* O M E 2 3 7 3 0 D 0 0 ** O M E 2 3 7 3 0 D 0 0 *
IMPORTANT NOTICES
• No part of this manual may be copied or reproduced without written permission.
• If this manual is lost or worn, contact your dealer about replacement.
• The contents of this manual and equipment specifications are subject to change without
notice.
• The example screens (or illustrations) shown in this manual may not match the screens
you see on your display. The screen you see depends on your system configuration and
equipment settings.
• This manual is intended for use by native speakers of English.
• FURUNO will assume no responsibility for the damage caused by improper use or
modification of the equipment or claims of loss of profit by a third party.
• Dispose of the equipment according to appropriate regulations.
i
SAFETY INSTRUCTIONS
WARNING
ELECTRICAL SHOCK HAZARD
Do not open the equipment.
Only qualified personnel
should work inside the
equipment.
Immediately turn off the power at the
switchboard if water leaks into the
equipment or something is dropped in
the equipment.
Continued use of the equipment can cause
fire or electrical shock. Contact a FURUNO
agent for service.
Do not disassemble or modify the
equipment.
Fire, electrical shock or serious injury can
result.
Do not place liquid-filled containers on
the top of the equipment.
Fire or electrical shock can result if a liquid
spills into the equipment.
Immediately turn off the power at the
switchboard if the equipment is emitting
smoke or fire.
WARNING
Keep heater away from equipment.
A heater can melt the equipment's power
cord, which can cause fire or electrical
shock.
Use the proper fuse.
Fuse rating is shown on the equipment.
Use of a wrong fuse can result in damage
to the equipment.
Do not operate the equipment with wet
hands.
Electrical shock can result.
Warning Labels
Warning labels are attached to the equipment. Do not remove the labels. If a
label is missing or damaged, contact
a FURUNO agent or dealer about replacement.
WARNING
To avoid electrical shock, do not
remove cover. No user-serviceable
parts inside.
Continued use of the equipment can cause
fire or electrical shock. Contact a FURUNO
agent for service.
Make sure no rain or water splash leaks
into the equipment.
Fire or electrical shock can result if water
leaks in the equipment.
ii
Electrical shock hazard.
Do not remove cover.
No user-serviceable parts inside.
Turn off power before servicing.
And wait until charge lamp (green)
goes off and then remove cover.
Do not transmit when the transducer is
out of water.
The transducer may become damaged.
The picture does not advance when the
picture advance speed is set for
"STOP".
A dangerous situation may result if the
vessel is navigated while monitoring
the depth indication since it is not updated
when the picture is stopped.
Set the gain properly.
No picture appears if the gain setting is
too low, and noise appears when the gain
is too high. If the gain is set improperly,
the depth indication may be wrong,
resulting in a potentially dangerous situation
if the vessel is navigated by monitoring
the depth indication.
Gain too high Proper gain Gain too low
iii
v
TABLE OF CONTENTS
FOREWORD ........................................................................................................ vi
SYSTEM CONFIGURATION ............................................................................... vii
FURUNO Electric Company thanks you for considering and purchasing the FCV-30 Color
Sounder. We are confident you will discover why the FURUNO name has become
synonymous with quality and reliability.
For over 50 years FURUNO Electric Company has enjoyed an enviable reputation for
quality and reliability throughout the world. This dedication to excellence is furthered by our
extensive global network of agents and dealers.
Your equipment is designed and constructed to meet the rigorous demands of the marine
environment. However, no machine can perform its intended function unless properly
installed and maintained. Please carefully read and follow the operation and maintenance
procedures set forth in this manual.
We would appreciate feedback from you, the end-user, about whether we are achieving our
purposes.
Thank you for considering and purchasing FURUNO.
Features
The FCV-30 is a high-performance echo sounder designed for fishing as well as
oceanographic research. Its newly developed multi-beam transducer provides
multi-directional and long-range detection. Consisting of a processor unit, control unit,
transceiver unit and transducer, the FCV-30 provides high definition underwater images on
the monitor of your choice.
The main features of the FCV-30 are
• Stabilization compensation in all directions (all directions ±20°) provides stable images in
rough seas.
• Fish distribution and size can easily be found using the fish histogram display.
• Detection in multi-direction or desired direction to help discern fish school density.
• Target graph lets you monitor fish movement.
• Bottom hardness and roughness graph provides intuitive data on bottom composition.
• User programmable display (three types) divides the screen in two, three, four or five
alarms requires appropriate sensor. The vertical temperature alarm additionally requires a
net sonde or trawl sonar.)
• USB port provided to save data to USB hard disk.
vi
SYSTEM CONFIGURATION
Monitor Unit
Control Unit
CV-301
F1
F2
F3
F4
100-240 V AC
φ
, 50/60 Hz
1
Processor Unit
CV-300
POWER
Transceiver Unit
CV-302
MU-151C/MU-201C (option) OR
commercially available monitor
Navigator
Satellite Compass
Net Sonde
Hard Disk
USB Memory
100-120/200-240 V AC
φ
, 50/60 Hz
1
Network Sounder
ETR-30N
Junction Box
(option)
CV-304
Transducer
CV-303
System configuration
.
vii
This page intentionally left blank.
viii
1. OPERATIONAL OVERVIEW
1.1 Controls
Thumbwheel
Shifts cursor position in
Left button
Confirms menu selection.
Turns the power
on/off.
setting window.
Right button
Displays pop-up
menu.
User-programmable
function keys.
F1
F2
F3
F4
Trackball
Moves trackball
pointer.
Control unit
1-1
1. OPERATIONAL OVERVIEW
1.2 Menu Operation
The FCV-30 is controlled from a menu system. Two types of menus are available:
Main menu: All menu options are contained in pull-down menus.
Pop-up menu: Contains most frequently used menu functions.
1.2.1 How to use the main menu
Close button
Menu barTitle bar
File Disp Setting System Window Marker line Monday, April 10, 2006 15:01
Trackball pointer
Date and time
Menu bar
1. Roll the trackball to choose menu desired (File, Disp, Setting, System, etc.) from the
menu bar at the top of the screen and the click it with the left button.
Ellipsis indicates a
dialog box is available
Setting menu
2. Roll the trackball to choose menu item desired and then click with the left button. A
menu item which contains an ellipsis (…) indicates a dialog box is available.
Tenkey
button
Multi beam dialog box
1-2
1. OPERATIONAL OVERVIEW
3. Change settings as appropriate, referring to the information below.
4. After changing settings, roll the trackball to place the pointer on the OK button and then
push the left button to save settings and close the dialog box. (See Note 3 below.)
Note 1: To restore previous settings, click the Cancel button instead of the OK button at
step 4.
Note 2: To see the result of a change without closing a dialog box, click the Apply button.
Note 3: Unless otherwise noted, the phrase “click…” means to click the object with the left
mouse button.
Changing setting using radio buttons
EnabledDisabled
Radio buttons
Click the item desired. A black dot appears when the item selected is enabled.
Changing setting using a list box
Normal
List box
1. Click ▼ in the list box or click the box itself to show a list of options.
2. Click the desired setting.
Changing setting using a spin box
Spin box
A setting may be changed from a spin box two ways depending on contents of the spin box:
1) Click ◄ or ► successively in the spin box to choose the setting desired. Each click
increments the setting downward or upward depending on arrow clicked.
2) Spin the thumbwheel to choose value with the cursor.
Cursor
Spin box
Place the pointer on the “tenkey” button () at the lower left-hand corner and then push the
left button to show the tenkey panel.
1-3
1. OPERATIONAL OVERVIEW
Tenkey
Close
Tenkey panel
Place the cursor on the desired numeric key and then push the left button. Repeat to enter
all data required.
Changing setting using a check box
Bottom alarm
1. Place the pointer on the check box of the item you want to enable or disable.
2. Push the left button to remove or insert the check mark as appropriate. An item is
enabled when the check mark is present.
Changing setting using a toggle button
Enabled (highlighted in blue)Disabled
1. Click the item desired.
2. Each click enables or disables your selection alternately. When the item selected is
enabled it is highlighted in dark blue and when disabled it is colored gray.
1-4
1. OPERATIONAL OVERVIEW
1.2.2 How to use the pop-up menus
1. Right-click anywhere on the screen, and a pop-up menu appears.
Indicates an
option menu
is available
Pop-up menu
2. Click the item desired. An item with ► indicates an option window; “…” (ellipsis) a
setting window. Currently selected option is marked with a filled circle.
Option window for "Mode".
Sliding
bar
Track
bar
Close button
Pop-up menu and shift setting window
3. Do one of the following according to item selected.
Operation from an options window
Roll the trackball to execute appropriate operation and then push the left button.
Operation from a setting window
There are three settings methods:
• Drag the sliding bar: Place the pointer on the sliding bar and hold down the left button
while rolling the trackball.
• Place the pointer on the track bar and then click with the left button
• With the track bar selected, spin the thumbwheel and roll the thumbwheel downward to
increment figure upward; upward to increment figure downward. If the drag bar is not
selected, place the pointer on it and push the left button.
Finally, click the Close button.
1-5
1. OPERATIONAL OVERVIEW
1.3 Turning the Power On/Off
Turning on the power
1. Turn on the monitor.
2. Open the power switch cover on the Control Unit and push the
comes on, Windows
display appears.
Windows is the registered trademark of Microsoft, Inc.
Note that the Windows
Turning off the power
®
* starts up, the FCV30 application launches and then the last-used
®
Help feature is not available with this equipment.
switch. The power
There are two methods to turn off the power.
a) Press the
switch. Do not press the switch more than four seconds; last-used
settings will not be memorized.
b) Click the close button (
). The following window appears. Then, click the Yes button to
quit.
Confirmation
Are you sure to quit?
[Yes]-> Shutdown
[No ]-> Cancel
Yes No
Shutdown confirmation window
Note: Power is automatically turned off anytime the processor unit is disconnected from the
control unit.
1-6
1. OPERATIONAL OVERVIEW
1.4 Transmitting, Receiving
Follow the procedure below to start transmitting and receiving, after turning on the power.
1. Click File.
File menu
2. Click TX Start. Transmission starts and images from underwater appear.
To stop transmission, choose TX Stop at step 2. It is recommended that, after stopping
transmission, wait three seconds before using the menu.
1-7
1. OPERATIONAL OVERVIEW
1.5 Displays
1.5.1 Choosing a display
Six display modes are available. Choose the display mode which matches your current
needs.
1. Click Disp.
Disp menu
2. Click Mode.
Mode dialog box
3. In the Display Mode window choose the mode you wish to use. See the next several
sections for display descriptions. To divide the screen, go to step 4. Otherwise go to step
5.
4. In the Screen Layout window, click the screen division you wish to use. The Preview
window shows the results of your selection. Note that the split-beam display cannot be
divided.
5. Click the OK button to finish.
1-8
1. OPERATIONAL OVERVIEW
1.5.2 Split-beam display
The split-beam display shows the underwater images captured with the split-beam. With
display of appropriate window you can discern single fish distribution (with the fish size
histogram) and monitor fish movement (with the target graph).
Timer marker
(each 30 sec.)
Measuring range
Elapsed time and postion
of vertical VRM
Mode indicator
Color bar
Temperature scale
Horizontal VRM
Depth of horizontal
TS
4.0
2.0
0.0
-2.0
-4.0
−
−
−
−
−
Fish echo
Bottom echo
60
00:00:28
34 42.616' N
135 19.740' E
0
20
40
60
Zero line
Temperature graph
Net marker
Depth scale
V ertical VRM
VRM marker
80
Split-beam display
Note: When the display range for the split-beam presentation is 500 m or higher, the
indication “Deep mode” appears in the menu bar. To enable deep mode detection,
click Deep Mode on the menu bar followed by clicking ON. The default deep mode
detection range is 500 m. For further details, contact your dealer.
Deep mode
Deep mode detection options
Menu for turning
deep mode
on or off
1-9
1. OPERATIONAL OVERVIEW
1.5.3 Three-beam display
The three-beam display shows the images captured from the port, downward and starboard
directions.
* Beam direction
B: Bearing (angle from bow direction)
T: Tilt angle (angle from downward direction)
<3>
20
0
Horizontal range
from transducer
Dividing
B:270
T: 10
20
0
B: 0
T: 0
0
B: 90
T: 10
<0><3>
20
15
line
40
60
40
60
40
60
* Beam direction can
be shown with an
80
Beam 3
(Port)
80
Beam 2
(Downward)
80
Beam 1
(Starboard)
icon ( ). See
paragraph 3.2.
Three-beam display
Changing display area
The width of each display can be adjusted as follows:
1. Place the trackball pointer on a dividing line between displays, and the pointer changes
to a double-ended arrow (
).
2. Hold the left button while dragging the double-ended arrow with the trackball to adjust
Note: Normally, pulses are fired from beam 1, beam 2 and beam 3 in that order and
displayed on the screen as the port, downward and starboard images. When the
display range of the 3-beam display is higher than 1000 m, the menu bar shows
“Simul. TX”. To fire pulses from beam 1 – 3 simultaneously click the Simul TX
indication to show ON. When this is done the picture advance speed quickens.
However, the display becomes susceptible to unwanted echoes. The default Simul
TX detection range is 1000 m. This value can be set between 100 and 5000 m. For
further details, see page 3-6.
Menu for turning
simultaneous
transmission on or off
Simul. TX
Simultaneous transmission options
1.5.4 Split-beam + two-beam display
The split-beam + two-beam display provides a split-beam plus images captured from the
port and starboard sides.
Split-beam display
0
TS
B:270.0
T: 10.0
0
B: 90.0
20
T: 10.0
40
60
80
20
40
60
80
0
20
40
60
80
Beam 2
(Port)
Beam 1
(Starboard)
Split-beam + two-beam display
Note: When the display range for the split beam + two beam presentation is 1000 m or
higher, “Deep Mode” and “Simul Tx” appear in the menu bar. To enable the deep
mode or simultaneous transmission, click the appropriate item on the menu bar
followed by clicking ON.
1-11
1. OPERATIONAL OVERVIEW
Deep mode, Simul. TX
Menu bar
1.5.5 User 1, User 2, User 3 display
These displays show the product of displays created with “Display mode change” button in
the Mode window. The default settings are as follows:
User 1: Split beam display
User 2: 3-beam (port, starboard and downward directions)
User 3: Split beam + two-beam (port and starboard directions)
For how to customize the User displays see paragraph 3.1.
1-12
1. OPERATIONAL OVERVIEW
1.6 Choosing a Display Range
Choose the detection range (from the transducer to the bottom), in 12 preset choices. The
default ranges are shown in the table below.
Display ranges
Unit
1 2 3 4 5 6 7 8 9 10 11 12
m 10 20 40 80 150300 500 10001500 2000 25003000
Note: You may set ranges as desired. For further details, see Chapter 3.
1. Click Setting.
Setting menu
2. Click Echo image.
Display Range
Echo image dialog box
1-13
1. OPERATIONAL OVERVIEW
3. Click ▼ in the Range box.
Range options
4. Click the range you wish to use. For example, choosing “80”, the display will show
underwater conditions from the transducer down to 80 m.
5. Click the OK button.
1-14
1. OPERATIONAL OVERVIEW
1.7 Shifting the Display Range
You may shift the display in order to look at a shallower or deeper depth without changing
the current range.
Shift the start
depth to watch
shallow or deep
Shift principle
1. Click Setting.
2. Click Echo image.
3. Click ◄ or ► in the Shift box to choose shift value.
4. Click the OK button.
Picture
1-15
1. OPERATIONAL OVERVIEW
1.8 Adjusting the Gain
Adjust the gain according to signal strength. Adjust so noise just disappears from the
screen.
CAUTION
Use the proper gain setting.
Incorrect gain may produce wrong depth
indication, possibly resulting in a
dangerous situation.
1. Click Setting.
2. Click Echo image.
3. Click ◄ or ► in the Gain box to choose the gain setting desired.
Gain too high Gain proper Gain too low
Examples of gain settings and resulting pictures
4. Click the OK button.
1-16
1. OPERATIONAL OVERVIEW
1.9 Find Depth and Position of a Fish Echo
You may measure the depth to a fish school or bottom with the horizontal line marker, called
the horizontal VRM (Variable Range Marker) marker. Further, you may measure elapsed
time with the vertical VRM. For example, you can measure how many minutes before the
current time a fish echo appeared.
Elapsed time and
postion* to vertical VRM
0
20
40
Vertical
VRM
60
80
Horizontal VRM
Depth to
horizontal VRM
TS
40
* Position shown in red when position data is lost.
00:00:28
34 42.616' N
135 19.740' E
Measuring depth, elapsed time and position with the VRMs
Measuring depth
1. Roll the trackball to place the pointer on the horizontal VRM. The shape of the pointer
changes to (
).
2. Hold down the left button while rolling the trackball upward or downward to place the
horizontal VRM on the object that you wish to measure the depth.
3. Read the depth indication (on the horizontal VRM).
Measuring elapsed time
1. Roll the trackball to place the pointer on the vertical VRM.
2. Hold down the left button while rolling the trackball leftward or rightward to place the
VRM on the object that you wish to find position.
3. Read the elapsed time and position (on the vertical VRM).
Note 1: The color of the VRMs can be changed. See paragraph 3.5
Note 2: You may change the location of the depth indication on the horizontal VRM. For
further details see paragraph 3.2.
1-17
1. OPERATIONAL OVERVIEW
1.10 Inscribing Lines
You may inscribe vertical lines on the display to mark fish schools, shoals, etc.
When an echo of interest appears, click Marker line on the menu bar to inscribe a vertical
line on the display. If TIMER is turned on in the Status window, elapsed time is counted from
the moment the mark is inscribed.
Note that color of the marker line can be changed. For details, see paragraph 3.5.
1.11 Fish Size Histogram
The fish size histogram display shows single fish size within the area measured. You may
set up this display as follows:
1. Display the split-beam display. See paragraph 1.5.
2. Click Disp.
3. Click Display Window.
Display window dialog box
4. Click Fish size histogram. The toggle button is then colored in purple (ON state). Each
click turns the display on or off alternately.
5. Click the OK button. The fish size histogram appears and measuring begins.
1-18
Vert. axis
1. OPERATIONAL OVERVIEW
Number of fish within vertical axis
Number of fish within measuring area
Area of measurement
0
20
40
60
80
Fish size
histogram
window
125/247
Hor. axis
40
00:00:28
34 42.616' N
135 19.740' E
Fish size histogram window
Note 1: The fish size histogram window can be moved as desired. For details, see Chapter
2.
Note 2: When the fish size histogram window is closed, all data measured is deleted.
How to read the fish histogram
The bar graph shows size and proportion of fish in measuring area. The vertical axis shows
fish length (default setting) and the horizontal axis shows distribution. In the fish size
histogram display above, more than 50% of the fish in the measuring area are 10 cm in
length.
1-19
1. OPERATIONAL OVERVIEW
1.12 Setting Measuring Area
There are four ways to set the measurement area for the fish size histogram:
• Measure fish at specific location
• Measuring depth shallower than bottom in all directions
• Measure specific depth range
• Measure bottom fish
1.12.1 Measuring fish in a specific location
1. Click Setting.
2. Click Fish size. Click the Fish size histogram tab if necessary to show the fish size
dialog box.
Fish size dialog box, measurement
If the Unit or Display dialog box is shown, click the Measurement tab.
3. Click ▼ in the Measuring area box.
Area options
4. Click Manual.
1-20
Loading...
+ 80 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.