinder connection.
Do this step again. If the leak continues, do not use the system.
3-13
wwww WWWWAAAARRRRNNNNIIIINNNNGGGGDo not leave gas cylinder valves open if the pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve supply in case of pipeline failure.
1006-0939-000 3-13
Page 46
Aestiva
How to install the gooseneck lamp (12 V)
wWWWWAAAARRRRNNNNIIIINNNNGGGGUse ONLY 2.4 Watt bulbs. Higher wattage bulbs can get hot enough to burn
you.
3-14
1. Align the connector with the cable towards you.
AA.96.173
2. Push the connector firmly into the socket until you hear a click.
3. Move the light into position.
3-14 1006-0939-000
Page 47
How to attach equipment to the top shelves
The basic system has two straps for equipment. Accessory shelves use clips and straps.
w CCCCAAAAUUUUTTTTIIIIOOOONNNNEach accessory shelf has a weight limit of 23 kg (50 lb).
w Systems without accessory shelves have a weight limit of 23 kg (50 lb).
1. Install or find the clips:
•With an accessory shelf, install the clips in the two slots nearest to the equipment.
3 Setup and Connections
3-15
• Systems without accessory shelves have the clips attached.
AA.96.130
AA71.145
1006-0939-000 3-15
Page 48
3-16
Aestiva
2. Install the straps:
•With an accessory shelf:
A
Large equipment.
Loop one time
(A, B, C)
A
Small equipment.
Loop two times
(A, B, C, D)
B
C
B
C
AA.50.015
D
A
A
B
B
C
C
•Without an accessory shelf:
D
C
B
A
3. Fully tighten the straps.
4. Make sure that the straps hold the equipment in position.
AA.96.135
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
3-16 1006-0939-000
If you do not fully tighten the strap, equipment can fall off the shelf.
Page 49
How to install equipment on the foldout shelf
This is a hidden shelf for equipment that you do not need to look at, adjust, or touch. To access the shelf, you must remove the cylinders.
1. Make sure you have a cylinder wrench (if necessary).
2. Close the valves on all gas cylinders.
3. Remove all gas cylinders.
4. Loosen the thumbscrews and open the rear door.
3 Setup and Connections
3-17
AA.96.118
AA.96.119
AA.96.120
5. Put the equipment on the shelves.
6. Make sure the item does not hang over the raised edge.
7. Connect all cables and route them through the cutouts in the door.
8. Close the door.
9. Test the equipment.
AA.96.114
10.Install the cylinders.
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
1006-0939-000 3-17
Items that hang over the raised edge can fall off.
Page 50
Aestiva
Installation notes
When the system is installed the service representative will check these set­tings and change them if necessary.
3-18
wWWWWAAAARRRRNNNNIIIINNNNGGGGThese settings can only be changed by Qualified Service personnel.
• Language
• Power up defaults: When you turn the system off, should it save the current settings or go back to the facility defaults.
• Automatic calculation of
• Altitude
• Ventilator drive gas
• Heliox mode availability
E alarm limits during mechanical ventilation.
3-18 1006-0939-000
Page 51
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
4 User Maintenance
TO PREVENT FIRES:
• Use lubricants approved for anesthesia or O2 equipment, such as Krytox
• Do not use lubricants that contain oil or grease. They burn or explode in high O
concentrations.
2
• All covers used on the system must be made from antistatic (conductive) materials. Static electricity can cause fires.
Obey infection control and safety procedures. Used equipment may contain blood and body fluids.
®
4-1
1
.
In this section Repair policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Manifold maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Expiratory valve maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Receiver maintenance (active gas scavenging only) . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Flow sensor maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Breathing circuit maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Bellows maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Bellows tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
sensor calibration - 21% O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
O
2
sensor calibration - 100% O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
O
2
Flow sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
How to prevent water build-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
1.
® Krytox is a registered trademark of Dupont de Nemours E.I. & Company Inc
1006-0939-000 4-1
Page 52
4-2
Aestiva
Repair policy
CCAAAAUUUUTTTTIIIIOOOONNNN
w CC
Do not use malfunctioning equipment. Make all necessary repairs or have the equipment serviced by an authorized Datex-Ohmeda service representative. After repair, test the equipment to ensure that it is functioning properly, in accordance with the manufacturer’s published specifications.
To ensure full reliability, have all repairs and service done by an authorized Datex­Ohmeda service representative. If this cannot be done, replacement and maintenance of those parts listed in this manual may be undertaken by a competent, trained individual having experience in the repair of devices of this nature.
No repair should ever be attempted by anyone not having experience in the repair of devices of this nature.
Replace damaged parts with components manufactured or sold by Datex-Ohmeda. Then test the unit to ascertain that it complies with the manufacturer’s published specifications.
Contact the Datex-Ohmeda Customer Service Center for service assistance. In all cases, other than where Datex-Ohmeda’s warranty is applicable, repairs will be made at Datex­Ohmeda’s current list price for the replacement part(s) plus a reasonable labor charge.
4-2 1006-0939-000
Page 53
Manifold maintenance
Step 1
Complete the basic disassembly (Refer to the Cleaning and Sterilization section).
4 User Maintenance
AB.23.094
4-3
Step 2 (APL valve)
Remove the diaphragm. Replace it if:
• The disk, cage, or diaphragm is damaged.
• The diaphragm looks worn.
• The diaphragm fits over the rim.
1
Step 3 (Seals)
Loosen the thumbscrew and remove the plate.
• Replace the seal if damaged or worn.
AB.23.069
AB.23.070
1006-0939-000 4-3
Page 54
Aestiva
Step 4 (Seals con.)
Loosen the thumbscrews.
4-4
AB.23.073
Step 5 (Seals con.)
Open the manifold (top middle, and bottom parts).
AB.23.074
Step 6 (Seals con.)
Replace the parts if they are damaged or worn.
Top
4-4 1006-0939-000
Bottom
Side
AB.23.071
Page 55
Step 7
(Bag/Vent)
Replace the parts if they are damaged or worn.
2
4 User Maintenance
4-5
AB.23.077
Step 8
Put the parts together
•Make sure all gaskets are fully seated.
• Fully tighten the thumbscrews.
AB.23.075
1. When you turn the APL knob, you change the pressure that is necessary to push the disk up and permit gas to exit.
2. The Bag/Vent switch pushed down plungers to close the other path. In the Bag position, the vent path is closed, etc.
Before you use the system, complete the preoperative test procedure. Refer to
"Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
1006-0939-000 4-5
Page 56
4-6
Aestiva
Expiratory valve maintenance
Disassemble the manifold and examine the seals at 12 autoclave cycles.
These conditions indicate a possible leak in the exhalation valve:
•A decrease in airway pressure during the inspiratory pause in the volume control mode of ventilation.
Alarms for low tidal volume.
The ventilator circuit has a leak, but the bellows pass the bellows tests in this section.
Step 1
Remove the exhalation valve (Cleaning and Sterilization section).
AB.23.089
Step 2
Remove the top cover.
AB.23.093
4-6 1006-0939-000
Page 57
Step 3
Unscrew the lock ring (one quarter turn) and remove the cap.
Step 4
4 User Maintenance
AB.23.072
4-7
Put the parts together
• Align the dimples in the cap with the valve seat.
• To install the top, align it with the base and push down.
AB.23.021
Dimples
Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
1006-0939-000 4-7
Page 58
Aestiva
Receiver maintenance (active gas scavenging only)
A blocked filter decreases flow through the scavenging system. In extreme cases this can open the high-pressure relief valve and vent to room air.
Note: Passive gas scavenging does not require maintenance.
Step 1
4-8
Remove the receiver:
• Loosen the thumbscrew.
• Remove the rear cover.
• Pull out and up on the receiver.
Step 2
Replace the filter (friction fit with gasket).
AB.23.091
AB.23.092
Filter
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
This system operates correctly at the electrical interference levels of IEC 601-1-2. Higher levels can cause nuisance alarms that may stop mechanical ventilation.
• Always make sure that the rear cover is installed.
• Completely tighten the thumbscrew.
Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
4-8 1006-0939-000
Page 59
Flow sensor maintenance
Step 1
Remove the flow sensor module (Cleaning and Sterilization section). Push in the tab and remove the cover.
Step 2
4 User Maintenance
4-9
AA.96p.037
Install the new sensor with the arrows up.
Keep the tubes straight.
Step 3
Slide on the cover. Do not pinch the tubes.
AA.96p.039AA.96p.038
Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
1006-0939-000 4-9
Page 60
Aestiva
Breathing circuit maintenance
Disassemble the patient circuit module to replace check valves or seals.
Step 1
4-10
Remove the patient circuit module (Cleaning and Sterilization section).
AA.96p.017
Step 2
Hold in the tab.
AA.96p.041
Step 3
Remove the rear cover. Note: The rear seal is keyed to fit a specific module.
AA.96p.042
4-10 1006-0939-000
Page 61
Step 4
Push in the tabs and remove the top cover.
Step 5
Remove the seals. Turn and pull to remove the check valves.
4 User Maintenance
AA.96p.043
4-11
AA.96p.044
AA.96p.045
Do the steps in the opposite order to assemble the module.
Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
1006-0939-000 4-11
Page 62
4-12
Aestiva
Bellows maintenance
Step 1
Remove the bellows from the breathing system.
AA.96p.023
Step 2
Turn the housing counter-clockwise and lift.
Step 3
Remove the bottom edge of the bellows from the rim.
AA.96p.051
AA.96p.052
4-12 1006-0939-000
Page 63
Step 4
Remove the disk from the bellows.
4 User Maintenance
AA.96p.031
Step 5
Remove the ring from inside the top of the bellows.
Step 6
Push the latch toward the center and remove the rim.
4-13
AA.96p.032AA.96p.033
Step 7
Remove the pressure relief assembly.
AA.96p.034
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
1006-0939-000 4-13
Do not disassemble the pressure relief valve. This can damage the seat or diaphragm and cause injury to the patient.
Page 64
Aestiva
Step 8
Push the latch towards the center and remove the locking tabs.
4-14
AA.96p.035
Step 9
Remove the seal.
AA.96p.036
Do these steps in the opposite order to assemble the bellows. If you see a dust-like powder on the housing or the bellows, apply a thin layer of KRYTOX lubricant to the ribs of the bellows housing. Make sure the lubricant is applied smoothly and there are no lumps.
Make sure that:
• The arrow on the seal points up.
• You hear a double click when you install the rim.
• The rim is locked in position.
• The inner ring is correctly installed inside the top of the bellows.
• Only the bottom ring of the bellows fits over the rim.
• The housing is locked in position. You cannot lift it off.
4-14 1006-0939-000
Page 65
Bellows tests
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
4 User Maintenance
Objects in the breathing system can stop gas flow to the patient. This can cause injury or death:
Do not use a test plug that is small enough to fall into the breathing system.
4-15
Make sure that there are no test plugs or other objects caught in the breathing system.
The bellows assembly test does not replace the preoperative tests. Always complete the tests in the section Preoperative Tests before you use the system with a patient.
Step 1
Hold the bellows assembly vertical and close the ports (A and B).
This test makes sure that all components are correctly assembled. It is not an alternative to a complete system checkout.
If the bellows operates correctly, install it in the system.
If there is a problem, disassemble the bellows. Look for and replace damaged parts.
A B
1006-0939-000 4-15
Page 66
4-16
Aestiva
Step 2
Invert the bellows. They must not fall more than 100 mL/min.
If it does:
• The ports are not tightly sealed.
• The bellows is incorrectly installed.
• The seal inside the bellows is not correctly installed (with its groove pointed up).
• Parts are damaged.
Step 3
Remove the plugs from the ports. Permit the bellows to fully extend.
Step 4
Close port C.
C
AA.96p.124
4-16 1006-0939-000
Page 67
Step 5
The bellows must not fall more than 100 mL/min.
If it does:
The port is not tightly sealed.
The bellows or the pressure relief valve is not correctly installed.
Parts are damaged.
4 User Maintenance
4-17
If the result for all the bellows tests was “passed,” install it in the system.
Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
1006-0939-000 4-17
Page 68
Aestiva
4-18
O2 sensor calibration - 21% O
Note: The 21% O2 calibration must be done before the 100% O2 calibration. During O2
calibration the screen replaces O perform calibration.
This procedure takes three minutes or less.
Step 1
Push the menu key.
AB.29.013
Step 2
Turn the knob to select Setup/Calibration (highlight).
AB.90.036
2
data with - - -. Mechanical ventilation must be off to
2
AB29.002
Step 3
Push the knob to show the next screen.
AB.90.037
AB.90.091
4-18 1006-0939-000
Page 69
Step 4
Tur n, then push the knob to select O
2 Sensor Cal.
Step 5
Select 21%. Then, push the knob.
4 User Maintenance
AB29.002
AB.90.048
AB.29.012
4-19
AB.90.049
Step 6
Complete the steps shown on the screen.
Remove the O
sensor from
2
the circuit.
• Do not twist or stress the
cable.
•Make sure the cable is
connected.
Step 7
Select Start Cal. Then, push the knob.
AB.48.080
The screen shows “Calibrating. . .”, followed by the result (“Complete” or “Failure”).
1006-0939-000 4-19
Page 70
Aestiva
If the calibration fails:
• perform the calibration again.
• do a 100% O
sensor calibration. If this passes, repeat the 21% O2 calibration.
2
4-20
If the calibration passes, install the O
sensor. If necessary, do the 100% O2 calibration.
2
Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
4-20 1006-0939-000
Page 71
O2 sensor calibration - 100% O2
Note: If w appears on the screen, the 21% O2 calibration must be completed before the
100% O
This procedure takes three minutes or less.
Step 1
calibration. Mechanical ventilation must be off to perform calibration.
2
4 User Maintenance
Push the menu key.
Step 2
Turn the knob to select Setup/Calibration (highlight).
Step 3
Push the knob to show the next screen.
4-21
AB.29.003
AB29.002
AB90.036
AB90.037
1006-0939-000 4-21
Page 72
Aestiva
Step 4
Tur n, then push the knob to select O
2 Sensor Cal.
AB29.002
AB90.048
4-22
Step 5
Select 100%. Then, push the knob.
Step 6
With the O2 sensor in the
circuit, fill the circuit with 100% O2:
• Push the flush button.
• Then flow 100% O2 at
5 L/min. (circuit should be open).
AB.29.046
AB.90.050
AB.90.088
Step 7
Select Start Cal. Then, push the knob.
AB48.081
4-22 1006-0939-000
Page 73
4 User Maintenance
The screen shows “Calibrating . . . ”, followed by the result (“Complete” or “Failure”).
If the calibration fails,
• perform the calibration again.
• decrease the airway pressure, and repeat the calibration.
Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
4-23
1006-0939-000 4-23
Page 74
4-24
Aestiva
Flow sensor calibration
Minimum Frequency: Weekly
The system automatically corrects for zero offset when you unplug the flow sensor connectors with power on. You must stop mechanical ventilation before you calibrate the flow sensors.
Step 1
Push up on the latch under the flow sensor module.
AA.96p.010
Remove the flow sensor module. You will feel some resistance. Continue to pull.
Step 2
When calibration is complete, the screen shows, "No Insp flow sensor" and "No Exp flow sensor".
No Insp Flow Sensor No Exp Flow Sensor
Step 3
• Install the flow sensor module.
4-24 1006-0939-000
Page 75
How to prevent water build-up
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
During the cleaning of the liquid collecting tray, avoid skin or eye contact with the contents of the absorber. In the event of skin or eye contact, immediately rinse the affected area with water and seek medical assistance.
4 User Maintenance
Why is water buildup a
problem?
How much water is too
much?
Where does the water
come from?
Solutions • Drain the absorber each morning.
Pooled water in the sensor or water in the sensing lines causes false alarms.
A thin layer of water or a foggy look in the flow sensors is OK. Drops of water is too much.
Water comes from exhaled gas and a chemical reaction between CO2 and the soda lime in the absorber.
At lower fresh gas flows more water builds up because less gas is scavenged and:
•More CO
•More moist, exhaled gas stays in the absorber.
• Drain all hoses as necessary.
• Flow sensor tubes must point up so that they do not collect water.
•With a circle breathing circuit, push the drain button before every case (listed in preoperative procedures).
• If Check Flow Sensors alarm occurs during a VERY LONG case, replace the flow sensors. Allow the original flow sensors to dry before you use them again.
stays in the absorber to react and produce water.
2
4-25
1006-0939-000 4-25
Page 76
4-26
Aestiva
4-26 1006-0939-000
Page 77
CCAAAAUUUUTTTTIIIIOOOONNNN
wCC
5 Alarms and Troubleshooting
5-1
No repair should ever be attempted by anyone not having experience in the repair of devices of this nature.
In this section About alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Alphabetical list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Breathing system problems (no alarm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
Electrical problems (power failure, etc.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16
Pneumatic problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-18
1006-0939-000 5-1
Page 78
5-2
Aestiva
About alarms
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
If an alarm occurs, safeguard the patient first, before troubleshooting
or repair procedures.
Two areas on the screen show alarms. The area at the top of the display shows most alarms. If there are more than 4 alarms at the same time, the lower priority alarms cycle every two seconds.
During severe malfunctions that prevent mechanical ventilation and/or monitoring, the area under the waveform shows minimum system messages. During normal operation, this area shows instructions (push the knob, etc.).
Alarms
Alarm priority depends on the level of danger to the patient. High priority alarms require immediate attention.
Priority Alarm tone Alarm silence Note
High 10 tones,
10 second pause, repeat
Medium 3 tones,
25 second pause, repeat
Low Single tone Tone does not repeat ---
120 seconds or cannot be silenced
120 seconds ---
Reverse video
AB.90.025
5-2 1006-0939-000
Page 79
5 Alarms and Troubleshooting
Alarm messages have three general causes.
• Malfunctions: Some malfunctions cause reduced function (for example, no PEEP). Others prevent mechanical ventilation (Minimum shutdown).
• Patient monitoring: These are high and low limit settings that you adjust.
• Informational: Control settings or system conditions can change operation. For example, if the audible circuit leak alarm is Off, the screen shows “Circuit Leak Audio Off” as a low priority alarm.
5-3
1006-0939-000 5-3
Page 80
5-4
Aestiva
Alphabetical list
The instructions in this section tell you what you can do:
• During a case to protect the patient
• After the case to repair a problem
This table does not include operator instructions.
There are two special types of alarms:
•Minimum monitoring alarms stop mechanical ventilation.
Minimum shutdown alarms stop mechanical ventilation and monitoring.
Message Priority Cause Action/Concerns Repair
+15V Analog Out­of-Range
-15V Analog Out­of-Range
12 Hour Test Low System in use for more
A/D Converter Failure
Absorber panel open
Adjust Low Ve Limit Medium The audible circuit leak
Min. shutdown (High)
Min. shutdown (High)
Min. shutdown (High)
Medium The top panel is not
Ventilator malfunction. Ventilate manually.
Ventilator malfunction. Ventilate manually.
than 12 hours without a power-up self test.
Ventilator malfunction. Ventilate manually.
completely closed.
alarm is Off (Alarm menu) but the low VE alarm is not set.
Contact a qualified
Monitoring is not reliable.
Monitoring is not reliable.
To do the test, move the system switch from Standby to On.
Monitoring is not reliable.
Close the panel. - - -
Set the low VE alarm. - - -
service representative.
Contact a qualified service representative.
Not necessary, informational.
Contact a qualified service representative.
VE alarm is Off in SIMV or PSVPro modes.
5-4 1006-0939-000
Page 81
5 Alarms and Troubleshooting
Message Priority Cause Action/Concerns Repair
Apnea Alarm Standby
Low Normal condition after
End Case, power-up, or ACGO change from On to Off
Apnea Alarm Off Low The cardiac bypass option
is selected (alarm limit menu).
Aux Gas Outlet On Medium
(low after acknowledged)
The outlet selection switch is set to the auxiliary common gas outlet.
Backup Mode Active
Battery Charger Fail
Low SIMV-PC + PSV mode
entered.
Low The current in the battery
charging circuit is too high.
Monitoring resumes after first breath (mechanical) or 2 breaths within 30 seconds (non-mechanical).
Apnea alarms are normally turned off when this option is selected.
Connect the patient circuit to the auxiliary outlet. For mechanical ventilation or manual ventilation with monitoring, select the common gas outlet.
Spontaneous breath rate fell below the set breath rate
The system is operational, but may fail later depending on what caused this alarm.
- - -
- - -
- - -
5-5
- - -
Contact a qualified service representative.
Battery Charging Low The battery is not fully
charged. If power fails, the total backup time will be less than 30 minutes.
Battery Current High
Battery Failure High
Low Battery current >
6 amps
for 10 seconds.
Low Battery voltage > 16 V for
10 seconds.
Battery Failure Low Low The battery voltage is too
low (<7 V) to supply the system if power fails.
Leave the system plugged in to charge the battery.
The system continues to operate, but may fail.
The system continues to operate, but may fail.
The battery does not have enough charge to power the equipment if power fails. Leave the system plugged in to charge the battery.
- - -
Contact a qualified service representative.
Contact a qualified service representative.
If the battery does not charge in 24 hours, contact a service representative.
1006-0939-000 5-5
Page 82
Aestiva
Message Priority Cause Action/Concerns Repair
5-6
Cal Flow Sensors Low The last flow sensor
calibration failed.
Calibrate O2
Low O
%>110% Does the sensor measure
2
Sensor
Canister open Medium The canister release is
open (large leak) without a bypass mechanism.
Cannot Drive Bellows
Low The internal manifold
pressure is higher than Paw + tolerance.
Cardiac Bypass Low The alarm limit settings
are set for a patient on cardiac bypass. Apnea alarms are off.
Check Flow Sensors
Medium (low after acknowledged)
No flow or negative flow on inspiratory sensor during inspiration in a circle system or negative flow on expiratory sensor in expiration (for 6 breaths in a row).
Calibrate the flow sensors. Look for water in the flow sensor tubes. Dry if
Contact a qualified service representative.
necessary.
sensor.
2
21% O
in room air?
2
Calibrate O
Close the canister release. A switch checks the
release position.
Fill the bellows if empty. - - -
Use the alarm limits menu to
- - -
change this setting.
Are the flow sensors correctly installed?
Inspect one way valves (breathing
circuit module). Water build-up in the flow sensor tubes?
Replace flow sensor
module with the Is a flow sensor tube cracked or broken?
spare. Check the
condition of the flow
sensor and its tubing.
Circuit Leak Audio Off
Low Control setting on the
Alarm limit menu.
This message tells you that the audio alarm for circuit
- - -
leaks was turned off.
Connect O
Sensor Low The O2 sensor is not
2
connected to the cable.
Connect the sensor. Contact a qualified
service
representative to
replace the cable.
5-6 1006-0939-000
Page 83
5 Alarms and Troubleshooting
Message Priority Cause Action/Concerns Repair
Control Settings Input has Failed
Minimum monitoring (Medium)
CPU Failure Minimum
shutdown (High)
CPU Internal Error Minimum
shutdown (High)
Display Voltage Out-Of-Range
Minimum shutdown (High)
Exp Flow Sensor
Low The system cannot read
Fail
Exp Reverse Flow Medium
(low after acknowledged)
Ventilator malfunction. Ventilate manually.
Monitoring is still available.
Ventilator malfunction. Ventilate manually.
Monitoring is not reliable.
Ventilator malfunction. Ventilate manually.
Monitoring is not reliable.
Ventilator malfunction. Ventilate manually.
Monitoring is not reliable.
Operation continues with the calibration data stored in the sensor.
Flow through the
default values.
Replace the flow sensor.
Look at the check valves. expiratory sensor during inspiration (for 6 breaths in a row).
Water build-up in the flow
sensor tubes?
Is a flow sensor tube cracked
or broken?
Contact a qualified service representative.
Contact a qualified service representative.
Contact a qualified service representative.
Contact a qualified service representative.
- - -
Replace the expiratory check valve.
Check the condition of the flow sensor.
5-7
Flow Valve (DAC) Failure Flow Valve (current)
Minimum monitoring (Medium)
Ventilator malfunction. Ventilate manually.
Monitoring is still available.
Contact a qualified service representative.
Failure
Gas Inlet Valve Failure
Hardware Watchdog Failure
1006-0939-000 5-7
Minimum shutdown
1
(High)
Minimum shutdown (High)
Ventilator malfunction. Ventilate manually.
Monitoring is still available.
Ventilator malfunction. Ventilate manually.
Monitoring is not reliable.
Contact a qualified service representative.
Contact a qualified service representative.
Page 84
Aestiva
Message Priority Cause Action/Concerns Repair
5-8
Heliox Mode is On Low Control setting on
ventilation setup menu.
High O
2
Medium O2% > alarm high limit
setting.
High Paw High Paw is greater than Plimit.
The ventilator cycles to expiration.
High Ve Medium The minute volume is
greater than the set high limit. This alarm is suspended for 9 breaths after you change the ventilator settings.
When Heliox is used, the ventilator must adjust volume calculations.
Is the limit set correctly? What is the O
flow?
2
Did you just push Flush? Does the sensor see 21% O2in
room air?
Are Plimit and other controls set correctly? Look for blockages. Check the patient connection.
Check patient for spontaneous breathing. Adjust control settings.
- - -
Calibrate O
Replace O
sensor.
2
sensor.
2
Calibrate the flow sensors. Replace the receiver filter.
- - -
High Vte Medium VTE is greater than high
alarm limit. This alarm is suspended for 9 breaths after you change the ventilator settings.
Insp Flow Sensor Fail
Low The system cannot read
the calibration data stored in the sensor.
Inspiration Stopped
Insp Reverse Flow Medium
High Drive gas safety switch
activated (high pressure)
Flow through the (low after acknowledged)
inspiratory sensor during
expiration (for 6 breaths
in a row).
Check patient for spontaneous breathing. Check ventilator and alarm settings.
Operation continues with default values. Replace the flow sensor.
Adjust controls. Check systems for blockages.
Look at the check valves.
Water build-up in the flow sensor tubes?
Is a flow sensor tube cracked or broken?
- - -
- - -
- - -
Replace the inspiratory check valve.
Check the condition of the flow sensor.
5-8 1006-0939-000
Page 85
5 Alarms and Troubleshooting
Message Priority Cause Action/Concerns Repair
Internal Ventilator Clock Too Fast
Minimum shutdown
Ventilator malfunction. Ventilate manually.
(High)
Internal Ventilator Clock Too Slow
Minimum shutdown
Ventilator malfunction. Ventilate manually.
(High)
Invalid Circuit Module
Low The system does not
recognize the type of circuit module installed. Normally the system uses the ID tabs to Identify circuits.
Limit Task Light Use Low The system is running on
battery power. Turn off the light to save power.
Loss of Backup Audio
Medium (low after
The audio alarm will not sound for a CPU failure.
acknowledged)
Contact a qualified
Monitoring is not reliable.
service representative.
Contact a qualified
Monitoring is not reliable.
service representative.
Make sure the module is correctly installed. Look for broken ID tabs or tape on the
Contact a qualified service representative
tabs
Turn off the light to extend
- - -
battery backup.
Monitor system operation. Contact a qualified
service representative.
5-9
Low Battery Voltage
Medium Voltage is <11.65V while
using battery power.
Low Drive Gas Pres Medium The ventilator did not
detect a rise in internal pressure when the flow valve opened.
Low O
2
High O2% < alarm low limit
setting
Manually ventilate the patient to save power.
Manually ventilate the patient.
Is the limit set correctly? Is the O
flow sufficient? Does
2
the sensor see 21% O
in
2
room air?
Make sure power is connected and circuit breakers are closed. Check ventilator fuse.
Make sure that the appropriate gas supplies (O
or Air)
2
are connected and pressurized.
Calibrate O2 sensor.
Replace O
sensor.
2
As sensors wear out, the measured % O
2
decreases.
1006-0939-000 5-9
Page 86
Aestiva
Message Priority Cause Action/Concerns Repair
5-10
Low Paw Medium Paw does not rise at least
4 cm from the lowest
pressure measured
during the last 20
seconds.
Low Ve Medium Exhaled minute volume
<low limit alarm setting.
This alarm is suspended
for 9 breaths after you
change the ventilator
settings.
Low Vte Medium Exhaled tidal volume <low
limit alarm setting. This
alarm is suspended for 9
breaths after you change
the ventilator settings.
Manifold Pressure Sensor Failure
Minimum monitoring
Ventilator malfunction. Ventilate manually. Contact a qualified
(Medium)
Are circuit connections OK? Look at the Paw gauge on the absorber.
Check patient condition. Check tubing connections. Check alarm settings.
Check patient condition. Check tubing connections. Check alarm settings.
Look for circuit disconnection.
- - -
- - -
service representative.
Memory (EEPROM) Fail
Memory (flash) Failure
Memory (RAM) Failure
Memory (Redundant Storage) Fail
Memory (video) Failure
Minimum shutdown (High)
Minimum shutdown (High)
Minimum monitoring (Medium)
Minimum shutdown (High)
The system cannot access
some stored values.
Default settings are used. Ventilation is still possible but service is necessary.
Ventilator malfunction. Ventilate manually.
Monitoring is not reliable.
Ventilator malfunction. Ventilate manually.
Monitoring is not reliable.
Ventilator malfunction. Ventilate manually.
Monitoring is still available.
Ventilator malfunction. Ventilate manually.
Monitoring is not reliable.
Contact a qualified service representative.
Contact a qualified service representative.
Contact a qualified service representative.
Contact a qualified service representative.
Contact a qualified service representative.
5-10 1006-0939-000
Page 87
5 Alarms and Troubleshooting
Message Priority Cause Action/Concerns Repair
Monitoring Only Medium A severe malfunction
prevents mechanical ventilation. Other alarms may also occur.
No Circuit Module Low The patient circuit module
is not installed.
No CO
Absorption Medium
2
(low after acknowledged)
The canisters are open (out of the circuit) but the bypass mechanism prevents a leak (optional feature).
No Exp Flow Sensor
No Insp Flow
Medium (low after acknowledged)
Electrical signals show the flow sensor is not connected.
Sensor
Ventilate manually. Cycle system power (On- Standby­On). If the alarm clears, restart mechanical ventilation.
Install a module. Refer to the Setup section.
User setting. Close the canister release to remove CO
from exhaled gas
2
Connect the flow sensors. Make sure the flow sensor module is on all the way.
Contact a qualified service representative.
Optical sensors look for tabs on the back of the module. Is the module assembled? are the sensors dirty?
- - -
- - -
5-11
No message, only specific shutdown message
High A severe malfunction
prevents mechanical ventilation and monitoring. Other alarms may also occur.
Ventilate manually. Use a stand-alone monitor. Cycle system power (On- Standby­On). If the alarm clears, restart mechanical
Contact a qualified service representative
ventilation.
No O
Pressure High
2
(cannot be silenced)
The O2 supply has failed. Air flow will continue.
Ventilate manually if necessary. Connect a pipeline supply or install an O
- - -
2
cylinder.
Flush Failure Low The pressure switch that
O
2
detects flush flow has seen a very long flush (≥30 seconds).
This alarm occurs if you hold down the Flush button for more than 30 seconds.
If the alarm occurs when flush is not in use, contact a qualified service representative.
1006-0939-000 5-11
Page 88
Aestiva
Message Priority Cause Action/Concerns Repair
5-12
O2 Sensor out of
circ
On Battery - Power OK?
Patient Circuit Leak?
Paw < -10 cmH
O High Subatmospheric pressure
2
Low O2 sensor not installed in
breathing circuit module;
sensor not measuring gas
in breathing circuit.
Medium (low after acknowledged)
The mains supply is not
connected or has failed
and the system is using
battery power.
Medium Exhaled volume <50% of
inspired volume for at
least 30 seconds
(mechanical ventilation).
(<-10 cmH
O).
2
Install a breathing circuit module and an O
sensor.
2
Ventilate manually to save power. At full charge, the battery permits approx. 30 minutes of mechanical ventilation.
Check breathing circuit and flow sensor connections.
Check patient condition, spontaneous activity? Increase fresh gas flow. Look for high flow through gas scavenging.
- - -
Make sure power is connected and circuit breakers are closed. Check ventilator fuse.
- - -
Calibrate the flow
2
sensors.
With active scavenging, check the negative relief valve on the receiver.
PEEP Not Achieved Low Pmin does not reach
within 2 cmH
O of PEEP
2
by the end of mechanical expiration for 6 consecutive breaths.
Positive SIB Vref Out-of- Range
Minimum shutdown
Ventilator malfunction. Ventilate manually.
(High)
Pres Mode Not Avail
Medium (Pressure, PSVPro and SIMV modes)
Manifold pressure not tracking airway pressure, or manifold pressure ≤ -15 cmH
O.
2
Low (Volume mode)
Check tubing connections. Rate and/or I:E ratio may prevent ventilator from reaching desired PEEP level.
Monitoring is not reliable.
Pressure control mode and PEEP are not available. Switch to Volume mode or ventilate manually.
- - -
Contact a qualified service representative.
Contact a qualified service representative.
5-12 1006-0939-000
Page 89
5 Alarms and Troubleshooting
Message Priority Cause Action/Concerns Repair
Pres/Vol Mon Inactive
Pressure Limit Switch Failure
Replace O
Sensor Low O2% < 5% Makes sure patient receives
2
Medium Outlet selection switch is
set to auxiliary gas outlet.
Minimum monitoring (Medium)
A pressure safety switch activated at a Paw <90 cmH
O and Pmanifold
2
<80 cm H
O.
2
Select Gas Outlet Medium Fresh gas may not flow to
the patient. Auxiliary gas outlet is On, but flow sensors have seen 3 breaths in patient circuit during the last 30 seconds.
Connect the patient circuit to the auxiliary gas outlet or set the switch to the common gas outlet for normal operation.
Ventilate manually. Monitoring is still available. Extreme control combinations may cause this alarm. Check control settings.
O
. Does the sensor see 21%
2
in room air? Use different
O
2
monitor.
Select the common gas outlet or connect the patient circuit to the auxiliary outlet.
- - -
Contact a qualified service representative.
Calibrate O2 sensor.
Replace O
sensor.
2
Note: the bag arm will not ventilate a patient at the auxiliary outlet.
5-13
Service Calibration Low Internal calibrations are
necessary for maximum accuracy.
Software Error Minimum
Ventilator malfunction. Ventilate manually. shutdown (High)
Software Watchdog Failure
Minimum shutdown
Ventilator malfunction. Ventilate manually.
(High)
Sustained Airway Pressure
Minimum shutdown
Paw > 100 cmH
seconds.
O for 10
2
The system is operational. Contact a qualified
service representative.
Contact a qualified
Monitoring is not reliable.
service representative.
Contact a qualified
Monitoring is not reliable.
service representative.
Check tubing for kinks, blockages, disconnects.
Calibrate the flow sensors.
(High)
1006-0939-000 5-13
Page 90
Aestiva
Message Priority Cause Action/Concerns Repair
5-14
Sustained Paw High Paw > sustained pressure
limit for 15 seconds
3
.
System Leak? Medium Delivered volumes do not
match set volumes.
Vaux_ref Out-of­Range
Minimum shutdown
Ventilator malfunction. Ventilate manually.
(High)
Vext_ref Out-of­Range
Minimum shutdown
Ventilator malfunction. Ventilate manually.
(High)
Volume Apnea Medium No mechanical breaths or
spontaneous breaths >20 mL in last 30 seconds.
Check tubing for kinks, blockages, disconnects.
If you are using Heliox, select Heliox on the ventilator setup menu. Look for leaks in the manifold. Compare set to delivered volumes.
Monitoring is not reliable.
Monitoring is not reliable.
Check patient. Bag as needed. Check for disconnects. If the patient is on a heart lung machine, select Cardiac Bypass on the main menu.
Calibrate the flow sensors.
Calibrate the flow sensors. Drain water buildup from the breathing system.
Contact a qualified service representative.
Contact a qualified service representative.
- - -
Vol Apnea > 2 min High No mechanical breaths or
See above. - - ­spontaneous breaths >20 mL in last 120 seconds.
Vt Not Achieved Low Tidal volume measured by
inspiratory flow sensor < set value 6 breaths in a row after the first minute
Adjust controls to supply
adequate tidal volumes.
Check I:E; Plimit; and volume
settings.
Possible leak.
of mechanical ventilation.
Vte > Insp Vt Medium Expired volume > inspired
Check patient condition. - - ­volume for 6 breaths with a circle module.
1. When power is first turned on.
2. Flow sensors are also used to measure pressures.
3. The sustained pressure threshold is calculated from the pressure limit setting. When mechanical ventilation is on, the sustained limit is calculated as follows: for pressure limits < 30 cmH2O, the sustained pressure limit is 6 cmH2O; for Plimit between 30 and 60 cmH2O, the sustained limit is 20% of
the pressure limit (Plimit); for pressure limits > 60 cmH the sustained pressure limit increases by PEEP - 2 cmH pressure limit is calculated as follows: for pressure limits ≤ 60 cmH limits > 60 cmH
5-14 1006-0939-000
O, the sustained pressure limit is 30 cmH2O.
2
O, the sustained pressure limit is 12 cmH2O. If both PEEP and mechanical ventilation are on,
2
O (the compensated weight of the bellows). When mechanical ventilation is off, the sustained
2
O, the sustained pressure limit is 50% of the pressure limit (Plimit); for pressure
2
Page 91
Breathing system problems (no alarm)
Symptom Problem Solution(s)
Gas scavenging flow is too low. Suction supply problem. Use a different suction
5 Alarms and Troubleshooting
supply.
The bellows fills when the Bag/ Vent switch is set to Bag or the bag fills when the switch is set to Vent.
The ventilator does not read the position of the Bag/Vent switch. Use manual ventilation, if necessary.
The ventilator does not correctly identify the breathing circuit module.
Filter blockage. Active systems have a flow indicator to show this.
Leak through Bag/Vent switch. Causes include dirt or damage to the o-ring.
Ventilator or absorber malfunction.
Broken identification tabs. Remove the module and look
Dirt on the optical sensors that read the tabs.
Ventilator malfunction. Ask a qualified service
Replace the filter. Refer to “Disassemble the scavenging path” in Section2, Cleaning and Sterilization.
Refer to “Manifold maintenance” in Section 4, User Maintenance.
Ventilate manually. Ask a qualified service representative to repair the system.
at the tabs on the rear. If there is damage, use a different module.
Use a DAMP alcohol prep to gently clean. Squeeze dry the prep before you use it.
representative to repair the system.
5-15
APL valve does not operate correctly.
Bellows falls > 100 mL/min during bellows tests.
1006-0939-000 5-15
APL valve problem. Replace APL Valve seal and
diaphragm - Refer to User Maintenance.
Leak in the breathing system.
Check/clean/reposition the pressure relief valve. If the problem persists, replace the pressure relief valve, bellows base, or bellows assembly.
Page 92
Aestiva
Electrical problems (power failure, etc.)
wwwwWWWWAAAARRRRNNNNIIIINNNNGGGGIf a circuit breaker opens frequently, do not use the system. Have
an approved service representative repair the system
Symptom Problem Solution
5-16
Mains indicator is not on. The electrical power
cable is not connected.
The inlet circuit breaker (toggle switch) is open.
The power cable is damaged.
The electrical socket the power cable connects to has no power.
An internal fuse is open. Have a service
The electrical outlets do not have power but the mains indicator is on.
One electrical outlet does not have power.
A circuit breaker opens frequently.
The main outlet circuit breaker is open.
The outlet circuit breaker is open.
Equipment connected to the outlet(s) uses more current than the circuit breaker rating (Figure 5-1).
The equipment connected to the outlet has a short.
Connect the power cable.
Close the circuit breaker (Figure 5-1).
Replace the power cable.
Use a different electrical socket.
representative repair the system.
Close the circuit breaker.
Close the circuit breaker.
Use a different power supply for some of the equipment.
Do not use the equipment until it is repaired.
Tec 6 has no power. Not plugged into outlet. Connect power cable.
5-16 1006-0939-000
Page 93
5 Alarms and Troubleshooting
Out
(No Power)
In
(OK)
AA.96.109
123
Open
(No Power)
Closed
(OK)
AA.96.108
5-17
AB.29.007
5
4
AA.96.117
1. Outlet CIrcuit Breakers*
2. Master Circuit Breaker (all outlets)*
3. Inlet Circuit Breaker*
4. Ventilator Fuse
5. Mains Indicator
* Labels show ratings.
Figure 5-1 • Circuit breakers, fuse, and the mains indicator
1006-0939-000 5-17
Page 94
Aestiva
Pneumatic problems
Symptom Problem Solution
High-pressure leak Controls are not set correctly. Set the system switch to
Standby and the auxiliary flowmeter to Off.
5-18
Equipment connected to pneumatic outlet.
High-pressure leak Incorrect cylinder connection (cylinder
yokes).
Incorrect cylinder connection (DIN connection.)
Low-pressure leak with a vaporizer on.
The vaporizer is not correctly installed. Correctly install the
The vaporizer fill spout is loose (fill port type vaporizer).
Vaporizer port o-rings (external) are damaged or not installed.
A vaporizer malfunction (the leak stops if you use a different vaporizer in the same position).
A port valve malfunction (the leak continues if you use a different vaporizer in the same manifold position).
Disconnect the equipment.
Make sure that there is only one cylinder gasket, the gasket is in good condition, and the T handle is tight.
Make sure the nut is tight.
vaporizer.
Tighten the fill spout.
Install new o-rings.
Send the vaporizer to a Datex-Ohmeda Service Center.
Have an approved service person repair the vaporizer manifold.
CCAAAAUUUUTTTTIIIIOOOONNNN::::
wCC
No repair should ever be attempted by anyone not having experience in the repair of devices of this nature.
5-18 1006-0939-000
Page 95
6 Illustrated Parts
6-1
In this section Breathing System Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Main manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Test tools and system parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-10
1006-0939-000 6-1
Page 96
Aestiva
Breathing system parts
Top level
Item Description (Figure 6-1) Stock Number
1 Flow sensor module (cover & holder) 1406-8208-000
a. Cover only 1406-3401-000
b. Holder only 1406-3400-000
2 Flow sensor (plastic) 1503-3856-000
Flow sensor (metal - autoclavable) 1503-3244-000
6-2
3O
4Breathing circuit module, circle 1406-8102-000
5 Bag arm long (adjustable) 1406-3350-000
6 Main manifold assembly 1406-8209-000
7Autoclavable Bellows Assembly 1406-8105-000
8 Blank, pop-off valve 1406-3240-000
9 Exhalation valve assembly 1503-8114-000
10 Rear column cover 1406-3455-000
11 Gas scavenging (4 options) --------
sensor, cell 6050-0004-110
2
O-ring 1406-3466-000
Breathing circuit module, Mapleson/Bain 1406-8211-000
Bag arm short (not adjustable) 1406-3380-000
Active gas scavenging assembly, 6.35 mm (ANSI, Not European Community)
Active gas scavenging assembly, 30 mm threaded outlet
1406-8216-000
1406-8214-000
Passive gas scavenging assembly, 30 mm ISO taper
Passive gas scavenging assembly, 25 mm taper (DEU)
12 AGSS nylon filter, 225 µ 1406-3521-000
6-2 1006-0939-000
1406-8215-000
1406-8217-000
Page 97
6 Illustrated Parts
Item Description (Figure 6-1) Stock Number
13 Upper dish (2 options) ------
Upper dish absorber plastic (seal not
1406-3413-000
included)
bypass assembly metal (seal not
CO
2
1406-3500-000
included)
14 Upper seal 1406-3414-000
15 Tube corrugated 210 mm 1400-3009-000
16 Elbow transfer tube 1406-3576-000
17 Transfer tube 1406-3575-000
18 Drain dish assembly 1406-8218-000
19 Plug, sight glass-drain 0229-2080-100
O-ring, plug 0210-0594-300
20 Canister 0229-3015-800
--- Not Shown
Upper seal, upper dish, corrugated tube,
1406-8207-000
drain dish, canisters
Barb cap, passive scavenging 1406-3542-000
6-3
1006-0939-000 6-3
Page 98
Aestiva
6-4
6
7
5
AB.23.095
8
9
4
3
2
1a
1b
* Active scavenging systems only
Figure 6-1 • Breathing system top assemblies
10
1112*13141516171820 19
6-4 1006-0939-000
Page 99
Main manifold
2c
2b 2a
3
6 Illustrated Parts
11
AB.23.096
10
9
8
6-5
7
6
4
5
2
2
2
1
AB.23.071
Top Bottom
Figure 6-2 • Main manifold
1006-0939-000 6-5
Page 100
Aestiva
Item Description (Figure 6-2) Stock Number
1 Seal, main manifold, APL/AGSS 1406-3316-000
2 Seal, bag/vent assy (a, b, c)
a. O-ring (Qty 2) 1406-3278-000
b. Poppet (Qty 2) 1406-3279-000
c. Seal only 1406-3314-000
3 Plate, top, main manifold 1406-3300-000
4 Plate, bottom main manifold 1406-3303-000
5 Seal, Main Manifold, Absorber 1406-3315-000
6 Body manifold, main 1406-3301-000
6-6
7 APL diaphragm assembly 1406-3331-000
8Manifold APL/AGSS 1406-3302-000
9 Split ring (Qty 3) 1406-3319-000
10 Thumbscrew, M6x43 mm w/10 mm Head 1406-3306-000
11 M6 x28.5 mm Thumbscrew (Qty 2) 1406-3305-000
6-6 1006-0939-000
Loading...